Hach-Lange 2100Q, 2100Qis User Manual

DOC022.52.80041
2100Q and 2100Qis
04/2013
User Manual
Table of Contents
Specifications..................................................................................................................................................................................3
General information.....................................................................................................................................................................3
Safety information..............................................................................................................................................................................4
Precautionary labels..........................................................................................................................................................................4
Certification........................................................................................................................................................................................4
Product overview...............................................................................................................................................................................5
Product components..........................................................................................................................................................................5
Installation.........................................................................................................................................................................................6
Install the battery................................................................................................................................................................................6
User interface and navigation...............................................................................................................................................7
User interface.....................................................................................................................................................................................7
Display description.............................................................................................................................................................................8
Navigation..........................................................................................................................................................................................8
Startup.................................................................................................................................................................................................8
Turn the meter on and off..................................................................................................................................................................8
Change the language ........................................................................................................................................................................9
Change the date and time .................................................................................................................................................................9
Standard operation......................................................................................................................................................................9
Use a sample ID................................................................................................................................................................................9
Use an operator ID.............................................................................................................................................................................9
Calibrate the turbidimeter with StablCal® Standards.......................................................................................................................10
Turbidity measurement....................................................................................................................................................................10
Measurement notes..................................................................................................................................................................10
Turbidity measurement procedure............................................................................................................................................11
Data management.......................................................................................................................................................................12
About stored data.............................................................................................................................................................................12
View data log...................................................................................................................................................................................12
Delete data log.................................................................................................................................................................................12
1
Table of Contents
Send stored data..............................................................................................................................................................................12
Advanced operation..................................................................................................................................................................12
Display contrast...............................................................................................................................................................................12
Power management.........................................................................................................................................................................13
Set the sound options......................................................................................................................................................................13
Security options...............................................................................................................................................................................13
Turn security options on...........................................................................................................................................................13
View meter information....................................................................................................................................................................13
Calibration........................................................................................................................................................................................13
Calibration options....................................................................................................................................................................14
Calibration standard overview..................................................................................................................................................14
StablCal® RapidCal™ calibration..............................................................................................................................................15
Verification options...................................................................................................................................................................16
Calibration verification (Verify Cal)...........................................................................................................................................16
Reading modes................................................................................................................................................................................17
Apply silicone oil to a sample cell....................................................................................................................................................17
Indexing a single cell........................................................................................................................................................................17
Maintenance...................................................................................................................................................................................19
Clean the meter...............................................................................................................................................................................19
Store the sample cells......................................................................................................................................................................19
Replace the battery..........................................................................................................................................................................19
Replace the lamp.............................................................................................................................................................................19
Troubleshooting..........................................................................................................................................................................22
Replacement parts and accessories...............................................................................................................................23
Replacement parts ..........................................................................................................................................................................23
Accessories .....................................................................................................................................................................................23
Index....................................................................................................................................................................................................25
2

Specifications

Specifications are subject to change without notice.
Specification Details
Measurement method
Regulatory 2100Q: Meets EPA Method 180.1
Lamp source 2100Q: Tungsten filament lamp
Range 0–1000 NTU (FNU)
Accuracy ±2% of reading plus stray light from 0–1000 NTU (FNU)
Repeatability ±1% of reading or 0.01 NTU (FNU), whichever is greater
Resolution 0.01 NTU on lowest range
Stray light ≤ 0.02 NTU (FNU)
Signal averaging Selectable on or off
Detector Silicon Photodiode
Reading modes Normal (Push to Read), Signal Averaging or Rapidly Settling
Calibration options
Calibration logger Records the last 25 successful calibrations
Verification logger Logs the last 250 successful verifications
Data logger 500 records
Ratio turbidimetric determination using a primary nephelometric light scatter signal (90°) to the transmitted light scatter signal.
2100Qis: Meets ISO 7027
2100Qis: Light-emitting diode (LED) at 860 nm
Turbidity
Single step RapidCal™ for Low-Level Regulatory Reporting from 0–40 NTU (FNU)
Full range calibration from 0–1000 NTU (FNU)
Calibration to degrees of turbidity
Specification Details
Power requirement
Operating conditions
Storage conditions
Interface Optional USB
Sample required 15 mL (0.5 oz.)
Sample cells Round cells 60 x 25 mm (2.36 x 1 in.) borosilicate glass with
Dimensions 22.9 x 10.7 x 7.7 cm (9.0 x 4.2 x 3.0 in.)
Weight 530 g (1.17 lb) without batteries
Meter enclosure rating
Protection class Power supply: Class II
Certification CE certified
Warranty 1 year (EU: 2 years)
AC 100–240 V , 50/60 Hz (with power or USB/power module)
4 AA alkaline batteries
Rechargeable NiMH (for use with USB/power module)
Temperature: 0 to 50 °C (32 to 122 °F)
Relative Humidity: 0–90% at 30 °C, 0–80% at 40 °C, 0–70% at 50 °C, noncondensing
–40 to 60 °C (–40 to 140 °F), instrument only
screw caps
620 g (1.37 lb) with four AA alkaline batteries
IP67 (closed lid, battery and module compartment excluded)

