General information.....................................................................................................................................................................3
Use of hazard information..................................................................................................................................................................4
Install the battery................................................................................................................................................................................6
User interface and navigation...............................................................................................................................................7
User interface.....................................................................................................................................................................................7
Turn the meter on and off..................................................................................................................................................................8
Change the language ........................................................................................................................................................................9
Change the date and time .................................................................................................................................................................9
Standard operation......................................................................................................................................................................9
Use a sample ID................................................................................................................................................................................9
Use an operator ID.............................................................................................................................................................................9
Calibrate the turbidimeter with StablCal® Standards.......................................................................................................................10
Data management.......................................................................................................................................................................12
About stored data.............................................................................................................................................................................12
View data log...................................................................................................................................................................................12
Delete data log.................................................................................................................................................................................12
Power management.........................................................................................................................................................................13
Set the sound options......................................................................................................................................................................13
View meter information....................................................................................................................................................................13
Calibration standard overview..................................................................................................................................................14
Apply silicone oil to a sample cell....................................................................................................................................................17
Indexing a single cell........................................................................................................................................................................17
Clean the meter...............................................................................................................................................................................19
Store the sample cells......................................................................................................................................................................19
Replace the battery..........................................................................................................................................................................19
Replace the lamp.............................................................................................................................................................................19
Replacement parts and accessories...............................................................................................................................23
Replacement parts ..........................................................................................................................................................................23
Specifications are subject to change without notice.
SpecificationDetails
Measurement
method
Regulatory2100Q: Meets EPA Method 180.1
Lamp source2100Q: Tungsten filament lamp
Range0–1000 NTU (FNU)
Accuracy±2% of reading plus stray light from 0–1000 NTU (FNU)
Repeatability±1% of reading or 0.01 NTU (FNU), whichever is greater
Resolution0.01 NTU on lowest range
Stray light≤ 0.02 NTU (FNU)
Signal averagingSelectable on or off
DetectorSilicon Photodiode
Reading modesNormal (Push to Read), Signal Averaging or Rapidly Settling
Calibration
options
Calibration logger Records the last 25 successful calibrations
Verification logger Logs the last 250 successful verifications
Data logger500 records
Ratio turbidimetric determination using a primary
nephelometric light scatter signal (90°) to the transmitted
light scatter signal.
2100Qis: Meets ISO 7027
2100Qis: Light-emitting diode (LED) at 860 nm
™
Turbidity
Single step RapidCal™ for Low-Level Regulatory Reporting
from 0–40 NTU (FNU)
Full range calibration from 0–1000 NTU (FNU)
Calibration to degrees of turbidity
SpecificationDetails
Power
requirement
Operating
conditions
Storage
conditions
InterfaceOptional USB
Sample required15 mL (0.5 oz.)
Sample cellsRound cells 60 x 25 mm (2.36 x 1 in.) borosilicate glass with
Dimensions22.9 x 10.7 x 7.7 cm (9.0 x 4.2 x 3.0 in.)
Weight530 g (1.17 lb) without batteries
Meter enclosure
rating
Protection classPower supply: Class II
CertificationCE certified
Warranty1 year (EU: 2 years)
AC 100–240 V , 50/60 Hz (with power or USB/power
module)
4 AA alkaline batteries
Rechargeable NiMH (for use with USB/power module)
Temperature: 0 to 50 °C (32 to 122 °F)
Relative Humidity: 0–90% at 30 °C, 0–80% at 40 °C, 0–70%
at 50 °C, noncondensing
–40 to 60 °C (–40 to 140 °F), instrument only
screw caps
620 g (1.37 lb) with four AA alkaline batteries
IP67 (closed lid, battery and module compartment excluded)
General information
In no event will the manufacturer be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or
omission in this manual. The manufacturer reserves the right to make
changes in this manual and the products it describes at any time, without
notice or obligation. Revised editions are found on the manufacturer’s
website.
English
3
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or
misuse of this product including, without limitation, direct, incidental and
consequential damages, and disclaims such damages to the full extent permitted
under applicable law. The user is solely responsible to identify critical application
risks and install appropriate mechanisms to protect processes during a possible
equipment malfunction.
Please read this entire manual before unpacking, setting up or operating
this equipment. Pay attention to all danger and caution statements.
Failure to do so could result in serious injury to the operator or damage
to the equipment.
