Die HAMEG Instruments GmbH bescheinigt die Konformität für das Produkt
The HAMEG Instruments GmbH herewith declares conformity of the product
HAMEG Instruments GmbH déclare la conformite du produit
Bezeichnung / Product name / Designation:
Oszilloskop/Oscilloscope/Oscilloscope
Typ / Type / Type: HM504-2
mit / with / avec: –
Optionen / Options / Options: –
mit den folgenden Bestimmungen / with applicable regulations / avec les
directives suivantes
EMV Richtlinie 89/336/EWG ergänzt durch 91/263/EWG, 92/31/EWG
EMC Directive 89/336/EEC amended by 91/263/EWG, 92/31/EEC
Directive EMC 89/336/CEE amendée par 91/263/EWG, 92/31/CEE
Niederspannungsrichtlinie 73/23/EWG ergänzt durch 93/68/EWG
Low-Voltage Equipment Directive 73/23/EEC amended by 93/68/EEC
Directive des equipements basse tension 73/23/CEE amendée par 93/68/CEE
KONFORMITÄTSERKLÄRUNG
DECLARATION OF CONFORMITY
DECLARATION DE CONFORMITE
EN 61000-3-2/A14
Oberschwingungsströme / Harmonic current emissions / Émissions de courant
harmonique: Klasse / Class / Classe D.
EN 61000-3-3
Spannungsschwankungen u. Flicker / Voltage fl uctuations and fl icker / Fluctuations
de tension et du fl icker.
Datum / Date / Date Unterschrift / Signature / Signatur
25.6.2003
G. Hübenett
Product Manager
B.
General information regarding the CE marking
HAMEG instruments fulfi ll the regulations of the EMC directive. The conformity test made by HAMEG is based on the actual
generic- and product standards. In cases where different limit values are applicable, HAMEG applies the severer standard.
For emission the limits for residential, commercial and light industry are applied. Regarding the immunity (susceptibility)
the limits for industrial environment have been used. The measuring- and data lines of the instrument have much infl uence
on emmission and immunity and therefore on meeting the acceptance limits. For different applications the lines and/or
cables used may be different. For measurement operation the following hints and conditions regarding emission and immunity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and external devices, (computer, printer etc.) suffi ciently
screened cables must be used. Without a special instruction in the manual for a reduced cable length, the maximum cable
length of a dataline must be less than 3 meters and not be used outside buildings. If an interface has several connectors
only one connector must have a connection to a cable. Basically interconnections must have a double screening. For IEEEbus purposes the double screened cables HZ72S and HZ72L from HAMEG are suitable.
cable HZ72 from HAMEG is suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and instrument should be as short as possible. Without
instruction in the manual for a shorter length, signal lines must be less than 3 meters and not be used outside buildings.
Signal lines must screened (coaxial cable - RG58/U). A proper ground connection is required. In combination with signal
generators double screened cables (RG223/U, RG214/U) must be used.
3. Infl uence on measuring instruments
Under the presence of strong high frequency electric or magnetic fi elds, even with careful setup of the measuring equipment an infl uence of such signals is unavoidable. This will not cause damage or put the instrument out of operation. Small
deviations of the measuring value (reading) exceeding the instruments specifi cations may result from such conditions in
individual cases.
4. RF immunity of oscilloscopes
4.1 Electromagnetic RF fi eld
The infl uence of electric and magnetic RF fi elds may become visible (e.g. RF superimposed), if the fi eld intensity is high. In
most cases the coupling into the oscilloscope takes place via the device under test, mains/line supply, test leads, control
cables and/or radiation. The device under test as well as the oscilloscope may be effected by such fi elds. Although the interior of the oscilloscope is screened by the cabinet, direct radiation can occur via the CRT gap. As the bandwidth of each
amplifi er stage is higher than the total –3dB bandwidth of the oscilloscope, the infl uence RF fi elds of even higher frequencies may be noticeable.
4.2 Electrical fast transients / electrostatic discharge
Electrical fast transient signals (burst) may be coupled into the oscilloscope directly via the mains/line supply, or indirectly
via test leads and /or control cables. Due to the high trigger and input sensitivity of the oscilloscopes, such normally high
signals may effect the trigger unit and/or may become visible on the CRT, which is unavoidable. These effects can also be
caused by direct or indirect electrostatic discharge.
HAMEG Instruments GmbH
Subject to change without notice
3
Page 4
Subject to change without notice
4
HM504-2
� 2 Channels with deflection coefficients 1mV/div.…20V/div.
� Time Base 50ns /div.…0.5s/div.,
with X Magnification to 10ns/div.
� Low Noise Measuring Amplifiers with high pulse fidelity
� Triggering 0…100MHz from 5mm signal level
� Time Base delay provide high X Magnification
of any portion of the signal
� 100MHz 4-Digit Frequency Counter,
Cursor and Automatic Measurement
� Save/Recall Memories for Instrument Settings
� Readout, Autoset, no Fan
� Yt, XY and component-test modes
� RS-232 Interface (for parameter queries and control only)
5 0 M H z An a lo g O s c i ll o s co p e
H M 5 04 - 2
HM504-2
Optimum deflection
linearity
Rise-time measurement
with cursor
Full screen display of
50 MHz sine wave
Page 5
50MHz Analog Oscilloscope HM504-2
All data valid at 23 °C after 30 minute warm-up
Vertical Deflection
Operating Modes:Channel 1 or 2 only
Channels 1 and 2 (alternate or chopped)
Sum or Difference of CH 1 and CH 2
Hold-Off Time:variable to approx. 10 : 1
XY
Bandwidth X amplifier:0…3 MHz (-3dB)
XY Phase shift ‹ 3 °:‹120 kHz
Operation/Readout/Control
Manual:via controls
Autoset:automatic signal related parameter settings
Save and Recall:9 instrument parameter settings
Readout:display of menu, parameters, cursors and
results
Autom. Measurement:Freq./Period, V
dc
, Vpp, Vp+, Vp-,
Trigger Level
Cursor Measurement:Δt, 1/Δt, tr, ΔV, V to GND, Gain, Ratio X and Y
Frequency counter:4 digit (0.01 % ± 1 digit) 0.5 Hz…100MHz
Interface:RS-232
1)
Component Tester
Test Voltage:approx. 7V
rms
(open circuit)
Test Current:max. 7 mA
rms
(short-circuit)
Test Frequency:approx. 50Hz
Test Connection:2 banana jacks 4 mm Ø
One test circuit lead is grounded via protective earth (PE)
Miscellaneous
CRT:D14-363GY, 8 x 10 div. with internal graticule
Acceleration Voltage:approx. 2kV
Trace Rotation:adjustable on front panel
Z-input (Intens. modulation): max. + 5 V (TTL)
Calibrator Signal (Square Wave):0.2V, 1 Hz…1MHz (tr ‹ 4 ns), DC
Power Supply (Mains):105…253 V, 50/60 Hz ± 10 %, CAT II
Power Consumption:approx. 34Watt at 230 V/50 Hz
Safety class:Safety class I (EN61010-1)
Operating temperature:+5…+40°C
Storage temperature:-20…+70°C
Rel. humidity:5…80% (non condensing)
Dimensions (W x H x D):285 x 125 x 380 mm
Weight:approx. 5.4kg
1)
Device control and Parameter query, no CRT content transfer possible.
Accessories supplied: Line Cord, Operators Manual and Software for Windows
on CD-ROM, 2 Probes 1:1 /10:1 (HZ154), RS-232 Interface
Optional accessories:
HZ14Interface cable (serial) 1:1
HZ20Adapter, BNC to 4mm banana
HZ33Test cable 50Ω, BNC/BNC, 0,5m
HZ34Test cable 50Ω, BNC/BNC, 1m
HZ4319''-Rackmount Kit 3RU
HZ51Probe 10:1 (150MHz)
HZ52Probe 10:1 RF (250MHz)
HZ53Probe 100:1 (100MHz)
HZ56-2 AC/DC Current probe
HZ70Opto Interface (with optical fiber cable)
HZ100 Differential probe 20:1 / 200:1
HZ109 Differential probe 1:1 / 10:1
HZ115 Differential probe 100:1 / 1000:1
HZ200 Probe 10:1 with auto attenuation ID (250MHz)
HZ350 Probe 10:1 with automatically identification (350MHz)
HZ355 Slimline probe 10:1 with automatically identification (500MHz)
HZO20 High voltage probe 1000:1 (400MHz,1000Vrms)
HZO30 Active probe 1GHz (0,9pF, 1MΩ, including many accessories)
HZO50 AC/DC Current probe 20A, DC...100kHz
HZO51 AC/DC Current probe 1000A, DC...20kHz
Automatic(Peak to Peak):20Hz…100MHz (≥5mm)Normal with Level Control:0…100MHz (≥5mm)Slope:Rising or fallingSources:Channel1or2,CH1/CH2alternate
(≥8mm),LineandExternal
Coupling:AC (10Hz…100MHz), DC (0…100MHz),
HF (50kHz…100MHz), LF (0…1.5kHz)
Trigger Indicator:LEDTriggering after Delay:with Level Control and Slope selectionExternal Trigger Signal:≥0.3V
pp
(0…50MHz)
Active TV sync. separator:Field and Line, +/-
Horizontal Deflection
Time Base:50ns/div.…0.5s/div. (1-2-5 Sequence)
Accuracy:± 3%Variable (uncalibrated):› 2.5:1 to › 1.25s/div.
X Magnification x 10:up to 10ns/div. (± 5%)
Accuracy:± 5%
Delay(selectable):200
ns…140ms (variable)
Hold-Off Time:variable to approx. 10 : 1XYBandwidth X amplifier:0…3MHz (-3dB)XY Phase shift ‹3°:‹120kHz
Operation/Readout/Control
Manual:via controlsAutoset:automatic signal related parameter settingsSave and Recall:9 instrument parameter settingsReadout:display of menu, parameters, cursors and
Test Frequency:approx. 50HzTest Connection:2 banana jacks 4mm Ø
One test circuit lead is grounded via protective earth (PE)
Miscellaneous
CRT:D14-363GY, 8 x 10div. with internal graticuleAccelerationVoltage:approx. 2kVTraceRotation:adjustable on front panelZ-input (Intens. modulation):max. + 5V (TTL)Calibrator Signal (Square Wave):0.2V, 1Hz…1MHz (tr ‹ 4ns), DCPower Supply (Mains):105…253V, 50/60Hz ±10%, CAT IIPower Consumption:approx. 34Watt at 230V/50HzSafety class:Safety class I (EN61010-1)Operating temperature:+5…+40°CStoragetemperature:-20…+70°CRel. humidity:5…80% (non condensing)Dimensions (W x H x D):285 x 125 x 380mmWeight:approx. 5.4kg
1)
Device control and Parameter query, no CRT content transfer possible.
Accessories supplied:Line Cord, Operators Manual and Software for Windowson CD-ROM, 2 Probes 1:1/10:1(HZ154), RS-232 Interface
Optional accessories:
HZ14Interface cable (serial) 1:1HZ20Adapter, BNC to 4mm bananaHZ33Test cable 50Ω, BNC/BNC, 0,5m HZ34Test cable 50Ω, BNC/BNC, 1m HZ4319''-Rackmount Kit 3RUHZ51Probe 10:1 (150MHz)HZ52Probe 10:1 RF (250MHz)HZ53Probe 100:1 (100MHz)HZ56-2AC/DC Current probeHZ70Opto Interface (with optical fiber cable)HZ100Differential probe 20:1 / 200:1HZ109Differential probe 1:1 / 10:1HZ115Differential probe 100:1 / 1000:1HZ200Probe 10:1 with auto attenuation ID (250MHz)HZ350Probe 10:1 with automatically identification (350MHz)HZ355Slimline probe 10:1 with automatically identification (500MHz)HZO20High voltage probe 1000:1 (400MHz,1000Vrms)HZO30Active probe 1GHz (0,9pF, 1MΩ, including many accessories)HZO50AC/DC Current probe 20A, DC...100kHzHZO51AC/DC Current probe 1000A, DC...20kHz
Specifications
Subject to change without notice
5
Page 6
General information
General information
Please check the instrument for mechanical damage or loose
parts immediately after unpacking. In case of damage we advise
to contact the sender. Do not operate.
B
C
B
T
A
List of symbols used
Consult the manual High voltage
Important note Ground
Positioning the instrument
As can be seen from the fi gures, the handle can be set into different positions:
A and B = carrying
C = horizontal operating
D and E = operating at different angles
F = handle removal
T = shipping (handle unlocked)
Attention!
When changing the handle position, the instru-
ment must be placed so that it can not fall (e.g.
placed on a table). Then the handle locking knobs
must be simultaneously pulled outwards and
rotated to the required position. Without pulling
the locking knobs they will latch in into the next
locking position.
C
D
F
E
D
E
A
PUOPFGkT
PUOPFGkT PUOPFGkT
PUOPFGkT
PUOGkT
PUOPFGkT
PUOPFGkT
HM507
PUOPFGkT
PUOPFGkT
PUOPFGkT PUOPFGkT PUOPFGkT PUOPFGkT
PUOPFGkT
PUOPFGkT PUOPFGkT
PUk PUk PUk PUkPUk PUk
PUkT
HGOPFFD
B
PUOPFGkT
PUOPFGkT
PUkT
PUkT
PUkT
INPUT CHI
OPK
HJ
PUkT
VBN
PUOPFGkT
HJKL
PUOPFGkT
PUkT
PUOPFGkT
HGOFFD
PUkT
PUkT
PUkT
INPUT CHI
INPUT CHI
HAMEG
OPK
OPK
HJ
HJ
VBN
VBN
PUOPFGkT
HJKL
HJKL
T
Handle mounting/dismounting
The handle can be removed by pulling it out further, depending on
the instrument model in position B or F.
Safety
The instrument fulfi ls the VDE 0411 part 1 regulations for
electrical measuring, control and laboratory instruments and
was manufactured and tested accordingly. It left the factory in
perfect safe condition. Hence it also corresponds to European
Standard EN 61010-1 resp. International Standard IEC 1010-1.
In order to maintain this condition and to ensure safe operation
the user is required to observe the warnings and other directions
for use in this manual. Housing, chassis as well as all measuring terminals are connected to safety ground of the mains.
All accessible metal parts were tested against the mains with
200 V
The oscilloscope may only be operated from mains outlets with a
safety ground connector. The plug has to be installed prior to connecting any signals. It is prohibited to separate the safety ground
connection.
Most electron tubes generate X-rays; the ion dose rate of this instrument remains well below the 36 pA /kg permitted by law.
In case safe operation may not be guaranteed do not use the instrument any more and lock it away in a secure place.
. The instrument conforms to safety class I.
DC
T
Safe operation may be endangered if any of the following
was noticed:
– in case of visible damage.
– in case loose parts were noticed
– if it does not function any more.
– after prolonged storage under unfavourable conditions (e.g.
like in the open or in moist atmosphere).
– after any improper transport (e.g. insuffi cient packing not
conforming to the minimum standards of post, rail or transport
company)
Proper operation
Please note: This instrument is only destined for use by personnel
well instructed and familiar with the dangers of electrical measurements.
For safety reasons the oscilloscope may only be operated from
mains outlets with safety ground connector. It is prohibited to
separate the safety ground connection. The plug must be inserted
prior to connecting any signals.
6
Subject to change without notice
Page 7
General information
CAT I
This oscilloscope is destined for measurements in circuits not
connected to the mains or only indirectly. Direct measurements,
i.e. with a galvanic connection to circuits corresponding to the
categories II, III, or IV are prohibited!
The measuring circuits are considered not connected to the mains
if a suitable isolation transformer fulfi lling safety class II is used.
Measurements on the mains are also possible if suitable probes
like current probes are used which fulfi l the safety class II. The
measurement category of such probes must be checked and
observed.
Measurement categories
The measurement categories were derived corresponding to the
distance from the power station and the transients to be expected
hence. Transients are short, very fast voltage or current excursions
which may be periodic or not.
Measurement CAT IV:
Measurements close to the power station, e.g. on electricity
meters
Measurement CAT III:
M e a su r e me n t s in t h e in t e r io r o f b u il d i n gs ( p o w e r d i s t ri b u t i on i n s ta l lations, mains outlets, motors which are permanently installed).
Measurement CAT II:
Measurements in circuits directly connected to the mains (household appliances, power tools etc).
Measurement CAT I:
Electronic instruments and circuits which contain circuit breakers
resp. fuses.
Warranty and repair
HAMEG instruments are subjected to a strict quality control. Prior
to leaving the factory, each instrument is burnt-in for 10 hours.
By intermittent operation during this period almost all defects
are detected. Following the burn-in, each instrument is tested for
function and quality, the specifi cations are checked in all operating
modes; the test gear is calibrated to national standards.
The warranty standards applicable are those of the country in
which the instrument was sold. Reclamations should be directed
to the dealer.
Only valid in EU countries
In order to speed reclamations customers in EU countries may
also contact HAMEG directly. Also, after the warranty expired, the
HAMEG service will be at your disposal for any repairs.
Return material authorization (RMA):
Prior to returning an instrument to HAMEG ask for a RMA number
either by internet (http://www.hameg.com) or fax. If you do not
have an original shipping carton, you may obtain one by calling the
HAMEG service dept (++49 (0) 6182 800 500) or by sending an
email to service@hameg.com.
Maintenance
Clean the outer shell using a dust brush in regular intervals. Dirt can
be removed from housing, handle, all metal and plastic parts using
a cloth moistened with water and 1 % detergent. Greasy dirt may
be removed with benzene (petroleum ether) or alcohol, there after
wipe the surfaces with a dry cloth. Plastic parts should be treated
with an antistatic solution destined for such parts. No fl uid may
enter the instrument. Do not use other cleansing agents as they
may adversely affect the plastic or lacquered surfaces.
Environment of use.
The oscilloscope is destined for operation in industrial, business,
manufacturing, and living sites.
Environmental conditions
Operating ambient temperature: +5 °C to +40 °C. During transport
or storage the temperature may be –20 °C to +70°C.
Please note that after exposure to such temperatures or in case of
condensation proper time must be allowed until the instrument has
reached the permissible temperature, resp. until the condensation
has evaporated before it may be turned on! Ordinarily this will be
the case after 2 hours.
The oscilloscope is destined for use in clean and dry environments.
Do not operate in dusty or chemically aggressive atmosphere or if
there is danger of explosion.
The operating position may be any, however, suffi cient ventilation
must be ensured (convecti on cooling). P rolonged operation requires
the horizontal or inclined position.
Do not obstruct the ventilation holes!
Specifi cations are valid after a 30 minute warm-up period between
15 and 30 degr. C. Specifi cations without tolerances are average
values.
Line voltage
The instrument has a wide range power supply from 105 to 253 V,
50 or 60 Hz ±10%. There is hence no line voltage selector.
