Hameg HMC8012 User Manual

Page 1
DIGITAL Multimeter HMC8012
User Manual
Benutzerhandbuch / User ManualBenutzerhandbuch / User Manual
Benutzerhandbuch / User Manual
Page 2
35
HMC8012 Brief Introduction
Introduction
The following chapter is an introduction to the most impor­tant HMC8012 multimeter features and settings allowing you to use the measurement instrument promptly. A more detailed description of basic operating steps is given in the manual.
3.1 Setting Parameters
Each function and operating mode of the multimeter can be selected with the keys on the front panel of the instru­ment. Use the respective function keys to select basic func­tions such as voltage, current or resistance measurement. Advanced functions are managed by use of soft menu keys to the right of the display. Pressing the SHIFT key activates the numeric keypad.
3.2 Measuring a Direct/Alternating Voltage
An example for a basic measurement that can be per­formed with a multimeter is the measurement of a DC voltage, for instance. The following measurement example shows the individual steps to perform this measurement effectively with the HMC8012. A DC power supply unit, e.g. the HMP2030, is used.
To measure a DC voltage, you can connect a power supply unit to the HMC8012 multimeter via silicon test leads, for instance. Use the COM and V connectors on the bottom right. Once the DC V key on the front panel of the instru­ment is used, the DC voltage value will be shown on the multimeter display.
The measurement of alternating voltage applies the same principle. In contrast to the DC voltage measurement, the AC voltage measurement requires the use of the AC V key. Please note that for an AC measurement, RMS values will be shown on the display.
3.3 Automatic / Manual Selection of the Measure­ment Range
You can use the soft menu keys Range Up or Range Down to switch the measurement range. Pressing Range Down will switch to the next lower measurement range, pressing Range Up will switch to the next higher measurement range. The automatic selection of the measurement range (Auto Range) is deactivated. If the function Auto Range is activated, the measurement instrument automatically selects the optimal measurement range. If the selected measurement value is too great while the manual selection of the measurement range is activa­ted, the message OVER RANGE appears on the display.
3.4 Displaying Several Measurement Values
If you would like to view a second measurement value on the display, select the soft menu 2nd Function and select a second measurement value via knob. In the current ex­ample, the second measurement value could be AC V. The second measurement value (2nd) will be displayed in blue above the display of the main measurement value (Main). If you select the option None via knob, the second measure­ment value will be deactivated.
3.5 Resistance Measurement with Zero Point
For the resistance measurement, connect measurement cables with clamps to the respective connectors on the front panel of the instrument. A two-wire measurement only requires the COM and V connectors. A four-wire measurement also requires the LO/HI connectors (SENSE). After connecting the cables and connecting with a resis­tance, press the Ω key. The resistance value will show on the display. During a resistance measurement, the cable resistance (cable offset) requires particular attention. Espe­cially for small resistances which may be close to the cable offset it is recommended to zero the measurement section.
A zero measurement is also called a relative measurement, i.e. each measurement value indicates the difference bet­ween a saved zero value and the input signal.
In this case it is necessary to short circuit the measurement cables and then press the NULL key. This triggers the offset correction of the entire measurement section. This value will be shown on the display below the main measurement value. The line resistance of the measurement cable, con­tact resistances and thermoelectric voltages between diffe­rent metals have been intentionally eliminated by this offset correction. The compensation values will not be stored after the HMC8012 has been switched off and have to be redetermined.
3.6 Statistic / Measuring Limits
In addition to the basic functions of the multimeter, ma­thematical functions are also available. These can be ac­cessed by pressing the MEAS key on the front panel of the instrument.
The soft menu Stats allows the user to activate individual statistic functions (Min/Max, Mean, StdDev, Pk to Pk, Count). Use the soft menu key STATS to activate (On) or deactivate (Off) statistic values. The statistic functions will be displayed below the main measurement value. You can use the soft menu key #MEAS. to determine how many measurement values are to be included in the statistic. You can select the number of measurement values via knob or via numeric keypad. If #MEAS is set to „0“, the statistic will be performed on all measurement values. If #MEAS is set to „7“, for instance, the statistic will be performed on 7 measurement values. The soft menu key Reset resets the statistic.
Page 3
36
HMC8012 Brief Introduction
4 Setting Parame-
ters
Each function and operating mode of the multimeter can be selected with the keys on the front panel of the instrument. Use the respective function key to select the measurement function. An active measurement function is highlighted by an illuminated white LED. Subsequent set­tings refer to the selected measurement function.
4.1 Soft Menu Keys
The soft menu keys on the upper right of the screen allow
you to use the shown menu eld in the display. Use the
numeric keypad or the knob to set the respective selec-
ted parameter. If a menu eld has been selected via soft
menu keys, this function will be marked in yellow and will be activated to set the parameter and function. If a spe-
cic setting makes an instrument setting unavailable, the
respective soft menu key will be deactivated and the label will be displayed in gray.
4.2 Numeric Keypad
The easiest way to enter a value precisely and promptly is to use the numeric keypad. Pressing the SHIFT key activa­tes the numeric keypad. When entering a numeric value via keypad, you can accept the entry by pressing a soft
menu key with the corresponding unit. Before conrming
the parameter unit, you can delete any value that has been entered incorrectly by pressing the key
(SHIFT + SETUP key). The ESC key allows you to cancel the operation to en­ter parameters. This will close the editing window.
4.3 Display of Measurement Values
The HMC8012 multimeter includes a brilliant TFT color display with 5-3/4 digits. The display can show up to 3 measurement values simultaneously. The display of the third measurement value is intended for the display of mathematical functions, such as statistic or limit display. The signal displays indicate the measurement units and the operating environment of the multimeter. Figure 4.2 shows an overview of the screen layout.
Fig. 4.1: Numeric keypad and editing keys
The soft menu Limit allows you to select the High Limit and Low Limit for a measurement. Based on these limits, a test can be evaluated as passed or not passed. The limits can be activated (On) or deactivated (Off) via soft menu key LIMITS. The limits will be displayed below the main measurement value. If a value is exceeded or falls
below the set limit during a test, it will be reected by the
color display of the measurement value. A measurement value that exceeds or falls below a limit is marked in red; a measurement value within a limit is marked in green. In addition, you can activate (On) or deactivate (Off) an error tone (soft menu key Beeper) which will sound when the value exceeds or falls below the set limits.
3.7 Storing Data
The HMC8012 multimeter can store two different types of data:
❙ Instrument settings ❙ Screen displays
Out of these data types, screen displays can only be stored on a USB stick. Instrument settings can be stored on a USB stick or internally in the instrument to non-volatile storage media.
