General Information
Use of tilt handle
Safety.
Operating conditions
Warranty
Maintenance
Mains/Line voltage change
Type of Signal
Amplitude Measurements
Time Measurements
Connection of Test Signal
Operating
First Time Operation
Trace Rotation TR
DC Balance Adjustment
Use and Compensation of Probes
Operating Modes of the Y Amplifier
X-Y Operation
X-Y Phase Measurements
Dual Trace Phase Difference Measurements
Measurement of an amplitude modulation
Triggering and
Triggering of video signals
Function of variable HOLD OFF control
Sweep Delay /After Delay Triggeringe
Delay Mode Indication
Component Tester
Miscellaneous
Test Patterns
Short Instruction
Front Panel Elements
Folder with Front View
Test Instructions
General
Cathode-Ray Tube: Brightness, Focus,
Astigmatismus Check
Symmetry and Drift of
Calibration of the Vertical Amplifier
Transmission Performance of the Vertical Amplifier
............................T2
Operating Modes: CH I/II-TRIG.
Triggering Checks
Timebase
SweepDelay
Component Tester
Trace Alignment
Miscellaneous
....................
.................
......................
.....................
...................
...................
.....................
.................
...................
Timebase
................
..................
...................
.......................
Linearity, Raster Distortions . T 1
CHOP., INV. l/II and XY-Betrieb . T 2
.................
......................
....................
.................
..................
...................
...............
................
............
............
...............
............
...............
.............
............
.............
............
..............
..............
...............
thevertical
........
.......
......
......
Amplifier
.........
l/II,
DUAL, ADD,
....
. .
....
M 10
P
1
Z 1
M 1
M 1
M 1
M 2
M 2
M 2
M 2
M 3
M 3
M 4
M 5
M 6
M 7
M 7
M 7
M 8
M 9
Ml 0
Ml 0
Ml 1
Ml 1
Ml 2
Ml3
Ml 3
Ml 5
Ml 5
Ml 7
Ml 8
K 1.
K
2
T 1
T 1
T 1
T 1
T 3
T 3
T 4
T 4
T 4
T 4
Oscilloscope
HM 604
Service Instructions
General .......................S 1
Instrument Case Removal............. S 1
Operating Voltages ................. S 1
Minimum Brightness ................ S 1
Astigmatismus control ............... S 1
Trouble Shooting the Instrument ..........S 2
Replacement of Components and Parts ......S 2
Replacement of the Power Transformer ......S 2
Adjustments ....................S 3
Circuit Diagrams
Block Diagram
Wiring Diagram
Identification of Components
Y Input, Attenuator, Preamplifier CH.
Y intermediate Amplifiers, Trigger Pre-Amplifiers,
Component Tester
Y Final Amplifier
Post Trigger, Field Selector
Timebase (analog)
Timebase
Tim,ebase Generator
X Final Amplifier, Calibrator
CRT and HV circuit
Power Supply
Component Locations
XY Board
TB Board
PTFS Board
TBG, CAL, YF Boards
CO, EY, Z Boards
...................
...................
..................
(digital)
..................
....................
......................
......................
.....................
..................
................
.............
.................
................
.............
.................
................
............
l/II
......
D 1
D 2
D 3
D 4
D 5
D 6
D 7
D 8
D 9
DlO
Dl
D12
D13
D14
D15
D16
D17
D18
1
Subject to change without notice
9.88. 604
Page 2
Specification
Vertical
Operating modes: Channel I or Ch. II separate,
Channel I and II: alternate or chopped.
(Chopper frequency approx.
Sum or difference of Ch. I and Ch. II,
(with invert buttons for both Channels).
XY-Mode: via Channel I and Channel II.
Frequency range: 2x DC to 6OMHz(- 3dB).
Risetime: approx. 5.8ns. Overshoot: II %.Deflection coefficients: 12 calibrated steps
from 5 mV/div. to
variable 2.5: 1 to min.
Accuracy in calibrated position:
Y-Magnification x5 (calibrated) to 1 mV/div.
(Frequency range DC to
input impedance: 1 MQ II
Input coupling: DC-AC-GD (Ground)
Input voltage: max. 400V (DC + peak AC).
Y-output from CH I or CH II, = 50
Delay Line: approx. 90ns.
Mhtion
ZOV/div
50V/cm.
0.5MHz).
in l-Z-5 sequence,
+3%.
20MHz. -
3OpF.
3dB.
mV,Jdiv.
(50
Trigger System
With automatic
normal with level control from DC- 100 MHz.
LED indication for trigger action.
Slope: positive or negative.
Sources: Ch. I, Ch. II, line, external.
Coupling: AC (21
Threshold: external
Active TV-Sync-Separator for line and frame.
Slope positive or negative.
2nd. Triggering
trolled (independent from slope direction).
+ selection for TV mode.
Threshold: 1 div; typlcal
Trigger bandwidth:
lOHz-IOOMHz (P5mm
OHz
LF
HF(?50kHzmlOOMHz).
to approx. 20 MHz), DC
(DC -
550 kHz),
r50mV.
(Del. Trig.):
0.5div.
225
Hz to 60 MHz.
autom. or slope con-
height)
52)
60 MHz Universal Oscilloscope
2 Channels, 1
Timebase: 2.5s/div. to 5ns/div. including
Triggering:
mV/div.
DC-lOOMHz,
Sensitivity, Delay Line, Component Tester
x10
Magnifier&Sweep Delay
TV Sync Separator, After-Delay Trigger
Time coefficients: 23 calibrated steps
from
50ns/div.
variable 2.5: 1 to min.
accuracy in calibrated position:
with X-Magnifier
Hold-Off time: variable (2 5 : 1).
Delay: 7 decade steps
from 1 OOns
Bandwidth X-Amplifier:
Input X-Amplifier via Channel
sensitivity see Ch. II specification.
X-Y phase shift:
Ramp output: approx 5V. positive going
Test voltage: max.
Test current: max.
Test frequency: 50 - 60 Hz (line frequency).
Cathode-ray tube:
rectangular screen, internal graticule 8x10 cm.
Total acceleration voltage: 12
Trace rotation: adjustable on front panel.
Calibrator: square-wave generator switchable
from
Output voltage: 0.2V and 2V fl
Protective system Safety Class I
Linevoltage: 110. 125, 220,
Line frequency: 50 to 60Hz.
Power consumption: = 40 Watt.
Weight: approx. 8kg. Colour: techno-brown.
Cabinet: W 285, H 145, D 380mm.
Lockable tilt handle.
to 1
s/div
in l-2-5 sequence,
2.5s/div,
x10 (k
to 0.1 s, variable approx 10 : 1 to 1 s.
<3”
below 120
8.5V,,,
8mA,,,
150CTB31
=
1
kHz
to 1 MHz
+3%.
5%) to =
DC-5MHz (-3dB).
II,
(open circuit).
(shorted).
P43l123,
(t,
approx. 3ns).
24OV- *IO%.
5nsIdiv..
kHz.
kV.
%.
(IEC
348).
With its variety of operating and trigger modes, the HM604 is a new in-
novative general purpose oscilloscope satisfying a wide range of exacting re-
quirements in laboratory, production, and service. The dual-channel measure-
ment amplifier ensures highly faithful waveform transfer characteristics,
which can be readily checked on the built-in fast-risetime 1 MHz Calibrator
from probe tip to CRT screen! Using Y-axis magnification, the instrument’s
high sensitivity enables stable displays of very small signals as low as 0.5
mV.
An analog output is provided for connecting multimeters or counters. Another
important feature is the internal delay line for observations of the leading edge
of a signal. As in dual-time base oscilloscopes, the HM604 features a calibrated sweep delay mode, allowing smallest waveform sections to be expanded
up to 1000 times.
The
HM604’s
most outstanding feature, however, is the unique, newly
developed automatic After-Delay Trigger mode to ensure extremely stable
displays and jitter-free measurements of asynchronous signal sections and
bursts or pulse trains, independent of amplitude fluctuations. An active
TV-
Sync-Separator further enhances trigger quality of video frame and line signals. In the alternate trigger mode, two signals of different frequencies can be
compared.
With this state-of-the-art oscilloscope, HAMEG again sets a new price/
performance standard which is not likely to be met by others in this category.
Users will be particularly impressed by the instrument’s outstanding versatility
and ease of operation. These features are possible in the HM604 due to
HAMEG’s
meticulous attention to detail and many decades of successful
design experience.
-
Subject to change
without
notice
Page 3
Test
Cable Banana - BNC
Coaxial test cable; length 1.15 m, characteristic impedance 5Ofi.
Cable capacitance
12OpF.
Input voltage max.
5OOV,.
HZ32
Modular Probes
The clear advantage over ordinary probes are field replaceable
parts and the HF-compensation feature on the 10: 1 attenuator pro-
bes For the first time, probes in this price range allow adjustments
of their HF-characteristics to match individually the input impedance of each scope. This is particularly important for scopes with
higher bandwidths (>SOMHz), as otherwise strong overshoot or
rounding may occur, when measuring fast-rising square waves.
An exact HF-compensation, however, is only possible with square-
wave generators having a
already feature such a calibration generator. For other oscillo-
scopes, it is available as accessory item
following Modular Probes are available (HZ36 without HF-com-
pensation):
risetime <5ns.
Most
HZ60-2.
HAMEG
scopes
At present the
Test Cable BNC-BNC
Coaxial test cable; length 1 m, characteristic impedance
Cable capacitance 126pF. Input voltage max.
5OOV,.
Adapter Banana - BNC
Two 4mm binding posts
male plug. input voltage max.
5Ofz
Through-Termination
For terminating systems with
Maximum load 2 W. Max. voltage 1
(19mm
between
5OOV,.
5OS2
centers)
characteristic impedance.
OV,,,.
HZ34
500.
HZ20
to standard BNC
HZ22
Carrying Cases
For HM 103
For HM203, HM204. HM205, HM208, HM408. HM604,
HM605 and HM 1005
Viewing Hood
For HM203, HM204, HM205, HM208, HM408, HM604, HM605
and HM 1005
Scope-T-tar
HZ95
HZ96
HZ47
HZtiOa
Type
Attenuation Ratio
Bandwidth min. (MHz)
Risetime
(ns)
Inp. Capacitance
Inp. Resistance (MR)
Inp. Voltage max.
Cable Length (m)
Spare Cable for HZ36
Spare Cable for HZ51, HZ54
Sparepart Kit (2 sprung
Special probe for AM-demodulation and wobbulator measure-
ments. HF-Bandwidth
age 250mV -
peak AC. Cable length 1.2m.
High
For measurement of voltages up to
approx. 500 mQ. Recommended load resistance1
(switchable). Attenuation ratio 1000 : 1. Bandwidth 1 MHz. Cable
length 1.5 m. BNC connector.
5OV,,,.
Vdtaget
HZ36
selectable
l:l/lO:l 1O:l
IO/
3513.5
(pF)
47/18
(V,)
hooks, 2 screw tips, 1 ground cable)
I’mbe
HZ51HZ52HZ53
IO:1
100
150
<2
16
l/10
600600
1.51.2
IOOkHz
DC isolation Voltage 200V DC including
10
-5OOMHz
(HF) 1OO:l
250
tl.4
16
10100
6001200
1.51.5
(+ldB).
150
<2
65
AC
HZ54
selectable
1:1/10:1
IO/
150
35/<2
40/I
l/10
600
1.2
HZ39
HZ57
HZ46
InputVolt-
Hz58
15kV,,.
Input resistance
Ma/l
0
MQ
For Checking the Y amplifier, timebase, and compensation of all
probes, the
square-wave generator with switchable frequencies of DC, 1
IOOHz,
nals of 25
100x probes); accuracy +q%. Battery-powered.
8
I-IO-I
mV,,
HZ6O-2
is a crystal-controlled, fast rising (typ. 3ns)
OOkHz,
and 1 MHz. Three BNC outputs provide sig-
into 50 M,
0.25V,,
and
2.5V,,
(open circuit for 1 Ox and
Component-Tester
Indispensable for trouble-shooting in electronic circuits. Single
component and in-circuit tests are both possible. The HZ65 operates with all scopes, which can be switched to X-Y operation (ext.
horizontal deflection). Non-destructive tests can be carried out on
almost all semiconductors, resistors, capacitors, and coils. Two
sockets provide for quick testing of the 3 junction areas in any small
power transistor. Other components are connected by using 2
banana jacks. Test leads supplied.
Examples of Test Displays:
ShortcircuIt
Capacitor
33~F
Junction E-CZ-Diode
-lO-
HZ65
t8V
Printed in West Germany 5/90
Zl
Page 4
Operating Instructions
General Information
This oscilloscope is easy to operate. The logical arrange-
ment of the controls allows anyone to become familiar with
the operation of the instrument after a short time, however,
experienced users are also advised to read through these
instructions so that all functions are understood.
Immediately after unpacking, the instrument should be
checked for mechanical damage and loose parts in the interior. If there is transport damage, the supplier must be informed immediately. The instrument must then not be put
into operation.
Check that the instrument is set to the correct mains/line
voltage. If not, refer to instructions on page M2.
Use of tilt handle
To view the screen from the best angle, there are three dif-
ferent positions (C, D, E) for setting up the instrument. If the
instrument is set down on the floor after being carried, the
handle remains automatically in the upright carrying posi-
tion (A).
In order to place the instrument onto a horizontal surface,
the handle should be turned to the upper side of the oscilloscope (C). For the D position
should be turned in the opposite direction out of the carry
ing position until it locks in place automatically underneath
the instrument. For the E position (20” inclination), the
handle should be pulled to release it from the D position and
swing backwards until it locks once more.
The handle may also be set to a position for horizontal carry-
ing by turning it to the upper side to lock in the B position. At
the same time, the instrument must be moved upwards,
because otherwise the handle will jump back.
B
6
(IO’
inclination), the handle
E
cict
20”
Safety
This instrument has been designed and tested in accordance with
Electronic Measuring Apparatus, and has left the factory
in a safe condition. The present instruction manual contains
important information and warnings which have to be followed by the user to ensure safe operation and to retain the
oscilloscope in safe condition. The case, chassis and all
measuring terminals are connected to the protective earth
contact of the appliance inlet. The instrument operates according to Safety C/ass I (three-conductor power cord with
protective earthing conductor and a plug with earthing contact). The mains/line plug shall only be inserted in a socket
outlet provided with a protective earth contact. The protective action must not be negated by the use of an extension
cord without a protective conductor.
Warning! Any interruption of the protective conductor
inside or outside the instrument or disconnection of the
protective earth terminal is likely to make the instru-
ment dangerous. Intentional interruption of the protec-
tive earth connection is prohibited. The mains/line plug
should be inserted before connections are made to
measuring circuits.
