Hanna Instruments BL 7916 User Manual

Instruction Manual
BL 7916 • BL 7917
pH and ORP
Measuring & Dosing
System
www.hannainst.com
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the pump. If you need additional technical information, do not hesitate to e-mail
us at tech@hannainst.com. These instruments are in compliance with the directives.
WARRANTY
All Hanna Instruments pumps are warranted for one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed main­tenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Num­ber from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection.
TABLE OF CONTENTS
PRELIMINARY EXAMINATION ........................................................ 3
GENERAL DESCRIPTION ................................................................ 4
FLOW RATE CHART .......................................................................5
FUNCTIONAL DESCRIPTION BL 7916 ............................................ 6
FUNCTIONAL DESCRIPTION BL 7917 ............................................ 7
SPECIFICATIONS........................................................................... 8
VALVE / HOSE ASSEMBLY DIAGRAM............................................. 9
MECHANICAL DIMENSIONS ........................................................ 10
INSTALLATION ........................................................................... 11
OPERATIONAL GUIDE ................................................................. 18
pH CALIBRATION ....................................................................... 22
TROUBLESHOOTING GUIDE ........................................................ 23
MAINTENANCE ........................................................................... 24
ELECTRODE CONDITIONING AND MAINTENANCE ........................... 26
TAKING REDOX MEASUREMENTS ................................................. 30
CHEMICAL COMPATIBILITY GUIDE................................................ 32
ACCESSORIES ............................................................................. 34
CE DECLARATION OF CONFORMITY.............................................. 38
PRELIMINARY EXAMINATION
Each pump is supplied complete with:
• discharge and suction valves
• LDPE tubing, 7 m (23')
• power cord
• instruction manual
Note: Save all packing material until you are sure that the pump
functions correctly. Any defective item must be returned in the original packaging together with the supplied accessories.
READ THE INSTRUCTIONS ATTENTIVELY
BEFORE INSTALLING OR OPERATING YOUR PUMP
The BL electronic dosing pumps are easy to use. We recommend, however, that you read the entire manual before using the pump. Familiarity with the features and controls of the unit will give you a better idea of the dosing potential and help reduce operator errors. Please operate the pump only as directed in the instruction manual. Follow all general safety guidelines during operation.
Remember: electrical devices are potentially hazardous. Check that the voltage of the installation matches the voltage indicated on the specification label on the back of the pump.
Note: It is the responsibility of the user to install and ground the
pump properly; it is highly recommended to install an external switch.
Each pump is protected by a 1 A/250 V fuse that is located together with 1 spare in a drawer on the power socket under the pump.
Always store chemicals in safe, out of reach places. Follow the directions for use with each chemical. Do not assume chemicals are the same because they look alike. Hanna Instruments cannot be held responsible for the misuse of chemicals or the pump.
Always wear protective clothing (gloves and safety glasses) when working near chemical dosing pumps. When pumping chemicals, make sure all tubes are securely attached to the fittings. It is recommended that tubing is shielded to prevent possible injury in case of rupture or accidental damage.
Avoid using a pipe wrench or pliers on plastic parts and connectors.
2
3
These are best tightened with an open end or crescent wrench. Avoid overtightening these parts as this could cause damage to the seats and threads.
If a hose is used, it should be securely fastened to columns, walls, braces, etc. This will ensure that the hose connection will remain tight and leak free. Shield the hose from direct sunlight. Sunlight can cause an autocatalytic reaction with some chemicals and weaken the hose walls.
The arrow on the pump head indicates the direction of chemical flow and should always point upwards (vertically). Never position the pump horizontally with suction and discharge valves horizontal. Locate the pump in an area out of the reach of children and pets.
All pumps undergo stringent tests to ensure that they comply with their stated specifications and are calibrated at the maximum rated pressure.
Unplug the instrument from the power supply before replacing the fuse or making any electrical connections.
FLOW RATE CHART
The following chart shows the inverse relationship between flow rate and pressure.
GENERAL DESCRIPTION
The BL 7916 and BL 7917 Control/Pump System offer respectively a pH and ORP monitoring system with proportional control of a diaphragm pump and an LCD readout.
Features include:
• Two advanced instruments in one compact unit
• Proportional control for precisely maintained set-points
• Rugged construction with a one-piece casing and a transpar­ent cover to protect controls and terminals
• Chemically resistant non-clogging pump head and superior materials for all components in contact with the chemicals being dosed (see page 32 for details)
• Convenient installation with all controls on front panel
• A solenoid-driven pump
• Automatic overheat protection and a built-in LCD display
• Alarm output: the alarm of the BL 7916 will be activated if the measured pH value is 2 pH units higher or lower than the setpoint. BL 7917's alarm will activate if the mV value is 200mV higher or lower than the setpoint.
• Auxiliary dosing contacts. This will drive other equipment such as mixers, priming pumps, etc.
4
The table below shows typical reduction of the flow rate with an increase of pressure. The pump supplied with the system has a capacity of 13.3 LPH (3.5 GPH) at 0.5 BAR (7.4 PSI).
