Hach-Lange Sigma SD900 User Manual

Page 1
Catalog Number DOC026.53.00742
Sigma SD900 Portable Sampler
USER MANUAL
June 2007, Edition 1
© HACH Company, 2007. All rights reserved. Printed in the U.S.A.
Page 2
Page 3

Table of Contents

Section 1 Specifications.................................................................................................................... 5
1.2 Bottle and retainer configurations ................................................................................................ 9
Section 2 General Information....................................................................................................... 11
2.1 Safety information...................................................................................................................... 11
2.1.1 Use of hazard information................................................................................................. 11
2.1.2 Precautionary labels ......................................................................................................... 11
2.1.3 Confined space precautions ............................................................................................. 12
2.2 Sampler overview ...................................................................................................................... 13
2.2.1 Sampler components........................................................................................................ 13
Section 3 Installation........................................................................................................................ 15
3.1 Unpack the sampler ...................................................................................................................15
3.2 Installation guidelines................................................................................................................. 17
3.2.1 Manhole installation .......................................................................................................... 19
3.3 Sampler base preparation.......................................................................................................... 19
3.3.1 Clean the sample bottles .................................................................................................. 19
3.3.2 Single bottle installation .................................................................................................... 20
3.3.2.1 Full bottle shut-off installation................................................................................... 20
3.3.3 Multiple bottle installation .................................................................................................. 21
3.3.3.1 Bottle number 1 position .......................................................................................... 21
3.3.3.2 Two or four bottle installation ................................................................................... 22
3.3.3.3 Eight bottle installation ............................................................................................. 23
3.3.3.4 24 bottle installation ................................................................................................. 24
3.3.3.5 Distributor installation............................................................................................... 26
3.4 Intake tubing and strainer installation ........................................................................................ 27
3.5 Electrical installation .................................................................................................................. 28
3.5.1 Power installation .............................................................................................................. 29
3.5.1.1 Battery installation.................................................................................................... 29
3.5.1.2 AC power installation ............................................................................................... 30
3.5.2 Flow meter installation ...................................................................................................... 31
3.5.3 Communications installation ............................................................................................. 31
Section 4 Sampler operation.......................................................................................................... 33
4.1 Power on and power off ............................................................................................................. 33
4.2 Controller overview.................................................................................................................... 33
4.2.1 Keypad description ........................................................................................................... 33
4.2.2 Navigation.........................................................................................................................34
4.2.3 Parameter selection.......................................................................................................... 34
4.3 Main menu overview.................................................................................................................. 35
4.4 Sampler programs ..................................................................................................................... 35
4.4.1 Sampler program overview............................................................................................... 36
4.4.2 Create a sampler program ................................................................................................ 37
4.4.2.1 Tips and techniques ................................................................................................. 37
4.4.3 Modify a program .............................................................................................................. 38
4.4.4 Review a program............................................................................................................. 38
4.4.5 Restore default settings .................................................................................................... 39
4.4.6 Store programs as presets................................................................................................ 39
4.4.6.1 Load a stored program............................................................................................. 39
4.5 Start or stop a program.............................................................................................................. 40
1
Page 4
Table of Contents
4.6 Manual operation........................................................................................................................40
4.6.1 Collect grab samples .........................................................................................................40
4.6.2 Move the distributor arm....................................................................................................41
4.6.3 Start or stop the pump.......................................................................................................41
4.7 View data....................................................................................................................................41
4.7.1 Status screen.....................................................................................................................41
4.7.2 Status for main program ....................................................................................................42
4.7.2.1 Ready to start ...........................................................................................................42
4.7.2.2 Running/halted .........................................................................................................42
4.7.2.3 Complete ..................................................................................................................42
4.7.3 Status for stormwater program..........................................................................................42
4.7.4 Sample history...................................................................................................................43
4.7.5 Event log............................................................................................................................43
4.7.5.1 View the event log ....................................................................................................44
4.7.5.2 Erase the event log...................................................................................................44
4.8 Volume calibration......................................................................................................................44
4.8.1 Volume calibration using the liquid sensor ........................................................................45
4.8.1.1 Reset the calibration.................................................................................................45
4.8.2 Volume calibration based on time .....................................................................................46
4.8.3 Verify sample volume........................................................................................................47
4.9 Controller settings ......................................................................................................................47
4.9.1 Liquid sensor calibration....................................................................................................48
4.9.2 Set password.....................................................................................................................48
4.9.3 Tubing life indicator ...........................................................................................................49
4.9.3.1 Enable the tubing life indicator .................................................................................49
4.9.3.2 Tubing life status ......................................................................................................49
4.9.3.3 Tubing life cycle limit ................................................................................................49
4.9.3.4 Reset tubing cycles ..................................................................................................50
Section 5 Advanced sampling........................................................................................................51
5.1 Advanced sampling overview.....................................................................................................51
5.2 Advanced sampling menu..........................................................................................................51
5.3 Advanced sampling instructions.................................................................................................51
5.3.1 Send output signal at program completion ........................................................................52
5.3.2 Send output signal with sample cycle (special output) ......................................................52
5.3.3 Operate from external signal (setpoint sample).................................................................53
5.3.4 Set multiple start and stop times .......................................................................................53
5.3.5 Collect first flush stormwater samples ...............................................................................54
5.3.6 Set variable intervals.........................................................................................................55
Section 6 Maintenance.....................................................................................................................57
6.1 Upgrades, Repairs, General Maintenance .................................................................................57
6.2 Sampler cleaning........................................................................................................................57
6.3 Pump maintenance ....................................................................................................................57
6.3.1 Pump tube life....................................................................................................................57
6.3.2 Pump tube replacement ....................................................................................................58
6.3.3 Rotor removal and cleaning...............................................................................................60
6.4 Distributor arm tubing replacement ............................................................................................62
6.5 Desiccant replacement...............................................................................................................62
Section 7 Troubleshooting ..............................................................................................................65
7.1 General troubleshooting .............................................................................................................65
7.2 Error messages and notifications ...............................................................................................66
2
Page 5
Table of Contents
7.3 Diagnostic tests.......................................................................................................................... 67
7.3.1 Distributor diagnostic ........................................................................................................ 67
7.3.2 Keypad diagnostic............................................................................................................. 68
7.3.3 LCD diagnostic.................................................................................................................. 68
7.3.4 Liquid sensor diagnostic ...................................................................................................68
Section 8 Replacement Parts and Accessories......................................................................... 69
8.1 Bottle kits ................................................................................................................................... 69
8.2 Containers and bottle sets......................................................................................................... 69
8.3 Replacement parts..................................................................................................................... 69
8.4 Accessories................................................................................................................................ 70
Section 9 Contact information....................................................................................................... 73
Appendix A Wiring for non-Sigma flow meters ......................................................................... 75
Index...................................................................................................................................................... 79
3
Page 6
Table of Contents
4
Page 7

Section 1 Specifications

Specifications are subject to change without notice.
Sampler body dimensions (see Figure 1 on page 8)
Diameter: 50.5 cm (19.9 in.)
Standard base
Compact base
Composite base
Sample bottle capacity (see Figure 2 on page 9)
Standard base
Compact base
Height: 69.4 cm (27.3 in.) Weight: with (24) 1-L polyethylene bottles—15 kg (35.6 lb);
weight with (1) 2.5-gal polyethylene container—14.8 kg (32.6 lb)
Diameter: 44.1 cm (17.4 in.) Height: 61 cm (24 in.) Weight: with (24) 575-mL polyethylene bottles— 12.2 kg (27 lb);
weight with (1) 2.5-gal polyethylene container—12.9 kg (28.3 lb)
Diameter: 50.28 cm (19.8 in.) Height: 79.75 cm (31.4 in.) Weight: with (12) 950-mL glass bottles—15 kg (36 lb)
(24) 1 L polyethylene and/or 350-mL glass bottles (8) 2.3 L (0.6 gal) polyethylene and/or 1.9 L (0.5 gal) glass bottles (4) 3.8 L (1 gal) polyethylene and/or (4) 3.8 L (1 gal) glass bottles (2) 3.8 L (1 gal) polyethylene and/or (2) 3.8 L (1 gal) glass bottles (1) 21 L (5.5 gal) polyethylene composite container or (1) 15 L (4 gal)
polyethylene composite container or (1) 20 L (5.25 gal) polyethylene or (1) 10 L (2.5 gal) polyethylene or (1) 10 L (2.5 gal) glass
(24) 575 mL polyethylene bottles (8) 950 mL glass bottles (1) 10 L (2.5 gal) polyethylene bottle (1) 10 L (2.5 gal) glass bottle
Composite base (1) 21 L (5.5 gal) polyethylene bottle
Sampler
Sampler housing
Sample temperature range 0–60°C (32–140 °F)
Strainers
Sample intake tubing 9.5 mm (
SD900 controller
Enclosure
Power requirements 12 VDC supplied by optional AC power converter or battery
Overload protection 6 amp DC line fuse for pump
Pump Peristaltic high speed, with spring-mounted rollers
Pump rollers Nylatron,
Pump enclosure
Pump tubing 9.5 mm I.D. x 15.9 O.D. mm (3/8 in. x5/8 in.) silicone
Impact-resistant ABS, 3-section construction. Double-walled base with 2.54 cm (1 in.) insulation—direct bottle contact with ice.
316 stainless steel in standard size, high velocity or low profile for shallow depth applications and Teflon
3
/8 in.) I.D. vinyl or Teflon®-lined polyethylene
High-Impact, injection-molded PC/ABS blend; submersible, watertight, dust-tight, corrosion, and ice resistant; NEMA 4X, 6, IP 67
impact/corrosion resistant
Track is injection-molded polyphenylene sulfide. Cover is polycarbonate, high impact resistant. Pump enclosure rated IP37.
®
/316 stainless steel in standard size
5
Page 8
Specifications
20,000 sample cycles under the following conditions:
1 L sample volume
1 rinse
Pump tubing life
Tubing replacement time < 1 minute using pre-cut pump tube
Maximum vertical lift to draw sample
Pump flow rate 1.25 gpm (4.8 L/min) at 3 ft (1 m) vertical lift using
Typical sample volume repeatability
Typical sample volume accuracy
Typical transfer velocity
Liquid sensor Ultrasonic
6 minute pacing interval
16 ft of
3
/8 in. intake tube
15 ft of vertical lift
70 °F sample temperature
Minimum of 28 ft, using 29 ft of
3
/8-in. vinyl intake tube at sea level at 20–25 °C
(68–77 °F)
± 5% of 200 mL sample volume using uncalibrated liquid detect with 15 feet vertical lift,16 feet of
3
/8-in. vinyl intake tube configured for a single bottle using full
bottle shut-off at room temperature and 5000 ft elevation ± 10% of 200 mL sample volume using uncalibrated liquid detect with 15 feet
3
vertical lift,16 feet of
/8-in. vinyl intake tube configured for a single bottle using full
bottle shut-off at room temperature and 5000 ft elevation
2.9 ft/s (0.9 m/s) with 15 ft (4.6 m) vertical lift, 16 ft of 70 °F (21 °C) and 5000 ft elevation
3
/8-in. intake tube
3
/8-in. vinyl intake tubing,
Liquid sensor body Ultem® NSF ANSI standard 51 approved, USP Class VI compliant
Internal battery Lithium
Internal clock Indicates real time and date
Storage temperature –30 to 60 °C (–22 to 140 °F)
Operating temperature 0 to 50 °C (32 to 122 °F)
Storage/operating humidity 100% condensing
Graphics display
Status display
Graphic dot matrix, 128 x 64 pixel with LED backlight. Self prompting, menu-driven program.
Indicates the number of samples collected, the number of missed samples, inhibit mode, bottle position, time or counts to next sample and battery voltage
Multiple bottle mode: after complete revolution of distributor arm (unless
Automatic shutdown
Continuous Mode is selected). Composite mode: after preset number of samples have been delivered to
composite container, from 1 to 999 samples, or upon full container.
User interface
Embossed keypad with one power key, four function keys, and eight navigation keys; LED indicator
Stores up to 255 entries for sample time stamp, bottle number and sample status
Sample history
(success, bottle full, rinse error, user abort, distributor error, pump fault, purge fail, sample timeout, power fail and low main battery)
Records Power On, Power Fail, Firmware Updated, Pump Fault, Distributor Arm
Event log
Error, Low Memory Battery, Low Main Battery, User On, User Off, Program Started, Program Resumed, Program Halted, Program Completed, Grab Sample, Tube Change Required
Connections Power, auxiliary, serial communications and distributor
3
Fittings Barbed fittings for
Wetted materials
Typical materials in contact with sample: stainless steel, PE, Teflon, Ultem, Silicon or approved materials that can be tested for leaching properties
/8-in. I.D. flexible tubing
Weight 4.2 kg (9 lb, 5 oz)
6
Page 9
Specifications
Dimensions 10-3/8 in. (26.4 cm) L x 11-½ in. (29.2 cm) W x 6-¾ in. (17.1 cm) H
Programming features
Password protection 6-character; protect changes to program and system settings
Multiple programs Stores up to three sampling programs
Cascade programs
Synchronized sampling Ability to simultaneously take two samples with input from a single flow meter
Sample volume Programmed in 10-mL increments from 100 to 10,000 mL
Air purge
Intake rinse option Option to rinse intake line with source liquid prior to each sample, 1 to 3 rinses.
Sample distribution Composite, samples per bottle or bottles per sample.
Set point sampling Ability to start and/or stop a sample program based on an external trigger.
User start/stop times Up to 12 user-defined start/stop times/dates, with option to restart at position 1.
Storm water program Ability to run time-based, first flush program in parallel with main sample program.
Two samplers used in combination. The second sampler is initiated after the first sampler completes the program.
Air purged automatically before and after each sample; duration automatically compensates for varying intake line lengths.
Current status Display parameters relevant to main or storm water programs.
Units of measure Volume: gallons or mL; length: feet (ft) or cm
Sample retries
Manual grab sample Ability to manually deliver a grab sample to a specific bottle location
Run modes Continuous or non-continuous with user-entered number of samples.
Time pacing Uniform or variable time intervals.
Flow pacing Uniform or variable flow intervals.
Auxiliary connector
Program delay
Communication
Firmware updates Ability to perform field upgrades using Sample View software
Serial interface
Option to repeat sample collection cycle from 1 to 3 times if sample not obtained on initial attempt.
Power to Sigma 9XX, SD900, flow pulse input, external inhibit, special output, bottle number output and program complete output.
Two formats: 1) 1–9,999 flow pulses (in one unit increments);
2) Programmable start time/date
RS232 compatible; allows on-site collection of stored data including event log and sample history. Ability to configure remotely. Supports Modbus for SCADA connectivity.
7
Page 10
Specifications

