HAMEG HM507 User Manual

Oscilloscope
HM507
Manual
English
Contents
General information regarding the CE marking ............ 4
Symbols ......................................................................... 6
Use of tilt handle ........................................................... 6
Safety ............................................................................. 6
Intended purpose and operating conditions ................ 6
Warranty ......................................................................... 7
Maintenance .................................................................. 7
Protective Switch-Off .................................................... 7
Power supply ................................................................. 7
Type of signal voltage ...................................................... 8
Amplitude Measurements ............................................ 8
Total value of input voltage ........................................... 9
Time Measurements ..................................................... 9
Connection of Test Signal ........................................... 10
Controls and readout ..................................................... 11
Menu ................................................................................ 29
First Time Operation ...................................................... 30
Trace Rotation TR ........................................................ 30
Probe compensation and use ..................................... 30
Adjustment at 1kHz ..................................................... 30
Adjustment at 1MHz ................................................... 31
Operating modes of the vertical
amplifiers in Yt mode .................................................. 31
X-Y Operation ............................................................... 32
Phase comparison with Lissajous figures .................. 32
Phase difference measurement
in DUAL mode (Yt) ....................................................... 32
Phase difference measurement in DUAL mode........ 32
Measurement of an amplitude modulation ............... 33
Triggering and time base ............................................... 33
Automatic Peak (value) -Triggering ............................. 33
Normal Triggering ....................................................... 34
(Slope) .................................................................... 34
Trigger coupling ........................................................... 34
Triggering of video signals .......................................... 34
Line/Mains triggering (~) ............................................. 35
Alternate triggering ..................................................... 35
External triggering ....................................................... 35
Trigger indicator “TR” ................................................. 36
HOLD OFF-time adjustment ....................................... 36
Delay / After Delay Triggering ..................................... 36
Auto Set .......................................................................... 38
3
Oscilloscope
HM507
Component Tester (analog mode) ................................ 38
General ........................................................................ 38
Using the Component Tester ..................................... 39
Test Procedure ............................................................ 39
Test Pattern Displays .................................................. 39
Testing Resistors ........................................................ 39
Testing Capacitors and Inductors ............................... 39
Testing Semiconductors ............................................. 39
Testing Diodes ............................................................ 39
Testing Transistors ...................................................... 39
In-Circuit Tests ............................................................. 40
Storage Mode ................................................................. 40
Signal capture modes .................................................. 40
Raltime sampling ......................................................... 40
Radom sampling .......................................................... 41
Raltime sampling ......................................................... 41
Signal display and recording modes ........................... 41
Vertical resolution ........................................................ 41
Horizontal resolution ................................................... 41
Alias signal display ...................................................... 42
Operating modes of the vertical amplifiers ................ 42
Adjustments .................................................................... 42
RS232 Interface - Remote Control ................................ 42
Safety ........................................................................... 42
Operation ..................................................................... 42
RS-232 Cable ............................................................... 42
RS-232 protocol ........................................................... 43
Baud-Rate Setting ........................................................ 43
Data Communication .................................................. 43
Mean Value Display ........................................................ 38
2
Front Panel HM507 ......................................................... 43
Subject to change without notice
KONFORMITÄTSERKLÄRUNG DECLARATION OF CONFORMITY DECLARATION DE CONFORMITE
Herstellers HAMEG Imstruments GmbH Manufacturer Industriestraße 6 Fabricant D-63533 Mainhausen
Die HAMEG Instruments GmbH bescheinigt die Konformität für das Produkt The HAMEG Instruments GmbH herewith declares conformity of the product HAMEG Instruments GmbH déclare la conformite du produit
Bezeichnung / Product name / Designation:
Oszilloskop/Oscilloscope/Oscilloscope
Typ / Type / Type: HM507
mit / with / avec: -
Optionen / Options / Options: -
mit den folgenden Bestimmungen / with applicable regulations / avec les directives suivantes
EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE
Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE
General information regarding the CE marking
Angewendete harmonisierte Normen / Harmonized standards applied / Normes harmonisées utilisées
Sicherheit / Safety / Sécurité
EN 61010-1: 1993 / IEC (CEI) 1010-1: 1990 A 1: 1992 / VDE 0411: 1994 EN 61010-1/A2: 1995 / IEC 1010-1/A2: 1995 / VDE 0411 Teil 1/A1: 1996-05 Überspannungskategorie / Overvoltage category / Catégorie de surtension: II Verschmutzungsgrad / Degree of pollution / Degré de pollution: 2
Elektromagnetische Verträglichkeit / Electromagnetic compatibility / Compatibilité électromagnétique
EN 61326-1/A1 Störaussendung / Radiation / Emission: Tabelle / table / tableau 4; Klasse / Class / Classe B. Störfestigkeit / Immunity / Imunitee: Tabelle / table / tableau A1.
EN 61000-3-2/A14 Oberschwingungsströme / Harmonic current emissions / Émissions de courant harmoni­que: Klasse / Class / Classe D.
EN 61000-3-3 Spannungsschwankungen u. Flicker / Voltage fl uctuations and fl icker / Fluctuations de tension et du fl icker.
Datum /Date /Date Unterschrift / Signature /Signatur
15.01.2001
E. Baumgartner Technical Manager /Directeur Technique
General information regarding the CE marking
HAMEG instruments fulfi ll the regulations of the EMC directive. The conformity test made by HAMEG is based on the actual generic- and product standards. In cases where different limit values are applicable, HAMEG applies the severer standard. For emission the limits for residential, commercial and light industry are applied. Regarding the immunity (susceptibility) the limits for industrial environment have been used. The measuring- and data lines of the instrument have much infl uence on emmission and immunity and therefore on meeting the acceptance limits. For different applications the lines and/or cables used may be different. For measurement operation the following hints and conditions regarding emission and im­munity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and external devices, (computer, printer etc.) suffi ciently screened cables must be used. Without a special instruction in the manual for a reduced cable length, the maximum cable length of a dataline must be less than 3 meters and not be used outside buildings. If an interface has several connectors only one connector must have a connection to a cable. Basically interconnections must have a double screening. For IEEE­bus purposes the double screened cables HZ72S and HZ72L from HAMEG are suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and instrument should be as short as possible. Without instruction in the manual for a shorter length, signal lines must be less than 3 meters and not be used outside buildings. Signal lines must screened (coaxial cable - RG58/U). A proper ground connection is required. In combination with signal generators double screened cables (RG223/U, RG214/U) must be used.
3. Infl uence on measuring instruments
Under the presence of strong high frequency electric or magnetic fi elds, even with careful setup of the measuring equip­ment an infl uence of such signals is unavoidable. This will not cause damage or put the instrument out of operation. Small deviations of the measuring value (reading) exceeding the instruments specifi cations may result from such conditions in individual cases.
4. RF immunity of oscilloscopes
4.1 Electromagnetic RF fi eld The infl uence of electric and magnetic RF fi elds may become visible (e.g. RF superimposed), if the fi eld intensity is high. In most cases the coupling into the oscilloscope takes place via the device under test, mains/line supply, test leads, control cables and/or radiation. The device under test as well as the oscilloscope may be effected by such fi elds. Although the in­terior of the oscilloscope is screened by the cabinet, direct radiation can occur via the CRT gap. As the bandwidth of each amplifi er stage is higher than the total –3dB bandwidth of the oscilloscope, the infl uence RF fi elds of even higher frequen­cies may be noticeable.
cable HZ72 from HAMEG is suitable.
4.2 Electrical fast transients / electrostatic discharge Electrical fast transient signals (burst) may be coupled into the oscilloscope directly via the mains/line supply, or indirectly via test leads and/or control cables. Due to the high trigger and input sensitivity of the oscilloscopes, such normally high signals may effect the trigger unit and /or may become visible on the CRT, which is unavoidable. These effects can also be caused by direct or indirect electrostatic discharge.
HAMEG Instruments GmbH
Subject to change without notice
3
Subject to change without notice
4
HM507
100MSa /s Real Time Sampling, 2GSa /s Random Sampling
2kPts Memory per Channel
2 Channels
Deflection coefficients 1mV/div.…20V/div.,
Time Base 20ns/div.…100 s /div.
