CE DECLARATION OF CONFORMITY . . . . . . . . . . 113
3
WARRANTY
All Hanna Instruments meters are guaranteed for two
years (sensors, electrodes and probes for six months)
against defects in workmanship and materials when used for
their intended purpose and maintained according to instructions. This warranty is limited to repair or replacement free of
charge.
Damage due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model
number, date of purchase, serial number and the nature of
the failure. If the repair is not covered by the warranty, you
will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Returned
Goods Authorization number from the Customer Service department and then send it with shipping costs prepaid. When
shipping any instrument, make sure it is properly packaged
for complete protection.
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance notice.
MODEL IDENTIFICATION
Dear Customer,
Thank you for choosing a Hanna Product.
Please read this instruction manual carefully before using the
instrument. It will provide you with the necessary information
for correct use of the instrument, as well as a precise idea of
its versatility.
If you need additional technical information, do not hesitate
to e-mail us at tech@hannainst.com.
This instrument is in compliance with directives.
The models HI 504XYZ-α are pH/ORP controllers.
The meaning of the last letters is according to the following
All rights are reserved. Reproduction in whole or in part is prohibited without the
written consent of the copyright owner, Hanna Instruments Inc., 584 Park East Drive,
Woonsocket, Rhode Island, 02895 , USA.
5
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred
during shipping. If there is any noticeable damage, notify
your Dealer or the nearest Hanna Customer Service Center
immediately.
NoteSave all packing materials until you are sure that the instru-
ment functions correctly. Any damaged or defective items must
be returned in their original packing materials together with
the supplied accessories.
GENERAL DESCRIPTION
The product is a real time microprocessor-based pH/ORP
controller. It provides accurate measurements, flexible
control capabilities and dual alarm signals.
The system is composed of a case inside which the signal
conversion circuitry, the microprocessor circuitry and the output
power drivers are contained.
MAIN FEATURES
• Configurations: for measure and control of pH or ORP, with
two separate sets of control settings. Thus no loss of settings
occurs when changing from pH to ORP and vice versa.
• Display: large LCD with 4 ½ 17 mm digits and 3 ½ 10 mm
digits.
• LEDs: a green LED (alarm relay) and a red LED for signaling
the device status.
• Alarm relay: one output relay for alarm condition (fail safe
mode: COM and NC are connected).
• Daily programmable control timing.
• Diagnostic features.
• pH/ORP probe check.
• Temperature sensor (Pt100 or Pt1000 type) with automatic
recognition and damage test.
• Hold management, including hold digital input to enter
hold mode with an external trigger.
• Logging of the last 100 error, configuration, calibration and
cleaning events.
6
• Calibration and Setup procedures are password protected.
• Calibration: pH calibration at 2 points with two possible sets
of buffers (either 7.01 - 4.01 - 10.01 or 6.86 - 4.01 - 9.18),
or at 1 point with user-selectable value.
• Temperature compensation of the HANNA standard buffers.
• GLP features: old probe check, dead probe check, calibration time-out, last calibration data internally recorded (nonvolatile EEPROM memory): calibration date and time, pH
offset, pH slope, number of calibration points and correspondent pH values.
• Solution compensation of the pH reading.
• Manual temperature setting when the temperature probe is
not inserted or temperature is out of range.
• Input: pH/ORP electrode with BNC connector.
• Input: pH/ORP/Temperature Digital Transmitter (HI
504910) which can be connected up to 1.2 Km (0.5 miles)
far away.
• Output (0-20 or 4-20 mA): one analog output for recording of pH/ORP or controlling (only for models with PID).
• RS485 communication with Short Messaging Service (SMS)
for GSM phones.
• Data logging: 6000 samples pH/°C or ORP downloadable
through RS485 and HI92500 application software.
• Real time clock.
OPTIONAL FEATURES (depending on model)
• LEDs: four yellow LEDs are provided for signaling the energizing of relay 1, relay 2 (optional), relay 3 (optional) and
relay 4 (optional).
• Relays: 1, 2 or 4 output relays. The relays LEDs are switched
ON whenever the correspondent relay is energized (NO
and COM connected).
• Advanced Cleaning is provided through two cleaning commands
and variety of trigger events (relays 3 and 4 only).