1.1 Dimensions

Figure 1 Portable sampler dimensions
1 Portable sampler with compact base 3 Portable sampler with composite base 2 Portable sampler with standard base
8
Page 11

1.2 Bottle and retainer configurations

with
8582
8584
8580
2347
1502
2189
2190
[18.75 in]
ASSEMBLY
ARM (8583)
and 476.3 mm
DISTRIBUTOR
with
[16.25 in]
ASSEMBLY
ARM (8585)
and 412.8 mm
DISTRIBUTOR
with
[9.75 in]
ASSEMBLY
ARM (8581)
and 247.7 mm
DISTRIBUTOR
RETAINER
POSITIONER/
SUPPORT
CONTAINER
RETAINER
RETAINER
TUBE (8578)
TUBE (8579)
TUBE (8577)
Specifications
1422
RETAINER
8996
RETAINER\
8561
STANDARD
8976
STANDARD
8975
COMPACT
FULL
CONTAINER
FOR 5.5 GAL.
INSULATED BASE
BASE
INSULATED
BASE
INSULATED
SHUT-OFF
POLYETHYLENE
with
247.7 mm [9.75 in]
CONTAINER
TUBE (8998)
8925SD
SD900 PORTABLE
WASTEWATER SAMPLER
WITH COVER (8890)
1918 2.5 GAL.
POLYETHYLENE CONTAINER
6559 2.5 GAL.
GLASS CONTAINER
1369 (24) 575 mL
POLYETHYLENE BOTTLES
2348 (8) 950 mL
GLASS BOTTLES
1367 4 GAL.
POLYETHYLENE CONTAINER
6498 5.5 GAL.
POLYETHYLENE CONTAINER
6494 5.5 GAL.
POLYETHYLENE CONTAINER
2215 (2) 1 GAL.
Figure 2 Bottle and retainer configurations
POLYETHYLENE BOTTLES
2214 (2) 1 GAL.
GLASS BOTTLES
2217 (4) 1 GAL.
POLYETHYLENE BOTTLES
2216 (4) 1 GAL.
GLASS BOTTLES
657 (8) 2.3 LITER
POLYETHYLENE BOTTLES
1118 (8) 1.9 LITER
GLASS BOTTLES
737 (24) 1 LITER
POLYETHYLENE BOTTLES
732 (24) 350 mL
GLASS BOTTLES
9
Page 12
Specifications
10
Page 13

Section 2 General Information

2.1 Safety information

Please read this entire manual before unpacking, setting up, or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
To make sure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this manual.

2.1.1 Use of hazard information

DANGER Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION Indicates a potentially hazardous situation that may result in minor or moderate injury.

2.1.2 Precautionary labels

This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or electrocution exists.
This symbol, if noted on the product, indicates the need for protective eye wear.
This symbol, when noted on the product, identifies the location of the connection for Protective Earth (ground).
Important Note: Information that requires special emphasis.
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol, if noted on the instrument, will be included with a danger or caution statement in the manual.
This symbol, when noted on the product, identifies the location of a fuse or current limiting device.
This symbol, when noted on the product, indicated the presence of devices sensitive to Electro-static Discharge (ESD) and indicated that care must be taken to prevent damage with the equipment.
This symbol, if noted on the product, indicates a pinch hazard. Keep hands and fingers clear.
11
Page 14
General Information

2.1.3 Confined space precautions

Important Note: The following information is provided to guide users of Sigma SD900 Portable Samplers on the dangers and risks associated with entry into confined spaces.
On April 15, 1993, OSHA's final ruling on CFR 1910.146, Permit Required Confined Spaces, became law. This new standard directly affects more than 250,000 industrial sites in the U.S.A. and was created to protect the health and safety of workers in confined spaces.
Definition of a confined space
A confined space is any location or enclosure that presents or has the immediate potential to present one or more of the following conditions:
An atmosphere with less than 19.5% or greater than 23.5%
oxygen and/or more than 10 ppm Hydrogen Sulfide (H
An atmosphere that may be flammable or explosive due to
gases, vapors, mists, dusts or fibers.
Toxic materials which upon contact or inhalation, could result in
injury, impairment of health or death.
2
S).
Confined spaces are not designed for human occupancy. They have restricted entry and contain known or potential hazards. Examples of confined spaces include manholes, stacks, pipes, vats, switch vaults, and other similar locations.
Standard safety procedures must always be followed prior to entry into confined spaces and/or locations where hazardous gases, vapors, mists, dusts or fibers may be present. Before entering any confined space check with your employer for procedures related to confined space entry.
12
Page 15

2.2 Sampler overview

2.2.1 Sampler components

General Information
DANGER This sampler is designed for collection of aqueous samples only. Non-aqueous samples may damage the equipment and could result in fire and chemical hazards.
The SD900 sampler automatically collects and preserves liquid samples. The sampler is suitable for collection of conventional and toxic pollutants and suspended solids.
The sampler consists of three main sections (Figure 3). The center section contains the controller. Use the controller to program the sampler and for manual operation. The controller contains the following components:
Pump—operates in the forward or reverse direction to
collect a sample, rinse and purge the intake tube.
Liquid sensor—allows the sampler to dispense accurate
sample volumes into the sample bottle(s). Can be calibrated to specific sample conditions in the field.
Cable connectors—for power, flow meter or
communication.
Desiccant—absorbs internal moisture in the controller and
prevents corrosion.
Figure 3 Sampler components
1 Bottle/base section 3 Power source 5 Controller 2 Center section 4 Top cover
13
Page 16
General Information
14
Page 17

Section 3 Installation

3.1 Unpack the sampler

CAUTION Only qualified personnel should conduct the tasks described in this section of the manual.
WARNING Explosion hazard. This product is not designed for hazardous locations where combustible environments may exist.
The sampler can be set up for single-bottle or multiple-bottle collection. The components for each configuration will vary (see
Figure 4 for single-bottle or Figure 5 on page 17 for multiple-bottle
collection).
Check the sampler packages for the following components:
Sampler—includes cover, base and midsection with controller
and distributor or full bottle shut-off
Bottle(s)—1, 2, 4, 8 or 24
Power source—battery or AC power supply
Pump tubing—replacement
Sample intake tubing—vinyl or Teflon-lined
Silicone grease packet—for desiccant maintenance
Strainer—stainless steel or Teflon/stainless steel
User documentation
15
Page 18
Installation
Figure 4 Sampler with single bottle
1 Sampler base 7 Top cover (Cat. No. 8890) 2 Container support (Cat. No. 1502) 3 Bottle 9 Intake tubing, vinyl or teflon-lined 4 Full bottle shut-off (Cat. No. 8996) 10 AC power supply (optional) 5 Sampler center section (Cat. No. 8922) 11 Battery recharger (optional) 6 Battery (Cat. No. 8754400, optional)
1
The container support for a 2.5-gal bottle with the standard base is shown. Other bottle sizes may not use a container support.
1
8 Strainer
16
Page 19
Installation
Figure 5 Sampler with multiple bottles
1 Sampler base 7 Intake tubing, vinyl or teflon-lined 2 Sample bottles 8 Strainer 3 Retainer 9 Distributor assembly 4 Sampler center section 10 AC power supply (optional) 5 Battery (Cat. No. 8754400, optional) 11 Battery recharger (optional) 6 Top cover (Cat. No. 8890)

3.2 Installation guidelines

WARNING Explosion hazard. This product is not designed for hazardous locations where combustible environments may exist.
Refer to the following guidelines and Figure 6 when evaluating site location.
If the site is located in a confined space such as a manhole,
refer to section 2.1.3 on page 12 for safety information.
Make sure that the site has a level surface or a place to hang
the suspension harness, support bracket or spanner bar (Figure 6).
17
Page 20
Installation
Make sure that the temperature at the site is within the
operating temperature range that is specified for the sampler.
Make sure that the intake tube is as short as possible.
Maximize the vertical slope of the intake tube from the sampler
to the sample source so that the tube drains completely. This prevents cross-contamination of samples and freezing the tube.
Note: See Specifications on page 5 for limitations on transport velocity and maximum vertical lift.
Note: If site conditions do not permit the intake tube to have a downward slope, or if the line is pressurized, disable the liquid sensor and calibrate the sample volume (section 4.8.2 on page 46).
Install the strainer in the middle of the sample stream (not near
the surface or bottom) to make sure that a representative sample is collected.
Figure 6 Mounting configuration
1 Strainer 4 Vertical lift 2 Intake tubing 5 Mounting surface 3 Suspension harness
18
Page 21

3.2.1 Manhole installation

Installation
WARNING Explosion hazard. This product is not designed for hazardous locations where combustible environments may exist.
Installation in a manhole requires suspending the sampler above the sample water. Refer to section 2.1.3 on page 12 for safety information in confined spaces.
Prerequisites:
Use the following hardware to mount the sampler in a manhole.
Spanner bar or support bracket. A spanner bar is placed inside
the manhole and is supported by pressure against the walls. A support bracket has the same width as the manhole cover and is placed directly under the cover for support.
Suspension harness
Note: See Accessories on page 70 for ordering information.
Installation procedure
Complete the following steps to mount the sampler in a manhole.

3.3 Sampler base preparation

3.3.1 Clean the sample bottles

1. Connect the suspension harness to the latches on the
sampler cover.
2. Hang the suspension harness on the spanner bar or
support bracket.
3. Mount the spanner bar or support bracket in the manhole.
CAUTION Biological sample hazard. Follow safe handling protocols during contact with sample bottles and sampler components. Disconnect the sampler from power to disable the pump before handling.
The sampler is set up at the factory with a single bottle or multiple bottle configuration. To change the bottle configuration:
Use Figure 2 on page 9 to determine the components that
are required for the selected configuration.
Install the full bottle shut-off device (section 3.3.2.1 on page 20)
or distributor assembly (section 3.3.3.5 on page 26).
Clean the sample bottles and caps using a brush and water with a mild detergent before setting up the sampler. Rinse the containers with fresh water followed by a distilled water rinse. Glass bottles may also be autoclaved.
19
Page 22
Installation

3.3.2 Single bottle installation

Use a single bottle when one composite sample is needed. A full bottle shut-off signals the controller when to stop sample collection. Refer to Figure 4 on page 16 for a diagram of required components.
Prerequisites:
One plastic or glass bottle
Retainer (Cat. No. 2190)
Note: Refer to Figure 2 on page 9 to make sure that the correct sample bottle is used with the correct sampler base and retainer.
Installation procedure:
1. Clean the sample bottle as described in section 3.3.1.
2. If using a 2.5-gallon bottle with the standard base, place a
container support (Cat. No. 1502) in the sampler base.
3. Place the sample bottle in the center of the base.
4. Fill the sampler base with ice after the bottle is in place to
preserve samples.
3.3.2.1 Full bottle shut-off installation
Note: The time that it takes the ice to melt varies with the amount of ice used, the external temperature and sample temperature.
5. Place the midsection over the base, making sure that the full
bottle shut-off is inside the neck of the bottle.
6. Program the sampler and install in the field.
The full bottle shut-off is typically installed at the factory and signals the controller when the bottle is full. Complete the following steps to replace or install a new full bottle shut-off.
Prerequisites:
Full bottle shut-off device
Installation procedure:
1. Slide the slot in the full bottle shut-off base plate under the
retainer pin inside the center section (Figure 7). Make sure that the tube and cable exit through the openings in the full bottle shut-off assembly.
2. Position the thumbscrew over the threaded hole and hand
tighten to hold the device in place.
20
3. Connect the cable on the full bottle shut-off to the bottom of the
controller (Figure 7). Turn to tighten.
4. Connect the free end of the full bottle shut-off tubing to the
inside tube fitting (Figure 7).
Page 23
Installation
Figure 7 Full bottle shut-off installation
1 Inside tube fitting 4 Slot in full bottle shut-off base plate 2 Threaded hole 5 Thumb screw on full bottle shut-off assembly 3 Retainer pin

