1.1 Technical data on the 1200-S sc - pH / ORP sensor................................................................................................. 3
Section 2 General information ..................................................................................................................................... 5
2.1 General handling instructions .................................................................................................................................... 5
Section 3 General safety instructions ......................................................................................................................... 7
3.1 Possible sources of hazards...................................................................................................................................... 7
3.3 Electrical safety measures and fire prevention measures ......................................................................................... 8
3.4 Chemical safety measures ........................................................................................................................................ 9
3.5 Safety measures related to the flow of sample.......................................................................................................... 9
5.3 Sensor data logger .................................................................................................................................................. 15
5.4 The commands under SENSOR DIAG.................................................................................................................... 15
5.5 The commands under SENSOR SETUP ................................................................................................................ 16
5.6.1 Calibration using a buffer solution.................................................................................................................. 17
5.6.2 Calibration using two buffer solutions ............................................................................................................ 17
5.6.3 Calibration using a sample solution ............................................................................................................... 18
5.6.4 Calibration using two sample solutions.......................................................................................................... 18
6.2 Cleaning the sensor................................................................................................................................................. 21
7.3 Important service data ............................................................................................................................................. 24
Section 9 Warranty and liability..................................................................................................................................27
Appendix A ModBUS Register Information ..............................................................................................................31
II
Section 1Specifications
1.1Technical data on the 1200-S sc - pH / ORP sensor
MaterialsStainless steel metal housing
Enclosure ratingIP 68; stainless steel metal housing
pH measuring range0 pH ... 14 pH
ORP measuring range–1,500 ... 1,500 mV
Temp. measuring range–5 °C ... 50 °C
Storage temperature
Sensor and controller
pH / ORP response time<15s; T90
Temp. response time<2min; T90
pH measuring accuracy± 0.02 pH
ORP measuring accuracy± 1.2 mV
Temp. measuring accuracy± 0.2 °C
Reproducibility± 0.5 % of the end value of the measuring range
Sensitivity± 0.5 % of the end value of the measuring range
Calibration, pH
Calibration, ORPOne point, comparative measurement
Calibration, tempOne point, comparative measurement
Max. immersion depth / pressure
for the sensor
Maximum flow speed4 m/s
Sensor interfaceMODBUS
–20 °C ... 60 °C; 95 % relative humidity, non-condensing
One or two point, standard buffer solution (automatic),
One or two point, comparative measurement
20 m / 2 bar over-pressure
Sensor cable10 m, hard wired, polyurethane
Sensor weight<1 kg
Sensor dimensions (Ø × L) 42 × 504 mm
Fastening
Service life measuring electrodeApprox. 1 year
Sensor power< 7 W
• Immersed pipe
• Chain
Specifications are subject to change without notice.
3
Specifications
4
Section 2General information
2.1General handling instructions
CAUTION!
The sensor will only work correctly when the tip of the probe is completely
immersed in liquid. The measuring probe must not be stored dry for more
than approx. 10 minutes or protrude from the medium to be measured. Use
the transport cap during the installation of the sensor, during maintenance
and when transporting, and fill the cap with 3 mol. KCl or pH 4 buffer
solution.
CAUTION!
Potential danger with contact with chemical/biological substances. Working
with chemical samples, standards and reagents can be dangerous. Make
yourself familiar with the necessary safety procedures and the correct
handling of the chemicals before use and read and follow all relevant safety
data sheets.
2.2Applications
The sensor makes it possible to simply and exactly measure pH directly in
aqueous solutions.
2.3Measuring principle
2.3.1 pH measurement
Typical applications include
•Inlet and / or outlet of a sewage treatment plant
•Activated sludge tanks or
•Surface water (rivers, streams or lakes).
Various different possible installations enable the system to be adapted to a very
wide range of conditions.
pH is the negative logarithm of the hydrogen ion activity as well as a
measurement of the acidic or alkaline content of a solution:
pH = –log10 [aH+]
The pH value is normally measured using a glass electrode and a reference
electrode. In this case the glass electrode acts as the transducer that converts the
chemical energy (hydrogen ion activity) into electrical energy (measured in
millivolts). The reaction is symmetrical and the electrical circuit is closed by the ion
flow from the reference solution to the test solution.
Together with the reference solution, the electrode generates a voltage (EMF).
The magnitude of the voltage is dependent on the type of reference electrode, the
internal design of the glass electrode, the pH value of the solution as well as the
temperature.
5
General information
This voltage is expressed by the following Nernst equation:
E = Eo – (2.3 RT/F)xlog a[H+]
E = Eo – (gradient) xlog a[H+]
Where the following applies:
2.3.2 REDOX measurement
E
Eo
R
T
a[H+]
F
=
EMF value of the cell
=
Zero voltage (isopotential) of the system (depends on the internal
design of the glass electrode and reference electrode)
=
Gas constant
=
Temperature in Kelvin
=
Activity of the hydrogen ions
(corresponds to the concentration of the hydrogen ions)
=
Faraday constant
With every unit change in the pH (or decimal change in the ion concentration), the
EMF produced by the pair of electrodes changes by 59.16 mV at 25 °C. This
value is termed the gradient of the electrode.