The line fuse is accessible on the rear panel and part of the line input
connector. Prior to exchanging a fuse the line cord must be pulled
out. Exchange is only allowed if the fuse holder is undamaged, it
can be taken out using a screwdriver put into the slot. The fuse
can be pushed out of its holder and exchanged.
The holder with the new fuse can then be pushed back in place
against the spring. It is prohibited to ”repair“ blown fuses or to
bridge the fuse. Any damages incurred by such measures will
void the warranty.
Type of fuse:
Size 5 x 20 mm; 250V~, C;
IEC 127, Bl. III; DIN 41 662
(or DIN 41 571, Bl. 3).
Cut off: slow blow (T) 0,8A.
Subject to change without notice
7
Page 8
Type of signal voltage
Type of signal voltage
The oscilloscope HM504-2 allows examination of DC voltages
and most repetitive signals in the frequency range up to at least
50 MHz (–3 dB).
The Y amplifiers have been designed for minimum overshoot and
therefore permit a true signal display.
The display of sinusoidal signals within the bandwidth limits
causes no problems, but an increasing error in measurement due
to gain reduction must be taken into account when measuring
high frequency signals. This error becomes noticeable at approx.
14 MHz. At approx. 30 MHz the reduction is approx. 10% and the
real voltage value is 11% higher. The gain reduction error can not
be defined exactly as the –3 dB bandwidth of the Y amplifiers
differs between 50 MHz and 55 MHz.
When examining square or pulse type waveforms, attention
must be paid to the harmonic content of such signals. The
repetition frequency (fundamental frequency) of the signal must
therefore be significantly smaller than the upper limit frequency
of the Y amplifiers.
Displaying composite signals can be difficult, especially if they
contain no repetitive higher amplitude content that can be used
for triggering. This is the case with bursts, for instance. To obtain
a well triggered display in this case, the assistance of the variable
holdoff function or the delayed time base may be required.
Television video signals are relatively easy to trigger using the
built in TV Sync Separator (TV).
For optional operation as a DC or AC voltage amplifier, each Y
amplifier input is provided with a DC/AC switch. DC coupling should
only be used with a series connected attenuator probe or at very
low frequencies, or if the measurement of the DC voltage content
of the signal is absolutely necessary.
Voltage values of a sine curve
= effective value; Vp = simple peak or crest value;
V
rms
= peak to peak value; V
V
pp
= momentary value.
mom
The minimum signal voltage which must be applied to the Y input
for a trace of 1div height is 1mV
(± 5%) with this deflection
pp
coefficient displayed on the screen (readout) and the vernier
switched off (VAR-LED dark). However, smaller signals than this
may also be displayed. The deflection coefficients are indicated
in mV/div or V/div (peak to peak value).
The magnitude of the applied voltage is ascertained by multiplying
the selected deflection coefficient by the vertical display height
in div. If an attenuator probe x10 is used, a further multiplication
by a factor of 10 is required to ascertain the correct voltage value.
This factor can be entered into the oscilloscope’s memory for
automatic calculation.
For exact amplitude measurements, the variable control (VAR)
must be set to its calibrated detent CAL position.
With the variable control activated the deflection sensitivity can
be reduced up to a ratio of 2.5 to 1 (
readout”
). Therefore any intermediate value is possible within
please note ”controls and
the 1-2-5 sequence of the attenuator(s).
When displaying very low frequency pulses, the flat tops may be
sloping with AC coupling of the Y amplifier (AC limit frequency
approx. 1.6 Hz for 3dB). In this case, DC operation is preferred,
provided the signal voltage is not superimposed on too high a DC
level. Otherwise a capacitor of adequate capacitance must be
connected to the input of the Y amplifier with DC coupling. This
capacitor must have a sufficiently high breakdown voltage rating.
DC coupling is also recommended for the display of logic and
pulse signals, especially if the pulse duty factor changes constantly.
Otherwise the display will move upwards or downwards at each
change. Pure direct voltages can only be measured with DC
coupling.
Amplitude Measurements
In general electrical engineering, alternating voltage data normally
refers to effective values (rms = root mean square value).
However, for signal magnitudes and voltage designations in
oscilloscope measurements, the peak to peak voltage (V
) value
pp
is applied. The latter corresponds to the real potential difference
between the most positive and most negative points of a signal
waveform.
If a sinusoidal waveform, displayed on the oscilloscope screen,
is to be converted into an effective (rms) value, the resulting peak
to peak value must be divided by 2 x √2 = 2.83. Conversely, it
should be observed that sinusoidal voltages indicated in Vrms
) have 2.83 times the potential difference in Vpp. The
(V
eff
relationship between the different voltage magnitudes can be
seen from the following figure.
With direct connection to the Y input, signals up to 400 V
pp
may be displayed (attenuator set to 20 V/div, variable
control to 2.5:1).
With the designations
H = display height in div,
U = signal voltage in V
at the Y input,
pp
D = deflection coefficient in V/div at attenuator switch,
the required value can be calculated from the two given quantities:
However, these three values are not freely selectable. They have
to be within the following limits (trigger threshold, accuracy of
reading):
H between 0.5 and 8div, if possible 3.2 to 8div,
U between 0.5 mV
and 160 Vpp,
pp
D between 1 mV/div and 20 V/div in 1-2-5 sequence.
Examples:
Set deflection coefficient D = 50 mV/div 0.05 V/div,
observed display height H = 4.6 div,
required voltage U = 0.05x4.6 = 0.23 V
pp
.
8
Subject to change without notice
Page 9
Type of signal voltage
Input voltage U = 5 Vpp,
set deflection coefficient D = 1 V/div,
required display height H = 5:1 = 5 div.
Signal voltage U = 230 V
x 2√2 = 651 V
rms
pp
(voltage > 160 Vpp, with probe 10:1: U = 65.1 Vpp),
desired display height H = min. 3.2 div, max. 8 div,
max. deflection coefficient D = 65.1:3.2 = 20.3 V/div,
min. deflection coefficient D = 65.1:8 = 8.1 V/div,
adjusted deflection coefficient D = 10 V/div.
The previous examples are related to the CRT graticule reading.
The results can also be determined with the aid of the DV cursor
measurement (
please note ”controls and readout”
).
The input voltage must not exceed 400 V, irrespective of polarity.
If an AC voltage which is superimposed on a DC voltage is
applied, the maximum peak value of both voltages must not
exceed + or –400 V. So for AC voltages with a mean value of zero
volt the maximum peak to peak value is 800 Vpp.
If attenuator probes with higher limits are used, the probes
limits are valid only if the oscilloscope is set to DC input
coupling.
If DC voltages are applied under AC input coupling conditions the
oscilloscope maximum input voltage value remains 400 V.
The attenuator consists of a resistor in the probe and the
1 MOhm input resistor of the oscilloscope, which is disabled by
the AC input coupling capacity when AC coupling is selected. This
also applies to DC voltages with superimposed AC voltages.
It also must be noted that due to the capacitive reactance of the
AC input coupling capacitor, the attenuation ratio depends on the
signal frequency. For sine wave signals with frequencies higher
than 40 Hz this influence is negligible.
Apart from the above listed exceptions, HAMEG 10:1 probes can
be used for DC measurements up to 600 V or AC voltages (with
a mean value of zero volt) of 1200 V
allows for use up to 1200 V DC or 2400 V
. The 100 :1 probe HZ53
pp
for AC.
pp
It should be noted that its AC peak value is derated at higher
frequencies. If a normal x10 probe is used to measure high
voltages, there is the risk that the compensation trimmer bridging
the attenuator series resistor will break down, causing damage to
the input of the oscilloscope.
Total value of input voltage
The dotted line shows a voltage alternating at zero volt level. If
superimposed on a DC voltage, the addition of the positive peak
and the DC voltage results in the max. voltage (DC + ACpeak).
Time Measurements
As a rule, most signals to be displayed are periodically repeating
processes, also called periods. The number of periods per second
is the repetition frequency. Depending on the time base setting
(TIME/DIV. knob) indicated by the readout, one or several signal
periods or only a part of a period can be displayed. The time
coefficients are stated in ms/div, µs/div or ns/div. The following
examples are related to the CRT graticule reading. The results can
also be determined with the aid of the Dt and 1/Dt cursor
measurement (
please note ”controls and readout”
The duration of a signal period or a part of it is determined by
multiplying the relevant time (horizontal distance in div) by the
(calibrated) time coefficient displayed in the readout.
Uncalibrated, the time base speed can be reduced until a maximum
factor of 2.5 is reached. Therefore any intermediate value is
possible within the 1-2-5 sequence.
With the designations
L = displayed wave length in div of one period,
T = time in seconds for one period,
F = recurrence frequency in Hz of the signal,
Tc = time coefficient in ms, µs or ns/div and the relation
F = 1/T, the following can be stated:
).
However, if for example only the residual ripple of a high voltage
is to be displayed on the oscilloscope, a normal x10 probe is
sufficient. In this case, an appropriate high voltage capacitor
(approx. 22 - 68nF) must be connected in series with the input tip
of the probe.
With Y-POS. control (input coupling to GD) it is possible to use a
horizontal graticule line as reference line for ground potential
before the measurement. It can lie below or above the horizontal
central line according to whether positive and/or negative
deviations from the ground potential are to be measured.
Subject to change without notice
However, these four values are not freely selectable. They
have to be within the following limits:
L between 0.2 and 10 div, if possible 4 to 10 div,
T between 10 ns and 5 s,
F between 0.5 Hz and 100 MHz,
Tc between 100 ns/div and 500 ms/div in 1-2-5 sequence
(with X-MAG. (x10) inactive), and
Tc between 10 ns/div and 50 ms/div in 1-2-5 sequence (with
X-MAG. (x10) active).
9
Page 10
Type of signal voltage
Examples:
Displayed wavelength L = 7 div,
set time coefficient Tc = 100 ns/div,
-9
thus period T = 7 x 100 x 10
thus freq. F = 1/(0.7 x 10
= 0.7 µs
-6
) = 1.428 MHz.
Signal period T = 1s,
set time coefficient Tc = 0.2 s/div,
thus wavelength L = 1/0.2 = 5 div.
Displayed ripple wavelength L = 1 div,
set time coefficient Tc = 10 ms/div,
-3
thus ripple freq. F = 1/(1 x 10 x 10
) = 100 Hz.
TV Line frequency F = 15625 Hz,
set time coefficient Tc = 10 µs/div,
required wavelength L = 1/(15,625 x 10-5) = 6.4 div.
Sine wavelength L = min. 4 div, max. 10 div,
Frequency F = 1 kHz,
max. time coefficient Tc = 1/(4 x 10
min. time coefficient Tc = 1/(10 x 10
3
) = 0.25 ms/div,
3
) = 0.1 ms/div,
set time coefficient Tc = 0.2 ms/div,
required wavelength L = 1/(103 x 0.2 x 10-3) = 5 div.
Displayed wavelength L = 0.8 div,
set time coefficient Tc = 0.5 µs/div,
pressed X-MAG. (x10) button: Tc = 0.05 µs/div,
-6
thus freq. F = 1/(0.8 x 0.05 x 10
thus period T = 1/(25 x 10
6
) = 25 MHz,
) = 40 ns.
If the time is relatively short as compared with the complete
signal period, an expanded time scale should always be applied
(X-MAG. (x10) active). In this case, the time interval of interest
can be shifted to the screen center using the X-POS. control.
Rise Time Measurement
When investigating pulse or square waveforms, the critical
feature is the rise time of the voltage step. To ensure that
transients, ramp-offs, and bandwidth limits do not unduly influence
the measuring accuracy, the rise time is generally measured
between 10% and 90% of the vertical pulse height. For
measurement, adjust the Y deflection coefficient using its variable function (uncalibrated) together with the Y-POS. control so that
the pulse height is precisely aligned with the 0% and 100% lines
of the internal graticule. The 10% and 90% points of the signal
will now coincide with the 10% and 90% graticule lines. The
risetime is given by the product of the horizontal distance in div
between these two coincident points and the calibrated time
coefficient setting. The fall time of a pulse can also be measured
by using this method.
The following figure shows correct positioning of the oscilloscope
trace for accurate rise time measurement.
With a time coefficient of 10 ns/div (X x10 magnification active),
the example shown in the above figure results in a total measured
risetime of
t
= 1.6 div x 10 ns/div = 16 ns
tot
When very fast risetimes are being measured, the risetimes of
the oscilloscope amplifier and of the attenuator probe have to be
deducted from the measured time value. The risetime of the
signal can be calculated using the following formula.
2
2
= √ t
t
r
In this t
– t
tot
osc
is the total measured risetime, t
tot
oscilloscope amplifier (approx. 7 ns), and t
probe (e.g. = 2 ns). If t
2
– t
p
is the risetime of the
osc
p
is greater than 100 ns, then t
tot
the risetime of the
can be
tot
taken as the risetime of the pulse, and calculation is unnecessary.
Calculation of the example in the figure above results in a signal
risetime:
= √162 – 72 – 22 = 14.25 ns
t
r
The measurement of the rise or fall time is not limited to the trace
dimensions shown in the above diagram. It is only particularly
simple in this way. In principle it is possible to measure in any
display position and at any signal amplitude. It is only important
that the full height of the signal edge of interest is visible in its full
length at not too great steepness and that the horizontal distance
at 10% and 90% of the amplitude is measured. If the edge shows
rounding or overshooting, the 100% should not be related to the
peak values but to the mean pulse heights. Breaks or peaks
(glitches) next to the edge are also not taken into account. With
very severe transient distortions, the rise and fall time
measurement has little meaning. For amplifiers with approximately
constant group delay (therefore good pulse transmission
performance) the following numerical relationship between rise
time tr (in ns) and bandwidth B (in MHz) applies:
Connection of Test Signal
In most cases, briefly depressing the AUTOSET causes a useful
signal related instrument setting. The following explanations
refer to special applications and/or signals, demanding a manual
instrument setting.
in the section ”controls and readout”.
Caution:
When connecting unknown signals to the oscilloscope input,
always use automatic triggering and set the input coupling switch to AC. The attenuator should initially be set to
20 V/div.
The description of the controls is explained
10
Sometimes the trace will disappear after an input signal has been
applied. Then a higher deflection coefficient (lower input sensitivity)
must be chosen until the vertical signal height is only 3 – 8 div.
With a signal amplitude greater than 160 V
and the deflection
pp
coefficient (VOLTS/DIV.) in calibrated condition, an attenuator
probe must be inserted before the Y input. If, after applying the
signal, the trace is nearly blanked, the period of the signal is
probably substantially longer than the set time deflection
coefficient (TIME/DIV.). It should be switched to an adequately
larger time coefficient.
Subject to change without notice
Page 11
Controls and Readout
The signal to be displayed can be connected directly to the Y input
of the oscilloscope with a shielded test cable such as HZ32 orHZ34, or reduced through a x10 or x100 attenuator probe. The
use of test cables with high impedance circuits is only
recommended for relatively low frequencies (up to approx. 50 kHz).
For higher frequencies, the signal source must be of low
impedance, i.e. matched to the characteristic resistance of the
cable (as a rule 50 Ohm). Especially when transmitting square and
pulse signals, a resistor equal to the characteristic impedance of
the cable must also be connected across the cable directly at the
Y-input of the oscilloscope. When using a 50 Ohm cable such as
the HZ34, a 50 Ohm through termination type HZ22 is available
from HAMEG. When transmitting square signals with short rise
times, transient phenomena on the edges and top of the signal
may become visible if the correct termination is not used. A
terminating resistance is sometimes recommended with sine
signals as well. Certain amplifiers, generators or their attenuators
maintain the nominal output voltage independent of frequency
only if their connection cable is terminated with the prescribed
resistance. Here it must be noted that the terminating resistor
HZ22 will only dissipate a maximum of 2 Watts. This power is
reached with 10 V
attenuator probe is used, no termination is necessary. In this
case, the connecting cable is matched directly to the high
impedance input of the oscilloscope. When using attenuator
probes, even high internal impedance sources are only slightly
loaded (approx. 10 MOhm II 12 pF or 100 MOhm II 5 pF with
HZ53). Therefore, if the voltage loss due to the attenuation of the
probe can be compensated by a higher amplitude setting, the
probe should always be used. The series impedance of the probe
provides a certain amount of protection for the input of the Y
amplifier. Because of their separate manufacture, all attenuator
probes are only partially compensated, therefore accurate
compensation must be performed on the oscilloscope
compensation )
Standard attenuator probes on the oscilloscope normally reduce
its bandwidth and increase the rise time. In all cases where the
oscilloscope bandwidth must be fully utilized (e.g. for pulses with
steep edges) we strongly advise using the probes HZ51 (x10)
HZ52 (x10 HF) and HZ54 (x1 and x10). This can save the purchase
of an oscilloscope with larger bandwidth.
The probes mentioned have an HF-adjustment in addition to low
frequency calibration adjustment. Thus a group delay correction
to the upper limit frequency of the oscilloscope is possible with
the aid of a 1 MHz calibrator, e.g. HZ60.
In fact the bandwidth and rise time of the oscilloscope are not
noticeably changed with these probe types and the waveform
reproduction fidelity can even be improved because the probe
can be matched to the oscilloscope’s individual pulse response.
If a x10 or x100 attenuator probe is used, DC input coupling
must always be used at voltages above 400 V. With AC
coupling of low frequency signals, the attenuation is no
longer independent of frequency, pulse tops can show pulse
tilts. Direct voltages are suppressed but charge the
oscilloscope input coupling capacitor concerned. Its voltage
rating is max. 400 V (DC + peak AC). DC input coupling is
therefore of quite special importance with a x100 attenuation probe which usually has a voltage rating of max.
1200 V (DC + peak AC). A capacitor of corresponding
capacitance and voltage rating may be connected in series
with the attenuator probe input for blocking DC voltage
(e.g. for hum voltage measurement).
(28.3 Vpp) with sine signal. If a x10 or x100
rms
.
(see Probe
With all attenuator probes, the maximum AC input voltage must
be derated with frequency, usually above 20 kHz. Therefore the
derating curve of the attenuator probe type concerned must be
taken into account.
The selection of the ground point on the test object is important
when displaying small signal voltages. It should always be as
close as possible to the measuring point. If this is not done,
serious signal distortion may result from spurious currents through
the ground leads or chassis parts. The ground leads on attenuator
probes are also particularly critical. They should be as short and
thick as possible. When the attenuator probe is connected to a
BNC-socket, a BNC adapter should be used. In this way ground
and matching problems are eliminated. Hum or interference
appearing in the measuring circuit (especially when a small
deflection coefficient is used) is possibly caused by multiple
grounding because equalizing currents can flow in the shielding
of the test cables (voltage drop between the protective conductor
connections, caused by external equipment connected to the
mains/line, e.g. signal generators with interference protection
capacitors).
Controls and Readout
A: Basic settings
The following description assumes that:
1. “Component Tester” is switched off.
