Hanna Instruments HI 504 User Manual

HI 504
Panel-mounted, Microprocessor-based pH/ORP Process Controller
Instruction Manual
TABLE OF CONTENTS
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
PRELIMINARY EXAMINATION . . . . . . . . . . . . . . . . . . 6
GENERAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . 6
FUNCTIONAL DESCRIPTION . . . . . . . . . . . . . . . . . . 8
SPECIFICATIONS. . . . . . . . . . . . . . . . . . . . . . . . . . 10
MECHANICAL DIMENSIONS . . . . . . . . . . . . . . . . . 11
INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
CALIBRATION MODE . . . . . . . . . . . . . . . . . . . . . . 14
SETUP MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
CONTROL MODE . . . . . . . . . . . . . . . . . . . . . . . . . 42
IN-LINE CLEANING . . . . . . . . . . . . . . . . . . . . . . . 51
IDLE MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
HOLD MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
ANALOG OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . 56
PC COMMUNICATION . . . . . . . . . . . . . . . . . . . . . 58
SHORT MESSAGING SERVICE (SMS) . . . . . . . . . . . . 71
MODEM CONNECTION . . . . . . . . . . . . . . . . . . . . 79
pH/ORP PROBE CHECK . . . . . . . . . . . . . . . . . . . . 81
SOLUTION COMPENSATION . . . . . . . . . . . . . . . . . 82
TEMPERATURE COMPENSATION . . . . . . . . . . . . . . 83
LAST CALIBRATION DATA . . . . . . . . . . . . . . . . . . . . 85
OFFSET AND SLOPE DIRECT SELECTION . . . . . . . . 87
EVENT LOG FILE SCROLLING . . . . . . . . . . . . . . . . 89
FAULT CONDITIONS . . . . . . . . . . . . . . . . . . . . . . . 91
SELFTEST PROCEDURES . . . . . . . . . . . . . . . . . . . . . 92
ALARM - ERROR CONFIGURATION . . . . . . . . . . . . 100
pH VALUES AT VARIOUS TEMPERATURES . . . . . . . . 103
ELECTRODE CONDITIONING AND MAINTENANCE . 104
DEFINITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . . . . 108
CE DECLARATION OF CONFORMITY . . . . . . . . . . 113
3
WARRANTY
All Hanna Instruments meters are guaranteed for two years (sensors, electrodes and probes for six months)
against defects in workmanship and materials when used for their intended purpose and maintained according to instruc­tions. This warranty is limited to repair or replacement free of charge.
Damage due to accident, misuse, tampering or lack of pre­scribed maintenance are not covered.
If service is required, contact the dealer from whom you pur­chased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Customer Service de­partment and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without ad­vance notice.
MODEL IDENTIFICATION
Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the
instrument. It will provide you with the necessary information for correct use of the instrument, as well as a precise idea of its versatility.
If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
This instrument is in compliance with directives.
The models HI 504XYZ-α are pH/ORP controllers. The meaning of the last letters is according to the following
scheme:
© 2003 Hanna Instruments
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., 584 Park East Drive, Woonsocket, Rhode Island, 02895 , USA.
5
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and exam­ine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer or the nearest Hanna Customer Service Center immediately.
Note Save all packing materials until you are sure that the instru-
ment functions correctly. Any damaged or defective items must be returned in their original packing materials together with the supplied accessories.
GENERAL DESCRIPTION
The product is a real time microprocessor-based pH/ORP controller. It provides accurate measurements, flexible control capabilities and dual alarm signals. The system is composed of a case inside which the signal conversion circuitry, the microprocessor circuitry and the output power drivers are contained.
MAIN FEATURES
• Configurations: for measure and control of pH or ORP, with two separate sets of control settings. Thus no loss of settings occurs when changing from pH to ORP and vice versa.
• Display: large LCD with 4 ½ 17 mm digits and 3 ½ 10 mm digits.
• LEDs: a green LED (alarm relay) and a red LED for signaling the device status.
• Alarm relay: one output relay for alarm condition (fail safe mode: COM and NC are connected).
• Daily programmable control timing.
• Diagnostic features.
• pH/ORP probe check.
• Temperature sensor (Pt100 or Pt1000 type) with automatic recognition and damage test.
• Hold management, including hold digital input to enter hold mode with an external trigger.
• Logging of the last 100 error, configuration, calibration and cleaning events.
6
• Calibration and Setup procedures are password protected.
