Appendix A Data Communications...............................................................................................35
Index ......................................................................................................................................................37
4
Section 1Specifications
Specifications are subject to change without notice.
General
Dimensions (W x L x H)114.30 mm x 209.60 mm x 57.20 mm (4.5" x 8.25" x 2.25")
Weight1.0 kg (2.2 lb)
Number of size channels2
Standard
IAQ0.5, 5.0 µm
Custom0.3 to 20 µm available
Flow rate0.1 cfm (2.83 L/min)
Light sourceLaser diode; index guided (26,000 MTBF at 25 °C)
CalibrationPSL particles in air (NIST traceable)
Counting efficiency50% at 0.3 µm; 100% for particles > 0.45 µm (per JIS B9921:1997)
Zero count1 count/5 minute (per JIS B9921:1997)
Coincidence loss5% at 2,000,000 particles per ft
Relative humidity20% to 90%, non-condensing
Temperature± 3 °C, 10 °C to 40 °C (50 °F to 104 °F)
Data storage100 sample records (rotating buffer)
Data recordedCounts, calibration error
DisplayLCD, 4 x 16 character display
AlarmsLow battery, calibration error
Count modesConcentration, totalize, audio
Delay time0 to 24 hours
Sample inletIsokinetic probe
InterfaceRS232 through RJ45
Vacuum sourceInternal pump, flow controlled
HousingInjection molded plastic
PowerAC Adapter, 12 VDC at 2.5 A, 90 to 250 VAC, 50 to 60 Hz
Rechargeable batteryNiMH, 4.8 V at 4.5 Ah; replaceable
Recharge time2 hours
Continuous Operating Time8 hours
StandardsCE, JIS B9921: 1997
Environment
0.3, 0.5 µm
0.3, 1.0 µm
3
Operating temperature10 °C to 40 °C (50 °F to 104 °F)
Operating relative humidity20% to 90%, non-condensing
Storage temperature–10 °C to 50 °C (14 °F to 122 °F)
Storage relative humidityUp to 90%, non-condensing
Compliance
JIS B 9921–1997—Light Scattering Automatic Particle Counter
Standards
Federal Standard 209E—Standard Practice for Defining Size Calibration,
Resolution and Counting Accuracy of an Air-Borne Particle Counter Using
Near-Monodispersed Spherical Particulate Material.
5
Specifications
6
Section 2General Information
2.1Safety information
Read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious
injury to the operator or damage to the equipment.
To make sure that the protection provided by this equipment is not impaired, do not use
or install this equipment in any manner other than that specified in this manual.
2.1.1 Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation which, if not avoided, will
result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that can result in minor or
moderate injury.
Important Note: Indicates a situation which, if not avoided, can cause damage to the
instrument. Information that requires special emphasis.
Note: Information that supplements points in the main text.
2.1.2 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol, if noted on the instrument, will be
included with a danger or caution statement in the manual.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol cannot be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC),
European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no
charge to the user.
Note: To return for recycling, contact the equipment producer or supplier for instructions on how to return end-of-life
equipment, producer-supplied electrical accessories and all auxiliary items for proper disposal.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or
electrocution exists.
This symbol, if noted on the product, indicates the need for protective eye wear.
This symbol, when noted on the product, identifies the location of a fuse or current limiting device.
This symbol indicates a laser device is used in the equipment.
This symbol, when noted on the product, indicated the presence of devices sensitive to Electro-static Discharge
(ESD) and indicated that care must be taken to prevent damage with the equipment.
7
General Information
2.1.3Electrostatic Discharge (ESD) considerations
Important Note: To minimize hazards and ESD risks, maintenance procedures not
requiring power to the analyzer should be performed with power removed.
Delicate internal electronic components can be damaged by static electricity, resulting in
degraded instrument performance or eventual failure.
The manufacturer recommends taking the following steps to prevent ESD damage to
the instrument:
•Before touching any instrument electronic components (such as printed circuit cards
and the components on them) discharge static electricity. This can be accomplished
by touching an earth-grounded metal surface such as the chassis of an instrument,
or a metal conduit or pipe.
•To reduce static build-up, avoid excessive movement. Transport static-sensitive
components in anti-static containers or packaging.
•To discharge static electricity from the body and keep it discharged, wear a wrist
strap connected by a wire to earth ground.
Handle all static-sensitive components in a static-safe area. If possible, use anti-static
floor pads and work bench pads.
2.1.4 Battery safety information
WARNING
An explosion can occur if the internal battery is replaced incorrectly.
Figure 1 shows the label that is displayed on the battery for the safety of the user.