2. The following settings are present under MAIN MENU
> SETUP & INFO > MISCELLANEOUS:
2.1 CONTROL BEEP and ERROR BEEP activated (x),
2.2 QUICK START not activated.
3. The screen Readout is visible.
The LED indicators on the large front panel facilitate operation
and provide additional information. Electrical end positions of
controls are indicated by acoustic signal (beep).
All controls, except the power switch (POWER), are electronically
set and interrogated. Thus, all electronically set functions and
their current settings can be stored and also remotely controlled.
B: Menu Display and Operation
Operation of some pushbuttons activates the display of menus.
There are Standard and Pulldown Menus.
Standard menus:
When a standard menu is displayed, all other readout information
(e.g. parameter settings) are switched off. The readout then
consists of the menu headline, and the respective menu functions.
At the bottom of the graticule are displayed symbols and
commands which can be operated by the pushbuttons related to
them below.
“Esc” CT pushbutton [37] switches one step back in the menu
hierarchy.
“Exit” SELECT – ON/OFF pushbutton [34] closes the menu and
switches back to the operating conditions present before calling
the menu.
The pushbuttons underneath the triangle symbols pointing
upwards UNIT CAL.SEL. [35] and downwards SOURCE GLUE
[33] enable you to select one item which becomes highlighted.
Subject to change without notice
11
Page 12
Controls and Readout
A
FO
O
M
“SET” MAIN MENU-pushbutton [31] calls the selected menu
item, starts a function or switches a function on/off.
Pulldown menus:
After pressing a pushbutton which calls a Pulldown menu, the
instrument parameter settings are still displayed. The readout
only changes in respect to the called parameter (e.g. input
coupling) and now shows all selectable parameter options (in
case of input coupling: AC, DC and GND). The previously displayed
parameter doesn‘t change but is displayed highlighted. Each time
the pushbutton is briefly pressed the next parameter becomes
active and highlighted, as long as the Pulldown menu is displayed.
Without further pressing the pushbutton, the Pulldown menu
extinguishes after a few seconds and the selected parameter is
displayed in the normal way.
C: READOUT Information
The readout alphanumerically displays the scope parameter
settings, measurement results and CURSOR lines. Which
information is displayed depends on the actual instrument settings.
The following list contains the most important display information.
Top of the graticule from left to right:
1st time deflection coefficient
2nd trigger source, slope and coupling
3rd operating condition of delay time base
4th measuring results
Bottom of the graticule from left to right:
1st probe symbol (x10), Y deflection coefficient and input
coupling channel I
2nd “+” symbol (addition)
3rd probe symbol (x10), Y deflection coefficient and input
coupling channel II
4th channel mode
The trigger point symbol is displayed at the left graticule border
line. The CURSOR lines can take any position within the graticule.
[2] AUTOSET
Briefly pressing this pushbutton results in an automatic
instrument setting selecting Yt mode as the default. The
instrument is set to the last used Yt mode setting (CH I, CH II
or DUAL).
The instrument is set automatically to normal (undelayed)
time base mode, even if the previous Yt mode was present
in combination with search (“sea”), delay (“del”) or triggered
delay (“dTr”) time base mode.
Please also note ”AUTO-
SET” in section “First Time Operation”.
Automatic CURSOR positioning:
If CURSOR lines are displayed and AUTOSET is chosen the
CURSOR lines are set automatically under suitable conditions
and the readout briefly displays “SETTING CURSOR”.
If the signal height is insufficient, the CURSOR lines do not
change. In DUAL mode the CURSOR lines are related to the
signal which is used for internal triggering.
Voltage CURSOR
If voltage measurement is present, the CURSOR lines are
automatically set to the positive and negative peak value of
the signal. The accuracy of this function decreases with
higher frequencies and is also influenced by the signal‘s
pulse duty factor.
Time/Frequency CURSOR
If complex waveforms such as video signals are applied, the
cursor lines may not align exactly with one period and give a
false reading.
[3] INTENS/FOCUS – Knob with associated LEDs and TRACE
ROT.-pushbutton.
If the readout (RO) is not switched off, briefly pressing the
READOUT pushbutton switches over the INTENS/FOCUS
knob function indicated by a LED in the sequence A, FOC,
RO, A. In condition READOUT deactivated, the switching
sequence is A, FOC, A.
Description of Controls
The large front panel is, as usual with Hameg oscilloscopes,
marked with several fields.
The following controls and LED indicators are located on the top,
to the right of the screen, above the horizontal line.
1345
2
POWER
AUTOSET
[1] POWER – Pushbutton and symbols for ON (I) and OFF (O).
INTENS / FOCUS
!
C
R
TRACE
ROT.
Instruments
50 MHz
R
ANALOG OSCILLOSCOPE
HM504-2
RECALL
SAVE
After the oscilloscope is switched on, all LEDs are lit and an
automated instrument test is performed. During this time the
HAMEG logo and the software version are displayed on the
screen. After the internal test is completed successfully, the
overlay is switched off and the normal operation mode is
present. Then the last used settings become activated and
LED [3] indicates the ON condition.
“A”:
The INTENS/FOCUS control knob adjusts the signal(s)
intensity. Turning this knob clockwise increases the intensity.
Only the minimum required trace intensity should be used,
depending on signal parameters, oscilloscope settings and
light conditions.
“FOC”:
The INTENS/FOCUS control knob adjusts both the trace and
the readout sharpness. Note: The electron beam diameter
gets larger with a higher trace intensity and the trace sharpness
decreases. This can be corrected to a certain extent. Assuming
that the trace sharpness was set to optimum in the screen
centre, it is unavoidable that the trace sharpness decreases
with an increasing distance from the centre.
Since the settings of the signal(s) intensity (A) and the
READOUT (RO) are usually different, the FOCUS should be
set for optimum signal(s) sharpness. The sharpness of the
READOUT then can be improved by reducing the READOUT
intensity.
“RO”:
The INTENS/FOCUS control knob adjusts the READOUT
intensity. Turning this knob clockwise increases and counter
clockwise decreases the intensity. Only the minimum required
intensity should be used.
12
Subject to change without notice
Page 13
Controls and Readout
V
R
V
NM
HO
V
A
FO
O
M
TRACE ROT. (trace rotation) is selected by pressing and
holding the pushbutton. This causes the display “Trace Rot.
with Int.” (trace rotation by use of INTENS/FOCUS knob) and
allows you to compensate the influence of the Earth’s
magnetic field on the trace deflection. To avoid misadjustment
due to unavoidable deflection non linearities, set the trace to
the graticule centre (Y-POS/CURS.I and X-POS. control).
After the trace has been set parallel to the horizontal graticule
line using INTENS/FOCUS, press “SAVE” to store the last
setting.
Further information can be found in section “First Time
Operation” under “Trace Rotation”.
1345
2
POWER
AUTOSET
INTENS / FOCUS
!
C
R
TRACE
ROT.
Instruments
50 MHz
R
ANALOG OSCILLOSCOPE
HM504-2
RECALL
SAVE
[4] RM
The remote control mode can be switched on or off (”RM”
LED dark) via the RS-232 interface. When the ”RM” LED is
lit, all electronically selectable controls on the front panel are
inactive. This state can be cancelled by depressing the
AUTO SET pushbutton provided it was not deactivated via
the interface.
[5] SAVE / RECALL – Pushbutton for instrument settings
The instrument contains 9 non volatile memories. These
can be used by the operator to save instrument settings and
to recall them.
SAVE:
Press and hold the RECALL/SAVE button to start a storage
process. This causes the SAVE menu (Standard menu, note
“B: Menu-Display and Operation”) to be displayed. Choose
the memory location cipher (highlighted) by pressing a
pushbutton underneath the triangle symbols. Briefly press
the pushbutton underneath “SET” to store the last instrument
setting and return from menu display to previous mode. If the
SAVE function was called inadvertently, it can be switched
off with “Esc”.
Switching the instrument off automatically stores the current
settings in memory location 9 (PWR OFF = Power Off), with
the effect that different settings previously stored in this
location get lost. To prevent this, RECALL 9 before switching
the instrument off.
RECALL:
Briefly pressing calls the RECALL menu. You can select the
required memory location using a “triangle” pushbutton.
Recall the previously stored instrument settings by briefly
pressing the “SET” pushbutton or briefly press “Esc” if the
function was called inadvertently.
Attention:
When an instrument setting is recalled, the current
signal may not be optimally displayed unless similar
(frequency, amplitude) to that used when the setting
was stored.
The setting controls and LED’s for the Y amplifiers, modes,
triggering and time base are located underneath the sector of the
front panel described above.
6
Y-POS / CURS.I
VOLTS / DIV.
AR
20V 1mV20V 1mV0.5s 50ns
CH ICH IIDEL.MODE Z-ON/OFF
VAR .
14
15
79811131210
Y-POS / CURS.IILEVELX-POS.
TRIG.
SOURCE
DUAL
MENU
CU
CURSOR
POS
16
VOLTS / DI V.
INV.
17
NM
AT
AR
MENU
TRIG. MODE
DEL.POS.
VAR .
19
20
18
PUSH
BOTH
PUSH
LONG
TIME / DIV.
ON
OFF
21
23
22
x10
X-
MAG.
AR
VAR .
24
[6] Y-POS/CURS. I – Control knob with two functions.
This knob allows position control of channel I trace or CURSOR line(s). Briefly pressing the CURSOR POS pushbutton
[7] selects the function. If the CURSOR line(s) are not
displayed the CURS. I function is not selectable.
Y-POS:
The vertical trace position of channel I can be set with this
control knob, if the CURSOR POS LED isn’t lit. In addition
(“add”) mode both Y-POS/CURS. I [6] and Y-POS/CURS. II[8] control knobs are active. If the instrument is set to XY
mode this control knob is inactive and the X-POS. [12] knob
must be used for horizontal positioning.
DC voltage measurement:
If no signal is applied at the INPUT CH I [25], the vertical trace
position represents 0 Volt. This is the case if INPUT CH I [25]
or in addition (ADD) mode, both INPUT CH I [25] and INPUTCH II [28], are set to GND (ground) [26; 29] and automatic
triggering AT [9] is present to make the trace visible.
The trace can then be set to the vertical position best suited
for the following DC voltage measurement. After switching
GND (ground) off and selecting DC input coupling, a DC
signal applied at the input changes the trace position in
vertical direction. The DC voltage then can be determined by
taking the deflection coefficient, the probe factor and the
trace position change with respect to the previous 0 Volt
position into account.
”0-Volt”-Symbol:
The READOUT indicates the “0-Volt” trace position of channel
⊥⊥
I by a ”
⊥” symbol to the left of the screen‘s vertical centre line
⊥⊥
in CHI and DUAL mode. When Y position is used, this symbol
changes to an “arrow” symbol pointing outside the graticule
just before the trace goes outside the graticule limits.
If addition mode (“add”) is present just one ”⊥” symbol is
visible. In XY mode the “0-Volt” trace position for channel I (X)
and channel II (Y) is symbolised by “triangle” symbols at the
right graticule border (Y) and above the Y deflection coefficient
display. The “triangle” symbol(s) point(s) outside the graticule
when the “0-Volt” trace position is outside the graticule.
CURS.I:
The CURSOR lines marked by the symbol “I” can be shifted
by the Y-POS/CURS. I control knob, if the CURSOR POS
LED [7] lit.
Subject to change without notice
13
Page 14
Controls and Readout
V
R
V
NM
HO
V
[7] CURSOR POS – Pushbutton and LED.
Briefly pressing this pushbutton determines the function of
the Y-POS/CURS.I [6] and Y-POS/CURS.II [8] controls.
If the LED is not lit the Y position control function is active.
Provided that the CURSOR lines are activated, the LED can
be switched on by briefly pressing the CURSOR POSpushbutton. Then the controls [6] and [8] are switched over
from Y position to CURSOR position control CURS.I [6] and
CURS.II [8]. Briefly pressing this pushbutton once again
switches back to the Y position control function.
6
Y-POS / CURS.I
VOLTS / DIV.
AR
20V 1mV20V 1mV0.5s 50ns
CH ICH IIDEL.MODE Z-ON/OFF
VAR .
14
15
79811131210
Y-POS / CURS.IILEVELX-POS.
TRIG.
SOURCE
DUAL
MENU
CU
CURSOR
POS
16
VOLTS / DIV.
INV.
17
18
NM
AT
AR
MENU
TRIG. MODE
DEL.POS.
VAR .
19
20
PUSH
BOTH
PUSH
LONG
TIME / DIV.
AR
ON
OFF
21
23
22
[8] Y-POS/CURS. II – Control knob with two functions.
This knob enables position control of channel II trace or
CURSOR line(s). Briefly pressing the CURSOR POSpushbutton [7] selects the function. If the CURSOR line(s)
are not displayed the CURS. I function is not selectable.
Y-POS:
The vertical trace position of channel II can be set with this
control knob, if the CURSOR POS LED isn’t lit. In addition
(“add”) mode both, Y-POS/CURS. I [6] and Y-POS/CURS. II[8] control knobs are active. If the instrument is set to XY
mode, this control knob is inactive and the X-POS.-knob [12]
must be used for horizontal positioning.
x10
X-
MAG.
If addition mode (“add”) is present just one ”^” symbol is
visible.
In XY mode the “0 Volt” trace position for channel I (X) and
channel II (Y) is symbolised by “triangle” symbols at the right
graticule border (Y) and above the Y deflection coefficient
display. The “triangle” symbol (s) point (s) outside the graticule
when the “0-Volt” trace position is outside the graticule.
CURS. II:
The CURSOR lines marked by the symbol “II” can be shifted
by the Y-POS/CURS. II control knob, if the CURSOR POS
LED [7] lit.
[9] NM AT – Pushbutton with a double function and asso-
ciated NM LED.
NM / AT selection:
Press and hold the pushbutton to switch over from automatic
(peak value) to normal triggering (NM LED above the
pushbutton lit) and vice versa. If the LED is dark, automatic
or automatic peak value triggering is selected.
VAR .
AT:
Automatic triggering can be carried out with or without peak
capture. In both cases the LEVEL control [11] is effective and
24
the trace is visible even if no signal is applied or trigger
settings are unsuitable. Signal frequencies below the
automatic trigger frequency can not be triggered as the
automatic trigger cycle starts to early for such signals.
In the automatic peak value triggering condition the LEVEL
control [11] range is limited to the trigger signal positive and
negative peak values. Automatic triggering without peak
value detection enables the trigger point to be set outside the
signal amplitude range. In the latter case, although untriggered,
there is still a signal display.
Whether the peak value detection is active or not depends on
the operating mode and the selected trigger coupling. The
actual state is recognised by the behaviour of the trigger
point symbol when changing the LEVEL setting.
DC voltage measurement:
If no signal is applied at the INPUT CH II [28], the vertical
trace position represents 0 Volt. This is the case if INPUT CH
II [28] or in addition (ADD) mode, both, INPUT CH I [25] and
INPUT CH II [28], are set to GND (ground) [26; 29] and
automatic triggering AT [9] is present to make the trace
visible.
The trace can then be set to the vertical position best suited
for the following DC voltage measurement. After switching
GND (ground) off and selecting DC input coupling, a DC
signal applied at the input changes the trace position in
vertical direction. The DC voltage then can be determined by
taking the deflection coefficient,
the probe factor and the
trace position change with respect to the previous 0 Volt
position into account.
”0-Volt”-Symbol:
The READOUT indicates the “0-Volt” trace position of channel
II by a ”^” symbol to the right of the screen‘s vertical centre
line in CH II and DUAL mode. When Y-position is used, this
symbol changes to an “arrow” symbol pointing outside the
graticule just before the trace goes outside the graticule
limits.
14
NM:
Normal triggering disables both the automatic trigger and the
peak value detection so even low frequency signals can be
displayed in a stable manner. Without suitable input signal
height, trigger coupling and LEVEL settings, no trace will be
displayed.
The last LEVEL setting of the time base is stored, then the
control again becomes active when selecting triggering after
delay DEL.MODE (“dTr”) time base mode (quasi 2
base). In combination with In “dTr” mode the LEVEL control
is operative for the “2nd time base”.
/ \(Slope selection):
Each time this pushbutton is briefly pressed, the slope
direction switches from falling edge to rising edge and vice
versa. The current setting is displayed in the readout by a
slope symbol. The last setting in undelayed time base mode
is stored and still active if triggered delay (“dTr”) time base
mode is selected. This allows for a different slope setting for
the triggered DELAY (DTR) time base mode.
[10] TR – Trigger indicator LED.
The TR LED is lit in Yt mode if the triggering conditions are
met for the first trigger unit used in undelayed time base
Subject to change without notice
nd
time
Page 15
Controls and Readout
mode. Whether the LED flashes or is lit constantly depends
on the frequency of the trigger signal.
[11] LEVEL – Control knob.
Turning the LEVEL knob causes a different trigger point
setting (voltage). The trigger unit starts the time base when
the edge of a trigger signal crosses the trigger point. In most
Yt modes the trigger point is displayed in the readout by the
symbol on the left vertical graticule line. If the trigger point
symbol would overwrite other readout information or would
be invisible when being set above or below the screen, the
symbol changes and an arrow indicates in which vertical
direction the trigger point has left the screen.
The trigger point symbol is automatically switched off in
those modes where there is no direct relation between the
trigger signal and the displayed signal. The last setting in
undelayed time base mode is stored and still active if triggered
delay (“dTr”) time base mode is selected. This allows for a
different level setting for the triggered delay (“dTr”) time
base mode.
[12] X-POS. – Control knob.
This control knob enables an X position shift of the signal(s)
in Yt and XY mode. In combination with X magnification x10
(Yt mode) this function makes it possible to shift any part of
the signal on the screen.
[13] X-MAG. x10 – Pushbutton and LED.
Each time this pushbutton is pressed the x10 LED located
above is switched on or off. If the x10 LED is lit, the signal
display is expanded 10 fold in all time deflection settings >
50ns/div. At 50 ns/div only the expansion is 5 fold and yields
10ns/div. As the X expansion results in a higher time base
speed (lower time deflection coefficient), all time and
frequency relevant information in the readout is switched
over.
After activating X MAG. x10, the visible part of the signal is
that which was previously at the graticule centre. The
interesting part of the signal can be made visible with aid of
the X-POS. [12] control.
This pushbutton is not operative in XY mode.
[15] CH I VAR. – Pushbutton with two functions.
Pressing and holding this pushbutton selects the VOLTS/
DIV. [14] control knob function between attenuator and
vernier (variable). The current setting is displayed by the
VAR-LED located above the knob.
CH I mode:
Briefly pressing the CHI button sets the instrument to channel
I (Mono CH I) mode. The deflection coefficient displayed in
the readout indicates the current conditions (”Y1...”). If
neither external nor line (mains) triggering was active, the
internal trigger source automatically switches over to channel
I and the READOUT displays “Y1, trigger slope, trigger
coupling”. The last function setting of the VOLTS/DIV [14]
knob remains unchanged.
