Hamilton Conducell UPW Arc Operating Instructions Manual

Conducell UPW Arc Sensors
Operating Instructions
TABLE OF CONTENTS
Table of Contents
TABLE OF CONTENTS
1.1 Intended Use .............................................................. 4
1.2 About this Operating Instruction ................................. 5
3.1 General Precautions ................................................... 5
3.2 Operating of Conducell UPW Arc Sensor ................... 6
3.3 Electrical Safety Precautions ...................................... 6
3.4 Earthing ...................................................................... 7
3.5 Chemical, Radioactive or Biological
Hazard Precautions .................................................... 7
4.1 Hardware Description ................................................. 8
4.2 Conducell UPW Arc with Micro-Transmitter inside ...... 8
5.1 Unpacking .................................................................. 9
5.2 Configuring the Conducell UPW Arc with ArcAir ......... 9
5.2.1 Installing ArcAir Basic on the Computer ......... 10
5.2.2 Connecting a Conducell UPW Arc
Sensor to ArcAir ............................................. 10
5.2.3 Create User Accounts .................................... 10
5.2.4 Configuring the Conducell UPW Arc
Sensor Parameters ..........................................11
5.2.5 Configuring the calibration settings .................11
5.2.6 Configuring the temperature settings
of SIP / CIP process ....................................... 12
5.2.7 Configuring the analog interface for
your process control system .......................... 12
5.2.8 Defining a measuring point name for
identification of the process ............................ 13
5.3 Install Conducell UPW Arc sensor in
your Measuring Loop ................................................ 13
5.3.1 Mechanical Process Connection .................... 13
5.3.2 VP 8 Pin Designation ...................................... 13
5.3.3 Controlling 4–20 mA current interface
signals by pulse-width modulation.................. 14
5.3.4 Electrical Connection for Analog
4–20 mA Connection ..................................... 15
5.3.5 Electrical connection for the
digital RS485 interface ................................... 16
5.4 Interface for the signals according to USP <645> ......17
6 Operation .............................................................. 20
6.1 Typical Installation e.g. Triclamp Version ................... 20
7 Maintenance ......................................................... 22
7.1 Verify Sensor Status ................................................. 22
7.2 Calibration ................................................................ 23
7.2.1 Automatic Standard Calibration
with ArcAir ...................................................... 23
7.2.2 One Point Calibration in Conductivity
Standard Solutions ......................................... 23
7.2.3 Product calibration ......................................... 24
7.2.4 Verification according to USP <645> .............. 25
8 Troubleshooting ................................................... 26
8.1 Sensor Self-Diagnostic ............................................. 26
8.1.1 War n in g s ........................................................ 26
8.1. 2 Er r o rs ............................................................. 27
8.2 Getting Technical Support ........................................ 27
8.3 Returning Conducell UPW Arc sensor for Repair ...... 27
9 Disposal ................................................................ 27
10 Ordering Information ........................................... 28
10.1 Conducell UPW Arc ................................................ 28
10.2 Parts and Accessories ............................................ 28
10.3 Services ................................................................. 32
Hamilton Warranty
Please refer to the General Terms of Sales (GTS).
Important note
Copyright © 2019 Hamilton Bonaduz AG, Bonaduz Switzerland. All rights reserved. The reproduction of any part of this document in any form is forbidden without the express written agreement of Hamilton Bonaduz AG.
Contents of this manual can be modified without previous announcement. Technical modifications reserved. Greatest possible care was used on the correctness of the information in this manual. If errors should be discovered nevertheless, Hamilton Bonaduz AG is pleased to be in­formed about it. Regardless of this, Hamilton Bonaduz AG cannot assume liability for any errors in this manual or for their consequences.
3Operating Instructions Conducell UPW Arc Sensors
GENERAL INFORMATION

SAFETY PRECAUTIONS AND HAZARDS

LIABILITY

1 General Information

This manual refers to the Conducell UPW Arc sensor from Ham­ilton Bonaduz AG. These sensors are compatible with all other components of the Hamilton Arc System, including a complete family of intelligent sensors for pH, dissolved oxygen and conduc­tivity measurements in a process control.
Hamilton Arc sensors are quality products manufactured ac­cording to the very latest research findings. Follow the instruc­tions given here to be sure of optimal safety and durability.
ATTENTION! These instructions must be read, understood and followed by all staff using Conducell UPW Arc sensors. Hamilton can assume no respon­sibility for damage and operational disruption arising from failure to observe these instructions.

