HAMEG HM604-3 User Manual

Oscilloscope data sheet with technical details . .... 4
Operating Instructions
General Information ................................................. 5
Symbols ................................................................... 5
Use of tilt handle .....................................................5
Safety ....................................................................... 5
Operating conditions ............................................... 5
EMC .........................................................................6
Warranty .................................................................. 6
Maintenance ............................................................ 6
Protective Switch-Off .............................................. 6
Power supply ........................................................... 6
Type of signal voltage..............................................7
Amplitude Measurements ....................................... 7
Total value of input voltage .....................................8
Time Measurements ............................................... 8
Connection of Test Signal........................................9
First Time Operation..............................................1 0
Trace Rotation TR ..................................................10
Probe compensation and use ...............................11
Adjustment at 1kHz ...............................................11
Adjustment at 1MHz ............................................. 11
Operating modes of the vertical
amplifiers in Yt mode. ...........................................12
X-Y Operation .........................................................1 2
Phase comparison with Lissajous figures ............ 1 3
Phase difference measurement
in DUAL mode .......................................................13
Measurement of an............................................... 13
amplitude modulation ............................................1 3
Triggering and time base ....................................... 14
Automatic Peak-Triggering .................................... 14
Normal Triggering ..................................................15
Slope ...................................................................... 1 5
Trigger coupling ...................................................... 15
Line triggering (~) .................................................. 16
Alternate triggering................................................ 16
External triggering ................................................. 1 6
Trigger indicator ..................................................... 16
Holdoff-time adjustment ....................................... 16
Delay / After Delay Triggering ...............................17
AUTO SET ..............................................................18
SAVE/RECALL........................................................19
Component Tester ................................................. 1 9
Using the Component Tester ................................ 19
Test Procedure ....................................................... 19
Test Pattern Displays............................................. 19
Testing Resistors ................................................... 20
Testing Capacitors and Inductors..........................20
Testing Semiconductors ........................................2 0
Testing Diodes ....................................................... 20
St.1196-Hüb/Ros
Testing Transistors ................................................. 2 0
In-Circuit Tests ....................................................... 20
Table of contents
Oscilloscope
HM604-3
Test Instructions
General ...................................................................23
Symmetry and Drift of the Vertical Amplifier ....... 23
Calibration of the Vertical Amplifier ....................... 23
Transmission Performance of the
Vertical Amplifier.................................................... 2 3
Triggering Checks .................................................. 24
Timebase................................................................ 2 4
Holdoff time ...........................................................25
Component Tester ................................................. 2 5
Trace Alignment .....................................................25
Service Instructions
General ................................................................... 25
Instrument Case Removal .....................................25
Operating Voltages ................................................ 26
Maximum and Minimum Brightness .................... 26
Astigmatism control .............................................. 26
Trigger Threshold ................................................... 26
Trouble-Shooting the Instrument .......................... 26
Adjustments ........................................................... 26
RS232 Interface - Remote Control .......................27
Baud-Rate Setting ..................................................27
Data Communication ............................................. 27
Command definition .............................................. 27
Command Chart: ................................................... 27
Instrument Data Field with Single Commands ....28
Short instruction for HM304 ................................. 29
Switching on and initial setting ............................. 2 9
Vertical amplifier mode.......................................... 29
Triggering mode .....................................................29
Measurements ...................................................... 2 9
Component tester mode.......................................29
Front Panel Elements HM604-
(Brief Description - Front View) ............................. 30
3
Subject to change without notice
Printed in Germany
1
General information regarding the CE marking
HAMEG instruments fulfill the regulations of the EMC directive. The conformity test made by HAMEG is based on the actual generic- and product standards. In cases where different limit values are applicable, HAMEG applies the severer standard. For emission the limits for residential, commercial and light industry are applied. Regarding the immunity (susceptibility) the limits for industrial environment have been used.
The measuring- and data lines of the instrument have much influence on emmission and immunity and therefore on meeting the acceptance limits. For different applications the lines and/or cables used may be different. For measurement operation the following hints and conditions regarding emission and immunity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and external devices, (computer, printer etc.) sufficiently screened cables must be used. Without a special instruction in the manual for a reduced cable length, the maximum cable length of a dataline must be less than 3 meters long. If an interface has several connectors only one connector must have a connection to a cable.
Basically interconnections must have a double screening. For IEEE-bus purposes the double screened cables HZ72S and HZ72L from HAMEG are suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and instrument should be as short as possible. Without instruction in the manual for a shorter length, signal lines must be less than 3 meters long.
Signal lines must screened (coaxial cable - RG58/U). A proper ground connection is required. In combination with signal generators double screened cables (RG223/U, RG214/U) must be used.
3. Influence on measuring instruments.
Under the presence of strong high frequency electric or magnetic fields, even with careful setup of the measuring equipment an influence of such signals is unavoidable. This will not cause damage or put the instrument out of operation. Small deviations of the measuring value (reading) exceeding the instruments specifications may result from such conditions in individual cases.