All channel I related controls are active if the input [25] is not
set to GND [26].
VAR.:
After switching the VAR-LED [14] on, the deflection coefficient
is still calibrated. Turning the VOLTS/DIV. [14] control knob
counter clockwise reduces the signal height and the deflection
coefficient becomes uncalibrated.
The readout then displays e.g. ”Y1>...” indicating the
uncalibrated condition instead of ”Y1:...”. Pressing and holding
the CHI pushbutton again switches the LED off, sets the
deflection coefficient into calibrated condition and activates
the attenuator function. The previous vernier setting will not
be stored.
[16] DUAL MENU – Pushbutton with multiple functions.
Switchover on DUAL (two channel), ADDITION and XY
operation:
Briefly pressing selects DUAL mode if channel I (mono) or
channel II (mono) mode had been present before. Then the
deflection coefficients of both channels and the channel
switchover mode (alt or chp) become visible on the READOUT.
The last trigger conditions (source, slope and coupling) remain
unchanged, but can be changed.
Pressing and holding the DUAL pushbutton switches directly
to XY mode if channel I (mono) or channel II (mono) mode had
been present before. On condition XY mode pressing the
pushbutton switches back to DUAL mode.
[14] VOLTS/DIV. – Control knob for channel I with a double
function.
This control is active only if channel I is enabled and it‘s input
coupling (AC or DC) is activated. Channel I is active in CH I
(Mono), DUAL, Addition (“add”) and XY mode. The knob is
automatically disabled if the channel related to it is switched
off, or if the input coupling is set to GND (ground).
Y-deflection coefficient setting (input attenuator): This function
is available if the VAR. LED is dark.
Turning the control knob clockwise increases the sensitivity
(decreases the deflection coefficient) in a 1-2-5 sequence
and decreases the sensitivity (increases the deflection
coefficient) if turned in the opposite direction (ccw.). The
available range is from 1mV/div up to 20V/div.
The deflection coefficients and additional information
regarding the active channel(s) are displayed in the readout,
e.g. ”Y1: deflection coefficient, input coupling”. The ”:”
symbolizes calibrated measuring conditions and is replaced
by the ”>” symbol in uncalibrated conditions.
VAR.: The vernier (variable) function is described under item
VAR [15].
Subject to change without notice
Choosing the channel switch over or sub menu:
Once DUAL mode is active, briefly pressing the (Dual)
pushbutton opens a Pulldown menu with the current
mode displayed highlighted. The menu offers “chp” (chopped
DUAL), “alt” (alternate DUAL), “add” (Addition) and “XY”
mode.
As long as the Pulldown menu is displayed, briefly pressing
the pushbutton selects the next mode and highlights the
actual setting. Please note “B: Menu Display and Operation“.
If “add” (Addition) mode is activated, briefly pressing the
pushbutton switches over to DUAL mode, without displaying
the Pulldown menu.
All channel related controls are effective as long as the input
coupling is not set to GND [26, 29].
DUAL mode:
On the right of the channel II (Y2:...) deflection coefficient the
READOUT displays the channel switch over mode. “alt”
indicates alternate and „chp“, chopped switch over. The
channel switch over is automatically selected by the time
15
Page 16
Controls and Readout
V
R
V
NM
HO
V
base setting, but can be changed in the pulldown menu. The
oscilloscope automatically determines the channel switching
mode after a change of the time base setting.
“chp” (Chopped):
Indicates chopped mode, whereby the channel switching
occurs constantly between channel I and II during each
sweep. This channel switching mode occurs when any time
base setting between 500 ms/div and 500 µs/div has been
chosen.
“alt” (Alternate):
Indicates alternate channel switching. After each time base
sweep the instrument internally switches over from channel
I to channel II and vice versa. This channel switching mode
is automatically selected if any time coefficient from 200 µs/
div to 50 ns/div is active.
“add” (Addition):
The readout indicates this mode by a ”+” sign located
between both channel deflection coefficients. In addition
mode, two signals (channel I and II) are displayed as one
signal. The Y position of the signal can be influenced by both
Y-POS/CURS.I [6] and Y-POS/CURS.II [8] controls. For
correct measurements the deflection coefficients for both
channels must be equal. While the trigger mode is not
affected, the trigger point symbol is switched off.
Whether the algebraic sum (addition) or the difference
(subtraction) of both input signals is displayed, depends on
the phase relationship and the INV (invert function) setting.
XY mode:
In XY mode the deflection coefficients are displayed as
”X...” for channel I and ”Y...” for channel II, followed by
”XY”. The following READOUT information is switched off:
time deflection coefficient, trigger source, slope, coupling
and trigger point symbol.
In addition to all trigger and time base related controls, the Y-POS/CURS.I-knob [7] is deactivated. For X-position alteration,
the X-POS.-knob [12] can be used.
The front panel shows that the DUAL MENU push button [16]
can also be depressed together with the CH II pushbutton
[19]. Please note item [19].
the signal used for triggering originates. The measuring
amplifiers (internal triggering) or the BNC socket which
serves as an input for externally applied signals (external
triggering) can be used as a trigger source.
Single channel operation (CH I or CH II):
Briefly pressing switches the trigger source over without
displaying the Pulldown menu. During single channel operation
the internal trigger signal (originating from channel I or
channel II) or the external trigger signal can be chosen.
DUAL and Addition mode:
Briefly pressing opens the trigger source Pulldown menu
with the actual setting highlighted. Please note “B: Menu
Display and Operation”.
The following listing shows the possible trigger sources and
how they are indicated by the READOUT. Their availability
depends on the actual channel operation mode.
„Y1“: The measurement amplifier of channel I serves as
Pulldown trigger source.
„Y2“: The measurement amplifier of channel II serves as
trigger source.
„alt“: Alternate triggering can be chosen if DUAL mode is
present. In alternate trigger mode, the switch over of the
internal trigger sources “Y1” and “Y2” is carried out
synchronously with the alternate channel switching and the
trigger point symbol is switched off.
As alternate triggering requires alternate channel operation,
alternate channel switching is set automatically. A change of
the time coefficient then has no affect regarding the channel
switching mode.
The following trigger coupling settings can not be chosen in
combination with alternate triggering: TVL, TVF and line
(mains).
If “add” (addition) or delayed time base mode (“sea”, “del”
or “dTr”) is present, alternate triggering is not available.
Therefore alternate triggering is automatically switched off if
one of these modes has been chosen.
6
Y-POS / CURS.I
VOLTS / DIV.
AR
20V 1mV20V 1mV0.5s 50ns
CH ICH IIDEL.MODE Z-ON/OFF
VAR .
14
15
TRIG.
SOURCE
DUAL
MENU
CU
CURSOR
POS
16
[17] TRIG. SOURCE – Pushbutton.
This pushbutton is for trigger source selection and deactivated
if line (mains) triggering is selected or XY operation is chosen.
The term “trigger source” describes the source from which
16
79811131210
Y-POS / CURS.IILEVELX-POS.
NM
AT
VOLTS / DIV.
AR
MENU
TRIG. MODE
INV.
VAR .
19
17
18
20
DEL.POS.
PUSH
BOTH
PUSH
LONG
ON
OFF
21
23
22
TIME / DIV.
AR
x10
X-
MAG.
“ext”: External trigger mode is available in all time base and
trigger coupling modes except line/mains triggering. Then
the TRIG.EXT. BNC-socket [30] serves as the external trigger
signal input. On external triggering mode, the intensity
modulation (Z), which might have been present before, is
automatically switched off.
[18] VOLTS/DIV. – Control knob for channel II with a double
function.
This control is active only if channel II is enabled and it‘s input
coupling (AC or DC) is activated. Channel II is active in CH II
(Mono), DUAL, Addition (“add”) and XY mode. The knob is
VAR .
automatically disabled if the channel related to it is switched
off, or if the input coupling is set to GND (ground).
This function is available if the VAR. LED is dark.
Turning the control knob clockwise increases the sensitivity
(decreases the deflection coefficient) in a 1-2-5-sequence
and decreases the sensitivity (increases the deflection
Subject to change without notice
Page 17
Controls and Readout
V
R
V
NM
HO
V
coefficient) if turned in the opposite direction (ccw.). The
available range is from 1mV/div up to 20V/div.
The deflection coefficients and additional information
regarding the active channel(s) are displayed in the readout,
e.g. ”Y2: deflection coefficient, input coupling”. The ”:”
symbolizes calibrated measuring conditions and is replaced
by the ”>” symbol in uncalibrated conditions.
VAR.: The vernier (variable) function is described under item
VAR [19].
[19] CH II - VAR. – Pushbutton with several functions.
Pressing and holding this pushbutton selects the VOLTS/DIV. [18] control knob function between attenuator and
vernier (variable). The current setting is displayed by the
VAR-LED located above the knob.
CH II mode:
Briefly pressing the CH II-button sets the instrument to
channel II (Mono CH II) mode. The deflection coefficient
displayed in the readout indicates the current conditions
(”Y2...”). If neither external nor line (mains) triggering was
active, the internal trigger source automatically switches
over to channel II and the READOUT displays “Y2, trigger
slope, trigger coupling”. The last function setting of the
VOLTS/DIV-knob [18] remains unchanged.
All channel II related controls are active if the input [28] is not
set to GND [29].
VAR.:
After switching the VAR-LED [18] on, the deflection coefficient
is still calibrated. Turning the VOLTS/DIV.-control knob [18]
counter clockwise reduces the signal height and the deflection
coefficient becomes uncalibrated.
6
Y-POS / CURS.I
VOLTS / DIV.
AR
20V 1mV20V 1mV0.5s 50ns
CH ICH IIDEL.MODE Z-ON/OFF
VAR .
14
15
79811131210
Y-POS / CURS.IILEVELX-POS.
TRIG.
SOURCE
DUAL
MENU
CU
CURSOR
POS
16
VOLTS / DI V.
INV.
17
18
NM
AT
AR
MENU
TRIG. MODE
DEL.POS.
VAR .
19
20
PUSH
BOTH
PUSH
LONG
TIME / DIV.
ON
OFF
21
23
22
x10
X-
MAG.
AR
VAR .
24
TVL: TV signal, line pulse triggering,
trigger point symbol switched off,
TVF: TV signal, frame pulse triggering,
trigger point symbol switched off.
~:line/mains triggering,
trigger point symbol switched off.
Line/mains triggering inactivates the TRIG. SOURCE-pushbutton [17].
In some trigger modes such as alternate triggering, some
trigger coupling modes are automatically disabled and can
not be selected.
[21] DEL.POS. HO LED – Control knob with double function and
related HO LED.
The DEL.POS. -knob has two functions which depend on the
actual time base mode.
The readout then displays e.g. ”Y2>...” indicating the
uncalibrated condition instead of ”Y2:...”. Pressing and holding
the CHI pushbutton again switches the LED off, sets the
deflection coefficient into calibrated condition and activates
the attenuator function. The previous vernier setting will not
be stored.
INV.:
Briefly and simultaneously pressing the CHII and the DUALMENU (16) pushbutton switches the channel II invert function
on or off. The invert ”on” condition is indicated by the
readout with a horizontal bar above ”Y2” (Yt mode). The
invert function causes the signal display of channel II to be
inverted by 180°.
[20] TRIG. MODE – Pushbuttons.
Pressing one of these pushbuttons opens the trigger coupling
Pulldown menu with the actual setting highlighted. Briefly
pressing a pushbutton selects the trigger coupling. Please
note “B: Menu Display and Operation”.
The term “trigger coupling” describes the way the trigger
signal is connected to the trigger unit.
AC: DC content suppressed,
DC: peak value detection inactive,
HF:high pass filter cuts off frequencies below
approx. 50 kHz, trigger point symbol switched off
LF:low pass filter cuts off frequencies above
approx. 1.5 kHz,
1. DEL.POS. (delay position).
In the delay time base modes “sea” (search), “del”(delay)
and “dTr” (triggered after delay) a delay time between the
trigger event and the start of the trace can be set with the
DEL.POS. control.
See DEL.MODE [23].
2. HO (holdoff time).
The holdoff time function can be activated if normal
(undelayed) time base mode is present. On condition that the
HO LED is not lit the holdoff time is set to minimum. The HO
LED lights up and the holdoff time increases as the knob is
rotated clockwise. A signal sounds on reaching the maximum
holdoff time. Similarly in the opposite direction until minimum
holdoff time is reached (HO LED extinguishes).
The holdoff time is automatically set to minimum when the
time base is changed. (For the application of holdoff time
setting see the paragraph with the same heading).
[22] TIME/DIV. – Control knob with a double function.
This control can be used for time coefficient selection in 1-25-sequence under calibrated condition when the VAR LED is
not lit or as a vernier (variable) control (VAR LED lit).
Time coefficient setting:
In undelayed, delayed (“del”) and as delayed triggered (“dTr”)
time base mode, this control serves as a time deflection
coefficient selector. Rotating anticlockwise increases the
deflection coefficient and rotating clockwise decreases the
coefficient. The setting is displayed at the top left in the
Subject to change without notice
17
Page 18
Controls and Readout
V
R
V
NM
HO
V
Readout (e.g. “10 µs”). Depending on the time base mode
the following ranges are available without taking X x10 into
account.
st
1
Undelayed: 500 ms/div – 50 ns/div.
The following (principle) values depend on the time base
setting in undelayed mode, as the undelayed time base
always has to have the higher coefficient.
nd
2
Search (“sea”): 20 ms/div – 100 ns/div.
rd
3
Delay (“del”): 20 ms/div. – 50 ns/div.
th
Triggered after delay (“dTr): 20 ms/div – 50 ns/div.
4
VAR.:
The vernier (variable) function is described under item Z ON/
OFF VAR [24].
6
Y-POS / CURS.I
VOLTS / DIV.
AR
20V 1mV20V 1mV0.5s 50ns
CH ICH IIDEL.MODE Z-ON/OFF
VAR .
14
15
79811131210
Y-POS / CURS.IILEVELX-POS.
TRIG.
SOURCE
DUAL
MENU
CU
CURSOR
POS
16
VOLTS / DIV.
INV.
17
18
NM
AT
AR
MENU
TRIG. MODE
DEL.POS.
VAR .
19
20
PUSH
BOTH
PUSH
LONG
TIME / DIV.
AR
ON
OFF
21
23
22
x10
X-
MAG.
VAR .
Functions
“sea”:
In SEARCH mode, the holdoff time is automatically set to
minimum and for the first few divisions the trace is blanked.
The point at which the trace is unblanked can be varied with
the DEL.POS [21] control (fine adjustment) from about 2 to
7 divisions. The blanked section serves as a guide to the
delay time. The delay time is based on the current time
deflection coefficient setting and can also be coarsely set
with the TIME/DIV control (range: 20 ms to 100 ns).
The signal position at which the unblanking occurs marks the
trace start position which is present after switching over
from “sea” to “del”. This enables lower time deflection
coefficient settings for signal expansion.
“del”:
In DELAY mode, a trigger event does not start the trace at
once but only initiates the delay time. After the delay time has
elapsed the trace is started. Selecting lower time deflection
coefficients (higher time base speed) cause a signal expansion
in X direction.
The DEL.POS [21] control can still be used for correcting the
signal start position affected by the TIME/DIV setting.
Note: With higher expansion rates the trace intensity may
reduce drastically.
“dTr”:
In triggered DELAY mode the first trigger unit, used for
triggering in undelayed time base mode, starts the delay time
as in “del” mode. After the delay time has elapsed the delay
time base must be triggered by the second trigger unit, to
24
start and unblank the trace. The latter requires suited
instrument settings (LEVEL, SLOPE) to enable triggering.
[23] DEL.MODE ON/OFF – Pushbutton with two functions.
ON/OFF function:
Pressing and holding this pushbutton switches over between
delayed and undelayed time base mode. The actual setting is
indicated by the READOUT. The delayed time base operation
enables a magnified display in X direction which is otherwise
only possible with a second time base.
a) Undelayed time base mode.
If on the right of the trigger READOUT information (source,
slope, coupling) neither “sea”, “del” nor “dTr” is indicated,
undelayed time base mode is present.
Note: When the intensity modulation function is switched
on, the letter “Z” is visible in this position on the screen.
b) Delayed time base mode.
Is indicated by the READOUT showing “sea”, “del” or
“dTr”. If intensity modulation was chosen before switching
over to delay time base mode, this function is automatically
switched off and consequently the letter “Z” deleted.
Switching over from undelayed to delayed time base mode
automatically selects “sea” (search) mode. Briefly pressing
the pushbutton then opens a Pulldown menu for operating
mode selection. Please note “B: Menu Display and Operation”.
The following description assumes that in undelayed time
base mode the trace starts at left edge of the graticule, with
x10 X MAG. switched off.
Note: The trigger indicator LED (TR) [10] only indicates the
trigger condition of the first trigger unit. It may be lit although
the trigger conditions for the second time base are not met
and the trace remains blanked.
As in “del” mode the DEL.POS [21] control can still be used.
In contrast to complex signals the effect of this function may
not be noticed with simple repetitive signals as the trigger
point ‘hops’ from cycle to cycle, each being the same.
[24] Z ON/OFF VAR. – Pushbutton with two functions.
Z ON/OFF:
Briefly pressing the pushbutton switches over the function
of the TRIG.EXT. BNC-socket [30] from external trigger input
to intensity modulation input and vice versa. In connection
with external triggering, delay time base (“sea”, “del” or
“dTr”) and “Component Tester” mode, Z modulation can not
be enabled.
Z modulation is shown on the READOUT to the right of
“trigger source, slope and coupling” indicated by the letter
“Z”. High TTL level (positive logic) gives blanking, dark, low
level gives unblanking, bright. No higher voltages than +5
Volt are permitted.
VAR.:
Pressing and holding the button changes the function of the
TIME/DIV. -knob [22] from time deflection coefficient switch
to time vernier (fine adjustment) control and vice versa.
18
Subject to change without notice
Page 19
Controls and Readout
The current function is indicated by the VAR LED. The TIME/
DIV. knob functions as a vernier when the VAR LED is
switched on, but the time base setting remains calibrated
until the (vernier) knob is operated. The readout now indicates
e.g. ”>10ms” instead of ”10ms”. Rotating further
anticlockwise increases the time deflection coefficient
(uncalibrated) until the maximum is reached indicated by a
beep. Rotating the knob clockwise has the opposite effect.
Now, the vernier is again in the calibrated position and the
symbol ”>” extinguishes.
Underneath the front panel sector described above, the BNC
sockets and two pushbuttons are located.
INPUT CHI (X)
1MΩ II
18pF
!!!
max.
400Vp
25
x1 / x10
AC/DC/GND
2627
CAT
I
INPUT CHI (X)
1MΩ II
18pF
max.