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.
English
3

Safety information

N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any manner other than that specified in this manual.

Use of hazard information

D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation that may result in minor or moderate injury.
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis.
W A R N I N G
C A U T I O N
N O T I C E

Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential injury. If on the instrument, refer to the instruction manual for operation or safety information.
This symbol indicates that a risk of electrical shock and/or electrocution exists.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.

Certification

Canadian Radio Interference-Causing Equipment Regulation, IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne sur les équipements provoquant des interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules. Operation is subject to the following conditions:
1. The equipment may not cause harmful interference.
4
English
2. The equipment must accept any interference received, including
interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:
1. Disconnect the equipment from its power source to verify that it is or
is not the source of the interference.
2. If the equipment is connected to the same outlet as the device
experiencing interference, connect the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the
interference.
5. Try combinations of the above.

Product overview

The 2100Q and 2100Qis portable turbidimeters measure turbidity from 0 to 1000 NTU (FNU). Primarily for field use, the portable meter operates on four AA batteries. Data can be stored and transferred to a printer, computer or USB storage device.
Figure 1 Product overview
1 Power on or off 5 Alignment arrow
2 Backlight keys (+ and -) 6 Module
3 Sample cell holder with lid 7 Lamp compartment
4 Attachment for lanyard 8 Battery compartment

Product components

Refer to Figure 2 to make sure that all components have been received. If any of these items are missing or damaged, contact the manufacturer or a sales representative immediately.
English 5
Figure 2 2100Q and 2100Qis components
1 2100Q or 2100Qis turbidimeter 6 Silicone oil
2 Carrying case 7 20, 100 and 800 NTU StablCal
3 User manual and Quick reference
guide
4 Oiling cloth 9 StablCal 10 NTU verification
5 1" sample cell (10 mL) with cap
(pk/6)
calibration standards
8 AA alkaline batteries (pk/4)
standard

Installation

Install the battery

W A R N I N G
Explosion hazard. An expired battery can cause hydrogen gas buildup inside the instrument. Replace the battery before it expires. Do not store the instrument for long periods with a battery installed.
Potential fire hazard. Use only alkaline or nickel metal hydride batteries (NiMH) in the meter. Other battery types or incorrect installation can cause a fire. Never mix battery types in the meter.
The battery compartment is not waterproof. If the battery compartment becomes wet, remove and dry the batteries and dry the interior of the compartment. Check the battery contacts for corrosion and clean them if necessary.
When using nickel metal hydride (NiMH) batteries, the battery icon will not indicate a full charge after freshly charged batteries have been inserted (NiMH batteries are 1.2 V versus 1.5 V for alkaline batteries). Even though the icon does not indicate complete charge, 2300 mAH NiMH batteries will achieve 90% of instrument operation lifetime (before recharge) versus new alkaline batteries.
To avoid potential damage to the meter from battery leakage, remove the meter batteries prior to extended periods of non-use.
The meter can be powered with AA alkaline or rechargeable NiMH batteries. To conserve battery life, the meter will power off after 10 minutes of inactivity, the backlight powers off after 30 seconds. This time can be changed in the Power Management menu.
Note: Rechargeable batteries will only be recharged with the USB/power module. Refer to the module documentation for further information.
For battery installation refer to Figure 3.
W A R N I N G
N O T I C E
N O T I C E
N O T I C E
6
English
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
1. Remove the battery cover.
2. Install 4 AA alkaline or 4 AA nickel metal hydride (NiMH) batteries.
Make sure that the batteries are installed in the correct orientation.
3. Replace the battery cover.
Figure 3 Battery installation

