Contents of this manual can be modified without previous
announcement. Technical modifications reserved. Greatest
possible care was used on the correctness of the
information in this manual. If errors should be discovered
nevertheless, Hamilton Bonaduz AG is pleased to be informed about it. Regardless of this, Hamilton Bonaduz AG
cannot assume liability for any errors in this manual or for
their consequences.
3Operating Instructions Conducell UPW Arc Sensors
GENERAL INFORMATION
SAFETY PRECAUTIONS AND HAZARDS
LIABILITY
1 General Information
This manual refers to the Conducell UPW Arc sensor from Hamilton Bonaduz AG. These sensors are compatible with all other
components of the Hamilton Arc System, including a complete
family of intelligent sensors for pH, dissolved oxygen and conductivity measurements in a process control.
Hamilton Arc sensors are quality products manufactured according to the very latest research findings. Follow the instructions given here to be sure of optimal safety and durability.
ATTENTION! These instructions must be read,
understood and followed by all staff using Conducell
UPW Arc sensors. Hamilton can assume no responsibility for damage and operational disruption arising
from failure to observe these instructions.
1.1 Intended Use
Conducell UPW Arc sensors were developed for the measurement of conductivity in pure and ultrapure water.
Other applications and uses are not recommended. Hamilton
will not take liability if damage to persons or equipment arises by
improper use. For more details about liability, please refer to the
section entitled «Liability».
Conducell UPW Arc sensors provide both standard analog
(4 – 20 mA) and digital Modbus interfaces. These are built into
each sensor and are supported directly from the sensor. They
do not require any additional equipment such as amplifier or
tr a n s m i t te r.
The Conducell UPW Arc sensor consists of a conductivity cell
part with a VP 11 head and an Arc Module with a VP 8 head.
Conductivity cell and Arc Module will be shipped as one unit with
identical serial numbers. Hereafter the unit will be called sensor.
1.2 About this Operating Instruction
These Operating Instructions are designed to support the integration, operation and qualification of the Conducell UPW Arc sensors. To achieve this, it describes the features of Conducell UPW
Arc and its integration in Process Control Systems (PCS). Both
the hardware and the communication between the Conducell
UPW Arc sensor and Process Control Systems are detailed in this
manual. After reading this manual the user should be capable of
installing and operating Conducell UPW Arc sensors.
ATTENTION! Essential information for avoiding
personal injury or damage to equipment.
NOTE: Important instructions or interesting information.
2 Liability
The liability of Hamilton Bonaduz AG is detailed in the document
«General Terms and Conditions of Sale and Delivery».
Hamilton is expressly not liable for direct or indirect losses arising
from use of the sensors. It must in particular be insured in this
conjunction that malfunctions can occur on account of the inherently limited useful life of sensors contingent upon their relevant
applications. The user is responsible for the calibration, maintenance and regular replacement of the sensors. In the case of
critical sensor applications, Hamilton recommends using back-up
measuring points in order to avoid consequential damages. The
user is responsible for taking suitable precautions in the event of
a sensor failure.
3 Safety Precautions
Additionally, for warnings and alarms according to USP <645>,
two digital outputs are integrated into the sensor.
ATTENTION! The Conducell UPW Arc sensor has a
built-in temperature sensor (NTC 22 kOhm). This
temperature sensor is to be used only for monitoring
the sensor conditions, not for controlling the process
temperature.
ATTENTION! Conducell UPW Arc is not intended
for hazardous atmospheres.
ATTENTION! The measurement values transmitted
over wireless communication are not intended to be
used for process control.
and Hazards
ATTENTION! Read the following safety instructions carefully before installing and operating the
Conducell UPW Arc sensor.
3.1 General Precautions
For safe and correct use of Conducell UPW Arc sensor, it is essential that both operating and service personnel follow generally
accepted safety procedures as well as the safety instructions
given in this document, the Conducell UPW Arc sensor operation
instruction manual. The specification given in the «Specification Sheet» as regards temperature, pressure etc. may under no
circumstances be exceeded. Inappropriate use or misuse can be
dangerous.
5Operating Instructions Conducell UPW Arc Sensors
SAFETY PRECAUTIONS AND HAZARDSPRODUCT DESCRIPTION
The lifetime of the Conducell UPW Arc sensor highly depends on
the specific conditions of the application. Temperature, pressure,
chemicals used may accelerate the ageing of the sensor. See
chapter 7 for replacement conditions.
