Hach-Lange TITRALAB Short-form Reminders User Manual

Page 1
ShorShor
Shor
ShorShor
Electrodes - Install/Replace
Select method/sequence
pH electrode calibration
Reagent calibration
Running methods
Running sequences
Direct measurements
Operations during analyses
Manual dosing
Maintenance recommendations
Storage recommendations
t-Fort-For
t-For
t-Fort-For
m Remindersm Reminders
m Reminders
m Remindersm Reminders
Spare parts & accessories
Electrodes
Calibration standards & solutions
D21T047 • Printed by Radiometer Analytical SAS • 2008-09H
Short-Form Reminders 1
Page 2
Short-Form Reminders 2
Page 3
Only one key press to
check the electrodes in use!
Select method or sequence

Electrodes - Install/Replace

Form no. 1
3
XYZ
1.
and
Enter Electrode window
2.
VW
2
then
1
STU
3. Press enter serial number
For a method
Install electrode system
4. Connect electrode
Socket Electrode
REF Single reference electrode TEMP Temperature sensor
✓✓
to
✓✓
GND Single metal electrode Pt-Pt Double metal electrode E1/E2 Other electrode types EC Conductivity cells
Example: pHC2001 to address TIM/E1 (socket E1 on rear panel).
and
VW
2
For a sequence
EC
Pt-Pt Ref E1E2
empGND
T
5.
STU
1
Confirm
Page 4
Replace with electrode of same type and ID
1.
2.
3.
1.
2.
VW
VW
Enter Electrode window
2
then
3
XYZ
4. Disconnect old electrode and confirm
1
STU
5.
Connect new electrode
✓✓
Press
to
✓✓
6. Enter serial no. and confirm
select electrode
Disconnect electrode system
Enter Electrode window
2
then
2
VW
3. Disconnect electrode and press
✓✓
to confirm
✓✓
1.
2.
XYZ
Check electrode system
Enter Electrode window
3.
Press select
✓✓
to
✓✓
electrode
3
To view parameters
4.
Esc
To return to the Electrode window
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 5
To replace empty reagent bottles,
use the bottle exchange procedure

Reagents - Install/Replace

Select method or sequence
Form no. 2
and
XYZ
3
For a method
and
VW
2
For a sequence
Install reagent system
1.
Enter Reagent window
3.
✓✓
Press
to
✓✓
enter batch no., expiry
2.
VW
2
then
1
STU
4.
date then confirm
If required, rinse burette
5. Install reagent bottle and
✓✓
press
to flush
✓✓
Remove a reagent
Run the Rinse procedure in the Burette functions screen. This procedure is required when you no longer wish to use the reagent or replace it by another one.
1.
Enter Reagent window
2.
ABC
7
then
4
JKL
Page 6
Check reagent system
1.
2.
1.
2.
3.
XYZ
VW
select
✓✓
to
✓✓
Enter Reagent window
3. Press
reagent
3
To view parameters
4.
Esc
To return to the reagent window
Bottle exchange: with one of same type and ID
Enter Reagent window
2
then
VW
2
4.
1
STU
5. Enter batch no. and expiry date of
Confirm
new reagent and confirm
6. Remove bottle
✓✓
Press
to
select
✓✓
7. Install new bottle and press flush and fill
✓✓
to
✓✓
reagent
Replace reagent: with different type and ID
1.
Enter Reagent window
2.
VW
2
then
1
STU
3. Enter batch no. and expiry date
4.
Confirm replace
5.
6. Remove old, install new reagent bottle, press
✓✓
to flush
✓✓
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 7
Straightforward
method selection

Select method/sequence

Form no. 3
Select method
1.
Enter main menu
2.
3.
4.
XYZ
Select Method mode
3
Select method from list
Confirm
Page 8
Select sequence
1.
Enter main menu
2.
3.
4.
5.
2
VW
Esc
Select Sequence mode
then
Enter Sequence ID confirm
Check icons
You can run the method/sequence
You can run the method/sequence. However a calibration and/or a reagent bottle exchange must be performed soon
Check the electrodes and reagents for installation or calibration purposes
Programming error due to missing parameters
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 9
Calibrate your electrode

pH electrode calibration

Select electrode
Form no. 4
1.
2.
3.
1.
Enter Electrode window
Confirm
STU
1
then
to select electrode from list
Run
1
STU
To run the calibration
2.
Enter User ID and temperature, if required, and confirm
1
STU
Page 10
3. Rinse electrode(s) and dip in beaker
5. The measurement starts
4. Confirm
2
VW
1
STU
Press 1 to save results
then
2
VW
Calibration can be performed on up to 5 buffers.
Automatic result display
Results: zero pH and sensitivity
View statistics
More cycles/End of analysis
Stop
1
STU
“Resume analysis”
Emergency stop
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 11
Calibrate your reagent
Select reagent

