Hanna Instruments HI 93748 User Manual

Instruction Manual
HI 93748
Manganese Low Range
ISM
http://www.hannainst.com
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
HI 93748 is warranted for two years against defects in workman­ship and materials when used for its intended purpose and maintained according to instructions.
This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered. If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur­chase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service de­partment and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protec­tion. To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instru­ments Inc., Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
This Instrument is in
Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct operation of the meter. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with
directives
EN 50081-1 and EN 50082-1.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If there is any damage, notify your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective item must be returned in its original packing.
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
The HI 93748 meter measures the manganese content in water and wastewater in the 0 to 300 µg/L (ppb) range.
The meter uses an exclusive positive-locking system to ensure that the cuvet is in the same place every time it is placed into the measurement cell.
The reagents are in liquid and powder form and are supplied in bottles and in packets. The amount of reagent is precisely dosed to ensure maximum repeatability.
Display codes aid the user in routine operations. The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
s
ins
rum n
HI 93748
READ
DIRECT
Range: 0 to 300 ppb
ON
OFF
MANGANESE LOW RANGE
ZERO
READ
TIMED
e
CUVET HOLDER
CUVET ALIGNMENT INDICATOR
ON/OFF
ZERO
READ TIMED
READ DIRECT
SPECIFICATIONS
Range 0 to 300 µg/L Resolution 1 µg/L Accuracy ±2 µg/L ±3% of reading Typical EMC ±1 µg/L
Deviation Light Source Light Emitting Diode @ 555 nm Method Adaptation of the PAN Method. The reac-
tion between manganese and the reagents causes an orange tint in the sample
Light Detector Silicon Photocell Environment 0 to 50C (32 to 122F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours Auto-Shut off After 10' of non-use Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8") Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity
HI 93748A-0 Ascorbic acid 2 packets HI 93748B-0 Alkaline-cyanide sol. 0.40 mL HI 93748C-0 0.1% PAN indicator 2 mL HI 93703-51 Dispersing Agent 4-6 drops (only
when necessary, see note)
REAGENT SETS
HI 93748-01 Reagents for manganese LR (100 pcs of
HI93748A-0, 20 mL of HI93748B-0, 100 mL of HI93748C-0, 20 mL of HI93703-51)
HI 93748-03 Reagents for manganese LR (300 pcs of
HI93748A-0, 60 mL of HI93748B-0, 300 mL of HI93748C-0, 60 mL of HI93703-51).
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
This indicates that the meter is in a ready state and zeroing can be per­formed.
Sampling in Progress. This prompt ap­pears each time the meter is performing a measurement.
This indicates that the meter is in a zeroed state and measurement can be performed.
A zero reading was not taken. Insert a sample before adding reagent and press ZERO.
Under range. A blinking "0.00" indicates that the sample absorbs less light than the zero reference. Check the procedure and make sure you use the same cuvet for reference (zero) and measurement.
Over range. A flashing value higher than the maximum concentration readable (see specifications) indicates that the sample absorbs too much light, meaning that the concentration is too high. Dilute the sample.
Light over range. The cuvet is not in­serted correctly and an excess ambient light is reaching the detector. If the cover is properly installed, then contact your dealer or the nearest Hanna Customer Service Center.
Light under range. The zero sample is too dark for proper zeroing. If this is not the case, contact your dealer or the nearest Hanna Customer Service Center.
V
The "V" indicates that the battery volt­age is getting low and the battery needs to be replaced.
This indicates that the battery is dead and must be replaced.
Note: once this indication is displayed, the meter will lockup. Change the bat­tery to restart.
OPERATIONAL GUIDEOPERATIONAL GUIDE
m
m
9V
SLIDE
OFF
OPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
ON
OFF
Note: for best results perform your tests between 20-24C.
• When the LCD displays "- - -", it is ready.
• Fill one cuvet up to 1.5 cm (¾")
1.5 c
below the rim with 10 mL of deion­ized water.
1.5 c
• Fill a second cuvet up to 1.5 cm (¾") below the rim with 10 mL of sample.
A
8 4 7 3
I9
• Add the content of one packet of
H
HI 93748A ascorbic acid to each cuvet, replace the caps and shake gently.
• Add 0.2 mL of the HI 93748B alkaline-cya­nide reagent solution to each cuvet, replace the
B
caps and shake gen­tly.
• Add 1 mL of the HI 93748C 0.1% PAN in­dicator solution to each cuvet, replace the caps
C
and shake gently.
• Place the cuvet with the re­acted deionized water (blank) into the holder and ensure that the notch on the cap is positioned securely into the groove.
• Press READ TIMED and the display will show the count­down prior to zeroing the blank. Alternatively wait for 2 minutes and press ZERO. In both cases "SIP" will appear during zeroing.
READ
TIMED
ZERO
2’
12
9
3
6
• Wait for a few seconds and the dis­play will show "-0.0-". Now the meter is zeroed and ready for measure­ment.
• Insert the second cuvet with the reacted sample into the instrument.
• Press READ DIRECT. "SIP" will appear during measure­ment.
READ
DIRECT
• The instrument directly displays concentration in µg/L of manganese on the display.
Note: a temperature above 30C may cause turbidity. In
this case, before zeroing and taking readings, add 2-3 drops of Dispersing Agent (HI93703-51) to each cuvet and swirl until complete dissolution of the turbidity.
INTERFERENCES
Interference may be caused by: Aluminum above 20 mg/L Cadmium above 10 mg/L Calcium above 200 mg/L as CaCO Cobalt above 20 mg/L Copper above 50 mg/L Iron above 10 mg/L Lead above 0.5 mg/L Magnesium above 100 mg/L as CaCO Nickel above 40 mg/L Zinc above 15 mg/L.
3
3
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENT
MEASUREMENTMEASUREMENT
The instruction listed below should be carefully followed during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the zeroing and the measuring phases, it is necessary to close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell, it must be completely free of fingerprints, oil or dirt. Wipe it thoroughly with HI 731318 or a lint-free cloth prior to insertion.
• It is important that the sample does not contain any debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it is recommended that a zero reading be taken for each sample and that the same cuvet is used for zeroing and measurement.
• It is important to discard the sample immediately after the reading is taken because the glass might become permanently stained.
• Shaking the cuvet can generate bubbles in the sample, causing higher readings. To obtain accurate measure­ments, remove such bubbles by swirling or by gently tapping the vial.
• All the reaction times reported in this manual are referred to 20C (68F). As a general rule of thumb, they should be doubled at 10C (50F) and halved at 30C (86F).
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard­ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter. Detach the battery from the terminals and attach a fresh 9V battery while paying attention to the correct polarity. Replace the battery and the cover.
ACCESSORIESACCESSORIES
ACCESSORIES
ACCESSORIESACCESSORIES
REAGENT SETS
HI 93748-01 Reagents for manganese LR (100 pcs of
HI93748A-0, 20 mL of HI93748B-0, 100 mL of HI93748C-0, 20 mL of HI93703-51)
HI 93748-03 Reagents for manganese LR (300 pcs of
HI93748A-0, 60 mL of HI93748B-0, 300 mL of HI93748C-0, 60 mL of HI93703-51)
OTHER ACCESSORIES
HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery (10 pcs) HI 731318 Tissue for wiping cuvets (4 pcs) HI 731321 Glass cuvets (4 pcs) HI 731325 Caps for cuvets (4 pcs) HI 93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATIONCE DECLARATION
CE DECLARATION
CE DECLARATIONCE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
OF CONFORMITYOF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environ­ment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.