3.3.3 Multiple bottle installation

Use multiple bottles to collect samples into separate bottles or into more than one bottle. A distributor positions the sample tube over each bottle. Position the bottles in the sampler base as shown in
section 3.3.3.1. Refer to Figure 5 on page 17 for a diagram of
required components.
3.3.3.1 Bottle number 1 position
Position the first sample bottle (number 1) under the label in the sampler base (see Figure 8). Place the remaining bottles in increasing number in the direction indicated by the label.
21
Page 24
Installation
Figure 8 Bottle number 1 location
1 Bottle #1 location for 24 bottles 6 Standard base 2 Bottle #1 location for 8 bottles 7 Bottle #1 3 Bottle #1 location for 2 or 4 bottles 8 Elastic straps 4 Bottle #1 location in compact base 9 Retainer 5 Compact base 10 Sample bottle
3.3.3.2 Two or four bottle installation
Use 2 or 4 bottles to collect samples into one-gallon bottles.
Prerequisites:
Two or four 1-gallon plastic or glass bottles
Retainer (Cat. No. 2190)
Note: Refer to Figure 2 on page 9 to make sure that the correct sample bottles are used with the correct sampler base and retainer.
Installation procedure:
1. Clean the sample bottles as described in section 3.3.1.
2. Place the bottles in the base as shown in Figure 9. Arrange the
bottles so that the first bottle is located under the label at the point marked 2 or 4 BOT (Figure 8 on page 22).
22
3. Place the retainer over the bottles with the knobs pointing up.
Secure with the straps (Figure 9).
Page 25
Installation
4. To keep samples cool, fill the sampler base with ice after the
bottles are in place.
Note: The time that it takes the ice to melt varies with the amount of ice used, the external temperature and sample temperature.
5. Place the midsection over the base, program the sampler and
install in the field.
Figure 9 Two and four bottle installation
1 Standard sampler base (Cat. No. 8976) 3 Retainer (Cat. No. 2190) 2 1-gallon plastic or glass bottles (2x) 4 1-gallon plastic or glass bottles (4x)
3.3.3.3 Eight bottle installation
Use 8 bottles to collect samples into one-liter or two-liter bottles.
Prerequisites:
8 plastic or glass bottles
Retainer (see Figure 2 on page 9 for catalog number)
Note: Refer to Figure 2 on page 9 to make sure that the correct sample bottles are used with the correct sampler base and retainer.
Installation procedure:
1. Clean the sample bottles as described in section 3.3.1.
2. Place the bottles in the base (see Figure 8 on page 22):
Compact base first bottle position: under the arrow label.
Standard base first bottle position: under the 8 BOT label.
3. Place the retainer over the bottles with the knobs pointing up.
Secure with the straps.
Note: The retainer for the compact base has two pieces. Place the larger piece in the bottom of the base.
23
Page 26
Installation
4. To keep samples cool, fill the sampler base with ice after the
bottles are in place.
Note: The time that it takes the ice to melt varies with the amount of ice used, the external temperature and sample temperature.
5. Place the midsection over the base, program the sampler and
install in the field.
Figure 10 Eight bottle installation
1 Standard sampler base (Cat. No. 8976) 5 950-mL glass bottles (8x) 2 1.9-L glass bottles or 2.3-L plastic bottles (8x) 6 Retainer bottom (Cat. No. 2347) 3 Retainer (Cat. No.1422) 7 Compact sampler base (Cat. No. 8975) 4 Retainer top (Cat. No.2347)
3.3.3.4 24 bottle installation
Use 24 bottles to collect samples into 350-mL, 575-mL or 1-liter bottles.
Prerequisites:
24 plastic or glass bottles
Retainer
Note: Refer to Figure 2 on page 9 to make sure that the correct sample bottles are used with the correct sampler base and retainer.
24
Page 27
Installation
Installation procedure:
1. Clean the sample bottles as described in section 3.3.1.
2. Place the bottles in the base:
Compact base: arrange the bottles so that the first bottle is
located under the arrow label (Figure 8 on page 22).
Standard base: arrange the bottles so that the first bottle is
located under the label marked 24 BOT (Figure 8 on
page 22).
3. Place the retainer over the bottles with the knobs pointing up.
Secure with the straps (Figure 11).
Note: If using 350-mL glass bottles with the standard base, place the bottles around the outside of the retainer.
4. To keep samples cool, fill the sampler base with ice after the
bottles are in place.
Note: The time that it takes the ice to melt varies with the amount of ice used, the external temperature and sample temperature.
5. Place the midsection over the base, program the sampler and
install in the field.
Figure 11 24 bottle installation
1 Standard sampler base (Cat. No. 8976) 5 Retainer (Cat. No. 1422) 2 350-mL glass bottles (24x) 6 Compact sampler base (Cat. No. 8975) 3 Retainer (Cat. No. 2189) 7 575-mL plastic bottles (24x) 4 1-L plastic bottles (24x)
25
Page 28
Installation
3.3.3.5 Distributor installation
CAUTION Pinch hazard. Make sure the sampler is powered off before removing or installing the distributor assembly.
The distributor automatically moves the sample tube over each bottle during multiple bottle sampling. The distributor assembly is typically installed at the factory. Complete the following steps to install a new or different assembly.
Prerequisites:
Distributor assembly—three assemblies are available. Refer to
Figure 2 on page 9 to make sure that the correct distributor
assembly is used.
Installation procedure:
1. Slide the two slots in the distributor assembly housing under
the retainer pins located on the top inside surface of the center section (Figure 12).
2. When fully seated, hand tighten the thumbscrew on the
distributor assembly to hold the distributor in place.
3. Push the distributor tubing on the sampler fitting on the top
inside surface of the center section (Figure 12).
4. To make sure that the arm has sufficient freedom of movement,
hand-rotate the arm in a circle (both clockwise and counterclockwise) until it hits the arm stop. If the arm does not rotate freely, reposition the tubing to the other side of the arm so that the arm can rotate.
Note: Do not force the arm past the arm stop. The arm stop prevents the arm from rotating more than 360 degrees and pinching the tubing.
5. Install the mid section on the base.
6. To make sure that the distributor is aligned properly, run the
manual distributor diagnostic (see section 7.3 on page 67).
26
Page 29
Installation
Figure 12 Distributor assembly installation
1 Thumbscrew 3 Retainer pins 2 Arm stop 4 Slots

3.4 Intake tubing and strainer installation

Place the intake tubing and strainer directly in the sample source to collect samples. Refer to section 3.2 on page 17 to make sure that samples are free from contamination and are representative of the sample source.
Prerequisites:
Vinyl or Teflon-lined tubing
Connection kit (Cat. No. 2186) (for Teflon-lined tubing only)
Strainer
Note: See Accessories on page 70 for ordering information.
Installation procedure:
1. Connect one end of the tubing to the liquid sensor fitting.
Note: Use Cat. No. 2186 (connection kit) if using Teflon-lined tubing.
2. Press the tubing into the strain relief (Figure 13).
3. Route the tubing through the sampler housing so that the
tubing is not pinched when the cover is on.
4. Connect the other end of the tubing to the strainer.
Note: Use Cat. No. 2186 (connection kit) if using Teflon-lined tubing.
27
Page 30
Installation
5. Place the intake tube and strainer in the main stream of the
sample source where the water is turbulent and well-mixed. Make sure that the intake tube does not have kinks or loops.
Figure 13 Intake tubing installation
1 Strainer 3 Strain relief 2 Intake tubing 4 Liquid sensor

3.5 Electrical installation

Connect cables to the controller as shown in Figure 14 for the following components:
Power (section 3.5.1)
Flow meter or other device (section 3.5.2 on page 31)
Communications (section 3.5.3 on page 31)
28
Page 31
Figure 14 Side view of controller—connectors
1 Power connector 3 Serial connector 2 Auxiliary connector 4 Distributor/full bottle shut-off connector
Installation

3.5.1 Power installation

3.5.1.1 Battery installation
The sampler can be powered by a battery (section 3.5.1.1) or by AC power (section 3.5.1.2).
Prerequisites:
12 VDC gel electrolyte (lead acid) battery plus recharger.
Note: A 2-pin to 3-pin adaptor cable (Cat. No. 8739400) may be used with older batteries containing a 2-pin connector to power the sampler.
Installation procedure:
1. Place the battery directly behind the controller (Figure 3 on
page 13).
2. Pull the rubber straps up and over the clips at each end of the
battery to secure it to the sampler body.
3. Connect the cable from the battery to the power connector on
the controller.
29
Page 32
Installation
3.5.1.2 AC power installation
Optional battery cable
If the external battery cable with bare leads (Cat. No. 2198) is used, the following requirements must be met:
The cable must not be altered to more than 3 meters long.
The cable can only be connected to a 12 VDC battery.
The maximum voltage cannot exceed 18 VDC.
Failure to observe proper polarity when connecting the
clips to the battery will result in a blown fuse in the cable.
WARNING Electric shock hazard. The temperature of the power supply will increase when the sampler pump is operating. If the time interval between sample cycles is too short, the power supply can become overheated and result in a electric shock hazard or power supply failure. Refer to the power supply instruction sheet (DOC306.53.00801) to determine the time that the pump must stay off between sample cycles to prevent overheating.
Prerequisites:
Important Note: Use only the AC power supply that is specified for this sampler. The power supply connector must have three pins. An adapter cannot be used to convert a older two-pin power supply for use with this sampler.
AC power supply (Cat. No. 8754500US)
Installation procedure:
1. Place the power supply directly behind the controller (Figure 3
on page 13).
2. Pull the rubber straps up and over the clips at each end of the
power supply to secure it to the sampler body.
3. Connect the cable from the power supply to the power
connector on the controller.
Electrical transients
When connected to an AC power source, the SD900 controller may be temporarily upset by strong electrical transients caused by sources such as lightning or large electrical motors. These upsets can result in a missed sample, but the SD900 controller will recover and continue with the sample program.
A missed sample will be recorded and can be viewed in the status, event log or sample history screens. Remedial actions may involve the use of a power line filter or connecting the controller to a different branch circuit.
30
Page 33

3.5.2 Flow meter installation

Installation
The sampler can be connected to a flow meter to start or stop sampling based on the flow volume.
Prerequisites:
Multi-purpose full cable for Sigma flow meters (or 980 half
cable for the model 980 flow meter). To connect to another brand of flow meter, use a multi-purpose half cable and follow the wiring instructions in Appendix A on page 75.
Optional splitter (Figure 15) to provide additional connections.
Two or more splitters may be connected in series.
Note: See Accessories on page 70 for ordering information.
Installation procedure:
1. Connect one end of the cable to the flow meter. For connection
to the model 980 flow meter, follow the instructions in the model 980 user manual.
2. Connect the other end of the cable to the controller.
Note: For connection to a flow meter with an existing 6-pin cable, use the auxiliary adapter cable (6-pin to 7-pin adapter)

3.5.3 Communications installation

Figure 15 Splitter
The sampler can be connected to a PC or Modbus device for data transfer or programming.
Prerequisites:
Serial cable (Figure 16)
Note: See Accessories on page 70 for ordering information.
Installation procedure:
1. Connect one end of the cable to the PC or Modbus device.
2. Connect the other end of the cable to the controller. The pin
assignments for the serial cable are detailed in Table 1.
31
Page 34
Installation
Figure 16 Serial connector cable
Table 1 Serial cable reference
7-pin connector assignment letter Signal description DB-9 pin assignment number
B Signal ground 5 DRCD3 FTXD2 G Earth ground
32
Page 35

Section 4 Sampler operation

4.1 Power on and power off

Important Note: The AC power supply will become overheated if the sampler pump is operated continuously. Refer to the power supply instruction sheet (DOC306.53.00801) to determine the time interval that must be allowed between sample cycles to prevent damage to the AC power supply and/or to prevent a burn hazard.

4.2 Controller overview

4.2.1 Keypad description

Power on: press
Power off: press
POWER (Figure 17). The LED will blink.
POWER and select YES. The LED will stay off.
Sleep mode: automatic after two minutes of inactivity. The LED will
blink but the display will be blank. Press any key to restore the display.
The keypad and description is detailed in Figure 17.
Figure 17 SD900 keypad
1 POWER: press to turn controller on or off 7 BACK: cancel or return to previous screen
VOLUME CALIBRATION: enter volume calibration menu 8 ARROW KEYS: moves cursor or scrolls through values
2
STOP: stops the pump or distributor if running 9 STATUS: displays current status of the sample program
3
LED: blinks when controller is powered on 10 RUN/HALT PROGRAM: starts or stops a sample program
4
MENU: displays the main menu 11 MANUAL OPERATION: manual operation of pump or
5
ENTER: selects the highlighted or entered value
6
distributor
33
Page 36
Sampler operation

4.2.2 Navigation

Use the controller to access all sampler operations. Use the
ARROW keys, ENTER key and BACK key to move from one screen
to another. An arrow on the display indicates that more screens are available (Figure 18).
Example:
1. Press the
2. Press the
Press
MAIN MENU key.
DOWN ARROW key to highlight diagnostics.
ENTER. The diagnostics menu will be shown.
3. Note the down arrow on the bottom of the display (Figure 18).
Press the down arrow until additional options are shown.
4. Press the
BACK key or MAIN MENU key to return to the
main menu.
Figure 18 Screen navigation
1 Left arrow 3 Up arrow 2 Right arrow 4 Down arrow

4.2.3 Parameter selection

Parameter selection is made in one of two ways (Figure 19):
Selection from a list
Enter a value using the arrow keys
Figure 19 Parameter selection
1 Select item from list 2 Enter value with arrows 3 Acceptable range
34
Page 37

4.3 Main menu overview

The SD900 sampler main menu contains four options for sampler operation, monitoring or data management. Each option is detailed in Table 2.
Table 2 SD900 sampler menu overview
Menu option Description
Program setup Create, review, or set up a sampling program
Modify all Create or edit a sampling program Modify selected Edit a portion of a sampling program Review Review a sampling program Presets Enter or select up to three program templates Restore defaults Restore the program default selections
Status Display the status of the current program
Diagnostics Review and manage data or test the operation of components
Event log Display or delete the event log Sample history Review the sampling history
Distrib diag
Keypad diag Test the operation of each key LCD diag Test the operation of the LCD display screen Liquid sensor diag Display calibration data for the liquid sensor
System setup Change controller settings or calibrate the liquid sensor
Test the ability of the distributor to detect each of the 24 bottle positions
Note: Only active when the sample program is configured for more than 1 bottle.
Sampler operation
Time/date setup Set the time (24 hour) and date
Communication
Setup base Specify whether the standard or compact base is used Language Select one of the available languages Liq sensor cal Calibrate the liquid sensor Set contrast Adjust the contrast of the LCD display screen Password setup Create or disable a password Tubing life Enable optional alarm to indicate when pump tubing needs to be changed
Set the baud rate (19200, 38400, 57600 or 115200) and protocol (Modbus RTU or ASCII) for the serial port

4.4 Sampler programs

Create sampler programs to operate the sampler automatically. Once programmed, install the sampler at the sampling site. Return to collect the samples at regular intervals or when the sampling program is complete. Sampler programs can be stored for later recall (4.4.6 on page 39).
35
Page 38
Sampler operation