December 1995
HAMEG GmbH
KONFORMITÄTSERKLÄRUNG DECLARATION OF CONFORMITY DECLARATION DE CONFORMITE
Name und Adresse des Herstellers HAMEG GmbH Manufacturer´s name and address Kelsterbacherstraße 15-19 Nom et adresse du fabricant D - 60528 Frankfurt
HAMEG S.a.r.l. 5, av de la République F - 94800 Villejuif
Die HAMEG GmbH / HAMEG S.a.r.l bescheinigt die Konformität für das Produkt The HAMEG GmbH / HAMEG S.a.r.l herewith declares conformity of the product HAMEG GmbH / HAMEG S.a.r.l déclare la conformite du produit
®
Instruments
Bezeichnung / Product name / Designation: Typ / Type / Type: mit / with / avec: Optionen / Options / Options:
mit den folgenden Bestimmungen / with applicable regulations / avec les directives suivantes
EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE
Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE
Angewendete harmonisierte Normen / Harmonized standards applied / Normes harmonisées utilisées
Sicherheit / Safety / Sécurité
EN 61010-1: 1993 / IEC (CEI) 1010-1: 1990 A 1: 1992 / VDE 0411: 1994 Überspannungskategorie / Overvoltage category / Catégorie de surtension: II Verschmutzungsgrad / Degree of pollution / Degré de pollution: 2
Elektromagnetische Verträglichkeit / Electromagnetic compatibility / Compatibilité électromagnétique
Oszilloskop/Oscilloscope/Oscilloscope
HM604-3
-
-
EN 50082-2: 1995 / VDE 0839 T82-2
ENV 50140: 1993 / IEC (CEI) 1004-4-3: 1995 / VDE 0847 T3 ENV 50141: 1993 / IEC (CEI) 1000-4-6 / VDE 0843 / 6
EN 61000-4-2: 1995 / IEC (CEI) 1000-4-2: 1995 / VDE 0847 T4-2: Prüfschärfe / Level / Niveau = 2 EN 61000-4-4: 1995 / IEC (CEI) 1000-4-4: 1995 / VDE 0847 T4-4: Prüfschärfe / Level / Niveau = 3 EN 50081-1: 1992 / EN 55011: 1991 / CISPR11: 1991 / VDE0875 T11: 1992
Gruppe / group / groupe = 1, Klasse / Class / Classe = B
Datum /Date /Date Unterschrift / Signature /Signatur
14.12.1995
Dr. J. Herzog
Technical Manager
Directeur Technique
Specifications
Vertical Deflection
Operating modes: Channel I or II separate,
Channel I and II: alternate or chopped. (0.5MHz chopper frequency, approx.) Sum or difference with Ch. I and Ch. II (both channels invertable).
XY-Mode: via channel I and channel II Frequency range: 2xDC to 60MHz (-3dB)
Risetime: <5.9ns. Overshoot max. 1%. Deflection coefficients: 14 calibrated steps from 1mV/div. to 20V/div. (1-2-5 sequence) with variable 2.5:1 up to 50V/div.
Accuracy in calibrated position:
1mV/div. to 2mV/div.: ±5% (0 to 10MHz (-3dB))
5mV/div. to 20V/div.: ±3% Input impedance: 1M II 20pF. Input coupling: DC-AC-GD (ground). Input voltage: max. 400V (DC + peak AC).
Delay line: approx. 90ns
Triggering
OSCILLOSCOPES
Automatic: Normal with level control: DC-100MHz (0.5div.)
Slope: positive or negative, ALT. Triggering; LED indicator for trigger action Sources: Channel I or II, CH. I alternating CH II,
Coupling: AC (10Hz to 100MHz), DC (0 Active TV-Sync-Separator (pos. and neg.)
External:0.3Vpp from DC to 60MHz 2nd triggering (Del. Trig.): normal with level
(peak to peak) <20Hz-100MHz ( 0.5div.)
line and external HF (1.5kHz to100MHz), LF (0 to 1.5kHz)
control DC to 100 MHz
to
100MHz),
Horizontal Deflection
Time coefficients: 22 calibrated steps
from 0.5s/div. to 50ns/div. in 1-2-5 sequence Accuracy in calibrated position: ±3%. variable 2.5:1 up to 1.25s/div., with X-Mag. x10: 5ns/div. ±5%
Holdoff time: variable to approx. 10:1 Delay: 50ms - 100ns, variable 6:1 up to 300ms Bandwidth X-amplifier: 0-3MHz (-3dB).
Input X-Amplifier via Channel II, (sensitivity see Channel II specification) X-Y phase shift: <3° below 120kHz.
Operation / Control
Auto Set (automatic parameter selection) Manual (Front Panel switches) Memory for 6 user-defined parameter settings Remote control with built-in RS-232 interface
Component Tester
Test voltage: approx. 8.5V
Test current: approx. 7mA Test frequency: approx. 50Hz Test connection: 2 banana jacks 4mmØ One test lead is grounded (Safety Earth)
(open circuit).
rms
(shorted).
rms
General Information
CRT: D14-372GH, rectangular screen (8x10div.)
internal graticule
Acceleration voltage: approx 14kV Trace rotation: adjustable on front panel Calibrator: square-wave generator (tr <4ns) 1kHz/1MHz; Output: 0.2V ±1% and 2V Line voltage: 100-240V AC ±10%, 50/60Hz Power consumption: approx. 40 Watt at 50Hz.