Cleaning, assembly and maintenance should be performed by
personnel trained in such work. Before removing the sensor from
the measuring setup, always make sure that no process medium
can be accidentally spilled. When removing and cleaning the
sensor, it is recommended to wear safety goggles and protective
gloves. The sensor can not be repaired by the operator and has
to be sent back to Hamilton for inspection.
Necessary precautions should be taken when transporting the
sensors. For repair or shipment the sensor should be sent back
in the original reusable packaging box. Every sensor sent back for
repair must be decontaminated.
If the conditions described in these operating instructions manual
are not adhered to or if there is any inappropriate interference
with the equipment, all of our manufacturer’s warranties become
obsolete.
3.2 Operating of Conducell UPW Arc Sensor
Conducell UPW Arc sensors must be used for their intended applications and in optimum safety and operational conditions. The
specifications (such as temperature or pressure) defined in the
section entitled ‘Technical Specification’ must not be exceeded
under any circumstances. Potential hazards can exist if the sensor is not operated correctly or appropriately.
Assembly and maintenance must be performed only by trained
personnel.
Make sure that the PG13,5 thread and the O-ring are not damaged when screwing the sensor into the process (PG 13,5 version
only). O-rings are consumable parts which must be exchanged
regularly (at least once per year). Even when all required safety
measures have been complied with, potential risks still exist with
respect to leaks or mechanical damage to the armature. Wherever there are seals or screws, gases or liquids can leak out
undetected.
Before removing the sensor from its measurement setup, always
make sure that no process medium can be accidentally spilled.
The built-in temperature sensor can only be used for monitoring
the sensor conditions, not for controlling the process temperature.
Keep the Conducell UPW Arc sensor away from other equipment
which emits electromagnetic radio frequency fields, and minimize
static electricity in the immediate environment of the measuring
parts. Carefully follow all the instructions in chapter 5.3 to avoid
electrical damage to the sensor. The contacts must be clean and
dry before sensor is connected to the cable.
ATTENTION! Switch off the power supply and unplug the connector before dismounting the Conducell
UPW Arc sensor.
ATTENTION! If the power supply (230VAC/24VDC)
is switched off or disconnected the reading on the
PCS is wrong.
3.4 Earthing
The sensor has to be mounted at the mounting location with an
electrostatically conductivity of < 1MOhm. It is recommended to
assign the VP 8 cable shield to earth (PE potential equalization)
especially in electromagnetically noisy environments. This significantly improves noise immunity and signal quality.
3.5 Chemical, Radioactive or Biological
Hazard Precautions
Selection of the appropriate safety level and implementation of
the required safety measures for working with Conducell UPW
Arc sensor is the sole responsibility of the user.
If working with hazardous liquids observe and carry out the
maintenance procedures, paying particular attention to cleaning
and decontamination. If Conducell UPW Arc sensor becomes
contaminated with biohazardous, radioactive or chemical material, it should be cleaned. Failure to observe and carry out the
maintenance procedures may impair the reliability and correct
function of the measuring module.
ATTENTION! When unscrewing the PG13,5 thread
connection never turn the sensor at the connector
head.
3.3 Electrical Safety Precautions
Do not connect the sensor to a power source of any voltage beyond the range stated in the power rating Technical Specifications
(see www.hamiltoncompany.com).
Always use Hamilton VP 8 cables for safe connection. Cables are
available in a broad range of lengths (Chapter 11). Make sure the
cable is intact and properly plugged to avoid any short circuit.
4 Product Description
Conducell UPW Arc sensors provide both standard analog (4 –
20 mA) and digital Modbus interfaces. These are built into each
sensor and are supported directly from the sensor. They do not
require any additional equipment such as amplifier or transmitter. The Conducell UPW Arc sensor consists of a conductivity
cell part with a VP 11 head and an Arc Module with a VP 8 head.
Conductivity cell and Arc Module will be shipped as one unit with
identical serial numbers. Hereafter the unit will be called sensor.
The sensor can monitor the water quality according to USP
<645> criteria. Additionally, for warnings and alarms according
to USP <645>, two digital outputs are integrated into the sensor.
For details, see the chapter «Interface for the signals according
to USP <645>.» Full details, including serial number and most
7Operating Instructions Conducell UPW Arc Sensors
PRODUCT DESCRIPTIONINSTALLATION
important specifications, can be found on the certificate provided
with each sensor. A temperature compensation in the range between 0 and 10% per K is possible and can be set manually.
The Conducell UPW sensor has been carefully tested and is
ready for use.
4.1 Hardware Description
The Conducell UPW Arc sensor consists of a sensor head with
integrated electronic and a sensor tip in contact with the measured medium. All materials in contact with the solution meet the
FDA requirement.
