Hanna Instruments HI 4115 User Manual

HI 4115
SILVER/SULFIDE
COMBINATION
HI 4015
Instruction Manual
HI 4015HI 4015
HI 4015
HI 4015HI 4015 HI 4115HI 4115
HI 4115
HI 4115HI 4115
Silver/Sulfide Ion
Selective Electrode
Half-cell Combination
1
HI 4015 Silver/ Sulfide Half-cellHI 4015 Silver/ Sulfide Half-cell
HI 4015 Silver/ Sulfide Half-cell
HI 4015 Silver/ Sulfide Half-cellHI 4015 Silver/ Sulfide Half-cell HI 4115 Silver/ Sulfide CombinationHI 4115 Silver/ Sulfide Combination
HI 4115 Silver/ Sulfide Combination
HI 4115 Silver/ Sulfide CombinationHI 4115 Silver/ Sulfide Combination ElectrodeElectrode
Electrode
ElectrodeElectrode
I. I.
Introduction:Introduction:
I.
Introduction:
I. I.
Introduction:Introduction: The Hanna HI 4015 and HI 4115 are ion selective elec­trodes designed for the measurement of Silver/ Sulfide ions in aqueous solutions. The HI 4015 is a solid state half-cell sensor that requires a separate reference. The HI 4115 is a combination ion selective electrode.
IIII
SpecificationsSpecifications
II.
Specifications
IIII
SpecificationsSpecifications
Type: Solid State electrode with
a Silver/ Sulfide pellet.
Ion(s) measured: Silver (Ag+), Sulfide (S2-)
Measurement range: Silver 1.0 M to 1 X 10
-6
M
(107,900 to 0.11 ppm)
Sulfide 1.0 M to 1 X 10
-7
M
(32,100 to 0.003 ppm)
Theory of OperationTheory of Operation
III. III.
Theory of Operation
III.
Theory of OperationTheory of Operation
III. III.
::
:
:: The HI 4015 or HI 4115 Silver/ Sulfide electrodes are po­tentiometric devices used for the rapid determination of free silver or sulfide ions. They can be used as an indicator following the progress of a halide titration with silver ni­trate. Sulfide ion determinations are made in ground wa­ters, soils and paper liquors. The electrode functions as a sensor or ionic conductor. The HI 4015 requires a separate reference electrode to complete its electrolytic circuit. The HI 4115 has a double junction reference electrode incorpo­rated in its design. The silver/ silver- sulfide pellet is prac­tically insoluble in the test solutions being measured and produces a potential change due to changes in the sample’s ion activity. When the ionic strength of the sample is fixed by the addition of ISA, the voltage is proportional to the concentration of silver or sulfide ions in solution and the electrode follows the Nernst equation.
E= E
+ 2.3 RT/nF log A
a
ion
E= observed potential
Ea= Reference and fixed internal voltages
Interfering ions: Mercury must be absent.
Ions that complex with measured species will re­duce measured ion con­centrations.
Operating Temperature: 0-80°C
Operating pH: 2-8 pH (For Silver)
12-14 pH (For Sulfide)
Dimensions: 12 mm (OD) X 120 mm
nominal insertion (0.47” X 4.72”)
Connection: BNC
2
R= gas constant (8.314 J/K Mol)
n= Charge on ion (2- for sulfide), (1+ for silver)
A i=ion activity in sample
T= absolute temperature in K
F= Faraday constant (9.648 x 10
4
C/mol)
3
IV. IV.
Design elements of the HI 4015 and HI 4115Design elements of the HI 4015 and HI 4115
IV.
Design elements of the HI 4015 and HI 4115
IV. IV.
Design elements of the HI 4015 and HI 4115Design elements of the HI 4015 and HI 4115
electrodeselectrodes
electrodes
electrodeselectrodes
Cap
HI 4015
SILVER/SULFIDE
Sensor Handle
V. Equipment required:V. Equipment required:
V. Equipment required:
V. Equipment required:V. Equipment required:
Hanna HI 5315 Double Junction Reference Electrode
with HI 7072 Fill Solution for use with HI 4015.
Hanna HI 4222 pH/ISE/mV meter or other suitable
ion or pH/mV meter. (Note: log/linear graph paper is useful if an ISE (ion) meter is not available).
Hanna HI 180 Magnetic Stirrer or equivalent with
Teflon coated stirring bars (HI 731320). (Note: iso­late beakers from stirrer motor heat by placing insu­lating material such as foam or cork between them).
Hanna HI 76404 Electrode Holder or equivalent.
Plastic beakers (HI 740036P) or other suitable mea-
surement vessel.