BL 7916 / BL 7917
FLOW / PRESSURE
BAR (PSI) LPH (GPH)
0.5 (7.4)
1.0 (14.7)
2.0 (29.4)
3.0 (44.1)
4.0 (58.8)
13.3 (3.46)
11.7 (3.04)
10.1 (2.63)
9.0 (2.33)
7.8 (2.03)
5
FUNCTIONAL DESCRIPTION BL 7916
FUNCTIONAL DESCRIPTION BL 7917
1. Discharge Valve Assembly
2. Pump head
3. Suction Valve Assembly
4. Hose
5. Liquid Crystal Display
6. Offset Calibration Trimmer
7. Setpoint Adjustment Trimmers (FINE and COARSE)
8. Slope Calibration Trimmer
9. Acid/Base Selection Switch
10. Display Mode Selection Switch (SET or MEASURE)
11. Terminal Connections
12. BNC Connector for pH electrode
13. Overheating LED
14. Power Socket and Fuse Holder
15. Cable Glands
Unplug the instrument from the power supply before replacing the fuse or making any electrical connections.
6
1. Discharge Valve Assembly
2. Pump head
3. Suction Valve Assembly
4. Hose
5. Liquid Crystal Display
6. Setpoint Adjustment Trimmers (FINE and COARSE)
7. Reduction/Oxidation Selection Switch
8. Operating Mode Selection Switch (SET or MEASURE)
9. Terminal Connections
10. BNC Connector for ORP electrode
11. Overheating LED
12. Power Socket and Fuse Holder
13. Cable Glands
Unplug the instrument from the power supply before replacing the fuse or making any electrical connections.
7
SPECIFICATIONS
BL 7916D BL 7916U
Range 0.00 to 14.00 pH
Resolution 0.01 pH Accuracy (@20°C/68°F) ±0.01 pH
Typical EMC Deviation ±0.1 pH Input Impedance 10¹² Ohm Dosage Proportional, acid or base, user selectable
Dosing Contact Isolated, 2A, Max. 240V, resistive load, 1,000,000 strokes Alarm Contact Isolated, 2A, Max. 240V, resistive load, 1,000,000 strokes Calibration Offset: ±1 pH with trimmer
Slope: 85 to 115% with trimmer
Recorder Output 4 to 20 mA (isolated) Power Supply 230 Vac ±15% 115 Vac ±15%
50/60 Hz (40 W) 50/60 Hz (40 W)
Environment 0 to 50°C (32 to 122°F);
RH max 95% non-condensing
Dimensions 221 x 142 x 181 mm (8.7 x 5.6 x 7.1'')
Weight Approximately 5 kg (11 lb.)
BL 7917D BL 7917U
Range -999 to +999 mV Resolution 1 mV Accuracy (@20°C/68°F) ±5 mV Typical EMC Deviation ±6 mV Input Impedance 10¹² Ohm Dosage Proportional, oxidizing or reducing, user selectable Dosing Contact Isolated, 2A, Max. 240V, resistive load, 1,000,000 strokes Alarm Contact Isolated, 2A, Max. 240V, resistive load, 1,000,000 strokes Recorder Output 4 to 20 mA (isolated)
Power Supply 230 Vac ±15% 115 Vac ±15%
50/60 Hz (40 W) 50/60 Hz (40 W)
Environment 0 to 50°C (32 to 122°F);
RH max 95% non-condensing
Dimensions 221 x 142 x 181 mm (8.7 x 5.6 x 7.1) Weight Approximately 5 kg (11 lb.)
VALVE / HOSE ASSEMBLY DIAGRAM
NECK
8
9
MECHANICAL DIMENSIONS
110mm
m "
1
m "
110mm
8.70"
INSTALLATION
The Controller/Pump series of instruments are enclosed in a modular housing for maximum protection. The dimensioned illustrations show the layout of the Controller/Pumps and how they utilize the one-piece polypropylene, injection-molded housing. Since there are no joints or screws holding different sections of the housing together, the case is extremely rugged and sturdy.
BOTTOM VIEW
181m
164mm
6.45"
17mm
.67"
142m
5.60
7.13
29mm
5.08"
99mm
3.90"
5mm
0.19"
4.33"
55mm
2.16"
204mm
221mm
4.33"
FRONT VIEW
pH
8.03"
Materials Needed
• LDPE hose (7 meter/22 feet) (included) or other type of tubing (PTFE, for example) more suitable for a specific application (optional)
• Power cord (included)
Optional Accessories
• 4 each, ceramic weights (HI 721008)
• 1 each, foot valve assembly (HI 721005)
• 1 each, injection valve assembly (HI 721004)
Location
A suitable location should:
• be near to a power source
• be conveniently close to the injection point
• allow easy access to the flow rate control and pipe or hose connections
• be no more than 1.5 meters (5 feet) above the operating position of the suction valve assembly.
Dimensions for Installation
BlackStone Pumps are designed for permanent installation. The pump can be mounted directly on a wall or tank (see page 10 for
the specific mounting dimensions).
Power Requirements
BlackStone pumps are designed to operate to specifications within the following voltage ranges:
100 - 130 Volts for 115 Vac models 200 - 250 Volts for 230 Vac models
10
11
To ensure maximum performance, check the voltage at the point of
-
supply to verify that it is sufficient. It is recommended that you install a 1 Amp circuit breaker between the pump and the power supply. This will give additional protection to the internal circuit and provide a convenient way to disconnect the power supply prior to servicing the pump, if needed.
Injection Point
• Choose an injection point that allows you to mount the injection valve assembly vertically.
• The spring in the injection valve assembly (HI 721004) adds approximately 1.5 bar of back pressure. If pumping into a high back pressure, the spring should be removed.
Other Considerations
• If you are mounting the system to a wall, column, etc., be sure it is strong enough to support the weight of the entire system.
• The ambient temperature of the pump, when in operation, should be between 0 and 50°C (32 to 122°F) and should be protected from direct exposure to outdoor elements (direct sun­light, rain, extreme temperatures, high humidity, etc.).