The grounded accessible metal parts (case, sockets, jacks)
and the mains/line supply contacts (line, neutral) of the in-
strument have been tested against insulation breakdown
with 2000 Vr.m.s.
Under certain conditions, 50 Hz or 60Hz hum voltages can
occur in the measuring circuit due to the interconnection
with other mains/line powered equipment or instruments.
This can be avoided by using an isolation transformer
(Safety Class II) between the mains/line outlet and the
power plug of the instrument. When displaying waveforms
where the “low-level” side of the signal is at a high potential, even with the use of a protective isolation transformer,
it should be noted that this potential is connected to the os-
cilloscope’s case and other accessible metal parts. High
voltages are dangerous. In this case, special safety precau-
tions are to be taken, which must be supervised
personnel if the voltage is higher than 42V.
Most cathode-ray tubes develop X-rays. However, the
dose equivalent rate falls far below the maximum per-
missible value of
Whenever it is likely that protection has been impaired, the
instrument shall be made inoperative and be secured
against any unintended operation. The protection is Ii kely to
be impaired if, for example, the instrument
-
shows visible damage,
-
fails to perform the intended measurements,
-
has been subjected to prolonged storage under un-
favourable conditions (e.g. in the open or in moist envi-
ronments),
-
has been subject to severe transport stress (e.g. in poor
packaging).
IECPublication346,SafetyRequirementsfor
(5OHz).
by qualified
36pA/kg (0.5mRlh).
Subject to change without notice
Ml
Page 5
Operating conditions
Maintenance
The instrument has been designed for indoor use.
The permissible ambient temperature range during operation is + 15°C
. . .
+3O”C. It may occasionally be subjected to
temperatures between + 10°C and - 10°C without degrad-
ing its safety. The permissible ambient temperature range
for storage or transportation is -40°C .
+70X.
The maximum operating altitude is up to 2200m (non-
operating 15000m). The maximum relative humidity is up
to 80%.
If condensed water exists in the instrument it should be
acclimatized
before switching on. In some cases (e.g.
extremely cold oscilloscope) two hours should be allowed
before the instrument is put into operation. The instrument
should be kept in a clean and dry room and must not be
operated in explosive, corrosive, dusty, or moist environ-
ments The oscilloscope can be operated in any position,
but the convection cooling must not be impaired. The
wen-
tilation holes may not be covered. For continuous operation the instrument should be used in the horizontal position, preferably tilted upwards, resting on the tilt handle.
The specifications stating tolerances are only valid if
the instrument has warmed up for 30 minutes at an
ambient temperature between
+15C”
and +3OC9 Values not stating tolerances are typical for an average
instrument.
Warranty
Each instrument runs through a quality test with 10 hour
burn-in before leaving the production. Practically every early
failure is detected in intermittent operation by this method.
However, it is possible that a component fails only after a
lengthy operating period. Therefore a functional guaran-
tee of 2 years is given for all units. The condition for this is
that no modifications have been made in the instrument. In
the case of shipments by post, rail or carrier it is recom-
mended that the original packing is carefully preserved.
Transport damages and damage due to gross negligence
are not covered by the guarantee.
In the case of a complaint, a label should be attached to the
housing of the instrument which describes briefly the faults
observed. If at the same time the name and telephone
number (dialing code and telephone or direct number or
department designation) is stated for possible queries, this
helps towards speeding up the processing of guarantee
claims.
Various important properties of the oscilloscope should be
carefully checked at certain intervals. Only in this way is it
largely certain that all signals are displayed with the accuracy on which the technical data are based. The test
methods described in the test plan of this manual can be
performed without great expenditure on measuring instruments. However, purchase of the new HAMEG scope tester HZ 60, which despite its low price is highly suitable for
tasks of this type, is very much recommended.
The exterior of the oscilloscope should be cleaned regularly
with a dusting brush. Dirt which is difficult to remove on the
casing and handle, the plastic and aluminium parts, can be
removed with a moistened cloth (99% water +I % mild
detergent). Spirit or washing benzine (petroleum ether) can
be used to remove greasy dirt. The screen may be cleaned
with water or washing benzine (but not with spirit (alcohol)
or solvents), it must then be wiped with a dry clean lint-free
cloth. Under no circumstances may the cleaning fluid get
into the instrument. The use of other cleaning agents can
attack the plastic and paint surfaces.
Switching over the mains/line voltage
The instrument is set for 220V (240V U.K.) line voltage on
delivery. It can be switched over to other voltages at the
fuse holder combined with the 3-pole appliance inlet at the
rear of the instrument. Firstly the fuse holder printed with
the voltage values is removed using a small screw driver
and - if required - provided with another fuse. Refer to the
table below for the prescribed value of the fuse. Then
replace the fuse holder so that the impressed white triangle
points to the desired voltage. Here pay attention that the
cover plate is also correctly engaged. The use of repaired
fuses or short circuiting the fuse holder is not allowed. Damage arising because of this is not covered by the guarantee.
Fuse type: Size 5 x 20 mm; 250 V-, C;
IEC 127, Sheet III; DIN 41 662 (possibly DIN
sheet 3).
Cutoff: time lag
Line voltage
llOV-flO%
125V- &IO%
22ov- &IO%
24OV- &IO%
(T).
Fuse rating
TO.63 A
TO.63 A
T0.315A
T0.315A
.41
571
M2
Subject to change
wlthout
notice
Page 6
Type of Signal
All types of signals with a frequency spectrum below
60 MHz can be displayed on the HM 604. The display of sim-
ple electrical processes such as sinusoidal RF and AF signals or ripple poses no problems. However, when square or
pulse-shaped signals are displayed it must be remembered
that their harmonic content must also be transmitted. In
this case, the bandwidth of the vertical amplifier must be
considerably higher than the repetition frequency of the sig-
nal. In view of this, accurate evaluation of such signals with
the HM 604 is only possible up to a maximum repetition rate
of
6MHz.
Operating problems can sometimes occur when
composite signals are to be displayed, especially if they do
not contain any suitable level components and repetition
frequency which can be used for triggering. This occurs, for
example, with burst signals. To obtain a stably triggered display in these cases, it may be necessary to use Normal Triggering, HOLD OFF time control, and/or
control.
TIME/DIV.
variable
If a sinusoidal waveform, displayed on the oscilloscope screen, is to be converted into an effective (rms) value, the resulting peak-to-peak value must be divided by
2x-
=
2.83. Conversely, it should be observed that sinusoidal voltages indicated in
V,,,
(V,,,)
have 2.83 times the potential difference in V,,. The relationship between the different voltage magnitudes can be seen from the following figure.
Voltage values of a sine curve
v
= effective value;
rms
V,,
= peak-to-peak value;
V,
= simple peak or crest value;
V,,,
= momentary value.
Video signals are easily triggerable by the aid of the active
TV sync separator (TV SEP. switch).
For optional operation as a DC or AC voltage amplifier, each
channel is provided with a DC-AC coupling switch. The DC
position should only be used with an attenuator probe or at
very low frequencies or if the determination of DC voltage
content of the signal is absolutely necessary.
However, when investigating very low-frequency pulses,
misleading ramp-offs may occur with AC coupling. In this
case, DC operation is to be preferred if the signal voltage is
not superimposed on a too high DC voltage level. Otherwise, a capacitor of adequate capacitance must be con-
nected before the input of the vertical amplifier (switched to
DC coupling). It should be remembered that this capacitor
must have a sufficiently high breakdown voltage. DC opera-
tion is also recommended for the display of logic and pulse
signals, particularly if their pulse duty factor changes perma-
nently during operation. Otherwise, the display will move
up and down with any change. DC voltages can only be
measured in the DC position.
Amplitude Measurements
The minimum signal voltage required at the vertical amplifier
input for a display of 1 cm is approximately
achieved with the attenuator control set at
7mV,,.
5mV/cm,
This is
its var-
iable control in the fully clockwise position and pulled
out. However, smaller signals than this may also be dis-
played. The deflection coefficients on the input attenuators
are indicated in
mV/cm
or V/cm (peak-to-peak value).
The magnitude of the applied voltage is ascertained by
multiplying the selected deflection coefficient by the
vertical display height in cm.
If an attenuator probe x 70 is used, a further multiplication by a factor of 70 is required to ascertain the correct
voltage value.
For exact amplitude measurements the variable con-trol on the attenuator switch must be set to its calibra-
ted detent CAL. When turning the variable control ccw
the sensitivity will be decreased by a factor of 2.5.
Therefore every intermediate value is possible within
the 7-2-5 sequence.
With direct connection to the vertical input, signals up to
4OOV,, may be displayed (attenuator set to
ZOV/cm,
variable control ccw).
When pulling the variable control knob (MAG x5), the sen-
sitivity is increased by a factor of 5. Hence follows a min. deflection coefficient of 1
mV/cm
(reduced bandwidth).
In general electrical engineering, alternating voltage data
normally refers to effective values (rms = root-mean-
square value). However, for signal magnitudes and voltage
designations in oscilloscope measurements, the peak-to-
peakvoltage (V,,) value is applied. The latter corresponds to
the real potential difference between the most positive and
most negative points of a signal waveform.
Subject to change without notice
With the designations
H = display height in cm,
= signal voltage in
U
V,,
at the vertical input,
D = deflection coefficient in V/cm at attenuator switch,
the required quantity can be calculated from the two given
quantities:
=
D-H
U
H=;
D+
M3 604
Page 7
However, these three values are not freely selectable. They
have to be within the following limits (trigger threshold, ac-
curacy of reading):
H between 0.5 and
U between 1
D between
5mV/cm
D between 1 mV/cm and
(with
pulled MAG x5 knob).As a rule, all signals to be displayed are periodically repeat-
mV,,
and 16OV,,,
8cm,
if possible 3.2 to
and
20V/cm
4V/cm
8cm,
in l-2-5 sequence.
in l-2-5 sequence
Examples:
Set deflection coefficient D = 50 mV/cm 2 0.05 V/cm,
observed display height H = 4.6 cm,
required voltage U =
Input voltage U =
5V,,,
0.05.4.6
= 0.23 V,,.
set deflection coefficient D = 1 V/cm,
required display height H = 5: 1 = 5cm
Signal voltage U =
22OV,,;2.fl=
622
V,,
(voltage > 16OV,,, with probe X 10 : U = 62.2 V,,),
desired display height H = min.
3.2cm,
max.
8cm.
max. deflection coefficient D = 62.2 : 3.2 = 19.4V/cm,
min. deflection coefficient D = 62.2 : 8 = 7.8V/cm,
adjusted deflection coefficient D =
If
the applied signal is superimposed on a DC (direct
lOV/cm
voltage) level the total value (DC + peak value of the alternating voltage) of the signal across the Y-input must
not exceed
*4OOV(see
figure). This same limit applies to
normal x 10 attenuator probes, the attenuation ratio of
which allows signal voltages up to approximately 1
to be evaluated. Voltages of up to approximately
,OOOV,,
2,4OOV,,
may be measured by using the HZ53 high voltage probe
which has an attenuation ratio of 100: 1. It should be noted
that its AC
peakvalue
is derated at higher frequencies. If a
normal x 10 probe is used to measure high voltages there is the
risk that the compensation trimmer bridging the attenuator
series resistor will break down causing damage to the input
of the oscilloscope. However, if for example only the residual ripple of a high voltage is to be displayed on the oscil-
loscope, a normal x 10 probe is sufficient. In this case, an appropriate high voltage capacitor (approx. 22-68nF) must be
connected in series with the input tip of the probe.
Voltage
’
DC +
AC,,,k
=
I
peak
F@u
AC
DC
DC
.-
/\
/
/
I
Total value of input voltage ’ -
The dotted line shows a voltage alternating at zero volt level. When superim-
posed a DC level, the addition of the positive peak and the DC voltage results
in the max. voltage (DC + AC,,,,).
4OOV,,,.
-.
-1
/
‘\
AC
\
\.
\
1
1’
/
!
/
’
\
\
Time
\
\
‘\
\
\
/
\,
It is very important that the oscilloscope input coupling is
set to DC, if an attenuator probe is used for voltages higher
than 400V (see page M6: Connection of Test Signal).
Time Measurements
ing processes and can also be designated as periods. The
number of periods per second is the recurrence frequency
or repetition rate. One or more signal periods or even part of
a period may be shown as a function of the adjustment of
the
TIMEIDIV.
DIV. switch are indicated in s/cm,
cordingly, the dial is subdivided into three sectors. The duration of a signal period or a portion of the waveform is
ascertained by multiplying the relevant time (horizon-
tal distance in cm) by the time coefficient selected on
the
TIME/DIV.
knob on the
detent CAL. for accurate measurement (arrow horizontal
and pointing to the right).
With the designations
L
= displayed wave length in cm of one period,
T= time in seconds for one period,
F
= recurrence frequency in Hz of the signal,
T,
= time coefficient in s/cm on
and the relation F =
stated
:
T =
L.T,
F
-
=
L.Tc
With X-MAG.
be divided by 10.
However, these four values are not freely selectable. They
have to be within the following’limits:
L
between 0.2 and 1 Ocm, if possible 4 to 1 Ocm,
Tbetween 5 ns and 1 OS,
F
between 0.1 Hz and 60 MHz,
T,
between
(with X MAG. x 10
T,
between 5 ns/cm and
(with pushed X MAG.
Examples:
Displayed wavelength L = 7 cm,
set time coefficient T, = 0.5 ps/cm,
required period T =
c
required rec. freq. F = 1:(3.5.1
/
Signal period T =
set time coefficient T, = 0.2 s/cm,
required wavelength L = 0.5 : 0.2 =
switch. The time coefficients on the TIME/
ms/cm,
and ps/cm. Ac-
switch. The time variable control (small
TIME/DIV.
switch) must be in its calibrated
timebase
l/T,
the following equations can be
1
L
xl0
button depressed the T, value must
50ns/cm
and 1 s/cm in l-2-5 sequence
=
1
F.Tc
in out position), and
lOOms/cm
x10
7.0.5.1
in l-2-5 sequence
button).
O-” =
3.5~s
OP6)
= 286
0.5s,
2.5cm.
switch
kHz.
T,
T,
=
=
;
&
.
M4 604
Subject to change without notice
Page 8
Displayed ripple wavelength L = 1 cm,
set time coefficient T, = 10 ms/cm,required ripple freq. F = 1 : (1 .10.10-3) = 100Hz.
If magnification is used, this product must be divided by 10.
The
fall
time of a pulse can also be measured by using this
method.
TV-line frequency F = 15 625 Hz,
set time coefficient T, = 10 @cm,
required wavelength L = 1: (15 625.1
Sine wavelength L = min.
