Die HAMEG Instruments GmbH bescheinigt die Konformität für das Produk t
The HAMEG Instruments GmbH herewith declares conformity of the product
HAMEG Instruments GmbH déclare la conformite du produit
HAMEG Instruments GmbH certifi ca la conformidad para el producto
Bezeichnung: Klirrfaktor Messbrücke
Product name: Distortion Meter
Designation: Distorsiomètre
Descripción: Medidor de distorsión
Typ / Type / Type / Tipo: HM80 27
mit / with / avec / con: HM80 01-2
Optionen / Options /
Options / Opciónes: –
mit den folgenden Bestimmungen / with applicable regulations /
avec les directives suivantes / con las siguientes directivas:
EMV Richtlinie 89/ 336/EWG ergänzt durch 91/263 /EWG, 92 /31/EWG
EMC Directive 89/ 336/EEC amended by 91/263/ EWG, 92/ 31/EEC
Directive EMC 89/ 336/CEE amendée par 91/263/EWG, 92/31/ CEE
Directiva EMC 89/ 336/CEE enmendada por 91/263 /CEE, 92 /31/CEE
Niederspannungsrichtlinie 73/23 /EWG ergänzt durch 93/68 /EWG
Low-Voltage Equipment Directive 73/23 /EEC amended by 93 /68/ EEC
Directive des equipements basse tension 73 /23/ CEE amendée par 93/68/CEE
Directiva de equipos de baja tensión 73 /23/ CEE enmendada por 93 /68/ EWG
EN 61000-3 -2/A14: Oberschwingungsströme / Harmonic current emissions
/ Émissions de courant harmonique / emisión de corrientes armónicas:
Klasse / Class / Classe / clase D.
EN 61000-3 -3: Spannungsschwankungen u. Flicker / Voltage fl uctuations
and fl icker / Fluctuations de tension et du fl icker / fl uctuaciones de tensión
y fl icker.
HAMEG instruments fulfi ll the regulations of the EMC directive. The
conformity test made by HAMEG is based on the actual generic- and
product standards. In cases where different limit values are applicable,
HAMEG applies the severer standard. For emission the limits for
residential, commercial and light industry are applied. Regarding the
immunity (susceptibility) the limits for industrial environment have
been used.
The measuring- and data lines of the instrument have much infl uence
on emmission and immunity and therefore on meeting the acceptance
limits. For different applications the lines and/or cables used may
be different. For measurement operation the following hints and
conditions regarding emission and immunity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and
external devices, (computer, printer etc.) suffi ciently screened cables
must be used. Without a special instruction in the manual for a reduced
cable length, the maximum cable length of a dataline must be less than
3 meters and not be used outside buildings. If an interface has several
connectors only one connector must have a connection to a cable.
Basically interconnections must have a double screening. For IEEE-bus
purposes the double screened cables HZ72S and HZ72L from HAMEG
are suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and
instrument should be as short as possible. Without instruction in the
manual for a shorter length, signal lines must be less than 3 meters
and not be used outside buildings.
Signal lines must screened (coaxial cable - RG58/U). A proper ground
connection is required. In combination with signal generators double
screened cables (RG223/U, RG214/U) must be used.
3. Infl uence on measuring instruments.
Under the presence of strong high frequency electric or magnetic fi elds,
even with careful setup of the measuring equipment an infl uence of
such signals is unavoidable.
This will not cause damage or put the instrument out of operation. Small
deviations of the measuring value (reading) exceeding the instruments
specifi cations may result from such conditions in individual cases.
HAMEG Instruments GmbH
G. Hübenett
Product Manager
2
Subject to change without notice
Content
Declaration of Conformity 2
General information regarding the CE-marking 2
Distortion meter HM8027 4
Specifi cations 5
Control elements 6
Operational check 6
Ajustment 7
Circuit and layout diagrams 8
Subject to change without notice
3
HM8027
Distortion Meter
HM8027
Frequency range 20 Hz to 20 kHz
Resolution up to 0.01 %
Low inherent distortion of 0.005%
Automatic frequency fine tuning (capture range 15 %)
Control output for residual distortion measurement
A complete measurement system for AF measurement
equipment in combination with the HM8037
20 Hz - 20 kHz
subdivided into 3 ranges
variable frequency control 10:1, overlapping
ranges
Distortion measurement range
0.01 % - 50 %, subdivided into 2 ranges
Full range:10 % and 100 %
Display resolution:100 % range: 0.1 %
10 % range: 0.01 %
Display accuracy
100 % range:±5 % ±1 digit for k ≤ 10 %
10 % range: ±5 % ±1 digit for k ≤ 1%
Inherent distortion + noise
≤ 0.5 digit
Fundamental rejection
30 dB greater than the measured distortion factor
or ≥ 70 dB in the 100% range
or ≥ 90 dB in the 10% range
Input voltage
min. for 100 % calibration: 300 mV
max. for 100 % calibration:50 V
Input impedance
100 kΩ
Subject to change without notice
5
Control elements
Control elements
1kHz HIGH-PA SS
(Pushbutton switch)
1 kHz high-pass fi lter with a roll off of 12 dB/octave for re-
jection of low fre quency hum and noise.
Display(7-segment LEDs)
3-digit display for indication of the mea sured distortion
factor in %.
FREQUENCY RANGE
(Pushbutton switches)
Selection of frequency range for signal under test (20 Hz-
200 H z, 200 Hz-2 kHz , 2 kHz-20 kHz).
OUTPUT (BNC connector)
Monitor output for distortion factor (Residual distortion).
Output voltage is 1 mV/digit.
TUNING INDICATOR (LEDs)
If the built-in fi lter is incorrectly tuned, one of the two LEDs
will indicate in which direction the fi lter frequency de viates
from the input frequency. Turn the tuning knob in the opposite direc tion until the LED goes out.
TUNING (Adjusting knob)
Permits tuning of the built-in fi lter for maximum rejection
of the fundamental wave. Fine tuning is automatic with a
capture range of approx. 15%. If both LEDs
are off, the
fi lter is properly synchronised.
DISTORTION (Pushbutton switches)
Range selection for 10% or 100% fuII scale.
100%-CALIBRATION
(Pushbutton switch)
Selection of calibration mode. Adjust ment for 100% reading
with LEVEL
.
LEVEL (Adjusting knob)
Continuous attenuation of input signal up to max. 15 dB to
achieve 100% reading when in the calibration mode.
INPUT (BNC connector)
Input for measurement signal. The permissible input vol-
tage range is 0.3 V - 50 V for a valid measurement.
ATTENUATOR (Pushbutton switch)
Input signal attenuation with two push button switches of
20 dB or 10 dB attenuation, respectively. They can be used
separately. Both pushbutton swi t ches activated, together
with the variable attenuator
mu st enab le a 10 0% read ing
when in the calibration mode, otherwise the input voltage
should be adjusted.
Operational check
Measuring equipment required
– 2 Low Distortion Generators (e.g. HM8037)
– Digital multimeter with 200 mV
Check of input sensitivity
Setting:
2k Cal. max. 0dB released
1) Adjust the output voltage of the HM8037 to 300 mV at
1 kHz.
2) Connect the output of the HM8037 to the input
HM8027.
3) With the above setting the display
indicate a value
≥
100.
Check of input signal attenuation
Setting: 2k Cal. max. odB released
1) Connect the output of the HM8037 to the in put
HM8027.