Press the key SAVE/RECALL to open the store and load menu. The soft menu Device Settings allows you to load or store instrument settings. Select the submenu Save to store the current instrument settings. Select the storage
location and the le name, then press the soft menu key
Save to save the current instrument settings. This le may be reloaded at a later time.
The menu item Default Settings in the main menu also al­lows you to load the factory default settings.
Page 4
37
Setting Parameters
4.4 Displaying Several Measurement Values
The HMC8012 offers the option (depending on the selected measurement function) to simultaneously show multiple measurement values on the display. The main measure­ment value is labeled as „Main“. A second measurement value can be activated via soft menu key 2nd Function (depending on the measurement function). If you press the soft menu key 2nd Function, you can use the soft menu key SELECT and the knob to select a second measurement value. The second measurement value (2nd) will be dis­played in blue above the display of the main measurement value (Main). If you select the option None via knob, the second measurement value will be deactivated.
4.5 Automatic / Manual Selection of the Measure­ment Range
Depending on the selected measurement value, you can change the measurement range via soft menu keys Range Up or Range Down. Pressing Range Down will switch to the next lower measurement range, pressing Range Up will switch to the next higher measurement range. This will deactivate the automatic selection of the measurement range (Auto Range).
Main (main measure­ment value)
2nd (secondary measurement value)
DC V AC V DC I AC I Fre-
quency
dB dBm
DC V
AC V
DC I
AC I
Tab. 4.1: Displaying Several Measurement Values
If the function Auto Range is activated, the measurement instrument automatically selects the optimal measurement range. Activating the automatic mode enables you to switch to a higher measurement range when 90% of the respective range end value have been reached. Operation will switch to the lower range if the value falls below 10% of the range end value.
If the set measurement value is too large (whether the measure­ment range is selected manually or automatically), the display shows the message OVER RANGE.
Description of the Operating Elements
soft menu keys
Display for math. function (e.g.: statistic)
Range (2nd) Second measurement value
(2nd Function)
Range (Main) Main measurement value Zero measurement (Null)
ADC
Rate Trigger
Local/
Remote
Log-
ging
Inter-
face
Fig. 4.2: Screen layout
Page 5
38
Instrument Functions
5 Instrument
Functions
5.1 Measuring a DC / AC Voltage
A basic measurement that can be performed with a multi­meter is the measurement of a DC or an AC voltage.
To measure a DC or AC voltage, you can connect a voltage source to the HMC8012 multimeter via silicon test leads, for instance. Use the COM and V connectors on the bot­tom right. Once the DC V / AC V key on the front panel of the instrument is used, the DC or AC voltage value will be shown on the multimeter display.
In addition, it is possible to show a second function (2nd) on the display. The soft menu key 2nd Function and the knob allow you to select AC V, DC I, dB or dBm. The se­lected function is shown above the main measurement value display. The menu display changes depending on the selected second function. The soft menu key Remove 2nd allows you to hide the second function from the display.
You can select the option Slow, Medium or Fast via soft menu key AC Filter and knob. This applies to a low pass lter. A lower setting (Slow) ensures more stable measure­ment values. However, the measurement is slow and signal elements of higher frequency will be attenuated. Depen­ding on the input frequency, this function will smoothen the displayed value. The setting Medium is saved as default value. Select the setting Slow for an input frequency of
<50kHz, and select Fast for a setting of >1kHz.
To compensate for the line resistance of measurement cables, it is necessary to short circuit the measurement cables and press the NULL key. This triggers the offset correction of the entire measurement section. This value will be shown next to the soft menu key NULL and on the display below the main measurement value. The line
ESC
ENTER
89
456
123
0
.
--
7
SETUP
SAVE
RECALL
HELP
SHIFT
DC V AC V
CAP
MEAS
AC I
DC I
SENSOR
HOLD
TRIGNULL
max.
600 V
pk
F10H250V
max.
10 A
CAT II 600 V
max. 200 V
pk
max. 1000 V
pk
max.
1000 V
pk
Fused
SENSE
4 W
- +
voltage source
(AC/DC)
Fig. 5.1: Measuring a DC or AC voltage
Please note that for an AC measurement, RMS values will be shown on the display.
resistance of the measurement cable, contact resistances and thermoelectric voltages between different metals have been intentionally eliminated by this offset correction. The compensation values will not be stored after the HMC8012 has been switched off and have to be redetermined. In addition, you may also enter the NULL value numerically or via knob.
The soft menu key ADC Rate and the knob allow you to set the display refresh rate to Slow (5 measurements per second), MEDIUM (10 measurements per second) or FAST (200 measurements per second).
The function Auto Zero on page 2|2 of the DC V main menu may be activated (ON) or deactivated (OFF). If this function is activated, the instrument automatically compen­sates for the connected measurement cables and deducts this offset from the measurement result (automatic zero measurement).
The input impedance can be set via soft menu key Input
Imp. to 10 MOhm or >10 GOhm.
If the second function (2nd Function) is activated (dB or dBm), you have the option to dene a reference value. For the function dB, the reference value is the voltage (Ref. Value). You can set the reference value numerically or via knob. If a reference value is dened, you can reset the pre­viously selected reference value via soft menu key Null.
For the function dBm, the reference value is 1mW and is often used for RF signal measurements. If the dBm func­tion is activated, the multimeter performs a measurement and calculates the power supplied to a comparator resis­tance. The reference value (Ref. Value) in this case is 50,
75 or 600 Ω. In addition, you can set a user-dened value
(User) either numerically or via knob.
5.2 Measuring a DC / AC Current
To measure a DC or AC current, you can connect a power supply unit to the HMC8012 multimeter via silicon test leads, for instance. Use the COM and A connectors on the bottom. Once the DC I / AC I key on the front panel of the instrument is used, the DC or AC current value will be shown on the multimeter display.
In addition, it is possible to show a second function (2nd) on the display. The soft menu key 2nd Function and the knob allow you to select AC I, DC V, dB or dBm. The se­lected function is shown above the main measurement value display. The menu display changes depending on the selected second function. The soft menu key Remove 2nd allows you to hide the second function from the display. To compensate for the line resistance of measurement cables, it is necessary to short circuit the measurement cables and press the NULL key. This triggers the offset
The maximum accuracy in measurements is achieved when using the setting Slow (ADC Rate).
The impedance setting is dependent on the measurement range
Page 6
39
Instrument Functions
correction of the entire measurement section. This value will be shown next to the soft menu key NULL and on the display below the main measurement value. The line resistance of the measurement cable, contact resistances and thermoelectric voltages between different metals have been intentionally eliminated by this offset correction. The compensation values will not be stored after the HMC8012 has been switched off and have to be redetermined. In addition, you may also enter the NULL value numerically or via knob.
