Hach-Lange SONATAX sc User Manual

Page 1
DOC023.52.00117

SONATAX sc

User manual
04/2012, Edition 3A
© HACH-LANGE GmbH, 2006, 2010, 2012. All rights reserved. Printed in Germany
Page 2
Page 3
Table of contents
Section 1 Technical data ........................................................................................................................................ 5
2.1 Safety information .............................................................................................................................................. 7
2.1.1 Use of hazard information ........................................................................................................................ 7
2.1.2 Precautionary labels ................................................................................................................................. 7
2.2 General sensor information................................................................................................................................ 8
2.2.1 Property rights note .................................................................................................................................. 8
2.2.2 Usage areas ............................................................................................................................................. 8
2.3 Device overview................................................................................................................................................. 8
2.4 Functional principle ............................................................................................................................................ 9
Section 3 Installation ............................................................................................................................................ 11
3.1 Unpack sensor ................................................................................................................................................. 11
3.1.1 Handling of the immersed probe ............................................................................................................ 11
3.2 Connect sensor to an sc controller .................................................................................................................. 11
3.2.1 Connect sensor with quick connection ................................................................................................... 11
3.3 Function test .................................................................................................................................................... 12
3.4 Install sensor.................................................................................................................................................... 13
3.4.1 Select the measurement location and pre-set the device....................................................................... 13
3.4.1.1 Determine the distance to the tank edge ....................................................................................... 15
3.4.1.2 Determine measurement location .................................................................................................. 15
3.4.2 Installation of the sensor......................................................................................................................... 18
3.5 Advanced settings............................................................................................................................................ 18
Section 4 Operation .............................................................................................................................................. 23
4.1 Use of the sc controller .................................................................................................................................... 23
4.2 Sensor data logging ......................................................................................................................................... 23
4.3 Sensor setup.................................................................................................................................................... 23
4.3.1 Change of the sensor name ................................................................................................................... 23
4.4 SENSOR STATUS menu................................................................................................................................. 23
4.5 SENSOR SETUP menu ................................................................................................................................... 23
Section 5 Maintenance ......................................................................................................................................... 27
5.1 Maintenance tasks ........................................................................................................................................... 27
5.2 Wiper change................................................................................................................................................... 27
5.3 Cleaning tasks ................................................................................................................................................. 28
Section 6 Troubleshooting................................................................................................................................... 29
6.1 Operating state LED ........................................................................................................................................ 29
6.2 Error messages................................................................................................................................................ 29
6.3 Warnings.......................................................................................................................................................... 30
6.4 SLUDGE DOCTOR, (diagnostic software for SONATAX sc)........................................................................... 30
Section 7 Replacement parts and accessories ................................................................................................ 31
7.1 Replacement parts........................................................................................................................................... 31
7.2 Accessories...................................................................................................................................................... 31
3
Page 4
Table of contents
Section 8 Warranty and liability........................................................................................................................... 33
Section 9 Contact ................................................................................................................................................. 35
Modbus register .................................................................................................................................................. 37
4
Page 5

Section 1 Technical data

Changes reserved!
General
Measurement method Ultrasonic measurement (750–1250 kHz)
Measuring range 0.2 m–12 m (0.7 ft–40 ft) sludge level
Resolution 0.03 m (0.1 ft) sludge level
Accuracy 0.1 m (0.33 ft)
Response time 10–600 s (adjustable)
calibration One time on commissioning
Environmental conditions
Ambient temperature \> 0–50 °C (\> 0–122 °F)
Temperature compensation Automatic
Flow speed Max. 3 m/s
Pressure range
Sensor specifications
Dimensions 130 mm
Earth Approximately 3.5 kg (123.5 oz) (without struts)
Maintenance requirements < 1 hour/month, typically
Cable length 10 m (33 ft), maximum 100 m (330 ft) with extension cable
Power consumption 12 V, 2.4 W, (200 mA)
Protection type IP68 (
Compliance CE, TÜV GS, UL/CSA
Materials
Probe body Stainless steel 1.4581
Base plate and wiper POM
Wiper magnet casting compound Epoxy resin
Wiper rubber Silicone rubber
Housing seals NBR (acrylonitrile butadiene rubber)
Light guide seal Polyurethane
0.3 bar or 3 m ( 43.55 psi or 10 ft)
× 185 mm (5 in. × 7.3 in.) (H × Ø)
1 bar (14.5 psi))
Light guide LEXAN polycarbonate
Sensor connection cable (fixed connection)
Sensor connection plug (fixed connection)
Cable gland Stainless steel 1.4571
Cable gland insert TPE-V
Cable gland O-ring NBR, silicone
1 cable pair AWG 22 / 12 VDC twisted, 1 cable pair AWG 24 / data twisted, common cable shield, Semoflex (PUR)
Type M12, protection type IP67
5
Page 6
Technical data
6
Page 7

Section 2 General information

The information in this manual has been carefully checked and is believed to be accurate. However, the manufacturer assumes no responsibility for any inaccuracies that may be contained in this manual. In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual, even if advised of the possibility of such damages. In the interest of continued product development, the manufacturer reserves the right to make improvements in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s web site.

2.1 Safety information

Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger, warning and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this manual.

2.1.1 Use of hazard information

2.1.2 Precautionary labels

DANGER
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate injury.
Notice
Indicates a situation that, if not avoided, could result in damage to the instrument. Information that requires special emphasis.
Note: Information that supplements points in the main text.
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
7
Page 8
General information
Electrical devices marked with this symbol may after August 12, 2005 no longer be disposed of in unsorted domestic or industrial waste Europe-wide. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of life equipment to the manufacturer for disposal This is free for the consumer.
Note: Obtain instructions on the correct disposal of all (marked and not marked) electrical products that were supplied or manufactured by the manufacturer at the relevant sales office.

