Hamilton C2 User manual

Operator‘s Manual
PN 624131/06
Software version 2.x
October 2010
Preface
This manual may be revised or replaced by HAMILTON MEDI­CAL at any time and without notice. Ensure that you have the most current applicable version of this manual; if in doubt, contact HAMILTON MEDICAL AG Marketing Department. While the information set forth is believed to be accurate, it is not a substitute for the exercise of professional judgment.
Nothing in this manual shall limit or restrict in any way HAMIL­TON MEDICAL’s right to revise or otherwise change or modify the equipment (including its software) described herein, with­out notice. In the absence of an express, written agreement to the contrary, HAMILTON MEDICAL has no obligation to furnish any such revisions, changes, or modifications to the owner or user of the equipment (including software) described herein.
The equipment must be operated and serviced by trained pro­fessionals only. HAMILTON MEDICAL’s sole responsibility with respect to the equipment and its use is as stated in the Limited Warranty provided in this manual.
Product and company names mentioned herein may be the trademarks of their respective owners.
HAMILTON MEDICAL will make available on request circuit dia­grams, component parts lists, descriptions, calibration instruc­tions, or other information that will assist the user’s authorized trained personnel to repair those parts of the equipment deemed by HAMILTON MEDICAL to be repairable.
624131/06 iii
Manufacturer Distributor in USA
HAMILTON MEDICAL AG Via Crusch 8 CH-7402 Bonaduz Switzerland Phone: (+41) 81 660 60 10 Fax: (+41) 81 660 60 20 info@hamilton-medical.com www.hamilton-medical.com
HAMILTON MEDICAL, Inc. 4990 Energy Way P.O. Box 30008 Reno, NV 89520 Phone: (775) 858-3200 Toll-free: (800) 426-6331 Fax: (775) 856-5621 marketing@hamilton-medical.net
iv 624131/06
HAMILTON-C2 software information
The software version for the HAMILTON-C2 is visible in the
System -> Info
window. The software version for the VUP (ventilator unit processor) (that is, the digit to the left of the decimal point for VUP) should match the version on the title page of this manual. See section 3.3.1 for details.
Definitions
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in minor or moderate injury.
NOTE:
Emphasizes information of particular importance.
Applies only when the neonatal option is installed.
624131/06 v
Applies only when the CO2 sensor option is installed
Applies only when the TRC option is installed
vi 624131/06
General cautions and notes
Intended use
The HAMILTON-C2 ventilator is intended to provide positive pressure ventilatory support to adults, pediatrics, infants and neonates.
Intended areas of use:
• In the intensive care ward or in the recovery room
• During secondary transport from one hospital to another
• During transfer of ventilated patients within the hospital The HAMILTON-C2 ventilator is a medical device intended for
use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifica­tions.
CAUTION
(USA only): Federal law restricts this device to sale by or on the order of a physician.
NOTE:
Not all options are available in all countries.
General operation notes
• The displays shown in this manual may not exactly match
what you see on your own ventilator.
• Familiarize yourself with this operator’s manual before using
the ventilator on a patient.
• Displayed information that is ghosted is not active and may
not be selected.
• Dashes displayed in place of monitored data indicate that
valid values are not yet available or do not apply.
• If a ventilator control does not respond when selected by
touch or by the turn of a knob, the control is not active in this particular instance or the function is not implemented.
624131/06 vii
Monitoring and alarms
• The HAMILTON-C2 is not intended to be a comprehensive vital sign monitor for patients on life-support equipment. Patients on life-support equipment should be appropriately monitored by qualified medical personnel and suitable monitoring devices. The use of an alarm monitoring system does not give absolute assurance of warning for every form of malfunction that may occur with the ventilator. Alarm messages may not exactly pinpoint a problem; the exercise of clinical judgment is necessary.
• An alternative means of ventilation shall be available when­ever the ventilator is in use. If a fault is detected in the ven­tilator or its life-support functions are in doubt, disconnect the HAMILTON-C2 from the patient and immediately start ventilation with such a device (for example, a resuscitation bag), using PEEP and/or increased oxygen concentration when appropriate. The ventilator must be removed from clinical use and serviced by a HAMILTON MEDICAL autho­rized service engineer.
• It is recommended that additional independent monitoring devices be used during mechanical ventilation. The opera­tor of the ventilator must still maintain full responsibility for proper ventilation and patient safety in all situations.
• Do not silence the audible alarm when leaving the patient unattended.
• Do not use the exhaust port of the expiratory valve for spirometry. Due to the HAMILTON-C2’s base flow, the exhaust gas output is larger than the patient’s actual exhaled volume.
• Do not put a vessel filled with a liquid on the ventilator. If a liquid enters the product, a fire and/or electric shock may occur.
Fire and other hazards
• To reduce the risk of fire or explosion, do not place the ven­tilator in a combustible or explosive environment (for exam­ple, around flammable anesthetics or other ignition
viii 624131/06
sources). Do not use it with any equipment contaminated with oil or grease.
• To reduce the risk of fire, do not use high-pressure gas
hoses that are worn or contaminated with combustible materials like grease or oil.
• To reduce the risk of fire, use only breathing circuits
intended for use in oxygen-enriched environments. Do not use antistatic or electrically conductive tubing.
• In case of fire, immediately secure the patient’s ventilatory
needs, switch off the ventilator, and disconnect it from its gas and electrical sources.
Service and testing
• To ensure proper servicing and to prevent possible physical
injury, only HAMILTON MEDICAL authorized service person­nel should attempt to service the ventilator.
• To reduce the risk of electrical shock, disconnect electrical
power from the ventilator before servicing.Be aware that battery power remains even after the mains is discon­nected. Be aware that if the power switch is off, some parts still carry high voltage.
• Do not attempt service procedures other than those speci-
fied in the service manual.
• Use replacement parts supplied by HAMILTON MEDICAL
only.
• Any attempt to modify the ventilator hardware or software
without the express written approval of HAMILTON MEDI­CAL automatically voids all warranties and liabilities.
• The preventive maintenance program requires a general
service every 5000 hours or yearly, whichever comes first.
• To ensure the ventilator’s safe operation, always run the
tests and calibrations prescribed in Section 3 before using the ventilator on a patient.If the ventilator fails any tests, remove it from clinical use immediately. Do not use the ven­tilator until necessary repairs are completed and all tests have passed.
624131/06 ix
• The manufacturer can only be responsible for the safety, reliability, and performance of the ventilator if:
– appropriately trained personnel carry out assembly
operations, extensions, readjustments, modifications or repairs;
– the electrical installation of the relevant room complies
with the appropriate requirements; and
– the ventilator system is used in accordance with the
operator’s manual.
Electromagnetic susceptibility
The HAMILTON-C2 complies with the IEC 60601-1-2 EMC (Electro Magnetic Compatibility) Collateral Standard. It is intended for use in the electromagnetic environment described in Table A-17 through Table A-19. Do not use the HAMILTON­C2 in an environment with magnetic resonance imaging (MRI) equipment.
Units of measure
NOTE:
In this manual pressure is indicated in cmH2O, PCO2 in mmHg and length in cm.
On the HAMILTON-C2 pressures are indicated in cmH2O, mbar or hPa. Hectopascals (hPa) are used by some institutions instead. Since 1 mbar equals 1 hPa, which equals
1.016 cmH
indicated in mmHg, Torr or kPa and length in cm or inch.
O, the units may be used interchangeably. CO2 is
2
Disposal
Dispose of all parts removed from the device according to your institution’s protocol. Follow all local, state, and federal regula­tions with respect to environmental protection, especially when disposing of the electronic device or parts of it (for example oxygen cell, batteries).
x 624131/06
Year of manufacture
The year of manufacture is shown on the serial number label on the HAMILTON-C2 ventilation unit.
624131/06 xi
xii 624131/06
Table of contents
1 General information. . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.2 Functional description . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1.2.1 System overview . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1.2.2 Gas supply and delivery . . . . . . . . . . . . . . . . . . . . . 1-5
1.2.3 Gas monitoring with the Flow Sensor. . . . . . . . . . . 1-7
1.3 Physical description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
1.3.1 Breathing circuits and accessories. . . . . . . . . . . . . . 1-9
1.3.2 Ventilator unit . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
1.3.3 Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20
1.4 Symbols used on device labels and packaging . . . . . . . . 1-22
2 Preparing for ventilation . . . . . . . . . . . . . . . . . . . . 2-1
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.2 Installing the patient tubing support arm . . . . . . . . . . . . . 2-4
2.3 Installing the humidifier . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2.4 Installing the option board . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2.5 Installing the patient breathing circuit . . . . . . . . . . . . . . . 2-6
2.6 Installing the pneumatic nebulizer . . . . . . . . . . . . . . . . . 2-17
2.7 Setting up for monitoring by the optional CO2 sensor . . 2-19
2.7.1 CO2 mainstream measurement . . . . . . . . . . . . . . 2-20
2.7.2 CO2 sidestream measurement . . . . . . . . . . . . . . . 2-24
2.8 Installing the optional Aeroneb Pro nebulizer . . . . . . . . . 2-26
2.9 Using an expiratory filter . . . . . . . . . . . . . . . . . . . . . . . . 2-26
2.10 Connecting to primary power source . . . . . . . . . . . . . . . 2-27
2.10.1Connecting to AC power . . . . . . . . . . . . . . . . . . . 2-28
2.10.2Connecting to DC power . . . . . . . . . . . . . . . . . . . 2-28
2.11 About the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30
2.12 Connecting the oxygen supply . . . . . . . . . . . . . . . . . . . . 2-32
2.13 Connecting to an external patient monitor or other device. . . 2-33
2.14 Starting up the ventilator . . . . . . . . . . . . . . . . . . . . . . . . 2-34
2.15 Shutting down the ventilator . . . . . . . . . . . . . . . . . . . . . 2-35
2.16 Display navigation guidelines . . . . . . . . . . . . . . . . . . . . . 2-35
3 Tests, calibrations and utilities . . . . . . . . . . . . . . . 3-1
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3.2 Running the preoperational check . . . . . . . . . . . . . . . . . . 3-4
3.3 System functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3.3.1 Info: Viewing device-specific information . . . . . . . . 3-6
624131/06 xiii
Table of contents
3.3.2 Tests & calib: Running sensor calibrations and the tight-
ness test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
3.3.3 Sensors on/off: Enabling/disabling
O2 and CO2 monitoring . . . . . . . . . . . . . . . . . . . . 3-15
3.3.4 Setting day and night . . . . . . . . . . . . . . . . . . . . . .3-16
3.3.5 Setting date and time . . . . . . . . . . . . . . . . . . . . . . 3-17
3.4 Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-18
3.4.1 Configuration: Configuring the ventilator . . . . . . .3-18
3.4.2 Data transfer: Copying event log data to a USB memory device .3-19
3.5 Alarm tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
3.5.1 High pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-21
3.5.2 Low minute volume . . . . . . . . . . . . . . . . . . . . . . . 3-22
3.5.3 Low oxygen alarm. . . . . . . . . . . . . . . . . . . . . . . . .3-22
3.5.4 Disconnection on patient side . . . . . . . . . . . . . . . . 3-22
3.5.5 Loss of external power . . . . . . . . . . . . . . . . . . . . .3-23
3.5.6 Exhalation obstructed . . . . . . . . . . . . . . . . . . . . . .3-23
3.5.7 Apnea. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-23
4 Ventilator settings . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.2 Patient grouping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
4.3 Quick start-up settings . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.4 Patient setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