Figure 1 Battery safety label
8
2.2General product information
The MET ONE Handheld Airborne Particle Counter–2 (HHPC–2) (refer to Figure 2) is a
portable instrument that measures, counts, stores and reports airborne particulates. The
standard MET ONE HHPC–2 holds 100 samples in the memory. The data records
include date, time, counts, sample labels, volume, alarm flags, temperature and relative
humidity. The user can download the data to a personal computer (PC) using the
included interface cable and software. The internal battery powers up the MET ONE
HHPC–2 when mobile or when connected to the AC power adapter.
Note: The AC power adapter provided with the unit is used for extended or stationary sampling.
The MET ONE HHPC–2 is used:
•To monitor the air quality in clean-rooms
•To monitor the air quality in manufacturing processes
•To monitor the air quality in pharmaceutical production
General Information
Figure 2 MET ONE HHPC–2
1Front view of the MET ONE HHPC–22Back view of the MET ONE HHPC–2
1Bottom view of the MET ONE HHPC–2
9
10
Section 3Installation
DANGER
Only qualified personnel should conduct the tasks described in this section of
the manual.
3.1Unpack the instrument
Remove the components from the shipping container and inspect for damage. Verify that
all the items listed in Figure 3 are included. If any items are missing or damaged, contact
the manufacturer or sales representative.
Retain the original packaging materials. Use the original packaging material to store or
ship the instrument to protect against damage during storage or transportation.
Figure 3 List of packaged items
1MET ONE HHPC–2 66¼”–OD high purity tubing (VP792002)
2Zero count filter7MET ONE HHPC–2 Utility software (CS200025)
3Intake nozzle, 1/8” hose barb (MP000026–01)8RS232 modular cable (VP894408)
4Black cap/Intake nozzle cover9RJ45 to DB9 adaptor (SA000070–01)
5US power cord (CS200025)10 AC/DC power supply (VP624005)
11
Installation
3.2Installation guidelines for MET ONE HHPC-2
The MET ONE HHPC–2 is ready to use when shipped and has default configuration
settings. Refer to the following general guidelines during installation.
1. Lift the MET ONE HHPC–2 from the shipping container or the carrying case and
place it on a flat surface.
2. Remove the layer of protective plastic from the LCD screen.
3. Remove the isokinetic probe cover (black plastic cap) before the startup of the MET
ONE HHPC–2.
Note: The pump and other internal electrical components will be damaged if the cap is left on.
Note: The battery is charged before shipping, but may need to be charged before operation (refer to
section 6.1 on page 27 for more information).
3.3Mechanical installation
The MET ONE HHPC–2 is shipped with two different intake devices:
•Isokinetic probe
•Intake hose barb
Important Note: Do not use a wrench to connect or disconnect the intake nozzle or
isokinetic probe. Only finger-tighten the connection.
3.3.1 Install the isokinetic probe
The isokinetic probe (refer to Figure 4 on page 13) is used to sample ambient air
samples. The isokinetic probe is pre-installed and ready for use as soon as the MET ONE
HHPC–2 is unpacked. To install the isokinetic probe:
1. Attach the isokinetic probe to the intake nozzle on the MET ONE HHPC–2 by turning
the isokinetic probe to the right.
2. Remove the black plastic cap attached to the isokinetic probe.
Note: Make sure that the cap is attached to the isokinetic probe when the probe is not in use.
The cap protects the MET ONE HHPC–2 from debris and contamination.
To sample areas that are difficult to reach, connect longer tubing to the isokinetic probe
with a hose barb end.
Note: Use a maximum of four feet of high purity or approved tubing only. Incorrect size or type of
tubing will affect the sample accuracy.
3.3.2 Install the intake hose barb
The intake hose barb (refer to Figure 4 on page 13) is connected to the MET ONE
HHPC–2 for zero counting and purging or to sample areas that are difficult to reach.
To install the intake hose barb:
1. Unscrew the isokinetic probe from the MET ONE HHPC–2.
12
2. Attach the hose barb.
3. Remove the protective cap from the intake hose barb.
Installation
Figure 4 Isokinetic probe with tubing and zero count filter
1Hose barb fitting3Tubing
2Isokinetic probe4
MET ONE HHPC–2
3.3.3 Install the zero count filter
In a clean-room application, the zero count filter (refer to Figure 3 on page 11, item 2)
verifies if the MET ONE HHPC–2 is not counting electrical noise signals from internal
components and is not subject to external interference. In indoor air quality and other
applications, the zero count filter cleans out the sensor immediately after a high
concentration sampling.
To connect the zero count filter:
1. Remove the isokinetic probe from the MET ONE HHPC–2.
2. Attach the intake hose barb (MP000026–01).
3. Secure the high purity tubing (VP792002) to the hose barb.
4. Secure the opposite end of the tubing to the zero count filter (VP212808).
Note: Make sure that the arrow of airflow direction or the indicator on the zero count filter points
towards the tubing.
