Hanna Instruments HI 4113 Instruction Manual

HI 4113
NITRATE
COMBINATION
HI
4000-50
SENSOR HANDLE
Instruction Manual
HI 4013
HI 4113
Nitrate Ion
Selective Electrode
Half-cell
Combination
1
HI 4013 Nitrate Half-cellHI 4013 Nitrate Half-cell
HI 4013 Nitrate Half-cell
HI 4013 Nitrate Half-cellHI 4013 Nitrate Half-cell HI 4113 Nitrate Combination ElectrodeHI 4113 Nitrate Combination Electrode
HI 4113 Nitrate Combination Electrode
HI 4113 Nitrate Combination ElectrodeHI 4113 Nitrate Combination Electrode
I. I.
Introduction:Introduction:
I.
Introduction:
I. I.
Introduction:Introduction: The Hanna HI 4013 and HI 4113 are ion selective elec­trodes designed for the measurement of nitrate ions in aqueous solutions. They utilize a replaceable sensing mod­ule that contains an organic polymer membrane that is sensitive to nitrate ions. The HI 4013 is a half-cell electrode that requires a separate reference. The HI 4113 is a combi­nation ion selective electrode.
IIII
SpecificationsSpecifications
II.
Specifications
IIII
SpecificationsSpecifications
Type: PVC membrane with or-
ganic ion exchanger Ion measured: Nitrate (NO Measurement range: 0.1 M to 1X 10
-
)
3
-5
M
6200 to 0.62 ppm Interference: Organic solvents and cationic detergents must be absent. Ratio of interfering ion to NO indicated below:
300 for F
-
must be less than the ratio
3
-
fluoride
100 for Cl-chloride
-
2-
3
-
2
-
4
carbonate
nitrite
iodide
perchlorate
4 for CO 2 for NO
0.01 for I
0.0045 for ClO
Operating Temperature: 0-40°C Operating pH: 3 to 8 pH (see Section
XIII) Dimensions: 12 mm (OD) X 120 mm
nominal insertion
(0.47” X 4.72”) Connection: BNC
2
III. III.
Theory of OperationTheory of Operation
III.
Theory of Operation
III. III.
Theory of OperationTheory of Operation
::
:
:: The HI 4013 and HI 4113 nitrate electrodes are potentio­metric devices used for the rapid determination of free ni­trate ions in water, emulsified foods and plant samples. The electrode functions as a sensor or ionic conductor. The HI 4013 requires a separate reference electrode to complete its electrolytic circuit. The HI 4113 is a combination elec­trode with a Ag/AgCl reference electrode with gel stabilized Cl- electrolyte in its inner chamber. The external reference chamber is refillable. The PVC membrane used on the sensor is impregnated with the organic ion exchanger. This organic ion exchanger is considered a carrier ionophore in that it is capable of shielding and carrying the charged nitrate ion in its polar cage freely through the apolar regions of the membrane. A charge imbalance developes between the test solution and internal cell of the sensor. This voltage changes in response to the sample’s ion activ- ity. When the ionic strength of the sample is fixed, the voltage is proportional to the concentration of nitrate ions in solution. The sensor follows the Nernst Equation:
E = E
+ 2.3 RT/nF log A
a
ion
E = observed potential Ea = Reference and fixed internal voltages R = gas constant (8.314 J/K Mol) n = Charge on ion (-1) A
= ion activity in sample
ion
T = absolute temperature in K F = Faraday constant (9.648 x 104 C/equivalent)
3
IV. IV.
Design elements of the HI 4013 and HI 4113Design elements of the HI 4013 and HI 4113
IV.