4.4.1 Sampler program overview

Table 3 outlines the menu for creating a basic sampling program.
Table 3 Basic program setup
Menu option Description
Bottles Enter bottle information.
Bottle quantity Select the number of bottles in the sampler (1, 2, 4, 8, 12 or 24). Bottle volume Enter the volume capacity of each bottle in milliliters or gallons (0.5–99.8 gal or 50–65000 mL).
Intake tubing Enter intake tubing information.
Tubing length
Intake tube type Select the size and type of tubing (¼ in. vinyl, 3/8 in. vinyl or 3/8 in. teflon)
Program delay
Enable/disable Select enable to use a program delay or disable to omit it.
Date and time If enabled, enter the date and time when the program will start (24 hour format).
Counts
Enter the length of the intake tube from
100–3000 cm). An accurate length is necessary to obtain accurate sample volumes.
Delay starting the sampling program until specified time and day or until specified number of counts is reached.
If enabled, enter the counts from the flow meter when the program will start (1–9999 counts). If counts is selected and the sample pacing is later set to time, the program delay will be disabled.
the strainer to the liquid sensor (3–99 ft or
Sample pacing/collection Specify whether to collect samples at regular time intervals or regular flow volume.
Time based Collect samples at regular time intervals.
Pacing interval Enter the time interval, in hours and minutes (0:01–999:00). Take first sample Select whether the program will start immediately or after the first time interval has passed.
Flow based Collect samples at a specified flow volume (requires external flow meter)..
Take sample every Enter the flow interval that must expire between sample cycles, in counts (1–9999 counts). Override time Select enable to force a sample to be collected if the flow volume is unusually low.
Time
Take first sample Select whether the program will start immediately or after the first flow interval has passed.
Sample distribution For multiple bottle sampling, specify how the samples are delivered to the bottles.
Deliver samples to all Select whether each sample will be delivered to all bottles or not.
Yes Each sample will be delivered to all bottles.
End after last Program will stop after the last sample has been collected. Enter number of samples (1–999). Continuous Sampler will continue collecting samples until stopped manually.
No Samples will be delivered to a subset of bottles (see Figure 20 on page 38).
Samples/bottle Enter the number of samples to be collected in each bottle (1–999). Bottles/sample Enter the number of bottles that will contain the same sample.
Liquid sensor Enable/disabled. If disabled, the sample volume must be calibrated by time.
If enabled, enter the maximum time between samples (0:01–999:00). Timer is reset each time a sample is collected based on flow counts.
Sample volume
Intake rinses Enter the number of times the intake tube will be rinsed before each sample is collected (0–3).
Sample retries Enter the number of sample retries to perform when a failure occurs (0–3).
Site ID
Advanced sampling Access the advanced sampling menu.
Done Finished creating basic sampling program. Prompts user to start or cancel the program.
Enter the volume to be collected per sample. If bottles/sample mode is selected, each bottle receives a full sample volume (100–10,000 mL).
Enter a name for the sampling location (enter up to 12 characters). The site ID is used as the preset name if the program is saved.
36
Page 39

4.4.2 Create a sampler program

Sampler operation
Create a sampler program to set up the sampler for automatic sampling routines.
Procedure:
1. Select
2. Select
PROGRAM SETUP from the main menu.
MODIFY ALL. The first parameter, bottle quantity, will be
shown.
3. Select the number of bottles in the sampler. The next
parameter, bottle volume, will be shown.
4. Enter the volume of the individual bottle(s). Use the
RIGHT
ARROW keys to move the cursor left or right. Use the UP
AND DOWN
ARROW keys to change the value. Change the
LEFT AND
units if necessary (gal or mL) with the arrow keys.
5. Continue to select or enter values for each parameter until
complete. Refer to Table 3 on page 36 for a complete list of all available options for a basic sample program.
4.4.2.1 Tips and techniques
Refer to Table 4 and Figure 20 for help with sample programming.
Table 4 Tips and techniques for the basic sampling program
Parameter Tips and techniques
When 1 bottle is selected, the full bottle shut-off device is enabled and the distributor
Bottle quantity
Intake tubing
Program delay
Sample pacing—flow Each time a sample is collected based on flow counts, the override timer is reset.
is disabled. When more than 1 bottle is selected, the distributor diagnostic is enabled.
The sampler uses the intake tubing information to determine sample volume. If the intake tubing information is not accurate, the sample volume will not be accurate.
If the program delay is set to counts and sample pacing is set to time, the program delay will be disabled.
If the bottle quantity is 1, the sample distribution option will not be available. See Figure 20 for a description of samples per bottle vs. bottles per sample.
Sample distribution
Liquid sensor If the liquid sensor is disabled, the sample volume must be calibrated manually.
Sample volume
Bottles per sample—use bottles per sample when the sample volume is larger than one bottle can hold.
Samples per bottle—the distributor arm will not advance to the next bottle until the specified number of samples has been delivered to the current bottle.
The sample volume must be entered as mL (1 gal.= 3785.4 mL).
Make sure that the sample volume does not exceed the bottle volume. Sample volumes are rounded to the nearest 10 mL.
37
Page 40
Sampler operation
Figure 20 Two samples per bottle (left) vs. two bottles per sample (right)

4.4.3 Modify a program

Use the modify selected option to change an individual parameter for the currently loaded program.
Procedure:

4.4.4 Review a program

1. Select
2. The selected parameters are grouped within the following
subheadings. Select one of the available subheadings using the up and down arrow keys.
Bottles—quantity and volume
Intake tubing—length and type
Program delay
Sample pacing/collection—time or flow based
Sample distribution—for multiple bottle sampling
Liquid sensor—enable/disable
Sample volume—enter volume
Intake rinses—enter number
Sample retries—enter number
Site ID—enter ID
Advanced sampling
3. Change the parameter as shown in section 4.2.3 on page 34.
PROGRAM SETUP>MODIFY SELECTED.
38
Use the review option to view the selected parameters without making any changes to the program.
Procedure:
1. Select
2. Press
PROGRAM SETUP>REVIEW from the main menu.
ENTER to view each entry.
Page 41

4.4.5 Restore default settings

Sampler operation
Use the restore option to set all program parameters to the default settings.
Procedure:

4.4.6 Store programs as presets

4.4.6.1 Load a stored program
1. Select
2. Select
PROGRAM SETUP>RESTORE from the main menu.
YES to restore the factory default settings.
Use the presets option to store up to three sampler programs for later use. If any changes are made to a preset program after it is loaded, the program must be stored again to save the changes.
Procedure:
1. Select
2. The three available presets are displayed as
PROGRAM SETUP>PRESETS from the main menu.
P1, P2 and P3.
Select one of the presets.
Note: If a program is already stored, the site ID for that program will appear to the right of the preset number. If no preset program has been stored, the preset field will be empty.
3. Select STORE CURRENT to store the current program as
apreset.
Use the load program option to use a preset program as the current program.
Procedure:
1. Select
2. Select one of the presets (
3. Select
PROGRAM SETUP>PRESETS from the main menu.
P1, P2 or P3).
LOAD PROGRAM to load the selected preset as the
current program. The selected preset program will be loaded and the display will return to the main menu. The program can be modified or used as is.
If a preset program is loaded and then modified, the program must be stored again (store current) to save the changes.
39
Page 42
Sampler operation

4.5 Start or stop a program

Use the RUN/HALT PROGRAM key to start or stop a program.
Procedure:

4.6 Manual operation

1. Press
2. Select one of the available options:
Note: A program must be stopped before the program setup or system setup menus can be modified.
Use manual operation to collect a grab sample, move the distributor arm or operate the pump.
RUN/HALT PROGRAM to run, halt, resume or end a
sampler program.
•Start: starts the currently loaded program. The status
changes to running.
•Halt: stops the program temporarily. The status changes
to halted.
•Resume: if halted, the program resumes from the point at
which it was halted. The status changes to running.
•Start over: if halted, the program starts from the beginning.
The status changes to running.
End program: ends the currently running program. The
program status changes to complete.

4.6.1 Collect grab samples

Grab samples can be collected to verify sample volume or to collect samples without running a sampling program.
Procedure:
1. Press the
2. Select
3. Place the intake tube into the sample water. Remove the outlet
pump tube from the fitting on the sampler and place the tube into a sample container.
4. Enter the volume to be collected and press
Note: The grab sample volume can only be modified if the liquid sensor is enabled. if the liquid sensor is disabled, the grab sample volume will reflect the volume defined using timed calibration.
5. The pump will purge the intake tube and then collect the
specified volume of sample. The pump will then purge the intake tube.
To stop the pump at any time during the sample cycle, press the
MANUAL OPERATION key.
GRAB SAMPLE.
ENTER.
STOP key.
40
Page 43

4.6.2 Move the distributor arm

Sampler operation
When configured for multiple bottles, the distributor arm can be moved manually to distribute a grab sample to a specific bottle.
Procedure:

4.6.3 Start or stop the pump

1. Press the
2. Select
3. The current bottle number will be displayed. Enter the bottle
number over which to position the distributor arm. Press
4. The distributor arm will move to the selected bottle. To stop the
distributor arm at any time, press the
The pump can be operated outside of a sampling program to collect a sample or purge the intake tubing.
Procedure:
1. Press the
2. Select
3. Select the direction to run the pump:
Run forward: pump operates in the forward direction to
Run backward: pump operates in the reverse direction to
MANUAL OPERATION key.
MOVE DISTRIB.
ENTER.
STOP key.
MANUAL OPERATION key.
OPERATE PUMP.
collect a sample and deposit into a sample container.
purge the intake line.

4.7 View data

4.7.1 Status screen

4. The screen will display pumping if the forward direction was
selected or purging if the reverse direction was selected.
5. The pump will operate until the
Data can be viewed from one of the status screens or from the diagnostics menu.
The Status screen can be accessed by pressing the STATUS key or by selecting
The status menu will be shown if the main and stormwater programs run simultaneously. The status of either main program or the stormwater program can be viewed by selecting
STORM from the status menu.
STATUS from the main menu.
STOP key is pressed.
MAIN or
41
Page 44
Sampler operation

4.7.2 Status for main program

4.7.2.1 Ready to start
4.7.2.2 Running/halted
The status screen for the main program gives information about the program depending on the status of the main program. The status information depends on whether the program is in the ready to start, running/halted or complete mode.
The status reads READY when a program is ready to start and also displays the power supply voltage and the current time.
The status reads RUNNING or HALTED when a program is currently running or has been halted by the user. Use the to access information about the program currently running.
RIGHT ARROW key
4.7.2.3 Complete
The information displayed includes:
Power supply voltage
Inhibit mode (none, setpoint trigger, program delay, user
start, storm trigger, full bottle, setpoint delay)
Sample summary (number of samples tried, missed and
remaining)
Next sample information (sample number, bottle number of
next sample)
Time before next sample
Program start time
The status reads COMPLETE when a program ends after all sample cycles have been completed or when ended by the user. Use the
RIGHT ARROW key to access information about the program that
just ended. The display includes:
Power supply voltage
Program end time
Sample summary (number of samples tried and missed)

4.7.3 Status for stormwater program

42
The status screen shows information about the stormwater program similar to that shown for the main program (Ready, Running, Halted or Complete).
Use the the program.
RIGHT ARROW key to access additional information about
Page 45

4.7.4 Sample history

Sampler operation
Sample history displays the sample number, bottle number, result and time/date for each sample. The following results can be shown:
Success—the sample was drawn successfully
Bottle full—the full bottle shut-off was activated
Rinse error—an error occurred during the rinse cycle
User abort—a user pressed the
Arm faulty—the distributor arm did not move correctly
Pump fault—a fault occurred when the pump was operating
•Purge fail—an error occurred during the purge cycle
Sample time out—liquid not detected within the timeout period
•Power fail—a power failure occurred during sampling
If a program is running or has been halted, the sample history is displayed for the currently running sample program. If the program has been completed, the sample history is displayed for the most recently completed sampling program. The sample history is automatically erased when a new program is started.
Procedure:
1. Select
to view the sample history.
2. The sample history will list the sample numbers, bottle numbers
and brief result for each sample. Select a sample from the list and press ENTER to view complete details.
3. The complete details for the selected sample will be shown.
DIAGNOSTICS>SAMPLE HISTORY from the main menu
STOP key to end the program

4.7.5 Event log

The event log records information on the following events:
Power on—power was connected to the controller
Power fail—power was intentionally or unintentionally
disconnected from the controller without first turning power off
Firmware update—a new version of firmware was installed
Pump fault—a fault occurred when the pump was operating
Arm faulty—the distributor arm did not move correctly
Low memory battery—the internal battery should be replaced
•User on—a user turned the power on using the power key
•User off—a user turned the power off using the power key
Program start—the sample program was started
Program resume—a program was resumed from the halt state
Program halt—a program was halted
Program complete—a program was completed
•Grab sample—a grab sample was taken
Change tube—the maximum pump cycle counts have expired
43
Page 46
Sampler operation
4.7.5.1 View the event log
View the event log to see details about the events that have been recorded.
Procedure:
4.7.5.2 Erase the event log
1. Select
the event log.
2. Select
3. The event log will list the date and event. Select an event from
the list and press ENTER to see complete details.
The top bar will show the event number and total number of events in the event log. For example, 01/80 will be displayed for event number 1 and a total number of 80 events.
4. The time/date, event description and any additional data for the
selected event will be displayed.
Erase the event log after event details have been reviewed to reduce the number of entries that are shown.
Procedure:
1. Select
the event log.
2. Select
3. Select
DIAGNOSTICS>EVENT LOG from the main menu to view
DISPLAY.
DIAGNOSTICS>EVENT LOG from the main menu to erase
ERASE.
YES to confirm the deletion.