User interface and navigation

User interface

Figure 4 Keypad description
1 SETTINGS key: select menu
options for setting up the meter
2 CALIBRATION key: shows
calibration screen, start calibration, select cal options
3 DOWN key: scroll through menus,
enter numbers and letters
4 LEFT key (contextual): access for
calibration verification, cancels or exits the current menu screen to the previous menu screen
5 UP key: scroll through menus, enter
numbers and letters
6 RIGHT key (contextual): read
turbidity sample, selects or confirms options, opens/jumps to sub-menus
7 DATA MANAGEMENT key: view,
delete or transfer stored data
English 7

Display description

The measurement screen shows the turbidity, unit, calibration status, date and time, operator ID (if setup) and sample ID (if setup). Refer to
Figure 5.
Figure 5 Single screen display
1 Operator identification 9 NTU (Nephelometric Turbidity Unit)
2 Sample identification 10 Reading mode: Rapidly Settling
3 Stability or display lock indicator 11 Reading mode: Signal Average
4 Calibration status indicator
(Calibration OK=pass)
5 Calibration status indicator
(Calibration ?=fail)
6 Parameter title 14 Read (contextual: OK, Select)
7 AC power icon 15 Options (contextual)
8 Battery icon 16 Verification calibration
or FNU (Formazin Turbidity Unit)
Turbidity (Target icon)
(X-bar icon)
12 Time
13 Date

Navigation

The meter contains a Settings menu, Reading Options menu, Calibration Options menu and Calibration Verification Options menu to change various options. Use the UP and DOWN keys to highlight different options. Push the RIGHT key to select an option. There are two ways to change options:
1. Select an option from a list: Use the UP and DOWN keys to select an
option. If check boxes are shown, more than one option can be selected. Push the LEFT key under Select.
Note: To deselect check boxes, push the LEFT key under Deselect.
2. Enter an option value using the arrow keys: Push the UP and DOWN keys to enter or change a value.
3. Push the RIGHT key to advance to the next space.
4. Push the RIGHT key under OK to accept the value.

Startup

Turn the meter on and off

Push the ON/OFF key to turn on or turn off the meter. If the meter
does not turn on, make sure that the batteries, or the module, are properly installed or that the AC power supply is properly connected to an electrical outlet.
Note: The Auto-Shutoff option can also be used to turn off the meter. Refer to
Power management on page 13.
8 English

Change the language

There are three options to set the language:
• The display language is selected when the meter is powered on for the first time.
• The display language is selected when the power key is pushed and held.
• The language can be changed from the Settings menu.
1. Select a language from the list. Confirm with OK.
2. Push Done when the update is complete.

Change the date and time

The date and time can be changed from the Date & Time menu.

Standard operation

Use a sample ID

The sample ID tag is used to associate readings with a particular sample location. If assigned, stored data will include this ID.
1. Select Sample ID in the Settings menu.
2. Select, create or delete a sample ID:
Option Description
Current ID Select an ID from a list. The current ID will be
associated with sample data until a different ID is selected.
Create a New SampleIDEnter a name for a new sample ID.
1. Push the SETTINGS key and select Date & Time.
2. Update the time and date information:
Option Description
Format Select one of the formats for the date and time:
yyyy-mm-dd 24h
yyyy-mm-dd 12h
dd-mm-yyyy 24h
dd-mm-yyyy 12h
mm/dd/yyyy 24h
mm/dd/yyyy 12h
Date Enter the current date
Time Enter the current time
The current date and time will be shown on the display.
After the date and time setup, the meter is ready to take a reading.
Delete Sample ID Delete an existing sample ID.

Use an operator ID

The operator ID tag associates readings with an individual operator. All stored data will include this ID.
1. Select Operator ID in the Settings menu.
2. Select, create or delete an operator ID:
Option Description
Current ID Select an ID from a list. The current ID will be
Create a New Operator ID
Delete Operator ID Delete an existing operator ID.
associated with sample data until a different ID is selected.
Enter a name for a new operator ID (maximum 10 names can be entered).
English 9

Calibrate the turbidimeter with StablCal® Standards

Note: For best accuracy use the same sample cell or four matched sample cells for all readings during calibration. Insert the sample cell in the instrument cell compartment so the diamond or orientation mark aligns with the raised orientation mark in front of the cell compartment.
1. Push the CALIBRATION key to
enter the Calibration mode. Follow the instructions on the display.
Note: Gently invert each standard before inserting the standard.
2. Insert the 20 NTU StablCal Standard and close the lid.
Note: The standard to be inserted is bordered.
3. Push Read. The display shows Stabilizing and then shows the result.
4. Repeat Step 2 and 3 with the 100 NTU and 800 NTU StablCal Standard.
Note: Push Done to complete a 2 point calibration.
5. Push Done to review the calibration details.
6. Push Store to save the results. After a calibration is complete, the meter automatically goes into the Verify Cal mode. Refer to Verification
calibration.

