EN 61000-4-4: 1995 / IEC (CEI) 1000-4-4: 1995 / VDE 0847 T4-4:
Prüfschärfe / Level / Niveau = 3
EN 50081-1: 1992 / EN 55011: 1991 / CISPR11: 1991 / VDE0875 T11: 1992
Gruppe / group / groupe = 1, Klasse / Class / Classe = B
Datum /Date /DateUnterschrift / Signature /Signatur
02.03.1998
Dr. J. Herzog
Technical Manager/Directeur Technique
®
HAMEG instruments fulfill the regulations of the EMC directive. The conformity test made by HAMEG is based
on the actual generic- and product standards. In cases where different limit values are applicable, HAMEG
applies the severer standard. For emission the limits for residential, commercial and light industry are applied.
Regarding the immunity (susceptibility) the limits for industrial environment have been used.
The measuring- and data lines of the instrument have much influence on emmission and immunity and therefore
on meeting the acceptance limits. For different applications the lines and/or cables used may be different. For
measurement operation the following hints and conditions regarding emission and immunity should be observed:
1. Data cables
For the connection between instruments resp. their interfaces and external devices, (computer, printer etc.)
sufficiently screened cables must be used. Without a special instruction in the manual for a reduced cable
length, the maximum cable length of a dataline must be less than 3 meters long. If an interface has several
connectors only one connector must have a connection to a cable.
Basically interconnections must have a double screening. For IEEE-bus purposes the double screened cables
HZ72S and HZ72L from HAMEG are suitable.
2. Signal cables
Basically test leads for signal interconnection between test point and instrument should be as short as possible.
Without instruction in the manual for a shorter length, signal lines must be less than 3 meters long.
Signal lines must screened (coaxial cable - RG58/U). A proper ground connection is required. In combination with
signal generators double screened cables (RG223/U, RG214/U) must be used.
3. Influence on measuring instruments.
Under the presence of strong high frequency electric or magnetic fields, even with careful setup of the measuring
equipment an influence of such signals is unavoidable.
This will not cause damage or put the instrument out of operation. Small deviations of the measuring value
(reading) exceeding the instruments specifications may result from such conditions in individual cases.
HAMEG GmbH
2
Subject to change without notice
Specifications
Vertical Deflection
Operating modes: Channel I or CH II separate,
Channel I and II: alternate or chopped
(Chopper Frequency approx. 0.5MHz)
Sum or Difference from Channel I and ± Ch. II,
XY-Mode: via CH I (X) and CH II (Y).
Frequency range: 2x DC to 40MHz (−3dB).
Risetime: <8.75ns. Overshoot: ≤1%.
Deflection coefficient: 14 calibrated positions
variable 2.5:1 to min. 50V/div.
1mV/div and 2mV/div: ±5% (0 to 10MHz (-3dB))
5mV/div to 20V/div: ±3% (1-2-5sequence).
Input impedance: 1MΩ II 15pF.
Input coupling: DC - AC - GD (Ground)
Input voltage: max. 400V (DC + peak AC).
Triggering
Automatic(peak to peak):Normal: DC-100MHz, LED for trigger indication.
Slope: positive or negative.
Sources: CH I or II, line, ext.
Coupling: AC (≥10Hz -100MHz), DC (0-100MHz),
Triggering ext.: ≥0.3Vpp from DC to 100MHz
Active TV-Sync-Separator (field & line, pos, neg.)
2nd triggering (Del. Trig.): normal with level
Time coefficients: 1-2-5 sequence, Accuracy ±3%
Analog: 22 cal. positions from 0.5s - 50ns/div.
Digital:
Variable (analog) 2.5:1 up to 1.25s/div.
X-MAG. x10: analog to 10ns/div., dig. to 0.1µs/di v ±5%
Delay: 120ms - 200ns, variable,
Hold-off time (analog): variable to approx. 10:1.
Bandwidth X-amplifier (analog): 0-3MHz (−3dB).
Input X-amplifier via Channel II, Sensitivity see
X-Y-phase shift : <3° below 120kHz.
Operating modes: Refresh, Roll, Single, XY,
Envelope, Average (2 to 512 waveforms).
Automatic Dot Join function
Sample Rate: max. 100MS/s (8 bit)
Refresh rate: max. 180/s
Record length: 2048 x 8 bit per channel.
Manual (front panel switches);
Auto Set (automatic parameter selection).
Save/Recall of 9 user-defined parameter settings
RS232 interface for remote control via a PC.Remote control (Option) HZ68.
Multifunction- Interface HO79-6(Option): RS232,
Readout: Display of parameter settings.
Cursor measurement of ∆V, ∆t or ∆1/t
separate or in tracking mode.
Test voltage: approx. 7V
Test current: max. 7mA
Test frequency: approx.50Hz
One test lead is grounded (Safety Earth).
CRT: D14-364GY/123 or ER151-GH/-,rectangular
screen (8x10cm) internal graticule
Acceleration voltage: approx 2000V
Trace rotation: adjustable on front panel
Calibrator: square-wave generator (tr <4ns)
≈1kHz/1MHz; Output: 0.2V ±1%.
Analog Intensitymodulation, max. +5V (TTL).
Line voltage: 100-240V AC ±10%, 50/60Hz
Power consumption: approx. 42 Watt at 50Hz.
Min./Max. ambient temperature: 0°C...+40°C
Protective system: Safety class I (IEC1010-1)
Weight: approx. 5.6kg, color: techno-brown
Cabinet: W 285, H 125, D 380 mm 3/98
Storage 2 x 2048 x 8 bit, Reference Memory, Post/Pre-Trigger
.
The worldwide success of HAMEG´s HM205 and HM305 has led to the
introduction of the new microprocessor controlled HM407 Analog/Digital
oscilloscope. This instrument offers much more performance and specifications
over its predecessores. The HM407 incorporates a microprocessor-based system
that extensively automates operation. The majority of signals can be displayed by
simply pressing the “Autoset“ button. A “Save/Recall“ function is available for
storing frequently used setup parameters.
The increased maximum sampling rate of 100MS/s now allows to capture a
10MHz signal in “Single“ mode with 10 samples (dots) per period. The automatic
Dot-Join function provides linear connections between the captured points,
ensuring that all digitized signals are displayed without gaps. New features are the
two reference memories, allowing their contents to be compared with the live
signal at any time. Cursors can be activated for waveform measurements. All
important parameter settings are displayed on the CRT screen. The built-in RS232-Interface enables remote control operation and signal processing via a PC.
Unique in its price range is also the analog section of the HM407. The increased
bandwidth of 40MHz (-3dB) allows the stable display of signals up to 100MHz. As
always, the Component Tester with one-button control is a standard feature in
the HM407. This is also true for the switchable 1kHz/1MHz Calibrator which
permits you to check the transient characteristics from probe tip to the screen at
any time.
All in all, the new HM407 presents itself as a practical hands-on oscilloscope for
today’s progressive measurement requirements offering a price/performance ratio that sets new standards world-wide.
Screen photo of stored sinewave signals. Screen shot of measurement software.
Storage Modes: Refresh, Single, Roll, Average and Envelope
Accessories supplied: Line Cord, Operators Manual, 2 Probes1:1/ 10:1
Subject to change without notice
3
General Information
General Information
This oscilloscope is easy to operate. The logical arrangement
of the controls allows anyone to quickly become familiar with
the operation of the instrument, however, experienced users
are also advised to read through these instructions so that all
functions are understood.
Immediately after unpacking, the instrument should be checked
for mechanical damage and loose parts in the interior. If there
is transport damage, the supplier must be informed
immediately. The instrument must then not be put into
operation.
Symbols
ATTENTION - refer to manual
Danger - High voltage
Protective ground (earth) terminal
Use of tilt handle
To view the screen from the best angle, there are three
different positions (C, D, E) for setting up the instrument. If the
instrument is set down on the floor after being carried, the
handle automatically remains in the upright carrying position
(A). In order to place the instrument onto a horizontal surface,
the handle should be turned to the upper side of the oscilloscope
(C). For the D position (10° inclination), the handle should be
turned to the opposite direction of the carrying position until it
locks in place automatically underneath the instrument. For
the E position (20° inclination), the handle should be pulled to
release it from the D position and swing backwards until it
locks once more. The handle may also be set to a position for
horizontal carrying by turning it to the upper side to lock in the
B position. At the same time, the instrument must be lifted,
because otherwise the handle will jump back.
The case, chassis and all measuring terminals are connected
to the protective earth contact of the appliance inlet. The
instrument operates according to Safety Class I (threeconductor power cord with protective earthing conductor and
a plug with earthing contact).
The mains/line plug shall only be inserted in a socket outlet
provided with a protective earth contact. The protective action
must not be negated by the use of an extension cord without
a protective conductor.
The mains/line plug must be inserted before connections
are made to measuring circuits.
The grounded accessible metal parts (case, sockets, jacks)
and the mains/line supply contacts (line/live, neutral) of the
instrument have been tested against insulation breakdown
with 2200V DC.
Under certain conditions, 50Hz or 60Hz hum voltages can
occur in the measuring circuit due to the interconnection with
other mains/line powered equipment or instruments. This can
be avoided by using an isolation transformer (Safety Class II)
between the mains/line outlet and the power plug of the
device being investigated.
Most cathode-ray tubes develop X-rays. However, the dose
equivalent rate falls far below the maximum permissible value
of 36pA/kg (0.5mR/h).
Whenever it is likely that protection has been impaired, the
instrument shall be made inoperative and be secured against
any unintended operation. The protection is likely to be impaired if, for example, the instrument
• shows visible damage,
• fails to perform the intended measurements,
• has been subjected to prolonged storage under unfavorable
conditions (e.g. in the open or in moist environments),
• has been subject to severe transport stress (e.g. in poor
packaging).
Safety
This instrument has been designed and tested in accordance
with IEC Publication 1010-1 (overvoltage category II, pollution
degree 2), Safety requirements for electrical equipment for
measurement, control, and laboratory use. The CENELEC
regulations EN 61010-1 correspond to this standard. It has left
the factory in a safe condition. This instruction manual contains
important information and warnings which have to be followed
by the user to ensure safe operation and to retain the
oscilloscope in a safe condition.
Intended purpose and operating conditions
This instrument must be used only by qualified experts who
are aware of the risks of electrical measurement.
The instrument is specified for operation in industry, light
industry, commercial and residential environments.
Due to safety reasons the instrument must only be connected
to a properly installed power outlet, containing a protective
earth conductor. The protective earth connection must not be
broken. The power plug must be inserted in the power outlet
while any connection is made to the test device.
The instrument has been designed for indoor use. The
permissible ambient temperature range during operation is
+10°C (+50°F) ... +40°C (+104°F). It may occasionally be
subjected to temperatures between +10°C (+50°F) and -10°C
(+14°F) without degrading its safety. The permissible ambient
temperature range for storage or transportation is -40°C (40°F) ... +70°C (+158°F). The maximum operating altitude is
up to 2200m (non-operating 15000m). The maximum relative
humidity is up to 80%.
If condensed water exists in the instrument it should be
acclimatized before switching on. In some cases (e.g. extremely
cold oscilloscope) two hours should be allowed before the
instrument is put into operation. The instrument should be
kept in a clean and dry room and must not be operated in
4
Subject to change without notice
General Information
explosive, corrosive, dusty, or moist environments. The
oscilloscope can be operated in any position, but the convection
cooling must not be impaired. The ventilation holes may not be
covered. For continuous operation the instrument should be
used in the horizontal position, preferably tilted upwards,
resting on the tilt handle.
The specifications stating tolerances are only valid if
the instrument has warmed up for 30minutes at an
ambient temperature between +15°C (+59°F) and +30°C
(+86°F). Values without tolerances are typical for an
average instrument.
EMC
This instrument conforms to the European standards regarding
the electromagnetic compatibility. The applied standards are:
Generic immunity standard EN50082-2:1995 (for industrial
environment) Generic emission standard EN50081-1:1992
(for residential, commercial and light industry environment).
This means that the instrument has been tested to the highest
standards.
Please note that under the influence of strong electromagnetic
fields, such signals may be superimposed on the measured
signals.
Under certain conditions this is unavoidable due to the
instrument’s high input sensitivity, high input impedance and
bandwidth. Shielded measuring cables, shielding and earthing
of the device under test may reduce or eliminate those effects.
Warranty
HAMEG warrants to its Customers that the products it
manufactures and sells will be free from defects in materials
and workmanship for a period of 2 years. This warranty shall
not apply to any defect, failure or damage caused by improper
use or inadequate maintenance and care. HAMEG shall not be
obliged to provide service under this warranty to repair damage
resulting from attempts by personnel other than HAMEG
representatives to install, repair, service or modify these
products.
In order to obtain service under this warranty, Customers
must contact and notify the distributor who has sold the
product. Each instrument is subjected to a quality test with 10
hour burn-in before leaving the production. Practically all early
failures are detected by this method. In the case of shipments
by post, rail or carrier it is recommended that the original
packing is carefully preserved. Transport damages and damage
due to gross negligence are not covered by the guarantee.
In the case of a complaint, a label should be attached to the
housing of the instrument which describes briefly the faults
observed. If at the same time the name and telephone number
(dialing code and telephone or direct number or department
designation) is stated for possible queries, this helps towards
speeding up the processing of guarantee claims.
recommended. The exterior of the oscilloscope should be
cleaned regularly with a dusting brush. Dirt which is difficult to
remove on the casing and handle, the plastic and aluminium
parts, can be removed with a moistened cloth (99% water
+1% mild detergent). Spirit or washing benzine (petroleum
ether) can be used to remove greasy dirt. The screen may be
cleaned with water or washing benzine (but not with spirit
(alcohol) or solvents), it must then be wiped with a dry clean
lint-free cloth. Under no circumstances may the cleaning fluid
get into the instrument. The use of other cleaning agents can
attack the plastic and paint surfaces.
Protective Switch-Off
This instrument is equipped with a switch mode power supply.
It has both overvoltage and overload protection, which will
cause the switch mode supply to limit power consumption to
a minimum. In this case a ticking noise may be heard.
Power supply
The oscilloscope operates on mains/line voltages between
100VAC and 240VAC. No means of switching to different input
voltages has therefore been provided.
The power input fuses are externally accessible. The fuseholder
is located above the 3-pole power connector. The power input
fuses are externally accessible, if the rubber connector is
removed. The fuseholder can be released by pressing its
plastic retainers with the aid of a small screwdriver. The
retainers are located on the right and left side of the holder and
must be pressed towards the center. The fuse(s) can then be
replaced and pressed in until locked on both sides.
Use of patched fuses or short-circuiting of the fuseholder is
not permissible; HAMEG assumes no liability whatsoever for
any damage caused as a result, and all warranty claims
become null and void.
Fuse type:
Size 5x20mm; 0.8A, 250V AC fuse;
must meet IEC specification 127,
Sheet III (or DIN 41 662
or DIN 41 571, sheet 3).
Time characteristic: time-lag (T).
Attention!
There is a fuse located inside the instrument within the
switch mode power supply:
Size 5x20mm; 0.8A, 250V AC fuse;
must meet IEC specification 127,
Sheet III (or DIN 41 662
or DIN 41 571, sheet 3).
Time characteristic: fast (F).
This fuse must not be replaced by the operator!
Maintenance
Various important properties of the oscilloscope should be
carefully checked at certain intervals. Only in this way is it
largely certain that all signals are displayed with the accuracy
on which the technical data are based. The test methods
described in the test plan of this manual can be performed
without great expenditure on measuring instruments. However,
purchase of the HAMEG scope tester HZ 60, which despite its
low price is highly suitable for tasks of this type, is very much
Subject to change without notice
5
Type of signal voltage
Type of signal voltage
The oscilloscope HM407 allows examination of DC voltages
and most repetitive signals in the frequency range up to at
least 40MHz (-3dB).
The vertical amplifiers have been designed for minimum
overshoot and therefore permit a true signal display.
The display of sinusoidal signals within the bandwidth limits
causes no problems, but an increasing error in measurement
due to gain reduction must be taken into account when
measuring high frequency signals. This error becomes
noticeable at approx. 14MHz. At approx. 18MHz the reduction
is approx. 10% and the real voltage value is 11% higher. The
gain reduction error can not be defined exactly as the -3dB
bandwidth of the amplifiers differ between 40MHz and 42MHz.
For sinewave signals the -6dB limit is approx. 50MHz.
When examining square or pulse type waveforms, attention
must be paid to the harmonic content of such signals. The
repetition frequency (fundamental frequency) of the signal
must therefore be significantly smaller than the upper limit
frequency of the vertical amplifier.
Displaying composite signals can be difficult, especially if they
contain no repetitive higher amplitude content which can be
used for triggering. This is the case with bursts, for instance.
To obtain a well-triggered display in this case, the assistance
of the variable holdoff function or the delayed time base may
be required. Television video signals are relatively easy to
trigger using the built-in TV-Sync-Separator (TV).
The relationship between the different voltage magnitudes
can be seen from the following figure.
Voltage values of a sine curve
Vrms = effective value; Vp = simple peak or crest value;
Vpp = peak-to-peak value; Vmom = momentary value.
The minimum signal voltage which must be applied to the Y
input for a trace of 1div height is 1mVpp (± 5%) when this
deflection coefficient is displayed on the screen (readout) and
the vernier is switched off (VAR-LED dark). However, smaller
signals than this may also be displayed. The deflection
coefficients are indicated in mV/div or V/div (peak-to-peak
value).
The magnitude of the applied voltage is ascertained by
multiplying the selected deflection coefficient by the vertical
display height in div. If an attenuator probe x10 is used, a
further multiplication by a factor of 10 is required to ascertain
the correct voltage value.
For optional operation as a DC or AC voltage amplifier, each
vertical amplifier input is provided with a DC/AC switch. DC
coupling should only be used with a series-connected attenuator
probe or at very low frequencies or if the measurement of the
DC voltage content of the signal is absolutely necessary.
When displaying very low frequency pulses, the flat tops may
be sloping with AC coupling of the vertical amplifier (AC limit
frequency approx. 1.6 Hz for 3dB). In this case, DC operation
is preferred, provided the signal voltage is not superimposed
on a too high DC level. Otherwise a capacitor of adequate
capacitance must be connected to the input of the vertical
amplifier with DC coupling. This capacitor must have a
sufficiently high breakdown voltage rating. DC coupling is also
recommended for the display of logic and pulse signals,
especially if the pulse duty factor changes constantly. Otherwise
the display will move upwards or downwards at each change.
Pure direct voltages can only be measured with DC-coupling.
The input coupling is selectable by the AC/DC pushbutton. The
actual setting is displayed in the readout with the ” = ” symbol
for DC- and the ”
~ ” symbol for AC coupling.
Amplitude Measurements
In general electrical engineering, alternating voltage data
normally refers to effective values (rms = root-mean-square
value). However, for signal magnitudes and voltage designations
in oscilloscope measurements, the peak-to-peak voltage (Vpp)
value is applied. The latter corresponds to the real potential
difference between the most positive and most negative
points of a signal waveform.
If a sinusoidal waveform, displayed on the oscilloscope screen,
is to be converted into an effective (rms) value, the resulting
peak-to-peak value must be divided by 2x√2 = 2.83. Conversely,
it should be observed that sinusoidal voltages indicated in
Vrms (Veff) have 2.83 times the potential difference in Vpp.
For exact amplitude measurements, the variable control (VAR)
must be set to its calibrated detent CAL position.
With the variable control activated the deflection sensitivity
can be reduced up to a ratio of 2.5 to 1 (please note “controls
and readout”). Therefore any intermediate value is possible
within the 1-2-5 sequence of the attenuator(s).
With direct connection to the vertical input, signals up
to 400Vpp may be displayed (attenuator set to 20V/div,
variable control to 2.5:1).
With the designations
H= display height in div,
U= signal voltage in Vpp at the vertical input,
D= deflection coefficient in V/div at attenuator switch,
the required value can be calculated from the two given
quantities:
However, these three values are not freely selectable. They
have to be within the following limits (trigger threshold,
accuracy of reading):
H between 0.5 and 8div, if possible 3.2 to 8div,
U between 1mVpp and 160Vpp,
D between 1mV/div and 20V/div in 1-2-5 sequence.
Examples:
Set deflection coefficient D = 50mV/div 0.05V/div,
observed display height H = 4.6div,
required voltage U = 0.05x4.6 = 0.23Vpp.
Input voltage U = 5Vpp,
set deflection coefficient D = 1V/div,
required display height H = 5:1 = 5div.
6
Subject to change without notice
Type of signal voltage
Signal voltage U = 230Vrmsx2√2 = 651Vpp
(voltage > 160Vpp, with probe 10:1: U = 65.1Vpp),
desired display height H = min. 3.2div, max. 8div,
max. deflection coefficient D = 65.1:3.2 = 20.3V/div,
min. deflection coefficient D = 65.1:8 = 8.1V/div,
adjusted deflection coefficient D = 10V/div.
The previous examples are related to the CRT graticule reading.
The results can also be determined with the aid of the DV
cursor measurement (please note “controls and readout”).
The input voltage must not exceed 400V, independent from
the polarity.
If an AC voltage which is superimposed on a DC voltage is
applied, the maximum peak value of both voltages must not
exceed + or - 400V. So for AC voltages with a mean value of
zero volt the maximum peak to peak value is 800Vpp.
If attenuator probes with higher limits are used, the probes
limits are valid only if the oscilloscope is set to DC input
coupling.
If DC voltages are applied under AC input coupling conditions
the oscilloscope maximum input voltage value remains 400V.
The attenuator consists of a resistor in the probe and the 1MΩ
input resistor of the oscilloscope, which are disabled by the AC
input coupling capacity when AC coupling is selected. This also
applies to DC voltages with superimposed AC voltages. It also
must be noted that due to the capacitive resistance of the AC
input coupling capacitor, the attenuation ratio depends on the
signal frequency. For sinewave signals with frequencies higher
than 40Hz this influence is negligible.
peak and the DC voltage results in the max. voltage (DC +
ACpeak).