4.5 Modes window: Setting the ventilation mode . . . . . . . . . . 4-6
4.6 Controls windows: Setting controls including apnea backup
ventilation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
4.6.1 Adjusting and confirming control settings without
mode change . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
4.6.2 Adjusting and confirming control settings after mode
change . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
4.6.3 About apnea backup ventilation . . . . . . . . . . . . . .4-12
4.6.4 Table of control settings, mode additions and ranges..
4-15
4.7 Alarms windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
4.7.1 Limits 1 and Limits 2: Setting alarm limits . . . . . . . 4-23
4.7.2 Loudness: Adjusting alarm loudness . . . . . . . . . . . 4-25
4.7.3 Buffer: Viewing alarm information . . . . . . . . . . . . 4-26
4.7.4 Table of alarm limit settings and ranges . . . . . . . .4-27
4.7.5 TRC: Setting tube resistance compensation . . . . . .4-29
xiv 624131/06
5 Neonatal ventilation . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.1 Breathing circuit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.2 Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.3 Testing and calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5.4 Ventilation modes and mode additions . . . . . . . . . . . . . . 5-3
5.5 Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5.5.1 Ti max . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.5.2 Flowtrigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.5.3 P-ramp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
5.6 Others . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
6 Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
6.2 Values window: Viewing numeric patient data . . . . . . . . . 6-4
6.3 Graphics window: Selecting second screen graphic . . . . . 6-7
6.4 About graphic types. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6.4.1 Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6.4.2 Dynamic Lung . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
6.4.3 Vent Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
6.5 Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
6.6 Loops. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
6.7 Table of monitored parameters . . . . . . . . . . . . . . . . . . . 6-14
6.8 Freeze and cursor measurement. . . . . . . . . . . . . . . . . . . 6-23
7 Intelligent Panels . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
7.2 Dynamic Lung panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
7.2.1 Tidal volume (Vt) . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
7.2.2 Compliance (Cstat) . . . . . . . . . . . . . . . . . . . . . . . . 7-3
7.2.3 Patient triggering: Muscle . . . . . . . . . . . . . . . . . . . 7-3
7.2.4 Resistance: Bronchial tree. . . . . . . . . . . . . . . . . . . . 7-4
7.3 Vent Status panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
8 Responding to alarms . . . . . . . . . . . . . . . . . . . . . . 8-1
8.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
8.2 How to respond to an alarm . . . . . . . . . . . . . . . . . . . . . . 8-6
8.3 Alarm buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
8.4 Events window: Reviewing the event log . . . . . . . . . . . . . 8-9
8.5 Alarm troubleshooting table. . . . . . . . . . . . . . . . . . . . . . 8-10
624131/06 xv
Table of contents
9 Special functions. . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
9.1 Standby . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
9.2 O2 enrichment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
9.3 Suctioning tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
9.4 Manual breath/inspiratory hold . . . . . . . . . . . . . . . . . . . . . 9-7
9.5 Nebulizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
9.6 Print screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9
10 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
10.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-2
10.2 Cleaning, disinfection and sterilization . . . . . . . . . . . . . .10-2
10.2.1General guidelines for cleaning . . . . . . . . . . . . . . .10-7
10.2.2General guidelines for chemical disinfection . . . . . 10-7
10.2.3General guidelines for autoclave, ETO or plasma steril-
ization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8
10.3 Preventive maintenance . . . . . . . . . . . . . . . . . . . . . . . . .10-9
10.3.1Servicing the air intake and fan filters . . . . . . . . . 10-10
10.3.2Replacing the batteries . . . . . . . . . . . . . . . . . . . .10-13
10.3.3Charging and calibrating the batteries . . . . . . . .10-14
10.3.4Replacing the oxygen cell . . . . . . . . . . . . . . . . . .10-14
10.4 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-15
10.5 Repacking and shipping . . . . . . . . . . . . . . . . . . . . . . . . 10-15
A Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
A.1 Physical characteristics. . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A.2 Environmental requirements . . . . . . . . . . . . . . . . . . . . . . A-3
A.3 Pneumatic specifications . . . . . . . . . . . . . . . . . . . . . . . . . A-3
A.4 Electrical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
A.5 Control settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5
A.6 Monitored parameters . . . . . . . . . . . . . . . . . . . . . . . . . A-13
A.7 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-27
A.8 Configuration specifications . . . . . . . . . . . . . . . . . . . . . A-31
A.9 Ventilator breathing system specifications . . . . . . . . . . . A-33
A.10 Other technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . A-34
A.11 Standards and approvals . . . . . . . . . . . . . . . . . . . . . . . . A-36
A.12 EMC declarations (IEC 60601-1-2). . . . . . . . . . . . . . . . . A-37
A.13 Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-42
A.14 Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-43
xvi 624131/06
B Modes of ventilation . . . . . . . . . . . . . . . . . . . . . . . B-1
B.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
B.2 The biphasic concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
B.3 Mandatory modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-8
B.3.1 (S)CMV+ mode or APVcmv . . . . . . . . . . . . . . . . . . B-8
B.3.2 PCV+ mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-11
B.4 Spontaneous modes (SPONT and NIV) . . . . . . . . . . . . . . B-13
B.5 SIMV modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-19
B.5.1 SIMV+ mode or APVsimv . . . . . . . . . . . . . . . . . . . B-20
B.5.2 PSIMV+ and NIV-ST modes . . . . . . . . . . . . . . . . . B-23
B.6 nCPAP-PS mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-28
B.6.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-28
B.6.2 Controls of nCPAP-PS . . . . . . . . . . . . . . . . . . . . . B-30
B.7 DuoPAP (Duo positive airway pressure). . . . . . . . . . . . . . B-31
B.7.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-31
B.7.2 The many faces of DuoPAP . . . . . . . . . . . . . . . . . B-32
B.7.3 Pressure support in DuoPAP breaths . . . . . . . . . . B-33
B.7.4 Synchronization . . . . . . . . . . . . . . . . . . . . . . . . . . B-34
B.7.5 Controls of DuoPAP . . . . . . . . . . . . . . . . . . . . . . . B-34
B.8 APRV (Airway pressure release ventilation) . . . . . . . . . . . B-36
B.8.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-36
B.8.2 Initialization of APRV . . . . . . . . . . . . . . . . . . . . . . B-37
B.8.3 Sustained high pressure recruitment manoeuvres. B-39
B.8.4 Controls of APRV . . . . . . . . . . . . . . . . . . . . . . . . . B-39
B.9 SAFETY mode and ambient state . . . . . . . . . . . . . . . . . . B-41
C ASV (adaptive support ventilation) . . . . . . . . . . . C-1
C.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
C.2 ASV use in clinical practice . . . . . . . . . . . . . . . . . . . . . . . . C-3
C.3 Detailed functional description of ASV . . . . . . . . . . . . . . C-15
C.3.1 Normal minute ventilation . . . . . . . . . . . . . . . . . . C-15
C.3.2 Targeted minute ventilation . . . . . . . . . . . . . . . . . C-16
C.3.3 Lung-protective rules strategy . . . . . . . . . . . . . . . C-17
C.3.4 Optimal breath pattern . . . . . . . . . . . . . . . . . . . . C-20
C.3.5 Dynamic adjustment of lung protection . . . . . . . . C-24
C.3.6 Dynamic adjustment of optimal breath pattern . . C-25
C.4 Minimum work of breathing (Otis’ equation) . . . . . . . . . C-26
C.5 ASV technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-28
C.6 ASV Start up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-31
C.7 References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-32
624131/06 xvii
Table of contents
D NIV (non invasive ventilation) . . . . . . . . . . . . . . . .D-1
D.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2
D.2 Benefits of noninvasive ventilation, . . . . . . . . . . . . . . . . . D-3
D.3 Required conditions for use . . . . . . . . . . . . . . . . . . . . . . . D-4
D.4 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-4
D.5 Potential adverse reactions . . . . . . . . . . . . . . . . . . . . . . . D-5
D.6 Selecting a patient interface . . . . . . . . . . . . . . . . . . . . . . D-5
D.7 Control settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-6
D.8 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-7
D.9 Monitored parameters . . . . . . . . . . . . . . . . . . . . . . . . . . D-8
D.10 Additional notes about using noninvasive ventilation. . . . D-9
D.11 References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-11
E CO2 sensor option: Volumetric capnography . . . E-1
E.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-2
E.2 CO2 elimination (V’CO2). . . . . . . . . . . . . . . . . . . . . . . . . . E-3
E.3 End-tidal CO2 (PetCO2 and FetCO2) . . . . . . . . . . . . . . . . . E-5
E.4 Airway dead space (VDaw) . . . . . . . . . . . . . . . . . . . . . . . . E-5
E.5 Alveolar minute ventilation (V’alv) . . . . . . . . . . . . . . . . . . . E-6
E.6 Capnogram shape (slope of the alveolar plateau, slopeCO2) .
E-7
E.7 Formulas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-9
E.8 References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-10
F Low-pressure oxygen. . . . . . . . . . . . . . . . . . . . . . . F-1
G Pneumatic diagram. . . . . . . . . . . . . . . . . . . . . . . . .G-1
H Parts and accessories . . . . . . . . . . . . . . . . . . . . . . .H-1
I Communications interface . . . . . . . . . . . . . . . . . . . I-1
I.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-2
I.2 Patient monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-2
I.2.1 Patient data management system (PDMS) or other com-
puter system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-4
I.2.2 Connector pin assignments. . . . . . . . . . . . . . . . . . . I-6
I.3 Nurse call . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-8
I.3.1 Inspiratory:expiratory (I:E) timing outlet . . . . . . . . . . I-8
I.3.2 Remote alarm outlet . . . . . . . . . . . . . . . . . . . . . . . . I-9
I.3.3 Connector pin assignments. . . . . . . . . . . . . . . . . . I-10
xviii 624131/06
J Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . J-1
J.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .J-2
J.2 Accessing configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . J-2
J.3 General: Selecting the language, units of measure and oxygen
source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . J-3
J.3.1 Language: Selecting the default language . . . . . . . .J-3
J.3.2 Units: Selecting the default unit of measure for pressure
display, CO2 and length. . . . . . . . . . . . . . . . . . . . . .J-3
J.3.3 More: Selecting the oxygen source and enabling the
communications interface . . . . . . . . . . . . . . . . . . . . J-4
J.4 Graphics window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .J-5
J.4.1 MMP: Selecting the default main monitoring parameter
display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . J-5
J.5 Setup window (quick start-up settings). . . . . . . . . . . . . . . . J-6
J.5.1 Use setups: Define the default quick start-up settings .
