General information.....................................................................................................................................................................6
Use of hazard information..................................................................................................................................................................6
Put paper in the printer......................................................................................................................................................................9
User interface................................................................................................................................................................................10
Turn the instrument on.....................................................................................................................................................................12
Turn the keypad sound off (optional)...............................................................................................................................................12
Set the date and time.......................................................................................................................................................................12
Show the current time (optional)......................................................................................................................................................12
Standard operation....................................................................................................................................................................12
Calibrate the turbidimeter with StablCal® Standards.......................................................................................................................12
Prepare the StablCal standards................................................................................................................................................13
Using Gelex secondary standards...................................................................................................................................................15
Optical system check................................................................................................................................................................17
Prepare a sample cell......................................................................................................................................................................17
1
Table of Contents
Clean the sample cell...............................................................................................................................................................18
Indexing a single sample cell....................................................................................................................................................19
Prepare the sample..........................................................................................................................................................................23
Prepare a representative sample..............................................................................................................................................23
Remove air bubbles from the sample.......................................................................................................................................23
Apply a vacuum.................................................................................................................................................................23
Use an ultrasonic bath.......................................................................................................................................................23
Prevent condensation on a sample cell....................................................................................................................................24
Absorbance and transmittance measurement.................................................................................................................................27
Absorbance and transmittance measurement procedure.........................................................................................................28
Color measurement.........................................................................................................................................................................29
Color measurement and calibration procedure.........................................................................................................................30
Manual or automatic ranging....................................................................................................................................................32
Signal averaging on or off.........................................................................................................................................................32
Ratio on or off...........................................................................................................................................................................32
Using the air purge system.......................................................................................................................................................33
Using a flow cell........................................................................................................................................................................33
Install a flow cell................................................................................................................................................................33
Clean a flow cell assembly................................................................................................................................................34
Using a manual flow cell....................................................................................................................................................35
Using an automated flow cell.............................................................................................................................................35
Static or dynamic measurement procedure.......................................................................................................................36
Use a cell adapter.....................................................................................................................................................................37
Install a cell adapter...........................................................................................................................................................38
Remove a cell adapter.......................................................................................................................................................38
Using the optional filter assemblies..........................................................................................................................................38
Application development using alternate wavelengths......................................................................................................38
Connect to a printer or computer.....................................................................................................................................................39
Configure the printer output.............................................................................................................................................................39
Configure the RS232 connection.....................................................................................................................................................39
Connect to a data recorder..............................................................................................................................................................40
Configure the data recorder output..................................................................................................................................................40
Calibrate the turbidimeter with formazin standards..........................................................................................................................40
Making 4000-NTU formazin stock solution...............................................................................................................................44
Calibrate the turbidimeter with user-selected formazin standards...................................................................................................44
Prepare formazin standards – user selected............................................................................................................................44
Change the calibration points...................................................................................................................................................44
Special research applications..........................................................................................................................................................45
Application specific methods............................................................................................................................................................45
Application specific calibration.........................................................................................................................................................45
Program new ASC data............................................................................................................................................................46
Make a print of ASC data points...............................................................................................................................................46
Change an ASC unit name or one ASC data point...................................................................................................................47
Delete one ASC data point.......................................................................................................................................................47
3
Table of Contents
Delete all the ASC data points..................................................................................................................................................48
Clean the instrument........................................................................................................................................................................48
Change the filter assembly..............................................................................................................................................................48
Clean the filter assembly..................................................................................................................................................................48
Replace the lamp.............................................................................................................................................................................48
Replace a fuse.................................................................................................................................................................................51
Replacement parts and accessories...............................................................................................................................53
4
Specifications
Specifications are subject to change without notice.
SpecificationDetails
Measurement methodNephelometric
RegulatoryMeets EPA Method 180.1
ASTM D7315 - Standard Test Method for
Determination of Turbidity Above 1 Turbidity Unit (TU)
in Static Mode
ASTM D6655 - Standard Test Method for
Determination of Turbidity Below 5 NTU in Static Mode
In no event will the manufacturer be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or
omission in this manual. The manufacturer reserves the right to make
changes in this manual and the products it describes at any time, without
notice or obligation. Revised editions are found on the manufacturer’s
website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or
misuse of this product including, without limitation, direct, incidental and
consequential damages, and disclaims such damages to the full extent permitted
under applicable law. The user is solely responsible to identify critical application
risks and install appropriate mechanisms to protect processes during a possible
equipment malfunction.
