1.1 Technical data for the 3798-S sc conductivity sensor .............................................................................................. 3
Section 2 General information ..................................................................................................................................... 5
2.1 General handling instructions .................................................................................................................................... 5
Section 3 General safety instructions ......................................................................................................................... 7
3.1 Possible sources of hazards...................................................................................................................................... 7
3.3 Electrical safety measures and fire prevention measures ......................................................................................... 8
3.4 Chemical safety measures ........................................................................................................................................ 9
3.5 Safety measures related to the flow of sample.......................................................................................................... 9
5.3 Sensor data logger .................................................................................................................................................. 15
5.4 The commands under SENSOR DIAG.................................................................................................................... 15
5.5 The commands under SENSOR SETUP ................................................................................................................ 16
5.6.1 Calibration in air (ZERO CAL)........................................................................................................................ 17
5.6.2 Calibration in air (ELECTRIC. SPAN) ............................................................................................................ 17
5.6.3 Calibration in the process (PROCESS SPAN)............................................................................................... 17
5.7 Adjusting the Temperature ...................................................................................................................................... 18
5.8 Calibrating two sensors simultaneously .................................................................................................................. 18
6.2 Cleaning the sensor................................................................................................................................................. 19
7.3 Important service data ............................................................................................................................................. 22
Section 8 Spare parts .................................................................................................................................................. 23
Section 9 Warranty and liability ................................................................................................................................. 25
Appendix A ModBUS Register Information ............................................................................................................. 29
1
Page 4
Table of contents
2
Page 5
Section 1Technical data
1.1Technical data for the 3798-S sc conductivity sensor
MaterialsStainless steel metal housing, PEEK
Enclosure ratingIP 68; stainless steel metal housing
Storage temperature
Sensor and controller
Cell constantK=2.35cm
Measuring range conductivity250 µS/cm ... 2.5 S/
Temp. measuring range–5 °C ... 50 °C
Sensor operating temperature–20 °C ... 50 °C
Conductivity response time< 2 s; T90
Temp. response time<2min; T90
Conductivity measuring accuracy± 1 % of the measured value displayed or ± 0.004 mS/
Temp. measuring accuracy± 0.2 °C
Reproducibility< 0.2 %
Sensitivity± 0.5 % of the end value of the measuring range
Sensor power< 7 W
Calibration
Max. immersion depth / pressure
for the sensor
–20 °C ... 60 °C; 95 % relative humidity, non-condensing
–1
cm
Zero value calibration in air.
Fixed value calibration with defined resistance or with standard solution
20 m / 2 bar
cm
Maximum flow speed4 m/s
Sensor interfaceMODBUS
Sensor cable10 m, hard wired, polyurethane
Sensor weight< 1 kg
Sensor dimensions (Ø × L) 43 × 370 mm
Fastening
• Immersed pipe
• Chain
Specifications are subject to change without notice.
3
Page 6
Technical data
4
Page 7
Section 2General information
2.1General handling instructions
Attention!
The sensor will only work correctly when the tip of the probe is completely
immersed in liquid. Ensure the tip of the probe always remains underwater
even when the water level fluctuates.
2.2Applications
CAUTION!
Potential danger with contact with chemical/biological substances. Working
with chemical samples, standards and reagents can be dangerous. Make
yourself familiar with the necessary safety procedures and the correct
handling of the chemicals before use and read and follow all relevant safety
data sheets.
The sensor facilitates the straightforward and exact determination of the
conductivity of soiled to heavily soiled aqueous solutions. The system is specially
designed for use in municipal and industrial waste water.
2.3Basic principles
Typical applications include
•Inlet and / or outlet of a sewage treatment plant
•Surface water if within the measuring range (> 250
Various different possible installations enable the system to be adapted to a very
wide range of conditions.
The electrolytic conductivity is the ability of a liquid to conduct an electrical current
(conductivity is the opposite of resistance). In metals, electrical currents are
passed by the movement of electrons, in liquids by the movement of ions. The
conductivity of a liquid depends on the one hand on the ionic concentration, on
the other hand on the temperature of the liquid.
To obtain the real conductivity of the liquid (in
measured
geometry of the probe and that is called the "cell constant or K" (
To be able to make a comparison between the measurements made at different
temperatures, the measurement must be converted to a reference temperature
(as a rule 25 °C).
