Hach-Lange MTC10130 User Manual

DOC022.52
User Manual
Gel filled ORP/Redox Probe: Model MTC10101, MTC10103, MTC10105, MTC10110, MTC10115 or MTC10130

Safety information

Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.

Specifications

Note: Specifications are subject to change without notice.
.80033
Specifications Details
Probe type Digital combination electrode with a non-refillable Ag/AgCl reference and a
built-in temperature sensor
Range ±1200 mV
Resolution 0.1 mV
Temperature accuracy ±0.3 °C (±0.54 °F)
Operating temperature range 0 to 80 °C (32 to 176 °F)
Storage temperature range 5 to 40 °C (41 to 104 °F)
Junction Open junction
Reference potential versus Standard Hydrogen Electrode
Reference electrolyte Electrode filling solution, 3M KCl
Reference type Ag/AgCl (3 M KCl)
Minimum sample depth 20 mm
Dimensions (standard) Diameter: 12 mm (0.47 in.)
Dimensions (rugged) Diameter: 46 mm (1.81 in.)
207 mV at 25 °C
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Length: 223 mm (8.73 in.)
Cable length: 5, 10, 15 or 30 m (16.40, 32.81, 49.21 or 98.42 ft)
Cable connection M12 digital output and connector compatible with HQd meters

Product overview

The MTC101 series probe is a non-refillable, gel-filled combination oxidation reduction potential (ORP/Redox) probe with a built-in temperature sensor (Figure 1). The MTC10101 or MTC10103 probe is available with a 1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The MTC10105, MTC10110, MTC10115 or MTC10130 probe
1
is available with a 5, 10, 15 or 30 m (16.40, 32.81, 49.21 or 98.42 ft) cable and is intended for field use. The probe measures absolute mV values in wastewater, drinking water and general applications.
Figure 1 Probe overview
1 Shroud (rugged model) 6 Locking ring (rugged model)
2 Probe storage cap 7 Rugged probe (5, 10, 15 or 30 meter cable)
3 Probe soaker bottle 8 Standard probe (1 or 3 meter cable)
4 Platinum electrode and temperature sensor 9 Probe storage cap or soaker bottle holder
5 Reference junctions and protective tape (rugged model)

Preparation for use

To prepare the probe for initial use:
1. If a rugged probe, remove the shroud and probe storage cap (refer to Remove the
shroud on page 11
2. If a standard probe, turn the probe soaker bottle cap counter-clockwise to loosen the cap. Remove the soaker bottle from the probe.
3. If a rugged probe, remove the protective tape from the reference junctions (refer to
Figure 1 on page 2). Dispose of the protective tape.
4. If the probe will be used immediately, prepare the probe for calibration or sample measurement.
5. If the probe will not be used immediately, store the probe (refer to Storage on page 11).
To prepare the probe for calibration or sample measurement:
1. If a rugged probe, remove the probe storage cap.
2. If a standard probe, turn the probe soaker bottle cap counter-clockwise to loosen the
cap. Remove the soaker bottle from the probe.
3. Rinse the reference junctions and electrode thoroughly with deionized water to remove the 3 M KCl solution completely. Blot dry with a lint-free cloth.
4. If a rugged probe, make sure that the shroud is installed before field use (refer to
Install the shroud on page 11).
Note: Damage to the sensing elements can occur if the shroud is not installed during field use. Damage under these conditions is not covered by the product warranty.
).
2

Calibration

Before calibration:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is configured to use the same calibration options.
To view the current calibration, push , select View Probe Data, then select View Current Calibration.
If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the Calibrate option.
Prepare the probe for use (refer to Preparation for use on page 2).
If a rugged probe, remove the shroud from the probe (refer to Remove the shroud on page 11).
Calibration notes:
Additional standards can be selected in the Calibration Options.
Do not dilute ORP/Redox standards. Use fresh ORP/Redox standard for calibration.
ZoBell’s redox potential is temperature dependent. The HQd calibration routine factors in this temperature dependency allowing accurate calibrations within the temperature range of 0 to 30 °C (32 to 86 °F). Light’s solution should be read at 25 °C (77 °F). Custom ORP/Redox calibration solution values and temperature are user­defined.
The calibration is recorded in the probe and the data log. The calibration is also sent to a PC, printer or flash memory stick if connected.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a calibration error occurs, refer to Troubleshooting on page 12.
1. Connect the probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn on the meter.
Calibration procedure:
2. Push Calibrate. The
display shows the ORP/Redox standard solution that is necessary for calibration.
3. Add the fresh ORP/Redox standard solution to a beaker or an appropriate container.
4. Rinse the probe with deionized water. Blot dry with a lint­free cloth.
5. Put the probe in the standard solution so that the temperature sensor is completely submerged. Stir gently. Shake the probe from side to side in the sample to refresh the reference junction.
3
6. Push Read.
Stir gently. The display will show "Stabilizing" and a progress bar as the probe stabilizes in the sample. The display shows the standard solution value and the mV offset when the reading is stable.
7. Push Done to view the calibration summary.
8. Push Store to accept the calibration and go back to the measurement mode. If a rugged probe, install the shroud on the probe (refer to
Install the shroud
on page 11).

Sample measurement

Before measurement:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
If complete traceability is necessary, enter a sample ID and operator ID before measurement. Refer to the HQd meter manual for more information.
Regular calibration is required for the best measurement accuracy (refer to Calibration on page 3).
Prepare the probe for use (refer to Preparation for use on page 2).
Make sure that the platinum disc is clean and smooth (refer to Clean the probe on page 11).
To deploy a rugged probe at a distance, toss the probe body with a gentle underhand throw. Do not swing the probe by the cable as this may cause injury to the user, will cause severe strain on the cable and will shorten the service life of the cell. Damage under these conditions is not covered by the product warranty.
Measurement notes:
Data is automatically stored in the data log when Press to Read or Interval is selected in the Measurement Mode. When Continuous is selected, data will only be stored when Store is selected.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a measurement error occurs, refer to Troubleshooting on page 12.
Measurement—direct method procedure:
4
1. Connect the
probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn the meter on.
2. To significantly reduce the stabilization time for reducing-type samples, put the platinum disc in Reducing Solution for ORP Electrodes for 3-10 minutes before the initial sample measurement.
3. Rinse the probe with the sample.
4. Put the probe in the sample and stir gently. Make sure that the reference junctions are completely submerged. Do not put the probe on the bottom or sides of the container. Shake the probe from side to side in the sample to refresh the reference junction.
5. Push Read. The display will show "Stabilizing" and a progress bar as the probe stabilizes in the sample. The display will show the lock icon when the reading stabilizes. If necessary for the application, record the sample pH and temperature.
6. Repeat steps 3-6 for additional measurements.
Temperature (°C) Electrode potential in mV (E
7. When
measurements are done, store the probe (refer to
Storage
on page
11).
Measurement—conversion to SHE reference procedure:
For some applications, it is customary to report redox potential readings relative to the standard hydrogen electrode (SHE), also called normal hydrogen electrode (NHE). To do this, select the value in Table 1
that corresponds to the temperature of the solution
measured. Substitute that value Eref into the equation and solve for Eh:
Eh = E + E
ref
where:
Eh = oxidation reduction potential of the sample relative to the SHE
E = potential developed by the ORP/Redox electrode
E
= potential developed by the reference electrode portion relative to the SHE
ref
(Table 1).
Table 1 shows the potentials, E
, developed by the reference electrode portion relative
ref
to the SHE at various temperatures.
Table 1 Standard potential of reference electrode
)
ref
80 163.1
75 167.7
5
Loading...
+ 11 hidden pages