Min./Max. ambient temperature: -10°C...+40°C Protective system: Safety class I (IEC1010-1) Weight: approx. 5.6kg (12.4lbs), color: Cabinet: W 285, H 125, D 380 mm Lockable tilt handle
Subject to change without notice. 10/95
techno-brown
(11.1x4.9x14.8 inches)
60MHz Multi-Function Oscilloscope HM 604
-3
with Auto-Set, Save and Recall (6 Setup Memories) Remote control via built-in RS-232 Interface
Vertical: 2 Channels, 1 mV/div. - 20 V/div., Comp.-Tester, 1MHz Calibrator Time Base: 0.5 s/div. to 5 ns/div.; Trigger-after-delay; Alternate Trigger Triggering: DC to 100 MHz; Automatic Peak to Peak; TV-Sync-Separator
The HM604-3 is HAMEG‘s 60 MHz microprocessor controlled analog oscil- loscope. The internal microprocessor has the capability of automatically configuring the oscilloscope parameters, when used in the "Auto Set" mode, so as to present a display of three cycles of the input signal, with all of the proper control settings automatically configured: a single trace will be displayed with 4 to 6 divisions amplitude, while each signal will be 3 to 4 divisions high in dual trace mode. At a touch of the controls, the user can over-ride the automatic settings. Manual control is simple with easy to see indicators denoting uncalibrated operation by blinking. Switch/parameter settings are easily visible as bright LED scale value indications. One powerful feature is the ability to store user-defined test scenarios, that can be called up on demand for repeated measurement tasks. A total of 6 setups can be saved and recalled as many times as required. Additional setup information is possible via the RS-232 port, which can be controlled from the serial port of a PC. Although the instrument‘s vertical bandwidth is specified as 60 MHz, signals to 100 MHz can be displayed and triggered. Signal expansion up to a factor of 1,000 times can be obtained in the "Delay" and "Trigger after Delay" modes. A switching mode power supply minimises power and results in a lightweight unit weighing only 12.4 pounds (5.6kg). The scope is ideal for fast troubleshooting. The Auto Set feature permits the rapid inspection of test points, without the need to constantly adjust the scope for each new test mode. The ability to Save/Recall facilitates the use of this unit for final inspections, where multiple, repeated scope settings are required. These capabilities result in labor saving operations which translate to more efficiency and cost savings to the customer.
Accessories supplied: Line Cord, Operators Manual, 2 Probes 10:1
4
Subject to change without notice
General Information
This oscilloscope is easy to operate. The logical arrangement of the controls allows anyone to quickly become familiar with the operation of the instrument, however, experienced users are also advised to read through these instructions so that all functions are understood. Immediately after unpacking, the instrument should be checked for mechanical damage and loose parts in the interior. If there is transport damage, the supplier must be informed immediately. The instrument must then not be put into operation.
Symbols
ATTENTION - refer to manual Danger - High voltage
contains important information and warnings which have to be followed by the user to ensure safe operation and to retain the oscilloscope in a safe condition.
The case, chassis and all measuring terminals are connected to the protective earth contact of the appliance inlet. The instrument operates according to Safety Class I (three-conductor power cord with protective earthing conductor and a plug with earthing contact).
The mains/line plug shall only be inserted in a socket outlet provided with a protective earth contact. The protective action must not be negated by the use of an extension cord without a protective conductor.
The mains/line plug should be inserted before connections are made to measuring circuits.
Protective ground (earth) terminal
Use of tilt handle
To view the screen from the best angle, there are three dif­ferent positions (C, D, E) for setting up the instrument. If the instrument is set down on the floor after being carried, the handle automatically remains in the upright carrying position (A). In order to place the instrument onto a horizontal surface, the handle should be turned to the upper side of the oscilloscope (C). For the D position (10° inclination), the handle should be turned to the opposite direction of the carrying position until it locks in place automatically underneath the instrument. For the E position (20° inclination), the handle should be pulled to release it from the D position and swing backwards until it locks once more. The handle may also be set to a position for horizontal carrying by turning it to the upper side to lock in the B position. At the same time, the instrument must be lifted, because otherwise the handle will jump back.
The grounded accessible metal parts (case, sockets, jacks) and the mains/line supply contacts (line/live, neutral) of the instrument have been tested against insulation breakdown with 2200V DC.
Under certain conditions, 50Hz or 60Hz hum voltages can occur in the measuring circuit due to the interconnection with other mains/line powered equipment or instruments. This can be avoided by using an isolation transformer (Safety Class II) between the mains/line outlet and the power plug of the device being investigated.
Most cathode-ray tubes develop X-rays. However, the dose
equivalent rate falls far below the maximum permissible value of 36pA/kg (0.5mR/h).
Whenever it is likely that protection has been impaired, the instrument shall be made inoperative and be secured against any unintended operation. The protection is likely to be impaired if, for example, the instrument
• shows visible damage,
• fails to perform the intended measurements,
• has been subjected to prolonged storage under unfavourable conditions (e.g. in the open or in moist environments),
Safety
This instrument has been designed and tested in accordance with IEC Publication 1010-1 (overvoltage category II, pollution degree 2), Safety requirements for electrical equipment for measurement, control, and laboratory use. The CENELEC regulations EN 61010-1 correspond to this standard. It has left the factory in a safe condition. This instruction manual
Subject to change without notice
• has been subject to severe transport stress (e.g. in poor packaging).