J-6
J.5.2 Use setups: Configure the quick start-up settings. . .J-7
J.6 Transfer window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .J-12
J.7 Options windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .J-13
Glossary. . . . . . . . . . . . . . . . . . . . . . . . . . . . Glossary-1
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Index-1
User Notes. . . . . . . . . . . . . . . . . . . . . . . . User Notes-1
624131/06 xix
Table of contents
xx 624131/06
List of Figures
1-1 Gas delivery in the HAMILTON-C2 . . . . . . . . . . . . . . . . . . . . . . . 1-6
1-2 Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
1-3 HAMILTON-C2 with accessories . . . . . . . . . . . . . . . . . . . . . . . . 1-10
1-4 Front view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
1-5 Rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
1-6 Left side view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17
1-7 Right side view. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-18
1-8 Default (basic) screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-20
2-1 Installing the patient tubing support arm and humidifier . . . . . . 2-4
2-2 Option board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
2-3 Patient breathing circuit for use with inspiratory heater wire (Pediatric/
Adult) - (In place of the flex tube shown, a 15 x 22 adapter may be
used to attach the Flow Sensor to the ET tube.) . . . . . . . . . . . . . 2-9
2-4 Patient breathing circuit for use without heater wires (Pediatric/Adult)
- (In place of the flex tube shown, a 15 x 22 adapter may be used to
attach the Flow Sensor to the ET tube.) . . . . . . . . . . . . . . . . . . 2-10
2-5 Patient breathing circuit for use with HME (Pediatric/Adult) -
(In place of the flex tube shown, a 15 x 22 adapter may be used to
attach the Flow Sensor to the HME or ET tube.) . . . . . . . . . . . . 2-11
2-6 Patient breathing circuit with inspiratory heater wire for use with nC-
PAP-PS (non-invasive ventilation neonatal) - (Use a 15M x 15F adapter
to connect the infant flow sensor.) . . . . . . . . . . . . . . . . . . . . . 2-12
2-7 Patient breathing circuit with inspiratory heater wire (invasive ventila-
tion neonatal) - (Use a 15M x 15F adapter to connect the infant flow
sensor.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
2-8 LiteCircuit (single-limb) patient breathing circuit (non-invasive ventila-
tion Pediatric/Adult) - (For use with NIV or NIV-ST) . . . . . . . . . . 2-14
2-9 Installing the expiratory valve . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
2-10 Installing the Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2-11 Installing a pneumatic nebulizer . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2-12 Connecting the CO2 sensor . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
2-13 Attaching the CO2 sensor to the airway adapter . . . . . . . . . . . 2-22
2-14 Connecting the CO2 sensor/airway adapter to the patient circuit . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
2-15 Inserting the sample cell into the receptacle . . . . . . . . . . . . . . . 2-25
2-16 Attaching the CO2 sensor to the airway adapter . . . . . . . . . . . 2-25
2-17 Installing the Aeroneb Pro nebulizer . . . . . . . . . . . . . . . . . . . . . 2-26
2-18 Car adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-29
624131/06 xxi
List of Figures
2-19 Power source symbols and battery charge indicator. . . . . . . . . 2-31
2-20 Oxygen inlet fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
2-21 Power switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-34
3-1 Info window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
3-2 Tests & calib window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-3 Disconnection of the mainstream sensor for calibration . . . . . . 3-14
3-4 Disconnection of the sidestream sensor for calibration. . . . . . . 3-14
3-5 Sensor on/off window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3-6 Day/Night window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
3-7 Date&Time window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
3-8 Data transfer window 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
3-9 Data transfer window 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
4-1 Quick start-up settings (Example). . . . . . . . . . . . . . . . . . . . . . . . 4-4
4-2 Patient setup window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4-3 Patient setup window neonatal . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
4-4 Modes window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4-5 Basic (Controls) window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
4-6 More window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
4-7 TRC window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4-8 Basic window during mode change (ASV mode change) . . . . . 4-12
4-9 Automatic button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
4-10 Limits 1 window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
4-11 Limits 2 window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
4-12 Loudness window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
4-13 TRC window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
4-14 Ptrach and Paw waveforms (with TRC active). . . . . . . . . . . . . . 4-31
6-1 HAMILTON-C2 screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
6-2 Values window 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
6-3 Values window 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6-4 Values window 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
6-5 CO2 window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
6-6 Graphics window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6-7 Pressure waveform display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
6-8 Trends window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
6-9 Trend display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
6-10 Loop window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
6-11 Loop display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
6-12 Freeze function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-24
7-1 Dynamic Lung panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
7-2 Compliance shown by the Dynamic Lung . . . . . . . . . . . . . . . . . 7-3
xxii 624131/06
7-3 Patient triggering shown by the Dynamic Lung muscle . . . . . . . . 7-4
7-4 Rinsp shown by the bronchial tree of the Dynamic Lung. . . . . . . 7-4
7-5 Vent Status panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
8-1 Visual alarm indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
8-2 Safety ventilation screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
8-3 Ambient state . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
8-4 Alarm buffer with active alarms . . . . . . . . . . . . . . . . . . . . . . . . . 8-8
8-5 Alarm buffer with inactive alarms . . . . . . . . . . . . . . . . . . . . . . . . 8-9
8-6 Events window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
9-1 Special function keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
9-2 Activate Standby window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
9-3 Standby window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
10-1 Removing the filter cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-11
10-2 Removing the air intake filters . . . . . . . . . . . . . . . . . . . . . . . . 10-12
10-3 Removing battery 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-13
10-4 Replacing the oxygen cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-15
A-1 HAMILTON-C2 dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
B-1 Conventional pressure-controlled ventilation in a passive patient.
Flow to patient during inspiration (I); flow from patient during exha-
lation (E) only. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
B-2 Conventional pressure-controlled ventilation in an active patient
when the trigger is off. Pressure increases when the patient tries to exhale (E) and pressure decreases when the patient tries to inspire (I),
as valves are closed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-7
B-3 Biphasic PCV+ in an active patient when trigger is off. The patient can
freely inspire and exhale during any phase of ventilation (+). . . . B-8
B-4 (S)CMV+ basic controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-9
B-5 (S)CMV+ more controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-10
B-6 Breath delivery by the adaptive volume controller. . . . . . . . . . . B-11
B-7 PCV+ basic controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-12
B-8 PCV+ more controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-13
B-9 SPONT basic controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-14
B-10 SPONT more controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-15
B-11 SPONT apnea controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-16
B-12 NIV basic controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-17
B-13 NIV more controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-18
B-14 NIV apnea controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-19
B-15 Breath timing in SIMV+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-20
B-16 SIMV+ basic controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-21
B-17 SIMV+ more controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-22
624131/06 xxiii
List of Figures
B-18 SIMV+ apnea controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-23
B-19 Breath timing in PSIMV+ and NIV-ST . . . . . . . . . . . . . . . . . . . . B-24
B-20 PSIMV+ basic controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-25
B-21 PSIMV+ more controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-26
B-22 NIV-ST basic controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-27
B-23 NIV-ST more controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-28
B-24 Breath timing in nCPAP-PS . . . . . . . . . . . . . . . . . . . . . . . . . . . B-29
B-25 nCPAP-PS basic controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-30
B-26 nCPAP-PS more controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-31
B-27 DuoPAP pressure curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-32
B-28 Pressure support in DuoPAP . . . . . . . . . . . . . . . . . . . . . . . . . . B-33
B-29 DuoPAP basic controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-34
B-30 DuoPAP more controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-35
B-31 DuoPAP apnea controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-36
B-32 APRV breath timing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-37
B-33 APRV basic controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-39
B-34 APRV more controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-40
B-35 APRV apnea controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-41
B-36 Display SAFETY mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-42
C-1 Clinical use of ASV. The numbers in parentheses are step numbers,
which are explained in the next subsections. . . . . . . . . . . . . . . C-4
C-2 ASV basic controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-5
C-3 ASV more controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-6
C-4 Hypothetical example of high %MinVol setting incompatible with the
lung-protective rules strategy. The open circle denotes the actual tar-
get, the closed triangle (never shown on the ventilator) denotes the
(energetically) optimal target according to Otis’ equation. The HAMIL-
TON-C2 will alarm and inform the user that the ASV target cannot be
achieved. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-11
C-5 ASV target graphics window . . . . . . . . . . . . . . . . . . . . . . . . . . C-12
C-6 Normal minute ventilation as a function of ideal body weight (IBW).
For adult patients, minute ventilation is calculated as 0.1 l/kg * IBW
(solid line). For pediatric patients, the value indicated by the dotted
line is used. Minute ventilation for a 15 kg patient thus is calculated as
0.2 l/kg * 15 kg = 3 l/min. . . . . . . . . . . . . . . . . . . . . . . . . . . . C-15
C-7 MinVol = 7 l/min. All possible combinations of Vt and f which result
in a minute ventilation of 7 l/min lie on the bold line. . . . . . . . C-16
C-8 Lung-protective rules strategy to avoid high tidal volumes and pressu-
res (A), low alveolar ventilation (B), dynamic hyperinflation or breath
stacking (C), and apnea (D) . . . . . . . . . . . . . . . . . . . . . . . . . . C-17
xxiv 624131/06
C-9 Anatomy of the ASV target graphics window. The rectangle shows
the safety limits; the circle shows the target breath pattern. . . C-21
C-10 Example of a situation after the three initial breaths. The patient sym-
bol marks the actual measured values for Vt and rate. . . . . . . . C-23
C-11 Lung-protective limits are changed dynamically and according to the
respiratory system mechanics. However, the limits derived from the
operator input are never violated. . . . . . . . . . . . . . . . . . . . . . . C-24
C-12 Changes of target values in broncho-constriction. For clarity, the sa-
fety limits are omitted. For clinical examples, see Belliato 2000. C-26
C-13 Three different relationships between rate and WOB are plotted for a
hypothetical lung: (+) purely resistive load causes WOB to rise with ra­te, (x) purely elastic load creates highest load at low rates, (o) the total lung shows a clear minimum which can be calculated according to
the equation below. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-27
E-1 Typical capnogram of patient on pressure-controlled ventilation, sho-
wing fractional concentration of CO2 plotted against time. . . . . E-3
E-2 Typical spirogram of a patient on pressure-controlled ventilation (sa-
me breath as shown in Figure E-1). . . . . . . . . . . . . . . . . . . . . . . E-4
E-3 Combination of capnogram and spirogram (that is, fractional end-ti-
dal CO2 concentration plotted against volume). . . . . . . . . . . . . E-4
E-4 Interpretation of volumetric capnogram. . . . . . . . . . . . . . . . . . . E-6
G-1 Pneumatic diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-1
H-1 Ventilator parts and accessoires - standard trolley. . . . . . . . . . . . H-1
H-2 Ventilator parts and accessoires - standard trolley. . . . . . . . . . . . H-2
I-1 HAMILTON-C2 connected to a patient monitor . . . . . . . . . . . . . I-3
I-2 HAMILTON-C2 connected to a computer system . . . . . . . . . . . . I-5
I-3 RS-232 connector pinout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-6
I-4 RS-232 cable (PN 157354) wiring diagram . . . . . . . . . . . . . . . . . I-7
I-5 HAMILTON-C2 connected to an external device through the Special
connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-8
I-6 Remote alarm relay positions . . . . . . . . . . . . . . . . . . . . . . . . . . I-10
I-7 Interface connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-11
J-1 Language configuration window . . . . . . . . . . . . . . . . . . . . . . . . J-3
J-2 Units configuration window . . . . . . . . . . . . . . . . . . . . . . . . . . . . J-4
J-3 More configuration window. . . . . . . . . . . . . . . . . . . . . . . . . . . . J-5
J-4 MMP configuration window. . . . . . . . . . . . . . . . . . . . . . . . . . . . J-6
J-5 Default setups configuration window . . . . . . . . . . . . . . . . . . . . . J-7
J-6 Setup > patient configuration window . . . . . . . . . . . . . . . . . . . . J-8
J-7 Mode controls configuration window. . . . . . . . . . . . . . . . . . . . . J-9
J-8 Mode controls configuration window
624131/06 xxv
List of Figures
(Vt/IBW) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . J-9
J-9 Alarms configuration window . . . . . . . . . . . . . . . . . . . . . . . . . . J-10
J-10 Vent Status configuration window . . . . . . . . . . . . . . . . . . . . . . J-10
J-11 Vent Status intelligent panel . . . . . . . . . . . . . . . . . . . . . . . . . . . J-11
J-12 Transfer window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . J-12
J-13 Software options window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . J-13
J-14 Hardware options window . . . . . . . . . . . . . . . . . . . . . . . . . . . . J-14
xxvi 624131/06
List of Tables
1-1 Compatible parts and accessories . . . . . . . . . . . . . . . . . . . . . . . 1-11
1-2 Symbols used on device labels and packaging . . . . . . . . . . . . . 1-22
2-1 Breathing circuit parts according to patient height or IBW . . . . . 2-8
2-2 Tracheal tubes and CO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
3-1 When to perform tests and calibrations . . . . . . . . . . . . . . . . . . . 3-2
4-1 Patient grouping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4-2 Control settings, mode additions and ranges . . . . . . . . . . . . . . 4-15
4-3 Alarm limit settings and ranges . . . . . . . . . . . . . . . . . . . . . . . . 4-27
6-1 Monitored parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
7-1 Dynamic Lung normal values . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
7-2 Vent Status parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
8-1 Alarm indications in HAMILTON-C2 . . . . . . . . . . . . . . . . . . . . . . 8-3
8-2 Alarms and other messages . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11
10-1 Decontamination methods for
HAMILTON-C2 parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
10-2 Preventive maintenance schedule . . . . . . . . . . . . . . . . . . . . . . . 10-9
A-1 Physical characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A-2 Environmental requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
A-3 Pneumatic specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
A-4 Electrical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
A-5 Control setting ranges and resolutions (adult). . . . . . . . . . . . . . . A-5
A-6 Control setting ranges and resolutions (neonatal) . . . . . . . . . . . . A-8
A-7 Controls active in HAMILTON-C2 ventilation modes . . . . . . . . . A-12
A-8 Monitored parameter ranges, resolutions
and accuracies (adult). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-13
A-9 Monitored parameter ranges, resolutions
and accuracies (neonatal) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-19
A-10 Real-time curves and loops (adult) . . . . . . . . . . . . . . . . . . . . . . A-25
A-11 Real-time curves and loops (neonatal). . . . . . . . . . . . . . . . . . . . A-26
A-12 Adjustable alarm ranges and resolutions (adult) . . . . . . . . . . . . A-27
A-13 Adjustable alarm ranges and resolutions (neonatal) . . . . . . . . . A-29
A-14 Configuration specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . A-31
A-15 Ventilator breathing system specifications . . . . . . . . . . . . . . . . A-33
A-16 Other technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-34
A-17 Guidance and manufacturer's declaration – electromagnetic emis-
sions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-37
A-18 Guidance and manufacturer's declaration – electromagnetic immuni-
ty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-38
624131/06 xxvii
List of Tables
A-19 Recommended separation distances between portable and mobile RF
communications equipment and the HAMILTON-C2 ventilatorA-41
B-1 Classification of HAMILTON-C2 ventilation
modesB-3 B-2 Control parameters for initialization of APRVB-38 B-3 Safety mode settingsB-43 C-1 Blood gas and patient conditions and possible adjustments for ASV
C-10 C-2 Interpretation of breathing pattern at 100 % MinVol settingC-13 C-3 Interpretation of breathing pattern at much higher than 100% Min-
Vol settingC-13 C-4 Interpretation of breathing pattern at much lower than 100% MinVol
settingC-14 C-5 ASV technical dataC-28 C-6 Initial breath pattern for Adult settingsC-31 C-7 Initial breath pattern for Pediatric settingsC-31 E-1 Examples of "normal" or expected values in mechanically ventilated
patientsE-7 H-1 Ventilator parts and accessoriesH-3 I-1 Interfacing hardware for patient monitorsI-4 I-2 Requirements for interfacing PDMSsI-5 I-3 Interface connector pin assignmentsI-11
xxviii 624131/06
1
General information
1.1 Introduction 1-2
1.2 Functional description 1-4
1.2.1 System overview 1-4
1.2.2 Gas supply and delivery 1-5
1.2.3 Gas monitoring with the Flow Sensor 1-7
1.3 Physical description 1-9
1.3.1 Breathing circuits and accessories 1-9
1.3.2 Ventilator unit 1-13
1.3.3 Screen 1-20
1.4 Symbols used on device labels and packaging 1-22
624131/06 1-1
1 General information
1.1 Introduction
NOTE:
The infant and neonatal module as well as tube resis­tance compensation (TRC) may not be installed in older devices (sold before August 2010).