• Generally speaking, the shorter the suction distance, the more efficient the pump operates.
• The pump should be placed in a conventional location that will allow easy access to the control and connections. It should be placed so that regular visual inspections of the connections and hoses are facilitated.
Vertical Surface Mounting
Once you have selected the best installation site, simply screw or bolt the unit into a wall or mounting panel above the chemical feed tank.
The 4 mounting screw holes on the pump will accommodate up to a 5 mm (3/16") screw or bolt (remember to use heavy screws or bolts to secure the system). Be sure you do not over tighten and cause excessive stress on the mounting holes.
Make sure to leave a slight overhang in front to allow for the connection cable.
Power Supply
Connect the power cord to the female socket of the pump and by doing so also grounding it. The power socket contains a 1 A/250 V fuse. Since there is no on/off switch, it is suggested to install an outside switch.
Probe connections
Connect the pH/ORP electrode to the BNC socket of the pump.
Permanent Connection using 3/8" PVC pipe
All piping for the pump feed and dis­charge should be plumbed to the location of the pump.
The threads on both valve assemblies allow the use of standard 3/8" (Euro­pean) pipe fittings for permanent pipe connections.
Diagram for Rigid Pipe Hose
UNION C ONNECTION
ADAPTER
DISCHARGE VALVE
PUMP
SUCTION V ALVE
ADAPTER
FOOT V ALVE
FILTER
The foot valve assembly (HI 721005) should always hang vertically and not lay horizontal on the bottom of the tank or drum.
A vertical assembly will ensure that the valve is positioned properly and prevent loss of prime.
For the US standard installations, use PVC adapters to connect the suction and discharge valves to the PVC pipe.
12
13
Hose Connections
• Cut a long enough section of the hose to reach the suction valve of the pump head from the feed tank. Al­low some slack in the hose and be sure it is not kinked or twisted.
• Slip a hose connector onto the hose over the head valve and up to the bottom of the threads ensuring it is fully seated.
• Slide the connector up to the threads and tighten to form a seal.
• Slip the ceramic weight (HI 721008) and a connector over the other end of the hose.
• Repeat the same installation procedure for the hose connections on the discharge end with the injection assembly (HI 721004)
• Secure the hose so that its movement is minimized when the pump is operating. Excessive hose movement could cause the connectors to loosen and result in leakage.
Assembling the Hose to the Valve
The end of the valve is specially tapered to form a leak free seal when the hose is properly installed.
Be sure to seat the hose completely so that there is no gap. Push the hose until it covers the end of the valve completely.
• Attach the foot valve assembly (HI 721005) to the hose and slide the connector up to the threads and tighten to form a seal.
14
Suction and Discharge Valves
The suction and discharge valves located on the pump head should not be inter­changed as they are different internally. The discharge valve is fitted with a valve guide and will not function properly if used on the suction side.
15
LEGEND
HOSE PIPE
CONNECTION
RESEVOIR
MANUAL SHUTOFF VALVE
PUMP
FILTER / FOOT VALVE
CHECK VALVE
EXAMPLE OF TYPICAL INSTALLATIONS
Flooded Suction Installation
Suggested Installation for consis­tent output when using a low stroke rate. Also suggested for highly viscous chemicals.
A slight suction pressure avoids self-priming problems, especially with high viscosity liquids.
Suction Lift Installation
Suggested installation for most in-line applications with nominal output and pressures.
The maximum self-priming height is
1.5 m (5 ft.). It is advisable to install
a level controller in order to stop the pump when feed tank (reservoir) liquid level is low.
Downhill Installation
Suggested installation when pumping from one container to another, each at different levels and with only nominal pressure.
START-UP
At start-up, purge all chemical gases and air from the suction tubing, valves and pump head. Start the pump.
When all the air or gas is vented, the solution being metered will appear in the output line.
Note: Only when operating under pressure, the pump must be
started unloaded.
Uphill Installation
Suggested installation whenever the supply is located higher than the discharge point; typically a waste water application.
It is important to install the Injection valve to prevent siphoning.
16
17
OPERATIONAL GUIDE
• Unscrew and remove the transparent front panel cover and gasket seal for access to the terminals.
• Remove the protective plastic plate covering the terminals by removing the 2 screws on both sides of the plate.
+ -
+ -
TERMINAL CONNECTIONS BL 7917
Pt100 (for BL 7916 only)
A 2-wire Pt100 can be connected to provide automatic temperature compensation of the pH measurements. The pump is supplied with a 110 Ohm resistor connected to the 2 Pt100 terminals. This delivers a fixed temperature compensation of 25°C (77°F). The Pt100 is recom­mended only in special instances where very high accuracy is absolutely necessary since the error is only 0.03 pH for a temperature difference of 10°C in the pH 6 to 8 range. The error will consequently be less than 0.09 pH at readings from pH 4 to 10 when the temperature is in the 15°C (59°F) to 35°C (95°F) range.
PROPORTIONAL SETTING
The pump is supplied with a 10K resistor connected to these terminals. With this value the pump works at 100% of the rating when the difference between measure and setpoint is more than 150 mV (BL 7917) or 1.5 pH (BL 7916). It goes into proportional dosing for values less than the above.