1
/R (in S) must be multiplied by a coefficient that depends on the
C= K/R (S/cm)
S
/cm), the conductivity value
µS
/cm).
1
/
).
cm
5
Page 8
General information
1
4
5
2
3
6
7
This temperature-dependence, expressed in [% / °C], is termed the temperature
coefficient (α).
2.4Measuring principle
C
= CT [1 + α (T – T
Tref
C
C
T
α: coefficient of the liquid temperature (
: conductivity adjusted to the reference temperature
Tref
: conductivity measured at T
T
: reference temperature (as a rule 25 °C)
ref
ref
)]
– 1
%
/°C)
The tip of the probe comprises two coils that are completely insulated from the
surrounding medium.
Primary coil
(transmitter):
An AC voltage is applied to the primary coil and produces an
alternating electromagnetic field in the surrounding liquid. This
magnetic field generates an electrical current in the liquid.
Secondary coil
(receiver):
The secondary coil determines the current produced by the
ion movements and calculates the conductivity of the liquid
from the current.
1.Receiver
2.Oscillator
3.Cable
4.Insulation
5.Receiver coil
6.Transmitter coil
7.Field line
The electrical insulation between liquid and sensor (magnetic coupling) has
advantages compared to the conventional method of using metal electrodes:
•no polarisation, for this reason the measuring range is larger
•high mechanical and chemical resistance
•possibility of taking measurements in soiled liquids
6
Page 9
Section 3General safety instructions
Prior to unpacking, commissioning or operating the instrument, read all of this
manual.
Please pay particular attention to all instructions on hazards and safety. Otherwise
there is a risk of serious injury to the operator or damage to the instrument, or
pollution.
The sensor is only allowed to be installed and used as per the instructions in this
manual.
3.1Possible sources of hazards
During the operation or calibration of the sensor, there exist the following sources
of hazards if the safety instructions are not observed:
•Potentially hazardous materials (buffer solutions, flow of sample)
In all circumstance observe the safety data sheets and the applicable health and
safety instructions.
3.2Safety symbols
This symbol, if present on the instrument, refers to information in the operating instructions on safe operation and / or
instructions that provide safety information.
This symbol, if present on a housing or a protective cover, identifies the risk of an electric shock (which may under
certain circumstances be fatal). Only personnel qualified for working on hazardous voltages are allowed to open the
enclosure or remove the protective cover.
This symbol, if present on the instrument, identifies the location of a fuse or current limit.
This symbol, if present on the instrument, identifies a part that may become hot and must not be touched without
taking precautions.
This symbol, if present on the instrument, indicates the presence of components that could be damaged by
electrostatic discharge. Appropriate precautions are to be taken.
This symbol, if present on the instrument, indicates the presence of dangerous chemical substances. Chemicals are
only allowed to be handled and maintenance on devices for supplying chemicals is only allowed to be performed by
personnel qualified and trained for working with chemicals.
All stickers and labels on the instrument are to be observed. Otherwise injuries,
pollution or damage to the instrument may occur.
7
Page 10
General safety instructions
This symbol, if present on the instrument, indicates that safety glasses must be worn.
This symbol, if present on the instrument, identifies the location of the connection for the protective earth (ground).
As of 12 August 2005, electrical appliances marked with this symbol are no longer allowed to be disposed of in
Europe in unsorted household or industrial waste. As per the applicable regulations, from this date on consumers in
the EU must return old appliances to the manufacturer for disposal. This disposal is free of charge for the consumer.
Note: You can obt ain instructions on the correct disposal of all (marked and unmarked) electrical products that have
been supplied or manufactured by Hach-Lange from your local Hach-Lange sales office.
3.3Electrical safety measures and fire prevention measures
The following safety instructions must be observed during the installation
and repair of cables that carry electrical power:
DANGER!
Sensors and controller are designed for compliance with the U.S. and
Canadian NEC as well as the European low voltage directive. No internal
electrical or electronic parts are allowed to be modified in any way, as this
could render the CE conformity void.
WARNING!
Only qualified experts may perform the tasks described in this section of the
manual, while adhering to all locally valid safety regulations.
•Prior to maintenance or repair of the instrument, isolate it from the power
supply.
•When making electrical connections, all applicable local and national
regulations are to be met.
•The use of earth leakage trips is strongly recommended.
•The instrument must be correctly earthed for correct operation.
8
Page 11
3.4Chemical safety measures
CAUTION!
