GNDSingle metal electrode
Pt-PtDouble metal electrode
E1/E2Other electrode types
ECConductivity cells
Example: pHC2001 to address TIM/E1 (socket E1 on
rear panel).
and
VW
2
For a
sequence
EC
Pt-PtRefE1E2
empGND
T
5.
STU
1
Confirm
Page 4
Replace with electrode of same type and ID
1.
2.
3.
1.
2.
VW
VW
Enter Electrode
window
2
then
3
XYZ
4. Disconnect old electrode
and confirm
1
STU
5.
Connect
new
electrode
✓✓
Press
✓ to
✓✓
6. Enter serial no. and confirm
select
electrode
Disconnect electrode system
Enter Electrode
window
2
then
2
VW
3. Disconnect electrode
and press
✓✓
✓ to confirm
✓✓
1.
2.
XYZ
Check electrode system
Enter Electrode
window
3.
Press
select
✓✓
✓ to
✓✓
electrode
3
To view
parameters
4.
Esc
To return to the
Electrode window
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 5
To replace empty reagent bottles,
use the bottle exchange procedure
Reagents - Install/Replace
Select method or sequence
Form no. 2
and
XYZ
3
For a
method
and
VW
2
For a
sequence
Install reagent system
1.
Enter Reagent
window
3.
✓✓
Press
✓ to
✓✓
enter batch
no., expiry
2.
VW
2
then
1
STU
4.
date then
confirm
If required,
rinse burette
5. Install reagent bottle and
✓✓
press
✓ to flush
✓✓
Remove a reagent
Run the Rinse procedure in the Burette functions screen. This procedure is required when you no longer wish to
use the reagent or replace it by another one.
1.
Enter Reagent
window
2.
ABC
7
then
4
JKL
Page 6
Check reagent system
1.
2.
1.
2.
3.
XYZ
VW
select
✓✓
✓ to
✓✓
Enter Reagent
window
3.Press
reagent
3
To view
parameters
4.
Esc
To return to the
reagent window
Bottle exchange: with one of same type and ID
Enter Reagent
window
2
then
VW
2
4.
1
STU
5. Enter batch no. and expiry date of
Confirm
new reagent and confirm
6. Remove bottle
✓✓
Press
✓ to
select
✓✓
7. Install new bottle and press
flush and fill
✓✓
✓ to
✓✓
reagent
Replace reagent: with different type and ID
1.
Enter Reagent
window
2.
VW
2
then
1
STU
3. Enter batch no. and expiry date
4.
Confirm
replace
5.
6. Remove old, install new reagent
bottle, press
✓✓
✓ to flush
✓✓
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 7
Straightforward
method selection
Select method/sequence
Form no. 3
Select method
1.
Enter main
menu
2.
3.
4.
XYZ
Select Method mode
3
Select method from list
Confirm
Page 8
Select sequence
1.
Enter main
menu
2.
3.
4.
5.
2
VW
Esc
Select Sequence mode
then
Enter Sequence
ID confirm
Check icons
You can run the method/sequence
You can run the method/sequence. However a calibration
and/or a reagent bottle exchange must be performed soon
Check the electrodes and reagents for installation or
calibration purposes
Programming error due to missing parameters
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 9
Calibrate your electrode
pH electrode calibration
Select electrode
Form no. 4
1.
2.
3.
1.
Enter Electrode
window
Confirm
STU
1
then
to select electrode from list
Run
1
STU
To run the calibration
2.
Enter User ID and temperature, if required,
and confirm
1
STU
Page 10
3. Rinse electrode(s) and dip
in beaker
5. The measurement starts
4. Confirm
2
VW
1
STU
Press 1 to save results
then
2
VW
Calibration can be performed
on up to 5 buffers.
Automatic result display
Results: zero pH and sensitivity
View statistics
More cycles/End of analysis
Stop
1
STU
“Resume analysis”
Emergency stop
D21T047• Printed by Radiometer Analytical SAS • France • 2008-09H
Page 11
Calibrate your reagent
Select reagent
Reagent calibration
Form no. 5
1.
2.
3.
1.
Enter Reagent
window
Confirm
STU
1
then
to select reagent from list
Run
1
STU
To run the calibration
2.
Enter User ID and temperature, if required,
and confirm
1
STU
Page 12
3.