Frequency F = 1
max. time coefficient T, = 1 :
min. time coefficient T, = 1
set time coefficient T, = 0.2
required wavelength L = 1: (1
Displayed wavelength L =
set time coefficient T, = 0.5
pressed MAG X 10 button: T, = 0.05
required rec. freq. F = 1: (0.8.0.05.1
required period T = 1: (25.1
If the time is relatively short as compared with the complete
signal period, an expanded time scale should always be
applied (X MAG
tained time values have to be divided by
intervals at optional points of the signal can be measured
more exactly with the aid of the sweep delay. With it, the
display and measurement of time intervals, which are smal-
ler than 1 % of the full signal period, are possible. The smallest measurable time interval is, on the whole, dependent on
the obtainable brightness of the CRT. The limit is an expansion of approximately 1000 times. Using a Viewing Hood
HZ47, more expansion is possible, provided that the time
coefficient set on the
S@cm
basic period. Otherwise, the fastest sweep speed deter-
mines the greatest possible expansion.
When investigating pulse or square waveforms, the critical
feature is the
transients, ramp-offs, and bandwidth limits do not unduly
influence the measuring accuracy, the
measured between 10% and 90% of the vertical pulse
height. For peak-to-peak signal amplitude of
which are symmetrically adjusted to the horizontal center
line, the internal graticule of the CRT has two horizontal dotted lines
tenuator switch with its variable control together with the
Y-POS. control so that the pulse height is precisely aligned
with the 0 and 100 % lines. The 10 % and 90 % points of the
signal will now coincide with the two lines, which have a
distance of
additional subdivision of
the product of the horizontal distance in cm between
these two coincidence points and the time coefficient
setting.
(and using the X MAG x 10 facility) for the signal’s
risetime
&2.4cm
f2.4cm
kHz,
x10
from the center line. Adjust the Y at-
4cm,
max. 1 Ocm,
(4.1 03)
:(I O-1 03)
ms/cm,
03-
0.8cm,
ys/cm,
06)
= 40 ns.
button pushed). In this case, the ascer-
TIME/DIV.
of the voltage step. To ensure that
from the horizontal center line and an
0.2cm.
The
0p5)
=
6.4cm.
=
0.25ms/cm,
= 0.1
m&m,
0.2 - 1
0p3)
= 5cm.
@cm,
Ov6)
= 25 MHz,
70.
Very small time
switch is greater than
risetime
risetime
is generally
6cm
height,
is given by
100%
90%
-I
qot t-
The above figure shows correct positioning of the oscilloscope trace for accurate
With a time coefficient of O.O5ys/cm and pushed X MAG
x10
button the example shown in the above figure results
in a measured total
ttor
=
1.6cm.O.O5@cm:
When very fast risetimes are being measured, the rise-
times of the oscilloscope amplifier and the attenuator probe
have to be deducted from the measured time value. The
risetime
formula.
In this
of the oscilloscope amplifier (approx. 5.8ns), and t, the
risetime
then t,,, can be taken as the
lation is unnecessary.
Calculation of the example in
signal
of the signal can be calculated using the following
t,
=
v
ttot2 - t
ttot
is the total measured risetime,
of the probe (e.g. = 2 ns). If
risetime
t,
= V 8* -
risetime
risetime
2 -
osc
5.8* - 2*
measurement.
of
10 = 8ns
t
2
P
to,,
is the
ttot
is greater than 42 ns,
risetime
= 5.1 ns
of the pulse, and calcu-
tie
figure above results in a
risetime
Connection of Test Signal
Caution: When connecting unknown signals to the oscillo-
scope input, always use automatic triggering and set the
DC-AC input coupling switch to AC. The attenuator switch
should initially be set to POV/cm.
Sometimes the trace will disappear after an input signal has
been applied. The attenuator switch must then be turned
back to the left, until thevertical signal height is
With a
probe must be inserted before the oscilloscope’s vertical
input. If, after applying the signal, the trace is nearly
blanked, the period of the signal is probably substantially
signa!
amplitude greater than 16OV,,, an attenuator
only3-8cm.
Subject to change
wlthout
notlce
Page 9
longer than the set value on the
should be turned to the left to an adequately greater time
coefficient.
The signal to be displayed should be fed to the vertical input
of the oscilloscope by means of a shielded test cable, e.g.
the HZ32 or HZ34, or by a x 10 or x 100 attenuator probe.
The use of these shielded cables with high impedance cir-
cuits is only recommended for relatively low frequencies
(up to approx. 50kHz). For higher frequencies, and when
the signal source is of low impedance, a cable of matched
characteristic impedance (usually
In addition, and especially when investigating square or
pulse waveforms, a resistor equivalent to the characteristic
impedance of the cable must also be connected to the cable
directly at the input of the oscilloscope. When using a 509
cable, such as the HZ34, a
HZ22 is available from HAMEG. When investigating square
or pulse waveforms with fast risetimes, transient
phenomena on both the edge and top of the signal may become visible if the correct termination is not used. It must
be remembered that the
dissipate a maximum of 2 watts. This power consumption
is reached with 1
If a x 10 or x 100 attenuator probe is used, no termination is
necessary. In this case, the connecting cable is matched di-
rectly to the high impedance input of the oscilloscope.
When using attenuator probes even high internal imped-
ance sources are only slightly loaded by approximately
10 MQ I I 16 pF or 100 MQ I I7 pF respectively. Therefore,
when the voltage loss due to the attenuation of the probe
can be compensated by a higher sensitivity setting on the
HM 604, the probe should always be used. Also it should be
remembered that the series impedance of the probe provides a certain amount of protection for the input of the os-
cilloscope amplifier. It should be noted that all attenuator
probes must be compensated in conjunction with the oscil-
loscope (see: Probe Adjustment, page M8).
If a x IO or x 100 attenuator probe is used at voltages
higher than 400 V, the DC input coupling must always
be set. With AC coupling, the attenuation is frequency-de-
pendent, the pulses displayed can exhibit ramp-off, DC-volt-
age contents are suppressed - but loads the respective
input coupling capacitor of the oscilloscope. The electric
strength of which is maximum 400V (DC + peak AC). For
the suppression of unwanted DC voltages, a capacitor of
adequate capacitance and electric strength may be con-
nected before the input tip of the probe (e.g. for ripple
measurements).
It is important to remember that when low voltage signals
are being investigated the position of the ground point on
the test circuit can be critical. This ground point should always be located as close as possible to the measuring
point. If this is not done, serious signal deformation may
OV,,,
50R
5OQ
through-termination will only
or with
TIMEIDIV.
5OQ)
is recommended.
through-termination type
28V,,
sine signal.
switch. It
result from any spurious currents through the ground leads
or test chassis parts. This comment also applies to the
ground leads on attenuator probes which ideally should be
as short and as thick as possible. For connection of a probe
to a BNC socket, a BNC-adapter should be used. It forms
often a part of the probe accessory. Grounding and matching problems are then eliminated.
Hum or interference voltage appearing in the measuring circuit (especially with a small deflection coefficient) is possibly caused by multiple grounding, because equalizing currents can flow in the shielding of the measuring cables (voltage drop between non-fused earthed conductors of other
line powered devices, which are connected to the oscilloscope or test object, e.g. signal generators with anti-interference capacitors).
Operating
For a better understanding of these Operating Instructions
the front panel picture at the end of these instructions can
be unfolded for reference alongside the text.
The front panel is subdivided into three sections according
to the various functions. The INTENS., FOCUS and TR
(trace rotation) controls are arranged on the left directly
below the screen of the cathode-ray tube (CRT). Continuing
towards the right are the horizontal magnification button (X
MAG. x10), the switch for calibrator frequency selection
(1
kHz/l
MHz) and calibrator output sockets
(CAL.). The COMPONENT TESTER pushbutton and its
measuring socket are located on the right side.
The X-Section, located on the upper right, next to the screen,
contains the red POWER pushbutton and indicating LED, all
controls for
zontal trace position (X-POS.), sweep delay (DELAY), TV
separator (TV SEP.) together with the field select button
(FIELD
doff adjustment (HOLD OFF).
The lower Y-Section contains the controls for the vertical
deflection system. On the right and left in this section are located: vertical input connector, DC-AC-GD input coupling
slide switch, Y-POS. control, INVERT pushbutton, attenuator switch with variable control, and ground jack. All
these controls and connectors exist in duplicate for each of
the Channels I and II. Three pushbuttons for selecting the
operating mode are arranged below the attenuator
switches: CH
These are explained later.
The instrument is so designed that even incorrect operation
will not cause serious damage. The pushbuttons control
only minor functions, and it is recommended that before
commencement of operation all pushbuttons are in the
“out” position. After this the pushbuttons can be operated
depending upon the mode of operation required.
timebase
l/II),
the XYmode button (XV), and the knob for
l/II
(TIME/DIV.),
triggering (TRIG.), hori-
-TRIG l/II, DUAL and ADD.
0.2V/2V
hol-
M6 604
Subject to change without notice
Page 10
The HM 604 accepts all signals from DC (direct voltage) up
to a frequency of at least
voltages the upper frequency limit will be
ever, in this higher frequency range the vertical display
height on the screen is limited to approx. 6cm. The time resolution poses no problem. For example, with 100 MHz and
the fastest adjustable sweep rate (5ns/cm), one cycle will
be displayed every 2cm. The tolerance on indicated values
amounts to
be measured can therefore be determined relatively accu-
rately. However, from approximately 25 MHz upwards the
measuring error will increase as a result of loss of gain. At
40MHz
11 % should be added to the measured voltage at this frequency. As the bandwidth of the amplifiers differ (normally
between 65 and 70 MHz), the measured values in the upper
limit range cannot be defined exactly. Additionally, as al-
ready mentioned, for frequencies above
dynamic range of the display height steadily decreases. The
vertical amplifier is designed so that the transmission performance is not affected by its own overshoot.
f3%
in both deflection directions. All values to
this reduction is about 10%. Thus, approximately
60MHz (-3dB).
For
80MHz.
60MHz
sinewave
How-
the
First Time Operation
Check that the instrument is set to the correct mains/
line voltage. (Refer to page
Before applying power to the oscilloscope it is recom-
mended that the following simple procedures are performed:
-
Check that all pushbuttons are in the out position, i.e. released.
-
Rotate the three variable controls with arrows to their
calibrated detent.
-
Set the variable controls with marker lines to their mid-
range position (marker lines pointing vertically).
-
The LEVEL control knob should be on its left stop (AT).
-
The three lever switches in the X-Section should be set
to their uppermost position.
-
Both input coupling slide switches for CH.1 and
the Y-Section should be set to the GD position.
Switch on the oscilloscope by depressing the red POWER
pushbutton. An LED will illuminate to indicate working
order. The trace, displaying one baseline, should be visible
after a short warm-up period of 10 seconds. Adjust
and X-POS. controls to center the baseline. Adjust
TENS. (intensity) and FOCUS controls for medium bright-
ness and optimum sharpness of the trace. The oscilloscope
is now ready for use.
If only a spot appears (CAUTION! CRT phosphor can be
damaged.), reduce the intensity immediately and check
that the X-Y pushbutton is in the released (out) position. If
the trace is not visible, check the correct positions of all
knobs and switches (particularly LEVEL knob in AT position
and DELAY MODE lever switch to OFF).
M2).
CH.11
Y-P0S.I
IN-
in
To obtain the maximum life from the cathode-ray tube, the
minimum intensity setting necessary for the measurement
in hand and the ambient light conditions should be used.
Particular care is required when a single spot is displayed, as a very high intensity setting may cause damage
to the fluorescent screen of the CRT. Switching the oscilloscope off and on at short intervals stresses the cathode of
the CRT and should therefore be avoided.
Trace Rotation TR
In spite of Mumetal-shielding of the CRT, effects of the
earth’s magnetic field on the horizontal trace position
cannot be completely avoided. This is dependent upon
the orientation of the oscilloscope on the place of work.
A centred trace may not align exactly with the horizon-
tal center line of the graticule. A few degrees of misalignment can be corrected by a potentiometer acessi-
ble through an opening on the front panel marked
TR.
DC Balance Adjustment
The vertical preamplifiers for
matched dual
After long periods of use the FET characteristics may
change which can alter the DC balance of the vertical
amplifier. A quick check of DC Balance can be made on each
channel by pulling the fine amplitude control MAG x5 and
pushing it back. If the trace moves from the vertical position
(up or down) more than 1 mm, the DC Balance will require
readjustment. This check should be made after a 20-minute
warm-up period.
FETs
connected as input source followers.
Adjustment procedure
The following instructions should be performed to obtain
the correct DC balance adjustment of both channels.
-
Remove all input cables and adjust oscilloscope controls
to display the baseline.
-
Center the baseline using Y-POS. and X-POS. controls.
-
Set attenuator switches to
switches to GD.
-
Release all pushbuttons in the Y-Section.
-
Place the oscilloscope so that it rests
right position) and locate DC balance adjustment potentiometer access holes - marked
CH.11 -
which are found underneath the instrument.
-
Insert a screwdriver (blade approx.
20 mm) in
is located behind the hole.
-
Pull and push the
just balance pot so that the baseline no longer moves up
or down. When the trace remains steady, correction of
CH.1
-
Depress CH I/II-TRIG.
procedure for
CH.1
hole. A plastic guide with slotted bottom
is completed.
CH.II.
CH.1
variable control MAG x5 and ad-
CH.1
and
CH.11
contain
5mV/cm
l/II
button. Repeat adjustment
and input coupling
firmlyon
CH.1
its back (up-
DC-BALANCE
3mm,
length min.
Subject to change without notice
M7 604
Page 11
Use and Compensation of Probes
To
display an undistorted waveform on an oscilloscope, the
probe must be matched to the individual input impedance
of the vertical amplifier.
small insulated non-metallic screwdriver or trimming tool,
the trimmer has to be adjusted slowly until the tops of the
squarewave signal are exactly parallel to the horizontal
graticule lines. (See Fig. above for 1
amplitude shown should be 4cm + 1.2 mm (= 3 %), During
this adjustment, the signal edges will remain invisible.
The HM604’s built-in calibration generator provides a
squarewave signal with a very low
switch-selectable frequencies of approx.
at two output sockets below the CRT screen. One output
provides
0.2V,, *I
% for 10: 1 probes, and
present at the other, for 100: 1 probes.
risetime
1 kHz and 1
(<5ns), and
MHz
2V,,
+I % are
Adjustment at 1 MHz
Probes
They incorporate resonance deemphasizing networks
trimmer in conjunction with inductances and capacitors)
which permit - for the first time - probe compensation in
the range of the upper frequency limit of the vertical oscillo-
When the attenuator switches are set to 5mV/cm vertical
deflection coefficient, these calibration voltages correspond to a screen amplitude of
The output sockets have an internal diameter of
accommodate the internationally accepted shielding tube
diameter of modern
Modular
4cm.
