Hamilton Conducell 4US-T150-50, Conducell 4US, Conducell 4US-G125, Conducell 4US-T150-100 Operating Instructions Manual

Conducell™ 4US Sensors Датчики Conducell ™ 4US
Operating Instructions Bedienungsanleitung Инструкция по эксплуатации
Important note
All efforts have been made to ensure the accuracy of the contents of this manual. However, should any errors be detected, Hamilton Bonaduz AG would greatly appreciate being informed of them. The above notwithstanding, Hamilton Bonaduz AG can assume no responsibility for any errors in this manual or their consequences.
Copyright © 2015 Hamilton Bonaduz AG, Bonaduz Switzerland. All rights reserved.
TABLE OF CONTENTS
Table of Contents
1 Liability ................................................................................. 4
2 Intended use ........................................................................ 5
3 Safety ................................................................................... 5
4 Initial operation .................................................................... 5
5 Electrical wiring ................................................................... 6
6 Assembly ............................................................................. 8
7 Diagrams with dimensions ................................................. 9
8 Calibration .......................................................................... 10
9 Cleaning .............................................................................. 11
10 Disposal .............................................................................. 11
11 Technical data .................................................................... 11
12 Instructions for use in potentially
explosive atmospheres ..................................................... 12
13 Connection to transmitter ................................................ 13
14 Accessories ....................................................................... 14
4US Sensors Operating Instructions
3Conducell
LIABILITY
Operating Instructions
These operating instructions are intended for the Hamilton Conducell 4US sensors.
Ref Description Thread ATEX
237700-OP Conducell 4US-G125 G 11⁄4” 237750 Conducell 4US-T150-50 Triclamp 1.5” 237760 Conducell 4US-T150-100 Triclamp 1.5”
Hamilton Conducell 4US sensors are quality products manufactured with the latest technology.
These instructions should be read, understood and followed by all staff using the device. Hamilton can assume no responsibility for damage and operational disruptions arising from failure to observe these instructions.
1 Liability
The Liability of Hamilton Bonaduz AG is detailed in the document “General Terms and Conditions of Sale and Delivery (GTS)”, chapter 12.
Hamilton is expressly not liable for direct or indirect losses arising from use of the sensors. It must in particular be insured in this conjunction that malfunctions can occur on account of the inherently limited useful life of sensors contingent upon their relevant applications. The user is responsible for the calibration, maintenance and regular replacement of the sensors. In the case of critical sensor applications, Hamilton recommends using back-up measuring points in order to avoid consequential damages. The user is responsible for taking suitable precautions in the event of a sensor failure.
INTENDED USE • SAFETY • INITIAL OPERATION
2 Intended use
With the Conducell 4US line, Hamilton has developed a family of in-line conductivity sensors which perform under demanding conditions in biotechnology and the general chemical industry. During development, special attention was paid to an optimum sanitary design. All materials in contact with the solution are FDA approved.
The sensors are based on the 4-electrode principle which excludes polarization phenomena normally observed with 2-electrode sensors. The engineered design guarantees an excellent linearity over a conductivity range of more than 6 decades.
3 Safety
This sensor is to be used only as intended and is to be kept in a condition that ensures com­plete safety. The specifications given in the section “Technical Data” as regards temperature, pressure etc. may under no circumstances be exceeded.
Inappropriate use or misuse can be dangerous.
Assembly and maintenance may only be done by trained personnel.
During assembly, ensure that the positioning is correct and that the O-ring is undamaged. O-rings are consumable items and must be replaced regularly – at least once a year.
The built-in temperature sensor should only be used for compensation of conductivity, not for controlling the process temperature.
Conducell 4US sensors are used to determine electrolyte conductivity in solutions. If these sensors are used in potentially explosive atmospheres, see the section “Instructions for use in potentially explosive atmospheres”.
4 Initial operation
This Conducell 4US sensor has been carefully tested and is ready for use. During unpacking, please check for any mechanical defects. In case of a complaint please return the Conducell 4US in its original packing to your Hamilton representative and describe the problem.
