Hach MP-4, MP-6, MP-6p User Manual

DOC022.53.80072

MP Series Portable Meters

USER MANUAL
October 2009, Edition 1
© Hach Company, 2009. All rights reserved. Printed in the U.S.A. hh/dk
2

Table of contents

Section 1 Specifications .................................................................................. 7
2.1 Safety information ..................................................................................... 11
2.1.1 Use of hazard information .............................................................. 11
2.1.2 Precautionary labels ....................................................................... 11
2.2 General product information .....................................................................12
2.2.1 Overview ......................................................................................... 12
2.2.2 Features common to all models ...................................................... 12
2.2.3 User mode features ....................................................................... 12
2.3 Conductivity and pH/ORP sensor cups .....................................................13
Section 3 Operation ........................................................................................ 15
3.1 System startup ..........................................................................................15
3.2 Display description ....................................................................................15
3.3 Keypad description ................................................................................... 16
3.4 Take a measurement ................................................................................ 17
3.5 Measure conductivity ................................................................................17
3.6 Measure resistivity (MP-4 and MP-6 models) ........................................... 17
3.7 Measure mineral/salt (MP-6p model only) ................................................ 18
3.8 Measure TDS ............................................................................................ 18
3.9 Measure ORP/Redox (MP-6 and MP-6p models) ..................................... 18
3.10 Measure pH (MP-6 and MP-6p models) ................................................. 18
3.11 Select a solution .....................................................................................19
3.11.1 Temperature compensation ..........................................................19
3.12 Change the user-selected temperature compensation factor ................. 20
3.12.1 Disable temperature compensation ............................................. 20
3.13 Change the user-selected conductivity/TDS ratio ................................... 20
3.14 Settings ................................................................................................... 21
3.14.1 Store a value in the memory ......................................................... 21
3.14.2 View the memory recall ................................................................ 22
3.14.3 Clear all records ...........................................................................22
3.15 Time and date ......................................................................................... 22
3.15.1 Set the time .................................................................................. 22
3.15.2 Set the date .................................................................................. 23
3.15.3 Set the date format ....................................................................... 24
3.16 Temperature format ................................................................................ 24
3.17 Return to factory settings ........................................................................25
3.18 Cell check ...............................................................................................25
3.19 Auto off ................................................................................................... 26
3.20 User mode calibration Linc
3.20.1 Calibrate meter for User mode ..................................................... 27
3.20.2 Set User mode calibration Linc ..................................................... 27
3.20.3 Cancel User mode calibration Linc ...............................................28
3.21 Download stored data ............................................................................. 29
Section 4 Calibration ...................................................................................... 31
TM
function .................................................... 27
3
Table of contents
4.1 Calibration intervals ...................................................................................31
4.2 Calibration limits ........................................................................................31
4.3 Calibration records ....................................................................................31
4.4 Calibrate the meter ....................................................................................31
4.5 Exit calibration mode .................................................................................32
4.6 Calibrate conductivity, mineral/salt or TDS ................................................32
4.7 Calibrate resistivity ....................................................................................33
4.8 Reset factory calibration—conductivity, mineral/salt or TDS .....................33
4.9 pH calibration .............................................................................................33
4.10 Set multiple point pH calibrations ............................................................34
4.11 ORP calibration .......................................................................................35
4.12 Temperature calibration ...........................................................................35
Section 5 Maintenance ...................................................................................37
5.1 Temperature extremes ..............................................................................37
5.2 Battery replacement ..................................................................................37
5.2.1 Maintain the conductivity cup ..........................................................38
5.2.2 Maintain the pH/ORP sensor cup ....................................................38
5.3 pH/ORP sensor replacement .....................................................................38
5.4 Clean the sensors ......................................................................................38
5.4.1 Clean the conductivity/resistivity/TDS sensor .................................38
5.4.2 Clean the pH/ORP sensor ...............................................................39
Section 6 Troubleshooting ............................................................................41
Section 7 Contact Information ......................................................................43
Section 8 Replacement parts and accessories ......................................45
8.1 Replacement parts ....................................................................................45
8.2 Accessories ...............................................................................................45
8.3 Consumables .............................................................................................45
8.4 Recommended cleaning consumables ......................................................46
Section 9 Limited Warranty ...........................................................................47
Appendix A Temperature compensation ..................................................49
A.1 Compensation to 25 ºC .............................................................................49
A.2 Changes in temperature compensation ....................................................49
A.3 Graph of comparative error .......................................................................50
A.4 Other solutions ..........................................................................................50
Appendix B Conductivity conversion .........................................................51
B.1 How conductivity conversion works ..........................................................51
B.2 Solution characteristics .............................................................................51
Appendix C Temperature compensation and TDS derivation ...........53
C.1 Conductivity characteristics ......................................................................53
C.2 Temperature compensation of unknown solutions ...................................53
C.2.1 Find temperature compensation by calculation ..............................53
C.2.2 Find temperature compensation by adjustment ..............................53
C.3 TDS ratio of unknown solutions ................................................................54
4
Table of contents
Appendix D Additional information on pH and ORP (MP-6
and MP-6p models) .........................................................................................55
D.1 pH ............................................................................................................. 55
D.1.1 pH as an indicator .......................................................................... 55
D.1.2 pH units .......................................................................................... 55
D.1.3 pH sensor ....................................................................................... 55
D.1.4 Sources of error .............................................................................56
D.1.5 Temperature compensation ........................................................... 57
D.2 Oxidation Reduction Potential/Redox (ORP) .............................. 57
D.2.1 ORP as an indicator ....................................................................... 57
D.2.2 ORP units ....................................................................................... 57
D.2.3 ORP sensor ................................................................................... 57
D.2.4 Sources of error .............................................................................57
5
Table of contents
6