Probes and F-series
4.9mm
slimline
probes. Only this type of construction ensures the
extremely short ground connections which are essential for
an undistorted waveform reproduction of non-sinusoidal
scope amplifier. Only this compensative adjustment
ensures optimum utilisation of the full bandwidth, together
with constant group delay at the high frequency end,
thereby reducing characteristic transient-distortion near the
to
leading signal edge (e.g. overshoot, rounding, ringing, holes
or bumps) to an absolute minimum.
Using the probes HZ51, 52, and 54, the full bandwidth of
the HM 604 can be utilized without risk of unwanted waveform distortion.
high frequency signals.
Prerequisite for this HF-adjustment is a squarewave
Adjustment at 1
The C-trimmer adjustment compensates the capacitive
loading on the oscilloscope input (approx.
HM604). By this adjustment, the capacitive division
assumes the same division ratio as the ohmic voltage
kHz
3OpF
with the
generator with fast
impedance (approx.
quency of approx. 1 MHz. The calibrator output of the
HM604 meets these requirements when the pushbutton
1 MHz is depressed.
divider to ensure an equal division ratio for high and low frequencies, as for DC. (For
set to 1: 1, this adjustment is neither required nor possible).
A baseline exactly parallel to the horizontal graticule lines is
a major condition for accurate probe adjustments. (See also
‘Trace Rotation
TR’,
Connect the probes (Types HZ51, 52, 53,
CH.1 input. All pushbuttons should be released (in the ‘out’
position), and all push-pull knobs pushed ‘in’. Set the input
coupling switch to DC, the attenuator switch to 5mV/cm,
and the
controls to CAL. position. Plug the probe tip into the appropriate calibrator output socket, i.e. IO:1 probes into the
0.2V
1
TIME/DIV.
socket, 100: 1 probes into the 2.OV socket.
kHz
incorrectcorrect
Approximately 2 complete waveform periods are displayed
on the CRT screen. Now the compensation trimmer has to
be adjusted. Normally, this trimmer is located in the probe
1: 1
page M7.)
switch to
probes or switchable probes
54,
or HZ37) to
0.2ms/cm,
and all variable
incorrect
Connect the probe
the calibrator pushbutton
should be released (‘out’ position). Set the input coupling
switch to DC, attenuator switch to
DIV. switch to 0.1
position.
Insert the probe tip into the output socket marked
waveform will be displayed on the CRT screen, with leading
and trailing edges clearly visible. For the HF-adjustment
now to be performed, it will be necessary to observe the ris-
ing edge as well as the upper left corner of the pulse top. To
gain access to the HF-compensation trimmer, the plastic
cover of the probe connecting box has to be slid off after
unscrewing the probe cable. The connecting boxes of the
HZ51 and HZ54 contain one R-trimmer screw, each, while
that of the HZ52 provides three. These R-trimmers have to
be adjusted in
top is as straight as possible. Overshoot or excessive round-
ing are unacceptable. This is relatively easy on the HZ51
and HZ54, but slightly more difficult on the HZ52. The rising
edge should be as steep as possible, with the pulse top
remaining as straight and horizontal as possible.
head. On the 100: 1 probe HZ53, however, it is located in
the connecting box at the other end of the cable. Using a
On the HZ52, each of the three trimmers has a clearly
kHz.)
The signal
HZ51,52,
and 54 will also allow for HF-adjustments.
risetime
(HZ51,52,
l&cm.
(typical 4ns). and low output
5OQ).
providing
or 54) to CH.1 input. Depress
1MHz.
Set all variable controls to CAL.
0.2V
and 2V at a fre-
All other pushbuttons
5mV/cm,
and TIME/
such a manner that the beginning of the pulse
0.2V.
(R-
A
M8 604
Subject to change without notice
Page 12
defined area of influence on the waveform shape (see Fig.),
offering the added advantage of being able to ‘straighten
out’ waveform aberrations near the leading edge.
The adjustment sequence must be followed in the order
described, i.e. first at 1
frequencies should not be used for timebase calibrations.
The pulse duty cycle deviates from 1 : 1 ratio.
Adjustment points of the probes
HZ51, HZ54
Prerequisites for precise and easy probe adjustments, as
well as checks of deflection coefficients, are straight horizontal pulse tops, calibrated pulse amplitude, and
osc.
I
Tz (I+)
1
CAL.
potential at the pulse base. Frequency and duty cycle are
relatively uncritical.
(NF) T,
response,
generator outputs are of particular importance.
kHz,
then at 1 MHz. The calibrator
zero-
For interpretations of transient
fast pulse risetimes and low-impedance
(I-F) T,
T3 T, T,
T,_:
alters the middle frequencies
Ti: alters the leading edge
T,: alters the lower frequencies
T,
(NF) 1
-
IOnskm
-
HZ52
T, (LF)
After completion of the HF-adjustment, the signal
amplitude displayed on the CRT screen should have the
same value as during the 1
kHz
adjustment.
Probes other than those mentioned above, normally have a
larger tip diameter and may not fit into the calibrator outputs Whilst it is not difficult for an experienced operator to
build a suitable adapter, it should be pointed out that most
of these probes have a slower risetime with the effect that
the total bandwidth of scope together with probe may fall
far below that of the HM604. Furthermore, the
HF-adjust-
ment feature is nearly always missing so that waveform distortion can not be entirely excluded.
Providing these essential features, as well as switch-select-
able output-frequencies, the calibrator of the HM 604 can,
under certain conditions, replace expensive squarewave
generators when testing or compensating
wideband-
attenuators or -amplifiers. In such a case, the input of an
appropriate circuit will be connected to one of the
CAL.-out-
puts via a suitable probe.
The voltage provided at a high-impedance input (I
5OpF)
will correspond to the division ratio of the probe used
(10:
1 =
20mV,,,
Suitable probes are
For low-impedance inputs (e.g. 50
100:
HZ51,
1 =
also
20mV,,
52, 53, and 54.
Q),
from 2V output).
a 1: 1 probe can be
MSJ
II
15-
employed which, however, must be fully terminated with a
5052
through-termination. Suitable probe types are HZ50
and HZ54. The latter must be switched to the 1: 1 position,
and the HF-trimmer in the connecting box turned fullycoun-
terclockwise.
When connected to the
0.2V
CAL. socket, and using the
HZ50, this arrangement will provide approx. 40mV,, at
50Q circuit input, and approx.
24mV,,
if the HZ54 is used.
The voltages given here will have larger tolerances than 1 %
since operation of a 1: 1 probe together with a
5OQ
load is
very uncommon.
incorrect
Adjustment
1 MHz
Subject to change without notice
correct
incorrect
Using the 2V CAL. socket under similar conditions is only
possible with the HZ54 probe. The potential obtained at the
5OQ
input will then be approx.
190mV,,,
but with almost
twice the risetime. Accurate readings of the available input
voltage can be shown directly on the HM604 when con-
necting a
5OQ
through-termination between the BNC plug
of the probe and the input of the oscilloscope.
Operating Modes of the Y Amplifier
The required operating modes are selected on three
pushbuttons located in the Y-Section. For Mono operation
all pushbuttons should be in the out position, the instrument is then operating on Channel/only.
M9
604
Page 13
For Mono operation with Channel
pushbutton has to be pressed. When the DUAL button is
depressed, the HM604 is in
mode, the channels are displayed consecutively (alternate
mode). This mode is not suitable for the display of very low
frequency signals
or jump. Under these conditions,
depressed additionally selecting chopped mode. In this
position, both channels then share the trace during each
sweep period. For the display of high frequency signals, the
type of channel switching selected is less important.
To select the add mode only the ADD button should be
depressed. The signals on both channels are then added
together. If in this mode one channel is inverted (pushbutton INVERT depressed), then the difference between the
two channels is displayed. For both of these operating
modes, the vertical position of the trace depends on the setting of the Y-POS. controls of both channels.
(<l kHz),
as the tracewill appear to flicker
II,
the CH
Dualchannel
the ADD
l/II
-TRIG.
operation. In this
button should be
l/II
nal frequencies up to 120
quency the inherent phase difference between the vertical
and horizontal system makes accurate measurements difficult. In this mode, one of the
horizontal deflection (X) while the other signal provides the
vertical deflection (Y).
0”
The phase angle between the two signals can be deter-
mined from the Lissajous pattern as follows:
cos
cp
35”
=
1/ I-@
kHz.
sin cp =
However, above this fre-
sinewave
signals provides
90”180”
E
Differentialmeasurements
urement of the voltage drop across floating components
(both ends above ground). Two identical probes should be
used for both vertical inputs. Using a separate ground connection and notconnecting the probe or cable shields to the
circuit under test avoid ground loops (hum, common-mode
disturbances).
X-Y Operation
For X-Yoperation, the pushbutton in the X-Section marked
X-Y must be depressed. The X signal is then derived from
the
Channe///(HOR.
during X-Y operation is determined by the setting of
the Channel II input attenuator and variable control.
This means that the sensitivity ranges and input impedances are identical for both the X and Y axes. However, the
Y-POS.II
is taken over by the X-POS. control. It is important to note
that the X MAG.
the sweep, should not be operated in the X-Y mode. It
should also be noted that the bandwidth of the X amplifier
is approximately
phase difference between both axes is noticeable from
50
The Y-Input signal may be inverted by using the INVERT
(channel I) facility.
control is disconnected in this mode. Its function
kHz
upwards.
INP.). The calibration
x10
facility, normally used for expanding
5MHz
X-Y Phase Measurements
The X-Y phase measurement method can be used to meas-
ure the phase difference between two signals of the sameence signal occupies exactly 10 divisions (see next figure).
frequency. This provides a method of measurement for sig-
techniques allow direct meas-
ofthexsignal
(-3dB), and therefore an increase in
= arc sin
%J
This simple formula works for angles less than
independent from both deflection amplitudes on the
screen.
Caution!
If a single spot appears (both deflection voltages are missing)
reduce the intensity immediately, as a high intensity setting
may cause damage to the fluorescent screen of the CRT.
f
b
90”,
it is
Dual-Trace Phase Difference Measurements
Phase comparison between two signals of the same fre-
quency can be made using the dual-trace feature (DUAL
button depressed). This method of phase difference meas-
urement can be used up to the frequency limit of the vertical
system. To make the comparison, use the following procedure:
Set the Input Coupling switches to the same position, and
the CH. I/II-TRIG.
reference signal (Phase 0”) is connected. Select ALT. chan-
nel switching for frequencies above 1
frequencies below 1
time delay to connect the signals to the input connectors.
Set the Input Attenuator switches and the CH I and CH II var-
iable controls so that the displays are approximately equal
and about five divisions in amplitude. Set the TIME/DIV.
switch to a sweep rate which displays about one cycle of
the waveform. Move the waveforms to the
graticule with the Y-P0S.I and
Turn the Variable Time Control until one cycle of the refer-
Each division represents 36” of the cycle.
l/II
pushbutton to the channel where the
kHz,
and CHOP. for
kHz.
Use probes which have equal
center
Y-POS.II
controls.
of the
Ml0 604
Subject to change without notice
Page 14
Figure 2
Amplitude modulated oscillation: F = 1 MHz; f = 1 kHz;
m = 50 %
; UT =
28.3
mVrms.
Dual-Trace Phase Difference Measurements
T =
Horizontal distance
t
= Horizontal distance of zero-crossing points (cm).
foroneperiod(cm).
Assume a horizontal difference of 3 divisions (t = 3cm) and
a period of
10
divisions (T = 1 Ocm), the phase difference
91
can be calculated using the following formula:
=
108”
l.885rad
or
arcg,
-2~t =$ a2~~
=f
respectively.
Measurement of an amplitude modulation
The momentary amplitude u at time t of a HF-carrier voltage, which is amplitude modulated without distortion by a
sinusoidal AF voltage, is in accordance with the equation
u
=
U,.
sinQt
+
0,5m
. UT - cos(S&w)t -
where
= unmodulated carrier amplitude
U,
S2= 2~rF
0
=2nf
= angular carrier frequency
= modulation angular frequency
m= modulation factor (S 1 P 100
The lower side frequency F-fand the upper side frequency
F+f
arise because of the modulation apart from the carrier
frequency
F.
T
0,5m*
Figure 1
Amplitude and frequency spectrum for AM display (m = 50%)
UT
4
I
F-f F F+f
The display of the amplitude-modulated
evaluated with the oscilloscope provided the frequency
spectrum is inside the oscilloscope bandwidth. The time
base is set so that several cycles of the modulation frequency are visible. Strictly speaking, triggering should be external with modulation frequency (from the AF generator or a
demodulator). However, internal triggering is frequently possible with normal triggering using a suitable LEVEL setting
and possibly also using the time variable adjustment.
0,5m
- UT -
cos(l(r+o)t
%).
UT
0,5m
-
UT
4
HF
oscillation can be
Oscilloscope setting for a signal according to figure 2:
Depress no buttons.
TIM
E/DIV. :
0.2
Triggering: NORMAL with LEVEL-setting; internal (or
Y:
ms/div.
CH. I; 20mV/div; AC.
external) triggering.
If the two values
ulation factor is calculated from
a-
m =
a+b
aand
b
resp. m =
bare read from the screen, the mod-
a*.
100
[%J
where a = UT (l+m) and b = UT (l-m).
The variable controls for amplitude and time can be set arbitrarily in the modulation factor measurement. Their position does not influence the result.
Triggering and Timebase
With the LEVEL knob in locked position (turned ccw to AT
position = automatic triggering), a baseline is displayed con-
tinuously even when no signal is present. In this position it is
possible to obtain stable displays of virtually all uncomplicated, periodically repeating signals above 30 Hz. Adjustments of the
With normal triggering (LEVEL knob not in AT position) and
LEVEL adjustment, triggering of time/div. deflection can be
set in any point of a given signal. The triggering range which
can be set with the LEVEL control depends greatly on the
amplitude of the displayed signal. If it is less than 1 div, then
the range is quite small and performance of settings requires a delicate touch.
If the
LEVEL
ble.
In order to obtain a satisfactory stable display, the
must be triggered synchronously with the test signal. The
trigger signal can be derived from the test signal itself,
when internal triggering is selected, or from a frequency related signal applied to the external trigger input.
timebase
then are limited to
timebase
setting.
control is incorrectly set, no trace will be visi-
timebase
Subject to change without notice
Ml1 604
Page 15
Triggering can be selected on either the rising or falling edge
of the trigger signal depending on whether the SLOPE +/pushbutton (next to
out position, triggering from the positive-going edge is
selected. The correct slope setting is important in obtaining
a display when only a portion of a cycle is being displayed.
With internal
the Y amplifier, the trigger signal is derived from the respective channel in use. In the Dualchannelmode, the internal
trigger signal may be selected from either Channel I or
Channel//using the
sition, the trigger signal is derived from Channel I. However,
it is always preferable to trigger from the less complicated
signal.