• Calibration: pH calibration at 2 points with two possible sets of buffers (either 7.01 - 4.01 - 10.01 or 6.86 - 4.01 - 9.18), or at 1 point with user-selectable value.
• Temperature compensation of the HANNA standard buffers.
• GLP features: old probe check, dead probe check, calibra­tion time-out, last calibration data internally recorded (non­volatile EEPROM memory): calibration date and time, pH offset, pH slope, number of calibration points and corre­spondent pH values.
• Solution compensation of the pH reading.
• Manual temperature setting when the temperature probe is not inserted or temperature is out of range.
• Input: pH/ORP electrode with BNC connector.
• Input: pH/ORP/Temperature Digital Transmitter (HI
504910) which can be connected up to 1.2 Km (0.5 miles) far away.
• Output (0-20 or 4-20 mA): one analog output for record­ing of pH/ORP or controlling (only for models with PID).
• RS485 communication with Short Messaging Service (SMS) for GSM phones.
• Data logging: 6000 samples pH/°C or ORP downloadable through RS485 and HI92500 application software.
• Real time clock.
OPTIONAL FEATURES (depending on model)
• LEDs: four yellow LEDs are provided for signaling the ener­gizing of relay 1, relay 2 (optional), relay 3 (optional) and relay 4 (optional).
• Relays: 1, 2 or 4 output relays. The relays LEDs are switched ON whenever the correspondent relay is energized (NO and COM connected).
• Advanced Cleaning is provided through two cleaning commands and variety of trigger events (relays 3 and 4 only).
• Output (0-20 or 4-20 mA): second independent analog output for recording of temperature or controlling (only for models with PID).
• PID control.
7
FUNCTIONAL DESCRIPTION
FRONT PANEL
1. Liquid Crystal Display
2. LCD key enters and exits the event scrolling mode. During pH calibration, alternately displays pH buffer value or current temperature
3. SETUP key enters and exits setup mode
4. CAL DATA key last calibration data viewing (enters and exits)
5. CAL key initiates and exits calibration mode
6. key increases the blinking digit/letter by one when selecting a parameter. Advances forward while in last calibration data viewing mode or event scrolling mode. Increases the tem­perature setting when temperature probe is not inserted
7. key decreases the blinking digit/letter by one when selecting a parameter. Reverts backward while in last calibration data viewing mode or event scrolling mode. Decreases the temperature setting when temperature probe is not inserted
8. key moves to the next digit/letter (circular buffer) when select­ing a parameter. Same as key during last calibration data viewing mode. Move to the next piece of information for the current event while in event scrolling mode
9. CFM key confirms current choice (and skips to the next item) ENTER
10. LEDs Relay’s 1, 2, 3, 4 activation
11. LEDs Controller Status, alarm activation
8
REAR PANEL
1. BNC Socket for pH or ORP electrode
2. Connection for Potential Matching Pin
3. Connection for electrode reference
4. Connections for Pt 100/Pt 1000 temperature sensor
5. Not Connected, for future use
6. Digital Transmitter input
7. HOLD input
8. Advanced Cleaning input (optional)
9. Power supply input
10. Alarm terminal
11. Relay 2 - second dosing terminal (optional)
12. Second analog output (optional)
13. Relay 1 - first dosing terminal
14. First analog output
15. HOLD output
16. Relay 4 - for Advanced Cleaning feature (optional)
17. Relay 3 - for Advanced Cleaning feature (optional)
18. RS485 output terminal
Unplug the meter before starting any electrical connections.
Disconnect the Potential Matching Pin when setup item I.04=”OFF”
9
SPECIFICATIONS
Range -2.00 to 16.00 pH
Resolution 0.01 pH
Accuracy ±0.02 pH
(@20°C/68°F) ±2 mV
Typical EMC Deviation ±0.2 pH
Analog Input 4 to 20 mA Digital Input for RS485
the pH/ORP/°C Transmitter Other Digital Insulated Inputs • 2 digital insulated inputs: 1 for hold and 1 for
Digital Insulated Output A digital insulated contact closed upon hold mode
-2000 to 2000 mV
-30 to 130.0 ºC
1 mV
0.1 ºC above -10 ºC; 1 ºC below
±0.5 ºC (-9.9 to 130.0 ºC); ±1 ºC (-30 to -10 ºC)
±10 mV ±0.5 ºC
the advanced cleaning
• ON state: 5 to 24 VDC
10
Temperature compensation Automatic or manual from -30 to 130 ºC Temperature probe With 3-wire or 2-wire Pt100/Pt1000 sensor
(with automatic recognition and damage test)
Installation Category II Power Supply 230±10% VAC, 115±10% VAC or 100±10% VAC;
(depending on model) 50/60 Hz
Power Consumption 10 VA Over Current Protection 400 mA 250V Quick Blow Fuse Max. Oscillation Frequency 8 MHz Relays 1, 2, 3, 4 Electromechanical relay SPDT contact outputs,
5A-250 VAC, 5A - 30 VDC (resistive load) Fuse protected: 5A, 250V Quick Blow Fuse
Alarm Relay Electromechanical Relay SPDT contact output,
Analog Output • 2 independent outputs
Analog Output Resolution 0.1‰ f.s. Analog Output Accuracy ± 2‰ f.s. Data logging 6000 pH/°C or ORP samples Environment 0 to 50 ºC; max 85% RH non-condensing Casing IP20 (housing); IP54 (front panel) Enclosure single case ½ DIN Weight approximately 1.6 kg (3.5 lb.)