The soft menu key ADC Rate and the knob allow you to set the display refresh rate to Slow (5 measurements per se­cond), Medium (10 measurements per second) or Fast (200 measurements per second).
You can select the option Slow, Medium or Fast via soft menu key AC Filter (main function AC I) and knob. This
applies to a low pass lter. A lower setting (Slow) ensures
more stable measurement values. However, the measure­ment is slow and signal elements of higher frequency will be attenuated. Depending on the input frequency, this function will smoothen the displayed value. The setting Medium is saved as default value. Select the setting Slow for an input frequency of
<50 kHz, and select Fast for a set-
ting of
>1 kHz.
If the second function (2nd Function) is activated (dB or dBm), you have the option to dene a reference value. For the function dB, the reference value is the current (Ref. Value). You can set the reference value numerically or via knob. If a reference value is dened, you can reset the pre­viously selected reference value via soft menu key „Null.
For the function dBm, the reference value is 1mW and is often used for RF signal measurements. If the dBm func­tion is activated, the multimeter performs a measurement and calculates the power supplied to a comparator resis­tance. The reference value (Ref. Value) in this case is 50,
75 or 600 Ω. In addition, you can set a user-dened value
(User) either numerically or via knob.
The maximum accuracy in measurements is achieved when using the setting „Slow (ADC Rate).
5.3 Measuring Frequency
The frequency display can be added in the main function AC V and AC I as the second measurement function (2nd).
The measurement time can be set via soft menu key Gate Time. The measurement time can vary between 10ms, 100 ms and 1s and is shown below the frequency informa­tion on the display. The display refresh rate is dependent on the selected ADC rate in the AC V or AC I mode. For the setting SLOW, the measurement rate of the frequency measurement is 1 measurement per second, for MEDIUM it is 10 measurements per second and for FAST it is 100 measurements per second.
5.4 Measuring Resistance
For the resistance measurement, connect measurement cables with clamps to the respective connectors on the front panel of the instrument. A two-wire measurement only requires the COM and V connectors. A four-wire measurement also requires the LO/HI connectors (SENSE). After connecting the cables and connecting with a resis­tance, press the Ω key. The resistance value will show on the display. During a resistance measurement, the cable resistance (cable offset) requires particular attention. Espe­cially for small resistances which may be close to the cable offset it is recommended to zero the measurement section.
A zero measurement is also called a relative measurement, i.e. each measurement value indicates the difference be­tween a saved zero value and the input signal.
ESC
ENTER
89
456
123
0
.
--
7
SETUP
SAVE
RECALL
HELP
SHIFT
DC V AC V
CAP
MEAS
AC I
DC I
SENSOR
HOLD
TRIGNULL
max.
600 V
pk
F10H250V
max.
10 A
CAT II 600 V
max. 200 V
pk
max. 1000 V
pk
max.
1000 V
pk
Fused
SENSE
4 W
+ -
frequency source
Fig. 5.3: Measuring frequency
ESC
ENTER
89
456
123
0
.
--
7
SETUP
SAVE
RECALL
HELP
SHIFT
DC V AC V
CAP
MEAS
AC I
DC I
SENSOR
HOLD
TRIGNULL
max.
600 V
pk
F10H250V
max.
10 A
CAT II
600 V
max. 200 V
pk
max. 1000 V
pk
max.
1000 V
pk
Fused
SENSE
4 W
Fig. 5.4: Measuring resistance (2-wire)
- +
resistance 2-wire
(Ω)
ESC
ENTER
89
456
123
0
.
--
7
SETUP
SAVE
RECALL
HELP
SHIFT
DC V AC V
CAP
MEAS
AC I
DC I
SENSOR
HOLD
TRIGNULL
max.
600 V
pk
F10H250V
max.
10 A
CAT II 600 V
max. 200 V
pk
max. 1000 V
pk
max.
1000 V
pk
Fused
SENSE
4 W
+ -
current source
(AC/DC)
Fig. 5.2: Measuring a DC or AC current
Page 7
40
Instrument Functions
In this case it is necessary to short circuit the measurement cables and then press the NULL key. This triggers the offset correction of the entire measurement section. This value will be shown on the display below the main measurement value. The line resistance of the measurement cable, con­tact resistances and thermoelectric voltages between diffe­rent metals have been intentionally eliminated by this offset correction. The compensation values will not be stored after the HMC8012 has been switched off and have to be redetermined. In addition, you may also enter the NULL value numerically or via knob.
On page 2|2 of the Ω main menu, you can use the soft
menu key Mode to distinguish between 2-wire operation (2w) and 4-wire operation (4w - (4w - connect SENSE cables).
The soft menu key ADC Rate and the knob allow you to set the display refresh rate for the 2-wire operation (2w) to Slow (5 measurements per second), Medium (10 measure­ments per second) or „Fast (50 measurements per second). For the 4-wire operation (4w), you can set the display re­fresh rate to Slow (5 measurements per second), Medium (10 measurements per second) or Fast (25 measurements per second).
5.5 Measuring a Capacity
To measure a capacity, you can connect a capacitor with the HMC8012 multimeter via silicon measurement cables and clamp-type test probes, for instance. Use the COM and V connectors on the bottom. Once the CAP key on the front panel of the instrument is used, the capacity value will be shown on the multimeter display.
To compensate for the line resistance of measurement cables, it is necessary to short circuit the measurement cables and press the NULL key. This triggers the offset correction of the entire measurement section. This value will be shown next to the soft menu key NULL and on the display below the main measurement value. The line resistance of the measurement cable, contact resistances
The maximum accuracy in measurements is achieved when using the setting „Slow (ADC Rate).
and thermoelectric voltages between different metals have been intentionally eliminated by this offset correction. The compensation values will not be stored after the HMC8012 has been switched off and have to be redetermined. In addition, you may also enter the NULL value numerically or via knob.
5.6 Measuring Sensors (PT100 / PT500 / PT1000)
Sensor measurements can be performed with various tem­perature probes. The temperature probes HZ812 (2w) or
HZ887 (4w) are available as optional accessories. Depen­ding on the type used, you can select „2w“ or „4w“ with the respective soft menu key. The soft menu key PT Type allows you to select PT100, PT500 or PT1000 as probe type. The measurement value may be displayed in °C, K or °F (Unit).
To perform precise measurements, it is necessary to com­pensate for the measurement section with the NULL func­tion, especially for the two-wire measurement. PT sensors include a lead resistance which is typically indicated on the data sheet. For an optimally adjusted measurement section, it is necessary to know the exact lead resistance. The measurement rate in this case is 10 measurements per second.
5.7 Continuity / Diode Testing
Pressing the key
opens the main menu for the diode
and continuity testing If the soft menu key is marked in
Fig. 5.7: 2-wire sensor measurement (temperature)
ESC
ENTER
89
456
123
0
.