Section 1 Specifications

Specifications are subject to change without notice.
General
Display 4-digit LCD Dimensions (L x W x H) 196 x 68 x 64 mm (7.7 x 2.7 x 2.5 in.) Weight 352 g (12.4 oz) Case material VALOX COND/RES/TDS cell material VALOX COND/TDS electrodes (4) 316 stainless steel COND/RES/TDS cell cup capacity 5 mL (0.2 oz) pH /ORP sensor cup capacity 1.2 mL (0.04 oz) Power 9V alkaline battery Battery life >100 hours (5000 readings) Operating/Storage Temperature 0 to 55 ºC (32 to 132 ºF) Protection Ratings IP67/NEMA 6
Warranty
Ranges
pH (MP-6 and MP-6p models) 0 to 14 pH ORP (MP-6 and MP-6p models) ±999 mV
Conductivity
TDS
Mineral/Salt (MP-6p model only)
Resistivity (MP-6 and MP-6p models) 10 K to 30 M Temperature 0 to 71 ºC (32 to 160 ºF)
®1
MP Series Meter Warranty: Two years from date of shipment (see Section 9 on page 47)
pH/ORP Sensor Warranty: Six months from date of shipment (see Section 9 on page 47)
0 to 9999 µS 10 to 200 mS/cm in 5 autoranges
0 to 9999 ppm 10 to 200 ppt in 5 autoranges
0 to 9999 ppm 10 to 200 ppt in 5 autoranges
/cm
7
Specifications
Resolution
pH ±0.01 pH ORP ±1 mV
0.01 (<100 µ
0.1 (<1000 µ
Conductivity
TDS
Mineral/Salt
Resistivity
Temperature 0.1 ºC/ºF
1.0 (<10 mS)
0.01 (<100 mS)
0.1 (<200 mS)
0.01 (<100 ppm)
0.1 (<1000 ppm)
1.0 (<10 ppt)
0.01 (<100 ppt)
0.1 (<200 ppt)
0.01 (<100 ppm)
0.1 (<1000 ppm)
1.0 (<10 ppt)
0.01 (<100 ppt)
0.1 (<200 ppt)
0.01 (<100 KΩ)
0.1 (<1000 KΩ)
0.1 (>1 MΩ)
Accuracy
pH ±0.01 pH ORP ±1 mV Conductivity ±1% of reading TDS ±1% of reading Mineral/salt ±1% of reading Resistivity ±1% of reading Temperature ±0.1 ºC
Auto temperature compensation
pH 0 to 71 ºC (32 to 160 ºF) Conductivity 0 to 71 ºC (32 to 160 ºF) TDS 0 to 71 ºC (32 to 160 ºF) Mineral/Salt 0 to 71 ºC (32 to 160 ºF) Resistivity 0 to 71 ºC (32 to 160 ºF)
Adjustable temperature compensation
Conductivity 0 to 9.99%/ºC TDS 0 to 9.99%/ºC Mineral/Salt 0 to 9.99%/ºC Resistivity 0 to 9.99%/ºC
S) S)
2
8
Specifications
COND/TDS ratios pre-programmed
Conductivity TDS Mineral/salt
Adjustable COND/TDS ratio factor
Conductivity
Mineral/salt
1
Trademark of SABIC Innovative Plastics IP BV
2
± 0.2 pH in presence of RF fields 3 V/m and >300 MHz.
3
Trademark of Myron L Company
KCl, NaCl, 442™
0.20 to 7.99TDS
3
9
Specifications
10