MECHANICAL DIMENSIONS
FRONT VIEW SIDE VIEW
5A - 250 VAC, 5A - 30 VDC (resistive load) Fuse protected: 5A, 250V Quick Blow Fuse
• 0 - 22 mA (configuring as 0-20 mA or 4-20 mA)
11
INSTALLATION
Refer to diagram on page #9
12
Input power: Connect a 3-wire power cable to the terminal strip line (L), earth (PE) and neutral (N) terminal connections.
Power: 100VAC-120mA / 115VAC-100mA / 230VAC-50mA.
Line Contact: 400mA fuse inside.
PE must be connected to ground; leakage current 1mA.
Electrode: Connect the pH or ORP electrode to the BNC socket (#1 at page 9). To benefit from the differential (symmetrical) input, connect the proper electrode wire (if available) or a cable with a potential matching pin (grounding bar) to the relevant ter­minal (#2 at page 9) and enable the differential input through the setup code I.04.
Note When it is not possible to immerse the Potential Matching Pin
together with the pH electrode in the solution (asymmetrical input), disable the differential input through the setup code I.04.
In this case, detach the Potential Matching Pin from its con­nector. Failure to do so may cause damage of the reference electrode.
Pt 100/Pt 1000 Terminals: these contacts (#4 at page 9)
connect the Pt 100/Pt 1000 temperature sensor for auto­matic temperature compensation of pH measurement. In the case of shielded wire, connect the shield to pin 4. In the case of a 2-wire sensor con­nect the Pt 100/Pt 1000 to pins 1 and 3, and short pins 2 and 3 with a jumper wire. If the Pt 100/Pt 1000 has more than 2 wires, connect the two wires of one end to pins 2 and 3 (pin 2 is an auxiliary input to compensate for the cable resistance) and one wire from the other end to pin 1. Leave the fourth wire unconnected, if present.
Note The instrument automatically recognizes the sensor type (Pt
100 or Pt 1000).
Analog Outputs: when using shielded cable, if the shield is not connected at the other end of the cable, then con­nect it to the “+” terminal, otherwise leave it floating.
Note All cables connected to rear panel should end with cable lugs. Note A circuit breaker (rated 10A max.) must be connected in
close proximity to the equipment, and in a position easy to reach by the operator, for disconnection of the instrument and of all the devices connected to the relays.
13
CALIBRATION MODE
The calibration mode allows to calibrate the pH/ORP input, temperature input and 4-20 mA analog outputs.
The controller is factory calibrated for all these parameters. Periodical calibration of the instrument is recommended, in particular when greatest accuracy is required and at least bi­yearly.
It is possible to calibrate the electrode over only one point, but it is always good practice to perform a 2-point calibra­tion.
To enter the calibration mode press the CAL key.
Enter the correct password and press the CFM key. If a wrong password is entered, the instrument returns to the previous mode.
Note If the meter is set as an ORP controller (setup item G.00), the
pH calibration, pH reading offset adjustment, and pH offset and slope are not available. If the meter is selected as a pH controller, no ORP calibration can be selected.
Note Any calibration procedure can be aborted at any time by
pressing the CAL key, and the instrument returns to the previ­ous mode.
14
The pH/ORP controller can be calibrated through a one­point or two-point calibration.
The pH calibration can not be performed if the pH electrode is broken or leaking or the reference electrode is broken or dirty and an error is active. The ORP calibration can not be performed if the “Reference electrode broken or dirty” error is active. For pH calibration the temperature probe should be connected to the process meter.
