HI 710005/8Voltage adapter from 115 VAC / 12 VDC 800 mA (USA plug)
HI 710006/8Voltage adapter from 230 VAC / 12 VDC 800 mA (European plug)
ChecktempCPocket-size thermometer (range –50.0 to 150.0 ºC)
HI 76404NElectrode holder
HI 8427pH and ORP electrode simulator with 1 m (3.3') coaxial cable ending in female BNC
connectors
HI 931001pH and ORP electrode simulator with LCD and 1 m (3.3') coaxial cable ending in female BNC
connectors
HI 76312Platinum 4-ring conductivity/TDS probe with temperature sensor and 1 m (3.3') cable
HI 7662-TTemperature probe with 1 m (3.3') cable
HI 92000Windows® compatible software
HI 9200109 to 9-pin RS232 cable
RECOMMENDATIONS FOR USERS
Before using these products, make sure they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV
equipment, requiring the operator to follow all necessary steps to correct interferences.
The glass bulb at the end of the pH electrode is sensitive to electrostatic discharges. Avoid touching this glass
bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic
discharges.
Any variation introduced by the user to the supplied equipment may degrade the instruments’ EMC
performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed
24 VAC or 60 VDC.
To avoid damage or burns, do not perform any measurement in microwave ovens.
Plastic-body, gel-filled, combination platinum ORP electrode.
Use: general purpose.
Dear Customer,
Thank you for choosing a Hanna Instruments product. This manual will provide you with the necessary
information for correct use of the instrument.
Please read this instruction manual carefully before using the instrument.
These instruments are in compliance with directives.
WARRANTYWARRANTY
WARRANTY
WARRANTYWARRANTY
HI 4430B
Plastic-body, gel-filled, combination gold ORP electrode.
Use: general purpose.
Consult the Hanna General Catalog for more electrodes with screw-type or BNC connectors.
EXTENSION CABLE FOR SCREW-TYPE ELECTRODES
(SCREW TO BNC ADAPTER)
HI 7855/1Extension cable 1 m (3.3') long
HI 7855/3Extension cable 3 m (9.9') long
114
HI 4521 and HI 4522 are warranted for two years against defects in workmanship and materials when used
for their intended purpose and maintained according to instructions. Electrodes and probes are guaranteed for
six months. This warranty is limited to repair or replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report
the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered
by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna
Instruments, first obtain a Returned Goods Authorization number from the Technical Service Department and
then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for
complete protection.
To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of
purchase.
Plastic-body, double junction, refillable, combination pH electrode.
Use: general purpose.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully to make sure that no damage has
occurred during shipping. If there is any damage, notify your dealer.
The meters are supplied complete with:
• HI 1131B Glass-body Combination pH Electrode
• HI 76312 Four-ring Conductivity Probe with built-in temperature sensor and ID
• HI 7662-T Temperature probe
• pH 4.01 & 7.01 Buffer solutions, 20 mL each
• HI 700661 Cleaning solution, 2 x 20 mL
• HI 7071S Electrolyte solution
• HI 76404N Electrode Holder
• 12 Vdc Power Adapter
• Instruction Manual
HI 4521 and HI 4522 are supplied with 12 Vdc/230 Vac adapter.
HI 4521-01 and HI 4522-01 are supplied with 12 Vdc/115 Vac adapter.
Plastic-body, double junction, refillable, combination pH electrode.
Use: general purpose for food industry.
112
Note: Save all packing material until you are sure that the instrument works properly. Any defective item
must be returned in the original packing with the supplied accessories.
5
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
HI 4521 and HI 4522 are professional bench meters with color graphic LCD for pH, ORP (Oxidation Reduction
Potential), ISE (HI 4522 only), conductivity, resistivity, TDS, salinity and temperature measurements.
The display can be configured as a single channel or dual channel display in various modes: Basic information
only, GLP information, Graph and Log History mode.
The main features of the instruments are:
• Two input channels;
• Eight measurement parameters: pH, mV, ISE (HI 4522 only), conductivity, resistivity, TDS, salinity and
temperature;
• Manual selection, automatic and semiautomatic pH calibration in up to five points, with standard (1.68,
3.00, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45) and custom buffers (up to 5 custom buffers);
• Manual Selection and Custom Standard ISE calibration in up to five points, with standard (up to five standard
solutions for each measurement unit) and custom solutions (up to 5 custom solutions), with or without
temperature compensation (HI 4522 only);
• Pure water checking using the USP <645> standard;
• Conductivity probe automatic recognition;
• Automatic or custom standard conductivity calibration in up to four points, probe offset calibration;
• One fixed point salinity calibration (Percent Scale only);
• AutoHold feature to freeze first stable reading on the LCD;
• Two selectable alarm limits (for pH, mV, ISE, conductivity, resistivity, TDS, salinity);
• Three selectable logging modes: Automatic, Manual, AutoHold logging;
• Up to 100 logging lots for automatic or manual modes and up to 200 USP reports, up to 100 ISE methods
reports;
• Selectable area and sampling period feature for automatic logging;
• GLP feature;
• Online and offline graph;
• User-friendly interface on large color graphic LCD (240 x 320 pixels);
• Opto-isolated PC interface via RS232, respectively USB.
HI 1131B
Glass-body, single junction, refillable, combination pH electrode.
Use: general purpose.
9) BNC electrode connector for pH/ORP/Ion measurements
10) Reference input socket
7
KEYBOARD DESCRIPTION
FUNCTION KEYS
To enter/exit calibration mode;
To select the desired measurement mode, pH, mV, Rel mV, ISE (HI 4522 only), Conductivity,
Resistivity, TDS, Salinity;
HI 7035M 111800 µS/cm (µmho/cm), 230 mL bottle
HI 7035L 111800 µS/cm (µmho/cm), 500 mL bottle
HI 8035L 111800 µS/cm (µmho/cm), 500 mL FDA approved bottle
HI 7037L 100% NaCl sea water standard solution, 500 mL
ELECTRODE STORAGE SOLUTIONS
To enter Setup (System Setup, pH Setup, mV Setup, ISE Setup (HI 4522 only), Conductivity
Setup, Resistivity Setup , TDS Setup or Salinity Setup) and to access Log Recall function;
To obtain general information about the selected option / operation.
VIRTUAL KEYS
The upper row keys are assigned to the virtual keys placed on the bottom of the LCD, which allow you to
,
perform the displayed function, depending on the current menu (e.g.
Measure
mode).
Note: All the virtual keys are assigned to the focused channel (with
LCD GENERAL DESCRIPTION
and in
key exception).
HI 70300LStorage Solution, 500 mL bottle
HI 80300LStorage Solution in FDA approved bottle, 500 mL
ELECTRODE AND PROBE CLEANING SOLUTIONS
HI 70000P Electrode Rinse Sachets, 20 mL, 25 pcs
HI 7061LGeneral Purpose Solution, 500 mL bottle
HI 7073LProtein Cleaning Solution, 500 mL bottle
HI 7074LInorganic Cleaning Solution, 500 mL bottle
HI 7077LOil & Fat Cleaning Solution, 500 mL bottle
HI 8061LGeneral Purpose Solution in FDA approved bottle, 500 mL
HI 8073LProtein Cleaning Solution in FDA approved bottle, 500 mL
HI 8077LOil & Fat Cleaning Solution in FDA approved bottle, 500 mL
ELECTRODE REFILL ELECTROLYTE SOLUTIONS
HI 70713.5M KCl + AgCl Electrolyte, 4x30 mL, for single junction electrodes
HI 70721M KNO3 Electrolyte, 4x30 mL
HI 70823.5M KCl Electrolyte, 4x30 mL, for double junction electrodes
HI 80713.5M KCl + AgCl Electrolyte in FDA approved bottle, 4x30 mL, for single junction electrodes
HI 80721M KNO3 Electrolyte in FDA approved bottle, 4x30 mL
HI 80823.5M KCl Electrolyte in FDA approved bottle, 4x30 mL, for double junction electrodes
HI 80931M KCl + AgCl Electrolyte in FDA approved bottle, 4x30 mL
ORP PRETREATMENT SOLUTIONS
HI 7020LTest Solution 200-275 mV, 500 mL bottle
HI 7021LTest Solution 240 mV, 500 mL bottle
HI 7022LTest Solution 470 mV, 500 mL bottle
HI 7091LReducing Pretreatment Solution, 500 mL
HI 7092LOxidizing Pretreatment Solution, 500 mL
HI 6001pH 1.679 Buffer Solution, 500 mL bottle
HI 6003pH 3.000 Buffer Solution, 500 mL bottle
HI 8004LpH 4.01 Buffer Solution in FDA approved bottle, 500 mL
HI 6004pH 4.010 Buffer Solution, 500 mL bottle
HI 8006LpH 6.86 Buffer Solution in FDA approved bottle, 500 mL
HI 6068pH 6.862 Buffer Solution, 500 mL bottle
HI 8007LpH 7.01 Buffer Solution in FDA approved bottle, 500 mL
HI 6007pH 7.010 Buffer Solution, 500 mL bottle
HI 6091pH 9.177 Buffer Solution, 500 mL bottle
HI 8009LpH 9.18 Buffer Solution in FDA approved bottle, 500 mL
HI 8010LpH 10.01 Buffer Solution in FDA approved bottle, 500 mL
HI 6010pH 10.010 Buffer Solution, 500 mL bottle
HI 6124pH 12.450 Buffer Solution, 500 mL bottle
CONDUCTIVITY BUFFER SOLUTIONS
HI 70033P 84 µS/cm (µmho/cm), 20 mL sachets (25 pcs.)
HI 7033M 84 µS/cm (µmho/cm), 230 mL bottle
HI 7033L 84 µS/cm (µmho/cm), 500 mL bottle
HI 8033L 84 µS/cm (µmho/cm), 500 mL FDA approved bottle
HI 70031P 1413 µS/cm (µmho/cm), 20 mL sachets (25 pcs.)
HI 7031M 1413 µS/cm (µmho/cm), 230 mL bottle
HI 7031L 1413 µS/cm (µmho/cm), 500 mL bottle
HI 8031L 1413 µS/cm (µmho/cm), 500 mL FDA approved bottle
HI 70039P 5000 µS/cm (µmho/cm), 20 mL sachets (25 pcs.)
HI 7039M 5000 µS/cm (µmho/cm), 230 mL bottle
HI 7039L 5000 µS/cm (µmho/cm), 500 mL bottle
HI 8039L 5000 µS/cm (µmho/cm), 500 mL FDA approved bottle
HI 70030P 12880 µS/cm (µmho/cm), 20 mL sachets (25 pcs.)
HI 7030M 12880 µS/cm (µmho/cm), 230 mL bottle
HI 7030L 12880 µS/cm (µmho/cm), 500 mL bottle
HI 8030L 12880 µS/cm (µmho/cm), 500 mL FDA approved bottle
HI 7034M 80000 µS/cm (µmho/cm), 230 mL bottle
HI 7034L 80000 µS/cm (µmho/cm), 500 mL bottle
HI 8034L 80000 µS/cm (µmho/cm), 500 mL FDA approved bottle
The resistance of glass electrodes partially depends on the temperature. The lower the temperature, the higher the
resistance. It takes more time for the reading to stabilize if the resistance is higher. In addition, the response time
will suffer to a greater degree at temperatures below 25 ºC (77 ºF).
10
Since the resistance of the pH electrode is in the range of 50 – 200 Mohm, the current across the membrane is in the pico
Ampere range. Large currents can disturb the calibration of the electrode for many hours.
For these reasons high humidity environments, short circuits and static discharges are detrimental to a stable
pH reading.
The pH electrode’s life also depends on the temperature. If constantly used at high temperatures, the
electrode life is drastically reduced.
Typical Electrode Life
Ambient Temperature1 – 3 years
90 ºC (194 ºF)Less than 4 months
120 ºC (248 ºF)Less than 1 month
Alkaline Error
High concentrations of sodium ions interfere with readings in alkaline solutions. The pH at which the interference
starts to be significant depends upon the composition of the glass. This interference is called alkaline error and
causes the pH to be underestimated. Hanna’s glass formulations have the indicated characteristics.
Sodium Ion Correction for the Glass at 20-25 ºC (68-77 ºF)
ConcentrationpH Error
Plug the 12 VDC adapter into the power supply socket.
Note: These instruments use non volatile memory to retain the meter settings, even when unplugged.
ELECTRODE AND PROBE CONNECTIONS
For pH or ORP measurements connect a pH / ORP electrode with internal reference to the BNC connector
located on the rear panel of the instrument.
For ISE measurements (HI 4522) connect, an ISE electrode with internal reference to the BNC connector
located on the rear panel of the instrument.
For electrodes with a separate reference connect the electrode’s BNC to the BNC connector and the electrode’s
reference to the reference input socket.
For temperature measurements and automatic temperature compensation connect the temperature probe to the
appropriate socket (mV channel only).
For conductivity, resistivity, TDS or salinity measurements connect a conductivity probe to the DIN connector
located on the rear panel of the instrument.
INSTRUMENT START UP
• Turn the instrument on from the power switch located on the rear panel of the instrument.
• Please wait until the instrument finishes the initialization process.
Note: It is normal for the loading process to take a few seconds. If the instrument doesn’t display the next
screen, restart the meter using the power switch. If the problem persists, contact your dealer.
If the electrode does not respond to pH changes, the battery run down and the electrode should be replaced.
MEASURE
Rinse the pH electrode tip with distilled water. Immerse the tip (bottom 4 cm /1½”) in the sample and
stir gently for a few seconds.
For a faster response and to avoid cross-contamination of the samples, rinse the electrode tip with a few drops
of the solution to be tested, before taking measurements.
STORAGE PROCEDURE
To minimize clogging and assure a quick response time, the glass bulb and the junction of pH electrode should
be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of HI 70300 or HI 80300 Storage Solution or, in its
absence, Filling Solution (HI 7071 or HI 8071 for single junction and HI 7082 or HI 8082 for double junction
electrodes). Follow the Preparation Procedure on page 103 before taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable used for connection to the instrument must be intact and there
must be no points of broken insulation on the cable or cracks on the electrode stem or bulb. Connectors must
be perfectly clean and dry. If any scratches or cracks are present, replace the electrode. Rinse off any salt
deposits with water.
pH Probe Maintenance
• Press while in
Measure
mode to access channel
selection menu. Four available options will be displayed:
Channel 1, Channel 2, or multi-channel with the first or the
second channel focused. The “Choose Channel Configuration” message is displayed in the Reminder messages area
when
is pressed.