--
7
SETUP
SAVE
RECALL
HELP
SHIFT
DC V AC V
CAP
MEAS
AC I
DC I
SENSOR
HOLD
TRIGNULL
max.
600 V
pk
F10H250V
max.
10 A
CAT II 600 V
max. 200 V
pk
max. 1000 V
pk
max.
1000 V
pk
Fused
SENSE
4 W
- +
temperature (2-wire)
ESC
ENTER
89
456
123
0
.
--
7
SETUP
SAVE
RECALL
HELP
SHIFT
DC V AC V
CAP
MEAS
AC I
DC I
SENSOR
HOLD
TRIGNULL
max.
600 V
pk
F10H250V
max. 10 A
CAT II 600 V
max. 200 V
pk
max. 1000 V
pk
max.
1000 V
pk
Fused
SENSE
4 W
Fig. 5.5: Measuring resistance (4-wire)
- +
resistance 4-wire
(Ω)
ESC
ENTER
89
456
123
0
.
--
7
SETUP
SAVE
RECALL
HELP
SHIFT
DC V AC V
CAP
MEAS
AC I
DC I
SENSOR
HOLD
TRIGNULL
max.
600 V
pk
F10H250V
max.
10 A
CAT II 600 V
max. 200 V
pk
max. 1000 V
pk
max.
1000 V
pk
Fused
SENSE
4 W
Fig. 5.6: Measuring the capacity
- +
capacity
Page 8
41
Instrument Functions
yellow, the respective function „Diode or Continuity is activated.
For both functions, you can dene / set a so-called thres-
hold via soft menu key „Threshold. In addition, the soft menu key Beep allows you to activate (On) or deactivate (Off) an acoustic signal to be used in case of a threshold violation. For the diode testing, the measurement rate is 10 measurements and for the continuity testing 200 measure­ments per second.
5.8 Measuring with HOLD Function
The HOLD function allows you to maintain the current measurement results on the display. If you press the HOLD key (white LED is illuminated), the measurement result(s) will be frozen on the display. Press the HOLD key again (LED is off) to deactivate the function.
5.9 Trigger Function
Pressing and holding the TRIG key or pressing the SETUP key (submenu TRIGGER) opens the trigger menu. The soft menu key Mode and the knob allow you to select the respective trigger mode. There are three different options to trigger a measurement:
Manual:
If the trigger mode „Manual“ is selected, press the key TRIG to start recording measurement values. The TRIG key blinks while measurement values are captured. Press the TRIG key again to stop the capture of measurement values.
ESC
ENTER
89
456
123
0
.
--
7
SETUP
SAVE
RECALL
HELP
SHIFT
DC V AC V
CAP
MEAS
AC I
DC I
SENSOR
HOLD
TRIGNULL
max.
600 V
pk
F10H250V
max.
10 A
CAT II
600 V
max. 200 V
pk
max. 1000 V
pk
max.
1000 V
pk
Fused
SENSE
4 W
Fig. 5.9: Diode testing
- +
Diode
The TRIG key will stop blinking. The label „Armed“ indica­tes that the instrument is waiting for a trigger (waiting for the TRIG key to be pressed again) and is consequently not triggering at the moment.
Single:
In the trigger mode Single, the soft menu key „Interval“ and the knob allow you to select a measurement interval. The measurement interval describes the time between the recorded measurements. For instance, if the function Inter­val is set to 2, all 2s will be included in the measurements. In addition, the soft menu key Count and the knob enable you to set the number of measurement values to be captu­red. Pressing the TRIG key starts the capture of measure­ment values. For instance, if you set an Interval of 2s and a Count of 5, 5 measurement values will be captured in inter­vals of 2s. During the capture of the measurement values, the LED for the TRIG key is blinking and the label „Trig“ is displayed on the upper right of the screen. Alternatively, the label „Armed“ indicates that the instrument is waiting for a trigger (e.g. waiting for the TRIG key to be pressed again) and is consequently not triggering at the moment.
Auto (Default Setting):
In the trigger mode Auto, the instrument does not wait for a specic event. Instead, it performs permanent measure­ments. The number of measurements is dependent on the selected ADC rate. In addition, you may dene a threshold. The measurement value display changes depending on the selected treshold settings. If the function Continue is activated, the display runs continuously. For the function Above/Below Threshold a measurement value is not cap­tured until the current value falls below or exceeds the set treshold. If the instrument triggers permanently, the label „Trig“ will be displayed on the upper right of the screen.
Fig. 5.10: Trigger menu Auto
Depending on which settings for the time interval and the number
of measurement values to be captured have been conrmed in the
trigger menu, these settings will be synchronized with the logging menu (see chapter 6: Mathematical Functions).
ESC
ENTER
89
456
123
0
.
--
7
SETUP
SAVE
RECALL
HELP
SHIFT
DC V AC V
CAP
MEAS
AC I
DC I
SENSOR
HOLD
TRIGNULL
max.
600 V
pk
F10H250V
max.
10 A
CAT II
600 V
max. 200 V
pk
max. 1000 V
pk
max.
1000 V
pk
Fused
SENSE
4 W
Fig. 5.8: 4-wire sensor measurement (temperature)
- +
temperature (2-wire)
Page 9
42
Mathematical Functions
6 Mathematical
Functions
In addition to the basic functions of the multimeter,
mathematical functions are also available. These can be accessed by pressing the MEAS key on the front panel of the instrument.
Tab. 6.1 describes which mathematical functions can be used with which measurement functions.
6.1 Statistic
The soft menu Stats allows the user to activate the statistic functions (Min/Max, Mean, StdDev, Pk to Pk, Count). Use the soft menu key STATS to activate (On) or deactivate (Off) statistic values. The statistic functions will be dis­played below the main measurement value (Main). You can use the soft menu key #MEAS. to determine how many measurement values are to be included in the statistic. You can select the number of measurement values via knob or via numeric keypad (SHIFT key). The statistic may be performed on a maximum of 50,000 measurement values. If #MEAS is set to „0“, the statistic will be performed on all measurement values. If #MEAS is set to „7“, for instance, the statistic will be performed on 7 measurement values. The soft menu key RESET resets the statistic. Additionally, the soft menu AUTO REST is available. If the Auto Reset function is activated (ON), the statistic measurement points (Count) will be not reset in Auto Range mode.