All calibrations refer to the process controller. No calibration for the Digital Transmitter can be performed by the process controller, but the Digital Transmitter Calibrator must be used for that.
pH CALIBRATION
Initial Preparation
The pH and ORP calibrations can not be initiated while the process controller is configurated to take measurements from the Digital Transmitter.
It is recommended to perform pH calibration when the probe is replaced and after any cleaning action.
To perform any pH calibration procedure, the instrument has to be set as pH controller.
Pour small quantities of pH 7.01 (HI 7007) and pH 4.01 (HI 7004) or pH 10.01 (HI 7010) solutions into individual beakers. If possible, use plastic beakers to minimize any EMC interference.
For accurate calibration use different beakers for each buffer solution, the first one for rinsing the electrode and the second one for calibration. By doing this, contamination between buffers is minimized.
RINSE
CALIBRATION
HI 7007
HI 7007
• Enter the calibration mode, select the pH calibration (by moving through the menu with the and keys), then press the CFM key.
• Choose the pH buffer set between the two available ones: the standard set (4.01, 7.01, 10.01) and the NIST set (4.01,
6.86, 9.18). For the standard set confirm the “Std”; for the NIST set confirm the “niSt” (use the and keys for select between the two options).
15
Two-point calibration
Note The electrode should be submerged ap-
Note When it is not possible to immerse the Potential Matching Pin
The default buffer set is the one used for last calibration,
even if the procedure was not completed.
• Once confirmed the set of buffer val­ues, the primary LCD shows the measured pH value, while the second­ary LCD displays the first required buffer value.
• Remove the protective cap from the pH electrode and immerse it into the buffer solution (e.g. pH 7.01) together with the Potential Matching Pin and the tem­perature probe, then stir gently.
proximately 4 cm (11/2’’) in the solution. The temperature probe has to be located as close as possible to the pH electrode.
together with the pH electrode in the solution, disable the differential input by setting setup item I.04 to “OFF”.
• When the reading becomes stable, the probe indicator will stop flashing (after about 30 seconds), and if the pH value is close to the selected buffer, the “CFM” indicator will start blinking, oth­erwise the “WRONG” indicator, the pot and the BUF tags will start blinking.
• In the first case press the CFM key to confirm calibration. The meter will show the scrolling message “Press CFM again to confirm the current buffer or right to escape” (to prevent from confirming the calibration point inadvertently).
16
Pressing again CFM, the secondary LCD will display the
second expected buffer value.
• In the second case (pH value not close to the buffer) the meter will remain in the same state until the reading be­comes unstable or the calibration mode is quitted.
• For the second buffer value it is possible to choose be­tween pH 4.01 and pH 10.01 (or pH 4.01 and pH 9.18 if the NIST set has been selected). Use the or key to switch between the two possibilities.
• Once selected the buffer, the procedure is the same as for the first calibration point.
Note A time-out of 1.5 minutes is present for the pH electrode
response time. During calibration, if the pH reading is not stable after 1.5 minutes, the device displays twice the scroll­ing message “time-out”, then shows “WRONG” and it is not possible to complete calibration.
• At the end of calibration, with the meter set as pH control­ler, the instrument checks if the offset is between -30 and 30 mV and the slope between 53.5 and 62 mV/pH. If the values are not within these ranges, the message “OLd ProbE” scrolls twice across the LCD. The electrode is still working, but it is necessary to perform a cleaning proce­dure (see “Electrode conditioning and maintenance” and “In-line Cleaning” sections) or replace it.
If the offset is outside the -60 to 60 mV range, the “dEAd
ProbE” message will scroll across the LCD and the corre­sponding error is activated. The electrode has to be replaced as soon as possible because there is no reliability on the measured pH values.
One-point calibration
A one-point calibration at a value different from the standard buffer values is possible by entering the desired calibration value. This is the actual pH value at the current calibration temperature.
• Enter the pH calibration mode (no matter if the standard or NIST buffer set is selected), then press the SETUP key while the secondary LCD is dis­playing the first buffer value (pH 7.01 or pH
6.86).
17
Note If SETUP key is pressed instead of CFM, the calibration value
ORP CALIBRATION
• The pH calibration value will switch to 7.00, the first digit starts blinking and it is possible to change its value simply using the or key.
• Once selected the first digit value, press the key: the first digit will be fixed and the second one will start blinking. Pressing of the key repeatedly will result on circularly moving on the secondary LCD.
• When the desired calibration value is reached (must be within 0.00 to 16.00 pH), press the CFM key to confirm and the calibration will proceed as described above.