• Output (0-20 or 4-20 mA): second independent analog
output for recording of temperature or controlling (only for
models with PID).
• PID control.
7
FUNCTIONAL DESCRIPTION
FRONT PANEL
1. Liquid Crystal Display
2. LCD keyenters and exits the event scrolling mode. During pH
calibration, alternately displays pH buffer value or current
temperature
3. SETUP keyenters and exits setup mode
4. CAL DATA keylast calibration data viewing (enters and exits)
5. CAL keyinitiates and exits calibration mode
6. keyincreases the blinking digit/letter by one when selecting a
parameter. Advances forward while in last calibration data
viewing mode or event scrolling mode. Increases the temperature setting when temperature probe is not inserted
7. keydecreases the blinking digit/letter by one when selecting a
parameter. Reverts backward while in last calibration data
viewing mode or event scrolling mode. Decreases the
temperature setting when temperature probe is not inserted
8. keymoves to the next digit/letter (circular buffer) when selecting a parameter. Same as key during last calibration
data viewing mode. Move to the next piece of information
for the current event while in event scrolling mode
9. CFM keyconfirms current choice (and skips to the next item) ENTER
10. LEDsRelay’s 1, 2, 3, 4 activation
11. LEDsController Status, alarm activation
8
REAR PANEL
1. BNC Socket for pH or ORP electrode
2. Connection for Potential Matching Pin
3. Connection for electrode reference
4. Connections for Pt 100/Pt 1000 temperature sensor
5. Not Connected, for future use
6. Digital Transmitter input
7. HOLD input
8. Advanced Cleaning input (optional)
9. Power supply input
10. Alarm terminal
11. Relay 2 - second dosing terminal (optional)
12. Second analog output (optional)
13. Relay 1 - first dosing terminal
14. First analog output
15. HOLD output
16. Relay 4 - for Advanced Cleaning feature (optional)
17. Relay 3 - for Advanced Cleaning feature (optional)
18. RS485 output terminal
Unplug the meter before starting any electrical connections.
Disconnect the Potential Matching Pin when setup item I.04=”OFF”
9
SPECIFICATIONS
Range-2.00 to 16.00 pH
Resolution0.01 pH
Accuracy±0.02 pH
(@20°C/68°F)±2 mV
Typical EMC Deviation±0.2 pH
Analog Input4 to 20 mA
Digital Input forRS485
the pH/ORP/°C Transmitter
Other Digital Insulated Inputs • 2 digital insulated inputs: 1 for hold and 1 for
Digital Insulated OutputA digital insulated contact closed upon hold mode
-2000 to 2000 mV
-30 to 130.0 ºC
1 mV
0.1 ºC above -10 ºC; 1 ºC below
±0.5 ºC (-9.9 to 130.0 ºC); ±1 ºC (-30 to -10 ºC)
±10 mV
±0.5 ºC
the advanced cleaning
• ON state: 5 to 24 VDC
10
Temperature compensation Automatic or manual from -30 to 130 ºC
Temperature probeWith 3-wire or 2-wire Pt100/Pt1000 sensor
(with automatic recognition and damage test)
Installation CategoryII
Power Supply230±10% VAC, 115±10% VAC or 100±10% VAC;
(depending on model)50/60 Hz
Power Consumption10 VA
Over Current Protection400 mA 250V Quick Blow Fuse
Max. Oscillation Frequency 8 MHz
Relays 1, 2, 3, 4Electromechanical relay SPDT contact outputs,
Analog Output Resolution0.1‰ f.s.
Analog Output Accuracy± 2‰ f.s.
Data logging6000 pH/°C or ORP samples
Environment0 to 50 ºC; max 85% RH non-condensing
CasingIP20 (housing); IP54 (front panel)
Enclosuresingle case ½ DIN
Weightapproximately 1.6 kg (3.5 lb.)
• Input power: Connect a 3-wire power cable
to the terminal strip line (L), earth (PE) and
neutral (N) terminal connections.
Power: 100VAC-120mA / 115VAC-100mA / 230VAC-50mA.
Line Contact: 400mA fuse inside.
PE must be connected to ground; leakage current 1mA.
• Electrode: Connect the pH or ORP electrode to the BNC
socket (#1 at page 9).