• Select the desired option by pressing the appropriate key:
, , or . The instru-
ment will enter in the selected option
Measure
mode.
For refillable electrodes:For refillable electrodes:
For refillable electrodes:
For refillable electrodes:For refillable electrodes:
Refill the reference chamber with fresh electrolyte (HI 7071 or HI 8071 for single junction or HI 7082 or
HI 8082 for double junction electrodes). Allow the electrode to stand upright for 1 hour. Follow the Storage
Procedure above.
pH CLEANING PROCEDURE
General
•
•
Protein
•
Inorganic
Oil/grease
•
Soak in Hanna HI 7061 or HI 8061 General Cleaning Solution for approximately ½ hour.
Soak in Hanna HI 7073 or HI 8073 Protein Cleaning Solution for 15 minutes.
Soak in Hanna HI 7074 Inorganic Cleaning Solution for 15 minutes.
Rinse with Hanna HI 7077 or HI 8077 Oil and Fat Cleaning Solution.
IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thoroughly with distilled
water, refill the reference chamber with fresh electrolyte (not necessary for gel-filled electrodes) and soak the
electrode in HI 70300 or HI 80300 Storage Solution for at least 1 hour before taking measurements.
104
13
DISPLAYING MODESDISPLAYING MODES
DISPLAYING MODES
DISPLAYING MODESDISPLAYING MODES
For each measurement mode (pH, mV, Rel mV, Ion, Conductivity, Resistivity, TDS or Salinity) the following
display configurations are available: Basic, Good Laboratory Practice (GLP), Graph and Log History.
Basic
Accessing this option, the measured value and its units are
displayed on the LCD, along with the temperature value, temperature compensation mode, and minimal GLP data.
To choose the Basic displaying mode:
• Press
Configuration” message will be displayed in the Reminder
messages area.
• Press
mation for the selected measurement mode.
GLP
Accessing this option, a detailed GLP data will be displayed on
the LCD for pH / ISE: Last Calibration Date and Time, Offset and
Slope values, Calibration Buffers, Electrode Condition and for
conductivity / salinity: Last Calibration Date and Time, Calibration Standards, Cell Constant, Probe Offset, Reference Temperature,
Compensation Coefficient, Temperature Compensation.
Note: If only a one-point pH calibration is performed or the current
calibration does not include at least two consecutive standard
buffers of pH 4.01, 7.01 (6.86) and 10.01 (9.18) buffers, the
Electrode Condition will be unknown.
To access the GLP displaying mode:
• Press while in
Configuration” message will be displayed in the Reminder
messages area.
• Press
data.
while in
. The instrument will display the basic infor-
. The instrument will display the detailed GLP
Measure
mode. The “Choose Display
Measure
mode. The “Choose Display
ELECTRODE CONDITIONINGELECTRODE CONDITIONING
ELECTRODE CONDITIONING
ELECTRODE CONDITIONINGELECTRODE CONDITIONING
& MAINTENANCE& MAINTENANCE
& MAINTENANCE
& MAINTENANCE& MAINTENANCE
PREPARATION PROCEDURE
Remove the protective cap of the pH electrode.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal with electrodes. They will disappear
when rinsed with water.
During transport, tiny bubbles of air may form inside the glass bulb affecting proper functioning of the electrode.
These bubbles can be removed by “shaking down” the electrode as you would do with a glass thermometer.
If the bulb and/or junction is dry, soak the electrode in HI 70300 or HI 80300
For refillable electrodes:For refillable electrodes:
For refillable electrodes:
For refillable electrodes:For refillable electrodes:
If the filling solution (electrolyte) is more than 2½ cm (1”) below the fill hole, add HI 7082 or HI 8082 3.5M
KCl Electrolyte Solution for double junction or HI 7071 or HI 8071 3.5M KCl+AgCl Electrolyte Solution for
single junction electrodes.
For faster response, unscrew the fill hole screw during measurements.
Rinse the conductivity probe with distilled water. Immerse the tip (bottom 4 cm /1½”) in the sample and
stir gently for a few seconds.
For a faster response and to avoid cross-contamination of the samples, rinse the probe with a few drops of the
solution to be tested, before taking measurements.
PERIODIC MAINTENANCE
Inspect the probe and the cable. The cable used for connection to the instrument must be intact and there must
be no points of broken insulation on the cable. Connectors must be perfectly clean and dry. Rinse off any salt
deposits with water.
If more cleaning is required, remove the probe sleeve and clean the probe with a cloth or a nonabrasive
detergent. Make sure to reinsert the sleeve onto the probe properly and in the right direction. After cleaning
the probe, recalibrate the instrument.
The platinum rings are sustained with glass. Take great care while handling the probe.
Graph
Accessing this option, the online graph with the currently logged
values (pH, mV, Rel mV, ISE respectively Conductivity, Resistivity, TDS, Salinity vs. Seconds) will be displayed.
If there is no active log, the previously logged data for the
selected parameter will be plotted.
Notes: • If no data were logged, the graph displaying mode
will not be accessible.
• If no automatic log is saved, the offline graph will
not be available.
To access the offline / online graph:
• Press
Configuration” message will be displayed in the Reminder
messages area.
• Press
When the onlinegraph is displayed:
• Use and to move the graph along horizontal (
while in
.
Measure
mode. The “Choose Display
Time
) axis.
IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thoroughly with distilled
water.
102
• Press to access the zoom menu for the vertical (
zooming vertical axis.
• Press to return to the main menu.
When the offlinegraph is displayed:
• Use the arrow keys to move the graph along the horizontal and vertical axes.
• Press to access the zoom menu for horizontal and vertical axes. Use , or /
/ / respectively / / / to switch
between the active zooming axes. Press or to zoom the selected axis.
Note: While in zoom graph menu the key is not accessible.
• Press to return to the main menu.
15
Parameter
) axis. Use or for
Log History
PMETSREFFUBHp
CºKºFº976.1000.3010.4268.6010.7771.9010.01454.21
037223076.1270.3700.4289.6031.7954.9613.01973.31
587214076.1150.3200.4949.6890.7193.9542.01871.31
0138205176.1330.3000.4129.6070.7823.9081.01589.21
5188295376.1910.3100.4798.6640.7372.9811.01997.21
0239286576.1800.3400.4878.6720.7222.9260.01126.21
5289277976.1000.3010.4268.6010.7771.9010.01054.21
0330368386.1599.2710.4158.6899.6731.9269.9682.21
5380359886.1199.2620.4248.6989.6801.9919.9821.21
04313401396.1099.2730.4738.6389.6960.9188.9879.11
54813311007.1099.2940.4438.6979.6040.9748.9438.11
05323221707.1199.2260.4438.6879.6410.9718.9796.11
55823131517.1399.2670.4638.6979.6099.8397.9665.11
06333041427.1599.2190.4938.6289.6969.8377.9244.11
56833941437.1899.2701.4448.6789.6849.8757.9323.11
07343851447.1000.3321.4058.6399.6929.8647.9112.11
57843761557.1200.3931.4758.6100.7019.8047.9401.11
08353671767.1300.3651.4568.6010.7198.8837.9300.11
58853581087.1200.3271.4378.6910.7178.8047.9809.01
09363491397.1000.3781.4088.6920.7158.8847.9918.01
59863302708.1699.2202.4888.6040.7928.8957.9437.01
Accessing this option, last logged records will be displayed on the LCD.
The log history list also contains the appropriate main parameter
values, the logged temperature, the temperature compensation
source / mode, as well as the records time stamp.
To access the Log History displaying mode:
• Press
while in
Measure
mode. The “Choose Display
Configuration” message will be displayed in the Reminder
messages area.
• Press. The instrument will display the log history
regarding the selected measure mode.
Notes: • When an alarm condition is active, all logged records will have an exclamation mark (!).
• When a meter is in auto-hold, the logged records will have an “H” symbol.
• If another measure mode is selected, the Log History will be cleared.
• If the temperature unit is changed, all logged temperature values will be automatically displayed
in the new temperature unit.
pp
H BUFFER TEMPERATUREH BUFFER TEMPERATURE
p
H BUFFER TEMPERATURE
pp
H BUFFER TEMPERATUREH BUFFER TEMPERATURE
DEPENDENCEDEPENDENCE
DEPENDENCE
DEPENDENCEDEPENDENCE
Temperature has an effect on pH. The calibration buffer solutions are affected by temperature changes to a
lower degree than normal solutions.
During calibration the instrument will automatically calibrate to the pH value corresponding to the measured
or set temperature.
16
During calibration the instrument will display the pH buffer value at 25 ºC.
101
PC INTERFACEPC INTERFACE
PC INTERFACE
PC INTERFACEPC INTERFACE
SYSTEM SETUPSYSTEM SETUP
SYSTEM SETUP
SYSTEM SETUPSYSTEM SETUP
Data transmission from the instrument to the PC can be done with the HI 92000 Windows® compatible
software (optional). HI 92000 also offers graphing and on-line help feature.
Data can be exported to the most popular spreadsheet programs for further analysis.
HI 4521 and HI 4522 instruments have two available serial interfaces: RS232 and USB. The desired serial
interface can be selected from the settings window of the HI 92000 software.
If choosing the RS232 serial interface, use the optional Hanna HI 920010 cable connector to connect your
instrument to a PC. Make sure that your instrument is switched off and then plug one connector to the
instrument RS232 socket and the other one to the serial port of your PC.
Note: Other cables than HI 920010 may use a different configuration. In this case, communication between
instrument and PC may not be possible.
If choosing the USB serial interface, use a standard USB cable to connect your instrument to the PC.
For both serial interfaces, make sure that the instrument and the HI 92000 software have the same baud rate
and the appropriate communication port.
The System Setup menu allows the user to customize the user interface, consult the meter information, set the
external serial communication interface and to restore the manufacturer settings.
Accessing System Setup
• Press
• Press. The system setup options will be displayed
on the LCD.
To access a System Setup option:
• Use or to highlight the desired option.
• Press to access the selected option.
The following is a detailed description of the System Setup option
screen.
Beeper
This option allows the user to enable or disable the beeper. When the beeper is enabled, a specific beep will
be heard when the reading becomes stable, when an alarm condition is reached,
a wrong key is pressed.
while in
Measure
mode.
when pressing a key or if
Stability Indicator
When the reading becomes stable, the instrument delivers a medium beep only if this option is ON, along with the
“Stable” indicator on the LCD.
Alarm
If this option is ON, a continuous double beep will be heard each time the set limits in
exceeded, along with the “Alarm” indicator on the LCD.
Measure
mode are
Key Pressed
If this option is ON, a short beep will be heard each time a valid key is pressed.
Wrong Key
If this option is ON, a long beep will be heard when an incorrect key is pressed.
100
17
To set the Beeper:
• Use
• Press and use or to highlight the
desired beeper associated parameter you want to modify.
• Press and use or to highlight the
beeper status option.
• Press to confirm your selection and return to the Beeper
menu or press
or to select the Beeper option.
to return without changing.
highlighted lot. The “Please wait...” message will be displayed on the LCD for a short period. The user customised
report will be displayed on the LCD.
Note: For automatic logging only, it is possible to view the
plotted graph.
• Press to display the graph.
• By pressing it is possible to move the graph
along the horizontal or vertical axis with the arrow keys.
• If pressing while the graph is displayed, the
zoom menu for the horizontal and vertical axes will be
accessed. Press
, or / /
Saving Confirmation
When enabling this option, a prompt will appear on the LCD alerting
the user to save the modified values by pressing
without saving by pressing or canceling the saving operation
and return to the editing mode by pressing . If disabled, the
modified values will be saved automatically.
To enable / disable the saving confirmation:
• Use or to select the
• Press and use or to choose enabled
/ disabled.
• Press to confirm your selection or press
cancel operation.
GLP Data
This option allows the user to set general information which will appear in the log reports. The options can have
a max of 10 characters.
Operator ID
Instrument ID
Company Name
– this option allows you to edit the name of the operator.
– this option allows you to edit an identification name/number for the instrument.
– this option allows you to edit the company name.
Additional Info 1 & Additional Info 2
Saving Confirmation
– for general purpose notations.
, exiting
option.
to
/ / / /
/ to switch between the active
zooming axes and then zoom in or out on the selected
axis by pressing the appropriate virtual key.
• Press to return to the previous menu at any time.
To delete lots:
• Press while in
• Press or to access delete or delete all mode.
Otherwise, press to return to
• After selecting one of the deleting modes, use or
to select one lot and then press or
to delete the selected lot or all lots. The “Please
wait...” message will be displayed on the LCD until the
selected lot or all lots are deleted.
• Press and then press to exit deleting mode and
return to
Log Recall
• Press
Note: Logged lots should also be deleted whenever “Please Delete
Old Log Files” or “Low Data Logging Space” message appears
on the LCD, in the Reminder messages area.
to exit
view mode.
Log Recall
Log Recall
mode.
mode and return to
Log Recall
view mode.
Measure
mode.
18
99
set as Direct/AutoHold in order to use this logging mode.
To log data using this mode:
• Press while in
frozen on the LCD, after was pressed and the stability criterion reached.
• To store another frozen value, press to return to normal logging mode and then again.
• To stop the logging session, press or / .
Notes: • For the automatic logging, if the maximum logging time (24h) has been reached, a warning pop-
up will be displayed on the LCD in order to stop the current log and start another one in a new lot.
• If 100 lots have been saved or maximum 10000 records have been manually stored, a warning
pop-up will be displayed on the LCD in order to delete one lot or to select a new lot for the
manual logging to log other records.
LOG RECALL
This feature allows the user to view all stored data. If no data were
logged, the “No records were found” message will be displayed on
the LCD in the Log Recall screen. Otherwise, the instrument will
display all the memorized lots in accordance with the selected option:
Automatic Log, Manual Log ISE Method Report (HI 4522), or USP
Report.
To view the memorized data:
• Press
• Press. The “Choose Log Report Type” message will
be displayed in the Reminder messages area.