Turbidity measurement

W A R N I N G
Potential explosion and fire hazard. This turbidimeter is designed for water based samples. Do not measure solvent or combustible based samples.
Readings can be taken with the Normal reading mode, Signal Average mode or in the Rapidly Settling Turbidity mode. Refer to Reading modes on page 17 for more information. For accurate turbidity readings use clean sample cells and remove air bubbles (degassing).

Measurement notes

Proper measurement techniques are important in minimizing the effects of instrument variation, stray light and air bubbles. Use the following measurement notes for proper measurements.
Instrument
• Make sure that the meter is placed on a level, stationary surface during the measurement.
Note: Do not hold the meter in the hand during measurement.
• Always close the sample compartment lid during measurement, calibration and storage.
10
English
• Remove sample cell and batteries from the instrument if the instrument is stored for an extended time period (more than a month).
• Keep the sample compartment lid closed to prevent the entry of dust and dirt.
Sample cells
• Always cap the sample cell to prevent spillage of the sample into the instrument.
• Always use clean sample cells in good condition. Dirty, scratched or damaged cells can cause inaccurate readings.
• Make sure that cold samples do not “fog” the sample cell.
• Store sample cells filled with distilled or deionized water and cap tightly.
Measurement
• Measure samples immediately to prevent temperature changes and settling. Before a measurement is taken, always make sure that the sample is homogeneous throughout.
• Avoid sample dilution when possible.
• Avoid operation in direct sunlight.

Turbidity measurement procedure

Note: Before a measurement is taken, always make sure that the sample is homogeneous throughout.
1. Collect a representative sample in a clean container. Fill a sample cell to the line (about 15 mL). Take care to handle the sample cell by the top. Cap the cell.
2. Wipe the cell with a soft, lint-free cloth to remove water spots and fingerprints.
3. Apply a thin film of silicone oil. Wipe with a soft cloth to obtain an even film over the entire surface (Apply silicone
oil to a sample cell).
4. Push the Power key to turn the meter on. Place the instrument on a flat, sturdy surface.
Note: Do not hold the instrument while making measurements.
5. Gently invert and then insert the sample cell in the instrument cell compartment so the diamond or orientation mark aligns with the raised orientation mark in front of the cell compartment. Close the lid.
6. Push Read. The display shows Stabilizing then the turbidity in NTU (FNU). The result is shown and stored automatically. Additional information is available on the manufacturer's website.
English 11

Data management

About stored data

The following types of data are stored in the data log:
• Reading Log: stores automatically each time a sample reading is taken (500 records).
• Calibration Log: stores only when Store is selected at the end of a calibration (25 records).
• Verify Cal Log: stores only after Done is selected at the end of a verification calibration (250 records).
When the data log becomes full, the oldest data point is deleted when more data is added to the log.

View data log

The data log contains Reading Log, Calibration Log and Verify Cal log. All logs can be sorted by date.
1. Push the DATA MANAGEMENT key.
2. Select View Data Log to view the stored data.
3. Push Select to view additional information.
Option Description
Reading Log Reading Log—shows the date, time and reading mode
and associated calibration data.
Calibration Log Calibration Log—shows the date and time of calibration
data and additional information about the calibration.
Verify Cal Log Verify Cal Log—shows the calibration verification date
and time and additional information about the verification.
All Logs by Date The most recent data and additional information is
shown. The icons show whether the data is from a reading, calibration or calibration verification and identifies the reading mode, if applicable.

Delete data log

There are two possibilities to delete stored readings in the Data Management menu:
1. Push the DATA MANAGEMENT key and select Delete Data Log.
Option Description
Delete Last Reading Only the last reading stored can be deleted until a
new reading is taken and stored.
Delete All Logs The entire Reading Log can be deleted at once.

Send stored data

Data can be stored and transferred to a printer, computer or USB storage device. The data will be formatted as an XML file. Install the USB/power module to the meter and to AC power. Refer to the module documentation for more information.