The HAMILTON-C2 is designed for intensive care ventilation of adult, pediatric, infant and neonatal patients.
Ventilation modes. This full-functioned intensive care ventila­tor offers a complete range of modes. PCV+, PSIMV+, and SPONT are conventional pressure-controlled modes. (S)CMV+ and SIMV+, delivered by an adaptive volume controller, com­bine the attributes of pressure-controlled with volume-targeted ventilation. DuoPAP and APRV are two related forms of pres­sure ventilation designed to support spontaneous breathing on two alternating levels of CPAP. ASV tion) guarantees that the patient receives the selected minute ventilation with the optimal breath pattern (lowest pressure and volume, optimal rate to minimize work of breathing and intrinsic PEEP). NIV (noninvasive ventilation) and NIV-ST (spon­taneous/timed noninvasive ventilation) provide pressure sup­port ventilation through a mask or other noninvasive interface. nCPAP-PS provides pressure support ventilation through nasal interface for infants and neonates.
Patient-triggered breaths are flow triggered. Monitoring. The HAMILTON-C2 offers a variety of monitoring
capabilities. It displays monitored parameters as numbers. You can also see this data graphically, as a combination of real-time waveforms (curves), Loops, Trends and special Intelligent Pan­els. These Intelligent Panels include the Dynamic Lung, which shows the lung’s activity, and the Vent Status, which indicates the patient’s level of ventilator dependency. The HAMILTON­C2’s monitored data is based on pressure and flow measure­ments collected by the HAMILTON MEDICAL proximal Flow Sensor, between the Y-piece and the patient, and on FiO2 measurements by the integral oxygen monitor.
®
(adaptive support ventila-
1-2 624131/06
Alarms. The HAMILTON-C2’s operator-adjustable and non­adjustable alarms help ensure your patient’s safety.
User interface. The ventilator’s ergonomic design, including a
10.4 in. color touchscreen, a press-and-turn knob, and keys, lets you easily access the ventilator settings and monitored parameters. You can tilt the graphical user interface up to 45°.
Customizability. You can customize the HAMILTON-C2 so that it starts up with institution-defined settings. Per patient group three settings can be predefined.
Power. The HAMILTON-C2 uses as its primary power source AC mains (100 to 240 V AC, 50/60 Hz) or a DC supply (+12 to +24 V). If the primary power source fails, the ventilator power source automatically switches to backup batteries. The stan­dard battery (battery 1) powers the HAMILTON-C2 typically for 3 h, and the optional, hot-swappable battery (battery 2) increases the running time up to 6.5 h.
Mounting variations for the HAMILTON-C2 include a stan­dard trolley, a compact transport solution, and a shelf mount. The trolley has space for oxygen cylinders. With adapter plate for the HAMILTON-C2, the device can be locked up to standard transport trolley.
Nebulization function. The nebulization function lets your HAMILTON-C2 power a pneumatic nebulizer connected to the nebulizer outlet.
The communications interface provides an RS-232 port for connection to a remote monitor, patient data management system (PDMS), or other computer system.
The CO2 sensor continuously monitors airway carbon dioxide and reports etCO2 and inhaled/exhaled CO
for display and
2
alarm purposes.
TRC. Tube resistance compensation Neonatal ventilation. Ventilation of infant and neonates
from a tidal volume of 2 ml.
624131/06 1-3
1 General information
Options1. The following options are available for your HAMILTON-C2:
nCPAP-PS. Is designed to apply nasal continuous positive airway pressure with additional pressure support to infants and neonates.
CO2 and nurse call interface. Gives the possibilty to con­nect CO2 and nurse call on the HAMILTON-C2.
1.2 Functional description
The following paragraphs describe the operation of the HAMILTON-C2 ventilator from a hardware perspective.
1.2.1 System overview
The HAMILTON-C2 is an electronically controlled pneumatic ventilation system with an integrated air compressing system. It is powered by AC or DC with battery backup to protect against power failure or unstable power and to facilitate intra-hospital transport. The HAMILTON-C2’s pneumatics deliver gas, and its electrical systems control pneumatics, monitor alarms, and dis­tribute power.
The user provides inputs to the HAMILTON-C2 microprocessor system through a touchscreen, keys, and a press-and-turn knob. These inputs become instructions for the HAMILTON-C2’s pneumatics to deliver a precisely controlled gas mixture to the patient. The HAMILTON-C2 receives inputs from the proximal Flow Sensor and other sensors within the ventilator. Based on this monitored data, the HAMILTON-C2 adjusts gas delivery to the patient. Monitored data is also dis­played by the graphic user interface.
1. Not all options are available in all markets
1-4 624131/06
The HAMILTON-C2’s microprocessor system controls gas deliv­ery and monitors the patient. The gas delivery and monitoring functions are cross-checked by an alarm controller. This cross­checking helps prevent simultaneous failure of these two main functions and minimizes the possible hazards of software fail­ure.
A comprehensive system of visual and audible alarms helps ensure the patient’s safety. Clinical alarms can indicate an abnormal physiological condition. Technical alarms, triggered by the ventilator’s self-tests, including ongoing background checks, can indicate a hardware or software failure. In the case of some technical alarms, a special safety mode ensures a basic minute ventilation while giving the user time for corrective actions. When a condition is critical enough to possibly com­promise safe ventilation, the HAMILTON-C2 is placed into the ambient state. The ambient and expiratory valves are opened, letting the patient inspire room air through the ambient valve and exhale through the expiratory valve.
The HAMILTON-C2 has several means to ensure that safe patient or respiratory pressures are maintained. The maximum working pressure is ensured by the high pressure alarm limit. If the set high pressure limit is reached, the ventilator cycles into exhalation. The ventilator pressure cannot exceed 60 cmH
O.
2
1.2.2 Gas supply and delivery
The HAMILTON-C2 uses room air and low- or high-pressure oxygen (Figure 1-1). Air enters through a fresh gas intake port and is compressed together with the oxygen by the blower. Oxygen enters through a high
1. High pressure oxygen: Maximal Pressure 600kPa / Maximal Flow 120l/min
2. Low Pressure oxygen: Maximal Pressure 600kPa / Maximal Flow 15 l/min
624131/06 1-5
1
- or low-pressure2 inlet.
1 General information
Figure 1-1. Gas delivery in the HAMILTON-C2
Within the ventilator, the gas enters the HAMILTON-C2’s pneu­matic system. If high-pressure oxygen is supplied, a mixer valve provides for the operator-set concentration. If low-pressure oxygen is supplied, the delivered oxygen concentration is determined by the flow of the source oxygen.
Gas is supplied to the patient via the inspiratory valve. The microprocessor controls the size of the inspiratory valve open­ing and the length of time it is open to meet the user settings.
The HAMILTON-C2 delivers gas to the patient through the inspiratory limb breathing circuit parts, which may include an inspiratory filter, flex tubes, the humidification system, water traps, the Y-piece, and the Flow Sensor. An internal pneumatic nebulizer supplies the nebulizer flow.
1-6 624131/06
Gas exhaled by the patient passes through the expiratory limb breathing circuit parts, including flex tubes, the Flow Sensor, the Y-piece, a water trap, and an expiratory valve cover and membrane. Gas is vented through the expiratory valve cover such that no exhaled gas comes into contact with any internal components of the HAMILTON-C2. Measurements taken at the Flow Sensor are used in the pressure, flow, and volume mea­surements.
An oxygen cell (sensor) monitors the oxygen concentration of the gas to be delivered to the patient. This galvanic cell gener­ates a voltage proportional to the partial pressure of oxygen in the delivered gas. This oxygen measurement is compensated for changes in pressure.
The operations of the inspiratory and expiratory valves are coordinated to maintain system pressure levels.
1.2.3 Gas monitoring with the Flow Sensor
The HAMILTON-C2 accurately measures flow, volume, and pressure in the patient’s airway with the HAMILTON MEDICAL Flow Sensor. This proximal Flow Sensor lets the HAMILTON-C2 sense even weak patient breathing efforts. Between its highly sensitive flow trigger and fast response time, the HAMILTON­C2 helps minimize the patient’s work of breathing.
The Flow Sensor contains a thin, diamond-shaped membrane within the outer housing and has a pressure port on either side. The membrane allows bidirectional flow through its vari­able orifice (Figure 1-2).
624131/06 1-7
1 General information
The area of the orifice changes depending on the flow rate. It opens progressively as the flow increases, creating a pressure drop across the orifice. The pressure difference is measured by a high-precision differential pressure sensor inside the ventila­tor. The pressure difference varies with flow (relationship deter­mined during Flow Sensor calibration), so the patient’s flow is determined from the pressure drop. The HAMILTON-C2 calcu­lates volume from the flow measurements.
The Flow Sensor is highly accurate even in the presence of secretions, moisture, and nebulized medications. The HAMIL­TON-C2 continuously flushes the sensing tubes with mixed gases (rinse flow) to prevent blockage.
Figure 1-2. Flow Sensor
1-8 624131/06
1.3 Physical description
1.3.1 Breathing circuits and accessories
NOTE:
To ensure proper ventilation operation, use only parts and accessories specified in Table 1-1.