Operating conditions
The instrument has been designed for indoor use. The permissible ambient temperature range during operation is +10°C (+50°F) ... +40°C (+104°F). It may occasionally be subjected to temperatures between +10°C (+50°F) and ­10°C (+14°F) without degrading its safety. The permissible ambient temperature range for storage or transportation is
-40°C (-40°F) ... +70°C (+158°F).
The maximum operating altitude is up to 2200m (non­operating 15000m). The maximum relative humidity is up to 80%. If condensed water exists in the instrument it should be acclimatized before switching on. In some cases (e.g. extremely cold oscilloscope) two hours should be allowed before the instrument is put into operation. The instrument should be kept in a clean and dry room and must not be operated in explosive, corrosive, dusty, or moist environments. The oscilloscope can be operated in any position, but the convection cooling must not be impaired. The ventilation holes may not be covered. For continuous
5
operation the instrument should be used in the horizontal position, preferably tilted upwards, resting on the tilt handle.
The specifications stating tolerances are only valid if the instrument has warmed up for 30minutes at an ambient temperature between +15°C (+59°F) and +30°C (+86°F). Values without tolerances are typical for an average instrument.
removed with a moistened cloth (99% water +1% mild detergent). Spirit or washing benzine (petroleum ether) can be used to remove greasy dirt. The screen may be cleaned with water or washing benzine (but not with spirit (alcohol) or solvents), it must then be wiped with a dry clean lint-free cloth. Under no circumstances may the cleaning fluid get into the instrument. The use of other cleaning agents can attack the plastic and paint surfaces.
Protective Switch-Off
EMC
This instrument conforms to the European standards regarding the electromagnetic compatibility. The applied standards are: Generic immunity standard EN50082-2:1995 (for industrial environment) Generic emission standard EN50081-1:1992 ( for residential, commercial und light industry environment). This means that the instrument has been tested to the highest standards. Please note that under the influence of strong electromagnetic fields, such signals may be superimposed on the measured signals. Under certain conditions this is unavoidable due to the instrument‘s high input sensitivity, high input impedance and bandwidth. Shielded measuring cables, shielding and earthing of the device under test may reduce or eliminate those effect‘s.
Warranty
HAMEG warrants to its Customers that the products it manufactures and sells will be free from defects in materials and workmaship for a period of 2 years. This warranty shall not apply to any defect, failure or damage caused by improper use or inadequate maintenance and care. HAMEG shall not obliged to provide service under this warranty to repair damage resulting from attempts by personnel other than HAMEG represantatives to install, repair, service or modify these products. In order to obtain service under this warranty, Customers must contact and notify the distributor who has sold the product. Each instrument is subjected to a quality test with 10 hour burn-in before leaving the production. Practically all early failures are detected by this method. In the case of shipments by post, rail or carrier it is recommended that the original packing is carefully preserved. Transport damages and damage due to gross negligence are not covered by the guarantee. In the case of a complaint, a label should be attached to the housing of the instrument which describes briefly the faults observed. If at the same time the name and telephone number (dialing code and telephone or direct number or department designation) is stated for possible queries, this helps towards speeding up the processing of guarantee claims.
This instrument is equipped with a switch mode power supply. It has both overvoltage and overload protection, which will cause the switch mode supply to limit power consumption to a minimum. In this case a ticking noise may be heard.
Power supply
The oscilloscope operates on mains/line voltages between 100VAC and 240VAC. No means of switching to different input voltages has therefore been provided.
The power input fuses are externally accessible. The fuseholder is located above the 3-pole power connector. The power input fuses are externally accessible, if the rubber conector is removed. The fuseholder can be released by pressing its plastic retainers with the aid of a small screwdriver. The retainers are located on the right and left side of the holder and must be pressed towards the center. The fuse(s) can then be replaced and pressed in until locked on both sides.
Use of patched fuses or short-circuiting of the fuseholder is not permissible; HAMEG assumes no liability whatsoever for any damage caused as a result, and all warranty claims become null and void.
Fuse type: Size 5x20mm; 0.8A, 250V AC fuse; must meet IEC specification 127, Sheet III (or DIN 41 662 or DIN 41 571, sheet 3). Time characteristic: time-lag (T).
Attention! There is a fuse located inside the instrument within the switch mode power supply:
Size 5x20mm; 0.5A, 250V AC fuse; must meet IEC specification 127, Sheet III (or DIN 41 662 or DIN 41 571, sheet 3). Time characteristic: fast (F).
This fuse must not be replaced by the operator!
Maintenance
Various important properties of the oscilloscope should be carefully checked at certain intervals. Only in this way is it largely certain that all signals are displayed with the accuracy on which the technical data are based. The test methods described in the test plan of this manual can be performed without great expenditure on measuring instruments. However, purchase of the HAMEG scope tester HZ 60, which despite its low price is highly suitable for tasks of this type, is very much recommended.