To benefit from the differential (symmetrical) input, connect
the proper electrode wire (if available) or a cable with a
potential matching pin (grounding bar) to the relevant terminal (#2 at page 9) and enable the differential input
through the setup code I.04.
NoteWhen it is not possible to immerse the Potential Matching Pin
together with the pH electrode in the solution (asymmetrical
input), disable the differential input through the setup code I.04.
In this case, detach the Potential Matching Pin from its connector. Failure to do so may cause damage of the reference
electrode.
• Pt 100/Pt 1000 Terminals: these contacts (#4 at page 9)
connect the Pt 100/Pt 1000 temperature sensor for automatic temperature compensation of pH measurement. In
the case of shielded wire, connect the shield to pin 4.
In the case of a 2-wire sensor connect the Pt 100/Pt 1000 to pins 1
and 3, and short pins 2 and 3 with
a jumper wire.
If the Pt 100/Pt 1000 has more than
2 wires, connect the two wires of
one end to pins 2 and 3 (pin 2 is
an auxiliary input to compensate
for the cable resistance) and one
wire from the other end to pin 1.
Leave the fourth wire unconnected, if present.
NoteThe instrument automatically recognizes the sensor type (Pt
100 or Pt 1000).
• Analog Outputs: when using shielded cable, if the shield
is not connected at the other end of the cable, then connect it to the “+” terminal, otherwise leave it floating.
NoteAll cables connected to rear panel should end with cable lugs.
NoteA circuit breaker (rated 10A max.) must be connected in
close proximity to the equipment, and in a position easy to
reach by the operator, for disconnection of the instrument
and of all the devices connected to the relays.
13
CALIBRATION MODE
The calibration mode allows to calibrate the pH/ORP input,
temperature input and 4-20 mA analog outputs.
The controller is factory calibrated for all these parameters.
Periodical calibration of the instrument is recommended, in
particular when greatest accuracy is required and at least biyearly.
It is possible to calibrate the electrode over only one point,
but it is always good practice to perform a 2-point calibration.
To enter the calibration mode press the CAL key.
Enter the correct password and press the CFM key. If a wrong
password is entered, the instrument returns to the previous
mode.
NoteIf the meter is set as an ORP controller (setup item G.00), the
pH calibration, pH reading offset adjustment, and pH offset
and slope are not available. If the meter is selected as a pH
controller, no ORP calibration can be selected.
NoteAny calibration procedure can be aborted at any time by
pressing the CAL key, and the instrument returns to the previous mode.
14
The pH/ORP controller can be calibrated through a onepoint or two-point calibration.
The pH calibration can not be performed if the pH electrode
is broken or leaking or the reference electrode is broken or
dirty and an error is active. The ORP calibration can not be
performed if the “Reference electrode broken or dirty” error is
active. For pH calibration the temperature probe should be
connected to the process meter.
All calibrations refer to the process controller. No calibration
for the Digital Transmitter can be performed by the process
controller, but the Digital Transmitter Calibrator must be used
for that.
pH CALIBRATION
Initial Preparation
The pH and ORP calibrations can not be initiated while the
process controller is configurated to take measurements from
the Digital Transmitter.
It is recommended to perform pH calibration when the probe
is replaced and after any cleaning action.
To perform any pH calibration procedure, the instrument has
to be set as pH controller.
Pour small quantities of pH 7.01 (HI 7007) and pH 4.01
(HI 7004) or pH 10.01 (HI 7010) solutions into individual
beakers. If possible, use plastic beakers to minimize any EMC
interference.
For accurate calibration use different beakers for each buffer
solution, the first one for rinsing the electrode and the second
one for calibration. By doing this, contamination between
buffers is minimized.
RINSE
CALIBRATION
HI 7007
HI 7007
• Enter the calibration mode, select the pH calibration (by
moving through the menu with the and keys), then
press the CFM key.
• Choose the pH buffer set between the two available ones:
the standard set (4.01, 7.01, 10.01) and the NIST set (4.01,
6.86, 9.18). For the standard set confirm the “Std”; for the
NIST set confirm the “niSt” (use the and keys for select
between the two options).
15
Two-point calibration
NoteThe electrode should be submerged ap-
NoteWhen it is not possible to immerse the Potential Matching Pin
The default buffer set is the one used for last calibration,
even if the procedure was not completed.