Reagent calibration

Form no. 5
1.
2.
3.
1.
Enter Reagent window
Confirm
STU
1
then
to select reagent from list
Run
1
STU
To run the calibration
2.
Enter User ID and temperature, if required, and confirm
1
STU
Page 12
3.
Enter Standard batch number and standard amount then confirm
4. Rinse electrode(s) and dip in beaker
1
STU
5. Confirm
2
VW
1
STU
Press 1 to save result
then
2
VW
1
STU
Automatic result display
View statistics
More tests/End of analysis
Stop
1
STU
“Resume analysis”
Emergency stop
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 13
Run your analyses
Run

Running methods

Form no. 6
1.
2.
3.
To run your selected method
1
STU
Enter User ID and temperature, if required, and confirm
1
STU
Enter Sample ID, sample amount and confirm
1
STU
4. Rinse electrode(s) and dip electrode(s) in beaker then confirm
1
STU
Page 14
Press 1 to save results
Automatic result display
View statistics
VW
Stop
2
then
2
VW
More tests/End of analysis
Emergency stop
1
“Resume analysis”
STU
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 15
Use your Sample stack
for daily analyses
Preparing the sample stack

Running sequences

Form no. 7
1.
2.
Enter a new sample ID and confirm
3.
Check the sample test quantity or enter the sample test quantity to place in beaker no.1
Confirm4.
2
VW
1
STU
+
Locate beaker no.1
5.
6.
Repeat steps 2 to 4 for each beaker in the sequence
7.
Esc
then
Esc
+
Page 16
Run
1.
2.
Press 1 to save results
2
then
VW
2
VW
1
STU
3. Rinse the electrodes, and dip in beaker 1, confirm
Enter User ID and temperature, if required, confirm
1
STU
Automatic result display (manual)
View statistics (manual)
1
STU
Emergency stop
“Resume analysis”
Stop
1
STU
SAC Sequence
For automatic sample handling, using a sample changer, proceed as follows:
1.
2.
1
STU
Select SAC Sequence
Run Sequence
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 17
A “?” in the Electrode tab?
Check your programming
1.
Connect the electrode(s). See form no. 1
2. Dip electrode(s) in beaker

Direct measurements

Run
3.
Live measurements
To select a connected electrode and display measurements
5
To display measurements
MNO
Form no. 8
1
STU
To start/stop stirring
Esc
To stop measurements
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 18
Page 19
Perform other operations while
your analysis is running

Operations during analyses

Instructions
Form no. 9
1.
Esc
2. Perform the necessary operations
Run in progress
1
then
Esc
STU
then
2
VW
Stirring in progress
In Routine mode, you can:
• View the method, reagent, electrode parameters,
• View the GLP-archives,
• Change the stirring speed in the Cell window.
In Supervisor mode, you can additionally:
• Create a method, a reagent, and an electrode.
In pH-Stat mode, you can:
• Access certain regulation parameters
To enter the Edit mode
You can change the parameters of the active method/sequence. These changes will be taken into account for the next run.
To return to the analysis
Page 20
1.
User action required during analyses
To return to analysis window
2.
1.
2.
Esc
Stop
Esc
To return to edition window
Emergency stop
then
then
2
VW
1
STU
To return to the analysis window
To resume analysis, if desired. Calculations are not performed
3.
Del
To end an analysis and get results.
Calculations are performed, if result found.
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 21
Deliver a preset volume of
reagent and follow
Volume = f(time) or
measured signal = f(volume)
Running the manual dosing
1. Select method. See form no. 3.

