Dear Customer,
Thank you for choosing a Hanna product.
Please read this instruction manual carefully before using the instru-
ment.
This manual will provide you with the necessary information for correct
use of the instrument, as well as a more precise idea of its versatility.
If you need additional technical information, do not hesitate to e-mail
us at tech@hannainst.com.
These instruments are in compliance with the directives.
WARRANTY
All Hanna Instruments meters are warranted for two years against defects in
workmanship and materials when used for their intended purpose and maintained
according to instructions. Theprobes are warranted for a period of six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accidents, misuse, tampering or lack of prescribed maintenance are
not covered.
If service is required, contact the dealer from whom you purchased the instrument. If
under warranty, report the model number, date of purchase, serial number and the
nature of the failure.
First obtain a Returned Goods Authorization number from the Customer Service department, then return the instrument with the Authorization number included along with
shipment costs prepaid.
If the repair is not covered by the warranty, you will be notified of the charges.
When shipping any instrument, make sure it is properly packaged for complete
CE DECLARATION OF CONFORMITY.............................................. 14
TEMPERATURE PROBES .............................................................. 15
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to
make sure that no damage has occurred during shipping. If there is
any damage, notify your dealer or the nearest Hanna Customer
Service Center.
Each meter is supplied complete with:
• Temperature probe
• 3 x 1.5V AA alkaline batteries
• Instruction manual.
Note: Save all packing material until you are sure that the
instrument functions correctly. All defective items must be
returned in the original packing with the supplied accessories.
GENERAL DESCRIPTION
HI 93510, HI 93510N, HI 93512 and HI 93522 are powerful and
flexible thermistor thermometers, which have been designed using the
latest microprocessor technology to provide reliable and accurate
measurements.
Standard features include waterproof casing, dual-level LCD, low
battery detection, long battery life and two-year warranty.
HI 93510 has been designed for laboratory and field use, while HI
93510N is suitable for high accuracy (meter and probe can be
calibrated in an ice bath at 0°C) or dimly lit areas (the backlight
feature can be easily activated through the keyboard).
HI 93512 is a two-channel thermometer, ideal for monitoring two
samples at once, while HI 93522 offers all the features of HI 93512
and more: calibration of meter and probe at 0°C, backlight, userselectable auto-off time period, and capability to store and recall a
reading.
32
SPECIFICATIONS OF
HI 93510 & HI 93510N
SPECIFICATIONS
Range-50.0 to 150.0°C
-58.0 to 302.0°F
Resolution0.1°C
0.1°F (-58.0 to 230.0°F) / 0.2°F (outside)
Accuracy±0.4°C / ±0.8°F
for one year, excluding probe error
Typical EMC ±0.5°C / ±1°F
Deviation with HI 762BL probe
Environment-10 to 50°C (14 to 122°F); RH 100%
Dimensions150 x 80 x 36 mm (5.9 x 3.1 x 1.4")
Weight235 g (8.3 oz.)
Main features:
• HI 762 probes series
• Available HI 762 accuracy test plugs
• Display of temperature in degrees Celsius or Fahrenheit
• Range -50 to 150°C or -58 to 302°F
• HOLD function
• HI and LO (Max and Min) temperature values always displayed
on the LCD
• Remaining battery life indication / low battery detection
• Waterproof casing
• Auto-off capability (HI 93510 only)
• Backlight feature (HI 93510N only)
• User calibration at 0°C (HI 93510N only)
HI 93510HI 93510N
Keyboard Functions:
ON/OFF : turn the meter ON and OFF.
HOLD : freeze the reading on display.
°C/°F : change reading unit (°C or °F).
CLR : clear the HI and LO values.
BL (HI 93510N only) : toggle the backlight ON and OFF.
CAL (HI 93510N only) : press and hold for 5 seconds to enter the
Calibration mode (with reading within ±3°C range); press to
abort the Calibration mode.