• Once confirmed the set of buffer values, the primary LCD shows the
measured pH value, while the secondary LCD displays the first required buffer
value.
• Remove the protective cap from the pH
electrode and immerse it into the buffer
solution (e.g. pH 7.01) together with
the Potential Matching Pin and the temperature probe, then stir gently.
proximately 4 cm (11/2’’) in the solution.
The temperature probe has to be located
as close as possible to the pH electrode.
together with the pH electrode in the solution, disable the
differential input by setting setup item I.04 to “OFF”.
• When the reading becomes stable, the
probe indicator will stop flashing (after
about 30 seconds), and if the pH value
is close to the selected buffer, the
“CFM” indicator will start blinking, otherwise the “WRONG” indicator, the pot
and the BUF tags will start blinking.
• In the first case press the CFM key to confirm calibration.
The meter will show the scrolling message “Press CFM again
to confirm the current buffer or right to escape” (to prevent
from confirming the calibration point inadvertently).
16
Pressing again CFM, the secondary LCD will display the
second expected buffer value.
• In the second case (pH value not close to the buffer) the
meter will remain in the same state until the reading becomes unstable or the calibration mode is quitted.
• For the second buffer value it is possible to choose between pH 4.01 and pH 10.01 (or pH 4.01 and pH 9.18 if
the NIST set has been selected). Use the or key to
switch between the two possibilities.
• Once selected the buffer, the procedure is the same as for
the first calibration point.
NoteA time-out of 1.5 minutes is present for the pH electrode
response time. During calibration, if the pH reading is not
stable after 1.5 minutes, the device displays twice the scrolling message “time-out”, then shows “WRONG” and it is not
possible to complete calibration.
• At the end of calibration, with the meter set as pH controller, the instrument checks if the offset is between -30 and
30 mV and the slope between 53.5 and 62 mV/pH. If the
values are not within these ranges, the message “OLd
ProbE” scrolls twice across the LCD. The electrode is still
working, but it is necessary to perform a cleaning procedure (see “Electrode conditioning and maintenance” and
“In-line Cleaning” sections) or replace it.
If the offset is outside the -60 to 60 mV range, the “dEAd
ProbE” message will scroll across the LCD and the corresponding error is activated. The electrode has to be replaced
as soon as possible because there is no reliability on the
measured pH values.
One-point calibration
A one-point calibration at a value different from the standard
buffer values is possible by entering the desired calibration
value. This is the actual pH value at the current calibration
temperature.
• Enter the pH calibration mode (no matter if the standard or
NIST buffer set is selected), then press the
SETUP key while the secondary LCD is displaying the first buffer value (pH 7.01 or pH
6.86).
17
NoteIf SETUP key is pressed instead of CFM, the calibration value
ORP CALIBRATION
• The pH calibration value will switch
to 7.00, the first digit starts blinking
and it is possible to change its value
simply using the or key.
• Once selected the first digit value, press the key: the first
digit will be fixed and the second one will start blinking.
Pressing of the key repeatedly will result on circularly
moving on the secondary LCD.
• When the desired calibration value is reached (must be
within 0.00 to 16.00 pH), press the CFM key to confirm
and the calibration will proceed as described above.
• If the selected value is outside boundaries, the confirmation is not accepted and the first digit keeps blinking (waiting
for confirmation of a valid value).
selection is aborted and the meter reverts back to a two-point
calibration.
To perform ORP calibration it is necessary to connect an HI931001 or HI 8427 simulator to the BNC socket.
The meter has to be set as ORP controller.
18
• Set item I.04 to “OFF” to disable the Matching Pin.
• A two-point calibration has to be performed: the first point
value is 0 mV and the second one can be chosen between
350 mV and 1900 mV.
• Enter the calibration mode, select the ORP calibration (use
the and keys to move through setup menu) and press
the CFM key.
• Set the HI 931001 or HI 8427 simulator to 0 mV.
• The primary LCD will display the current mV measure and the secondary
LCD will show the first calibration
point (0 mV).
• When the reading becomes stable, if the ORP value is
close to the calibration point, the “CFM” indicator starts
blinking; otherwise the “WRONG” indicator blinks and
the “CAL” is fixed on.
• In the first case press CFM to confirm calibration. The
meter will proceed showing the scrolling message “Press
CFM again to confirm the current buffer or right to escape” (to prevent from confirming the calibration point
inadvertently). Pressing again CFM the secondary LCD will
display the second calibration point.
