Hanna Instruments HI 93828 User Manual

Instruction Manual
HI 93828
Nitrate ISM
www.hannainst.com
HI 93828 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions.
This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered. If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
This Instrument is in
Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct operation of the meter. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with directives EN 50081-1 and EN 50082-1.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If there is any damage, notify your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective item must be returned in its original packing.
GENERAL DESCRIPTION
The HI 93828 meter measures the nitrate (NO water and wastewater in the 0 to 100 mg/L (ppm) range.
The meter uses an exclusive positive-locking system to ensure that the cuvet is in the same position every time it is placed into the measurement cell.
The reagents are in powder form and are supplied in packets. The amount of reagent is precisely dosed to ensure maximum repeatability.
Display codes aid the user in routine operations. The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
) content in
3
SPECIFICATIONS
CUVET HOLDER
CUVET ALIGNMENT INDICATOR
ON/OFF
ZERO
READ TIMED
READ DIRECT
SPECIFICATIONS
Range 0 to 100 mg/L Resolution 1 mg/L Accuracy ±5 mg/L ±5% of reading Typical EMC ±1 mg/L
Deviation Light Source Light Emitting Diode @ 555 nm Method Adaptation of the cadmium reduction
method. The reaction between nitrate and the reagent causes an amber tint in the sample
Light Detector Silicon Photocell Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours Auto-Shut off After 10' of non-use Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8") Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity
HI 93728-0 Powder reagent 1 packet
REAGENT SETS
HI 93728-01 Reagents for 100 tests HI 93728-03 Reagents for 300 tests
DISPLAY CODE GUIDE
This indicates that the meter is in a ready state and zeroing can be performed.
Sampling in Progress. This prompt appears each time the meter is performing a measurement.
This indicates that the meter is in a zeroed state and measurement can be performed.
A zero reading was not taken. Insert a sample before adding reagent and press ZERO.
Under range. A blinking "0.00" indicates that the sample absorbs less light than the zero reference. Check the procedure and make sure you use the same cuvet for reference (zero) and measurement.
Over range. A flashing value higher than the maximum concentration readable (see specifications) indicates that the sample absorbs too much light, meaning that the concentration is too high. Dilute the sample.
Light over range. The cuvet is not inserted correctly and an excess ambient light is reaching the detector. If the cover is properly installed, then contact your dealer or the nearest Hanna Customer Service Center.
Light under range. The zero sample is too dark for proper zeroing. If this is not the case, contact your dealer or the nearest Hanna Customer Service Center.
V
The "V" indicates that the battery voltage is getting low and the battery needs to be replaced.
This indicates that the battery is dead and must be replaced.
Note: once this indication is displayed, the meter will lockup. Change the battery to restart.
Loading...