6.2 Measuring Limits (Limits)
The soft menu LIMIT allows you to select the HIGH LIMIT and LOW LIMIT for a measurement. Based on these li-
If the instrument shows OVER RANGE on the display, the statistic measurement values can be not captured. This values will be dis­played as INVALID COUNT in red.
mits, a test can be evaluated as passed or not passed. The limits can be activated (ON) or deactivated (OFF) via soft menu key LIMITS. The limits will be displayed below the main measurement value (Main). If a value falls below or exceeds the limits during a test, it will be reected by the color display of the measurement value. If the limits are ob­served, the measurement value is displayed in green. If the value falls below or exceeds the set limits, the limit value will be displayed in red. In addition, you can activate (ON) or deactivate (OFF) an error tone (soft menu key BEEPER) which will sound when the value exceeds or falls below the set limits.
6.3 Recording Data (Logging)
The soft menu LOGGING allows you to start the capture of measurement values and to select various settings. Use the soft menu key LOGGING to activate (ON) or deactivate (Off) the capture and the storage of measurement values. Use the soft menu Storage to select the storage location (Internal/USB stick), the le name (File Name) and the le format (CSV/TXT). The soft menu key Interval and the knob allow you to select a measurement interval. The measure­ment interval describes the time between the recorded measurements. For instance, if the function Interval is set to 2, all 2s will be included in the measurements.
The soft menu MODE offers three different logging modes.
Select the function U if you intend to perform an innite
data capture. The limiting factor in this context is the size of the internal storage (maximum of 50,000 measurement points) or of the connected USB stick. If the function N is
activated, the soft menu key Count and the knob enable you to set the number of measurement values to be captu­red. For instance, if you set an interval of 2 s and a count of 5, 5 measurement values will be captured in intervals of 2 s. If the function T is activated, the soft menu key TIME and the knob enable you to set the duration of the capture of the measurement values.
Depending on which settings for the time interval and the number
of measurement values to be captured have been conrmed in the
logging menu, these settings will be synchronized with the trigger menu (see chapter 5: Instrument Functions).
Fig. 6.1: Logging menu setting
Measurement function
Mathematic functions
Stats Limits HOLD NULL
DC V
AC V
DC I
AC I
Ω
CAP
SENSOR
Tab. 6.1: Measurement functions with possible math options
Page 10
43
Documentation, Storage and Recall
Date Format Example for a Logging File
# HAMEG Device under test HM -Log-File ; # Date: 2013-05-27 ; # Start Time:, 14:27:07; # Stop Time:, 14:27:11 ; # Settings: ; # ADC Rate : Slow; # AC Filter : ------; # Input Imp. : 10M; ; ; ; DCV[V],ACV[V],Flag;
15.005159, 0.000186 ;
15.005088, 0.000195 ;
15.004896, 0.000200
;
6.4 Power Display (Power)
The soft menu key Power activates the power display. The function Power is only available if the measurement of DC / AC current is activated. For instance, if you select the func­tion DC V as the main measurement value (Main) and DC I as secondary measurement value (2nd Function) (or vice versa), the power will be shown in the lower display section
after conrming your selection via soft menu key Power.
If the power display is activated, the soft menu key Power is marked in yellow. To deactivate the power display, press the soft menu key Power again.
6.5 Measuring AC+DC (True RMS)
You can show the true RMS value (True RMS) below the main measurement value (Main) on the display via soft menu key AC+DC. This function is only available if the measurement of AC voltage or AC current is activated. Otherwise, the function is grayed out.
7 Documentation,
Storage and Recall
The multimeter HMC8012 enables users to store all screen­shots and user settings. Instrument settings may be saved internally. This data can also be stored on a connected USB stick. Screenshots may only be stored on a USB stick. You can access the main menu to store and load functions by pressing the SAVE/RECALL key.
7.1 Instrument Settings
Use the soft menu Device Settings to save current in­strument settings and to load previously saved settings.
Press the soft menu key Save to open the „Save“ menu. You can use the soft menu key Storage to select a possible location (Internal or Front) where you would like to save the instrument settings. Selecting the respective storage
location and conrming the selection via soft menu key Ac­cept opens the le system manager. The FILE NAME can
be changed or adjusted to the corresponding setting (SET is the default label). You can use the soft menu key Com-
ment to enter a comment which will be displayed in the le manager footer once a le has been selected. Instrument
settings are saved in the HDS format (binary). The format may not be changed. The option Save allows you to store the settings.
To reload stored preference les, press the respective soft menu key to open the soft menu Load. This open the le
manager where you can use the knob to select the respec-
tive le.
Fig. 7.1: Menu for storing instrument settings
Instrument settings from a previous rmware version cannot be loaded with a new rmware version.
Page 11
44
Documentation, Storage and Recall
Once the storage location and the respective settings le has been selected, you can load the le by pressing the soft menu key Load. To remove les that are no longer
required you can use the knob to select the respective set­tings le and remove it by pressing the soft menu key Re­move File. If a USB stick is connected, you can also change or delete directories.
The menu item Default Settings also allows you to load the factory default settings.
7.2 Screenshot
The most important format to store information for docu­mentation purposes is the screenshot. A screenshot is an
image le which shows the current screen content at the
time that storage takes place.
Screenshots may only be stored to a USB stick. If a USB stick is connected, you can also change, create or delete
directories. Press Accept to conrm the location of the
target directory.
The FILE NAME can be changed or adjusted to the corres­ponding setting (SCR is the default label).
The le format of a graphics le determines the color depth and the type of compression. The quality of the various formats is identical for the multimeter graphics. You can
choose from the following le formats in the soft menu
Format:
❙ BMP = Windows Bitmap Format ❙ PNG = Portable Network Graphic
Use the soft menu key Color Mode and the knob to select between Grayscale, Color and Inverted. If Grayscale is selected, the colors are converted to gray scales when the data is stored, if Color is selected, the data is stored as it displays in the screen, and if Inverted is activated, data will be stored in color with a white back-ground.
If you press the key Save, the current screen will be saved immediately to the selected storage location with the selec­ted name and format.
Fig. 7.2: Screenshot menu
The soft menu key Print allows you to print a screenshot immediately to a connected printer (e.g. PCL or PCLX as „printer language“). If a printer is detected, the soft menu key Print will no longer be grayed out. The free software HMScreenshot (software module of the HMExplorer soft­ware) enables the transfer of screenshots in bitmap or PNG format from the HMC8012 via interface to a connected PC where the screenshots may then be saved or printed. For additional information on the software, refer to the internal HMExplorer help at www.hameg.com.
Pressing and holding the HELP key allows you to save a screens­hot to a connected USB stick.
Page 12
45
General Instrument Settings
8 General Instru-
ment Settings
Important general settings, such as basic settings or in­terface settings may be selected via SETUP key. Press the
key to move up one level.
