Instruction Manual
HI 93701
Free Chlorine ISM
www.hannainst.com
Warranty
HI 93701 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained
according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
This Instrument is in
Compliance with the CE Directives
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with directives.
Preliminary Examination
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
General Description
The HI 93701 portable microprocessor meter measures the
free chlorine (Cl2) content in water and wastewater in the
0.00 to 2.50 mg/L (ppm) range.
The meter uses an exclusive positive-locking system to ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid or powder form and are supplied
in bottles or in packets. The amount of reagent is precisely
dosed to ensure maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
Specifications Display Code Guide
This indicates that the meter is in a
ready state and zeroing can be performed.
Sampling in Progress. This prompt appears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
SPECIFICATIONS
Range 0.00 to 2.50 mg/L
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading
Typical EMC
Deviation ±0.01 mg/L
Light Source Light Emitting Diode @ 555 nm
Method Adaptation of the EPA recommended
DPD method 330.5. The reaction between
free chlorine and the DPD reagent causes
a pink tint in the sample
Light Detector Silicon Photocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 80 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.)
REQUIRED REAGENTS
POWDER:
Code Description Quantity
HI 93701-0 DPD 1 packet
LIQUID:
Code Description Quantity
HI 93701A-F DPD1 Indicator 3 drops
HI 93701B-F DPD1 Buffer 3 drops
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
the maximum readable concentration (see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not inserted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery voltage is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the battery to restart.
Operational Guide Tips for an Accurate Measurement
• The instrument directly displays concentration in mg/L of
free chlorine on the Liquid Crystal Display.
MEASUREMENT PROCEDURE
• Turn the meter on by pressing
ON/OFF.
• When the LCD displays “- - -”, it
is ready.
• Fill the cuvet with 10 mL of
unreacted sample, up to the
mark, and replace the cap.
10 mL
ON
OFF
Liquid reagents procedure
• To an empty cuvet add
3 drops of HI 93701A-F
DPD1 indicator, 3 drops
of HI 93701B-F DPD1
buffer. Swirl gently to
mix and immediately add
10 mL of unreacted
sample.
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
• Place the cuvet into the holder
and ensure that the notch on
the cap is positioned securely into
the groove.
• Press ZERO and “SIP” will appear on the display.
• Replace the cap and
shake gently again.
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
ZERO
• Reinsert the cuvet into
the instrument.
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Wait for a few seconds and the
display will show “-0.0-”. Now
the meter is zeroed and ready for
measurement.
• Remove the cuvet.
• Wait for 1 minute and then press READ DIRECT. The
display will show “SIP” during measurements.
READ
DIRECT
• Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measurements, remove such bubbles by swirling or by gently
tapping the vial.
• All the reaction times reported in this manual are
referred to 20°C (68°F). As a general rule of thumb,
Powder reagent procedure
• Add the content of one packet of
• The instrument directly displays concentration in mg/L of
free chlorine on the Liquid Crystal Display.
they should be doubled at 10°C (50°F) and halved at
30°C (86°F).
HI 93701 DPD reagent. Replace
the cap and shake gently for 20
seconds (or 2 minutes in case of
seawater analysis).
• Replace the cuvet into the holder
and ensure that the notch on
the cap is positioned securely
into the groove.
• Wait for 1 minute and press READ DIRECT. The display
will show “SIP” during measurements.
INTERFERENCES
Interference may be caused by:
Bromine Iodine
Chlorine dioxide Ozone
Hydrogen peroxide
Oxidized Chromium (Cr6+) and Manganese (Mn4+, Mn7+).
Alkalinity above 250 mg/L or acidity above 150 mg/L will
not reliably develop the full amount of color or it may
rapidly fade. To resolve this, neutralize the sample with
Battery Replacement
Battery replacement must only take place in a non-hazardous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
diluted HCl or NaOH.
READ
DIRECT
In case of water with hardness greater than 500 mg/L
CaCO3, shake the sample for approximately 1 minute after
adding the powder reagent.
Accessories
REAGENT SETS
HI 93701-F Reagents for 300 tests (liquid)
HI 93701-01 Reagents for 100 tests (powder)
HI 93701-03 Reagents for 300 tests (powder)
OTHER ACCESSORIES
HI 710009 Blue rubber boot
HI 710010 Orange rubber boot
HI 721310 9V battery (10 pcs)
HI 731318 Tissue for wiping cuvets (4 pcs)
HI 731321 Glass cuvets (4 pcs)
HI 731325 Caps for cuvets (4 pcs)
HI 93703-50 Cuvets cleaning solution (230 mL)
CE Declaration of Conformity
Recommendations for Users
Before using these products, make sure that they are entirely suitable for
the environment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to
take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may
degrade the instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.