Hanna Instruments HI 3880 User Manual

Instruction Manual
HI 3880
pH 4.0-6.5
Test Kit
www.hannainst.com
Dear Customer, Thank you for choosing a Hanna Product. Please read the instructions carefully before using the
chemical test kit. It will provide you with the necessary information for correct use of the kit.
Remove the chemical test kit from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable damage, notify your Dealer or the nearest Hanna office immediately.
Each kit is supplied with:
HI 3880-0 pH 4.0-6.5 Reagent, 1 bottle with
• dropper (25 mL);
1 color comparator cube.
Note: Any damaged or defective item must be returned in
its original packing materials.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
Range 4.0 to 6.5 as pH unit Smallest Increment 0.5 as pH unit Analysis Method Colorimetric Sample Size 5 mL Number of Tests 100 Case Dimensions 115x102x82 mm (4.5x4.0x3.2")
Shipping Weight 110 g (3.9 oz.)
SIGNIFICANCE AND USESIGNIFICANCE AND USE
SIGNIFICANCE AND USE
SIGNIFICANCE AND USESIGNIFICANCE AND USE
pH represents acidity or alkalinity of an aqueous solution and is proportional to the hydrogen-ion concentration of the solution. Under neutral conditions water is dissociated into the OH- and H+ ions in equal ratio and hence it has a pH of 7. When bases or acids are added to a water solution they ionize, increasing the concentration of OH- or H+, respectively. Thus solutions with a pH of 1-3 contain strong acids, whereas those with a pH of 4-6 contain weak acids. Weak bases result in solutions of pH 8-10 and strong bases in pH of 11-13.
Examples of pH value for some liquids: LiquidLiquid
Liquid
LiquidLiquid sea water 7.8-8.2 gastric juices 1.7 milk 6.5-7 soil 6-7 (optimum for crops)
CHEMICAL REACTIONCHEMICAL REACTION
CHEMICAL REACTION
CHEMICAL REACTIONCHEMICAL REACTION
HI 3880-0 Hanna Reagent reacts in contact with the aqueous solution changing its color according to the hydrogen­ion concentration (pH) in the given range.
pH ValuepH Value
pH Value
pH ValuepH Value
INSTRUCTIONSINSTRUCTIONS
INSTRUCTIONS
INSTRUCTIONSINSTRUCTIONS
READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
Note:
always shake the reagent bottle before use.
• Fill the color comparator cube with 5 mL of the sample.
• Add 3 drops of reagent HI 3880-0.
• Replace the cap and mix the solution by inverting the cube several times.
• Determine which color best matches the solution in the cube and record the result as pH unit.
• It is better to match the color with a white sheet at about 10 cm behind the color comparator cube.
Note: in case of a concentration of chlorine above 50 ppm
require the HI 3880/O Test Kit. To measure pH in the 6.0-8.5 range use the HI
3881 pH 6.0-8.5 Test Kit. To measure pH in the 7.5-10.0 range use the HI
3886 pH 7.5-10.0 Test Kit.
REFERENCESREFERENCES
REFERENCES
REFERENCESREFERENCES
Vogel's Quantitative Chemical Analysis 5th Ed. Longman Scientific & Technical.
HEALTH AND SAFETYHEALTH AND SAFETY
HEALTH AND SAFETY
HEALTH AND SAFETYHEALTH AND SAFETY
The chemicals contained in this kit may be hazardous if improperly handled. Read Health and Safety Data Sheet before performing this test.
ISTR3880R2 02/03-L PRINTED IN ITALY