8.1 Update (Instrument Firmware / Help)
Firmware and help (if available) are packed in one ZIP le. Depending on the required extent of the update, the ZIP le
might include either all updates or simply the instrument
rmware, for instance. After downloading the ZIP le, un-
pack the data to the base directory of a USB stick. Then connect the USB stick with the USB port of the multimeter and press the SETUP key. Press the soft menu key Update to access the update menu. Selecting this menu item opens a window displaying version number, date and build
information of the currently installed rmware.
Pressing the soft menu key Firmware to update the instru-
ment rmware will result in a search for the corresponding le on the USB stick. The information for the new rmware
to be installed will then be displayed on the stick below the row labeled NEW:. The version number will be displayed
in red in case the existing rmware on the instrument is
identical to the latest version; otherwise the version num­ber will be shown in green. Only if this is the case, press the soft menu Execute to start the update. To update the help, select the menu item Help in the update menu. The information window will now display all installed langua­ges, the date and relevant information about the languages available on the stick. The soft menu allows you to add, remove or update languages. Please note the date format (YYYY-MM-DD) in compliance with multilingual help ISO 8601 standards.
8.2 Interface Setting
The soft menu Interface enables you to select the settings for:
❙ VCP (Virtual COM Port) ❙ USB
Fig. 8.1: Firmware update menu
❙ Ethernet (IP address, sub-net mask etc.) and ❙ IEEE-488 GPIB interface (GPIB address)
Select the respective interface for the communication via respective soft menu key. Set the required interface para­meters via soft menu item Parameter. Please refer to chap­ter 9 „Remote Control via Interfaces“ for more information about the used interfaces.
8.3 Trigger Settings
See chapter „Instrument Functions“ for a description of the trigger menu.
8.4 Default Settings
The soft menu key „Default Settings“ allows you to load the factory default settings.
8.5 General Settings (Misc)
Device Infos
This soft menu key allows you to retrieve instrument infor­mation such as serial number, software version etc.
Fig. 8.2: Device Infos
Date & Time
The soft menu key Date&Time allows you to set the time and date and add a date and time stamp to printouts and saved data records. The user can reset the date and time. Date and time can be set via knob. The respective soft menu item is activated when it is marked in yellow. Press Save to accept the date and time parameters.
SOUND
The HMC8012 offers the option to issue a signal in the event of an error (or simply as a control measure). This signal can be activated (On) or deactivated (Off) via soft menu key Error Beep or Control Beep.
Display
The soft menu Display and the soft menu key Backlight allows you to set the screen intensity via knob from 10% to 100%.
Key Brightness
The soft menu key „Key Bright“ allows you to set the key brightness via knob from 0% to 100%.
Page 13
46
General Instrument Settings
9 Remote Control
via Interfaces
As a standard feature, the HMC8012 is equipped with a built-in interface card which includes an Ethernet and a USB connection.
9.2 USB
In addition to a LAN interface, the HMC8012 includes a USB device port. For this interface, the user can select if the instrument is accessed via virtual COM port (VCP) or via USB TMC class. The traditional version of the VCP allows the user to communicate with the HMC using any terminal program via SCPI commands once the corres­ponding Windows drivers have been installed. These com­mands are generally compatible with the Agilent multime­ters 34401A and 34410A. In addition, you may use the free HAMEG software „HMExplorer“. This Windows application offers HMC8012 instruments a terminal function, the op­tion to create screenshots and to sort the measured data storage.
A modern alternative to the virtual COM port (VCP) is the control via USB TMC class. TMC stands for „Test & Mea­surement Class“ which indicates that the connected measurement instrument can be recognized without special Windows drivers if VISA drivers are installed and that it can be used directly in corresponding environments. The GPIB interface serves as model to the structure of the
TMC design. A major benet of the USB TMC class is that by sampling specic registers the user can determine if
commands have been terminated and if they have been processed correctly. However, the communication via VCP requires analysis and polling mechanisms within the cont­rolling software which may signicantly strain the interface of the measurement instruments. The TMC status registers solve this problem with the USB TMC in the same manner as is the case with the GPIB interface for the hardware, namely via corresponding control lines.
9.1 Ethernet
In addition to the USB interface, the interface card includes an Ethernet interface. Select Ethernet as interface and press the soft menu key Parameter to then determine the settings for the necessary parameters directly within the
To enable communication, the selected interface and the respec­tive settings in the instrument must be identical to the selections for the PC.
All currently available USB drivers have been fully tested and released for Windows XP™, Windows VISTA™ and Windows 7™ (32 + 64 bit).
Device Name
In this menu, you can select an instrument name. Pressing the soft menu key DEVICE NAME opens a keypad. Use the
knob to select the letters. You can conrm each letter via
ENTER key (SHIFT). Press the soft menu key ACCEPT to
conrm the entered instrument name.
Page 14
47
Remote Control via Interfaces
multimeter. You can specify all parameters and assign a
xed IP address. You can also assign a dynamic IP address
with the activated DHCP function. Please contact your IT
management to congure the settings properly.
If the device has an IP address, it can be accessed via web browser at this IP since the Ethernet interface inclu­des an integrated web server. Enter the IP address in the browser‘s address bar (http://xxx.xxx.xxx.xxx). This will open a window including the instrument type and the serial number.
9.3 IEEE 488.2 / GPIB
In addition to the GPIB functions which are available via USB TMC class, the HMC8012 is optionally available with an integrated GPIB interface. This solution is particularly attractive for customers who already have an existing GPIB environment. With minimum efforts, an old instrument can be replaced by a HMC8012 model.
If DHCP is used and the system cannot assign an IP address to the HMC8012 (for instance, if no Ethernet cable is connected or the network does not support DHCP), it may take up to three minutes until a timeout allows the interface to be congured again.
Fig. 9.2: Web server
Congure the settings in the HMC8012 for all necessary
parameters after you select IEEE488 as interface and press the soft menu key Parameter.
The optional IEEE 488 interface (GPIB) can only be factory-tted as it is necessary for this purpose to open the instrument and break the guarantee seal.
Fig. 9.1: Ethernet-Settings
Page 15
48
Specications
Digital Multimeter HMC8012
All data valid at 23°C ±5K after 90 minutes warm-up and with 5 3/4 digits.