8-Bit Low Noise Flash A/D Converters
Programmable Mathematical Signal Processing
Acquisition modes: Single, Refresh, Envelope, Average, Roll
RS-232 interface for control and signal data transfer,
incl. Windows
®
software
optional: Multifunction Interface HO79-6
See HM504-2 for analog mode
5 0 M H z C o m b i S c ope
®
H M 5 0 7
HM507
Signal processing with userdefined formulas
Cursor measurement
Automatic measurements
50MHz CombiScope®HM507
All data valid at 23 °C after 30 minute warm-up
Vertical Deflection
Operating Modes: Channel 1 or 2 only
Channels 1 and 2 (alternate or chopped) Sum or Difference of CH 1 and CH 2
Invert: CH 2 XY Mode: via CH 1 (X) and CH 2 (Y) Bandwidth: 2 x 0…50 MHz (-3dB) Rise Time: ‹ 7 ns Deflection Coefficients: 1-2-5 Sequence
1…2 mV/div.: ± 5 % (0…10 MHz (-3dB)) 5 mV/div.…20 V/div.: ± 3 % (0…50 MHz (-3dB)) Variable (uncalibrated): ›2.5: 1to › 50 V/div.
Input Impedance: 1 MΩ II 15 pF Coupling: DC, AC, GND (ground) Max. Input Voltage: 400 V (DC + peak AC)
Triggering
Automatic (Peak to Peak): 20 Hz…100 MHz (5mm) Normal with Level Control: 0…100 MHz (5 mm) Slope: Rising or falling Sources: Channel 1 or 2, CH 1/CH 2 alternate (8mm)
Line and External
Coupling: AC (10 Hz…100 MHz), DC (0…100MHz),
HF (50 kHz…100MHz), LF (0…1.5 kHz)
Trigger Indicator: with LED Triggering after Delay: with Level Control and Slope selection External Trigger Signal: 0.3 V
pp
(0…50 MHz)
Active TV sync. separator: Field and Line, +/-
Horizontal Deflection (analog and digital)
Anal
og
Time Base: 50 ns/div.…0.5 s/div. (1-2-5 Sequence)
Accuracy: ± 3 % Variable (uncalibrated): › 2.5 :1 to › 1.25 s/div.
X-Magnification x 10: up to 10 ns/div. (± 5 %)
Accuracy: ± 5 %
Delay (selectable): 200
ns…140 ms (variable)
Hold-Off Time: variable to approx. 10 : 1 XY Mode Bandwidth X amplifier: 0…3MHz (-3 dB) XY Phase shift ‹ 3°: ‹ 120 kHz Digit
al
Time Base: 100 ns/div.…100 s/div. (1-2-5 Sequence)
Accuracy: ± 2 %
X-Magnification x 10: up to 20 ns/div.
Accuracy: ± 2 % XY Mode Bandwidth X Amplifier : 0…50 MHz (-3 dB) XY Phase shift ‹ 3°: ‹ 10 MHz
Digital Storage
Operating Modes: Refresh, Roll, Single, XY, Envelope,
Average, Random Sampling
Interpolation: Linear Dot Join Function Sampling Rate (Real Time): max 100MSa/s, 8 bit Flash A/D Converter Sampling Rate (Random): 2 GSa/s relative Post/Pre-Trigger: -10.…+10 div. (continuous) Display Refresh Rate: max. 180/s Bandwidth: 2 x 0…50 MHz (-3dB) Signal Memory: 3x 2 k x 8 bit Reference Signal Memory: 3x 2 k x 8 bit Mathematical Signal Memory:3 x 2 k x 8 bit Resolution (dots/div.) Yt Mode: X: 200/div., Y: 25/ div. Resolution (dots/div.) XY Mode: X: 25/div., Y: 25/div.
Operation/Readout /Control
Manual: via controls Autoset: automatic signal related parameter settings Save and Recall: 9 user defined parameter settings Readout: display of menu, parameters, cursors
and results
Auto Measurements:
Analog mode: Frequency, Period, V
DC
, Vpp, Vp+, Vp-,
also in digital mode: V
rms
, V
average
Cursor Measurements:
Analog mode: ΔV, Δt, 1/Δt (f), tr, V to GND, ratio X and Y also in digital mode: Pulse count, Vt related to Trigger Point,
Peak to Peak, Peak+, Peak-
Frequency counter: 4 digit (0.01 % ±1 digit) 0.5 Hz…100MHz Interface (standard fitting): RS-232 (Control, Signal Data) Interface Option: HO79-6 (IEEE-488, RS-232, Centronics)
Component Tester
Test Voltage: approx. 7V
rms
(open circuit)
Test Current: max. 7 mA
rms
(short-circuit)
Test Frequency: approx. 50Hz Test Connection: 2 banana jacks 4 mm Ø
One test circuit lead is grounded via protective earth (PE)
Miscellaneous
CRT: D14-363GY, 8 x 10 div. with internal graticule Acceleration Voltage: approx. 2 kV Trace Rotation: adjustable on front panel Z-Input (Intens. modulation, analog): max. + 5V (TTL) Calibrator Signal (Square Wave): 0.2 V, 1 Hz…1MHz (tr ‹ 4 ns), DC Power Supply (Mains): 105…253V, 50/60 Hz ±10%, CAT II Power Consumption: approx. 42 Watt at 230 V/50 Hz Safety class: Safety class I (EN61010-1) Operating temperature: +5…+40°C Storage temperature: -20…+70°C Rel. humidity: 5…80% (non condensing) Dimensions (W x H x D): 285 x 125 x 380 mm Weight: approx. 6.0kg
Accessories supplied: Line Cord, Operators Manual and Software for Windows on CD-ROM, 2 Probes 1:1 / 10:1 (HZ154), RS-232 Interface
Optional accessories:
HO79-6 Multifunction Interface HZ14 Interface cable (serial) 1:1 HZ20 Adapter, BNC to 4mm banana HZ33 Test cable 50Ω, BNC/BNC, 0,5m HZ34 Test cable 50Ω, BNC/BNC, 1m HZ43 19''-Rackmount Kit 3RU HZ51 Probe 10:1 (150MHz) HZ52 Probe 10:1 RF (250MHz) HZ53 Probe 100:1 (100MHz) HZ56-2 AC/DC Current probe HZ70 Opto Interface (with optical fiber cable) HZ72 GPIB-Cable 2m HZ100 Differential probe 20:1 / 200:1 HZ109 Differential probe 1:1 / 10:1 HZ115 Differential probe 100:1 / 1000:1 HZ200 Probe 10:1 with auto attenuation ID (250MHz) HZ350 Probe 10:1 with automatically identification (350MHz) HZ355 Slimline probe 10:1 with automatically identification (500MHz) HZO20 High voltage probe 1000:1 (400MHz,1000Vrms) HZO30 Active probe 1GHz (0,9pF, 1MΩ, including many accessories) HZO50 AC/DC Current probe 20A, DC...100kHz HZO51 AC/DC Current probe 1000A, DC...20kHz
®
· DQS-certified in accordance with DIN EN ISO 9001:2000, Reg.-No.: DE-071040 QM
w w w . h a m e g . c o m
50MHz CombiScope®HM507
All data valid at 23 °C after 30 minute warm-up
Vertical Deflection
Channels 1 and 2 (alternate or chopped) Sum or Difference of CH 1 and CH 2
1…2 mV/div.: ± 5 % (0…10 MHz (-3dB)) 5 mV/div.…20 V/div.: ± 3 % (0…50 MHz (-3dB)) Variable (uncalibrated): 2.5: 1to 50 V/div.
Triggering
Line and External
HF (50 kHz…100MHz), LF (0…1.5 kHz)
pp
(0…50 MHz)
Horizontal Deflection (analog and digital)
og
Accuracy: ± 3 % Variable (uncalibrated): 2.5 :1 to 1.25 s/div.