• Press, ,
the desired Log / ISE/ USP Report type. All logged lots for the
selected Log Report type will be displayed on the LCD.
• To filter the displayed lots, press and then the
desired unit (,,
lected measurement unit lots will be displayed on the LCD.
• Select the desired lot with
while in
, ,
to display the logged / report data from the
Measure
mode to start the logging session. The logged values are only the ones
Measure
mode.
,
or
or ). Only the se-
or and press
to select
(HI 4522),
To set the GLP Data:
• Use
• Press and use or to highlight the
desired option.
• Press to edit the desired information. The Text Editor menu
will be displayed on the LCD.
• Enter the desired information by using and to
highlight the desired character. It is also possible to delete the last
character by positioning the cursor on the Backspace character ( )
and pressing.
• Press to return to the
Confirmation
option, to escape without saving or to return
to the editing mode. Otherwise, the modified options are saved
automatically.
Date & Time
This option allows the user to set the current date & time and the format in which they appear.
or to select the
is enabled, press to accept the modified
GLP Data
GLP Data
option.
menu. If the
Saving
Set Date and Time
This option allows you to set the current date (year/month/day) and time (hour/minute/second).
Notes: • Only years starting with 2000 are accepted.
• The time is set using the selected time format. For 12 Hour time format only, the AM/PM can also
be selected with or .
Set Time Format
This option allows you to choose between 12 Hour (AM/PM) time format and 24 Hour time format from the
displayed pop-up menu.
Set Date Format
This option allows you to choose the desired date format from 6 available options: DD/MM/YYYY; MM/DD/YYYY;
YYYY/MM/DD; Mon DD, YYYY; DD-MM-YYYY and YYYY-Mon-DD.
To set the Date & Time:
• Use
• Press and use or to highlight the desired option you want to modify.
or to select the
Date& Time
option.
98
19
• Press to confirm your selection. Use and
and then use to modify the value with
or (for
other two options press to confirm your selection and
select one of the displayed formats with or .
• Press to confirm your selection and return to the
options.
& Time
• Press to return to the previous mode.
Notes: • For the
LCD Setup
This option allows the user to set the
parameter can be adjusted within 7 steps, while the
be set from 1 to 60 minutes or it can be OFF (disabled). All the changes are visible on the LCD for each parameter.
Note: If the instrument backlight is turned off after the set period of
To set the LCD Setup:
• Use or to select the
• Press and use key to highlight the desired
parameter.
• Use
• Press
Set Date and Time
enabled, press to accept the modified option,
to escape without saving or to return to the editing mode.
Otherwise, the modified option is saved automatically.
• If the time is changed with more than one hour before
last pH/ION calibration, a pop-up warning will appear on the LCD, notifying the user that a date/time
conflict has occured and some time-dependent modes could work improperly (e.g.
Log
).
time, press any key to turn it back on.
or to adjust the selected parameter.
to return to the
Set Date and Time
option, if the
Contrast
Saving Confirmation
, the
LCD Setup
System Setup
option). For the
Date
is
Backlight
Backlight
menu with saving.
of the LCD and the
parameter within 4 steps. The
option.
Backlight Saver
Backlight Saver
Measure, GLP
. The
Contrast
can
The “Logging”, sampling period and “AutoHold” indicators will be displayed on the LCD.
• To store another frozen value, press
• To stop the logging session, press / .
LOGGING MODE 3
This logging mode can be used for any sample measurements. By
choosing this logging mode, /
able in
Measure
mode.
To log data using this mode:
• Press while in
record. The “Logged” indicator will be displayed on the LCD.
• The records will be stored in one lot. In order to change the
logging lot, see the measured parameter setup for details, Log
option, New Lot generation.
,
LOGGING MODE 4
This logging mode can be used for multiple samples measurement. By choosing this logging mode,
/
To log data using this mode:
• Press /
• To store another frozen value, press to return to normal logging mode and then again.
• The records will be stored in one lot. In order to change the logging lot, see the measured unit
LOGGING MODE 5
This logging mode can be used for multiple samples measurement. By choosing this logging mode,
and will be available in
Notes: • The
and / will be available in
when the key was pressed. When the measured value is frozen on the LCD by pressing /
and the stability criterion is reached, the logged value is the one that has been frozen on the LCD.
details, Log option, New Lot generation.
mode, depending on the focused channel.
• If the Reading Mode option is set as Direct and the Logging Mode 5 session is started, a warning
pop-up will be displayed on the LCD, informing the user that the Reading Mode option must be
Measure
mode to manually log a
while in
Measure
Measure
or and or will be available in multichannel
/ again.
will be avail-
Measure
mode.
mode to manually log a record. Each value is logged at the time
mode.
Setup
Measure
for
20
97
• To stop the logging session, press / . The
Log Save screen will display thelog lot ID, the settable log
interval / sampling:
• Press to adjust the log interval and/or the log
sampling or press to save the current log in
the displayed format.
• Press to enter log interval edit menu and use
or to adjust the logging start-stop
time or the log sampling. Press to save the
current value and use or to adjust
next / previous parameter.
• Press
press to save the current log.
• While the instrument is saving the data, a “Please
wait...” pop-up message will be displayed on the LCD.
LOGGING MODE 2
This logging mode can be used for multiple samples measurement.
By choosing this logging mode, / and
to exit log interval edit menu and then
Language
This option allows the user to choose the desired language in which
all information will be displayed.
To select the Language:
• Use
• Press and use or to highlight the
desired language.
• Press to confirm your selection and return to the
Setup
menu without changing.
Serial Communication
This option allows the user to set the desired speed for the serial
communication (baud rate) between the instrument and PC from
1200, 2400, 4800 or 9600
To set the serial communication:
• Use or to select the
option.
• Press and use or to highlight the
desired baud rate.
• Press to confirm your selection and return to the
Setup
or to select the Language option.
menu or press
menu or press
to return to the System Setup
.
to return without changing.
System
Serial Communication
System
/ will be available in
To log data using this mode:
• Press / while in
logging session. When the measured value is frozen on the LCD
by pressing
reached, the logged value is the one that has been frozen on the
LCD until returning to normal logging mode by pressing .
/ and the stability criterion is
Measure
Measure
mode.
mode to start the
96
Note: The meter and the PC program must have the same baud rate.
21
Meter Information
edoMgniggoLepyTgniggoLedoMgnidaeR
1citamotuAtceriD
2citamotuAdloHotuA/tceriD
3launaMtceriD
4launaMdloHotuA/tceriD
5dloHotuAdloHotuA/tceriD
This option provides general information about the instrument serial
number (each instrument has an unique identification serial number),
the software version and the factory calibration date and time (for mV,
conductivity and temperature).
Note: All the instruments are factory calibrated for mV, conductivity
and temperature. After one year following factory calibration,
the “Factory Calibration Due” message will appear on the
LCD, in the Reminder messages area, notifying the user that
the instrument should be taken to the nearest Hanna Customer
Service for factory calibration.
To view the meter information:
• Use
or to select the
Meter Information
option.
LOGGINGLOGGING
LOGGING
LOGGINGLOGGING
This feature allows the user to log pH, mV, ISE (HI 4522), conductivity, resistivity, TDS, salinity and
temperature. The logging behaviour is dependent on the
Logging Type
and
Reading Mode
options from the
parameter setup.
The
Logging Data Configuration
options from the appropriate unit setup must be set first in order to be saved
into the log report.
The maximum number of logged records is 10.000/lot, the maximum logging time is 24h and up to 100 lots
can be saved.
Regarding data logging, the available logging modes are shown in the table below:
• Press to confirm and to view the meter information or press to return to the
menu.
Restore Factory Settings
This option allows the user to reset the instrument to the default
factory settings.
To restore the factory settings:
• Use or to select the
Restore Factory Settings
option.
• Press to confirm your selection. A pop-up box will be
displayed, asking for confirmation.
• Press to confirm your selection and return to the
or press to return without restoring defaults.
Setup
System
System Setup
LOGGING MODE 1
This logging mode can be used to monitor a chemical reaction. By choosing this logging mode,
be available in
Measure
mode.
To log data using this mode:
• Press / while in
Measure
mode to start the
logging session. The “Logging” and the Sampling Period indicators will be displayed on the LCD and data will be stored at the
set sampling period.
Note: While automatic logging is running, the measured pa-
rameter setup is not available. A warning message will
be displayed if the setup is accessed.
• If accessing Graph option while logging, the online graph can
be visualized on the LCD (see
Display Mode
section).
• If accessing Log History option while logging, last logged data
can be visualized on the LCD (see
Display Mode
).
will
22
95
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATION
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
pHpH
SETUP SETUP
pH
SETUP
pHpH
SETUP SETUP
The temperature user calibration menu can be accessed at the meter startup by pressing simultaniously three
keys as in the below drawing:
Note: The temperature user calibration is performed in three points: around 0°C, 50°C, 100°C.
To perform temperature user calibration:
• Select the desired temperature channel by pressing
(the temperature channel is switched between temperature EC
channel and temperature pH channel).
• Press to start the temperature calibration. Adjust the
temperature preset value using or when
necessary.
• Insert the probe in the beaker with water at 0°C.
• Wait to stabilize and then press to confirm the
calibration point.
• Repeat the previous three steps for 50°C and 100°C.
• Save the calibration.
• Press to return to measure mode.
The pH Setup menu allows the user to set the parameters associated with pH measurement and calibration.
Accessing pH Setup
• Press while in
select pH range for the desired channel.
• Press and then to access pH Setup menu.
To access a pH Setup option:
• Use or to highlight the desired option.
• Press
The following is a detailed description of the pH Setup option
screens.
Profile
Choosing this option the measuring and the calibration mode can be customized. Up to 10 profiles can be
defined by the user.
The available options are:
Save Current Profile:
Load Profile:
Delete Profile:
load from available profiles.
delete a profile.
Measure
mode and then to
to access the selected option.
save the current profile.
Save Current Profile
To save the current profile:
• Use or to select the
Profile
option.
Note: Press
if you want to clear the temperature user calibration.
94
• Press and then select
Text Editor
box will be displayed on the LCD.
• Enter the desired profile name by using
to highlight the desired character and then press to add it to
the text bar. It is also possible to delete the last character by positioning
the cursor on the Backspace character (
• Press to return to the Profile options.
Save Current Profile
and
) and pressing .
option. The
23
Note: The saved profile will automatically become the current profile.
A list with all customised profiles will be displayed
Profile
option.
on the screen.
• Use or to select the desired profile and press
to
confirm or
to exit without selecting.
Delete Profile
To delete one of the existing profiles:
• Use or to select the
• Press
Delete Profile
• Press
and use or to highlight the
option.
.
A list with all customised profiles will appear on
Profile
option.
the screen.
• Use or to select the desired profile and press
.
• Press
to return to the previous menu.
According to the definition, salinity of a sample in psu (practical salinity units) is calculated using the
following formula:
where:
RT - coefficient;
CT(sample) - uncompensated conductivity at T °C;
C(35,15)= 42914 µS/cm - the corresponding conductivity of KCl solution containing a mass of 32.4356 g
KCl / 1 Kg solution;
rT - temperature compensation polynom
Note: The formula can be applied for salinity values between 0 and 42 psu.
The formula can be applied for temperatures between -2 °C and 35 °C.
93
SALINITY CALIBRATIONSALINITY CALIBRATION
SALINITY CALIBRATION
SALINITY CALIBRATIONSALINITY CALIBRATION
Salinity calibration is a one-point calibration procedure at 100.0% NaCl. Use the HI 7037L calibra-
tion solution (sea water solution) as a 100% NaCl standard solution.
To enter salinity calibration:
• Set the meter for salinity range;
• Select the
• Rinse the probe with some of the calibration solution or deionized water;
• Immerse the probe into HI 7037L solution. The sleeve holes must be completely submerged. Tap
the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.
• Enter in calibration mode by pressing ;
• Wait to stabilize;
• Press
Three measurement scales are available for salinity (Natural Sea Water Scale, Practical Salinity Scale and
Percent Scale).
NATURAL SEA WATER SCALE (UNESCO 1966)NATURAL SEA WATER SCALE (UNESCO 1966)
NATURAL SEA WATER SCALE (UNESCO 1966)
NATURAL SEA WATER SCALE (UNESCO 1966)NATURAL SEA WATER SCALE (UNESCO 1966)
According to the definition, salinity of a sample in ppt is calculated using the following formula:
Percent Scale
to finish salinity calibration or press to cancel calibration.
(see
Salinity Setup
SALINITY MEASURESALINITY MEASURE
SALINITY MEASURE
SALINITY MEASURESALINITY MEASURE
section);
MENTMENT
MENT
MENTMENT
Temperature
The temperature has a direct influence on pH. This option allows the user to choose the temperature source and
units, as well as the desired manual temperature for manual temperature compensation mode.
Temperature Source
If using a temperature probe, Automatic Temperature Compensation will be performed relative to the displayed
temperature, with the “ATC 1 / 2” indicator displayed on the LCD. The ATC option can be selected for Channel 1
or Channel 2, in accordance with the configurated channel utilizing a temperature probe. If no temperature probe
is detected, Manual Temperature Compensation will be performed, with the “MTC” indicator on the LCD.
Temperature Unit
Accessing this option, the desired temperature unit can be chosen (Celsius, Fahrenheit or Kelvin degrees) and the
meter will automatically make the conversion for the selected unit.
Manual Temperature
If no temperature probe is connected or the
set manually.
To set one of the temperature options:
• Use or to select the
• Press and use or to highlight the
desired temperature option you wish to modify.
• Press and use or to highlight the
desired option (for
or to adjust the temperature value between
the displayed limits (for
• Press to confirm your selection (for
Temperature Source
Manual Temperature
Temperature Source
Temperature
&
Unit options
option.
) or use
option).
Temperature Source
is set as
&
Manual
, the desired temperature can be
where:
RT - coefficient;
CT(sample) - uncompensated conductivity at T °C;
C(35,15)= 42914 µS/cm - the corresponding conductivity of KCl solution containing a mass of 32.4356 g
KCl / 1 Kg solution;
rT - temperature compensation polynom.
Note: The formula can be applied for temperatures between 10 °C and 31 °C.
92
Unit
options) or press to save the current value (for
Temperature
Calibration
This option allows the user to set all the data regarding the pH calibration process.
option). Otherwise, press
to cancel operation.
