HI 4101 Ammonia ElectrodeHI 4101 Ammonia Electrode
HI 4101 Ammonia Electrode
HI 4101 Ammonia ElectrodeHI 4101 Ammonia Electrode
I. I.
Introduction:Introduction:
I.
Introduction:
I. I.
Introduction:Introduction:
The Hanna HI 4101 Ammonia gas selective electrode is a
combination electrode designed for the measurement of
ammonia in aqueous solutions such as waste water samples,
wine, beer. Ammonium ions are also measured by conversion to ammonia gas upon ISA addition.
II.II.
SpecificationsSpecifications
II.
Specifications
II.II.
SpecificationsSpecifications
Type:NH3 gas sensing
electrode with glass pH
internal, Ag/ AgCl
reference and gas
permeable PTFE
membrane.
Species Measured:NH
+
, NH
4
3
Measurement Range:1.0 M to 1x 10-6M
17000 to 0.02 ppm
Interfering ions:Surfactants, wetting
agents, volatile amines.
Operating Temperature:0 to 40°C
Operating pH:>11 pH
Dimensions:12 mm (OD) X 120
mm (insertion)
0.47”x 4.72”
Wetted materials:Delrin®, body and cap
PTFE gas membrane
Connection:BNC
3
III.III.
Theory of OperationTheory of Operation
III.
Theory of Operation
III.III.
Theory of OperationTheory of Operation
::
:
::
The ammonia electrode is a complete potentiometric cell
that contains both a silver/silver chloride (Ag/AgCl) reference and a pH measurement element. These elements are
housed within a thermoplastic body in a chloride ion-containing electrolyte, and are isolated from the sample by a
gas permeable membrane made of polytetrafluoroethylene
(PTFE).
Dissolved gas in the sample solution diffuses into the membrane and changes the pH in the thin film of electrolyte on
the surface of the pH glass. Diffusion continues until the
partial pressures of the gas in the sample and thin film are
equal. The change in pH is proportional to the concentration of dissolved gas in the sample solution.
The Nernst expression for an ammonia sensor is expressed
in the equation below. Note that the potential is a function
of the ammonia gas, which in turn is related to the hydroxyl
ion concentration. The glass internal, Ag/AgCl reference,
and Henry’s law constant are rolled into the E’ and E
constants. The Nernst equation for the sensor becomes the
equation noted below:
o
E = E’–2.3RT/nF log [A]= Eo–0.059 log [OH-]
E = observed potential
E’ = Reference and fixed internal voltages
R = gas constant (8.314 J/K Mol)
n= Charge on ion (equivalents/mol)
A
= ion activity in sample
ion
T = absolute temperature in K
F = Faraday constant (9.648 x 104 C/equivalent)
The mV should decrease in a Nernstian manner as the
ammonia partial pressure increases in the sample.
4
IV.IV.
Design ElementsDesign Elements
IV.
Design Elements
IV.IV.
Design ElementsDesign Elements
The Hanna HI 4101 ammonia gas sensor has 3 main
parts. These are the membrane/membrane cap, outer probe
body with antirotation key and the pH/reference assembly
which includes the outer electrode cap, spring, inner cap
and pH/reference electrode assembly.
pH/reference electrode assembly
cable
outer electrode
cap
spring
antirotation key inner cap
Outer probe body
reference
electrode
pH
sensitive
membrane
membrane/membrane cap
5
V.V.
Equipment Required:Equipment Required:
V.
Equipment Required:
V.V.
Equipment Required:Equipment Required:
•Hanna HI 4222 pH/ISE/mV meter or other suitable
ion or pH/mV meter. (Note: log/linear graph paper is
useful if an ISE meter is not available).
•Hanna HI 180 magnetic stirrer or equivalent with
stirring bars. (Note: Isolate beakers from stirrer motor
heat by placing insulating material such as foam or
cork between them).
•2 or 3 necked flask with stoppers or
•Hanna HI 76404 electrode holder or equivalent with
Beakers or other suitable measurement vessel with
plastic sealing film or wrap.
VI. VI.
Solutions Required for Calibration:Solutions Required for Calibration:
VI.
Solutions Required for Calibration:
VI. VI.
Solutions Required for Calibration:Solutions Required for Calibration:
Ionic Strength Adjuster (ISA):HI 4001-00
Hanna 0.1 M standard:HI 4001-01
Hanna 100 ppm N standard:HI 4001-02*
Hanna 1000 ppm N standard:HI 4001-03*
*Please Note: These calibration standards are ppm as
NH3-N.
See Section XVII for additional solutions used for maintenance.
Using volumetric pipettes and glassware make dilutions of
the standard to bracket the concentration of the samples.
Standards with concentrations less than 10-3M should be
prepared fresh daily. Store solution in a tightly sealed
bottle without ISA added. 2 mL of HI 4001-00 ISA should
be added to each 100 mL sample of standard and samples
just prior to measurement. ISA adjusts the pH of the sample
or standard to about pH 11 thus converting ammonium ion
to ammonia. It also provides samples and standards a
constant ionic strength background that stabilizes the solutions activity coefficient and permits concentration to be
measured directly. The ISA provides color indication to verify
it has been added to the solution and a complexing agent
to remove metal ions (i.e copper, zinc) from solution. These
6
ions are capable of reducing the ammonia concentration.
If other volumes of sample/standard are used, add ISA at 2
parts per 100 parts standard/sample.
VIIVII
General GuidelinesGeneral Guidelines
VII.
General Guidelines
VIIVII
General GuidelinesGeneral Guidelines
•Calibration standards and sample solutions should
have the same ionic strength. ISA should be added to
both samples and standards immediately before taking measurements.
• Calibration standards and sample solutions should be
at the same temperature. Thermally insulate solution
vessel from magnetic stirrer with cork or other insulating medium.