Time Measurements
As a rule, most signals to be displayed are periodically repeating
processes, also called periods. The number of periods per
second is the repetition frequency. Depending on the time
base setting (TIME/DIV.-knob) indicated by the readout, one or
several signal periods or only a part of a period can be
displayed. The time coefficients are stated in ms/div, µs/div or
ns/div. The following examples are related to the CRT graticule
reading. The results can also be determined with the aid of the
∆T and 1/∆T cursor measurement (please note “ controls and
readout”).
The duration of a signal period or a part of it is determined by
multiplying the relevant time (horizontal distance in div) by the
(calibrated) time coefficient displayed in the readout.
Uncalibrated, the time base speed can be reduced until a
maximum factor of 2.5 is reached. Therefore any intermediate
value is possible within the 1-2-5 sequence.
With the designations
L = displayed wave length in div of one period,
T = time in seconds for one period,
F = recurrence frequency in Hz of the signal,
Tc= time coefficient in ms, µs or ns/div and the relation
F = 1/T, the following equations can be stated:
With the above listed exceptions HAMEG 10:1 probes can be
used for DC measurements up to 600V or AC voltages (with
a mean value of zero volt) of 1200Vpp. The 100:1 probe HZ53
allows for 1200V DC or 2400Vpp for AC.
It should be noted that its AC peak value is derated at higher
frequencies. If a normal x10 probe is used to measure high
voltages there is the risk that the compensation trimmer
bridging the attenuator series resistor will break down causing
damage to the input of the oscilloscope. However, if for
example only the residual ripple of a high voltage is to be
displayed on the oscilloscope, a normal x10 probe is sufficient.
In this case, an appropriate high voltage capacitor (approx. 2268nF) must be connected in series with the input tip of the
probe.
With Y-POS. control (input coupling to GD) it is possible to use
a horizontal graticule line as reference line for ground potential
before the measurement. It can lie below or above the horizontal central line according to whether positive and/or negative
deviations from the ground potential are to be measured.
Total value of input voltage
However, these four values are not freely selectable. They
have to be within the following limits:
Lbetween 0.2 and 10div, if possible 4 to 10div,
Tbetween 10ns and 5s,
Fbetween 0.5Hz and 100MHz,
Tcbetween 100ns/div and 500ms/div in 1-2-5 sequence
(with X-MAG. (x10) inactive), and
Tcbetween 10ns/div and 50ms/div in 1-2-5 sequence
(with X-MAG. (x10) active).
Examples:
Displayed wavelength L = 7div,
set time coefficient Tc = 100ns/div,
required period T = 7x100x10
required rec. freq. F = 1:(0.7x10
-9
= 0.7µs
-6
) = 1.428MHz.
Signal period T = 1s,
set time coefficient Tc = 0.2s/div,
required wavelength L = 1:0.2 = 5div.
Displayed ripple wavelength L = 1div,
set time coefficient Tc = 10ms/div,
required ripple freq. F = 1:(1x10x10
-3
) = 100Hz.
TV-line frequency F = 15625Hz,
set time coefficient Tc = 10µs/div,
required wavelength L = 1:(15 625x10-5) = 6.4div.
The dotted line shows a voltage alternating at zero volt level.
If superimposed on a DC voltage, the addition of the positive
Subject to change without notice
Sine wavelength L = min. 4div, max. 10div,
Frequency F = 1kHz,
max. time coefficient Tc = 1:(4x10
3
) = 0.25ms/div,
min. time coefficient Tc = 1:(10x103) = 0.1ms/div,
set time coefficient Tc = 0.2ms/div,
required wavelength L = 1:(10
3
x0.2x10-3) = 5div.
7
Type of signal voltage
Displayed wavelength L = 0.8div,
set time coefficient Tc = 0.5µs/div,
pressed X-MAG. (x10) button: Tc = 0.05µs/div,
required rec. freq. F = 1:(0.8x0.05x10
required period T = 1:(25x10
6
) = 40ns.
-6
) = 25MHz,
If the time is relatively short as compared with the complete
signal period, an expanded time scale should always be applied
(X-MAG. (x10) active). In this case, the time interval of interest
can be shifted to the screen center using the X-POS. control.
When investigating pulse or square waveforms, the critical
feature is the risetime of the voltage step. To ensure that
transients, ramp-offs, and bandwidth limits do not unduly
influence the measuring accuracy, the risetime is generally
measured between 10% and 90% of the vertical pulse height.
For measurement, adjust the Y deflection coefficient using its
variable function (uncalibrated) together with the Y-POS.
control so that the pulse height is precisely aligned with the 0%
and 100% lines of the internal graticule. The 10% and 90%
points of the signal will now coincide with the 10% and 90%
graticule lines. The risetime is given by the product of the
horizontal distance in div between these two coincident points
and the calibrated time coefficient setting. The fall time of a
pulse can also be measured by using this method.
The following figure shows correct positioning of the
oscilloscope trace for accurate risetime measurement.
measure in any display position and at any signal amplitude. It
is only important that the full height of the signal edge of
interest is visible in its full length at not too great steepness and
that the horizontal distance at 10% and 90% of the amplitude
is measured. If the edge shows rounding or overshooting, the
100% should not be related to the peak values but to the mean
pulse heights. Breaks or peaks (glitches) next to the edge are
also not taken into account. With very severe transient
distortions, the rise and fall time measurement has little
meaning. For amplifiers with approximately constant group
delay (therefore good pulse transmission performance) the
following numerical relationship between rise time tr (in ns)
and bandwidth B (in MHz) applies:
Connection of Test Signal
In most cases briefly depressing the AUTO SET causes a
useful signal related instrument setting. The following
explanations refer to special applications and/or signals,
demanding a manual instrument setting. The description of
the controls is explained in the section “controls and readout”.
Caution:
When connecting unknown signals to the oscilloscope
input, always use a x10 probe, automatic triggering and
set the input coupling switch to DC (readout). The
attenuator should initially be set to 20V/div.
With a time coefficient of 10ns/div (X x10 magnification
active), the example shown in the above figure results in a total
measured risetime of
= 1.6div x 10ns/div = 16ns
t
tot
When very fast risetimes are being measured, the risetimes of
the oscilloscope amplifier and of the attenuator probe has to
be deducted from the measured time value. The risetime of
the signal can be calculated using the following formula.
In this t
the oscilloscope amplifier (approx. 8.75ns), and t
of the probe (e.g. = 2ns). If t
can be taken as the risetime of the pulse, and calculation is
is the total measured risetime, t
tot
is greater than 100ns, then t
tot
is the risetime of
osc
the risetime
p
unnecessary.
Calculation of the example in the figure above results in a signal
risetime
= √162 - 8.752 - 22 = 13.25ns
t
r
The measurement of the rise or fall time is not limited to the
trace dimensions shown in the above diagram. It is only
particularly simple in this way. In principle it is possible to
Sometimes the trace will disappear after an input signal has
been applied. Then a higher deflection coefficient (lower input
sensitivity) must be chosen until the vertical signal height is
only 3-8div. With a signal amplitude greater than 160Vpp and
the deflection coefficient (VOLTS/DIV.) in calibrated condition,
an attenuator probe must be inserted before the vertical input.
If, after applying the signal, the trace is nearly blanked, the
period of the signal is probably substantially longer than the set
time deflection coefficient (TIME/DIV.). It should be switched
to an adequately larger time coefficient.
The signal to be displayed can be connected directly to the Yinput of the oscilloscope with a shielded test cable such as
HZ32 or HZ34, or reduced through a x10 or x100 attenuator
probe. The use of test cables with high impedance circuits is
only recommended for relatively low frequencies (up to approx.
50kHz). For higher frequencies, the signal source must be of
low impedance, i.e. matched to the characteristic resistance
of the cable (as a rule 50Ω). Especially when transmitting
square and pulse signals, a resistor equal to the characteristic
impedance of the cable must also be connected across the
cable directly at the Y-input of the oscilloscope. When using a
50Ω cable such as the HZ34, a 50Ω through termination type
HZ22 is available from HAMEG. When transmitting square
signals with short rise times, transient phenomena on the
edges and top of the signal may become visible if the correct
termination is not used. A terminating resistance is sometimes
recommended with sine signals as well. Certain amplifiers,
generators or their attenuators maintain the nominal output
tot
voltage independent of frequency only if their connection
cable is terminated with the prescribed resistance. Here it
must be noted that the terminating resistor HZ22 will only
dissipate a maximum of 2Watts. This power is reached with
10Vrms or at 28.3Vpp with sine signal. If a x10 or x100
attenuator probe is used, no termination is necessary. In this
case, the connecting cable is matched directly to the high
impedance input of the oscilloscope. When using attenuators
probes, even high internal impedance sources are only slightly
loaded (approx. 10MΩ II 12pF or 100MΩ II 5pF with HZ53).
Therefore, if the voltage loss due to the attenuation of the
8
Subject to change without notice
Controls and readout
probe can be compensated by a higher amplitude setting, the
probe should always be used. The series impedance of the
probe provides a certain amount of protection for the input of
the vertical amplifier. Because of their separate manufacture,
all attenuator probes are only partially compensated, therefore
accurate compensation must be performed on the oscilloscope
(see Probe compensation ).
Standard attenuator probes on the oscilloscope normally reduce
its bandwidth and increase the rise time. In all cases where the
oscilloscope bandwidth must be fully utilized (e.g. for pulses
with steep edges) we strongly advise using the probes HZ51
(x10) HZ52 (x10 HF) and HZ54 (x1 and x10). This can save the
purchase of an oscilloscope with larger bandwidth.
The probes mentioned have a HF-calibration in addition to low
frequency calibration adjustment. Thus a group delay correction
to the upper limit frequency of the oscilloscope is possible with
the aid of an 1MHz calibrator, e.g. HZ60.
In fact the bandwidth and rise time of the oscilloscope are not
noticeably changed with these probe types and the waveform
reproduction fidelity can even be improved because the probe
can be matched to the oscilloscopes individual pulse response.
If a x10 or x100 attenuator probe is used, DC input
coupling must always be used at voltages above 400V.
With AC coupling of low frequency signals, the
attenuation is no longer independent of frequency,
pulses can show pulse tilts. Direct voltages are
suppressed but load the oscilloscope input coupling
capacitor concerned. Its voltage rating is max. 400 V
(DC + peak AC). DC input coupling is therefore of quite
special importance with a x100 attenuation probe which
usually has a voltage rating of max. 1200 V (DC + peak
AC). A capacitor of corresponding capacitance and
voltage rating may be connected in series with the
attenuator probe input for blocking DC voltage (e.g. for
hum voltage measurement).
With all attenuator probes, the maximum AC input voltage
must be derated with frequency usually above 20kHz. Therefore
the derating curve of the attenuator probe type concerned
must be taken into account. The selection of the ground point
on the test object is important when displaying small signal
voltages. It should always be as close as possible to the
measuring point. If this is not done, serious signal distortion
may result from spurious currents through the ground leads or
chassis parts. The ground leads on attenuator probes are also
particularly critical. They should be as short and thick as
possible. When the attenuator probe is connected to a BNCsocket, a BNC-adapter, should be used. In this way ground and
matching problems are eliminated. Hum or interference
appearing in the measuring circuit (especially when a small
deflection coefficient is used) is possibly caused by multiple
grounding because equalizing currents can flow in the shielding
of the test cables (voltage drop between the protective
conductor connections, caused by external equipment
connected to the mains/line, e.g. signal generators with
interference protection capacitors).
Controls and readout
The following description assumes that the operating mode
“COMPONENT TEST” is switched off.
parameter settings are displayed in the screen readout when
the oscilloscope is on.
The LED indicators on the large front panel facilitate operation
and provide additional information. Electrical end positions of
controls are indicated by acoustic signal (beep).
All important measuring
All controls, except the power switch (POWER), the calibration
frequency pushbutton (CAL. 1kHz/1MHz), the FOCUS control
and the trace rotation control, are electronically set and
interrogated. Thus, all electronically set functions and their
current settings can be stored and also remotely controlled.
Some controls are only operative in the digital mode or have a
different function. Explanations pertaining to them are indicated
with the hint “storage mode only”.
The large front panel is, as is usual with Hameg oscilloscopes,
is marked with several fields.
The following controls and LED indicators are located on the
top, to the right of the screen, above the horizontal line:
(1) POWER - Pushbutton and symbols for ON (I) and OFF (O).
After the oscilloscope is switched on, all LEDs are lit and
an automated instrument test is performed. During this
time the HAMEG logo and the software version are
displayed on the screen. After the internal test is completed
successfully, the overlay is switched off and the normal
operation mode is present. Then the last used settings
become activated and one LED indicates the ON condition.
It is possible to modify certain functions (SETUP) or to call
automatic calibration procedures (CALIBRATE).
relating to this see section “MENU”
(2) AUTOSET - Pushbutton
Briefly depressing this pushbutton results in an automatic
instrument setting automatically selecting Yt mode. The
instrument is set to the last used Yt mode setting (CH I,
CH II or DUAL).
SEARCH (SEA) and DELAY (DEL and DTR) mode is
automatically switched off.
Please note “AUTO SET”
.
For details
.
STORAGE MODE ONLY
Additionally, AUTOSET automatically selects refresh
mode (RFR) when SINGLE (SGL) or ROLL (ROL) function
is in operation.
(3) RM - LED
The remote control mode can be switched on or off
(”RM” LED dark) via the RS232 interface. On condition
that the “RM” LED is lit, all electronically selectable
controls on front panel are inactive. This state can be left
by depressing the AUTOSET pushbutton provided it was
not deactivated via the interface.
STORAGE MODE ONLY
The RM-LED is lit during data transfer via the built in
RS232 interface. At this time the controls are inactive.
(4) INTENS - READOUT - Control knob with associated
pushbutton and LEDs.
This control knob is for adjusting the trace (A) and readout
intensity (RO). Turning this knob clockwise increases and
turning it counterclockwise decreases the intensity.
The READOUT pushbutton below is for selecting the
function in two ways.
Subject to change without notice
9
Controls and readout
If the readout (RO) is not switched off, briefly pressing
the READOUT pushbutton switches over the INTENS
knob function indicated by a LED in the sequence:
Yt (time base) mode: A - RO - A
XY mode: A - RO - A.
Component Test: A - RO - A.
Pressing and holding the READOUT pushbutton switches
the readout on or off. In readout off condition the INTENS
knob function can consequently not be set to RO.
Switching the readout off, may be required if interference
is visible on the signal(s). Such interference may also
originate from the chopper generator if the instrument is
operated in chopped DUAL mode.
With the exception of the letters “CT” all other READOUT
information is switched off in COMPONENT TEST mode.
All INTENS settings are stored after the instrument is
switched off.
The AUTOSET function switches the readout on. The
INTENS setting for each function is automatically set to
the mean value, if less intensity was previously selected.
Attention!
The time base ranges are dependent on the operating
mode Analog or Digital (storage). The following data
relate to operation without X magnification (X-MAG.
x10).
Analog mode:
Time base from 500ms/cm to 50ns/cm
(without trace delay).
With trace delay, from 20ms/cm to 50ns/cm.
Delay ranges from 20ms/cm to 100ns/cm.
Digital mode:
Time bases from 100s/cm to 1µs/cm.
This results in the following behavior when switched
from analog to digital mode and vice versa:
1. If in analog mode, the time base has been selected
between 500ns/cm and 50ns/cm, then on switching
to digital mode the lowest available time coefficient will
be automatically selected, i.e. 1µs/cm. If now one
switches back to analog mode without having made
any time base changes in the digital mode, then the last
time base selected in the analog mode is again active
(e.g. 500ns/cm).
If on the other hand, the time base is changed after
switching over to digital mode (e.g. to 2µs/cm). Then,
when switched back to analog mode, the time base in
analog mode will be set to the value selected in the
digital mode (e.g. 2µs/cm).
(5) T R - Trimming potentiometer.
The trace rotation control can be adjusted with a small
screwdriver (
(6) FOCUS - Control knob.
This control knob effects both the trace and the readout
sharpness.
(7) STOR. ON / HOLD - Pushbutton with two functions.
STOR. ON
Pressing and holding the button switches from analog (Yt
or XY) to storage mode and vice versa. If CT (Component
Tester) mode is present (only available in analog mode),
it must be switched off first to enable switching over to
storage mode.
The oscilloscope is in analog mode if none of the LED’s
associated with the STOR.MODE (9) pushbuttons are lit
and a pre- or post-trigger value (PT...%) is not indicated by
the readout. Pressing and holding the STOR. ON button
switches over to the digital mode, but without changing
the channel operating mode (CH I, CH II, DUAL, ADD andXY). The actual signal capture mode is indicated by one
of the STOR. MODE-LED‘s (RFR - ENV - AVM - ROL) and
in addition displayed by the readout. In digital XY mode
the RFR-LED is lit and the readout indicates XY.
If digital SINGLE event (SGL) capture mode is selected,
all STOR. MODE-LED‘s are dark, but the readout displays
the pre- or post-trigger value (PT...%).
please note “trace rotation TR”
)
2.If a time base between 100s/cm and 1s/cm has been
set in the digital mode and the mode is switched to
analog, then the time base in analog mode is automatically
set to 500ms/cm. The rest is as described before.
The X-MAG x10 setting remains unchanged when
switched from analog to digital mode and vice versa.
STORAGE MODE ONLY
If by pressing and holding the STOR. ON / HOLD button,
the mode is switched to digital, then one of the associated
LED’s lights up. Which one it is, depends on the last
selected digital operation.
Exception
Switching over from analog SINGLE mode to digital
mode sets the instrument automatically to digital
SINGLE mode.
Attention
The possibilities of delayed trace and the related
operations with delayed time base are not available in
digital mode.
For additional information regarding the digital mode, see
section STORAGE OPERATION.
HOLD
STORAGE MODE ONLY
Briefly pressing the STOR. ON / HOLD pushbutton
switches over between protected and unprotected mode
of the current memory contents.
The current contents of the memory are protected against
overwriting when HLD (HOLD) instead of channel
10
Subject to change without notice
Controls and readout
information (e.g. Y1... ) is displayed in the readout. This
prevents a change in the Yt mode setting, but it is possible
to select between DUAL (Yt) and XY display by pressing
the DUAL (22) pushbutton if one of these modes was
selected before activating HOLD.
If HOLD is switched off, one can observe how the
existing memory contents are successively overwritten
by new data especially with slow time base settings and
refresh mode. Protecting the memory contents in the
middle of a data acquisition process can result in an
irregularity at the junction of old (right) and new data (left).
This can be avoided by recording in single shot mode
(SGL), even though the input signal is repetitive. At the
end of a sweep, one can use HOLD to protect the
contents against being overwritten by an unintentional
actuation of RESET (RES).
The signal in each of the current memory can be shifted
in the vertical direction (+/- 4cm) with the corresponding
Y-POS rotary knob when HOLD is operative.
The original trace position will be lost when shifted
vertically, but this can be found again. To this end the Y-POS knob in question must be rotated quickly. Once the
original position is reached, the trace does not shift
anymore although the knob is rotated further.
Simultaneously a signal tone sounds. To shift the trace
vertically again it will be required to stop rotating the knob
for at least about 2 seconds.
Attention!
The dynamic range limits of the A/D converter may
become visible if a Y-position shift is performed after
storage. This can affect those signal parts which were
originally above or below the screen.
PRETRIGGER
0% PRETRIGGER (readout “PT0%”) means that the
signal display starts with the trigger event. The trigger
point symbol indicates this position. If the X-POS. control
is not in center position, an arrow pointing to the left may
be displayed. Then the X-POS. control must be turned
clockwise until the arrow is no longer visible.
25% PRETRIGGER (readout “PT25%”) is achieved after
pressing the PTR button once. The signal display starts
with 25% pre-history and the trigger point symbol is
shifted 2.5 divisions to the right.
Each time the PTR button is pressed the PRETRIGGER
value increases by 25% until 100% is reached. If in 100%
condition an arrow symbol is displayed in addition to the
trigger point symbol, the X-POS. control should be turned
ccw. to make the trigger point visible on the screen.
The duration of the prehistory is determined by multiplying
the time coefficient by the pretrigger value (in divisions).
E.g. 20ms/div x 7,5 div (= 75% pretrigger) = 150ms.
POSTTRIGGER
In POSTTRIGGER condition the trigger point is always to
the left of the screen and therefore not visible. The trigger
point symbol then only indicates the LEVEL setting. An
additional arrow symbol which points to the left is displayed
to indicate post trigger operation. In POSTTRIGGER
condition the arrow symbol does not indicate a wrong XPOS. setting. A minus sign (-) placed in front of the
percentage value, is displayed by the readout for
POSTTRIGGER mode indication.
Proceeding from 100% pre-trigger, the instrument switches over to 75% POSTTRIGGER (“PT-75%”) after the
PTR button is pressed. Then the trigger point is 7.5 div to
the left of the trace start on the screen. This means that
the signal capture starts 7.5 x time deflection coefficient
after the trigger event occurred.
(8) PTR - Pushbutton for PRE and POST Trigger selection.
This function is not available in analog mode.
The PRETRIGGER function is used to capture signals that
occur prior to a trigger event, making the pre-history visible.
In contrast to this function, the POSTTRIGGER is used to
capture signals occurring after the trigger event, which
could not be captured in 0% Pretrigger condition. Due to the
dependence on trigger events, neither function is available
in the trigger independent modes XY and ROLL.
The actual PRE or POSTTRIGGER value is displayed by
the readout and changes, each time the PTR button is
pressed, in the following sequence: PT0%, PT25%,
PT50%, PT75%, PT100%, PT-75%, PT-50%, PT-25%
and back to PT0%. The values refer to the X-axis (graticule)
of the screen display (10% = 1div).
The following description assumes that the X magnifier
(x10) is inactive and the signal display starts on the
leftmost vertical graticule line. It is also assumed that a
trigger mode (source, coupling) is chosen, in which the
trigger point symbol is displayed. In contrast to analog
mode, using pre-trigger the trigger point symbol can be
shifted in X-direction.