Accuracy: ± 5 %
ns…140 ms (variable)
al
Accuracy: ± 2 %
Accuracy: ± 2 %
Digital Storage
Average, Random Sampling
Operation/Readout /Control
Manual: via controls Autoset: automatic signal related parameter settings Save and Recall: 9 user defined parameter settings Readout: display of menu, parameters, cursors
and results
Auto Measurements:
Analog mode: Frequency, Period, V
DC
, Vpp, Vp+, Vp-,
also in digital mode: V
rms
, V
average
Cursor Measurements:
Analog mode: ΔV, Δt, 1/Δt (f), tr, V to GND, ratio X and Y also in digital mode: Pulse count, Vt related to Trigger Point,
Peak to Peak, Peak+, Peak-
Frequency counter: 4 digit (0.01 % ±1 digit) 0.5 Hz…100MHz Interface (standard fitting): RS-232 (Control, Signal Data) Interface Option: HO79-6 (IEEE-488, RS-232, Centronics)
Component Tester
Test Voltage: approx. 7V
rms
(open circuit)
Test Current: max. 7 mA
rms
(short-circuit)
Test Frequency: approx. 50Hz Test Connection: 2 banana jacks 4 mm Ø
One test circuit lead is grounded via protective earth (PE)
Miscellaneous
CRT: D14-363GY, 8 x 10 div. with internal graticule Acceleration Voltage: approx. 2 kV Trace Rotation: adjustable on front panel Z-Input (Intens. modulation, analog): max. + 5V (TTL) Calibrator Signal (Square Wave): 0.2 V, 1 Hz…1MHz (tr 4 ns), DC Power Supply (Mains): 105…253V, 50/60 Hz ±10%, CAT II Power Consumption: approx. 42 Watt at 230 V/50 Hz Safety class: Safety class I (EN61010-1) Operating temperature: +5…+40°C Storage temperature: -20…+70°C Rel. humidity: 5…80% (non condensing) Dimensions (W x H x D): 285 x 125 x 380 mm Weight: approx. 6.0kg
Accessories supplied: Line Cord, Operators Manual and Software for Windows on CD-ROM, 2 Probes 1:1 / 10:1 (HZ154), RS-232 Interface
Optional accessories:
HO79-6 Multifunction Interface HZ14 Interface cable (serial) 1:1 HZ20 Adapter, BNC to 4mm banana HZ33 Test cable 50Ω, BNC/BNC, 0,5m HZ34 Test cable 50Ω, BNC/BNC, 1m HZ43 19''-Rackmount Kit 3RU HZ51 Probe 10:1 (150MHz) HZ52 Probe 10:1 RF (250MHz) HZ53 Probe 100:1 (100MHz) HZ56-2 AC/DC Current probe HZ70 Opto Interface (with optical fiber cable) HZ72 GPIB-Cable 2m HZ100 Differential probe 20:1 / 200:1 HZ109 Differential probe 1:1 / 10:1 HZ115 Differential probe 100:1 / 1000:1 HZ200 Probe 10:1 with auto attenuation ID (250MHz) HZ350 Probe 10:1 with automatically identification (350MHz) HZ355 Slimline probe 10:1 with automatically identification (500MHz) HZO20 High voltage probe 1000:1 (400MHz,1000Vrms) HZO30 Active probe 1GHz (0,9pF, 1MΩ, including many accessories) HZO50 AC/DC Current probe 20A, DC...100kHz HZO51 AC/DC Current probe 1000A, DC...20kHz
Subject to change without notice
Specifications
5
General information
General information
Please check the instrument for mechanical damage or loose parts immediately after unpacking. In case of damage we advise to contact the sender. Do not operate.
B
C
B
T
A
List of symbols used
Consult the manual High voltage
Important note Ground
Positioning the instrument
As can be seen from the fi gures, the handle can be set into diffe­rent positions: A and B = carrying C = horizontal operating D and E = operating at different angles F = handle removal T = shipping (handle unlocked)
Attention!
When changing the handle position, the instru-
ment must be placed so that it can not fall (e.g. placed on a table). Then the handle locking knobs must be simultaneously pulled outwards and rotated to the required position. Without pulling the locking knobs they will latch in into the next locking position.
C
D
F
E
D
E
A
PUOPFGkT
PUOPFGkT PUOPFGkT
PUOPFGkT
PUOGkT
PUOPFGkT
PUOPFGkT
HM507
PUOPFGkT
PUOPFGkT
PUOPFGkT PUOPFGkT PUOPFGkT PUOPFGkT
PUOPFGkT
PUOPFGkT PUOPFGkT
PUk PUk PUk PUk PUk PUk
PUkT
HGOPFFD
B
PUOPFGkT
PUOPFGkT
PUkT
PUkT
PUkT
INPUT CHI OPK HJ
PUkT
VBN
PUOPFGkT
HJKL
PUOPFGkT
PUkT
PUOPFGkT
HGOFFD
PUkT
PUkT
PUkT
INPUT CHI
INPUT CHI
HAMEG
OPK
OPK
HJ
HJ
VBN
VBN
PUOPFGkT
HJKL
HJKL
T
Handle mounting/dismounting
The handle can be removed by pulling it out further, depending on the instrument model in position B or F.
Safety
The instrument fulfi ls the VDE 0411 part 1 regulations for electrical measuring, control and laboratory instruments and was manufactured and tested accordingly. It left the factory in perfect safe condition. Hence it also corresponds to European Standard EN 61010-1 resp. International Standard IEC 1010-1. In order to maintain this condition and to ensure safe operation the user is required to observe the warnings and other directions for use in this manual. Housing, chassis as well as all measu­ring terminals are connected to safety ground of the mains. All accessible metal parts were tested against the mains with 200 V
The oscilloscope may only be operated from mains outlets with a safety ground connector. The plug has to be installed prior to con­necting any signals. It is prohibited to separate the safety ground connection.
Most electron tubes generate X-rays; the ion dose rate of this instrument remains well below the 36 pA/kg permitted by law.
In case safe operation may not be guaranteed do not use the in­strument any more and lock it away in a secure place.
. The instrument conforms to safety class I.
DC
T
Safe operation may be endangered if any of the following was noticed:
– in case of visible damage. – in case loose parts were noticed – if it does not function any more. – after prolonged storage under unfavourable conditions (e.g.
like in the open or in moist atmosphere).
– after any improper transport (e.g. insuffi cient packing not
conforming to the minimum standards of post, rail or transport company)
Proper operation
Please note: This instrument is only destined for use by personnel well instructed and familiar with the dangers of electrical mea­surements.
For safety reasons the oscilloscope may only be operated from mains outlets with safety ground connector. It is prohibited to separate the safety ground connection. The plug must be inserted prior to connecting any signals.
6
Subject to change without notice
General information
CAT I
This oscilloscope is destined for measurements in circuits not connected to the mains or only indirectly. Direct measurements, i.e. with a galvanic connection to circuits corresponding to the categories II, III, or IV are prohibited!
The measuring circuits are considered not connected to the mains if a suitable isolation transformer fulfi lling safety class II is used. Measurements on the mains are also possible if suitable probes like current probes are used which fulfi l the safety class II. The measurement category of such probes must be checked and observed.
Measurement categories
The measurement categories were derived corresponding to the distance from the power station and the transients to be expected hence. Transients are short, very fast voltage or current excursions which may be periodic or not.
Measurement CAT IV: Measurements close to the power station, e.g. on electricity meters
Measurement CAT III: M e a su r e me n t s in t h e in t e r io r o f b ui l d i ng s ( p o w er d i s tr i b u tio n i n s t al ­lations, mains outlets, motors which are permanently installed).
Measurement CAT II: Measurements in circuits directly connected to the mains (house­hold appliances, power tools etc).
Measurement CAT I: Electronic instruments and circuits which contain circuit breakers resp. fuses.
Warranty and repair
HAMEG instruments are subjected to a strict quality control. Prior to leaving the factory, each instrument is burnt-in for 10 hours. By intermittent operation during this period almost all defects are detected. Following the burn-in, each instrument is tested for function and quality, the specifi cations are checked in all operating modes; the test gear is calibrated to national standards.
The warranty standards applicable are those of the country in which the instrument was sold. Reclamations should be directed to the dealer.