The exterior of the oscilloscope should be cleaned regularly with a dusting brush. Dirt which is difficult to remove on the casing and handle, the plastic and aluminium parts, can be
6
Subject to change without notice
Type of signal voltage
With the HM604-3, most repetitive signals in the frequency range up to at least 60MHz (-3dB) can be examined.
Sinewave signals of 100MHz are displayed with a height of approx. 50% (-6dB). However when examining square or pulse type waveforms, attention must be paid to the harmonic content of such signals. The repetition frequency (fundamen­tal frequency) of the signal must therefore be significantly smaller than the upper limit frequency of the vertical amplifier.
Displaying composite signals can be difficult, especially if they contain no repetitive higher amplitude content which can be used for triggering. This is the case with bursts, for instance. To obtain a well-triggered display in this case, the assistance of the variable holdoff and/or delay function may be required. Television video signals are relatively easy to trigger using the built-in TV-Sync-Separator (TV).
For optional operation as a DC or AC voltage amplifier, the vertical amplifier input is provided with a DC/AC switch. The DC position should only be used with a series-connected attenuator probe or at very low frequencies or if the measurement of the DC voltage content of the signal is absolutely necessary.
The minimum signal voltage which must be applied to the Y input for a trace of 1div. height is 1mVpp when the 1mV LED is lit and the vernier is set to CAL by turning the fine adjustment knob within the VOLTS/DIV. section fully clockwise. However, smaller signals than this may also be displayed. The deflection coefficients on the input attenuators are indicated in mV/div. or V/div. (peak-to-peak value). The magnitude of the applied voltage is ascertained by multiplying the selected deflection coefficient by the vertical display height in div. If an attenuator probe x10 is used, a further multiplication by a factor of 10 is required to ascertain the correct voltage value.
For exact amplitude measurements, the variable control ( VAR. 2.5:1) must be set to its calibrated detent CAL position. When turning the variable control ccw, the
deflection coefficient LED will start to blink and the sensitivity will be reduced until a maximum factor of 2.5 is reached. Therefore any intermediate value is possible within the 1-2-5 sequence.
With direct connection to the vertical input, signals up to 400Vpp may be displayed (attenuator set to 20V/ div., variable control to left stop).
With the designations
When displaying very low frequency pulses, the flat tops may be sloping with AC coupling of the vertical amplifier (AC limit frequency approx. 1.6 Hz for 3dB). In this case, DC operation is preferred, provided the signal voltage is not superimposed on a too high DC level. Otherwise a capacitor of adequate capacitance must be connected to the input of the vertical amplifier with DC coupling. This capacitor must have a sufficiently high breakdown voltage rating. DC coupling is also recommended for the display of logic and pulse signals, especially if the pulse duty factor changes constantly. Otherwise the display will move upwards or downwards at each change. Pure direct voltages can only be measured with DC-coupling.
Amplitude Measurements
In general electrical engineering, alternating voltage data normally refers to effective values (rms = root-mean-square value). However, for signal magnitudes and voltage designations in oscilloscope measurements, the peak-to-peak voltage (Vpp) value is applied. The latter corresponds to the real potential difference between the most positive and most negative points of a signal waveform.
If a sinusoidal waveform, displayed on the oscilloscope screen, is to be converted into an effective (rms) value, the resulting peak-to-peak value must be divided by 2x2 = 2.83. Conversely, it should be observed that sinusoidal voltages indicated in Vrms (Veff) have 2.83 times the potential difference in Vpp. The relationship between the different voltage magnitudes can be seen from the following figure.
H = display height in div., U= signal voltage in Vpp at the vertical input, D = deflection coefficient in V/div. at attenuator switch,
the required value can be calculated from the two given quantities:
However, these three values are not freely selectable. They have to be within the following limits (trigger threshold, accuracy of reading):
H between 0.5 and 8div., if possible 3.2 to 8div., U between 1mVpp and 160Vpp, D between 1mV/div. and 20V/div. in 1-2-5 sequence.
Examples:
Set deflection coefficient D = 50mV/div. 0.05V/div., observed display height H = 4.6div., required voltage U = 0.05x4.6 = 0.23Vpp.
Input voltage U = 5Vpp, set deflection coefficient D = 1V/div., required display height H = 5:1 = 5div.
Signal voltage U = 230Vrmsx22 = 651Vpp (voltage > 160Vpp, with probe 10:1: U = 65.1Vpp), desired display height H = min. 3.2div., max. 8div., max. deflection coefficient D = 65.1:3.2 = 20.3V/div., min. deflection coefficient D = 65.1:8 = 8.1V/div., adjusted deflection coefficient D = 10V/div.
Voltage values of a sine curve
Vrms = effective value; Vp = simple peak or crest value; Vpp = peak-to-peak value; Vmom = momentary value.
Subject to change without notice
The input voltage must not exceed 400V, independent from the polarity.
If an AC voltage which is superimposed on a DC voltage is applied, the maximum peak value of both voltages must not exceed + or -400V. So for AC voltages with a mean value of zero volt the maximum peak to peak value is 800Vpp.