Every time the PTR button is pressed the POSTTRIGGER
value changes in 25% steps until PTR-25% is active.
When the PTR button is pressed again, both post and pretrigger are switched off and the readout indicates “PT0%”.
Attention!
In time base settings from 100s/div to 50ms/div the preor post-trigger is automatically switched off (“PT0%) if
refresh (RFR), envelope (ENV) or average (AVM) mode is
active. This is to avoid excessive waiting times.
If the pre- or post-trigger function is required in combination
with those time coefficients, SINGLE (SGL) mode operation must be used.
(9) STOR. MODE - Pushbuttons with associated LEDs.
These functions are not available in analog mode.
If digital SINGLE (SGL) mode has not been chosen, one
of the associated LEDs is lit. The signal capture and
display mode can be selected by pressing one of the
buttons. The mode setting is indicated by one of the LEDs
(RFR, ENV, AVM and ROL) and also displayed by the
readout. The only exception is in XY storage mode. Then
the RFR-LED is lit and the readout displays XY. No other
signal capture and display mode can be chosen in XY
mode.
Subject to change without notice
11
Controls and readout
The desired Yt signal capture mode can be selected by
pressing the upper or lower STOR. MODE button.
The following description presumes that HOLD (HLD) is
not activated and the trigger conditions are met.
(9 ) RFR - stands for refresh operation. In this mode, as in
analog mode, periodically repeating signals can be captured
and displayed.
The signal acquisition is started by triggering the digital
time base. Then the previously captured and displayed
signal will be overwritten with the current signal. This will
be displayed until the digital time base is triggered again.
This is in contrast to analog operation where the screen
remains blank when the time base is not triggered.
In refresh mode, the signal acquisition can be effected
with pre-triggering or post-triggering when a time base
between 20ms/cm and 1µs/cm is selected. The pretriggering or post-triggering will be automatically switched
off (PT0%), with larger time coefficients (100s/cm to
50ms/cm) in order to avoid excessive waiting times. If it
is required to measure with pretrigger or post-trigger in
this time base range, one should select single shot
(SINGLE = SGL).
In XY digital mode the RFR-LED lights. It indicates a
continuous, trigger independent signal acquisition. The
trigger circuit is switched off.
The accuracy of the mean value evaluation increases as
the number the number of signal acquisition scans used
for evaluation is increased. One can select the number
between 2 and 512. The selected setting is displayed in
the readout. Of course, with increasing accuracy the
time required for this also increases.
To select a different value briefly press both STOR.MODE pushbuttons simultaneously. The AV... display in
the readout flashes indicating the setting mode. Now, the
value can be changed by briefly pressing the upper or
lower STOR. MODE button. The setting mode can be
exited by again briefly pressing the two buttons
simultaneously. The setting mode will also be switched
off automatically if none of the two buttons is actuated
during about 10 seconds.
The averaging begins anew after briefly pressing the
SINGLE (10) pushbutton (RESET-function).
Attention!
The pretrigger or post-trigger will be automatically
switched off (PT0%) in the time base range from 100s/
cm to 50ms/cm.
(9) ROL - indicates ROLL mode.
In ROLL mode the ROL-LED is lit and the readout displays
“ROL”.
In this mode, the memory contents and thus also the
signal display, are continuously updated. Because signal
capture is untriggered, no idle states arise while waiting
for a new trigger event to start signal capture. With each
signal sampling the new value is shown on the right-hand
edge of the screen, while the previously captured data
are shifted to the left. The leftmost value is shifted out of
the memory and lost.
(9) ENV - is the abbreviation for ENVELOPE operation.
In this mode the minimum and maximum values of the
signal during several signal acquisitions will be determined
and displayed. Except for this display, the ENVELOPE
operation is identical to the refresh operation.
Changes in the signal are easier to measure and are more
visible in ENVELOPE operation. This is valid not only for
amplitude changes but also for frequency variations
(Jitter).
The ENVELOPE evaluation begins anew when the SIN-
GLE (10) button is pressed briefly, to actuate the RESET
(RES) function.
Attention!
The pretrigger or post-trigger will be automatically
switched off (PT0%) in the time base range from 100s/
cm to 50ms/cm.
(9) AVM - indicates Average (mean value) mode.
This operation is effective when the AVM-LED lights up
and the readout displays AV... .
In this case also several signal acquisition scans are
required; hence, it is similar to Refresh operation. The
signal is averaged over the several acquisitions so that
amplitude variations ( e.g. noise) and frequency variations
(Jitter) are minimized or eliminated in the display. The
basic mode ”AV4” is effective when the oscilloscope is
switched on.
The recording can be stopped at any time by selecting the
HOLD (7) function.
ROLL mode can only be used with time coefficients from
100s/div to 50ms/div, as lower time coefficients (faster
time base speeds) are impractical.
If the time base is set to values between 20ms/div and
1µs/div and ROLL mode is selected, the time base will be
automatically set to 50ms/cm. The time deflection
coefficient set previously before switching to ROLL mode
will be internally stored (e.g. 20ms/cm). If ROLL mode
has been selected inadvertently and the TIME/DIV. knob
has not been changed, the time base will be automatically
set to the internally stored coefficient when switching
from ROLL to AVERAGE mode.
(10) SINGLE - Pushbutton with two functions and associated
LEDs.
SINGLE
Pressing and holding the SINGLE pushbutton switches
between SINGLE and:
1. storage mode Yt (time base) or XY operation or
2. analog mode Yt (time base) operation,
dependent on the actual instrument setting.
In this operating mode a single signal acquisition process
or sweep can be started with a trigger, providing the
trigger circuit has been previously activated with RESET.
12
Subject to change without notice
Controls and readout
SINGLE automatically switches to normal triggering (NM
LED lights up). Otherwise the trigger automatic would
start the signal acquisition processes without an input
(trigger) signal.
STORAGE MODE ONLY
SINGLE mode is indicated by the SGL-LED (lit).
If the readout displays additionally the pre- or post-trigger
value (PT...), SINGLE mode in combination with Yt (time
base) storage mode is selected. In connection with XY
storage mode the readout displays the sampling rate (e.g.
40MS/s) and replaces the pre- or post-trigger value (PT...)
by SGL.
When switched over to SGL, single signal acquisition is
in operation. The signal capture currently in progress and
not yet finished will not be terminated but continued to
the end.
Attention!
If SINGLE mode is present in combination with DUAL
mode, the minimum time coefficient is 5µs/div instead
of 1µs/div. Similarly if X-MAG. x10 is operative, then
500ns/div replaces 100ns/div.
ANALOG MODE ONLY
Selecting SINGLE mode switches the current sweep off
and blanks the screen.
Attention!
If time coefficients between 100s/div and 50ms/div are
present the signal acquisition becomes visible at once
as a ROLL display, but the signal acquisition has nothing to do with ROLL mode.
B:
Briefly pressing the SINGLE pushbutton (RESET function)
is also effective if (instead of SINGLE) ENVELOPE or
AVERAGE mode is selected.
In both modes the evaluation / averaging begins anew.
ANALOG MODE ONLY
Capturing single events can also be carried out in analog
mode (e.g. photographing).
Briefly pressing the SINGLE pushbutton activates the
RES-LED in SINGLE mode. The next trigger event then
unblanks the beam and causes one time base sweep.
SINGLE mode is indicated by the lighting SGL-LED and
the readout displaying SGL next to the / (SLOPE) symbol.
Two signals can be displayed during a single sweep only
when it is continuously switched between channel I and
channel II (chopper operation).
RESET (RES)
Briefly pressing the SINGLE pushbutton causes a RESET.
The result depends on the current signal capture mode.
See DUAL (22)
.
STORAGE MODE ONLY
A:
In combination with SINGLE, briefly pressing the SINGLE pushbutton activates the RESET function. Then both
LEDs (SGL and RES) are lit. Whether the RES-LED
flashes once or is lit constantly, depends on:
1. the presence or absence of a trigger signal,
2. the selected time coefficient (time base) and
3. the pre- or post-trigger setting.
After the RESET function is switched on, the signal
acquisition will be effective at once if the HOLD function
is not active. If the pretrigger function is active, the
prehistory must elapse before the trigger event becomes
effective. The signal capture terminates with the trigger
event only with 100% pretrigger setting.
With all other pretrigger and post-trigger settings, the
signal acquisition is not complete when the trigger occurs
and will only be terminated later. After termination the
RES-LED extinguishes but the signal display remains.
Briefly pressing the SINGLE pushbutton (RESET function)
again restarts a new single event capture which then
overwrites the previously recorded display.
Single events recorded in DUAL mode can also be
displayed in the XY mode when switched over to XY
operation.
(11) REFERENCE - Pushbutton with 2 functions, associated
with 2 LED’s (only in Yt (time base) storage mode).
The oscilloscope contains 2 non volatile reference memories.
One reference signal can be displayed together with the
actual signal. The contents of the reference memories are
not erased when the instrument is switched off.
The LEDs associated with the REFERENCE pushbutton
indicate the display of a reference signal and from its
origin. The reference signal is displayed in addition to the
actual signal. With the exception of DUAL mode, there is
no direct relationship between the reference memories (
I or II) and the channels (I or II).
Display
Briefly pressing the REFERENCE pushbutton switches
the reference indicator LEDs and consequently the signal
display in the following sequence:
reference memory off reference memory I reference memory II reference memory off.
Overwrite
After the reference memory has been selected (briefly
pressing) press and hold the REFERENCE pushbutton
until an acoustical signal indicates that the previously
stored reference signal has been overwritten by the
actual signal. During this procedure the instrument can
but must not be set to HOLD (HLD).
In DUAL mode only, REFERENCE memory I relates to
channel I and consequently REFERENCE memory II
relates to channel II. In this case the overwriting procedure
must be performed twice. At first REFRENCE memory I
must be displayed and then overwritten.
Then REFERENCE memory II must be chosen and
overwrittten.
Subject to change without notice
13
Controls and readout
Attention!
As the reference signal is stored and displayed in the
same position as the current signal, it is not immediately
noticeable in most cases.
(12) SAVE / RECALL - Pushbuttons.
The instrument contains 9 non volatile memories. These
can be used by the operator to save instrument settings
and to recall them. This relates to all controls which are
electronically selected.
SAVE
Press the SAVE pushbutton briefly to start the save
procedure. The readout then indicates the letter “S”
followed by a cipher between 1 and 9, indicating the
memory location. If the instrument settings stored in this
memory location must not be overwritten, briefly press
the SAVE or the RECALL pushbutton to select another
memory location. Each time the SAVE pushbutton is
briefly pressed the memory location cipher increases
until the location number 9 is reached. The RECALL
pushbutton function is similar but decreases the memory
location cipher until 1 is reached.
Press and hold SAVE for
approx. 3 seconds to write the instruments settings in the
memory.
channels. After switching GD off and selecting DC input
coupling it is possible to determine the DC content of a
signal by comparing the actual Y position with the
previously determined 0 Volt Y position.
Y-POS. I Symbol
Provided that the readout is displayed and ”DC REF = ON”
is selected in the ”SETUP” submenu ”MISCELLANEOUS”
the 0 Volt reference position is indicated by a ground ( )
symbol. For channel I this symbol is displayed on the left
of the vertical (graticule) center line, if the Y-POS. I trace
position is set within the screen. This allows you to
determine the 0 Volt reference position at any time.
Attention!
In XY mode the 0 Volt reference symbol is automatically
switched off.
STORAGE MODE ONLY
In XY mode the Y-POS. I knob is operative as the X
position control and the X-POS. knob is inactive.
The Y-POS. I knob can be used for shifting the position
of a signal stored with HOLD.
Additional information
relating to this operation is described under HOLD (7)
.
RECALL
To recall a front panel setup, start that procedure by
briefly pressing the RECALL pushbutton. The readout
then indicates the letter “R” and the memory location
number. If required, select a different memory location as
described above. Recall the settings by pressing and
holding the RECALL pushbutton for approx. 3 seconds.
Attention:
Make sure that the signal to be displayed is similar to
the one that was present when the settings were
stored. If the signal is different (frequency, amplitude)
to the one during storage then a distorted display may
result.
If the SAVE or the RECALL pushbutton was depressed
inadvertently, briefly press both pushbuttons at the same
time or wait approx. 10 seconds without pressing either
pushbutton to exit that function.
The setting controls and LED’s for the Y amplifiers, modes,
triggering and time base are located underneath the sector of
the front panel described before.
(13) Y-POS. I - Control knob.
The vertical trace position of channel I can be set with this
control knob. In ADD (addition) mode both (Y-POS. I andY-POS. II) control knobs are active. If the instrument is
set to analog XY mode this control knob is inactive and the
X-POS. knob must be used for a horizontal position shift.
(14) Y-POS. II - Control knob
The vertical trace position of channel II can be set with
this control knob. In ADD (addition) mode both (Y-POS. I
and Y-POS. II) control knobs are active.
If automatic triggering (AT) is present and the input is set
to GD (37), the vertical trace position corresponds with 0
Volt (reference) at the input and can be set to any suitable
position. In ADD mode these conditions apply to both
channels. After switching GD off and selecting DC input
coupling it is possible to determine the DC content of a
signal by comparing the actual Y position with the
previously determined 0 Volt Y position.
Y-POS. II Symbol
Provided that the readout is displayed and ”DC REF =
ON” is selected in the ”SETUP” submenu ”MISCELLANEOUS” the 0 Volt reference position is indicated by
a ground (
on the right of the vertical (graticule) center line, if the Y-POS. II trace position is set within the screen. This allows
you to determine the 0 Volt reference position at any
time.
⊥⊥
⊥) symbol. For channel II this symbol is displayed
⊥⊥
If automatic triggering (AT) is present and the input is set
to GD (33), the vertical trace position corresponds with 0
Volt (reference) at the input and can be set to any suitable
position. In ADD mode these conditions apply to both
14
Attention!
In XY mode the 0 Volt reference symbol is automatically
switched off.
Subject to change without notice
Controls and readout
STORAGE MODE ONLY
The Y-POS. II knob can be used for shifting the position
of a signal stored with HOLD. Additional information
relating to this operation are described under HOLD (7).
(15) NM - AT -
associated NM-LED.
The following description assumes that Yt (time base)
mode has been chosen.
NM - AT selection
Press and hold the pushbutton to switch over from
automatic (peak value) to normal triggering (NM-LED
above the pushbutton lit) and vice versa. If the LED is
dark, automatic (peak value) triggering is selected.
Whether the peak value detection in automatic trigger
mode is automatically activated or not, depends on the
trigger coupling setting (TRIG.MODE (26)). The way the
trigger point symbol in the readout responds on different
LEVEL control knob settings indicates the situation:
1. If the trigger symbol can not be shifted in the vertical
direction when a signal is not applied or the signal height
is not sufficient, the peak value detection is active.
2.Under the condition that the trigger point symbol cannot
be shifted in such a way that it leaves the signal display
on the screen, the peak value detection is active.
3. The peak value detection is switched off if the trigger
point can be set outside the maximum peak values of
the signal, thus causing an untriggered signal display.
Slope selection
Briefly pressing this pushbutton selects which slope of
the signal is used for triggering the time base generator.
Each time this pushbutton is briefly pressed, the slope
direction switches from falling edge to rising edge and
vice versa. The current setting is displayed in the readout
by a slope symbol.
- Pushbutton with a double function and
ANALOG MODE ONLY
The last setting in undelayed time base mode is stored
and still active if triggered DELAY (DTR) time base mode
is selected. This allows for a different slope setting for the
triggered DELAY (DTR) time base mode.
(16) TR - Trigger indicator LED.
ANALOG MODE ONLY
The last setting in undelayed time base mode is stored
and still active if triggered DELAY (DTR) time base mode
is selected. This allows for a different level setting for the
triggered DELAY (DTR) time base mode.
STORAGE MODE ONLY
In storage mode the trigger point symbol also indicates
the post or pre-trigger condition by a horizontal position
Please note PTR (8)
shift.
(18) X-POS. - Control knob.
This control knob enables an X position shift of the
signal(s) in Yt (time base) and analog XY mode. In
combination with X magnification x10 (Yt mode) this
function makes it possible to shift any part of the signal
on the screen.
.
STORAGE MODE ONLY
In XY mode the X-POS. knob is inoperative. The Y-POS.
I (13) must be used for X-position shift.
(19) X-MAG. x10 - Pushbutton and LED.
Each time this pushbutton is pressed the x10 LED located
above is switched on or off. If the x10 LED is lit, the signal
display in all Yt (time base) modes is expanded 10 fold and
consequently only a tenth part of the signal curve is
visible. The interesting part of the signal can be made
visible with aid of the X-POS. (18) control. As the X
expansion results in a higher time base speed (lower time
deflection coefficient), all time and frequency relevant
information in the readout is switched over.
The TR LED is lit in Yt (time base) mode if the triggering
conditions are met. Whether the LED flashes or is lit
constantly depends on the frequency of the trigger signal.
(17) LEVEL - Control knob.
Turning the LEVEL knob causes a different trigger point
setting (voltage). The trigger unit starts the time base
when the edge of a trigger signal crosses the trigger
point. In most Yt modes the trigger point is displayed in
the readout by the symbol on the left vertical graticule
line. If the trigger point symbol would overwrite other
readout information or would be invisible when being set
above or below the screen, the symbol changes and an
arrow indicates in which vertical direction the trigger
point has left the screen.
The trigger point symbol is automatically switched off in
those modes where there is no direct relation between
the trigger signal and the displayed signal.
Subject to change without notice
This pushbutton is not operative in XY mode.
ANALOG MODE ONLY
The expansion is 5 fold if the time base is set to 50ns/div.
Consequently the lowest time deflection coefficient is
10ns/div.
(20) VOLTS/DIV. - Control knob and associated LED.
This control knob for channel I has a double function. The
following description relates to the input attenuator
function (VAR-LED dark).
This control knob is operative in those modes where
channel I is active (CH I, DUAL, ADD and XY) and the
input coupling is not set to ground (GD (33)).
Turning the control knob clockwise increases the sensitivity
(decreases the deflection coefficient) in a 1-2-5 sequence
and decreases the sensitivity (increases the deflection
15
Controls and readout
coefficient) if turned in the opposite direction (ccw.). The
available range is from 1mV/div up to 20V/div. The knob
is automatically switched inactive if the channel related to
it is switched off, or if the input coupling is set to GD
(ground).
The deflection coefficients and additional information
regarding the active channel(s) are displayed in the readout,
i.e. “Y1:...”deflection coefficient, input coupling” (”X:...”
in XY mode). The “:” symbolizes calibrated measuring
conditions and is replaced by the “>” symbol in uncalibrated
conditions.
The previous trigger setting stays as it was, but can be
changed.
All controls related to both channels are active, if the
inputs (31) and (35) are not set to GD (33) (37).
Whether alternated or chopped channel switching is
present depends on the actual time base setting, and is
displayed in the readout.
ALT
displayed in the readout, indicates alternate channel
switching. After each time base sweep the instrument
internally switches over from channel I to channel II and
vice versa. This channel switching mode is automatically
selected if any time coefficient from 200µs/div to 50ns/
div is active.
CHP
indicates chopper mode, whereby the channel switching
occurs constantly between channel I and II during each
sweep. This channel switching mode occurs when any
time base setting between 500ms/div and 500µs/div has
been chosen. The actual channel switching can be changed
to the opposite mode by briefly pressing both CH I (21)
and DUAL (22) simultaneously. If afterwards the time
coefficient is changed, the channel switching is
automatically set to the time coefficient related mode.
(21) CH I - VAR. - Pushbutton with several functions.
CH I
Briefly pressing the CH I button sets the instrument to
channel I (Mono CH I) mode. The deflection coefficient
displayed in the readout indicates the current conditions
(“Y1...”). If neither external nor line (mains) triggering
was active, the internal trigger source automatically
switches over to channel I (TRIG.-LED (23) CH I lits). The
last function setting of the VOLTS/DIV (20) knob remains
unchanged.
All channel I related controls are active if the input (31) is
not set to GD (33).
VAR.
Pressing and holding this pushbutton selects the VOLTS/
DIV. (20) control knob function between attenuator and
vernier (variable). The current setting is displayed by the
VAR-LED located above the knob.
After switching the VAR-LED (20) on, the deflection
coefficient is still calibrated. Turning the VOLTS/DIV.(20) control knob counter clockwise reduces the signal
height and the deflection coefficient becomes
uncalibrated.
The readout then displays i.e. “Y1>...” indicating the
uncalibrated condition instead of “Y1:...”. Pressing and
holding the CH I pushbutton again switches the LED off,
sets the deflection coefficient into calibrated condition
and activates the attenuator function. The previous vernier
setting will not be stored.
The CH I pushbutton can also be pressed simultaneously
with the DUAL (22) button.
(22) DUAL - XY - Pushbutton with multiple functions.
DUAL mode
Briefly pressing this button switches over to DUAL
mode. Both deflection coefficients are then displayed.
Please note item (22).
STORAGE MODE ONLY
The signal acquisition is carried out with both A/D
converters in the two channel (DUAL) digital mode. Since
there is no need to switch channels as in analog mode,
the readout display shows the signal acquisition mode
instead of ALT or CHP.
ADD mode
Addition mode can be selected by briefly pressing the
DUAL (22) and CH II (25) pushbuttons simultaneously.
Whether the algebraic sum (addition) or the difference
(subtraction) of both input signals is displayed, depends
on the phase relationship and the INV (37) setting. As a
result both signals are displayed as one signal. For correct
measurements the deflection coefficients for both
channels must be equal.
Please note “Operating modes
of the vertical amplifiers in Yt mode”.
The readout indicates this mode by a “+” sign located
between both channel deflection coefficients. While the
trigger mode is not affected, the trigger point symbol is
switched off.
The Y-position of the signal can be influenced by both Y-
POS controls (13) and (14).