You can vary this hysteresis band by simply changing the resistor as follows:
BL 7916 BL 7917
0.50 pH 1.91 K 50 mV 1.91 K
1.00 pH 4.87 K 100 mV 4.87 K
1.50 pH 10 K 150 mV 10 K
2.00 pH 21 K 200 mV 21 K
TERMINAL CONNECTIONS BL 7916
18
LEVEL CONTROL
The pump is supplied with these two terminals short-circuited. A contact coming from a level controller (e.g. HI 7871, HI 7873) can be connected to these terminals. If this contact is closed, the pump works normally. If it opens, the unit will not pump even if the controller commands it.
19
For example, a level controller can be placed in the tank of the liquid being dosed so that the pump is stopped when the chemicals are exhausted.
MANUAL
An auxiliary switch can be connected to these terminals to force the unit to pump whatever value is measured by the controller. This might be needed for example when the pump needs to be primed.
AUXILIARY CONTACT
The controller can drive a relay when the measure overtakes the setpoint. The contact available on the terminals can be normally open or closed. It can be used as a switch to drive an external mixer when the pump is dosing, or can activate an auxiliary pump or any other device. With BL 7916, the contact can be selected to remain open when the pH value is within the set limits (the pump is not dosing) and closed when the pump is dosing or viceversa.
By shorting the Common (middle) terminal with the NO or NC terminal on the driving circuit inside the controller section, a normally open or a normally closed state can be achieved, as shown below:
ALARM CONTACT
If the reading drifts from the setpoint by more than ±2.00 pH in BL 7916 and ±200 mV in BL 7917, an external alarm can be activated. The contact is rated at 2 amps at 220VAC. No power is supplied through the system and it is an open/closed contact only.
With BL 7916, the contact can be selected to be closed when the alarm is activated and open when the alarm is off or viceversa. Short the middle terminal located on the driving circuit inside the controller section with the NO or NC terminal (see above).
4-20 mA ISOLATED OUTPUT
A 4-20 mA proportional output is available in these connections:
4 mA=0 pH; 20 mA=14 pH (BL 7916)
and
4 mA=-999 mV; 20 mA=+999 mV (BL 7917).
The output is optically isolated.
20
SETPOINT ADJUSTMENTS
• Move the display selector switch to SET.
• With a small screwdriver adjust the SET COARSE trimmer to display a value close to the desired setpoint e.g. 7.00.
• Adjust the SET FINE trimmer to display the exact setpoint value.
• Once the desired setpoint is achieved, move the display selector switch back to MEASURE.
ACID OR BASE SELECTION (BL 7916 only)
If an acid chemical is to be dosed, move the ACID/BASE switch to ACID. This means that the pump will dose when pH measurements are higher than the setpoint.
If an alkaline chemical is to be dosed, move the ACID/BASE switch to BASE. This means that the pump will dose when pH measure­ments are lower than the setpoint.
REDUCING OR OXIDIZING SELECTION (BL 7917 only)
If a reducing chemical is to be dosed, move the RED/OXID switch to RED. This means that the pump will dose when redox measure­ments are higher than the setpoint.
If an oxidizing chemical is to be dosed, move the RED/OXID switch to OXID. This means that the pump will dose when the redox measurements are lower than the setpoint.
The wires and cables needed for all connections can be fed through the two cable fittings located below the terminals. It is important to attach these cables properly because pump vibrations could loosen them. Replace the transparent cover and fasten the screws.
21
pH CALIBRATION
TROUBLESHOOTING GUIDE
BL 7916 only:
• Ensure that the display selector switch is set to MEASURE and dip the electrode tip in a neutral buffer solution (pH 7.01).
• Adjust the OFFSET trimmer to display 7.01 on the LCD.
• Rinse the electrode with clean water and dip the electrode tip in pH 4.01 (acidic) or pH 10.01 (alkaline) buffer solution.
• Adjust the slope trimmer to read 4.01 or 10.01 on the LCD.
The slope calibration with pH 4.01 buffer is suggested if the controller will be used for acid measurements or pH values below 7, pH 10.01 is in turn recommended for alkaline measurements or pH values above 7.
22
ELECTRICAL
The pump does not operate when turned ON:
• Check the power supply and connections. Voltage should be between 100 - 130 VAC for 115 V models and between 200
- 250VAC for 230 V models.
• See Installation section on page 11 or call your dealer or the nearest Hanna Customer Service Center for technical assis­tance.
OPERATING
Display does not indicate:
Check that the pump is properly plugged in
• Check the fuse.
Display shows 1 on the far left hand side:
Check electrode and/or electrode cable.
"No Dosage" LED is lit:
Check the tank where the level controller is working or verify
the connection on the terminals (they must be short for the pump to dose).
LIQUID
The pump operates but does not prime:
Check for a clogged or loose filter on the suction valve
assembly. Retighten if necessary.
Check to see if the pump is too high above the foot valve assembly (HI 721005) in the feed tank. This vertical distance should not exceed 1.5 meters (5 feet). Either lower the pump or raise the feed tank.
Check the pump head, suction and discharge valves for blockage.
Pump flow rate is reduced:
Check the pump head, discharge and injection valve assembly
for any clogging. Clean and reassemble.
Check for any additional back pressure created since the last flow rate was conducted.
Check for any changes in the viscosity of the chemical being used.
Be sure that valves have been properly installed in the pump head.
23
Leakage at the connections:
Be sure that the hose is fully seated and hose connectors are
tight.
Be sure that valves are tight and O-rings are in place.
Leakage around the pump head:
Be sure that the valves are tight and O-rings are in place and
the head screws (hex bolts) are tight.
MAINTENANCE
Your BlackStone Pump is designed to give you years of trouble-free service. Maintenance should be the preventative type, that is, periodic cleaning and inspecting for any damage or leakage.