Section 2 General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or
consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.

2.1 Safety information

Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger, warning and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this manual.

2.1.1 Use of hazard information

DANGER
Indicates a potentially or imminently hazardous situation which, if not avoided, will
result in death or serious injury.
WA RN IN G
Indicates a potentially or imminently hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in
minor or moderate injury.
NOTICE
Indicates a situation that is not related to personal injury.

2.1.2 Precautionary labels

Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
11
General information

2.2 General product information

The MP-4, MP-6 and MP-6p (see Figure 2 on page 14) handheld meters allow users to test water for pH, ORP, conductivity, resistivity, TDS (Total Dissolved Solids), mineral/salt concentration and temperature.

2.2.1 Overview

The MP Series portable meters measure various parameters in water. Data can be stored and (with the optional MP-Dock) transferred to a printer, PC, or USB storage device.
•MP-4—Measures conductivity, resistivity, TDS and temperature
•MP-6p—Measures pH, ORP, conductivity, mineral/salt concentration, TDS and
temperature. The mineral/salt measurement is a TDS value based on an NaCl profile.
•MP-6—Measures pH, ORP, conductivity, resistivity, TDS and temperature

2.2.2 Features common to all models

4-digit LCD User-adjustable conductivity/TDS
IP67 rating Accuracy of ±1% of reading or
Internal electrode sensors for
maximum protection
Time and date-stamped data logging Memory stores 100 readings
Automatic temperature
compensation
Download capability with optional
MP-Dock
conversion ratio
better
Autorange
conductivity/TDS/resistivity
Factory-stored calibrations
Adjustable auto shut-off

2.2.3 User mode features

Adjustable conductivity/TDS conversion
factor
Programmable temperature
compensation factor
12
General information

2.3 Conductivity and pH/ORP sensor cups

Figure 1 Model MP-6 conductivity and pH/ORP sensor cups
1 Temperature sensor 5 pH glass electrode 2 Conductivity cup (built-in
electrodes)
3 ORP electrode 7 pH/ORP sensor protective cap 4 pH/ORP sensor cup (replaceable
sensor)
6 Reference junction under glass pH
bulb
13
General information
Figure 2 Model MP-6
1 Conductivity cup 4 Display 2 pH/ORP sensor cup 5 Keypad 3 pH/ORP sensor protective cap 6 Wrist strap slot (customer supplied)
14

Section 3 Operation

3.1 System startup

There is no ON key or OFF key. Press any measurement key to power on the meter. After 15 seconds of inactivity, the meter turns off (60 seconds in CAL mode). Users can adjust the automatic shut off time up to 75 seconds (see section 3.19 on
page 26).