XY mode
This mode can be switched on or off by pressing and
holding the DUAL button (22).
In XY mode the deflection coefficients are displayed as
“X...” for channel I and “Y...” for channel II, followed by
“XY”. Consequently INPUT CH I (31) serves as an X-input and INPUT CH II (35) is used for Y-deflection. The
X-MAG x10 (19) function is automatically switched off.
The cursor lines may be active, but the trigger point and
the 0 Volt reference symbols are switched off.
ANALOG MODE ONLY
All trigger and time base related controls as well as the YPOS. I (13) knob are deactivated. For X position alteration,the X-POS. (18) knob can be used.
16
Subject to change without notice
Controls and readout
STORAGE MODE ONLY
XY mode is indicated by the readout display ”XY” and the
RFR-LED is lit. No other STOR. MODE (9) can be chosen.
The readout displays the sampling rate (e.g. ”40MS/s”)
which can be selected by the TIME/DIV. (28) knob.
There are gaps in the display of Lissajous figures when
the sampling rate is too high. Too low a sampling rate can
result in a display which does not permit the frequency
ratio of the two signals to be determined. The selection
of a suitable sampling rate is simplified if both the signals
are first displayed in the refresh DUAL mode. The TIME/DIV control should be then so set that at least one period
of each signal is displayed. After this one can switch to the
XY Digital mode.
Attention!
Note the following differences compared to the analog
XY mode:
The Y POS. I (13) control functions as X-Position control
and the X-POS (18) control is disabled.
ALT:
Pressing and holding the button selects alternate triggering
in DUAL mode. Under these conditions both TRIG CH I
and CH II LEDs are lit. As alternate triggering requires
alternate channel operation, alternate channel switching
is set automatically. A change of the time coefficient then
has no affect regarding the channel switching mode. In
addition to the deflection coefficients display, “ALT” is
displayed by the readout instead of “CHP”.
In alternate trigger mode the trigger point symbol is
switched off.
Alternate triggering is not available or automatically
switched off under the following conditions:
ADD (addition) mode,
TVL, TVF and line (mains) trigger coupling and
the time base modes which are available only in analog
mode (search (SEA), delayed (DEL and DTR)).
STORAGE MODE ONLY
Alternate triggering automatically selects ”PT0%” (post
and pre-trigger switched off).
(24) VOLTS/DIV. - Control knob and associated LED.
This control knob for channel II has a double function. The
following description relates to the input attenuator
function (VAR-LED dark).
(23) TRIG. - Pushbutton with double function for trigger
source selection and associated LEDs.
The button and the LEDs are deactivated if line (mains)
triggering is selected or XY operation is chosen. In the
latter case this also applies to all trigger related controls
and LEDs.
With the aid of this button, the trigger source can be
chosen. There are three trigger sources available:
channel I,
channel II (both designated as internal trigger sources)
and the TRIG. EXT. (33) input for external triggering.
The availability of the internal sources depends on the
actual channel mode. The actual setting is indicated by
the associated LED(s).
Briefly pressing the button switches over in the following
sequence:
I - II - EXT - I in DUAL and ADD (addition) mode,
I - EXT - I if mono channel I is present,
II - EXT - II under mono channel II conditions.
Each condition is indicated by the associated LED. The trigger
point symbol is switched off in external trigger condition.
STORAGE MODE ONLY
In combination with ROLL mode, all controls and LEDs
regarding trigger functions are disabled.
This control knob is operative in those modes where
channel II is active (CH II, DUAL, ADD and XY) and the
input coupling is not set to ground (GD (37)).
Turning the control knob clockwise increases the sensitivity
(decreases the deflection coefficient) in a 1-2-5 sequence
and decreases the sensitivity (increases the deflection
coefficient) if turned in the opposite direction (ccw.). The
available range is from 1mV/div up to 20V/div. The knob
is automatically switched inactive if the channel related to
it is switched off, or if the input coupling is set to GD
(ground).
The deflection coefficients and additional information
regarding the active channel(s) are displayed in the readout,
i.e. “Y2:...”deflection coefficient, input coupling” (”Y:...”
in XY mode). The “:” symbolizes calibrated measuring
conditions and is replaced by the “>” symbol in uncalibrated
conditions.
(25) CH II - VAR. - Pushbutton with several functions.
CH II
Briefly pressing the CH I button sets the instrument to
channel II (Mono CH I) mode. The deflection coefficient
displayed in the readout indicates the current conditions
(“Y2...”). If neither external nor line (mains) triggering
was active, the internal trigger source automatically
switches over to channel II (TRIG.-LED (23) CH I lits). The
last function setting of the VOLTS/DIV (24) knob remains
unchanged.
All channel I related controls are active if the input (35) is
not set to GD (37).
VAR.
Pressing and holding this pushbutton selects the VOLTS/
DIV. (24) control knob function between attenuator and
vernier (variable). The current setting is displayed by the
VAR-LED located above the knob.
Subject to change without notice
17
Controls and readout
After switching the VAR-LED (24) on, the deflection
coefficient is still calibrated. Turning the VOLTS/DIV.(24) control knob counter clockwise reduces the signal
height and the deflection coefficient becomes
uncalibrated.
The readout then displays i.e. “Y2>...” indicating the
uncalibrated condition instead of “Y2:...”. Pressing and
holding the CH II pushbutton again switches the LED off,
sets the deflection coefficient into calibrated condition
and activates the attenuator function. The previous vernier
setting will not be stored.
The CH II pushbutton can also be pressed simultaneously
with the DUAL(22) button.
Please note item (22)
.
is automatically set to minimum when the time base is
changed. (For the application of hold off time setting see
the paragraph with the same heading).
The start of the trace can be set with the DEL.POS.
control in the time base modes SEA. (SEARCH) or DEL.(DELAY).
See SEA. / DEL. - ON / OFF (29).
STORAGE MODE ONLY
The DEL.POS. control and the HO-LED are disabled,
since in this mode the hold off time is set to minimum.
The hold off time last used in the analog mode is not
saved. Consequently, the hold off time is set to minimum
when it is switched back to analog mode.
The DEL.POS. control cannot be used for delay setting
since neither SEA. (SEARCH) nor DEL. (DELAY) time
base operation is available.
(28) TIME/DIV. - Control knob.
The time base is set with this knob in the TIME/DIV. field,
and the setting is displayed at the top left in the readout
(e.g. ”T:10µs”). This knob acts as the time base step
switch when the VAR-LED above it is not lit. Then, the
time deflection coefficient can be set in a 1-2-5 sequence
and the time base is calibrated. Rotating anticlockwise
increases the deflection coefficient and rotating clockwise
decreases the deflection. The control acts as a vernier
(fine adjustment) when the VAR-LED is lit.
(26) TRIG. MODE
Pushbuttons and associated indicator LEDs.
Pressing the upper or lower button selects the trigger
coupling. The actual setting is indicated by a TRIG.MODE-LED (26).
Each time the lower TRIG. MODE pushbutton is pressed
the trigger coupling changes in the sequence:
AC DC content suppressed,
DC DC content effective (peak value detection inactive),
HFhigh-pass filter cuts off frequencies below
approx. 50kHz (trigger point symbol switched off),
LFlow-pass filter cuts off frequencies above
approx. 1.5kHz, in combination with automatic
triggering AC or normal triggering DC coupled,
TVL TV signal, line pulse triggering,
trigger point symbol switched off,
TVF TV signal, frame pulse triggering,
trigger point symbol switched off.
~line/mains triggering, trigger point symbol
and TRIG. LED (23) are switched off.
In some trigger modes such as alternate triggering, some
trigger coupling modes are automatically disabled and
can not be selected.
(27) DEL.POS.
Rotary knob with two functions and related HO-LED.
The DEL.POS. knob functions as a hold off time control,
when the time base is not working in the SEA. (SEARCH)
or in DEL. (DELAY) mode. The HO-LED is not lit when the
hold off time is set to minimum. The HO-LED lights up and
the hold off time increases as the knob is rotated clockwise.
A signal sounds on reaching the maximum hold off time.
Similarly in the opposite direction until minimum hold off
time is reached (HO-LED extinguishes). The hold off time
The following description refers to the function as a time
base switch.
Attention
The different deflection coefficient ranges between the
analog and digital time bases require special consideration when switching between analog and digital
modes. These are described in section (7).
ANALOG MODE ONLY
Time deflection coefficients between 500ms/div. and
50ns/div. in a 1-2-5 sequence can be selected without XMAG. x10 magnification. Time delay between 120ms and
200ns can be selected in the ”SEA” (SEARCH) mode.
Time deflection coefficient range in the ”DEL” (DELAY)
mode extends from 20ms/div. to 50ns/div.
STORAGE MODE ONLY
The time base can be set to deflection coefficients
between 100s/div. and 1µs/div in storage mode, if XMAG x10 is inactive.
(29) SEA./DEL. - ON/OFF pushbutton
The pushbutton is operative only in analog mode.
This button is used to switch between delayed and undelayed
time base. The delayed time base operation enables a
magnified display in X-direction which is otherwise only
possible with a second time base. Pressing and holding the
button switches from normal time base mode to SEA.
(SEARCH), when currently neither ”SEA” (SEARCH) nor
”DEL” (DELAY) operation or ”DTR” (triggered DELAY) iseffective. Afterwards, it can be switched between SEA. and
DEL. by briefly pressing the button.
These operating modes are indicated in the readout to the
right of the trigger slope indication thus:
In case of SEARCH , ”SEA” will be displayed;
In untriggered DELAY mode, ”DEL” (DEL.) and
in triggered DELAY mode, ”DTR” (DEL.TRIG.).
18
Subject to change without notice
Controls and readout
None of these will appear in the readout in undelayed time
base operation.
When ”SEA”, ”DEL” or ”DTR” mode is effective, pressing
and holding the button switches over to undelayed time
base.
Provided that neither ”SEA” nor ”DEL” resp. ”DTR” are
active, the following description assumes that:
1. X-MAG. x10 is switched off,
2. the trace starts on the left vertical graticule line and
3. the part of signal to be expanded must be displayed
within the (horizontal) range 2 up to 6 divisions after
trace start position.
SEA
In SEA. (SEARCH) mode, the hold off time is automatically
set to minimum and for the first few divisions the trace
is blanked. The trace will then be unblanked. The point at
which the trace starts can be varied with DEL.POS (fine
adjustment) from about 2 to 6 divisions. The blanked
section serves as a guide to the delay time. The delay time
is based on the current time deflection coefficient setting
and can also be coarsely set with the TIME/DIV control
(range: 20ms to 100ns).
DEL
Pressing the button briefly switches over from ”SEA” to
”DEL” (DELAY) mode. Now the trace starts at the left
vertical graticule line, beginning with that part of the
signal that was previously unblanked first when ”SEA”
was active. From that position, the signal display can be
expanded in the X-direction by rotating the TIME/DIV
control clockwise and thus decreasing the time deflection
coefficient. If a part of the signal of interest goes beyond
the right edge, it can be brought within the screen and
made visible by DEL.POS knob. Increasing the time
deflection coefficient beyond that used in the ”SEA”(SEARCH) mode is not possible.
In the case of untriggered ”DEL” (DELAY) mode (29),
briefly pressing the button switches over to ”DTR”
(triggered DELAY mode). Thereby, the previously active
settings, Automatic/normal triggering (15), trigger
LEVEL (17), trigger slope (15) and trigger coupling
(26) will be stored.
The instrument will be automatically switched to normal
triggering (NM) and DC trigger coupling in ”DTR” mode.
Subsequently the trigger LEVEL setting and the trigger
slope should be so adjusted that the signal for delayed
trigger can trigger the time base. Without triggering the
screen will remain blank.
Briefly pressing the button again switches back to
(untriggered) DEL. operation.
VAR.
Pressing and holding the pushbutton changes the function
of the TIME/DIV. knob.
The TIME/DIV. knob (28) can function as a time deflection
coefficient switch (1-2-5 sequence) or as a time vernier
(fine adjustment). The current function is indicated by the
VAR-LED. The TIME/DIV. knob functions as a vernier
when the VAR-LED is switched on, but the time base
setting remains calibrated until the (vernier) knob is
operated. The readout now indicates ”T>...” instead of
”T:...”. Rotating further anticlockwise increases the time
deflection coefficient (uncalibrated) until the maximum is
reached indicated by a beep. Rotating the knob clockwise
has the opposite effect. Now, the vernier is again in the
calibrated position and the symbol ” >” will be replaced
by symbol ” :”.
The function of the knob can be switched back to normal
(calibrated) time base, pressing and holding the button.
Underneath the front panel sector described above, the BNC
sockets and four pushbuttons are located.
In the untriggered ”DEL” (DELAY) mode, a trigger event
does not start the trace at once but only starts the delay
time. After the delay time has elapsed the trace is started.
DTR
In triggered DELAY mode (DTR), to start the sweep, a
signal suitable for triggering must appear after the delay
time. The trace will be started if the instrument settings
(e.g. LEVEL setting) enable a triggering.
- VAR. (30)
(30) DEL.TRIG. - VAR. - Pushbutton with two functions
The pushbutton is operative only in analog mode.
.
See DEL.TRIG.
(31) INPUT CH I - (HOR. INP. (X)) - BNC socket.
This BNC socket is the signal input for channel I (INPUTCH I). In XY mode, signals at this input are used for the Xdeflection. The outer (ground) connection is galvanically
connected to the instrument ground and consequently to
the safety earth contact of the line/mains plug.
(32) AC / DC - Pushbutton with two functions.
Input coupling:
Briefly pressing this pushbutton switches over from AC
~ symbol) to DC ( = symbol) input coupling and vice versa,
(
if the input is not switched to GD (33). The AC/DC setting
is displayed in the readout with the deflection coefficient.
Probe factor:
Pressing and holding the pushbutton selects the indicated
deflection coefficient of channel I displayed in the readout,
between 1:1 and 10:1. In condition 10:1 the probe factor
is thus indicated by a probe symbol displayed by the
readout in front the channel information (e.g. “probesymbol”, Y1...). In the case of cursor voltage measurement, the probe factor is automatically included.
Subject to change without notice
19
Controls and readout
Please note:
The probe symbol should not be activated unless a x10
(10:1) attenuator probe is used.
(33) GD - Pushbutton
Each time this pushbutton is pressed briefly, the input is
switched from active to inactive and vice versa. It is
displayed in the readout as an earth (ground) symbol
instead of the deflection coefficient and the ~ (AC) or =
(DC) symbol.
The GD setting disables the input signal, the AC/DC (32)
pushbutton and the VOLTS/DIV (20) knob. In automatic
trigger mode the undeflected trace is visible representing
the 0 Volt trace position.
(34) Ground socket
4mm banana socket galvanically connected to safety
earth.
This socket can be used as reference potential connection
for DC and low frequency signal measurement purposes
and in COMPONENT TEST mode.
(35) INPUT CH II - BNC socket.
This BNC socket is the signal input for channel II (INPUTCH II). In XY mode, signals at this input are used for the
Y-deflection. The outer (ground) connection is galvanically
connected to the instrument ground and consequently to
the safety earth contact of the line/mains plug.
(36) AC / DC - Pushbutton with two functions.
Input coupling:
Briefly pressing this pushbutton switches over from AC
~ symbol) to DC ( = symbol) input coupling and vice versa,
(
if the input is not switched to GD (37). The AC/DC setting
is displayed in the readout with the deflection coefficient.
See Y-POS. I (13).
trigger mode the undeflected trace is visible representing
the 0 Volt trace position.
INV.
Pressing and holding this pushbutton switches the channel
II invert (INV.) function on or off. The invert “on” condition
is indicated by the readout with a horizontal bar above
“Y2” (Yt mode) or ”Y” (XY mode). The invert function
causes the signal display of channel II to be inverted by
180°.
(38) TRIG. EXT. / INPUT (Z) - BNC socket with two functions.
The outer (ground) connection is galvanically connected
to the instrument ground and consequently to the safety
earth contact of the line/mains plug. The input impedance
is approx. 1MΩ II 20pF.
TRIG. EXT.
This BNC socket is the external trigger signal input, if
external triggering is selected. Briefly pressing the TRIG.(23) pushbutton, until the TRIG. “EXT” -LED (23) is lit,
switches the external trigger input active.
The trigger coupling depends on the TRIG. MODE (26)
setting.
See Y-POS. II (14).
ANALOG MODE ONLY
Z- Input
If neither COMPONENT TEST nor external trigger coupling
(TRIG. EXT.) is chosen, the socket is operative as a Z
(trace intensity modulation) input.
High TTL level (positive logic) effects blanking, low level
gives unblanking. No higher voltages than +5 Volt are
permitted.
Below the CRT there are the controls for the readout, the
component tester and the squarewave calibrator with their
outputs.
Probe factor:
Pressing and holding the pushbutton selects the indicated
deflection coefficient of channel II displayed in the readout,
between 1:1 and 10:1. In condition 10:1 the probe factor
is thus indicated by a probe symbol displayed by the
readout in front the channel information (e.g. “probesymbol”, Y2...). In the case of cursor voltage measurement, the probe factor is automatically included.
Please note:
The probe symbol should not be activated unless a x10
(10:1) attenuator probe is used.
(37) GD - INV. - Pushbutton with two functions.
GD
Each time this pushbutton is pressed briefly, the input is
switched from active to inactive and vice versa. It is
displayed in the readout as an earth (ground) symbol
instead of the deflection coefficient and the
(DC) symbol.
The GD setting disables the input signal, the AC/DC (36)
pushbutton and the VOLTS/DIV (24) knob. In automatic
~ (AC) or =
20
Please note:
The following description of the cursor related controls
assumes that the readout is visible and the component
tester is switched off.
(39) PRINT / MENU - Pushbutton with 2 functions.
STORAGE MODE ONLY
PRINT
Briefly pressing the pushbutton starts a documentation
(hardcopy) if the following preconditions are met:
1. The oscilloscope must be connected to the external
HAMEG interface HO79-6.
2.The software version installed in HO79-6 should not be
< V2.00.
The device used for documentation (e.g. printer, plotter)
must be connected with one of the HO79-6 ports. The
documentation includes the signal display, the graticule,
the measurement parameters and additional information
such as oscilloscope type and HO79-6 software version.
Subject to change without notice
Controls and readout
The PRINT function replaces the actuation of the HO79-
6 ”START” pushbutton, which may not be accessible
(e.g. rack mount).
For further information please note the HO79-6 manual.
ANALOG and DIGITAL MODE
MENU
Pressing and holding the pushbutton activates the display
of the MAIN MENU. It contains the submenus SETUP,CALBRATE and HO79 if connected.
Once a menu is displayed, the following pushbuttons are
of importance:
1. SAVE and RECALL (12) pushbutton.
Briefly pressing selects the submenu or an item within
the submenu.
2. SAVE (12) pushbutton SET function.
Pressing and holding the SAVE (12) pushbutton calls
(set‘s) the menu or the previously selected item. In
those cases where the item is marked with ON / OFF
the setting changes from ON to OFF or vice versa.
In some cases the called function is not performed at
once and a warning is displayed to protect from calling
the function inadvertently. Then the function can be
called by pressing and holding the SAVE pushbutton
(SET function) again, otherwise if the function was
called inadvertently the proceeding can be cancelled by
pressing the AUTOSET (3) pushbutton.
3. AUTOSET (3) pushbutton.
Each time the AUTOSET pushbutton is pressed the
menu is switched back one step until MAIN MENU is
displayed. Then pressing the AUTOSET pushbutton
again switches the menu operation off and the AUTO-SET is automatically set to the normal function.
applied causing an X and a Y deflection. The deflection
coefficient selected for each channel may be different,
thus as in DUAL mode the
requires a channel selection. Under channel I (X signal)
measuring condition the cursor lines are displayed as
vertical lines and the readout displays “
pressing the pushbutton changes to channel II (Y signal)
voltage measurement. Then the cursor lines are displayed
as horizontal lines and the readout indicates “
In CH I or CH II mono mode, only one deflection coefficient
is present and there is no requirement to select between
different deflection coefficients. Consequently this
function is disabled in combination with
∆∆
1/
∆t:
∆∆
Briefly pressing the button selects between time and
frequency measurement, if the
In calibrated time base condition the readout displays
∆∆
“
∆t:...” if time measurement is chosen. After briefly
∆∆
pressing the pushbutton “f:...” (frequency) is displayed.
If the time base is uncalibrated the readout displays
∆∆
“
∆t>...” or “f<...”.
∆∆
∆∆
∆V cursor measurement
∆∆
∆∆
∆VX...”. Briefly
∆∆
∆∆
∆VY...”.
∆∆
∆∆
∆V measurements.
∆∆
∆∆
∆V function is not present.
∆∆
Attention!
Time and frequency measurements are not possible in
XY operation, since in this mode the time base is
switched off.
(41) TRK - CURSOR - Track function.
Briefly pressing simultaneously both buttons ON/OFF CH I/II - 1/
from single cursor line operation to track mode and vice
versa.
In TRK (track) mode both cursor lines are indicated as
active (both dotted lines uninterrupted).
∆∆
∆t (40) and I/II -
∆∆
∆∆
∆∆
∆V/
∆t (42) switches over
∆∆
∆∆
(40) ON/OFF - CH I/II - 1/
Pushbutton with several functions.
ON/OFF:
Pressing and holding the pushbutton switches both cursor
lines on or off. As the cursor lines are part of the readout,
they are visible only if the readout is switched on.
CH I/II
This function is required and available only in DUAL and
XY mode in combination with
it is only in these modes that different deflection
coefficients (VOLTS/DIV.) may be selected. The probe
factor setting (32) (36) is automatically taken into account.
Briefly pressing the button selects between the deflection
coefficients of channel I and channel II. The measured
result is displayed by the readout with “
The cursor settings relate to the signal of the selected
channel.
The position of the active CURSOR line(s), adjustable by
the CURSOR lever (43), can be determined by activating
the I/II- (42) resp. TRK (41) function.