Figure 1-3 shows the HAMILTON-C2 with its breathing circuit and accessories. Contact your HAMILTON MEDICAL represen­tative for details on breathing circuits and accessories supplied by HAMILTON MEDICAL. See Table 1-1 for information on other compatible breathing circuits and accessories.
624131/06 1-9
1 General information
Figure 1-3. HAMILTON-C2 with accessories
1 Graphic user interface (GUI) 2 Support arm 3 Breathing circuit 4 HAMILTON-HC humidifier 5 Trolley 6 Breathing circuit connections
1-10 624131/06
Table 1-1. Compatible parts and accessories
Part Use...
Patient breath­ing circuit
• HAMILTON MEDICAL patient breathing circuits
• Other circuits that meet the ventilator breathing system specifications in Appendix A.
Mask • HAMILTON MEDICAL reusable face masks
• Other face or nasal masks, except those incorporating an expiratory valve.
Inspiratory filter • HAMILTON MEDICAL inspiratory bacteria filter
• Other filters that have a 22 mm female conical inlet con­nector and a 22 mm male conical outlet connector, and that meet the ventilator breathing system specifications in Appendix A.
Humidification device
• HAMILTON-HC (HC 180 or HC 200)
• Any active humidifier with a flow capability of up to 120 l/min that is approved for the intended use. Humid­ifiers must comply with EN ISO 9360-1.
• Heat and moisture exchanger (HME). HMEs must comply with EN ISO 9360-1.
Flow Sensor HAMILTON MEDICAL parts only (marked with the HAMIL-
TON "H")
Expiratory valve
HAMILTON MEDICAL parts only membrane and cover
Nebulizer • Internal nebulizer: Pneumatic nebulizer specified for
8 l/min
• External nebulizer: Pneumatic (small-volume) nebulizer powered by an external gas source, or a standalone ultrasonic or electronic (piezo) micropump nebulizer such as the Aerogen
®
Aeroneb® Pro nebulizer system
Oxygen cell HAMILTON MEDICAL parts only
Battery HAMILTON MEDICAL parts only
624131/06 1-11
1 General information
Table 1-1. Compatible parts and accessories
Part Use...
CO2 sensor (Mainstream)
CO2 airway adapter (Main­stream)
CO2 sensor (Sidestream)
CO2 airway adapter (Side­stream)
HAMILTON MEDICAL CAPNOSTAT 5TM sensor
• HAMILTON MEDICAL CO2 airway adapter
• Philips Respironics CAPNOSTAT 5 airway adapter
TM
HAMILTON MEDICAL LoFlow
• HAMILTON MEDICAL sidestream CO2 airway adapter
• Philips Respironics LoFlow airway adapter
sensor
1-12 624131/06
1.3.2 Ventilator unit
Figure 1-4 through Figure 1-6 show the controls, indicators, and other important parts of the ventilator unit.
When a key is pressed and the selected function is active, the LED beside the key is lit.
Figure 1-4. Front view
624131/06 1-13
1 General information
Item Description
1 Touchscreen
2 Alarm lamp. Entire lamp lights when an alarm is active (red =
high-priority alarm, yellow = medium- or low-priority alarm). In addition, a red LED in the middle is continuously lit when alarm silence is active. This red LED flashes when an alarm silence is inactive but an alarm is active.
3 Battery charge indicator. Lights to show that the batteries can
be charged. It is lit whenever the ventilator is connected to AC power or to > 20 V DC, whether or not power is switched on.
4 Power/standby switch. Powers the ventilator on and off
and accesses standby.
To put the ventilator into standby, press and quickly release the switch, then select
Activate Standby
(For details on
standby, see Section 9.1). To switch off ventilator power, press the switch quickly to access
standby, then press the switch again for > 3 s; or, if there is a technical fault, press and hold the switch for > 10 s.
5 Screen lock/unlock key. Prevents inadvertent touchscreen
entries.
6O
enrichment key.
2
Infant/Neonatal option: Delivers 125% of the last oxygen setting for 2 min. The backlit color changes to green and the currently applied oxygen concentration is displayed on the oxygen knob (green). Pushing the key a second time or manually changing the oxygen concentration (FiO2) ends the oxygen enrichment period.
Adults and Pediatric: Delivers 100% oxygen for 2 min. The back­lighting changes color to green and the actually applied oxygen concentration is displayed on the oxygen control (green). Pushing the key a second time or manually changing the oxygen concen­tration (FiO2) ends the 100% oxygen enrichment period.
7 Manual breath/inspiratory hold key. Triggers a mandatory
breath when pressed and released during exhalation. Triggers an inspiratory hold when held down during any breath phase. For details see Section 9.3.
1-14 624131/06
Item Description
8 Nebulizer on/off key. Activates pneumatic nebulizer, during the
inspiration phase if high-pressure oxygen is connected. The indi­cator is lit whenever nebulization is active. Nebulization stops automatically after 30 min. You can switch it off earlier by press­ing the key again. For details, see Section 9.4.
9 Print screen key. Saves a JPG file of the used ventilator screen to
a USB memory key.
10 Alarm silence key. Silences the main ventilator audible alarm for
2 min. Pushing a second time cancels the alarm silence. The red LED beside the key flashes when an alarm is active but not muted. It is continuously lit while the alarm silence is active.
11 Press-and-turn (P&T) knob. Selects and adjusts ventilator set-
tings and selects monitored data. A green ring around the knob is lit when power is switched on.
12 Expiratory valve cover and membrane
13 From patient port. The expiratory limb of the patient breathing
circuit and the expiratory valve are connected here.
14 To patient port. The inspiratory filter and the inspiratory limb of
the patient breathing circuit are connected here.
15 Flow sensor connection. Always attach the blue tube to the
blue connector and the clear tube to the silver connector.
16 Pneumatic nebulizer output connector
17 Oxygen cell with cover
624131/06 1-15
1 General information
Figure 1-5. Rear view
Item Description
1 Serial number label
2 RS-232 connector
3 Fresh air intake and cooling fan vents
4 AC power cord with retaining clip
5 DC power connector
6 AC power receptacle
1-16 624131/06
Item Description
7 Low-flow oxygen connector
8 High-pressure oxygen DISS or NIST inlet fitting
9 Option slot
Figure 1-6. Left side view
624131/06 1-17
1 General information
Item Description
1 Graphical user interface tilt assembly
2 Expiratory valve cover exhaust port
Figure 1-7. Right side view
1-18 624131/06
Item Description
1USB connector. For software update, event log, and configura-
tion setting export and import.
NOTE:
The USB connector is intended for passive memory devices only.
2 Battery door
624131/06 1-19
1 General information
1.3.3 Screen
Directly access all the windows for mode, controls, alarms, and monitoring from the screen during normal ventilation. The default screen is shown (Figure 1-8).
Figure 1-8. Default (basic) screen
Item Description
1 Active mode and patient group.
2 Main controls. The most important controls. Open the
trols
controls.
3 Window buttons (tabs). Open the associated windows.
window via the
Controls
button to show all ventilator
Con-
1-20 624131/06
Item Description
4 Input power. Shows all available power sources. The framed
symbol indicates the current source (AC = mains, DC = DC power supply, 1 = battery 1, 2 = battery 2 (optional). The green part of each battery symbol shows the level of battery charge, while the red shows the level of discharge.
5 Alarm silence countdown. Shows if alarm silence has been
activated. Displays the remaining silence time.
6 Graphic display. Shows the pressure/time waveform (curve)
plus one additional user-selected graphic, including another real-time waveform or an Intelligent Panel.
7 Trigger symbol. Indicates the patient is triggering a breath.
8 Main monitoring parameters (MMP). You can view other
numeric parameters from the monitored parameter windows. If the patient’s condition becomes critical, the colour of the numeric parameters change to red and a high priority alarm appears and to yellow for a medium priority alarm.
9Message bar. Displays alarm messages. If an alarm is active,
view the alarm buffer by touching the message bar. See Section 8 for further information.
10 Maximum Pressure Indication Line
11 Pressure limitation. Maximum Pressure - 10 cmH
O.
2
12 Inactive alarm indicator. Indicates that there is information
about inactive alarms in the alarm buffer. View the alarm buffer by touching the inactive alarm indicator.
624131/06 1-21
1 General information
1.4 Symbols used on device labels and packaging
Table 1-2. Symbols used on device labels and packaging
Symbol Definition
Power on/off switch
Manufacturer
Date of manufacture
Type B applied part (classification of medical electri­cal equipment, type B, as specified by IEC 60601-1)
Type BF applied part (classification of medical elec­trical equipment, type BF, as specified by IEC 60601-1)
Consult operator’s manual. Refer to the operator’s manual for complete information. This label on the device points the user to the operator’s manual for complete information. In the operator’s manual, this symbol cross-references the label.
1-22 624131/06
Table 1-2. Symbols used on device labels and packaging (continued)
Symbol Definition
CE Marking of Conformity, seal of approval guaran­teeing that the device is in conformance with the Council Directive 93/42/EEC concerning medical devices
Indicates the degree of protection against electric shock according to IEC 60601-1. Class II devices have double or reinforced insulation, as they have no provision for protective grounding.
"The TÜV NRTL mark with the indicators “C“ and “US“ means that the product complies with Cana­dian requirements and the requirements of US authorities for safety."
Dispose according to Council Directive 2002/96/EC or WEEE (Waste Electrical and Electronic Equip­ment)
Serial number
This way up at transport and storage
Fragile, handle with care at transport and storage
624131/06 1-23
1 General information
Table 1-2. Symbols used on device labels and packaging (continued)
Symbol Definition
Keep dry at transport and storage
Temperature limitations at transport and storage
Humidity limitations at transport and storage
Atmospheric pressure limitations at transport and storage
Stacking limitations at transport and storage
Recyclable materials
Read the operator’s manual
1-24 624131/06
2
Preparing for ventilation
2.1 Introduction 2-2
2.2 Installing the patient tubing support arm 2-4
2.3 Installing the humidifier 2-5
2.4 Installing the option board 2-5
2.5 Installing the patient breathing circuit 2-6
2.6 Installing the pneumatic nebulizer 2-17
2.7 Setting up for monitoring by the optional CO2 sensor 2-19
2.7.1 CO2 mainstream measurement 2-20
2.7.2 CO2 sidestream measurement 2-24
2.8 Installing the optional Aeroneb Pro nebulizer 2-26
2.9 Using an expiratory filter 2-26
2.10 Connecting to primary power source 2-27
2.10.1Connecting to AC power 2-28
2.10.2Connecting to DC power 2-28
2.11 About the batteries 2-30
2.12 Connecting the oxygen supply 2-32
2.13 Connecting to an external patient monitor or other device 2-33
2.14 Starting up the ventilator 2-34
2.15 Shutting down the ventilator 2-35
2.16 Display navigation guidelines 2-35
624131/06 2-1
2 Preparing for ventilation
2.1 Introduction
WARNING
• Additional equipment connected to medical electrical equipment must comply with the respective IEC or ISO standards. Furthermore, all configurations shall comply with the require­ments for medical electrical systems (see IEC 60601-1-1 or clause 16 of edition 3 of IEC 60601­1, respectively). Anybody connecting additional equipment to medical electrical equipment con­figures a medical system and is therefore responsible that the system complies with the requirements for medical electrical systems. Also be aware that local laws take priority over the above mentioned requirements. If in doubt, consult your local representative or Technical Support.
• To prevent possible patient injury, do not block the holes at the back of the ventilator. These holes are vents for the fresh air intake and the cooling fan.
• To prevent back pressure and possible patient injury, do not attach a spirometer, tube, or other device to the exhaust port of the exhala­tion valve housing.
• To prevent interrupted operation of the ventila­tor or any accessories, use only accessories or cables that are expressly stated in this manual.
• To prevent interrupted operation of the ventila­tor due to electromagnetic interference, avoid using it adjacent to or stacking other devices on it. If adjacent or stacked use is necessary, verify the ventilator’s normal operation in the config­uration in which it will be used.