Enter Standard batch number and standard
amount then confirm
4. Rinse electrode(s) and dip in beaker
1
STU
5. Confirm
2
VW
1
STU
Press 1 to save result
then
2
VW
1
STU
Automatic result display
View statistics
More tests/End of analysis
Stop
1
STU
“Resume analysis”
Emergency stop
D21T047• Printed by Radiometer Analytical SAS • France • 2008-09H
Page 13
Run your analyses
Run
Running methods
Form no. 6
1.
2.
3.
To run your selected method
1
STU
Enter User ID and temperature, if required,
and confirm
1
STU
Enter Sample ID, sample amount
and confirm
1
STU
4. Rinse electrode(s) and dip electrode(s) in beaker then confirm
1
STU
Page 14
Press 1 to save results
Automatic result display
View statistics
VW
Stop
2
then
2
VW
More tests/End of analysis
Emergency stop
1
“Resume analysis”
STU
D21T047• Printed by Radiometer Analytical SAS • France • 2008-09H
Page 15
Use your Sample stack
for daily analyses
Preparing the sample stack
Running sequences
Form no. 7
1.
2.
Enter a new sample ID
and confirm
3.
Check the sample test quantity or
enter the sample test quantity to
place in beaker no.1
Confirm4.
2
VW
1
STU
+
Locate beaker no.1
5.
6.
Repeat steps 2 to 4 for
each beaker in the
sequence
7.
Esc
then
Esc
+
Page 16
Run
1.
2.
Press 1 to save results
2
then
VW
2
VW
1
STU
3. Rinse the electrodes, and dip in
beaker 1, confirm
Enter User ID and
temperature, if
required, confirm
1
STU
Automatic result display (manual)
View statistics (manual)
1
STU
Emergency stop
“Resume analysis”
Stop
1
STU
SAC Sequence
For automatic sample handling, using a sample changer, proceed as follows:
1.
2.
1
STU
Select SAC
Sequence
Run Sequence
D21T047• Printed by Radiometer Analytical SAS • France • 2008-09H
Page 17
A “?” in the Electrode tab?
Check your programming
1.
Connect the electrode(s).
See form no. 1
2. Dip electrode(s) in beaker
Direct measurements
Run
3.
Live measurements
To select a connected electrode
and display measurements
5
To display measurements
MNO
Form no. 8
1
STU
To start/stop stirring
Esc
To stop measurements
D21T047• Printed by Radiometer Analytical SAS • France • 2008-09H
Page 18
Page 19
Perform other operations while
your analysis is running
Operations during analyses
Instructions
Form no. 9
1.
Esc
2. Perform the necessary operations
Run in progress
1
then
Esc
STU
then
2
VW
Stirring in progress
In Routine mode, you can:
• View the method, reagent, electrode
parameters,
• View the GLP-archives,
• Change the stirring speed in the Cell
window.
In Supervisor mode, you can additionally:
• Create a method, a reagent, and an
electrode.
In pH-Stat mode, you can:
• Access certain regulation parameters
To enter the
Edit mode
You can change the
parameters of the active
method/sequence. These
changes will be taken into
account for the next run.
To return to the
analysis
Page 20
1.
User action required during analyses
To return to analysis window
2.
1.
2.
Esc
Stop
Esc
To return to edition window
Emergency stop
then
then
2
VW
1
STU
To return to the analysis window
To resume analysis, if desired. Calculations
are not performed
3.
Del
To end an analysis and get results.
Calculations are performed, if result found.
D21T047• Printed by Radiometer Analytical SAS • France • 2008-09H
Page 21
Deliver a preset volume of
reagent and follow
Volume = f(time) or
measured signal = f(volume)
Running the manual dosing
1. Select method. See form no. 3.
Manual dosing
Form no. 10
2.
3.
Enter Reagent menu then .
Select reagent and electrode.
Enter increment size and addition speed.
Enter the maximum volume to dose.
Define the type of measurement to perform.
Confirm .
5
MNO
1
STU
Manual dosing with measurements
Adjust size of increments.
Manual or automatic addition of increments.
Volume dispensed and live
measurements.
Direct curve
Measured signal =
f(dispensed volume ).
1
To add one increment or
STU
start/stop adding increments
automatically.
Page 22
Manual dosing with no measurement
Adjust size of increments.
Manual or automatic addition of increments.
Volume dispensed.
Direct curve
Dispensed volume (in ml) = f(time).
To add one increment or start/stop adding increments
1
STU
automatically.
Displaying direct or derivative curve - Identifying curve points
Del
Stop
Del
or
1
then .