In XY mode the instrument is automatically set to
measurement. In this mode two signals are normally
∆∆
∆t
∆∆
∆∆
∆V (42) measurement, as
∆∆
∆∆
∆V1...” or “
∆∆
∆∆
∆V2...”.
∆∆
∆∆
∆V
∆∆
∆∆
(42) I/II -
I/II:
Briefly pressing this button changes the active (controllable) cursor in the sequence I - II - I, if TRK (track) mode
is not active. The active cursor is indicated by a continuously
dotted line. An interrupted dotted line indicates the
inactive cursor.
∆∆
∆V /
∆∆
Pressing and holding this pushbutton changes from voltage
to time (or frequency) measurement and vice versa. In XY
mode the instrument is automatically set to
time base is disabled and consequently time or frequency
measurements can not be performed.
∆∆
∆V:
∆∆
∆∆
∆V /
∆t - Pushbutton with two functions.
∆∆
∆∆
∆∆
∆t:
∆∆
∆∆
∆V, as the
∆∆
Please note!
∆∆
In all
∆
V (voltage) measurement conditions, the division
∆∆
ratio of the probe(s) must be taken into account. The
voltage value displayed in the readout must be
multiplied by 100 if e.g. a x100 (100:1) probe is used. In
case of x10 (10:1) probes, the probe factor can be
automatically included (see item (32) and (36)).
item 1:
Time base mode (CH I or CH II single channel mode,
DUAL and ADD). In the voltage measurement condition,
the cursors are displayed as horizontal lines and the result
is displayed in the readout.
Subject to change without notice
21
Menu
Single channel mode (channel I or channel II)
∆∆
The
∆V measuring result is automatically related to the
∆∆
deflection coefficient of the active channel. The readout
displays “
Dual mode
The CURSOR lines must be set on the channel I or
channel II signal. As the deflection coefficients may be
different, it will be required to select between the deflection
coefficients of channel I and II.
(40)
Addition (ADD) mode
In ADD (addition) mode normally two input signals are
displayed as one signal (sum or difference). As the result
can only be determined if both (calibrated) deflection
coefficients are equal, the CH I/II (40) selection function
is deactivated. In that case the readout indicates “
without any additional channel information. Different
deflection coefficient settings or uncalibrated deflection
coefficient(s) are indicated in the readout as “Y1<>Y2”.
item 2: XY mode
In XY mode the instrument is automatically set to
measurement. The deflection coefficient selected for
each channel may be different, thus as in DUAL mode the
∆∆
∆V cursor measurement requires a channel selection.
∆∆
Under channel I (X signal) measuring condition the cursor
lines are displayed as horizontal lines and the readout
displays “
pushbutton selects channel II (Y signal) measuring. Then
the cursor lines are displayed as vertical lines and the
readout indicates “
∆∆
∆t:
∆∆
In time or frequency measurement condition two vertical
cursor lines are displayed. The measurement result is
indicated as “
(frequency measurement).
∆∆
∆V1...” or “
∆∆
∆∆
∆V2...”.
∆∆
Please note item CH I/II
.
∆∆
∆VX...”. Briefly pressing the CH I/II (40)
∆∆
∆∆
∆VY...” .
∆∆
∆∆
∆t...” (time measurement) or “f...”
∆∆
Please note item (40) 1/∆t.
∆∆
∆V...”
∆∆
∆∆
∆V
∆∆
NOTE:
For frequency measurement, the distance between the
cursors must equal exactly one signal period.
If the pushbutton is released it is approx. 1kHz and can be
switched over (depressed) to approx. 1MHz. The pulse
duty factor may deviate from 1:1 and the frequency is not
calibrated.
(45) CT - Pushbutton and banana jack.
Pressing the pushbutton switches the instrument over
from oscilloscope to component test mode and vice
versa.
In component test mode, all controls are deactivated with
the exception of the CT button, AUTOSET (2) and
INTENS (4). All LEDs except “A” or “RO” (associated
with the INTENS knob) are dark. The readout displays
only “CT” in this condition.
One test lead is connected to the CT socket. The second
test lead uses the ground socket (34).
Please note
“Component Tester”.
The maximum test voltage is approx. 20Vpp under open
circuit conditions, while the max. test current under short
circuit condition is approx. 20mApp.
Menu
The instrument software contains several menus. The controls
regarding the menus are described under item (39) PRINT /MENU in section “Controls and Readout”.
The following menus, submenus and items within the
submenus are available:
1. MAIN MENU
1.1 CALIBRATE
Information regarding this can be found in the section ”Service
Instructions” item ”Adjustments”.
1.2 SETUP
This menu allows changes to the default settings regarding the
instrument behavior during operation. The SETUP menu contains
the submenus ”MISCELLANEOUS” and ”FACTORY”.
In XY mode the time base is switched off. Consequently
time or frequency measurement is disabled.
(43) CURSOR - Center biased lever
The active cursor line (in track mode: both lines) can be
shifted in the required direction, until the graticule limits
are reached. The directions are marked on the front panel
and depend on the selected measurement (
∆∆
∆t = f).
∆∆
Which of two shift speeds is used, depends on how far
the CURSOR knob is pressed. With slight knob pressure
the cursor line(s) moves slowly. If the knob is pressed to
the full extent the cursor moves fast.
(44) CAL. - Pushbutton and concentric socket.
A squarewave signal of 0.2Vpp ±1% is available from the
socket for probe adjustment purposes.
The signal frequency depends on the pushbutton setting.
∆∆
∆V,
∆∆
∆∆
∆t or 1/
∆∆
22
1.2.1 MISCELLANEOUS contains:
1.2.1.1 CONTROL BEEP ON/OFF.
In OFF condition the acoustic signals actuated by the control
limits are switched off.
Note:
The default setting is ON. If different conditions are
required the setting must be performed each time after
switching the oscilloscope on.
1.2.1.2 ERROR BEEP ON/OFF.
Acoustic signals indicating faulty control operation are
suppressed in OFF condition.
Note:
The default is ON. If different conditions are required
the setting must be performed each time after switching
the oscilloscope on.
1.2.1.3 QUICK START ON/OFF.
In condition ON the HAMEG logo and the menus will not be
displayed after switching the instrument on. Then the
instrument is quickly ready for operation. To change from
QUICK START OFF to ON, press and hold the AUTOSET
button when switching the instrument on until the menus
become visible and change the setting from ON to OFF .
Subject to change without notice
First Time Operation
1.2.1.4 DC REF ON/OFF.
If ON is selected and Yt (time base) mode is present, the
readout displays a ground symbol (
evaluation and determination of DC contents by indicating the
0 Volt reference position.
1.2.1.5 TRIG SYMBOL ON/OFF.
In most of the Yt (time base) modes the readout displays a trigger
point symbol which will not be displayed in condition OFF.
1.2.2 FACTORY contains
1.2.2.1 LOAD S/R DEFAULT (S/R = SAVE/RECALL).
This overwrites all SAVE/RECALL memories with the following
instrument setting: Single channel operation (“Y1:500mV~”),
time base (“T:100µs”), automatic peak value triggering and
AC trigger coupling.
1.2.2.2 RESTORE FACTORY DEFAULT.
If inadvertently an adjustment was performed in CALIBRATE
MENU condition, which was not stored by OVERWRITE
FACTORY DEFAULT, the factory adjustments can be recalled
by this function.
1.2.2.3 OVERWRITE FACTORY DEFAULT.
⊥⊥
⊥). This symbol eases the
⊥⊥
Attention!
Calling this function causes the factory adjustments to
be overwritten by new data. The factory adjustment
then is lost and can not be recalled by RESTORE
FACTORY DEFAULT.
This function is only intended for those cases in which
an adjustment can be performed with very expensive
”0% error” calibrators, to adjust the instrument for
optimum tolerance when operated under extreme environmental conditions.
1.3 HO79
This information will only be displayed if the interface HO79-6 is connected to the oscilloscope.
For further information
please note the HO79-6 manual which is supplied with the
interface.
If the AUTO SET function was not used and only a spot
appears (CAUTION! CRT phosphor can be damaged), reduce
the intensity immediately and check that the XY mode is not
selected (XY not displayed in the readout).
To obtain the maximum life from the cathode-ray tube, the
minimum intensity setting necessary for the measurement in
hand and the ambient light conditions should be used.
Particular care is required when a single spot is displayed, as
a very high intensity setting may cause damage to the
fluorescent screen of the CRT. Switching the oscilloscope off
and on at short intervals stresses the cathode of the CRT and
should therefore be avoided.
The instrument is so designed that even incorrect operation
will not cause serious damage.
Trace Rotation TR
In spite of Mumetal-shielding of the CRT, effects of the earth’s
magnetic field on the horizontal trace position cannot be
completely avoided. This is dependent upon the orientation of
the oscilloscope on the place of work. A centered trace may
not align exactly with the horizontal center line of the graticule.
A few degrees of misalignment can be corrected by a
potentiometer accessible through an opening on the front
panel marked TR.
Probe compensation and use
To display an undistorted waveform on an oscilloscope, the
probe must be matched to the individual input impedance of
the vertical amplifier.
For this purpose a square wave signal with a very fast rise time
and minimum overshoot should be used, as the sinusoidal
contents cover a wide frequency range.
The built-in calibration generator provides a square wave signal
with a very fast risetime (<4ns), and switch-selectable
frequencies of approx. 1kHz and 1MHz from the output socket
below the CRT screen.
First Time Operation
The following text assumes that the “SAFETY” section of this
manual has been read carefully and understood.
Each time before the instrument is put into operation check that
the oscilloscope is connected to protective earth. For that reason
the power cable must be connected to the oscilloscope and the
power outlet. Then the test lead(s) must be connected to the
oscilloscope input(s). Check that the device under test is switched
off and connect the test lead(s) to the test point(s). Then switch
on the instrument and afterwards the device under test.
The oscilloscope is switched on by depressing the red POWER
pushbutton. After a few seconds the HAMEG logo and the
instrument software release is displayed on the screen. As
long as the HAMEG logo is visible different internal checks are
made. Thereafter the instrument will revert to its last used
operating mode.
If after approx. 20 seconds no trace is visible, the AUTO SET
pushbutton should be pressed briefly. This selects the Yt
mode and medium trace and readout intensity (please note
“AUTO SET”). Adjust Y-POS.I and X-POS. controls to center
the baseline. Adjust INTENS. (intensity) and FOCUS controls
for medium brightness and optimum sharpness (input(s)
grounded) of the trace. The oscilloscope is now ready for use.
Subject to change without notice
As the squarewave signals are used for probe compensation
adjustments, neither the frequency accuracy nor the pulse
duty factor are of importance and therefore not specified.
The output provides 0.2Vpp ±1% (tr <4ns) for 10:1 probes.
When the Y deflection coefficient is set to 5mV/div, the
calibration voltage corresponds to a vertical display of 4
divisions (10:1 probe).
The output socket has an internal diameter of 4.9mm to
accommodate the internationally accepted shielding tube
diameter of modern Probes and F-series slimline probes. Only
this type of construction ensures the extremely short ground
connections which are essential for an undistorted waveform
reproduction of non-sinusoidal high frequency signals.
Adjustment at 1kHz
The C-trimmer adjustment (low frequency) compensates the
capacitive loading on the oscilloscope input. By this adjustment,
the capacitive division assumes the same ratio as the ohmic
voltage divider to ensure the same division ratio for high and
low frequencies, as for DC. (For 1:1 probes or switchable
probes set to 1:1, this adjustment is neither required nor
possible). A baseline parallel to the horizontal graticule lines is
essential for accurate probe adjustments. (
rotation TR”
).
See also “Trace
23
First Time Operation
Connect the probes (Types HZ51, 52, 54, or HZ36) to the CH
I input. Set the deflection coefficient to 5mV/div and the input
coupling to DC. The time deflection coefficient should be set
to 0.2ms/div. All deflection coefficients should be calibrated.
Plug the probe tip into the calibrator output socket.
Approximately 2 complete waveform periods are displayed on
the CRT screen. The compensation trimmer should be adjusted.
The location of the low frequency compensation trimmer can
be found in the probe information sheet. Adjust the trimmer
with the insulated screw driver provided, until the tops of the
square wave signal are exactly parallel to the horizontal graticule
lines (see 1kHz diagram). The signal height should then be 4div
± 0.16div (= 4% (oscilloscope 3% and probe 1%). During this
adjustment, the signal edges will remain invisible.
Adjustment at 1MHz
Probes HZ51, 52 and 54 can also be HF-compensated. They
incorporate resonance de-emphasing networks (R-trimmer in
conjunction with inductances and capacitors) which permit
probe compensation in the range of the upper frequency limit
of the vertical oscilloscope amplifier. Only this compensative
adjustment ensures optimum utilization of the full bandwidth,
together with constant group delay at the high frequency end,
thereby reducing characteristic transient distortion near the
leading edge (e.g. overshoot, rounding, ringing, holes or
bumps) to an absolute minimum.
After completion of the HF-adjustment, the signal amplitude
displayed on the CRT screen should have the same value as
during the 1kHz adjustment.
Probes other than those mentioned above, normally have a
larger tip diameter and may not fit into the calibrator output.
Whilst it is not difficult for an experienced operator to build a
suitable adapter, it should be pointed out that most of these
probes have a slower risetime with the effect that the total
bandwidth of scope together with probe may fall far below that
of the oscilloscope. Furthermore, the HF-adjustment feature
is nearly always missing so that waveform distortion can not
be entirely excluded. The adjustment sequence must be
followed in the order described, i.e. first at 1kHz, then at
1MHz. The calibrator frequencies should not be used for time
base calibration. The pulse duty cycle deviates from 1:1 ratio.
Prerequisites for precise and easy probe adjustments, as well
as checks of deflection coefficients, are straight horizontal
pulse tops, calibrated pulse amplitude, and zero-potential at
the pulse base. Frequency and duty cycle are relatively uncritical.
For interpretation of transient response, fast pulse risetimes
and low-impedance generator outputs are of particular
importance.
Providing these essential features, as well as switch-selectable
output-frequencies, the calibrator of the instrument can, under
certain conditions, replace expensive squarewave generators
when testing or compensating wideband-attenuators or amplifiers. In such a case, the input of an appropriate circuit will
be connected to the CAL.-output via a suitable probe.
The voltage provided at a high-impedance input (1MΩ II 1530pF) will correspond to the division ratio of the probe used
(10:1 = 20mVpp output). Suitable probes are HZ51, 52, and 54.
Using the probes HZ51, 52 and 54, the full bandwidth of the
oscilloscope can be utilized without risk of unwanted waveform
distortion.
Prerequisite for this HF compensation is a square wave
generator with fast risetime (typically 4ns), and low output
impedance (approx. 50Ω), providing 0.2V at a frequency of
approx. 1MHz. The calibrator output of this instrument meets
these requirements when the CAL. pushbutton is depressed.
Connect the probe to CH.I input. Depress the CAL. pushbutton
for 1MHz. Operate the oscilloscope as described under 1kHz
but select for 0.2µs/div time deflection coefficient setting.
Insert the probe tip into the output socket. A waveform will be
displayed on the CRT screen, with leading and trailing edges
clearly visible. For the HF-adjustment now to be performed, it
will be necessary to observe the rising edge as well as the
upper left corner of the pulse top. The location of the high
frequency compensation trimmer(s) can also be found in the
probe information sheet. These R-trimmer(s) have to be
adjusted such that the beginning of the pulse is as straight as
possible. Overshoot or excessive rounding are unacceptable.
The adjustment is relatively easy if only one adjusting point is
present. In case of several adjusting points the adjustment is
slightly more difficult, but causes a better result. The rising
edge should be as steep as possible, with a pulse top remaining
as straight and horizontal as possible.
Operating modes of the vertical
amplifiers in Yt mode
The most important controls regarding the operation modes of
the vertical amplifiers are the pushbuttons: CH I (21), DUAL(22) and CH II (25).
“ Controls and readout”
In most cases oscilloscopes are used to display signals in Yt
mode. Then the signal amplitude deflects the beam in vertical
direction while the time base causes an X deflection (from left
to right) at the same time. Thereafter the beam becomes
blanked and fly back occurs.
The following Yt operation modes are available:
Single channel operation of channel I (Mono CH I).
Single channel operation of channel II (Mono CH II).
Two channel operation of channel I and channel II (DUAL).
Two channel operation of channel I and channel II -displaying
the algebraic result as the sum or difference - (ADD).
The way the channel switching is determined in DUAL mode
depends on the time base setting and is described in the
section “Controls and readout”.
In ADD mode the signals of both channels are algebraically
added and displayed as one signal. Whether the resulting
display shows the sum or difference is dependent on the
phase relationship or the polarity of the signals and on the
invert function.
Their functions are described in the section
.
24
In ADD mode the following combinations are possible for
In-phase input voltages:
No invert function active = sum.
Channel II invert function active = difference.
Subject to change without notice
Operating modes of the vertical amplifiers in Yt mode
Antiphase input voltages:
No invert function active = difference.
Channel II invert function active = sum.
In the ADD mode the vertical display position is dependent
upon the Y-POS. setting of both channels. The same Y
deflection coefficient is normally used for both channels with
algebraic addition.
Please note that the Y-POS. settings are also added but are not
affected by the INV setting.
Differential measurement techniques allow direct measurement of the voltage drop across floating components (both
ends above ground). Two identical probes should be used for
both vertical inputs. In order to avoid ground loops, use a
separate ground connection and do not use the probe ground
leads or cable shields.
X-Y Operation
The important control for this mode is the pushbutton labeled
DUAL and XY (22). The following description refers to the XY
analog mode.
In XY mode the time base is deactivated. The signal applied to
the input of channel I - front panel marking HOR. INP. (X) -
causes the X deflection. The input related controls (AC/DC,
GD pushbutton and the VOLTS/DIV knob) consequently
affect the X deflection. For X position alteration, the X-POS.
control knob must be used, as the Y-POS. I control is
automatically inactivated. The input deflection coefficient ranges
are the same for both channels, because the X-MAG x10
function is inactive in XY mode.
The bandwidth of the X amplifier, is lower than the Y amplifier
and the phase angle which increases with higher frequencies,
must be taken into account (please note data sheet).
The inversion of the X-input signal is not possible.
Lissajous figures can be displayed in the X-Y mode for certain
measuring tasks:
• Comparing two signals of different frequency or bringing
one frequency up to the frequency of the other signal. This
also applies for whole number multiples or fractions of the
one signal frequency.
• Phase comparison between two signals of the same
frequency.
Phase comparison with Lissajous figures
The following diagrams show two sine signals of the same
frequency and amplitude with different phase angles.
The following must be noted here:
• Because of the periodic nature of the trigonometric functions,
the calculation should be limited to angles ≤90°. However
here is the advantage of the method.
• Due to phase shift, do not use a too high test frequency.
• It cannot be seen as a matter of course from the screen
display if the test voltage leads or lags the reference
voltage. A CR network before the test voltage input of the
oscilloscope can help here. The 1 MΩ input resistance can
equally serve as R here, so that only a suitable capacitor C
needs to be connected in series. If the aperture width of the
ellipse is increased (compared with C short-circuited), then
the test voltage leads the reference voltage and vice versa.
This applies only in the region up to 90° phase shift.
Therefore C should be sufficiently large and produce only a
relatively small just observable phase shift.
Should both input voltages be missing or fail in the X-Y
mode, a very bright light dot is displayed on the screen.
This dot can burn into the phosphor at a too high
brightness setting (INTENS. knob) which causes either
a lasting loss of brightness, or in the extreme case,
complete destruction of the phosphor at this point.
Phase difference measurement
in DUAL mode (Yt)
Phase differences between two input signals of the same
frequency and shape can be measured very simply on the
screen in Dual mode. The time base should be triggered by the
reference signal (phase position 0). The other signal can then
have a leading or lagging phase angle. In alternate triggering
condition, phase difference measurement is not possible. For
greatest accuracy adjust the time base for slightly over one
period and approximately the same height of both signals on
the screen. The Y deflection coefficients, the time base
coefficients and the trigger level setting can be used for this
adjustment, without influence on the result. Both base lines
are set onto the horizontal graticule center line using the YPOS. knobs before the measurement. With sinusoidal signals,
observe the zero (crossover point) transitions; the sine peaks
are less accurate. If a sine signal is noticeably distorted by even
harmonics, or if a DC voltage is present, AC coupling is
recommended for both channels. If it is a question of pulses
of the same shape, read off at steep edges.
It must be noted that the phase difference cannot be determined
if alternate triggering is selected.
Phase difference measurement in DUAL mode
Calculation of the phase angle or the phase shift between the
X and Y input voltages (after measuring the distances a and b
on the screen) is quite simple with the following formula, and
a pocket calculator with trigonometric functions. Apart from
the reading accuracy, the signal height has no influence on the
result.
Subject to change without notice
t = horizontal spacing of the zero transitions in div
T = horizontal spacing for one period in div
25
Triggering and time base
In the example illustrated, t = 3div and T = 10div The phase
difference in degrees is calculated from
or expressed in radians
Relatively small phase angles at not too high frequencies can
be measured more accurately in the X-Y mode with Lissajous
figures.
Measurement of an amplitude modulation
The momentary amplitude u at time t of a HF-carrier voltage,
which is amplitude modulated without distortion by a sinusoidal
AF voltage, is in accordance with the equation
where
The lower side frequency F-f and the upper side frequency F+f
arise because of the modulation apart from the carrier frequency F.
= unmodulated carrier amplitude
U
T
Ω= 2πF = angular carrier frequency
ω= 2πf = modulation angular frequency
m= modulation factor (i.a. 1 100%).
• Triggering: Normal; with LEVEL-setting; internal
(or external) triggering.
If the two values a and b are read from the screen, the
modulation factor is calculated from
where a = UT (1+m) and b = UT (1-m).
The variable controls for amplitude and time can be set
arbitrarily in the modulation factor measurement. Their position
does not influence the result.
Triggering and time base
All controls regarding trigger and time base are located on the
right of the VOLTS/DIV. knobs.
section “Controls and readout”.