2-2 624131/06
• To prevent possible per-
sonal injury and equip­ment damage, make sure the ventilator is secured to the trolley or shelf with the quick­locking mechanism.
• To prevent possible
equipment damage, avoid tipping over the ventilator when crossing thresholds.
• To prevent possible
equipment damage, lock the trolley’s wheels when parking the venti­lator.
CAUTION
• Before using the ventilator for the first time,
HAMILTON MEDICAL recommends that you clean its exterior and sterilize its components as described in Section 10.
• To electrically isolate the ventilator circuits from
all poles of the supply mains simultaneously, disconnect the mains plug.
624131/06 2-3
2 Preparing for ventilation
2.2 Installing the patient tubing support arm
WARNING
To prevent possible patient injury due to accidental extubation, check the support arm joints and secure as necessary.
Install the patient tubing support arm on either side of the HAMILTON-C2 trolley. The arm snaps into place.
Figure 2-1. Installing the patient tubing support arm and
humidifier
1 Support arm mount 2 Humidifier slide bracket
2-4 624131/06
2.3 Installing the humidifier
WARNING
• To prevent possible patient injury and possible
water damage to the ventilator, make sure the humidifier is set to appropriate temperature and humidification settings.
• To prevent possible patient injury and equip-
ment damage, do not turn the humidifier on until the gas flow has started and is regulated. Starting the heater or leaving it on without gas flow for prolonged periods may result in heat build-up, causing a bolus of hot air to be deliv­ered to the patient. Circuit tubing may melt under these conditions. Turn the heater power switch off before stopping gas flow.
Install a humidifier to the HAMILTON-C2 using the slide bracket on the trolley column. Prepare the humidifier as described in the manufacturer’s operation manual.
2.4 Installing the option board
Install the option board as follows:
1. Remove the cover of the option slot.
2. Insert the option board containing the interfaces and fix it with the screws.
624131/06 2-5
2 Preparing for ventilation
Figure 2-2. Option board
2.5 Installing the patient breathing circuit
WARNING
• To minimize the risk of bacterial contamination or physical damage, handle bacteria filters with care.
• To prevent patient or ventilator contamination, always use a bacteria filter between the venti­lator and the inspiratory limb of the patient breathing circuit.
• To reduce the risk of fire, use only breathing cir­cuits intended for use in oxygen-enriched envi­ronments. Do not use antistatic or electrically conductive tubing.
2-6 624131/06
NOTE:
• For optimal ventilator operation, use HAMILTON
MEDICAL breathing circuits or other circuits that meet the specifications given in Appendix A. When altering the HAMILTON MEDICAL breathing circuit configurations (for example, when adding accesso­ries or components), make sure not to exceed these inspiratory and expiratory resistance values of the ventilator breathing system, as required by IEC 60601-2-12: neonates, 6 cmH
O at 30 l/min.
2
• Any bacteria filter, HME, or additional accessories in
the expiratory limb may substantially increase flow resistance and impair ventilation.
• To ensure that all breathing circuit connections are
leak-tight, perform the tightness test every time you install a circuit or change a circuit part.
• Regularly check the water traps and the breathing
circuit hoses for water accumulation. Empty as required.
• Do not combine the neonatal CO2 airway adapter
and the Adult Flow sensor. Artefacts during the mea­surement are possible.
624131/06 2-7
2 Preparing for ventilation
1. Install the breathing circuit as follows:
1. Select the correct breathing circuit parts for your patient from Table 2-1 and Table 2-2.
Table 2-1. Breathing circuit parts according to patient
height or IBW
Patient height (cm)
30 to 150 (11 to 59 in.)
> 130 (51 in.)
-- 0.2 to 3 ≤ 5 10 infant/
-- 3 to 30 3 to 7 15
IBW (kg)
3 to 48 3 to 7 15 Pediatric/
> 30 ≥ 5 22
Trach tube ID (mm)
Breathing circuit tube ID (mm)
Flow Sensor
adult
neonatal
Table 2-2. Tracheal tubes and CO2
Trach tube ID (mm) CO2 airway adapter
< 4 Neonatal/pediatric (sidestream)
≥ 4 Pediatric/adult (mainstream)
2. Assemble the patient breathing circuit. Figure 2-3 through Figure 2-7 show six typical circuit configurations; for order­ing information, contact your HAMILTON MEDICAL repre­sentative. Follow the specific guidelines for the different parts.
3. Properly position the breathing circuit after assembly. Make sure the hoses will not be pushed, pulled, or kinked during patient’s movement, nebulization, or other procedures.
2-8 624131/06
Figure 2-3. Patient breathing circuit for use with inspira-
tory heater wire (Pediatric/Adult) - (In place of the flex tube
shown, a 15 x 22 adapter may be used to attach the Flow Sen-
sor to the ET tube.)
624131/06 2-9
2 Preparing for ventilation
Figure 2-4. Patient breathing circuit for use without
heater wires (Pediatric/Adult) - (In place of the flex tube
shown, a 15 x 22 adapter may be used to attach the Flow Sen-
sor to the ET tube.)
2-10 624131/06
Figure 2-5. Patient breathing circuit for use with HME
(Pediatric/Adult) -
(In place of the flex tube shown, a 15 x 22 adapter may be
used to attach the Flow Sensor to the HME or ET tube.)
624131/06 2-11
2 Preparing for ventilation
Figure 2-6. Patient breathing circuit with inspiratory
heater wire for use with nCPAP-PS (non-invasive ventila-
tion neonatal) - (Use a 15M x 15F adapter to connect the
infant flow sensor.)
2-12 624131/06
Figure 2-7. Patient breathing circuit with inspiratory
heater wire (invasive ventilation neonatal) - (Use a 15M x
15F adapter to connect the infant flow sensor.)
624131/06 2-13
2 Preparing for ventilation
Figure 2-8. LiteCircuit (single-limb) patient breathing cir-
cuit (non-invasive ventilation Pediatric/Adult) - (For use
with NIV or NIV-ST)
Expiratory valve membrane: Holding the expiratory valve hous­ing (Figure 2-9) upside-down, seat the silicone membrane onto the housing. The metal plate goes toward the ventilator. Posi­tion the housing and twist clockwise until it locks into place.
2-14 624131/06
Figure 2-9. Installing the expiratory valve
1 Expiratory valve housing 2 Expiratory valve membrane 3 Metal plate toward ventilator
Flow sensor: Insert a Flow Sensor between the Y-piece of the breathing circuit and the patient connection (Figure 2-10). Connect the blue and colorless tubes to the Flow Sensor con­nectors in the front panel. The blue tube goes to the blue con­nector. The colorless tube goes to the silver connector.
624131/06 2-15
2 Preparing for ventilation
NOTE:
To prevent inaccurate Flow Sensor readings, make sure the Flow Sensor is correctly installed:
• The Flow Sensor tubings must not be kinked.
• The Flow Sensor tubings must be secured with clamp
(included with Flow Sensor).
• In the nCPAP-PS mode the correct position of the
flow sensor is at the expiratory valve.
Figure 2-10. Installing the Flow Sensor
2-16 624131/06
2.6 Installing the pneumatic nebulizer
WARNING
• Do not use an expiratory filter or HME in the patient’s breathing circuit during nebulization. Nebulization can cause an expiratory side filter to clog, substantially increasing flow resistance and impairing ventilation.
• Connect the nebulizer in the inspiratory limb per your institution’s policy and procedures. Connecting the nebulizer between the Flow Sensor and the endotracheal tube increases dead space ventilation.
• To prevent the expiratory valve from sticking due to nebulized medications, use only medica­tions approved for nebulization and regularly check and clean or replace the expiratory valve membrane.
• Be aware that nebulization affects delivered oxygen concentration.
NOTE:
• To change the nebulizer settings open the configura­tion mode.
• Nebulization is disabled during the neonatal applica­tion.
The nebulization feature provides a stable driving pressure to power a pneumatic nebulizer connected to the nebulizer out­let, optimally specified for 6 to 7 l/min flow.
Connect the nebulizer and accessories as shown in Figure 2-11. Table 1-1 has information about compatible nebulizers.
624131/06 2-17
2 Preparing for ventilation
Figure 2-11. Installing a pneumatic nebulizer
1 Inspiratory limb 2 Nebulizer 3 Connector 4 Tube
2-18 624131/06
2.7 Setting up for monitoring by the optional CO2 sensor
WARNING
• Always ensure the integrity of the patient breathing circuit after insertion of the airway adapter by verifying a proper CO (capnogram) on the ventilator display.
• If the capnogram appears abnormal, inspect the CO
airway adapter and replace if needed.
2
• Monitor the capnogram for higher-than­expected CO2 levels during ventilation. These can be caused by sensor or patient problems.
• Use the correct adapter. In adult patients small geometrics may induce low tidal volumes and intrinsic PEEP. In neonatal patients large geo­metrics detain effective CO2 removal.
• Do not use the CO2 sensor if it appears to have been damaged or if it fails to operate properly. Refer servicing to HAMILTON MEDICAL autho­rized personnel.
• To reduce the risk of explosion, do not place the CO2 sensor in a combustible or explosive envi­ronment (for example, around flammable anes­thetics or other ignition sources).
• Do not operate the CO
sensor when it is wet or
2
has exterior condensation.
• Avoid permanent direct contact of the CO2 sen­sor with the body.
• Nebulization may influence the CO2 measure­ments.
waveform
2
624131/06 2-19
2 Preparing for ventilation
CAUTION
• Position airway adapters with windows in a vertical, not a horizontal, position. This helps keep patient secretions from pooling on the windows.
• To prevent premature failure of the CO HAMILTON MEDICAL recommends that you remove it from the circuit whenever an aero­solized medication is delivered. This is due to the increased viscosity of the medication, which may contaminate the airway adapter window.
• All devices are not protected against reanima­tion with a defibrillator.
• Disconnect the CO2 sensor before using a defi­brillator to the patient.
CO2 monitoring is used for various applications in order to gain information such as the assessment of the patient’s airway integrity or the proper endotracheal tube placement. For the HAMILTON-C2 two possibilities to monitor CO2 are possible:
• mainstream CO2 measurement
• sidestream CO2 measurement Whether, mainstream or sidestream CO2 is used to monitor
end-tidal CO2 depends on the clinical setting. A volumetric capnogram as described in chapter “Volumetric Capnography” is only possible with a mainstream CO2 sensor.
sensor,
2
2.7.1 CO2 mainstream measurement
WARNING
In NIV and neonatal ventilation with uncuffed tubes leaks may influence the volumetric capno­gram and the deduced numerical monitoring parameters.
2-20 624131/06
2.7.1.1 Introduction
The optional mainstream CO2 sensor is a solid-state infrared sensor, which is attached to an airway adapter that connects to an endotracheal (ET) tube or other airway and measures bases flowing through these breathing circuit components.
The sensor generates infrared light and beams it through the airway adapter or sample cell to a detector on the opposite side. CO airway adapter or aspirated into the sample cell, absorbs some of this infrared energy. The HAMILTON-C2 determines the CO concentration in the breathing gases by measuring the amount of light absorbed by gases flowing through the airway or sam­ple cell.
The HAMILTON-C2 can display measurements derived from the
sensor as numeric values, waveforms, trends, and loops.
CO
2
The waveform is a valuable clinical tool that can be used to assess patient airway integrity and proper endotracheal (ET) tube placement.
The CO C2 ventilator to another, even "on the fly", during ventilation.
from the patient, flowing through the mainstream
2
sensor can be easily transferred from one HAMILTON-
2
2
2.7.1.2 Connecting the CO2 mainstream sensor
Set up the HAMILTON-C2 for CO2 monitoring, as follows:
1. Plug the sensor cable into the CO2 module connector (Fig­ure 2-12), observing the orientation of the indexing guides on the connector body. The cable should snap into place.
2. Attach the airway adapter to the CO2 sensor:
A. Verify that the adapter windows are clean and dry.
Clean or replace the adapter if necessary.