• If the selected value is outside boundaries, the confirma­tion is not accepted and the first digit keeps blinking (waiting for confirmation of a valid value).
selection is aborted and the meter reverts back to a two-point calibration.
To perform ORP calibration it is necessary to connect an HI 931001 or HI 8427 simulator to the BNC socket.
The meter has to be set as ORP controller.
18
• Set item I.04 to “OFF” to disable the Matching Pin.
• A two-point calibration has to be performed: the first point value is 0 mV and the second one can be chosen between 350 mV and 1900 mV.
• Enter the calibration mode, select the ORP calibration (use the  and keys to move through setup menu) and press the CFM key.
• Set the HI 931001 or HI 8427 simulator to 0 mV.
• The primary LCD will display the cur­rent mV measure and the secondary LCD will show the first calibration point (0 mV).
• When the reading becomes stable, if the ORP value is close to the calibration point, the “CFM” indicator starts blinking; otherwise the “WRONG” indicator blinks and the “CAL” is fixed on.
• In the first case press CFM to confirm calibration. The meter will proceed showing the scrolling message “Press CFM again to confirm the current buffer or right to es­cape” (to prevent from confirming the calibration point inadvertently). Pressing again CFM the secondary LCD will display the second calibration point.
• In the second case (blinking “WRONG”) the meter will remain in the WRONG state until the reading becomes unstable or the calibration mode is exited by pressing the CAL key.
• For the second calibration point it is possible to choose between 350 mV or 1900 mV. Pressing the or  key the value on the secondary LCD will switch between the two possibilities.
• Once selected the second calibration point, set the HI 931001 (350 mV) or HI 8427 (350 or 1900 mV) simula­tor to the same value and the calibration proceeds as for the first point.
Note In ORP calibration there is no time-out.
ANALOG OUTPUT CALIBRATION
The instrument can be provided with one or two analog out­puts, each of them can be calibrated.
• Connect an Ammeter or the HI 931002 calibrator to the analog output to measure the current erogated by the meter.
• Enter the calibration mode and move through the menu using the or key until “Out 1” or “Out 2” message appears on the primary LCD.
19
• Once visualized the selected output, press the key to choose the range of the analog output (0-20 mA or 4-20 mA); then press the CFM key to confirm the choice.
• Once selected and confirmed the range for the analog output, the secondary LCD shows the first point of calibration (1 or 4 mA) and the primary LCD displays the range of the actual calibration.
• With the tester check the real current value provided by the instrument at the output port. If this value is different from the expected one, it is possible to change the current erogated by the instrument pressing the or key.
• Adjust the value until it matches with the first point of cali­bration, then press CFM to confirm. The instrument will turn to the second calibration point (20 mA).
• The calibration procedure is the same as for the first point.
• Once the instrument is calibrated for both points (and con­firmed), it will exit from the calibration menu.
TEMPERATURE CALIBRATION
• Prepare a beaker containing crushed ice and water at 0°C (32°F) and another one with hot water at 25°C (77°F) or 50°C (122°F).
• Use a Checktemp or another cali­brated thermometer with a resolution of 0.1° as a reference thermometer.
• Immerse the temperature probe in the beaker with ice and water as near to the Checktemp as possible.
20
0 °C
(32 °F)
°C
0 °C
(32 °F)
50 ºC
(122 ºF)
°
C
°C
Note The instrument can support Pt100 or Pt1000 temperature
sensor and calibration can be performed with anyone of these two probes.
• After entering the calibration mode, move through the menu (using the or key) to choose the temperature and the correct kind of used probe; the pri­mary LCD shows “°C” and the secondary LCD gives indication about the kind of probe (“100” in­dicates a Pt100 probe, while “1000” stands for a Pt1000 probe).
Press the CFM key to confirm selection.
• The calibration has to be performed over two points: the first point has to be 0°C and the second one can be chosen between 25°C and 50°C.
Once confirmed the type of calibration, the primary LCD
will display the current temperature measure and the sec­ondary LCD will show the first calibration point (0°C).
• When the reading becomes stable, if the temperature value is close to the calibration point the CFM tag starts blink­ing, otherwise the WRONG indicator will flash.
• In the first case press the CFM key to confirm calibration. The meter will proceed showing the scrolling message “Press CFM again to confirm the current buffer or right to es­cape” (to prevent from confirming the calibration point inadvertently).
• Pressing again the CFM key, the sec­ondary LCD will display the second calibration point.
• In the case the measured temperature is not close to the calibration value, the meter remains in the WRONG status until the reading becomes unstable or the calibration mode
21
VOLT CALIBRATION
is exited (by pressing the CAL key).