3.2 Display description

The meter display shows the temperature, units, parameter, test values, user mode, memory recall, memory store, calibration, date and time (Figure 3).
Figure 3 Model MP-6 display
1 Test value—Shows the test value. 2 Units of measurement—Shows the units of measurements. 3 Parameters—Shows the parameters being measured. 4 Multiple value readout—Shows the temperature value readout, user
temperature compensation or conductivity/TDS ratio. Memory record location numbers or pH calibration. Also shows same date readout as the time and date indicator.
5 Time and date—Shows the time and date. 6 Selected solution—Shows the solution profile that is selected.
15
Operation

3.3 Keypad description

The MP-6 meter is used as an example for keypad description and function.
16
Figure 4 Keypads on MP Series meters
Operation
1COND
Turns on the meter, measures conductivity, and exits any function
2RES1 (MP-4 and MP- 6 only)—Turns on the meter, measures resistivity, and exits
any function
3TDS—Turns on the meter, measures TDS, and exits any function 4ORP (MP-6 and MP-6p only)—
function
Turns on the meter, measures pH, and exits any
5UP/MSScrolls up and stores value to memory 6 MR/DOWNScrolls down and recalls stored memory information 7 CAL/CMC LREnters the calibration mode, clears the memory, and provides
confirmation
1
The MP-6p meter has a MIN/SALT key instead of the RES key. The mineral/salt
measurement is a TDS value based on an NaCl profile.

3.4 Take a measurement

To take a measurement:
1. Rinse the sensor cup with test solution three times and refill.
Note: If testing solutions that are highly concentrated or at extreme temperatures, more rinsing is required.
2. Push the desired measurement key.
Note: To prevent auto shut off, push the measurement key again and as needed.
3. Observe or record the value displayed, or push UP/MS to store the reading.

3.5 Measure conductivity

To measure conductivity:
1. Rinse the conductivity cup three times with the sample to be measured. This
conditions the temperature compensation sensor and prepares the cell.
2. Fill the conductivity cup with the solution.
3. Push the COND key.
4. Observe or record the value displayed, or push UP/MS to store the reading. A
display of [----] indicates an over range condition.
Note: Carefully fill conductivity cup to ensure that the air bubbles do not cling to the cell wall.

3.6 Measure resistivity (MP-4 and MP-6 models)

Resistivity is measured in low conductivity solutions. In the conductivity cup, the value can drift due to trace contaminants or absorption from atmospheric gasses. Therefore, measuring a flowing sample is recommended.
1. Make sure the pH/ORP sensor protective cap is secure to avoid contamination
(MP-6 model).
17
Operation
2. Hold the meter at a 30 degree angle and let the sample flow into the
conductivity cup continuously with no aeration.
3. Push the RES key.
4. Observe or record the value displayed.
Note: If reading is lower than 10 k, [- - - -] is shown. Measure conductivity for these samples.

3.7 Measure mineral/salt (MP-6p model only)

To measure mineral/salt:
1. Rinse the conductivity cup three times with the sample to be measured. This
conditions the temperature compensation sensor and prepares the cell.
2. Fill the conductivity cup with the solution.
3. Push the MIN/SALT key.
4. Observe or record the value displayed, or push UP/MS to store the reading.

3.8 Measure TDS

To measure TDS:
1. Rinse the conductivity cup three times with the sample to be measured. This
conditions the temperature compensation sensor and prepares the cell.
2. Fill the conductivity cup with the solution.
3. Push the TDS key.
4. Observe or record the value displayed, or push UP/MS to store the reading.

3.9 Measure ORP/Redox (MP-6 and MP-6p models)

To measure the ORP/Redox:
1. Remove the protective pH/ORP sensor cap. Squeeze the sides and pull up.
2. Rinse the sensor cup three times with the sample to be measured.
3. Shake the meter after each rinse to remove residual liquid.
4. Fill both sensor cups with the sample.
5. Push the ORP key.
6. Observe or record the value displayed, or push UP/MS to store the reading.
Important Note: After the test, fill the pH/ORP sensor cup with pH Storage Solution
and replace the protective cap. Do not allow the pH/ORP sensor cup to dry out.

3.10 Measure pH (MP-6 and MP-6p models)

To measure pH:
1. Remove the protective pH/ORP sensor cap. Squeeze the sides and pull up.
2. Rinse the pH/ORP sensor cup three times with the sample to be measured.
3. Shake the meter after each rinse to remove residual liquid.
4. Fill both sensor cups with the sample.
18
Operation
5. Push the pH key.
6. Observe or record the value displayed, or push UP/MS to store the reading.
Important Note: After the test, fill the pH/ORP sensor cup with pH Storage Solution
and replace the protective cap. Do not allow the pH/ORP sensor cup to dry out.