If attenuator probes with higher limits are used, the probes limits are valid only if the oscilloscope is set to
7
DC input coupling. If DC voltages are applied under AC input
coupling conditions the oscilloscope maximum input voltage value remains 400V. The attenuator consists of a resistor in the probe and the 1M input resistor of the oscilloscope, which are disabled by the AC input coupling capacity when AC coupling is selected. This also applies to DC voltages with superimposed AC voltages. It also must be noted that due to the capacitive resistance of the AC input coupling capacitor, the attenuation ratio depends on the signal frequency. For sinewave signals with frequencies higher than 40Hz this influence is negligible.
In the GD (ground coupling) setting, the signal path is interrupted directly beyond the input. This causes the attenuator to be disabled again, but now for both DC and AC voltages.
With the above listed exceptions HAMEG 10:1 probes can be used for DC measurements up to 600V or AC voltages (with a mean value of zero volt) of 1200Vpp. The 100:1 pro­be HZ53 allows for 1200V DC or 2400Vpp for AC.
It should be noted that its ACpeak value is derated at higher frequencies. If a normal x10 probe is used to measure high voltages there is the risk that the compensation trimmer bridging the attenuator series resistor will break down causing damage to the input of the oscilloscope. However, if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope, a normal x10 probe is sufficient. In this case, an appropriate high voltage capacitor (approx. 22-68nF) must be connected in series with the input tip of the probe.
Total value of input voltage
The duration of a signal period or a part of it is determined by multiplying the relevant time (horizon­tal distance in div.) by the time coefficient indicated on the TIME/DIV. LED scales.
The variable time control (identified with an arrow knob cap) must be in its calibrated position CAL. (arrow pointing horizontally to the right). For exact time measurements, the variable control ( VAR. 2.5:1) must be set to its calibrated detent CAL position. When turning the variable control ccw, the time coefficient indicator LED starts blinking and the timebase speed will be reduced until a maximum factor of 2.5 is reached. Therefore any intermediate value is possible within the 1-2-5 sequence.
With the designations
L = displayed wave length in div. of one period, T = time in seconds for one period, F = recurrence frequency in Hz of the signal, Tc = time coefficient in s/div. on timebase switch and
the relation F = 1/T, the following equations can be stated:
= ⋅
=
=
=
=
=
With active X-MAG (x10) indicated by the x10 LED lit, the Tc value must be divided by 10.
However, these four values are not freely selectable. They have to be within the following limits:
The dotted line shows a voltage alternating at zero volt level. If superimposed on a DC voltage, the addition of the positive peak and the DC voltage results in the max. voltage (DC + ACpeak).
With Y-POS. control (input coupling to GD) it is possible to use a horizontal graticule line as reference line for ground potential before the measurement. It can lie below or above the horizontal central line according to whether positive and/ or negative deviations from the ground potential are to be measured.
Time Measurements
As a rule, most signals to be displayed are periodically repeating processes, also called periods. The number of periods per second is the repetition frequency. Depending on the time base setting indicated by one of the TIME/DIV. LED‘s, one or several signal periods or only a part of a period can be displayed. The time coefficients are stated in s/div. when the red sec-LED and the 0.5 or 0.2 LED (ms/div scale) are lit. The ms/div. or µs/div. time coefficients are indicated by one of the LED‘s on the ms or µs scale.
L between 0.2 and 10div., if possible 4 to 10div., T between 0.01µs and 5s, F between 0.5Hz and 35MHz, Tc between 0.05µs/div. and 0.5s/div. in 1-2-5 sequence
(with X-MAG. (x10) inactive), and
Tc between 5ns/div. and 20ms/div. in 1-2-5 sequence
(with X-MAG. (x10) active).
Examples:
Displayed wavelength L = 7div., set time coefficient Tc = 0.1µs/div., required period T = 7x0.1x10 required rec. freq. F = 1:(0.7x10
-6
= 0.7µs
-6
) = 1.428MHz.
Signal period T = 1s, set time coefficient Tc = 0.2s/div., required wavelength L = 1:0.2 = 5div..
Displayed ripple wavelength L = 1div., set time coefficient Tc = 10ms/div., required ripple freq. F = 1:(1x10x10
-3
) = 100Hz.
TV-line frequency F = 15625Hz, set time coefficient Tc = 10µs/div., required wavelength L = 1:(15 625x10-5) = 6.4div..
Sine wavelength L = min. 4div., max. 10div., Frequency F = 1kHz, max. time coefficient Tc = 1:(4x10
3
) = 0.25ms/div., min. time coefficient Tc = 1:(10x103) = 0.1ms/div., set time coefficient Tc = 0.2ms/div., required wavelength L = 1:(10
3
x0.2x10-3) = 5div.
8
Subject to change without notice
Displayed wavelength L = 0.8div., set time coefficient Tc = 0.5µs/div., pressed X-MAG. (x10) button: Tc = 0.05µs/div., required rec. freq. F = 1:(0.8x0.05x10 required period T = 1:(25x10
6
) = 40ns.
-6
) = 25MHz,
If the time is relatively short as compared with the complete signal period, an expanded time scale should always be applied (X-MAG. (x10) active). In this case, the ascertained time values have to be divided by 10. The time interval of interest can be shifted to the screen center using the X-POS. control.