54
SPECIFICATIONS OF
HI 93512 & HI 93522
SPECIFICATIONS
Range-50.0 to 150.0°C
-58.0 to 302.0°F
Resolution0.1°C
0.1°F (-58.0 to 230.0°F) / 0.2°F (outside)
Accuracy±0.4°C / ±0.8°F
for one year, excluding probe error
Typical EMC ±0.5°C / ±1°F
Deviation with HI 762BL probe
Environment-10 to 50°C (14 to 122°F); RH 100%
Dimensions150 x 80 x 36 mm (5.9 x 3.1 x 1.4")
Weight235 g (8.3 oz.)
HI 93512HI 93522
Main features:
• Two independent measurement channels, T1 and T2
• HI 762 probes series
• Available HI 762 accuracy test plugs
• Display of temperature in degrees Celsius or Fahrenheit
• Range -50 to 150°C or -58 to 302°F
• HOLD function
• HI and LO (Max and Min) temperature values always displayed
on the LCD
• DISP function to select the information to be displayed (Normal,
Relative or T1/T2)
• Remaining battery life indication / low battery detection
• Waterproof casing
• Auto-off capability (HI 93522 only)
• MR function for recalling memorized reading (HI 93522 only)
• Backlight feature (HI 93522 only)
• User calibration at 0°C (HI 93522 only)
Keyboard Functions:
ON/OFF : turn the meter ON and OFF.
T1/T2 : select the reading channel (T1, T2 or T1-T2).
HOLD : freeze the reading on display. In HI 93522, the frozen value
is also stored in non-volatile memory.
CLR : clear the HI and LO values, reset the relative measurement.
DISP : select Normal, Relative or T1/T2 measuring mode.
ALT (HI 93522 only) : enable the second function keys; the "ALT" tag
turns on to indicate that the second functions are enabled.
Note: The ALT key can be released before pressing the second
function key for 1-hand operation.
ALT/ CAL : press and hold for about 5 seconds to enter the
Calibration mode (with reading within ±3°C range).
(ALT +) AOFF : set auto power off delay (8min, 60min, disabled).
(ALT +) °C/°F : change reading units (°C or °F).
(ALT +) MR : recall memorized value.
(ALT +) BL : toggle the backlight ON and OFF.
76
OPERATIONAL GUIDE
To switch the instrument ON, press the ON/OFF key.
The thermometer will carry out a self diagnostic test routine, the LCD
will show all segments for a few seconds (or as long as ON/OFF is
held), followed by the percentage indication of the remaining battery
life.
Note: Although the display is frozen, internally the meter continues
measuring and updating Hi/Lo and relative values.
In HI 93522 only, the held value is also stored in non-volatile
memory and can be recalled by pressing (ALT+) MR key. While the
MR key is held, the "HOLD" tag lights up and the meter displays the
stored value; when the MR key is released, after 1 second the meter
returns to normal measurement mode.
The thermometer then enters normal measurement mode.
If a temperature probe is plugged in, the meter displays the
measured temperature.
If no probe is plugged in, or if the reading is
over-range, the display shows flashing dashes.
If a measurement is slightly over the range
of the meter specifications, the display will
flash the closest full-scale value.
To switch the meter OFF, press the ON/OFF key.
°C/°F SELECTION
Measurements can be displayed in either degrees Celsius or Fahrenheit. The meter is factory set to °C scale; to change the scale, press
°C/°F key or (ALT+) °C/°F keys,
HI 93510 / 510N
HI 93522
or set the switch located in the battery compartment (HI 93512 only, see
figure on page 10).
HOLD and MR FUNCTIONS
The HOLD function is activated by pressing the HOLD key.
The measured temperature is held on the
display until HOLD is pressed again. The
"HOLD" tag blinks on the display while in
HOLD mode.
HIGH/LOW TEMPERATURES
The maximum and minimum temperatures are continuously monitored and
displayed on the lower portion of the LCD.