CLEANING THE SUCTION, DISCHARGE AND INJECTION VALVES
Remove the valves from the pump head, the injection fitting and the feed.
Keep the suction and discharge valves separated as they are not interchangeable.
Disassemble each valve and clean it with a neutral liquid. Inspect the PVDF springs. After cleaning the glass balls, inspect them for any excessive wear due to abrasion from the chemical. Replace if necessary with parts from HI 721102, HI 721105, HI 721004 and HI 721005 (see page 34 for listing).
When reinstalling the valves into the pump head, tighten by hand first and then with a wrench ¼ to ½ turn.
SCHEDULED MAINTENANCE
After 50 hours
Tight the pump head screws with a torque force of 2.5 Nm (22" lbf).
After 12 months
It is recommended to replace HI 721102, HI 721103 (suction and discharge valves assemblies) as well as the O-rings. The LDPE hose can also deteriorate over time and, for safety reasons, should also be changed with HI 720032.
After 24 months
It is recommended to replace HI 721102, HI 721103, HI 720032 and HI 721106.
INSPECTING THE HOSE (if used as supplied with the pump)
Inspect to see if the hose has worn out or weakened due to the chemicals. Pay particular attention for any signs of abrasion or discol­oration. Also check the connectors to ensure they are tight.
Replace if necessary with parts from HI 720032.
CLEANING THE PUMP HEAD
The pump head should be cleaned at regular intervals and at least once a year. Remove the deposits that form in the cavities with a solution that is neutral to the chemical the pump has been dosing. Inspect the head for any cracks or worn areas.
Replace if necessary with parts from HI 721106 (pump head).
24
25
ELECTRODE CONDITIONING
AND MAINTENANCE
These bubbles can be removed by "shaking down" the electrode as you would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in HI 70300 storage solution for at least one hour.
For refillable electrodes**:
If the refill solution (electrolyte) is more than 2½ cm (1") below the fill hole, add HI 7082 3.5M KCl electrolyte solution for double junction or HI 7071 3.5M KCl+AgCl electrolyte solution for single junction electrodes.
For AmpHel® electrodes:
If the electrode does not respond to pH changes, the battery is run down and the electrode should be replaced.
TEST MEASUREMENT
Rinse the electrode tip with distilled water. Immerse the tip (bottom 4 cm / 1½") in the sample and stir gently
for approximately 30 seconds. For a faster response and to avoid cross contamination of the samples,
rinse the electrode tip with the solution to be tested, before taking your measurements.
STORAGE
To minimize clogging and assure a quick response time, the glass bulb and the junction should be kept moist and not allowed to dry out. This can be achieved by installing the electrode in such a way that it is constantly in a well filled with the sample (stream or tank).
When not in use, replace the solution in the protective cap with a few drops of HI 70300 storage solution.
Follow the Preparation Procedure above before taking measurements. NEVER STORE THE ELECTRODE WITH DISTILLED OR DEIONIZED WATER.
* Only available with refillable electrodes. For industrial applications, gel-filled
electrodes are preferable due to lesser maintenance requirements.
PREPARATION
Remove the protective cap. DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal with electrodes and they will disappear when rinsed
with water. During transport tiny bubbles of air may have formed inside the glass
bulb. The electrode cannot function properly under these conditions.
26
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable used for the connection to the controller must be intact and there must be no points of broken insulation on the cable or cracks on the electrode stem or bulb.
Connectors must be perfectly clean and dry. If any scratches or cracks are present, replace the electrode. Rinse off any salt deposits with water.
**For industrial applications, gel-filled electrodes are preferable due to lesser maintenance requirements.
AmpHel® is a registered Trademark of "Hanna Instruments"
27
For refillable electrodes**: Refill the electrode with fresh electrolyte (HI 7071 for single junction or
HI 7082 for double junction electrodes). Allow the electrode to stand
upright for 1 hour. Follow the Storage Procedure above.
CLEANING PROCEDURE
General Soak in HI 7061 general cleaning solution for
approximately 30 minutes.
Removal of films, dirt or deposits on the membrane/junction: Protein Soak in HI 7073 protein cleaning solution for 15
minutes.
Inorganic Soak in HI 7074 inorganic cleaning solution for
15 minutes.
Oil/ grease Rinse with HI 7077 Oil & Fat cleaning solution. IMPORTANT: After performing any of the cleaning procedures rinse
the electrode thoroughly with distilled water, drain and refill the reference chamber with fresh electrolyte (not necessary for gel-filled electrodes), and soak the electrode in HI 70300 storage solution for at least 1 hour before reinstalling it.
TROUBLESHOOTING
Evaluate your electrode performance based on the following.
Noise (Readings fluctuate up and down) could be due to:
- Clogged/Dirty Junction: Refer to the Cleaning Procedure above.
- Loss of shielding due to low electrolyte level (refillable electrodes only): refill with HI 7071 for single junction or HI 7082 for double junction electrodes.
Dry Membrane/Junction: Soak in HI 70300 storage solution for at least 1 hour. Check to make sure the installation is such as to create a well for the electrode bulb to constantly remain moist.
Drifting: Soak the electrode tip in warm HI 7082 solution for one hour and rinse tip with distilled water (if necessary, refill with fresh HI 7071 for single junction electrodes and HI 7082 for double junction electrodes).
Low Slope: Refer to the cleaning procedure above.
No Slope: - Check the electrode for cracks in glass stem or
bulb (replace the electrode if cracks are found).
- Make sure cable and connections are not dam­aged nor lying in a pool of water or solution.