• When performing the second calibration point, it is pos­sible to choose between two values, 25°C and 50°C. Pressing the  or key the value on the secondary LCD will switch between the two possibilities.
• Once selected the value, immerse the temperature probe in the second beaker as near as possible to the Checktemp and the calibration procedure will be the same as for the first point.
The instrument is factory calibrated; however the user may also perform the Volt calibration, which is a procedure avail­able for the meter configured as pH controller only.
During pH measurements the instrument reads a mV value from the electrode and then converts it into a pH value.
• Set item I.04 to “OFF” to disable the Matching Pin.
• Connect a HI 931001 or HI 8427 simulator to the BNC socket.
• Once entered the calibration mode, move through the menu with or key, select the Volt calibration (the primary LCD shows “UOLt” mes­sage) and confirm it by pressing the CFM key.
• After confirmation the primary LCD will show the actual mV value and the secondary LCD will display the first calibration point.
• The calibration is performed over two points, 0 and 350 mV.
• Set the HI 931001 or HI 8427 simulator to 0 mV.
• When the reading becomes stable, if the measured value is close to the calibration point the CFM tag starts blink­ing, otherwise the WRONG indicator will flash.
• In the first case press the CFM key to confirm calibration.
22
The meter will proceed showing the scrolling message “Press CFM to confirm the current buffer or right to escape” (to prevent from confirming the calibration point inadvertently).
• Pressing again CFM the secondary LCD will display the second calibration point.
• Set the HI 931001 or HI 8427 simulator to 350 mV and follow the same procedure as for the first point.
• If the measured value is not close to the calibration value, the meter will remain in the WRONG status until the read­ing becomes unstable or the calibration mode is exited (by pressing the CAL key).
23
SETUP MODE
The Setup Mode allows the user to set all needed character­istics of the meter.
To enter the mode, press the SETUP key and enter the password when the device is in idle or control mode.
If the correct password is not entered, the user can only view the setup parameters (except for passwords) without modify­ing them (and the device remains in control mode). An exception is certain setup items, or flags, which can activate special tasks, when set and confirmed.
To each setup parameter (or setup item) is assigned a three characters (one letter followed by two digits) setup code which is entered and displayed on the secondary LCD. The first character identifies the group of setup items, while the two digits identify the particular item within that group.
The setup codes can be selected after password is entered and CFM key is pressed. When CFM is pressed, the current setup item is saved on EEPROM and the following item is displayed.
The possible transitions in setup mode are the following:
ENTERING THE PASSWORD
• Press SETUP to enter the setup mode. The primary LCD will
• Enter the first digit of the pass-
• Then move to the next digit with and
24
display “0000”, while the secondary LCD shows “PAS”. The first digit of the primary LCD will blink.
word by using the or key.
enter a digit as described above. Continue for the last two digits.
Note The default password is set at “0000”.
ENTERING SETUP ITEMS
• When the whole password has been in­serted, press CFM to confirm it.
After confirmation of the password the primary LCD will show the name of the first setup group (see table) while the secondary LCD will display the setup code of the first item of the group.
• By pressing the or key it is possible to cycle through the setup groups; the secondary LCD will always show the code of the first item of the group.
• Once a group is selected, it is pos­sible to choose an item manually. Pressing the key, the first digit of the setup code will start blinking. Now it is possible to change its value by pressing the or key.
• Pressing the key again, the first digit will be fixed while the second digit starts blinking and its value can be changed as described above.
• By pressing the key again all the digits will be fixed.
• If CFM key is pressed, the selected item is confirmed; the secondary LCD will show the setup item code while the primary LCD will display the current item value.
If a fixed set of values is available for the selected item, use the or  key to switch between them.
25
Otherwise, if a numeric value has to be entered for the item, use the or key to change the value of the blink­ing digit and the key to cycle through the number’s digits.
• Once a value is set, press the CFM key to confirm. The instrument will turn to the next item and the new item’s value will be displayed on the primary LCD.
• If a wrong value is confirmed, the WRONG indicator starts blinking, the new value is not accepted and the instrument will not switch to the next item until a correct value will be confirmed.
• Instead of selecting the item manually, it is possible to cycle through all the items of a selected group by pressing re­peatedly CFM key. The procedure to modify the item’s value is the same described above.
If the last item of the group is reached, by pressing the CFM key again the primary display will show the group name and it will be possible to change the group by press­ing the or key.