Time related amplitude changes on a measuring signal (AC
voltage) are displayable in Yt-mode. In this mode the signal
voltage deflects the beam in vertical direction (Y) while the
time base generator moves the beam from the left to the right
of the screen (time deflection = t).
Normally there are periodically repeating waveforms to be
displayed. Therefore the time base must repeat the time
deflection periodically too. To produce a stationary display, the
time base must only be triggered if the signal height and slope
condition coincide with the former time base start conditions.
A DC voltage signal can not be triggered as it is a constant
signal with no slope.
They are described in the
Figure 1
Amplitude and frequency spectrum for AM display (m = 50%)
The display of an amplitude-modulated HF oscillation can be
evaluated with the oscilloscope provided the frequency
spectrum is inside the oscilloscope bandwidth. The time base
is set so that several cycles of the modulation frequency are
visible. Strictly speaking, triggering should be external with
modulation frequency (from the AF generator or a demodulator).
However, internal triggering is frequently possible with normal
triggering using a suitable trigger level setting and possibly also
using the time vernier (variable) adjustment.
Figure 2
Amplitude modulated oscillation: F = 1MHz; f = 1kHz;
m = 50%; UT = 28.3 mVrms
Triggering can be performed by the measuring signal itself
(internal triggering) or by an external supplied but synchronous
voltage (external triggering).
The trigger voltage should have a certain minimum amplitude.
This value is called the trigger threshold. It is measured with
a sine signal. Except when external trigger is used the trigger
threshold can be stated as vertical display height in div, at
which the time base generator starts, the display is stable, and
the trigger indicator LED lights or flashes.
The internal trigger threshold of the oscilloscope is given as £
5div. When the trigger voltage is externally supplied, it can be
measured in Vpp at that input. Normally, the trigger threshold
may be exceeded up to a maximum factor of 20.
The instrument has two trigger modes, which are characterized
as Automatic Peak and Normal triggering.
Automatic Peak (value) -Triggering
Instrument specific information’s can be drawn from the
items
NM - AT - (15)
“Controls and readout”.
This trigger mode is automatically selected after the AUTO-SET pushbutton was pressed. As the peak value detection
makes no sense in combination with DC and TV (television)
signals, it is switched off automatically in DC, TVL and TVF
trigger coupling conditions as well as in alternate trigger mode.
In this case the automatic is still present, but a wrong trigger
level setting causes an untriggered display.
and
LEVEL (17)
in the section
Oscilloscope setting for a signal according to figure 2:
• Y: CH. I; 20mV/div; AC.
• TIME/DIV.: 0.2ms/div.
26
In automatic trigger mode the sweep generator can run
without test signal or external trigger voltage. A base line will
always be displayed even with no signal. With an applied AC
signal the peak value triggering enables the user to select the
Subject to change without notice
Triggering and time base
voltage point on the trigger signal (trigger point), by the
adjustment of the trigger level control. The control range
depends on the peak to peak value of the signal. This trigger
mode is therefore called Automatic Peak (Value)- Triggering.
Operation of the scope needs only correct amplitude and time
base settings, for a constantly visible trace. Automatic mode
is recommended for all uncomplicated measuring tasks.
However, automatic triggering is also the appropriate operation
mode for the “entry” into difficult measuring problems, e.g.
when the test signal is unknown relating to amplitude, frequency
or shape. Presetting of all parameters is now possible with
automatic triggering; the change to normal triggering can
follow thereafter.
The automatic triggering works above 20Hz. The failure of
automatic triggering at frequencies below 20Hz is abrupt.
However, it is not signified by the trigger indicator LED this is still
blinking. Break down of triggering is best recognizable at the left
screen edge (the start of the trace in differing display height).
The automatic peak (value) triggering operates over all variations
or fluctuations of the test signal above 20Hz. However, if the
pulse duty factor of a square-wave signal exceeds a ratio of
100:1, switching over to normal triggering will be necessary.
Automatic triggering is practicable with internal and external
trigger voltage.
Normal Triggering
Information specific to the instrument are given in the articles
NM - AT (15), LEVEL (17) and TRIG. MODE (26) in the section
”Controls and readout”. The time fine adjustment (VAR.) and
the hold off time setting assist in triggering under specially
difficult signal conditions. The following description refers to
the Yt (time base) analog mode.
With normal triggering, the sweep can be started by AC signals
within the frequency range defined by the trigger coupling
setting.
In the absence of an adequate trigger signal or when the trigger
controls (particularly the trigger LEVEL control) are misadjusted,
no trace is visible, i.e. the screen completely blanked.
When using the internal normal triggering mode, it is possible
to trigger at any amplitude point of a signal edge, even with
very complex signal shapes, by adjusting the trigger LEVEL
control. If the signal applied at the Y input is used for triggering
(internal trigger source), its adjusting range is directly dependent
on the display height, which should be at least 0.5div. If it is
smaller than 1div, the trigger LEVEL adjustment needs to be
operated with a sensitive touch. In the external normal triggering
mode, the same applies to approx. 0.3Vpp external trigger
voltage amplitude.
Other measures for triggering of very complex signals are the
use of the time base variable control and HOLD OFF time
control, hereinafter mentioned.
(Slope)
Please note item
instrument specific information.
The actual slope setting is displayed in the readout. The setting
is not changed by the AUTOSET function. The slope setting
can be changed in delay mode for the delay time base trigger
unit if the delay trigger function is active. The previous slope
setting for the undelayed time base trigger is stored and still
For further information please note “Controls and
active.
readout”.
(15)
in section “Controls and readout” for
The time base generator can be triggered by a rising or falling
edge of the test signal. Whether the rising or the falling edge
is used for triggering, depends on the slope direction setting.
This is valid with automatic and normal triggering. The positive
slope direction means an edge going from a negative potential
and rising to a positive potential. This has nothing to do with
zero or ground potential and absolute voltage values. The
positive slope may also lie in a negative part of a signal.
However the trigger point may be varied within certain limits
on the chosen edge using the LEVEL control. The slope
direction is always related to the input signal and the non
inverted display.
Trigger coupling
Instrument specific information regarding this item can be
noted in the “Data Sheet”.
(26))
and indication are described under “Controls and readout”.
As the automatic triggering does not work below 20Hz, normal
triggering should be used in DC and LF trigger coupling mode.
The coupling mode and accordingly the frequency range of the
trigger signal should meet the signal requirements.
AC: This is the most frequently used trigger mode. The
trigger threshold increases below and above the
frequency limits mentioned in the data sheet. This filter
cuts off both the DC content of the trigger signal and the
lowest frequency range.
DC: In this coupling mode the trigger signal is coupled
galvanically to the trigger unit. Therefore there is no low
frequency limit.
DC triggering is recommended if the signal is to be
triggered with quite slow processes or if pulse signals
with constantly changing pulse duty factors have to be
displayed.
HF: In this coupling mode the transmission range equals a
high pass filter. It cuts off the DC content of the trigger
signal and the lower frequency range.
LF: LF trigger coupling has a low pass filter characteristic. As
in DC trigger coupling, there is no low limit for the pass
frequency range (galvanic coupling) in connection with
normal triggering. The trigger signal is coupled through
a capacitor in the automatic (peak value) triggering and
LF trigger coupling. This results in a low frequency limit
which , however, does not disturb as the limit is below
the repetition frequency of the trigger automatic.
The LF trigger coupling is often more suitable for low
frequency signals than DC trigger coupling because the
noise components of the trigger signals are strongly
suppressed. This avoids or reduces , under borderline
conditions, jitter or double traces especially with very
low signal voltages. The trigger threshold raises
continuously above the pass band.
TV-L: The built-in active TV-Sync-Separator provides the
separation of line sync pulses from the video signal. Even
distorted video signals are triggered and displayed in a
stable manner. This mode is described under paragraph
“Triggering of video signals”.
TV-F: The built-in active TV-Sync-Separator also provides the
separation of frame sync pulses from the video signal.
Even distorted video signals are triggered and displayed
in a stable manner. This mode is described under
paragraph “Triggering of video signals”.
The coupling setting
( TRIG. MODE
Subject to change without notice
27
Triggering and time base
Triggering of video signals
In TV-L and TV-F trigger coupling mode the instrument is
automatically set to automatic triggering and the trigger point
indicator is switched off. As only the separated synchronization
pulses are used for triggering the relationship between the
displayed signal and the trigger signal is lost. In TV-F mode
interference may occur if chopped DUAL mode is chosen or
the readout is active.
Video signals are triggered in the automatic mode. The internal
triggering is virtually independent of the display height, but the
sync pulse must exceed 0.5div height.
The polarity of the synchronization pulse is critical for the slope
selection. If the displayed sync pulses are above the picture
(field) contents (leading edge positive going), then the slope
setting for positive going edges must be chosen. In the case
of sync pulses below the field/line, the leading edge is negative
and consequently the slope selection must be set for falling
edges. Since the invert function may cause a misleading
display, it must not be activated.
On the 2ms/div setting and field TV triggering selected, 1 field
is visible if a 50 fields/s signal is applied. If the hold off control
is in fully ccw position, it triggers without line interlacing
affects caused by the consecutive field.
The display can be expanded by switching on the X-MAG. x10
function so that individual lines are recognizable. Commencing
with a frame synchronizing pulse, the display can also be
expanded with the knob TIME/DIV. But note that this can
result in an apparently unsynchronized display as each frame
(half picture) triggers. This is due to the offset of half a line
between frames.
The influence of the integrating network which forms a trigger
pulse from the vertical sync pulses may become visible under
certain conditions. Due to the integrating network time constant
not all vertical sync pulses starting the trace are visible.
On the 10µs/div setting and line TV triggering selected, approx.
1½ lines are visible. Those lines originate from the odd and
even fields at random.
This trigger mode is present if the ~ LED is lit (26). The trigger
point symbol is inactive in line/mains trigger mode as there is
no direct amplitude relationship between the trigger voltage
and the signal voltage.
A voltage originating from mains/line (50 to 60Hz) is used for
triggering purposes if the trigger coupling is set to ~. This
trigger mode is independent of amplitude and frequency of the
Y signal and is recommended for all mains/line synchronous
signals. This also applies within certain limits, to whole
number multiples or fractions of the line frequency. Line
triggering can also be useful to display signals below the
trigger threshold (less than 0.5div). It is therefore particularly
suitable for measuring small ripple voltages of mains/line
rectifiers or stray magnetic field in a circuit. In this trigger mode
the slope direction pushbutton selects the positive or negative
portion of the line/mains sinewave. The trigger level control
can be used for trigger point adjustment.
Magnetic leakage (e.g. from a power transformer) can be
investigated for direction and amplitude using a search or pickup coil. The coil should be wound on a small former with a
maximum of turns of a thin lacquered wire and connected to
a BNC connector (for scope input) via a shielded cable.
Between cable and BNC center conductor a resistor of at least
100Ω should be series-connected (RF decoupling). Often it is
advisable to shield statically the surface of the coil. However,
no shorted turns are permissible. Maximum, minimum, and
direction to the magnetic source are detectable at the measuring
point by turning and shifting the coil.
Alternate triggering
This trigger mode can be selected in DUAL mode by pressing
and holding the TRIG. pushbutton (23) if the preconditions are
please note “Controls and readout”
met (
chopped DUAL mode, selecting alternate trigger mode
automatically sets the instrument to alternate DUAL mode. If
the trigger coupling is set to TV-L or TV-F, selecting alternate
triggering automatically results in AC coupling conditions.
Under line/mains triggering condition alternate triggering can
not be chosen. Thus only the following trigger coupling modes
are available in alternate trigger mode: AC, DC, HF and LF. The
trigger point symbol is internally inactivated.
). In the case of
The sync-separator-circuit also operates with external triggering.
It is important that the voltage range (0.3Vpp to 3Vpp) for
external triggering should be noted. Again the correct slope
setting is critical, because the external trigger signal may not
have the same polarity or pulse edge as the test signal
displayed on the CRT. This can be checked, if the external
trigger voltage itself is displayed first (with internal triggering).
In most cases, the composite video signal has a high DC
content. With constant video information (e.g. test pattern or
color bar generator), the DC content can be suppressed easily
by AC input coupling of the oscilloscope amplifier. With a
changing picture content (e.g. normal program), DC input
coupling is recommended, because the display varies its
vertical position on screen with AC input coupling at each
change of the picture content.
The DC content can be compensated using the Y-POS. control
so that the signal display lies in the graticule area. Then the
composite video signal should not exceed a vertical height of
6div.
Line triggering (~)
The instrument specific information regarding this mode is
part of the section “Controls and readout” paragraph
MODE (26).
28
TRIG.
With alternate triggering it is possible to trigger two signals
which are different in frequency (asynchronous). In this case
the oscilloscope must be operated in DUAL alternate mode
and internal triggering each input signal must be of sufficient
height to enable trigger. To avoid trigger problems due to
different DC voltage components, AC input coupling for both
channels is recommended. The internal trigger source is
switched in alternate trigger mode in the same way as the
channel switching system in DUAL alternate mode, i.e. after
each time base sweep. Phase difference measurement is not
possible in this trigger mode as the trigger level and slope
setting are equal for both signals. Even with 180° phase
difference between both signals, they appear with the same
slope direction.
If signals are applied with a high frequency ratio (difference),
the trace intensity then becomes reduced if the time base is
set to smaller time coefficients (faster sweep). This happens
as the number of sweeps does not increase because it
depends on the lower frequency signal, but with a faster
sweep the phosphor becomes less activated.
External triggering
The external trigger input is activated with the aid of the TRIG.
(23) pushbutton (
see “Controls and readout”
Subject to change without notice
), if the trigger
Triggering and time base
coupling is not set to line/mains trigger coupling. Then the
internal trigger source is deactivated. As the external trigger
signal applied at the TRIG. EXT socket normally has no relation
to the signal height of the displayed signal, the trigger point
symbol is switched off. The external trigger voltage must have
a minimum amplitude of 0.3Vpp and should not increase above
3Vpp. The input impedance of the TRIG. EXT. socket is
approx. 1MΩ II 15pF.
The maximum input voltage of the input circuit is 100V
(DC+peak AC).
The external trigger voltage may have a completely different
form from the test signal voltage, but must be synchronous
with the test signal. Triggering is even possible in certain limits
with whole number multiples or fractions of the test frequency.
It must be noted that a different phase angle between the
measuring and the triggering signal may cause a display not
coinciding with the slope selection setting.
The trigger coupling selection can also be used in external
triggering mode.
Trigger indicator “TR”
The following description applies to the “TR” LED.
item (16) under “Controls and readout”.
Please note
signal, required for evaluation, is easily obtainable by expanding
the hold off time until one signal is displayed.
A double display is possible with certain pulse signals, where
the pulses alternately show a small difference of the peak
amplitudes. Only a very exact trigger level adjustment makes
a single display possible. The use of the holdoff control
simplifies the right adjustment.
After specific use the holdoff control should be reset into its
calibration detent (fully ccw), otherwise the brightness of the
display is reduced drastically. The function is shown in the
following figures.
An LED on condition indicates that the trigger signal has a
sufficient amplitude and the trigger level control setting is
correct. This is valid with automatic and with normal triggering.
By observing the trigger LED, sensitive trigger level adjustment
is possible when normal triggering is used, particularly at very
low signal frequencies. The indication pulses are of only
100ms duration.
Thus for fast signals the LED appears to glow continuously, for
low repetition rate signals, the LED flashes at the repetition rate
or at a display of several signal periods not only at the start of
the sweep at the left screen edge, but also at each signal period.
In automatic triggering mode the sweep generator starts
repeatedly without test signal or external trigger voltage. If the
trigger signal frequency decreases the trigger automatic
repetition frequency the sweep generator starts without
awaiting the trigger pulse. This causes an untriggered display
and a flashing trigger LED.
HOLD OFF-time adjustment
(Only in analog mode)
For instrument specific information please note
HO (27)
If it is found that a trigger point cannot be found on extremely
complex signals, even after careful adjustment of the trigger
level control, a stable display may often be obtained using the
holdoff control. This facility varies the holdoff time between
two sweep periods approx. up to the ratio 10:1. Pulses or other
signal waveforms appearing during this off period cannot
trigger the time base.
Particularly with burst signals or aperiodic pulse trains of the
same amplitude, the start of the sweep can be delayed until
the optimum or required time.
A very noisy signal or a signal with a higher interfering
frequency is at times displayed double. It is possible that
trigger level adjustment only controls the mutual phase shift,
but not the double display. The stable single display of the
in section “Controls and readout”.
DEL.POS. -
Fig. 1 shows a case where the holdoff control is in the minimum
position and various different waveforms are overlapped on the
screen, making the signal observation unsuccessful.
Fig. 2 shows a case where only the desired parts of the signal are stably
displayed.
Delay / After Delay Triggering
(Only in analog mode)
The instrument specific information regarding this mode is
part of the section “Controls and readout” paragraph
- HO (27), SEA./DEL. - ON/OFF (29)
.
(30)
As mentioned before, triggering starts the time base sweep
and unblanks the beam. After the maximum X deflection to the
right, the beam is blanked and flies back to the (left) start
position. After the hold off period the sweep is started
automatically by the automatic trigger or the next trigger
signal. In normal triggering mode the automatic trigger is
switched off and will only start on receipt of a trigger signal.
As the trigger point is always at the trace start position, trace
expansion in X direction with the aid of the time base is limited
to the display on the left of the trace. Parts of the signal to be
expanded which are displayed near the trace end (right side of
the screen) are lost when the time base speed is increased
(time coefficient reduced).
The delay function delays the trace start by a variable time from
the trigger point. This allows the sweep to begin on any portion
of a signal. The time base speed can then be increased to
expand the display in X direction. With higher expansion rates,
the intensity reduces and within certain limits this can be
compensated by the INTENS knob setting.
and
DEL.TRIG. - VAR.
DEL.POS.
Subject to change without notice
29
Triggering and time base
If the display shows jitter, it is possible to select for (second)
triggering after the elapsed delay time (DELay TRIGger =
DTR). As mentioned before, it is possible to display video
signals using the frame sync pulses for triggering (TV-F). After
the delay time set by the operator, the next line sync pulse or
the line content may be used for triggering. So data lines and
test lines can be displayed separately.
Operation of the delay function is relatively simple. Without
delay function set the time coefficient setting (TIME/DIV) until
1 to 3 signal periods are displayed. Display of less the one
period should be avoided as it limits the selection of the signal
section to be expanded, and may cause trigger problems.
The X-MAG. x10 function should be switched off and the time
variable control should be CAL position. The signal must be
triggered and stable. The following explanation assumes that
the trace starts on the left vertical graticule line.
Pressing and holding the SEA./DEL. - ON/OFF button switches
over to SEARCH mode (SEA displayed by the readout).
In all delay modes, the DEL. POS. knob assumes the function
of DEL. POS. (delay position), and the hold off time defaults to
minimum. Now the function of this knob (DEL. POS.) is to
adjust the delay time, indicated as a blanked part of the screen.
The length of the blanked sector depends on the DEL. POS.
setting and can be set between approx. one and six division
after the normal trace start position. As the trace right end is
not effected, the visible trace length is reduced. In delay (DEL)
mode, the trace will start from the normal left end where the
blanking starts. If the maximum delay is not sufficient, the time
coefficient must be increased (TIME/DIV knob) and the DEL.POS. knob set to the later starting point. To return to normal
(undelayed) time base operation, press and hold the SEA./DEL pushbutton again.
Photo 2 shows that the delay time can be measured. It is
identical with the displacement of the start of the trace. One
can calculate this by multiplying the blanked out section
(horizontal) by the time deflection coefficient setting.
Example:
The SEARCH setting selected in figure 2 is 5ms/cm. The
display in DELAY mode, also with 5ms/cm is delayed but
unexpanded (1:1). A further increase in the deflection
coefficient, e.g. 10ms/cm would be meaningless and
therefore automatically blocked.
Please note that the previous time coefficient chosen in DEL
and DTR mode is stored and automatically set after activating
one of those modes. If the stored time coefficient in DEL or
DTR mode was higher than the actual value in SEA (search)
mode, the time coefficient in DEL or DTR mode is automatically
set to the value used during SEA (search) operation.
Reducing the time coefficient (increasing the time base speed)
now expands the signal. If the signal start position is not set
to the optimum, it can still be shifted in the X direction by
turning the DEL. POS. knob. Photo 4 shows a 50 fold X
magnification caused by setting the time coefficient to 0.1ms/
div (5ms/div : 0.1ms/div = 50). The reading accuracy also
increases with higher X magnification. As already mentioned,
the time variable control must be in CAL position when
measurements are taken.
The delayed and expanded signal display can be triggered
again if a signal slope suitable for triggering appears after the
delay time. For this, one must switch to DEL.TRIG. (2nd
triggering after the expiry of the delay time - after Delay
Triggering). The settings selected before switching, automatic
Peak value triggering / Normal triggering, trigger coupling, the
trigger LEVEL setting and slope setting, remain valid and
trigger the start of the delay time.
The ”After Delay” Triggering automatically switches to nor-
mal triggering (indicated by the NM-LED) and DC trigger
coupling. These default conditions cannot be changed. But the
trigger level (LEVEL) and the trigger slope direction can be
altered in order to enable the triggering at the desired signal
section. The trace does not start and the screen remains blank
if the signal amplitude is not sufficient for triggering or if the
setting of the trigger LEVEL is unsuitable.
The total trace starting with the previously selected section will
be visible when switched from SEARCH to DELAY, providing
the (stored) current time deflection coefficient is not too small.
If the trace is invisible or hardly visible because of too high
expansion (too small deflection coefficient), the time deflection
coefficient must be increased with TIME / DIV. knob. A larger
deflection coefficient than in the SEARCH mode cannot be set.
30
The expanded display can also be displaced in the X direction
with DELAY position control (DEL.POS.) under suitable
settings. However, the displacement is not continuous as in
the untriggered DELAY operation but jumps from one trigger
slope to another - with most signals this is not evident. This
means, in the case of TV Triggering, that it is possible to trigger
not only with line synchronizing pulses but also on suitable
slopes occurring within the line.