Advanced operation

Display contrast

1. Push the SETTINGS key and select Display Contrast.
2. Use the UP and DOWN key to adjust the contrast of the display and
push OK.
12 English

Power management

Use power management to change the backlight option and the battery saving auto-shutoff option.
Note: Power management is not active when the meter is connected to AC power.
1. Push the SETTINGS key and select Power Management.
2. Select which display option to change.
Option Description
Backlight The display is illuminated. To maximize battery life, select a
time period after which the backlight will automatically power off if no key is pushed: 10 s, 20 s, 30 s,1 min, 2 min, 5 min
Note: The Backlight keys (Product overview) will turn the backlight on and off.
Auto-Shutoff To maximize battery life, set a time period after which the
meter will automatically power off if no key is pushed: 1 min, 2 min, 5 min, 10 min, 30 min, 1 h

Set the sound options

The meter can make an audible sound when a key is pushed, when a reading is complete or when the calibration reminder is due.
1. Push SETTINGS and select Sounds.
2. Select which events will produce an audible sound. Multiple items
can be selected.
Option Description
Key Press The meter will make an audible sound whenever a key
is pushed.
Reading complete The meter will make an audible sound whenever a
reading is completed.
Reminders The meter will make an audible sound when a
calibration is due.

Security options

The Security Options menu is used to protect the meter setup.
The Setup Date and Time, Delete Data Log, Restoring Factory Defaults and Restore Factory Cal screens are not accessible without a password.
Store the password in a safe and accessible place. If the specified password is forgotten and Security Options is turned on, the operator is locked out of the restricted menus. Contact technical support if the password is lost.

Turn security options on

The security options and the set password options are used together to prevent access to restricted menus.
1. Push the SETTINGS key and select Security Options.
2. Select Edit Password and use the UP and DOWN keys to set a
password.
3. Select Security On to enable the password setting. The requirement
for the password entry is controlled by setting Security Options on or off.
Note: Set the Security to Off to disable the password setting.
4. Push the ON/OFF key to turn off and on the meter to activate the
password settings.

View meter information

The instrument information menu shows specific information such as the meter name, model number, software version, serial number and available Operator and Sample IDs. 10 Operator IDs and 100 Sample IDs are available.
1. Push the SETTINGS key and select Meter Information.

Calibration

The portable turbidimeter is calibrated with Formazin Primary Standards at the factory. The meter should be calibrated upon receipt for best results. The manufacturer recommends calibration with a primary standard such as StablCal® Stabilized Standards or with formazin standards every three months.
English
13
Note: Set Cal Reminder Repeat in the Calibration Options menu for periodical calibration. Verify the calibration once a week.

Calibration options

The calibration options contain Calibration History, Calibration Curves, Cal Reminder Repeat and Restore Factory Calibration.
1. Push the CALIBRATION key and then the UPand DOWN key.
Option Description
Calibration History
Cal.Curve Select one of the calibration curves for calibration:
Cal Reminder Repeat
Restore Factory Calibration
The calibration history shows a list of the times when the meter was calibrated. Select a date and time to view a summary of the calibration data.
StablCal® RapidCal™ (0–40 NTU)
StablCal® (0–1000 NTU)
Formazin RapidCal™ (0–40 NTU)
Formazin (0–1000 NTU)
Degrees (0–100 mg/L)
SDVB (0–1000 NTU)
Custom (0–1000 NTU)
The meter will make an audible sound when calibration is due. Select one of the following options for time interval and push OK: Off, 1 d, 7 d, 30 d, 60 d, 90 d
All user calibrations will be deleted. The original factory calibration is restored.

Calibration standard overview

Refer to Calibration standard overview for the calibration standard overview.
Table 1 Calibration overview
Type of calibration Required standards
StablCal® RapidCal (0–40 NTU)
StablCal® (0– 1000 NTU)
Formazin RapidCal (0–40 NTU)
Formazin (0– 1000 NTU)
Degrees (0–100 mg/L) Typically deionized or
SDVB (0–1000 NTU) Typically deionized or
Custom (0–1000 NTU) Typically deionized or
1
The water must have a turbidity <0.5 NTU to prepare the calibration standards.
20 NTU –
20 NTU 100 NTU 800 NTU
Typically deionized or distilled waterFootnote
Typically deionized or distilled waterFootnote
distilled waterFootnote
distilled waterFootnote
distilled waterFootnote
20 NTU –
20 NTU 100 NTU 800 NTU
20 NTU 100 NTU –
20 NTU 100 NTU 800 NTU
Select values
14 English