400Vp
282930
x1 / x10
AC/DC/GND
TRIG. EXT.
INP.
(Z)
max.
100Vp
[25] INPUT CH I (X) – BNC socket.
This BNC socket is the signal input for channel I. The outer
(ground) connection is galvanically connected to the
instrument ground and consequently to the safety earth
contact of the line/mains plug. The AC/DC/GND pushbutton
[26] is assigned to the input.
In XY mode, signals at this input are used for the X deflection.
[26] AC/DC/GND x1/x10 – Pushbutton with several functions.
AC/DC/GND:
Briefly pressing this pushbutton opens the input coupling
Pulldown menu if a channel mode is present in which
channel I is activated.
The following input couplings are available: AC, DC and GND
(ground). Please note “B: Menu Display and Operation”.
After the Pulldown menu has extinguished, the READOUT
displays the present input coupling at the bottom right hand
of ”Y1: deflection coefficient”; the “~“ symbol indicates
AC, the “=” symbol DC and “GND” is for ground.
The GND setting disables the input signal and the VOLTS/DIV [14] knob. Then in automatic trigger mode (Yt) the
undeflected trace is visible representing the 0 Volt trace
position; in XY mode the X deflection is deactivated.
x1/x10:
Probe factor selection is performed by pressing and holding
the pushbutton. This selects the indicated deflection
coefficient of channel I displayed in the readout, between 1:1
and 10:1. In condition 10:1, the probe factor is thus indicated
by a probe symbol displayed by the readout in front the
channel information (e.g. ”probe symbol”, Y1...). In the
case of cursor voltage measurement, the probe factor is
automatically included.
Please note:
The symbol should not be activated unless a x10 (10:1)
attenuator probe is used.
to safety earth. This socket can be used as reference potential connection for DC and low frequency signal measurement
purposes and in “Component Tester” mode.
[28] INPUT CH II – BNC socket.
This BNC socket is the signal input for channel II. The outer
(ground) connection is galvanically connected to the
instrument ground and consequently to the safety earth
contact of the line/mains plug. The AC/DC/GND pushbutton
(29) is assigned to the input.
In XY mode, signals at this input are used for the Y deflection.
[29] AC/DC/GND x1/x10 – Pushbutton with several functions.
AC/DC/GND:
Briefly pressing this pushbutton opens the input coupling
Pulldown menu if a channel mode is present in which
channel II is activated.
The following input couplings are available: AC, DC and GND
(ground). Please note “B: Menu Display and Operation”.
After the Pulldown menu has extinguished, the READOUT
displays the present input coupling at the bottom right hand
of ”Y2: deflection coefficient”; the “~“ symbol indicates
AC, the “=” symbol DC and “GND” is for ground.
The GND setting disables the input signal and the VOLTS/DIV (18) knob. Then in automatic trigger mode (Yt) the
undeflected trace is visible representing the 0 Volt trace
position; in XY mode the Y deflection is deactivated.
x1/x10:
Probe factor selection is performed by pressing and holding
the pushbutton. This selects the indicated deflection
coefficient of channel II displayed in the readout, between
1:1 and 10:1. In condition 10:1 the probe factor is thus
indicated by a probe symbol displayed by the readout in front
the channel information (e.g. ”probe symbol”, Y2...). In the
case of cursor voltage measurement, the probe factor is
automatically included.
Please note:
The symbol should not be activated unless a x10 (10:1)
attenuator probe is used.
[30] TRIG. EXT. / INPUT (Z) – BNC socket with two functions.
The outer (ground) connection is galvanically connected to
the instrument ground and consequently to the safety earth
contact of the line/mains plug. The input impedance is
approx. 1 MOhm II 20pF.
Briefly pressing the Z ON/OFF VAR [24] pushbutton switches
over the function of this socket.
TRIG. EXT:
The BNC socket serves as external trigger signal input, if
external triggering is selected.
The trigger coupling depends on the TRIG. MODE [20]
setting.
Z-Input:
If neither “Component Tester”, delayed time base mode
(“sea”, “del” or “dTr”) nor external trigger coupling (“ext”)
is chosen, the socket is operative as a Z (trace intensity
modulation) input.
High TTL level (positive logic) affects blanking, low level
gives unblanking. No higher voltages than +5 Volt are
permitted.
Below the CRT are the controls for the readout, the component
tester and the squarewave calibrator with their outputs.[31]
Subject to change without notice
19
Page 20
Controls and Readout
[31] MAIN MENU – READOUT – Pushbutton with double
function.
MAIN MENU
Briefly pressing calls the MAIN MENU. It contains the
submenus ADJUSTMENT and SETUP & INFO partly containing further submenus. A menu description can be found
under „E: MAIN MENU“.
Although self explanatory, a description of the menu selection
and other operating functions can be found in this part of the
manual under „B: Menu Display and Operation“.
READOUT
Pressing and holding the pushbutton switches the readout
on or off. With the readout switched off, the INTENS/FOCUS
function can consequently not be set to RO.
It may be required to switch off the readout if interference is
visible on the signal(s). Such interference may also originate
from the chopper generator if the instrument is operated in
chopped DUAL mode.
All intensity and focus (INT./FOC.) settings are stored after
the instrument is switched off and present when switching
the instrument on again; however the READOUT will always
be switched on.
For voltage measurement, AC or DC trigger coupling is
required. DC voltage measurement assumes DC input
coupling.
In the case of high frequency signals, the different frequency
responses of the trigger amplifier and Y amplifier cause a
reduction of the measurement accuracy.
If relatively low frequency signals (<20 Hz) are present, the
measurement value continuously changes, following the
waveform. The pulse duty factor of such signals also affects
the measurement result as well as the trigger slope setting.
Frequency and Period measurement assume that the trigger
condition is met (TR LED lit) and normal triggering active for
signals >20 Hz. Very low frequency signals require a
measuring time of several seconds.
To avoid measuring errors the complete signal must be
displayed within the vertical graticule limits; i.e. without
overranging.
DC shows the DC Mean Value (please note “Mean Value
Display”)
Frequency for frequency measurement. Depending on the
trigger point setting, complex signals cause different results.
INPUT CHI (X)
1MΩ II
18pF
!!!
max.
400Vp
25
x1 / x10
AC/DC/GND
2627
CAT
I
INPUT CHI (X)
1MΩ II
18pF
max.
400Vp
282930
x1 / x10
AC/DC/GND
TRIG. EXT.
INP.
(Z)
max.
100Vp
[32] MEASURE SET – Pushbutton with double function.
a) MEASURE:
Briefly pressing calls the “AUTO MEASURE” menu, if CURSOR lines are not activated. Otherwise the “CURSOR
MEASURE” menu is displayed. Pressing and holding the
SELECT - ON/OFF-pushbutton [34] activates or deactivates
the CURSOR lines.
Applicability of measuring functions
Where a measuring function is not supported in conjunction
with an operating mode, instead of a measuring value the
READOUT indicates “n/a” (not applicable). For example the
READOUT displays “∆t: n/a” if ∆t measurement is selected
in combination with XY mode.
Uncalibrated Settings / Overflow Indication
If the deflection coefficient is uncalibrated the READOUT
indicates e.g. “Y1>2V= or >500 µs. Such conditions are
indicated by a “>” or “<” sign automatically put in front of the
displayed measuring value.
Measurement range overflow (exceeding) is indicated in
front of the measuring value by the “>” sign.
Non executability of measurements
A question mark (?) is displayed if the measuring unit can‘t
find a useful value (e.g. frequency measurement without a
signal).
b) AUTO MEASURE
The measurement results of the various menu items explained
under 32.x are related to the signal used for triggering.
Period measures the period length. Depending on the trigger
point setting, complex signals cause different results.
Peak + determines the positive peak value of an AC voltage.
The DC content is considered if DC input coupling is present.
Peak – determines the negative peak value of an AC voltage.
The DC content is considered if DC input coupling is present.
Peak Peak measures the difference voltage (AC) between
the positive and the negative peak values.
Trigger Level displays the trigger comparator reference
voltage which must be exceed for triggering.
Off automatic measurement switched off.
c) CURSOR MEASURE:
Briefly pressing the MEASURE SET pushbutton on condition
CURSOR ON [34] calls this menu. The measurement results
of the different menu items then are related to the CURSOR
settings referring to the signal.
The Y-POS/CURS.I- [6] and Y-POS/CURS.II-knobs [8] enable
CURSOR line setting if the CURSOR POS LED is lit. Then
each CURSOR line is marked by a symbol (“I“, “II”) indicating
the relationship between each Y POS/CURS. knob and CURSOR line. In cases where more than two CURSOR lines or
additionally “+” symbols are displayed, the SELECT [34]
function switches over the assignment. If both CURSOR
lines or “+” symbols have the same marking, both can be
shifted simultaneously (Tracking function).
∆∆
∆t (display “∆t: measured value“)
∆∆
Enables time measurement by aid of two vertical CURSOR
lines in Yt mode (not in XY mode). Briefly pressing the UNIT
(35) pushbutton directly switches over from
∆∆
∆t to 1/
∆∆
∆∆
∆t
∆∆
(frequency) measurement and vice versa.
20
Subject to change without notice
Page 21
∆∆
1/
∆t (display “1/∆t: measured value“)
∆∆
Two vertical CURSOR lines enable frequency measurement
in Yt mode (not in XY mode). Briefly pressing the UNIT [35]
∆∆
pushbutton directly switches over from 1/
∆t to
∆∆
∆∆
∆t (time)
∆∆
measurement and vice versa.
Rise Time (display “tr 10: measured value”)
Rise time measurement by aid of two horizontal CURSOR
lines and two “+” symbols which have the following meaning.
st
Lower CURSOR line = 0%.
1
nd
Lower “+” symbol = 10% of the CURSOR lines
2
distance.
3rdUpper “+” symbol = 90% of the CURSOR lines
distance.
4thUpper CURSOR line = 100%.
SET[32] enables an automatic signal related CURSOR line
setting (in DUAL mode related to the signal used for triggering),
which can later be changed manually.
The distance between the “+” symbols and the CURSOR
lines are set automatically. For rise time measurement the
horizontal position of the “+” symbols must be set manually
to the signal slope. This requires that the CURSOR POS is
active and each “+” symbol is marked (“I”, “II”) by the aid
of the SELECT [34] function.
Controls and Readout
The unit to be displayed must be selected by briefly pressing
the UNIT [35] pushbutton to call the UNIT menu. Then the
following units are being offered: ratio, %, ° (angle unit:
degree of angle and pi.
The long CURSOR line in the left position always serves as
reference line. A “–“ (minus) sign indicates measurement
results if the short CURSOR line is placed left of the reference
line.
Ratio:
Enables the measurement of pulse duty ratio. The distance
between the long CURSOR lines is equivalent to1 (whole
cycle).
Example for a pulse signal with 4 div. pulse and 1 div. space:
The long CURSOR lines must coincide with the start position
of first and the second pulse (distance = 5 div.) as the
reference distance (1). Then the “I” symbol must be assigned
to the short CURSOR line SELECT [34] which must then be
set to the pulse end position (4 div. after the pulse start).
Corresponding to the ratio of pulse duration to period length
(4:5 = 0.8) “0.8” will be displayed.
%:
Same function as described before under “Ratio”. The
measurement result is displayed in % (unit).
Note: For maximum “+” symbol positioning and measuring
accuracy first set the signal slope to the screen centre XPOS. [12] and then activate X-magnifier X-MAG. x10 [13].
To avoid CURSOR line and “+” symbol changes after each
change of a signal position in X and/or Y direction, a fixed
relation between signal and CURSOR display can be made by
activating the GLUE (33) function. GLUE is indicated by a
reduced number of dots in the CURSOR lines and the “+”
symbols.
Further information about this item can be found in this
manual under “Type of signal voltage” in section “Rise Time
Measurement”.
∆∆
∆V (display ∆V: channel, measured value)
∆∆
CURSOR supported voltage measurement.
In Yt (time base) mode two horizontal CURSOR lines are
displayed: Single channel (CH I or CH II) mode automatically
relates one signal and the CURSOR lines. The measurement
value is connected with the Y deflection coefficient.
DUAL mode requires selection between channel I and II with
the SOURCE -pushbutton [33]. The CURSOR line must be
placed on the signal (channel) chosen by the SOURCE
function.
Addition (“add”) mode requires equal Y deflection coefficients
for both channels.
XY mode causes the display of two vertical or horizontal
CURSOR lines: The SOURCE-pushbutton [33] allows selection
between X (CH I) and Y (CH II) voltage measurement. In the
case of X voltage measurement, vertical CURSOR lines are
displayed.
V to GND (display V: channel, measured value)
One CURSOR line is displayed for voltage measurement
related to the trace 0 Volt position. This is the only exception
to the description of item ∆V.
Ratio X (display “ratio:X, measured value, unit”)
Ratio X measurement causes the display of two long and one
short CURSOR lines and is enabled in Yt (time base) mode only.
°:
Angle measurement referring to the CURSOR line distances.
The distance between the long CURSOR lines should cover
one signal period, equivalent to 360°. Angle measurement
then can be performed by shifting the short CURSOR line.
Additional information can be found in section “Operating
modes of the vertical amplifiers in Yt mode” under “Phase
difference measurement in DUAL mode”.
pi:
Determination of the value for “pi” referring to the CURSOR
line distances. The equivalent for “2 pi” is one sine wave
period; thus the distance between the long CURSOR lines
must be 1 period. If the distance between the long CURSOR
in left hand position and the short CURSOR line referring to
it is 1.5 periods, “3 pi” will be indicated.
Ratio Y (display “ratio:Y, measured value, unit”)
Ratio Y measurement causes the display of two long and one
short CURSOR lines and is enabled only in Yt (time base)
mode.
Briefly pressing the UNIT-pushbutton [35] switches over
between the ratio (unnamed) and ratio in %.
The long CURSOR line in the lower position always serves as
the reference line. A “–“ (minus) sign indicates measurement
results if the short CURSOR line is placed below the
reference line.
Ratio: The distance between both long CURSOR lines is
equivalent to1.
Example:
If the distance between the long CURSOR lines is
6 div. and the short CURSOR line is activated SELECT [34]
and set 4 div. above the reference CURSOR line, the ratio is
4:6, causing “0.667” (without unit) to be displayed.
%:
The only difference between previous item “Ratio” and “%”
is that the distance between the long CURSOR lines is
equivalent to100% and the measuring result is displayed as
a % value.
Subject to change without notice
21
Page 22
Controls and Readout
Gain (display “gain: measured value, unit”)
Ratio measurement of signal voltages by the aid of two long
and two short CURSOR lines; enabled only in Yt (time base)
mode.
Briefly pressing the UNIT-pushbutton [35] selects ratio
(unnamed), % or dB.
The application of Gain measurement depends whether one
or two signals are displayed.
1th: One signal (CH I, CH II or “add” mode).
A measurement can be made on one signal before and after
a signal frequency change.
The distance between the long CURSOR lines serves as the
reference value. The measured value is calculated from the
distance between the short CURSOR lines compared to the
reference value.
This method is suitable to determine e.g. the oscilloscope‘s
frequency response.
2nd: DUAL mode.
Enables two port measurements (amplifier, attenuator) by
determination of the ratio of input and output voltages. For
correct measuring results you must determine which channel
is applied to the input and output ports respectively.
Both long CURSOR lines must be placed on the channel I
signal while the short CURSOR lines must be set on the
channel II signal.
Briefly pressing the SOURCE pushbutton calls a menu which
offers “g1→2:“ and “g2→1:“. The selection of the required
setting can then be made by briefly pressing the SOURCE
pushbutton until the setting is highlighted. If channel I is
connected to the input and channel II the output of the two
ports, “g1→2:“ must be chosen. Conversely if the channels
are reversed choose “g2→1:“.
SET
Pressing and holding SET in condition CURSOR supported
voltage measurement, causes an automatic signal related
CURSOR line setting within certain limits. As it is the trigger
signal that is measured, (trigger source CH I or CH II) the
trigger coupling affects the measuring result. Without a
signal or with an untriggered signal, the CURSOR lines do not
change. SET is activated on condition that:
1. The CURSOR lines are visible.
2. A CURSOR MEASURE menu function must have been
chosen which causes the display of horizontal CURSOR
lines (Rise Time, ∆V, V to GND, Ratio Y and Gain).
3. CH I, CH II or DUAL mode is activated.
CT
37
0.2 Vpp
UNIT
SELECTSOURCE
CAL.
SEL.
36373534333231
ON
OFF
CURSOR
MEASURE MAIN MENU
GLUESET
[33] SOURCE GLUE – Pushbutton with double function.
SOURCE
Briefly pressing selects the source (channel) the measurement
display refers too. Consequently this function is inactive
during one channel operation (CH I, CH II or “add“ mode).
READ
OUT
“∆V“ and “V to GND“) two long CURSOR lines are
displayed. Briefly pressing SOURCE selects the channel
and it`s deflection coefficient for the measurement. The
CURSOR lines must be set to the signal according to the
selected channel.
b) DUAL mode in combination with “Gain“ (two port)
measurement allows you to determine the input and
output voltage ratio with two long and two short CURSOR
lines being visible. A correct measurement requires the
input of the conditions which channel is connected to the
input and output respectively.
GLUE
Pressing and holding switches this function on or off, which
is indicated by the way the CURSOR lines are displayed. In
GLUE on condition the number of dots from which CURSOR
lines and “+” symbols consist is reduced.
GLUE combines the CURSOR lines and “+” symbol position
with the Y and X position controls. Y and X position changes
then affect both the signal and the CURSOR lines and “+”
symbols.
[34] SELECT ON OFF – Pushbutton with double function.
ON OFF
Pressing and holding switches the CURSOR lines on or off.
When the CURSOR lines are activated, the READOUT displays
the last activated measuring function of the CURSOR
MEASURE menu. Briefly pressing MEASURE [32] opens this
menu.
Switching the CURSOR lines off additionally switches over
to last used AUTO MEASURE function. Briefly pressing
MEASURE [32] opens this menu.
SELECT
If the CURSOR lines are visible (CURSOR MEASURE) and
the CURSOR POS function [7] is active, the symbols “I” and
“II” are assigned to CURSOR lines or “+” symbols. The “I”
and “II” symbols indicate by which Y-POS/CURS. (I or II)
control the CURSOR line(s) position can be changed. Briefly
pressing the SELECT pushbutton changes the assignment.
Only the CURSOR lines and “+” symbols which are assigned
can be shifted. Tracking mode is present when two CURSOR
line or “+”symbols have the same assignment; i.e. they are
shifted simultaneously by the same control.
[35] UNIT CAL. SEL – Pushbutton with double function.
UNIT
Briefly pressing changes the unit of the displayed measuring
value. If CURSOR MEASURE is active (CURSOR lines visible) and more then two units are selectable, a menu opens;
otherwise the switch over appears directly without a menu.