1.1 Intended Use

Conducell UPW Arc sensors were developed for the measure­ment of conductivity in pure and ultrapure water.
Other applications and uses are not recommended. Hamilton will not take liability if damage to persons or equipment arises by improper use. For more details about liability, please refer to the section entitled «Liability».
Conducell UPW Arc sensors provide both standard analog (4 – 20 mA) and digital Modbus interfaces. These are built into each sensor and are supported directly from the sensor. They do not require any additional equipment such as amplifier or tr a n s m i t te r.
The Conducell UPW Arc sensor consists of a conductivity cell part with a VP 11 head and an Arc Module with a VP 8 head. Conductivity cell and Arc Module will be shipped as one unit with identical serial numbers. Hereafter the unit will be called sensor.

1.2 About this Operating Instruction

These Operating Instructions are designed to support the integra­tion, operation and qualification of the Conducell UPW Arc sen­sors. To achieve this, it describes the features of Conducell UPW Arc and its integration in Process Control Systems (PCS). Both the hardware and the communication between the Conducell UPW Arc sensor and Process Control Systems are detailed in this manual. After reading this manual the user should be capable of installing and operating Conducell UPW Arc sensors.
ATTENTION! Essential information for avoiding
personal injury or damage to equipment.
NOTE: Important instructions or interesting information.

2 Liability

The liability of Hamilton Bonaduz AG is detailed in the document «General Terms and Conditions of Sale and Delivery».
Hamilton is expressly not liable for direct or indirect losses arising from use of the sensors. It must in particular be insured in this conjunction that malfunctions can occur on account of the inher­ently limited useful life of sensors contingent upon their relevant applications. The user is responsible for the calibration, main­tenance and regular replacement of the sensors. In the case of critical sensor applications, Hamilton recommends using back-up measuring points in order to avoid consequential damages. The user is responsible for taking suitable precautions in the event of a sensor failure.
3 Safety Precautions
Additionally, for warnings and alarms according to USP <645>, two digital outputs are integrated into the sensor.
ATTENTION! The Conducell UPW Arc sensor has a built-in temperature sensor (NTC 22 kOhm). This temperature sensor is to be used only for monitoring the sensor conditions, not for controlling the process temperature.
ATTENTION! Conducell UPW Arc is not intended for hazardous atmospheres.
ATTENTION! The measurement values transmitted over wireless communication are not intended to be used for process control.
and Hazards
ATTENTION! Read the following safety instruc­tions carefully before installing and operating the Conducell UPW Arc sensor.

3.1 General Precautions

For safe and correct use of Conducell UPW Arc sensor, it is es­sential that both operating and service personnel follow generally accepted safety procedures as well as the safety instructions given in this document, the Conducell UPW Arc sensor operation instruction manual. The specification given in the «Specifica­tion Sheet» as regards temperature, pressure etc. may under no circumstances be exceeded. Inappropriate use or misuse can be dangerous.
5Operating Instructions Conducell UPW Arc Sensors
SAFETY PRECAUTIONS AND HAZARDS PRODUCT DESCRIPTION
The lifetime of the Conducell UPW Arc sensor highly depends on the specific conditions of the application. Temperature, pressure, chemicals used may accelerate the ageing of the sensor. See chapter 7 for replacement conditions.
Cleaning, assembly and maintenance should be performed by personnel trained in such work. Before removing the sensor from the measuring setup, always make sure that no process medium can be accidentally spilled. When removing and cleaning the sensor, it is recommended to wear safety goggles and protective gloves. The sensor can not be repaired by the operator and has to be sent back to Hamilton for inspection.
Necessary precautions should be taken when transporting the sensors. For repair or shipment the sensor should be sent back in the original reusable packaging box. Every sensor sent back for repair must be decontaminated.
If the conditions described in these operating instructions manual are not adhered to or if there is any inappropriate interference with the equipment, all of our manufacturer’s warranties become obsolete.