Make sure that the protection provided by this equipment is not impaired.
Do not use or install this equipment in any manner other than that
specified in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will
result in death or serious injury.
Indicates a potentially or imminently hazardous situation which, if not avoided,
could result in death or serious injury.
Indicates a potentially hazardous situation that may result in minor or moderate
injury.
Indicates a situation which, if not avoided, may cause damage to the instrument.
Information that requires special emphasis.
W A R N I N G
C A U T I O N
N O T I C E
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed. A symbol on the
instrument is referenced in the manual with a precautionary statement.
This is the safety alert symbol. Obey all safety messages that follow
this symbol to avoid potential injury. If on the instrument, refer to the
instruction manual for operation or safety information.
This symbol indicates that a risk of electrical shock and/or
electrocution exists.
Electrical equipment marked with this symbol may not be disposed of
in European public disposal systems after 12 August of 2005. In
conformity with European local and national regulations (EU Directive
2002/96/EC), European electrical equipment users must now return
old or end-of-life equipment to the Producer for disposal at no charge
to the user.
Note: For return for recycling, please contact the equipment producer or supplier
for instructions on how to return end-of-life equipment, producer-supplied
electrical accessories, and all auxiliary items for proper disposal.
Certification
Canadian Radio Interference-Causing Equipment Regulation,
IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian
Interference-Causing Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la
réglementation canadienne sur les équipements provoquant des
interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device
complies with Part 15 of the FCC Rules. Operation is subject to the
following conditions:
1. The equipment may not cause harmful interference.
4
English
2. The equipment must accept any interference received, including
interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by
the party responsible for compliance could void the user's authority to
operate the equipment. This equipment has been tested and found to
comply with the limits for a Class A digital device, pursuant to Part 15 of
the FCC rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference
to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference, in which case the user will be
required to correct the interference at their expense. The following
techniques can be used to reduce interference problems:
1. Disconnect the equipment from its power source to verify that it is or
is not the source of the interference.
2. If the equipment is connected to the same outlet as the device
experiencing interference, connect the equipment to a different
outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the
interference.
5. Try combinations of the above.
Product overview
The 2100Q and 2100Qis portable turbidimeters measure turbidity from
0 to 1000 NTU (FNU). Primarily for field use, the portable meter operates
on four AA batteries. Data can be stored and transferred to a printer,
computer or USB storage device.
Figure 1 Product overview
1 Power on or off5 Alignment arrow
2 Backlight keys (+ and -)6 Module
3 Sample cell holder with lid7 Lamp compartment
4 Attachment for lanyard8 Battery compartment
Product components
Refer to Figure 2 to make sure that all components have been received.
If any of these items are missing or damaged, contact the manufacturer
or a sales representative immediately.
English 5
Figure 2 2100Q and 2100Qis components
1 2100Q or 2100Qis turbidimeter6 Silicone oil
2 Carrying case7 20, 100 and 800 NTU StablCal
3 User manual and Quick reference
guide
4 Oiling cloth9 StablCal 10 NTU verification
5 1" sample cell (10 mL) with cap
(pk/6)
calibration standards
8 AA alkaline batteries (pk/4)
standard
Installation
Install the battery
W A R N I N G
Explosion hazard. An expired battery can cause hydrogen gas buildup
inside the instrument. Replace the battery before it expires. Do not
store the instrument for long periods with a battery installed.
Potential fire hazard. Use only alkaline or nickel metal hydride batteries (NiMH) in
the meter. Other battery types or incorrect installation can cause a fire. Never mix
battery types in the meter.
The battery compartment is not waterproof. If the battery compartment becomes
wet, remove and dry the batteries and dry the interior of the compartment. Check
the battery contacts for corrosion and clean them if necessary.
When using nickel metal hydride (NiMH) batteries, the battery icon will not
indicate a full charge after freshly charged batteries have been inserted (NiMH
batteries are 1.2 V versus 1.5 V for alkaline batteries). Even though the icon does
not indicate complete charge, 2300 mAH NiMH batteries will achieve 90% of
instrument operation lifetime (before recharge) versus new alkaline batteries.
To avoid potential damage to the meter from battery leakage, remove the meter
batteries prior to extended periods of non-use.