Manual
Buffer Entry Type
Three entry modes for the pH buffers used for calibration are available:
Automatic– the instrument automatically selects the closest buffer to the measured pH value from the edited
buffer group.
Semiautomatic– the instrument automatically selects the closest buffers to the measured pH value from all
25
available buffers and you can choose the one used.
Manual Selection– the desired pH buffer is manually selected from all available buffers, regardless of measured value.
To select the Buffer Entry Type:
• Use
or to select the Calibration option.
TDS MEASURETDS MEASURE
TDS MEASURE
TDS MEASURETDS MEASURE
Make sure the TDS factor has been set before taking TDS measurements (see
MENTMENT
MENT
MENTMENT
TDS Setup
section).
• Press and use or to highlight the
Buffer Entry Type
• Press and use or to highlight the
desired option.
• Press to confirm your selection or press
cancel operation.
option.
to
1st Cal. Point
Two options are available for the 1st Cal.: Point and Offset.
If Point option is selected, the slope values adjacent to the calibration points will be reevaluated (normal calibration).
If at least a two-point calibration has been performed and an offset correction of the electrode is wanted
(maintaining the existing slope values), perform a one-point calibration using the Offset option.
To set the 1st Cal. Point:
• Use or to select the
• Press and use or to highlight the
1st Cal. Point
• Press and use or to highlight the
desired option.
• Press to confirm your selection or press
cancel operation.
option.
Calibration
option.
to
DIRECT MEASUREMENT
To measure the TDS of a sample using the
• Press and then to select TDS measure mode.
• Select the
• Proceed as for the conductivity measurement (see
Measurement
DIRECT/AUTOHOLD MEASUREMENT
To measure TDS of a sample using the
• Select the
section).
• Proceed as for the conductivity measurement. (see
ity Measurement
Direct
reading mode (see
section).
Direct / AutoHold
section)
reading mode (see
Direct
TDS Setup
section).
Direct / AutoHold
reading mode:
Conductivity
reading mode:
TDS Setup
Conductiv-
Edit Custom Buffers
If you want to use other buffers than the ones already memorized (standard buffers), the Edit Custom Buffers option
is available, allowing you to set the desired pH buffers. Up to five pH custom buffers can be set.
26
91
RESISTIVITY MEASURERESISTIVITY MEASURE
RESISTIVITY MEASURE
RESISTIVITY MEASURERESISTIVITY MEASURE
Make sure the instrument has been calibrated before taking resistivity measurements.
DIRECT MEASUREMENT
To measure the resistivity of a sample using the
mode:
• Press and then to select resistivity measure mode.
• Select the
• Proceed as for the conductivity measurement (see
Measurement
Direct
reading mode (see
section).
Resistivity Setup
Direct
reading
section).
Conductivity
MENTMENT
MENT
MENTMENT
To edit the Custom Buffers:
• Use
• Press and use or to highlight the
Edit Custom Buffers
• For a previous set value, press to set the custom buffer value
to “----” if desired and confirm the setting by pressing ,
otherwise press to edit the selected custom buffer.
• While in edit custom buffer menu press to set the
custom buffer value to 7.000 pH and then use or
• Press to exit custom buffer edit menu. If the Saving Confirmation is enabled, press to
or to select the
option.
to set the desired custom buffer value.
Calibration
option.
DIRECT/AUTOHOLD MEASUREMENT
To measure resistivity of a sample using the
mode:
• Select the
section).
• Proceed as for the conductivity measurement. (see
Measurement
Direct / AutoHold
section)
reading mode (see
Direct / AutoHold
Resistivity Setup
reading
Conductivity
accept the modified option,
Otherwise, the modified option is saved automatically.
• Use
key to select the next custom buffer to be set or press to return to Calibration options.
to escape without saving or to return to the editing mode.
Edit Buffer Group
Accessing this option the user can edit the desired group of five pH buffers for automatic buffer recognition
(Automatic Buffer Entry Type). If the Buffer Group already contains five pH buffers, at least one pH buffer has to
be removed in order to add another buffer.
To set the Buffer Group:
• Use or to select the
• Press and use or to highlight the
Edit Buffer Group
• Press and use and to choose the
pH buffer to be included in the buffer group.
• Press or to add / remove the selected pH
buffer to / from the buffer group.
• Press
option.
to return to Calibration options and to save the changes.
Calibration
option.
90
27
Calibration Reminder
Hp
ytivitcudnoC
)mc/Sµ(
Hp
ytivitcudnoC
)mc/Sµ(
0.57.41.64.2
1.51.42.65.2
2.56.33.64.2
3.53.34.63.2
4.50.35.62.2
5.58.26.61.2
6.56.27.66.2
7.55.28.61.3
8.54.29.68.3
9.54.20.76.4
0.64.2
In order to have accurate readings, the instrument must be calibrated frecquently. Three options are available for
the calibration reminder: Daily, Periodic or Disabled.
To set the Calibration Reminder:
• Use or to select the Calibration option.
• Press and use or to highlight the
Calibration Reminder
option.
• Press and use or to highlight the
desired option.
• Press to confirm your selection or press
to
cancel operation.
Set Reminder Period
If choosing Daily or Periodic options for the Calibration Reminder, the Set Remind Period must be accessed in order to set
the time interval until next calibration. The time interval between two calibrations can be set up to 1 day / 1 year for
Daily / Periodic options.
Note: If Set Reminder Period parameter is accessed and the Calibration Reminder is disabled, a warning
message appears on the LCD informing the user that the reminder period can be set only if the Calibration
Reminder is set as Daily or Periodic.
To set the Reminder Period:
• Use or to select the
Calibration
option.
• Press and use or to highlight the
Set Reminder Period
option.
• Press and use / to select next /
previous entry to be edit.
• Press and use or to set the desired
value, then press
• Press to return to the Calibration options. If the Saving
Confirmation is enabled, press
option, to escape without saving or to return to the editing mode. Otherwise, the modified
option is saved automatically.
to save the modified value.
to accept the modified
28
Stage 3 - this is an off-line test.
To perform this test:
• Take the water sample from the previous test and increase its ionic strength for a pH measurement at
25 °C;
•
Record the pH and round it to the nearest 0.1 pH;
• Look up the corresponding conducivity value measured in Stage 2 above;
• If the conductivity is lower than the conductivity from the below table, then the sample has met the USP
requirements. Otherwise, the water didn’t meet the USP requirements.
To access the USP menu:
• Select the
Direct / USP
reading mode (see
Conductivity Setup
);
• Return to measure mode;
• Press (USP) and then select the desired USP stage.
In this measure mode the user can check for ultra pure water using the United States Pharmacopeia standard
(USP <645>).
This USP standard consists of three stages (one in-line and two off-line tests) as following:
Stage 1 - this is an in-line test.
To perform this test:
• Measure the temperature of the water and the uncompensated conductivity readings. The measurement
may be performed in a suitable container or as in-line measurement.
• The temperature will be round down to the nearest 5 °C and look up the corresponding conductivity value
in the below table.
• If the measured conductivity is lower than the conductivity in the table, then the water meets the USP
requirements.
• Otherwise, proceed to Stage 2 testing.
Clear Calibration
Accessing this option, the existent pH calibration can be cleared. If the calibration is cleared, another calibration
has to be performed.
To clear Calibration:
• Use or to select the Calibration option.
• Press and use or to highlight the
Clear Calibration
option.
• Press to clear calibration. A pop-up menu will be
displayed asking for confirmation.
• Press to confirm or press to escape without
saving and return to the Calibration options.
Sample ID
This option allows the user to give to the measured samples an identification number/name. Two Sample ID
options are available:
ID Increment Mode
and
Edit Sample ID
.
ID Increment Mode
Two increment modes for the sample ID can be selected:
None – the sample ID will be fixed and it can be set alphanumerically.
Automatic – the sample ID will be increased with 1 for each new log lot.
To set the ID Increment Mode:
• Use or to select the
Sample ID
option.
Stage 2 - this is an off-line test.
To perform this test:
• Store the water sample in an enclosed clean container that has been rinsed previously with water of the
same quality.
• Adjust the sample’s temperature to 25 °C and agitate the sample to ensure that it has equilibrated with
ambient CO2.
• If the measured conductivity is less than 2.1 µS/cm, then the sample has met the USP requirements.
• Otherwise, proceed to Stage 3 testing.
88
• Press and use or to highlight the
ID Increment Mode
option.
• Press and use or to highlight the
desired option.
• Press
to confirm your selection or press
cancel operation.
29
to
Edit Sample ID
This option allows the user to edit the sample ID.
Note: If the ID Increment Mode is set to None, the sample ID can be set alphanumerically, otherwise it can be
set only numerically.
To edit the Sample ID:
• Use or to select the Sample ID option.
• Press and use or to highlight the
Edit Sample ID
• Press to confirm your selection.
• If the selected increment mode is None, the Text Editor menu will be
displayed on the LCD, allowing you to enter the desired sample
number/name by accepting the highlighted character which is added
in the text bar, using
help the user to select the desired character. It is also possible to
delete the last character by positioning the cursor on the Backspace
character and pressing.
• Press to return to Sample ID options. If the Saving
Confirmation is enabled, press to accept the modified
option, to escape without saving or to return
to the editing mode. Otherwise, the modified options are saved
automatically.
• If the selected increment mode is Automatic, the desired sample ID
value can be set using or .
• Press to save the current value or press
cancel operation.
option.
. The and keys
to
CONDUCTIVITY MEASUREMENTCONDUCTIVITY MEASUREMENT
CONDUCTIVITY MEASUREMENT
CONDUCTIVITY MEASUREMENTCONDUCTIVITY MEASUREMENT
Make sure the instrument has been calibrated before taking conductivity measurements.
DIRECT MEASUREMENT
To measure the conductivity of a sample using the Direct reading
mode:
• Press and then to select conductivity
measure mode.
• Select the
• Submerge the conductivity probe and tapit repeatedly to
remove any air bubbles that may be trapped inside the sleeve.
Allow time for the reading to stabilize.
• The measured conductivity value will be displayed.
Note: If the reading is out of range, “-----” will be displayed.
DIRECT/AUTOHOLD MEASUREMENT
To measure conductivity of a sample using the Direct / AutoHold
reading mode:
• Select the Direct / AutoHold reading mode (see
Setup
• If pressing , the “AutoHold” indicator will start blinking on the display until the stability criterion is reached. The
conductivity value will be frozen on the display, along with
“AutoHold” indicator.
• To return to normal measure mode press.
Direct
reading mode (see
Conductivity Setup
).
Conductivity
).
Stability Criteria
This option allows the user to select the signal stability criterion for the measured parameter (pH, mV, ISE):
Fast – this criterion will give faster results with less accuracy.
Medium – this criterion will give medium speed results with medium accuracy.
Accurate – this criterion will give slower results with high accuracy.
30
87
• Insert and rinse the probe in the first beaker in order to decontaminate it;
• Insert the probe in the second beaker;
• Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.
• Enter in calibration mode by pressing
• Wait to stabilize;
• When the automatic standard recognition is selected, the calibration point will be automatically selected
from the Hanna standard list (84 µS, 1413 µS, 5.0 mS,12.88 mS, 80.0 mS, 111.8 mS). The user can
also select the desired standard value by usingand;
• Otherwise (user standard), the pop-up will prompt for the custom standard value.
• Pressto finish the calibration orto abort calibration.
;
To set the Stability Criteria:
• Use
• Press and use or to highlight the
desired option.
• Press to confirm your selection or press
cancel operation.
or to select the
Stability Criteria
option.
to
Note: The calculated cell constant will be used for the whole range.
Multi Point Calibration
• Up to 4 calibration points can be performed in order to
increase the measurement accuracy on the wide range;
• Select the multi point calibration (see
Calibration
• Repeat the steps from the single point calibration for each
calibration point. The meter will calculate a cell constant
corresponding to each calibration point;
• Pressto exit calibration mode.
Note: For each range the corresponding cell constant will be displayed.
CELL CONSTANT CALIBRATION (edited by the user)
• A known value of the probe cell constant can be set by the user for the whole range (see
Setup
Note: When the user cell constant is used the old calibration (in solution) will be cleared.
);
->
Cell Constant
section)
Conductivity Setup
->
Conductivity
Reading Mode
This option allows the user to select between
pH reading mode. If choosing the second option, the current reading can be frozen on the LCD when is
pressed and if the stability criterion is reached.
To set the Reading Mode:
• Use or to select the
• Press and use or to highlight the
desired option.
• Press to confirm your selection or press to
cancel operation.
Log
This option allows the user to edit the logging settings:
and
New Lot.
Direct
and
Reading Mode
Direct / AutoHold
option.
Logging Type, Logging Data Configuration, Sampling Period
Logging Type
Three logging types are available:
logged automatically at constant time intervals while with the Manual type the user has the capability to log a
measured value by pressing . With the AutoHold type the stored values are only the ones frozen on the LCD
Automatic, Manual
and
AutoHold
. With the Automatic type the values are
86
31
after the logging session was started.
To set the Logging Type:
• Use
• Press and use or to highlight the
Logging Type
• Press and use or to highlight the
desired option.
• Press to confirm your selection or press to
cancel operation.
or to select the
option.
Log
option.
Logging Data Configuration
This option allows the user to select the parameters that accompany a logged value:
and
Sample ID, Instrument ID, Operator ID, Company Name, Aditional Info 1
To set the Logging Data Configuration:
• Use or to select the Log option.
• Press and use or to highlight the
Logging Data Configuration
• Press and use or to highlight the
desired parameter to be logged in file.
• Press to confirm your selection and use or
to enable the parameter by selecting Yes or to disable
it by selecting No.
• Press to confirm your selection or press
cancel operation.
option.
to
Aditional Info 2.
Date / Time, Calibration Data,
CONDUCTIVITY CALIBRATIONCONDUCTIVITY CALIBRATION
CONDUCTIVITY CALIBRATION
CONDUCTIVITY CALIBRATIONCONDUCTIVITY CALIBRATION
It is recommended to calibrate the instrument frequently, especially if high accuracy is required.
The conductivity range should be recalibrated:
• Whenever the conductivity probe is replaced.
• At least once a week.
• Before USP measurements.