With
internalalternate
X-Section depressed) in the DUAL channel alternate mode
of the Y amplifier, the trigger voltage is derived alternately
from Channel I and Channel
ticularly useful when two asynchronous signs/s are being
investigated. Normal triggering should be preferable in this
mode. The display of one signal only is not possible on the
alternate trigger mode.
For
exfernaltriggering,
tion must be depressed. The sync. signal
must then be fed to the TRIG. INP. input socket.
Coupling mode and frequency range of the trigger signal
are selected with the TRIG. lever switch in the X-Section for
internal and external triggering, provided that the TV SEP.
switch is in off position. The HM 604 has 4 coupling modes:
AC, DC, LF, HF. The AC coupling mode is mainly used. DC
trigger coupling is only recommended, when very low fre-
quency signals are being investigated and triggering at a
particular value is necessary, or when pulses, which significantly change in duty cycle during observation time,
have to be displayed. If DC coupling is selected, it is advisable to use the normal triggering mode. In the HF coupling
mode, a high pass filter is switched into the trigger
amplifier. This filter cuts off the DC content of the trigger
signal and the lower frequency range.
LEVEL)is
triggering
in the out or in position. In the
in the Mono channel mode on
CHMI-TRIG.I/II
triggering (ALT pushbutton in the
the EXT. pushbutton in the X-Sec-
button; in the out po-
II.
This trigger mode is par-
(0.05V,,-0.5VJ
frequency component in complex waveforms. Therefore it
is especially suited for the measurement of small ripple volt-
ages from power supply rectifiers or of magnetic or static
leakage
In some countries, the standard power plug has symmetri-
cally arranged plugs (interchanging of Line and Neutral is
possible). In such cases, the SLOPE
indicate the wrong polarity compared with the display (triggering with falling edge instead of rising edge). For correction, the power plug of the instrument has to be turned.
fields~in
a circuit.
+/-
pushbutton may
Triggering of video signals
The built-in active TV-Sync-Separator separates the sync
pulses from the video signal, permitting the display of distorted video signals either in line (H = horizontal) or in frame
(V = vertical) trigger mode. The TV lever switch has five po-
sitions: the OFF position is for normal operation.
The TV: H+ and H- positions (horizontal= line) and the
TV: V+ and V- (vertical
video triggering. In these four positions the TRIG. coupling
switch and the LEVEL control (in NORM. trigger mode) are
inoperative. In the TV: V+ and V- positions (frame trigger-
ing), a low-pass filter or integrating network is connected
into circuit, which forms a trigger pulse sequence with
frame frequency from thevertical sync pulses (incl. pre-and
postequalizing pulses).
When in V mode, it is possible to select field I or II by releas-
ing or depressing FIELD
For correct video triggering, the + and - positions at V and
H must be selected corresponding to the video input signal.
If the sync pulses are placed above the picture content, H+
or V+ should be in use. For sync pulses below the picture
content of the input signal, correct triggering, without any
influence from changing picture contents, will be possible
only in V- or H-setting. The INVERT pushbutton only
changes the display on the CRT, not the input signal.
=frame)
l/II
pushbutton.
positions are used for
In the LF coupling mode, a low-pass
the trigger amplifier. This filter cuts off any amplifier noise
and the frequency range of the trigger signal above 50
For the purpose of line triggering (TRIG. lever switch in the
X-Section) to N, a (divided) secondary voltage of the power
transformer is used as a trigger signal. This trigger mode is
independent of the signal amplitude or display height and al-
lows a display below the (internal) trigger threshold. Line
triggering is recommended for all signals which are time-re-
lated (multiple or submultiple) to the mains/line frequency
or when it is desirable to provide a stable display of a line-
Ml2 604
filter
is switched into
kHz.
In TV: H trigger mode, the trigger point lies on the starting
edge of a sync pulse if SLOPE button is in + position. As
mentioned before, in TV: V mode an integrating network is
additionally added to the sync separator which delays the
formed trigger pulse by about 50~s.
Video signals are triggered in the automatic mode. Therefore the adjustment of the trigger point is superfluous. The
internal triggering is virtually independent of the display
height, which may differ from 0.8 to 8div. As opposed to AT
mode, when in normal mode, the screen is blanked without
signal at the input (turning the LEVEL knob is ineffectual).
Subject to change without notice
Page 16
Function of var. HOLD-OFF control
If it is found that a trigger point cannot be located on extremely complex signals even after repeated and careful adjustment of the LEVEL control in the Norma/ Triggering
mode, a stable display may be obtained using the HOLD-
OFF control (in the X-Section). This facility varies the hold-
off time between two sweep periods up to the ratio >5 :
Pulses or other signal waveforms appearing during this off
period cannot trigger the timebase. Particularly with burst
signals or aperiodic pulse trains of the same amplitude, the
start of the sweep can be shifted to the optimum or required moment. After specific use the HOLD-OFF control
should be re-set into its calibration detent min., otherwise
the brightness of the display is reduced drastically.
,_
~
heavy parts
cyc,e
“;e
displayed
In DEL. TRIG. mode
andN
SEP. switch in H or V position,
after delay triggering to the next following line is possible.
Therefore discrete lines are representable. The slope is appointed by the TV+ or N- position of the delay switch.
When not in TV-trig. mode the susceptibility to interference
(sense) can be influenced (A = normal, v = reduced).
Operation of the sweep delay is relatively easy, as normally
1.
only 3 controls in the X-Section need to be used: the DELAY
operating mode lever switch (OFF-SEARCH-DELAY-DEL.
TRIG.), the DELAY rotary switch (delay time range), and its
variable control VAR. IO:1 (small knob on the DELAY
switch). The latter, a twenty-turn precision potentiometer
with overwind protection, can increase the delay time range
tenfold. An LED near the DELAY mode switch indicates the
operating mode.
For reliable operation of the sweep delay, it is recommended that the following procedure should always be
adopted; also reference to the accompanying figures will
be of assistance.
Figure 1
HOLD-‘OFF time
I
v
Fig. 1 shows a case where the HOLD-OFF knob is in the Xl position and var-
ious
different
observation unsuccessful.
Fig. 2 shows a case where only the desired parts of the signal are stably dis-
played.
waveforms are overlapped on the screen, making the signal
Sweep Delay / After Delay Triggering
With the sweep delay, the start of the sweep can be de-
layed from the trigger point by a selectable time (IOOns to
maximum 1 s). It is therefore possible to start the sweep at
practically any point of a waveform. The interval, which follows the start of the sweep, can be greatly expanded by the
increase of the sweep speed. From the 5 p/cm
range downwards to slower sweep speeds, an expansion
of at least 100 times, and with the aid of the X MAG. x10 ex-
pansion of 1000 times, is possible. With time coefficients
higher than
5@cm,
the maximum expansion increases
proportionally. However, with increasing the expansion, the
display brightness decreases. Under very high ambient light
conditions a Viewing Hood like HZ47 can overcome this
problem. It should be noted that there are some difficulties
with higher expansions, if the test signal has inherent jitter.
To reduce or eliminate this jitter, expanded parts of a signal
can be triggered again “after delay” provided there is
another suitable edge (DEL. TRIG.).
TIMEIDIV.
MODE
TIME/DIV.
LED
:
NORM.
:
0.5
ms/cm
:
off
Initially, the sweep delay mode lever switch should be set in
the OFF position. In this mode, the complete waveform
under investigation will be displayed as for normal oscilloscope operation. The mode indicator LED is not illuminated
in OFF mode. The time coefficient on the
TIME/DIV.
switch
is selected so that 1 to 3 basic periods of the signal are
displayed. A larger number unnecessarily decreases the
brightness and maximum expansion. The display of only a
portion of a period limits the choice of the expanded time interval and possibly complicates the triggering. On the other
hand, the range of 1 to 3 basic periods can always be set unconstrainedly with the
TIME/DIV.
switch. In doing so, the
x10 expansion must be switched off temporarily (X
MAG.
x10
button is in out position). In the X-Section, the
HOLD-OFF control should be set to min. and the variable
control to CAL.The LEVEL control is adjusted so that stable
triggering is ensured (TRIG. LED is illuminated).
The mode switch should now be set to the SEARCH position; it will be seen that the start of the display will shift to
the right. The amount of shift indicates the exact delay time.
Subject to change without notice
Ml3 604
Page 17
If a display is not obtained in this mode, then a lower delay
time range should be selected. For example, when investigating the waveform shown in the figures, a display could
not be obtained with a delay time setting of
IOms,
as the
display is completely blanked. However, as a result of setting the DELAY rotary switch to 0.1
ps,
the shifting is not
visible. The DELAY range switch should then be rotated
clockwise until the display starts just prior to the short time
interval to be investigated. The precise adjustment to the
start is done with the VAR.
1O:l
delay time control. The
rotating range of the latter has no stop. On the range limits
a certain snapping noise is audible. Initially, this control
should be set in the left start position. In the SEARCH
mode, the LED indicator will flash.
Figure 2
If the
timebase
sweep speed is increased (rotate TIME/
DIV. switch clockwise), then the short time interval will be
expanded. It may be found that, as the amount of expansion
is increased, the trace will tend to shift. If this happens, the
VAR. delay time control can be readjusted - also subsequently at any time - to enable the exact point of interest
to be displayed.
In the example shown in figure 4, it can be seen that an expansion of
sweep speed from
x10
was obtained by increasing the
O.Sms/cm
to 5Ops/cm. Also the pre-
timebase
cise measurement for the delayed portion of the waveform
is possible. In the example, this was found to be
250~s
on
multiplication of the horizontal length in cm (of an optional
signal section) by the time coefficient just adjusted.
Figure
4
MODE
DELAY
range
TIME/DIV.
LED
Delay time = 2.5cm
: SEARCH
:lms
: 0.5
mskm
:
flashing
.
0.5mskm
= 1.25ms
In figure 2 it can be seen that the delay time is also measurable as the blanked portion or apparent shift of the start of
the trace. This time can be determined by multiplication of
(the horizontal shifting in cm) by (the time coefficient set on
the
TIME/DR!.
switch).
Now the mode switch can be set to DELAY. In this mode,
the LED is permanently illuminated. The display will now
shift to the left and the trace will commence in the same po-
sition as for a normal display; however, the short time inter-
val under investigation now starts on the first or left vertical
graticule line.
Figure 3
MODE
DELAY
range
TIME/DIV.
LED
Expansion
T =
5cm. 50@cm
T-
:
DELAY
:lms
: 50
@cm
:
illuminated
:o.5~10-3:50~10-6=10
=
250~s
Operation of the sweep delay requires a constant trigger
point. All signals, which have a constant phase shift between the expanded section and trigger point, pose no
problems. This means all electrical signal shapes, which
contain signal edges of the same polarity and with
trigger-
able level values, which are constantly repeated with the recurring frequency.
If there is no constant phase shift, the triggering may fail
after switching from the SEARCH to DELAY position or
with changing of the time coefficient. It is best to attempt to
find a trigger point, which has a constant phase shift up to
the signal section to be expanded in the OFF mode. With
complicated composite signals, the display of the basic
period could become superimposed by other signal portions. These dissappear as a rule when the sweep is in-
creased. Otherwise, a stable expanded display is obtained
by adjusting the LEVEL and the variable sweep control or by
means of the HOLD-OFF control.
Ml4 604
MODE
DELAY
range
TIME/DIV.
LED
:
DELAY
:lms
:
0.5
ms/cm
:
illuminated
Using the X MAG.
desired signal section is possible without
x10
button, a tenfold expansion of the
any
change of triggering or timebase. This can be of assistance with complicated or difficult-to-trigger signals.
Subject to change without notice
Page 18
Operation of the sweep delay needs some experience, par-
ticularly with composite signals. However, the display of
sections from simple signal waveforms is easily possible. It
is recommended to operate only the sequence
SEARCH-DELAY,
because otherwise location of the short
OFF-
time interval to be investigated will be relatively difficult.
and the timebase generator inoperative. A shortened horizontal trace will be observed. It is not necessary to disconnect scope input cables unless in-circuit measurements are
to be carried out. In the
be operated are
controls and settings have no influence on the test operation.
The sweep delay facility can be used in the following
modes: Mono, Dual, and
Algebraic
Addition
(+l+ll).
For the component connection, two simple test leads with
4mm @ banana plugs, and with test prod, alligator clip or
sprung hook, are required. The test leads are connected to
Delay Mode Indication
the insulated CT socket and the adjacent ground socket in
the Y-Section. The component can be connected to the test
Both operating modes of the sweep delay are indicated
leads either way round.
with an LED, located to the right of the DELAY mode lever
switch. In
dication of the temporary operating state. The
SEARCH
position, the LED will flash. This is an in-
DELAY
posi-
After use, to return the oscilloscope to normal operation, re-
lease the
tion is indicated by constant lighting of the LED. However,
should this be noted, and normal operating mode is required then the change-over of the lever switch to its
OFF
position has been overlooked. This could cause errors in displaying a signal by complete or partial blanking. This indication, therefore, should be closely observed.
Test Procedure
Caution! Do not testanycomponent in live circuitry - re-
move all grounds, power and signals connected to the
component under test. Set up Component Tester as
stated above. Connect test leads
tested. Observe oscilloscope display.
CTmode,
INTENS.,
COMPONENT TESTER
the only controls which can
FOCUS,
pushbutton.
and
X-POS..
acoss
component to be
All other
Component Tester
General
The HM 604 has a built-in electronic Component Tester (abbreviated CT), which is used for instant display of a test pat-
tern to indicate whether or not components are faulty. The
CT can be used for quick checks of semiconductors (e.g.
diodes and transistors), resistors, capacitors, and inductors.
Certain tests can also be made to integrated circuits. All
these components can be tested in and out of circuit.
The test principle is fascinatingly simple. The power transformer of the oscilloscope delivers a sine voltage, which is
Only discharged capacitors should be tested!
A built-in quick-acting fuse protects the
scope against mis-operation, e.g. device under test not dis-
connected from mains/line supply. In that case the fuse will
blow. For fuse replacement the oscilloscope has to be
opened (see service instruction page Sl “Instrument Case
Removal”). The fuse is located on the bottom side of the in-
strument (close to the CT pushbutton). Make sure that only
fuses of the specified type are used for replacement:
5x20mm,
41661).
applied across the component under test and a built-in fixed
resistor. The sine voltage across the test object is used for
the horizontal deflection, and the voltage drop across the resistor (i.e. current through test object) is used for vertical deflection of the oscilloscope. The test pattern shows a cur-
rent-voltage characteristic of the test object.