The meter can be powered with AA alkaline or rechargeable NiMH
batteries. To conserve battery life, the meter will power off after
10 minutes of inactivity, the backlight powers off after 30 seconds. This
time can be changed in the Power Management menu.
Note: Rechargeable batteries will only be recharged with the USB/power module.
Refer to the module documentation for further information.
For battery installation refer to Figure 3.
W A R N I N G
N O T I C E
N O T I C E
N O T I C E
6
English
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.
1. Remove the battery cover.
2. Install 4 AA alkaline or 4 AA nickel metal hydride (NiMH) batteries.
Make sure that the batteries are installed in the correct orientation.
3. Replace the battery cover.
Figure 3 Battery installation
User interface and navigation
User interface
Figure 4 Keypad description
1 SETTINGS key: select menu
options for setting up the meter
2 CALIBRATION key: shows
calibration screen, start calibration,
select cal options
3 DOWN key: scroll through menus,
enter numbers and letters
4 LEFT key (contextual): access for
calibration verification, cancels or
exits the current menu screen to the
previous menu screen
5 UP key: scroll through menus, enter
numbers and letters
6 RIGHT key (contextual): read
turbidity sample, selects or confirms
options, opens/jumps to sub-menus
7 DATA MANAGEMENT key: view,
delete or transfer stored data
English 7
Display description
The measurement screen shows the turbidity, unit, calibration status,
date and time, operator ID (if setup) and sample ID (if setup). Refer to
3 Stability or display lock indicator11 Reading mode: Signal Average
4 Calibration status indicator
(Calibration OK=pass)
5 Calibration status indicator
(Calibration ?=fail)
6 Parameter title14 Read (contextual: OK, Select)
7 AC power icon15 Options (contextual)
8 Battery icon16 Verification calibration
or FNU (Formazin Turbidity Unit)
Turbidity (Target icon)
(X-bar icon)
12 Time
13 Date
Navigation
The meter contains a Settings menu, Reading Options menu, Calibration
Options menu and Calibration Verification Options menu to change
various options. Use the UP and DOWN keys to highlight different
options. Push the RIGHT key to select an option. There are two ways to
change options:
1. Select an option from a list: Use the UP and DOWN keys to select an
option. If check boxes are shown, more than one option can be
selected. Push the LEFT key under Select.
Note: To deselect check boxes, push the LEFT key under Deselect.
2. Enter an option value using the arrow keys:
Push the UP and DOWN keys to enter or change a value.
3. Push the RIGHT key to advance to the next space.
4. Push the RIGHT key under OK to accept the value.
Startup
Turn the meter on and off
Push the ON/OFF key to turn on or turn off the meter. If the meter
does not turn on, make sure that the batteries, or the module, are
properly installed or that the AC power supply is properly connected to
an electrical outlet.
Note: The Auto-Shutoff option can also be used to turn off the meter. Refer to
Power management on page 13.
8 English
Change the language
There are three options to set the language:
• The display language is selected when the meter is powered on for
the first time.
• The display language is selected when the power key is pushed and
held.
• The language can be changed from the Settings menu.
1. Select a language from the list. Confirm with OK.
2. Push Done when the update is complete.
Change the date and time
The date and time can be changed from the Date & Time menu.
Standard operation
Use a sample ID
The sample ID tag is used to associate readings with a particular sample
location. If assigned, stored data will include this ID.
1. Select Sample ID in the Settings menu.
2. Select, create or delete a sample ID:
OptionDescription
Current IDSelect an ID from a list. The current ID will be
associated with sample data until a different ID is
selected.
Create a New SampleIDEnter a name for a new sample ID.
1. Push the SETTINGS key and select Date & Time.
2. Update the time and date information:
OptionDescription
FormatSelect one of the formats for the date and time:
yyyy-mm-dd 24h
yyyy-mm-dd 12h
dd-mm-yyyy 24h
dd-mm-yyyy 12h
mm/dd/yyyy 24h
mm/dd/yyyy 12h
DateEnter the current date
TimeEnter the current time
The current date and time will be shown on the display.
After the date and time setup, the meter is ready to take a reading.
Delete Sample IDDelete an existing sample ID.
Use an operator ID
The operator ID tag associates readings with an individual operator. All
stored data will include this ID.
1. Select Operator ID in the Settings menu.
2. Select, create or delete an operator ID:
OptionDescription
Current IDSelect an ID from a list. The current ID will be
Create a New
Operator ID
Delete Operator IDDelete an existing operator ID.
associated with sample data until a different ID is
selected.