DC Specifications DC Accuracy in ± (% of reading + % of range): Function Range
1)
Test Current Voltage drop
Input Impedance 1 Year 23 °C ± 5 K Temp. Coefficient
0-18 °C, 28-55 °C
DC Voltage 400.000 mV
4.00000 V
40.0000 V
400.000 V
1000.00 V
10 MΩ/>10 GΩ 10 MΩ/>10 GΩ 10 MΩ 10 MΩ 10 MΩ
0.015+0.002
0.015+0.002
0.020+0.002
0.020+0.002
0.025+0.002
0.0010+0.0005
0.0008+0.0003
0.0010+0.0005
0.0015+0.0005
0.0015+0.0005
Resistance (2/4-wire)
2)
400.000 Ω
4.00000 kΩ
40.0000 kΩ
400.000 kΩ
4.00000 MΩ
40.0000 MΩ
250.000 MΩ
1 mA 1 mA 100 µA 10 µA 1 µA 100 nA
460 nA || 10 MΩ (parallel)
0.050+0.005
0.015+0.002
0.015+0.002
0.030+0.003
0.060+0.005 0,250+0.003
2.000+0.010
0.0020+0.0005
0.0020+0.0002
0.0020+0.0002
0.0020+0.0002
0.0020+0.0002
0.0080+0.0005 0,200+0.0005
DC Current
4)
20.0000 mA
200.000 mA
2.00000 A
10.0000 A
3)
<0,30 V <0,27 V <0,2 V <0,60 V
0.05+0.010
0.05+0.010 0,25+0.070 0,25+0.070
0.008+0.0010
0.008+0.0010
0.012+0.0015
0.010+0.0010 Continuity 4000 Ω 1 mA 0.05+0.010 0.005+0.0005 Diode Test 5 V 1 mA 0.05+0.010 0.005+0.0005 Notes: 1) 240.000 / 480.000 counts exept in 1000 V and 10 A range
2) Specifications are for 4-wire measurement; 2-wire measurement using NULL function.
3) Maximum current load at >5 A is 30 seconds, followed by a pause of >30 seconds
4) At 250 V maximum
AC Specifications AC Accuracy in ± (% of reading + % of range): Function Range
1)
Frequency 1 Year 23 °C ± 5 K Temp. Coefficient 0-18 °C, 28-55 °C
AC Voltage
2)
400.000 mV
4.00000 V
40.0000 V
400.000 V
750.00 V
6)
10 Hz–20 Hz 20 Hz–45 Hz 45 Hz–20 kHz 20 kHz–50 kHz 50 kHz–100 kHz
3.0+0.05
1.5+0.05
0.3+0.05
1.0+0.05
3.0+0.05
0.01+0.01
0.01+0.01
0.01+0.01
0.02+0.01
0.05+0.01
AC Current
5)
20.0000 mA
200.000 mA
2.00000 A
10.0000 A
4)
20 Hz–40 Hz 40 Hz–1 kHz 1 kHz–5 kHz 5 kHz–10 kHz
3)
1.5+0.05
0.5+0.05
1.5+0.05
2.5+0.05
0.01+0.01
0.01+0.01
0.01+0.01
0.01+0.01
Notes: 1) 240.000 / 480.000 counts exept in 750 V and 10 A range
2) Specifications are for sinusoidal curves. Input impedance is 1 MΩ parallel <100 pF
3) Except 10 A range
4) Maximum current load at >5 A is 30 seconds, followed by a pause of >30 seconds
5) At 250 V maximum
6) For ACV measurements and frequencies above 50 kHz the user is required to choose an appropriate measurement range
Frequency Counter Specifications Frequency Accuracy in ± (% of reading): Function Range
1)
Frequency 1 Year 23 °C ± 5 K Temp. Coefficient 0-18 °C, 28-55 °C
AC Voltage
2)
all ranges 5 Hz–700 kHz 0.01 0.005
AC Current
2)
20 mA, 200 mA 2 A, 10 A
5 Hz–10 kHz 5 Hz–5 kHz
0.01
0.01
0.005
0.005
Anmerkungen: 1) Display of frequency available as 2nd measurent for main functions ACV and ACI
2) Input sensitivity: >7,5 % of full scale (5 Hz-400 kHz) resp. >20 % (400-700 kHz)
Frequency Counter Resolution Setting Measurement time Display range Resolution
Slow 1 s 999,999 kHz 1 Hz Medium 100 ms 999,99 kHz 10 Hz Fast 10 ms 999,9 kHz 100 Hz
Capacitance Specifications Capacitance Accuracy in ± (% of reading + % of range): Function Range 1 Year 23 °C ± 5 K Temp. Coefficient 0-18 °C, 28-55 °C
Capacitance 5.000 nF
50.00 nF
500.0 nF
5.000 µF
50.00 µF
500.0 µF
2.0+2,5
1.0+2.0
1.0+0,5
1.0+0,5
1.0+0,5
2.0+1.0
0.02+0.002
0.02+0.002
0.02+0.002
0.02+0.002
0.02+0.002
0.02+0.002
10 Specications
Page 16
49
Specications
All data valid at 23°C ±5K after 90 minutes warm-up and with 5 3/4 digits
Reading Rates
Function Setting Resolution Display Reading
Rates (per sec.)
AC Voltage Slow 5 3/4 400.000 5
Medium 4 3/4 40.000 10 Fast 4 3/4 40.000 200
DC Voltage Slow 5 3/4 400.000 5
Medium 4 3/4 40.000 10 Fast 4 3/4 40.000 200
AC Current Slow 5 3/4 200.000 5
Medium 4 3/4 20.000 10 Fast 4 3/4 20.000 200
DC Current Slow 5 3/4 200.000 5
Medium 4 3/4 20.000 10
Fast 4 3/4 20.000 200 Resistance Slow 5 3/4 400.000 5 (2-wire) Medium 4 3/4 40.000 10
Fast 4 3/4 40.000 50 Resistance Slow 5 3/4 400.000 5 (4-wire) Medium 4 3/4 40.000 10
Fast 4 3/4 40.000 25 Frequency Slow 6 999,999 1
Medium 5 99,999 10
Fast 4 9,999 100 Diode 4 3/4 40.000 10 Continuity 4 3/4 40.000 200 Temperature 4 999,9 10
Additional Specications
DC voltage
Measuring Method Sigma Delta analog/digital converter Input Resistance
>10 G Ω (selectable for ranges 400 mV / 4 V 10 MΩ (on all ranges)
CMRR 120 dB for VCM <500 V, 1 kΩ unbalance in the
LO-line and 5 measurements/second
SMRR
>60 dB for 50 or 60 Hz ±0,1% and 5 measurements/second
Input current 60 pA at 25 °C Overload protection 1000 V for all ranges
AC voltage
Measuring Method AC coupled True RMS measurement Input Resistance 1 MΩ parallel <100 pF (on all ranges) Crest Faktor Max. 10 (0,5% additional measurement
uncertainty)
CMRR
>60 dB 1 kΩ in the LO -line and frequency <60 Hz
Overload protection 750 V
rms
(for all ranges)
AC current / DC current
Shunt Resistance 13,75 Ω at 20 mA; 1,25 Ω at 200 mA; 25 mΩ bei
2 A, 10 A
Overload protection Fuse: F10H250V on the front panel
Resistance
Measuring Method 2-wire and 4-wire Overload protection 1000 V for all ranges
Continuity
Measuring Method 1 mA constant current Threshold value Adjustable in 1Ω steps Response Time 200 measurements/second Overload protection 1000 V
Diode Test
Measuring Method 1 mA constant current Threshold value Adjustable in 10 mV steps Response Time 10 measurements/second Overload protection 1000 V
Temperature
Measuring Method Resistance measurement with platinum
sensor
Sensor types PT100, PT500, PT1000 Connection 2-wire, 4-wire Overload protection 1000 V
Math Functions
Statistics Min/Max/Average/Standard deviation Relative measurement NULL key, offset via keyboard Logarithmic functions dB Reference level via keyboard or NULL key
dBm Reference impedance 50/75/600Ω or
freely selectable
Accessories supplied:
Line cord, printed operating manual, HZ15 Silicon test lead with safety connector and test probe, 1m (black + red), Software-CD
Recommended accessories:
HZ812 PT100 Temperature probe 2-wire HZ887 PT100 Temperature probe 4-wire HZC95 19“ rackmount kit 2RU for HMC series
General Specications
Power supply: 115 V / 230 V ±10%, adjustable voltage selector
on rear panel; Frequency 50Hz / 60Hz
Power consumption: 25 W max., 12 W typ. Ambient temperature: 0 …+55 °C (operating); –40 …+70 °C (storage) EMC compliance: DIN EN 61326-1, DIN EN 55011 Safety compliance: DIN EN 61010-1, CAN/CSA-C22.2 No.
