HAMAMATSU L7047, L2570, L2273, L2482, L2175 Datasheet

...
NEW
HIGH PRECISION LAMP HOUSING E7536
FEATURES
TLSXF0129
High accuracy
The E7536 lamp housing was designed to extract the full capabilities of lamps. One example of this is the stationary emission point which ensures high accuracy and easy opti­cal design. The E7536 also delivers high output stability, making it ideally suited for high-precision photometric appli­cations. Light output stability(p-p) is better than 0.3 % Typ. (when used with a Hamamatsu Xenon lamps) or 1.0 % Typ. (when used with a Hamamatsu Mercury-Xenon lamps).
High output intensity and collimated light flux
A reflective mirror and focus lens are incorporated. The mir­ror collects light from a lamp very efficiently to obtain a high output of over 1.5 times. The focus lens can be used to pro­vide collimated light output.
Built-in starter for improved lamp start-up operation
The built-in starter prevents high-voltage leakage, thus improving start-up operation of the lamp and also minimiz­ing noise which might otherwise occur at lamp start-up.
Adjustable light flux
The focus lens can be moved back and forth by turning the screw on the front flange (with the accessory tool), to adjust the output light flux. The desired light flux can be easily obtained to match your specific application.
Easy optical axis adjustment during lamp replacement
The optical axis can be easily adjusted by using the acces­sory tool, without removing any lamp housing covers. Three adjustment screws are provided in easy-to-access posi­tions: one on the side of the housing (for right-left move­ment of the lamp), one at the upper part on the rear (for ver­tical movement of the lamp), and one in the center on the rear (for back and forth movement of the mirror).
Interlock function
If the side cover becomes loose or the temperature in the lamp housing rises to an abnormal level, the interlock function triggers automatically to stop lamp operation.
High safety
The built-in cooling fan maintains the surface temperature of the lamp housing to a safe level below +40 °C.
HIGH SAFETY
The internal cooling fan ensures safe operation.
Built-in reflective mirror
A starter is incorporated.
The built-in focus lens allows light flux adjustment.
Stationary emission point ensures high accuracy.
INTERLOCK FUNCTION
HIGH OUTPUT INTENSITY
IMPROVED START-UP OPERATION
EASY OPTICAL AXIS ALIGNMENT DURING LAMP REPLACEMENT
HIGH ACCURACY
ADJUSTABLE OUTPUT LIGHT
PATENT PENDING
TLSX1030E02 NOV. 1999 SI (500)
* PATENT PENDING: JAPAN 2
TLSXA0083EB
TLSXC0041EB
TLSXB0107EB
HIGH PRECISION LAMP HOUSING
Light sources in various experiments
Light sources in spectrophotometry
FEATURES
DIMENSIONAL OUTLINES (Unit: mm) APPLICATIONS
Block Diagram for Measurement
STABILITY
HAMAMATSU PHOTONICS K.K., Electron Tube Center 314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205
U.S.A.:
Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218
Germany:
Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658
France:
Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10
United Kingdom:
Hamamatsu Photonics UK Limited: Lough Point, 2 Gladbeck Way, Windmill Hill, Enfield, Middlesex EN2 7JA, United Kingdom, Telephone: 44(20)8-367-3560, Fax: 44(20)8-367-6384
North Europe:
Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01
Italy:
Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©1999 Hamamatsu Photonics K.K
100
COOLING FAN
FOCUS LENS BACK AND FORTH ADJUSTMENT SCREW M3 HEX SOCKET HEAD BOLT
M37 P=1
OUTPUT LENS ( 30)
50
140
280
510
60
77
5 100 5
78
39
78
39
FRONT VIEW
BOTTOM VIEW
SIDE VIEW
LAMP RIGHT-LEFT ADJUSTMENT SCREW M3
HEX SOCKET HEAD BOLT
SIDE COVER MOUTING SCREW 4-M3 HEX SOCKET HEAD BOLT
SIDE COVER (REMOVE THIS COVER WHEN REPLACING LAMP)
LAMP VERTICAL ADJUSTMENT SCREW
MIRROR BACK AND FORTH ADJUSTMENT SCREW
M4
4-M4 PLASTIC FEET ARE ATTACHED TO THE BOTTOM OF THE UNIT BUT CAN BE REMOVED TO USE THEIR HOLES FOR INSTALLATION OF THE LAMP HOUSING.
COMPUTERMETERPOWER SUPPLY
C7535
LAMP
HOUSING
E7536
300 mm
SILICON PHOTODIODE
150 W XENON LAMP L2274
DIFFUSER
0.1 % ND FILTER 10 % ND FILTER
APERTURE 1 × 1 mm
TIME
LIGHT OUTPUT
1.0 %
5 min
Parameter Description/Value
150 W
Xenon Lamps
L2175 L2195
Lamp
Optical
Starter Output Voltage Cooling Function Operating Ambient Temperature Recommended Operating Temperature Operating Humidity Dimensions (W × H × D) Weight Suitable Power Supply Type No.
Fused Silica Window Ozone-free Silica Window
Material Dimension Focal Point Light Output Material Dimension Curvature
Suitable Lamps (Rating and kind) Suitable Lamps
Type No. Lamp Current Lamp Voltage
Focus Lens
Reflective Mirror
Optical Axis Height
150 W
Mercury-Xenon Lamps
L2482 L7047
150 W (GS)
Xenon Lamps
L2273 L2274
8.5 A
18 V Typ.
C4264, C7535
Fused Silica
30 mm
F = 30 mm (550 nm)
Collimating
BK7 Al + MgF
2 Coating (All reflection)
30 mm
R = 30 mm
140 mm (exclude plastic feet height)
30 kV Typ.
Forced Air Cooling (Built-in automatic cooling fan
A
)
0 °C to +40 °C
+5 °C to +35 °C
30 % to 90 % (without moisture condensation)
100 mm × 100 mm × 280 mm (exclude projection part)
Approx. 3.5 kg
7.5 A
20 V Typ.
C4263, C7535
200 W
Mercury-Xenon Lamps
L2423 L2570
8.0 A
24 V Typ.
C7535
NOTE: A The cooling fan automatically starts when the temperature rises inside the lamp housing.
Loading...