Enter a name for a new operator ID (maximum
10 names can be entered).
English 9
Calibrate the turbidimeter with StablCal® Standards
Note: For best accuracy use the same sample cell or four matched sample cells for
all readings during calibration. Insert the sample cell in the instrument cell
compartment so the diamond or orientation mark aligns with the raised orientation
mark in front of the cell compartment.
1. Push the
CALIBRATION key to
enter the Calibration
mode. Follow the
instructions on the
display.
Note: Gently invert
each standard before
inserting the standard.
2. Insert the 20 NTU
StablCal Standard and
close the lid.
Note: The standard to
be inserted is bordered.
3. Push Read. The
display shows
Stabilizing and then
shows the result.
4. Repeat Step 2 and
3 with the 100 NTU and
800 NTU StablCal
Standard.
Note: Push Done to
complete a 2 point
calibration.
5. Push Done to
review the calibration
details.
6. Push Store to save
the results.
After a calibration is
complete, the meter
automatically goes into
the Verify Cal mode.
Refer to Verification
calibration.
Turbidity measurement
W A R N I N G
Potential explosion and fire hazard. This turbidimeter is designed for water based
samples. Do not measure solvent or combustible based samples.
Readings can be taken with the Normal reading mode, Signal Average
mode or in the Rapidly Settling Turbidity mode. Refer to Reading modes
on page 17 for more information. For accurate turbidity readings use
clean sample cells and remove air bubbles (degassing).
Measurement notes
Proper measurement techniques are important in minimizing the effects
of instrument variation, stray light and air bubbles. Use the following
measurement notes for proper measurements.
Instrument
• Make sure that the meter is placed on a level, stationary surface
during the measurement.
Note: Do not hold the meter in the hand during measurement.
• Always close the sample compartment lid during measurement,
calibration and storage.
10
English
• Remove sample cell and batteries from the instrument if the
instrument is stored for an extended time period (more than a month).
• Keep the sample compartment lid closed to prevent the entry of dust
and dirt.
Sample cells
• Always cap the sample cell to prevent spillage of the sample into the
instrument.
• Always use clean sample cells in good condition. Dirty, scratched or
damaged cells can cause inaccurate readings.
• Make sure that cold samples do not “fog” the sample cell.
• Store sample cells filled with distilled or deionized water and cap
tightly.
Measurement
• Measure samples immediately to prevent temperature changes and
settling. Before a measurement is taken, always make sure that the
sample is homogeneous throughout.
• Avoid sample dilution when possible.
• Avoid operation in direct sunlight.
Turbidity measurement procedure
Note: Before a measurement is taken, always make sure that the sample is
homogeneous throughout.
1. Collect a
representative sample
in a clean container. Fill
a sample cell to the line
(about 15 mL). Take
care to handle the
sample cell by the top.
Cap the cell.
2. Wipe the cell with a
soft, lint-free cloth to
remove water spots and
fingerprints.
3. Apply a thin film of
silicone oil. Wipe with a
soft cloth to obtain an
even film over the entire
surface (Apply silicone
oil to a sample cell).
4. Push the Power key
to turn the meter on.
Place the instrument on
a flat, sturdy surface.
Note: Do not hold the
instrument while
making measurements.
5. Gently invert and
then insert the sample
cell in the instrument
cell compartment so the
diamond or orientation
mark aligns with the
raised orientation mark
in front of the cell
compartment. Close the
lid.
6. Push Read. The
display shows
Stabilizing then the
turbidity in NTU (FNU).
The result is shown and
stored automatically.
Additional information is
available on the
manufacturer's website.
English 11
Data management
About stored data
The following types of data are stored in the data log:
• Reading Log: stores automatically each time a sample reading is
taken (500 records).
• Calibration Log: stores only when Store is selected at the end of a
calibration (25 records).
• Verify Cal Log: stores only after Done is selected at the end of a
verification calibration (250 records).
When the data log becomes full, the oldest data point is deleted when
more data is added to the log.
View data log
The data log contains Reading Log, Calibration Log and Verify Cal log.
All logs can be sorted by date.
1. Push the DATA MANAGEMENT key.
2. Select View Data Log to view the stored data.
3. Push Select to view additional information.
OptionDescription
Reading LogReading Log—shows the date, time and reading mode
and associated calibration data.