Design elements of the HI 4013 and HI 4113
IV. IV.
Design elements of the HI 4013 and HI 4113Design elements of the HI 4013 and HI 4113
electrodeselectrodes
electrodes
electrodeselectrodes
Cap
HI
4000-50
HANDLE R
SENSO
Sensor Handle
O-Ring Sensing
module
Upper Cap
Upper Threads
O-Ring
Ceramic Junction on Inner Stem
Liquid junction
Spring
Fill Hole
O-Ring and Plug
Outer Sleeve
Sensing Module
4
V. V.
Equipment required:Equipment required:
V.
Equipment required:
V. V.
Equipment required:Equipment required:
The HI 4013 requires the Hanna HI 5315 Double
junction reference electrode with HI 7078 as external electrolyte.
Hanna HI 4222 pH/ISE/mV meter or other suitable
ion or pH/mV meter (Note: log/linear graph paper is useful if an ISE (ion) meter is not available).
Hanna HI 180 magnetic stirrer or equivalent with
magnetic stirring bars (Note: Isolate beakers from stir­rer motor heat by placing insulating material such as foam or cork between them).
Hanna HI 76404 electrode holder or equivalent.
Plastic beakers (HI 740036P) or other suitable mea-
surement vessel.
VI. VI.
Solutions RequiredSolutions Required
VI.
Solutions Required
VI. VI.
Solutions RequiredSolutions Required
Standards for Nitrate Measurements
0.1 M sodium nitrate standard, 500 mL HI 4013-01 100 ppm nitrate standard, 500 mL HI 4013-02 1000 ppm nitrate standard, 500 mL HI 4013-03 Ionic Strength Adjuster ISA, 500 mL HI 4013-00 Using volumetric pipettes and glassware make dilutions to bracket the concentration of the samples. Store samples in plastic bottles. Standards with concentrations < 10
-3
should be prepared daily. Two mL of Hanna ISA HI 4013-00 should be added to 100 mL sample or standard. ISISA is Interferent suppressent ISA. Dissolve 17.32g Al2(S04)3.H20, 3.43.g Ag2SO4, 1.28g H3BO4, 2.52g H2NSO3H in approximately 800 mL deionized water. Adjust pH to 3 with 0.1 N NaOH. Dilute to 1 liter. Store in a dark colored container. 50 mL of ISISA should be added to each 50 mL sample or standard.
M
VIIVII
General GuidelinesGeneral Guidelines
VII.
General Guidelines
VIIVII
General GuidelinesGeneral Guidelines
Ensure the o-ring is installed on modules before screw-
ing into the sensor handle or inner stem.
5
O-Ring
Due to shipping or storage the internal solution inside the PVC modules may have developed an air pocket near the membrane. Gently shaking the sensor down (like the old style mercury thermometer) will place the internal solution next to the membrane.
Presoaking the Nitrate sensor in a 10
-2
M standard without ISA for at least half-hour before calibration will help to optimize the sensor response.
Do not leave your sensors in standard or samples with ISA or ISISA for long periods of time.
Calibration standards and sample solutions should have the same ionic strength. ISA should be added to both samples and standards.
Calibration standards and sample solutions should be at same temperature.
Thermally insulate solution vessel from magnetic stir­rer.
Calibration standards and sample solutions should be stirred at the same rate using identical sized TFE coated stir bars.
Rinse electrodes with distilled or deionized water be­tween samples and gently dab dry with lab wipe or other soft disposable absorbent toweling. Do not rub the sensing surface.
Check for gas bubbles that may form near sensing surface (due to solution temperature changes). Tap off gently.
Avoid large changes in temperature (thermal shock) as it may damage the sensor.
6
Additional HI 4113 guidelines
Remove the protective plastic wrap that covers the ceramic junction before assembling sensor for the first time.
Add reference HI 7078 fill solution to bottom of fill hole or empty and refill fill solution daily before us­ing.
During measurement always operate electrode with the fill hole open.
During normal use, fill solution will slowly drain out of the tapered cone junction at the lower portion of the electrode. Excessive loss (>4 cm drop within 24 hours) is not normal. If this occurs verify cap is tightened and the interface between the internal cone and outer body is free of debris.