Slow Response/Excessive Drift: Soak the tip in Hanna Solu­tion HI 7061 for 30 minutes, rinse thoroughly in distilled water and then follow the Cleaning Procedure above.
For ORP Electrodes: polish the metal tip with a lightly abrasive paper (paying attention not to scratch the surface) and wash thoroughly with water.
Note: with industrial applications, it is always recommended to keep
at least one spare electrode handy. When anomalies are not resolved with a simple maintenance, change the electrode (and recalibrate the controller) to see if the problem is alleviated.
**For industrial applications, gel-filled electrodes are preferable due to lesser maintenance requirements.
28
29
TAKING REDOX MEASUREMENTS
Redox measurements allow the quantification of the oxidizing or reducing power of a solution, and are commonly expressed in mV.
Oxidation may be defined as the process during which a molecule (or an ion) loses electrons and reduction as the process by which electrons are gained.
Oxidation is always coupled together with reduction so that as one element gets oxidized, the other is automatically reduced, therefore the term oxidation-reduction is frequently used.
Redox potentials are measured by an electrode capable of absorbing or releasing electrons without causing a chemical reaction with the elements with which it comes into contact.
The electrodes most usually available for this purpose have gold or platinum surfaces; gold possesses a higher resistance than platinum in conditions of strong oxidation such as cyanide, while platinum is preferred for the measurements of oxidizing solutions containing halides and for general use.
When a platinum electrode is immersed in an oxidizing solution a monomolecular layer of oxygen is developed on its surface. This layer does not prevent the electrode from functioning, but it increases the response time. The opposite effect is obtained when the platinum surface absorbs hydrogen in the presence of reducing mediums. This phenomenon is rough on the electrode.
To make correct redox measurements the following conditions must prevail:
– The surface of the electrode must be cleaned and smooth. – The surface of the electrode must undergo a pretreatment in
order to respond quickly.
Because the Pt/PtO system depends on the pH, the pretreatment of the electrode may be determined by the pH and the redox potential values of the solution to be measured.
As a general rule, if the ORP mV reading corresponding to the pH value of the solution is higher than the values in the table below, an oxidizing pretreatment is necessary; otherwise a reducing pretreat­ment is necessary: pH mV pH mV pH mV pH mV pH mV
0 990 1 920 2 860 3 800 4 740 5 680 6 640 7 580 8 520 9 460 10 400 11 340 12 280 13 220 14 160
Reducing pretreatment: immerse the electrode in HI 7091 solution for a few minutes.
Oxidizing pretreatment: immerse the electrode in HI 7092 solution for a few minutes.
If the pretreatment is not performed, the electrode will take signifi­cantly longer to respond.
As with pH electrodes, gel-filled redox electrodes are more suitable for industrial applications due to lesser maintenance requirements. How­ever, if working with refillable electrodes, the electrolyte level should not fall more than 2½ cm (1") below the fill hole and topped up if necessary. Use HI 7071 refill solution for single junction and HI 7082 for double junction electrodes.
In the event that measurements are performed with solutions contain­ing sulfides or proteins, the cleaning of the diaphragm of the reference electrode must be performed more often.
In order to have a correct functioning of the ORP electrode, immerse it into HI 7020 and measure the response; the obtained value should be within 200 and 275 mV.
After this functional test, it is suggested to wash the electrode thoroughly with water and proceed to the oxidizing or reducing pretreatment before taking measurements.
When not in use, the electrode tip should be kept moist and far from any type of mechanical stress which might cause damage. This can be achieved by installing the electrode in such a way that it is constantly in a well filled with the sample (stream or tank). The protective cap can also be filled with HI 70300 storage solution if the electrode is not being used at all.
Note: With industrial applications, it is always recommended to keep
at least one spare electrode handy. When anomalies are not resolved with a simple maintenance, change the electrode to see if the problem is alleviated.
30
31
CHEMICAL COMPATIBILITY GUIDE
PARTIAL LISTING OF CHEMICALS THAT CAN BE USED WITH BLACKSTONE PUMPS
(Rated for 45°C. For higher temperatures consult your dealer or nearest Hanna Customer Service Center)