Only valid in EU countries
In order to speed reclamations customers in EU countries may also contact HAMEG directly. Also, after the warranty expired, the HAMEG service will be at your disposal for any repairs.
Return material authorization (RMA):
Prior to returning an instrument to HAMEG ask for a RMA number either by internet (http://www.hameg.com) or fax. If you do not have an original shipping carton, you may obtain one by calling the HAMEG service dept (++49 (0) 6182 800 500 ) or by sending an email to service@hameg.com.
Maintenance
Clean the outer shell using a dust brush in regular intervals. Dirt can be removed from housing, handle, all metal and plastic parts using a cloth moistened with water and 1 % detergent. Greasy dirt may be removed with benzene (petroleum ether) or alcohol, there after wipe the surfaces with a dry cloth. Plastic parts should be treated with an antistatic solution destined for such parts. No fl uid may enter the instrument. Do not use other cleansing agents as they may adversely affect the plastic or lacquered surfaces.
Environment of use.
The oscilloscope is destined for operation in industrial, business, manufacturing, and living sites.
Environmental conditions
Operating ambient temperature: +5 °C to +40 °C. During transport or storage the temperature may be –20 °C to +70°C.
Please note that after exposure to such temperature s or in case of condensation proper time must be allowed until the instrument has reached the permissible temperature, resp. until the condensation has evaporated before it may be turned on! Ordinarily this will be the case after 2 hours. The oscilloscope is destined for use in clean and dry environments. Do not operate in dusty or chemically aggressive atmosphere or if there is danger of explosion.
The operating position may be any, however, suffi cient ventilation must be ensured (convecti on cooling). P rolonged operation requires the horizontal or inclined position.
Do not obstruct the ventilation holes!
Specifi cations are valid after a 30 minute warm-up period between 15 and 30 degr. C. Specifi cations without tolerances are average values.
Line voltage
The instrument has a wide range power supply from 105 to 253 V, 50 or 60 Hz ±10%. There is hence no line voltage selector.
The line fuse is accessible on the rear panel and part of the line input connector. Prior to exchanging a fuse the line cord must be pulled out. Exchange is only allowed if the fuse holder is undamaged, it can be taken out using a screwdriver put into the slot. The fuse can be pushed out of its holder and exchanged.
The holder with the new fuse can then be pushed back in place against the spring. It is prohibited to ”repair“ blown fuses or to bridge the fuse. Any damages incurred by such measures will void the warranty.
Type of fuse:
Size 5 x 20 mm; 250V~, C; IEC 127, Bl. III; DIN 41 662 (or DIN 41 571, Bl. 3). Cut off: slow blow (T) 0,8A.
Subject to change without notice
7
Type of signal voltage
Type of signal voltage
The oscilloscope HM507 allows examination of DC voltages and most repetitive signals in the frequency range up to at least 40MHz (-3dB).
The vertical amplifiers have been designed for minimum overshoot and therefore permit a true signal display.
The display of sinusoidal signals within the bandwidth limits causes no problems, but an increasing error in measurement due to gain reduction must be taken into account when measuring high frequency signals. This error becomes noticeable at approx. 14MHz. At approx. 18MHz the reduction is approx. 10% and the real voltage value is 11% higher. The gain reduction error can not be defined exactly as the -3dB bandwidth of the amplifiers differ between 40MHz and 42MHz. For sinewave signals the -6dB limit is approx. 50MHz.
When examining square or pulse type waveforms, attention must be paid to the harmonic content of such signals. The repetition frequency (fundamental frequency) of the signal must therefore be significantly smaller than the upper limit frequency of the vertical amplifier.
Displaying composite signals can be difficult, especially if they contain no repetitive higher amplitude content which can be used for triggering. This is the case with bursts, for instance. To obtain a well-triggered display in this case, the assistance of the variable holdoff function or the delayed time base may be required. Television video signals are relatively easy to trigger using the built-in TV-Sync-Separator (TV).
For optional operation as a DC or AC voltage amplifier, each vertical amplifier input is provided with a DC/AC switch. DC coupling should only be used with a series-connected attenuator probe or at very low frequencies or if the measurement of the DC voltage content of the signal is absolutely necessary.
When displaying very low frequency pulses, the flat tops may be sloping with AC coupling of the vertical amplifier (AC limit frequency approx. 1.6 Hz for 3dB). In this case, DC operation is preferred, provided the signal voltage is not superimposed on a too high DC level. Otherwise a capacitor of adequate capacitance must be connected to the input of the vertical amplifier with DC coupling. This capacitor must have a sufficiently high breakdown voltage rating. DC coupling is also recommended for the display of logic and pulse signals, especially if the pulse duty factor changes constantly. Otherwise the display will move upwards or downwards at each change. Pure direct voltages can only be measured with DC-coupling.
Voltage values of a sine curve
Vrms = effective value; Vp = simple peak or crest value; Vpp = peak-to-peak value; Vmom = momentary value.
The minimum signal voltage which must be applied to the Y input for a trace of 1div height is 1mVpp (± 5%) when this deflection coefficient is displayed on the screen (readout) and the vernier is switched off (VAR-LED dark). However, smaller signals than this may also be displayed. The deflection coefficients are indicated in mV/div or V/div (peak-to-peak value).
The magnitude of the applied voltage is ascertained by multiplying the selected deflection coefficient by the vertical display height in div. If an attenuator probe x10 is used, a further multiplication by a factor of 10 is required to ascertain the correct voltage value.
For exact amplitude measurements, the variable control (VAR) must be set to its calibrated detent CAL position.
With the variable control activated the deflection sensitivity can be reduced up to a ratio of 2.5 to 1 (please note “controls and readout”). Therefore any intermediate value is possible within the 1-2-5 sequence of the attenuator(s).
With direct connection to the vertical input, sig­nals up to 400Vpp may be displayed (attenuator set to 20V/div, variable control to 2.5:1).
With the designations
H = display height in div, U = signal voltage in Vpp at the vertical input, D = deflection coefficient in V/div at attenuator switch,
the required value can be calculated from the two given quantities:
The input coupling is selectable by the AC/DC pushbutton. The actual setting is displayed in the readout with the = symbol for DC- and the
~ symbol for AC coupling.
Amplitude Measurements
In general electrical engineering, alternating voltage data normally refers to effective values (rms = root-mean-square value). However, for signal magnitudes and voltage designations in oscilloscope measurements, the peak-to-peak voltage (Vpp) value is applied. The latter corresponds to the real potential difference between the most positive and most negative points of a signal waveform.
If a sinusoidal waveform, displayed on the oscilloscope screen, is to be converted into an effective (rms) value, the resulting peak­to-peak value must be divided by 2x2 = 2.83. Conversely, it should be observed that sinusoidal voltages indicated in Vrms (Veff) have 2.83 times the potential difference in Vpp. The relationship between the different voltage magnitudes can be seen from the following figure.
8
However, these three values are not freely selectable. They have to be within the following limits (trigger threshold, accuracy of reading):
H between 0.5 and 8div, if possible 3.2 to 8div, U between 0.5mVpp and 160Vpp, D between 1mV/div and 20V/div in 1-2-5 sequence.
Examples: Set deflection coefficient D = 50mV/div 0.05V/div, observed display height H = 4.6div, required voltage U = 0.05x4.6 = 0.23Vpp. Input voltage U = 5Vpp, set deflection coefficient D = 1V/div, required display height H = 5:1 = 5div. Signal voltage U = 230Vrmsx2 (voltage
> >
> 160Vpp, with probe 10:1: U = 65.1Vpp),
> >
√√
2 = 651Vpp
√√
desired display height H = min. 3.2div, max. 8div,
Subject to change without notice
Type of signal voltage
max. deflection coefficient D = 65.1:3.2 = 20.3V/div, min. deflection coefficient D = 65.1:8 = 8.1V/div, adjusted deflection coefficient D = 10V/div.
The previous examples are related to the CRT graticule reading. The results can also be determined with the aid of the DV cursor measurement (please note controls and readout). The input voltage must not exceed 400V, independent from the polarity.
