Hanil Component wR Quick Manual

Page 1
Quick Guide
Component wR
Cat. No. : CP-wR
Display & Control Panel
Display Screen
Status of RPM / RCF / Time / Temperature / Rotor ID / ACC / DEC / Program
Setting Speed
Press once : Setting RPM Press twice : Setting RCF
Setting ACC / DEC Rate
ACC (0 ~ 9 steps, Max.9) DEC (0 ~ 9 steps, Natural DEC.0)
Setting Time
Up < 100 hr Continuous run (00:00)
xg
Setting Temperature
Setting temperature ( -20℃ ~ 40℃) Setting temp. limit
Lid Open
Setting the Rotor ID manually
Enter the rotor ID. manually, when rotor id is not identified automatically
PULSE
2 31
5 64
8 97
±
0
Pulse
After reaching the set speed, the centrifugation is decelated and stopped
Max. RPM : 4,500 rpm Max. RCF : 5,096 xg
View Settings
Verifying the setting parameters during centrifugation
VIEW
SETTINGS
Program Save / Call
Saving a program (up to 100)
Calling up a program
Setting Run Parameters
(RPM / RCF / Time / Temperature / ACC / DEC)
Press a parameter
to be changed
Enter the desired values
on the keypad
Pulse
Enter the desired speed
(See ‘Setting Run Parameters’)
Pulse button
Enter button
Press the
Press the
Saving / Calling Up a Program
Press the
Save or call button
Enter the desired
program number on
the keypad
Setting Temp. Limit
Press the
Temp. limit button
Temperature range that stops operation when the temperature
is lower or higher than the set temperature.
Enter the desired
temperature range
on the keypad
Press the
Enter button
Press the
Enter button
Serial No. Installation Date
Page 2
Quick Guide
Cautions for Use
Component wR
Cat. No. : CP-wR
Before Operation After Operation
Check the Accessories for Abnormalities
Check the rotor, buckets and tubes are free of scratch and crack before operation.
Load Tubes Symmetrically
Load tubes symmetrically into the rotor with identical amount.
Maintenance
30 cm
Balancing the Centrifuge and Rotor
After relocating the device or loading the rotor, balance them by using a leveler. If necessary, level the device by adjusting the ‘Leveling foot’ on the bottom of the device.
Check the Lid
Place the lid on the wind shield rotor before centrifugation.
30 cm
30 cm
Keep the Chamber Dry
Wipe up the moisture in the chamber with a dry cloth after use. Keep the lid open when not in use.
Cleaning
Clean the chamber and rotor by wiping with a cloth soaked in mild detergent and dry completely.
Emergency Lid Open
If the centrifuge lid cannot be opened, insert the provided T-wrench into the emergency release hole on the right side of the device and press it. (Before activat­ing emergency open, switch off the device and wait for the rotor to stop.)
Allow a Safety Clearance of 30 cm
Allow a safety clearance of 30 cm around the centrifuge during operation and storage.
Care of Swing-Out Rotor
Spread the grease lightly on the bucket where it contacts the rotor and the pivot.
Remove Dust Empty the Condensation
Remove any dust from the ventilation slits on the right side of the device.
Water
Empty the condensation water regularly by removing the drain cap on the left side of the device.
Electrical Safety
3.0KVA 220~240V 50/60Hz
Check the Line Voltage Grounded Outlet Power Connection Switch Off the Device
Connect the device to voltage sources, which correspond to the electrical requirements on the label attached to the device.
Rotors
WS750-6 6 x 750 mL / Max. 4.5k
Check the wall outlet is earth­grounded, surge-protected and can supply enough power.
Connect the power cable to the distribution box, or install the plug and socket that are packed with the device.
Switch off the device after use. Unplug the power plug before cleaning or unused for a long period of time.
Order & Inquiry l +82-2-3472-0727 l [email protected] l www.ihanil.com Service l +82-2-3452-8966 l [email protected]
Proudly made by Jiyoung Shim, Hyewon Jeong
Rev.1
Loading...