Adipic Acid Alcohol Amyl Alcohol, Diacetone Alcohol, Isoproyl Alcohol, Methyl Aluminium, Ammonium Sulfate Aluminium Chloride Aluminium Sulfate Alums Ammonium Carbonate Ammonium Chloride Ammonium Fluoride Ammonium Hydroxide Ammonium Nitrate Ammonium Phosphate Ammonium Sulfate Aqua Ammonia Arsenic Acid Barium Carbonate Barium Chloride Barium Hydroxide Barium Sulfate Beer Beet Sugar Liquors Bismuth Carbonate Back Liquor Bleach Borax Boric Acid Bromic Acid Butyric Acid Calcium Bisulfite Calcium Carbonate Calcium Chlorate Calcium Chloride Calcium Hydroxide Calcium Hypochlorite Calcium Nitrate Calcium Sulfate Carbonic Acid Castor Oil
Caustic Soda Chloral Hydrate Chromic Acid 50% Citric Acid Copper Chloride Copper Cyanide Copper Nitrate Copper Sulfate Corn Oil Cottonseed Oil Cresylic Acid Crude Oil Dextrose Detergents (general) Diesel Fuel Dictyl Phthalate Disodium Phosphate Ethanol (1-95%) Ethylene Dichloride Ethylene Glycol Fatty Acids Ferric Chloride Ferric Nitrate Ferric Sulfate Ferrous Chloride Ferrous Sulfate Fluoboric Acid Fluosilicic Acid Formaldehyde Fruit Juice Pulp Fuel Oil Gallic Acid Gasoline, Refined Glucose Glycerine or Glycerol Glycolic Acid 30% Hexane Hydrazine Hydrobromic Acid 20% Hydrochloric Acid (Concentrated) Hydrochloric Acid (Diluted) Hydrofluoric Acid 60%
32
Hydrogen Sulfide Aqueous Solution Hypochlorous Acid Kerosene Lactic Acid Lard Oil Lauric Acid Lead Acetate Linoleic Acid Linseed Oil Lithium Salts Magnesium Carbonate Magnesium Chloride Magnesium Hydroxide Magnesium Nitrate Magnesium Oxide Magnesium Sulfate Maleic Acid Malic Acid Mercuric Chloride Methanol Methyl Sulfate Milk Mineral Oils Noptha Petroleum Nickel Chloride Nickel Sulfate Nitric Acid 50% Oils and Fats Oleic Acid Olive Oil Oxalic Acid Palmitric Acid Perchloric Acid 70% Perchloroethylene Petroleum Oils (sour) Phenol Phosphoric Acid Photographic Solutions Plating Solutions Potassium Carbonate Potassium Bromide Potassium Chlorate Potassium Chloride Potassium Cyanide Potassium Ferrocyanide Potassium Hydroxide Potassium Nitrate Potassium Permanganate 10% Potassium Phosphate Potassium Sulfate Propyl Alcohol Propylene Dichloride Sea Water
Silver Nitrate Silver Plating Solutions Soaps Sodium Acetate Sodium Bicarbonate Sodium Bisulfate Sodium Bisulfite Sodium Borate Sodium Chlorate Sodium Chloride Sodium Cyanide Sodium Fluoride Sodium Hexametaphosphate Sodium Hydroxide 50% Sodium Hypochlorite 18% Sodium Metaphosphate Sodium Nitrate Sodium Peroxide Sodium Phosphate Sodium Silicate Sodium Sulfate Sodium Sulfide Sodium Sulfite Sodium Thiosulfate Sour Crude Oil Stannic Chloride Stannous Chloride Stearic Acid Sulfur Sulfuric Acid Concentration Sulfurous Acid Tannic Acid Tanning Liquors Tartaric Acid Tetrachlorethane Tetraethyl Lead Tetralin Tin Salts Vegetable Oils Vinegar Water Acid, Mine Water, Fresh Water, Distilled Water, Salt Whiskey Wines Zinc Chloride Zinc Sulfate
33
ACCESSORIES
®
T
Y
SPARE PARTS
HI 721102 Discharge Valve (Glass Ball, Valve O-Ring, Hose Con-
nector)
HI 721101 Pump head, O-Ring, 6 screws and washers HI 721106 Pump head, Large PTFE Diaphragm, Aluminum
Piston and Aluminum Disk
TUBE NUT HEAD NIPPLE SPACER
CHECK
BALL
VALVE
SEAT
VITON O-RING
HI 721103 Suction Valve (Glass Ball, Valve O-Ring, Hose Connec-
tor)
CHECK
VITON® O-RING
BALL
VALVE
SEAT
HEAD NIPPLE TUBE NU
HI 721003 10 x Glass Balls and 10 x Valve O-Rings HI 721004 Injection Valve Assembly
CHECK
PVDF
SPRING
BALL
VALVE ASSEMBLYINJECTION NIPPLE KYNAR®
HI 721005 Foot Valve Assembly
FILTER FILTER HOLDER
VALVE ASSEMBL
HI 721006 PVDF Springs, 4 pcs HI 720032 LDPE Hose - 100 m (330') HI 721008 Ceramic Weights, 4 pcs
PUMP HEAD DIAPHRAGM
pH CALIBRATION SOLUTIONS
HI 7004M pH 4.01 buffer solution, 230 mL bottle HI 7004L pH 4.01 buffer solution, 500 mL bottle HI 7006M pH 6.86 buffer solution, 230 mL bottle HI 7006L pH 6.86 buffer solution, 500 mL bottle HI 7007M pH 7.01 buffer solution, 230 mL bottle HI 7007L pH 7.01 buffer solution, 500 mL bottle HI 7009M pH 9.18 buffer solution, 230 mL bottle HI 7009L pH 9.18 buffer solution, 500 mL bottle HI 7010M pH 10.01 buffer solution, 230 mL bottle HI 7010L pH 10.01 buffer solution, 500 mL bottle
ORP SOLUTIONS
HI 7020M 200-275mV test solution, 230 mL bottle HI 7020L 200-275mV test solution, 500 mL bottle HI 7091M Reducing pretreatment solution, 230 mL HI 7091L Reducing pretreatment solution, 500 mL HI 7092M Oxidizing pretreatment solution, 230 mL HI 7092L Oxidizing pretreatment solution, 500 mL
MAINTENANCE SOLUTIONS
HI 70300M Storage solution, 230 mL botte HI 70300L Storage solution, 500 mL bottle HI 7061M General cleaning solution, 230 mL botte HI 7061L General cleaning solution, 500 mL bottle HI 7073M Protein cleaning solution, 230 mL botte HI 7073L Protein cleaning solution, 500 mL bottle HI 7074M Inorganic cleaning solution, 230 mL botte HI 7074L Inorganic cleaning solution, 500 mL bottle HI 7077M Oil & Fat cleaning solution, 230 mL bottle
34
35
HI 7077L Oil & Fat cleaning solution, 500 mL bottle HI 7071 3.5M KCl+AgCl electrolyte solution, 4x50 mL, for
single junction electrodes
HI 7072 1M KNO3 electrolyte solution, 4x50 mL HI 7082 3.5M KCl electrolyte solution, 4x50 mL, for double
junction electrodes
RECOMMENDED pH ELECTRODES
All electrodes are gel-filled and with ceramic junction unless otherwise indicated.