Reference and standard solutions are used for the calibration. Some of
these compounds are toxic or caustic.
Potential danger with contact with chemical/biological substances. Working
with chemical samples, standards and reagents can be dangerous. Make
yourself familiar with the necessary safety procedures and the correct
handling of the chemicals before use and read and follow all relevant safety
data sheets.
Physical contact with a calibration solution and inhalation of vapours from a
calibration solution are to be avoided or limited to an absolute minimum.
3.5Safety measures related to the flow of sample
The assessment of the possible hazards from the individual sample flows is the
responsibility of the user. Suitable safety measures are to be taken to avoid any
unnecessary contact with a flow of sample of unknown composition in relation to
the hazards due to traces of chemicals, radiation or biological effects.
General safety instructions
9
Page 12
General safety instructions
10
Page 13
Section 4Installation
1
2
3
4
5
6
4.1Connecting sensor cable
You can connect the sensor cable to the controller very easily using the plug.
Retain the protective cap for the socket in case you need to remove the sensor in
the future. Connecting cables are available in the lengths 5 m, 10 m, 15 m, 20 m,
30 m and 50 m. From a length of 100 m a bus termination box must be integrated
(see Section 8 Spare parts).
Fig. 1Connection of the sensor plug to the controller
Notice!
The sensor will only work correctly when the tip of the probe is completely
immersed in liquid. Ensure the tip of the probe always remains underwater
even when the water level fluctuates.
CAUTION!
Potential danger with contact with chemical/biological substances. Working
with chemical samples, standards and reagents can be dangerous. Make
yourself familiar with the necessary safety procedures and the correct
handling of the chemicals before use and read and follow all relevant safety
data sheets.
Requirements
•Ensure the sensor does not collide with other instruments or objects in the
tank. In this way you will avoid damaging the sensor.
•Fasten the sensor to the nearest wall with a minimum spacing of 0.5 m.
4.2.1 Installation dimensions
12
Page 15
Fig. 3Installation examples
Installation
1.Rim mounting chain bracket, PVC or stainless steel2.Rim mounting immersed tube, PVC or stainless steel
13
Page 16
Installation
14
Page 17
Section 5Operation
5.1Operating the sc controller
The sensor can be operated with all sc controllers. Prior to using the sensor,
familiarise yourself with the principle of operation of your controller. Learn how to
navigate in the menus and run appropriate functions.
5.2Sensor setup
When you connect the sensor for the first time, the serial number of the sensor is
displayed as the sensor name. You can change the sensor name as follows:
1. Open the MAIN MENU.
2. Choose SENSOR SETUP and accept.
3. Choose the related sensor and accept.
4. Choose CONFIGURE and accept.
5. Choose EDIT NAME and accept.
6. Edit the name and accept to return to the SENSOR SETUP menu.
In the same way complete your system configuration using the commands as per
table 5.5 "The commands under SENSOR SETUP".
5.3Sensor data logger
A data memory and event memory per sensor are available via the sc controller.
While measured data are saved in the data memory at stipulated intervals, the
event memory collects numerous events such as configuration changes, alarms
and warning conditions. Both the data memory and the event memory can be
read out in CSV format. For information on how you can download the data,
please see the controller manual.
5.4The commands under SENSOR DIAG
SENSOR DIAG
SELECT SENSOR (for several sensors)
ERROR LISTList of all errors that have occurred (see Section 7.1 "Error messages")
WARNING LIST
List of all warnings that have occurred
(see Section 7.2 "Warnings")
15
Page 18
Operation
5.5The commands under SENSOR SETUP
SENSOR SETUP
SELECT SENSOR (for several sensors)
CALIBRATE
ZERO CAL
ELECTRIC. SPAN
PROCESS SPAN
PROCESS TEMP
CAL CONFIG
OUTPUT_MODE
CAL REMINDER
CORR FACTORS
SET CAL DEFLTReturn to the factory settings after prompt for confirmation.
CONFIGURE
EDIT NAMEEnter a 10-character name.
PARAMETERSChoose a parameter.
DEGREES C-FChoose between degrees Celsius or Fahrenheit.
T-COMPENSATIONChoose between LINEAR and NONE.
T-SENSOR
AUTOMATIC
MANUALEnter a value.
FILTEREnter a value.
LOG. DELAY
LOG INTERVALChoose between the values available or DISABLED.