13
Installation
3.3.4 Connect the AC/DC power supply
Important Note: Do not use a power supply unit that does not supply the required voltage
and charging current (polarity) to operate the MET ONE HHPC–2.
To connect the external power supply:
1. Turn off power to the MET ONE HHPC–2 unit.
2. Plug one end of the power cord into the external power supply and the other end into
an electrical wall receptacle.
3. Plug the external power supply jack into the MET ONE HHPC–2 power port (refer to
Figure 5, item 1).
Figure 5 Location of power port
1I/O port2Power port
14
Section 4Operation
4.1Install the MET ONE HHPC–2 utility software
To install the MET ONE HHPC–2 utility software:
1. Connect the MET ONE HHPC–2 to the communications port on the PC using the
RS232 modular cable (VP894408).
2. Connect the DB9 to the RJ45 converter (SA000070–01).
3. Insert the MET ONE HHPC–2 utility software (CSVDownloadUtility) diskette into the
drive.
4. Copy CSVDownloadUtility.exe and CSVDownloadUtility.ini from the diskette to a
working directory on the PC. If the communications port to be used is other than
Com1, open CSVDownloadUtility.ini and change Port=Com1 to Port=ComX (where X
represents the number of the port to use.)
5. To open the utility software, double-click the CSVDownloadUtility.exe file. The
CSVDownloadUtility window is displayed (refer to Figure 6 on page 15).
6. Follow the instructions in the Readme.txt file on the diskette to install the software.
Figure 6 MET ONE HHPC-2 Utility (CSVDownloadUtility) window
4.2Start up
Press and hold the POWER button on the right of the display of the MET ONE HHPC–2 to
turn on the unit. When the MET ONE HHPC–2 is powered on, the MET ONE Opening
screen is displayed briefly followed by the Main screen.
4.3Display and keypad
The MET ONE HHPC–2 features a Liquid Crystal Display (LCD) with a keypad at the
bottom. Power on the unit and the MET ONE Opening screen displays briefly, followed by
the Main screen.
Note: If the MET ONE HHPC–2 does not power up, double-check the AC connection or the battery
power. To charge the battery, refer to section 5.1 on page 25.
15
Operation
4.3.1 Keypad
Figure 7 shows the display of the Main screen of MET ONE HHPC–2. Table 1 on page
16 provides information about the keypad available at the bottom of the screen and the
MODESelects modes and saves data when the unit is counting
POWERToggles the power on and off
16
4.3.2 Screens
Operation
Press the corresponding button from the keypad at the bottom of the LCD to invoke the
following screens:
•Main screen—The main screen displays:
•Channel sizes
•Particle counts (only displayed when counting or in Buffer mode)
•Status of the counting process (Run or Wait)
•Battery charge indicator (amount of charge remaining in the battery, refer to Table
4 on page 25)
•Select mode screen
•Select volume screen
•Buffer count and Buffer options screen
•Display buffer screen
•Contrast screen
•Date and Clock setup screen
4.4Configuration setup
Once the MET ONE HHPC–2 is unpacked and connected to the appropriate intake
device, use the default settings to conduct a 1-minute count cycle. Download the data to
a PC and print the results.
To perform a 1-minute count cycle:
1. Remove the isokinetic probe cover (black plastic cap).
2. Press and hold the
turns on.
3. Press
START/STOP on the Main screen to start counting. The count cycle begins and
the bottom line of display changes from “Wait” to “Run”. After one minute, the count
cycle ends and the display clears.
Note: The results are saved automatically in the print buffer.
Refer to section 4.8 on page 22 to transfer the data to a PC and print the results.
4.5Select custom settings
To choose a setting, select an option that is displayed at the bottom of the Main screen
when the
through the options or save selections.
MODE key is pressed, then use the START/STOP and MODE keys to scroll
POWER key until the MET ONE HHPC–2 sounds two beeps and
4.5.1 Set the date and time
To set the date and time:
1. Press the
and date.
2. Press and hold the
and Time screen is displayed with the date selected.
3. Use the
MODE key six times on the Main screen. The bottom line displays the time
MODE key for one second until two beeps are heard. The Date
MODE key to select the digit. The cursor moves to the next digit.
17
Operation
4. When the date, month, year or time is set, press the START/STOP key to change the
numbers for that field.
5. To select another unit of time, press the
START/STOP key to change the numbers.
6. When the date and time are set, press and hold the
save the new setting and return to the Main screen.
4.5.2 Select the count mode
To select the count mode:
1. Press the
different modes available. The bottom line displays the current mode (refer to
Figure 8 on page 18).
2. Press the
3. When the required mode is displayed on the screen, press and hold the
for one second to save the new setting and return to the Main screen.
MODE key to select the next field. Press the
MODE key for one second to
MODE key once on the Main screen. Refer to Table 2 on page 18 for the
START/STOP or the MODE key.