Subject to change without notice
Auto Set • Save/RecallComponent Tester (analog mode)
Of course, the magnification is not restricted to a factor 50 as
mentioned in the example. The limit is given by the increasing
loss of trace intensity as the magnification is increased.
The manipulation of time delay requires a certain experience,
especially with complicated signal combinations which are
difficult to display. The recording of signal sections of simple
signals is, in contrast, from the beginning fairly easy. The time
delayed display is also possible in the dual channel, addition
and difference modes.
Attention:
The display may be in chopped DUAL mode with
DELAY operation in dual channel mode (DUAL). This is
the case when in SEARCH, a time base between 500ms/
div and 500µs/div is set, which automatically switches
in chopped DUAL mode.
The chopped DUAL mode is effective even if
subsequently the time base is switched to between
0,2ms/cm and 50ns in triggered or untriggered DELAY
operation. With highly expanded displays the channel
switching will be visible during a sweep (alternately
switching between channels I and II). One can then
switch to alternate DUAL mode. A renewed change in
the time base setting will again result in chopped
display but this can be overcome again.
Auto Set
The instrument specific information regarding this function is
part of the section “Controls and readout” paragraph AUTOSET (2). As also mentioned in that section all controls are
electronically selected with the exception of the POWER and
calibrator frequency (CAL 1kHz/1MHz) pushbuttons as well as
the FOCUS and TR (trace rotation) control.
Thus automatic signal related instrument set up in Yt (time
base) mode is possible. In most cases no additional manual
instrument setting is required.
• X x10 magnifier switched off
• Automatic X and Y position settings
• Trace and readout visible.
If DC trigger coupling had been selected, AC trigger coupling
will not be chosen and the automatic trigger is operative
without the peak value detection.
The X position is set to the CRT center as well as the Y position
in CH I or CH II mode. In DUAL mode the channel I trace is set
to the upper half and the channel II trace to the lower half of
the CRT.
The 1mV/div and 2mV/div deflection coefficient will not be
selected by AUTO SET as the bandwidth is reduced on these
settings.
Attention!
If a signal is applied with a pulse duty factor of approx.
400:1 or larger, an automatic signal display can not be
performed. The pulse duty factor causes too low a Y
deflection coefficient (sensitivity too high) and too high a
time deflection coefficient (time base speed to slow) and
results in a display in which only the baseline is visible.
In such cases it is recommended to select normal triggering
and to set the trigger point approx. 0.5div above or below the
trace. If under one of these conditions the trigger indicator LED
is lit, this indicates the presence of a signal. Then both the time
coefficient and Y deflection coefficient should be reduced.
Please note that a reduction in intensity may occur, which
could result in a blank screen when the physical limits are
reached.
STORAGE MODE ONLY
In contrast to analog mode, intensity reduction does not occur.
But it must be noted that pulses of less than 20ns width are
displayed with too small amplitude. This is because of too few
samples per pulse, as in 100MS/s condition (10ns sampling
interval) only 4 samples are taken.
Briefly pressing the AUTO SET pushbutton causes the
instrument to switch over to the last Yt mode settings
regarding CH I, CH II and DUAL. If the instrument was
operated in Yt mode, the actual setting will not be affected
with the exception of ADD mode which will be switched off.
At the same time the attenuator(s) (VOLTS/DIV) are
automatically set for a signal display height of approx. 6 div in
mono channel mode or if in DUAL mode for approx. 4 div
height for each channel. This and the following explanation
regarding the automatic time coefficient setting assumes that
the pulse duty factor of the input signal is approx. 1:1.
The time deflection coefficient is also set automatically for a
display of approx. 2 signal periods. The time base setting
occurs randomly if complex signals consisting several
frequencies e.g. video signals are present.
AUTO SET sets the instrument automatically to the following
operating conditions:
• The input coupling is not effected while AC or DC is used,
but if GD wasselected the last used input coupling (AC
or DC) will be chosen.
• Internal triggering (channel I or channel II)
• Automatic peak (value) triggering
• Trigger level in electrical midrange position
• Y deflection coefficient(s) calibrated
• Time base deflection coefficient calibrated
• AC trigger coupling (except DC trigger coupling)
• SEARCH and DELAY time base mode switched off
If SINGLE (SGL) or ROLL (ROL) mode are active, AUTO SET
sets the instrument automatically to REFRESH (RFR) mode.
Save/Recall
The instrument specific information regarding the controls and
their use are part of item
(12)
in section “Controls and
readout”.
The instrument contains a non volatile memory which can be
used by the operator to save 9 different instrument settings
and to recall them. This relates to all settings with the
exception of FOCUS, TR (trace rotation) and the calibrator
frequency pushbutton.
Component Tester (analog mode)
General
The instrument specific information regarding the control and
terminals are part of item
readout”.
The instrument has a built-in electronic Component Tester,
which is used for instant display of a test pattern to indicate
whether or not components are faulty. It can be used for quick
checks of semiconductors (e.g. diodes and transistors),
resistors, capacitors, and inductors. Certain tests can also be
made to integrated circuits. All these components can be
tested individually, or in circuit provided that it is unpowered.
(45)
in section “Controls and
Subject to change without notice
31
Component Tester (analog mode)
The test principle is fascinatingly simple. A built-in generator
delivers a sine voltage, which is applied across the component
under test and a built-in fixed resistor. The sine voltage across
the test object is used for the horizontal deflection, and the
voltage drop across the resistor (i.e. current through test
object) is used for vertical deflection of the oscilloscope. The
test pattern shows a current-voltage characteristic of the test
object.
The measurement range of the component tester is limited
and depends on the maximum test voltage and current (please
note data sheet). The impedance of the component under test
is limited to a range from approx. 20 Ohm to 4.7k Ohm. Below
and above these values, the test pattern shows only shortcircuit or open-circuit. For the interpretation of the displayed
test pattern, these limits should always be born in mind.
However, most electronic components can normally be tested
without any restriction.
Using the Component Tester
After the component tester is switched on, the vertical
preamplifier and the time base generator are inoperative. A
shortened horizontal trace will be observed. It is not necessary
to disconnect scope input cables unless in-circuit measurements are to be carried out.
For the component connection, two simple test leads with
4mm Ø banana plugs, and with test prod, alligator clip or
sprung hook, are required.
The test leads are connected as
described in section “Controls and readout”.
Test Procedure
Caution! Do not test any component in live circuitry - remove
all grounds, power and signals connected to the component
under test. Set up Component Tester as stated. Connect test
leads across component to be tested. Observe oscilloscope
display.
Only discharged capacitors should be tested!
A horizontal ellipse indicates a high impedance or a relatively
small capacitance or a relatively high inductance.
A vertical ellipse indicates a small impedance or a relatively
large capacitance or a relatively small inductance.
A sloping ellipse means that the component has a considerable
ohmic resistance in addition to its reactance.
The values of capacitance of normal or electrolytic capacitors
from 0.1µF to 1000µF can be displayed and approximate
values obtained. More precise measurement can be obtained
in a smaller range by comparing the capacitor under test with
a capacitor of known value. Inductive components (coils,
transformers) can also be tested. The determination of the
value of inductance needs some experience, because inductors
have usually a higher ohmic series resistance. However, the
impedance value (at 50Hz) of an inductor in the range from 20Ω
to 4.7kΩ can easily be obtained or compared.
Testing Semiconductors
Most semiconductor devices, such as diodes, Z-diodes,
transistors, FETs can be tested. The test pattern displays vary
according to the component type as shown in the figures
below.
The main characteristic displayed during semiconductor testing
is the voltage dependent knee caused by the junction changing
from the conducting state to the non conducting state. It
should be noted that both the forward and the reverse
characteristic are displayed simultaneously. This is a twoterminal test, therefore testing of transistor amplification is
not possible, but testing of a single junction is easily and quickly
possible. Since the test voltage applied is only very low, all
sections of most semi-conductors can be tested without
damage. However, checking the breakdown or reverse voltage
of high voltage semiconductors is not possible. More important
is testing components for open or short-circuit, which from
experience is most frequently needed.
Testing Diodes
Test Pattern Displays
This section contains some typical patterns displayed by the
various components under test.
- Open circuit is indicated by a straight horizontal line.
- Short circuit is shown by a straight vertical line.
Testing Resistors
If the test object has a linear ohmic resistance, both deflecting
voltages are in the same phase. The test pattern expected
from a resistor is therefore a sloping straight line. The angle of
slope is determined by the resistance of the resistor under
test. With high values of resistance, the slope will tend
towards the horizontal axis, and with low values, the slope will
move towards the vertical axis. Values of resistance from 20Ω
to 4.7kΩ can be approximately evaluated. The determination
of actual values will come with experience, or by direct
comparison with a component of a known value.
Testing Capacitors and Inductors
Capacitors and inductors cause a phase difference between
current and voltage, and therefore between the X and Y
deflection, giving an ellipse-shaped display. The position and
opening width of the ellipse will vary according to the impedance
value (at 50Hz) of the component under test.
Diodes normally show at least their knee in the forward
characteristic. This is not valid for some high voltage diode
types, because they contain a series connection of several
diodes. Possibly only a small portion of the knee is visible.
Zener diodes always show their forward knee and, depending
on the test voltage, their zener breakdown forms a second
knee in the opposite direction. If the breakdown voltage is
higher than the positive or negative voltage peak of the test
voltage, it can not be displayed.
The polarity of an unknown diode can be identified by
comparison with a known diode.
Testing Transistors
Three different tests can be made to transistors: base-emitter,
base-collector and emitter-collector. The resulting test patterns
are shown below.
32
Subject to change without notice
Component Tester (analog mode)Storage Mode
The basic equivalent circuit of a transistor is a Z-diode between
base and emitter and a normal diode with reverse polarity
between base and collector in series connection. There are
three different test patterns:
For a transistor the figures b-e and b-c are important. The
figure e-c can vary; but a vertical line only shows short circuit
condition.
These transistor test patterns are valid in most cases, but
there are exceptions to the rule (e.g. Darlington, FETs). With
the COMPONENT TESTER, the distinction between a P-N-P
to an N-P-N transistor is discernible. In case of doubt,
comparison with a known type is helpful. It should be noted
that the same socket connection (COMP. TESTER or ground)
for the same terminal is then absolutely necessary. A connection
inversion effects a rotation of the test pattern by 180 degrees
round about the center point of the scope graticule.
In-Circuit Tests
Caution!
During in-circuit tests make sure the circuit is dead. No
power from mains/line or battery and no signal inputs
are permitted. Remove all ground connections including
Safety Earth (pull out power plug from outlet). Remove
all measuring cables including probes between
oscilloscope and circuit under test. Otherwise both
COMPONENT TESTER leads are not isolated against
the circuit under test.
In-circuit tests are possible in many cases. However, they are
not well defined. This is caused by a shunt connection of real
or complex impedances - especially if they are of relatively low
impedance at 50Hz - to the component under test, often
results differ greatly when compared with single components.
In case of doubt, one component terminal may be unsoldered.
This terminal should then not be connected to the ground
socket avoiding hum distortion of the test pattern.
Another way is a test pattern comparison to an identical circuit
which is known to be operational (likewise without power and
any external connections). Using the test prods, identical test
points in each circuit can be checked, and a defect can be
determined quickly and easily. Possibly the device itself under
test contains a reference circuit (e.g. a second stereo channel,
push-pull amplifier, symmetrical bridge circuit), which is not
defective.
Storage Mode
In contrast to analog mode, the storage mode offers the
following advantages:
One time events can be captured easily. Even very low
frequency signals can be displayed as a complete curve.
Narrow pulses with low repetition rates do not cause intensity
reduction. Documentation and processing of captured signals
is easily possible.
In comparison with analog mode, the disadvantages of
storage mode are:
The reduced X and Y resolution and a lower update rate.
Danger of alias signal display, caused by a sampling rate (time
base setting) which is relatively too low with respect to the
current signal.
The analog mode offers an unsurpassed faithful signal display.
With the combination of analog and digital oscilloscope, HAMEG
enables the user to select the most suitable mode for the
specific measurement.
The HM407 contains two 8 bit flash A/D converters with a
maximum sampling rate of 50MS/s each. Except in the
combination of DUAL mode and SINGLE event recording with
a maximum of 40MS/s (time coefficient 5µs/div) for each
channel, in all other modes 100MS/s is available if the lowest
time coefficient is chosen.
Subject to change without notice
33
Storage Mode
Apart from the factors above, there is no principle difference
between capturing repetitive signals and one-time events. The
signal is always displayed with a linear connection between
the dots (Dot Join function).
Signals captured and stored in storage mode can be called via
the built-in RS232 interface for documentation purposes. For
further information please note section “RS232 Interface Remote Control”.
Signal recording modes
Signals can be recorded and displayed in six different modes:
REFRESH mode (RFR LED lit, readout indicates RFR)
ENVELOPE mode (ENV LED lit, readout indicates ENV)
AVERAGE mode (AVM LED lit, readout indicates AV...)
SINGLE mode (SGL LED lit, readout indicates SGL)
ROLL mode (ROL LED lit, readout indicates ROL)
XY mode (RFR LED lit, readout indicates XY and the sampling rate)
Except ROLL and XY mode, a signal recording in all other
modes requires a trigger signal.
In REFRESH, ENVELOPE and AVERAGE modes the
instruments behaves like an analog oscilloscope. The trigger
circuit starts a recording, overwriting the previous recording
from the left to the right side of the screen. After the recording
has been finished, the next trigger event starts the same
procedure. This can also be caused in automatic trigger mode
without an applied signal by the automatic circuitry. Then only
the trace (Y-POS. setting) is recorded.
In contrast to automatic trigger mode, in normal trigger mode
the automatic system is switched off and consequently only
a trigger signal can start a recording. Unlike analog mode
where the screen is dark until a trigger signal starts the time
base, in store mode the last recorded signal remains visible as
long as no new recording is triggered by an input signal.
AVERAGE and ENVELOPE are REFRESH sub-modes and
described in section “Controls and readout” under item (9).
SINGLE mode (SGL) enables the capture of one-time events,
started by a suitable trigger signal. It is recommended, to
select input ground (GD) condition and set the trace on a
graticule line which is then used as the 0 Volt position
(reference) line. In SINGLE mode the instrument is
automatically set to normal trigger mode. After SINGLE (SGL)
mode has been selected, the trigger point symbol should be
set above or below the 0 Volt position line, according to the
expected voltage of the event to be captured. Whether the
slope selection is set for a rising or falling slope depends on the
measurement task. After this procedure AC or DC input
coupling must be selected and the signal capture is started
after activating the RESET function. The following example is
for additional explanation.
If a voltage drop on a +5 Volt line is expected, the 0 Volt line
may be set to the horizontal center graticule line by using the
Y-POS. control. Then the trigger point symbol may be set 2
division above the 0 Volt position using the LEVEL control.
If the deflection coefficient is set to 200mV and DC input
coupling selected, using a x10 probe, the trigger point is at +4
Volts. A voltage drop from +5 Volts below +4 Volts then is used
for triggering, if negative slope setting is selected. If the slope
selection is positive the end of the event (voltage drop ended)
is used for triggering.
For explanations regarding ROLL mode, please note this item
(9)
in section “Readout and Controls”.
Vertical resolution
The dot density in each operation mode is 8 bits = 28 = 256 dots
displayed over a height of roughly 10 divisions. The instrument
is adjusted for 25 dots per division. This eases processing and
cursor measurement.
Insignificant differences between the (analog) screen display
and the (digital) data are unavoidable.
This concerns signal height as well as the position. The trace
position is defined in respect to the following horizontal
graticule lines:
Center line = 10000000 (binary) = 80 (hex) = 128 (dec).
Top line= 11100100 (binary) = E4 (hex) = 228 (dec).
Bottom line = 00011100 (binary) = 1C (hex) = 28 (dec).
In contrast to analog mode with its theoretically unlimited
resolution, the vertical resolution has 25 possible trace positions
per division.
If the signal is superimposed by noise or a critical Y-POS.
setting is used, the least significant bit (LSB) may change
continuously. This additionally reduces the vertical resolution
in storage mode, but is unavoidable. In contrast to the expensive flash A/D converters used in this instrument, other
converters such as CCD cause more noise.
Horizontal resolution
The maximum number of signals to be displayed simultaneously
is three. Each signal consists of 2048 (211) byte (samples).
Referred to the horizontal raster, the resolution is 200 samples
per division.
Pure (only) digital oscilloscope with VGA monitor type CRTs
offer only 50 samples per division. If LCD displays are used the
current resolution is 25 samples per division. For a given time
base setting the HAMEG instrument samples at a 4 (compared
to VGA) or 8 ( referred to LCD) times higher sampling rate. The
higher number of samples/div results in a shorter sampling
interval. For the following example it must be kept in mind, that
the time base setting is related to the signal period duration and
consequently should enable the display of one complete signal
period. If e.g. a 50Hz signal has to be displayed the time base
should be set to 2ms/div. The maximum signal frequency of a
superimposed sinewave signal, which must be sampled with at
least 10 samples per period, depends on the horizontal resolution:
samples/div sampling intervalsampling rate max frequency
During X magnifier mode the record length is still 2048 byte.
The CRT displays a tenth of the recording (20 byte/division)
plus 180 byte/div calculated in linear interpolation operation by
a RISC processor. The smallest available time coefficient then
is 50ns/div instead of 500ns/div without the X magnifier.
Maximum signal frequency in storage mode
The highest capturable signal frequency cannot be exactly
defined, since it depends to a large extent on the waveform.
With the start of each recording, the signal voltage at the
input(s) of the analog-to-digital converters is briefly measured
(sampled), converted to an 8-bit value and written to an
address in RAM. The next sampled value is converted in the
same way, but stored at the next RAM address.
34
Subject to change without notice
Test Instructions
The maximum sampling rate is 100MS/s. This yields a sampling
interval of 10ns. Assuming that 10 measurements (samples)
per signal period are sufficient for a sinewave signal, the
maximum frequency to be captured is 10MHz (100ns per
period).
Alias signal display
If, due to the time base setting, the sampling rate is too low,
the display of an alias signal may occur.
For example:
A sinewave signal may be sampled only once per period. If
the sinewave signal frequency is accidentally in phase with
the sampling frequency and each sample is taken at the
positive peak value, a straight line is displayed in this
position.
Alias signal display may also occur in the form of an apparent
untriggered waveform display of different frequency from the
true signal. Another aliasing condition is the display of signals
seeming to be amplitude modulated.
The easiest way to recognize alias signals is to switch to analog
mode, where the true waveform is displayed. Transfer from
analog to store mode without changing time base range must
produce the same frequency display.
Operating modes of the vertical amplifiers
In principle, the instrument can operate in digital storage mode
with the same operating modes as in analog mode. Thus, the
following can be displayed:
- Channel I by itself
- Channel II by itself
- Channel I and II simultaneously
- The sum or difference of both channels
- XY mode
Storage mode differs from analog operation of the scope in the
following respects:
- In DUAL mode (for simultaneous display of both channels),
both input signals are also simultaneously captured, since
each channel has its own A/D converter. In contrast to
analog mode therefore, it is not necessary to switch between
chopped and alternating mode.
- Because of the high display rate (approx. 80Hz) of the screen,
no flicker can occur.
- The storage time base is also active in XY storage mode.
Test Instructions
General
positions. This is the case if AUTO SET had been pressed. It
is recommended to switch on the instrument for about 20
minutes prior to the commencement of any check.
Cathode-Ray Tube:
Brightness and Focus,
Linearity, Raster Distortion
Normally, the CRT of the instrument has very good brightness.
Any reduction of this brightness can only be judged visually.
Decreased brightness may be the result of reduced high
voltage. In this case the sensitivity of the vertical amplifiers is
significantly increased .
The control range for maximum and minimum brightness
(intensity) must be such that the beam just disappears before
reaching the left hand stop of the INTENS. control (particularly
when in XY mode), while with the control at the right hand stop
the focus and the line width are just acceptable.
With maximum intensity the time base fly-back must on no
account be visible. Visible trace fault without input signal:
bright dot on the left side or decreasing brightness from left to
right or shortening of the baseline. (Cause: Incorrect Unblanking
Pulse.) It should be noted that with wide variations in brightness,
refocusing is always necessary. Moreover, with maximum
brightness, no “pumping” of the display must occur. If pumping
does occur, it is normally due to a fault in the regulation
circuitry for the high voltage supply. The presetting pots for the
high voltage circuit, minimum and maximum intensity, are only
accessible inside the instrument.
A certain out-of-focus condition in the edge zone of the screen
must be accepted. It is limited by standards of the CRT
manufacturer. The same is valid for tolerances of the
orthogonality, the undeflected spot position, the non-linearity
and the raster distortion in the marginal zone of the screen in
accordance with international standards (see CRT data book).
These limit values are strictly supervised by HAMEG. The
selection of a cathode-ray tube without any tolerances is
practically impossible.
Astigmatism Check
Check whether the horizontal and vertical sharpness of the
display are equal. This is best seen by displaying a square-wave
signal with the repetition rate of approximately 1MHz. Focus
the horizontal tops of the square-wave signal at normal intensity,
then check the sharpness of the vertical edges. If it is possible
to improve this vertical sharpness by turning the FOCUS
control, then an adjustment of the astigmatism control is
necessary. A potentiometer of 47k Ohm is provided inside the
instrument for the correction of astigmatism. A certain loss of
marginal sharpness of the CRT is unavoidable; this is due to the
manufacturing process of the CRT.
These Test Instructions are intended as an aid for checking the
most important characteristics of the instrument at regular
intervals without the need for expensive test equipment.
Resulting corrections and readjustments inside the instrument,
indicated by the following tests, are described in the “Service
Manual”. They should only be undertaken by qualified
personnel.
A chargeable “Service Manual” is available from HAMEG. The
“Service Manual” describes the complete adjustment of the
instrument in English language and contains the circuit diagrams
as well as the component locations.