On condition AUTO MEASURE the UNIT function selects
between frequency and period or PEAK+ and PEAK–.
CAL. SEL.
Pressing and holding opens the CAL. FREQUENCY menu,
which offers DC and AC (1Hz to 1MHz) voltage signals. In
setting “dependent on TB” the output signal frequency
depends on the time coefficient setting TIME/DIV. [22].
a) If DUAL or XY mode is present in combination with
CURSOR voltage measurement (CURSOR MEASURE:
22
The “0.2V
” [36] marked socket serves as an output for the
pp
selected signal.
Subject to change without notice
Page 23
Main Menue
1 Hz – 1 MHz
These AC square wave signals can be used for probe
adjustment and judgement of the frequency response. As
the frequency and the pulse duty factor accuracy are not
important for such purposes, these values are not specified
and are therefore relatively inaccurate.
Dependent on TB
The square wave signal’s pulse duty factor differs widely in
most settings from 1:1. In time deflection ranges from
500 ms/div to 1µs/div. the period length and the time deflection
coefficient are equal; this allows you to check the time base
accuracy. The smallest period length is 1 µs (1 MHz) and
consequently time coefficients <1µs/div do not change it.
CT
37
[36] 0.2 V
0.2 Vpp
Concentric socket
pp
UNIT
SELECTSOURCE
CAL.
SEL.
36373534333231
ON
OFF
CURSOR
MEASURE MAIN MENU
GLUESET
This socket serves as the output for the signals described
under item CAL. SEL. [35]. The output impedance is approx.
READ
OUT
50 Ohm. For high impedance loads (Oscilloscope approx.
1 M Ohm, Digital Voltmeter approx. 10 MOhm) the output
voltage is either 0.2 Volt DC or 0.2 V
(AC, square wave).
pp
Under “First Time Operation” section “Probe compensation
and use” the most important applications of this signal can
be found.
[37] CT – Pushbutton and 4mm banana jack
Briefly pressing the pushbutton switches the instrument
over from oscilloscope to “Component Tester” mode and
vice versa.
This mode is indicated by the READOUT which displays
”Component Tester”.
One test lead is connected to the CT socket. The second test
lead uses the ground socket [27].
Please note ”Component
Tester”.
The maximum test voltage is approx. 20 Vpp under open
circuit conditions, while the max. test current under short
circuit condition is approx. 20 mApp.
Briefly pressing the CT-pushbutton switches back to the
previous oscilloscope operating conditions.
MAIN MENU
The instrument software contains several menus. The following
menus, submenus and menu items are available:
1.ADJUSTMENT contains the submenus:
1.1AUTO ADJUSTMENT with the menu items
1.1.1 SWEEP START POSITION
1.1.2 Y AMP
1.1.3 TRIGGER AMP
1.1.4 X MAG POS
1.1.5 CT X POS
Calling one of these menu items requires that no signal is applied
on any input. For further information note „Adjustments“.
1.2MANUAL ADJUSTMENT contains menu items that are
only accessible for HAMEG workshops.
2.SETUP & INFO contains the submenus:
Subject to change without notice
2.1MISCELLANEOUS
Active functions are indicated by „x“. SET switches over
from active to inactive and vice versa.
2.1.1 CONTROL BEEP. Acoustic signal confirms error free
operation.
2.1.2 ERROR BEEP. Sounds in case of operating error.
2.1.3 QUICK START. Reduces the waiting time after POWER
ON, as neither the HAMEG logo nor the check and
initialisation are displayed.
2.2FACTORY
Submenus are available only for HAMEG authorized service
workshops
2.3INFO
Displays information regarding the instrument’s hardware
and software.
23
Page 24
First Time Operation
First Time Operation
The following text assumes that the ”SAFETY” section of this
manual has been read carefully and understood.
Each time before the instrument is put into operation check that
the oscilloscope is connected to protective earth. For that reason
the power cable must be connected to the oscilloscope and the
power outlet. Then the test lead(s) must be connected to the
oscilloscope input(s). Check that the device under test is switched
off and connect the test lead(s) to the test point(s). Then switch
on the instrument and afterwards the device under test.
The oscilloscope is switched on by depressing the red POWER
pushbutton. After a few seconds the HAMEG logo and the
instrument software release is displayed on the screen, if this
function is active. As long as the HAMEG logo is visible different
internal checks are made. Thereafter the instrument will revert to
its last used operating mode.
If after that no trace is visible, the AUTO SET pushbutton should
be pressed briefly. This selects the Yt mode and medium trace
and readout intensity (
POS.I and X-POS. controls to centre the baseline. Set INTENS.
for suitable brightness (intensity) and FOCUS for optimum
sharpness (input(s) grounded) of the trace. The oscilloscope is
now ready for use.
If the AUTOSET function was not used and only a spot appears
(CAUTION! CRT phosphor can be damaged), reduce the intensity
immediately and check that the XY mode is not selected (XY not
displayed in the readout).
To obtain the maximum life from the cathode ray tube, the
minimum intensity setting necessary for the measurement in
hand and the ambient light conditions should be used.
please note ”AUTOSET”
). AdjustY-
The built in calibration generator provides a square wave signal
with selectable frequencies and a very fast rise time (<4 ns) from
the output socket below the CRT screen.
As the square wave signals are used for probe compensation
adjustments, neither the frequency accuracy nor the pulse duty
factor are of importance and therefore not specified. The output
provides 0.2 Vpp ±1% (tr <4 ns) for 10:1 probes. When the Ydeflection coefficient is set to 5 mV/div, the calibration voltage
corresponds to a vertical display of 4 divisions (10:1 probe).
The output socket has an internal diameter of 4.9 mm to
accommodate the internationally accepted shielding tube diameter
of modern probes and F series slimline probes. Only this type of
construction ensures the extremely short ground connections
which are essential for an undistorted waveform reproduction of
non sinusoidal high frequency signals.
Adjustment at 1 kHz
The C-trimmer adjustment (low frequency) compensates the
capacitive loading on the oscilloscope input. By this adjustment,
the capacitive division assumes the same ratio as the ohmic
voltage divider to ensure the same division ratio for high and low
frequencies, as for DC. (For 1:1 probes or switchable probes set
to 1:1, this adjustment is neither required nor possible). A
baseline parallel to the horizontal graticule lines is essential for
accurate probe adjustments.
Connect the 10:1 probe to the input of the channel it is to be
adjusted for and don‘t mix up the probes later (always use that
particular probe with the same channel). Set the deflection
coefficient to 5mV/div and the input coupling to DC. The time
deflection coefficient should be set to 0.2ms/div. All deflection
coefficients should be calibrated (Variable controls at CAL position).
Plug the probe tip into the calibrator output socket.
(See also ”Trace rotation TR”).
Particular care is required when a single spot is displayed, as a
very high intensity setting may cause damage to the fluorescent
screen of the CRT. Switching the oscilloscope off and on at short
intervals stresses the cathode of the CRT and should therefore be
avoided.
The instrument is so designed that even incorrect operation will
not cause serious damage.
Trace Rotation TR
In spite of Mumetal shielding of the CRT, effects of the Earth’s
magnetic field on the horizontal trace position cannot be completely
avoided. This is dependent upon the orientation of the oscilloscope
on the place of work. A centred trace may not align exactly with
the horizontal center line of the graticule. A few degrees of
misalignment can be corrected. Please note “Controls and
Readout” section “E: MAIN MENU item 1. TRACE ROT”.
Probe compensation and use
To display an undistorted waveform on an oscilloscope, the probe
must be matched to the individual input impedance of the Y
amplifier.
For this purpose a square wave signal with a very fast rise time
and minimum overshoot should be used, as the sinusoidal
contents cover a wide frequency range.
Approximately 2 complete waveform periods are displayed on
the CRT screen. The compensation trimmer should be adjusted.
The location of the low frequency compensation trimmer can be
found in the probe information sheet. Adjust the trimmer with the
insulated screwdriver provided, until the tops of the square wave
signal are exactly parallel to the horizontal graticule lines (see
1kHz diagram). The signal height should then be 4div ± 0.16div (=
4 % (oscilloscope 3% and probe 1%). During this adjustment, the
signal edges will remain invisible.
Adjustment at 1 MHz
Probes HZ51, 52 and 54 can also be HF compensated. They
incorporate resonance de-emphasing networks (R-trimmer in
conjunction with capacitor) which permit probe compensation in
the range of the upper frequency limit of the Y amplifier. Only this
compensation adjustment ensures optimum utilization of the full
bandwidth, together with constant group delay at the high
frequency end, thereby reducing characteristic transient distortion
near the leading edge (e.g. overshoot, rounding, ringing, holes or
bumps) to an absolute minimum.
Using the probes HZ51, 52 and 54, the full bandwidth of the
oscilloscope can be utilized without risk of unwanted waveform
distortion.
24
Subject to change without notice
Page 25
Operating modes oft the Y amplifiers in Yt mode
Prerequisite for this HF compensation is a square wave generator
with fast risetime (typically 4 ns), and low output impedance
(approx. 50 Ohm), providing 0.2 V at a frequency of approx. 1MHz.
The calibrator output of this instrument meets these requirements.
Connect the probe to the input previously used when 1 kHz
adjustment was made. Select 1 MHz output frequency. Operate
the oscilloscope as described under 1 kHz but select for 0.2 µs/
div time deflection coefficient setting.
Insert the probe tip into the output socket. A waveform will be
displayed on the CRT screen, with leading and trailing edges
clearly visible. For the HF-adjustment now to be performed, it will
be necessary to observe the rising edge as well as the upper left
corner of the pulse top. The location of the high frequency
compensation trimmer(s) can also be found in the probe
information sheet. These R-trimmer(s) have to be adjusted such
that the beginning of the pulse is as straight as possible. Overshoot
or excessive rounding is unacceptable. The adjustment is relatively
easy if only one adjusting point is present. In case of several
adjusting points the adjustment is slightly more difficult, but
causes a better result. The rising edge should be as steep as
possible, with a pulse top remaining as straight and horizontal as
possible.
After completion of the HF adjustment, the signal amplitude
displayed on the CRT screen should have the same value as
during the 1 kHz adjustment.
Probes other than those mentioned above, normally have a larger
tip diameter and may not fit into the calibrator output. Whilst it is
not difficult for an experienced operator to build a suitable
adapter, it should be pointed out that most of these probes have
a slower rise time with the effect that the total bandwidth of
scope together with probe may fall far below that of the
oscilloscope. Furthermore, the HF adjustment feature is nearly
always missing so that waveform distortion can not be entirely
excluded. The adjustment sequence must be followed in the
order described, i.e. first at 1 kHz, then at 1 MHz.
Operating modes
of the Y amplifiers in Yt mode
The most important controls regarding the operating modes of
the Y amplifiers are the pushbuttons: CH I [15], DUAL [16] and
CH II [19]. Their functions are described in the section ”Controls
and Readout”.
In most cases oscilloscopes are used to display signals in Yt
mode. Then the signal amplitude deflects the beam in vertical
direction while the time base causes an X deflection (from left to
right) at the same time. Thereafter the beam becomes blanked
and fly back occurs.
The following Yt operation modes are available:
st
Single channel operation of channel I (Mono CH I).
1
nd
2
Single channel operation of channel II (Mono CH II).
rd
Two channel operation of channel I and channel II (DUAL).
3
th
Two channel operation of channel I and channel II displaying
4
the algebraic result as the sum or difference (“add”).
The way the channel switching is determined in DUAL mode
depends on the time base setting and is described in the section
”Controls and Readout”.
In ADD mode the signals of both channels are algebraically added
and displayed as one signal. Whether the resulting display shows
the sum or difference is dependent on the phase relationship or
the polarity of the signals and on the invert function.
In ADD mode the following combinations are possible for
In phase input voltages:
Channel II invert function inactive = sum.
Channel II invert function active = difference.
Antiphase input voltages:
Channel II invert function inactive = difference.
Channel II invert function active = sum.
Prerequisites for precise and easy probe adjustments, as well as
checks of deflection coefficients, are straight horizontal pulse
tops, calibrated pulse amplitude, and zero-potential at the pulse
base. Frequency and duty cycle are relatively uncritical. For
interpretation of transient response, fast pulse rise times and low
impedance generator outputs are of particular importance.
Providing these essential features, as well as selectable output
frequencies, the calibrator of the instrument can, under certain
conditions, replace expensive square wave generators when
testing or compensating wideband attenuators or amplifiers. In
such a case, the input to an appropriate circuit will be connected
to the calibrator output via a suitable probe.
The voltage provided by the probe to a high impedance input
(1 MOhm II 15 - 30 pF) will correspond to the division ratio of the
probe used (10:1 = 20 mV
output). Suitable probes are HZ51,
pp
52 and 54.
In the ADD mode the vertical display position is dependent upon
the Y position setting of both channels. The same Y deflection
coefficient is normally used for both channels with algebraic
addition.
Please note, that the Y-position settings are also added but
are not affected by the invert function.
Differential measurement techniques allow direct measurement
of the voltage drop across floating components (both ends above
ground). Two identical probes should be used for both Y inputs.
In order to avoid ground loops, use a separate ground connection
and do not use the probe ground leads or cable shields.
X-Y Operation
The important control for this mode is the pushbutton labelled
DUAL and MENU [16].
In XY mode the time base is deactivated. The signal applied to the
input of channel I front panel marking INPUT CHI (X) causes the
X deflection. The input related controls (AC/DC/GND pushbutton
and the VOLTS/DIV knob) consequently affect the X deflection.
For X position alteration, the X-POS.-control knob must be used,
as the Y-POS./CURS.I control is automatically deactivated. The
input deflection coefficient ranges are the same for both channels,
because the X x10 magnifier is inactive in XY mode.
Subject to change without notice
25
Page 26
Operating modes oft the Y amplifiers in Yt mode
The bandwidth of the X amplifier, is lower than the Y amplifier
and the phase angle which increases with higher frequencies,
must be taken into account (please note data sheet).
The Y signal applied at INPUT CHII can be inverted.
Lissajous figures can be displayed in the X-Y mode for certain
measuring tasks:
I Comparing two signals of different frequency or bringing
one frequency up to the frequency of the other signal.
This also applies for whole number multiples or fractions
of the one signal frequency.
I Phase comparison between two signals of the same
frequency.
Phase comparison with Lissajous figures
The following diagrams show two sine signals of the same
frequency and amplitude with different phase angles.
Phase difference measurement
in DUAL mode (Yt)
Phase differences between two input signals of the same
frequency and shape can be measured very simply on the screen
in Dual mode. The time base should be triggered by the reference
signal (phase position 0). The other signal can then have a leading
or lagging phase angle. In alternate triggering condition, phase
difference measurement is not possible.
For greatest accuracy, adjust the time base for slightly over one
period and set approximately the same height of both signals on
the screen. The Y-deflection coefficients, the time base coefficient
and the trigger level setting can be used for this adjustment,
without influence on the result. Both base lines are set onto the
horizontal graticule center line using the Y-POS.-knobs before the
measurement. With sinusoidal signals, use the zero (crossover
point) transitions; the sine peaks are less accurate. If a sine signal
is noticeably distorted by even harmonics, or if a DC voltage is
present, AC coupling is recommended for both channels. If it is
a question of pulses of the same shape, read off at steep edges.
It must be noted that the phase difference cannot be determined
if alternate triggering is selected.
Phase difference measurement in DUAL mode
Calculation of the phase angle or the phase shift between the X
and Y input voltages (after measuring the distances a and b on the
screen) is quite simple with the following formula, and a pocket
calculator with trigonometric functions. Apart from the reading
accuracy, the signal height has no influence on the result.
The following must be noted here:
I Because of the periodic nature of the trigonometric functions,
the calculation should be limited to angles ≤90° However
here is the advantage of the method.
I Due to phase shift, do not use too high a test frequency.
I It cannot be seen as a matter of course from the screen display
if the test voltage leads or lags the reference voltage. A CR
network before the test voltage input of the oscilloscope can
help here. The 1MOhm input resistance can equally serve as
R here, so that only a suitable capacitor C needs to be
connected in series. If the aperture width of the ellipse is
increased (compared with C short-circuited), then the test
voltage leads the reference voltage and vice versa. This applies
only in the region up to 90° phase shift. Therefore C should
be sufficiently large and produce only a relatively small, just
observable phase shift.
Should both input voltages be missing or fail in the XY mode, a
very bright light dot is displayed on the screen. This dot can burn
into the phosphor at too high a brightness setting (INTENS.
setting) which causes either a lasting loss of brightness, or in the
extreme case, complete destruction of the phosphor at this point.
t = horizontal spacing of the zero transitions in div
T = horizontal spacing for one period in div
In the example illustrated, t = 3 div and T = 10 div, the phase
difference in degrees is calculated from
or expressed in radians
Relatively small phase angles at not too high frequencies can be
measured more accurately in the X-Y mode with Lissajous
figures.
Measurement of an amplitude modulation
The momentary amplitude u at time t of a HF carrier voltage,
which is amplitude modulated without distortion by a sinusoidal
AF voltage, is in accordance with the equation
where:UT= unmodulated carrier amplitude
Ω= 2 pF = angular carrier frequency
ω= 2 pf = modulation angular frequency
m= modulation factor.
As well as the carrier frequency F, a lower side frequency F-f and
upper side frequency F+f arise because of the modulation.
26
Subject to change without notice
Page 27
Figure 1:
Amplitude and frequency spectrum for AM display (m = 50%)
Triggering and time base
base generator moves the beam from the left to the right of the
screen (time deflection = t).
Normally there are periodically repeating waveforms to be
displayed. Therefore the time base must repeat the time deflection
periodically too. To produce a stationary display, the time base
must only be triggered if the signal height and slope condition
coincide with the former time base start conditions. A DC voltage
signal can not be triggered as it is a constant signal with no slope.
The display of an amplitude modulated HF oscillation can be
evaluated with the oscilloscope provided the frequency spectrum
is inside the oscilloscope bandwidth. The time base is set so that
several cycles of the modulation frequency are visible. Strictly
speaking, triggering should be external with modulation frequency
(from the AF generator or a demodulator). However, internal
triggering is frequently possible with normal triggering using a
suitable trigger level setting and possibly also using the time
vernier (variable) adjustment.
Oscilloscope setting for a signal according to figure 2:
I Y: CH. I; 20mV/div; AC.
I TIME/DIV.: 0.2ms/div.
I Triggering: Normal; with LEVEL-setting; internal
(or external) triggering.
Figure 2: Amplitude modulated oscillation
(F = 1 MHz; f = 1 kHz; m = 50 %; U
If the two values a and b are read from the screen, the modulation
factor is calculated from
where
a = UT (1+ m) and b = UT (1- m).
The variable controls for amplitude and time can be set arbitrarily
in the modulation factor measurement. Their position does not
influence the result.