StablCal® RapidCal™ calibration

1. Push the CALIBRATION key to
enter the Calibration mode. Follow the instructions on the display.
Note: Gently invert each standard before inserting the standard.
7. Push Store to save the results. After a calibration is complete, the meter automatically goes into the Verify Cal mode, refer to Verification
calibration.
2. Push the UP and DOWN key to access
Cal Options and then select Cal.Curve.
3. Select StablCal RapidCal™ from the list and push OK.
®
4. Insert the 20 NTU StablCal Standard and close the lid.
Note: The standard to be inserted is bordered.
5. Push Read. The display shows Stabilizing and then shows the result.
6. Push Done to review the calibration details.
English 15

Verification options

The Verification Options contain: Set Verification Standard, Set Acceptance Criteria and Verification Reminder.
1. Push the Left key (Verify Cal) and then the UP and DOWN keys.
Option Description
Set Verification Standard
Set Acceptance Criteria
Verification Reminder
To change the verification standard use the UP and DOWN keys to enter a new standard value.
Range 0.50–20.0 NTU (Default setting: 10.00 NTU)
Range 0–20 NTU for RapidCal™ (0–40 NTU)
0–800 NTU for calibration curves with a range from 0– 1000 NTU
Enter the Acceptance Criteria for comparison against the initial calibration verification reading to determine passing or failing. Range 1–50% (Default setting: 10%)
Verification Reminder—The meter will make an audible sound when verification is due. Select one of the following options for time interval and push OK:Off, 30 min (Default setting), 2 h, 4 h, 8 h, 24 h Allow Defer—Push Allow Defer and select Yes or No to postpone the verification due time

Calibration verification (Verify Cal)

The manufacturer recommends a calibration verification once a week. After a calibration is complete, the meter automatically goes into the Verify Cal mode.
Make sure that the sample cell is clean. Oil the sample cell with silicone oil, refer to Apply silicone oil to a sample cell. Check the standard
solution. Prepare a formazin standard at the same value and read the value.
1. Push Verify Cal to enter the Verify menu.
4. Push Done to return to the reading display. Repeat the calibration verification if the verification failed.
2. Gently invert the standard. Insert the
10.0 NTU (or other defined value) Verification Standard and close the lid.
3. Push Read. The display shows Stabilizing and then shows the result and tolerance range.
16 English

Reading modes

1. Push the UP or DOWN key to enter the Reading Options menu.
2. Select Reading Mode to select one of the following options:
Option Description
Normal (Default setting)
Signal Average
Rapidly Settling Turbidity (RST)
The normal mode reads and averages three readings. The result is shown after the reading.
The Signal Average mode compensates for reading fluctuations caused by drifting of sample particles through the light path.
The X-bar icon is shown on the display when signal averaging is on.
The Signal Average mode measures 12 times and starts to show the average after three readings. The final result is the average of all 12 readings.
The Rapidly Settling Turbidity (RST) mode calculates and continuously updates the turbidity reading of the sample to
a confidence of 95%, based on the accumulated trend of the real time measured values.
The RST mode is best used on samples that settle rapidly and continuously change in value. The reading is based on a correctly prepared sample that is homogeneous at the beginning of the reading. It is best applied to samples that are greater than 20 NTU. The sample must be mixed thoroughly by inversion immediately before inserting it into the meter.
The target icon is shown on the display when the Rapidly Settling Turbidity is on.
The Rapidly Settling Turbidity reads and calculates five readings while showing intermediate results.

Apply silicone oil to a sample cell

Sample cells and caps must be extremely clean and free from significant scratches. Apply a thin coating of silicone oil on the outside of the sample cells to mask minor imperfections and scratches that may contribute to light scattering.
Note: Use only the provided silicone oil. This silicone oil has the same refractive index as the sample cell glass.
1. Clean the inside and outside of the cells and caps by washing with a laboratory glass cleaning detergent. Follow with multiple rinses with distilled or demineralized water.
2. Apply a small bead of silicone oil from the top to the bottom of the cell.
3. Use the provided oiling cloth to spread the oil uniformly. Wipe off the excess so that only a thin coat of oil is left. Make sure that the sample cell is almost dry with little or no visible oil.
Note: Store the oiling cloth in a plastic storage bag to keep the cloth clean.

