Pharmaceutical Refrigerator
Operation Manual
Model:
HYC-68
HYC-68A
HYC-118
Certificate of Quality
Checker:
Haier Medical and Laboratory Products Co., Ltd.
Address: Haier Industrial Park, Economic Technology Development
Zone. Qingdao 266510.P.R.China
Web:www.haiermedical.com
HYC-118A
Version:2nd,2017
Dedicated code:0270501000
V13026
0270501000
Read the Operation Manual carefully before using your appliance.
Keep the Operation Manual in a safe place.
Appearance , color and layout of the door may vary.
Translation of the original instruction.
Haier pharmaceutical refrigerators are applicable for the storage of pharmaceutical
products, medicines, and other laboratory products that need to be stored at a
temperature of 2
℃ to 8℃.
Packing List
EC Declaration of Conformity
according to the Machinery Directive 2006/42/EC
For the following equipment :
Product : Pharmaceutical Refrigerator
Type Designation/Trademark:
Manufactures Name : Haier Medical and Laboratory Products Co.,Ltd.
Manufactures Address : Haier Industrial Park, Economic Technology
Development Zone. Qingdao 266510.P.R.China
Refer to in this declaration confirms with the following directive(s)/standards:
HYC-68
HYC-68A
Haier
HYC-118
HYC-118A
Model
Operation Manual
Plastic bag
Key
Drain hole cap
Note: Technical information may vary from the appliance that you have just purchased due to
technical improvements. Please take care of the key, if lost, please contact the seller.
HYC-68
1
1
2
1
HYC-68A
1
1
2
1
HYC-118
1
1
2
1
HYC-118A
1
1
2
1
Global Warming Potential
Model
HYC-68 220~240 50/60 0.072
HYC-118 220~240 50/60 0.129
HYC-118A 220~240 50/60 0.122
into the atmosphere.
Rated voltage
(
Refrigerant type GWP
R134a 1430
)
VAC
GWP=global warming potential
Rated frequency
(Hz)
CO
equivalent
2
(Tonnes)
2006/42/EC, EN60335-1, EN60335-2-89 EN62233
21
Name
Pharmaceutical Refrigerator
Model
Exterior Dimensions
(W×D×H)
(mm)
Interior Dimensions
(W×D×H)
(mm)
Effective Volume
Door
Insulation
Compressor
Shelves/Basket
Load
Refrigerating Method
Exterior / Interior
Condenser/Evaporimeter
Temperature Control
HYC-118
597×635×835
515×415×630
118L
Solid door
CFC-Free foamed-in-place urethane
High Quality Hermetically sealed compressor
3/1
Max:10kg/m²
Force-air cooling circulation
sprayed steel plate /PS plate sprayed steel plate /PS plate
Wire and tube / Plate type
Microprocessor controlled
Glass door with electric heat
Force-air cooling circulation
HYC-118A
597×635×835
515×415×630
118L
3/1
Max:10kg/m²
Wire and tube / Plate type
Microprocessor controlled
The product is applicable to pharmacies, pharmaceutical factories, quarantine stations, health
centers and hospitals, used to store biological products and those need to be stored at a
temperature between 2~8℃.
Temperature
Control
Safety
System
Microprocessor control, digital display of temperature, temperature
display accuracy as 1℃ and temperature range as 2℃~8℃.
Multi-fault alarm (high and low temperature alarm, power failure
alarm, sensor error alarm and door ajar alarm,low battery alarm)
Two alarm methods (audible buzzer and visual)
All independent components are safely grounded
Refrigerant
Lamp
Net Weight
Temperature Range
Voltage
Rated Power/Current
Noise Level(Lp)
Foaming Agent
Climate Type
Anti-shock Safety Classification
Power connection type
Alarm system
Battery duration for alarm system
Note: Climate Type 4 means the temperature is +30℃,the relative humidity is 55%.Technical information may vary from
the appliance that you have just purchased due to technical improvements.
High & low temperature alarm, sensor error alarm, power failure alarm,
R134a 90g
LED 0.3W
41kg
2℃~8℃
220V-240V~/50/60Hz
135W/1.03A
32dB(A)
CP/IP
4
I
Y
door ajar alarm,low battery alarm
at least 8h (when the battery is fully charged)
R134a 85g
LED 0.3W
46kg
2℃~8℃
220V-240V~/50/60Hz
165W/1.16A
32dB(A)
CP/IP
4
I
Y
Optimized refrigeration system, hermetically sealed refrigeration
Refrigeration
System
User-friendly
Design
compressor
Automatic defrost maintains optimum cooling capacity
User-friendly design, Microprocessor control, smart and carefree,
adjustment not required
High-performance thermal insulation layer, with excellent insulation
effect
Automatic condensate removal, no drain lines required
Note: Technical information might be somewhat different on your refrigerator than
published due to continuous improvement.
20
1
Specification
Specification
Content:
Product Feature........................................................
Content.....................................................................
Application guidelines...............................................
Precautions for Use..................................................
Product Installation ...................................................
Component Names Control Panel...........................
Application method ...................................................
1
2
3
6
7
11
13
Name
Model
Exterior Dimensions
(W×D×H)
(mm)
Interior Dimensions
(W×D×H)
(mm)
Effective Volume
Door
Insulation
Compressor
Shelves/Basket
Load
Refrigerating Method
Exterior / Interior
Condenser/Evaporimeter
Pharmaceutical Refrigerator
HYC-68
495×580×660
415×385×505
68L
Solid door
CFC-Free foamed-in-place urethane
High Quality Hermetically sealed compressor
2/1
Max:10kg/m² Max:10kg/m²
Force-air cooling circulation
sprayed steel plate /PS plate
Wire and tube / Plate type
HYC-68A
495×580×660
415×385×505
68L
Glass door
2/1
Force-air cooling circulation
sprayed steel plate /PS plate
Wire and tube / Plate type
Cleaning and Maintenance.......................................
FAQ..........................................................................
Refrigeration Schematic & Wiring Diagram..............
Specification.............................................................
15
16
17
19
Temperature Control
Lamp
Net Weight
Temperature Range
Voltage
Rated Power/Current
Refrigerant
Noise Level(Lp)
Foaming Agent
Climate Type
Anti-shock Safety Classification
Power connection type
Alarm system
Battery duration for alarm system
Note: Climate Type 4 means the temperature is +30℃,the relative humidity is 55%.Technical information may vary from
the appliance that you have just purchased due to technical improvements.
Microprocessor controlled
LED 0.3W
38kg
℃~8℃
2
220V-240V~ /50/60Hz
84W/0.65A
R134a 50g
35dB(A)
CP/IP
4
I
Y
High & low temperature alarm, sensor error alarm, power failure alarm,
door ajar alarm,low battery alarm
at least 8h (when the battery is fully charged)
Microprocessor controlled
LED 0.3W
38kg
℃~8℃
2
220V-240V~/50/60Hz
84W/0.65A
R600a 22g
35dB(A)
CP/IP
4
I
Y
2