= 28.3 mV
T
rms
)
Triggering and time base
All controls regarding trigger and time base are located on the
right of the VOLTS/DIV.-knobs.
”Controls and Readout”.
Time related amplitude changes on a measuring signal (AC
voltage) are displayable in Yt mode. In this mode the signal
voltage deflects the beam in vertical direction (Y) while the time
They are described in the section
Triggering can be performed by the measuring signal itself
(internal triggering) or by an external supplied but synchronous
voltage (external triggering).
The trigger voltage should have a certain minimum amplitude.
This value is called the trigger threshold. It is measured with a
sine signal. Except when external trigger is used the trigger
threshold can be stated as vertical display height in div, at which
the time base generator starts, the display is stable, and the
trigger indicator LED lights or flashes.
The internal trigger threshold of the oscilloscope is given as £
0.5div. When the trigger voltage is externally supplied, it can be
measured in Vpp at that input. Normally, the trigger threshold
may be exceeded up to a maximum factor of 20.
The instrument has two trigger modes, which are characterized
as Automatic Peak and Normal triggering.
Automatic Peak (Value) Triggering
Instrument specific information can be drawn from the items NM
- AT - [9], LEVEL [11] and TRIG. MODE [20] in the section
”Controls and Readout”.
This trigger mode is automatically selected after the AUTO SET
pushbutton is pressed. As the peak value detection makes no
sense in combination with DC and TV (television) signals, it is
switched off automatically in DC, TVL and TVF trigger coupling
conditions as well as in alternate trigger mode. In this case the
automatic is still present, but a wrong trigger level setting causes
an untriggered display.
In automatic trigger mode the sweep generator can run without
an input signal or external trigger voltage. A base line will always
be displayed even with no signal. With an applied AC signal, peak
value triggering enables the user to select the trigger point on the
displayed signal, by the adjustment of the trigger level control.
The control range depends on the peak to peak value of the signal.
This trigger mode is therefore called Automatic Peak (Value)
Triggering.
Operation of the scope needs only correct amplitude and time
base settings, for a constantly visible trace. Automatic mode is
recommended for all uncomplicated measuring tasks. However,
automatic triggering is also the appropriate operation mode for
the ”entry” into difficult measuring problems, e.g. when the test
signal is unknown relating to amplitude, frequency or shape.
Presetting of all parameters is now possible with automatic
triggering; the change to normal triggering can follow thereafter.
The automatic triggering works above 20Hz. The failure of
automatic triggering at frequencies below 20Hz is abrupt.
However, it is not signified by the trigger indicator LED which may
still be blinking. Break down of triggering is best recognizable at
the left screen edge (the start of the trace in differing display
height).
Subject to change without notice
27
Page 28
Triggering and time base
The automatic peak (value) triggering operates over all variations
or fluctuations of the test signal above 20 Hz. However, if the
pulse duty factor of a square wave signal exceeds a ratio of
100 : 1, switching over to normal triggering will be necessary.
Automatic triggering is practicable with internal and external
trigger voltage.
Normal Triggering
Information specific to the instrument is given in the sections NM
- AT - [9], LEVEL [11] and TRIG. MODE [20] in the paragraphs
”Controls and Readout”. The time fine adjustment (VAR.), and
the holdoff time setting assist in triggering under specially difficult
signal conditions.
With normal triggering, the sweep can be started by AC signals
within the frequency range defined by the trigger coupling setting.
In the absence of an adequate trigger signal or when the trigger
controls (particularly the trigger LEVEL control) are misadjusted,
no trace is visible.
When using the internal normal triggering mode, it is possible to
trigger at any amplitude point of a signal edge, even with very
complex signal shapes, by adjusting the trigger LEVEL control. If
the signal applied at the Y input is used for triggering (internal
trigger source), its adjusting range is directly dependent on the
display height, which should be at least 0.5div. If it is smaller than
1div, the trigger LEVEL adjustment needs to be operated with a
sensitive touch. In the external normal triggering mode, the same
applies to approx. 0.3 V
Other measures for triggering of very complex signals are the use
of the time base variable control and HOLDOFF time control,
mentioned below.
external trigger voltage amplitude.
pp
As the automatic triggering does not work below 20 Hz, normal
triggering should be used in DC and LF trigger coupling mode. The
coupling mode and accordingly the frequency range of the trigger
signal should meet the signal requirements.
AC: This is the most frequently used trigger mode. The trigger
threshold increases below and above the frequency limits
mentioned in the data sheet. This filter cuts off both the DC
content of the trigger signal and the lowest frequency range.
DC: In this coupling mode the trigger signal is coupled galvanically
to the trigger unit if normal triggering (NM) is present.
Therefore there is no low frequency limit.
DC triggering is recommended if the signal is to be triggered
with quite slow processes or if pulse signals with constantly
changing pulse duty factors have to be displayed.
HF: In this coupling mode the transmission range equals a high
pass filter. It cuts off the DC content of the trigger signal and
the lower frequency range.
LF: LF trigger coupling has a low pass filter function characteristic.
As in DC trigger coupling, there is no limit for the pass
frequency range in connection with normal triggering.
The LF trigger coupling is often more suitable for low frequency
signals than DC trigger coupling because the noise
components of the trigger signals are strongly suppressed.
This avoids or reduces, under borderline conditions, jitter or
double traces especially with very low signal voltages. The
trigger threshold rises continuously above the pass band.
TVL: The built in active TV Sync Separator provides the separation
of line sync pulses from the video signal.
SLOPE
Please note item [9] in section ”Controls and Readout” for
instrument specific information.
The actual slope setting is displayed in the readout. The setting
is not changed by the AUTO SET function. The slope setting can
be changed for the delay time base trigger unit in delay mode if
the delay trigger function is active. The previous slope setting for
the undelayed time base trigger is stored and still active. For
further information please note ”Controls and Readout”.
The time base generator can be triggered by a rising or falling
edge of the test signal. Whether the rising or the falling edge is
used for triggering, depends on the slope direction setting. This
is valid with automatic and normal triggering. The positive slope
direction means an edge going from a negative potential and
rising to a positive potential. This has nothing to do with zero or
ground potential or absolute voltage values. The positive slope
may also lie in a negative part of a signal.
However the trigger point may be varied within certain limits on
the chosen edge using the LEVEL control. The slope direction is
always related to the input signal and the non inverted display.
Trigger coupling
Instrument specific information regarding this item can be noted
in the ”Data Sheet”. The coupling setting TRIG. MODE [20] and
indication are described under ”Controls and Readout”.
Even distorted video signals are triggered and displayed in a
stable manner. This mode is described under paragraph
”Triggering of video signals”.
TVF: The built in active TV Sync Separator also provides the
separation of frame sync pulses from the video signal. Even
distorted video signals are triggered and displayed in a stable
manner.
This mode is described under paragraph ”Triggering
of video signals”.
~: Indicates “line/mains triggering” and is described under the
paragraph of the same name.
Triggering of video signals
In TVL and TVF trigger coupling mode the instrument is
automatically set to automatic triggering and the trigger point
indicator is switched off. As only the separated synchronization
pulses are used for triggering the relationship between the
displayed signal and the trigger signal is lost. In TVF mode
interference may occur if chopped DUAL mode is chosen or the
readout is active.
Video signals are triggered in the automatic mode. The internal
triggering is virtually independent of the display height, but the
sync pulse must exceed 0.5div height.
The polarity of the synchronization pulse is critical for the slope
selection. If the displayed sync pulses are above the picture (field)
contents (leading edge positive going), then the slope setting for
28
Subject to change without notice
Page 29
Triggering and time base
positive going edges must be chosen. In the case of sync pulses
below the field/line, the leading edge is negative and consequently
the slope selection must be set for falling edges. Since the invert
function may cause a misleading display, it must not be activated.
On the 2 ms/div setting and field TV triggering selected, 1 field is
visible if a 50 fields/s signal is applied. If the holdoff control is in
fully ccw position, it triggers without line interlacing affects
caused by the consecutive field.
The display can be expanded by switching on the X-MAG. x10
function so that individual lines are recognizable. Commencing
with a frame synchronizing pulse, the display can also be expanded
with the knob TIME/DIV. But note that this can result in an
apparently unsynchronized display as each frame (half picture)
triggers. This is due to the off set of half a line between frames.
The influence of the integrating network which forms a trigger
pulse from the vertical sync pulses may become visible under
certain conditions. Due to the integrating network time constant
not all vertical sync pulses starting the trace are visible.
On the 10µs/div setting and line TV triggering selected, approx.
1½ lines are visible. Those lines originate from the odd and even
fields at random.
The sync-separator-circuit also operates with external triggering.
It is important that the voltage range (0.3 V
to 3 Vpp) for external
pp
triggering should be noted. Again the correct slope setting is
critical, because the external trigger signal may not have the
same polarity or pulse edge as the test signal displayed on the
CRT. This can be checked, if the external trigger voltage itself is
displayed first (with internal triggering).
In most cases, the composite video signal has a high DC content.
With constant video information (e.g. test pattern or colour bar
generator), the DC content can be suppressed easily by AC input
coupling of the oscilloscope amplifier. With a changing picture
content (e.g. normal program), DC input coupling is recommended,
because the display varies its vertical position on screen with AC
input coupling at each change of the picture content. The DC
content can be compensated using the Y-POS. control so that the
signal display lies in the graticule area. Then the composite video
signal should not exceed a vertical height of 6div.
In this trigger mode the slope direction pushbutton selects the
positive or negative portion of the line/mains sine wave. The
trigger level control can be used for trigger point adjustment.
Magnetic leakage (e.g. from a power transformer) can be
investigated for direction and amplitude using a search or pick up
coil. The coil should be wound on a small former with a maximum
of turns of a thin lacquered wire and connected to a BNC
connector (for scope input) via a shielded cable. Between cable
and BNC centre conductor a resistor of at least 100 Ohm should
be series connected (RF decoupling). Often it is advisable to
statically shield the surface of the coil. However, no shorted turns
are permissible. Maximum, minimum, and direction to the
magnetic source are detectable at the measuring point by turning
and shifting the coil.
Alternate triggering
This trigger mode can be selected in DUAL mode by the aid of the
TRIG. SOURCE [17] pushbutton (
Readout”
). In the case of chopped DUAL mode, selecting
alternate trigger mode automatically sets the instrument to
alternate DUAL mode.
Under TVL, TVF and line/mains triggering conditions alternate
triggering can not be chosen. Thus only the following trigger
coupling modes are available in alternate trigger mode: AC, DC,HF and LF. The trigger point symbol is not displayed in alternate
trigger mode.
With alternate triggering it is possible to trigger two signals from
different frequency sources (asynchronous). In this case the
oscilloscope must be operated in DUAL alternate mode with
internal triggering and each input signal must be of sufficient
height to enable trigger. To avoid trigger problems due to different DC voltage components, AC input coupling for both channels
is recommended.
The internal trigger source is switched in alternate trigger mode
in the same way as the channel switching system in DUAL
alternate mode, i.e. after each time base sweep. Phase difference
measurement is not possible in this trigger mode as the trigger
level and slope setting are equal for both signals. Even with 180°
phase difference between both signals, they appear with the
same slope direction.
please note ”Controls and
Line/Mains triggering (~)
The instrument specific information regarding this mode is part of
the section ”Controls and Readout” paragraph TRIG. MODE[20].
This trigger mode is present if the READOUT indicates the “~”
symbol instead of the “trigger source”, “slope” and “coupling”
information. The trigger point symbol is inactive in line/mains
trigger mode as there is no direct amplitude relationship between
the trigger voltage and the signal voltage.
A voltage originating from mains/line (50 to 60 Hz) is used for
triggering purposes if the trigger coupling is set to ~. This trigger
mode is independent of amplitude and frequency of the Y signal
and is recommended for all mains/line synchronous signals. This
also applies within certain limits, to whole number multiples or
fractions of the line frequency. Line triggering can also be useful
to display signals below the trigger threshold (less than 0.5 div).
It is therefore particularly suitable for measuring small ripple
voltages of mains/line rectifiers or stray magnetic field in a circuit.
Subject to change without notice
If signals are applied with a high frequency ratio (difference), the
trace intensity then becomes reduced if the time base is set to
smaller time coefficients (faster sweep). This happens as the
number of sweeps does not increase because it depends on the
lower frequency signal, but with a faster sweep the phosphor
becomes less activated.
External triggering
The external trigger input is activated with the aid of the TRIG.
SOURCE [17] pushbutton (see ”Controls and Readout”), if the
trigger coupling is not set to line/mains trigger coupling. Then the
internal trigger source is deactivated. As the external trigger
signal applied at the TRIG. EXT socket normally has no relation
to the signal height of the displayed signal, the trigger point
symbol is switched off. The external trigger voltage must have a
minimum amplitude of 0.3 V
. The input impedance of the TRIG. EXT. socket is approx.
3V
pp
and should not increase above
pp
1 MOhm II 20 pF.
The maximum input voltage of the input circuit is 100V (DC+peak
AC). The external trigger voltage may have a completely different
29
Page 30
Triggering and time base
form from the test signal voltage, but must be synchronous with
the test signal. Triggering is even possible in certain limits with
whole number multiples or fractions of the test frequency.
It must be noted that a different phase angle between the
measuring and the triggering signal may cause a display not
coinciding with the slope selection setting.
The trigger coupling selection can also be used in external
triggering mode.
Trigger indicator ”TR”
The following description applies to the ”TR” LED.
[10]
item
An LED on condition indicates that the trigger signal has a
sufficient amplitude and the trigger level control setting is correct.
This is valid with automatic and with normal triggering. By
observing the trigger LED, sensitive trigger level adjustment is
possible when normal triggering is used, particularly at very low
signal frequencies. The indication pulses are of only 100ms
duration. Thus for fast signals the LED appears to glow
continuously, for low repetition rate signals, the LED flashes at
the repetition rate or at a display of several signal periods not only
at the start of the sweep at the left screen edge, but also at each
signal period.
In automatic triggering mode the sweep generator starts
repeatedly without test signal or external trigger voltage. If the
trigger signal frequency decreases the sweep generator starts
without awaiting the trigger pulse. This causes an untriggered
display and a flashing trigger LED.
under ”Controls and Readout”.
Please note
HOLDOFF time adjustment
For instrument specific information please note DEL.POS. / HO
LED [21] in section ”Controls and Readout”.
If it is found that a trigger point cannot be found on extremely
complex signals, even after careful adjustment of the trigger level
control, a stable display may often be obtained using the holdoff
control. This facility varies the holdoff time between two sweep
periods approx. up to the ratio 10:1. Pulses or other signal
waveforms appearing during this off period cannot trigger the
time base.
Particularly with burst signals or aperiodic pulse trains of the
same amplitude, the start of the sweep can be delayed until the
optimum or required time.
A very noisy signal or a signal with a higher interfering frequency
is at times displayed double. It is possible that trigger level
adjustment only controls the mutual phase shift, but not the
double display. The stable single display of the signal, required for
evaluation, is easily obtainable by expanding the holdoff time until
one signal is displayed.
A double display is possible with certain pulse signals, where the
pulses alternately show a small difference of the peak amplitudes.
Only a very exact trigger level adjustment makes a single display
possible. The use of the holdoff control simplifies the right
adjustment.
After specific use the holdoff control should be reset into its
calibration detent (fully ccw), otherwise the brightness of the
display may be reduced drastically. The function is shown in the
following figures.
Fig. 1 shows a case where the holdoff control is in the minimum
position and various different waveforms are overlapped on the
screen, making the signal observation unsuccessful.
Fig. 2 shows a case where only the desired parts of the signal are
stably displayed.
Delay / After Delay Triggering
The instrument specific information regarding this mode is part of
the section ”Controls and Readout” paragraph DEL.POS. / HO
LED [21] and DEL.MODE / ON OFF [23].
As mentioned before, triggering starts the time base sweep and
unblanks the beam. After the maximum X deflection to the right,
the beam is blanked and flies back to the (left) start position. After
the holdoff period the sweep is started automatically by the
automatic trigger or the next trigger signal. In normal triggering
mode the automatic trigger is switched off and will only start on
receipt of a trigger signal.
As the trigger point is always at the trace start position, trace
expansion in X direction with the aid of the time base is limited to
the display on the left of the trace. Parts of the signal to be
expanded which are displayed near the trace end (right side of the
screen) are lost when the time base speed is increased (time
coefficient reduced).
The delay function delays the trace start by a variable time from
the trigger point. This allows the sweep to begin on any portion
of a signal. The time base speed can then be increased to expand
the display in X direction. With higher expansion rates, the
intensity reduces and within certain limits this can be compensated
by a higher intensity (INTENS) setting.
If the display shows jitter, it is possible to select for (second)
triggering after the elapsed delay time (“dTr”). As mentioned
before, it is possible to display video signals using the frame sync
pulses for triggering (TVF). After the delay time set by the
operator, the next line sync pulse or the line content may be used
for triggering. So data lines and test lines can be displayed
separately.
Operation of the delay function is relatively simple. Without delay
function set the time coefficient setting (TIME/DIV) until 1 to 3
signal periods are displayed. Display of less than two periods
should be avoided as it limits the selection of the signal section
to be expanded.
The X MAG (x10) function should be switched off in the beginning
but may be activated later. The signal must be triggered and
stable.
30
Subject to change without notice
Page 31
Triggering and time base
The following explanation assumes that the trace starts on the
left vertical graticule line.
Switching over from undelayed to delayed time base automatically
sets the holdoff time to minimum so that the HO LED extinguishes,
the DEL. POS. knob function changes from holdoff time to delay
time control and the READOUT indicates “sea”.
In search (“sea”) mode a part of the previously complete visible
trace becomes blank. The length of the blanked sector depends
on the delay time (DEL. POS.) setting and can be set between
approx. two and seven divisions after the normal trace start
position. Consequently the trace is displayed with reduced length.
If the maximum delay time is not sufficient, the time coefficient
must be increased (TIME/DIV-knob) and the DEL. POS.-knob set
to the later starting point.
Note:
Actually the trace start is not really delayed in “sea” (search)
condition, as the blank sector serves only as an adjusting indicator
making visible the delay time which will be active after selecting
“del” (delay time base) or “dTr” (delay time base in triggered
condition).
Photo 2
Please note that the previous time coefficient chosen in “del”
and “dTr” mode is stored and automatically set after activating
one of those modes. If the stored time coefficient in “del” /”dTr” mode was higher than the actual value in “sea” (search)
mode, the time coefficient in “del” / ”dTr” mode is automatically
set to the value used during “sea” (search) operation.
Reducing the time coefficient (increasing the time base speed)
now expands the signal. If the signal start position is not set to the
optimum, it can still be shifted in the X direction by changing the
delay time.