3.2 Operating of Conducell UPW Arc Sensor

Conducell UPW Arc sensors must be used for their intended ap­plications and in optimum safety and operational conditions. The specifications (such as temperature or pressure) defined in the section entitled ‘Technical Specification’ must not be exceeded under any circumstances. Potential hazards can exist if the sen­sor is not operated correctly or appropriately.
Assembly and maintenance must be performed only by trained personnel.
Make sure that the PG13,5 thread and the O-ring are not dam­aged when screwing the sensor into the process (PG 13,5 version only). O-rings are consumable parts which must be exchanged regularly (at least once per year). Even when all required safety measures have been complied with, potential risks still exist with respect to leaks or mechanical damage to the armature. Wher­ever there are seals or screws, gases or liquids can leak out undetected.
Before removing the sensor from its measurement setup, always make sure that no process medium can be accidentally spilled. The built-in temperature sensor can only be used for monitoring the sensor conditions, not for controlling the process temperature.
Keep the Conducell UPW Arc sensor away from other equipment which emits electromagnetic radio frequency fields, and minimize static electricity in the immediate environment of the measuring parts. Carefully follow all the instructions in chapter 5.3 to avoid electrical damage to the sensor. The contacts must be clean and dry before sensor is connected to the cable.
ATTENTION! Switch off the power supply and un­plug the connector before dismounting the Conducell UPW Arc sensor.
ATTENTION! If the power supply (230VAC/24VDC) is switched off or disconnected the reading on the PCS is wrong.

3.4 Earthing

The sensor has to be mounted at the mounting location with an electrostatically conductivity of < 1MOhm. It is recommended to assign the VP 8 cable shield to earth (PE potential equalization) especially in electromagnetically noisy environments. This signifi­cantly improves noise immunity and signal quality.

3.5 Chemical, Radioactive or Biological Hazard Precautions

Selection of the appropriate safety level and implementation of the required safety measures for working with Conducell UPW Arc sensor is the sole responsibility of the user.
If working with hazardous liquids observe and carry out the maintenance procedures, paying particular attention to cleaning and decontamination. If Conducell UPW Arc sensor becomes contaminated with biohazardous, radioactive or chemical mate­rial, it should be cleaned. Failure to observe and carry out the maintenance procedures may impair the reliability and correct function of the measuring module.
ATTENTION! When unscrewing the PG13,5 thread connection never turn the sensor at the connector head.

3.3 Electrical Safety Precautions

Do not connect the sensor to a power source of any voltage be­yond the range stated in the power rating Technical Specifications (see www.hamiltoncompany.com).
Always use Hamilton VP 8 cables for safe connection. Cables are available in a broad range of lengths (Chapter 11). Make sure the cable is intact and properly plugged to avoid any short circuit.

4 Product Description

Conducell UPW Arc sensors provide both standard analog (4 – 20 mA) and digital Modbus interfaces. These are built into each sensor and are supported directly from the sensor. They do not require any additional equipment such as amplifier or transmit­ter. The Conducell UPW Arc sensor consists of a conductivity cell part with a VP 11 head and an Arc Module with a VP 8 head. Conductivity cell and Arc Module will be shipped as one unit with identical serial numbers. Hereafter the unit will be called sensor.
The sensor can monitor the water quality according to USP <645> criteria. Additionally, for warnings and alarms according to USP <645>, two digital outputs are integrated into the sensor. For details, see the chapter «Interface for the signals according to USP <645>.» Full details, including serial number and most
7Operating Instructions Conducell UPW Arc Sensors
PRODUCT DESCRIPTION INSTALLATION
important specifications, can be found on the certificate provided with each sensor. A temperature compensation in the range be­tween 0 and 10% per K is possible and can be set manually.
The Conducell UPW sensor has been carefully tested and is ready for use.

4.1 Hardware Description

The Conducell UPW Arc sensor consists of a sensor head with integrated electronic and a sensor tip in contact with the mea­sured medium. All materials in contact with the solution meet the FDA requirement.
1
9
analog 4–20 mA. The micro transmitter located in the sensor head stores all relevant sensor data, including calibration and diagnostic information, simplifying calibration and maintenance.

