HAMAMATSU L7810 Datasheet

NEW
Calibrated Xenon Light Source L7810
is a logo mark for an accreditation system based on Japanese Meas­urement Law. Our Planning & Engi­neering Division is accredited in accordance with ISO/IEC Guide 25: 1990 in the field of photometry cali­bration. (Our accreditation number is 0081.) [JCSS: Japan Calibration Service System (The Japanese Goverment accreditation scheme for calibration laboratories and a signatory of APLAC Mutual Recognition Arrange­ment)]
LEFT: LAMP HOUSING, RIGHT: POWER SUPPLY
FEATURES
High Stability
Stability is one of the important characteristics of calibrated light sources. Hamamatsu’s excellent cathode technology and optimized lamp housing realize high stability and good repeatability.
TLSOB0065EA TLSOB0063EA
110 100
90 80 70 60 50 40 30
RELATIVE INTENSITY (%)
20 10
0
10 20
30 40 50 60 70 80 90 100 110
OPERATING TIME (h)
High Power
A Xenon lamp emits high intense UV, visible and IR light so that a reliable measurement can be possible even in the range of 300 nm to 400 nm, where either a deuterium lamp or a tungsten lamp doesn’t deliver enough light.
Comparison of Spectral Irradiance (Reference data)
: 50 cm away from light source
100
)
-1
·nm
-2
SPECTRAL IRRADIANCE (µW·cm
0.001
CALIBRATED XENON LIGHT SOURCE
10
1
0.1
0.01
200 250 300 350 400 450 500 550 600 650 700 750 800
CONVENTIONAL HALOGEN LAMP
CONVENTIONAL DEUTERIUM LAMP
WAVELENGTH (nm)
SPECIFICATIONS
Lamp Input Wattage Spectral Distribution Calibrated Range
A
Type of Calibration Calibration Unit Stability
Fluctuation (p-p)
Drift Window Material (Lamp Housing) Warm-up Time Recommended Calibration
B
Period Guaranteed Life
C
Xenon Lamp 150 W 185 nm to 2000 nm 250 nm to 800 nm Spectral Irradiance
-2
-1
µW.cm
.
nm
1.0 % Max. ±0.5 %/h Max. Synthetic Silica
10 min 100 hours or 1 year whichever comes first 500 hours
Input Voltage (switched manually) Power Consumption Cooling (Lamp Housing, Power Supply) Ambient Temperature Humidity Dimensions (W × H × D)
Lamp Housing Power Supply
D
Case Lamp Housing
Weight
Power Supply
D
Case
100 Vac/115 Vac/200 Vac/230 Vac (±10 %), 50/60 Hz 500 VA Forces Air Cooling (cooling fan is included) 0 °C to 40 °C 30 % to 90 % 100 mm × 280 mm × 100 mm 144 mm × 176 mm × 280 mm 595 mm × 425 mm × 345 mm Approx. 3.5 kg Approx. 10 kg Approx. 7.8 kg
NOTE: AThe calibrated value is the light power at 50 cm away from front surface of lamp housing.
BThis is the recommended calibration period when high reliability is required. Please contact to our sales office for re-calibration fee. CThis is the guaranteed time the unit satisfies requirements as calibrated light source. DThis is an accessory coming with the unit.
DIMENSIONAL OUTLINES (Unit: mm)
LAMP HOUSING POWER SUPPLY
COOLING FAN
M37 P=1
OUTPUT WINDOW ( 28)
CONNECTOR *
[FRONT VIEW]
50
100
[BOTTOM VIEW]
5
M4
280
140
SIDE COVER MOUNTING SCREW 4-M3 HEX SOCKET HEAD BOLT
[SIDE VIEW]
60
5 100 5
REFERENCE SURFACE OF CALIBRATION
SIDE COVER (REMOVE THIS COVER WHEN REPLACING LAMP)
77
280
XENON LAMP POWER SUPPLY
TIME COUNTER
hours
0.0
M 1A/125V
CURRENT SEL
7.5 8.0 8.5
LAMP ON
[Adc]
POWER
POWER INTER LOCK OVER HEAT
[FRONT VIEW] [SIDE VIEW] [REAR VIEW]
CONNECTOR *
144
SN.
OUTPUT
100/115/200/230
VOL. SEL
50/60 HZ
AC INPUT CONNECTOR
176
78
39
78
4-M4 PLASTIC FEET ARE ATTACHED TO THE BOTTOM OF THE UNIT BUT CAN BE REMOVED TO USE THEIR HOLES FOR INSTALLATION OF THE LAMP HOUSING.
39
* For cable to connect lamp housing with power supply, length: 2000
TLSOA0066EA TLSOA0067EA
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are
subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2000 Hamamatsu Photonics K.K
HOMEPAGE URL http://www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Electron Tube Center 314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205
U.S.A.:
Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: usa@hamamatsu.com
Germany:
Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: info@hamamatsu.de
France:
Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: france@hamamatsu.com
United Kingdom:
North Europe: Italy:
Hamamatsu Photonics UK Limited: Lough Point, 2 Gladbeck Way, Windmill Hill, Enfield, Middlesex EN2 7JA, United Kingdom, Telephone: 44(20)8-367-3560, Fax: 44(20)8-367-6384 E-mail: info@hamamatsu.co.uk
Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: info@hamamatsu.se
Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: info@hamamatsu.it
TLSO1039E01 JUN. 2000 SI Printed in Japan(1000)
Loading...