Add filling solution daily to maintain a good head pressure. For optimum response, this level should be maintained and not be allowed to drop more than 2
-3 cm (1-inch) below fill hole.
Do not use an electrode if crystallized salts are visible inside the electrode. Drain electrode, disassemble and rinse internal body with deionized water. Reassemble and refill with fresh fill solution.
If an erratic measurement occurs, check to see if for­eign matter is seen trapped near the internal cone. Drain by depressing the electrode cap then refill with fresh fill solution.
VIII.VIII.
Electrode PreparationElectrode Preparation
VIII.
Electrode Preparation
VIII.VIII.
Electrode PreparationElectrode Preparation HI 4013 The Hanna HI 4013 is a 2 piece design comprised of a sensor handle (HI 4000-50) and a sensing module (HI 4013-51). The sensor is shipped with two HI 4013-51 modules.
1. Remove sensing module from shipping vial. Do not
touch the sensing membrane with the “H” hole pat­tern on it.
2. Screw the module into the sensor handle finger tight.
Do not overtighten.
7
HI
4000-50
E L D N A H
R O S N E S
3. Holding the assembled electrode at the cable end, shake the sensor to ensure internal fill solution that may have separated during shipping is in contact with inner membrane surface.
4. Prepare HI 5315 reference electrode by filling electro­lyte reservoir with HI 7078 fill solution.
5. Place sensor and reference electrodes into electrode holder and connect cable connectors to meter.
6. Soak the Nitrate electrodes membrane in a Nitrate containing standard (0.001M) without ISA before calibration.
HI 4113 The Hanna HI 4113 is shipped disassembled. The sensor is shipped with two HI 4113-51 modules.
1. Unwrap Parafilm® seal found over ceramic junction on inner stem and discard. This is only used for ship­ping or long term storage.
2. Remove sensing cone from shipping vial. Do not touch the sensing membrane with the “H” hole pattern on it.
3. Screw the cone into the inner stem finger tight. Do not over tighten.
8
Remove
Water
Deionized
Parafilm
Install sensing module
4. Rinse inner stem with deionized water making certain to wet o-ring found on the inner stem.
5. Reassemble electrode by gently pushing the inner assembly into the outer body, sliding spring down cable, and screwing cap into place.
6. Remove fill hole cover and o-ring on fill hole spout.
Using the dropper pipette provided, add a few drops
HI 7078 fill solution to the electrode. Invert the elec­trode to wet the o-ring and rinse the fill solution cham­ber.
9
7. Holding the body of the electrode gently press upper cap with your thumb. This permits the fill solution to drain out of the body. Release cap and verify elec­trode returns to its original position. (You may need to gently assist for this to occur).
COMBINATION
NITRATE
HI 4113
8. Tighten the electrode cap onto the body and fill elec­trode body until fill solution volume is just below fill hole.
9. Position electrode in a Hanna HI 76404 electrode holder (or equivalent) and connect BNC connector to meter.
IX. IX.
Quick Check of Electrode SlopeQuick Check of Electrode Slope
IX.
Quick Check of Electrode Slope
IX. IX.
Quick Check of Electrode SlopeQuick Check of Electrode Slope
Connect electrode(s) to pH/mV/ISE meter
Place meter in mV mode.
Place 100 mL of deionized water into a beaker with
stir bar.
Place reference and measuring half-cell or combina­tion electrode into prepared sample.
Add 1 mL of a standard to beaker. Record the mV value when stable.
Add an additional 10 mL of standard to the
solution. Record the mV when reading has
10
stabilized. This value should be less than the previous value noted (more negative).
Determine the difference between the two mV values. An acceptable value for this slope is
56 ± 4 mV (20-25°C) X. X.
Corrective actionCorrective action
X.
Corrective action
X. X.
Corrective actionCorrective action
Verify module has been screwed into sensor handle or inner stem.
