Code Valid Values Default PW
14 Relay 1 deviation (D1) 0.5 to 10% f.s. 1% f.s. no
15 Relay 1 reset time 0.1 to 999.9 minutes 999.9 no
16 Relay 1 rate time 0.0 to 999.9 minutes 0.0 no
21 Relay 2 mode (M2) same as relay 1 0 no
22 Relay 2 setpoint (S2) 0.5 to 99.5% full scale 75% f.s. no
23 Relay 2 hysteresis (H2) 0 to 5% f.s. 1% f.s. no
24 Relay 2 deviation (D2) 0.5 to 10% f.s. 1% f.s. no
25 Relay 2 reset time 0.1 to 999.9 minutes 999.9 no
26 Relay 2 rate time 0.0 to 999.9 minutes 0.0 no
30 Relay 3 high alarm (HA) 0.5 to 99.5% full scale 95% f.s. no
HA-HysLA+Hys,Hys=1.5%f.s.,HAS1 or HAS2
31 Relay 3 low alarm (LA) 0.5 to 99.5% full scale 5% f.s. no
LA+HysHA-Hys,Hys=1.5%f.s.,LAS1 or LAS2
32 Proportional control 1 to 30 min 5 no
mode period
33 Maximum relay ON time 10 to 9999 min 60 no
(after which an alarm mode is entered)
34 Alarm mask time 00:00 to 30:00 00:00 no
40 Analog output selection 0: 0-1mA 2 no
1: 0-20 mA
2: 4-20 mA
3: 0-5 VDC
4: 1-5 VDC
5: 0-10 VDC
41 Analog output 0 to 100% full scale 0 no
lower limit (O_VARMIN) (O_VARMIN O_VARMAX - 5% f.s.)
42 Analog output 0 to 100% full scale 100% f.s. no
upper limit (O_VARMAX) (O_VARMIN O_VARMAX - 5% f.s.)
• After confirmation, the selected parameter
is displayed. The user can scroll through
the parameters by pressing CFM.
In order to directly set another parameter, press SETUP again and enter the
code or scroll to it by pressing CFM.
The following table lists the setup codes along with the description of the specific setup items, their valid values and whether
password is required to view that item (“PW” column):
Code Valid Values Default PW
00 Factory ID 0 to 9999 0000 no
01 Process ID 0 to 9 9 00 no
02 Control enable/disable 0: C.M. disabled 0 no
1: C.M. enabled
03 Range 1: 0.0-199.9 μS (or 100.0 ppm) 4 no
(depends on model) 2: 0-1999 μS (or 1000 ppm)
3: 0.00-19.99 mS (or 10.00 ppt)
4: 0.0-199.9 mS (or 100.0 ppt)
04 Reference Temperature 20°C or 25°C 25°C no
05 Temperature Coefficient 0.00 to 10.00 %/°C 2.00 no
06 Input Selection 0: conductivity probe 0 no
1: 4-20 mA input signal
07 Temperature ATC: Automatic ATC no
compensation User: Manual
08 TDS Factor (HI710 only) 0.00 to 1.00 0.50 no
11 Relay 1 mode 0: disabled 0 no
(M1) 1: ON-OFF high setpoint
2: ON-OFF low setpoint
3: PID, high setpoint
4: PID, low setpoint
12 Relay 1 setpoint (S1) 0.5 to 99.5% full scale 25% f.s. no
13 Relay 1 hysteresis (H1) 0 to 5% f.s. 1% f.s. no