Manual dosing

Form no. 10
2.
3.
Enter Reagent menu then .
Select reagent and electrode.
Enter increment size and addition speed.
Enter the maximum volume to dose.
Define the type of measurement to perform.
Confirm .
5
MNO
1
STU
Manual dosing with measurements
Adjust size of increments.
Manual or automatic addition of increments.
Volume dispensed and live measurements.
Direct curve Measured signal = f(dispensed volume ).
1
To add one increment or
STU
start/stop adding increments automatically.
Page 22
Manual dosing with no measurement
Adjust size of increments.
Manual or automatic addition of increments.
Volume dispensed.
Direct curve Dispensed volume (in ml) = f(time).
To add one increment or start/stop adding increments
1
STU
automatically.
Displaying direct or derivative curve - Identifying curve points
Del
Stop
Del
or
1
then .
STU
Direct or derivative curve.
Coordinates of that point of the curve.
To move along the curve.
1
or
STU
Stop
To quit.
Quit manual dosing
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 23
For burette air bubble removal,
run the Flush procedure

Maintenance recommendations

Form no. 11
Daily
Check and maintain electrodes according to manufacturer recommendations.
Check that there is no crystallisation at piston level, if so rinse the burette using appropriate solvent to remove crystals. If not sufficient, dismount burette to remove crystals.
Check that there is no crystallisation at addition tip end that may block the reagent to pour freely, if so use appropriate solvent to remove crystals.
Check the level of reagent in the reagent bottle
Empty waste bottle if installed.
Remove air bubbles from the burette and tubing by running a “Flush” procedure from the burette menu.
Weekly
glass
Reagent
cylinder
piston
Reagent
Flush the burette with appropriate solvent to ensure complete cleaning of burette and stopcock.
Change burette
stand
Check the piston for leakages.
The burette stand must be changed when at least 2 mm of liquid can be seen between the two sealing lips. Refer to diagram. Having 1 mm or less liquid between the two sealing lips does not require to change the burette stand.
Page 24
Annualy
Replace burette stand, 1, 5, 10, 25, 50 ml and the anti-diffusion tip, part. no.956-309.
Frequent use or use with aggressive or concentrated reagents occurring crystallisation risks may severely reduce the burette and anti-diffusion tip lifetime and may consequently require to replace them more frequently.
Replace burette stand set of 3 tubing, (stands 5, 10, 25, 50 ml, part number X31T115 and stand 1 ml, part number X31T116).
Replace damaged burette accessories: bottle stopper, desiccant tubing.
Contact your Radiometer Analytical representative for instrument revision and cleaning.
Recommendations of use
After installing new burette and/or tubing, before use, check that all tubing fittings are sufficiently tightened to prevent leakage, and that no tubings are crimped (do not over tight fittings)
The reagent should be at room temperature before use in the burette.
Do not leave burette installed with concentrated base, acid or salt solutions. Always rinse after use, at least daily, to increase stopcock and burette lifetime.
Organic solvents: the organic solvents given as reacting with and/or destroying the VITON® (Fluorinated polymer) should not be used with our burette stands (example: TetraHydroFurane). The solvents given as simply making inflate the VITON® are allowed. Please contact us if you wish to work with solvents destroying the VITON®.
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 25
Keep your system in
tiptop condition!

Storage recommendations

Form no. 12
Daily
Rinse electrodes and delivery tip throughly with distilled water after use.
Store the electrode(s) following the instructions given in the respective operating instructions.
Overnight and up to 1 week
In case of spillages, clean the exterior surface of the instrument with a soft cloth and tepid water. Avoid the accumulation of excess water around the keyboard area.
Long-term 1 week or more
Run the “Rinse” procedure from the Burette menu to remove reagent from burette and tubings.
If the system is to be left in the same location, store electrode(s) following the long-term storage instructions given in the respective operating instructions.
If the system is to be transported, dismount the burette stand, metal rod, and cell accessories before packing.
For transportation, always use the packaging supplied by the manufacturer.
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 26
Page 27
Ordering information
at a glance