2.2 General sensor information

2.2.1 Property rights note

The SONATAX sc is intended for the measurement of sludge levels in water. Use in other media without test of the materials (refer to
Section 1 Technical data on page 5) or consultation with the
manufacturer is regarded as explicitly not in accordance with requirements.
Any use other than use in accordance with requirements defined in the user manual leads to the loss of the warranty claims and can lead to personal injury and property damage, for which the manufacturer assumes no liability.
Parts of the device software are based on the work of the Independent JPEG Group.

2.2.2 Usage areas

2.3 Device overview

The SONATAX sc can be used anywhere that the solid/liquid separation layer must be monitored. For example in the area of final sedimentation or in sludge treatment (thickeners).
Figure 1 shows the device components of the SONATAX sc.
Figure 1 Device overview
1 Operating state LED (refer to Section 6.1) 3 Mount for the tank edge attachment 2 Wiper 4 Sensor head
8
Page 9

2.4 Functional principle

General information
In a tank where solid materials in water (or another fluid) can settle to the bottom, there is a boundary between the settled solids and the clear phase above. The distance from the water surface to the bottom is the sludge level. The sludge height is the distance from the tank floor.
More precisely, the sludge level (or the sludge height) indicates the place in a tank where (viewed from the water surface) the solid content first exceeds a defined limit. This limit value depends on the application. It will, for example, be higher in a pre-thickener of a wastewater treatment plant than in its final sedimentation, where the overlying fluid should be clear water.
The SONATAX sc measures the sludge level via an echo signal (refer to Figure 2) of an ultrasonic pulse. This echo signal is shown in the probe menu SENSOR SETUP > TEST / MAINT > SIGNALS in the echo list (refer to ECHO LIST on page 26). The depth and the echo strength are specified at the ultrasonic transducer in digits (1 digit approximately 1 µV).
Figure 2 Functional principle
Echos from layers that are further away are quieter (weaker) than those which are closer. The SONATAX sc compensates for this attenuation. The result is indicated on the SONATAX sc as a profile. The data is found in the probe menu SENSOR SETUP > TEST / MAINT > SIGNALS in the profile list (refer to PROFILE LIST on
9
Page 10
General information
page 26) ((2) Figure 3). The graphic display (for example sc1000)
shows the profile in the measurement operation as a graphic ((1)
Figure 3).
Figure 3 Profile as graphic and list
The example shows a typical profile curve of a sludge level measurement. The Y axis in this graphic reaches from the floor (which can be set in the probe menu SENSOR SETUP > CALIBRATE > TANK DEPTH on page 24) up to the water surface. The profile strength is given on the X axis. The profile increases at a sludge level. If the solid content below the sludge level remains constant, the profile strength decreases again due to the absorption of the ultrasound in the sludge. At the very bottom, the echo of the floor is normally detected.
The vertical line represents the threshold. The sludge level is detected by the SONATAX sc at the point at which the profile exceeds this threshold for the first time viewed from above (analogous to the abovementioned definition of the sludge level). In the graphic, the detected sludge level is identified by a mark on the Y axis.
The threshold is set automatically on the SONATAX sc (setting: THRESHOLD AUTO = xy % in the probe menu SENSOR SETUP > CALIBRATE > ADV.SETTINGS > THRESHOLD AUTO on
page 24). The automatic threshold function searches for the
maximum in the detected profile. Through an exact setting of the tank depth, a faulty measurement due to the echo of the floor is ruled out.
The background is that the echo signal of the floor in most tanks is the strongest in the profile. With a correct setting of the tank depth, the floor signal is not considered by the automatic threshold function during the determination of the sludge level.
For more information on the correct determination of the tank depth, refer to section Section 3.4.1 on page 13.
10
Page 11

Section 3 Installation

3.1 Unpack sensor

DANGER
The installation described in this section of the user manual should only be performed by qualified and specialized personnel. The sensor is not suitable for installation in hazardous areas.
The SONATAX sc can be used with an sc100, sc200 or with an sc1000 controller. Installation instructions are found in the handbook of the controller.
The SONATAX sc is supplied with the following components:
Sensor
User manual
Wiper blades set (5 pieces)
If one of the parts is missing or damaged, please contact the manufacturer or dealer.

3.1.1 Handling of the immersed probe

The immersed probe contains a sensitive ultrasound transducer. Make sure that this is subjected to no hard mechanical impacts. Do not install the probe hanging on the cable. Make sure before the insertion of the probe into the medium that all functions run undisrupted through a check of the system. Check the probe carefully for external damage.

3.2 Connect sensor to an sc controller

3.2.1 Connect sensor with quick connection

The sensor cable is very easy to connect to the controller using a quick connection (Figure 4). Keep the protective cap of the connection socket in case you remove the sensor later and must seal the socket. For longer sensor cable lengths, optional extension cables are available.
Figure 4 Connect sensor with quick connection
11
Page 12
Installation
1
2
3
4
5
6
Figure 5 Pin assignment of the quick connection
Number Assignment Wire color
1 +12 V= Brown 2EarthBlack 3 Data (+) Blue 4 Data (–) White
5Shield
6 Insulator notch
Shield (gray wire for existing quick
connection)

3.3 Function test

Directly after the connection of the sensor to the controller, perform a function check.
1. Connect the controller to the mains power.
2. If the controller does not recognize the new sensor, go to the
SCAN SENSORS menu (refer to the user manual of the controller).
3. Confirm any prompt and wait until the new sensor is found.
If the new sensor is found, the controller switches to the measurement operation.
A wipe process is triggered and the device switches to the measurement operation. If no other messages appear, the sensor is functional.
Note: Measurements in air lead to no measurement values. The error message SENSOR MEASURE is shown. This does not show a malfunction.
12
Page 13

3.4 Install sensor

Installation
The SONATAX sc is intended for installation with a tank edge attachment. For more information, refer to the installation instructions of the tank edge attachment.
Note: Do not begin with the installation of the tank edge attachment before the installation location is determined (refer to Section 3.4.1).
Figure 6 SONATAX sc scale drawing