Please read this entire manual before unpacking, setting up or operating
this equipment. Pay attention to all danger and caution statements.
Failure to do so could result in serious injury to the operator or damage
to the equipment.
Make sure that the protection provided by this equipment is not impaired.
Do not use or install this equipment in any manner other than that
specified in this manual.
Use of hazard information
Indicates a potentially or imminently hazardous situation which, if not avoided, will
result in death or serious injury.
Indicates a potentially or imminently hazardous situation which, if not avoided,
could result in death or serious injury.
D A N G E R
W A R N I N G
6 English
C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate
injury.
Indicates a situation which, if not avoided, may cause damage to the instrument.
Information that requires special emphasis.
N O T I C E
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed. A symbol, if noted
on the instrument, will be included with a danger or caution statement in
the manual.
This symbol, if noted on the instrument, references the instruction
manual for operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of
in European public disposal systems after 12 August of 2005. In
conformity with European local and national regulations (EU Directive
2002/96/EC), European electrical equipment users must now return
old or end-of-life equipment to the Producer for disposal at no charge
to the user.
Note: For return for recycling, please contact the equipment producer or supplier
for instructions on how to return end-of-life equipment, producer-supplied
electrical accessories, and all auxiliary items for proper disposal.
Certification
Canadian Radio Interference-Causing Equipment Regulation,
IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian
Interference-Causing Equipment Regulations.
Cet appareil numèrique de la classe A respecte toutes les exigences du
Rëglement sur le matériel brouilleur du Canada.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device
complies with Part 15 of the FCC Rules. Operation is subject to the
following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including
interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by
the party responsible for compliance could void the user's authority to
operate the equipment. This equipment has been tested and found to
comply with the limits for a Class A digital device, pursuant to Part 15 of
the FCC rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference
to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference, in which case the user will be
required to correct the interference at their expense. The following
techniques can be used to reduce interference problems:
1. Disconnect the equipment from its power source to verify that it is or
is not the source of the interference.
2. If the equipment is connected to the same outlet as the device
experiencing interference, connect the equipment to a different
outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the
interference.
5. Try combinations of the above.
Product overview
The 2100AN laboratory turbidimeter measures turbidity in NTUs
(nephelometric turbidity units), NEPs (nephelos) and EBCs (European
Brewing Convention units). NEPs and EBCs are calculated using the
conversion factors of 6.7 nephelos per 1.0 NTU and 0.245 EBCs per
1.0 NTU. The 2100AN turbidimeter also measures absorbance,
transmittance and color (APHA Pt-Co Method).
English
7
In addition, two user-defined measurement units can be specified. Refer
to Application specific methods on page 45. The application specific
mode of operation uses the nephelometric optical system and the NTU
measurement mode.
The turbidimeter has a built-in printer and an RS232 output for
connection to a printer, data logger or computer and a recorder output.
The turbidimeter contains a real-time clock with battery. The clock
provides a time-date stamp on all data transmitted to the built-in printer
or to external devices by way of the RS232 interface (i.e., measurements
and calibration records).
Figure 1 Front overview
Figure 2 Back overview
1 Power cord connector5 Air purge fitting
2 Fuse holder6 Recorder output jack for a chart
recorder (0 to 1 V output)
3 Power switch7 Remote cable jack for flow valve
module connection to the automatic
flow cell (low pressure)
4 DB9 connector for RS232 cable8 Lamp access cover
Product components
Refer to Figure 3 to make sure that all components have been received.
If any of these items are missing or damaged, contact the manufacturer
or a sales representative immediately.
Do not remove the plastic wrapper from the paper rolls until the paper is
installed.
standardization kit
11 Power cord
1
2
Installation
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.
Put paper in the printer
N O T I C E
Use only the provided thermal paper. Use of other thermal paper may cause poor
print quality and decrease the life of the print-head.