Since this circuit operates with mains/line frequency (50 or
60 Hz) and a voltage of
ing range of the
CTis
nent under test is limited to a range from
8.5V
max. (open circuit), the indicat-
limited. The impedance of the compo-
2OQ
to 4.7
Below and above these values, the test pattern shows only
short-circuit or open-circuit. For the interpretation of the dis-
played test pattern, these limits should always be borne in
mind. However, most electronic components can normally
Test Pattern Displays
Page M 18 shows typical test patterns displayed by the vari-
ous components under test.
-
-
Testing Resistors
If the test object has a linear ohmic resistance, both deflecting voltages are in the same phase. The test pattern expected from a resistor is therefore a sloping straight line. The
kS2.
angle of slope is determined by the resistance of the resistor
under test. With high values of resistance, the slope will tend
towards the horizontal axis, and with low values, the slope
will move towards the vertical axis.
be tested without any restriction.
Values of resistance from 20Q to
Using the Component Tester
The CT is switched on by depressing the COMPONENT
TESTER
pushbutton. This makes the vertical preamplifier
mately evaluated. The determination of actual values will
come with experience, or by direct comparison with a com-
ponent of a known value.
CTand
the oscillo-
quick-acting,
Open circuit is indicated by a straight horizontal line.
Short circuit is shown by a straight vertical line.
25OV, C.
50mA
#.7&Q
(IEC
127/ll
or DIN
can be approxi-
Subject to change without notice
I
Ml5 604
Page 19
Testing Capacitors and Inductors
Capacitors and inductors cause a phase difference between current and voltage, and therefore between the X
and Y deflection, giving an ellipse-shaped display. The position and opening width of the ellipse will vary according to
the impedance value (at 50 or
60Hz)
of the component
under test.
A horizontal ellipse indicates a high impedance or a relatively small capacitance or a relatively high inductance.
A vertical ellipse indicates a small impedance or a rela-
tively large capacitance or a relatively small inductance.
A sloping ellipse means that the component has a considerable ohmic resistance in addition to its reactance.
The values of capacitance of normal or electrolytic
capacitors from
U.If#to 1OUOpFcan
be displayed and approximate values obtained. More precise measurement
can be obtained in a smaller range by comparing the
capacitor under test with a capacitor of known value. Inductive components (coils, transformers) can also be tested.
The determination of the value of inductance needs some
experience, because inductors have usually a higher ohmic
series resistance. However, the impedance value (at 50 or
60 Hz) of an inductor in the range from 20 S2 to 4.7 kQ can
easily be obtained or compared.
diodes. Possibly
only
a small portion of the knee is visible.
Z-
diodes always show their forward knee and, up to approx.
1 OV, their Z-breakdown, forms a second knee in the oppo-
site direction. A Z-breakdown voltage of more than
12V
can
not be displayed.
i
I’
I
-I
Z-Diode 12 V
Cathode-Anode
(CT-GD)
Type:
Terminals:
Connections:
i
--
.t
I
Normal Diode HighVoltage Diode
Cathode-Anode
(CT-GD)(CT-GD)
-7
Cathode-Anode
i
I
The polarity of an unknown diode can be identified by comparison with a known diode.
Testing Transistors
Three different tests can be made to transistors: base-emit-
ter, base-collector and emitter-collector. The resulting test
patterns are shown below.
The basic equivalent circuit of a transistor is a Z-diode between base and emitter and a normal diode with reverse polarity between base and collector in series connection.
There are three different test patterns:
N-P-N Transistor:
__+_
-+
Testing Semiconductors
Most semiconductor devices, such as diodes, Z-diodes,
transistors,
FETs
can be tested. The test pattern displays
vary according to the component type as shown in the figures below.
The main characteristic displayed during semiconductor
testing is the voltage dependent knee caused by the junction changing from the conducting state to the non conduct-
ing state. It should be noted that both the forward and the
reverse characteristic are displayed simultaneously. This is
always a two-terminal test, therefore testing of transistor
amplification is not possible, but testing of a single junction
is easily and quickly possible. Since the CT test voltage
applied is only very low (max.
8.5V,,,),
all sections of most
semiconductors can be tested without damage. However,
checking the breakdown or reverse voltage of high voltage
semiconductors is not possible. More important is testing
components for open or short-circuit, which from experience is most frequently needed.
Testing Diodes
Diodes normally show at least their knee in the forward
characteristic. This is not valid for some high voltage diode
types, because they contain a series connection of several
II
Terminals:
Connections:
P-N-PTransistor:
Terminals:
Connections:
-+-
I
b-eb-c
(CT-GD)(CT-GD)
.J__
I
I
e-c
(CT-GDI
1 \
I
CC;-:D,
For a transistor the figures b-e and b-c are important. The figure e-c can vary; but a vertical line only shows short circuit
condition.
These transistor test patterns are valid in most cases, but
there are exceptions to the rule (e.g. Darlington,
FETs).
With
the CT, the distinction between a P-N-P to a N-P-N transistor is discernible. In case of doubt, comparison with a
known type is helpful. It should be noted that the same
socket connection (CTor ground) for the same terminal is
then absolutely necessary. A connection inversion effects a
rotation of the test pattern by 180 degrees round about the
center
point of the scope graticule.
Ml6 604
Subject to change without notice
Page 20
Pay attention to the usual caution with single
ponents relating to static charge or frictional electricity!
MOS-com-
In-Circuit Tests
Caution! During in-circuittests make sure the circuit is
dead. No power from mains/line or battery and no signal
inputs are permitted. Remove all ground connections including Safety Earth (pull out power plug from outlet).
Remove all measuring cables including probes between
oscilloscope and circuit under test. Otherwise the connection of both CT test leads is not recommended.
In-circuit tests are possible in many cases. However, they are
not so well-defjned. This is caused by a shunt connection of
real or complex impedances - especially if they are of relatively low impedance at 50 or 60Hz - to the component
under test, often results differ greatly when compared with
single components. In case of doubt, one component terminal may be unsoldered. This terminal should then be connected to the insulated
the test pattern.
Another way is a test pattern comparison to an identical circuit which is known to be operational (likewise without power
and any external connections). Using the test prods, identical
test points in each circuit can be checked, and a defect can be
determined quickly and easily. Possibly the device itself
under test contains a reference circuit (e.g. a second stereo
channel, push-pull amplifier, symmetrical bridge circuit),
which is not defective.
The test patterns on page M 18 show some typical displays
for in-circuit tests.
CTsocket
avoiding hum distortion of
Miscellaneous
A posifive-going sawtooth voltage of approximately
coincident with display’s sweep time is available at a BNC
output connector on the rear panel. This ramp output is
marked with M. The load impedance should not be less
than 10
not required, a capacitor should be connected in series with
the output. The ramp output can be used for different
measuring tasks in combination with the oscilloscope and
other instruments, triggering of signal sources, swept-frequency signal generators and so on.
The oscilloscope also contains a vertical output with BNC
connector marked Y on the rear panel. The output voltage is
?-50mV,,/cm
from the vertical amplifier like the trigger signal and it is
similarly switchable. Channel I or II is selected with the
CHI/II-TRIGI/II
switching (DUAL button in the Y-Section depressed) and alternate triggering (ALT. button in the X-Section depressed),
the vertical output is consecutively driven (in time with the
sweep period) from Channel I and Channel II. The vertical
output is not dependent on the vertical trace position. It
does not respond to the adjustment of the Y-P0S.I and Y-
POS.II
buttons. The vertical output is DC coupled and has approximately zero level to ground. The bandwidth of the output is
approx. 60 MHz (with 50 Q termination).
k!2
II 47
pF.
If the DC potential of the ramp output is
display height (into
pushbutton. With alternate channel
controls and to the depressing one of the INVERT
5OQ).
It is picked off
SV,
Subject to change without notice
Ml7 604
Page 21
Short
circuit
Resistor 510 Q
Single Transistors
Junction B-C
Test patterns
Junction B-E
Mains transformer prim.
Single Diodes
Z-diode under
8V
Capacitor 33
Z-diode beyond 12V
vF
Junction E-C
In-circuit Semiconductors
Diode paralleled by
6800
2 Diodes antiparallel
FET
Silicon diodeGermanium diode
Rectifier
Thyristor G + A together
Ml8 604
Diode in series with 51 QB-E paralleled by6808
B-E with 1 PF + 68052Si-diode with 10
Subject to change without notice
PF
Page 22
First Time Operation
Connect the instrument to power outlet. Switch on POWER pushbutton. No other button is depressed.
LED indicates operating condition.
Case, chassis, and all measuring connectors are connected
the Safety Earth Conductor (Safety Class I).
TRIG. selector switch to AC, TV SEP. switch to OFF, LEVEL knob in AT position (Automatic Triggering)
DELAY lever switch to OFF. and HOLD-OFF control
Adjust
INTENS.
Center trace on screen using X-POS. and Y-P0S.I controls. Then focus trace using FOCUS control.
control for average brightness.
Operating Modes of the Vertical System
Channel I: All pushbuttons in out position.
Channel
Triggering from positive-going signal edge: SLOPE
Triggering from negative-going signal edge: SLOPE
Internal triggering: select Channel with button
Internal alternate triggering: ALT button is depressed (only with alternate channel switching).
External triggering from
Line triggering:
Trigger coupling selected with
Video signals with line freq.: TV SEP. switch to H+ or H-.
Video signals with frame freq.: TV SEP. switch to V+ or V-.
First or second half of frame selection with
Pay attention to TRIG. indication LED!
Triggering:
This facility is important when only a portion of a cycle is being displayed.
External trigger signal: 50mV-0.5V,,, time-related to vertical input signal.
Trig. freq. range:
LEVEL
TRIG.
AC
turnded cw. Trace visible when triggered.
TRIG. INP.
selector switch in N position.
TRIG.
and
DC
to 20 MHz,
ATposition.
connector:
switch
HF
Short Instruction for HM604
to
min.
buttons depressed. Signals <I
(CH. I) buttons depressed.
(CH. II) buttons depressed.
Trace always visible.
+/-
button is in out position.
+/-
button depressed.
CH.I/II-TRIG.I/II.
EXT.
button depressed.
AC-DC-LF-HF.
above
20
MHz,
LF
below 50
FlELDI/II
switch.
kHz.
kHz
with CHOP.
Measuring
Connect test signal to
Compensate attenuator probe using
Select
AC
or DC input coupling. GD: Y amplifier is disconnected from input and grounded.
Adjust required display height of signal with attenuator switch and variable control.
Select sweep speed with
Adjust trigger point with
Calibrated amplitude measurement with attenuator variable control to
Calibrated time measurement with
Trigger complex or aperiodic signals using
Horizontal xl 0 expansion: X
External horizontal deflection: (X-Y operation) with X-Y button depressed (X input via
Sweep Delay Operation
OFF:
SEARCH:
DELAY:
DEL TRIG.:
normal oscilloscope operation. DELAY LED not illuminated.
use DELAY range switch and VAR.
displayed wave-form. DELAY LED flashing.
delayed signal now displayed. Expansion obtained by rotating
clockwise. Press X MAG. x10 button if necessary. DELAY LED illuminated.
After Delay Triggering; together with TV
Component Tester
Press
COMPONENT TESTER
/n-circuit test: Circuit under test must be disconnected to battery or power (pull out power plug),
signals and ground (earth). Remove all signal connections to HM 604 (cable, probe), then start testing.
CH.1
and/or
TIME/DIV.
LEVEL
CH.11
vertical input connector.
CAL.
square-wave signal.
switch and variable control.
control.
CAL.
TIME/DIV.
MAG.
x10
button. Connect both component terminals to
variable control to
HOLD-OFF
button depressed.
control in normal trigger mode.
1O:l
fine control to select point of interest on
SEP.:
CAL.
selection of trigger slope.
TIMEIDIV.
CTand
ground jacks.
CH.II).
switch
Subject to change without notice
Kl
604
Page 23
Test Instructions
General
These Test Instructions are intended as an aid for checking
the most important characteristics of the HM 604 at regular
intervals without the need for expensive test equipment.
Resulting corrections and readjustments inside the instrument, detected by the following tests, are described in the
Service Instructions or on the Adjusting Plan. They should
only be undertaken by qualified personnel.
Astigmatism Check
Check whether the horizontal and vertical sharpness of the
display are equal. This is best seen by displaying a
wave signal with the repetition rate of approximately
1
at normal intensity, then check the sharpness of the vertical
edges. If it is possible to improve this vertical sharpness by
turning the FOCUS control, then an adjustment of the astig-
matism control is necessary. A potentiometer of 50 kQ (see
As with the First Time Operation instructions, care should
be taken that all knobs with
arrows
are
set to their calibrated
positions. None of the pushbuttons should be depressed.
LEVEL knob out in AT position, TRIG. selector switch to
Adjusting Plan) is provided inside the instrument for the cor-
rection of astigmatism (see Service Instructions). A certain
loss of marginal sharpness of the CRT is unavoidable; this is
due to the manufacturing process of the CRT.
AC, DELAY slide switch to OFF. It is recommended to
switch on the instrument for about 30 minutes prior to the
Symmetry and Drift of the Vertical Amplifier
commencement of any check.
Both of these characteristics are substantially determined
Cathode-Ray Tube: Brightness and Focus,
Linearity, Raster Distortions
Normally, the CRT of the HM 604 has very good brightness.
Any reduction of this brightness can only be judged visually.
However, decreased brightness may be the result of
reduced high voltage. This is easily
greatly increased sensitivity of the vertical amplifier. The
control range for maximum and minimum brightness (intensity) must be such that the beam just disappears before
reaching the left hand stop of the
larly when the X-Y button is depressed), while with the control at the right hand stop the focus and the line width are
just acceptable.
recognized
INTENS.
by the
control (particu-
by the input stages of the amplifiers. The checking and
correction of the DC balance for the amplifiers should
be carried out as already described in the Operating
Instructions (page M 7).
The symmetry of Channel I and the vertical final amplifier
can be checked by inverting Channel I, depress INVERT
(CH I pushbutton). The vertical position of the trace should
not change by more than 5mm. However, a change of 1 cm
is just permissible. Larger deviations indicate that changes
have occurred in the amplifier.
A further check of the vertical amplifier symmetry is possible by checking the control range of the
sine-wave signal of 1 O-l 00
input. When the Y-POS. control is then turned fully in both
ness from left to right or shortening of the baseline. (Cause:
incorrect Unblanking Pulse.
timebase
-
or
fly-back must
-
decreasing bright-
directions from stop to stop with a display height of approximately
are visible should be approximately of the same height. Differences of up to 1 cm are permissible (input coupling
should be set to
It should be noted that with wide variations in brightness,
refocusing is always necessary. Moreover, with maximum
brightness, no “pumping” of the display must occur. If
pumping does occur, it is normally due to a fault in the regu-
lation circuitry for the high voltage supply. The presetting
pots for the high voltage circuit, minimum and maximum
intensity, are only accessible inside the instrument (see
Adjusting Plan and Service Instructions).