3.4.1 Select the measurement location and pre-set the device

In the determination of the sludge level over the duration of an ultrasonic echo, the entire space between the probe on the water surface and the tank bottom is required for the measurement. As solid bodies in this area disrupt the measurements, special attention is required in the selection of the measurement location. In the following, it is explained how the suitability of a possible measurement location can be tested. The tank depth should be known here.
A suitable measurement location must be found depending on tank type. The graphics that follow show typical examples for a round tank and a rectangular tank. The point (1) in Figure 7 and Figure 8 is a suitable measurement location.
13
Page 14
Installation
Figure 7 Round tank graphic
Figure 8 Rectangular tank graphic
At a suitable measurement location, there should be:
Sufficient sludge present for the measurement, and
14
A calm, clear phase above the sludge level
In the examples shown, area (A) shows only very little or no sludge, while in the area (C) above the sludge layer, sludge clouds may disrupt the measurement. The measurement points (A) and (C) are
Page 15
not suitable for the installation. Suitable measurement points can be found in between.
Note: Installation note: For round tanks with a rotating scraper bridge, install the probe in the direction of rotation on the back of the bridge (Figure 7).
3.4.1.1 Determine the distance to the tank edge
If the SONATAX sc is installed on the tank edge (i.e. not on a scraper bridge), make sure there is sufficient distance from the SONATAX sc to the tank edge.
This distance is dependent on the depth of the tank. Use this formula to find a guideline value for this distance:
0.20 m + (0.05 x tank depth in meters) = distance to tank edge
This value can vary depending on composition of the tank. How a possible installation location can be checked for its suitability is described in Section 3.4.1.2 Determine measurement location.
3.4.1.2 Determine measurement location
1. Connect the probe to the sc controller and carefully immerse
the probe in the water (approximately probe cable at the selected measurement location.
Installation
20 cm (7.9 in.)) on the
2. Select the probe menu SENSOR SETUP > CALIBRATE and do a measurement in the REFLEXLIST menu item.
After a few seconds, the reflex list is shown (refer to REFLEXLIST
on page 24). This list usually contains the tank floor as the
strongest signal.
If strong reflections between the water surface and the tank floor are present (for example due to pipes, plates etc.), another measurement location must be selected. Sometimes it is sufficient to move the probe by a few centimeters.
In the example (Table 1), there are two disruptive reflexes at a depth of 0.87 m and 2.15 m, and the tank floor is at a depth of
3.30 m.
Table 1 Reflex list
Meters Intensity
0.87 25 %
2.15 2 %
3.30 100 %
At an ideal measurement location, no disruptive reflections should occur above the floor in the reflex list after multiple measurements (Table 2).
Table 2 Reflex list
Meters Intensity
3.30 100 %
15
Page 16
Installation
If no measurement location without disruptive reflexes can be found, select the probe menu SENSOR SETUP > CALIBRATE and trigger a measurement in the PROFILE LIST menu item.
If there are increased values at the depth of the disruptive reflexes in the PROFILE LIST which influence the measurement value, another measurement location must be selected. Sometimes it is sufficient to move the probe by a few centimeters (refer to
Figure 9, right).
If the reflections do not disrupt the measurement value, this can be a suitable measurement location (refer to
Figure 9, left).
Figure 9 Determine measurement location
3. Then assemble the tank edge attachment there and install the probe.
Note: The probe should be installed approximately 20 cm (7.9 in.) deep at the installation location (somewhat more than the entire probe head) to make sure that it remains immersed during the entire measurement operation.
4. In the CALIBRATE probe menu, enter the actual plunger depth (Item A,
Figure 10) in the PLUNGERDEPTH menu item.
5. Trigger after approximately 2 minutes another measurement of the reflex list (to adjust the temperature sensor to the water temperature) (also refer to
6. In the CALIBRATE probe menu, enter the recorded value of the tank depth (Item C,
Figure 10) in the TANK DEPTH menu item.
Important note: The shown value of the tank depth is calculated from information shown:
Tank depth C = plunger depth A + measured distance to floor B This calculation is performed internally and leads to incorrect results if the plunger depth is entered incorrectly.
REFLEXLIST on page 24).
16
Page 17
Installation
Figure 10 Plunger depth–tank depth
Note: Should the reflex list not contain the tank depth (possible when there is a lot of sludge above the floor), the tank depth must be determined in another way (e.g. measuring).
Important note: The value entered in the TANK DEPTH menu item must in no case be larger than the actual tank depth. This could lead to incorrect measurement values.
In the determination of a suitable measurement location, the optionally available SLUDGE DOCTOR diagnostic software can be helpful (refer to 7.2 Accessories on page 31). The software enables the representation and saving of all graphic profiles of the SONATAX sc in a defined time interval (5 minutes–2 hours).
17
Page 18
Installation

3.4.2 Installation of the sensor

The software also shows and saves all important measurement and configuration parameters, such as the reflex list, measurement values, limit values and all advanced settings, such as response time, frequency, amplitude, angle and temperature.
Take further information from the SLUDGE DOCT OR user manual
DOC013.98.90411.
Note: Take further information on the installation from the installation
instructions.