MODE key
Figure 8 Select mode screen
Table 2 Count modes
ModeDescriptionSample unit
The Concentration mode is useful for taking a quick snapshot of
airborne particulate samples, especially in areas where
particulate levels are unknown and may exceed the operating
Concentration
Rate
To ta li ze
levels of the counter. The MET ONE HHPC–2 estimates count
per cubic foot or liter depending on the volume selected. The unit
updates the results by calculating completed sample counts and
volume compared to the balance of volume remaining.
In the Rate mode, counting is continuous. The sliding average of
ten samples is displayed.
In the Totalize mode, number of counts are displayed as they
accumulate, until the end of the sample. When the sample is
completed, the record is stored and the value is displayed on the
screen until the next sample is started.
Volume size
Volume unit
Volume size
18
4.5.3 Set the sample units and volume
Sample volume may be measured in either liters or cubic feet. The volume size can
range from 0.01 cubic foot or 1 liter up to 1 cubic foot or 28.3 liters.
4.5.3.1 Rate mode
In the Rate mode, the unit of the sample volume is selected.
To select the unit of sample volume:
1. Make sure the unit is set to Rate mode as described in section 4.5.2 on page 18.
Operation
2. Press the
MODE key twice on the Main screen. The bottom line displays the current
unit settings as shown in Figure 9.
3. Press and hold the
MODE key for one second, until two beeps are heard. The Select
Units screen displays the selected units.
4. To select the units, press either the
START/STOP key for cubic feet or the MODE key
for liters.
5. Press and hold the
MODE key for one second to save the new setting and return to
the Main screen. The units selected appear on the right column of the Main screen.
Figure 9 Select Units screen
4.5.3.2 Concentration or totalize mode
In the Totalize or Concentration modes, the volume size is selected.
To select the volume size in concentration or totalize mode:
1. Make sure the unit is set to Concentration or Totalize mode as described in
section 4.5.2 on page 18.
2. Press the
volume setting.
3. Press and hold the
Volume screen displays the selected volume as shown in Figure 10 on page 20.
4. To review the sample volume settings available, press either the
the
MODE key twice on the Main screen. The bottom line displays the current
MODE key. Refer to Table 3 for sample volume choices.
MODE key for one second, until two beeps are heard. The Select
START/STOP key or
19
Operation
Table 3 Sample volume settings
VolumeTime for sample (MM:SS)
0.01 cubic feet00:06
28.3 cubic liters10:00
10.0 cubic liters03:53
2.83 cubic liters01:00
1.0 cubic liter00:21
1.0 cubic foot10:00
0.10 cubic foot01:00
5. Once the required volume choice is displayed, press and hold the MODE key for one
second to save the new setting and return to the Main screen. The volume selected
is displayed in the right column of the Main screen.
4.5.4 Set the contrast
To set the contrast:
1. Press the
2. The bottom line displays the current contrast.
3. Press and hold the
4. Press the
5. Once the required contrast is achieved, press and hold the
Figure 10 Select volume screen
MODE key five times on the Main screen.
MODE key for one second, until two beeps are heard. The
Program Contrast screen is displayed (refer to Figure 11).
START/STOP key to increase the contrast. Press and hold the MODE key to
decrease the contrast.
MODE key for one second
to save the new setting and return to the Main screen.
20
4.6Counter operation
Once the MET ONE HHPC–2 unit configuration is set:
•Use the zero count to verify the operation of the filter.
Operation
Figure 11 Program contrast screen
•Collect air samples, transfer them to a computer for further analysis and print them.
•Review the number of samples in the print buffer and, if needed, clear the print buffer.
4.6.1 Verify the MET ONE HHPC–2 with a zero count filter
Make sure that the instrument is not falsely counting due to sensor leakage, internal or
external electrical interference or any other reason before the unit is used for monitoring
or verifying clean-room condition.
Clean the MET ONE HHPC–2 using the zero count filter before storage or shipment.
Purge the MET ONE HHPC–2 with a zero count filter before using it in a clean-room or a
clean manufacturing environment. This process meets the Japanese Industrial Standard
B9221-1997 (JIS).
To verify the unit with a zero count filter:
1. Remove the isokinetic probe from the unit.
2. Attach the intake hose barb (MP000026-01).
3. Secure the high purity tubing (VP 792002) to the hose barb.
4. Secure the opposite end of the tubing to the zero count filter (VP212808). Make sure
the arrow on the filter is pointing away from the tubing. Now the unit is ready to purge
and zero count.
5. Operate the unit for approximately 15 minutes.
6. Verify that the data on the display is:
•Not more than one particle greater than 0.3
•Not more than one particle per 0.5 cubic foot.
Note: Initial counts are typically from particles shedding from components used for zero
counting. The zero count process or guidelines set forth by applicable industry standard(s) is
continued.