Checking the drift is relatively simple. Ten minutes after
switching on the instrument,
the horizontal
must not change by more than 5mm during the following
hour. Larger deviations generally result from different
characteristics of the dual
Y amplifier. To some extent, fluctuations in drift are caused
by offset current on the gate. The drift is too high, if the verA certain out-of-focus condition in the edge zone of the
screen must be accepted. It is limited by standards of the
CRT manufacturer. The same is valid for tolerances of the
orthogonality, the
undeflected
spot position, the non-linear-
tical trace position drifts by more than
the appropriate attenuator switch through all 12 steps.
Sometimes such effects occur after long periods of opera-
tion.
ity and the raster distortion in the marginal zone of the
screen in accordance with international standards (see CRT
Calibration of the Vertical Amplifier
data book). These limit values are strictly supervised by
HAMEG.
The selection of a cathode-ray tube without any
tolerances is practically impossible.
Two square-wave voltages of
present at the output sockets of the calibrator
square-
MHz. Focus the horizontal tops of the square-wave signal
Y-POS.
kHz
is applied to the amplifier
8cm,
the upper and lower positions of the trace that
controls. A
AC).
set the baseline exactly on
center
line of the graticule. The beam position
FETs
in both channel inputs to the
g.ZmV,
0.5mm
and 2
on turning
VP,, f
1 %
(CAL.).
are
If a
Subject to change without notice
Tl
604
Page 24
direct connection is made between the 0.2 mV output and
the input of the vertical amplifier, the displayed signal in the
50mV/cm position (variable control to CAL.) should be
#cm
high (DC input coupling). Maximum deviations of
1.2 mm (3%) are permissible. If a x
between the
play height should result. With higher tolerances it should
first be investigated whether the cause lies, within the
amplifier or in the amplitude of the square-wave signal. On
occasions it is possible that the probe is faulty or incorrectly
compensated. If necessary the measuring amplifier can be
calibrated with an accurately known DC voltage (DC input
coupling). The trace position should then vary in accordance
with the deflection coefficient set.
With variable control at the attenuator switch fully counterclockwise, the input sensitivity is decreased at least by the
factor 2.5 in each position. In the
displayed calibrator signal height should vary from 4cm to at
least 1.6cm.
2V-output
socket and Y input, the same dis-
lllprobe
50mVkm
is connected
position, the
attenuators and readjust them as necessary. A compensated
2:1
series attenuator is also necessary, and this
must be matched to the input impedance of the oscilloscope. This attenuator can be made up locally. It is important that this attenuator is shielded. For local manufacture,
the electrical components required are a 1
tor and, in parallel with it, a trimmer 3-l 5 pF in parallel with
approx.
directly to the input connector of the vertical amplifier and
the other side is connected to the generator, if possible via
a low-capacitance coaxial cable. The series attenuator must
be matched to the input impedance of the oscilloscope in
the
coupling; square tops exactly horizontal; no ramp-off is permitted). This is achieved by adjusting the trimmer located in
the 2: 1 attenuator. Theshapeofthesguare-waveshould
then be the same in each input attenuator position.
Operating Modes:
ADD, CHOP.,
2OpF.
5mV/cm
One side of this parallel circuit is connected
position (variable control to CAL., DC input
CH.I/II -TRIG.I/II,
INV.I/II
and X-Y Operation
MS1 +I %
DUAL,
resis-
When pulling the Y-expansion x5 knob (MAG x5), the sensitivity is increased by the factor 5. In the
the displayed signal should change from 1 cm to 5cm by
pulling the MAG x5 knob.
O.PV/cm
position
Transmission Performance of the
Vertical Amplifier
The transient response and the delay distortion correction
can only be checked with the aid of a square-wave
generator with a fast
cable (e.g. HZ34) must be terminated at the vertical input of
the oscilloscope with a resistor equal to the characteristic
impedance of the cable (e.g. with HZ22). Checks should be
made at 1
deflection coefficient should be set at
input coupling (Y variable control in CAL. position). In so
doing, the square pulses must have a flat top without ramp-
off, spikes and glitches; no overshoot is permitted, especially at 1 MHz and a display height of
time, the leading top corner of the pulse must not be
rounded. In general, no great changes occur after the instru-
ment has left the factory, and it is left to the operator’s discretion whether this test is undertaken or not.
Of course, the quality of the transmission performance is
not only dependent on the vertical amplifier. The input
attenuators, located in the front of the amplifier, are fre-
quency-compensated in each position. Even small
capacitive changes can reduce the transmission perfor-
mance. Faults of this kind are as a rule most easily detected
with a square-wave signal with a low repetition rate (e.g.
1
kHz).
available, it is advisable to check at regular intervals the
deflection coefficients on all positions of the input
OOHz,
If a suitable generator with max. output of
1
risetime
kHz,
10
(max.
kHz,
100
5~).
The signal coaxial
kHz
and 1 MHz, the
5mV/cm
&km.
At the same
with DC
4OV,,
is
On depressing the DUAL pushbutton, two traces must
appear immediately. On actuation of the Y-POS. controls,
the trace positions should have no effect on each other.
Nevertheless, this cannot be entirely avoided, even in fully
serviceable instruments. When one trace is shifted vertically across the entire screen, the position of the other trace
must not vary by more than 0.5 mm.
A criterion in chopped operation is trace widening and
shadowing around and within the two traces in the upper or
lower region of the screen. Set
cm,
depress the DUAL and CHOP. pushbutton, set input
coupling of both channels to GD and advance the INTENS.
control fully clockwise. Adjust FOCUS for a sharp display.
With the Y-POS. controls shift one of the traces to a +2 cm,
the other to a -2cm vertical position from the horizontal
center
line of the graticule. Do not try to
chop frequency (500kHz)! Then alternately release and
depress the CHOP. pushbutton. Check for negligible trace
widening and periodic shadowing in the chopped mode.
It is important to note that in the
depressed) or the -I+11 difference mode INVERT (CH
button depressed in addition) the vertical position of the
trace can be adjusted by using both the Channel I and Chan-
nel II Y-POS. controls. If a trace is not visible in either these
modes, the overscanning
the trace.
In X-Y Operation (X-Y pushbutton depressed), the sensitivity in both deflection directions will be the same. When the
signal from the built-in square-wave generator is applied to
the input of Channel II, then, as with Channel I in the vertical
direction, there must be a horizontal deflection of
when the deflection coefficient is set to
TIME/DIV.
I+11
LEDs
will indicate the position of
switch to 1
synchronize
add mode (only ADD
50mV/cm
ps/
the
#cm
position
I)
T2 604
Subject to change without notice
Page 25
(variable control set to its CAL. position, X MAG. x10
released). The check of the mono channel display with the
CH
l/II
button is unnecessary; it is contained indirectly in the
tests above stated.
Triggering Checks
The internal trigger threshold is important as it determines
the display height from which a signal will be stably displayed. It should be approx. 5mm for the HM 604. An
increased trigger sensitivity creates the risk of response to
the noise level in the trigger circuit, especially when the
sensitivity of the vertical input is increased by pulling the
MAG. x5 knob. This can produce double-triggering with
two out-of-phase traces. Alteration of the trigger threshold
is only possible internally. Checks can be made with any
sine-wave voltage between 50 Hz and 1 MHz. The LEVEL
knob should be in AT position. Following this it should be
ascertained whether the same trigger sensitivity is also present with Normal Triggering (LEVEL knob not in AT posi-
tion).
On depressing the SLOPE
sweep changes from the positive-going to the
going edge of the trigger signal.
Internally the HM 604 should trigger perfectly with
dal signals up to IOOMHz at a display height of approx.
5mm , when the HF trigger coupling mode is set.
For external triggering (EXT. button depressed), the EXT.
TRIG. input connector requires a signal voltage of at least
50mV,,,
Checking of W triggering is possible with a video signal of
switchable polarity. A check of both polarities in V and H
mode should be made.
The display should not shift horizontally during a change of
the trigger coupling from AC to DC with a sine-wave signal
without DC offset. The basic requirement for this is a cor-
rect DC Balance Adjustment on the input of the vertical
amplifier (see Operating Instructions, page M7).
which is in synchronism with the Y input signal.
+/-
button, the start of the
negative-
sinosoi-
Checking of the line/mains frequency triggering (50-60 Hz)
is possible, when the input signal is time-related (multiple or
submultiple) to the power line frequency (TRIG. selector
switch to LINE). In this trigger mode, there is no trigger
threshold. Even very small input signals are triggered stably
(e.g. ripple voltage). For this check, use an input of approx.
1 V. The displayed signal height can then be varied by turn-
ing the respective input attenuator switch and its variable
control.
Timebase
Before checking the
the trace length is
the potentiometer for sweep amplitude (see Adjusting
Plan). This adjustment should be made with the
switch in a mid position (i.e. 5pdcm). Prior to the com-
mencement of any check set the time variable control to
CAL. and the HOLD-OFF time control to min. The X MAG.
x10
button should be released. This condition should be
maintained until the variation ranges of these controls are
checked.
If a precise marker signal is not available for checking the
Timebase
generator may be used. Its frequency tolerance should not
be greater than &l %. The
HM 604 is given as
ter than this. For the simultaneous checking of
linearity and accuracy at least 10 oscillations, i.e. 1 cycle
everycm, should always be displayed. For precise determi-
nation, set the peak of the first marker or cycle peak exactly
behind the first vertical graticule line using the X-POS. control. Deviation tendencies can be noted after some of the
marker or cycle peaks.
The 20 and
be checked very precisely with line frequency (5OHz only).
On the 20ms/cm range a cycle will be displayed every cm,
while on the
time coefficients, then an accurate sine-wave
lOms/cm
IOmskm
timebase
IUcm.
+3%,
ranges of the
it should be ascertained that
If not, it can be corrected with
timebase
but as a rule it is considerably bet-
range it will be every
accuracy of the
TIMEIDIV.
TIMEIDIV.
timebase
switch can
2cm.
In the
dualchannelmode
alternate channel switching and with alternate triggering (ALT. button in the X-Section depressed), two non-fre-
quency related signals (e.g. mains/line frequency signal and
calibrator signal) should reliably be triggered internally
dependent on the positions of the
pushbuttons. In the dual channel mode with chop channel
switching and depressed ALT. button, only triggering from
Channel I should be possible. Periodical signal blanks (due
to the chopper frequency 0.5 MHz) should not be visible.
If both vertical inputs are AC coupled to the same signal and
both traces are brought to coincide exactly on the screen,
when working in the alternate dual channel mode, then
no change in display should be noticeable, when the ALT.
button is depressed or released.
Subject to change without notice
(DUAL button depressed) with
CHI/II-TRIG. l/II
The following table shows which frequencies are required
for the particular ranges.
1
s/cm
0.5 s/cm
0.2 s/cm
0.1 s/cm - 10 Hz
50
ms/cm -
20
ms/cm -
10
ms/cm -
5
ms/cm -
2
ms/cm -
1
ms/cm -
0.5ms/cm -
0.2
ms/cm -
1 Hz
-
2 Hz
-
-
5 Hz
20 Hz
50 Hz
100 Hz
200 Hz
500 Hz
1
2
5
kHz
kHz
kHz
0.1
ms/cm -
@cm - 20
50
20
f&cm @cm -
10
@cm -
5
yslcm
2
1
f&cm -
0.5
@cm -
0.2
f&cm -
0.1
0.05
10
kHz
kHz
50
100
200
-
f&cm - 10MHz
f&cm - 20MHz
500
1 MHz
2MHz
5MHz
kHz
kHz
kHz
kHz
Page 26
The time variable control range can also be checked. The
sweep speed becomes slower by turning this variable control counter-clockwise to its left stop. 2.5 cycles at least
every cm should be displayed (with X MAG.
released; measurement in the 5pdcm range).
When the X MAG.
cycle peak will be displayed every 10 cm +5 % (with variable
control in CAL. position; measurement in the 5pdcm
range). The tolerance is better measureable in the
range (one cycle every 1 cm).
Check the ramp output voltage on rear panel (BNC connector
marked by
ting: I V/cm; Timebase to one step slower sweep speed
than on the HM 604 under test. The latter must have no input
and no trigger voltage (free-running sweep; input coupling
switch to GD). The sawtooth voltage is applied with a BNC
cable without termination from the ramp output connector
to the input of the Test Scope. The Test Scope should show
a positive-going linear sawtooth with an amplitude of approx.
SV,,.
trol min.-max. can be checked. The hold-off time variation
cannot be measured precisely with this method, because the
unblanking pulse of the
the ramp width.
/w)
At the same time the function of the HOLD-OFF con-
x10
button is depressed, a marker or
with a Test Oscilloscope. Test Scope set-
timebase
generator is smaller than
x10
button
50ps/cm
Sweep Delay
When the Sweep DELAY mode lever switch is set to the OFF
mode, the delay should have no effect on the display of the
lkHz
calibration signal. When the Sweep Delay is set in
the SEARCH mode (refer to Sweep Delay Operating Instructions), it is possible to check the delay time by means of a distance measurement of the blanked baseline. For this, the
DELAY VAR. 1O:l control must be set to xl (rotate coun-
terclockwise until a snap noise is audible). When DELAY
mode is selected, the
without any blanking.
Over the full range of adjustment of the DELAY VAR.
10: 1 control, the displayed waveform of the calibration
signal should be shifted without any jitter, jumping or
intermittent blanking.
Control settings: Connect calibrator socket
CH.1 input connector, DC input coupling, deflection coeffi-
cient 50mV/cm, TRIG. selector switch to AC, time coeffi-
cient 1 mdcm, no pushbutton depressed, all controls in cali-
brated position, DELAY mode switch in OFF position, LEVEL
knob in AT position. Now the calibrator signal is displayed
with a signal height of 4cm and approx. 1 cycle per cm. After
switching to SEARCH, the mode indication lamp blinks. Set
the DELAY range switch to 1 ms. Rotate the DELAY VAR.
control until the left half of the display is blanked. The delay
time is now 5ms. After switching to DELAY, the display is
again fully visible. The DELAY mode LED is illuminated continuously. The displayed signal can now be expanded. For
this purpose turn the TIME/DR/. switch clockwise to
cm. The expansion is now x200. With the DELAY VAR. control, the nearest edge of the calibration signal can be brought
in the screen center and checked on the above-mentioned
criteria. With x200 expansion, the display brightness
traGe
reverts to the full
IOcm
display
(0.2V/l kHz)
to
5@
nor-
mally needs increasing (with
However, larger expansions than x200 are quite possible, but
the decrease of brightness and the jitter restricts the evaluation.
When switching to DEL. TRIG. positions, every slope is
accepted for triggering after the delay time has elapsed.