If an AC voltage which is superimposed on a DC voltage is applied, the maximum peak value of both voltages must not exceed + or - 400V. So for AC voltages with a mean value of zero volt the maximum peak to peak value is 800Vpp.
If attenuator probes with higher limits are used, the probes limits are valid only if the oscilloscope is set to DC input coupling.
If DC voltages are applied under AC input coupling conditions the oscilloscope maximum input voltage value remains 400V. The attenuator consists of a resistor in the probe and the 1M input resistor of the oscilloscope, which are disabled by the AC input coupling capacity when AC coupling is selected. This also applies to DC voltages with superimposed AC voltages. It also must be noted that due to the capacitive resistance of the AC input coupling capacitor, the attenuation ratio depends on the signal frequency. For sinewave signals with frequencies higher than 40Hz this influence is negligible.
Time Measurements
As a rule, most signals to be displayed are periodically repeating processes, also called periods. The number of periods per second is the repetition frequency. Depending on the time base setting (TIME/DIV.-knob) indicated by the readout, one or several signal periods or only a part of a period can be displayed. The time coefficients are stated in ms/div, µs/div or ns/div. The following examples are related to the CRT graticule reading. The results can also be determined with the aid of the T and 1/ note controls and readout).
The duration of a signal period or a part of it is determined by multiplying the relevant time (horizontal distance in div) by the (calibrated) time coefficient displayed in the readout. Uncalibrated, the time base speed can be reduced until a maximum factor of 2.5 is reached. Therefore any intermediate value is possible within the 1-2-5 sequence.
With the designations
L = displayed wave length in div of one period, T = time in seconds for one period, F = recurrence frequency in Hz of the signal, Tc = time coefficient in ms, µs or ns/div and the relation F = 1/T, the following equations can be stated:
∆∆
T cursor measurement (please
∆∆
With the above listed exceptions HAMEG 10:1 probes can be used for DC measurements up to 600V or AC voltages (with a mean value of zero volt) of 1200Vpp. The 100:1 probe HZ53 allows for 1200V DC or 2400Vpp for AC.
It should be noted that its AC peak value is derated at higher frequencies. If a normal x10 probe is used to measure high voltages there is the risk that the compensation trimmer bridging the attenuator series resistor will break down causing damage to the input of the oscilloscope. However, if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope, a normal x10 probe is sufficient. In this case, an appropriate high voltage capacitor (approx. 22­68nF) must be connected in series with the input tip of the probe.
With Y-POS. control (input coupling to GD) it is possible to use a horizontal graticule line as reference line for ground potential before the measurement. It can lie below or above the horizontal central line according to whether positive and/or negative deviations from the ground potential are to be measured.
Total value of input voltage
However, these four values are not freely selectable. They have to be within the following limits:
L between 0.2 and 10div, if possible 4 to 10div, T between 10ns and 5s, F between 0.5Hz and 100MHz, Tc between 100ns/div and 500ms/div in 1-2-5 sequence
(with X-MAG. (x10) inactive), and
Tc between 10ns/div and 50ms/div in 1-2-5 sequence
(with X-MAG. (x10) active).
Examples:Examples:
Examples:
Examples:Examples: Displayed wavelength L = 7div,Displayed wavelength L = 7div,
Displayed wavelength L = 7div,
Displayed wavelength L = 7div,Displayed wavelength L = 7div, set time coefficient Tc = 100ns/div,set time coefficient Tc = 100ns/div,
set time coefficient Tc = 100ns/div,
set time coefficient Tc = 100ns/div,set time coefficient Tc = 100ns/div, required period T = 7x100x10required period T = 7x100x10
required period T = 7x100x10
required period T = 7x100x10required period T = 7x100x10 required rec. freq. F = 1:(0.7x10required rec. freq. F = 1:(0.7x10
required rec. freq. F = 1:(0.7x10
required rec. freq. F = 1:(0.7x10required rec. freq. F = 1:(0.7x10
Signal period T = 1s,Signal period T = 1s,
Signal period T = 1s,
Signal period T = 1s,Signal period T = 1s, set time coefficient Tc = 0.2s/div,set time coefficient Tc = 0.2s/div,
set time coefficient Tc = 0.2s/div,
set time coefficient Tc = 0.2s/div,set time coefficient Tc = 0.2s/div, required wavelength L = 1:0.2 = 5div.required wavelength L = 1:0.2 = 5div.
required wavelength L = 1:0.2 = 5div.
required wavelength L = 1:0.2 = 5div.required wavelength L = 1:0.2 = 5div.
Displayed ripple wavelength L = 1div,Displayed ripple wavelength L = 1div,
Displayed ripple wavelength L = 1div,
Displayed ripple wavelength L = 1div,Displayed ripple wavelength L = 1div, set time coefficient Tc = 10ms/div,set time coefficient Tc = 10ms/div,
set time coefficient Tc = 10ms/div,
set time coefficient Tc = 10ms/div,set time coefficient Tc = 10ms/div, required ripple freq. F = 1:(1x10x10required ripple freq. F = 1:(1x10x10
required ripple freq. F = 1:(1x10x10
required ripple freq. F = 1:(1x10x10required ripple freq. F = 1:(1x10x10
-9-9
-9
-9-9
= 0.7µs = 0.7µs
= 0.7µs
= 0.7µs = 0.7µs
-6-6
-6
-6-6
) = 1.428MHz.) = 1.428MHz.
) = 1.428MHz.
) = 1.428MHz.) = 1.428MHz.
-3-3
-3
-3-3
) = 100Hz.) = 100Hz.
) = 100Hz.
) = 100Hz.) = 100Hz.
The dotted line shows a voltage alternating at zero volt level. If superimposed on a DC voltage, the addition of the positive peak and the DC voltage results in the max. voltage (DC + ACpeak).
Subject to change without notice
TV-line frequency F = 15625Hz,TV-line frequency F = 15625Hz,
TV-line frequency F = 15625Hz,
TV-line frequency F = 15625Hz,TV-line frequency F = 15625Hz, set time coefficient Tc = 10µs/div,set time coefficient Tc = 10µs/div,
set time coefficient Tc = 10µs/div,
set time coefficient Tc = 10µs/div,set time coefficient Tc = 10µs/div, required wavelength L = 1:(15 625x10required wavelength L = 1:(15 625x10
required wavelength L = 1:(15 625x10
required wavelength L = 1:(15 625x10required wavelength L = 1:(15 625x10
Sine wavelength L = min. 4div, max. 10div,Sine wavelength L = min. 4div, max. 10div,
Sine wavelength L = min. 4div, max. 10div,
Sine wavelength L = min. 4div, max. 10div,Sine wavelength L = min. 4div, max. 10div, Frequency F = 1kHz,Frequency F = 1kHz,
Frequency F = 1kHz,
Frequency F = 1kHz,Frequency F = 1kHz, max. time coefficient Tc = 1:(4x10max. time coefficient Tc = 1:(4x10
max. time coefficient Tc = 1:(4x10
max. time coefficient Tc = 1:(4x10max. time coefficient Tc = 1:(4x10 min. time coefficient Tc = 1:(10x10min. time coefficient Tc = 1:(10x10
min. time coefficient Tc = 1:(10x10
min. time coefficient Tc = 1:(10x10min. time coefficient Tc = 1:(10x10 set time coefficient Tc = 0.2ms/div,set time coefficient Tc = 0.2ms/div,
set time coefficient Tc = 0.2ms/div,
set time coefficient Tc = 0.2ms/div,set time coefficient Tc = 0.2ms/div, required wavelength L = 1:(10required wavelength L = 1:(10
required wavelength L = 1:(10
required wavelength L = 1:(10required wavelength L = 1:(10 Displayed wavelength L = 0.8div,Displayed wavelength L = 0.8div,
Displayed wavelength L = 0.8div,
Displayed wavelength L = 0.8div,Displayed wavelength L = 0.8div, set time coefficient Tc = 0.5µs/div,set time coefficient Tc = 0.5µs/div,
set time coefficient Tc = 0.5µs/div,
set time coefficient Tc = 0.5µs/div,set time coefficient Tc = 0.5µs/div,
33
3
33
x0.2x10x0.2x10
x0.2x10
x0.2x10x0.2x10
-5-5
-5
-5-5
) = 6.4div.) = 6.4div.
