A Wide Lineup of 248 nm, 365 nm and 436 nm Type
The C6080 series is a compact UV power meter specifically designed to measure the intensity of UV radiation emitted from
the light guide output end of a UV spot light source. The C6080
uses a UV sensor that exhibits stable sensitivity even after long
exposure to UV radiation, allowing reliable measurement with
good repeatability. As a useful option, an adapter for clamping
the light guide output end is also available.
FEATURES
●Compact and lightweight
●Excellent accuracy
●Absolute value display of UV intensity (mW/cm
●Easy-to-read digital display
●Sensor head isolated from display unit allows flexible
measurement.
●200-hours continuous operation by 9 V battery (006P)
●External output (analog output)
2
)
UV POWER METERS
C6080 SERIES
APPLICATIONS
●UV intensity measurement of UV curing and UV ink light
sources
●UV intensity measurement of various UV spot light sources
Figure 1: Spectral Response Characteristics
■C6080-02 ■C6080-03, -13 ■C6080-04
TPT B0013EA
100
90
80
70
60
50
40
30
RELATIVE SENSITVITY (%)
20
10
0
220
230 240 250 260 270 280
WAVELENGTH (nm)
TPT B0006EB TPT B0014EA
100
90
80
70
60
50
40
30
RELATIVE SENSITVITY (%)
20
10
0
300
320 340 360 380 400 420
WAVELENGTH (nm)
100
90
80
70
60
50
40
30
RELATIVE SENSITVITY (%)
20
10
0
360
380 400 440420 460 480 500 520
WAVELENGTH (nm)
Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office.
Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are
subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2003 Hamamatsu Photonics K.K.
UV POWER METERS C6080 SERIES
SPECIFICATIONS
Parameter C6080-02 C6080-03 C6080-04 C6080-13
Sensitive Area
Sensitive Wavelength Range
Calibrated Wavelength
Optical Power
Measurement Range *
SENS. L
SENS. H
Display
6
230 to 270
248
3·1/2 digits Liquid crystal digital display
320 to 400
365
1 to 1999
0.1 to 199.9
Sampling Interval
External Output (analog)
Operating Temperature Range
Power Requirement
1.000 V per liquid crystal display "1000"
0 to +50
9 V battery (006P) 1 piece
Current Consumption
Weight
Sensor Head (including cable)
Display Unit (including battery)
Carrying Case Size (W × H × D)
* Sensitivity is calibrated at the specific calibration wavelength, so the display value may vary when used to measure light source emissions on a different spectrum.
This is the recommended calibration period (every 1 year) when high reliability is required. Please contact to our sales office for re-calibration fee.
Approx. 100
Approx. 200
330 × 70 × 230
380 to 480
2.5
2
1
436
320 to 400
365
10 to 19990
1 to 1999
Figure 2: Dimensional Outline (Unit: mm)
Unit
mm
nm
nm
mW/cm
—
times/s
—
°C
—
mA Max.
g
mm
2
SENSOR HEAD
39
SENSITIVE AREA
6 (-02 ONLY)
1 (OTHER TYPES)
CABLE LENGTH
1 m
20
DISPLAY UNIT
ZERO ADJUST
TRIMMER
POWER
ON/OFF
37
80
* External output connector plug: PJ-3057-P (Sato Parts) or equivalent
Figure 3: Light Guide Clamp Adapter (Option) Dimensional Outline (Unit: mm)
1
6
Bottom View Side View
39
SENSOR
HEAD
5
17
SENSOR
HEAD
CABLE
46
20
Top View
LIGHT GUIDE
CLAMP SCREW (M3)
7 (E8718-01)
6 (E8718-02)
LIGHT GUIDE
RECEPTACLE
SENSOR HEAD
CLAMP SCREW (M3)
SENSOR HEAD CONNECTOR
EXTERNAL OUTPUT
CONNECTOR*
mw
2
cm
Adapter
Type No.
E8718-01
E8718-02
( 3.5 EARPHONE JACK)
HOLD SWITCH
145
SENSITIVITY
SELECTOR
TPT A0021EC
Applicable Outer Diameter
of The Light Guide End
7 mm
6 mm
TPT A0027EA
* Adapters for furcated light guide are also available.
* This adapter is designed to measure UV intensity under constant conditions, at a point 10 mm away from the light guide output end.
HAMAMATSU PHOTONICS K.K., Electron Tube Center
314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, P. O. Box 6910, Bridgewater. N.J. 08807-0910, U.S.A., Telephone: (1)908-231-0960, Fax: (1)908-231-1218 E-mail: usa@hamamatsu.com
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: info@hamamatsu.de
France: Hamamatsu Photonics France S.A.R.L.: 8, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: (33)1 69 53 71 00, Fax: (33)1 69 53 71 10 E-mail: infos@hamamatsu.fr
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: info@hamamatsu.co.uk
North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E-mail: info@hamamatsu.se
Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: info@hamamatsu.it
WEB SITE http://www.hamamatsu.com
TPT 1005E06
MAY 2003 IP (0202)
Printed in Japan (1000)