Notes:
• Do not rub the thermal paper with a hard object.
• Do not use chemical paste on thermal paper.
• A red line on the edge of the thermal paper shows when the paper
supply is low.
1. Cut the end of the paper with scissors to make an arrow shape.
2. Open the printer cover.
3. Put the point of the thermal paper in the paper entrance slot.
4. Push the paper through until the point of the paper comes out the
exit slot.
5. Pull the paper out of the exit slot until the full width of the paper is
past the exit slot.
6. Put the paper roll in the printer.
7. Put the thermal paper through the slot in the printer cover, then close
the printer cover.
English
9
User interface
Figure 4 Keypad
1 ENTER key7 FLOW key
2 EDIT (arrow) keys8 SIGNAL AVG key
3 SAMPLE key9 PRINT key
4 RANGE key10 LINE FEED key
5 UNITS/Exit key11 SETUP key
6 RATIO key12 CAL/Zero key
Table 1 Key descriptions
KeyDescription
Enters the value on the display. Starts the measurement of a
calibration standard. Clears data from the buffer.
Table 1 Key descriptions (continued)
KeyDescription
Starts the changing of the sample number shown on the mode
display.
Selects automatic or manual ranging.
Selects the unit of measure. Exits Calibration or Setup mode
without saving changes.
Turns Ratio on or off.
Turns on or off the Flow mode of operation. Used only with the
automated flow cell.
Turns signal averaging on or off.
Sends the data that is on the display to a printer or computer.
Sends a calibration data report to a printer or computer when in
Calibration mode. Sends diagnostic results to a printer or computer
if held down when the instrument is turned on. Provides a print of
the setup commands when in Setup mode. Turns the print interval
feature on or off if the instrument has been configured with a printer
interval.
Moves the printer paper forward one line.
10English
Changes the numbers and/or letters on the display. Steps through
the calibration standards. The right arrow key moves the cursor to
the previous or next digit.
Turns on Setup mode and starts the selection of the setup number
on the mode display.
Starts a calibration when in NTU, NEP or EBC mode. Starts
analytical zeroing when in CU, %T or Abs mode.
Figure 5 Indicator lights
1 Lamp icon light7 PRINT light
2 "CAL?" light8 SIGNAL AVG light
3 "Manual" light9 CAL/Zero light
4 "Auto" light10 SETUP light
5 RATIO light11 SAMPLE light
6 FLOW light
Table 2 Light descriptions
LightDescription
Illuminated when the instrument light source is on.
Flashes when there is not sufficient light for measurement.
CAL?"CAL?" is shown during a calibration if the calibration data is not
within the acceptable range.
Flashes when the instrument should be calibrated.
Note: The CAL? light applies when the USEPA filter and a 25-mm sample cell
are used. Ignore the CAL? light if illuminated during calibration when a different
filter or a smaller sample cell is used. Push UNITS/Exit to start measurements.
Manual "MANUAL" is shown above the Range Mode label when the
RATIOIlluminated when Ratio is on.
FLOWIlluminated when the Flow mode of operation is selected.
PRINTIlluminated when the printer interval feature is selected.
SIGNAL
SETUP Illuminated when Setup mode is selected.
SAMPLE Illuminated when Sample mode is selected.
instrument is in manual ranging mode.
Auto"AUTO" is shown below the Range Mode label when the instrument is
in auto ranging mode.
Flashes when the flow cycle is done.
Flashes when a print interval has been selected but is not active.
Illuminated when signal averaging is on.
AVG
CAL
Illuminated when Calibration or Zeroing mode is selected.
Zero
English 11
Startup
Turn the instrument on
1. Put the instrument on a stable, level surface that is free of vibration.
Do not put in direct sunlight.
2. Make sure that there is air circulation around the instrument. Keep
the back and area below the instrument free of material that could
decrease air flow through the vents.
3. Connect the power cord to the power plug on the back of the
instrument.
4. Connect the power cord to a power socket with ground contact.
5. Push the power switch on the back of the instrument to turn the
instrument on.
