EMA 84 Instruction Manual
8
3 Start-up
3.1 Electrical connections
Use only IEC 6 LR 51 batteries.
Open the battery compartment on the rear of the instrument by sliding down the silver
locking button. Clip the battery to the battery connector, making sure that the battery
terminals match those of the connector. Place the battery in the compartment and
replace the cover.
3.2 Pneumatic connections
Connect the EMA 84 to the pressure source at the overpressure port (+) when
measuring overpressure, at the vacuum port (-) when measuring vacuum pressure, or
at both ports (making certain to connect the + and - ports appropriately) when
measuring differential pressure.
The two valves next to the pressure ports should always be kept open (by turning all
the way to the left) when transporting the instrument or when beginning a
measurement. Turning the valves to the right will close them.
The valve of the overpressure port (+) must be closed when measuring overpressure,
and the valve of the vacuum port (-) must be closed when measuring vacuum
pressure. Both valves should be closed when measuring differential pressure.
3.3 Operating the instrument
The battery voltage must be greater than 7 V for proper operation. The battery must
be replaced if this is not the case. “LO BAT” will be displayed when the battery charge
reaches a critical level.
Turn the instrument on by flipping switch D to the up position.
Use switch C to set the desired units (kPa – mbar).
Use switch A to set the desired response time. The lowest possible response time
(“FAST” or 20 ms) is generally the best option. If, however, the pressure being
measured fluctuates to the point where a precise reading cannot be taken, the
response time can be lengthened by flipping switch A up to “MED” (0.1s) or “SLOW”
(1 s).
Proceed as outlined in section 4 (“Zero-point calibration”) if the display does not read
000 when the valves are open and/or when the pressure ports are disconnected. The
instrument will then be ready for use.