Note: When the reading goes over-range or the probe is removed,
the Hi and Lo values display dashes until cleared.
CLEAR FUNCTION
Upon pressing the CLR key, the High/Low values may be cleared at
any time during measurement and the current reading is assigned to
the highest and lowest temperature values for the displayed channel
only.
AUTO SHUT-OFF (HI 93510 and HI 93522)
To save battery life, the HI 93510 and HI 93522 models are
provided with an auto-off feature, which switches the instrument off
after a certain period of nonuse.
The auto-off delay time can be changed by the user (8 minutes, 60
minutes, or disabled).
To access this feature, the HI 93510 is provided with an internal
button located in the battery compartment (see figure on page 10).
Unscrew the 4 screws on the back of the meter and select the desired
auto-off time period by pressing the button and viewing the value on
the LCD. Reattach the back making sure that the gasket is in place
and tighten the 4 screws to ensure a watertight seal.
98
HI 93522 allows the user to select the auto-off time period through
the keyboard; press (ALT+) AOFF to enter the mode, then set the
desired auto shut-off interval (8 min, 60 min, or disabled) with the
ALT key; release all buttons, wait a few seconds and the meter will
return to normal measurement mode.
2-CHANNEL MODELS (HI 93512 and HI 93522)
HI 93512 and HI 93522 models can monitor two samples through
two independent temperature channels (probes).
The display will show the actual, Hi and Lo (or T1 and T2) values of
the selected channel (T1, T2 or T1-T2). The corresponding tag will
light up to inform the user.
To select the desired channel, use the T1/T2 key.
BACKLIGHT FEATURE (HI 93510N and HI 93522)
HI 93510N and HI 93522 models are provided with a Backlight
feature, which can be easily activated through the keyboard by
pressing the BL key or (ALT+)BL keys.
HI 93510N
HI 93522
Note: The backlight automatically shuts off after approximately 1
minute with no buttons pressed.
BATTERY COMPARTMENT
The 2-channel models are also provided with the DISP function, which
allows the user to select the information to be displayed.
• While in T1 or T2, pressing DISP will switch the display between
standard and relative measurement modes.
In the standard mode, the main body of the LCD shows the
current temperature while the lower portion displays the Hi/Lo
limits for that channel.
When the meter enters relative mode, the current temperature for
that channel is set as the reference temperature. Pressing CLR will
also set the reference to the current temperature.
In relative mode, the main body of the LCD shows the difference
from the reference temperature. The lower portion displays the
current temperature for the selected channel.
• While in T1-T2, pressing DISP will switch between 3 displays:
Note: Pressing T1/T2 to change the channel will not alter the
information chosen to be displayed for each channel (DISP).
Note: Relative mode can not be entered if dashes are shown in the
main portion of the LCD.
1110
BATTERY REPLACEMENTUSER CALIBRATION
The HI 93510N and HI 93522 models can be calibrated at 0°C by
using an ice bath.
• Prepare an ice bath with approximately equal volumes of distilled
water and chopped ice made from distilled water.
• Immerse the temperature probe in the center of the ice bath,
taking care not to touch the ice with the probe tip.
• Ensure that the meter is measuring a temperature within ±3°C.
• To enter the Calibration mode, press and hold the CAL button for
about 5 seconds.
• The CAL tag turns on to indicate that
the Calibration mode has been entered.
Note: If the measurement is outside the
±3°C window, the meter does not
enter Calibration mode.
• When the meter reaches the stability condition, which is detected
when the measurement remains within ±0.2°C for 5 seconds, the
calibration is accepted and the reading becomes 0°C (32°F).
• The meter then automatically returns to normal mode.
Note: To exit the Calibration mode at any time, press the CAL key.
Note: User calibration cannot be entered in relative or in T1-T2 mode.
Note: User calibration is only performed on the current channel
displayed (T1 or T2).