• In the second case (blinking “WRONG”) the meter will
remain in the WRONG state until the reading becomes
unstable or the calibration mode is exited by pressing the
CAL key.
• For the second calibration point it is possible to choose
between 350 mV or 1900 mV. Pressing the or key the
value on the secondary LCD will switch between the two
possibilities.
• Once selected the second calibration point, set the HI931001 (350 mV) or HI 8427 (350 or 1900 mV) simulator to the same value and the calibration proceeds as for
the first point.
NoteIn ORP calibration there is no time-out.
ANALOG OUTPUT CALIBRATION
The instrument can be provided with one or two analog outputs, each of them can be calibrated.
• Connect an Ammeter or the HI 931002 calibrator to the
analog output to measure the current erogated by the meter.
• Enter the calibration mode and move through the menu
using the or key until “Out 1” or “Out 2” message
appears on the primary LCD.
19
• Once visualized the selected output, press the key to
choose the range of the analog output (0-20 mA or 4-20
mA); then press the CFM key to confirm the choice.
• Once selected and confirmed the
range for the analog output, the
secondary LCD shows the first point
of calibration (1 or 4 mA) and the
primary LCD displays the range of
the actual calibration.
• With the tester check the real current value provided by the
instrument at the output port. If this value is different from
the expected one, it is possible to change the current
erogated by the instrument pressing the or key.
• Adjust the value until it matches with the first point of calibration, then press CFM to confirm. The instrument will
turn to the second calibration point (20 mA).
• The calibration procedure is the same as for the first point.
• Once the instrument is calibrated for both points (and confirmed), it will exit from the calibration menu.
TEMPERATURE CALIBRATION
• Prepare a beaker containing
crushed ice and water at 0°C (32°F)
and another one with hot water at
25°C (77°F) or 50°C (122°F).
• Use a Checktemp or another calibrated thermometer with a
resolution of 0.1° as a reference
thermometer.
• Immerse the temperature probe in
the beaker with ice and water as
near to the Checktemp as possible.
20
0 °C
(32 °F)
°C
0 °C
(32 °F)
50 ºC
(122 ºF)
°
C
°C
Note The instrument can support Pt100 or Pt1000 temperature
sensor and calibration can be performed with anyone of these
two probes.
• After entering the calibration mode, move through the menu
(using the or key) to choose the temperature and the
correct kind of used probe; the primary LCD shows “°C” and the
secondary LCD gives indication
about the kind of probe (“100” indicates a Pt100 probe, while
“1000” stands for a Pt1000
probe).
Press the CFM key to confirm selection.
• The calibration has to be performed
over two points: the first point has to
be 0°C and the second one can be
chosen between 25°C and 50°C.
Once confirmed the type of calibration, the primary LCD
will display the current temperature measure and the secondary LCD will show the first calibration point (0°C).
• When the reading becomes stable, if the temperature value
is close to the calibration point the CFM tag starts blinking, otherwise the WRONG indicator will flash.
• In the first case press the CFM key to confirm calibration.
The meter will proceed showing the scrolling message “Press
CFM again to confirm the current buffer or right to escape” (to prevent from confirming the calibration point
inadvertently).
• Pressing again the CFM key, the secondary LCD will display the second
calibration point.
• In the case the measured temperature is not close to the
calibration value, the meter remains in the WRONG status
until the reading becomes unstable or the calibration mode
21
VOLT CALIBRATION
is exited (by pressing the CAL key).
• When performing the second calibration point, it is possible to choose between two values, 25°C and 50°C. Pressing
the or key the value on the secondary LCD will switch
between the two possibilities.
• Once selected the value, immerse the temperature probe
in the second beaker as near as possible to the Checktemp
and the calibration procedure will be the same as for the
first point.
The instrument is factory calibrated; however the user may
also perform the Volt calibration, which is a procedure available for the meter configured as pH controller only.
During pH measurements the instrument reads a mV value
from the electrode and then converts it into a pH value.
• Set item I.04 to “OFF” to disable the Matching Pin.
• Connect a HI 931001 or HI 8427 simulator to the BNC
socket.