7. Remove the zero count filter and attach the isokinetic probe to resume normal
operations.
µm in five minutes on average, or
21
Operation
4.7Collect samples
To collect samples:
1. Place the MET ONE HHPC–2 unit upright and remove the intake valve cover (the
black plastic cap) from the intake barb or the isokinetic probe.
2. Press the
3. To review each setting, press the
bottom line of the Main screen. Continue to press the
are reviewed.
4. Press the
POWER key to turn on the unit.
START/STOP key to start sampling on the Main screen.
4.8Transfer the sample data to the PC
Refer to section 3. on page 14 to install the MET ONE HHPC–2 Utility software.
To transfer sample data from MET ONE HHPC–2 to a PC:
1. Use the MET ONE HHPC–2 Utility software to download particle count data to a PC
in comma separated variable (CSV) file format. The particle count data can be
viewed in Microsoft
2. All parameters required by the download utility are stored in the initialization file that
is also on the diskette.
3. Double click the CSV_Download_Utility.exe to start the application.
®
Excel® and other spreadsheet software applications.
4.9Print samples
Print samples at the end of each run or print each sample as it is taken.
Note: Not all printers are compatible with the MET ONE HHPC–2. Make sure that the portable
printer is configured correctly to function with the MET ONE HHPC–2.
MODE key to display the current setting on the
MODE key until all the settings
To print the downloaded data from the PC:
1. Connect the MET ONE HHPC–2 to a serial printer using the RS232 cable (Catalog
No. VP894408) and the DB9 to the RJ45 convertor (Catalog No. SA000091-01)
supplied with the printer (Catalog No. SA000094-01).
2. Press
3. Press and hold the
4. To download the buffer records, press the
5. Select one of the following options:
6. Press
MODE three times on the Main screen of the particle counter. The bottom line
displays the number of records in the buffer.
MODE key for one second until two beeps are heard. The Buffer
Options screen displays the buffer count.
MODE key. The second Buffer Options
screen displays the download options.
•
MODE key—prints formatted data
•
START/STOP key—prints unformatted data
The Print Buffer screen is displayed.
START/STOP to print the records. The data from the sample run is printed.
22
4.10 Storing samples
The MET ONE HHPC–2 stores each measurement in a 100-record rotating buffer. Data
is stored in a ‘first in first out’ order. Thus, when record 101 is saved, the first record is
automatically deleted, leaving a total of 100 records.
4.10.1 Determine the number of samples
Press the MODE key three times on the Main screen. The bottom line displays the
number of records in the Buff Count field (refer to Figure 12).
Operation
4.10.2 Display buffer records
To display the buffer records:
1. Press the
2. Press and hold the
recent record in the buffer is displayed (refer to Figure 13).
3. To review the buffer records, press either the
•
MODE key—displays the first record
•
START/STOP key—scrolls backwards through the records from the most recent to
the oldest record
4. Press and hold the
and return to the Main screen.
Figure 12 Buffer count screen
MODE key four times on the Main screen.
MODE key for one second until two beeps are heard. The most
MODE key or the START/STOP key:
MODE key for one second to exit the Stored Buffer Record screen
Figure 13 Stored buffer record screen
23
Operation
4.10.3 Clear the buffer
To clear the buffer:
1. Press the
MODE key three times on the Main screen. The bottom line displays the
number of records in the Buff Count field (refer to Figure 14).
2. Press and hold the
MODE key for one second, until two beeps are heard.The Buffer
count will be displayed on the Buffer Options screen.
3. Press the
START/STOP key to delete the buffer records. The Delete Buffer screen is
displayed (refer to Figure 15 on page 24).
4. Press the
START/STOP key to confirm the deletion. The Main screen displays with the
buffer count shown at the bottom line.
Note: The buffer count should be equal to zero.
Figure 14 Buffer options screen
Figure 15 Delete Buffer screen
24
Section 5Maintenance
DANGER
Only qualified personnel should conduct the tasks described in this section of the
manual. MET ONE
5.1Battery maintenance
The battery status light indicates the relative battery charge status, described in Table 4.
The power supply provided with the instrument serves as battery charger and a direct
power source. Connect the power supply for continuous and long-term sampling.
Table 4 Battery status indicator
LightCharge mode
Steady greenFast charge
Flashing greenMaintenance charging mode
RedBattery is defective and needs to be replaced
No lightBattery is fully charged
To charge the battery:
1. Connect the external power supply (refer to section 3.3.4 on page 14).
2. Turn on the MET ONE HHPC–2.
3. Remove the power supply when no light is seen.
5.1.1 Charge a drained battery
To charge a drained battery:
1. Attach the external power supply to the MET ONE HHPC–2 unit.
2. Wait for five minutes. Turn on the MET ONE HHPC–2.
3. Leave the unit attached to the power supply until the battery charging indicator light
turns off.