FACTORY RECALIBRATION
When the battery level is below 10%, a warning
symbol will blink on the LCD to indicate a low
battery condition.
If the battery level is low enough to cause erroneous readings, the
Battery Error Prevention System (BEPS) turns the meter off.
Immediately replace the batteries with new ones.
The batteries are accessed by separating the front and the back
halves of the meter: unscrew the 4 screws on the back of the meter
and carefully replace the three batteries located in the battery
compartment, while paying attention to their polarity. Reattach the
back making sure that the gasket is in place and tighten the 4
screws to ensure a watertight seal.
Battery replacement must only take place in a non-hazardous area
using 1.5V AA (IEC LR6) alkaline batteries.
All Hanna thermometers have been accurately pre-calibrated at the
factory. It is generally recommended to have all thermometers recalibrated at least once a year. For an accurate recalibration, contact your
nearest Hanna Customer Service Center.
Hanna Accuracy Test Plugs provide a quick and easy way to check the
meter's accuracy by simply connecting these Test Plugs to the probe
connector of the meter. If the reading differs by more than ±0.4°C
(±0.8°F) from the Test Plug value, the meter is due for recalibration.
Choose the right Test Plug to suit your needs:
HI 762-18CCalibration test key, -18.0°C
HI 762000CCalibration test key, 0.0°C
HI 762070CCalibration test key, 70.0°C
HI 762-004F Calibration test key, -0.4°F
HI 762032FCalibration test key, 32.0°F
HI 762158FCalibration test key, 158.0°F
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the
environment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary
steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24VAC or 60 VDC.
To avoid damages or burns, do not perform any measurement in microwave ovens.
Note: To clean the meters, do not use aggressive detergents. It is
recommended to use water.
1312
CE DECLARATION OF CONFORMITY
All rights are reserved. Reproduction in whole or in part is prohibited without the
written consent of the copyright owner, Hanna Instruments.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
TEMPERATURE PROBES
Hanna offers a wide range of probes to meet your needs with air,
liquid, penetration and surface measurements.
All Hanna temperature probes are supplied pre-calibrated from the
factory and are ready to be used with your meter.
These interchangeable probes make it possible for you to switch from
one probe to another without wasting time and money by going
through tedious calibration procedures.
The Hanna probes are also available in different handle colors to
avoid cross contamination during testing:
HI 762AAir probe, w/1 m (3.3') cable & white handle
HI 762A/10Air probe, w/10 m (33') cable & white handle
HI 762BLGeneral purpose liquid probe, w/1 m (3.3') cable &
black handle
HI 762BL/10General purpose liquid probe, w/10 m (33') cable &
black handle
HI 762LGeneral purpose liquid probe, w/1 m (3.3') cable &
white handle
HI 762L/2General purpose liquid probe, w/2 m (6.6') cable &
white handle
HI 762L/10General purpose liquid probe, w/10 m (33') cable &
white handle
HI 762PBLGeneral purpose penetration probe w/1 m (3.3') cable
& blue handle
HI 762PBL/10 General purpose penetration probe w/10 m (33') cable
& blue handle
HI 762PGGeneral purpose penetration probe with 1 m
(3.3') cable and green handle
HI 762PG/10General purpose penetration probe w/10 m (33') cable
& green handle
HI 762PRGeneral purpose penetration probe w/1 m (3.3') cable
& red handle
HI 762PR/10General purpose penetration probe w/10 m (33') cable
& red handle
HI 762PWGeneral purpose penetration probe w/1 m (3.3') cable
& white handle
HI 762PW/10General purpose penetration probe w/10 m (33') cable
& white handle
HI 762PWLPenetration probe w/1 m (3.3') cable & white handle
HI 762WWire probe without handle for hard-to-reach places
w/1 m (3.3') cable
HI 762W/10Wire probe, without handle for hard-to-reach places
w/10 m (33') cable
1514
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and
Technical offices, please see www.hannainst.com
16
MANL93R3 01/03