3.11 Select a solution

Conductivity, resistivity and TDS (including mineral/salt) require temperature compensation to 25 °C. The solution profile selection determines the temperature compensation of conductivity and calculation of TDS and mineral/salt from compensated conductivity.
There are four solution types:
KCl
NaCl
442
User
On the left side of the display is the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion. By default, KCI is used for conductivity, NaCl is used for resistivity (and mineral/salt), and 442 (natural water characteristic) is used for TDS. The User selection allows a custom value to be entered for the temperature compensation of conductivity and the conversion ratio if measuring TDS.
Check the display to see if the solution profile displayed is the solution type desired for that measurement. To change a solution:
1. Push the COND key, the RES key, the MIN/SALT key or the TDS key to select
the parameter to change the solution type.
2. Push and hold the CAL/MCLR key for three seconds and wait for SEL to
appear in the display.
3. Push UP/MS or MR/DOWN to scroll to the desired solution type.
4. Push the CAL/MCLR key to accept the new solution.

3.11.1Temperature compensation

Electrical conductivity indicates solution concentration and ionization of the dissolved material. Because temperature affects ionization, conductivity measurements change with temperature and must be corrected to read at 25 ºC.
Temperature compensation uses the characteristics of salt solutions. The selected salt solution is displayed on the left side of the display. By default, the meter uses KCl for conductivity, NaCl for resistivity and 442 for TDS (see Appendix B on page
51).
The User mode customizes the temperature compensation and the conversion ratio if measuring TDS.
Note: Calibration of each solution type is performed separately and calibration of one solution does not affect the calibration of the other solution types.
19
Operation

3.12 Change the user-selected temperature compensation factor

Select the User mode to change the temperature compensation factor. This feature does not apply to pH or ORP. For user mode information, (see section 2.2.3 on
page 12).
1. Select the User mode (see section 3.11 on page 19).
2. Push the CAL/MCLR key.
3. Push UP/MS or MR/DOWN to adjust the temperature compensation factor
from 0-9.99%/ºC.
4. Push the CAL/MCLR key twice to skip calibration adjustment and accept the
new temperature compensation (three times if in TDS or MIN/SALT mode).
5. Measure samples with the new temperature compensation factor.

3.12.1 Disable temperature compensation

1. Select the User mode (see section 3.11 on page 19).
2. Push the CAL/MCLR key. Hold the MR/DOWN key until the temperature
compensation shows .00%/
ºC.
3. Push the CAL/MCLR key twice (three times for TDS or MIN/SALT).
4. Temperature compensation is now disabled (=0) for measurements in the User
mode.

3.13 Change the user-selected conductivity/TDS ratio

Select the User mode to change a custom conductivity/TDS conversion ratio in the range of 0.20 to 7.99.
To determine the conversion ratio for a custom solution of a known TDS ppm value, measure the solution conductivity at 25 ºC with the MP Series meter and divide the
20
Operation
ppm value by the µS value. For example, a solution of known 75 ppm TDS and measured
To enter a new conversion ratio:
1. Push the TDS key.
2. Select the User mode (see section 3.11 on page 19).
3. Push the CAL/MCLR key twice (to skip over temperature compensation
4. Push UP/MS or MR/DOWN until the new conversion ratio is displayed.
5. Push the CAL/MCLR key twice (to skip over calibration adjustment) to accept
6. Use the new conductivity/TDS ratio to measure samples.
100µS conductivity at 25ºC has a conversion ratio of 75/100 or 0.75.
adjustment) and the ratio will appear.
the new conversion ratio.