• After testing aggressive chemicals.
• When calibration reminder is activated (“Conductivity Cal Expired”).
• If the readings are far from the calibration point.
Note:TDS and Resistivity readings are automatically derived from the conductivity readings and no specific
calibration is needed.
OFFSET CALIBRATION
The meter allows the user to calibrate the probe for the offset.
• Set the meter for conductivity range;
• Select the automatic standard recognition (see
• Leave the dry probe in the air;
• Enter in calibration mode by pressing;
• Wait to stabilize. The 0.000 µS calibration point will appear on the screen;
• Press to finish the probe offset calibration.
• Press to exit calibration mode or continue calibration in the other standard solutions.
Note: The offset calibration can be performed only if it is perfomed first (no other calibration points present).
Clear the old calibration if it is present.
CELL CONSTANT CALIBRATION (in solution)
Conductivity Setup
->
Calibration
);
32
Single Point Calibration
• Select the single point calibration (see
• Pour a small quantity of the standard solution into a clean beaker. If possible, use plastic beakers to
minimize any EMC interferences.
• For accurate calibration and to minimize cross-contamination, use two beakers for each standard solution.
One for rinsing the probe and one for calibration.
Conductivity Setup
85
->
Calibration
);
Clear Calibration
Accessing this option, the existent salinity calibration (
To clear calibration:
• Use or to highlight the
• Press to clear calibration. A pop-up menu will be displayed to ask for confirmation.
• Press to confirm or press to cancel operation.
Salinity Scale
The meter uses three salinity scales: Natural Sea Water 1966, Practical Scale 1978, Percent Scale [%].
To select the salinity measurement scale:
• Press
• Press .
• Use or to select the
• Press and use or to highlight the
desired option.
• Press to confirm your selection or press
cancel operation.
while in salinity measure mode.
Salinity Scale
Percent Scale
Clear Calibration
option.
) can be cleared.
option.
to
Sampling Period
This option allows the user to select the desired sampling period for
automatic logging type.
To set the Sampling Period:
• Use or to select the
• Press and use or to highlight the
Sampling Period
• Press and use or to select the
desired logging time interval.
• Press to confirm your selection or press to
cancel operation.
option.
Log
option.
New Lot
Accessing this option, a new manual lot can be generated next time you want to manually log a measured value.
Note: If New Lot option is accessed and the Logging Type is Automatic,
a warning message appears on the LCD informing the user
that a new lot can be created only if the Logging Type is set
as Manual.
To generate a New Lot:
• Use or to select the
Log
option.
Sample ID - see
Log - see
Alarm - see
three
Conductivity Setup
Conductivity Setup
Conductivity Setup
section.
section.
section.
84
• Press and use or to highlight the
New Lot
option.
• Press to generate a new manual lot. A pop-up menu
will be displayed asking for confirmation.
• Press to confirm or press to escape without
saving and return to the Log options.
33
Alarm
This option allows the user to edit the alarm settings:
a continuous double beep will be heard, along with the “Alarm” indicator blinking on the LCD, each time the set
limits in
Measure
mode are exceeded.
Alarm State
and
Alarm Limits
. If the
Alarm
option is enabled,
SALINITYSALINITY
SALINITY
SALINITYSALINITY
The Salinity Setup menu allows the user to set the parameters related to Salinity measurement and
calibration.
SETUP SETUP
SETUP
SETUP SETUP
Alarm State
Three modes are available for the
Disabled
– the alarm will be disabled.
Inside Limts
Outside Limits
To set the Alarm State:
• Use or to select the
• Press and use or to highlight the
• Press and use or to highlight the
• Press to confirm your selection or press
– the alarm will notify the user when the measured value is inside the set limits.
– the alarm will notify the user when the measured value is outside the set limits.
Alarm State
desired option.
cancel operation.
option.
Alarm State
option:
Alarm
option.
to
Alarm Limits
This option allows the user to set the alarm limits for the measured value.
Note: The Alarm High value can not be lower than the Alarm Low value.
To set the Alarm Limits:
• Use or to select the
• Press and use or to highlight the
Alarm Limits
• Press and use / to select next /
previous entry to be edit.
• Press
option.
and use or to set the desired
Alarm
option.
Accessing Salinity Setup
• Press
select
• Press and then
menu.
To access an
• Use or to highlight the desired option.
• Press to access the selected option.
The following is a description of the
Profile - see
Reading Mode - see
Temperature
This option allows the user to choose the temperature source and units.
To access a Temperature options:
• Use or to highlight the Temperature option.
• Press to access Temperature option.
• Press to confirm your selection and use or to enable the parameter by selecting
Yes or to disable it by selecting No.
while in
Salinity
range.
Salinity Setup
Conductivity Setup
Temperature Source, Temperature Unit
Measure
option:
Salinity Setup
section.
Resistivity Setup
mode and then to
to access
section.
- see
Salinity Setup
option screens.
Conductivity Setup
section.
value, then press
• Press to return to the Alarm options. If the Saving
to save the modified value.
34
83
TDS factor
Confirmation is enabled, press to accept the modified option, to escape without saving
With this option the user can set the TDS factor.
• Use or to highlight the TDS options.
• Press to confirm your selection and use or
to increase / decrease the value.
• Press to confirm your selection or press to
cancel operation.
Sample ID - see
Log - see
Alarm - see
Conductivity Setup
Conductivity Setup
Conductivity Setup
section.
section.
section.
or to return to the editing mode. Otherwise, the modified option is saved automatically.
Isopotential Point
This option allows the user to edit the isopotential point of the electrode used for pH measurements. The isopotential point
is the mV reading for an electrode at which temperature has no effect on the measurement. The ideal electrode has an
isopotential point of 0.0 mV and 7.00 pH, while an actual electrode tipically deviates slightly from the ideal values.
If the actual isopotential pH for an electrode is known, it can be set by accessing this option. The isopotential pH
limits (low and high) will be displayed on the LCD.
Note: If the isopotential point has been modified, recalibration must be performed for the pH channel.
To set the Isopotential Point:
• Use or to select the
• Press
• Press to save the current value or press
cancel operation.
pH Resolution
Accessing this option, the desired pH resolution can be set, with one (x.x), two (x.xx) or three (x.xxx) decimals.
To set the pH Resolution:
• Use or to select the
and set the desired isopotential pH value using
or .
Isopotential Point
pH Resolution
option.
option.
to
82
• Press and use or to highlight the
desired option.
• Press to confirm your selection or press
cancel operation.
35
to
mVmV
SETUP SETUP
mV
SETUP
mVmV
SETUP SETUP
The mV Setup menu allows the user to set the parameters associated with mV and Relative mV
measurements.
TDSTDS
SETUP SETUP
TDS
SETUP
TDSTDS
SETUP SETUP
The TDS Setup menu allows the user to set the parameters related to the TDS measurement.
Accessing mV Setup
• Press
• Press and then to access mV Setup menu.
To access a mV Setup option:
• Use or to highlight the desired option.
• Press
The following is a detailed description of the mV Setup option
screens.
Profile – See
Temperature – See
Sample ID – See
Stability Criteria – See
Reading Mode – See
Log – See
Alarm – See
Note: The Alarm Limits (Low and High) are set in mV units.
while in
to select mV / mV Rel range.
pH Setup
pH Setup
pH Setup
Measure
mode and then or
to access the selected option.
section.
pH Setup
pH Setup
section.
section.
pH Setup
pH Setup
section.
section.
section.
section.
Accessing TDS Setup
• Press while in
select TDS range.
• Press and then
To access a TDS Setup option:
• Use or to highlight the desired option.
• Press to access the selected option.
The following is a description of the TDS Setup option screens.
Profile - see
Reading Mode - see
Temperature - see
Units
This option allows the user to set the TDS measuring range
ppm(mg/L), ppt(g/L), AutoRanging
To select the suitable unit:
• Use or to highlight the
Conductivity Setup
Measure
mode and then to
section.
Resistivity Setup
Conductivity Setup
to access TDS Setup menu.
section.
section.
TDS Units
.
36
• Press to confirm and then use or
to highlight the convenient measuring range option.
• Press
cancel operation.
to confirm your selection or press to
81
Temperature - see
Units
The user can choose between Ohm, KOhm, MOhm, AutoRanging
measuring modes.
To select the units:
• Use
• Press to confirm and then use or
to highlight the desired unit.
• Press
Conductivity Setup
or to highlight the
to confirm or press to cancel operation.
section.
Units
option.
ISEISE
SETUP (HI 4522 SETUP (HI 4522
ISE
SETUP (HI 4522
ISEISE
SETUP (HI 4522 SETUP (HI 4522
The ISE Setup menu allows the user to set the parameters regarding ISE measurement and calibration.
These parameters can be set specifically for each channel. The settings will be applied only to the focused
channel.
Accessing ISE Setup
• Press
select ISE range for the desired channel.
• Press and then
To access an
• Use
while in
ISE Setup
or to highlight the desired option.
option:
Measure
mode and then to
to access
ISE Setup
menu.
onlyonly
only
onlyonly
))
)
))
Sample ID - see
Log - see
Alarm - see
Conductivity Setup
Conductivity Setup
Conductivity Setup
section.
section.
section.
• Press
The following is a detailed description of the ISE Setup option
screens.
Reading Mode
This option allows the user to select the desired reading mode:
Subtraction, Analyte Addition
are incremental (see ISE Theory section for details).
to access the selected option.
and
Analyte Subtraction
Direct, Direct / AutoHold, Known Addition, Known
. Except Direct and Direct / AutoHold, all the other methods
Direct
If using Direct reading mode, the ion concentration can be read directly from the instrument. Make sure the
instrument was calibrated before taking measurements.
Direct / AutoHold
If using Direct/AutoHold reading mode, the ion concentration can be frozen on the LCD after starting the AutoHold
mode.
Known Addition
In the Known Addition method a standard with a known concentration of the measured ion is added to the sample. The
difference in mV potential is then used to calculate the concentration of the ion in the sample.
Known Subtraction
In the Known Subtraction method a standard with a known concentration is added to the sample, reacting with the
ion to be measured. The stoichiometric ratio between standard and sample must be known. The ion concentration
is then calculated using the difference in mV potential.
The method is especially usefull when there is no available ISE for the measured ion.
80
37
Analyte Addition
This method is similar to the Known Addition method, with the difference that an aliquot of sample is added to a
standard of known concentration. The sample and standard contain the same ion to be measured. The ion
concentration is then calculated using the difference in mV potential.
Analyte Subtraction
In the Analyte Subtraction method an aliquot of sample is added to a standard of known concentration, reacting
with the ion to be measured. The stoichiometric ratio between standard and sample must be known. The ion
concentration is then calculated using the difference in mV potential. The method is especially usefull when there
is no available ISE for the measured ion.
To set the Reading Mode:
• Use or to select the
• Press and use or to highlight the
desired option.
• Press
cancel operation.
to confirm your selection or press
Reading Mode
option.
to
RESISTIVITYRESISTIVITY
RESISTIVITY
RESISTIVITYRESISTIVITY
The Resistivity Setup menu allows the user to set the parameters related with the resistivity measurements.
Accessing Resistivity Setup
• Press while in
select resistivity range.
• Press and then to access
menu.
To access a
• Use or to select the desired option.
• Press to confirm your selection.
The following is a description of the Resistivity Setup option
screens.
Resistivity Setup
Measure
mode and then to
option:
SETUP SETUP
SETUP
SETUP SETUP
Resistivity Setup
Temperature – See
Note: The Temperature Compensation option is also included in the Temperature parameter.
pH Setup
section.
Temperature Compensation
This option allows the user to enable / disable the temperature compensation for the measurement of Ion
concentration.
To set the Temperature Compensation:
• Use or to select the
• Press and use or to highlight the
Temperature Compensation
• Press
the Temperature Compensation.
and use or to enable / disable
option.
Temperature
38
option.
Profile - see
Reading Mode
This option allows the user to select between
resistivity reading modes.
To set the
• Use or to select the
• Press and use or to highlight the
desired option.
• Press
cancel operation.
Conductivity Setup
Reading Mode:
to confirm your selection or press to
section.
Direct
and
Reading Mode
Direct/AutoHold
option.
79
Alarm
This option allows the user to define two alarm limits.
Alarm State
The following options are available:
Disabled
– the alarm will be disabled.
Inside Limits
value is inside the preset limits.
Outside Limits
value is outside the preset limits.
To set the alarm state:
• Use or to highlight the
• Press and use or to highlight the desired option.
• Press
– the alarm will notify the user when the measured
– the alarm will notify the user when the measured
Alarm State
to confirm your selection or press
option.
to cancel operation.
Alarm Limits
This option allows the user to set the alarm limits for the measured
value.
Note: The alarm high value cannot be lower than the alarm low
value.
To set the alarm limits:
• Hightlight the
Alarm Limits
option and then press
.
• Press
Note: If an ISE calibration was performed and the Temperature Compensation option is changed, a warning
message appears on the LCD informing the user to perform a new calibration or to set the previous option
in order to perform accurate measurements.
Calibration
This option allows the user to set all the data regarding the Ion calibration process.
to confirm your selection or press
to cancel operation.
Standard Entry Type
Two entry modes for the standard solutions used for calibration are available:
Manual Selection– when the calibration is started, the desired standard solution can be manually selected from
a standard solutions list (0.1 ppm, 1 ppm, 10 ppm, 100 ppm and 1000 ppm).
Custom Standard– when the calibration is started, each standard solution can be manually set (in ppm units)
in a displayed pop-up menu.
To set the Standard Entry Type:
• Use
• Press and use or to highlight the
Standard Entry Type
• Press and use or to highlight the
desired option.
• Press to confirm your selection or press
cancel operation.
Calibration Reminder
Set Reminder Period
Clear Calibration
or to select the Calibration option.
option.
– See
pH Setup
section.
– See
– See
pH Setup
pH Setup
section.
section.
to
• Use or to select the low /high alarm limit and
then press (Edit) .
• Use
alarm value.
• Press
or to increase / decrease the selected
to return to the Alarm options.
