1GHz Spectrum Analyzer HMS1000, HMS1010 (with TG)
3GHz Spectrum Analyzer HMS3000, HMS3010 (with TG)
All data valid at 23 °C after 30 minute warm-up
Frequency
Frequency range:
HMS1000, HMS1010 100kHz...1GHz
HMS3000, HMS3010 100kHz…3GHz
Temperature stability: ± 2ppm (0...30°C)
Aging: ± 1ppm/year
Frequency counter:
Resolution 1Hz
Accuracy ± (Frequency x failure of reference)
Span setting range:
HMS1000, HMS1010 0Hz (zero span) and 1kHz…1GHz
HMS3000, HMS3010 0Hz (zero span) and 100Hz…3GHz
Spectral purity, SSB phase noise:
30kHz from carrier
(500MHz, +20°C...30°C) ‹-85dBc/Hz
100kHz from carrier
(500MHz, +20°C...30°C) ‹-100dBc/Hz
1MHz from carrier
(500MHz, +20°C...30°C) ‹-120dBc/Hz
Sweep time:
Span = 0Hz 20ms...100s
Span › 0Hz 20ms...1000s, min. 20ms/600MHz
Resolution bandwidths (-3 dB):
HMS1000, HMS1010 1kHz...1MHz in 1–3 steps, 200kHz
HMS3000, HMS3010 100Hz...1MHz in 1–3 steps, 200kHz
Tolerance:
≤ 300 kHz ± 5% typ.
1MHz ± 10% typ.
Resolution bandwidths (-6 dB):
HMS1000, HMS1010 9kHz, 120kHz,1MHz
HMS3000, HMS3010 200Hz, 9kHz, 120kHz, 1MHz
Video bandwidths: 10Hz...1MHz in 1–3 steps
Amplitude
Display range: Average noise trace up to +20dBm
Amplitude measurement range: Typ. -114dBm...+20dBm
Max. permissible DC at HF input: 80V
Max. power at HF input: 20dBm, 30dBm für max. 3 Min.
Intermodulation free range:
IM3 products, 2 x –20dBm
(-10dBm ref. level) 66dB typ. (typ. +13dBm third-order intercept)
(at distance between
signals ≤ 2MHz) 60dB typ. (+10dBm TOI)
(at distance between
signals › 2MHz) 66dB typ. (typ. +13dBm TOI)
DANL (Displayed average noise level):
(RBW 1kHz, VBW 10Hz,
ref. level ≤ –30dBm
10MHz...1GHz resp. 3GHz) -105dBm, typ. -114dBm
With Preamp. -135dBm typ. (100Hz RBW)
Inherent spurious:
(ref. level≤ -20dBm,
f ‹ 30MHz, RBW≤ 100kHz) ‹ -80dBm
Input related spurious:
(Mixer level ≤ -40dBm,
carrier offset ›1MHz) -70dBc typ., -55dBc (2…3GHz)
2nd harmonic receive frequency
(mixer level -40dBm): -60dBc typ.
Level display:
Reference level -80dBm...+20dBm in 1dB steps
Display range 80dB, 40dB, 20dB, 10dB, linear
Logarithmic display scaling dBm, dBμV, dBmV
Linear display scaling μV, mV, V, nW, μW, mW, W
Measured curves: 1 curve and 1 memory curve
Trace Mathematics: A-B (curve-stored curve), B-A
Detectors: Auto-, Min-, Max-Peak, Sample, RMS,
Average, Quasi-Peak
Failure of level display: ‹ 1,5dB, typ. 0,5dB
(ref. level to ref. level-50dB, 20°C…30°C)
Marker / Deltamarker
Number of marker: 8
Marker functions: Peak, next peak, minimum, center = marker
frequency, reference level = marker level,
all marker on peak
Marker displays: Normal (level), noise marker,
(frequency) counter
Inputs / Outputs
HF Input N socket
Input Impedance: 50Ω
VSWR (10MHz…1GHz/ 3GHz): ‹ 1,5 typ.
Output tracking generator:
(HMS1010/HMS3010) N socket
Output Impedance: 50Ω
Frequency range: 5MHz…1GHz / 3GHz
Output level: -20dBm...0dBm in 1dB steps
Trigger and external
reference input: BNC socket, switchable
Trigger voltage TTL
Reference frequency 10MHz
Essential level (50Ω) 10dBm
Supply output for field probes: 6VDC, max. 100mA (2,5mm DIN jack)
Audio output (Phone): 3,5mm DIN jack
Demodulation AM and FM
Miscellaneous
Display: 6,5" TFT Color VGA Display
Save / Recall memory: 10 complete device settings
Trigger: Free run, Video Trigger, external Trigger
Interfaces: Dual-Interface USB/RS-232 (HO720),
USB-Stick (frontside),
USB-Printer (rear side) from SW 2.0,
DVI for ext. monitor
Power supply: 105/253 V, 50/60 Hz, CAT II
Power consumption: Max. 40Watt at 230V, 50 Hz
Protection class: Safety class I (EN61010-1)
Operating temperature: +5°C...+40°C
Storage temperature: -20°C...+70°C
Max. rel. humidity: 5%...80% (non condensing)
Dimensions (W x H x D): 285 x175x 220 mm
Weight: 3.6 kg
Accessories supplied:
Line cord, Operating manual, Dual-Interface
USB/RS-232 (HO720), CD, HZ21 Adapter plug (N plug to BNC socket)
Optional accessories:
HO3011 Preamplifier -135dBm DANL (100Hz RBW)
HO730 Dual-Interface Ethernet/USB
HO740 Interface IEEE-488 (GPIB), galvanically isolated
HZ547 3GHz VSWR Test Unit for HMS1010, HMS3010
HZ520 Plug-in Antenna with BNC connection
HZ530 Near-Field Probe Set 1GHz for EMV diagnostics
HZ540 Near-Field Probe Set 3GHz for EMV diagnostics
HZ560 Transient limiter
HZ575 75/50Ω Converter
HZ46 4RU 19'' Rackmount Kit
HM3000/3010E/061108/ce · Subject to alterations · © HAMEG Instruments GmbH®· DQS-certified in accordance with DIN EN ISO 9001:2000, Reg.-No.: DE-071040 QM
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 800 0 · Fax +49(0) 6182 800100 · www.hameg.com · info@hameg.com
www.hameg.com