INSTRUCTION MANUAL
HI 96752HI 96752
HI 96752
HI 96752HI 96752
Calcium and Magnesium
High Range
ISM
www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna product. This manual will provide you
with the necessary information for the correct use of the instrument. Please
read it carefully before using the meter. If you need additional technical
information, do not hesitate to e-mail us at tech@hannainst.com.
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is
not damaged. If any damage occured during shipment, please notify
your Dealer.
Each HI 96752 Ion Selective Meter is supplied complete with:
• Two Sample Cuvettes and Caps
• 9V Battery
• Instruction Manual
Note: save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.
For more details about spare parts and accessories
see “Accessories”.
Technical specifications:
Range Calcium 0 to 400 mg/L
Magnesium 0 to 150 mg/L
Resolution 1 mg/L
Accuracy Calcium ±10 mg/L ±5% of reading at 25°C
Magnesium ±3 mg/L ±3% of reading at 25°C
Typical EMC Dev. ±1 mg/L
Light Source Light Emitting Diode
Light Detector Silicon Photocell with narrow band interference
Method For Calcium: Adaptation of the Oxalate method.
Environment 0 to 50°C (32 to 122°F);
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use in measurement mode;
Dimensions 192 x 104 x 69 mm (7.6 x 4.1 x 2.7”)
Weight 360 g (12.7 oz.).
filter @ 466 nm
For Magnesium: Adaptation of the Calmagite method.
max 95% RH non-condensing
after 1 hour of non-use in calibration mode;
with last reading reminder.
Functional description:
1. RANGE/GLP/ key: press to change the paramter, press and hold
for three seconds to enter
the date and time.
2. CAL CHECK key: press to perform the validation of the meter, or
press and hold for three seconds to enter
3. ZERO/CFM key: press to zero the meter prior to measurement, to
confirm edited values or to confirm factory calibration restore.
4. READ/ /TIMER key: In
measurement, or press and hold for three seconds to start a
pre-programmed countdown prior to measurement. In
press to view the next screen.
5. ON/OFF key: to turn the meter on and off.
6. Liquid Crystal Display (LCD)
7. Cuvette alignment indicator
8. Cuvette holder
DISPLAY ELEMENTS DESCRIPTION:
1. The measuring scheme (lamp, cuvette, detector), appears during
different phases of zero or reading measurement
2. Error messages and warnings
3. The battery icon indicates the charge state of the battery
4. The hourglass appears when an internal check is in progress
5. Status messages
6. The chronometer appears when the reaction timer is running
7. The month, day and date icons appear when a date is displayed
8. Four digit main display
9. Measuring units
10. Four digit secondary display
GLP mode
. In
calibration mode
measurement mode
calibration mode
, press to make a
press to edit
.
GLP mode
Errors and warnings:
ON ZERO READING:
Light High: There is too much light to perform
a measurement. Please check the preparation of
the zero cuvette.
Light Low: There is not enough light to perform
a measurement. Please check the preparation of
the zero cuvette.
No Light: The instrument cannot adjust the
light level. Please check that the sample does
not contain any debris.
ON SAMPLE READING:
Inverted cuvettes: The sample and the zero
cuvette are inverted.
Zero: A zero reading was not taken. Follow the
instructions of the measurement procedure for
zeroing the meter.
Under range: A blinking “0” indicates that
the sample absorbs less light than the zero
reference. Check the procedure and make sure
you use the same cuvette for reference (zero)
and measurement.
Over Range: A flashing value of the maximum
concentration indicates an over range condition.
The concentration of the sample is beyond the
programmed range: dilute the sample and
re-run the test.
DURING CALIBRATION PROCEDURE:
Standard Low: The standard reading is less
than expected.
Standard High: The standard reading is higher
than expected.
OTHER ERRORS AND WARNINGS:
Cap error: Appears when external light enters
in the analysis cell. Assure that the cuvette cap
is present.
Cooling lamp: The instrument waits for the
lamp to cool down.
Battery low: The battery must be replaced
soon.
Dead battery: This indicates that the battery
is dead and must be replaced. Once this
indication is displayed, normal operation of the
instrument will be interrupted. Change the
battery and restart the meter.
Measurement procedure:
Measurement
2
3
4
5-6
8-9
10
11
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the LCD
displays dashes and “P1” (Calcium) and
“P2” (Magnesium) the meter is ready. The
code that appears on the secondary display is
the one of the last selected parameter. If
necessary, press RANGE/GLP/ to change
parameter. The blinking “ZERO” indicates
that the instrument needs to be zeroed first.
3• Calcium: Using a 5 mL syringe add exactly
3 mL of sample to the cuvette. Use the
pipette to fill the cuvette up to the 10 mL
mark with the HI 93752A-0 Ca Buffer
Reagent. Add 4 drops of Buffer Reagent.