26
Note At any time for exiting the setup mode press the SETUP key. If
no modification has been confirmed, no setup item is changed.
Note While in the setup mode, if no activity is performed for about
5 minutes after entering the setup mode, the mode is auto­matically exited and the instrument returns to the previous mode.
The below table lists the setup codes along with the description of the specific setup items, their valid values and whether the item is present for ORP mode. Titles in Italic character (eg. “SETPOINT 1 “) represent different subgroups within the same group. The subgroup name (eg. “Set1”) is showed when entering the first item of the sub­group. To scroll through subgroups, use the key. Depending on the device model, some of the below items or item values may not be available.
Code Valid Values Default Present
for ORP
GENERAL (“GEnE”)
G.00 pH/ORP input “PH”, “OrP” (see note 10) “PH” yes G.01 Temperature compensation “AtC”: Automatic “AtC” no
“USEr”: Manual (see notes 9, 25)
G.02 Manual or probe error -30 to 130.0 ºC 25.0 no
temperature (see notes 9, 25) G.10 Factory ID 0000 to 9999 (see note 20) 0000 yes G.11 Process ID alias RS485 address 0 0 to 99 (see note 20) 00 yes G.12 Hardware Identifier 7 digits identification number, read only yes
(see notes 4, 20) G.98 Calibration/hold password 0000 to 9999 (see notes 2, 20) 0000 yes G.99 General password 0000 to 9999 (see notes 2, 20) 0000 yes
CONTROL (“CtrL”)
C.00 Control enable “OFF”: Control mode disabled “OFF” yes
“On”: Control mode enabled
SETPOINT 1 (“SEt1”)
C.10 Setpoint 1 mode (M1) “OFF”: Disabled “OOHI” yes
(see note 1) “OOHI”: ON-OFF, high set point
“OOLO”: ON-OFF, low set point “PIdH”: P.I.D., high set point / “PIdL”: P.I.D., low set point
C.11 Setpoint 1 (S1) -2.00 to 16.00 pH or 8.00 pH or yes
(see note 1) -2000 to 2000 mV 500 mV
27
Code Valid Values Default Present
for ORP
C.12 Hysteresis for setpoint 1 (H1) 0.00 to 18.00 pH or 1 pH or yes
(see note 1) 0 to 4000 mV 50 mV C.13 Deviation for setpoint 1 (D1) 0.50 to 18.00 pH or 1 pH or yes
(see note 1) 25 to 4000 mV 50 mV C.14 Reset time for setpoint 1 0.1 to 999.9 minutes 999.9 mins yes C.15 Rate time for setpoint 1 0.0 to 999.9 minutes 0.0 mins yes
SETPOINT 2 (“SEt2”)
C.20 Setpoint 2 mode (M2) “OFF”: Disabled “OOLO” yes
(see note 1) “OOHI”: ON-OFF, high set point
“OOLO”: ON-OFF, low set point “PIdH”: P.I.D., high set point “PIdL”: P.I.D., low set point
C.21 Setpoint 2 (S2) -2.00 to 16.00 pH or 6.00 pH or yes
(see note 1) -2000 to 2000 mV -500 mV C.22 Hysteresis for setpoint 2 (H2) 0.00 to 18.00 pH or 1 pH or yes
(see note 1) 0 to 4000 mV 50 mV C.23 Deviation for setpoint 2 (D2) 0.50 to 18.00 pH or 1 pH or yes
(see note 1) 25 to 4000 mV 50 mV C.24 Reset time for setpoint 2 0.1 to 999.9 minutes 999.9 mins yes C.25 Rate time for setpoint 2 0.0 to 999.9 minutes 0.0 mins yes
ALARM (“ALAr”)
C.30 Alarm relay delta value 0.50 to 18.00 pH 1.00 pH yes
for setpoint 1 (A1) or or
(see notes 1, 22) 25 to 4000 mV 100 mV C.31 Alarm relay delta value 0.50 to 18.00 pH 1.00 pH yes
for setpoint 2 (A2) or or
(see notes 1, 22) 25 to 4000 mV 100 mV C.32 Maximum relay ON time 1-60 minutes 60 yes