4.8 Volume calibration

Samples can be collected with the liquid sensor enabled (recommended) or disabled. When the liquid sensor is enabled, the volume calibration is optional. When the liquid sensor is disabled, the volume for samples, rinses and stormwater must be calibrated manually.
When the liquid sensor is used for volume calibration, the sensor is adjusted to accurately measure all programmed volumes. When the sensor is disabled and the volume is calibrated by time, all programmed sample volumes must be calibrated individually.
Verify calibrations by measuring the volume from a grab sample (section 4.8.3 on page 47). If calibration using the liquid sensor does not give accurate volumes, the sensor can be calibrated (section 4.9.1 on page 48).
For typical applications, tap water can be used for volume calibration. If the sample composition is significantly different from typical water samples, calibrate the sampler using the liquid to be collected.
Calibration cannot occur while a sample program is running. All programs must be ended before calibration.
44
Page 47

4.8.1 Volume calibration using the liquid sensor

Use the volume calibration with the liquid sensor enabled to adjust sample volumes slightly.
Procedure:
Sampler operation
1. Make sure the liquid sensor is enabled by selecting
SETUP
2. Press
>MODIFY SELECTED>LIQUID SENSOR>ENABLE.
VOLUME CALIBRATION and select CALIBRATION.
PROGRAM
3. Remove the outlet pump tube from the fitting on the sampler
and place the tube into a graduated cylinder. Place the intake tube into the sample source or tap water.
Note: Use the sample source for best accuracy.
4. Select START.
5. The pump will purge the intake tube and then collect the
sample volume that is specified in the sampling program. The pump will then purge the intake tube.
To halt the calibration at any time, press the
6. When the sample has been collected, select
STOP key.
DONE. Compare
the volume collected in the graduated cylinder with the sample volume that is entered in the program setup menu.
Select
REPEAT to repeat the volume collection if necessary.
7. If the volume that was collected is different from the sample
volume in the current program, enter the volume that was actually collected. Press
ENTER. The sensor is then adjusted to
accurately measure all programmed volumes.
4.8.1.1 Reset the calibration
To verify sample volume, collect a grab sample (section 4.8.3
on page 47).
8. Select
START to start the sampling program or CANCEL to exit
the volume calibration menu.
Use the reset cal option to restore the volume calibration to the default factory settings.
Procedure:
1. Make sure the liquid sensor is enabled by selecting
SETUP
2. To undo the calibration, press
select
>MODIFY SELECTED>LIQUID SENSOR>ENABLE.
VOLUME CALIBRATION and
RESET CAL.
PROGRAM
3. The calibration will be reset to the default calibration and the
message “value has been reset to 0” will be shown.
45
Page 48
Sampler operation

4.8.2 Volume calibration based on time

When the liquid sensor is disabled, the main sample volume, the first flush storm volume (if storm water is enabled), and the rinse (if rinses are > 1) must be calibrated manually.
The sample volume is calibrated for the volume specified in the current program. If the sample volume is changed in the program, the sample volume must be recalibrated for the new volume.
Procedure:
1. Make sure the liquid sensor is disabled by selecting
SETUP
2. Press
>MODIFY SELECTED>LIQUID SENSOR>DISABLE.
VOLUME CALIBRATION. Select one of the displayed
PROGRAM
volumes to calibrate if more than one is listed.
3. If calibrating a sample volume, remove the outlet pump tube
from the fitting on the sampler and place the tube into a graduated cylinder. Place the intake tube into the sample source or tap water.
4. Select
START. The pump will purge the intake tube and then
begin to collect a sample.
5. Stop the pump at the selected volume or rinse location:
Sample volume: observe the volume in the graduated
cylinder and press the
STOP key when the volume
specified in the main program is collected.
Rinse: observe the liquid as it travels up the intake tube
and press the
STOP key when the liquid reaches the liquid
detector.
Storm volume: observe the volume in the graduated
cylinder and press the
STOP key when the volume
specified in the stormwater program is collected.
6. If the pump was stopped at the correct volume, select
repeat the calibration select
REPEAT.
DONE. To
46
7. Select another volume to calibrate or select
Note: All volumes must be calibrated before exiting the volume calibration menu.
DONE.
8. When finished, connect the outlet pump tube to the tube fitting
on the sampler.
9. Select
START to start the sampling program or CANCEL to exit.
Page 49

4.8.3 Verify sample volume

Sampler operation
To check the sample size after calibration use the Manual Operation key to take a grab sample. Do not go back into calibration to check the volume since the volume compensation is reset to zero at the start of a calibration.
Procedure:

4.9 Controller settings

1. Press the
2. Remove the outlet pump tube from the fitting on the sampler
and place the tube into a graduated cylinder. Place the intake tube into the sample source or tap water.
3. If the liquid sensor is enabled, enter the volume that needs to
be verified. If the liquid sensor is disabled, enter the volume specified in the program.
4. Press
5. Compare the volume that was collected in the graduated
cylinder to the volume specified in step 3. If the volume is acceptable, the sampler is calibrated correctly. If the volume is incorrect, repeat the volume calibration.
Use the system setup menu to change the following parameters:
Date and time
Communication
Sampler base
Liquid sensor calibration
MANUAL OPERATION key. Select GRAB SAMPLE.
ENTER. The pump cycle will start (purge-sample-purge).
Display contrast
Password
Pump tube replacement reminder
Procedure:
1. Select
2. Select one of the available parameters, for example time/date
setup. Refer to Table 2 on page 35 for an overview of the system setup menu.
3. Make selections or change values with the arrow keys. Use the
LEFT AND RIGHT ARROW keys to move the cursor left or right.
Use the
4. Select another parameter from the system setup menu, or
press
SYSTEM SETUP from the main menu.
UP AND DOWN ARROW keys to change the value.
BACK to return to the main menu.
47
Page 50
Sampler operation

4.9.1 Liquid sensor calibration

The accuracy of the liquid sensor can vary with the type of liquid that is being sampled. For example, the volume of a highly turbid sample may not be as accurate as that of clear water. In some applications, the liquid sensor may be calibrated using the source liquid to improve detector performance.
Procedure:
1. Make sure the liquid sensor is enabled by selecting
SETUP
2. Select
3. Select
>MODIFY SELECTED>LIQUID SENSOR>ENABLE.
SYSTEM SETUP>LIQ SENSOR CAL from the main menu.
PERFORM CAL.
PROGRAM
4. Place the intake tube into the sample source or tap water (if tap
water is representative of the type of samples to be collected).
Note: Use the sample source for best accuracy.
5. Select START. The pump will operate in reverse to purge the
intake tube.
6. The pump will then operate in the forward direction. Observe
the sample flow in the intake tube to verify that it passes the liquid sensor. When the sample passes through the liquid sensor and no bubbles are present, press the
STOP key.
7. The liquid sensor calibration is complete after the intake tube is
automatically purged. Select menu or
Note: To restore the default calibration for the liquid sensor, select SYSTEM SETUP>LIQ SENSOR CAL>RESTORE DEFAULT from the
main menu.
REPEAT to repeat the calibration.
DONE to exit the liquid sensor

4.9.2 Set password

A password can be enabled to restrict access to the program setup and system setup menus.
Procedure:
1. Select
SYSTEM SETUP>PASSWORD SETUP from the main
menu.
2. Select
3. Select
ENABLE to activate a password.
YES to set a new password or NO to activate the existing
password.
4. If yes was selected, enter the old password and press
ENTER. If
changing the password for the first time, enter 900900 as the old password.
5. Enter a new password and press
ENTER. The new password
will be required to access the menus for program setup and system setup.
48
Page 51

4.9.3 Tubing life indicator

4.9.3.1 Enable the tubing life indicator
Sampler operation
The tubing life option can be used to set a reminder to replace the pump tubing. The user defines a maximum number of pump cycles after which a message will be shown to replace the pump tubing. After replacing the tubing, the current number of cycles must be reset for the new tubing.
Enable the tubing life indicator to set a reminder for when the pump tubing should be replaced.
Procedure:
4.9.3.2 Tubing life status
4.9.3.3 Tubing life cycle limit
1. Select
SYSTEM SETUP>TUBING LIFE from the main menu to
access the tubing life option.
2. Select
ENABLE to activate the tubing life indicator.
Use the status option to see how many pump cycles have been completed.
Procedure:
1. If the tubing life option is not enabled, select
SYSTEM SETUP>TUBING LIFE>ENABLE>STATUS. If the tubing
life option is enabled, select
SYSTEM SETUP>TUBING LIFE>STATUS.
2. The current number of cycle counts on the pump since the
tubing life indicator was reset and the cycle limit set by the user are both shown. Press
ENTER or BACK to return to the system
setup menu.
Use the cycle limit option to set the number of pump cycles when the change tube reminder will be shown.
Procedure:
1. If the tubing life option is not enabled, select
SYSTEM SETUP>TUBING LIFE>ENABLE>CYCLE LIMIT. If the
tubing life option is enabled, select
SYSTEM SETUP>TUBING LIFE>CYCLE LIMIT.
2. Change the cycle limit to the number of pump cycles after
which the pump tubing should be replaced. After setting the limit, press
ENTER to view the current status and verify the
cycle limit.
3. When the number of actual cycle counts equals the cycle limit
set by the user, the “change tube” message will be shown when a key is pressed and an entry will be placed in the event log.
49
Page 52
Sampler operation
4.9.3.4 Reset tubing cycles
When the “change tube” message is shown, replace the pump tubing and then reset the pump cycles.
Procedure:
1. If the tubing life option is not enabled, select
SYSTEM SETUP>TUBING LIFE>ENABLE>RESET CYCLES. If the
tubing life option is enabled, select
SYSTEM SETUP>TUBING LIFE>RESET CYCLES.
2. Select
YES to reset the pump cycle count to 0. Press ENTER to
view the status screen to verify that the cycle count is set to 0.
Note: It is important to reset the pump cycles so that the change tube message will be shown at the correct time for the new tube.
50
Page 53

Section 5 Advanced sampling

5.1 Advanced sampling overview

Use the advanced sampling menu for the following operations:
Send a 12 VDC output signal to start a second sampler,
activate a relay or for communications
Set the sampler to start and stop from an external signal
Set multiple start and stop times
Set variable time intervals
Collect first flush samples from stormwater

5.2 Advanced sampling menu

Table 5 outlines the menu for the advanced sampling operations.
Table 5 SD900 advanced sampling menu
Menu option Description
Program complete O/P
Setpoint sample Start and stop the sampler from an external source
Start on setpoint
Start/stop
Special output Send a signal to a connected device.
Send a signal at the completion of a sampling program. The signal can be used to start a second sampler or activate a relay.
Start a sampling program when the external signal falls outside of the setpoint limits. Continue running even when the signal falls within the setpoint limits.
Start sampling when the external signal falls outside of the setpoint limits. Stop sampling when the signal falls within the setpoint limits.
Each sample Send signal at the completion of each sample cycle. Sample pumping Send signal during the sample intake. Rinse to purge Send signal during the entire sample cycle, including all purge and rinse cycles.
Start/stop times Start and stop a program at user specified dates and times.
Program start Set up to 12 start and stop times for any one program.
Stormwater Collect first flush samples at timed intervals. Can be initiated by external device.
Immediate Start program immediately. External Start program from an external device.
Variable intervals Sample at variable time or flow intervals.
Done Exit the advanced sampling menu.

5.3 Advanced sampling instructions

To access the advanced sampling menu, select PROGRAM
SETUP
>MODIFY SELECTED>ADV SAMPLING from the main menu.
51
Page 54
Advanced sampling

5.3.1 Send output signal at program completion

A 12 VDC signal can be sent through pin F of the auxiliary connector at the completion of a sampling program or when the full bottle shut-off is activated. The signal remains on for 61 seconds.
To enable the ability to transfer sample history to an external Sigma 950 flow meter, program complete must be disabled to enable the bottle number output. In addition, the special output must be enabled and configured to assert after the sample cycle is complete.
The output signal can be used for the following purposes:
Start another sampler. Cascading samplers require a
cascading cable (Cat. No. 8757300). The slave sampler must enable setpoint sampling and start on external trigger.
Activate a relay or other device when the program is complete.
Procedure:
1. Select
PROGRAM SETUP>MODIFY SELECTED>ADV SAMPLING
from the main menu.
2. Select
PGM COMPLT O/P>ENABLE.
3. The message “Bottle number output disabled on pin F” will be
shown. The bottle number output to send sample history to a Sigma 950 flow meter will be disabled.
4. Select
BACK or ENTER to activate the output signal.