STU
Direct or derivative curve.
Coordinates of that point of the curve.
To move along the curve.
1
or
STU
Stop
To quit.
Quit manual dosing
D21T047• Printed by Radiometer Analytical SAS • France • 2008-09H
Page 23
For burette air bubble removal,
run the Flush procedure
Maintenance recommendations
Form no. 11
Daily
Check and maintain electrodes according to manufacturer recommendations.
Check that there is no crystallisation at piston level, if so rinse the burette using
appropriate solvent to remove crystals. If not sufficient, dismount burette to remove
crystals.
Check that there is no crystallisation at addition tip end that may block the reagent to
pour freely, if so use appropriate solvent to remove crystals.
Check the level of reagent in the reagent bottle
Empty waste bottle if installed.
Remove air bubbles from the burette and tubing by running a “Flush” procedure from
the burette menu.
Weekly
glass
Reagent
cylinder
piston
Reagent
Flush the burette with appropriate solvent to
ensure complete cleaning of burette and
stopcock.
✓
Change burette
stand
Check the piston for leakages.
The burette stand must be changed when at
least 2 mm of liquid can be seen between
the two sealing lips. Refer to diagram.
Having 1 mm or less liquid between the two
sealing lips does not require to change the
burette stand.
Page 24
Annualy
Replace burette stand, 1, 5, 10, 25, 50 ml and the anti-diffusion tip, part. no.956-309.
Frequent use or use with aggressive or concentrated reagents occurring crystallisation risks
may severely reduce the burette and anti-diffusion tip lifetime and may consequently
require to replace them more frequently.
Replace burette stand set of 3 tubing, (stands 5, 10, 25, 50 ml, part number X31T115
and stand 1 ml, part number X31T116).
Contact your Radiometer Analytical representative for instrument revision and cleaning.
Recommendations of use
After installing new burette and/or tubing, before use, check that all tubing fittings are
sufficiently tightened to prevent leakage, and that no tubings are crimped (do not over
tight fittings)
The reagent should be at room temperature before use in the burette.
Do not leave burette installed with concentrated base, acid or salt solutions. Always rinse
after use, at least daily, to increase stopcock and burette lifetime.
Organic solvents: the organic solvents given as reacting with and/or destroying the
VITON® (Fluorinated polymer) should not be used with our burette stands (example:
TetraHydroFurane). The solvents given as simply making inflate the VITON® are allowed.
Please contact us if you wish to work with solvents destroying the VITON®.
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 25
Keep your system in
tiptop condition!
Storage recommendations
Form no. 12
Daily
Rinse electrodes and delivery tip throughly with distilled water after use.
Store the electrode(s) following the instructions given in the respective operating
instructions.
Overnight and up to 1 week
In case of spillages, clean the exterior surface of the instrument with a soft cloth and
tepid water. Avoid the accumulation of excess water around the keyboard area.
Long-term 1 week or more
Run the “Rinse” procedure from the Burette menu to remove reagent from burette and
tubings.
If the system is to be left in the same location, store electrode(s) following the long-term
storage instructions given in the respective operating instructions.
If the system is to be transported, dismount the burette stand, metal rod, and cell
accessories before packing.
For transportation, always use the packaging supplied by the manufacturer.