3.5 Advanced settings

18
Figure 11 Installation of the sensor
In the SENSOR SETUP > CALIBRATE > ADV. SETTINGS menu, special probe parameters can be found.
The default settings of these parameters are selected so that no correction is required in most applications. In exceptional cases, these parameters must be adapted depending on the application.
The examples that follow show typical sludge profiles.
Page 19
Installation
Example 1: Pre-thickener with clear separation layer, but
sporadically empty.
The image shows a thickener with a clear profile. The signals above the sludge level caused by turbid water do not disrupt the measurement.
The same thickener is empty in this image. The threshold now
justs itself automatically to the signals of the turbid water. The
ad recorded measurement value at 0.63 m is incorrect.
Recommended measure:
Increase the adjustment of the ADV. SETTINGS > LL THRESH. AUTO setting from 0.3 to 1.0.
Note: LL THRESH. AUTO gives the smallest possible value that the threshold can accept.
The image shows the empty thickener after the adjustment. The result 3.18 m is correct.
Example 2: Application with c signal.
Very low to absolutely no interference signals through turbid water above the separation layer. Strength of the echo 0.61 is above the LL THRESH. AUTO (pre-setting 0.3) and leads to correct measurement value 2.28 m.
lear separation layer, but weak
19
Page 20
Installation
Very low to absolutely no interference signals through turbid water above the separation layer. Strength of the maximum echo <0.3. Instead of the correct measurement value, only the tank floor is found.
Increase the adjustment of the ADV. SETTINGS > LL THRESH.
O setting from 0.3 to 0.1. The threshold 0.16 finds the correct
AUT measurement value 2.25 m.
Example 3: Thickene concentration up to just under the water surface; echo is absorbed in the uppermost sludge layer:
In this application, the floor signal is no longer detected due to the high absorption of the ultrasonic signal. The shown measurement value 0.73 is correct. No adjustment of the ADV. SETTINGS necessary.
If the sludge increases so far that the sludge level is above the
easurement range (the measurement range begins from 0.2 m
m beneath the sensor), it can occur that no sludge level is detected (SENSOR MEASURE error message). In this case, the application must be checked.
r or other application with very high sludge
20
Page 21
Installation
Example 4: Application with large tank depth and clear separation
layer, but high absorption of the ultrasound by turbid materials above the separation layer.
Absorption is so high that no sludge level is detected at a plunger depth of 0.2 m and LL THRESH. AUTO 0.1 (SENSOR MEASURE error message).
After adjustment of the plunger depth SENSOR SETUP > CALIBRA precisely determined at 3.30. The strength of the echo 1.41 shows that there is a clear separation layer.
Note: After adjustment of the plunger depth, the setting of the tank floor must be checked (SENSOR SETUP > CALIBRATE > TANK DEPTH)
TE > PLUNGERDEPTH to 3 m, the sludge level can be
Example 5: Application with clear separation layer, but interference signal above the separation layer (for example skimmer).
Ideal profile, correct measurement value.
The same measurement point with sporadic interference of the
o in a 0.67 m tank depth. The measurement value is incorrect.
ech
21
Page 22
Installation
Fade out this disruption in SENSOR SETUP > CALIBRATE > ADV. SETTINGS > FADE-OUT in the range of 0.5–0.8 m. The sensor now ignores all signals in this range and finds the correct measurement value at 2.13 m.
Example 6: Application cloud above the separation layer.
In this application, the separation layer should be measured, the sludge cloud should be ignored (for example pre-thickener).
No change of the default setting. The automatic threshold detects the separation layer; the echo of the sludge cloud is lower than the echo of the sludge cloud.
The image shows the same profile as before, but with another
reshold and measurement value.
th
In this application, the sludge cloud should be measured as an early warning system (for example sludge drift in the final clarification tank).
Change of the automatic threshold SENSOR SETUP > CALIBRATE > ADV. SETTINGS > THRESHOLD AUTO to 25 %. The automatic threshold detects the sludge cloud.
with clear separation layer and a sludge
22
The optionally available SLUDGE DOCTOR diagnostic software can be helpful for the setting of the special probe parameters (refer to 7.2 Accessories on page 31). The software enables the representation and saving of all graphic profiles of the SONATAX sc in a defined time interval (5 minutes–2 hours).
The software also shows and saves all important measurement and configuration parameters, such as the reflex list, measurement values, limit values and all advanced settings, such as response time, frequency, amplitude, angle and temperature.
Take further information from the SLUDGE DOCT OR user manual DOC013.98.90411.
Page 23

Section 4 Operation

4.1 Use of the sc controller

4.2 Sensor data logging

4.3 Sensor setup

Before using the sensor with an sc controller, make yourself familiar with the operating method of the controller. Learn to navigate through the menu and to use the menu functions. Further information is found in the user manual of the controller.
The sc controller provides one data log and one event log for each sensor. The data log stores the measurement data at selectable intervals. The event log stores a large number of events that occur on the devices, such as configuration changes, alarms and warnings etc. The data log and the event log can be read out in CSV format. Refer to the controller user manual for more information about the log download.
During initial sensor setup, select the parameter that corresponds to the applicable instrument.

4.3.1 Change of the sensor name

When a sensor is installed for the first time, the serial number is shown as the measurement location (or sensor name). The measurement location can be changed as follows:
1. From the main menu, select SENSOR SETUP and confirm the selection.
2. If more than one sensor is connected, mark the required sensor and confirm the selection.
3. Select CONFIGURE and confirm the selection.
4. Select EDITED NAME and edit the name. Return to the Sensor
Setup menu with confirm or cancel.

4.4 SENSOR STATUS menu

Select the SONATAX sc (if several sensors are connected)
SONATAX sc
ERRORS Lists error messages; refer to 6.2 Error messages on page 29 WARNINGS Lists warnings; refer to 6.3 Warnings on page 30