5.2Calibrate the MET ONE HHPC–2
Return the MET ONE HHPC–2 to the manufacturer annually for recalibration to comply
with JIS B9921–1989 standards. The calibration date is displayed on the manufacturer
decal on the back of the MET ONE HHPC–2. Refer to Section 8 on page 31 for MET
ONE HHPC–2
return procedures.
25
Maintenance
26
Section 6Troubleshooting
6.1Problems and recommendations
Table 5 lists the problems, reasons and the recommended actions that occur while using
MET ONE HHPC–2.
Table 5 Troubleshooting procedures
SymptomCauseSolution
MET ONE HHPC–2 does not
zero count.
The MET ONE HHPC–2 does not
turn on when the POWER key is
pressed.
MET ONE HHPC–2 unit does not
charge.
Unable to download data in the
buffer.
Unable to print data in the buffer.
MET ONE HHPC–2 display
shows the message, “CAL ALM.”
Zero count filter or
hose barb is not
properly attached to
the unit.
Internal air leak.Contact the customer service center for repair.
Cell contamination by
excessive dust
exposure.
The battery voltage too
low.
Connections to MET
ONE HHPC–2 unit is
proper.
No records in bufferMake sure there are records in the buffer.
Communication
parameter are not set.
The connections are
not proper
No records in bufferMake sure that there are records in the buffer.
Communication
parameter are not set
The connections are
not proper
The cell is
contaminated.
Attach the filter or hose barb.
Return the unit to a factory authorized service center.
Connect external power supply to MET ONE HHPC–2 for
operation and charging. Then, turn on the power again.
Disconnect the power jack from the unit and the power cord
from the AC power source (wall receptacle). Disconnect the
cord/plug from the power supply and reconnect the cord/plug to
the power supply firmly. Restore all connections to the unit and
power source. If the problem persists, contact the customer
service center for repair.
Make sure that the communication parameters of the computer
are set to 9600 baud, 8 data bits, no parity, 1 stop bit.
Check the connections between the MET ONE HHPC–2 and
the computer.
Make sure that the communication parameters of the printer are
set to 9600 baud, 8 data bits, no parity, 1 stop bit.
Make sure that the connections between the MET ONE
HHPC–2 and the printer.
Connect the zero count filter and program the MET ONE
HHPC–2 to run for at least five minutes.
If the problem persists, contact the customer service center for
repair.
27
Troubleshooting
28
Section 7Replacement Parts and Accessories
7.1Accessories
DescriptionQuantityCatalog no.
Intake nozzle, 1/8” hose barb1MP000026–01
AC/DC power supply1VP624005
US power cord1VP23501
RJ45 to DB9 adapter1SA000070–01
RS232 modular cable1VP894408
MET ONE HHPC–2 Utility software1CS00025
6¼” high purity tubing1VP792002
7.2Optional accessories
Figure 16 Optional accessories
1Carrying case (SA000089–01)4Isokinetic probe, hose barb (MP000120–01)
2Vinyl carrying case (SA000249–01)5Serial thermal printer (SA000094–01)
3RS232 serial interface extension cable (EP096010)6European power cord (VP625300)
29
Replacement Parts and Accessories
30
Section 9Limited Warranty
Hach Ultra warrants this instrument to be free of defects in materials and workmanship for a period of
one year from the shipping date. If any instrument covered under this warranty proves defective during
this period, Hach Ultra will, at its option either repair the defective part without charge for extra parts and
labor or provide an equivalent replacement in exchange for the defective product.
If any diode covered under this warranty proves defective during this period, Hach Ultra will, at its option,
either repair the defective diode without charge for parts and labor or provide an equivalent replacement
in exchange for the defective product.
To obtain service under this warranty, the customer must notify the nearest Hach Ultra service support
center on or before the expiration of the warranty period and follow their instructions for return of the
defective instrument. The customer is responsible for all costs associated with packaging and
transporting the defective unit to the service support center and must prepay all shipping charges. Hach
Ultra will pay for return shipping if the shipment is to a location within the same country as the service
support center.
This warranty shall not apply to any defect, failure or damage caused by improper use or maintenance or
by inadequate maintenance or care. This warranty shall not apply to damage resulting from attempts by
personnel other than Hach Ultra representatives or factory authorized and trained personnel, to install,
repair or service the instrument; to damage resulting from improper use or connection to incompatible
equipment; or to instruments that have been modified or integrated with other products when the effect
of such modification or integration materially increases the time or difficulty of servicing the instrument.
THIS WARRANTY IS GIVEN BY HACH ULTRA ANALYTICS WITH RESPECT TO THIS INSTRUMENT
IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED. HACH ULTRA ANALYTICS
AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS
FOR A PARTICULAR NON-CONTRACTUAL PURPOSE. HACH ULTRA ANALYTICS’
RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND
EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY.