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 26
Page 27
Ordering information
at a glance
Spare parts & accessories
Form no. 13
Part no. TypeDescription
Burette stands
X51T025B501 Burette Stand, 1 ml, UV protected with connection tubing
X51T026B505 Burette Stand, 5 ml, UV protected with connection tubing
X51T017B510 Burette Stand, 10 ml, UV protected with connection tubing
X51T027B525 Burette Stand, 25 ml, UV protected with connection tubing
X51T028B550 Burette Stand, 50 ml, UV protected with connection tubing
Burette maintenance parts
956-309Anti-diffusion tip PTFE
X31T115Maintenance Kit of 3 tubes for B5xx & B6xx (excl, 1 ml burette)
X31T116Maintenance Kit of 3 tubes for B501 & B601only
X51T010Delivery tip, glass, burette >5 ml
X51T011Delivery tip, glass, burette <5 ml
X31T070Titrant bottle stopper GL45 assembly with desiccant & suction tubing
X31T110Titrant bottle stopper GL45 assembly with desiccant & suction tubing
for B501
Polypropylene titration vessels
904-490Titration Vessel, PP, 40-100 ml (pack of 50 pcs)
904-489Titration Vessel, PP, 22-45 ml (pack of 50 pcs)
904-488Titration Vessel, PP, 8-25 ml (pack of 50 pcs)
904-487Titration Vessel, PP, 1-9 ml (pack of 50 pcs)
Page 28
Part no. TypeDescription
Notebook keyboard
X31T108Notebook Keyboard Mask
X16T005Notebook Keyboard 83-keys, PS/2 port, US version
X16T006Notebook Keyboard 83-keys, PS/2 port, French
Balance connecting cables
A95Z201 C201Z9pf/1m/25pm (Mettler PE+option 016; Mettler AE+option 012 or 011)
A95Z205 C205Z9pf/2m/25pm (Mettler PE+option 016; Mettler AE+option 012 or 011)
A95Z206 C206Z1m serial cable for Mettler AM, PM, SM, AT, MT, UMT, PJ, AJ
A95Z202 C202Z9bf/1m/25bm (Sartorius MasterPro, QC, MP8-4)
A95Z204 C204Z9bf/2m/25bm (Sartorius MasterPro, QC, MP8-4)
A95Z207 C207Z9pf/1.20m/25pm (Precisa, old models)
A95Z208 C208Z9pf/1m/RJ45 (Precisa, XB series)
For other balances, please consult.
S40 (for use with new TIM550 casing)
X31T051Syringe, plastic, 5 ml (10 pcs)
X41T006Needle, 120 mm, KF cell (10 pcs)
X31T038Silicone Septa for X31T036, set of 10
X31T165Sample Container with funnel, 5pcs
S72T004Molecular Sieve, 250 g
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Page 31
Ordering information
at a glance
Calibration standards & solutions
Part no.TypeDescription
IUPAC Series pH Standards
S11M001pH 1.679pH 1.679 ±0.010 at 25°C, 500 ml
S11M002pH 4.005pH 4.005 ±0.010 at 25°C, 500 ml
S11M003pH 6.865pH 6.865 ±0.010 at 25°C, 500 ml
S11M004pH 7.000pH 7.000 ±0.010 at 25°C, 500 ml
S11M005pH 7.413pH 7.413 ±0.010 at 25°C, 500 ml
S11M006pH 9.180pH 9.180 ±0.010 at 25°C, 500 ml
S11M007pH 10.012pH 10.012 ±0.010 at 25°C, 500 ml
S11M008pH 12.45pH 12.45 ±0.05 at 25°C, 500 ml
4-7-10 Series
S11M012pH 4pH 4.00 at 25°C, 500 ml
S11M013pH 7pH 7.00 at 25°C, 500 ml
S11M014pH 10pH 10.00 at 25°C, 500 ml
Form no. 15
Conductivity standards
S51M0011 D KCl111.3 mS/cm ±0.5 % at 25°C, 500 ml
S51M0020.1 D KCl12.85 mS/cm ±0.35 % at 25°C, 500 ml
S51M0030.01 D KCl1408 µS/cm ±0.5 % at 25°C, 500 ml
S51M0040.05 % NaCl1015 µS/cm ±0.5 % at 25°C, 500 ml
C20C2500.1 M KCl12.88 mS/cm ±1 % at 25°C, 500 ml
C20C2700.01 MKCl1413 µS/cm ±1 % at 25°C, 500 ml
C20C2800.001 MKCl146.9 µS/cm ±2.5 % at 25°C, 500 ml
S51M013NaCl 2525 µS/cm ±5 % at 25°C, 250 ml
Page 32
Electrode Maintenance Solutions
S16M001RENOVO•NNormal Cleaning Solution, 250 ml
S16M002RENOVO•XXtra Strong Cleaning Solution, 250 ml
C20C370KS400Pepsin in HCl Cleaning Solution, 250 ml
C20C380KS410Thiourea Solution, 250 ml, for junction cleaning
C20C300KS100KCl Solution, saturated, 500 ml, for Red Rod & Calomel reference
C20C500KS160K
C20C310KS120KCl Solution, saturated with AgCl, 500 ml, for Ag/AgCl reference
S21M001KCl•CKCl Crystals, 15 g
S91M001GK ANNEXElectrode Maintenance Kit for pH & reference electrodes
Solution, saturated, 500 ml, for Hg/Hg2SO4 reference
2SO4
D21T047 • Printed by Radiometer Analytical SAS • France • 2008-09H
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.