TEMP INTERVALChoose between the values available or DISABLED.
MAINS FREQ.Enter the mains frequency.
SET DEFAULTSReturns to the factory settings after a prompt for confirmation.
DIAG/TEST
PROBE INFOProvides information on driver, software and serial number
CAL DATAProvides information on GAIN, Offset (T) and GAIN CORR.
SIGNALSProvides information on raw data measured.
COUNTERSProvides information on the operating time since the last calibration
TEST/MAINTDisable OUTPUT during test or maintenance
SERVICEReset Service Counter
Eliminates sensor offset
See 5.6.1 "Calibration in air (ZERO CAL)".
Calibration with a defined resistance.
See 5.6.2 "Calibration in air (ELECTRIC. SPAN)".
Calibration with a reference solution.
See 5.6.3 "Calibration in the process (PROCESS SPAN)".
Calibration of the temperature.
See 5.7 "Adjusting the Temperature".
Choose between:
ACTIVE, output signal follows the input signal;
HOLD, last measured value and output signal are held;
TRANSFER, fixed value is output to the peripherals; and
CHOICE
You can set when the next calibration is to be performed. The
controller then automatically indicates when the next calibration is due.
Choose between
TEMP OFFSET,
GAIN CORR and
GAIN VALUE,
16
Page 19
5.6Sensor calibration (conductivity)
The sensor has been permanently calibrated and operates with great enough
precision and stability that calibration is rarely necessary.
Calibrate the sensor
•as required (measured value outside the permitted tolerance,
see Section 7.3 "Important service data") or
•in accordance or agreement with the authorities.
5.6.1 Calibration in air (ZERO CAL)
1. On the controller, open the menu SENSOR SETUP => CALIBRATE => ZERO
CAL and accept.
2. Remove the sensor from the tank, clean and dry.
3. Accept.
Operation
4. Wait until the controller displays CAL COMPLETE.
5. Confirm that you have replaced the sensor in the sample flow and accept.
5.6.2 Calibration in air (ELECTRIC. SPAN)
1. Remove the probe from the flow of sample, clean and dry.
2. Connect a defined resistance (in range 5 Ohms - 5 kOhms) and accept to
continue.
3. Enter the value for the resistance and accept.
4. Wait until the value has stabilised.
5. Accept the value.
6. Replace the probe in the flow of sample and accept.
5.6.3 Calibration in the process (PROCESS SPAN)
1. Hold the cleaned probe in the solution and accept to continue.
2. Accept when the measured value has stabilised.
3. Enter the value and accept.
4. Replace the probe in the flow of sample and accept.
17
Page 20
Operation
5.7Adjusting the Temperature
1. From the Main Menu, select SENSOR SETUP and confirm.
2. Highlight the appropriate sensor if more than one sensor is attached and
confirm.
3. Select CALIBRATE and confirm.
4. Select PROCESS TEMP and confirm.
5. Press ENTER when Stable, TEMP: XX.X is displayed. confirm to continue.
6. Adjust the Reading XX.X °C with the keypad and confirm.
7. CAL COMPLETE, OFFSET: X.X °C, confirm to continue.
8. MOVE PROBE TO PROCESS is displayed. Confirm.
5.8Calibrating two sensors simultaneously
1. Start by calibrating the first sensor and when you arrive at the point at which
you are prompted to "PRESS ENTER WHEN STABLE".
2. Press the BACK key.
3. Select EXIT and press accept. The instrument returns to the display of the
measurements. The measured value for the sensor to be calibrated starts to
flash.
4. Start the calibration of the other sensor and when you again arrive at the point
at which you are prompted to "PRESS ENTER WHEN STABLE".
5. Press the BACK key.
6. Select EXIT and press accept. The instrument returns to the display of the
measurements. The measured values for both sensors start to flash.
7. To return to the calibration menu for the individual sensors, press the Menu
key, select SENSOR SETUP and accept. Choose the required sensor and
accept.
When calibration is complete, accept.
18
Page 21
Section 6Maintenance
6.1Maintenance schedule
The following table reflects experience and may, depending on the sector and
application, vary significantly from actual requirements.
Maintenance task90 daysannual
Clean sensorx
Check sensor for damagex
Calibration (if necessary)
You can set the calibration interval in the sensor setup. The controller then
reminds you when calibration is due.