4US Sensors Operating Instructions
5Conducell
ELECTRICAL WIRING
5 Electrical wiring
Signal description Cable Conducell 4US
Pt1000 (low end) Grey Pt1000 (low end) White Pt1000 (high end) Blue Current electrode (high end) Pink Potential electrode (high end) Green Potential electrode (low end) Brown Current electrode (low end) Yellow Not connected on sensor Shield Not connected Red
Conducell Hamilton Knick Knick Stratos Knick Stratos e 4US H100 Cond Cond 2201 X 2401 Cond 2402 / 2405 Cond
Grey D 7 6 D White D 7 6 D Blue E 8 5 E Pink 1 1 1 1 Green 2 2 2 2 Brown 3 3 3 3 Yellow 4 4 4 4 Shield C 5 5 C Red - - - ­Comment Short 4 and 5 Short 4 and 5 Short 4 and 5 Short 4 and 5
ELECTRICAL WIRING
Conducell 4US Emerson Yokogawa SIEMENS 5081 C Model SC150, SC202, SIPAN SC402, SC450 32/32X
Grey 3 11 23 White 4 11 24 Blue 5 12 25 Pink 8 14 26 Green 11 13 29 Brown open 15 27 Yellow open 16 30 Shield 5 63 Shield Red - - ­Comment 2-electrode arrangement *
* For Yokogawa SC150, SC202, SC402, SC450 please be aware that the measuring range must stay within the specified input range of the instrument.
ATTENTION! The linearity of the system is strongly dependent on the instru­ment to which the sensor is attached. For applicable measuring range refer to Technical Specifications of the manufacturer of the instrument.
4US Sensors Operating Instructions
7Conducell
ASSEMBLY
6 Assembly
Please check if the process connection is sufficiently clean. Then fit the Conducell 4US. Please follow the instructions mentioned below.
Examples:
ATTENTION! The cell constant and the linearity of the sensor can vary with the fitting situation. A symmetrical setup is recommended. Leave an open space of 60 mm minimum diameter.
Partitions made of non-conductive material
Ø 60mm
min
should preferably be used.
ATTENTION! A symmetrical setup is recommended in order to ensure a high degree of linearity. To achieve high precision the cell con­stant should be calibrated in the final setup. Make sure that all 4 electrodes are completely and continuously immersed in the measuring sample.
DIAGRAMS WITH DIMENSIONS
Position marker
If the sensor is dismounted, the position marker helps to remount it in the original position.
7 Diagrams with dimensions
E = H + 37
H*
Ø 13
* H = O-ring position (22 mm ≤ OP ≤ 55 mm)
4US Sensors Operating Instructions
Ø 50.5
Ø 24
Ø 50.5
Ø 24
50
97
9Conducell
CALIBRATION
8 Calibration
There are two possibilities for calibration:
1) Calibration while sensor is disassembled from process:
• First, rinse the sensor thoroughly with deionized water.
• Use a beaker or container that has an (internal) diameter of at least 30 mm. (Conducell 4US fits perfectly into the Hamilton Conductivity Standards, and can be calibrated directly in the bottle.)
• Use a solution of known conductivity. Check the temperature table of the conductivity standard.
• Immerse the lower part of the sensor in the standard. Make sure that the 4 electrodes are completely immersed. The sensor should be placed in the center of the beaker.
• Leave the sensor for at least 5 minutes for equilibration, before initiating the calibration on the instrument.
For precise determination of the cell constant, it is recommended that you use an assembly situation similar or equal to the situation in the process. The cell constant may vary with the assembly situation.
2) Calibration in the process:
• Insert the sensor in the process.
• Leave conductivity and temperature at least 15 min. for equilibration.
• Take a process sample and perform an external measurement with a reference conductivity system. The best approach is to perform the measurement at a temperature equal to that of the process. If this is not possible, you need to know the temperature coefficient of your sample (similar to temperature table on Hamilton Conductivity Standards bottle).
• Manually adjust the cell constant to read the same conductivity value on the process instrument.
• You may switch off the temperature compensation of the process and the laboratory instru­ments to prevent any errors.