Calibration Log Calibration Log—shows the date and time of calibration
data and additional information about the calibration.
Verify Cal LogVerify Cal Log—shows the calibration verification date
and time and additional information about the verification.
All Logs by Date The most recent data and additional information is
shown. The icons show whether the data is from a
reading, calibration or calibration verification and
identifies the reading mode, if applicable.
Delete data log
There are two possibilities to delete stored readings in the Data
Management menu:
1. Push the DATA MANAGEMENT key and select Delete Data Log.
OptionDescription
Delete Last Reading Only the last reading stored can be deleted until a
new reading is taken and stored.
Delete All LogsThe entire Reading Log can be deleted at once.
Send stored data
Data can be stored and transferred to a printer, computer or USB
storage device. The data will be formatted as an XML file. Install the
USB/power module to the meter and to AC power. Refer to the module
documentation for more information.
Advanced operation
Display contrast
1. Push the SETTINGS key and select Display Contrast.
2. Use the UP and DOWN key to adjust the contrast of the display and
push OK.
12 English
Power management
Use power management to change the backlight option and the battery
saving auto-shutoff option.
Note: Power management is not active when the meter is connected to AC power.
1. Push the SETTINGS key and select Power Management.
2. Select which display option to change.
OptionDescription
BacklightThe display is illuminated. To maximize battery life, select a
time period after which the backlight will automatically power
off if no key is pushed: 10 s, 20 s, 30 s,1 min, 2 min, 5 min
Note: The Backlight keys (Product overview) will turn the
backlight on and off.
Auto-Shutoff To maximize battery life, set a time period after which the
meter will automatically power off if no key is pushed: 1 min,
2 min, 5 min, 10 min, 30 min, 1 h
Set the sound options
The meter can make an audible sound when a key is pushed, when a
reading is complete or when the calibration reminder is due.
1. Push SETTINGS and select Sounds.
2. Select which events will produce an audible sound. Multiple items
can be selected.
OptionDescription
Key PressThe meter will make an audible sound whenever a key
is pushed.
Reading complete The meter will make an audible sound whenever a
reading is completed.
RemindersThe meter will make an audible sound when a
calibration is due.
Security options
The Security Options menu is used to protect the meter setup.
The Setup Date and Time, Delete Data Log, Restoring Factory Defaults
and Restore Factory Cal screens are not accessible without a password.
Store the password in a safe and accessible place. If the specified
password is forgotten and Security Options is turned on, the operator is
locked out of the restricted menus. Contact technical support if the
password is lost.
Turn security options on
The security options and the set password options are used together to
prevent access to restricted menus.
1. Push the SETTINGS key and select Security Options.
2. Select Edit Password and use the UP and DOWN keys to set a
password.
3. Select Security On to enable the password setting. The requirement
for the password entry is controlled by setting Security Options on or
off.
Note: Set the Security to Off to disable the password setting.
4. Push the ON/OFF key to turn off and on the meter to activate the
password settings.
View meter information
The instrument information menu shows specific information such as the
meter name, model number, software version, serial number and
available Operator and Sample IDs. 10 Operator IDs and 100 Sample
IDs are available.
1. Push the SETTINGS key and select Meter Information.
Calibration
The portable turbidimeter is calibrated with Formazin Primary Standards
at the factory. The meter should be calibrated upon receipt for best
results. The manufacturer recommends calibration with a primary
standard such as StablCal® Stabilized Standards or with formazin
standards every three months.
English
13
Note: Set Cal Reminder Repeat in the Calibration Options menu for periodical
calibration. Verify the calibration once a week.
Calibration options
The calibration options contain Calibration History, Calibration Curves,
Cal Reminder Repeat and Restore Factory Calibration.
1. Push the CALIBRATION key and then the UPand DOWN key.
OptionDescription
Calibration
History
Cal.CurveSelect one of the calibration curves for calibration:
Cal Reminder
Repeat
Restore Factory
Calibration
The calibration history shows a list of the times when the
meter was calibrated. Select a date and time to view a
summary of the calibration data.