Spare parts & accessories

Form no. 13
Part no. Type Description
Burette stands
X51T025 B501 Burette Stand, 1 ml, UV protected with connection tubing X51T026 B505 Burette Stand, 5 ml, UV protected with connection tubing X51T017 B510 Burette Stand, 10 ml, UV protected with connection tubing X51T027 B525 Burette Stand, 25 ml, UV protected with connection tubing X51T028 B550 Burette Stand, 50 ml, UV protected with connection tubing
Burette maintenance parts
956-309 Anti-diffusion tip PTFE X31T115 Maintenance Kit of 3 tubes for B5xx & B6xx (excl, 1 ml burette)
X31T116 Maintenance Kit of 3 tubes for B501 & B601only X51T010 Delivery tip, glass, burette >5 ml
X51T011 Delivery tip, glass, burette <5 ml X31T070 Titrant bottle stopper GL45 assembly with desiccant & suction tubing
X31T110 Titrant bottle stopper GL45 assembly with desiccant & suction tubing
for B501
Polypropylene titration vessels
904-490 Titration Vessel, PP, 40-100 ml (pack of 50 pcs)
904-489 Titration Vessel, PP, 22-45 ml (pack of 50 pcs)
904-488 Titration Vessel, PP, 8-25 ml (pack of 50 pcs)
904-487 Titration Vessel, PP, 1-9 ml (pack of 50 pcs)
Page 28
Part no. Type Description
Notebook keyboard
X31T108 Notebook Keyboard Mask X16T005 Notebook Keyboard 83-keys, PS/2 port, US version X16T006 Notebook Keyboard 83-keys, PS/2 port, French
Balance connecting cables
A95Z201 C201Z 9pf/1m/25pm (Mettler PE+option 016; Mettler AE+option 012 or 011) A95Z205 C205Z 9pf/2m/25pm (Mettler PE+option 016; Mettler AE+option 012 or 011) A95Z206 C206Z 1m serial cable for Mettler AM, PM, SM, AT, MT, UMT, PJ, AJ A95Z202 C202Z 9bf/1m/25bm (Sartorius MasterPro, QC, MP8-4) A95Z204 C204Z 9bf/2m/25bm (Sartorius MasterPro, QC, MP8-4) A95Z207 C207Z 9pf/1.20m/25pm (Precisa, old models) A95Z208 C208Z 9pf/1m/RJ45 (Precisa, XB series) For other balances, please consult.
KF consumables
X31T048 KF Pump Tubing assembly, cell - bottle solvent X31T061 KF Pump Tubing assembly, cell - bottle waste X31T112 Desiccant Tube with tubing, Bottle - KF pump for bottle stopper GL45 &
S40 (for use with new TIM550 casing) X31T051 Syringe, plastic, 5 ml (10 pcs) X41T006 Needle, 120 mm, KF cell (10 pcs) X31T038 Silicone Septa for X31T036, set of 10 X31T165 Sample Container with funnel, 5pcs S72T004 Molecular Sieve, 250 g
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 29
Ordering information
at a glance

Electrodes

Form no. 14
Part no. Type Description
Combined pH/Reference electrodes – Acid/Base titrations
E16M313 pHC2001-8 Combined pH electrode, Red Rod, porous pin junction E16M400 pHC2401-8 Combined pH electrode, Red Rod, annular junction E16M500 pHC2005-8 Combined pH electrode, Red Rod, epoxy body protection E16M501 pHC2085-8 Combined pH electrode, Red Rod, with temperature probe E16M300 pHC3001-8 Combined pH electrode, Ag/AgCl, porous pin junction E16M302 pHC3005-8 Combined pH electrode, Ag/AgCl, epoxy body protection E16M305 pHC3081-8 Combined pH electrode, Ag/AgCl, with temperature probe
Single pH electrodes
E11M006 pHG201-8 Glass pH electrode, ø = 7.5 mm, general purpose E11M003 pHG211-8 Glass pH electrode, ø = 7.5 mm, alkaline samples
Combined Metal/Reference electrodes – RedOx & Precipitation titrations
E31M003 MC3051Pt-9 Combined Platinum electrode, Ag/AgCl Ref., Screw cap E34M004 MC6091Ag-9 Combined Silver electrode, Hg/Hg2SO4 Ref., Screw cap
*
For use with connection cable, type CL114, part no. A94L114
*
*
Single Metal electrodes
E34M003 M295Ag Silver Rod, ø = 7.5 mm, PTFE body, banana plug E31M001 M241Pt Platinum Plate, ø = 7.5 mm, glass body, banana plug
**
For use with adapter, type BNC-m/Banana-f, part no. A94P807
**
Temperature Sensor
E51M001 T201 Temperature Sensor, glass
**
Page 30
Part no. Type Description
Pt-Pt Metal electrodes – Karl Fischer titrations
E32M001 M231Pt2 Double platinum wire, 2 banana plugs E32M002 M241Pt2-8 Double platinum staples, short cable
***
For use with adapter, type BNC-f/2 Banana, part no. A94P801
***
ISE electrodes – Complexometric & Precipitation titrations
E41M002 ISE25Ca Ion Selective Electrode, Calcium, Screw cap E41M003 ISE25Cl Ion Selective Electrode, Chloride, Screw cap E41M006 ISE25Cu Ion Selective Electrode, Copper, Screw cap E41M007 ISE25F Ion Selective Electrode, Fluoride, Screw cap E41M017 ISEC301F Combined Ion Selective Electrode, Fluoride,
Ag/AgCl ref., Screw cap
*
For use with connection cable, type CL114, part no. A94L114
*
*
*
*
*
Single Reference electrodes
E21M009 REF201 Reference Electrode, Red Rod, ø = 7.5 mm, banana plug E21M011 REF401 Reference Electrode, Calomel, ø = 7.5 mm, banana plug E21M012 REF601 Reference Electrode, Mercurous Sulphate, ø = 7.5 mm, banana plug E21M001 REF251 Reference Electrode, Red Rod, double junction, banana plug
**
For use with adapter, type BNC-m/Banana-f, part no. A94P807
**
**
**
**
Conductivity cells
B15B001 CDC641T Conductivity Cell, 2-pole; platinised; built-in temp.sensor E61M010 CDC566T Conductivity Cell, 4-pole; not platinised; built-in temp.sensor E61M012 CDC741T Conductivity Cell, 2-pole; platinised; built-in temp.sensor; 3m cable E61M013 CDC745-9 Conductivity Cell, 2-pole; platinised; epoxy body, Screw cap E61M015 CDC866T Conductivity Cell, 4-pole; platinised; built-in temp.sensor
***
For use with cable, type CL136, part no. A94L136
***
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 31
Ordering information
at a glance

