HAMEG HMO3522, HMO3524 User Guide

4GSa /s Real time, 50GSa/s Random sampling,
low noise flash A/D converter (reference class)
2MPts memory per channel, memory oom up to 100,000:1
MSO (Mixed Signal Opt. HO3508) with 8 /16 logic channels
Vertical sensitivity 1mV...5V/div. (into 1MΩ/50Ω)
Offset control ±0,2...±20V
12div. x-axis display range
delayed, event
FFT for spectral analysis
6 digit counter, Autoset, automeasurement, formula editor
6.5” TFT VGA display, DVI output
3 x USB for mass storage, printer and remote ctrl.
optional IEEE-488 or Ethernet/USB
350MHz 2/4 Channel Digital Oscilloscope
HMO3522/HMO3524
HMO3524
1stQuarter
2009
The operating mode XYZ
Quickview provides all avai­lable automeasurements at the push of a button
8 Channel logic probe HO3508
HMO3522[3524]E/031108/ce · Subject to alterations · © HAMEG Instruments GmbH
®
· DQS-certified in accordance with DIN EN ISO 9001:2000, Reg.-No.: DE-071040 QM · www.hameg.com
350 MHz 2 [4] Channel Digital Oscilloscope
HMO3522 [HMO3524]
All data valid at 23 °C after 30 minute warm-up
Display
Display: 6,5” VGA Color TFT Resolution: 640 x 480 Pixel Backlight: LED 400cd/m
2
Display area for curves:
without menu 400 x 600 Pixel (8 x 12 div.)
with menu 400 x 500 Pixel (8 x 10 div.) Color depth: 256 colors Intensity steps per channel: 0…31
Vertical System
Channels:
DSO mode CH1, CH2 [CH1...CH4]
MSO mode CH1, CH2 LCH 0...15 (logic channels)
with 2 x Option HO3508
Auxiliary input: Frontside [Rear side]
Function Ext. Trigger
Impedance 1MΩ || 13pF ± 2pF
Coupling DC, AC
Max. input voltage 100V (DC + peak AC) XYZ-mode: All analog channels on individual choise Invert: CH 1, CH 2 [CH1...CH4] Y-bandwidth (-3dB): 350MHz (5mV...5V) / div.
100MHz (1mV, 2mV)/div.
Lower AC bandwidth: 2Hz Bandwidth limiter (switchable): approx. 20MHz Rise time (measured): ‹ 1ns DC gain accuracy 2% Input sensitivity: 12 calibrated steps
CH1, CH2 [CH1...CH4] 1mV/ div.…5V/div. (1-2-5 Sequence) Variable Between calibrated steps
Inputs CH1, CH2 [CH1…CH4]:
Impedance 1MΩ II 13pF ± 2pF (50Ω switchable)
Coupling DC, AC, GND
Max. input voltage 200V (DC + peak AC), 50Ω ‹5V
eff
Measuring circuits: Measuring Category I (CAT I) Position range ± 10Divs Offset control:
1mV, 2mV ± 0,2V
5mV…50mV ±1V
100mV…5V ±20V Logic channels With Option HO3508
Select. switching thresholds TTL, CMOS, ECL, 2 x User -2V…+8V
Impedance 100kΩ || ‹ 4pF
Coupling DC
Max. input voltage 40V (DC + peak AC)
Triggering
Analog channels: Automatic: Linking of peakdetection and triggerlevel
Min. signal height 0.5 div.
Frequency range 5Hz…400MHz
Level control range From peak- to peak+ Normal (without peak):
Min. signal height 0.5 div.
Frequency range 0…400MHz
Level control range –10div....+10div.
Operating modes: Flanke /Video/Logik /Pulse Slope: Rising, falling, both Sources: CH1, CH2, Line, Ext.
[CH1...CH4, Line, Ext.]