4.5 SENSOR SETUP menu

Select the SONATAX sc (if several sensors are connected)
WIPE
CALIBRATE
Plunger depth of the probe underside (refer to 3.4.1.2 Determine measurement
PLUNGERDEPTH
location on page 15).
Configurable: 0.1 m to 3 m (0.3 ft to 9.8 ft)
23
Page 24
Operation
4.5 SENSOR SETUP menu (Continued)
Select the SONATAX sc (if several sensors are connected)
Shows the reflex list. A new measurement can be triggered. A list is shown of all detected solid bodies that have clearly reflected the ultrasonic
impulse. The measurement depth is shown in m or ft and the strength of the
REFLEXLIST
TANK DEPTH
PROFILE LIST
ADV. SETTINGS
FACTOR
THRESHOLD AUTO
LL THRESH. AUTO
FADE-OUT
BEGIN
END
SET DEFAULTS
              
reflected signal in %, relative to the strongest signal in the list. This list mostly contains the tank floor. If strong reflections between the water surface and the tank floor are present (for example due to pipes, plates etc.), it should be checked whether another installation location offers better conditions.
Entry of the floor depth (refer to 3.4.1.2 Determine measurement location on
page 15).
Configurable: 1.00 m to 12 m (3.3 ft to 39.4 ft) A profile is calculated from the ultrasound echo and shown for the corresponding
depths respectively as profile strength. The profile is similar in its curve with the TS profile of the tank. At an average solid content, the values are in the order of 1. A new measurement can be triggered (refer to Figure 3 on page 10).
Correction factor for the speed of sound. Configurable: 0.3 to 3.0, default setting 1.0
A change of the default factor is only required if the local speed of sound in the fluid deviates from the speed of sound in water:
Factor (fluid) = speed of sound (fluid) / speed of sound (water)
Note: For applications in water, the factor should remain at 1.0.
With the automatic threshold function, the system constantly adapts the environmental conditions and automatically changes the sensitivity in order to guarantee maximum accuracy. Recommendation: 75% Configurable: 1-95%
LL THRESH. AUTO gives the smallest possible value that the threshold can accept. Configurable: 0.1 to 1.0, recommendation 0.3
If fixed installations or other influences interfere at certain tank depths, then this range can be faded out —·it is then completely ignored. Configurable: ON, OFF
Upper limit of the range that should be faded out. Only active when FADE-OUT = ON.
Lower limit of the range that should be faded out. Only active when FADE-OUT = ON.
Reset to factory configuration for all probe-specific parameters. This occurs only after a security prompt.
24
Page 25
4.5 SENSOR SETUP menu (Continued)
Select the SONATAX sc (if several sensors are connected)
KONFIGURIEREN (CONFIGURE)
Operation
EDIT NAME
PAR AMETER
MEAS UNITS
CLEAN. INTERVAL
RESPONSE TIME
LOGGER INTERVAL
SET DEFAULTS
TEST / MAINT
PROBE INFO
SENSOR NAME Display of the device name.
EDIT NAME
SERIAL NUMBER. Device number MODEL NUMBER Item no. of the sensor. HARDWARE-VERS Production status of main circuit board SOFTWARE-VERS Sensor software version
COUNTER
WIPER COUNTER
TOTAL TIME Operating hours counter MOTOR Forward counter for wiping processes. TEST / MAINT Date of the last performed maintenance.
REPLACE PROFILE
        
Freely editable (up to 16 characters) Factory setting: device number
The measurement result can be shown as the sludge level (as the distance of the sludge from the water surface) or as the sludge height (as the distance from the tank floor). To calculate the sludge height, the tank depth specified in the TANK DEPTH menu item is used. (Sludge height = tank depth – sludge level) Configurable: sludge level, sludge height
Dimension of the measurement result. Configurable: meters, feet
Wipe interval, Recommendation: 15 minutes Configurable: 1 minute to 1 hour
Damping of the measurement value. In the case of high measurement value fluctuations, a high damping of for example 300 seconds is recommended. Configurable: 10 to 1800 seconds
The interval for the internal data log. Configurable: 1, 2, 3, 4, 5, 6, 10, 15, 30 minutes
Reset to factory configuration for all menu items listed above. This occurs only after a security prompt.
Display of the freely selectable measurement location (factory setting: device number).
Backward counter for wiping processes of the wiper profile. After expiration of the counter, a warning message is shown. In the case of a wiper change, the counter should be reset again.
For the wiper profile change, the wiper arm moves to a central position. In this position, the wiper arm can then be removed and installed without problems.
25
Page 26
Operation
4.5 SENSOR SETUP menu (Continued)
Select the SONATAX sc (if several sensors are connected)
TEST / MAINT
SIGNALS
MOIST Indicator as to whether water is in the probe. TEMPERATURE Temperature of the surrounding water in °C or F°. SENSOR ANGLE Deviation of the probe axis from the perpendicular in degrees.
The received echo signal in digits (units of the AD converter) is shown at the
ECHO LIST
PROFILE LIST
REFLEXLIST
FREQUENCY The resonance frequency of the ultrasound transducer is shown. AMPL DIAG The resonance voltage of the ultrasound transducer is shown.
THRESHOLD
SHOW AMPL.
corresponding measurement depths in a list. The first element at 0 meters shows the strength of the transmission pulse. A new measurement can be started.
A profile is calculated from the ultrasound echo and shown for the corresponding depths respectively as profile strength. The profile is similar in its curve with the TS profile of the tank. At an average solid content, the values are in the order of 1. A new measurement can be started. (See Figure 3 on page 10)
Shows the reflex list. A new measurement can be triggered. Refer to REFLEXLIST
on page 24
To determine the sludge level, a profile is first calculated from the ultrasonic echo. This roughly returns the solid content depending on the tank depth. The sludge level is assigned to the tank depth where the profile exceeds the threshold for the first time.
On activation, the resonance profile of the ultrasound transducer is shown instead of the sludge profile (PROFILE LIST) as a graphic in the measurement window (only valid with sc1000). The resonance profile can be switched ON and OFF. After switch-off, the sludge profile (PROFILE LIST) is shown on the measurement window again.
26
Page 27

Section 5 Maintenance

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
No components that can be serviced by the user are in the inside of the probe. Opening of the probe by the user leads to the loss of the manufacturer guarantee and may cause malfunctions.
The cleanliness of the ultrasound transducer is crucial for the accuracy of the measurement results.
The installed wiper is able to remove all impurities under normal circumstances and if the selection of the wiper interval is not too large (30 minutes).
If the regular check of the probe head and of the wiper (monthly) indicates dirt, wear of the wiper rubber or a defect, then either the probe head must be cleaned, or the wiper profile and/or the defective part must be exchanged.

