Dear Customer, Thank you for choosing a Hanna Product. Please read the instructions carefully before using the
chemical test kit. It will provide you with the necessary information for correct use of the kit.
Remove the chemical test kit from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any noticeable dam­age, notify your Dealer or the nearest Hanna office immediately.
Each kit is supplied with:
1 color comparator cube;
• Reagent 1 (30 mL);
• Reagent 2 (20 mL).
Range 0 to 2.5 mg/L (ppm) Iodine Smallest Increment 0.5 mg/L (ppm) Iodine Analysis Method Colorimetric Sample Size 25 mL (average) Number of Tests 50 (average) Case Dimensions 220x145x55 mm (8.7x5.7x2.1") Shipping Weight 180 g (6.7 oz.)
SIGNIFICANCE AND USESIGNIFICANCE AND USE
SIGNIFICANCE AND USE
SIGNIFICANCE AND USESIGNIFICANCE AND USE
Iodine may be used as a disinfectant in drinking and swim­ming pool waters. Unlike chlorinated pools, water treated with iodine decrease eye irritation among swimmers and provides a disinfectant more stable to adverse conditions. However, iodine levels in the water must be frequently checked to maximize its given purpose.
The Hanna Iodine Test Kit employs a color cube to deter­mine the iodine content in water. The kit is compact and is practical for field use. No chlorine or bromine can be present in the water sample for this test to work properly.
CHEMICAL REACTIONCHEMICAL REACTION
CHEMICAL REACTION
CHEMICAL REACTIONCHEMICAL REACTION
The iodine is measured by a colorimetric method. The sample is initially treated with a phosphate buffer to a pH of approximately 6.3. The addition of DPD (N,N-diethyl-p­phenylenediamine) is immediately oxidized by iodine producing a reddish color. The color intensity of the solution determines the iodine concentration.
READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
• Add 7 drops of Reagent 1 and 3 drops of Reagent 2 in the color comparator cube and mix by carefully swirling the cube in tight circles.
• Fill the color cube with water sample to the 25 mL mark.
• Cap and invert several time to mix.
• Determine which color matches the solution in the cube and record the results as mg/L (or ppm) iodine.
Standard Methods for the Examination of Water and Waste-
th
water, 16
Edition, 1985.