Magnesium: Using a 1 mL syringe add
exactly 1 mL of HI 93752A-0 Mg Buffer
Reagent. Use the pipette to fill the cuvette
up to the 10 mL mark with the HI 93752B-0
Mg Indicator Reagent. Replace the cap
and invert several times to mix.
4• Place the cuvette into the holder and ensure
that the notch on the cap is positioned
securely into the groove.
5• Press ZERO/CFM and the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
6• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for measurement.
7• Remove the cuvette.
8• Calcium: Using a 1 mL syringe add exactly
1 mL of the HI 93752B-0 Ca Oxalate reagent.
Magnesium: Using another 1 mL syringe,
add to the cuvette exactly 0.5 mL of sample.
Note: Do not mix up the two 1 mL syringes!
9• Calcium: Replace the cap and invert the
cuvette 10 times to mix.
Magnesium: Replace the cap and invert
several times to mix.
10•Replace the cuvette into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
11•Press READ/ /TIMER for three seconds.
The display will show the countdown prior to
measurement. The beeper is playing a beep
at the end of countdown period.
Calcium only: invert again the cuvette 10
times replace the cuvette and press READ/
/TIMER.
Alternatively wait for:
Calcium: 5 minutes, then invert again
the cuvette 10 times
Magnesium:Wait for 15 seconds
then press READ/ /TIMER.
In all cases the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
12
12•The instrument directly displays concentration in mg/L of Calcium or
Magnesium on the LCD, depending on the selected parameter.
INTERFERENCES:
• Calcium: Concentration Reading
Acidity (as CaCO3) above 1000 mg/L lower
Alkalinity (as CaCO3) above 1000 mg/L higher
Magnesium (Mg2+) above 400 mg/L higher
• Magnesium: Concentration Reading
Acidity (as CaCO3) above 1000 mg/L lower
Alkalinity (as CaCO3) above 1000 mg/L higher
Calcium (Ca2+) above 200 mg/L higher
Iron - higher
Aluminum - higher
Copper - higher
Validation and Calibration procedures
Warning: do not validate or calibrate the instrument with standard
solutions other than the Hanna CAL CHECK™ Standards, otherwise
erroneous results will be obtained.
For accurate validation and calibration results, please perform tests at
room temperature (18 to 25°C; 64.5 to 77.0°F).
Use the Hanna CAL CHECK™ cuvettes (see
“Accessories”) to validate or calibrate instruments.
VALIDATION
Note: The validation is performed only for the
selected parameter. For full validation of
the instrument, the following procedure
must be performed for each parameter.
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
3• Place the CAL CHECK™ Standard Cuvette
A into the cuvette holder and ensure that
the notch on the cap is positioned securely
into the groove.
4• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the display,
depending on the measurement phase.
5• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for validation.
6• Remove the cuvette.
7• Place the specific CAL CHECKTM Standard
Cuvette B into the cuvette holder, for:
Calcium: B, HI 96752-11
Magnesium: B, HI 96754-11
Ensure that the notch on the cap is positioned
securely into the groove.
Validation
3
4-5
7
8• Press CAL CHECK key and the lamp,
cuvette and detector icons together with
“CAL CHECK” will appear on the display,
depending on the measurement phase.
9• At the end of the measurement the display
will show the validation standard value.
The reading should be within specifications as
reported on the CAL CHECK™ Standard
Certificate. If the value is found out of
specifications, please check that the cuvettes
are free of fingerprints, oil or dirt and repeat
validation. If results are still found out of
specifications then recalibrate the instrument.
CALIBRATION
Note: It is possible to interrupt the calibration
procedure at any time by pressing
CAL CHECK or ON/OFF keys.
When calibrating, only the selected range
is affected.
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
3• To change the range, simply press
RANGE/GLP/ .
4• Press and hold CAL CHECK for three seconds
to enter
calibration mode.
show “CAL” during calibration procedure. The
blinking “ZERO” asks for instrument zeroing.
5• Place the CAL CHECK™ Standard Cuvette
A into the cuvette holder and ensure that
the notch on the cap is positioned securely
into the groove.
6• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the display,
depending on the measurement phase.
7• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for calibration. The blinking “READ”
asks for reading calibration standard.
8• Remove the cuvette.
9• Place the specific CAL CHECKTM Standard
Cuvette B into the cuvette holder, for:
Calcium: B, HI 96752-11
Magnesium: B, HI 96754-11
Ensure that the notch on the cap is positioned
securely into the groove.
10•Press READ/ /TIMER and the lamp,
cuvette and detector icons will appear on
the display, depending on the measurement
phase.