Indexing a single cell

Precise measurements for very low turbidity samples require the use of a single cell for all measurements or optically matching the cells. Use one cell to provide the best precision and repeatability. When one cell is used, an index or orientation mark (other than the factory-placed diamond) can be placed on the cell so it is inserted into the instrument with the same orientation each time. When using a single cell, make an index or orientation mark on the cell as follows:
English
17
1. Fill the clean sample
cell to the line with high quality water (<
0.5 NTU) and cap
immediately. Let the sample cell degas for at least five minutes.
2. Wipe with lint-free cloth. Apply a thin film of silicone oil (Apply
silicone oil to a sample cell).
3. Push the POWER key to turn the meter on. Place the instrument on a flat, sturdy surface.
Note: Do not hold the instrument while making measurements.
4. Insert the sample cell in the instrument cell compartment so the diamond or orientation mark always aligns with the raised orientation mark in front of the cell compartment. Close the lid.
5. Push the UP and DOWN key to access
the Reading Options and then select Indexing Sample Cell.
Note: The instruments always stays in the last selected reading mode.
6. Push Read. The display shows Stabilizing then the turbidity in NTU. Record the cell position in the cell compartment and the reading result.
7. Remove the cell,
rotate it slightly approximately of a turn and insert it again into the cell compartment. Close the lid.
18 English
8. Push Read. Record
the cell position in the cell compartment and the reading result.
9. Repeat step 6 until the lowest reading is shown. Place an orientation mark on the cell marking band near the top of the cell so the cell can be consistently inserted in the position that yields the lowest reading.

Maintenance

Replace the battery

C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.

Clean the meter

The meter is designed to be maintenance-free and does not require regular cleaning for normal operation. Exterior surfaces of the meter may be cleaned as necessary.
Note: Do not clean the meter with solvents to avoid damaging the material.
1. Clean the meter with a dust- and lint-free dry or slightly damp cloth. A
mild soap solution can also be used for liposoluble contamination.

Store the sample cells

N O T I C E
Do not air dry the sample cells.
Note: Always store the sample cells with caps on to prevent the cells from drying.
1. Fill the sample cells with distilled or demineralized water.
2. Cap and store the sample cells.
3. Wipe the outside of the sample cells dry with the a soft cloth.
W A R N I N G
Explosion hazard. An expired battery can cause hydrogen gas buildup inside the instrument. Replace the battery before it expires. Do not store the instrument for long periods with a battery installed.
Potential fire hazard. Use only alkaline or nickel metal hydride batteries (NiMH) in the meter. Other battery types or incorrect installation can cause a fire. Never mix battery types in the meter.
For battery replacement refer to Install the battery on page 6.
1. Remove the battery cover.
2. Remove the batteries.
3. Install 4 AA alkaline or 4 AA nickel metal hydride (NiMH) batteries.
Make sure that the batteries are installed in the correct orientation.
4. Replace the battery cover.
W A R N I N G

Replace the lamp

C A U T I O N
Burn Hazard. Wait until lamp cools down. Contact with the hot lamp can cause burns.
English 19
1 2
3 4
20 English
5 6
7 8
English 21

Troubleshooting

Refer to the following table for common problem messages or symptoms, possible causes and corrective actions.
Error/Warning Description Solution
Close lid and push Read.
Low Battery! Battery is low.
ADC Failure! Hardware error causing
Detector signal too low! Insufficient light on the
Overrange! Turbidity too high-
Underrange! The measured
Please check the lamp! Signals are too low on
The lid is open or lid detection failed.
reading to fail.
180° detector.
caused probably by calibrating with RapidCal™ only.
absorbance is below the calibration range.
the 90° and 180° detector.
Make sure that the lid is closed during reading and re-read.
• Insert new batteries
• Connect USB/power module if rechargeable batteries are used
Repeat the reading.
• Check for obstructed light path.
• Check the lamp.
• Calibrate the upper range.
• Dilute the sample.
Repeat calibration
2100Q: The lamp is defective. Change the lamp (refer to Replace
the lamp on page 19).
2100Qis: Contact technical support.
Error/Warning Description Solution
Temperature too high! Switch off instrument.
RST: Average value! Solids are settling too
Confidence level is < 95%
Standard value out of range. Insert standard and push Read
ID already in use. Enter new ID
Error - Security Please set password before activating security
Please enter at least one character.
Password incorrect. Please retry.
Temperature has exceeded the meter limits (>60 °C or >140 °F).
slowly. The reading mode is not suitable for this sample.
The reading mode Rapidly Settling Turbidity did not meet the range of ≥ 95% confidence.
Used incorrect standard value for the reading.
The Operator or Sample ID is unavailable as it is already assigned.
No password is created. Create a new password.
Password must contain minimum of one character.
Incorrect password was entered.
Turn off the meter and let it cool down.
Select Normal or Signal Average reading mode.
• Invert the sample several times so that the solids allocate. Repeat the reading again.
• Switch to the Normal reading mode if the sample is stable and does not have settable solids.
Insert the appropriate standard and read again.
Create a new ID.
Create a password of at least one character.
Enter the appropriate password.
22 English
Error/Warning Description Solution
Please disconnect the USB cable from your computer.
USB module memory full. Delete data and try again.
Delete Last Reading Failed!
Delete Data Log failed!
Can't read data set!
Can't store data!
Can't store to the Reading Log!
Can't store to the Verify Cal Log!
Error storing data!
Error reading data!
Data storage does not respond while connected to the meter and the computer.
Data storage is full. 1. Connect USB/power
Error in the data storage. Turn the meter off and
Disconnect the USB cable from the meter and try sending data again.
module to the computer.
2. Download the stored data to the computer.
3. Delete Data Log on the module.
on. If the error message still occurs, contact technical support.
Replacement parts (continued)
Description Quantity Item no.
Insert, molded bottom 1 2971507
Sample cell oiling cloth 1 4707600
1" glass sample cell (10 ml ) w/cap (Turb) pkg/6 2434706
Carrying case (includes insert) 1 2971500
Battery set, AA alkaline batteries pkg/4 1938004
Lamp assy 1 4653900
Blank module 1 LZV797
Rubber foot set 1 LZV821
Lamp cover (includes screws) 1 LZV822
Battery cover (includes 2 feet) 1 LZV823
Module cover 1 LZV824
Connector cover for USB/power module 1 LZV825
Connector cover for power module 1 LZV826
Lid (includes magnet) 1 LZV827