Verify Parafilm® seal has been removed from ce- ramic junction (HI 4113 or HI 5315 reference).
Verify fill solution has been added to reference cham­ber.
Verify electrodes are connected properly to meter and meter is powered.
Verify dilute standards are freshly made and stored. Remake solutions if appropriate. Store in plastic bottles.
If the reading is jumpy or unstable, shake sensor down (see section VII).
If the sensor slope just misses the suggested slope window, soaking the sensor in a standard solution without ISA may solve the problem.
If the membrane is damaged, the response becomes extremely sluggish, or the slope of the electrode has decreased significantly, and procedures above have not helped, the module should be replaced.
For HI 4013
1. Dry off module and sensor handle.
2. Unscrew sensing module and replace with a new one. (HI 4013-51).
3. Soak new module in nitrate solution to condition it before calibration.
For HI 4113
1. Drain the fill solution by depressing cap. Rinse elec­trode with distilled or deionized water. Drain.
11
2. Unscrew upper cap and slide down cable toward con­nector.
3. Move spring and outer body down cable also.
4. Dry off inner stem and module with a soft tissue.
5. Hold inner stem and unscrew module and replace with a new one. (HI 4113-51).
6. Reassemble electrode (see section VII), and refill with electrolyte. Soak new membrane in nitrate solution without ISA to condition before calibration.
XI. XI.
Direct Calibration and MeasurementDirect Calibration and Measurement
XI.
Direct Calibration and Measurement
XI. XI.
Direct Calibration and MeasurementDirect Calibration and Measurement
This method is a simple procedure for measuring many samples. A direct reading ISE meter (HI 4222 or equiva­lent) determines concentration of the unknown by a direct reading after calibrating the meter with the standards. Add HI 4013-00 to adjust ionic strength at a dose of 2 mL of per 100 mL sample or standard. ISISA* may also be used at a dose of 50 mL for 50 mL of sample or standard. The meter is calibrated using freshly made standards that are in the measurement range of the unknowns. Unknowns are read directly. In the region where the electrode calibration be­comes less linear, more calibration points are needed, and calibration will need to be repeated more frequently.
A pH/ mV meter in mV mode and semi log graph paper may also be used. Two or more freshly prepared standards that are in the measurement range of the unknowns are measured in mV mode on the meter. These values are plotted on the semi-log paper and the points are connected to form a straight-line curve. When samples are measured, their mV values are converted to concentration by following the mV to the concentration axis on the semi-log plot.
*Note: ISISA is the recommended ISA used for procedure 4500-NO3-D. published in Standard Methods for the Ex­amination of Water and Wastewater.
12
Procedure: Follow sections VIII and IX to prepare electrodes for mea­surement.
1) Follow section VI to prepare standards/ solution. Stan­dards should bracket and fall within the range of interest. Standards and solutions should be at the same temperature.
2 mL of HI 4013-00 is added to 100 mL of both samples and standards. OR
50 mL of ISISA are added to 50 mL both samples or standards. Add stir bar and mix before taking measurements.
2) Follow section VII; General Guidelines to optimize test set-up.
3) During calibration it is best to start with lower concen­tration samples first. Wait for a stable reading before reading/ recording values. Permit longer equilibra­tion times at these levels (3 or 4 minutes).
To prevent carry over and contamination of samples, rinse sensors with deionized water and remove moisture with absorbant tissue between samples.
Typical HI 4013 and HI 4113 Linearity
0
50
100
150
mV
200
250
300
0123 4567
-L og of Molar Concentration
13
XII. XII.
Other Measurement TOther Measurement T
XII.
Other Measurement T
XII. XII.
Other Measurement TOther Measurement T Known additionKnown addition
Known addition
Known additionKnown addition
echniquesechniques
echniques
echniquesechniques
An unknown concentration can be determined by adding a known volume and concentration of NO
-
standard to the
3
sample. mV values are noted before and after the addition of standard (E). An ideal sensor slope can be used in the equation but actual determined slopes at the temperature of measurement should be used if known (S). This method is preprogrammed in the Hanna HI 4222 pH/ISE/mV meter, which simplifies the method greatly. Example: Nitrate ion determination with known addition.