HACH ULTRA ANALYTICS AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT,
SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES EVEN IF HACH ULTRA ANALYTICS OR
ITS VENDORS HAS BEEN GIVEN ADVANCED NOTICE OF THE POSSIBILITY OF SUCH
DAMAGES.
32
Section 10Certification
33
Certification
34
Appendix A Data Communications
A.1 Remote serial interface (I/O Port)
The MET ONE HHPC–2 comes with an RS232 communications port, an RS232 cable
and an RJ45 connector. The RS232 communications port is used to communicate with a
computer to download data and for remote operation.
The specifications and wire pin-outs for the I/O port are:
•Baud rate: 9600 baud
•Data bits: 8 bits
•Parity: None
•Stop bits: 1 bit
A.1.1 DB9 serial connection
If the computer or terminal has a 9-pin serial port with a female connector, use the
modular RJ45 to DB9 connector provided along with the RS232 cable. The MET ONE
HHPC–2 is configured as a Data Communication Equipment (DCE), so the use of a
null-modem cable is not required. The cable pin assignments are shown in Figure 17.
Figure 17 Pin layout for the RJ45 to DB9 connector cable
zero count filter........................................................ 13
37
Electrical equipment marked with this symbol may not be disposed of in European
public disposal systems after 12 August of 2005. In conformity with European local
and national regulations (EU Directive 2002/96/EC), European electrical equipment
users must now return old or end-of life equipment to the Producer for disposal at no
charge to the user. Note: For return for recycling, please contact the equipment
producer or supplier for instructions on how to return end-of-life equipment for proper
disposal. Important document. Retain with product records.
GERMAN Elektrogeräte, die mit diesem Symbol gekennzeichnet sind, dürfen in Europa nach
dem 12. August 2005 nicht mehr über die öffentliche Abfallentsorgung entsorgt werden. In
Übereinstimmung mit lokalen und nationalen europäischen Bestimmungen (EU-Richtlinie
2002/96/EC), müssen Benutzer von Elektrogeräten in Europa ab diesem Zeitpunkt alte bzw. zu
verschrottende Geräte zur Entsorgung kostenfrei an den Hersteller zurückgeben. Hinweis:
Bitte wenden Sie sich an den Hersteller bzw. an den Händler, von dem Sie das Gerät bezogen
haben, um Informationen zur Rückgabe des Altgeräts zur ordnungsgemäßen Entsorgung zu
erhalten. Wichtige Informationen. Bitte zusammen mit den Produktinformationen aufbewahren.
FRENCH A partir du 12 août 2005, il est interdit de mettre au rebut le matériel électrique
marqué de ce symbole par les voies habituelles de déchetterie publique. Conformément à la
réglementation européenne (directive UE 2002/96/EC), les utilisateurs de matériel électrique
en Europe doivent désormais retourner le matériel usé ou périmé au fabricant pour élimination,
sans frais pour l’utilisateur. Remarque : Veuillez vous adresser au fabricant ou au fournisseur
du matériel pour les instructions de retour du matériel usé ou périmé aux fins d’élimination
conforme. Ce document est important. Conservez-le dans le dossier du produit.
ITALIAN Le apparecchiature elettriche con apposto questo simbolo non possono essere
smaltite nelle discariche pubbliche europee successivamente al 12 agosto 2005. In conformità
alle normative europee locali e nazionali (Direttiva UE 2002/96/EC), gli utilizzatori europei di
apparecchiature elettriche devono restituire al produttore le apparecchiature vecchie o a fine
vita per lo smaltimento senza alcun costo a carico dell’utilizzatore. Nota: Per conoscere le
modalità di restituzione delle apparecchiature a fine vita da riciclare, contattare il produttore o il
fornitore dell’apparecchiatura per un corretto smaltimento. Documento importante. Conservare
con la documentazione del prodotto.
DANISH Elektriske apparater, der er mærket med dette symbol, må ikke bortskaffes i
europæiske offentlige affaldssystemer efter den 12. august 2005. I henhold til europæiske
lokale og nationale regler (EU-direktiv 2002/96/EF) skal europæiske brugere af elektriske
apparater nu returnere gamle eller udtjente apparater til producenten med henblik på
bortskaffelse uden omkostninger for brugeren. Bemærk: I forbindelse med returnering til
genbrug skal du kontakte producenten eller leverandøren af apparatet for at få instruktioner
om, hvordan udtjente apparater bortskaffes korrekt. Vigtigt dokument. Opbevares sammen
med produktdokumenterne.