StablCal® RapidCal™ (0–40 NTU)
StablCal® (0–1000 NTU)
Formazin RapidCal™ (0–40 NTU)
Formazin (0–1000 NTU)
Degrees (0–100 mg/L)
SDVB (0–1000 NTU)
Custom (0–1000 NTU)
The meter will make an audible sound when calibration is
due. Select one of the following options for time interval
and push OK: Off, 1 d, 7 d, 30 d, 60 d, 90 d
All user calibrations will be deleted. The original factory
calibration is restored.
Calibration standard overview
Refer to Calibration standard overview for the calibration standard
overview.
Table 1 Calibration overview
Type of calibrationRequired standards
StablCal® RapidCal
(0–40 NTU)
StablCal® (0–
1000 NTU)
Formazin RapidCal
(0–40 NTU)
Formazin (0–
1000 NTU)
Degrees (0–100 mg/L) Typically deionized or
SDVB (0–1000 NTU)Typically deionized or
Custom (0–1000 NTU) Typically deionized or
1
The water must have a turbidity <0.5 NTU to prepare the calibration
standards.
™
–20 NTU ––
–20 NTU 100 NTU 800 NTU
™
Typically deionized or
distilled waterFootnote
Typically deionized or
distilled waterFootnote
distilled waterFootnote
distilled waterFootnote
distilled waterFootnote
20 NTU ––
20 NTU 100 NTU 800 NTU
20 NTU 100 NTU –
20 NTU 100 NTU 800 NTU
Select values
14 English
StablCal® RapidCal™ calibration
1. Push the
CALIBRATION key to
enter the Calibration
mode. Follow the
instructions on the
display.
Note: Gently invert
each standard before
inserting the standard.
7. Push Store to save
the results.
After a calibration is
complete, the meter
automatically goes into
the Verify Cal mode,
refer to Verification
calibration.
2. Push the UP and
DOWN key to access
Cal Options and then
select Cal.Curve.
3. Select StablCal
RapidCal™ from the list
and push OK.
®
4. Insert the 20 NTU
StablCal Standard and
close the lid.
Note: The standard to
be inserted is bordered.
5. Push Read. The
display shows
Stabilizing and then
shows the result.
6. Push Done to
review the calibration
details.
English 15
Verification options
The Verification Options contain: Set Verification Standard, Set
Acceptance Criteria and Verification Reminder.
1. Push the Left key (Verify Cal) and then the UP and DOWN keys.
OptionDescription
Set Verification
Standard
Set Acceptance
Criteria
Verification
Reminder
To change the verification standard use the UP and
DOWN keys to enter a new standard value.
Range 0.50–20.0 NTU (Default setting: 10.00 NTU)
Range 0–20 NTU for RapidCal™ (0–40 NTU)
0–800 NTU for calibration curves with a range from 0–
1000 NTU
Enter the Acceptance Criteria for comparison against the
initial calibration verification reading to determine passing
or failing. Range 1–50% (Default setting: 10%)
Verification Reminder—The meter will make an audible
sound when verification is due. Select one of the following
options for time interval and push OK:Off, 30 min (Default
setting), 2 h, 4 h, 8 h, 24 h
Allow Defer—Push Allow Defer and select Yes or No to
postpone the verification due time
Calibration verification (Verify Cal)
The manufacturer recommends a calibration verification once a week.
After a calibration is complete, the meter automatically goes into the
Verify Cal mode.
Make sure that the sample cell is clean. Oil the sample cell with silicone
oil, refer to Apply silicone oil to a sample cell. Check the standard
solution. Prepare a formazin standard at the same value and read the
value.
1. Push Verify Cal to
enter the Verify menu.
4. Push Done to return
to the reading display.
Repeat the calibration
verification if the
verification failed.
2. Gently invert the
standard. Insert the
10.0 NTU (or other
defined value)
Verification Standard
and close the lid.
3. Push Read. The
display shows
Stabilizing and then
shows the result and
tolerance range.
16 English
Reading modes
1. Push the UP or DOWN key to enter the Reading Options menu.
2. Select Reading Mode to select one of the following options:
OptionDescription
Normal
(Default
setting)
Signal
Average
Rapidly
Settling
Turbidity
(RST)
The normal mode reads and averages three readings. The
result is shown after the reading.
The Signal Average mode compensates for reading
fluctuations caused by drifting of sample particles through
the light path.
The X-bar icon is shown on the display when signal
averaging is on.
The Signal Average mode measures 12 times and starts to
show the average after three readings. The final result is
the average of all 12 readings.