5.3.2 Send output signal with sample cycle (special output)

The special output option sends a 12 VDC signal through pin E of the auxiliary connector during or after a sample cycle. When program complete output is disabled, and program complete is disabled (thereby enabling the program complete output). The sample history will be transferred to the external Sigma 950.
There are three possible configurations:
After each sample—four second pulse at the end of each
sample cycle. This option is used to signal an external device that a sample cycle is complete.
During sample intake—during the sample intake portion of the
cycle only, ignoring all purges and rinse cycles.
52
During entire cycle—during the entire sample cycle including all
purge and rinse cycles.
Procedure:
1. Select
PROGRAM SETUP>MODIFY SELECTED>ADV SAMPLING
from the main menu.
2. Select
SPECIAL OUTPUT>ENABLE.
3. Select one of the output options:
AFTER CYCLE—at the completion of each sample cycle.
SAMPLE PUMPING—during sample intake only.
ENTIRE CYCLE—during entire sample cycle.
Page 55

5.3.3 Operate from external signal (setpoint sample)

The setpoint sample option starts and stops the sampler after receiving a signal from an external device such as a flow meter. The current status indicates when the sample program is being inhibited by an external device.
Note: Setpoint and stormwater sampling cannot be enabled at the same time.
Procedure:
Advanced sampling
1. Select
2. Select
3. Select one of the setpoint options:
4. Enter a delay time in hours and/or minutes that must expire

5.3.4 Set multiple start and stop times

The start/stop time option creates an intermittent sampling schedule. The program begins sampling at the first start time and continues until the first stop time. Up to 12 start and 12 stop times can be specified.
Procedure:
PROGRAM SETUP>MODIFY SELECTED>ADV SAMPLING
from the main menu.
SETPT SAMPLE>ENABLE.
START ON SETPTstarts a sample program when the
signal is received, ignoring all future transitions, and continues until the program is complete.
START/STOPstarts a sample program when the signal is
received and stops the program when the signal is no longer received.
before the sample program starts. The delay time will begin when the external signal is received. The program will begin after the delay time has passed.
1. Select
PROGRAM SETUP>MODIFY SELECTED>ADV SAMPLING
from the main menu.
2. Select
STRT/STOP TIME>ENABLE.
3. If the sample program is configured for multiple bottles, select
ENABLE to start sampling at bottle 1 when each start time is
reached. Select
DISABLE to continue the distribution sequence
from the last stop time.
4. Enter the time (24 hour) and date for the first start time.
5. Enter the time (24 hour) and date for the first stop time.
6. Repeat steps 4 and 5 until all start and stop times have
been entered.
53
Page 56
Advanced sampling

5.3.5 Collect first flush stormwater samples

The stormwater option collects first flush (FF) samples or initial runoff during storm conditions when pollutants are typically at a high concentration. The first flush samples are collected at timed intervals and can be initiated by an external device. A stormwater program can run at the same time as the main program. The stormwater option has the following capabilities:
Collection of grab samples at up to 24 different time intervals.
First flush and main program flow-weighted composite samples are automatically segregated. The first flush sample volume may be set independently of the sample volume for the flow-weighted composite.
Initiation by an external device. A dry contact closure across
pins B and D on the auxiliary connector, held closed for at least 61 seconds, is required.
When sampling with multiple bottles, the sampler can be
programmed to collect one large “first flush” sample (or small multiple samples) at timed interval(s). The number of bottles segregated for the first flush sample is selectable. Concurrently, flow-weighted samples are collected from the beginning of the storm until all remaining bottle(s) are filled, or after a user selected time has elapsed.
Procedure:
1. Select
from the main menu.
2. Select
3. Select whether the program will be started immediately or from
an external device:
Immediate—program starts when
External—program starts when signaled by an
4. If configured for a single bottle, enter the number of first flush
(FF) samples to collect.
5. If configured for multiple bottles, enter the number of bottles to
be used for first flush collection.
6. Enter the multi-bottle distribution mode (Figure 20 on page 38):
Samples per bottle: more than one sample is collected in
Bottles per sample: more than one bottle is used to collect
PROGRAM SETUP>MODIFY SELECTED>ADV SAMPLING
STORMWATER>ENABLE.
RUN is pressed and
START is selected.
external device.
each bottle
each sample.
54
7. Enter the number of samples per bottle or bottles per sample.
8. Enter the time interval between each first flush sample cycle.
Variable intervals may be entered, or the same time may be entered for each interval. The sum of the intervals is equal to
Page 57
Advanced sampling
the total length of the stormwater collection period. First flush samples are usually collected within 30 minutes of a storm.
Note: The number of intervals is determined by the number of bottles and the distribution mode.
9. Enter the volume of each first flush sample.

5.3.6 Set variable intervals

10. Select
ENABLE to set a time limit for stormwater collection.
Enter the program time limit. The program will stop when this time limit is reached.
Select
DISABLE to allow the program to run until complete
without a time limit.
NPDES requirements typically request monitoring for the first three hours of a storm. If the flow volume is lower than expected, flow-weighted sampling could continue for some time as flow rates drop and sample intervals become longer.
Variable time or flow intervals can be set for sample cycles. Intervals will be in hours and minutes if the sample pacing mode is time based. Intervals will be in counts if the sample pacing mode is flow based.
Procedure:
1. Select
PROGRAM SETUP>MODIFY SELECTED>ADV SAMPLING
from the main menu.
2. Select
VARIABLE INTVL>ENABLE.
3. Enter the time (or count) interval between each sample cycle.
Up to 99 intervals can be entered.
55
Page 58
Advanced sampling
56
Page 59

Section 6 Maintenance

CAUTION Only qualified personnel should conduct the tasks described in this section of the manual.
WARNING Potential shock hazard. Always disconnect power to the sampler before performing any maintenance or service.
CAUTION Biological sample hazard. Follow safe handling protocols during contact with sample bottles and sampler components. Disconnect the sampler from power to disable the pump before handling.

6.1 Upgrades, Repairs, General Maintenance

Only a qualified technician should service the sampler. For example, steps that require knowledge of CMOS electrostatic discharge precautions and advanced electronics training should only be performed by a qualified technician.

6.2 Sampler cleaning

6.3 Pump maintenance

6.3.1 Pump tube life

Sampler housing
Clean the interior and exterior of the sampler housing with a damp sponge and mild detergent. Do not use abrasive cleaners.
Sample bottles
Clean the bottles using a brush and water with a mild detergent, followed by a fresh water rinse and a distilled water rinse. Glass bottles may also be autoclaved.
CAUTION Pinch hazard. Always disconnect power to the sampler before opening the pump cover.
The tubing in the peristaltic pump will deteriorate over time from the action of the rollers against the tubing. Be sure to replace the pump tubing before it splits to prolong the life of the sampler and prevent contamination of the work area.
Pump tube life depends on several factors:
Distance from sample source. Locate the sampler as close to
the sample source as possible.
Vertical lift. Minimize the vertical lift as much as possible.
Intake rinses. Minimize the number of intake rinses prior to
sample collection.
Sample retries. Minimize the number of sample retries to
successfully collect a sample.
Sample volumes/duration. Minimize the volume of sample and
maximize the time between samples.
57
Page 60
Maintenance

6.3.2 Pump tube replacement

Constituents in the sample liquid. Grit and other abrasive solids
cause greater wear on the tube when they are squeezed through the pump rollers.
Important Note: Use of tubing other than that supplied by the manufacturer may cause excessive wear on mechanical parts and/or poor pump performance.
Inspect the pump tubing and rollers on a regular basis. Replace the tubing when deteriorated, at regular intervals or when the “change tube” message is displayed (section 4.9.3 on page 49).
Prerequisites:
Pump tubing—pre-cut or bulk (15 ft or 50 ft)
Note: See Replacement parts on page 69 for ordering information.
Removal procedure:
1. Turn off the controller power.
2. If using bulk pump tubing, cut a 23.25 inch piece and mark the
tube with alignment dots at the locations shown in Figure 21.
3. Disconnect the old pump tubing from the liquid sensor and from
the tube fitting on the sampler housing.
4. Loosen the thumbscrew on the pump cover and open
the cover.
5. Pull the tube from the pump tube track while turning the rotor to
remove the old tubing from the pump.
6. Clean the silicone residue from the interior of the pump housing
and from the rollers (see section 6.3.3 on page 60).
Figure 21 Pump tubing measurement
1 To liquid sensor 3 Inside pump body 2 Alignment dot 4 To tube fitting on center section
58
Page 61
Maintenance
Installation procedure:
1. Connect one end of the tubing to the liquid sensor fitting. Press
the tubing into the pump tube guide (Figure 22).
2. Press the tubing into the upper pump tube track. Make sure
that the alignment dot is in line with the upper pump tube track as shown in Figure 22.
3. Hold the tubing with one hand while rotating the pump rotor to
guide the tubing around the rotor (Figure 22). The guide pins will push the tubing into place.
4. Press the tubing into the lower pump tube track. Make sure that
the alignment dot is in line with the lower pump tube track as shown in Figure 22.
5. Close and secure the pump cover with the thumbscrew.
6. Connect the open tube end to the tube fitting on the sampler
(Figure 23).
Figure 22 Pump tubing installation
1 Pump tube guide 5 Rotor 2 Pump tube track (upper) 6 Pump tube track (lower) 3 Guide pin 7 Connection to liquid sensor 4 Alignment dot
59
Page 62
Maintenance
Figure 23 Pump tube installed
1 Pump tube to sampler tube fitting 2 Pump tube to liquid sensor

6.3.3 Rotor removal and cleaning

Remove the pump rotor to thoroughly clean the silicone residue from the pump tube.
Procedure:
1. Remove the pump tube (see section 6.3.2 on page 58).
2. Open the rubber flap on the end of the rotor and remove the
screw with a flat-blade screwdriver (Figure 24).
3. Remove the rotor. A collet connects the rotor to the pump shaft
and can come off when the rotor is removed (Figure 24).
4. Clean the rotor, pump tube tracks and pump housing with a
mild detergent.
5. If the collect was removed in step 3, align the recess in the
collet with the tang on the shaft and push the collet on the shaft (Figure 24).
6. Push the rotor on the shaft and secure with the screw. Do not
overtighten the screw. Close the rubber flap on the rotor.
7. Install the pump tube (section 6.3.2 on page 58).
60
Page 63
Maintenance
Figure 24 Pump rotor removal
1 Shaft 5 Rotor 2 Collet recess 6 Collet 3 Rubber flap 7 Tang 4 Screw
61
Page 64
Maintenance

6.4 Distributor arm tubing replacement

The distributor arm moves over each bottle during multiple bottle sampling. Inspect the tubing in the distributor arm on a regular basis. Replace the tubing when deteriorated or at regular intervals.
Prerequisites:
Tubing, distributor arm, pre-cut. Make sure that the correct tube
is used for the correct distributor and distributor arm. Refer to
Figure 2 on page 9 and to Replacement parts on page 69.
Procedure:
1. Remove the old tubing from the distributor arm.
2. Insert the new tubing into the distributor arm so that the end of
the tubing extends out of the nozzle no more than (Figure 25).
Note: Do not let the tubing extend more than 1/8 inch past the nozzle end of the arm. If the tubing extends too far, the tubing will get caught on the bottles and interfere with sampling.
3. To make sure that the distributor is aligned properly, run the
manual distributor diagnostic (see section 7.3 on page 67).
Figure 25 Distributor assembly
1 Nozzle 4 Shaft 2 Distributor arm 5 Tube 3 Distributor motor
1
/8 inch

6.5 Desiccant replacement

62
A desiccant cartridge is located inside the controller to absorb moisture and prevent corrosion. Over time the desiccant will become saturated with moisture and should be replaced.
Monitor the desiccant color through the clear plastic window (Figure 26). The color will change from yellow to green when the desiccant is saturated.
Prerequisites:
Desiccant
Silicone grease
Note: See Replacement parts on page 69 for ordering information.
Page 65
Maintenance
Procedure:
1. Unscrew and remove the desiccant holder from the controller
(Figure 26).
2. Remove the plug from the end of the desiccant tube (Figure 27)
and discard the spent desiccant.
3. Fill the desiccant tube with fresh desiccant.
Note: Fresh desiccant should be yellow.
4. Replace the plug.
5. Apply grease to the O-ring (Figure 27).
6. Replace the desiccant tube in the controller.
Figure 26 Side view of controller—pump and desiccant
1 Desiccant window 2 Peristaltic pump
Figure 27 Desiccant holder
1 Plug 3 O-ring 2 Desiccant tube
63
Page 66
Maintenance
64
Page 67

Section 7 Troubleshooting

7.1 General troubleshooting

Refer to the information in Table 6 when a problem with the sampler occurs.
Table 6 SD 900 general troubleshooting information
Problem Cause Solution
Instrument will not power on with AC power.
Instrument will not power on with DC power.
Battery life is short.
Sampler will not create sufficient lift.
Inaccurate sample volumes.
Problem may be with the controller, power supply or main power source.
Problem with circuit breaker.
Problem with electrical outlet.
Defective power supply. Replace power supply or use a battery. Defective controller. Contact the service center. Battery is not charged. Replace with a fully charged battery.
Battery does not hold charge.
Defective controller. Contact the service center.
Insufficient voltage.
Battery power drains quickly.
Incorrect sampler base configuration.
Strainer is not completely submerged.
Intake tube has a leak. Replace intake tube. Pump tube is worn. Replace pump tube. Pump roller assembly is worn. Contact the service center.
Incorrect volume calibration.
Incorrect intake tube length is programmed in the sampler.
Intake tube is not completely purging.
Stainer is intermittently submerged.
Worn pump tubing and/or roller assembly.
When the liquid sensor is disabled, variable suction heads will cause variable sample volumes.
Liquid sensor is not operating properly.
Isolate components to determine which component has the problem.
Check the circuit breaker for the main power.
Check to see if the outlet is receiving power.
Replace battery or use an AC power supply.
The lead-acid battery should hold
12.6 V to 13.4 V when fully charged. Fully charge the battery and allow to
stand for one hour. If the voltage drops below 12.5 V, replace the battery.
Set sampler base to portable option (Table 2 on page 35).
Make sure the strainer is completely submerged.
Repeat the volume calibration and take a grab sample to verify accuracy.
Measure the intake tube length and enter in the program setup menu.
Make sure the intake tube is on a downward slope as vertical as possible, with no low points where liquid can accumulate.
Install the shallow depth strainer (Cat. No. 2071 or 4652).
Replace pump tubing; contact service center for roller assembly.
Enable the liquid sensor and complete a volume calibration.
Calibrate the liquid sensor using the same liquid that is being sampled.
65
Page 68
Troubleshooting