• Once entered the calibration mode,
move through the menu with or
key, select the Volt calibration (the
primary LCD shows “UOLt” message) and confirm it by pressing the
CFM key.
• After confirmation the primary LCD
will show the actual mV value and
the secondary LCD will display the
first calibration point.
• The calibration is performed over two points, 0 and 350 mV.
• Set the HI 931001 or HI 8427 simulator to 0 mV.
• When the reading becomes stable, if the measured value
is close to the calibration point the CFM tag starts blinking, otherwise the WRONG indicator will flash.
• In the first case press the CFM key to confirm calibration.
22
The meter will proceed showing the scrolling message “Press
CFM to confirm the current buffer or right to escape” (to
prevent from confirming the calibration point inadvertently).
• Pressing again CFM the secondary LCD will display the
second calibration point.
• Set the HI 931001 or HI 8427 simulator to 350 mV and
follow the same procedure as for the first point.
• If the measured value is not close to the calibration value,
the meter will remain in the WRONG status until the reading becomes unstable or the calibration mode is exited (by
pressing the CAL key).
23
SETUP MODE
The Setup Mode allows the user to set all needed characteristics of the meter.
To enter the mode, press the SETUP key and
enter the password when the device is in idle or
control mode.
If the correct password is not entered, the user can only view
the setup parameters (except for passwords) without modifying them (and the device remains in control mode). An
exception is certain setup items, or flags, which can activate
special tasks, when set and confirmed.
To each setup parameter (or setup item) is assigned a three
characters (one letter followed by two digits) setup code which
is entered and displayed on the secondary LCD. The first
character identifies the group of setup items, while the two
digits identify the particular item within that group.
The setup codes can be selected after password is entered
and CFM key is pressed. When CFM is pressed, the current
setup item is saved on EEPROM and the following item is
displayed.
The possible transitions in setup mode are the following:
ENTERING THE PASSWORD
• Press SETUP to enter the setup mode. The primary LCD will
• Enter the first digit of the pass-
• Then move to the next digit with and
24
display “0000”, while the secondary LCD shows “PAS”.
The first digit of the primary LCD will blink.
word by using the or key.
enter a digit as described above. Continue
for the last two digits.
NoteThe default password is set at “0000”.
ENTERING SETUP ITEMS
• When the whole password has been inserted, press CFM to confirm it.
After confirmation of the password the
primary LCD will show the name of the
first setup group (see table) while the
secondary LCD will display the setup
code of the first item of the group.
• By pressing the or key it is possible to cycle through
the setup groups; the secondary LCD will always show the
code of the first item of the group.
• Once a group is selected, it is possible to choose an item manually.
Pressing the key, the first digit of
the setup code will start blinking.
Now it is possible to change its value
by pressing the or key.
• Pressing the key again, the first
digit will be fixed while the second
digit starts blinking and its value can
be changed as described above.
• By pressing the key again all the
digits will be fixed.
• If CFM key is pressed, the selected item is confirmed; the
secondary LCD will show the setup item code while the
primary LCD will display the current item value.
If a fixed set of values is available for the selected item, use
the or key to switch between them.
25
Otherwise, if a numeric value has to be entered for the
item, use the or key to change the value of the blinking digit and the key to cycle through the number’s
digits.
• Once a value is set, press the CFM key to confirm. The
instrument will turn to the next item and the new item’s
value will be displayed on the primary LCD.
• If a wrong value is confirmed, the
WRONG indicator starts blinking,
the new value is not accepted and
the instrument will not switch to the
next item until a correct value will
be confirmed.
• Instead of selecting the item manually, it is possible to cycle
through all the items of a selected group by pressing repeatedly CFM key. The procedure to modify the item’s value
is the same described above.
If the last item of the group is reached, by pressing the
CFM key again the primary display will show the group
name and it will be possible to change the group by pressing the or key.
26
NoteAt any time for exiting the setup mode press the SETUP key. If
no modification has been confirmed, no setup item is
changed.
NoteWhile in the setup mode, if no activity is performed for about
5 minutes after entering the setup mode, the mode is automatically exited and the instrument returns to the previous
mode.
The below table lists the setup codes along with the description of
the specific setup items, their valid values and whether the item is
present for ORP mode.
Titles in Italic character (eg. “SETPOINT 1 “) represent different
subgroups within the same group. The subgroup name (eg.