) = 6.4div.
) = 6.4div.) = 6.4div.
33
3
33
) = 0.25ms/div,) = 0.25ms/div,
) = 0.25ms/div,
) = 0.25ms/div,) = 0.25ms/div,
33
3
33
) = 0.1ms/div,) = 0.1ms/div,
) = 0.1ms/div,
) = 0.1ms/div,) = 0.1ms/div,
-3-3
-3
-3-3
) = 5div.) = 5div.
) = 5div.
) = 5div.) = 5div.
9
Type of signal voltage
pressed X-MAG. (x10) button: Tc = 0.05µs/div, required rec. freq. F = 1:(0.8x0.05x10
-6
) = 25MHz,
required period T = 1:(25x106) = 40ns.
If the time is relatively short as compared with the complete signal period, an expanded time scale should always be applied (X-MAG. (x10) active). In this case, the time interval of interest can be shifted to the screen center using the X-POS. control.
When investigating pulse or square waveforms, the critical feature is the risetime of the voltage step. To ensure that transients, ramp-offs, and bandwidth limits do not unduly influence the measuring accuracy, the risetime is generally measured between 10% and 90% of the vertical pulse height. For measurement, adjust the Y deflection coefficient using its varia­ble function (uncalibrated) together with the Y-POS. control so that the pulse height is precisely aligned with the 0% and 100% lines of the internal graticule. The 10% and 90% points of the signal will now coincide with the 10% and 90% graticule lines. The risetime is given by the product of the horizontal distance in div between these two coincident points and the calibrated time coefficient setting. The fall time of a pulse can also be measured by using this method.
The following figure shows correct positioning of the oscilloscope trace for accurate risetime measurement.
with approximately constant group delay (therefore good pulse transmission performance) the following numerical relationship between rise time tr (in ns) and bandwidth B (in MHz) applies:
Connection of Test Signal
In most cases briefly depressing the AUTO SET causes a useful signal related instrument setting. The following explanations refer to special applications and/or signals, demanding a manual instrument setting. The description of the controls is explained in the section controls and readout”.
Caution: When connecting unknown signals to the oscilloscope input, always use a x10 probe, automatic triggering and set the input coupling switch to DC (readout). The attenuator should initially be set to 20V/div.
Sometimes the trace will disappear after an input signal has been applied. Then a higher deflection coefficient (lower input sensitivity) must be chosen until the vertical signal height is only 3-8div. With a signal amplitude greater than 160Vpp and the deflection coefficient (VOLTS/DIV.) in calibrated condition, an attenuator probe must be inserted before the vertical input. If, after applying the signal, the trace is nearly blanked, the period of the signal is probably substantially longer than the set time deflection coefficient (TIME/DIV.). It should be switched to an adequately larger time coefficient.
With a time coefficient of 10ns/div (X x10 magnification active), the example shown in the above figure results in a total measured risetime of
= 1.6div x 10ns/div = 16ns
t
tot
When very fast risetimes are being measured, the risetimes of the oscilloscope amplifier and of the attenuator probe has to be deducted from the measured time value. The risetime of the signal can be calculated using the following formula.
In this t oscilloscope amplifier (approx. 8.75ns), and t probe (e.g. = 2ns). If t as the risetime of the pulse, and calculation is unnecessary.
is the total measured risetime, t
tot
is greater than 100ns, then t
tot
is the risetime of the
osc
the risetime of the
p
can be taken
tot
Calculation of the example in the figure above results in a signal risetime
= √162 - 8.752 - 22 = 13.25ns
t
r
The measurement of the rise or fall time is not limited to the trace dimensions shown in the above diagram. It is only particularly simple in this way. In principle it is possible to measure in any display position and at any signal amplitude. It is only important that the full height of the signal edge of interest is visible in its full length at not too great steepness and that the horizontal distance at 10% and 90% of the amplitude is measured. If the edge shows rounding or overshooting, the 100% should not be related to the peak values but to the mean pulse heights. Breaks or peaks (glitches) next to the edge are also not taken into account. With very severe transient distortions, the rise and fall time measurement has little meaning. For amplifiers
The signal to be displayed can be connected directly to the Y­input of the oscilloscope with a shielded test cable such as HZ32 or HZ34, or reduced through a x10 or x100 attenuator probe. The use of test cables with high impedance circuits is only recommended for relatively low frequencies (up to approx. 50kHz). For higher frequencies, the signal source must be of low impedance, i.e. matched to the characteristic resistance of the cable (as a rule 50). Especially when transmitting square and pulse signals, a resistor equal to the characteristic impedance of the cable must also be connected across the cable directly at the Y-input of the oscilloscope. When using a 50 cable such as the HZ34, a 50 through termination type HZ22 is available from HAMEG. When transmitting square signals with short rise times, transient phenomena on the edges and top of the signal may become visible if the correct termination is not used. A terminating resistance is sometimes recommended with sine signals as well. Certain amplifiers, generators or their attenuators maintain the nominal output voltage independent of frequency only if their connection cable is terminated with the prescribed resistance. Here it must be noted that the terminating resistor HZ22 will only dissipate a maximum of 2Watts. This power is reached with 10Vrms or at 28.3Vpp with sine signal. If a x10 or x100 attenuator probe is used, no termination is necessary. In this case, the connecting cable is matched directly to the high impedance input of the oscilloscope. When using attenuators probes, even high internal impedance sources are only slightly loaded (approx. 10M II 12pF or 100M II 5pF with HZ53). Therefore, if the voltage loss due to the attenuation of the probe can be compensated by a higher amplitude setting, the probe should always be used. The series impedance of the probe provides a certain amount of protection for the input of the vertical amplifier. Because of their separate manufacture, all attenuator probes are only partially compensated, therefore accurate compensation must be performed on the oscilloscope (see Probe compensation ).
Standard attenuator probes on the oscilloscope normally reduce its bandwidth and increase the rise time. In all cases where the oscilloscope bandwidth must be fully utilized (e.g. for pulses
10
Subject to change without notice
Controls and readout
with steep edges) we strongly advise using the probes HZ51 (x10) HZ52 (x10 HF) and HZ54 (x1 and x10). This can save the purchase of an oscilloscope with larger bandwidth.
The probes mentioned have a HF-calibration in addition to low frequency calibration adjustment. Thus a group delay correction to the upper limit frequency of the oscilloscope is possible with the aid of an 1MHz calibrator, e.g. HZ60.
In fact the bandwidth and rise time of the oscilloscope are not noticeably changed with these probe types and the waveform reproduction fidelity can even be improved because the probe can be matched to the oscilloscopes individual pulse response.
If a x10 or x100 attenuator probe is used, DC input coupling must always be used at voltages above 400V. With AC coupling of low frequency signals, the attenuation is no longer independent of frequency, pulses can show pulse tilts. Direct voltages are suppressed but load the oscilloscope input coupling capacitor concerned. Its voltage rating is max. 400 V (DC + peak AC). DC input coupling is therefore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max. 1200 V (DC + peak AC). A capacitor of corresponding capacitance and voltage rating may be connected in series with the attenuator probe input for blocking DC voltage (e.g. for hum voltage measurement).
With all attenuator probes, the maximum AC input voltage must be derated with frequency usually above 20kHz. Therefore the derating curve of the attenuator probe type concerned must be taken into account. The selection of the ground point on the test object is important when displaying small signal voltages. It should always be as close as possible to the measuring point. If this is not done, serious signal distortion may result from spurious currents through the ground leads or chassis parts. The ground leads on attenuator probes are also particularly critical. They should be as short and thick as possible. When the attenuator probe is connected to a BNC-socket, a BNC­adapter, should be used. In this way ground and matching problems are eliminated. Hum or interference appearing in the measuring circuit (especially when a small deflection coefficient is used) is possibly caused by multiple grounding because equalizing currents can flow in the shielding of the test cables (voltage drop between the protective conductor connections, caused by external equipment connected to the mains/line, e.g. signal generators with interference protection capacitors).