5.1 Maintenance tasks

DANGER
Take the maintenance intervals from Table 3.
Table 3 Maintenance schedule
Time interval Action
Monthly Visual inspection, if necessary, clean Annually, at the latest after 20,000 wipe cycles Change wiper rubber

5.2 Wiper change

CAUTION
Obey the locally applicable accident prevention regulations. Wear protective gloves where necessary during the change of the wiper rubber.
1. Go to SENSOR SETUP > TEST / MAINT > REPLACE
The wiper arm (refer to item 3 in Figure 12) moves for the wiper change into a central position.
2. Remove the guide screw (refer to item 2 in Figure 12) and take off the wiper arm.
PROFILE.
3. Pull the wiper profile (refer to item 1 in Figure 12) forward out of the wiper arm.
4. Slide a new wiper profile with the bevel at the front into the guide.
27
Page 28
Maintenance
5. Install the wiper arm again and tighten the guide screw again
by hand.
Note: Only tighten the screw hand-tight until the safety lugs have audibly engaged 2–3 times.
6. Select OK and the wiper arm automatically moves back to the initial position.
Figure 12 Wiper unit
1 Wiper profile 3 Wiper arm 2 Guide screw

5.3 Cleaning tasks

Use water and a suitable brush to remove heavy contamination on the probe or on the strut as necessary.
Clean the ultrasound transducer carefully with water and a lint-free cloth.
28
Page 29

Section 6 Troubleshooting

6.1 Operating state LED

The sensor is equipped on the top side with an LED that provides information on the operating state.
Figure 13 Operating state LED
1 Operating state LED
Table 4 Sensor status
Green LED No errors or warnings
Green/red flashing LED
Red LED Error
LED off No device function
Probe in operation, probe position deviates strongly from the perpendicular, the measurement value is retained, there is no error

6.2 Error messages

In the case of an error, an error message is shown on the controller. Find error messages and notes on the resolution of the errors in
Table 5.
Table 5 Error messages
Error message Reason Resolution
The PROFILE LIST and graphic on the sc1000 is less than the programmed value for
SENSOR MEASURE
POS. UNKNOWN
AMPL DIAG Internal error Call service MOIST Moisture value \> 10 Call service
SENSOR ANGLE
SYSTEM ERROR Faulty RAM Call service
LL THRESH.AUTO. over the entire tank depth, or the ultrasound transducer is contaminated or not immersed.
The wiper position is not detected. The wiper is in the central position (after a wiper change).
Defective light barrier plate Call service
Application with particles that block the wiper.
The probe is more than 20° from perpendicular and for longer than 180 seconds.
Incorrectly calibrated position sensor Call service
Check device data, also the plunger depth, floor depth and advanced settings, check installation and remove contamination.
Start a wiping process
Cleaning of the ultrasound transducer and of the wiper system.
Check installation
29
Page 30
Troubleshooting

6.3 Warnings

In the case of a warning, a warning message is shown on the controller. Find warnings and notes on the resolution of the warning in Table 6.
Table 6 Warnings
Warning message Reason Resolution
REPLACE PROFILE The counter for the wiper profile has expired Change wiper profile

6.4 SLUDGE DOCTOR, (diagnostic software for SONATAX sc)

SLUDGE DOCTOR is an optionally available diagnostic software for the SONATAX sc probe in connection with the controllers sc100, sc200 or sc1000. The software enables the representation and saving of all graphic profiles of the SONATAX sc in a defined time interval (5 minutes–2 hours).
The software also shows and saves all important measurement and configuration parameters, such as the reflex list, measurement values, limit values and all advanced settings, such as response time, frequency, amplitude, angle and temperature.
Take further information from the SLUDGE DOCT OR user manual DOC013.98.90411.
30
Page 31

Section 7 Replacement parts and accessories

7.1 Replacement parts

Description Number Order number
SONATAX sc 1 LXV431.99.00001 1 set of replacement wiper blades out of silicone
for SONATAX/SONATAX sc probes (5 pieces) Wiper arm 1 LZY344 Guide screw
(for stopping the wiper arm) User manual (xx=language code) 1 DOC023.xx.00117

7.2 Accessories

Description Order number
Pivot attachment set, 0.35 m (1.15 ft) LZX414.00.72000 Pivot attachment set, 1 m (3.3 ft) LZX414.00.71000 Rail assembly attachment set LZX414.00.73000 Tank edge attachment set LZX414.00.70000 Scraper bridge attachment set LZX414.00.74000 SONATAX sc chain stay LZX914.99.11300 SLUDGE DOCTOR, diagnostic software without interface cable LZY801.99.00000 SLUDGE DOCTOR, diagnostic software with sc200 interface cable LZY801.99.00010 SLUDGE DOCTOR, diagnostic software with sc1000 interface cable LZY801.99.00020
1LZX328
1LZY345
31
Page 32
Replacement parts and accessories
32
Page 33

