Hanna Instruments HI 4105 User Manual

H
I 4105
CARBON
Instruction Manual
HI 4105HI 4105
HI 4105
HI 4105HI 4105
Carbon Dioxide
Ion
Selective Electrode
1
HI 4105 Carbon Dioxide ElectrodeHI 4105 Carbon Dioxide Electrode
HI 4105 Carbon Dioxide Electrode
HI 4105 Carbon Dioxide ElectrodeHI 4105 Carbon Dioxide Electrode
I. I.
Introduction:Introduction:
I.
Introduction:
I. I.
Introduction:Introduction: The Hanna HI 4105 Carbon Dioxide gas sensing electrode is a combination electrode designed for the measurement of Carbon Dioxide in aqueous solutions such as water, soft drinks or wine, samples. Carbonate and Bicarbonate ions are also measured by conversion to Carbon Dioxide gas upon ISA addition.
II.II.
SpecificationsSpecifications
II.
Specifications
II.II.
SpecificationsSpecifications
Type: CO2 gas sensing
electrode with glass pH internal, Ag/ AgCl reference and gas permeable membrane.
2-
Species detected: CO
-
,HCO
, CO
3
3
2
CO2 Measurement Range: 0.02 M to 4 x 10-4M
880 to 17.6 ppm
2-
Interfering ions: SO2/SO
3
, H2S/S
2-
Operating Temperature: 0 to 40°C
Operating pH: 4-5 pH
Dimensions: 12 mm (OD) X 120
mm (insertion)
0.47x 4.72
Wetted materials: Delrin®, body and cap
with PTFE membrane
Connection: BNC
2
3
III.III.
Theory of OperationTheory of Operation
III.
Theory of Operation
III.III.
Theory of OperationTheory of Operation
::
:
:: The Carbon Dioxide electrode is a complete potentiometric cell that contains both a silver/silver chloride (Ag/AgCl) reference and a pH measurement element. These elements are housed within a thermoplastic body in a chloride ion­containing electrolyte, and are isolated from the sample by a PTFE membrane. ISA addition changes the pH of the sample to approxi­mately 4.7 pH and Bicarbonate (HCO
2-
(CO
) ions in the sample are converted to carbon dioxide
3
-
) and carbonate
3
(CO2). The CO2 in the sample solution diffuses through the PTFE membrane where it dissolves into the thin film of fill solution found between the membrane and the internal pH membrane. Here it converts back into bicarbonate and hydrogen ions. The pH changes proportionally with the concentration of dissolved gas in the sample solution. Dif­fusion of CO2 continues until the partial pressures of the gas in the sample and thin film are equal. The Nernst expression for an Carbon Dioxide sensor is ex­pressed in the equation below. Note that the potential is a function of the Carbon Dioxide gas, which in turn is related to the hydrogen ion concentration. The glass internal, Ag/AgCl reference, equilibrium constant and Henrys law constant are rolled into the E and Eo terms. The Nernst equation for the sensor becomes the equation noted below:
E = E’+2.3RT/nF log [CO2]= E
E = observed potential
E = Reference and fixed internal voltages
o
+0.059 log [H
+
]
IV.IV.
Design ElementsDesign Elements
IV.
Design Elements
IV.IV.
Design ElementsDesign Elements The Hanna HI 4105 Carbon Dioxide gas sensor has 3 main parts. These are the membrane/membrane cap, outer probe body with antirotation key and the pH/reference assembly which includes the outer electrode cap, spring, inner cap and pH/reference electrode assembly.
pH/reference electrode assembly
cable
outer electrode cap
spring
antirotation key inner cap
Outer probe body
Reference electrode
R = gas constant (8.314 J/K Mol)
n= Charge on ion (equivalents/mol)
T = absolute temperature in K
4
F = Faraday constant (9.648 x 10
C/equivalent)
The mV should increase in a Nernstian manner as the carbon dioxide partial pressure increases in the sample.
4
O-ring
pH
sensitive
membrane
Membrane/membrane cap
5
V.V.
Equipment Required:Equipment Required:
V.
Equipment Required:
V.V.
Equipment Required:Equipment Required:
Hanna HI 4222 pH/ISE/mV meter or other suitable
ion or pH/mV meter. (Note: log/linear graph paper is useful if an ISE meter is not available).
Hanna HI 180 magnetic stirrer or equivalent with
stirring bars. (Note: Isolate beakers from stirrer motor heat by placing insulating material such as foam or cork between them).
Hanna HI 4000-71 gas sensor test vessel or
Hanna HI 76404 electrode holder or equivalent with Beakers or other suitable measurement vessel with plastic sealing film or wrap.
VI. VI.
Solutions Required for Calibration:Solutions Required for Calibration:
VI.
Solutions Required for Calibration:
VI. VI.
Solutions Required for Calibration:Solutions Required for Calibration: Ionic Strength Adjuster (ISA), 500 mL: HI 4005-00 Hanna 0.1 M standard, 500 mL: HI 4005-01 Hanna 1000 ppm CO2 standard, 500 mL: HI 4005-03* *Please Note: This calibration standards is ppm as CaCO
See Section XVII for additional solutions and accessories used for maintenance.
Using volumetric pipettes and glassware prepare serial dilutions of the standard. Select concentrations that will approximately bracket the concentration of the samples to be measured. Standards with concentrations less than 10-3M should be prepared fresh daily. Store solution in a tightly sealed bottle without ISA added. 10 mL of HI 4005­00 ISA should be added to each 100 mL sample of stan­dard and samples just prior to measurement. ISA adjusts the pH of the sample or standard to about pH 4.7 thus converting carbonate and bicarbonate ion to carbon diox­ide. It also provides samples and standards a constant ionic strength background that stabilizes the solutions ac­tivity coefficient and permits concentration to be measured directly.
VIIVII
General GuidelinesGeneral Guidelines
VII.
General Guidelines
VIIVII
General GuidelinesGeneral Guidelines
Calibration standards and sample solutions should
have the same ionic strength. ISA should be added to both samples and standards immediately before tak­ing measurements.
Calibration standards and sample solutions should be
stirred at the same rate using identical sized stir bars. Stir thoroughly and continuously.
Calibration standards and sample solutions should be
at the same temperature. Thermally insulate solution vessel from magnetic stirrer with cork or other insulat­ing medium.
Wait until sensor value has stabilized before taking
reading (at least 5 minutes when going from more dilute to more concentrated samples, longer when the order is reversed).
Surface coating on the PTFE membrane will effect the
3
response. Inspect sensor before using. Wash off with a jet of deionized water or mild detergent to remove film.
Replace PTFE membrane if damage is evident or a
droplet of internal electrolyte is seen.
Rinse electrode with distilled or deionized water be-
tween samples and dab dry with lab wipe or other soft disposable absorbent toweling.
Check calibration every 1-2 hours. Recalibrate if nec-
essary.
Position sensors at an angle of approximately 20° to 30° from verticle to reduce possibility of trapping gas
bubbles on on the membrane cap. Gas bubbles also form from solution out-gassing due to temperature change. Gently tap body of sensor to dislodge them.
Close container with plastic wrap or use HI 4000-71
gas sensor test vessel to prevent gas from leaving.
If electrode was left out in air for a prolonged period,
gently pulling cable will permit an exchange of fill solution at the membrane surface. Re-Calibration is required.
6
7
Loading...
+ 7 hidden pages