When investigating pulse or square waveforms, the critical feature is the risetime of the voltage step. To ensure that transients, ramp-offs, and bandwidth limits do not unduly influence the measuring accuracy, the risetime is generally measured between 10% and 90% of the vertical pulse height. For measurement, adjust the Y deflection coefficient with its variable control together with the Y-POS. control so that the pulse height is precisely aligned with the 0% and 100% lines of the internal graticule. The 10% and 90% points of the signal will now coincide with the 10% and 90% graticule lines. The risetime is given by the product of the horizontal distance in div. between these two coincident points and the time coefficient setting. If X x10 magnification is used, this product must be divided by 10. The fall time of a pulse can also be measured by using this method.
The following figure shows correct positioning of the oscilloscope trace for accurate risetime measurement.
... .... .... .... .... .... .... .... .... ....
.
... .... .... .... .... .... .... .... .... ....
particularly simple in this way. In principle it is possible to measure in any display position and at any signal amplitude. It is only important that the full height of the signal edge of interest is visible in its full length at not too great steepness and that the horizontal distance at 10% and 90% of the amplitude is measured. If the edge shows rounding or overshooting, the 100% should not be related to the peak values but to the mean pulse heights. Breaks or peaks (glitches) next to the edge are also not taken into account. With very severe transient distortions, the rise and fall time measurement has little meaning. For amplifiers with approximately constant group delay (therefore good pulse transmission performance) the following numerical relationship between rise time tr (in ns) and bandwidth B
(in MHz) applies:
Connection of Test Signal
Caution: When connecting unknown signals to the oscilloscope input, always use automatic triggering and set the DC-AC input coupling switch to AC (DC not lit). The attenuator should initially be set to 20V/ div.
Sometimes the trace will disappear after an input signal has been applied. The attenuator must be switched to a higher deflection coefficient by pressing the left (<) arrow pushbutton in the VOLTS/DIV. section constantly or step by step, until the vertical signal height is only 3-8div. With a signal amplitude greater than 160Vpp, an attenuator probe must be inserted before the vertical input. If, after applying the signal, the trace is nearly blanked, the period of the signal is probably substantially longer than the set value on the TIME/DIV. scale. It should be switched to an adequately larger time coefficient by pressing the left (<) arrow pushbutton in the TIME/DIV section by pressing it constantly or step by step. In most cases the easiest way to adapt the instruments settings to the input signal is to depress the AUTO SET pushbutton for automatic instrument settings.
With a time coefficient of 0.05µs/div. and X x10 magnification, the example shown in the above figure results in a total measured risetime of
= 1.6div x 0.05µs/div. : 10 = 8ns
t
tot
When very fast risetimes are being measured, the risetimes of the oscilloscope amplifier and of the attenuator probe has to be deducted from the measured time value. The risetime of the signal can be calculated using the following formula.
=√
In this t of the oscilloscope amplifier (approx. 12ns), and t of the probe (e.g. = 2ns). If t can be taken as the risetime of the pulse, and calculation is
is the total measured risetime, t
tot
is greater than 100ns, then t
tot
unnecessary. Calculation of the example in the figure above results in a
signal risetime
is the risetime
osc
the risetime
p
=√ =
The measurement of the rise or fall time is not limited to the trace dimensions shown in the above diagram. It is only
The signal to be displayed can be connected directly to the Y ­input of the oscilloscope with a shielded test cable such as HZ32 or HZ34, or reduced through a x10 or x100 attenuator probe. The use of test cables with high impedance circuits is only recommended for relatively low frequencies (up to approx. 50kHz). For higher frequencies, the signal source must be of low impedance, i.e. matched to the characteristic resistance of the cable (as a rule 50). Especially when transmitting square and pulse signals, a resistor equal to the characteristic impedance of the cable must also be connected across the cable directly at the Y-input of the oscilloscope. When using a 50 cable such as the HZ34, a 50 through termination type HZ22 is available from HAMEG. When transmitting square signals with short rise times, transient phenomena on the edges and top of the signal may become visible if the correct termination is not used. A terminating resistance is sometimes recommended with sine signals as well. Certain amplifiers, generators or their attenuators
tot
maintain the nominal output voltage independent of frequency only if their connection cable is terminated with the prescribed resistance. Here it must be noted that the terminating resistor HZ22 will only dissipate a maximum of 2Watts. This power is reached with 10Vrms or at 28.3Vpp with sine signal.
If a x10 or x100 attenuator probe is used, no termination is necessary. In this case, the connecting cable is matched directly to the high impedance input of the oscilloscope. When using attenuators probes, even high internal impedance
Subject to change without notice
9
sources are only slightly loaded (approx. 10M II 16pF or 100M II 9pF with HZ53). Therefore, if the voltage loss due to the attenuation of the probe can be compensated by a higher amplitude setting, the probe should always be used. The series impedance of the probe provides a certain amount of protection for the input of the vertical amplifier. Because of their separa­te manufacture, all attenuator probes are only partially compensated, therefore accurate compensation must be performed on the oscilloscope (see Probe compensation ).
Standard attenuator probes on the oscilloscope normally reduce its bandwidth and increase the rise time. In all cases where the oscilloscope bandwidth must be fully utilized (e.g. for pulses with steep edges) we strongly advise using the probes HZ51 (x10) HZ52 (x10 HF) and HZ54 (x1 and x10). This can save the purchase of an oscilloscope with larger bandwidth.