6.2Cleaning the sensor
If necessary as per
agreement with the
authorities
Clean the sensor with a jet of water. If there is still soiling present, use a soft,
damp cloth.
19
Page 22
Maintenance
20
Page 23
Section 7Faults, causes, rectification
7.1Error messages
Possible sensor errors are displayed by the controller.
Table 1: Error messages
Error displayedCauseRectification
*****
SENSOR MISSING
FFFFFFFFFFFFF
TEMP TOO LOWMeasured temperature < –5 °C
TEMP TOO HIGH
COND TOO LOWConductivity < 100
COND TOO HIGHConductivity > 500 mS/
RES. TOO LOWResistance < 2
RES. TOO HIGHResistance > 10 k
No communication with the
controller
No communication with the
controller
Measured temperature >
+100 °C
µS
/
cm
cm
ΩPlease contact service.
ΩPlease contact service.
Check the connection to the controller
Check the cable to the controller
Check the connection to the controller
Check the cable to the controller
Ensure that the
medium temperature is > –5 °C.
Ensure that the
medium temperature is < +100 °C.
Ensure that the
conductivity is > 100 µS/cm.
Ensure that the
conductivity is < 500 mS/cm.
7.2Warnings
Possible warning messages are displayed by the controller.
Table 2: Warnings
Error displayedCauseRectification
CAL TOO OLD
HUMIDITY BAG
The last calibration was more than
180 days ago.
The desiccant bag is more than
1000 days old.
Calibrate the sensor
Please contact service.
21
Page 24
Faults, causes, rectification
7.3Important service data
DataMinimumMaximum
Electrical gain correction95 %105 %
CAL DATA
Signals
Counter
MODBUS STATSNumber of communication errors0< 1 %
Temperature offset correction– 5 °C+ 5 °C
Cell constant2.502.00
Output voltage
Raw measured data– 1 %+ 1 %
Desiccant bag
Operating time1000 days
User Manual ..................................................................................................................................DOC023.52.03252
HACH LANGE GmbH warrants that the product supplied is free of material and
manufacturing defects and undertakes the obligation to repair or replace any
defective parts at zero cost.
The warranty period for instruments is 24 months. If a service contract is taken out
within 6 months of purchase, the warranty period is extended to 60 months.
With the exclusion of the further claims, the supplier is liable for defects including
the lack of assured properties as follows: all those parts that, within the warranty
period calculated from the day of the transfer of risk, can be demonstrated to have
become unusable or that can only be used with significant limitations due to a
situation present prior to the transfer of risk, in particular due to incorrect design,
poor materials or inadequate finish will be improved or replaced, at the supplier's
discretion. The identification of such defects must be notified to the supplier in
writing without delay, however at the latest 7 days after the identification of the
fault. If the customer fails to notify the supplier, the product is considered
approved despite the defect. Further liability for any direct or indirect damages is
not accepted.
If instrument-specific maintenance and servicing work defined by the supplier is to
be performed within the warranty period by the customer (maintenance) or by the
supplier (servicing) and these requirements are not met, claims for damages due
to the failure to comply with the requirements are rendered void.
Any further claims, in particular claims for consequential damages cannot be
made.
Consumables and damage caused by improper handling, poor installation or
incorrect use are excluded from this clause.
HACH LANGE GmbH process instruments are of proven reliability in many
applications and are therefore often used in automatic control loops to provide the
most economical possible operation of the related process.
To avoid or limit consequential damage, it is therefore recommended to design the
control loop such that a malfunction in an instrument results in an automatic
change over to the backup control system; this is the safest operating state for the
environment and the process.