B. Align the arrow on the bottom of the adapter with the
arrow on the bottom of the sensor.
Press the sensor and the adapter together until they click (Figure 2-12).
624131/06 2-21
2 Preparing for ventilation
Figure 2-12. Connecting the CO2 sensor
Figure 2-13. Attaching the CO2 sensor to the airway adapter
1 CO2 sensor 2 Adapter
3. Connect the sensor/airway adapter to the patient circuit as follows (Figure 2-14):
A. Place the sensor/airway adapter assembly at the proxi-
mal end of the airway circuit as shown. Do not place the airway adapter between the ET tube and the elbow, as
2-22 624131/06
this may allow patient secretions to accumulate in the adapter.
B. Position the airway adapter with its windows in a verti-
cal, not a horizontal, position. This helps keep patient secretions from pooling on the windows. If pooling does occur, the airway adapter may be removed from the cir­cuit, rinsed with water and reinserted into the circuit. To prevent moisture from draining into the airway adapter, do not place the airway adapter in a gravity-dependent position.
Figure 2-14. Connecting the CO2 sensor/airway adapter
to the patient circuit
4. Check that connections have been made correctly by verify-
ing the presence of a proper CO
waveform (capnogram)
2
on the HAMILTON-C2 display. Monitor the capnogram for higher-than-expected CO2 levels. If CO2 levels are higher than expected, verify patient condition first. If you deter­mine that the patient’s condition is not contributing, cali­brate the sensor.
5. The sensor cable should face away from the patient. To
secure the sensor cable safely out of the way, attach sensor cable holding clips to the airway tubing, then connect the sensor cable to the clips.
624131/06 2-23
2 Preparing for ventilation
To remove the sensor cable, pull back on the connector sheath and disengage from connector.
2.7.2 CO2 sidestream measurement
2.7.2.1 Introduction
The optional side stream CO2 sensor samples gases over sam­pling port placed into the breathing circuit proximal to the patient. The gas passes through sampling tube to the sample cell. The sampling tube is water permeable in order to mini­mize cross interference effects and collision broadening. The sampling cell measures the gas components are using infrared spectroscopy at a wavelength of 4260 nm. The measured val­ues can be displayed by the HAMILTON-C2 as real-time wave­form, loops, and trends and as numeric values.
2.7.2.2 Connecting the CO2 sidestream sensor
WARNING
• Do not use patients that cannot tolerate the
removal of 50 ml/min from their total minute ventilation.
Set up the device for CO2 sidestream monitoring, as follows:
1. Plug the LoFlo
TM
CO2 Module cable into the CO2 option board connector (yellow), observing the orientation of the indexing guides on the connector body. The cable should snap into place.
2. The sample cell of the sampling kit must be inserted into
TM
the sample cell receptacle of the LoFlo
CO2 Module as shown in the figure below. A “click” will be heard when the sample cell is properly inserted.
2-24 624131/06
Figure 2-15. Inserting the sample cell into the receptacle
3. Inserting the sample cell into the receptacle automatically
starts the sampling pump. Removal of the sample cell turns the sample pump off.
4. Install the airway adapter between Flow Sensor and ET
tube.
5. The sample line should face away from the patient. To
secure the sample line safely out of the way, attach sensor cable holding clips to the airway tubing, then connect the sample line to the clips.
Figure 2-16. Attaching the CO2 sensor to the airway adapter
1 CO2 sensor 2 Adapter
624131/06 2-25
2 Preparing for ventilation
To remove the sampling kit sample cell from the sample cell receptacle, press down on the locking tab and pull the sample cell from the sample cell receptacle.
2.8 Installing the optional Aeroneb Pro nebulizer
The Aerogen Aeroneb Pro nebulizer system is available as an option for the HAMILTON-C2. Attach it to the mounting bracket. Consult the operating instructions supplied with the nebulizer for further installation and operating information.
Figure 2-17. Installing the Aeroneb Pro nebulizer
2.9 Using an expiratory filter
CAUTION
The use of an expiratory filter may lead to a signif­icant increase in expiratory circuit resistance. Excessive expiratory circuit resistance may compro­mise ventilation and increased patient work of breathing and/or AutoPEEP.
2-26 624131/06
NOTE:
Monitored parameters for increased expiratory resis­tance are not specific to the breathing circuit and may indicate increased patient airway resistance and/or increased resistance of the artificial airway (if used). Always check the patient and confirm adequate ventila­tion.
An expiratory filter is not required on the HAMILTON-C2, but you may use one according to your institution’s protocol. An expiratory filter is not required, because the expiratory valve design prevents internal ventilator components from contact with the patient’s exhaled gas.
If you do use an expiratory filter, place it on the patient side of the expiratory valve cover. Remove any expiratory filter or HME during nebulization. Monitor closely for increased expiratory circuit resistance. An
Exhalation obstruction
alarm may
also indicate excessive expiratory circuit resistance. If the
Exhalation obstruction
alarm occurs repeatedly, remove the expiratory filter immediately. If you otherwise suspect increased expiratory circuit resistance, remove the expiratory filter or install a new filter to eliminate it as a potential cause.
2.10 Connecting to primary power source
Either AC or DC can supply the primary power to the HAMIL­TON-C2.
624131/06 2-27
2 Preparing for ventilation
2.10.1 Connecting to AC power
NOTE:
• To prevent unintentional disconnection of the power
cord, make sure it is well seated into the ventilator socket and secured with the power cord retaining clip.
• The HAMILTON-C2 does not require protective earth
grounding, because it is a class II device, as classified according to IEC 60601-1.
Connect the HAMILTON-C2 to an outlet that supplies AC power between 100 and 240 V, 50/60 Hz. Always check the reliability of the AC outlet. The AC power symbol in the bot­tom right-hand corner of the screen is displayed with a frame around it.
2.10.2 Connecting to DC power
CAUTION
Connect the HAMILTON-C2 to the 12 to 24 V DC on­board power circuit of an ambulance vehicle only!
If the HAMILTON-C2 is connected to a DC power source the DC power symbol in the bottom right-hand corner of the screen is displayed with a frame around it.
2-28 624131/06
2.10.2.1 Car adapter
NOTE:
• Only the HAMILTON car adapter is allowed to be use
• Connect the HAMILTON car adapter to the DC jack
• The input power of the car adapter, from 11 to 32 V
• The HAMILTON car adapter is intended to be used
• The use of the device for secondary transport in air-
A car adapter is available in order to connect the HAMILTON­C2 to DC. The car adapter transfers input voltage to 24 V, and allows for charging the batteries.
with the HAMILTON-C2.
of the HAMILTON-C2.
ensures that the batteries 1+2 of HAMILTON-C2 are charged.
during secondary transport.
crafts is not permitted.
Figure 2-18. Car adapter
624131/06 2-29
2 Preparing for ventilation
2.11 About the batteries
NOTE:
• The use of one battery is mandatory. The battery is
used as internal backup battery.
• HAMILTON MEDICAL recommends that the ventila-
tor’s batteries be fully charged before you ventilate a patient. If the batteries are not fully charged and AC power fails, always pay close attention to the level of battery charge.
Two backup batteries, one mandatory and the other optional, protect the HAMILTON-C2 from low, or failure of, the primary power source. When the primary power source (either AC mains or a DC power supply) fails, the ventilator automatically switches to operation on backup battery with no interruption in ventilation. An alarm sounds to signal the switchover. Silence the alarm to confirm notification of the power system change; this resets the alarm. If the optional battery (battery 2) is avail­able and adequately charged, the ventilator switches to this battery first. When battery 2 is depleted or not installed, the ventilator switches to the standard battery (battery 1). The bat­teries power the ventilator until the primary power source is again adequate or until the battery is depleted. Two batteries power the ventilator typically for 6.5 h.
As a further safeguard, the HAMILTON-C2 provides a low bat­tery alarm. It also has a capacitor-driven backup buzzer that sounds continuously for at least 2 min when battery power is completely lost.
The ventilator charges the batteries whenever the ventilator is connected to either AC or DC > 20 V, with or without the ven­tilator power switch on. The battery charge indicator lights show that the batteries are being charged.
2-30 624131/06
Figure 2-19. Power source symbols and battery charge
indicator
1 Battery charge indicator 2 Crossed-out battery 1 means standard battery not available 3 AC mains symbol (or DC) 4 Frame indicates current power source
The power source symbols in the bottom right-hand corner of the screen show the available power sources. A frame around a symbol indicates the current ventilator power source. Green indicates the level of battery charge.
Check the battery charge level before putting the ventilator on a patient and before unplugging the ventilator for transport or other purposes. A green symbol indicates a fully charged bat­tery. A red and green symbol indicates a partially charged bat­tery. If battery symbol 1 is crossed out, the standard battery is discharged or defective. If battery symbol 2 is not shown, the
624131/06 2-31
2 Preparing for ventilation
optional battery is not installed. If a battery is not fully charged, recharge it by connecting the ventilator to the primary power source for a minimum of 4 h, until the battery charge level is 80 to 100%. Alternatively, the battery can also be charged with the external charger.
Section 9.3.2 describes how to replace the batteries.
2.12 Connecting the oxygen supply
CAUTION
• Always check the status of the oxygen cylinders
or other supply before using the ventilator dur­ing transport.
• Make sure oxygen cylinders are equipped with
pressure-reducing valves.
• To minimize the risk of fire, do not use high-
pressure gas hoses that are worn or contami­nated with combustible materials like grease or oil.
NOTE:
• To prevent damage to the ventilator, connect only
clean, dry medical-grade oxygen.
• Before starting ventilation, make sure the appropri-
ate oxygen source, either high-pressure oxygen (
) or low-pressure oxygen (
mode
selected during configuration, see Appendix J.
Oxygen for the HAMILTON-C2 can come from a high- or low­pressure source.
2-32 624131/06
LPO mode
HPO
), was
High-pressure oxygen (Flow: ≤ 120 l/min, Pressure: 2.8 to 6 bar/280 to 600 kPa/41 to 87 psi), provided by a central gas supply or a gas cylinder, is supplied through DISS or NIST male gas fittings. With the optional cylinder holder, you can mount oxygen cylinders to the trolley. If you use gases from cylinders, secure the cylinders to the trolley with the accompanying straps.
Low-flow oxygen (Flow: ≤ 15 l/min, Pressure: ≤ 6 bar/600 kPa/ 87 psi) is provided by a concentrator or liquid cylinder. For information about connecting low-pressure oxygen, see Appendix F.
Connect the oxygen hose to the HAMILTON-C2’s high-pressure or low-flow oxygen inlet fitting, shown in Figure 2-20.
Figure 2-20. Oxygen inlet fittings
1 Oxygen low pressure fitting 2 Oxygen high-pressure inlet fitting
2.13 Connecting to an external patient monitor or other device
NOTE:
All devices connected to the HAMILTON-C2 must be for medical use and meet the requirements of standard IEC 60601-1.
624131/06 2-33
2 Preparing for ventilation
You can connect your ventilator to a patient monitor, a Patient Data Monitoring System (PDMS), or a computer via the RS-232 port. See Appendix I for details on the communications inter­face.
2.14 Starting up the ventilator
CAUTION
To ensure the ventilator’s safe operation, always run the preoperational check before using the ven­tilator on a patient.If the ventilator fails any tests, remove it from clinical use immediately. Do not use the ventilator until necessary repairs are com­pleted and all tests have passed.
NOTE:
If the HAMILTON-C2 is new, be sure it has been properly configured for default language, alarms, and others (see Appendix J).
1. Switch on the ventilator power switch. The ventilator will run a self-test.
Figure 2-21. Power switch
1 Power switch
2-34 624131/06
2. After a short time, you will see the patient setup window.
Set up the ventilator as described in Section 4.2.
3. Run the preoperational check (Section 3.2).
2.15 Shutting down the ventilator
NOTE:
The ventilator remains connected to power when the power switch is switched off. This permits the batteries to charge. To completely disconnect the ventilator from power, unplug it from the mains power outlet or discon­nect it from the DC supply.