“Set1”) is showed when entering the first item of the subgroup. To scroll through subgroups, use the key.
Depending on the device model, some of the below items or
item values may not be available.
CodeValid ValuesDefault Present
for ORP
GENERAL (“GEnE”)
G.00 pH/ORP input“PH”, “OrP” (see note 10)“PH”yes
G.01 Temperature compensation“AtC”: Automatic“AtC”no
“USEr”: Manual (see notes 9, 25)
G.02 Manual or probe error-30 to 130.0 ºC25.0no
temperature(see notes 9, 25)
G.10 Factory ID0000 to 9999 (see note 20)0000yes
G.11 Process ID alias RS485 address 0 0 to 99 (see note 20)00yes
G.12 Hardware Identifier7 digits identification number, read onlyyes
(see notes 4, 20)
G.98 Calibration/hold password0000 to 9999 (see notes 2, 20) 0000yes
G.99 General password0000 to 9999 (see notes 2, 20) 0000yes
CONTROL (“CtrL”)
C.00 Control enable“OFF”: Control mode disabled“OFF”yes
“On”: Control mode enabled
SETPOINT 1 (“SEt1”)
C.10 Setpoint 1 mode (M1)“OFF”: Disabled“OOHI”yes
(see note 1)“OOHI”: ON-OFF, high set point
“OOLO”: ON-OFF, low set point
“PIdH”: P.I.D., high set point / “PIdL”: P.I.D., low set point
C.11 Setpoint 1 (S1)-2.00 to 16.00 pH or8.00 pH oryes
(see note 1)-2000 to 2000 mV500 mV
27
CodeValid ValuesDefault Present
for ORP
C.12 Hysteresis for setpoint 1 (H1) 0.00 to 18.00 pH or1 pH oryes
(see note 1)0 to 4000 mV50 mV
C.13 Deviation for setpoint 1 (D1)0.50 to 18.00 pH or1 pH oryes
(see note 1)25 to 4000 mV50 mV
C.14 Reset time for setpoint 10.1 to 999.9 minutes999.9 minsyes
C.15 Rate time for setpoint 10.0 to 999.9 minutes0.0 minsyes
SETPOINT 2 (“SEt2”)
C.20 Setpoint 2 mode (M2)“OFF”: Disabled“OOLO”yes
(see note 1)“OOHI”: ON-OFF, high set point
“OOLO”: ON-OFF, low set point
“PIdH”: P.I.D., high set point
“PIdL”: P.I.D., low set point
C.21 Setpoint 2 (S2)-2.00 to 16.00 pH or6.00 pH oryes
(see note 1)-2000 to 2000 mV-500 mV
C.22 Hysteresis for setpoint 2 (H2) 0.00 to 18.00 pH or1 pH oryes
(see note 1)0 to 4000 mV50 mV
C.23 Deviation for setpoint 2 (D2)0.50 to 18.00 pH or1 pH oryes
(see note 1)25 to 4000 mV50 mV
C.24 Reset time for setpoint 20.1 to 999.9 minutes999.9 minsyes
C.25 Rate time for setpoint 20.0 to 999.9 minutes0.0 minsyes
ALARM (“ALAr”)
C.30 Alarm relay delta value0.50 to 18.00 pH1.00 pHyes
for setpoint 1 (A1)oror
(see notes 1, 22)25 to 4000 mV100 mV
C.31 Alarm relay delta value0.50 to 18.00 pH1.00 pHyes
for setpoint 2 (A2)oror
(see notes 1, 22)25 to 4000 mV100 mV
C.32 Maximum relay ON time1-60 minutes60yes
(after that an alarm is generated)
C.33 Alarm mask time00:00 to 30:00 minutes00:30yes
PROGRAMMABLE CONTROL TIMING (“tIME”)
C.41 Hold time start00:00 to 23:59 (see note 8)00:00yes
C.42 Hold time stop00:00 to 2 3:59 (see note 8)00:00yes
28
CodeValid ValuesDefault Present
for ORP
C.51 Monday hold mode“OFF”: Disabled“OFF”yes
enable“On”: Enabled