5 Installation

5.1 Unpacking

1) Unpack carefully the Conducell UPW Arc sensor. Enclosed you will find the sensor, the Declaration of Quality, the sensor Instruction Manual, and the material Certificate.
2) Inspect the sensor for shipping damages or missing parts.
Conductivity Sensor
8
4
5
6
Figure 1: Components of Conducell UPW Arc exemplarily shown with a TriClamp process connection.
1 Socket head with VP 8 conne ctor
2 Arc logo, serial and reference number
3 TriClamp 1.5”
4 Insulation: PEEK
5 Inner electrode
6 Outer electrode
7 VP 11 connection
8 Conductivity cell
9 Arc Module
7
3
2
Figure 2: Conducell UPW Arc sensor delivery package.
5.2 Configuring the Conducell UPW Arc
with ArcAir
Arc sensors require application specific configuration. Following parts are required to configure and calibrate Arc sensors:
• Arc View Mobile or ArcAir computer Software Solution
• External Power supply with Arc USB Power Cable (Ref 243490-xx)
To configure and set up the Arc sensors at least ArcAir Basic is required. Below in this table you will find the different ArcAir licenses and its functionality:
ArcAir Read Calibrate Configure Documentation
Free
Basic
Advanced
– –

4.2 Conducell UPW Arc with Micro-Transmitter inside

With the micro-transmitter integrated,Conducell UPW Arc sensors offer fully compensated signal directly to the process control system. Communication protocols include standard
Arc Wire less
Conver ter BT
ArcAir
Figure 3: Conducell UPW configuration with ArcAir.
Arc Wi 1G/2G
Adapter BT
Conducell UPW
Arc USB Power Cable
9Operating Instructions Conducell UPW Arc Sensors
INSTALLATIONINSTALLATION

5.2.1 Installing ArcAir Basic on the Computer

1) Download the Zip file «ArcAir» from the Hamilton webpage www.hamiltoncompany.com (search for ArcAir).
2) Unpack the ZIP-File.
3) Do not plug in the Wireless Converter before the installation of ArcAir is completed.
4) Install «ArcAir» by double clicking «ArcAir.exe» and follow the instructions on the screen.
5.2.2 Connecting a Conducell UPW Arc
Sensor to ArcAir
1) Connect a sensors with the power supply, e.g. Arc USB Power Cable Ref 243490-xx
2) Switch on the mobile’s Bluetooth connection or connect a Wireless Converter BT to USB Port of your computer (only for wireless connection
3) The ArcAir application recognizes and displays the connected sensors automatically
ATTENTION! For automatic sensor login a unique and global Operator Level S password for all intelli­gent sensors is required. Please make sure you have added the same Operator Level S Password for all Arc sensors in the ArcAir application under Backstage/Settings/Operator Level S Password.

5.2.3 Create User Accounts

5.2.4 Configuring the Conducell UPW Arc Sensor Parameters
1) Start the ArcAir application
2) Select the desired sensor
3) Open the drawer «Settings» (make sure you have the «Sensor Settings» user right)
4) Configure the sensor
A description of the available settings is given below:
Parameter Default Name Description Value Configuration Location
COND These are the μS/cm Required Unit measurement Values physical units: μS/cm, mS/c m
T unit These are the °C Required Measurement/ temperature Values physical units °C, °F, K
T comp. Temperature 25°C Recommended Measurement/ Temp compensation default Values value value can be parameter set between 20°C and 25°C
T comp. Temperature 0%/K Required Measurement/ factor compensation Values value factor can be set between 0 %/K to 10 %/ K
Moving The sensor 2 average uses a moving default Values average over parameter the measuring points
Recommended
Measurement
Measurement/
/
1) Start ArcAir application on computer
2) Click on «Backstage» left upper corner
3) Select «User Management»
4) Click the «Add» Button for opening the user editor
5) Type in the user details and password
6) Select the specific rights for the user
ATTENTION! First user is the administrator and all
user rights are assigned as default.
NOTE: Initial operation of ArcAir is in the laboratory mode as long as no user account is created. Laboratory mode does not require a login password and enable all features in the installed license version.
5.2.5 Configuring the calibration settings
Parameter Default Name Description Value Configuration Location
Drift stability Higher drif t will 1%/Min Recommend Calibration/ interrupt the default Calibration calibration parameter Settings process. Warning comes up «drif t oxygen»
Higher drift will 0.5 K/min Recommend Calibration/ interrupt the default Calibration calibration parameter Settings process. Warning comes up «drift temperature»
11Operating Instructions Conducell UPW Arc Sensors