Turn the keypad sound off (optional)
By default, the instrument makes an audible sound when a key is
pushed. To turn the keypad sound off:
1. Push SETUP. The SETUP light turns on.
2. Use the arrow keys to select 00.
3. Push ENTER.
4. Use the arrow keys to select the sound option:
OptionDescription
BEEP ONAn audible sound is made when a key is pushed.
BEEP OFFNo sound is made when a key is pushed.
5. Push ENTER.
6. Push SETUP.
Set the date and time
1. Push SETUP. The SETUP light turns on.
2. Use the arrow keys to select an option:
OptionDescription
05Sets the hours and minutes (HH-MM).
06Sets the month and day (MM-DD).
07Sets the year (YY).
3. Push ENTER.
4. Use the arrow keys to change the value.
5. Push ENTER.
6. Push SETUP.
Show the current time (optional)
1. Push SETUP. The SETUP light turns on.
2. Use the arrow keys to select 08.
3. Push ENTER. The current time is shown on the display (HH-MM-
SS).
4. Push SETUP.
Standard operation
Calibrate the turbidimeter with StablCal® Standards
Calibrate the turbidimeter before it is used for the first time using the
StablCal® sealed vial standards provided. As an alternative, calibration
can be done with recently prepared formazin standards. Refer to
Calibrate the turbidimeter with formazin standards on page 40.
Calibrate the turbidimeter at least every 3 months or as specified by the
regulating authority when data is used for USEPA reporting.
The instrument is ready for calibration 60 minutes after start-up. Keep
the instrument on 24 hours a day if the instrument is used regularly.
Note: Unknown results may occur if standards other than the recommended
calibration points are used. The recommended calibration points (< 0.1, 20, 200,
1000, 4000 and 7500 NTU) provide the best calibration accuracy. Use of standards
other than StablCal, or user-prepared formazin, may result in less accurate
12
English
calibrations. The manufacturer cannot guarantee the performance of the instrument
if calibrated with co-polymer styrenedivinylbenzene beads or other suspensions.
Prepare the StablCal standards
When received and at intervals:
1. Clean the exterior surface of the StablCal vials with laboratory glass
cleaning detergent.
2. Rinse the vials with distilled or deionized water.
3. Dry the vials with a lint-free cloth.
Note: Never shake or invert the < 0.1 NTU standard. If the standard has been
mixed or shaken, do not move the vial for 15 minutes or more before using.
Note: Do not remove the caps from the sealed vials.
Make sure that the StablCal standards are at ambient instrument
temperature before use (and no greater than 40 °C (104 °F)).
Mix the standards before use:
1. Open the case lid. Remove the < 0.1 NTU standard from the plastic
case.
2. Leave the other standards in the case. Close the case lid.
3. Shake the case vigorously for at least 10 seconds.
4. Let the standards stand with no movement for 3–5 minutes before
use.
Calibration notes
• Make sure that the instrument is in the same ambient conditions as
where it is used.
• Make sure that the standards are at the same ambient temperature as
the instrument before use.
• Use only the provided silicone oil. This silicone oil has the same
refractive index as the vial glass and masks minor glass differences
and scratches.
• Store the oiling cloth in a plastic storage bag to keep the cloth clean.
• If power is lost during calibration, the new calibration data is lost and
the last calibration data is used. To exit a calibration and not save the
new values, push UNITS/Exit.
• In Calibration mode, automatic range and signal averaging on are
selected. When calibration is completed, all operational modes go
back to the last settings.
• All nephelometric (turbidity units of measure) calibrations are done at
the same time.
• Ratio-on and Ratio-off calibration data is measured and recorded at
the same time.
• The 7500-NTU standard does not have to be measured during
calibration if turbidity less than 4000 NTU will be measured. Push
CAL/Zero after the 4000 NTU standard is measured to complete the
calibration procedure.
• Clean the USEPA filter assembly before doing a primary calibration,
or at least every 3 months (which is the USEPA-recommended
primary calibration interval).
English 13
StablCal calibration procedure
1. Remove the filter
assembly. Refer to
Change the filter
assembly on page 48.
7. Use the oiling cloth
to apply the oil equally
to the surface of the
vial. Remove the
excess oil. Make sure
that the vial is almost
dry.
14 English
2. Clean the lens of the
USEPA filter assembly.