The Rapidly Settling Turbidity (RST) mode calculates and
continuously updates the turbidity reading of the sample to
™
a confidence of 95%, based on the accumulated trend of
the real time measured values.
The RST mode is best used on samples that settle rapidly
and continuously change in value. The reading is based on
a correctly prepared sample that is homogeneous at the
beginning of the reading. It is best applied to samples that
are greater than 20 NTU. The sample must be mixed
thoroughly by inversion immediately before inserting it into
the meter.
The target icon is shown on the display when the Rapidly
Settling Turbidity is on.
The Rapidly Settling Turbidity reads and calculates five
readings while showing intermediate results.
Apply silicone oil to a sample cell
Sample cells and caps must be extremely clean and free from significant
scratches. Apply a thin coating of silicone oil on the outside of the
sample cells to mask minor imperfections and scratches that may
contribute to light scattering.
Note: Use only the provided silicone oil. This silicone oil has the same refractive
index as the sample cell glass.
1. Clean the inside and
outside of the cells and
caps by washing with a
laboratory glass
cleaning detergent.
Follow with multiple
rinses with distilled or
demineralized water.
2. Apply a small bead
of silicone oil from the
top to the bottom of the
cell.
3. Use the provided
oiling cloth to spread
the oil uniformly. Wipe
off the excess so that
only a thin coat of oil is
left. Make sure that the
sample cell is almost
dry with little or no
visible oil.
Note: Store the oiling
cloth in a plastic
storage bag to keep the
cloth clean.
Indexing a single cell
Precise measurements for very low turbidity samples require the use of
a single cell for all measurements or optically matching the cells. Use
one cell to provide the best precision and repeatability. When one cell is
used, an index or orientation mark (other than the factory-placed
diamond) can be placed on the cell so it is inserted into the instrument
with the same orientation each time.
When using a single cell, make an index or orientation mark on the cell
as follows:
English
17
1. Fill the clean sample
cell to the line with high
quality water (<
0.5 NTU) and cap
immediately. Let the
sample cell degas for at
least five minutes.
2. Wipe with lint-free
cloth. Apply a thin film
of silicone oil (Apply
silicone oil to a sample
cell).
3. Push the POWER
key to turn the meter
on. Place the
instrument on a flat,
sturdy surface.
Note: Do not hold the
instrument while
making measurements.
4. Insert the sample
cell in the instrument
cell compartment so the
diamond or orientation
mark always aligns with
the raised orientation
mark in front of the cell
compartment. Close the
lid.
5. Push the UP and
DOWN key to access
the Reading Options
and then select
Indexing Sample Cell.
Note: The instruments
always stays in the last
selected reading mode.
6. Push Read. The
display shows
Stabilizing then the
turbidity in NTU. Record
the cell position in the
cell compartment and
the reading result.
7. Remove the cell,
rotate it slightly
approximately ⅛ of a
turn and insert it again
into the cell
compartment. Close the
lid.
18 English
8. Push Read. Record
the cell position in the
cell compartment and
the reading result.
9. Repeat step 6 until
the lowest reading is
shown. Place an
orientation mark on the
cell marking band near
the top of the cell so the
cell can be consistently
inserted in the position
that yields the lowest
reading.
Maintenance
Replace the battery
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.
Clean the meter
The meter is designed to be maintenance-free and does not require
regular cleaning for normal operation. Exterior surfaces of the meter may
be cleaned as necessary.
Note: Do not clean the meter with solvents to avoid damaging the material.
1. Clean the meter with a dust- and lint-free dry or slightly damp cloth. A
mild soap solution can also be used for liposoluble contamination.
Store the sample cells
N O T I C E
Do not air dry the sample cells.
Note: Always store the sample cells with caps on to prevent the cells from drying.
1. Fill the sample cells with distilled or demineralized water.
2. Cap and store the sample cells.
3. Wipe the outside of the sample cells dry with the a soft cloth.
W A R N I N G
Explosion hazard. An expired battery can cause hydrogen gas buildup
inside the instrument. Replace the battery before it expires. Do not
store the instrument for long periods with a battery installed.
Potential fire hazard. Use only alkaline or nickel metal hydride batteries (NiMH) in
the meter. Other battery types or incorrect installation can cause a fire. Never mix
battery types in the meter.
For battery replacement refer to Install the battery on page 6.