61010-1-12
Measuring Category: CAT II, 600 V; CAT I, 1000 V
DC
, 750 VAC
RMS
Dimensions: 222 x 88 x 280 mm Weight: approx. 2,7 kg Warm-up time: 90 minutes
Data logging
Number of measuring counts
Internal: 50.000; external: defined by USB stick capacity.
Rate Log Min.: 5 ms typ. (in accordance to measuring
function and resolution); Max: 3.600 s
Duration Log Internal: 250 s...50.000 h;
External: defined by USB stick capacity
Log Daten Main, 2nd, Time-Stamp Interfaces USB 2.0 (TMC and CDC/VCP), Ethernet
10/100 (LXI in Preparation), IEEE-488/GPIB optional
Programming SCPI, compatible with Agilent 34401A and
34410A
Page 17
50
Appendix
11 Appendix
11.1 List of Figures
Fig. 1.1: Operating positions ......................30
Fig. 1.2: Rear panel of the HMC with voltage selector and
connectors ............................32
Fig. 1.3: Connectors on the front panel of the
instrument .............................32
Fig. 1.4: Product labeling in accordance with
EN 50419 ..............................33
Fig. 2.1: Front panel of the HMC 8012 ..............34
Fig. 2.2: Rear panel of the HMC 8012 ..............34
Fig. 4.1: Numeric keypad and editing keys ..........36
Fig. 4.2: Screen layout ..........................37
Tab. 4.1: Displaying Several Measurement Values .....37
Fig. 5.1: Measuring a DC or AC voltage .............38
Fig. 5.2: Measuring a DC or AC current .............39
Fig. 5.3: Measuring frequency ....................39
Fig. 5.4: Measuring resistance (2-wire) .............39
Fig. 5.5: Measuring resistance (4-wire) .............40
Fig. 5.6: Measuring the capacity ..................40
Fig. 5.7: 2-wire sensor measurement (temperature) ...40 Fig. 5.8: 4-wire sensor measurement (temperature) ...41
Fig. 5.9: Diode testing ...........................41
Fig. 5.10: Trigger menu Auto ......................41
Fig. 6.1: Logging menu setting ....................42
Fig. 7.1: Menu for storing instrument settings ........43
Fig. 7.2: Screenshot menu .......................44
Fig. 8.1: Firmware update menu ..................45
Fig. 8.2: Device Infos ............................45
Fig. 9.1: Ethernet-Settings .......................47
Fig. 9.2: Web server ............................47
11.2 Glossary
A accuracy in measurements: 38, 39, 40 AC Filter: 38, 39, 43 AC voltage: 34, 35, 38, 43 ADC rate: 39, 41, 43 Alternating Voltage: 35 Auto Range: 35
B Bitmap: 44
C cable offset: 35, 39 capacity: 34, 40 continuity testing: 40, 41
D DC voltage: 30, 34, 35 Diode Testing: 40
E Ethernet: 46
F
le format: 42, 44
Firmware: 45 four-wire measurement: 35, 39 Frequency: 32, 39 frequency display: 39
G GPIB: 47 GPIB address: 45 GPIB interface: 45
I input frequency: 38, 39 instrument information: 45 instrument settings: 34, 36, 43 interface: 34, 44, 45, 46, 47
K keypad: 34, 35, 36, 42, 46 knob: 34, 35, 36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46
L limits: 32, 35, 36, 42 Logging: 42, 43 logging menu: 41, 42
M main measurement value: 35, 36, 37, 38, 39, 40, 42, 43 Maintenance: 31 measurement categories: 32 measurement interval: 41, 42 measurement rate: 39, 40, 41 measurement time: 39 Measuring category: 32
O OVER RANGE: 35, 37, 42
P power display: 43 Product Disposal: 33
R reference value: 38, 39 repair: 31 resistance: 31, 34, 35, 38, 39, 40 resistance measurement: 31, 35, 39 RMS: 35, 38, 43 RMS values: 35, 38
S Safety: 30 safety class: 32 screen displays: 36 second measurement value: 35, 37 statistic: 34, 35, 36, 42 statistic functions: 35, 42 store: 36, 43, 44
Page 18
51
Appendix
T treshold settings: 41 trigger mode: 41 two-wire measurement: 35, 39, 40
U USB: 34, 36, 42, 43, 44, 45, 46, 47 USB stick: 36, 42, 43, 44, 45
V VCP: 45, 46 Virtual COM Port: 45 voltage value: 35, 38
W Warranty: 31
Z zero measurement: 35, 38, 39 Zero Point: 35 zero value: 35, 39
Page 19
value-instruments.com
R&S® ist ein eingetragenes Warenzeichen der Rohde & Schwarz GmbH & Co. KG HAMEG Instruments® ist ein eingetragenes Warenzeichen der HAMEG Instruments GmbH Markennamen sind Warenzeichen der Eigentümer 05 / 2014 | © HAMEG Instruments GmbH | 48-8012-C000 In Deutschland gedruckt | Änderungen vorbehalten
www.hameg.com
HAMEG Instruments GmbH Industriestr. 6 | 63533 Mainhausen | Germany | Tel +49 (0) 6182 8000
Loading...