C.52 Tuesday hold mode“OFF”: Disabled“OFF”yes
enable“On”: Enabled
C.53 Wednesday hold mode“OFF”: Disabled“OFF”yes
enable“On”: Enabled
C.54 Thursday hold mode“OFF”: Disabled“OFF”yes
enable“On”: Enabled
C.55 Friday hold mode“OFF”: Disabled“OFF”yes
enable“On”: Enabled
C.56 Saturday hold mode“OFF”: Disabled“OFF”yes
enable“On”: Enabled
C.57 Sunday hold mode“OFF”: Disabled“OFF”yes
enable“On”: Enabled
C.60 Proportional control mode01:00 - 30:00 minutes05:00yes
period
C.70 Hold mode end delay00 to 99 seconds00yes
OUTPUT (“OutP”)
RELAYS (“rELA”)
O.01 Relay 1 mode“OFF”: Disabled“SEt1”yes
(see notes 19, 21)“SEt1”: Control, setpoint 1
“SEt2”: Control, setpoint 2
“SCLE”: Simple cleaning
“HOLd”: Energized in hold mode
O.02 Relay 2 mode“OFF”: Disabled“SEt2”yes
(see notes 19, 21)“SEt1”: Control, setpoint 1
“SEt2”: Control, setpoint 2
“SCLE”: Simple cleaning
“HOLd”: Energized in hold mode
O.03 Relay 3 mode“OFF”: Disabled“OFF”yes
(see notes 15, 19)“SCLE”: Simple cleaning
“ACLE”: Advanced cleaning / “HOLd”: Energized in hold mode
O.04 Relay 4 mode“OFF”: Disabled“OFF”yes
(see notes 15, 19)“SCLE”: Simple cleaning
“ACLE”: Advanced cleaning / “HOLd”: Energized in hold mode
29
CodeValid ValuesDefault Present
for ORP
0.05 Hold digital output“OFF”: Disabled“HOLd”yes
“HOLd”: Enabled upon hold mode
ANALOG OUTPUT #1 (“Out1”)
O.10 Analog output 1 mode“rECO”: Recorder“rECO”yes
(see notes 6, 21)“SEt”: Control, setpoint 1
O.11 Analog output 1 type“0-20”: 0-20 mA4-20 (6)yes
“4-20”: 4-20 mA
O.12 Measurement value for-2.00 to 16.00 pH0.00 pH oryes
analog output 1 minimumor -2000 to 2000 mV-2000 mV (6)
(O_VARMIN1)(O_VARMIN1 R O_VARMAX1-1 pH or 50 mV,
O_VARMIN1 R O_HOLD1R O_VARMAX1)
O.13 Measurement value for-2.00 to 16.00 pH14.00 pH oryes
analog output 1 maximum or -2000 to 2000 mV2000 mV (6)
(O_VARMAX1)(O_VARMIN1 R O_VARMAX1-1 pH or 50 mV,
O_VARMIN1 R O_HOLD1R O_VARMAX1)
O.14 Analog output 1 value upon “USEr”: User selected value“HOLd”yes
hold mode (see note 16)“HOLd”: Previous value is frozen
O.15 Analog output 1 value upon -2.00 to 16.00 pH7.00 pH oryes
hold mode if O.14=”USEr” or -2000 to 2000 mV0 mV (6)
(O_HOLD1)(it must be within the O_VARMIN1
to O_VARMAX1-1 interval)
ANALOG OUTPUT #2 (“Out2”)
O.20 Analog output 2 mode“rECO”: Recorder (temperature) “rECO” (pH)yes
(see notes 21, 26)“SEt”: Control, setpoint 2 (pH)or “OFF” (ORP)
or
“OFF”: Disabled
“SEt”: Control, setpoint 2 (ORP)
O.21 Analog output 2 type“0-20”: 0-20 mA“4-20”yes
“4-20”: 4-20 mA
O.22 Measurement value for-30 to 130.0 ºC0 ºCno
analog output 2 minimum(O_VARMIN2 R O_VARMAX2-10 ºC,
(O_VARMIN2)O_VARMIN2 R O_HOLD2R O_VARMAX2)
O.23 Measurement value for-30 to 130.0 ºC+100.0 ºCno
analog output 2 maximum (O_VARMIN2 R O_VARMAX2-10 ºC,
(O_VARMAX2)O_VARMIN2 R O_HOLD2R O_VARMAX2)
30
Loading...
+ 86 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.