In TVSEP. mode condition, the slope can be chosen
INTENS.
and FOCUS control).
(+/-).
Component Tester
After pressing the
tal straight line has to appear immediately, when the CTsoc-
ket is open. The length of this trace should be approx. 8cm.
With connection of the
in the Y-Section, a vertical straight line with approx. 6cm
height should be displayed. The above stated measurements
have some tolerances. They are dependent among other
things on the mains/line voltage.
COMPONHVT
CTsocket
TESTER button, a horizon-
to one of the ground jacks
Trace Alignment
The CRT has an admissible angular deviation
the X deflection plane Dl-D2 and the horizontal center line of
the internal graticule. This deviation, due to tube producion
tolerances (and only important after changing the CRT), and
also the influence of the earth’s magnetic field, which is
dependent on the instrument’s North orientation, are cor-
rected by means of the TR potentiometer. In general, the
trace rotation range is asymmetric. It should be checked,
whether the baseline can be adjusted somewhat sloping to
both sides round about the horizontal center line of the
graticule. With the HM 604 in its closed case, an angle of rotation
AO.57”
(1 mm difference in elevation per
length) is sufficient for the compensation of the earth’s
magnetic field.
+5”
IOcm
between
graticule
Miscellaneous
Y output
A check of the Y output (rear panel) is possible on the screen
using the dual channel mode by means of the calibrator signal. To this a connection is made from the calibrator socket
(0.2V/l kHz)
BNC cable and a 5OB through-termination - a second connection from the Y output to
tings:
switch to
coupling to GD, TIME/DIV. switch to
triggering (LEVEL knob in AT position), TRIG. selector to AC,
no button depressed. Now the square-wave signal is visible
with 4cm display height. With Y-P0S.I control, the tops of
the square-wave are adjusted to
center line of the graticule. Then the DUAL button has to be
pressed. The appearing second trace (without signal) is
adjusted to -2cm using the Y-POS.II control. Then the
input coupling is set to DC. Now the signal across the Y output appears with the same phase position as the calibrator
signal via Channel I. As well as the DC offset (e.g.
+80mV)
can be measured. In the example, the sensivity of the Y output can be calculated as
to the
CH.1
input connector and - using a BNC-
CH.11
CH.1
attenuator switch to
O.lV/cm, CH.1
the amplitude (e.g. 2cm P
input coupling to DC,
0.2V:4cm
input connector. Set-
50mV/cm, CH.11
attenuator
CH.11
0.5ms/cm,
+2cm
from the horizontal
0.2V,,)
=
SOmVkm.
automatic
+0.8cm &
of the Y output
input
CH.11
.
T4 604
Subject to change without notice
Page 27
tor. Settings: CH.1 attenuator switch to 50mV/cm,
attenuator switch to O.lV/cm, CH.1 input coupling to DC,
CH.11
input coupling to GD,
cm, automatic triggering (LEVEL knob in AT position),
TRIG. selector to AC, no button depressed. Now the
square-wave signal is visible with
TIMEIDIV.
4cm
switch to
display height. With
CH.11
0.5ms/
Y-P0S.I control, the tops of the square-wave are adjusted
to
+2cm
from the horizontal
Y-POS.II
Now the signal across the Y output appears with the same
phase position as the calibrator signal via Channel I. As well
as the DC offset (e.g. +0_8cm P
(e.g. 2cm &
example, the sensivity of
0.2V:4cm
control. Then the
CH.11
input coupling is set to DC.
0.2V,,)
of the Y output can be measured. In the
=
SOmV/cm.
center
+80mV)
the Y output can be calculated as
the amplitude
2 modulation (optional)
Checking the blanking facility on the 2 modulation connec-
tor (rear panel) is possible with a sine- or square-wave
generator. The sine-wave generator requires an output voltage control. The square-wave generator must deliver posi-
tive pulses to ground (chassis). Alternatively a small adjust-
able sine voltage from a (separate) power transformer may
be used. For the latter set the TIME/DIV. range to e.g.
lOms/cm. With the TIMEBASE variable control, the gap in
the baseline can be brought acceptably to a standstill. Line
frequency triggering is better (TRIG. selector switch to
The length ratio from bright to dark lines on the display is
dependent on the sine voltage amplitude. Without a modulating generator, the function of the Z modulation can be
checked coarsely by short-circuiting the 2 connector. Then
the baseline is blanked fully.
N
.
Subject to change without notice
T5 604
Page 28
Service Instructions
General
The following instructions are intended as an aid for the
electronic technician, who is carrying out readjustments on
the HM 604, if the nominal values do not meet the specifications. These instructions primarily refer to those faults,
which were found after using the Test Instructions. However, this work should only be carried out by properly qualified personnel. For any further technical information call or
write to HAMEG. Addresses are provided at the back of the
manual. It is recommended to use only the original packing
material, should the instrument be shipped to HAMEG for
service or repair (see also Warranty, page M2).
Instrument Case Removal
The rear cover can be taken off after unplugging the power
cord’s triple-contact connector and after two cross recessed pan head screws
have been removed. While the instrument case is firmly
held, the entire chassis with its front panel can withdrawn
forward. When the chassis is inserted into the case later on,
it should be noticed that the case has to fit under the flange
of the front panel. The same applies for the rear of the case,
on which the rear cover is put.
Caution
During opening or closing of the case, the instrument
must be disconnected from all power sources for
maintenance work or a change of parts or components.
If a measurement, trouble-shooting, or an adjustment
is unavoidable, this work must be done by a specialist,
who is familiar with the risk involved.
When the instrument is set into operation after the case
has been removed, attention must be paid to the acceleration voltage for the CRT operating voltages for both final amplifier stages
together
CRT socket, on the upper and the lower horizontal
PC&,
neck. High voltages of max. 2000 V are at the
and FOCUS potentiometers. They are highly dangerous
and therefore precautions must be taken. It should be
noted furthermore that shorts occuring on different
points of the CRT high voltage and unblanking circuitry
will definitely damage some semiconductors. For the
same reason it is very risky to connect capacitors to
these points while the instrument is on.
Capacitors in the instrument
when the instrument is disconnected from all voltage
sources. Normally, the capacitors are discharged 6 seconds after switching ofi. However, with a defective
instrument an interruption of the load is possible.
14OV.
and on the lateral
(M4x30mm)
Potentials of these voltages are on the
PCB
with two washers on it
12.5kV -
directly beside the CRT
may
still be charged, even
and to the
INTENS.
Therefore,
nect one by one all terminals of the check strip across
1 kQ to ground (chassis) for a period of 1 second.
Handling of the CRT needs utmost caution. The glass
bulb must not be allowed - under any circumstances
to come into contact with hardened tools, nor should it
undergo local superheating (e.g. by soldering iron) or
local undercooling (e.g. by cryogenic-spray). We recommend the wearing of safety goggles (implosion danger).
atter
switching off, it is recommended to con-
Operating Voltages
Besides the two AC voltages for the CRT heating (6.3V) and
graticule illumination, Component Tester and line triggering
(12V)
there are eight electronically regulated DC operating
voltages generated
+
14OV. - 18OOV,
voltages are fixed voltages, except the +
adjusted. The variation of the fixed voltages greater than
+2
% from the nominal value indicates a fault. This voltage
is measured on the checkpoint strip (located on XY Board)
with reference to ground. Measurements of the high voltage may only be accomplished by the use of a sufficient
highly resistive voltmeter
lutely sure that the electric strength of the voltmeter is suf-
ficiently high.
and + 10.4
(+12V, +5V, -5V, -12V. +68V,
kV).
These different operating
12V,
which can be
(>lOMQ).
You must make abso-
Minimum Brightness
The variable resistor VR601, located on the Z-PCB, is used
for this adjustment procedure. It may only be touched by a
properly insulating screwdriver (Caution! High voltage!).
Correct adjustment is achieved, when the trace can be
blanked while X-Y pushbutton is depressed and, in addi-
tion, when the requirement described in the Test Instruc-
tions are met.
-
Astigmatism control
The ratio of vertical and horizontal sharpness can be
adjusted by the variable resistor VR603, located on the
PCB (see Adjusting Plan). As a precaution however, the
voltage for the vertical deflecting plates (approx.
should firstly be checked, because this voltage will affect
the astigmastism correction. Use the 1 MHz square-wave
signal from the built-in calibrator via 10: 1 probe for this cor-
rection. Firstly adjust the sharpness of the horizontal
square-wave lines using the FOCUS control. The sharpness
of the verticallines should then be corrected with the aid of
the Astigm. pot VR626. Repeat the correction several
times in this sequence. Adjustment has been correctly car-
ried out when, on using FOCUS control only, the sharpness
is not improved in
eitherdirection.
-
Z-
+41.5V)
Subject to change without notice
Sl
604
Page 29
Trouble-Shooting the Instrument
Replacement of Components and Parts
For this job, at least an isolating variable mains/line transformer (protection class II), a signal generator, an adequate
precise multimeter, and, if possible, an oscilloscope are
needed. This last item is required for complex faults, which
can be traced by the display of signal or ripple voltages. As
noted before, the regulated high voltage and the supply
voltage for the final stages are highly dangerous. Therefore
it is recommended to use totally insulated extendedprobe tips, when trouble-shooting the instrument. Accidental contact with dangerous voltage potentials is then
unlikely. Of course, these instructions cannot thoroughly
cover all kinds of faults. Some common-sense will certainly
be required, when a complex fault has to be investigated.
If trouble is suspected, visually inspect the instrument
thoroughly after removal of the case. Look for loose or badly
contacted or
ing). Check to see that all circuit board connections are making good contact and are not shorting to an adjacent circuit,
Especially inspect the connections between the
the power transformer, to front chassis parts, to CRT socket, to trace rotation coil (inside of CRT’s shielding), to the 3
BNC connectors at the rear chassis, and to the control
potentiometers and switches on top of and beneath both
main-PCBs. Furthermore, the soldering connections of the
transistors and Fixed Three-Terminal Regulators resp. on
the rear chassis. This visual inspection can lead to success
much more quickly than a systematic fault location using
measuring instruments. Prior to any extensive trouble-
shooting, also check the external power source.
If the instrument fails completely, the first and most important step - after checking the mains/line voltage and
power fuse - will be to measure the deflecting plate volt-
ages of the CRT. In almost any case, the faulty section can
be located. The sections represent:
1.
Vertical deflection.
2. Horizontal Deflection.
3.
CRTcircuit.
4. Power supply.
While the measurement takes place, the position controls
of both deflection devices must be in mid-position. When
the deflection devices are operating properly, the separate
voltages of each plate pair are almost equal then (Y
and X
If the separate voltages of a plate pair are very different, the
associated circuit must be faulty.
An absent trace in spite of correct plate voltages means a
fault in the CRT circuit. Missing deflection plate voltages is
probably caused by a defect in the power supply.
=67V).
discolored
components (caused by overheat-
PCBs,
to
=42V
For the replacement of parts and components use only
parts of the same or equivalent type. Resistors without
specific data in the diagrams have a power dissipation of
0.25 Watt and a tolerance of 2 %. Resistors in the high voltage circuit must have sufficient electric strength.
Capacitors without a voltage value must be rated for an
operating voltage of 63V. The capacitance tolerance should
not exceed 20%. Many semiconductors are selected,
especially the gate-diodes 1 N4154, and all amplifier transistors, which are contained in push-pull circuits (including the
FETs).
If a selected semiconductor is defective, all gate-
diodes or both push-pull transistors of a stage should be
replaced by selected components, because otherwise
there are possibly deviations of the specified data or functions. The HAMEG Service Department can give you advice
for troubleshooting and replaceable parts. Replacement
parts can be ordered by letter or telephone from the nearest
HAMEG Service Office. Please supply the following infor-
mation: Instrument type and serial number, description of
the part (including function and location in the instrument),
quantity desired.
Replacement of the Power Transformer
Should it be necessary to replace the mains/line transformer, the correct terminal sequence (color identification)
for primary and secondary windings must be followed (see
diagram “Power Supply” and the figure below). In addition,
the relevant Safety Regulations must be observed. Here,
we refer only to those requirements relative to the parts
conductively connected to the supply mains:
-
The construction of the instrument shall be such as to
prevent any short-circuiting or bridging of the insulation,
clearances or creepage distances between those parts
connected to the supply mains and any accessible conductive parts due to accidental loosening or freeing of
the wiring, screws, etc.
-
The rigidity of the mains wiring connections, which may
be subject to mechanical stresses, shall not be dependent upon the soldering alone. To meet this requirement,
the bare ends of the wires must be pushed through the
holes in the respective soldering tab, bent over with a
pair of pliers, and subsequently fixed by soldering.
-
The minimum cross section of the protective earth con-
nection between the instrument’s power inlet and the
connecting soldering tab on the rear chassis must be
0.81
mm2
in North America and 0.75mm2 in Western
Europe. The connecting soldering tab on the rear chassis
has to be secured mechanically against loosening (e.g.
with lock washer).
S2 604
Subject to change without notice
Page 30
After replacing the power transformer, all remaining bits of
Adjustments
wire, solder and other foreign matter must be removed
from the
PCBs,
the vicinity of the power transformer and
from within the insulating connecting box by shaking,
brushing and blowing. Finally, the top plate of the insulating
connecting box has to be replaced. Before connecting the
instrument to the power supply, replace the possibly defective fuse, press the POWER button and make sure that
there is an adequate insulation state between chassis
safety earth conductor) on the one hand, and the live/line
pin as well as the neutral pin, on the other. Only after proper
insulation has been established may further function tests
with open chassis follow, but with appropriate precaution-
ary measures.
Connection of the safety
-
earth pin via long sold.
tab to rear chassis
Appliance Inlet
As advised in the Operating, Test and Service Instructions,
small corrections and adjustments are easily carried out
with the aid of the Circuit Diagrams and Adjusting P/an.
However, a complete recalibration of the scope should not
be attempted by an inexperienced operator, but only someone with sufficient expertise. Several precision measuring
instruments with cables and adapters are required, and only
(=
then should the pots and trimmers be readjusted, provided
that the result of each adjustment can be exactly deter-
mined. Thus for each operating mode and switch position,
a signal with the appropriate sine or square waveform, frequency, amplitude,
risetime
and duty cycle is required.
Safety Class
Plug with earthing contact
I
II
Rear View of Power Switch and Appliance
Inlet with Voltage Selector and Fuse
1
2+
2*
Pushbutton
power switch
4At24A
-_
I
250v
bk
_..
tl’
gr
(see diagram “POWER SUPPLY”)
J
Subject to change without notlce
Page 31
Page 32
Page 33
Page 34
Page 35
Page 36
Page 37
Page 38
Page 39
Page 40
Page 41
Page 42
Page 43
Page 44
Page 45
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.