Coupling: AC: 5Hz...400MHz
DC: 0...400MHz HF: 30kHz...400MHz LF: 0...5kHz Noise rejection: 100MHz LPF switchable
Video: Pos. / neg. sync. impulse
Standards 525 Line /60Hz systems
625 Line/50Hz systems
Fields Field 1, field 2, both
Line All, selectable line number
Source CH1, CH2, Ext. [CH1...CH4] Logic: AND, OR, TRUE, FALSE
Source LCH0…15
State LCH0…15 X, H, L Indicator for trigger action: LED Ext. Trigger via: Auxiliary input [Aux. input at rear side] Pre/Post Trigger: -100%...+200% related to 2MPts 2nd Trigger:
Slope Rising or falling
Min. signal height 0.5 div.
Accessories supplied: Line cord, Operating manual, 2 [4] Probes, 10:1 with attenuation ID (HZ350), Dual-Interface USB/RS-232 (HO720), CD
Optional accessories:
HO730 Dual-Interface Ethernet/USB HO740 Interface IEEE-488 (GPIB) galvanically isolated HZ46 4RU 19'' Rackmount Kit HO3508 8 Channel Logic Probe
Frequency range 0…400MHz
Level control range –10div....+10div.
Operating modes: after time 20ns…0.1s after incidence 1…2
16
Horizontal System
Domain representation: Time, Frequency (FFT), Voltage (XY) Representation Time Base: Main-window, main- and zoom-window Memory Zoom: Up to 100.000:1 Accuracy: 20ppm Time Base:
Refresh operating modes 20ms/ div.…1ns/ div. Roll operating modes 50s /div.…50ms/ div.
Digital Storage
Sampling rate (real time): 2x2GSa/s, 1x4GSa/s
[4 x 2GSa/ s, 2 x4GSa/s] Logic channels: 16 x 1GSa/s
Sampling rate (random): 50GSa/ s (n /a to logic channels) Memory: 2 x2MPts [4 x 2MPts] Operation modes: Refresh, Average, Envelope, Peak-Detect
Roll: free run/ triggered
Resolution (vertical) 8Bit Resolution (horizontal)
Yt Mode (50 Pts./div.)
XY Mode 8 Bit Interpolation: Sinx/x (CH1...CH4), Pulse (LCH0...15) Persistence: Off, 50ms...∞ Delay pretrigger: 0...2 Million x (1/samplerate)
posttrigger: 0...4 Million x (1/samplerate) Display refresh rate: Up to 2500 waveforms/ s Display: Dots (acquired points only), vectors (inter-
polation), ‘persistence’
Reference memories: 10
Operation / Measuring / Interfaces
Operation: Menu-driven (multilingual), Autoset,
help functions (multilingual)
Save / Recall memories: 10 complete instrument parameter settings Frequency counter:
0.5Hz...1MHz 5 Digit resolution
› 1MHz…350MHz 6 Digit resolution
Accuracy 20ppm Auto measurements: Frequency, Period, UDC, Upp, Up+, Up-,
U
RMS
, U
Avg
, t
Rise
, t
Fall
Cursor measurements: ΔV, Δt, 1/Δt (f), V to Gnd, Vtrelated to Trigger
point, ratio X and Y, pulse count, peak to peak, peak+, peak-
Interface: Dual-Schnittstelle USB/RS-232 (HO720)
USB-Stick (frontside) USB-Printer (rear side) with SW V2.0 DVI for ext. monitor
Optional: IEEE-488, Ethernet/USB
Mathematic functions
Number of formula sets: 10 formula sets with up to 5 formulas each Sources: All channels, references
and math. memories
Targets: Math. memories Functions: ADD, SUB, 1/X, ABS, MUL, DIV, SQ,
POS, NEG, INV
Display: Up to 4 math. memories
General Information
Probe ADJ Output: 1kHz/1MHz square wave signal ~0.2V
pp
(ta ‹ 4ns)
Line voltage: 105...253V, 50 / 60Hz, CAT II Power consumption: Max. 70Watt at 230V, 50Hz Protective system: Safety class I (EN61010-1) Operating temperature: +5°C...+40°C Storage temperature: -20°C...+70°C Max. rel. humidity: 5%...80% (non condensing) Dimensions (W x H x D): 285x175x220mm Weight: 3.6kg