7.2 Error messages and notifications

Messages will be shown in the controller display when errors occur or to confirm that an operation is complete. Refer to Table 7 for a description of the possible messages.
Table 7 SD 900 error and information messages
Message Reason Action
ERROR WRONG RANGE
BOTTLE NUMBER OUTPUT DISABLED ON PIN F
NOT ALLOWED CANNOT ENABLE SETPOINT SAMPLING AND STORM WATER
NOT PERMITTED. END THE PROGRAM FIRST TO ACCESS THIS FEATURE
NEED TO CAL. LIQUID SENSOR IS DISABLED. CAL IS REQUIRED!
The value that was entered is outside of the acceptable range.
Message is displayed when program complete output is enabled. The bottle number output on the auxiliary port (pin F) is automatically disabled.
An attempt was made to enable setpoint sampling and stormwater sampling at the same time.
An attempt was made to access a menu such as program setup, system setup, volume calibration or manual operation while a program was running.
The liquid sensor was disabled and a manual volume calibration is required.
Press number that is within the range specified on the bottom of the screen.
Press BACK or ENTER to keep program complete output enabled. If bottle number output is preferred, disable the program complete output.
Disable setpoint sampling or stormwater sampling, then enable the preferred option.
Press
END PROGRAM or CANCEL to continue
the program.
Calibrate all volumes as specified in
section 4.8 on page 44.
BACK or ENTER. Enter a valid
BACK or ENTER and select
NEED TO CAL. LIQUID SENSOR IS ENABLED. SENSOR REQUIRES CAL
DISTRIBUTOR MOVE IS COMPLETE
NOT ALLOWED. 1 BTL CONFIGURED DISTRIBUTOR NOT RELEVANT.
RESET CAL. VALUE HAS BEEN RESET TO 0.
NOT ALLOWED. ENABLE LIQUID SENSOR TO PERFORM CAL
BOTTLE NUMBER OUTPUT ENABLED ON PIN F
CANNOT ENABLE. PGM START/STOP AND STORM WATER AT SAME TIME
FAULT. PUMP FAULT
ALARM. CHANGE TUBE
FAULT. ARM FAULTY
The liquid sensor must be calibrated.
Confirmation that the distributor arm was moved to a specific bottle position and the move is complete.
An attempt was made to move the distributor when the program setup was set to a single bottle.
Confirmation that volume calibration with the liquid sensor enabled was reset to 0 in the volume calibration menu.
An attempt was made to calibrate the liquid sensor when the liquid sensor was disabled.
Calibrate the liquid sensor as specified in section 4.9.1 on page 48.
BACK or ENTER to acknowledge
Press the information.
Change the bottle number configuration in the program setup menu before moving the distributor.
Press BACK or ENTER to acknowledge the information. The liquid sensor is no longer calibrated.
Enable the liquid sensor in the program setup menu.
Confirmation that the bottle number output on pin F of auxiliary port is enabled when the program complete
BACK or ENTER to acknowledge
Press the information.
output is disabled.
An attempt was made to enable start/stop time and stormwater at the same time.
Disable start/stop time or stormwater in the advanced sampling menu, then enable the preferred option.
The pump failed to operate properly. Examine the pump for obstructions.
The specified maximum number of pump cycles have expired. The pump tube should be changed.
The distributor arm failed to operate properly.
Change the pump tube and reset the maximum number of pump cycles (section 4.9.3 on page 49).
Make sure the distributor arm is connected properly and can move freely in both directions.
66
Page 69
Table 7 SD 900 error and information messages (continued)
Message Reason Action
Troubleshooting
ERASING LOG. PLEASE WAIT
ERROR WRONG PWD
PGM DELAY DISABLED
STARTING PLEASE WAIT
ALARM LOW BATTERY
ALARM LOW MEMORY BATTERY
ERROR INVALID SELECTION

7.3 Diagnostic tests

Confirmation that the event log is being cleared.
An incorrect password was entered.
A program delay was set based on counts but sample pacing/collection was set based on time. The program delay is automatically disabled.
Confirmation that a program is starting. Wait for the program to start.
The main battery power is low.
The internal memory battery is low Contact the service center.
The selection is not compatible with the program setup. For example, the base selected in the system setup menu must be compatible with the number of bottles in the program setup menu.
Wait for the log to be erased.
Press
BACK or ENTER and enter the
correct password.
To use a program delay in counts, set the sample pacing/collection to flow based.
Replace the battery with a fully-charged battery.
Select a value that is compatible with the values specified in the program setup menu.
Use the diagnostics option to test for correct operation of the following components:

7.3.1 Distributor diagnostic

Distributor (section 7.3.1)
Keypad (section 7.3.2)
Display (section 7.3.3)
Liquid sensor (section 7.3.4)
An overview of the available diagnostic tests is shown in Table 2 on
page 35.
Use the distributor diagnostic to check for proper operation of the distributor. The test moves the distributor arm to all 24 bottle positions regardless of the number of bottles that are specified in the program.
Note: The distributor diagnostic is only active when the sample program is configured for multiple bottles.
Procedure:
1. Select
DIAGNOSTICS>DISTRIB DIAG from the main menu. The
distributor arm will begin moving to all positions.
2. The display will show
ARM MOVING and then return to the
diagnostics menu when the test is complete.
If the arm is not connected or is not able to move to all positions, the message “arm faulty” will be shown. Review
section 3.3.3.5 on page 26 and section 6.4 on page 62 to make
sure the distributor and arm are set up properly.
67
Page 70
Troubleshooting

7.3.2 Keypad diagnostic

Use the keypad diagnostic to make sure each key on the keypad is operating correctly.
Procedure:

7.3.3 LCD diagnostic

7.3.4 Liquid sensor diagnostic

1. Select
2. Press each key on the keypad (except the
display should show the name of the key that was pressed.
3. To exit the test, press
Use the LCD diagnostic to make sure the controller display is operating correctly.
Procedure:
1. Select
the test.
2. Observe the display. Each row will show SD900 and then each
pixel will darken. When complete, the display will briefly show “test complete”.
Use the liquid sensor diagnostic to show the current calibration information for the liquid sensor.
Procedure:
DIAGNOSTICS>KEYPAD DIAG from the main menu.
BACK key). The
BACK key.
DIAGNOSTICS>LCD DIAG from the main menu to begin
1. Select
2. The display will show the input frequency (MHz), the output
voltage (VDC) and the signal to noise ratio for the current calibration.
DIAGNOSTICS>LIQ SENSE DIAG from the main menu.
68
Page 71

Section 8 Replacement Parts and Accessories

8.1 Bottle kits

Description Catalog number
Single bottle kits (include full bottle shut off):
10-L (2.5-gal) glass bottle and full bottle shut off (for standard base) PS010025 10-L (2.5-gal) poly bottle and full bottle shut off (for standard base) PS010030 10-L (2.5-gal) glass bottle and full bottle shut off (for compact base) PC010025 10-L (2.5-gal) poly bottle and full bottle shut off (for compact base) PC010030 15-L (4.0-gal) poly bottle and full bottle shut off (for standard base) PS010040 21-L (5.5-gal) poly bottle and full bottle shut off (for standard base) PS010055
Multi-bottle kits (include retainer and distributor arm):
(24) 350-mL glass bottles, retainer and distributor arm (for standard base) PS240350 (24) 575-mL poly bottles, retainer and distributor arm (for compact base) PC240575 (24) 1-L poly bottles, retainer and distributor arm (for standard base) PS241000

8.2 Containers and bottle sets

Description Catalog number
Container, 10-L (2.5-gal) glass with cap 6559 Container, 10-L (2.5-gal) poly with cap 1918 Container, 15-L (4-gal) poly with cap 1367 Container, 21-L (5.5-gal) poly with cap 6498 Container, 20-L (5.25 -gal) poly with cap 6494 Bottle set, (24) 1-L poly with caps 737 Bottle set, (24) 350-mL glass with caps 732 Bottle set, (24) 575-mL poly with caps 1369 Bottle set, (8) 2.3-L (0.6-gal) poly with caps 657 Bottle set, (8) 1.9-L (0.5-gal) glass with caps 1118 Bottle set, (8) 950-mL glass with caps 2348 Bottle set, (4) 3.8-L (1-gal) poly with caps 2217 Bottle set, (4) 3.8-L (1-gal) glass with caps 2216 Bottle set, (2) 3.8-L (1-gal) poly with caps 2215 Bottle set, (2) 3.8-L (1-gal) glass with caps 2214

8.3 Replacement parts

Description Catalog number
Desiccant, refill, 1.5 pounds 8755500 Desiccant cap assembly 8754900 Desiccant tube 8742100 Desiccant tube assembly 8741500 Desiccant tube assembly with grease packet 8755600 Grease, silicone, 7 grams (0.25 oz) 000298HY Pump Cover/Door Assembly 8755400 Tubing for Distributor Arm, 16¼-inch, for assembly 8584 (arm 8585) 8579
69
Page 72
Replacement Parts and Accessories
8.3 Replacement parts (continued)
Description Catalog number
Tubing for Distributor Arm, 17-inch, for assembly 8580 (arm 8581) 8577 Tubing for Distributor Arm, 18¾-inch, for assembly 8582 (arm 8583) 8578 Tubing for Peristaltic Pump, pre-cut for portable sampler 8753800 Tubing for Peristaltic Pump, 15 ft 4600-15 Tubing for Peristaltic Pump, 50 ft 4600-50
3
Tubing, Teflon-lined intake Tubing, Teflon-lined intake 3/8-in., 25 ft 922 Tubing, Teflon-lined intake Tubing, vinyl intake 3/8-in., 25 ft 920 Tubing, vinyl intake Tubing, vinyl intake, 3/8-in., 500 ft 924 Tubing Connection Kit, for Teflon-lined PE tubing 2186
3
/8-in., 10 ft 921
3
/8-in., 100 ft 925
/8-in., 100 ft 923

8.4 Accessories

Description Catalog number
3-way Splitter Assembly 939 Battery, gel electrolyte (lead acid), with 3 pin connector, 12 VDC 8754400 Battery Charger, gel electrolyte, with 3 pin connector, 90–250 VAC 8753500US Base, standard assembly 8976 Base, compact assembly 8975 Base, 5.5-gal. poly container 8561 Cable, multi-purpose full, 7 pin, 10 ft 8757100 Cable, multi-purpose full, 7 pin, 25 ft 8757000 Cable, multi-purpose half, 7 pin, 10 ft 8756900 Cable, multi-purpose half, 7 pin, 25 ft 8756800 Cable for Cascade Sampling, 7 pin, 25 ft 8757300 Cable for 980 instrument, half, 7 pin, 10 ft 8758100 Cable for 980 instrument, half, 7 pin, 25 ft 8757200 Cable, battery adaptor (2 to 3 pin) 8739400 Cable, auxiliary adaptor (6 to 7 pin) 8758000 Cable, serial (DB9 to 7 pin), 3 meters 8758200 Distributor Assembly with Arm, 24 bottle, compact base 8580 Distributor Assembly with Arm, 24 bottle standard base 8582 Distributor Assembly with Arm, 2, 4, and 8 bottle base and 8 bottle compact base 8584 Distributor Arm for Assembly 8580 8581 Distributor Arm for Assembly 8582 8583 Distributor Arm for Assembly 8584 8585 Flow-thru Module 2471 Full Bottle Shut-off 8996 Manhole Spanner Bar, 18–28 inches 9542 Manhole Spanner Bar, 28–48 inches 9557 Manhole Support Bracket, 18–27 inches 5713000
70
Page 73
Replacement Parts and Accessories
8.4 Accessories (continued)
Description Catalog number
Power Supply, 3 pin connector, 100–120 VAC 8754500US Retainer, (8), (24) poly/glass bottles 1422 Retainer, (24) 350 mL glass bottles 2189 Retainer, (8) 950-mL glass bottles 2347 Retainer, 1-gal. glass/poly bottles 2190 Retrofit Kit, SD900 controller assembly for portable samplers (includes controller, pump tube, manual) 8759400 Sampler Locking Assembly 1354 SampleView CD with RS-232 Cable 8757500 Strainer, all 316 stainless steel, 6.0 in. long x 0.406 in. OD 2071 Strainer, all 316 stainless steel, 7.94 in. long, x 1.0 in. OD 2070 Strainer, stainless steel, 3.9 in. long x 0.406 in. OD 4652 Strainer, Teflon®/stainless steel, 5.5 in. long x 0.875 in. OD 926 Strainer, Teflon/stainless steel, 11.0 in. long x 0.875 in. OD 903 Support, container, for standard base, 2.5 gal. glass and 3 gal. poly 1502 Suspension Harness 1355
71
Page 74
Replacement Parts and Accessories
72
Page 75