HI 1090T Screw connector, external PG13.5 thread, double junc-
tion, glass-body
HI 1110S Screw connector, single junction, glass-body HI 1130B/3 BNC connector, 3 m (9.9') cable, single junction, glass-
body with external thread
HI 1110T Screw connector, external PG13.5 thread, double junc-
tion, glass-body with ground glass junction
HI 1114S Screw connector, double junction plastic-body HI 1134B/3 BNC connector, 3 m (9.9') cable, double junction,
plastic body with external thread
HI 1115S Screw connector, single junction, refillable with side-
arm, glass-body
HI 1135B/3 BNC connector, 3 m (9.9') cable, single junction,
refillable with side-arm, glass-body
HI 1210T Screw connector, external PG13.5 thread, double junc-
tion, plastic body, cloth junction
HI 1910B BNC connector, 1 m (3.3') cable, double junction,
plastic body with built-in amplifier and external thread
HI 1912B BNC connector, 1 m (3.3') cable, double junction,
plastic body with built-in amplifier and external thread
HI 1912B/5 BNC connector, 5 m (16.5') cable, double junction,
plastic body with built-in amplifier and external thread
HI 2114B/5 BNC connector, 5 m (16.5') cable, double junction,
plastic body with external thread and cloth junction
HI 2910B/5 BNC connector, 5 m (16.5') cable, double junction,
plastic body with built-in amplifier and cloth junction
PLATINUM ORP ELECTRODES
HI 2930B/5 BNC connector, 5 m (16.5') cable, double junction, Pt,
plastic body with built-in amplifier, external thread and cloth junction
HI 3110S Screw connector, single junction, Pt, glass-body
HI 3130B/3 BNC connector, 3 m (9.9') cable, Pt, glass-body with
external thread
HI 3110T Screw connector, external PG13.5 thread, double
junction, Pt, glass-body
HI 3115S Screw connector, single junction, Pt, refillable with
side-arm, glass-body
HI 3135B/3 BNC connector, 3 m (9.9') cable, single junction, Pt,
refillable with side-arm, glass-body
HI 3210T Screw connector, external PG13.5 thread, double junc-
tion, Pt, plastic body
HI 3410S Screw connector, double junction, Pt, plastic body HI 3430B/3 BNC connector, 3 m (9.9') cable, double junction, Pt,
plastic body with external thread
HI 3932B/5 BNC connector, 5 m (16.5') cable, double junction, Pt,
plastic body with built-in amplifier and external thread
GOLD ORP ELECTRODES
HI 4110S Screw connector, single junction, Au, glass-body HI 4130B/3 BNC connector, 3 m (9.9') cable, single junction, Au,
glass-body with external thread
HI 4932B/5 BNC connector, 5 m (16.5') cable, double junction,
Au, plastic body with built-in amplifier and external thread
OTHER ACCESSORIES
ChecktempC Electronic thermometer (range -50.0 to 150.0°C) ChecktempF Electronic thermometer (range -58.0 to 302°F) HI 8614 pH transmitter HI 8614L pH transmitter with LCD HI 8615 ORP transmitter HI 8615L ORP transmitter with LCD BL PUMPS Dosing pumps with flow rate from 1.5 to 20 LPH HI 7871 & HI 7873 Level controllers HI 6050 & HI 6051 Submersible electrode holders HI 6054 & HI 6057 Electrode holders for in-line applications HI 778P Screened coaxial cable with screw connectors HI 8427 pH/ORP electrode simulator HI 931001 pH/ORP electrode simulator with LCD
36
37
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environ­ment in which they are used.
Operation of these instruments in residential areas could cause unacceptable interfer­ence to radio and TV equipment.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
Unplug the instruments from power supply before replacing the fuse or making any electrical connections.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
SALES AND TECHNICAL SERVICE CONTACTS
Australia:
Tel. (03) 9769.0666 • Fax (03) 9769.0699
China:
Tel. (10) 88570068 • Fax (10) 88570060
Egypt:
Tel. & Fax (02) 2758.683
Germany:
Tel. (07851) 9129-0 • Fax (07851) 9129-99
Greece:
Tel. (210) 823.5192 • Fax (210) 884.0210
Indonesia:
Tel. (21) 4584.2941 • Fax (21) 4584.2942
Japan:
Tel. (03) 3258.9565 • Fax (03) 3258.9567
Korea:
Tel. (02) 2278.5147 • Fax (02) 2264.1729
Malaysia:
Tel. (603) 5638.9940 • Fax (603) 5638.9829
Singapore:
Tel. 6296.7118 • Fax 6291.6906
South Africa:
Tel. (011) 615.6076 • Fax (011) 615.8582
Taiwan:
Tel. 886.2.2739.3014 • Fax 886.2.2739.2983
Thailand:
Tel. 66.2619.0708 • Fax 66.2619.0061
United Kingdom:
Tel. (01525) 850.855 • Fax (01525) 853.668
USA:
Tel. (401) 765.7500 • Fax (401) 765.7575
For e-mail contacts and complete list of Sales and Technical offices, please see www.hannainst.com
MAN7916R3
07/05
Loading...