To shut the HAMILTON-C2 down, press and quickly release the power switch to access standby, then press the switch again for > 3 s; or, if there is a technical fault, press and hold the switch for > 10 s.
2.16 Display navigation guidelines
Use the touchscreen and the press-and-turn knob to access the HAMILTON-C2 ventilation parameters and monitored data. You typically use a select - activate or select - activate - adjust ­activate procedure.
To open a window, touch the window tab to select and activate it; or turn the knob to select the window tab (it is framed in yellow) and then press the knob to activate your selection.
To close a window, touch the window tab or the X in the upper left-hand corner to select and activate it; or turn the knob to select the X (it is framed in yellow) and then press the knob to activate your selection.
624131/06 2-35
2 Preparing for ventilation
To adjust a control, touch the control to select and activate it; or turn the knob to select the control (it is framed in yellow) and then press the knob to activate your selec­tion. The activated control turns red. Turn the knob to increment or decrement the value. Press the knob or touch the control to confirm the adjustment and deactivate.
To scroll through a log using the scroll bar or arrows, touch the scroll bar to select and
activate it; or turn the knob to select the scroll bar (it is framed in yellow) and then press it to activate your selection. Your selection becomes red when activated. Now turn the knob to scroll through the log. Touch the scroll bar or press the knob to deactivate.
2-36 624131/06
3
Tests, calibrations and utilities
3.1 Introduction 3-2
3.2 Running the preoperational check 3-4
3.3 System functions 3-6
3.3.1 Info: Viewing device-specific information 3-6
3.3.2 Tests & calib: Running sensor calibrations and the tightness test 3-7
3.3.3 Sensors on/off: Enabling/disabling O2 and CO2 monitoring 3-15
3.3.4 Setting day and night 3-16
3.3.5 Setting date and time 3-17
3.4 Utilities 3-18
3.4.1 Configuration: Configuring the ventilator 18
3.4.2 Data transfer: Copying event log data to a USB memory device 3-19
3.5 Alarm tests 3-21
3.5.1 High pressure 3-21
3.5.2 Low minute volume 3-22
3.5.3 Low oxygen alarm 3-22
3.5.4 Disconnection on patient side 3-22
3.5.5 Loss of external power 3-23
3.5.6 Exhalation obstructed 3-23
3.5.7 Apnea 3-23
624131/06 3-1
3 Tests, calibrations and utilities
3.1 Introduction
The tests and calibrations described in this section help verify the safety and reliability of the HAMILTON-C2. Perform the HAMILTON-C2’s tests and calibrations as described in Table 3-1. If a test fails, troubleshoot the ventilator as indicated or have the ventilator serviced. Make sure the tests pass before you return the ventilator to clinical use.
Table 3-1. When to perform tests and calibrations
When to perform Test or calibration
Before placing a new patient on the ventilator
CAUTION
To ensure the ventilator’s safe operation, always run the full preoperational check before using the ven­tilator on a patient. If the ventilator fails any tests, remove it from clinical use immediately. Do not use the ventilator until necessary repairs are completed and all tests have passed.
After installing a new or decontami­nated breathing circuit or component (including a Flow Sensor)
After installing a new oxygen cell or when a related alarm occurs
Preoperational check
Tightness test, Flow Sensor calibration
Oxygen cell calibration
3-2 624131/06
Table 3-1. When to perform tests and calibrations (continued)
When to perform Test or calibration
Required after installing a new, previ­ously unused CO
sensor or when a
2
CO2 sensor/adapter calibration (main-
stream/ sidestream) related alarm occurs, recommended after switching between different air­way adapter types
NOTE:
All calibration data is saved in the sensor head. Therefore, when a previously used sensor is reconnected, you need not recalibrate the sensor unless you have changed the adapter type.
As desired Alarm tests
624131/06 3-3
3 Tests, calibrations and utilities
3.2 Running the preoperational check
CAUTION
To prevent possible patient injury, disconnect the patient from the ventilator before running this test. Make sure another source of ventilatory sup­port is available.
When to perform: Before placing a new patient on the venti-
lator. Required materials: Use the setup below appropriate to your
patient type. To ensure that the ventilator also functions according to specifications on your patient, we recommend that your test circuit be equivalent to the circuit used for venti­lation.
Adult patients
Pediatric patients
Neonatal patients
• Breathing circuit, 22 mm ID with 22F connec­tors
• Flow Sensor, pediatric/adult
• Demonstration lung, 2 l, with adult ET tube between Flow Sensor and lung (PN 281637 or equivalent)
• Breathing circuit, 15 mm ID with 22F connec­tors
• Flow Sensor, pediatric/adult
• Demonstration lung, 0.5 l, with pediatric ET tube between Flow Sensor and lung (PN 151816 or equivalent)
• Breathing circuit, 10 mm ID with 10F connec­tors
• Flow Sensor, infant
• Lung model, neonatal, with neonatal ET tube between Flow Sensor and lung model (An Ing­Mar neonatal lung model is recommended)
3-4 624131/06
Procedure:
Do or observe... Verify... Notes
1. Connect ventilator to AC or DC power
Breathing circuit is
assembled correctly. and oxygen supply. Assemble the patient breathing circuit.
2. Switch on power. When ventilator is switched on, buzzer sounds and the red alarm lamp flashes. After the self-test is passed the alarm lamp flashes red again.
3. Make sure the venti-
lator is in standby, and select
check
Preop
from the
-
Patientsetup
window.
4. Open
System
Tests & calib
->
These tests pass. For details on running
window (Figure 3-2). Select and run the
ness Flow Sensor
Tight-
test, then the
cali­bration. Follow all prompts.
See Figure 2-2 through 2-7.
The buzzer sounds only briefly in the beginning.
these tests and calibra­tions, refer to Section
3.3.2.
These tests pass. See Section 3.3.2.4.
5. If necessary, run
O2
cell calibra-
. Close win-
tion
dow.
624131/06 3-5
3 Tests, calibrations and utilities
Do or observe... Verify... Notes
6. Generate an alarm (for example, by dis­connecting mains power).
7. Resolve the alarm situation (for exam­ple, reconnect mains power).
Corresponding alarm message in message bar (for example,
external power
Alarm is reset.
Loss of
).
During standby, patient alarms are suppressed.
Corrective action: If the ventilator does not pass the preoper­ational check, have it serviced.
3.3 System functions
NOTE:
The audible alarm is silenced during the calibration func­tions and for 30 s thereafter.
You can run tests and calibrations, view device-specific infor­mation, and perform other ventilator system functions from
System
the
3.3.1 Info: Viewing device-specific information
Open the information.
window.
System
->
Info
window to view device-specific
3-6 624131/06
Figure 3-1. Info window
3.3.2 Tests & calib: Running sensor calibrations and the tightness test
NOTE:
To enable or disable O2 and CO2 monitoring see chap­ter 3.3.3.
Open the tests and calibrations.
624131/06 3-7
System
->
Tests&calib
window to access the
3 Tests, calibrations and utilities
Figure 3-2. Tests & calib window
3.3.2.1 Tightness test
NOTE:
• Make sure another source of ventilatory support is
available during this test. The patient must be dis­connected from the ventilator during it.
• To cancel the tightness test while it is in progress,
select
Tightness
again.
3-8 624131/06
Description: This test checks for leakage in the patient breath­ing circuit and determines the circuit’s compliance compensa­tion factor. The ventilator is pressurized to 50 cmH
O. The
2
circuit is considered tight if this pressure can be maintained. If there is a leak, the pressure falls in proportion to the size of leak.
Procedure:
1. Set the ventilator up as for normal ventilation, complete
with the breathing circuit.
2. Activate
Tightness test
from the
dow.
3.
Disconnect patient
is now displayed. Disconnect the breathing circuit at the patient side of the Flow Sensor. Do not block the open end of the Flow Sensor.
Tighten patient system
4.
is now displayed. Block the opening (a finger covered with an alcohol pad may be used).
Connect patient
5.
is now displayed. Reconnect the
patient.
6. VERIFY that there is a green check mark in the box beside
Tightness
.
Corrective action:
Troubleshoot any alarms as described in Section 8.
3.3.2.2 Tightness test with the LiteCircuit
NOTE:
• Make sure another source of ventilatory support is
available during this test. The patient must be dis­connected from the ventilator during it.
• To cancel the tightness test while it is in progress,
select
Tightness
again.
Tests&calib
win-
624131/06 3-9
3 Tests, calibrations and utilities
Description: This test checks for leakage in the patient breath­ing circuit and determines the circuit’s compliance compensa­tion factor. The ventilator is pressurized to 50 cmH circuit is considered tight if this pressure can be maintained. If there is a leak, the pressure falls in proportion to the size of leak.
Procedure:
1. Set the ventilator up as for normal ventilation, complete with the LiteCircuit.
2. Disconnect the Whisper valve together with Flow Sensor from the circuit.
O. The
2
3. Activate
Tightness test
from the
Tests&calib
win-
dow.
4. The message line displays
Patient system
. Block the
opening with a clean gauze-covered finger.
5.
Connect patient
is now displayed. Reconnect the Whis-
per valve with Flow Sensor.
6. Repeat the
Tightness test
as described above (steps 3
to 4).
7. VERIFY that there is a green check mark in the box besides
Tightness
.
8. Reconnect the patient.
3-10 624131/06
3.3.2.3 Flow Sensor calibration
NOTE:
• Make sure another source of ventilatory support is available during this calibration. The patient must be disconnected from the ventilator during it.
• To cancel the Flow Sensor calibration while it is in progress, select
• Circuit resistance compensation measured during calibration.
• If you are using a LiteCircuit, block the opening of the whisper valve with a clean gauze-covered finger.
• Do not turn the infant Flow Sensor during calibra­tion.
• The flow sensor must be on the correct position. Intendíng to ventilate with nCPAP-PS, the flow sen­sor calibration has to be done with the flow sensor at the proximal side of the patient and not at the expi­ration valve of the HAMILTON-C2.
FlowSensor
again.
Description: This calibration checks and resets the calibration points specific to the Flow Sensor in use.
Procedure:
1. Set the ventilator up as for normal ventilation, complete with breathing circuit and Flow Sensor.
2. Activate
Flow Sensor
from the
Tests&calib
window.
3. If you have not already disconnected the patient, the mes­sage line displays
Disconnect patient
. Disconnect the
patient now.
4. Follow the instructions displayed in the message line, turn­ing the Flow Sensor as indicated.
5. VERIFY that the message line displays the green check mark. If the message line displays the red cross, rerun the test. If the second attempt fails, install a new Flow Sensor.
624131/06 3-11
3 Tests, calibrations and utilities
6. Reconnect the patient, as indicated.
Procedure for infant Flow Sensor:
1. Set the ventilator up as for normal ventilation, complete with breathing circuit, Flow Sensor, and expiratory mem­brane and cover. Make sure that the group is selected and that the infant Flow Sensor type is installed.
Neonatal
patient
2. Activate
3.
If you have not already disconnected the patient, the message line displays Disconnect patient
4. VERIFY that the message line displays the green tick. If the message line displays the red cross, rerun the test. If the second attempt fails, install a new Flow Sensor.
5. Reconnect the patient, and close the dow.
Corrective action: If the message bar displays
cal. needed
fails, install a new Flow Sensor.
Flow sensor
. Disconnect the patient now.
, run the calibration again. If the calibration still
from the
3.3.2.4 Oxygen cell calibration
NOTE:
• The oxygen cell calibration can be performed in the
standby mode only.
• The oxygen cell calibration requires that a HAMILTON
MEDICAL oxygen cell be installed and that the venti­lator’s oxygen monitoring be enabled. To check for an oxygen cell, see Section 10. To determine whether oxygen monitoring is enabled, check the
Sensorson/off
• If using the low-pressure-mode disconnect all O2-
supplies during calibration. After reconnecting the oxygen concentration is realised at 21 %.
window.
Tests&calib
Tests & calib
Flow Sensor
System
window.
win-
->
3-12 624131/06
Loading...