(after that an alarm is generated) C.33 Alarm mask time 00:00 to 30:00 minutes 00:30 yes
PROGRAMMABLE CONTROL TIMING (“tIME”)
C.41 Hold time start 00:00 to 23:59 (see note 8) 00:00 yes C.42 Hold time stop 00:00 to 2 3:59 (see note 8) 00:00 yes
28
Code Valid Values Default Present
for ORP
C.51 Monday hold mode “OFF”: Disabled “OFF” yes
enable “On”: Enabled C.52 Tuesday hold mode “OFF”: Disabled “OFF” yes
enable “On”: Enabled C.53 Wednesday hold mode “OFF”: Disabled “OFF” yes
enable “On”: Enabled C.54 Thursday hold mode “OFF”: Disabled “OFF” yes
enable “On”: Enabled C.55 Friday hold mode “OFF”: Disabled “OFF” yes
enable “On”: Enabled C.56 Saturday hold mode “OFF”: Disabled “OFF” yes
enable “On”: Enabled C.57 Sunday hold mode “OFF”: Disabled “OFF” yes
enable “On”: Enabled C.60 Proportional control mode 01:00 - 30:00 minutes 05:00 yes
period C.70 Hold mode end delay 00 to 99 seconds 00 yes
OUTPUT (“OutP”)
RELAYS (“rELA”)
O.01 Relay 1 mode “OFF”: Disabled “SEt1” yes
(see notes 19, 21) “SEt1”: Control, setpoint 1
“SEt2”: Control, setpoint 2 “SCLE”: Simple cleaning “HOLd”: Energized in hold mode
O.02 Relay 2 mode “OFF”: Disabled “SEt2” yes
(see notes 19, 21) “SEt1”: Control, setpoint 1
“SEt2”: Control, setpoint 2 “SCLE”: Simple cleaning “HOLd”: Energized in hold mode
O.03 Relay 3 mode “OFF”: Disabled “OFF” yes
(see notes 15, 19) “SCLE”: Simple cleaning
“ACLE”: Advanced cleaning / “HOLd”: Energized in hold mode O.04 Relay 4 mode “OFF”: Disabled “OFF” yes
(see notes 15, 19) “SCLE”: Simple cleaning
“ACLE”: Advanced cleaning / “HOLd”: Energized in hold mode
29
Code Valid Values Default Present
for ORP
0.05 Hold digital output “OFF”: Disabled “HOLd” yes
“HOLd”: Enabled upon hold mode
ANALOG OUTPUT #1 (“Out1”)
O.10 Analog output 1 mode “rECO”: Recorder “rECO” yes
(see notes 6, 21) “SEt”: Control, setpoint 1 O.11 Analog output 1 type “0-20”: 0-20 mA 4-20 (6) yes
“4-20”: 4-20 mA
O.12 Measurement value for -2.00 to 16.00 pH 0.00 pH or yes
analog output 1 minimum or -2000 to 2000 mV -2000 mV (6)
(O_VARMIN1) (O_VARMIN1 R O_VARMAX1-1 pH or 50 mV,
O_VARMIN1 R O_HOLD1R O_VARMAX1)
O.13 Measurement value for -2.00 to 16.00 pH 14.00 pH or yes
analog output 1 maximum or -2000 to 2000 mV 2000 mV (6)
(O_VARMAX1) (O_VARMIN1 R O_VARMAX1-1 pH or 50 mV,
O_VARMIN1 R O_HOLD1R O_VARMAX1)
O.14 Analog output 1 value upon “USEr”: User selected value “HOLd” yes
hold mode (see note 16) “HOLd”: Previous value is frozen O.15 Analog output 1 value upon -2.00 to 16.00 pH 7.00 pH or yes
hold mode if O.14=”USEr” or -2000 to 2000 mV 0 mV (6)
(O_HOLD1) (it must be within the O_VARMIN1
to O_VARMAX1-1 interval)
ANALOG OUTPUT #2 (“Out2”)
O.20 Analog output 2 mode “rECO”: Recorder (temperature) “rECO” (pH) yes
(see notes 21, 26) “SEt”: Control, setpoint 2 (pH) or “OFF” (ORP)
or “OFF”: Disabled “SEt”: Control, setpoint 2 (ORP)
O.21 Analog output 2 type “0-20”: 0-20 mA “4-20” yes
“4-20”: 4-20 mA
O.22 Measurement value for -30 to 130.0 ºC 0 ºC no
analog output 2 minimum (O_VARMIN2 R O_VARMAX2-10 ºC,
(O_VARMIN2) O_VARMIN2 R O_HOLD2R O_VARMAX2) O.23 Measurement value for -30 to 130.0 ºC +100.0 ºC no
analog output 2 maximum (O_VARMIN2 R O_VARMAX2-10 ºC,
(O_VARMAX2) O_VARMIN2 R O_HOLD2R O_VARMAX2)
30
Loading...
+ 86 hidden pages