1
9
analog 4–20 mA. The micro transmitter located in the sensor
head stores all relevant sensor data, including calibration and
diagnostic information, simplifying calibration and maintenance.
5 Installation
5.1 Unpacking
1) Unpack carefully the Conducell UPW Arc sensor. Enclosed
you will find the sensor, the Declaration of Quality, the sensor
Instruction Manual, and the material Certificate.
2) Inspect the sensor for shipping damages or missing parts.
Conductivity Sensor
8
4
5
6
Figure 1: Components of Conducell UPW Arc exemplarily shown with a TriClamp
process connection.
Arc sensors require application specific configuration. Following
parts are required to configure and calibrate Arc sensors:
• Arc View Mobile or ArcAir computer Software Solution
• External Power supply with Arc USB Power Cable
(Ref 243490-xx)
To configure and set up the Arc sensors at least ArcAir Basic
is required. Below in this table you will find the different ArcAir
licenses and its functionality:
ArcAir Read Calibrate Configure Documentation
Free
Basic
Advanced
– – –
–
4.2 Conducell UPW Arc with
Micro-Transmitter inside
With the micro-transmitter integrated,Conducell UPW Arc
sensors offer fully compensated signal directly to the process
control system. Communication protocols include standard
Arc Wire less
Conver ter BT
ArcAir
Figure 3: Conducell UPW configuration with ArcAir.
Arc Wi 1G/2G
Adapter BT
Conducell UPW
Arc USB Power Cable
9Operating Instructions Conducell UPW Arc Sensors
INSTALLATIONINSTALLATION
5.2.1 Installing ArcAir Basic on the Computer
1) Download the Zip file «ArcAir» from the Hamilton webpage
www.hamiltoncompany.com (search for ArcAir).
2) Unpack the ZIP-File.
3) Do not plug in the Wireless Converter before the installation of
ArcAir is completed.
4) Install «ArcAir» by double clicking «ArcAir.exe» and follow the
instructions on the screen.
5.2.2 Connecting a Conducell UPW Arc
Sensor to ArcAir
1) Connect a sensors with the power supply, e.g. Arc USB Power
Cable Ref 243490-xx
2) Switch on the mobile’s Bluetooth connection or connect a
Wireless Converter BT to USB Port of your computer (only for
wireless connection
3) The ArcAir application recognizes and displays the connected
sensors automatically
ATTENTION! For automatic sensor login a unique
and global Operator Level S password for all intelligent sensors is required. Please make sure you have
added the same Operator Level S Password for all
Arc sensors in the ArcAir application under
Backstage/Settings/Operator Level S Password.
5.2.3 Create User Accounts
5.2.4 Configuring the Conducell UPW Arc
Sensor Parameters
1) Start the ArcAir application
2) Select the desired sensor
3) Open the drawer «Settings» (make sure you have the «Sensor
Settings» user right)
4) Configure the sensor
A description of the available settings is given below:
ParameterDefaultNameDescription Value Configuration Location
COND These are the μS/cm Required
Unitmeasurement Valuesphysical units:
μS/cm,
mS/c m
T unit These are the °C Required Measurement/temperature Valuesphysical units
°C, °F, K
T comp. Temperature 25°C Recommended Measurement/Tempcompensation default Valuesvaluevalue can be parameter
set between
20°C and 25°C
T comp. Temperature 0%/KRequired Measurement/factorcompensation Valuesvalue factor can be
set between
0 %/K to
10 %/ K
Moving The sensor 2 average uses a moving default Valuesaverage over parameter
the measuring
points
Recommended
Measurement
Measurement/
/
1) Start ArcAir application on computer
2) Click on «Backstage» left upper corner
3) Select «User Management»
4) Click the «Add» Button for opening the user editor
5) Type in the user details and password
6) Select the specific rights for the user
ATTENTION! First user is the administrator and all
user rights are assigned as default.
NOTE: Initial operation of ArcAir is in the laboratory
mode as long as no user account is created. Laboratory
mode does not require a login password and enable all
features in the installed license version.
5.2.5 Configuring the calibration settings
ParameterDefaultNameDescription Value Configuration Location
Drift stability Higher drif t will 1%/Min Recommend Calibration/interrupt the default Calibrationcalibration parameter Settingsprocess.
Warning
comes up
«drif t oxygen»
Higher drift will 0.5 K/min Recommend Calibration/interrupt the default Calibrationcalibration parameter Settingsprocess.
Warning
comes up «drift
temperature»