Photo 4 shows a 50 fold X magnification caused by setting the
time coefficient to 0.1 ms/div (5 ms/div : 0.1 ms/div = 50). The
reading accuracy also increases with higher X magnification.
Figure 2 shows that the delay time can be measured. It is identical
with the displacement of the start of the trace. One can calculate
this by multiplying the blanked out section (horizontal) by the time
deflection coefficient setting.
The full length trace will be visible when switched from “sea”
(SEARCH) to “del” (DELAY), starting with the section previously
selected, providing the (stored) current time deflection coefficient
is not too small.
If the trace is invisible or hardly visible because of too much
expansion (too small deflection coefficient), the time deflection
coefficient must be increased with TIME / DIV knob. A larger
deflection coefficient than in the “sea” (SEARCH) mode cannot
be set.
Example: The SEARCH setting selected in figure 2 is 5 ms/cm.
The display in “del” (DELAY) mode, also with 5ms/div is delayed
but unexpanded (1:1). A further increase in the deflection
coefficient, e.g. 10ms/div would be meaningless and therefore
automatically blocked.
The delayed and expanded signal display can be triggered again
if a signal slope suitable for triggering appears after the delay
time. For this, one must switch to “dTr” (2nd triggering after the
expiry of the delay time - after Delay Triggering). The settings
selected before switching, automatic Peak value triggering /
Normal triggering, trigger coupling, the trigger LEVEL setting and
slope setting, remain valid and trigger the start of the delay time.
The ”After Delay” Triggering automatically switches to normal
triggering (indicated by the NM LED) and DC trigger coupling.
These default conditions cannot be changed. But the trigger level
(LEVEL) and the trigger slope direction can be altered in order to
enable the triggering at the desired signal section. The trace does
not start and the screen remains blank if the signal amplitude is
not sufficient for triggering or if the setting of the trigger LEVEL
is unsuitable.
The expanded display can also be displaced in the X direction by
changing the delay time (DEL.POS.) under suitable settings.
However, the displacement is not continuous as in the untriggered
“del” (DELAY) operation but jumps from one trigger slope to
another - with most signals this is not evident. This means, in the
case of TV Triggering, that it is possible to trigger not only with
line synchronizing pulses but also on suitable slopes occurring
within the line.
Of course, the magnification is not restricted to a factor 50 as
mentioned in the example. The limit is given by the increasing
loss of trace intensity as the magnification is increased.
Subject to change without notice
31
Page 32
AUTOSET
The manipulation of time delay requires a certain experience,
especially with complicated signal combinations which are
difficult to display. The display of sections of simple signals is, in
contrast, fairly easy. The time delayed display is also possible in
the dual channel, addition and difference modes.
In chopped DUAL mode, if after switching over to “del” or “dTr”,
the time deflection coefficient is reduced (TIME/DIV.), the channel
switching mode doesn`t change automatically to alternate.
Attention:
In chopped DUAL mode, using high expansion ratios in “del”
mode, chop interference may be visible. This can be overcome
by selecting alternate DUAL mode. A similar effect can be
caused by the READOUT with the result that parts of a signal
displayed in CH I, CH II or DUAL mode are blanked
(unsynchronised). In such a case the READOUT can be
switched off.
AUTOSET
The instrument specific information regarding this function is part
of the section ”Controls and Readout” paragraph AUTOSET [2].
As also mentioned in that section, all controls are electronically
selected with the exception of the POWER pushbutton.
Thus automatic, signal related instrument set up in Yt (time base)
mode is possible. In most cases no additional manual instrument
setting is required.
I X x10 magnifier switched off
I Optimum X and Y position settings
I Trace and readout visible.
If DC trigger coupling had been selected, AC trigger coupling will
not be chosen and the automatic trigger is operative without the
peak value detection.
The X position is set to the CRT centre as well as the Y position
in CH I or CH II mode. In DUAL mode the channel I trace is set to
the upper half and the channel II trace to the lower half of the CRT.
The 1 mV/div and 2 mV/div deflection coefficient will not be
selected by AUTOSET as the bandwidth is reduced on these
settings.
Attention!
If a signal is applied with a pulse duty factor of approx. 400 :1
or larger, an automatic signal display can not be performed.
The pulse duty factor causes too low a Y deflection coefficient
(sensitivity too high) and too high a time deflection coefficient
(time base speed to slow) and results in a display in which
only the baseline is visible.
In such cases it is recommended to select normal triggering and
to set the trigger point approx. 0.5div above or below the trace.
If under one of these conditions the trigger indicator LED is lit, this
indicates the presence of a signal. Then both the time coefficient
and Y deflection coefficient should be reduced. Please note that
a reduction in intensity may occur, which could result in a blank
screen when the physical limits are reached.
Briefly pressing the AUTOSET pushbutton causes the instrument
to switch over to the last Yt mode settings regarding CH I, CH II
and DUAL. If the instrument was operated in Yt mode, the actual
setting will not be affected with the exception of ADD mode
which will be switched off. At the same time the attenuator(s)
(VOLTS/DIV) are automatically set for a signal display height of
approx. 6 div in mono channel mode or if in DUAL mode for
approx. 4 div height for each channel. In the determination of the
time deflection coefficient, it is assumed that the pulse duty
factor of the input signal is approx. 1:1.
The time deflection coefficient is also set automatically for a
display of approx. 2 signal periods. The time base setting occurs
randomly if complex signals consisting several frequencies e.g.
video signals are present. If cursor voltage measurement is
selected, AUTOSET also affects the position of the CURSOR
Please note AUTOSET
lines.
[2]
in section ”Controls and
Readout”.
AUTOSET sets the instrument automatically to the following
operating conditions:
I AC or DC input coupling unaltered
or in GND condition the last used setting
I Internal triggering (channel I or channel II)
I Automatic triggering
I Trigger level in electrical midrange position
I Optimum calibrated Y deflection coefficient(s) 5 mV - 20 mv/div
I Optimum calibrated Time base deflection coefficient
I AC trigger coupling (except if DC trigger coupling last present
I Undelayed time base mode
Mean Value Display
The DC Mean Value is displayed in place of the cursor line
measurement, if the cursor lines are switched off, the AUTO
MEASURE menu function “DC” is activated and further condition
are met:
The signal to be measured (AC > 20 Hz) must be applied at input
CH I [25] or CH II [28] with its DC content at the measuring
amplifier; DC input coupling [26; 29] required. Yt (time base) mode
in combination with internal triggering (trigger source CH I or
CH II; not alternated triggering) must be present. AC or DC trigger
coupling must be selected.
If the above conditions are not met, ”n/a” will be displayed.
The mean value is acquired using the trigger signal amplifiers
used for internal triggering. With the exception of DUAL mode,
the indicated mean value is automatically related to the active
channel (CH I or CH II), as the channel selection also assigns the
trigger amplifier. In DUAL mode one can select between trigger
amplifier CH I or CH II for triggering. The indicated mean value
refers to the channel from which the trigger signal originates.
The DC mean value is displayed with an algebraic sign (e.g.
”dc:Y1 501mV” resp. ”dc:Y1 -501mV). Overranging is indicated
by ” < ” resp. ” > ” sign (e.g. ”dc:Y1 <1.80V” resp. ”dc:Y1
>1.80V”). Being dependent on a necessary time constant for
mean value creation, the display update requires a few seconds
after a voltage change.
The reading accuracy is dependent on the instrument
specifications (Y deflection tolerance max. 3% from 5 mV/div. to
20 V/div.). Although the tolerances are significantly smaller in
32
Subject to change without notice
Page 33
Component Tester
reality, other deviations such as unavoidable offset voltages must
be taken into account, which may cause a display deviating from
0 Volt without signal applied at the input.
The display shows the arithmetic (linear) mean value. The DC
content is displayed if DC or AC superimposed DC voltages are
applied. In case of square wave voltages, the mean value depends
on the pulse duty factor.
Component Tester
General
The instrument specific information regarding the control
and terminals are part of item [37] in section ”Controls and
Readout”.
The instrument has a built in electronic Component Tester, which
is used for instant display of a test pattern to indicate whether or
not components are faulty. It can be used for quick checks of
semiconductors (e.g. diodes and transistors), resistors, capacitors,
and inductors. Certain tests can also be made to integrated
circuits. All these components can be tested individually, or in
circuit provided that it is unpowered.
The test principle is fascinatingly simple. A built in generator
provides a sine voltage, which is applied across the component
under test and a built in fixed resistor. The sine voltage across the
test object is used for the horizontal deflection, and the voltage
drop across the resistor (i.e. current through test object) is used
for Y deflection of the oscilloscope. The test pattern shows the
current/voltage characteristic of the test object.
The measurement range of the component tester is limited and
depends on the maximum test voltage and current (please note
data sheet). The impedance of the component under test is
limited to a range from approx. 20 Ohm to 4.7 kOhm. Below and
above these values, the test pattern shows only short circuit or
open circuit. For the interpretation of the displayed test pattern,
these limits should always be born in mind. However, most
electronic components can normally be tested without any
restriction.
Using the Component Tester
After the component tester is switched on, the Y amplifier and
the time base generator are inoperative. A shortened horizontal
trace will be observed. It is not necessary to disconnect scope
input cables unless in circuit measurements are to be carried out.
For the component connection, two simple test leads with 4 mm
Ø banana plugs, and test prods, alligator clips or sprung hooks,
are required.
”Controls and Readout”.
Test Procedure
Caution!
Do not test any component in live circuitry, remove all
grounds, power and signals connected to the component
under test. Set up Component Tester as stated. Connect test
leads across component to be tested. Observe oscilloscope
display. –
The test leads are connected as described in section
Only discharged capacitors should be tested!
Test Pattern Displays
The following ”Test patterns” show typical patterns displayed by
the various components under test.
I Open circuit is indicated by a straight horizontal line.
I Short circuit is shown by a straight vertical line.
Testing Resistors
If the test object has a linear ohmic resistance, both deflecting
voltages are in the same phase. The test pattern expected from
a resistor is therefore a sloping straight line. The angle of slope is
determined by the value of the resistor under test. With high
values of resistance, the slope will tend towards the horizontal
axis, and with low values, the slope will move towards the vertical
axis. Values of resistance from 20 Ohm to 4.7 kOhm can be
approximately evaluated. The determination of actual values will
come with experience, or by direct comparison with a component
of known value.
Testing Capacitors and Inductors
Capacitors and inductors cause a phase difference between
current and voltage, and therefore between the X and Y deflection,
giving an ellipse shaped display. The position and opening width
of the ellipse will vary according to the impedance value (at 50Hz)
of the component under test.
I A horizontal ellipse indicates a high impedance or a relatively
small capacitance or a relatively high inductance.
I A vertical ellipse indicates a low impedance or a relatively
large capacitance or a relatively small inductance.
I A sloping ellipse means that the component has a considerable
ohmic resistance in addition to its reactance.
The values of capacitance of normal or electrolytic capacitors
from 0.1µF to 1000µF can be displayed and approximate values
obtained. More precise measurement can be obtained in a
smaller range by comparing the capacitor under test with a
capacitor of known value. Inductive components (coils,
transformers) can also be tested. The determination of the value
of inductance needs some experience, because inductors have
usually a higher ohmic series resistance. However, the impedance
value (at 50 Hz) of an inductor in the range from 20 Ohm to
4.7 kOhm can easily be obtained or compared.
Testing Semiconductors
Most semiconductor devices, such as diodes, Z-diodes, transistors
and FETs can be tested. The test pattern displays vary according
to the component type as shown in the figures below. The main
characteristic displayed during semiconductor testing is the
voltage dependent knee caused by the junction changing from
the conducting state to the non conducting state. It should be
noted that both the forward and reverse characteristic are displayed
simultaneously. This is a two terminal test, therefore testing of
transistor amplification is not possible, but testing of a single
junction is easily and quickly possible. Since the test voltage
applied is only very low, all sections of most semiconductors can
be tested without damage. However, checking the breakdown or
reverse voltage of high voltage semiconductors is not possible.
More important is testing components for open or short circuit,
which from experience is most frequently needed.
Subject to change without notice
33
Page 34
AUTOSET
Testing Diodes
Diodes normally show at least their knee in the forward
characteristic. This is not valid for some high voltage diode types,
because they contain a series connection of several diodes.
Possibly only a small portion of the knee is visible. Zener diodes
always show their forward knee and, depending on the test
voltage, their zener breakdown forms a second knee in the
opposite direction. If the breakdown voltage is higher than the
positive or negative voltage peak of the test voltage, it can not be
displayed.
The polarity of an unknown diode can be identified by comparison
with a known diode.
Testing Transistors
Three different tests can be made to transistors: base-emitter,
base-collector and emitter-collector. The resulting test patterns
are shown below. The basic equivalent circuit of a transistor is a
Z-diode between base and emitter and a normal diode with
reverse polarity between base and collector in series connection.
There are three different test patterns:
In-Circuit Tests
Caution!
During in circuit tests make sure the circuit is dead. No power
from mains/line or battery and no signal inputs are permitted.
Remove all ground connections including Safety Earth (pull
out power plug from outlet). Remove all measuring cables
including probes between oscilloscope and circuit under
test. Otherwise both COMPONENT TESTER leads are not
isolated against the circuit under test.
In-Circuit tests are possible in many cases. However, they are
not well defined. Complex displays may be caused by a shunt
connection of real or complex impedance, especially if they
are of relatively low impedance at 50Hz, to the component
under test, often results differ greatly when compared with
single components. In case of doubt, one component terminal
should be unsoldered. This terminal should then not be
connected to the ground socket avoiding hum distortion of the
test pattern.
Another way is a test pattern comparison to an identical circuit
which is known to be operational (likewise without power and any
external connections). Using the test prods, identical test points
in each circuit can be checked, and a defect can be determined
quickly and easily. Possibly the device under test itself may
contain a reference circuit (e.g. a second stereo channel, pushpull amplifier, symmetrical bridge circuit), which is not defective
and can therefore be used for comparison.
For a transistor the figures b-e and b-c are important. The figure
e-c can vary; but a vertical line only shows short circuit condition.
These transistor test patterns are valid in most cases, but there
are exceptions to the rule (e.g. Darlington, FETs). With the
COMPONENT TESTER, the distinction between a P-N-P to an NP-N transistor is discernible. In case of doubt, comparison with a
known type is helpful. It should be noted that the same socket
connection (CT or ground) for the same terminal is then absolutely
necessary. A connection inversion effects a rotation of the test
pattern by 180 degrees about the centre point of the scope
graticule.
Pay attention to the usual caution with single MOS
components relating to static discharge or frictional
electricity!
34
Subject to change without notice
Page 35
RS-232 Interface – Remote Control
Adjustments
After calling MAIN MENU > ADJUSTMENT > AUTO
ADJUSTMENT, several menu items are displayed. Each item can
be called and causes an automatic adjustment.
All items are subject to the instrument’s temperature response
under extreme environmental temperature conditions and results
are stored in a non volatile memory. Incorrect adjustment settings
can be caused by component failures as a result of the application
of excessive voltage inputs and therefore cannot be compensated
by the automatic adjustment procedure.
Before starting an automatic adjustment procedure a warm up
time of 20 minutes must be allowed. During these automatic
adjustments there must be no signal applied to any input.
The following items are available:
1. SWEEP START POSITIONS
In Yt (time base) mode the trace start position is affected by time
base setting. The automatic adjustment minimises such effects.
During execution the readout indicates “WORKING”.
2. Y AMP (measuring amplifier CH I and CH II)
Different Y deflection coefficient settings cause minor Y position
changes. Changes higher than ± 0.2 div (5 mV/div to 20 V/div)
become corrected. This value relates on open but screened
inputs.
The automatic adjustment affects both channels. After execution
the readout displays the AUTO ADJUSTMENT MENU.
3. TRIGGER AMP
This adjustment reduces trigger amplifiers dc offset to a minimum.
After completion the AUTO ADJUSTMENT MENU becomes
visible again.
4. X MAG POS
This adjustment coordinates the X-POS control setting range in
unmagnified and magnified (X-MAG. x10) condition.
5. CT X-POS
This adjustment adapts the setting range of the X-POS control
setting in “Component Tester” and Yt (X-MAG. x1) mode.
Operation
The oscilloscope is supplied with a serial interface for control
purposes. The interface connector (9 pole D SUB female) is
located on the rear of the instrument. Via this bidirectional port,
the instrument parameter settings can be transmitted to, or
received from a PC.
RS-232 Cable
The maximum connecting cable length must be less then 3
meters and must contain 9 screened lines connected 1:1. The
oscilloscope RS-232 connection (9 pole D SUB female) is
determined as follows:
Pin
2 Tx data (data from oscilloscope to external device)
3 Rx data (data from external device to oscilloscope)
7 CTS (clear to send)
8 RTS (request to send)
5 Ground (reference potential - connected via the
oscilloscope’s power cord with protective earth)
9 +5 V supply for external device (max. 400 mA).
The maximum voltage swing at pin 2, 3, 7 and 8 is ± 12 Volt.
RS-232 protocol
N-8-2 (no parity bit, 8 data bits, 2 stop bits, RTS/CTS hardware
protocol).
Baud-Rate Setting
After the first POWER UP (switching on of the oscilloscope ) and
the first command SPACE CR (20 hex, 0 Dhex) sent from the PC,
the baud rate is recognized and set automatically between 110
baud and 115200 baud. The oscilloscope is then switched over to
REMOTE control mode. The oscilloscope then transmits the
RETURNCODE: 0 CR LF to the PC. In this status all settings (with
the exception of those functions mentioned under ”Controls and
Readout”) can be controlled via the interface only.
The only ways to quit this status are:
I Switching the oscilloscope off,
or transmitting the command
I RM= 0 from the PC to the oscilloscope, or
I depressing the AUTOSET (LOCAL) pushbutton,
if in unlocked condition (command LK=1... was not sent)
RS-232 Interface - Remote Control
Safety
Caution:
All terminals of the RS-232 interface are galvanically
connected with the oscilloscope and subsequently with
protective (safety) earth potential.
Measurement on a high level reference potential is not permitted
and endangers operator, oscilloscope, interface and peripheral
devices.
In case of disregard of the safety warnings contained in this
manual, HAMEG refuses any liability regarding personal injury
and/or damage of equipment.
Subject to change without notice
After the remote state has been switched off the RM LED is dark.
Please note:
A minimum time must elapse between the commands RM=1...
(remote on) and RM=0... (remote off) and vice versa. The time
can be calculated with the formula:
t
= 2x(1/baud rate) + 60µs.
min
If at the beginning no SPACE CR command is recognizable, the
oscilloscope pulls the TxD line low for approx. 0.2ms and causes
a break on the PC.
Data Communication
After successfully being set to remote control mode, the
oscilloscope is prepared for command reception.
A CompactDisc with programming examples, a command list
and a program executable under Windows 95, 98, Me, 2000, NT
4.0 (with Service Pack 4 or higher) and XP is part of the delivery.