Section 9 Contact information

United States: Hach Company
P.O. Box 389 Loveland, Colorado 80539-0389 U.S.A. Tel 800-635-4567 Fax (970) 461-3919 orders@hach.com www.hach.com
Hach Lange EU subsidiaries:
HACH LANGE LTD
Pacific Way Salford GB-Manchester, M50 1DL Tel. +44 (0)161 872 14 87 Fax +44 (0)161 848 73 24 info@hach-lange.co.uk www.hach-lange.co.uk
DR. BRUNO LANGE AG
Juchstrasse 1 CH-8604 Hegnau Tel. +41(0)44 9 45 66 10 Fax +41(0)44 9 45 66 76 info@hach-lange.ch www.hach-lange.ch
Repair Service in the United States:
Authorization must be obtained before sending any items for repair.
HACH Company Tel 1 (800) 635-1230 or (970) 669-3050 Fax (970) 461-3924 techhelp@hach.com
HACH LANGE LTD
Unit 1, Chestnut Road Western Industrial Estate IRL-Dublin 12 Tel. +353(0)1 46 02 5 22 Fax +353(0)1 4 50 93 37 info@hach-lange.ie www.hach-lange.ie
HACH LANGE FRANCE S.A.S.
33, Rue du Ballon F-93165 Noisy Le Grand Tél. +33 (0)1 48 15 68 70 Fax +33 (0)1 48 15 80 00 info@hach-lange.fr www.hach-lange.fr
Canada, Latin America, sub-Saharan Africa, Asia, Australia and New Zealand:
Hach Company World Headquarters, P.O. Box 389 Loveland, Colorado, 80539-0389 U.S.A. Tel 1 (970) 669-3050 Fax 1 (970) 461-3939 intl@hach.com
DR. BRUNO LANGE GES. MBH
Industriestraße 12 A-3200 Obergrafendorf Tel. +43 (0)27 47 74 12 Fax +43 (0)27 47 42 18 info@hach-lange.at www.hach-lange.at
HACH LANGE SA
Motstraat 54 B-2800 Mechelen Tél. +32 (0)15 42 35 00 Fax +32 (0)15 41 61 20 info@hach-lange.be www.hach-lange.be
Europe, Middle East and Mediterranean Africa: Hach Lange GMBH
Willstätterstraße 11 D-40549 Düsseldorf Tel. +49 (0)2 11 52 88-320 Fax +49 (0)2 11 52 88-210 info@hach-lange.de www.hach-lange.de
DR. LANGE NEDERLAND B.V.
Laan van Westroijen 2a NL-4003 AZ Tiel Tel. +31(0)344 63 11 30 Fax +31(0)344 63 11 50 info@hach-lange.nl www.hach-lange.nl
HACH LANGE E.P.E.
27, Avlidos str GR-115 27 Athens Tel. +30 210 7777038 Fax +30 210 7777976 info@hach-lange.gr www.hach-lange.gr
HACH LANGE APS
Åkandevej 21 DK-2700 Brønshøj Tel. +45 36 77 29 11 Fax +45 36 77 49 11 info@hach-lange.dk www.hach-lange.dk
HACH LANGE LDA
Av. do Forte nº8 Fracção M P-2790-072 Carnaxide Tel. +351 214 253 420 Fax +351 214 253 429 info@hach-lange.pt www.hach-lange.pt
HACH LANGE KFT.
Hegyalja u. 7-13. H-1016 Budapest Tel. +36 (06)1 225 7783 Fax +36 (06)1 225 7784 info@hach-lange.hu www.hach-lange.hu
HACH LANGE AB
Vinthundsvägen 159A SE-128 62 Sköndal Tel. +46 (0)8 7 98 05 00 Fax +46 (0)8 7 98 05 30 info@hach-lange.se www.hach-lange.se
HACH LANGE SP.ZO.O.
ul. Opolska 143 a PL-52-013 Wrocław Tel. +48 (0)71 342 10-83 Fax +48 (0)71 342 10-79 info@hach-lange.pl www.hach-lange.pl
HACH LANGE S.R.L.
Str. Leonida, nr. 13 Sector 2 RO-020555 Bucuresti Tel. +40 (0) 21 201 92 43 Fax +40 (0) 21 201 92 43 info@hach-lange.ro www.hach-lange.ro
HACH LANGE S.R.L.
Via Riccione, 14 I-20156 Milano Tel. +39 02 39 23 14-1 Fax +39 02 39 23 14-39 info@hach-lange.it www.hach-lange.it
HACH LANGE S.R.O.
Lešanská 2a/1176 CZ-141 00 Praha 4 Tel. +420 272 12 45 45 Fax +420 272 12 45 46 info@hach-lange.cz www.hach-lange.cz
HACH LANGE
8, Kr. Sarafov str. BG-1164 Sofia Tel. +359 (0)2 963 44 54 Fax +359 (0)2 866 04 47 info@hach-lange.bg www.hach-lange.bg
HACH LANGE S.L.U.
Edif. Arteaga Centrum C/Larrauri, 1C- 2ª Pl. E-48160 Derio/Vizcaya Tel. +34 94 657 33 88 Fax +34 94 657 33 97 info@hach-lange.es www.hach-lange.es
HACH LANGE S.R.O.
Roľnícka 21 SK-831 07 Bratislava – Vaj nory Tel. +421 (0)2 4820 9091 Fax +421 (0)2 4820 9093 info@hach-lange.sk www.hach-lange.sk
HACH LANGE SU ANALÝZ SÝSTEMLERÝ LTD.ÞTÝ.
Hilal Mah. 75. Sokak Arman Plaza No: 9/A TR-06550 Çankaya/ANKARA Tel. +90 (0)312 440 98 98 Fax +90 (0)312 442 11 01 bilgi@hach-lange.com.tr www.hach-lange.com.tr
HACH LANGE D.O.O.
Fajfarjeva 15 SI-1230 Domžale Tel. +386 (0)59 051 000 Fax +386 (0)59 051 010 info@hach-lange.si www.hach-lange.si
ÇÁCH LANGE E.Ð.Å.
Αυλίδος 27 GR-115 27 Αθήνα Τηλ. +30 210 7777038 Fax +30 210 7777976 info@hach-lange.gr www.hach-lange.gr
73
Page 76
74
Page 77

Appendix A Wiring for non-Sigma flow meters

CAUTION Only qualified personnel should conduct the tasks described in this section of the manual.
Prerequisites:
Multi-purpose half cable
Auxiliary adapter cable (6-pin to 7-pin adapter)
Optional splitter (Figure 15 on page 31) to provide additional
connections. Two or more splitters may be connected in series.
See Accessories on page 70 for ordering information.
Procedure:
Use the auxiliary connector diagram in Figure 28 and wiring information in Table 8 to wire the flow meter to the controller.
Pin
A
Signal description
12 VDC power output
Figure 28 Auxiliary connector (7-pin)
Table 8 Half-cable wiring information for non-Sigma flow meters
Wire color
Red
Purpose Rating
1
Power supply positive output. Powers an external device or flow meter. Must be used in conjunction with pin B (common).
Power supply negative return. When using AC power, this pin is connected to earth ground.
This output is protected by a thermal current limiter. 12 VDC nominal output at 1 A maximum when using
battery 8754400 to power the sampler. 15 VDC regulated output at 0.5 A maximum when
using AC power to power the sampler.B Common Black
75
Page 78
Wiring for non-Sigma flow meters
Table 8 Half-cable wiring information for non-Sigma flow meters (continued)
Signal
Pin
description
C Pulse input Brown
Liquid level actuator/aux
D
iliary control input
Special
E
output
Wire color
Orange
Yell ow
Purpose Rating
1
This input triggers sample collection. The signal may be generated by:
An external flow meter or other
device with an appropriate control output.
Simple floating (dry) contact
closure.
The sampler is capable of accumulating from 1 to 9,999 input pulses. This count is advanced at the beginning (rising edge) of the input signal. Each contact closure output from a flow meter represents a specific flow volume.
The contact closure signal is sent more frequently for high flow rates, and less frequently for low flow rates. The sample frequency will increase as the flow rate increases and decrease as the flow rate decreases, thus taking samples in proportion to flow rate.
This input sends a signal to the microprocessor inside the sampler causing it to "wake up" and begin or resume its sampling program. Applications for this input include:
A simple float level switch (dry contacts) may be used to actuate the sampler when liquid is present.
Start this sampler after another sampler has finished its program.
Start the sampler in response to some user-defined condition from an external device that provides an appropriate control signal. For example a pH meter could be used to initiate sampling in response to a high or low pH condition.
Normally at 0 VDC, this line goes to +12 VDC upon any of the selected events described in section 5.3.2 on page 52.
This input responds to a positive pulse with respect to Common (pin B). It is terminated (pulled low) to Common (pin B) through a series 1 kΩ resistor and
7.5 V zener diode. Absolute maximum input: 0 to 15 VDC with respect
to Common (pin B). Signal required to activate input: 5–15 V positive going pulse with respect to common
(pin B) of at least 50 millisecond duration. Source impedance of driving signal should be less than 5kΩ.
Dry contact (switch) closure of at least 50 millisecond duration between pin A (12 VDC) and pin C (Pulse Input). Contact resistance should be less than 5 kΩ. Contact current will be less than 8 mA DC.
This input is terminated (pulled high) to an internal +5 V supply through an 11 kΩ resistance and protected by a series 1 kΩ resistor and 7.5 V zener diode terminated to Common (pin B). The input responds to a high to low voltage transition with a logic low pulse width of at least 50 milliseconds.
Absolute maximum input: 0 to 15 VDC with respect to Common (pin B). Signal
required to activate input: External logic signal having 5-15 VDC power source.
The drive signal should be normally high. The external driver should be capable of sinking 0.5 mA @ 1 VDC maximum at the logic low level.
A logic high signal from a driver with power source of greater than 7.5 V will source current into this input at the rate of: I = (V - 7.5)/1000 where:
I is the source current V is the power supply voltage of the driving logic. Dry contact (switch) closure of at least 50 millisecond
duration between pin D and pin B (common). Contact resistance should be less than 2 kΩ. Contact current will be less than 0.5 mA DC.
This output is thermally protected against short circuit currents to common (pin B). Limit external load current to 0.2 A.
The output is normally logic low (0 VDC) with respect to common (pin B). Active high output levels are:
15 VDC nominal when using AC power to power the sampler
12 VDC nominal when using battery 8754400 to power the sampler.
76
Page 79
Wiring for non-Sigma flow meters
Table 8 Half-cable wiring information for non-Sigma flow meters (continued)
Pin
Signal description
Wire color
Purpose Rating
1
Normally an open circuit, this line switches to ground for 90 seconds at the conclusion of the sampling program.
F
Program complete output
Green
Used to “wake up” another sampler to take over sampling or to signal an operator or data logger upon the completion of the sampling program. This pin is also used to signal the bottle full condition in a single bottle/continuous mode, and will
This is an open collector output with 18 V zener clamp diode for overvoltage protection. The output is active low with respect to common (pin B).
Absolute maximum ratings for output transistor: Sink current = 25 mA DC maximum. External pull-up voltage = 18 VDC maximum.
transmit the bottle number to a flow meter if the program complete signal is disabled.
Not intended for use as safety ground. Do not use as a current carrying conductor.
Cables greater than 3 meters length connected to the sampler auxiliary port should be shielded.
Regardless of length, if a shielded cable is connected to the auxiliary port, the shield (drain) wire should only be connected to earth ground at one end
GShield N/A
Provides connection to earth ground when the sampler is powered using AC power. This connection is only intended for use in controlling RF emissions and susceptibility.
of the cable to avoid ground loop currents.
1
Wire color refers to the colors of multi-purpose cables Cat. No. 8756800 and 8756900.
77
Page 80
Wiring for non-Sigma flow meters
78
Page 81

Index

A
AC power
installation accessories advanced sampling
menu overview
operate from external signal
output signal
stormwater
variable intervals auxiliary connector
output signal
.......................................................... 30
.............................................................. 69
.................................................. 51
.............................. 53
...................................................... 52
......................................................... 54
................................................ 55
...................................................... 52
B
battery installation .................................................... 29
bottles
bottle no. 1 position
cleaning
eight bottle installation
four bottle installation
list of
setting the number of
twenty four-bottle installation
two bottle installation bottles per sample
............................................................. 19
.................................................................. 69
........................................... 21
....................................... 23
......................................... 22
......................................... 38
............................. 24
......................................... 22
.................................................... 38
C
calibration
liquid sensor
manual calibration
sample volume cascade sampling cleaning
bottles
pump rotor communication
serial connector containers. See bottles
...................................................... 48
............................................. 46
.................................................. 44
.................................................... 52
................................................................ 19
......................................................... 60
................................................. 31
D
data
event log
sample history
viewing default settings desiccant
replacement diagnostic tests
............................................................ 43
................................................... 43
............................................................... 41
......................................................... 39
....................................................... 62
........................................................ 67
distributor
arm tube replacement diagnostic test installing manual operation of
.................................................... 67
............................................................. 26
........................................ 62
........................................... 41
E
error messages ........................................................ 66
event log
.................................................................. 43
F
flow meter
installation trigger sampler operation with
full bottle shut-off installation
.......................................................... 31
........................... 53
.................................... 20
G
grab samples ..................................................... 40, 47
I
ice for sample preservation ...................................... 20
intake tubing
installing specifying type and length
............................................................. 27
................................. 38
K
keypad ..................................................................... 33
L
LED indicator ........................................................... 33
liquid sensor
calibration of diagnostic test volume determination with
...................................................... 48
.................................................... 68
................................. 44
M
maintenance
desiccant replacement distributor tube replacement pump rotor cleaning
pump tube replacement manhole installation manual operation
distributor
grab samples
pump
.................................................................. 41
................................................. 19
........................................................... 41
..................................................... 40
....................................... 62
.............................. 62
........................................... 60
..................................... 58
O
output signal
at each sample cycle
at program completion
......................................... 52
....................................... 52
79
Page 82
Index
P
password .................................................................. 48
preserving samples presets pump
..................................................................... 39
cleaning the rotor manual operation of tube installation tube replacement
.................................................. 20
............................................... 60
........................................... 41
.................................................. 59
............................................... 58
R
rotor .......................................................................... 60
S
safety
confined space precautions
general information sample history sample program
basic menu overview
bottle number
cascade sampling
default settings
history
intake tubing
multiple start/stop times
reviewing
starting or stopping
status sampler
components
manhole installation
mounting
replacement parts sampler setup
eight bottles
twenty four bottles
two or four bottles
.......................................................... 43
..................................................... 38
.................................................. 39
................................................................ 43
...................................................... 38
........................................................... 38
................................................................. 41
....................................................... 13
............................................................ 17
....................................................... 23
............................................ 11
.............................................. 52
............................................ 40
........................................... 19
.............................................. 69
............................................. 24
.............................................. 22
............................... 12
......................................... 35
..................................... 53
samples per bottle SD900 controller
advanced software overview diagnostic tests keypad passwords software overview turning on or off
serial connector
pin assignments setpoint sampling spanner bar stormwater sampling
first flush samples
menu overview
............................................................... 33
.................................................... 38
............................. 51
.................................................. 67
......................................................... 48
.............................................. 35
................................................. 33
................................................. 31
..................................................... 53
.............................................................. 19
.............................................. 54
.................................................. 51
T
templates. See presets troubleshooting
diagnostic tests
table
................................................................... 65
tubing replacement reminder tubing, distributor tubing, intake
installing tubing, pump
installation
replacement
tubing life
............................................................. 27
.................................................. 67
................................... 49
..................................................... 62
.......................................................... 59
....................................................... 58
........................................................... 57
V
variable intervals ...................................................... 55
volume
bottle
.................................................................. 37
calibration
manual calibration
verify sample volume
.......................................................... 44
............................................. 46
......................................... 47
80
Loading...