Villa 14 – Rue 2 Casa
Plaisance
Quartier Racine Extension
MA-Casablanca 20000
Tél. +212 (0)522 97 95 75
Fax +212 (0)522 36 89 34
info-maroc@hach-lange.com
www.hach-lange.ma
28
Page 31
Appendix AModBUS Register Information
Table A-3 Sensor Modbus Registers
Tag NameRegister #Data Type
Measurement mS/cm40001Float2RConductivity in mS/cm
Measurement Ohm.cm40003Float2RResistivity Ohm.cm
Measurement temperature40005Float2RTemperature
Measurement uScm40007Float2RConductivity in uS/cm
Measurement S/m40009Float2RConductivity in S/m
Measurement mS/m40011Float2RConductivity in mS/m
Measurement KOhm.cm40013Float2RResistivity KOhm.cm
Measurement Ohm.m40015Float2RResistivity Ohm.m
Measurement Ohm.m (2)40017Float2RResistivity Ohm.m2
AutoRange S/cm40019Integer1RAuto Ranging redirection
AutoRange S/m40020Integer1RAuto Ranging redirection of Sm
AutoRange Ohm.cm40021Integer1RAuto Ranging redirection of Ohm.cm
AutoRange Ohm.m40022Integer1RAuto Ranging of Ohm.m
measurement raw temperature40023Float2RRaw Temperature
Conductivity unit40025Integer1RConductivity unit
Temperature unit40026Bit1R/WTemperature unit
Output Mode40027Integer1R/WOutputMode
Sensorname[0]40028Integer1R/Wsensorname[0]
Sensorname[1]40029Integer1R/Wsensorname[1]
Sensorname[2]40030Integer1R/Wsensorname[2]
Sensorname[3]40031Integer1R/Wsensorname[3]
Sensorname[4]40032Integer1R/Wsensorname[4]
Sensorname[5]40033Integer1R/Wsensorname[5]
Software Version (float)40034Float2R/WSoftware version
Driver Version (float)40036Float2R/WDriver version
Mains Frequency 50Hz40038Bit1R/WMain Frequency
Function code40039Integer1R/WFunction Code
Next state40040Integer1R/WNext Step
Password40041Password1R/WPassword
Serial number[1]40042Integer1R/WSerial number[0]
Serial number[2]40043Integer1R/WSerial number[1]
Serial number[3]40044Integer1R/WSerial number[2]
Conductivity parameter40045Bit1R/W&CMD_kunit
Temperature unit40046Bit1R/W&CMD_tunit
Offset correction40047Float2R/WResistivity Offset
Electrical Calibration Resistance40049Float2R/WResistivity Adjust vaue
Electrical Slope40051Float2R/WElectrical slope
Process Slope40053Float2R/WProcess slope
Main Calibration Adjust Value40055Float2R/WCal Conductivity Adjust Value
Second. Calibration Adjust Value40057Float2R/WCal Temperature Adjust Value
Temporary Meas.[0]40059Float2R/WTemporary Measurement[0]
Temporary Meas.[1]40061Float2R/WTemporary Measurement[1]
Lengt
h
R/WDescription
29
Page 32
ModBUS Register Information
Table A-3 Sensor Modbus Registers
Constant cell40063Float2R/WConstant cell
Temperature Compensation40065Bit1R/WTemperature Compensation
Coefficient Compensation40066Float2R/WCompensation Coefficient
Temperature Reference 40068Float2R/WTemperature Reference
AutomaticTemperature 40070Bit1R/WAutomaticTemperature
Manual Temperature40071Float2R/WManual Temperature
Temperature Offset40073Float2R/WTemperature Offset
---40075Integer1R/W&RS_tgMainMeas
---40076Integer1R/W&RS_tgSecondMeas
---40077Integer1R/W&RS_tgCalMainMeas
---40078Integer1R/W&RS_tgCalSecondMeas
---40079Integer1R/W&RS_tgCalMainAdjValue
---40080Integer1R/W&RS_tgCalSecondAdjValue
---40081Integer1R/W&RS_tgTemporary0
---40082Integer1R/W&RS_tgTemporary1
---40083Integer1R/W&RS_tgTempOffsetCorr
---40084Integer1R/W&RS_tgTempRef
---40085Integer1R/W&RS_tgTempManual
---40086Integer1R/WAnalogue Output Command
Serial Number String[0]40087Integer1R/W&RS_sn_string[0]
Serial Number String[2]40088Integer1R/W&RS_sn_string[2]
Serial Number String[4]40089Integer1R/W&RS_sn_string[4]
Serial Number String[6]40090Integer1R/W&RS_sn_string[6]
Serial Number String[8]40091Integer1R/W&RS_sn_string[8]
Serial Number String[8]40092Integer1R/W&RS_sn_string[10]
---40096Integer1R/W&MESS_cal_code
Delay from last Calibration40097Integer1RDelay from last Calibration
Time from Start up40098Integer1RTime from Start up
Time of Humidity Bag40099Integer1RTime of Humidity Bag
Conductivity Log Interval40100Integer1RConductivity Log Interval
Temperature Log Interval40101Integer1RTemperature Log Interval
30
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.