78
Electrode Type
This option allows the user to select the desired Ion Selective Electrode used for measurements from a list:
Ammonia, Bromide, Cadmium, Calcium, Carbon Dioxide, Chloride, Chlorine, Cupric, Cyanide, Fluoride, Iodide,
Lead, Nitrate, Potassium, Silver, Sodium, Sulfate, Sulfide and five custom ISE. For the standard ISE it is possible
to view the Ion constants (Name, Molar Weight and Electric Charge/Slope), while for the custom ISE all these
constants can be manually set.
39
To set the Electrode Type:
• Use or to select the Electrode Type option.
• Press and use or to select the
desired standard ISE or a custom one from the list.
For standard ISE:
• Press to visualize the Ion constants and then press
at any time to exit Ion Constants view mode.
• Press to confirm your selection and return to
options.
Setup
For custom ISE:
• Press to edit the Ion constants for the selected
ISE
Sampling Period
This option allows the user to select the desired sampling period for
automatic logging.
To set the sampling period:
• Use or to highlight the
• Press and use or to select the
desired option from
• Press to confirm your selection or press to
cancel operation.
1, 2, 5, 10, 30
Sampling Period
seconds.
option.
custom ISE. Use
desired Ion constant and press to enter Ion
constant edit mode or to cancel operation.
• For the Ion name the Text Editor menu will be displayed
on the LCD. Enter the desired information by accepting
the highlighted character which is added in the text bar,
using . The and keys help
the user to select the desired character. It is also possible
to delete the last character by positioning the cursor on
the Backspace character and pressing. Press
to return to the Ion Constants menu. If the
Saving Confirmation is enabled, press to accept
the modified option, to escape without saving
or to return to the editing mode. Otherwise, the
modified option is saved automatically.
• To set the appropriate Ion molar weight (in g/mol units) use
or and then press to save
the current value or press
• To select the appropriate Electric Charge/Slope use
or and then press . If the
Ion electric charge is None, its slope can be manually set
or to select the
to cancel operation.
New Lot
Accessing this option, the new manually logged readings will be put in a new log lot.
To generate a new lot:
• Use or to highlight the
• Press to generate a new manual lot. A pop-up menu will be displayed to ask for confirmation.
• Press to confirm or press to escape without saving and return to the Log options.
Note: If
New Lot
option is accessed and the
informing the user that a new lot can be created only if the Logging Type is set on
New Lot
Logging Type
option.
is
Automatic
, a warning message appears on the LCD
Manual
.
40
77
Log
This option allows the user to edit the settings related to the logging feature, as following:
Logging Type
Three logging types are available:
Automatic logging
time intervals (see
Manual logging
AutoHold logging
auto-hold event occured.
To set the sample logging type:
• Use
• Press and use or to choose from
Automatic, Manual
• Press to confirm your selection or press to
cancel operation.
- the readings are logged automatically at constant
Sampling Period
( log on demand)- the readings are logged each time
is pressed.
- the readings are logged automatically at each
or to highlight the
or
Auto Hold
Automatic, Manual
option).
Logging Type
.
and
AutoHold.
option.
Logging Data Configuration
This option allows the user to select the parameters that accompany
a logged value:
,
Operator ID, Company Name, Aditional Info 1
ID
To set the logging data configuration:
• Use or to highlight the
Configuration
• Press and then use or to enable
the parameter by selecting
by pressing
on the LCD, in which the slope value can be set using
or . Press to save the
modified value or press
Constants menu.
Note: If an ISE calibration was performed and a different Ion
Selective Electrode is selected (standard or custom), a
warning message appears on the LCD informing the user to
perform a new calibration or to select the previous ISE in
order to perform accurate measurements.
Concentration Unit
Accessing this option, the user can select the desired concentration unit for the measured ion or chemical compound.
The available concentration units are: ppt (g/L), ppm (mg/L), ppb (µg/L), mg/mL, M (mol/L), mmol/L and %w/v.
To set the Concentration Unit:
• Use
option.
• Press and use or to highlight the
desired option.
• Press to confirm your selection or press
cancel operation.
Sample ID – See
Stability Criteria – See
Log – See
Note: The Logging Data Configuration option includes also the Ion Constants parameter. If you want it to appear
in the log reports, it must be enabled.
Alarm – See
Note: The Alarm Limits (Low and High) are set in the selected concentration unit of the measured ion or chemical
compound and are displayed in a scientific format (value and value exponent).
or to select the Concentration Unit
pH Setup
pH Setup
section.
pH Setup
. A pop-up menu will be displayed
to return to the Ion
section.
pH Setup
section.
section.
to
76
41
Isopotential Point
This option allows the user to edit the isopotential point of the electrode used for ion measurements. The ion
selective electrodes have different isopotential points. If temperature compensation is desired for the ion
measurements, the isopotential point value is absolutely necessary.
The isopotential point is always edited in ppm (mg/L) units.
To set the Isopotential Point:
• Use or to select the Isopotential Point option.
• Press and use or to increase /
decrease the isopotential point value.
• Press key to save the modified value.
Sample ID
This option allows the user to give to the measured samples an identification number/name. Two
options are available:
ID Increment
and
Edit Sample ID
.
ID Increment
– the sample ID will be edited alphanumerically by the user.
None
Automatic
every new log lot notification.
To select the ID increment mode:
• Use
– the sample ID will be automatically incremented at
or to highlight the
ID Increment
option.
Sample ID
• Press
Confirmation is enabled, press
option,
the editing mode. Otherwise, the modified option is saved automatically.
Note: If an ISE calibration was performed and the Isopotential Point option is changed, a warning message
appears on the LCD informing the user to perform a new calibration or to set the previous option in order
to perform accurate measurements.
ISE Significant Digits
Accessing this option, the number of ISE significant digits can be set, with one (x), two (xx) or three (xxx)
significant digits.
To set the ISE Significant Digits:
• Use or to select the
option.
• Press and use or to highlight the
desired option.
• Press
cancel operation.
to return to the Ion Setup options. If the Saving
to accept the modified
to escape without saving or to return to
ISE Significant Digits
to confirm your selection or press to
• Press
the desired option.
• Press
cancel operation.
and then use or to highlight
to confirm your selection or press
to
Edit Sample ID
This option allows the user to edit the sample ID (numeric - autoincrement mode, alphanumeric - user editable).
To edit the Sample ID:
• Use or to select the Sample ID option.
• Press and use or to highlight the
• Edit numerically / alphanumerically the sample ID.
• Press to save the current sample ID or press
Edit Sample ID
to cancel operation.
option and then press .
42
75
Clear Calibration
Accessing this option, the existent conductivity calibration can be cleared. If the calibration is cleared, another
calibration has to be performed.
To clear calibration:
• Use or to highlight the
• Press to clear calibration. A pop-up menu will be displayed asking for confirmation.
• Press to confirm or press to escape without saving and return to the Calibration options.
Probe Type
This option allows the user to obtain some information about the connected conductivity probe: name, default cell
constant, range and rings number.
Units
The user can select the desired measurement unit. The available options are: µS, mS or
• Use or to highlight the
• Press to confirm your selection.
• Use or to select
• Press to confirm your selection or press to
cancel operation.
µS, mS
Clear Calibration
Units
option.
or
AutoRanging.
option.
AutoRanging
.
pH CALIBRATIONpH CALIBRATION
pH CALIBRATION
pH CALIBRATIONpH CALIBRATION
Calibrate the instrument often, especially if high accuracy is required.
The instrument should be recalibrated:
• Whenever the pH electrode is replaced.
• At least once a week.
• After testing aggressive chemicals.
• When “No pH Calibration” or “pH Calibration Expired” message appears on the LCD, in the Reminder messages area.
PREPARATION
Pour small quantities of the buffer solutions into clean beakers. If possible, use plastic beakers to minimize
any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two beakers for each buffer solution. One
for rinsing the electrode and one for calibration.
If you are measuring in the acidic range, use pH 7.01 or 6.86 as first buffer and pH 4.01/3.00 or 1.68 as
second buffer. If you are measuring in the alkaline range, use pH 7.01 or 6.86 as first buffer and pH 10.01/
9.18 or 12.45 as second buffer.
For extended range measurements (acidic and alkaline), perform a five-point calibration by selecting five of
the available buffers.
CALIBRATION PROCEDURE
Calibration has a choice of 8 memorized buffers: pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 and
up to 5 custom buffers.
For accurate measurements it is recommended to perform a five-point calibration. However, at least a two-point
calibration is suggested.
pH Calibration screen description
74
43
Three buffer entry types are available: Automatic, Semiautomatic and Manual Selection. The default option is
Manual Selection.
To calibrate the instrument using Manual Selection buffer entry type:
• Press . If the instrument was calibrated before and calibration was not cleared, the old
calibration can be cleared by pressing . After 10 seconds, will no longer be available
from calibration screen.
Note: It is very important to clear calibration history when a new electrode is used
because most errors and warning messages that appear during calibration
depend on calibration history.
• Immerse the pH electrode and the temperature probe approximately 4 cm
(1½”) into a buffer solution of your choice (pH 1.68, 3.00, 4.01, 6.86, 7.01,
9.18, 10.01, 12.45 or a custom buffer) and stir gently. The temperature probe
should be close to the pH electrode.
• Select the used pH calibration buffer using
wait...” message will appear on the LCD until the reading is stable or the buffer is validated.
• If the pH buffer is validated, will appear on the LCD. Press to update calibration. The
calibration buffer will be added to the Calibrated Buffers list.
• Immerse the pH electrode and the temperature probe into the next buffer solution and follow the above
procedure or press to exit calibration.
Notes: • The new added calibration point will replace an old one
if the difference between them is ±0.2 pH.
• If the existing stored calibration is full (five calibration
points), a pop-up menu will be displayed on the LCD in
which you can select with or keys
the buffer you want to replace with the current buffer.
Press to delete the selected buffer and then
press to update calibration with the new
buffer.
• When in MTC mode, if pressing after entering
pH calibration and while a HANNA pH calibration buffer
with a x.xxx resolution is selected, a pop-up menu will
be displayed on the LCD in which the temperature value can be adjusted using
or . The “Please
or
Calibration Reminder
This option allows the user to set the calibration reminder as
Periodic
or
Disabled
.
To set the calibration reminder:
• Use or to highlight the
option.
minder
• Press to confirm your selection and then use
or to choose the desired option
• Press to confirm your selection or press to
cancel operation.
.
Daily
Calibration re-
,
Set Reminder Period
reminder - the user can set the time of day when the reminder is to appear.
Daily
Periodic
reminder - the user can set the time from the last calibration (days, hours and minutes) after which the
reminder appears.
To set the reminder period:
• Use or to highlight the
option.
• Press and use / to select next /
previous entry to be edited.
• Press and use or to set the desired
value, then press to save the modified value.
• Press to return to the previous menu.
Set Reminder Period
. Press to save the new temperature value.
44
73
Calibration Points
The user can choose between
To set the calibration points:
• Use or to highlight the
option.
• Press to confirm your selection and then use
or to choose the desired option
• Press to confirm your selection or press to
cancel operation.
Single Point
and
Multi Points
Calibration Points
.
calibration.
Cell constant manual editing:
The conductivity probe can also be calibrated by entering the cell
constant value.
To edit the cell constant value:
• Use or to highlight the
Cell Constant
option.
• When in MTC mode, if pressing
calibration and while a custom pH calibration buffer is
selected, a pop-up menu will be displayed on the LCD in
which the custom buffer and the temperature value can
be ajusted by pressing and then or
keys. Press to save the modified
value and then / to select next /
previous value to be adjusted.
• When in ATC mode, if pressing after entering calibration and while a standard pH buffer (with a
x.xxx resolution) or a custom pH buffer is selected, a
pop-up menu will be displayed on the LCD in which
the buffer value can be adjusted using
buffer value.
• If the Automatic buffer entry type has been selected for the calibration procedure, the instrument
will automatically select the closest buffer to the measured pH value from the edit buffer group
(see
pH Setup
for details).
• If the Semiautomatic buffer entry type has been selected for the calibration procedure, the instrument
will automatically select the closest buffers to the measured pH value from all the available buffers
and the used buffer value can be selected with or .
after entering
or . Press to save the new
• Press and then use or to increase / decrease the value.
• Press to preset the cell constant value to the probe
auto-recognizable default cell constant.
• Press to confirm the new value or press to
exit without modifying.
72
CALIBRATION MESSAGES
• Wrong buffer. Please check the buffer: this message appears when the difference between the pH
reading and the value of the selected calibration buffer is significant. If this message is displayed, check
if you have selected the appropriate calibration buffer.
• Wrong buffer temperature: this message appears if the buffer temperature is out of the defined buffer
temperature range.
• Clean the electrode or check the buffer. Press <Accept> to update calibration: this message alerts the
user that some dirt or deposits could be on the electrode. Refer to the electrode Cleaning Procedure (see page 65).
• Slope too low. Please check the buffer / Slope too high. Please check the buffer: these messages appear
if the current slope is under 80% or over 110% of default slope. Recalibrate the instrument using fresh buffers.
• Slope too low. Press <Clear Cal> to clear old calibration / Slope too high. Press <Clear Cal>to clear old calibration: these messages appear as a result of an erroneous slope condition. Follow
displayed instructions.
45
• Unrecognized buffer. Please check the buffer or the buffer list (for Semiautomatic and Automatic buffer entry
type): this message appears if the current buffer value is not close to any of the buffers from the buffer list/group.
Check if the current buffer is present in the buffer list or the appropriate buffer group was selected.
• The current buffer was already calibrated. Press <Accept> to recalibrate in this point, or changethe buffer: this message appears when calibration is performed in one of the previous calibrated buffers.
Follow displayed instructions.
Compensation Coefficient
To set the compensation coefficient:
• Use or to highlight the
Coefficient
• Press
using or to increase/decrease the value.
option.
(only for
Linear
temperature compensation)
Compensation
and set the desired compensation coefficient
• Press to save the current value or press
cancel operation.
Calibration
Cell Constant
The conductivity probe can be calibrated using the conductivity standards or by entering the cell constant of the
probe by the user.
Using standard solutions:Using standard solutions:
Using standard solutions:
Using standard solutions:Using standard solutions:
The meter can be calibrated in a single or multi-points (up to four points), using 6 Hanna standards (84 µS, 1413
to
µS, 5.0 mS,12.88 mS, 80.0 mS, 111.8 mS) or using the custom standards.
The following options are available for calibration:
Standard Recognition
The user can choose between
standards available) or
used for calibration).