Calibration standards & solutions

Part no. Type Description
IUPAC Series pH Standards
S11M001 pH 1.679 pH 1.679 ±0.010 at 25°C, 500 ml S11M002 pH 4.005 pH 4.005 ±0.010 at 25°C, 500 ml S11M003 pH 6.865 pH 6.865 ±0.010 at 25°C, 500 ml S11M004 pH 7.000 pH 7.000 ±0.010 at 25°C, 500 ml S11M005 pH 7.413 pH 7.413 ±0.010 at 25°C, 500 ml S11M006 pH 9.180 pH 9.180 ±0.010 at 25°C, 500 ml S11M007 pH 10.012 pH 10.012 ±0.010 at 25°C, 500 ml S11M008 pH 12.45 pH 12.45 ±0.05 at 25°C, 500 ml
4-7-10 Series
S11M012 pH 4 pH 4.00 at 25°C, 500 ml S11M013 pH 7 pH 7.00 at 25°C, 500 ml S11M014 pH 10 pH 10.00 at 25°C, 500 ml
Form no. 15
Conductivity standards
S51M001 1 D KCl 111.3 mS/cm ±0.5 % at 25°C, 500 ml S51M002 0.1 D KCl 12.85 mS/cm ±0.35 % at 25°C, 500 ml S51M003 0.01 D KCl 1408 µS/cm ±0.5 % at 25°C, 500 ml S51M004 0.05 % NaCl 1015 µS/cm ±0.5 % at 25°C, 500 ml C20C250 0.1 M KCl 12.88 mS/cm ±1 % at 25°C, 500 ml C20C270 0.01 MKCl 1413 µS/cm ±1 % at 25°C, 500 ml C20C280 0.001 MKCl 146.9 µS/cm ±2.5 % at 25°C, 500 ml S51M013 NaCl 25 25 µS/cm ±5 % at 25°C, 250 ml
Page 32
Electrode Maintenance Solutions
S16M001 RENOVO•N Normal Cleaning Solution, 250 ml S16M002 RENOVO•X Xtra Strong Cleaning Solution, 250 ml C20C370 KS400 Pepsin in HCl Cleaning Solution, 250 ml C20C380 KS410 Thiourea Solution, 250 ml, for junction cleaning C20C300 KS100 KCl Solution, saturated, 500 ml, for Red Rod & Calomel reference C20C500 KS160 K C20C310 KS120 KCl Solution, saturated with AgCl, 500 ml, for Ag/AgCl reference S21M001 KCl•C KCl Crystals, 15 g S91M001 GK ANNEX Electrode Maintenance Kit for pH & reference electrodes
Solution, saturated, 500 ml, for Hg/Hg2SO4 reference
2SO4
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Loading...