Section 8 Warranty and liability

The manufacturer warrants that the supplied product is free of material and manufacturing defects, and undertakes to repair or to replace any defective parts without charge.
The warranty period is 24 months. If a maintenance contract is taken out within 6 months of purchase, the warranty period is extended to 60 months.
With the exclusion of further claims, the supplier is liable for defects, including the lack of assured properties, as follows: all parts that, within the warranty period calculated from the day of the transfer of risk, can be demonstrated to have become unusable or that can only be used with significant limitations owing to circumstances prior to transfer of risk, in particular due to incorrect design, substandard materials or inadequate finish, shall be repaired or replaced at the supplier's discretion. The identification of such defects must be reported to the supplier in writing as soon as possible, but no later than 7 days after the discovery of the fault. If the customer omits this notification, the product is considered approved despite the defect. Further liability for indirect or direct damages is not accepted.
If device-specific maintenance- or inspection work prescribed by the supplier is to be performed within the guarantee period by the customer (maintenance) or by the supplier (inspection) and these requirements are not met, claims for damages that result from non-observance of these requirements are void.
Further claims, in particular for consequential damages, cannot be made valid.
Wear and damage caused by improper handling, incorrect installation or non-designated use are excluded from this clause.
The process instruments of the manufacturer are of proven reliability in many applications and are therefore often used in automatic control loops to enable the most economical and efficient operation of the relevant process.
To avoid or limit consequential damage, it is therefore recommended that the control loop be designed such that an instrument malfunction results in an automatic changeover to the backup control system. This guarantees the safest operating condition both for the environment and the process.
33
Page 34
Warranty and liability
34
Page 35