The pH electrode pair is calibrated using solutions with a known and constant
hydrogen ion concentration. These solutions are termed buffer solutions. The
buffer solutions are used for the calibration of both the electrode isopotential and
the gradient.
In a Redox system the measurements are made using a balanced instrument that
comprises a Redox electrode and a reference electrode. The potential to be
measured is termed the Redox potential and depends on the relationship between
the activities of the two components of a Redox system and the number of
electrons transferred. In many cases, the pH value of the solution also affects the
potential.
The half-cell potential εB of the reference electrode has a large effect on the
potential E of the measuring chain. To rectify this effect, the potential of the
measuring electrode can be referred to the hydrogen electrode. If εB is the halfcell potential of the reference electrode used, the calculation is performed using
the following formula:
ε(H) = E + εB
ε(H)
εB
EMF value of the cell
=
=
Half-cell potential
Such standardised Redox potentials provide a certain amount of information on
the oxidisation or reduction potential of a Redox system. Increasing positive
values indicate increasing oxidation energy. The more negative the potential, the
higher the reduction energy. The area that is of interest in practice is between
+1500 and –1000 mV.
Standard potentials for a Redox system for aOx = aRed (a=activity) and for pH = 0
can be defined. This is turn corresponds to a standardised hydrogen ion activity of
aH+ = 1 mole per litre.
The stability and ability to reverse a Redox system have a significant effect on the
reproducibility of the Redox potential measured.
6
Section 3General safety instructions
Prior to unpacking, commissioning or operating the instrument, read all of this
manual.
Please pay particular attention to all instructions on hazards and safety. Otherwise
there is a risk of serious injury to the operator or damage to the instrument, or
pollution.
The sensor is only allowed to be installed and used as per the instructions in this
manual.
3.1Possible sources of hazards
During the operation or calibration of the sensor, there exist the following sources
of hazards if the safety instructions are not observed:
•Potentially hazardous materials (buffer solutions, flow of sample)
In all circumstance observe the safety data sheets and the applicable health and
safety instructions.
3.2Safety symbols
This symbol, if present on the instrument, refers to information in the operating instructions on safe operation and / or
instructions that provide safety information.
This symbol, if present on a housing or a protective cover, identifies the risk of an electric shock (which may under
certain circumstances be fatal). Only personnel qualified for working on hazardous voltages are allowed to open the
enclosure or remove the protective cover.
This symbol, if present on the instrument, identifies the location of a fuse or current limit.
This symbol, if present on the instrument, identifies a part that may become hot and must not be touched without
taking precautions.
This symbol, if present on the instrument, indicates the presence of components that could be damaged by
electrostatic discharge. Appropriate precautions are to be taken.
This symbol, if present on the instrument, indicates the presence of dangerous chemical substances. Chemicals are
only allowed to be handled and maintenance on devices for supplying chemicals is only allowed to be performed by
personnel qualified and trained for working with chemicals.
All stickers and labels on the instrument are to be observed. Otherwise injuries,
pollution or damage to the instrument may occur.
7
General safety instructions
This symbol, if present on the instrument, indicates that safety glasses must be worn.
This symbol, if present on the instrument, identifies the location of the connection for the protective earth (ground).
As of 12 August 2005, electrical appliances marked with this symbol are no longer allowed to be disposed of in
Europe in unsorted household or industrial waste. As per the applicable regulations (EU directive 2002/96/EC), from
this date on consumers in the EU must return old appliances to the manufacturer for disposal. This disposal is free of
charge for the consumer.
Note: You can obt ain instructions on the correct disposal of all (marked and unmarked) electrical products that have
been supplied or manufactured by Hach-Lange from your local Hach-Lange sales office.
3.3Electrical safety measures and fire prevention measures
The following safety instructions must be observed during the installation
and repair of cables that carry electrical power:
DANGER!
Sensors and controller are designed for compliance with the U.S. and
Canadian NEC as well as the European low voltage directive. No internal
electrical or electronic parts are allowed to be modified in any way, as this
could render the CE conformity void.
WARNING!
Only qualified experts may perform the tasks described in this section of the
manual, while adhering to all locally valid safety regulations.
•Prior to maintenance or repair of the instrument, isolate it from the power
supply.
•When making electrical connections, all applicable local and national
regulations are to be met.
•The use of earth leakage trips is strongly recommended.
•The instrument must be correctly earthed for correct operation.
8
3.4Chemical safety measures
CAUTION!
Reference and standard solutions are used for the calibration. Some of
these compounds are toxic or caustic.
Potential danger with contact with chemical/biological substances. Working
with chemical samples, standards and reagents can be dangerous. Make
yourself familiar with the necessary safety procedures and the correct
handling of the chemicals before use and read and follow all relevant safety
data sheets.
Physical contact with a calibration solution and inhalation of vapours from a
calibration solution are to be avoided or limited to an absolute minimum.
3.5Safety measures related to the flow of sample
The assessment of the possible hazards from the individual sample flows is the
responsibility of the user. Suitable safety measures are to be taken to avoid any
unnecessary contact with a flow of sample of unknown composition in relation to
the hazards due to traces of chemicals, radiation or biological effects.
General safety instructions
9
Loading...
+ 23 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.