Follow precautions and all the safety requirements described on this user manual
to prevent any damage and failure of device and loss of lives.
1. The device should be installed on at surface to maintain level.
2. Check the voltage to be used, before connecting the device to the power source.
3. Only use rotors, parts, and accessories provided by Hanil Scientic Inc.
Hanil Scientic Inc. is not responsible for damages of the device and accidents caused by
using parts and accessories not recommended.
4. Do not exceed the maximum rated speed of the rotor or buckets in use.
5. Make sure to prepare necessary safety measures before using samples that are toxic or
radioactive samples or pathogenic or samples or infectious blood.
6. Substances that may generate volatile or explosive vapor can not be deviced.
7. The balancing work of samples should done in advance before operation.
8. To ensure safe use of the device, do not expose the device to strong acids, strong bases,
cesium, salt, or alkaline detergents.
9. If the device is contaminated by toxic or radioactive samples or infectious blood
samples, remove contaminants completely and take needful actions such as ventilation
or isolation of device.
10. Before operation, rotor and chamber should be dry.
11. Do not attempt to slow or stop the spinning rotor by hand.
12. Only device with rotor and rotor door rmly tightened.
13. Do not block vents.
14. When serving the device, be sure to remove contaminants in advance.
15. Please contact the place of purchase or Hanil Scientic Inc. for product repairs.
16. According to IEC61010-2-020 maintain a 30cm “clearance envelope” around the
device while the rotor is spinning.
17. Turn the power switch o after using the device.
18. Unplug the power plug before cleaning or left unused for a long period of time.
19. Don't lean against the device
4/21
www.ihanil.com
Page 5
1. General Considerations
1.2 Transport & Storage
- The device and the accessories must only be stored in dry rooms.
- Only lift and transport the device with sucient number of helpers.
-Storage--Transport-
Ambient temperature 2℃~40℃
Maximum relative humidity 30%~85%
Air pressure 500~1060hpa
1.3 Safety label on the Device
1.4 Electric safety
Ambient temperature -10℃~40℃
Maximum relative humidity 10%~90%
Air pressure 500~1060hpa
Insert tube symmetrically.
Firmly tighten the rotor door.
Watch your gers when close the door.
1. It is recommended that switchgear or circuit breakers and overcurrent protection devices
be installed near the device.
2. Use a power cord only provided with the device.
3. Use sockets with a protective earth conductor and suitable power cord.
4. Do not use an extension cord.
5. Do not place anything on the powercable.
6. If you have the following emergencies, shut o the power supply and unplug the
powercord from outlet and contact your place of purchase.
- Unusual noises or smell from the equipment.
- Damage or wear of a power cord.
- Breakdown of circuit breaker, fuse or safety device.
2. Press the Left or Right button until the RPM/RCF display(rpm) ashes.
▶ RPM setting mode: RPM(rpm) display ashes.
▶ RCF setting mode: RCF(xg) display ashes.
3. Press Up or Down button until the required RPM/RCF is displayed.
4. Press the Setup/Enter button.
4.3 Setting Run Time
1. Press the Setup/Enter button.
2. Press the Left or Right button until the Time display(0' 00'') ashes.
▶ Setting run time in hours: 00' ashes
▶ Setting run time in minutes: 00'' ashes
3. Press Up or Down button until the required run time is displayed.
4. Press the Setup/Enter button.
14/21
www.ihanil.com
Page 15
4. Operation
4.4 Setting Acceleration/Deceleration
1. Press the Setup/Enter button.
2. Press the Left or Right button until the Accel/Decel display(0) ashes.
▶ Accel rate setting mode: Accel display ashes(0).
▶ Decel rate setting mode: Decel display ashes(0).
3. Press the Setup/Enter button.
4.5 Start/Stop a centrifugation
[Start]
1. Check all parameters are correct.
2. Press the Start/Stop/Pulse button.
[Stop]
1. Press the Start/Stop/Pulse button to stop centrifugation.
▶ The centrifugation stops when the set run time has been reached.
This funcion is used for short run centrifugation.
1. Set the desired run speed. [ Refer to. 4.2 Setting RPM/RCF]
2. Press the Start/Stop/Pulse button more than 3sec.
▶ The rotor accelerates at maximum rate to the set speed and stops automatically.
4.7 Saving/Calling up a Program
[Saving]
1. Set the RPM(or RCF) / Run time / Accel, Decel rate
2. Press the Left of Right button until Prog display ashes(0).
▶ You can save 10(0~9) programs.
3. Press Up or Down button until the required program number is displayed.
4. Press the Enter button.
[Calling up]
1. Press the Setup/Enter button.
2. Press the Left of Right button until Prog display ashes(0).
3. Press Up or Down button until the required program number is displayed.
4. Press the Enter button.
16/21
www.ihanil.com
Page 17
5. Maintenance
5.1 Care Instructions
1. Regularly inspect the rotor chamber for check the motor shaft is normal.
2. Regularly check the rotor and accessories to be sure there is no damage.
3. Rotate the motor shaft with your hand to make sure it turns smoothly.
4. To prevent the rotor from sticking, lubicate the rotor drive hole which contacts motor
shaft regularly.
5.2 Cleaning
• Before cleaning the centrifuge, be sure to switch o the device and
disconnect the power cord.
[Oustside of the device]
1 Clean the outside of deivice with a soft and dry cloth.
▶ If the device is contaminated, use a mild cleaning uid to clean.
2. Do not use aggressive chemicals on the device such as alchol, benzene, acetone or
phenol.
3. Make sure do not scratch the surface of deivice when cleaning it.
▶ Do not use a metal sponge.
▶ If the device is rusted, remove it with a mild detergent and wipe it with a dry cloth.
[Chamber]
1. If the rotor chamber is not dry, wipe moisture from the chamber with a dry cloth.
2. If the rotor chamber is dirty, clean it with a mild uid to clean.
[Rotor]
1. After centrifugation, remove the rotor from the chamber and keep it upside down.
2. If any sample is spilt inside the rotor, clean it and dry well.
[Disposal]
1. In case of product is to be disposed of, the local wastes laws and regulations are to be
observed.
If the device door cannot be opened, you can activte the emergency open manually.
1. Wait for rotor to stop before activating the emergency open.
2. Insert the T-wrench into the emergency open hole and turn it clockwise
until the door is opened.
Emergency open hole
20/21
www.ihanil.com
Page 21
7. Rotor
7.1 Rotor
RotorTube
Capacity
A0.2M-32
(PCR Strip Rotor)
A2.0M-18
A2.0M-24
A2.0MP-24
A5.0M-12
Hole angle : ∠45°
Max. Capacity : 32 x 0.2 mL
Size (ø x H) : ø172 x 38.5 mm
Hole angle : ∠45°
Max. Capacity : 18 x 1.5/2.0 mL
Size (ø x H) : ø147 x 38.5 mm
Max. height for tube fit : 53.5 mm
Incl. a lid
Hole angle : ∠45°
Max. Capacity : 24 x 1.5/2.0 mL
Size (ø x H) : ø172 x 38.5 mm
Max. height for tube t : 56 mm
Incl. a lid
A2.0MP-24 incl. plastic lid
Hole angle : ∠50°
Max. Capacity : 5 x 5.0 mL
Size (ø x H) : ø167.2 x 57.4 mm
Incl. a lid