• Use or to highlight the
option.
tion
• Use or to choose from
option
Standard
• Press to confirm your selection or press to
cancel operation.
.
Automatic
User Standard
recognition (from 6 Hanna
(when custom standards are
Standard Recogni-
Automatic
or
User
46
71
Temperature Unit
The user can choose from the
units.
To set the temperature unit:
• Use or to highlight the
option.
• Press and then use or to select
Celsius, Fahrenheit
• Press to confirm your selection or press to
cancel operation.
Celsius, Fahrenheit
or
Kelvin
degrees unit
or
Kelvin
temperature
Temperature Unit
.
pH MEASUREMENTpH MEASUREMENT
pH MEASUREMENT
pH MEASUREMENTpH MEASUREMENT
Make sure the instrument has been calibrated before taking pH measurements.
DIRECT MEASUREMENT
To measure the pH of a sample using the Direct reading mode:
• Press
• Select the Direct reading mode (see pH Setup for details).
• Submerge the electrode tip and the temperature probe
approximately 4 cm (1½”) into the sample to be tested.
Allow time for the electrode to stabilize.
• The measured pH value will be displayed on the LCD, together
with a short GLP information or “Default Calibration” message
if no pH calibration was performed.
Note: If the reading is out of range, “-----” will be displayed on
the LCD.
and then to select
pH Measure
mode.
Reference Temperature
To set the reference temperature:
• Use or to highlight the
option.
ture
• Press
/ decrease the value.
• Press to save or press
(only for
and then use
Linear
or
or to increase
Non-Linear
temperature compensation)
Reference Tempera-
to cancel operation.
70
DIRECT / AUTOHOLD MEASUREMENT
To measure pH of a sample using the Direct/AutoHold reading mode:
• Press
• Select the Direct/AutoHold reading mode (see
details).
• Submerge the electrode tip and the temperature probe
approximately 4 cm (1½”) into the sample to be tested.
Allow time for the electrode to stabilize.
• The measured pH value will be displayed on the LCD. If pressing
LCD until the stability criterion is reached. The pH value will be
frozen on the LCD, along with “AutoHold” indicator.
• To return to normal
Note: If the reading is out of range, “-----” will be displayed on
the LCD.
and then to select
, the “AutoHold” indicator will start blinking on the
Measure
mode press.
pH Measure
pH Setup
47
mode.
for
Outside Cal Range feature warns the user if the current reading is
out of the calibrated area. The calibrated area is that part of the pH
range in which the calibration point assures an accurate reading. If
the reading is taken out of the calibration area, the “Outside Cal
Range” message will start blinking on the LCD. The calibrated area
is calculated in according with the pH resolution used during the
reading. To avoid having this message, the calibration points have to
be well distributed in the desired measurement range.
If measurements are taken successively in different samples, it is
recommended to rinse the electrode thoroughly with deionized
water or tap water and then with some of the next sample to
prevent cross-contamination and to condition the electrode before
immersing it into the sample solution.
The pH reading is affected by temperature. In order to measure the pH accurately, the temperature effect must
be compensated. To use the Automatic Temperature Compensation (ATC) feature, connect and submerge the
temperature probe into the sample as close as possible to the electrode and wait for a few seconds.
If the temperature of the sample is known, Manual Temperature Compensation (MTC) can be performed by
disconnecting the temperature probe.
Notes: • When in MTC mode, the temperature can be modified by pressing
for mV/
• When in ATC mode for pH, respectively TEMP for mV/Rel mV, “-----” will be displayed on the LCD if
the measured temperature is under or over temperature range (–20.0 ºC to 120.0 ºC).
Rel mV Measure
from –20.0 ºC to 120.0 ºC. Press to save the new temperature value or press
to return to
Measure
mode. The temperature value can be ajusted with or
mode.
48
for
pH Measure
Table for non-linear temperature compensation:
mode and
69
Temperature Compensation
The user can choose from the following options:
Linear -
the meter will compensate automatically the conductivity using the following formula:
mV & Relative mV MEASUREmV & Relative mV MEASURE
mV & Relative mV MEASURE
mV & Relative mV MEASUREmV & Relative mV MEASURE
mV/ORPmV/ORP
mV/ORP
mV/ORPmV/ORP
Oxidation-reduction potential (ORP) measurements provide the quantification of the oxidizing or reducing power of the
tested sample.
To correctly perform a redox measurement, the surface of the ORP electrode must be clean and smooth.
MEASUREMENTS MEASUREMENTS
MEASUREMENTS
MEASUREMENTS MEASUREMENTS
MENTMENT
MENT
MENTMENT
SS
S
SS
where:
C
- conductivity at reference temperature
ref
- compensation coefficient
T - temperature
ref - reference temperature
Non-Linear -
Disabled -
To set the temperature compensation mode:
• Use or to highlight the
• Press and then use or to select
• Press to confirm your selection or press to cancel
operation.
Whatever form of compensation is used, the reading will not be as accurate as taking a reading of the sample’s
conductivity at the reference temperature.
recommended for measuring the conductivity of the natural water in accordance with the compensation
table on the next page.
the meter will display the conductivity with no temperature compensation.
Temperature Com-
pensation
Linear, Non-Linear
option.
or
Disabled
option.
DIRECT MEASUREMENT
To measure the mV of a sample using the Direct reading mode:
• Press and then to enter
• Select the Direct reading mode (see
• Submerge the tip of the ORP electrode (4 cm/1½”) into the
sample to be tested and allow a few seconds for the reading
to stabilize.
• The instrument will display the measured mV value on the LCD.
Note: If the reading is out of range, “-----” will be displayed
on the LCD.
DIRECT/AUTOHOLD MEASUREMENT
To measure mV of a sample using the Direct/AutoHold reading mode:
• Press
• Select the Direct/AutoHold reading mode (see
details).
• Submerge the tip of the ORP electrode approximately 4 cm
(1½”) into the sample to be tested and allow a few
seconds for the reading to stabilize.
• The measured mV value will be displayed on the LCD. If
pressing , the “AutoHold” indicator will start blinking
on the LCD until the stability criterion is reached. The mV value
will be frozen on the LCD, along with “AutoHold” indicator.
• To return to normal
Note: If the reading is out of range, “-----” will be displayed
on the LCD.
and then to select
Measure
mode press.
mV Setup
mV Measure
for details).
mV Measure
mV Setup
mode.
mode.
for
68
49
Relative mVRelative mV
Relative mV
Relative mVRelative mV
To measure the Relative mV of a sample:
• Press
Measure
• The instrument will display the measured Relative mV value
on the LCD, together with a short GLP information about the
last calibration or “Not Calibrated” message if no Rel mV
calibration was performed (no Rel mV offset set).
Notes: • If the measured mV potential is out of range, “-----”
• If pressing while in
• Before performing a Rel mV calibration, make sure
MEASUREMENTS MEASUREMENTS
MEASUREMENTS
MEASUREMENTS MEASUREMENTS
and then to enter
mode.
will be displayed on the LCD.
Relative mV
Relative mV Measure
mode, the Relative mV offset screen will be displayed
on the LCD. Press
Relative mV or use or to adjust
the Relative mV value and then press
the current Relative mV. Press at any time
to return to
that the electrode is immersed into the sample to be
measured.
Relative mV Measure
to clear the existing
mode.
to save
Temperature
From the
source and units, as well as the temperature compensation mode,
reference temperature and compensation coefficient.
To access a
• Use or to highlight the
• Press
Temperature
Temperature
tion from the
menu the user can choose the temperature
option:
Temperature
Conductivity Setup
to access the
menu.
Temperature
option.
op-
Temperature Source
To set the temperature source:
• Use or to highlight the
option.
• Press
Automatic
• Press
cancel operation.
and then use or to select
or
Manual
temperature source.
to confirm your selection or press to
Temperature Source
50
67
Delete Profile
To delete one of the existing profiles:
• Use or to select the
• Press
Delete Profile
• Press
the screen.
• Use or to select the desired profile and press
and use or to highlight the
option.
.
A list with all customised profiles will appear on
Profile
option.
.
• Press
Reading Mode
This option allows the user to select between
Direct/USP
To set the reading mode:
• Use or to select the
to return to the previous menu.
conductivity reading modes.
Direct, Direct / AutoHold,
Reading Mode
option.
ISE CALIBRATIONISE CALIBRATION
ISE CALIBRATION
ISE CALIBRATIONISE CALIBRATION
(HI 4522 only)(HI 4522 only)
(HI 4522 only)
(HI 4522 only)(HI 4522 only)
For greatest accuracy, it is recommended to calibrate the instrument frequently. The instrument should also be recalibrated
whenever “No ISE Calibration” or “ISE Calibration Expired” message appears on the LCD, in the Reminder messages area.
Due to electrode conditioning time, the electrode must be kept immersed a few seconds to stabilize. The user
will be guided step by step during calibration with easy to follow messages on the display. This will make
the calibration a simple and error-free procedure.
PREPARATION
Pour small quantities of the standard solutions into clean beakers. If possible, use plastic beakers to minimize
any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two beakers for each standard solution. One for
rinsing the electrode and one for calibration.
Note: For accurate measurements, add the appropriate ISAB (Ionic Strength Adjustment Buffer) to the
calibration standards.
CALIBRATION PROCEDURE
Up to five-point calibration is possible, using five memorized standard solutions: 0.1, 1, 10, 100, 1000 ppm
and 5 custom solutions.
The ISE calibration and measurement can be performed with or without temperature compensation. If the
temperature compensation option is enabled, the isopotential point of the electrode must be set in ISE Setup in
order to perform correct concentration measurements.
Before calibrating, make sure that the appropriate Electrode Type has been selected in ISE Setup according to the
measured Ion/compound.
ISE Calibration screen description
• Press and use or to highlight the
desired option.
• Press to confirm your selection or press
cancel operation.
66
to
51
Two standard entry types are available: ManualSelection and Custom Standard. The default option is Manual Selection.
A. To calibrate the instrument using
• Press. If the instrument was calibrated before and calibration was not cleared, the old calibration can be
cleared by pressing . After 10 seconds, will no longer be available.
Note: It is very important to clear calibration history when a new electrode is used
because most errors and warning messages that appear during calibration
depend on calibration history.
• Add ISAB to all standard solutions.
• Immerse the Ion Selective Electrode and the temperature probe approximately 4 cm
(1½”) into the less concentrated standard solution and stir gently.
• Select the used standard solution concentration with or. The “Please wait...” message
will appear on the LCD until the reading is stable or the concentration is validated.
Note: When in ATC mode, if pressing
concentration value can be adjusted using or . Press to save the new
concentration value.
• If the standard solution concentration is validated, will appear on the LCD. Press to update
calibration. The calibration point value will be added to the Calibrated Standards list.
• The “Please wait...” message will appear on the LCD for 10 seconds and until the reading is stable, time in
which immerse the ion selective electrode and the temperature probe into the next standard solution and
follow the above procedure or press to exit calibration.
B. To calibrate the instrument using
• Press. If the instrument was calibrated before and calibration was not cleared, the old
calibration can be cleared by pressing . After 10 seconds, will no longer be available.
Note: It is very important to clear calibration history when a new electrode is
used because most errors and warning messages that appear during
calibration depend on calibration history.
• Add ISAB to all standard solutions.
• Immerse the Ion Selective Electrode and the temperature probe approximately 4
cm (1½”) into the less concentrated standard solution and stir gently.
• When the stability condition is reached and 10 seconds elapsed, a pop-up menu will appear on the LCD
in which the standard concentration value can be set.
• Press
/ to select next/previous entry to be edit. Press and use or
Manual Selection
a pop-up menu will be displayed on the LCD in which the
Custom Standard
standard entry type:
entry type:
Save Current Profile
To save the current profile:
• Use or to select the
• Press and then select
Text Editor
box will be displayed on the LCD.
• Enter the desired profile name by using and
to highlight the desired character and then press to add it to
the text bar. It is also possible to delete the last character by positioning
the cursor on the Backspace character ( ) and pressing .
• Press to return to the Profile options.
Note: The saved profile will automatically become the current profile.
Profile
option.
Save Current Profile
option. The
Load Profile
To load one profile:
• Use or to select the
• Press
Load Profile
• Press
on the screen.
• Use or to select the desired profile and press
and use or to highlight the
option.
.
A list with all customised profiles will be displayed
to
confirm or
Profile
option.
to exit without selecting.
to set the desired value. Press to save the modified value and then press
to store the set concentration value.
52
65
ConductivityConductivity
Conductivity
ConductivityConductivity
The Conductivity Setup menu allows the user to set the parameters related to the conductivity measurement.
Accessing Conductivity Setup
• Press while in
select the
• Press and then to access
Setup
To access a conductivity setup options:
• Use or to highlight the desired option.
• Press
exit setup.
The following is a detailed description of the
option screens.
Profile
Choosing this option the measuring and the calibration mode can be customized. Up to 10 profiles can be
defined by the user.
The available options are:
Conductivity
menu.
Measure
mode and then to
measure mode.
to access the selected option or to
Conductivity Setup
SETUP SETUP
SETUP
SETUP SETUP
Conductivity
• If the standard solution concentration is validated,
calibration. The calibration point value will be added to the Calibrated Standards section.
• The “Please wait...” message will appear on the LCD until the reading is stable and 10 seconds elapsed,
time in which immerse the Ion Selective Electrode and the temperature probe into the next standard solution
and follow the above procedure or press to exit calibration.
Notes: • The new added calibration point will replace an old
one if the difference between them is less than 20%
of the standard solution.
• If the existing stored calibration is full (five calibration points), a pop-up menu will be displayed on the
LCD in which you can select with
the standard solution you want to replace with the
current one. Press
calibrated point and then press
calibration with the new standard solution.
• If the isopotential point of the electrode is unknown, the
ion calibration and measurements can be performed only
without temperature compensation (see ISE Setup, Temperature option for details, page 33). When the temperature
compensation option is disabled, at least a two-point
calibration must be performed in order to perform ISE
measurements. Otherwise, “-----” will appear on the LCD.
to delete the selected
will appear on the LCD. Press to update
or
to update
Save Current Profile:
Load Profile:
Delete Profile:
load from available profiles.
delete a profile.
save the current profile.