B: Menu Display and Operation
Operation of some pushbuttons activates the display of menus. There are Standard and Pulldown Menus.
Standard menus: When a standard menu is displayed, all other readout information (e.g. parameter settings) are switched off. The readout then consists of the menu headline, and the respective menu functions. At the bottom of the graticule are displayed symbols and commands which can be operated by the pushbuttons related to them below.
Esc switches one step back in the menu hierarchy.Exitcloses the menu and switches back to the operating
conditions present before calling the menu.
Set calls the selected menu item or starts a function.SAVEresults in storage.Editcalls the edit menu.
The pushbuttons below the triangle and arrow symbols select one item that is then highlighted. If in addition Use INT./FOC. knob to select is displayed, the INT./FOC. knob can be used to select within the item. Where a [ ] symbol appears in an activated line, a [x]/[ ] symbol is displayed with the other command symbols at the bottom of the screen. The pushbutton below the symbol is used for switchover (toggle).
Pulldown menus:
After pressing a pushbutton which calls a pulldown menu, the instrument parameter settings are still displayed. The readout only changes in respect to the called parameter (e.g. input coupling) and now shows all selectable parameter options (in case of input coupling: AC, DC and GND). The previously displayed parameter doesnt change but is displayed highlighted. Each time the pushbutton is briefly pressed the next parameter becomes active and highlighted, as long as the pulldown menu is displayed. Without further pressing of the pushbutton, the pulldown menu extinguishes after a few seconds and the selected parameter, the CURSOR line(s) and the measuring result are displayed in the normal way.
C: READOUT Information
Controls and Readout
A: Basic settings
The following description assumes that:
1. Component Tester” is switched off.
2. The following settings are present under
MAIN MENU > SETUP & INFO > MISCELLANEOUS:
2.1 CONTROL BEEP and ERROR BEEP activated (x),
2.2 QUICK START not activated.
3. The screen Readout is visible.
The LED indicators on the large front panel facilitate operation and provide additional information. Electrical end positions of controls are indicated by acoustic signal (beep).
All controls, except the power switch (POWER), are electronically set and interrogated. Thus, all electronically set functions and their current settings can be stored and also remotely controlled.
Subject to change without notice
The readout alphanumerically displays the scope parameter settings, measurement results and CURSOR lines. Which information is displayed depends on the actual instrument settings. The following list contains the most important display information.
Top of the graticule from left to right:
1. time deflection coefficient and additionally the sampling rate in digital mode.
11
Controls and readout
2. trigger source, slope and coupling.
3. operating condition of delay time base in analog mode; or in digital mode, pre or post trigger time.
4. measuring results.
Bottom of the graticule from left to right:
1. probe symbol (x10), Y deflection coefficient and input
coupling channel I.
2. “+” symbol (addition).
3. probe symbol (x10), Y deflection coefficient and input
coupling channel II.
4. channel mode (analog) or signal display mode (digital).
The trigger point symbol is displayed at the left graticule border line (analog mode). The CURSOR lines can take any position within the graticule.
D: Description of Controls
Preliminary note: For better identification all controls are numbered consecutively. A number within a square indicates a control which is for digital mode. The latter will be described at the end of the listing.
The large front panel is, as usual with Hameg oscilloscopes, marked with several fields.
The following controls and LED indicators are located on the top, to the right of the screen, above the horizontal line:
If the signal height is insufficient, the not change. In to the signal which is used for internal triggering.
Voltage CURSOR
If voltage measurement is present, the CURSOR lines are automatically set to the positive and negative peak value of the signal. The accuracy of this function decreases with higher frequencies and is also influenced by the signal‘s pulse duty factor.
Time/Frequency CURSOR
If complex waveforms such as video signals are applied, the cursor lines may not align exactly with one period and give a false reading.
DIGITAL MODE ONLY
If ROLL (rol) or SINGLE (sgl) is active, AUTOSET switches to the last used REFRESH mode.
(3) INT./FOC. Knob for intensity and focus setting, with associated LEDs and TRACE ROT. pushbutton.
3.1 Briefly pressing the TRACE ROT. pushbutton switches over the INT./FOC. knob to another function, which is indicated by an LED. If the readout off, the sequence is A, FOC, RO, A. In condition READOUT deactivated, the switching sequence is A, FOC, A.
3.1.1 “A”:
The INT./FOC. knob controls the signal(s) intensity. Turning this knob clockwise increases the intensity. Only the minimum required trace intensity should be used, depending on signal parameters, oscilloscope settings and light conditions.
DUAL
..
.
..
mode the
..
.
..
CURSOR
CURSOR
(RO)
lines do
lines are related
is not switched
(1) POWER
After the oscilloscope is switched on, all LEDs are lit and an automated instrument test is performed. During this time the HAMEG logo and the software version are displayed on the screen. After the internal test is completed successfully, the overlay is switched off and the normal operation mode is present. The last used settings and the readout then become activated. An LED (3) indicates the ON condition.
(2) AUTOSET
Briefly pressing this pushbutton results in an automatic instrument setting selecting Yt mode as the default. The instrument is set to the last used Yt mode setting (CH I,
CH II
intensity had been present before. The operation mode (analog or digital) will not be changed.
The instrument is set automatically to normal (undelayed) time base mode, even if the previous Yt mode included search (“sea”), delay (del) or triggered delay (“dTr) time base mode.
Please also note AUTOSET in section First Time Operation”.
Automatic If CURSOR CURSOR lines are set automatically under suitable conditions and the readout briefly displays SETTING CURSOR”.
Pushbutton and symbols for ON (I) and OFF (O).
or
DUAL)
and to a medium trace intensity, if less
CURSOR
lines are displayed and AUTOSET is chosen the
positioning:
3.1.2 FOC“:
The INT./FOC. knob controls both the trace and the readout sharpness. Note: The electron beam diameter gets larger with a higher trace intensity and the trace sharpness decreases. This can be corrected to a certain extent. Assuming that the trace sharpness was set to optimum in the screen center, it is unavoidable that the trace sharpness decreases with an increasing distance from the center.
Since the settings of the signal(s) intensity (A) and the READOUT (RO) are usually different, the FOCUS should be set for optimum signal(s) sharpness. The sharpness of the READOUT then can be improved by reducing the READOUT intensity.
(4) RM
The remote control mode can be switched on or off (RM LED dark) via the lit, all electronically selectable controls on the front panel are inactive. This state can be cancelled by depressing the
AUTOSET
the interface.
(5) RECALL / SAVE
The instrument contains 9 non volatile memories. These can be used by the operator to save instrument settings and to recall them.
SAVE:
Press and hold the RECALL/SAVE button to start a storage process. This causes the SAVE menu (Standard menu, note B: Menu-Display and Operation) to be displayed. Choose the memory location cipher (highlighted) by pressing a
RS232
interface. When the RM LED is
pushbutton provided it was not deactivated via
Pushbutton for instrument settings
12
Subject to change without notice
Controls and readout
pushbutton underneath the triangle symbols. Briefly press the pushbutton underneath SET to store the last instrument setting and return from menu display to previous mode. If the SAVE function was called inad­vertently, it can be switched off with “Esc”.
Switching the instrument off automatically stores the current settings in memory location 9 (PWR OFF = Power Off), with the effect that different settings previously stored in this location get lost. To prevent this, RECALL 9 before switching the instrument off.
RECALL:
Briefly pressing calls the RECALL menu. You can select the required memory location using a triangle push­button. Recall the previously stored instrument settings by briefly pressing the SET pushbutton or briefly press Esc if the function was called inadvertently.
RECALL also offers the item DEFAULTS, which covers all functions.
The setting controls and LED’s for the Y amplifiers, modes, triggering and time base are located under­neath the sector of the front panel described above.
coupling, a trace position in vertical direction. The DC voltage then can be determined by taking the deflection coefficient, the probe factor and the trace position change with respect to the previous 0 Volt position into account.