•Calibration standards and sample solutions should
be stirred at the same rate using identical sized stir
bars.
•Surface coating or “wetting“ the PTFE membrane will
effect the response. Inspect sensor before using. Replace PTFE membrane if damage is evident.
•Rinse electrode with distilled or deionized water between samples and dab dry with lab wipe or other soft
disposable absorbent toweling.
•Check calibration every 1-2 hours.
•Position sensors at an angle of approximately 20° to
lessen bubble adherence from solution out-gassing
due to temperature change.
• Close container with plastic wrap or use
2 or 3 necked flask to prevent gas from leaving.
•Gently pulling cable will permit an exchange of fill
solution at membrane surface. Re-Calibration is required.
7
VIII. VIII.
Inner Electrode CheckInner Electrode Check
VIII.
Inner Electrode Check
VIII. VIII.
Inner Electrode CheckInner Electrode Check
Before assembling the electrode for the
first timefirst time
first time or if
first timefirst time
reactivating it after storage, the inner electrode assembly
should be conditioned and then tested as a pH electrode.
Prepare pH test solutions HI 4000-47-4 and HI 4000-477 by mixing and dissolving each buffer packet in separate
containers with 50 mL deionized water. These pH solutions
contain chloride ions and pH buffers that are used to verify
the inner electrode (pH internal) is operational. See
Section XVII for replacement accessories and maintenance
items.
For a new sensor:
Remove the protective shipping cap from the glass inner
electrode.
Protective shipping cap
For existing sensor:
Unscrew the upper cap on the top of the electrode and
carefully withdraw the internal pH/reference assembly.
For stable readings, glass should be covered to the
bottom of the long black band.
Use test
tube or
Test buffer
can be
used as a
conditioning
solution
for the
pH internal
graduated
cylinder
(weighed
Bottom) when
testing or
conditioning
pH internal
8
•If sensor has been stored or shipped dry, it should be
“conditioned” by soaking the pH/reference assembly 1 hour or more in one of the pH test solutions.
• Avoid touching the pH glass with your fingers.
• Attention: The pH/reference assembly is fragile!
Support the upper portion of the internal cell while
immersing the glass and reference assembly. A tall
narrow container with weighted bottom is best. The
pH test solution should cover the bottom of the large
black ban.
Test: Connect the BNC connector on the electrode cable to a
pH/mV (mV or ORP mode) meter. Carefully immerse
the sensor assembly into one of the buffers. When the
measurment stabilizes record the mV generated. Rinse
sensor tip in deionized water and dab dry between
buffers to prevent solution carry-over. Do not rub the
glass. Take a measurement in the second buffer and
record mV. Pay attention to minus sign if present.
Calculate the difference in mV between the two solutions.
Example of typical values:
HI 4000-47- 7-90.2 mV
HI 4000-47-4 80.66 mV
Difference 170.8 mV= 80.6-(-90.2)
A calculated value equal or greater than 160 mV is acceptable for ambient temperatures between 20° and 25°C.
IX. IX.
Electrode PreparationElectrode Preparation
IX.
Electrode Preparation
IX. IX.
Electrode PreparationElectrode Preparation
1)Remove glass internal from sensor body and perform
inner electrode check. (See section VIII).
2)Install membrane on the outer probe body. Use twee-
zers provided and avoid touching working area of
membrane with your fingers as skin oil will change
the hydrophobic properties. Discard the paper backing (blue) found between white PTFE membranes.
Hold membrane at corner with tweezers and drape
over lower opening of outer probe body.
9
3)Hold one corner against the threads with thumb while
gently stretching membrane over opening and capturing opposite membrane corner against threads .
Smooth excess membrane material around the threads.
4)Screw outer membrane cap onto body
thus capturing the membrane between the cap and
outer body threads.
10
5.) Using dropper provided, add about 2 mL of ammonia internal electrolyte HI 4001-40 into outer probe
body.
6)Insert and position the inner glass/reference assembly into the outer body so that the anti-rotation key
sits in the cut out on the outer probe body.
HI 4101
IA
N
O
M
M
A
HI 4101
IA
N
O
M
M
A
7)Holding the electrode upright, slide spring and electrode cap down cable and screw cap on outer body
until fully engaged. Do not invert electrode. Do not
overtighten.
8)Install assembled electrode in gas sensor test vessel
or in electrode holder and connect cable connector to
pH/mV meter.
11
X. X.
Quick Check of Electrode SlopeQuick Check of Electrode Slope
X.
Quick Check of Electrode Slope
X. X.
Quick Check of Electrode SlopeQuick Check of Electrode Slope
•Connect BNC (connector) to pH/mV/ISE meter.
•Place meter in mV mode.
• Place 100 mL of deionized water into a vessel with
stir bar. Add 2 mL of ISA Hanna HI 4001-00.
•Place sensor into prepared sample.
•Add 1 mL of 1000 ppm or 0.1 M Ammonia standard
to beaker. Record the mV value when stable.
•Add an additional 10 mL of standard to the solution.
Record the mV when reading has stabilized. This
value should be less than the previous noted (more
negative).
•Determine the difference between the two mV values.
An acceptable value for this slope is 54±4 mV at
ambient temperatures between 20 and 25°C.
XI.XI.
Corrective actionCorrective action
XI.
Corrective action
XI.XI.
Corrective actionCorrective action
•Verify that the upper cap has been screwed in all the
way.
•Verify electrode is connected properly to meter and the
meter is is powered.
•Verify ISA has been added in the correct ratio to the
standard.
•Examine the white membrane and check for electrolyte that might have leaked through the PTFE film.
Replace membrane if damaged.
•If sensor does not change mV verify the the glass
assembly is operational (See section VIII).
12
Loading...
+ 28 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.