CLEANING • DISPOSAL • TECHNICAL DATA
9 Cleaning
Conducell 4US can be cleaned with standard cleaning agents. Never use scouring agents which contain hard particles. Please check the O-rings from time to time. When mounting the new O-ring take care not to damage the O-ring nut or the O-ring itself. Otherwise the tightness of seal can no longer be guaranteed.
10 Disposal
The design of Hamilton sensors minimizes environmental impact. According to the EU directive 2002/96/EC the Hamilton sensors should be disposed as waste of electrical and electronic equipment, and not in municipal waste, or it can be sent back to Hamilton for disposal.
11 Technical data
Measuring Range 0.1 μS/cm to 500 mS/cm Linearity (relative) ± 0.5 - 5 % Cell constant 0.147 cm-1 Process Temperature -20 to 135 °C Pressure Range 0 to 6 bar Temperature Sensor Pt1000 Wetted Par ts PEEK (FDA approved), Stainless Steel DIN 1.4435,
EPDM (FDA approved)
Surface Quality of Steel Ra < 0.4 μm (N5) O-ring Material EPDM (FDA approved) Process Connection G125: G 1-1/4” / T150 : Triclamp 1.5” Electrical Connector 5 m Open End
1) Uncertainty of ± 5 % arise when using only ONE single cell constant for the full range. ± 0.5% accuracy can be
achieved when calibration is performed in a conductivity range close to that of the sample. Remember also that linearity is strongly influenced by the transmit ter used.
2) Individual cell constant measured with Hamilton standard procedure printed on the label. The cell constant is
influenced by the assembly situation.
4US Sensors Operating Instructions
1)
2)
11Conducell
INSTRUCTIONS FOR USE IN POTENTIALLY EXPLOSIVE ATMOSPHERES
Fig. 1: Linearity of the Conducell 4US, connected to a typical process transmitter. One single cell constant dete rmined at 1413 μS/cm was used over the full range.
ATTENTION! The linearity of the system is strongly dependent on the
transmitter connected to the sensor.
12 Instructions for use in potentially explosive atmospheres
ATEX / IECEx marking: Gas: CE 0035 II 1/2 G Ex ia IIB T4/T5/T6 Ga/Gb Dust: CE 0035 II 1/2 D Ex ia IIIC T x °C Da/Db
Manufacturer: Hamilton Bonaduz AG, 7402 Bonaduz, Switzerland
EC type examination report: TÜV 03 ATEX 7005 X IECEx certificate of conformity: IECEx TUR 14.0001 X
EC type examination report and IECEx certificate of conformity can be downloaded from www.hamiltoncompany.com.
CONNECTION TO TRANSMITTER
The conditions described in the ATEX / IECEx certificate must be respected.
ATTENTION! In case a gas atmosphere and a dust atmosphere are or could be present at the same time, the risk of explosion must be examined carefully and special precautions may be necessary.
Assembly
1) The operator of equipment in potentially explosive atmospheres is responsible for ensuring that all components of the system are certified for that area classification and are compatible with each other.
2) In gas atmospheres, the Ex approval is not dependent on any conditions regarding mount­ing of the sensor. In dust atmospheres, however, special restrictions may apply when using armatures made of plastic material.
3) O-rings having a sealing function between Ex-zone 0 and 1 or Ex-zone 20 and 21, respectively, must be replaced at each dismantling of the sensor.
4) Sensors, transmitter and equipment required are to be set up within one balanced potential system.
5) When the sensor is assembled, the ATEX / IECEx sticker is to be attached to the cable in an easily visible place, as near as possible to the sensor itself. This sticker is to indicate that an ATEX / IECEx approved sensor is in use. The sticker should not be removed at any time.
13 Connection to transmitter
Conducell 4US is suitable for connection to an intrinsically safe electrical circuit with protection level ia. The operator of the equipment must ensure that the allowable electrical values for the sensor all exceed those for the transmitter. None of the values cited for electrical power, voltage and current may be exceeded in sum (measurement and temperature circuit together). Electric values for Conducell 4US are as follows:
U = 24 V; I = 173 mA; P: Function of process temperature. The maximum allowable electric power P of the transmitter (measurement and temperature circuit together) depends on the desired process temperature.