Form No 011229
SWEDISH Elektronikutrustning som är märkt med denna symbol kanske inte kan lämnas in på
europeiska offentliga sopstationer efter 2005-08-12. Enligt europeiska lokala och nationella
föreskrifter (EU-direktiv 2002/96/EC) måste användare av elektronikutrustning i Europa nu
återlämna gammal eller utrangerad utrustning till tillverkaren för kassering utan kostnad för
användaren. Obs! Om du ska återlämna utrustning för återvinning ska du kontakta tillverkaren
av utrustningen eller återförsäljaren för att få anvisningar om hur du återlämnar kasserad
utrustning för att den ska bortskaffas på rätt sätt. Viktigt dokument. Spara tillsammans med
dina produktbeskrivningar.
SPANISH A partir del 12 de agosto de 2005, los equipos eléctricos que lleven este símbolo no
deberán ser desechados en los puntos limpios europeos. De conformidad con las normativas
europeas locales y nacionales (Directiva de la UE 2002/96/EC), a partir de esa fecha, los
usuarios europeos de equipos eléctricos deberán devolver los equipos usados u obsoletos al
fabricante de los mismos para su reciclado, sin coste alguno para el usuario. Nota: Sírvase
ponerse en contacto con el fabricante o proveedor de los equipos para solicitar instrucciones
sobre cómo devolver los equipos obsoletos para su correcto reciclado. Documento importante.
Guardar junto con los registros de los equipos.
DUTCH Elektrische apparatuur die is voorzien van dit symbool mag na 12 augustus 2005 niet
meer worden afgevoerd naar Europese openbare afvalsystemen. Conform Europese lokale en
nationale wetgegeving (EU-richtlijn 2002/96/EC) dienen gebruikers van elektrische apparaten
voortaan hun oude of afgedankte apparatuur kosteloos voor recycling of vernietiging naar de
producent terug te brengen. Nota: Als u apparatuur voor recycling terugbrengt, moet u contact
opnemen met de producent of leverancier voor instructies voor het terugbrengen van de
afgedankte apparatuur voor een juiste verwerking. Belangrijk document. Bewaar het bij de
productpapieren.
POLISH Sprzęt elektryczny oznaczony takim symbolem nie może byćlikwidowany w
europejskich systemach utylizacji po dniu 12 sierpnia 2005. Zgodnie z europejskimi, lokalnymi i
państwowymi przepisami prawa (Dyrektywa Unii Europejskiej 2002/96/EC), użytkownicy
sprzętu elektrycznego w Europie muszą obecnie przekazywać Producentowi stary sprzęt lub
sprzęt po okresie użytkowania do bezpłatnej utylizacji. Uwaga: Aby przekazać sprzęt do
recyklingu, należy zwrócić siędo producenta lub dostawcy sprzętu w celu uzyskania instrukcji
dotyczących procedur przekazywania do utylizacji sprzętu po okresie użytkowania. Ważny
dokument. Zachować z dokumentacją produktu.
PORTUGESE Qualquer equipamento eléctrico que ostente este símbolo não poderá ser
eliminado através dos sistemas públicos europeus de tratamento de resíduos sólidos a partir
de 12 de Agosto de 2005. De acordo com as normas locais e europeias (Directiva Europeia
2002/96/EC), os utilizadores europeus de equipamentos eléctricos deverão agora devolver os
seus equipamentos velhos ou em fim de vida ao produtor para o respectivo tratamento sem
quaisquer custos para o utilizador. Nota: No que toca à devolução para reciclagem, por favor,
contacte o produtor ou fornecedor do equipamento para instruções de devolução de
equipamento em fim de vida para a sua correcta eliminação. Documento importante.
Mantenha junto dos registos do produto.
Form No 011229
Trademarks are property of their respective owners.
Specifications subject to change without notice.
Thank you for reading this data sheet.
Please contact us for a quotation:
Head Office (UK)
P.O. Box 2124
Chelmsford
Essex CM1 3UP
England
Tel: +44 (0)1245 600 560
Fax: +44 (0)1245 600 030
Email:
General Information/Sales:
info@keison.com
Accounts Department:
accounts@keison.com
USA Office
233 Rogue River Highway #425
Grants Pass
Oregon 97527-5429
USA
Switzerland Office
Zugerstrasse 46
6314 Unterägeri
Zug
Switzerland
Tel: +41 41 541 5877
Fax: +41 41 541 5878
Email:
General Information/Sales:
info@kesn.ch
Accounts Department:
accounts@kesn.ch
Latin America Office
World Trade Centre
Apartado 0832-00708
Ciudad de Panamá
República de Panamá
Tel: +1 503 206 0941
Fax: +1 541 610 1646
Email:
General Information/Sales:
info@keison.com
Accounts Department:
accounts@keison.com
Tel: +507 836 5304
Fax: +507 399 1804
Email:
General Information/Sales:
ventas@keison.com
Accounts Department:
cuentas@keison.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.