64
• When in MTC mode, if pressing after entering ISE calibration and while a standard solution is
selected, a pop-up menu will be displayed on the LCD
in which the concentration and the temperature value
can be ajusted by pressing and then
or keys. Press to save the modified value and then / to select
next /previous value to be adjusted.
CALIBRATION MESSAGES
• Wrong standard solution. Please check the standard solution: this message appears when the
difference between the ppm reading and the value of the selected standard solution concentration (ppm)
53
is significant. If this message is displayed, check if you have selected the appropriate calibration
standard.to be tested. Allow time for the electrode to stabilize.
• Wrong new slope. Please check the standard solution: this message appears if the current slope
exceeds slope window (50% to 120% of default slope for the corresponding ion charge - see ISE
Theorysection for details). Recalibrate the instrument using fresh standards.
• Wrong old slope. Press <Clear Cal> to clear old calibration: this message appears as a result of
an erroneous slope condition. Press
to clear old calibration and restart calibration.
• Difference between standards temperature is too high. Press <Accept> to update the calibration orclear old calibration: this message appears when the difference between two standards temperature is
greater than 5 ºC. Recalibrate the instrument while paying attention to the standards temperature.
where: C
- the sample concentration;
SAMPLE
C
- the standard concentration;
STD
V
V
and VT = V
- the sample volume;
SAMPLE
- the standard volume;
STD
SAMPLE
+ V
STD
∆E - the difference of potential from the electrode;
S - the electrode slope, determined in a
previous calibration;
f - the stoichiometric ratio between sample
and standard;
54
63
where: C
C
V
V
and VT = V
- the sample concentration;
SAMPLE
- the standard concentration;
STD
- the sample volume;
SAMPLE
- the standard volume;
STD
+ V
SAMPLE
STD
∆E - the difference of potential from the electrode;
S - the electrode slope, determined in a
previous calibration;
f - the stoichiometric ratio between sample
and standard;
Example 1
+
Ag
You have sulfide samples and you are adding
One mole sulfide sample reacts with 2 moles silver standard (
. The reaction is:
f
= ½).
Example 2
2+
Pb
You have sulfide samples and you are adding
One mole sulfide sample reacts with 1 mole lead standard (
. The reaction is:
f
= 1).
Analyte Addition and Analyte Subtraction
Analyte Addition and Subtraction are variations of the previous two methods.
ISE MEASUREMENTISE MEASUREMENT
ISE MEASUREMENT
ISE MEASUREMENTISE MEASUREMENT
(HI 4522 only)(HI 4522 only)
(HI 4522 only)
(HI 4522 only)(HI 4522 only)
Make sure the instrument has been calibrated before taking ISE measurements.
When using one of the incremental methods for measurement and no or only a one-point ISE calibration was
performed, a warning message will be displayed on the LCD, notifying the user that at least a two-point ISE
calibration must be performed.
For accurate measurements, add the appropriate ISAB (Ionic Strength Adjustment Buffer) to both samples and
standards, if necessary. If ISAB is added to the calibration standards, it is important to add ISAB also to the
samples to be measured.
DIRECT MEASUREMENT
To measure the concentration of a sample using the Direct
reading mode:
• Press
and then to select
• Select the Direct reading mode (see
ISE Setup
ISE Measure
for details).
mode.
• Add ISAB to the sample solution.
• Submerge the Ion Selective Electrode tip and the temperature probe approximately 4 cm (1½”) into the sample to
be tested. Allow time for the electrode to stabilize.
• The measured concentration value will be displayed on the LCD
in the selected units.
Note: If the reading is out of range, “-----” will be dis-
played on the LCD.
With Analyte Addition,
sample (analyte) is added
to an ion standard being measured. The standard and
sample contain the same ion. mV are taken before and after the sample addition. From the mV the analyte
concentration is determined.
With
Analyte Subtraction,
sample (analyte) is added to an ion standard
being measured. The analyte reacts
with the measured ion in a known manner thus removing measured ions from the solution. From the change
in mV the concentration of the analyte is determined.
62
DIRECT/AUTOHOLD MEASUREMENT
To measure the concentration of a sample using the Direct /
AutoHold reading mode:
• Press and then to select
ISE Measure
mode
for the selected channel.
• Select the Direct/AutoHold reading mode (see
ISE Setup
for
details).
• Add ISAB to the sample solution.
• Submerge the Ion Selective Electrode tip and the temperature probe approximately 4 cm (1½”) into the sample to
be tested. Allow time for the electrode to stabilize.
55
• The measured concentration value will be displayed on the LCD. If pressing
will start blinking on the LCD until the stability criterion is reached. The concentration value will be frozen on the
LCD, along with “AutoHold” indicator.
• To return to normal
Note: If the reading is out of range, “-----” will be displayed on the LCD.
Note: Press at any time to stop the measurement and return to
KNOWN ADDITION
To measure the concentration of a sample using the Known
Addition incremental method:
• Press and then to select
mode for the selected channel.
• Select the Known Addition method (see
• Press to start the measurement. The first step of the
method will be displayed on the LCD in which the user is notified
to add the appropriate ISA Buffer Volume (if different from 0 mL)
to the sample and to immerse the electrode into the Sample
Volume. The method parameters are also displayed on the LCD.
• Press to set the method parameters. Press
/ to select next/previous parameter to be edit, then
press and use or to set the
desired parameter value. Press to save the modified value
and then press to exit method parameters edit menu.
• Make sure that the set parameters values are the used ones and
submerge the Ion Selective Electrode tip approximately 4 cm
(1½”) into the sample to be tested.
• Press to take the first mV reading.
Measure
mode press.
ISE Setup
ISE Measure
for details).
, the “AutoHold” indicator
ISE Measure
mode.
Actual samples that are more concentrated have much smaller activity coefficients (
inert background salt to standards and samples stabilizes the activity coefficient so that concentration measurements
may be made directly. Hanna’s Ionic Strength Adjustment Buffer (ISAB) formulations also may optimize pH, and
complex interferences in addition to standardizing the ionic strength.
The Nernst equation can be rewritten:
ION SELECTIVE ANALYSIS METHODS
Direct AnalysisDirect Analysis
Direct Analysis
Direct AnalysisDirect Analysis
This method is a simple procedure for measuring multiple samples. It should only be used in the linear
working regions of the sensor. A direct reading instrument such as the HI 4522 determines concentration of
the unknown by a direct reading after calibrating the instrument with the standards. The instrument is
calibrated as described in “ISE CALIBRATION” section, with two or more freshly made standards that are in the
measurement range of the unknowns. Ionic strength adjustment is made to samples and standards. Unknowns
are measured directly by the instrument.
At lower concentrations, in non-linear regions of the electrode response, multiple calibration points will extend
measurements to a practical detection limit. Calibrations must be performed more frequently in these cases.
Incremental MethodsIncremental Methods
Incremental Methods
Incremental MethodsIncremental Methods
Incremental methods are useful for the measurement of samples whose constituents are variable or
concentrated. Incremental techniques can reduce errors from such variables as temperature, viscosity, or pH
extremes and will provide indirect analysis of ions for which there is no ISE sensor for a direct measurement.
There are four different incremental methods for sample measurement. They are Known Addition, Known
Subtraction, Analyte Addition and Analyte Subtraction. HI 4522 allows the analyst to use these techniques as
a simple routine procedure thus eliminating calculations or tables. The method once set up can be used for
repetitive measurements on multiple samples.
Known Addition and Known Subtraction
Known addition,
With
same ion. mV are taken before and after the standard addition. From the mV the sample concentration is
determined.
standard is added
to a sample being measured. The standard and sample contain the
<1). The addition of an
• When the reading is stable, press
mV reading. The second step of the method will be displayed
on the LCD in which the user is notified to add the Standard
Volume of set concentration to the sample. The method
parameters are also displayed on the LCD.
to store the first
56
With
Known subtraction,
with the measured ion in the sample in a known manner thus removing measured ions from the solution.
From the change in mV the concentration of the sample is determined.
a known standard is added to an ionic sample
61
being measured. The standard reacts
ISE THEORYISE THEORY
SEICEPS
EPOLS
)edaced/Vm(
noitactnelavonoM61.95+
noinatnelavonoM61.95–
noitactnelaviD85.92+
noinatnelaviD85.92–
ISE THEORY
ISE THEORYISE THEORY
An Ion Selective Electrode (ISE) is an electrochemical sensor that changes voltage with the activity or concentration
of ions in solutions. The change in voltage is a logarithmic relationship with concentration and is expressed by the
Nernst equation:
• Press
to change the desired method parameters as indicated in the first step of the method.
• Press to take the second mV reading.
• When the reading is stable, press to store the second mV reading. The ISE measurement
results will be displayed on the LCD.
• Press to log the current results into a ISE Method Report or press to return to
ISE Measure
mode.
where: E - the measured voltage;
o
- standard voltage and other standard system voltages;
E
a - the activity of the ion being measured;
S - the Nernst slope factor and is derived from thermodynamic principles:
R - the universal gas constant (8.314 J*K-1*mol-1);
T - the temperature in degrees Kelvin;
F - the Faraday’s constant (96,485 C*mol-1);
n - the ion charge.
The slope may be positive or negative depending upon the ion charge (n).
Activity and concentration are related by an “activity coefficient”, expressed as:
where: a - the activity of the ion being measured;
- the activity coefficient;
C - the concentration of the ion being measured.
In very dilute solutions
approaches 1 so activity and concentration are the same.
• If pressing , the method parameters can be changed and the sample concentration will be recalculated.
• Press to start another measurement.
Note: Press at any time to stop the measurement and return to
ISE Measure
mode.
KNOWN SUBTRACTION
To measure the concentration of a sample using the Known Subtraction incremental method:
• Press
and then to select
• Select the Known Subtraction method (see
ISE Measure
ISE Setup
mode for the selected channel.
for details).
• Press to start the measurement. The first step of the method will be displayed on the LCD in which the
user is notified to add the appropriate ISA Buffer Volume (if different from 0 mL) to the sample and to immerse
the electrode into the Sample Volume. The method parameters are also displayed on the LCD.
• Press to set the method parameters. Press / to select next / previous
parameter to be edit, then press and use or to set the desired parameter
value. Press to save the modified value and then press to exit method parameters edit
menu.
Note: For the Known Subtraction method, the stoichiometric ratio (factor) between standard and sample
must also be set in order to take accurate measurements.
• Make sure that the set parameters values are the used ones and submerge the Ion Selective Electrode tip
approximately 4 cm (1½”) into the sample to be tested.
• Press to take the first mV reading.
• When the reading is stable, press to store the first mV reading. The second step of the method
will be displayed on the LCD in which the user is notified to add the Standard Volume of set concentration
to the sample. The method parameters are also displayed on the LCD.
• Press
to change the desired method parameters as indicated in the first step of the method.
• Press
• When the reading is stable, press to store the second mV reading. The ISE measurement
60
to take the second mV reading.
57
results will be displayed on the LCD.
• Press
Report or press to return to
• If pressing, the method parameters can be changed
and the sample concentration will be recalculated.
• Press to start another measurement.
Note: Press at any time to stop the measurement and
return to
ANALYTE ADDITION
To measure the concentration of a sample using the Analyte
Addition incremental method:
• Press
• Select the Analyte Addition method (see
• Press to start the measurement. The first step of the method will be displayed on the LCD in
which the user is notified to add the appropriate ISA Buffer Volume (if different from 0 mL) to the standard
and to immerse the electrode into the Standard Volume of set concentration. The method parameters are
also displayed on the LCD.
• Press to set the method parameters. Press / to select next / previous parameter
to be edit, then press and use or to set the desired parameter value. Press
• Make sure that the set parameters values are the used ones and submerge the Ion Selective Electrode tip
approximately 4 cm (1½”) into the standard solution.
• Press to take the first mV reading.
• When the reading is stable, press to store the first mV reading. The second step of the method
will be displayed on the LCD in which the user is notified to add the Sample Volume to the standard
solution. The method parameters are also displayed on the LCD.
• Press to change the desired method parameters as indicated in the first step of the method.
• Press
• When the reading is stable, press
results will be displayed on the LCD.
• Press to log the current results into a ISE Method Report or press to return to
to log the current results into a ISE Method
ISE Measure
ISE Measure
and then to select
mode.
ISE Setup
to save the modified value and then press to exit method parameters edit menu.
to take the second mV reading.
mode.
ISE Measure
to store the second mV reading. The ISE measurement
mode.
for details).
ISE Measure
mode.
• If pressing
• Press to start another measurement.
Note: Press at any time to stop the measurement and return to
ANALYTE SUBTRACTION
To measure the concentration of a sample using the Analyte Subtraction incremental method:
• Press and then to select
• Select the Analyte Subtraction method (see
• Press to start the measurement. The first step of the method will be displayed on the LCD in
which the user is notified to add the appropriate ISA Buffer Volume (if different from 0 mL) to the standard
and to immerse the electrode into the Standard Volume of set concentration. The method parameters are
also displayed on the LCD.
• Press
to be edit, then press and use or to set the desired parameter value. Press
to save the modified value and then press to exit method parameters edit menu.
Note: For the Analyte Subtraction method, the stoichiometric ratio (factor) between standard and sample
must also be set in order to take accurate measurements.
• Make sure that the set parameters values are the used ones and submerge the Ion Selective Electrode tip
approximately 4 cm (1½”) into the standard solution.
• Press to take the first mV reading.
• When the reading is stable, press to store the first mV reading. The second step of the method
will be displayed on the LCD in which the user is notified to add the Sample Volume to the standard
solution. The method parameters are also displayed on the LCD.
• Press to change the desired method parameters as indicated in the first step of the method.
• Press to take the second mV reading.
• When the reading is stable, press to store the second mV reading. The ISE measurement
results will be displayed on the LCD.
• Press
• If pressing, the method parameters can be changed and the sample concentration will be recalculated.
• Press
Note: Press at any time to stop the measurement and return to
, the method parameters can be changed and the sample concentration will be recalculated.
ISE Measure
ISE Measure
ISE Setup
to set the method parameters. Press / to select next/previous parameter
to log the current results into a ISE Method Report or press to return to
to start another measurement.
mode.
for details).
ISE Measure
mode.
mode.
ISE Measure
mode.
58
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
59
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.