11•The instrument will show for three seconds
the CAL CHECK™ standard value.
The display will
8
9
Calibration
4
5
6-7
9
10-11
Note: If the display shows “STD HIGH”, the
standard value was too high. If the
display shows “STD LOW”, the standard
value was too low. Verify that both CAL
CHECK™ Standard Cuvettes, A and B
are free from fingerprints or dirt and
that they are inserted correctly.
12•Then the date of last calibration (e.g.:
“01.08.2009”) appears on the display, or
“01.01.2009” if the factory calibration
was selected before. In both cases the year
number is blinking, ready for date input.
13•Press RANGE/GLP/ to edit the desired
year (2009-2099). If the key is kept pressed,
the year number is automatically increased.
14•When the correct year has been set, press
ZERO/CFM or READ/ /TIMER to confirm.
Now the display will show the month blinking.
15•Press RANGE/GLP/ to edit the desired
month (01-12). If the key is kept pressed,
the month number is automatically increased.
16•When the correct month has been set, press
ZERO/CFM or READ/ /TIMER to confirm.
Now the display will show the day blinking.
17•Press RANGE/GLP/ to edit the desired
day (01-31). If the key is kept pressed,
the day number is automatically increased.
Note: It is possible to change the editing from
day to year and to month by pressing
READ/ /TIMER.
18•Press ZERO/CFM to save the calibration date.
19•The instrument displays “Stor” for one
second and the calibration is saved.
20•The instrument will return automatically
to
measurement mode
by displaying dashes
on the LCD.
GLP
In
GLP mode
, the last calibration date can be verified and the factory
calibration can be restored.
LAST CALIBRATION DATE
1• Press and hold RANGE/GLP/ for three
seconds to enter
month and day will appear on the main
display and the year on the secondary display.
2• If no calibration was performed, the factory
calibration message, “F.CAL” will appear on
the main display and the instrument returns
to
GLP mode
measurement mode
. The calibration
after three seconds.
12-14
15-16
17
18
19
1
2
or
or
Last
Calibration
Date
2
Factory
Calibration
Restore
FACTORY CALIBRATION RESTORE
It is possible to delete the calibration and restore
factory calibration.
1• Press and hold RANGE/GLP/ for three
seconds to enter
GLP mode
.
2• Press READ/ /TIMER to enter in the
factory calibration restore screen. The
instrument asks for confirmation of user
calibration delete.
3• Press ZERO/CFM to restore the factory
3-4
calibration or press RANGE/GLP/
again to abort factory calibration restore.
4• The instrument briefly indicates “donE”
upon restoration of factory calibration prior
to returning to
measurement mode
.
Battery management
To save the battery, the instrument shuts down after 10 minutes of
non-use in
measurement mode
mode.
If a valid measurement was displayed before
auto-shut off, the value is displayed when the
instrument is switched on. The blinking “ZERO”
means that a new zero has to be performed.
One fresh battery lasts for arround 750 measurements, depending on the
light level.
The remaining battery capacity is evaluated at the instrument startup
and after each measurement.
The instrument displays a battery indicator with three levels as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken any
more, the instrument shows “dEAd bAtt” and turns off.
To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
• Turn the instrument off by pressing ON/OFF.
• Turn the instrument upside down and remove the battery cover by
turning it counterclockwise.
• Extract the battery from its location and replace it with a fresh one.
• Insert back the battery cover and turn it clockwise to close.
and after 1 hour of non-use in
calibration
Accessories
REAGENT SETS
HI 93752-01 Reagents for 100 tests (50 tests Ca and 50 tests Mg)
HI 93752-03 Reagents for 300 tests (150 tests Ca and 150 tests Mg)
OTHER ACCESSORIES
HI 96752-11 CAL CHECKTM Standard Cuvettes for Calcium(1 set)
HI 96754-11 CAL CHECKTM Standard Cuvettes for Magnesium(1 set)
HI 721310 9V battery (10 pcs.)
HI 731318 Cloth for wiping cuvettes (4 pcs.)
HI 731331 Glass cuvettes (4 pcs.)
HI 731335 Caps for cuvettes
HI 93703-50 Cuvette cleaning solution (230 mL)
Warranty
HI 96752 is warranted for two years against defects in workmanship and
materials when used for its intended purpose and maintained according
to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact your dealer. If under warranty, report the
model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified
of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service Department
and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for your specific application and for
the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic equipments, this
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument
safety do not use or store the instrument in hazardous environments.
Hanna Instruments reserves the right to modify the
design, construction and appearance of its products
without advance notice.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find the Hanna Office in your area,
IST96752 11/10A
visit our web site
www.hannainst.com