This instruction manual describes the features of the P82R differential pressure
transducer and provides guidelines for its use.
Improper use of this instrument or failure to follow these instructions may cause injury
or equipment damage. Every person who uses the device must therefore read the
manual and understand the possible risks. The instruction manual, and in particular
the safety precautions contained therein, must be followed carefully. Contact the
manufacturer if you do not understand any part of this instruction manual.
Handle this manual with care:
• It must be readily available throughout the lifecycle of the instrument.
• It must be provided to any individuals who assume responsibility for operating the
instrument at a later date.
• It must include any supplementary materials provided by the manufacturer.
The manufacturer reserves the right to continue developing this instrument model
without documenting such development in each individual case. The manufacturer
will be happy to determine whether this manual is up-to-date.
Conformity
This instrument corresponds to the state of the art and meets all legal requirements
set forth in EC directives as evidenced by the CE label.
The manufacturer owns the copyright to this instruction manual. This manual contains
data, instructions and drawings pertaining to the features and usage of this
instrument; copying this manual in part or in full or distributing it to third parties is
prohibited.
3
P82R Instruction Manual
1 Safety precautions
1.1 Appropriate use
In addition to differential pressure data, the P82R differential pressure transducer also
records positive and negative overpressures.
Always observe the operating requirements—particularly the permissible supply
voltage—indicated on the rating plate and in the "Technical data" section of this
manual.
The instrument may only be handled as indicated in this manual. Modifications to the
instrument are prohibited. The manufacturer is not liable for damages caused by
improper use or failure to follow these instructions. Violations of this type render all
warranty claims null and void.
1.2 Shipping, assembly, electrical connections and start-up
Do not close the pressure input ports when shipping, as changes in barometric
pressure could damage instruments with low measuring ranges.
Only technical personnel who are appropriately trained and authorized by the
operator of the facility may assemble the instrument and set up its electrical
connections.
The instrument may only be operated by appropriately trained individuals who have
been authorized by the operator of the facility.
Pressurized air or breath is not to be used for performance tests, as this could
damage instruments with low measurement ranges.
Measurement errors may occur if the instrument is not kept protected from sunlight.
Specific safety precautions are given in individual sections of this manual.
The individual responsible for the electrical connections must be notified immediately
if the instrument is damaged or if errors occur that cannot be corrected as indicated in
section 6.
This individual must take the instrument out of service until the error has been
corrected and ensure that it cannot be used unintentionally.
Always unplug the power cord before opening the instrument!
This instrument requires no maintenance.
Only the manufacturer may perform repairs that require the housing to be opened.
4
Loading...
+ 9 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.