The probes mentioned have a HF-calibration in addition to low frequency calibration adjustment. Thus a group delay correction to the upper limit frequency of the oscilloscope is possible with the aid of an 1MHz calibrator, e.g. HZ60.
where a trace appears on the screen if the INTENS. knob is in center position, all LED‘s should remain unlit. The trace, displaying one baseline or the shorter COMP TESTER baseline, should be visible after a short warm-up period of approx. 10 seconds. If the COMP TESTER mode is active, depress the COMP TESTER pushbutton once to switch to XY or Yt mode. In XY mode the XY LED in the TIME/DIV section is lit, in this case depress the XY pushbutton once to switch over to Yt mode. Adjust Y-POS.I and X-POS. controls to center the baseline. Adjust INTENS. (intensity) and FOCUS controls for medium brightness and optimum sharpness of the trace. The oscilloscope is now ready for use.
• Rotate the variable controls with arrows, i.e. TIME/DIV. variable control, CH.I and CH.II attenuator variable controls, and HOLD OFF control to their calibrated detent.
• Set all controls with marker lines to their midrange position (marker lines pointing vertically).
••
• Depress the upper NORM. pushbutton until the AC
•• symbol on the trigger coupling scale is lit.
• Both GD input coupling pushbutton switches for CH.I and CH.II in the Y-field should be set to the GD position (GD lit).
In fact the bandwidth and rise time of the oscilloscope are not noticably changed with these probe types and the waveform reproduction fidelity can even be improved because the probe can be matched to the oscilloscopes individual pulse response.
If a x10 or x100 attenuator probe is used, DC input coupling must always be used at voltages above 400V . With AC coupling of low frequency signals, the attenuation is no longer independent of frequency, pulses can show pulse tilts. Direct voltages are suppressed but load the oscilloscope input coupling capacitor concerned. Its voltage rating is max. 400 V (DC + peak AC). DC input coupling is therefore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max. 1200 V (DC + peak AC). A capacitor of corresponding capacitance and voltage rating may be connected in series with the attenuator probe input for blocking DC voltage (e.g. for hum voltage measurement).
With all attenuator probes, the maximum AC input voltage must be derated with frequency usually above 20kHz. Therefore the derating curve of the attenuator probe type concerned must be taken into account.
The selection of the ground point on the test object is important when displaying small signal voltages. It should always be as close as possible to the measuring point. If this is not done, serious signal distortion may result from spurious currents through the ground leads or chassis parts. The ground leads on attenuator probes are also particularly critical. They should be as short and thick as possible. When the attenuator probe is connected to a BNC-socket, a BNC­adapter, should be used. In this way ground and matching problems are eliminated. Hum or interference appearing in the measuring circuit (especially when a small deflection coefficient is used) is possibly caused by multiple grounding because equalizing currents can flow in the shielding of the test cables (voltage drop between the protective conductor connections, caused by external equipment connected to the mains/line, e.g. signal generators with interference protection capacitors).
If only a spot appears (CAUTION! CRT phosphor can be damaged), reduce the intensity immediately and check that the XY mode is not selected (XY LED dark). If the trace is not visible, check the correct positions of all knobs and modes (particularly NM LED - normal triggering - LED on).
To obtain the maximum life from the cathode-ray tube, the minimum intensity setting necessary for the measurement in hand and the ambient light conditions should be used.
Particular care is required when a single spot is displayed, as a very high intensity setting may cause damage to the fluorescent screen of the CRT. Switching the oscilloscope off and on at short intervals stresses the cathode of the CRT and should therefore be avoided.
The instrument is so designed that even incorrect operation will not cause serious damage.
The HM604-3 accepts all signals from DC (direct voltage) up to a frequency of at least 60MHz (-3dB). For sinewave voltages the upper frequency limit will be 100MHz (-6dB). However, in this higher frequency range the vertical display height on the screen is limited to approx. 4-5div. The time resolution poses no problem. For example, with 100MHz and the fastest adjustable sweep rate (5ns/div.), one cycle will be displayed every 2div. The tolerance on indicated values amounts to ±3% in both deflection directions. All values to be measured can therefore be determined relatively accurately.
However, from approximately 10MHz upwards the measuring error will increase as a result of loss of gain. At 18MHz this reduction is about 10%. Thus, approximately 11% should be added to the measured voltage at this frequency. As the bandwidth of the amplifiers may differ slightly (normally between 60 and 78MHz), the measured values in the upper frequency range cannot be defined exactly. Additionally, as already mentioned, for frequencies above 60MHz the dynamic range of the display height steadily decreases. The vertical amplifier is designed so that the transmission performance is not affected by its own overshoot.
First Time Operation
Switch on the oscilloscope by depressing the red POWER pushbutton. The instrument will revert to its last used operating mode. Except in the case of COMP. TESTER mode,
10
Trace Rotation TR
In spite of Mumetal-shielding of the CRT , effects of the earths magnetic field on the horizontal trace position cannot be completely avoided. This is dependent upon the orientation
Subject to change without notice
Loading...
+ 23 hidden pages