1. Remove the battery cover.
2. Remove the batteries.
3. Install 4 AA alkaline or 4 AA nickel metal hydride (NiMH) batteries.
Make sure that the batteries are installed in the correct orientation.
4. Replace the battery cover.
W A R N I N G
Replace the lamp
C A U T I O N
Burn Hazard. Wait until lamp cools down. Contact with the hot lamp can cause
burns.
English 19
12
34
20 English
56
78
English 21
Troubleshooting
Refer to the following table for common problem messages or
symptoms, possible causes and corrective actions.
Error/WarningDescriptionSolution
Close lid and push
Read.
Low Battery!Battery is low.
ADC Failure!Hardware error causing
Detector signal too low! Insufficient light on the
Overrange!Turbidity too high-
Underrange!The measured
Please check the lamp! Signals are too low on
The lid is open or lid
detection failed.
reading to fail.
180° detector.
caused probably by
calibrating with
RapidCal™ only.
absorbance is below the
calibration range.
the 90° and 180°
detector.
Make sure that the lid is
closed during reading
and re-read.
• Insert new batteries
• Connect USB/power
module if
rechargeable batteries
are used
Repeat the reading.
• Check for obstructed
light path.
• Check the lamp.
• Calibrate the upper
range.
• Dilute the sample.
Repeat calibration
2100Q: The lamp is
defective. Change the
lamp (refer to Replace
the lamp on page 19).
2100Qis: Contact
technical support.
Error/WarningDescriptionSolution
Temperature too high!
Switch off instrument.
RST: Average value!Solids are settling too
Confidence level is <
95%
Standard value out of
range. Insert standard
and push Read
ID already in use. Enter
new ID
Error - Security Please
set password before
activating security
Please enter at least
one character.
Password incorrect.
Please retry.
Temperature has
exceeded the meter
limits (>60 °C or
>140 °F).
slowly. The reading
mode is not suitable for
this sample.
The reading mode
Rapidly Settling Turbidity
did not meet the range of
≥ 95% confidence.
Used incorrect standard
value for the reading.
The Operator or Sample
ID is unavailable as it is
already assigned.
No password is created.Create a new password.
Password must contain
minimum of one
character.
Incorrect password was
entered.
Turn off the meter and let
it cool down.
Select Normal or Signal
Average reading mode.
• Invert the sample
several times so that
the solids allocate.
Repeat the reading
again.
• Switch to the Normal
reading mode if the
sample is stable and
does not have settable
solids.
Insert the appropriate
standard and read again.
Create a new ID.
Create a password of at
least one character.
Enter the appropriate
password.
22 English
Error/WarningDescriptionSolution
Please disconnect the
USB cable from your
computer.
USB module memory
full. Delete data and try
again.
Delete Last Reading
Failed!
Delete Data Log failed!
Can't read data set!
Can't store data!
Can't store to the
Reading Log!
Can't store to the Verify
Cal Log!
Error storing data!
Error reading data!
Data storage does not
respond while connected
to the meter and the
computer.
Data storage is full.1. Connect USB/power
Error in the data storage. Turn the meter off and
Disconnect the USB
cable from the meter and
try sending data again.
module to the computer.
2. Download the stored
data to the computer.
3. Delete Data Log on
the module.
on. If the error message
still occurs, contact
technical support.
Replacement parts (continued)
DescriptionQuantityItem no.
Insert, molded bottom12971507
Sample cell oiling cloth14707600
1" glass sample cell (10 ml ) w/cap (Turb)pkg/62434706
Carrying case (includes insert)12971500
Battery set, AA alkaline batteriespkg/41938004
Lamp assy14653900
Blank module1LZV797
Rubber foot set1LZV821
Lamp cover (includes screws)1LZV822
Battery cover (includes 2 feet)1LZV823
Module cover1LZV824
Connector cover for USB/power module1LZV825
Connector cover for power module1LZV826
Lid (includes magnet)1LZV827
Accessories
Replacement parts and accessories
Replacement parts
DescriptionQuantityItem no.
StablCal ampule calibration kit12971205
10 NTU verification standard100 mL2961701
Silicone Oil15 mL126936
DescriptionQuantityItem no.
USB/power module (includes: universal
power supply, USB cable, instruction
sheet)
Power module (includes: universal power
supply, instruction sheet)