1. A 50 mL sample of unknown (V
) is placed in a
SAMPLE
clean plastic beaker with an electrode (s). Add 50 mL of ISISA* to sample. Mix. mV 1 is recorded.
2. 5 mL (V
STANDARD
) of 10
-1
M (C
) standard is added
STANDARD
to the beaker and the mV value decreases. (Note: for other concentration samples, add a known volume and concentration of standard to produce approxi­mately 30 mV change). The unknown Nitrate concentration in the original sample (C
) can then be determined by using the
SAMPLE
equation that follows.
3. The procedure can be repeated with second standard addition to verify slope and operation of the method.
C
C
=
sample
(V
T
(V
sample+Vstandard+VISA
(V
sample+VISA
standardVstandard
E/S
)10
- (VS)
)= V
)= V
S
V V
S
sample
T
*Note: ISISA is Interferent suppressent ISA see Section IV.
14
XIIIXIII
pH and Interferents pH and Interferents
XIII.
pH and Interferents
XIIIXIII
pH and Interferents pH and Interferents HI 4013/ HI4113 Nitrate electrodes can operate over a pH range of 3 to 8 but better results are found if pH is kept constant throughout calibration and test. Sulfuric acid or NaOH can be used for pH adjustment or use ISISA to surpress interferences and buffer pH. Limiting the length of time of exposure to samples containing interferences will prolong useful life of your electrode. If sensor has been exposed to ions above recommended levels, soaking in pure nitrate solutions without ISA will aid recovery of function.
XIV.XIV.
Storage and Care of the HI 4013 andStorage and Care of the HI 4013 and
XIV.
Storage and Care of the HI 4013 and
XIV.XIV.
Storage and Care of the HI 4013 andStorage and Care of the HI 4013 and
HI 4113 electrodesHI 4113 electrodes
HI 4113 electrodes
HI 4113 electrodesHI 4113 electrodes The HI 4013 sensor can be stored in standards that do not contain ISA for short time periods. For long term storage, unscrew sensing module from sensor handle and store dry in the shipping vial. The model HI 4113 combination electrode can be left in dilute standards that do not contain ISA for short time periods. If the electrode will be used frequently and needs to be ready for use, take measures to prevent evaporation of fill solution. Top off fill solution, replace o-ring, fill hole cover on the fill hole opening, and place sensor in dilute nitrate solution that does not contain ISA. Store electrode upright. Prior to use, drain electolyte chamber and refill with fresh HI 7078 fill solution. For long term storage, the electrode should be drained, disassembled and washed of salts with deionized water. Wrap the ceramic junction in Parafilm® or other sealing wrap. Unscrew the module and store dry in the shipping vial. Refrigeration of module will extend its life. Store dis­assembled electrode in storage box provided with electrode.
XV. XV.
Conversion tablesConversion tables
XV.
Conversion tables
XV. XV.
Conversion tablesConversion tables
--
-
--
For NOFor NO
For NO
For NOFor NO
33
3
33
Moles/L (M) NO
-
to ppm NO
3
Multiply by Multiply by
Multiply by
Multiply by Multiply by
-
(mg/L) 620 00
3
ppm (mg/L) to M (moles/L) 1.61 X 10
15
-5
MAN4113 07/06R1
WARRANTY WARRANTY
WARRANTY
WARRANTY WARRANTY Hanna Instruments Ion Selective Electrodes are warranted to be free of defects in material and workmanship for 6 months from date of purchase when used for their intended purpose and maintained according to instructions. If they fail to work when first used contact your dealer immediately. Damage due to accidents, misuse, misapplication, tampering or lack of prescribed maintenance is not covered.
Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice.
16
Loading...