0 Volt” Symbol:
The READOUT indicates the 0 Volt trace position of channel I by a center line in CHI and DUAL used, this symbol changes to an arrow symbol pointing outside the graticule just before the trace goes outside the graticule limits
If addition mode (add) is present just one ? is visible.
In XY mode the 0 Volt trace position for channel I (X) and channel II (Y) is symbolised by triangle symbols at the right graticule border (Y) and above the Y deflection coefficient display. The “triangle” symbol(s) point(s) outside the graticule when the “0 Volt” trace position is outside the graticule.
CURS.I:
The CURSOR lines marked by the symbol “I” can be shifted by the Y-POS/CURS. I control knob, if the CURSOR POS LED (7) is lit.
DCDC
DC signal applied at the input changes the
DCDC
””
⊥⊥
””
symbol to the left of the screens vertical
””
⊥⊥
””
mode. When Y position is
” symbol
(6) Y-POS/CURS. I Control knob with several functions.
This knob allows position control of channel I trace or CURSOR line(s). Briefly pressing the CURSOR POS pushbutton (7) selects the function. If the CURSOR line(s) are not displayed the CURS. I function is not selectable.
Y-POS:
The vertical trace position of channel I can be set with this control knob, if the CURSOR POS (7) LED isnt lit. In addition (add) mode both (Y-POS/CURS. I (6) and Y POS/ CURS. II) control knobs are active. If the instrument is set
XY XY
to
XY mode this control knob is
XY XY
(12)
knob must be used for horizontal positioning.
DC voltage measurement:
If no signal is applied at the INPUT CHI (25), the vertical trace position represents 0 Volt. This is the case if INPUT CHI (25) or in addition (ADD) mode, both INPUT CHI (25) and INPUT CHII (28), are set to GND (ground) (26) (29) and automatic triggering (AT (9)) is present to make the trace visible.
The trace can then be set to the vertical position best suited for the following DC voltage measurement. After switching GND
(ground) off and selecting DC input
inactiveinactive
inactive and the
inactiveinactive
X POS.
STORAGE MODE ONLY
In contrast to analog mode the Y-POS/CURS.I (6) knob must be used for X position shift in XY mode and the X-POS. knob is disabled.
The Y-POS/CURS.I (6) knob can also be used for shifting a signal position although it is stored by HOLD (hld).
If a REFERENCE or MATH (mathematic) signal is displayed and the M/R [38] LED is lit, the Y-POS/CURS.I (6) knob serves as a MATHEMATIC or REFERENCE position control.
(7) CURSOR POS
Briefly pressing this pushbutton determines the function of the Y POS/CURS.I (6) and Y POS/CURS.II (8) controls.
If the CUR LED is not lit the Y position control function is active.
Provided that the CURSOR lines are activated, the LED can be switched on by briefly pressing the CURSOR POS pushbutton. Then the controls (6) and (8) are switched over from Y position to CURSOR position control (CURS.I (6) and CURS.II (8)). Briefly pressing this pushbutton once again switches back to the Y position control function.
The CUR LED extinguishes after a (mathematic) signal is displayed and the M/R LED is switched on by the MATH/REF POS [38] pushbutton. Under these conditions the Y-POS/CURS.I (6) knob serves as a REFERENCE or MATH (mathematic) signal position control, while Y-POS/ CURS.II (8) affects the channel II signal if present.
(8) Y POS/CURS. II Control knob with two functions.
This knob enables position control of channel II trace or CURSOR line(s). Briefly pressing the CURSOR POS pushbutton (7) selects the function. If the CURSOR line(s) are not displayed the CURS. I function is not selectable.
Pushbutton and LED.
STORAGE MODE ONLY
REFERENCE or MATH
Subject to change without notice
13
Controls and readout
Y POS:
The vertical trace position of channel II can be set with this control knob, if the CURSOR POS LED isnt lit. In addition (“add”) mode both (Y POS/CURS. I (6) and
Y
POS/CURS. II) control knobs are active. If the instrument is set to POS. (12) POS. (12)
POS. (12) knob must be used for horizontal positioning.
POS. (12) POS. (12)
XY XY
XY mode, this control knob is
XY XY
inactiveinactive
inactive and the
inactiveinactive
X X
X
X X
DC voltage measurement:
If no signal is applied at the INPUT CHII (28), the vertical trace position represents 0 Volt. This is the case if INPUT CHII (28) or in addition (ADD) mode, both INPUT CHI (25) and INPUT CHII automatic triggering (AT (9))
(28), are set to GND (ground) (26) (29) and
is present to make the trace
visible.
The trace can then be set to the vertical position best suited for the following DC voltage measurement. After switching GND
(ground) off and selecting DC input
coupling, a DC signal applied at the input changes the trace position in vertical direction. The
DCDC
DC voltage then
DCDC can be determined by taking the deflection coefficient, the probe factor and the trace position change with respect to the previous 0 Volt position into account.
0 Volt” Symbol:
The READOUT indicates the 0 Volt trace position of channel II by a vertical center line in CHII and DUAL
⊥⊥
symbol to the right of the screens
⊥⊥
mode. When Y position is used, this symbol changes to an arrow symbol pointing outside the graticule just before the trace goes outside the graticule limits
If addition mode (add) is present just one symbol is visible.
In the automatic peak value triggering condition the LEVEL control (11) range is limited to the trigger signal positive and negative peak values. Automatic triggering without peak value detection enables the trigger point to be set outside the signal amplitude range. In the latter case, although untriggered, there is still a signal display.
Whether the peak value detection is active or not depends on the operating mode and the selected trigger coupling. The actual state is recognised by the behaviour of the trigger point symbol when changing the LEVEL setting.
NM:
Normal triggering disables both the automatic trigger and the peak value detection so even low frequency signals can be displayed in a stable manner. Without suitable input signal height, trigger coupling and LEVEL settings, no trace will be displayed.
Analog only: The last LEVEL setting of the time base is stored, then the control again becomes active when selecting triggering after delay (DEL.MODE (“dTr)) time base mode (quasi 2nd time base). In combination with In dTr mode the LEVEL control is operative for the “2
nd
time base”.
/ \ (Slope selection):
Each time this pushbutton is briefly pressed, the slope direction switches from falling edge to rising edge and vice versa. The current setting is displayed in the readout by a slope symbol. The last setting in undelayed time base mode is stored and still active if triggered delay
(dTr”)
time base mode is selected (analog only). This allows for a different slope setting for the triggered
DELAY (dtr)
time base mode.
In XY mode the 0 Volt trace position for channel I (X) and channel II (Y) is symbolised by triangle symbols at the right graticule border (Y) and above the Y deflection coefficient display. The “triangle” symbol(s) point(s) outside the graticule when the “0 Volt” trace position is outside the graticule.
CURS.II:
If the CUR (7) LED is lit, the CURSOR line(s) marked with the symbol II can be shifted by the Y-POS/CURS. II (8) control knob.
STORAGE MODE ONLY
The Y-POS/CURS.II (8) knob can also be used for shifting a signal position although it is stored by HOLD (hld).
(9) NM AT Pushbutton with a double function and associated
NM LED.
NM AT selection:
Press and hold the pushbutton to switch over from automatic (peak value) to normal triggering (NM LED above the pushbutton lit) and vice versa. If the LED is dark, automatic or automatic peak value triggering is selected.
AT:
Automatic triggering can be carried out with or without peak capture. In both cases the LEVEL control (11) is effective and the trace is visible even if no signal is applied or trigger settings are unsuitable. Signal frequencies below the automatic trigger frequency can not be triggered as the automatic trigger cycle starts to early for such signals.
(10) TR Trigger indicator LED.
The
TR
LED is lit in
Yt
mode if the triggering conditions are met for the first trigger unit used in undelayed time base mode. Whether the LED flashes or is lit constantly depends on the frequency of the trigger signal.
In XY mode the TR LED is switched off.
(11) LEVEL Control knob.
Turning the
LEVEL
knob causes a different trigger point setting (voltage). The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point. In most Yt modes the trigger point is displayed in the readout by the symbol on the left vertical graticule line. If the trigger point symbol would overwrite other readout information
14
Subject to change without notice
Loading...
+ 30 hidden pages