Section 9 Contact

HACH Company World Headquarters
P.O. Box 389 Loveland, Colorado 80539-0389 U.S.A. Tel (800) 227-HACH (800) -227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com
HACH LANGE GMBH
Willstätterstraße 11 D-40549 Düsseldorf Tel. +49 (0)2 11 52 88-320 Fax +49 (0)2 11 52 88-210 info@hach-lange.de www.hach-lange.de
HACH LANGE GMBH
Rorschacherstrasse 30a CH-9424 Rheineck Tel. +41 (0)848 55 66 99 Fax +41 (0)71 886 91 66 info@hach-lange.ch www.hach-lange.ch
Repair Service in the United States:
HACH Company Ames Service 100 Dayton Avenue Ames, Iowa 50010 Tel (800) 227-4224 (U.S.A. only) Fax (515) 232-3835
HACH LANGE LTD
Pacific Way Salford GB-Manchester, M50 1DL Tel. +44 (0)161 872 14 87 Fax +44 (0)161 848 73 24 info@hach-lange.co.uk www.hach-lange.co.uk
HACH LANGE FRANCE S.A.S.
8, mail Barthélémy Thimonnier Lognes F-77437 Marne-La-Vallée cedex 2 Tél. +33 (0) 820 20 14 14 Fax +33 (0)1 69 67 34 99 info@hach-lange.fr www.hach-lange.fr
Repair Service in Canada:
Hach Sales & Service Canada Ltd. 1313 Border Street, Unit 34 Winnipeg, Manitoba R3H 0X4 Tel (800) 665-7635 (Canada only) Tel (204) 632-5598 Fax (204) 694-5134 canada@hach.com
HACH LANGE LTD
Unit 1, Chestnut Road Western Industrial Estate IRL-Dublin 12 Tel. +353(0)1 460 2522 Fax +353(0)1 450 9337 info@hach-lange.ie www.hach-lange.ie
HACH LANGE NV/SA
Motstraat 54 B-2800 Mechelen Tel. +32 (0)15 42 35 00 Fax +32 (0)15 41 61 20 info@hach-lange.be www.hach-lange.be
Repair Service in Latin America, the Caribbean, the Far East, Indian Subcontinent, Africa, Europe, or the Middle East:
Hach Company World Headquarters, P.O. Box 389 Loveland, Colorado, 80539-0389 U.S.A. Tel +001 (970) 669-3050 Fax +001 (970) 669-2932 intl@hach.com
HACH LANGE GMBH
Hütteldorfer Str. 299/Top 6 A-1140 Wien Tel. +43 (0)1 912 16 92 Fax +43 (0)1 912 16 92-99 info@hach-lange.at www.hach-lange.at
DR. LANGE NEDERLAND B.V.
Laan van Westroijen 2a NL-4003 AZ Tiel Tel. +31(0)344 63 11 30 Fax +31(0)344 63 11 50 info@hach-lange.nl www.hach-lange.nl
HACH LANGE APS
Åkandevej 21 DK-2700 Brønshøj Tel. +45 36 77 29 11 Fax +45 36 77 49 11 info@hach-lange.dk www.hach-lange.dk
HACH LANGE LDA
Av. do Forte nº8 Fracção M P-2790-072 Carnaxide Tel. +351 214 253 420 Fax +351 214 253 429 info@hach-lange.pt www.hach-lange.pt
HACH LANGE KFT.
Vöröskereszt utca. 8-10. H-1222 Budapest XXII. ker.
Tel. +36 1 225 7783 Fax +36 1 225 7784 info@hach-lange.hu www.hach-lange.hu
HACH LANGE AB
Vinthundsvägen 159A SE-128 62 Sköndal Tel. +46 (0)8 7 98 05 00 Fax +46 (0)8 7 98 05 30 info@hach-lange.se www.hach-lange.se
HACH LANGE SP. ZO.O.
ul. Krakowska 119 PL-50-428 Wrocław Tel. +48 801 022 442 Zamówienia: +48 717 177 707 Doradztwo: +48 717 177 777 Fax +48 717 177 778 info@hach-lange.pl www.hach-lange.pl
HACH LANGE S.R.L.
Str. Căminului nr. 3, et. 1, ap. 1, Sector 2 RO-021741 Bucureşti Tel. +40 (0) 21 205 30 03 Fax +40 (0) 21 205 30 17 info@hach-lange.ro www.hach-lange.ro
HACH LANGE S.R.L.
Via Rossini, 1/A I-20020 Lainate (MI) Tel. +39 02 93 575 400 Fax +39 02 93 575 401 info@hach-lange.it www.hach-lange.it
HACH LANGE S.R.O.
Zastrčená 1278/8 CZ-141 00 Praha 4 - Chodov Tel. +420 272 12 45 45 Fax +420 272 12 45 46 info@hach-lange.cz www.hach-lange.cz
HACH LANGE
8, Kr. Sarafov str. BG-1164 Sofia Tel. +359 (0)2 963 44 54 Fax +359 (0)2 866 15 26 info@hach-lange.bg www.hach-lange.bg
HACH LANGE S.L.U.
Edificio Seminario C/Larrauri, 1C- 2ª Pl. E-48160 Derio/Vizcaya Tel. +34 94 657 33 88 Fax +34 94 657 33 97 info@hach-lange.es www.hach-lange.es
HACH LANGE S.R.O.
Roľnícka 21 SK-831 07 Bratislava – Vaj nory Tel. +421 (0)2 4820 9091 Fax +421 (0)2 4820 9093 info@hach-lange.sk www.hach-lange.sk
HACH LANGE SU ANALİZ SİSTEMLERİ LTD.ŞTİ.
Ilkbahar mah. Galip Erdem Cad. 616 Sok. No:9 TR-Oran-Çankaya/ANKARA Tel. +90312 490 83 00 Fax +90312 491 99 03 bilgi@hach-lange.com.tr www.hach-lange.com.tr
35
Page 36
Contact
HACH LANGE D.O.O.
Fajfarjeva 15 SI-1230 Domžale Tel. +386 (0)59 051 000 Fax +386 (0)59 051 010 info@hach-lange.si www.hach-lange.si
HACH LANGE OOO
Finlyandsky prospekt, 4A Business Zentrum “Petrovsky fort”, R.803 RU-194044, Sankt-Petersburg Tel. +7 (812) 458 56 00 Fax. +7 (812) 458 56 00 info.russia@hach-lange.com www.hach-lange.com
ΗΑCH LANGE E.Π.Ε.
Αυλίδος 27 GR-115 27 Αθήνα Τηλ . +30 210 7777038 Fax +30 210 7777976 info@hach-lange.gr www.hach-lange.gr
HACH LANGE D.O.O.
Ivana Severa bb HR-42 000 Varaždin Tel. +385 (0) 42 305 086 Fax +385 (0) 42 305 087 info@hach-lange.hr www.hach-lange.hr
HACH LANGE MAROC SARLAU
Villa 14 – Rue 2 Casa Plaisance Quartier Racine Extension MA-Casablanca 20000 Tél. +212 (0)522 97 95 75 Fax +212 (0)522 36 89 34 info-maroc@hach-lange.com www.hach-lange.ma
36
Page 37
Appendix A Modbus register
Table 7 Sensor Modbus register
Tag name Register # Data type Length R/W Description
SLUDGELEVEL m 40001 Float 2 R Measured sludge level in m SLUDGELEVEL ft 40003 Float 2 R Measured sludge level in ft SLUDGEHEIGHT m 40005 Float 2 R Measured sludge height in m SLUDGEHEIGHT ft 40007 Float 2 R Measured sludge height in ft PLUNGERDEPTH m 40009 Float 2 R/W Plunger depth in m PLUNGERDEPTH ft 40011 Float 2 R/W Plunger depth in ft BOTTOM m 40013 Float 2 R/W Tank depth in m BOTTOM ft 40015 Float 2 R/W Tank depth in ft
SET PARAMETER 40017 Unsigned integer 1 R/W
MEAS UNITS 40018 Unsigned integer 1 R/W Dimension set: meters, feet
FACTOR 40019 Float 2 R/W
WIPE 40021 Unsigned integer 1 R/W Wiper status ERROR 40022 String 8 R Error shown EDITED NAME 40022 String 8 R/W Name of the measuring location CLEAN. INTERVAL 40030 Unsigned integer 1 R/W Wiper interval RESPONSE TIME 40031 Unsigned integer 1 R/W Response time: 10–600 seconds FADE-OUT 40032 Unsigned integer 1 R/W Blanking: ON/OFF BEGIN m 40033 Float 2 R/W Blanking start in m BEGIN ft 40035 Float 2 R/W Blanking start in ft END m 40037 Float 2 R/W Blanking end in m END ft 40039 Float 2 R/W Blanking end in ft LOGGER INTERVAL 40041 Unsigned integer 1 R/W Logger interval THRESHOLD AUTO 40042 Unsigned integer 1 R/W Automatic threshold function: ON/OFF THRESHOLD 40043 Float 2 R/W Threshold value (manual): 0.1–50 WINDOW 40045 Unsigned integer 1 R/W Window PROFILE COUNTER 40046 Unsigned integer 1 R/W Wiper profile counter SERIAL NUMBER 40047 String 6 R Serial number TEST / MAINT 40053 Time2 2 R/W Date of the last maintenance PROGRAM 40055 Float 2 R Application version BOOTPROG. 40057 Float 2 R Boot loader version STRUCTURE 40059 Unsigned integer 1 R Structure driver version FIRMWARE 40060 Unsigned integer 1 R Firmware driver version CONTENT 40061 Unsigned integer 1 R Register driver version FormatMinSL m 40062 Float 2 R Lower sludge level limit in m FormatMaxSL m 40064 Float 2 R Upper sludge level limit in m FormatMinSL ft 40066 Float 2 R Lower sludge level limit in ft FormatMaxSL ft 40068 Float 2 R Upper sludge level limit in ft FormatMinSH m 40070 Float 2 R Lower sludge height limit in m FormatMaxSH m 40072 Float 2 R Upper sludge height limit in m FormatMinSH ft 40074 Float 2 R Lower sludge height limit in ft FormatMaxSH ft 40076 Float 2 R Upper sludge height limit in ft MOIST 40078 Unsigned integer 1 R Moisture signal
Type of measurement: sludge level, sludge height
Correction factor for the measured value:
0.9–1.1
37
Page 38
Table 7 Sensor Modbus register
Tag name Register # Data type Length R/W Description
TEMPERATURE 40079 Integer 1 R Temperature signal in °C SENSOR ANGLE 40080 Unsigned integer 1 R Probe position signal in degrees FREQUENCY 40081 Integer 1 R Resonance frequency signal in hertz AMPL DIAG 40082 Integer 1 R Resonance voltage signal in volts
38
Loading...