Sensing Membrane
Upper Cap
Upper Threads
O-Ring
Ceramic Junction on Inner Stem
Liquid junction
Spring
Fill Hole
O-Ring and Plug
Outer Sleeve
Sensing Membrane
VI. VI.
Solutions Required for CalibrationSolutions Required for Calibration
VI.
Solutions Required for Calibration
VI. VI.
Solutions Required for CalibrationSolutions Required for Calibration Select appropriate Hanna Instruments standard and ISA from the list below:
0.1 M Silver Standard, 500 mL HI 4015-01 ISA for Silver measurement, 500 mL HI 4000-00 SAOB for Sulfide measurement, 500 mL HI 4015-00 Note: For Sulfide standard, see directions that follow.
Silver Standards For Molar solutions: Using volumetric pipettes and glassware make serial dilu­tions of HI 4015-01 to approximately bracket the concen­tration of the samples. Standards with concentrations
-3
< 10
M should be prepared daily. Store in dark plastic bottles. Two mL of Hanna ISA for Halide electrodes (HI 4000-00) should be added to 100 mL of sample or standard. For ppm solutions: Prepare 1078.7 ppm Silver standard by diluting HI 4015­01: Pipette 100 mL standard to a 1 liter volumetric flask. Add deionized water to volume. Using additional pipettes and glassware make serial dilutions of this 1078.7 ppm standard to bracket the sample concentra-
4
5
tions. Standards with concentrations < 100 ppm should be made daily.
Solutions for Sulfide Measurements SAOB, Hanna HI 4015-00 : Add content of SAOB reagent B bottle to the bottle with SAOB reagent A. Cap bottle and mix well until completely dissolved. Reagent that is brown has been oxidized and should be discarded.
Sulfide Standards: Sulfide standards are susceptible to air oxidation and therefore are not commercially available. Sulfide stock solution: Weigh approximately 100 gram of reagent grade Sodium Sulfide nonahydrate (Na2S·9H2O) into a 1 liter volumetric flask that contains a volume of distilled or deionized /deareated water. Swirl to dissolve, bring to volume, then transfer to a tightly sealed 1 L container with minimal head space and allow solution to equilibrate overnight.
133.5 ppm standard (approximate concentration): Transfer 5.00 mL stock solution and 250 mL Sulfide Antioxidant buffer SAOB (Hanna HI 4015-00) to a to 500 mL volumetric flask. Mix well then bring to volume with distilled or deionized /deareated water. Hanna HI 4015-00 SAOB, should be clear to pale straw in color after mixing. Note: higher concentrations of standard can also be prepared and titrated if samples are greater in concentration. The sensor is useable from 32,100 to
0.003 ppm. Determine exact concentration of the prepared standard by doing a titration with Lead Perchlorate standard (user prepared) on a weekly basis. The exact concentration of the standard is found by titrating 10 mL of the standard with 0.01 M Lead Perchlorate standard; (Pb(ClO4)2 ). Pb2++ S
2-
> PbS Use the electrode pair as an indicator for S2- to determine the endpoint. See Section XII.
Calculate as follows: Weekly standard in Molar concentration
VV
= Volume of the standard sulfide solution used
V
VV
s. s.
s.
s. s.
C (moles/liter)= (0.01) *
(10 mL) VV
V
= Volume of the Lead Perchlorate titrant used to reach
VV
t t
t
t t
V
/ V
t
s.
endpoint (mL)) Weekly standard in ppm C (ppm)= (0.01) *32.07
V
/ V
t
s.
Using volumetric pipettes and glassware make dilutions daily from the titrated standard whose value has been determined. Choose standards to bracket the concentra­tion of the samples. 50 mL of HI 4000-00 ISA should be added to each 100 mL flask of standard before being brought to volume.
VIIVII
General GuidelinesGeneral Guidelines
VII.
General Guidelines
VIIVII
General GuidelinesGeneral Guidelines
Calibration standards and sample solutions should
have the same ionic strength. ISA or SAOB should be added to both samples and standards in the same ratio.
Calibration standards and sample solutions should
be at same temperature.
Thermally insulate beaker with standard or sample
from magnetic stirrer.
Calibration standards and sample solutions should
be stirred at the same rate using identically sized TFE coated stir bars.
Rinse electrode pair with distilled or deionized water
between samples and gently dab off using soft disposable absorbent toweling. Do not rub sensor sur­face.
Presoaking silver sensor in a dilute standard
(<10-3M) will help to optimize response.
A scratched, pitted, or tarnished pellet surface can
cause drift, poor repeatability or a loss of low level response. Restore optimum response can be by removing the damaged surface with the microabrasive strip HI 4000-70.
6
7
Loading...
+ 7 hidden pages