As with the First Time Operation instructions, care should be
taken that all variable functions are set to their calibrated
Subject to change without notice
Symmetry and Drift of the Vertical Amplifier
Both of these characteristics are substantially determined by
the input stages of the amplifiers.
The symmetry of both channels and the vertical final amplifier
can be checked by inverting Channel I and II (depress the
corresponding INV pushbutton). The vertical position of the
trace should not change by more than 0.5div. However, a
change of 1div is just permissible. Larger deviations indicate
that changes have occurred in
the amplifier.
A further check of the vertical amplifier symmetry is possible
by checking the control range of the Y-POS. controls. A sine-
35
Test Instructions
wave signal of 10-100kHz is applied to the amplifier input.
When the Y-POS. control is then turned fully in both directions
from stop to stop with a display height of approximately 8div,
the upper and lower positions of the trace that are visible
should be approximately of the same height. Differences of up
to 1div are permissible (input coupling should be set to AC).
Checking the drift is relatively simple. 20minutes after switching
on the instrument, set the baseline exactly on the horizontal
center line of the graticule. The beam position must not
change by more than 0.5div during the following hour.
Calibration of the Vertical Amplifier
A square-wave voltage of 0.2Vpp ±1% is present at the output
socket of the calibrator (CAL.) If a direct connection is made
between the 0.2V output and the input of the vertical amplifier
(e.g. using a x1 probe), the displayed signal in the 50mV/div
position (variable control to CAL.) should be 4div high (DC input
coupling).
Maximum deviations of 0.12div (3%) are permissible. If a x10
probe (1%) is connected between the 0.2V output and Y input,
the deflection coefficient should be set to 5mV/div. Then the
maximum deviation is 0.16div.
With higher tolerances it should first be investigated whether
the cause lies, within the amplifier or in the amplitude of the
square-wave signal. On occasions it is possible that the probe
is faulty or incorrectly compensated. If necessary the measuring
amplifier can be calibrated with an accurately known DC
voltage (DC input coupling). The trace position should then
vary in accordance with the deflection coefficient set.
With variable control in the attenuator sector fully counterclockwise, the input sensitivity is decreased at least by the
factor 2.5 in each position. In the 50mV/div position, the
displayed calibrator signal height should vary from 4div to at
least 1.6div.
Transmission Performance
of the Vertical Amplifier
The transient response and the delay distortion correction can
only be checked with the aid of a square-wave generator with
a fast risetime (max. 5ns). The signal coaxial cable (e.g. HZ34)
must be terminated at the vertical input of the oscilloscope
with a resistor equal to the characteristic impedance of the
cable (e.g. with HZ22). Checks should be made at 100Hz,
1kHz, 10kHz, 100kHz and 1MHz, the deflection coefficient
should be set at 5mV/div with DC input coupling. In so doing,
the square pulses must have a flat top without ramp-off,
spikes and glitches; no overshoot is permitted, especially at
1MHz and a display height of 4-5div. At the same time, the
leading top corner of the pulse must not be rounded. In
general, no great changes occur after the instrument has left
the factory, and it is left to the operators discretion whether
this test is undertaken or not. A suited generator for this test
is HZ60 from HAMEG.
Of course, the quality of the transmission performance is not
only dependent on the vertical amplifier. The input attenuators,
located in the front of the amplifier, are frequency-compensated
in each position. Even small capacitive changes can reduce the
transmission performance. Faults of this kind are as a rule most
easily detected with a square-wave signal with a low repetition
rate (e.g. 1kHz). If a suitable generator with max. output of
40Vpp is available, it is advisable to check at regular intervals the
deflection coefficients on all positions of the input attenuators
and readjust them as necessary. A compensated 2:1 series
attenuator is also necessary, and this must be matched to the
input impedance of the oscilloscope. This attenuator can be
made up locally. It is important that this attenuator is shielded.
For local manufacture, the electrical components required are
a 1MΩ ±1% resistor and, in parallel with it, a trimmer 3-15pF in
parallel with approx. 10pF. One side of this parallel circuit is
connected directly to the input connector of CH I or CH II and
the other side is connected to the generator, if possible via a
low-capacitance coaxial cable. The series attenuator must be
matched to the input impedance of the oscilloscope in the 5mV/
div position (variable control to CAL., DC input coupling; square
tops exactly horizontal; no ramp-off is permitted). This is achieved
by adjusting the trimmer located in the 2:1 attenuator. The
shape of the square-wave should then be the same in each input
attenuator position.
Operating Modes:
CH.I/II, DUAL, ADD, CHOP.,
INVERT and X-Y Operation
In DUAL mode two traces must appear immediately. On
actuation of the Y-POS. controls, the trace positions should
have minimal effect on each other. Nevertheless, this cannot
be entirely avoided, even in fully serviceable instruments.
When one trace is shifted vertically across the entire screen,
the position of the other trace must not vary by more than
0.5mm.
A criterion in chopped operation is trace widening and
shadowing around and within the two traces in the upper or
lower region of the screen. Set time coefficient to 0.5ms/div,
set input coupling of both channels to GD and advance the
INTENS. control fully clockwise. Adjust FOCUS for a sharp
display. With the Y-POS. controls shift one of the traces to a
+2div, the other to a -2div vertical position from the horizontal
center line of the graticule.
Do not try to synchronize (with the time variable
control) the chop frequency (0.5MHz)! Check for negligible trace widening and periodic shadowing when
switching between 0.5ms/div and 0.2ms/div.
It is important to note that in the I+II add mode or the I-II
difference mode the vertical position of the trace can be adjusted
by using both the Channel I and Channel II Y-POS. controls.
In X-Y Operation, the sensitivity in both deflection directions
will be the same. When the signal from the built-in squarewave generator is applied to the input of Channel II, then, as
with Channel I in the vertical direction, there must be a
horizontal deflection of 4div when the deflection coefficient is
set to 50mV/div position. The check of the mono channel
display is unnecessary; it is contained indirectly in the tests
above stated.
Triggering Checks
The internal trigger threshold is important as it determines the
display height from which a signal will be stably displayed. It
should be approx. 0.3-0.5div for the instrument. An increased
trigger sensitivity creates the risk of response to the noise
level in the trigger circuit. This can produce double-triggering
with two out-of-phase traces.
Alteration of the trigger threshold is not required. Checks can
be made with any sine-wave voltage between 50Hz and
1MHz. The instrument should be in automatic peak (value)
triggering (NM LED dark) and the LEVEL knob in electrical
midrange position. It should be ascertained whether the same
trigger sensitivity is also present with Normal Triggering (NMLED lights). In this trigger mode, LEVEL adjustment is absolutely
necessary.
36
Subject to change without notice
Service Instructions
The checks should show the same trigger threshold with the
same frequency. On changing the trigger slope, the start of
the sweep changes from the positive-going to the negativegoing edge of the trigger signal.
As described in the Operating Instructions, the trigger frequency
range is dependent on the trigger coupling selected. For lower
frequencies the LF coupling mode can be selected. In this
mode, triggering up to at least 1.5kHz (sine-wave) is possible.
Internally the instrument should trigger perfectly at a display
height of approx. 0.5div, when the appropriate trigger coupling
mode is set.
For external triggering, the external trigger input connector
requires a signal voltage of at least 0.3Vpp, which is in
synchronism with the Y input signal. The voltage value is
dependent on the frequency and the trigger coupling mode
(AC-DC-HF-LF).
Checking of the TV triggering is possible with a video signal of
any given polarity.
Use the TV-L or TV-F setting for video sync pulse separation.
The correct slope of the sync pulse (front edge) must be
selected and a suitable time coefficient setting must be
chosen. The slope is then valid for both sync frequencies.
Perfect TV triggering is achieved, when in both display modes
the amplitude of the complete TV signal (from white level to
the top of the line sync pulse) is limited between 0.8 and 6div
and sync pulses of more then 0.5 div height. The display should
not shift horizontally during a change of the trigger coupling
from AC to DC when displaying a sine-wave signal without DC
offset.
If both vertical inputs are AC coupled to the same signal and
both traces are brought to coincide exactly on the screen,
when working in the alternate dual channel mode, then no
change in display should be noticeable, when switching from
TRIG CH I to TRIG CH II or when the trigger coupling is
changed from AC to DC.
Checking of the line/mains frequency triggering (50 - 60Hz) is
possible, when the input signal is time-related (multiple or
submultiple) to the power line frequency ( ~ LED lights). There
is no trigger threshold visible in this trigger mode. Even very
small input signals are triggered stably (e.g. ripple voltage). For
this check, use an input of approx. 1V. The displayed signal
height can then be varied by turning the respective input
attenuator switch and its variable control.
Time base
Before checking the time base it should be ascertained that
the trace length is approx. 10div in all time ranges. The X-MAG. x10 LED should not light. This condition should be
maintained until the variation ranges of these controls are
checked. Check that the sweep runs from the left to the right
side of the screen (TIME/DIV. setting to 100ms/div). This
check is only necessary after changing the cathode-ray tube.
If a precise marker signal is not available for checking the Time
base time coefficients, then an accurate sine-wave generator
may be used. Its frequency tolerance should not be greater
than ±0.1%. The time base accuracy of the instrument is given
as ±3%, but it is considerably better than this. For the
simultaneous checking of time base linearity and accuracy at
least 10 oscillations, i.e. 1 cycle every div, should always be
displayed. For precise determination, set the peak of the first
marker or cycle peak exactly behind the first vertical graticule
line using the X-POS. control. Deviation tendencies can be
noted after some of the marker or cycle peaks. If a precise
Time Mark Generator is used for checking, Normal Triggering
and LEVEL control adjustment is recommended.
The signal frequency required for each time coefficient setting
can be determined easily by the aid of the readout. In time
measurement condition the distance of the vertical cursor
lines must be set to exactly 1 div. Then both readout information
regarding the time deflection coefficient and the time difference
show exactly the same value. After switching over from time
to frequency measurement, the readout displays the required
signal frequency in each time base setting.
Hold Off time (analog mode only)
The variation of the hold off time setting can not be measured
without opening the instrument. However, a visual check can
be made if the instrument is operated in undelayed time base
mode. Without an input signal, set time base to 50ns/div, use
automatic peak (value) triggering. With minimum hold off time
the trace should be bright. It should noticeably darken if the
hold off time is increased to the maximum.
Component Tester
After selecting component tester mode, a horizontal straight
line should appear immediately, when the test leads are not
connected (open circuit). The length of this trace should be
approx. 8div. With the test leads shorted together, a vertical
straight line with approx. 6div height should be displayed. The
above stated measurements have some tolerances.
Trace Alignment
The CRT has an admissible angular deviation ±5° between the
X deflection plane D1-D2 and the horizontal center line of the
internal graticule. This deviation, due to tube production
tolerances (and only important after changing the CRT), and
also the influence of the earth’s magnetic field, which is
dependent on the instruments North orientation, are corrected
by means of the TR potentiometer. In general, the trace
rotation range is asymmetric. It should be checked, whether
the baseline can be adjusted somewhat sloping to both sides
round about the horizontal center line of the graticule. With the
instrument in its closed case, an angle of rotation ±0.57°
(0.1div difference in elevation per 10div graticule length) is
sufficient for the compensation of the earth’s magnetic field.
Service Instructions
General
The following instructions are intended as an aid for the
electronic technician, who is carrying out readjustments on
the instrument, if the nominal values do not meet the
specifications. These instructions primarily refer to those
faults, which were found after using the Test Instructions.
However, this work should only be carried out by properly
qualified personnel.
call or write to HAMEG
of the manual. It is recommended to use only the original
packing material, should the instrument be shipped to for
service or repair (see also Warranty).
Instrument Case Removal
The rear cover can be taken off after unplugging the power
cords triple-contact connector and after two nuts have been
removed. If a cross recessed pan head screw is present on the
bottom of the instrument, it must be removed too. While the
instrument case is firmly held, the entire chassis with its front
For any further technical information
. Addresses are provided at the back
Subject to change without notice
37
Service Instructions
panel can withdrawn forward. When the chassis is inserted
into the case later on, it should be noticed that the case has to
fit under the flange of the front panel. The same applies for the
rear of the case, on which the rear cover is put.
Caution !
During opening or closing of the case, the instrument
must be disconnected from all power sources for
maintenance work or a change of parts or components.
If a measurement, trouble-shooting, or an adjustment
is unavoidable, this work must be done by a specialist,
who is familiar with the risk involved.
When the instrument is set into operation after the case has been
removed, attention must be paid to the acceleration voltage for
the CRT approx. -2025V and to the operating voltages for both
final amplifier stages 175V and 146V. Potentials of these voltages
are on the PS-Board, the CRT-PCB and on the main PCB. They are
highly dangerous and therefore precautions must be taken. It
should be noted furthermore that shorts occurring on different
points of the CRT high voltage and unblanking circuitry will
definitely damage some semiconductors. For the same reason it
is very risky to connect capacitors to these points while the
instrument is on.
Capacitors in the instrument may still be charged, even when the
instrument is disconnected from all voltage sources. Normally,
the capacitors are discharged approx. 6 seconds after switching
off. However, with a defective instrument an interruption of the
load is not impossible. Therefore, after switching off, it is
recommended to connect one by one all terminals across 1kΩ to
ground (chassis) for a period of 1 second.
Caution
Handling of the CRT needs utmost caution. The glass bulb
must not be allowed under any circumstances to come into
contact with hardened tools, nor should it undergo local
superheating (e.g. by soldering iron) or local undercooling (e.g.
by cryogenic-spray). We recommend the wearing of safety
goggles (implosion danger). If any work has been carried out,
the complete instrument (with case closed and POWER
button depressed) must undergo a safety test, including a
voltage test with 2200V DC, between accessible parts to both
mains/line supply terminals. This test is dangerous and requires
an adequately trained specialist. The maximum impedance
between each accessible metallic part and the safety earth
connector at the 3-pole power connector must be less than
0.1Ω.
Operating Voltages
adjustment is achieved, when the trace can be blanked in XY
analog mode and, in addition, when the requirement described
in the Test Instructions are met.
Astigmatism control
The ratio of vertical and horizontal sharpness can be adjusted
by the variable resistor of 47kΩ, located on the CRT PCB. As
a precaution however, the voltage for the vertical deflecting
plates (approx. +71V when the trace is in center position)
should firstly be checked, because this voltage will affect the
astigmatism correction. While the adjustment is being carried
out (with medium brightness and a 1MHz square-wave signal),
the upper horizontal square-wave tops are firstly focussed
with the FOCUS control. Then the sharpness of the vertical
lines are corrected with the 47kΩ Astigm. pot. The correction
should be repeated several times in this sequence. The
adjustment is optimized, when the FOCUS knob exclusively
brings no improvement of the sharpness in both directions.
Trigger Threshold
The internal trigger threshold should be in the range 0.3 to
0.5div display height.
Trouble-Shooting the Instrument
For this job, at least an isolating variable mains/line transformer
(protection class II), a signal generator, an adequate precise
multimeter, and, if possible, an oscilloscope are needed. This
last item is required for complex faults, which can be traced by
the display of signal or ripple voltages. As noted before, the
regulated high voltage and the supply voltages for the final
stages are highly dangerous. Therefore it is recommended to
use totally insulated extended probe tips, when troubleshooting the instrument. Accidental contact with dangerous
voltage potentials is then unlikely. Of course, these instructions
cannot thoroughly cover all kinds of faults. Some commonsense will certainly be required, when a complex fault has to
be investigated.
If trouble is suspected, visually inspect the instrument
thoroughly after removal of the case. Look for loose or badly
contacted or discolored components (caused by overheating).
Check to see that all circuit board connections are making good
contact and are not shorting to an adjacent circuit. Especially
inspect the connections between the PCBs. This visual
inspection can lead to success much more quickly than a
systematic fault location using measuring instruments. Prior
to any extensive trouble-shooting, also check the external
power source.
The operating voltages ( +12V, -6V, -2025V) are stabilized by
the switch mode power supply. The +12V supply is further
stabilized and used as a reference voltage for -6V stabilization.
These different operating voltages are fixed voltages, except
the +12V, which can be adjusted. The variation of the fixed
voltages greater than 5% from the nominal value indicates a
fault. Measurements of the high voltage may only be
accomplished by the use of a sufficient highly resistive voltmeter
(>10MΩ). You must make absolutely sure that the electric
strength of the voltmeter is sufficiently high. It is recommended
to check the ripple and also the interaction from other possible
sources. Excessive values might be very often the reason for
incomprehensible faults.
Maximum and Minimum Brightness
A variable resistor (100kΩ), located on the CRT PCB, is used
for this adjustment procedure. It may only be touched by a
properly insulating screwdriver (Caution! High voltage!). Correct
38
If the instrument fails completely, the first and important step
- after checking the power fuses - will be to measure the
deflecting plate voltages of the CRT. In almost any case, the
faulty section can be located. The sections represent:
1. Vertical deflection.
2. Horizontal deflection.
3. CRT circuit.
4. Power supply.
While the measurement takes place, the position controls of
both deflection devices must be in mid-position. When the
deflection devices are operating properly, the separate voltages
of each plate pair are almost equal then (Y approx. 71V and X
approx 90V). If the separate voltages of a plate pair are very
different, the associated circuit must be faulty. An absent
trace in spite of correct plate voltages means a fault in the CRT
circuit. Missing deflection plate voltages is probably caused by
a defect in the power supply.
Subject to change without notice
RS232 Interface - Remote Control
Adjustments
As advised in the Operating, Test and Service Instructions,
small corrections and adjustments are easily carried out with
the aid of the “Service Manual”. However, a complete
recalibration of the scope should not be attempted by an
inexperienced operator, but only someone with sufficient
expertise. Several precision measuring instruments with cables
and adapters are required, and only then should the pots and
trimmers be readjusted, provided that the result of each
adjustment can be exactly determined. Thus for each operating
mode and switch position, a signal with the appropriate sine or
square waveform, frequency, amplitude, risetime and duty
cycle is required.
As described in the “Menu” section, the instrument`s software
contains the submenu “CALIBRATE”.
be used by operators not equipped with precision instruments.
After the required warm-up time of 20 minutes, press and hold
the MENU pushbutton until the MAIN MENU becomes
visible. Then the submenu ”CALIBRATE” must be called.
In the “CALIBRATE” submenu three items can be chosen:
1: Y AMP (amplifier channel I and II)
2: TRIGGER AMP (trigger amplifier channel I and II)
3: STORE AMP (digital section)
The following items can
located on the rear of the instrument. Via this bi-directional
port, the instrument parameter settings and (in storage mode)
signal data, can be transmitted to, or received from a PC.
RS-232 Cable
The maximum connecting cable length must not exceed 3
meters and must contain 9 lines connected 1:1. The oscilloscope
RS232 connection (9 pole D-SUB female) is determined as
follows:
Pin
2Tx data (data from oscilloscope to external device)
3Rx data (data from external device to oscilloscope)
7CTS (clear to send)
8RTS (request to send)
5Ground (reference potential - connected via the
oscilloscope’s power cord with protective earth)
9+5V supply voltage for external devices (max. 400mA)
The maximum voltage swing at pin 2, 3, 7 and 8 is ± 12
Volt.
RS-232 protocol
N-8-2 (no parity bit, 8 data bits, 2 stop bits, RTS/CTS hardware
protocol).
Attention!
During the automatic adjustment procedure no signal
may be applied on any input.
Each item is called by pressing and holding the SAVE pushbutton.
Then the instrument automatically adjusts and stores the
adjustment values. In case of the Y amplifiers these are the field
effect transistor‘s operating points, the invert and the variable
gain balance. For the trigger amplifier it is regarding the DC
operating points and the trigger threshold. In storage mode the
adaptation of the storage display to the analog display.
The new adjustment values are always present after the
instrument is switched on. Therefore it is not required to
overwrite the factory adjustments by calling the OVERWRITEFACTORY DEFAULT function in the FACTORY submenu.
Note! The automatic adjustment should only be made if the
instrument‘s warm up time has passed and no faults such as
incorrect operating voltages are present. During some automatic
adjustment procedures the readout indicates “Working”.
RS232 Interface - Remote Control
Safety
Caution:
All terminals of the RS232 interface are galvanically
connected with the oscilloscope and subsequently with
protective (safety) earth potential.
Baud-Rate Setting
After the first POWER UP (switching on of the oscilloscope )
and the first command SPACE CR (20hex, 0Dhex) sent from
the PC, the baud rate is recognized and set automatically
between 110 baud and 19200 baud. The oscilloscope is then
switched over to REMOTE control mode. The oscilloscope
then transmits the RETURNCODE: 0 CR LF to the PC. In this
status all settings (with the exception of those functions
mentioned under “Controls and readout”) can be controlled
via the interface only.
The only ways to quit this status are:
• Switching the oscilloscope off, transmitting the command
• RM= 0 from the PC to the oscilloscope, or
• depressing the AUTO SET ( LOCAL ) pushbutton,
if in unlocked condition (command LK=1... was not sent)
After the remote state has been switched off the RM -LED (3)
is dark.
Please note:
A minimum time must elapse between the commands
RM=1... (remote on) and RM=0... (remote off) and vice
versa. The time can be calculated with the formula:
= 2/baud rate + 60µs.
t
min
Measurement on a high level reference potential is not
permitted and endangers operator, oscilloscope, interface and
peripheral devices.
In case of disregard of the safety warnings contained in this
manual, HAMEG refuses any liability regarding personal injury
and/or damage of equipment.
Operation
The oscilloscope is supplied with a serial interface for control
purposes. The interface connector (9 pole D- SUB female) is
Subject to change without notice
If at the beginning no SPACE CR command is recognizable, the oscilloscope pulls the TxD line low for approx.
0.2ms and causes a break on the PC.
Data Communication
After successfully being set to remote control mode, the
oscilloscope is prepared for command reception.
A disk with programming examples and a list of the instrument
commands is supplied with the oscilloscope.
39
Front Panel HM407
40
Subject to change without notice
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.