Accessories

Replacement parts and accessories

Replacement parts

Description Quantity Item no.
StablCal ampule calibration kit 1 2971205
10 NTU verification standard 100 mL 2961701
Silicone Oil 15 mL 126936
Description Quantity Item no.
USB/power module (includes: universal power supply, USB cable, instruction sheet)
Power module (includes: universal power supply, instruction sheet)
USB module with USB cable (2x) 1 LZV949.99.00001
StablCal 0.1 NTU Standard 100 mL 2723342
1 LZV813.99.00001
1 LZV804.99.00001
English 23
1
1
1
Accessories (continued)
Description Quantity Item no.
StablCal 0.3 NTU Standard 100 mL 2697943
StablCal 0.5 NTU Standard 100 mL 2698042
StablCal calibration kit 100 mL 2971210
StablCal calibration kit 500 mL 2971200
Gelex secondary standards set 1 2464105
Deionized water 4 vials 27217
Filter 0.2 micron 2323810
Formazin 500 ml 246149
Formazin 1000 ml 246142
Sample degassing kit 1 4397500
Sample degassing and filtration kit 1 4397510
Battery, NiMH AA pk/4 2971304
1
Not available in all regions
24 English
A
auto-shutoff ............................................................................................13
B
backlight .................................................................................................13
battery installation ..............................................................................6, 19

Index

M
menu navigation .......................................................................................8
meter components ...................................................................................5
meter specifications .................................................................................3
meters
troubleshooting .................................................................................. 22
C
calibration ...............................................................................................13
contrast, display .....................................................................................12
D
data, calibration log ................................................................................12
data, data log .........................................................................................12
data, reading log ....................................................................................12
data, stored data ....................................................................................12
data, verification log ...............................................................................12
date and time ...........................................................................................9
display ......................................................................................................8
I
Indexing a sample cell ...........................................................................17
K
keypad .....................................................................................................7
L
language ..................................................................................................9
N
normal reading .......................................................................................17
O
operator ID ...............................................................................................9
orientation mark .....................................................................................17
P
parts list ............................................................................................23, 24
power
battery installation .........................................................................6, 19
R
rapidly settling turbidity (RST) ................................................................17
reading mode .........................................................................................17
S
sample ID .................................................................................................9
security options ......................................................................................13
signal average reading ...........................................................................17
25
Index
silicone oil ..............................................................................................17
sound options .........................................................................................13
stored data, send ...................................................................................12
V
verification ..............................................................................................16
26
HACH COMPANY World Headquarters
P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. (970) 669-3050 (800) 227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com
©
Hach Company, 2009, 2013. All rights reserved. Printed in the U.S.A.
HACH LANGE GMBH
Willstätterstraße 11 D-40549 Düsseldorf, Germany Tel. +49 (0) 2 11 52 88-320 Fax +49 (0) 2 11 52 88-210 info@hach-lange.de www.hach-lange.de
HACH LANGE Sàrl
6, route de Compois 1222 Vésenaz SWITZERLAND Tel. +41 22 594 6400 Fax +41 22 594 6499
Loading...