Refer to Clean the filter
assembly on page 48.
8. Put the vial in the
sample cell holder with
the triangle on the vial
aligned with the
reference mark on the
sample cell holder.
Close the cover.
3. Hold the tab of the
USEPA filter assembly
so that the arrows point
toward the front of the
instrument. Push the
filter assembly fully in
the housing.
9. Push ENTER.
The instrument display
counts down, then
measures the standard.
The next expected
standard (e.g., 20.00) is
shown. The mode
display shows "01".
4. Push CAL/Zero.
The CAL/Zero light
turns on, and the mode
display shows "00". The
NTU value of the
dilution water that was
used in the previous
calibration is shown on
the display.
10. Remove the vial
from the sample cell
holder.
5. Get the < 0.1 NTU
vial. Clean the vial with
a soft, lint-free cloth to
remove water spots and
fingerprints. Do not
invert the vial.
11. Do steps 5–10 for
the other StablCal vials
(from lowest to highest
NTU standard).
The mode display
shows "00" after the
last vial is measured.
6. Apply a small bead
of silicone oil from the
top to the bottom of the
vial.
12. Push CAL/Zero.
The instrument saves
the new calibration data
and goes back to
Measurement mode.
StablCal standards storage
instrument due to small differences in glass and instrument optical
systems.
• Do not move a StablCal standard to a different container for storage.
Keep StablCal standards in the plastic case provided with the cover
closed.
• Store at 5 to 25 °C (41 to 77 °F).
• For long-term storage (more than one month between use), keep at
5 °C (41 °F).
• Do not keep a Gelex vial in the instrument for more time than is
necessary to complete measurement. The heat from the lamp can
change the turbidity value of a Gelex vial.
• Keep the Gelex standards at room temperature. Do not let Gelex
standards freeze or become warmer than 50 °C (122 °F). High
temperatures may cause Gelex suspensions to divide.
• Make sure that the Gelex standards are at ambient instrument
Using Gelex secondary standards
temperature before measurement.
The Gelex secondary standards are used when a calibration check or an
optical system check is done. Refer to Calibration verification
on page 17 and Optical system check on page 17.
Gelex notes
• Measure the Gelex secondary standards on the instrument on which
they will be used. The measured values can only be used for one
Measure the Gelex stray light standard
Measure the Gelex stray light standard when the instrument is first received. Record the value on the Gelex vial with a permanent marker one time.
1. Clean the stray light
standard with a soft,
lint-free cloth to remove
water spots and
fingerprints.
2. Apply a small bead
of silicone oil from the
top to the bottom of the
vial.
3. Use the oiling cloth
to apply the oil equally
to the surface of the
vial. Remove the
excess oil. Make sure
that the vial is almost
dry.
4. Push RANGE to
select automatic
ranging.
"AUTO" is shown below
the Range Mode label
on the instrument.
5. Push SIGNAL AVG
to turn signal averaging
off.
The SIGNAL AVG light
turns off.
6. Push UNITS/Exit to
select the NTU
measurement mode.
English 15
7. Push RATIO to turn
Ratio mode on.
8. Put the stray light
standard in the sample
cell holder with the
triangle on the vial
aligned with the
reference mark on the
sample cell holder.
Close the cover.
9. Read the value
when stable. Remove
the vial from the
instrument.
10. Record the value
on the white diamond
space on the vial using
a permanent marker.
Measure the Gelex secondary turbidity standards
Measure the Gelex secondary turbidity standards each time the instrument is calibrated and record the new values on the Gelex vials with a water
soluble marker.
1. Clean the Gelex
vials with a soft, lint-free
cloth to remove water
spots and fingerprints.
16 English
2. Apply a small bead
of silicone oil from the
top to the bottom of the
vial.
3. Use the oiling cloth
to apply the oil equally
to the surface of the
vial. Remove the
excess oil. Make sure
that the vial is almost
dry.
4. Push RANGE to
select automatic
ranging.
"AUTO" is shown below
the Range Mode label
on the instrument.
5. Push SIGNAL AVG
to turn signal averaging
off.
The SIGNAL AVG light
turns off.
6. Push UNITS/Exit to
select the NTU
measurement mode.
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.