3.14 Settings

3.14.1 Store a value in the memory

The MP series handheld meters have memory storage for up to 100 readings. Time and date is recorded with each stored reading.
To download this stored data to a computer, see section 3.21 on page 29.
1. Push the UP/MS key to record a value.
2. The MEMORY icon appears and the temperature display is briefly replaced by
a number (1-100) that shows the position of the record. Figure 5 shows a reading of 1806 µS stored in memory record # 4.
Figure 5
21
Operation

3.14.2 View the memory recall

To view the records in memory:
1. Push any measurement key.
2. Push the MR/DOWN key. The MEMORY icon appears, and shows the last
record stored.
3. Push UP/MS or MR/DOWN to scroll to the location desired.
Note: The temperature display alternates between temperature recorded and location number.
4. Push the CAL/MCLR key to show the time and date stamp.
5. Push any measurement key to leave memory recall.
3.14.2.1 Clear a single record
After the user recalls a specific record location, push and hold the CAL/MCLR key to clear that memory location. This memory location is used for the next stored memory record unless the user scrolls to another empty memory position before the recall sequence ends.

3.14.3 Clear all records

To clear all records in the memory:
1. Push the MR/DOWN key.
2. Scroll down until CLR ALL is displayed.
3. Push the CAL/MCLR key. This clears all records.

3.15 Time and date

Change the time and date for travel or for a battery replacement that takes longer than three minutes.

3.15.1 Set the time

Time is displayed in a 24-hour format.
1. Push the any measurement key.
2. Push the MR/DOWN key repeatedly until the time is displayed. To quickly scroll
through all stored memory records, hold down the MR/DOWN key.
22
Operation
3. Push the CAL/MCLR key to begin. The CAL icon shows the time.
4. Push UP/MS or MR/DOWN to change the time.
5. Push the CAL/MCLR key to accept the new time.

3.15.2 Set the date

To change the date format, refer to section 3.15.3 on page 24. The default format for the date is US (mo/dy/yr).
1. Push any measurement key. To quickly scroll through all stored memory records, hold down the MR/DOWN key.
2. Push the MR/DOWN key repeatedly until the date shows in the display. For example: 01.05/05 (January 5, 2005)
3. Push the CAL/MCLR key to begin. The CAL icon displays above the year.
4. Push UP/MS or MR/DOWN to change the year.
5. Push CAL/MCLR to accept the new setting for the year.
6. Push UP/MS or MR/DOWN to change the month.
7. Push CAL/MCLR to accept the new setting for the month.
23
Operation
8. Push UP/MS or MR/DOWN to change the day.
9. Push CAL/MCLR to accept the new setting for the day.

3.15.3 Set the date format

To set the date format:
1. Push any measurement key.
2. Push the MR/DOWN key repeatedly until US or Int displays. To quickly scroll through all stored memory records, hold down the MR/DOWN key.
3. Push CAL/MCLR to change the date format. The new format is now displayed.

3.16 Temperature format

To set the temperature format:
1. Push any measurement key.
2. Push the MR/DOWN key repeatedly until C or F is displayed. To quickly scroll through all stored memory records, hold down the MR/DOWN key.
24
Operation
3. Push the CAL/MCLR key to switch units.
4. Push any measurement key to accept the unit preference for all temperature
readings.
Note: Temperature compensation is always displayed in %/ ºC.

3.17 Return to factory settings

To set all calibrations to factory settings or to erase all records, follow the steps below.
1. Push any measurement key.
2. Push the MR/DOWN key repeatedly until FAC SEL is displayed. To quickly scroll through all stored memory records, hold down MR/DOWN.
3. Push the CAL/MCLR key to accept the factory reset. The meter returns to the
measurement mode.

3.18 Cell check

The cell check verifies the cleanliness of the conductivity/TDS/resistivity sensor. If the display shows .00 when the cell cup is dry, the sensor is probably clean.
In normal use, the conductivity cell may become dirty or coated and require cleaning. To perform a cell check:
1. Push the COND key.
2. Push the MR/DOWN key repeatedly until the display shows CELL ch. To quickly scroll through all stored memory records, hold down MR/DOWN.
3. Push the CAL/MCLR key to test. If the cell is clean, “Good” displays briefly. If
cell is dirty, “Cell cLn” displays. To clean the sensors, (see section 5.4 on
page 38).
25
Operation

3.19 Auto off

Auto off turns the meter off when there is no activity for a period of time after a key is pushed. The default time is 15 seconds, and 60 seconds in the CAL (calibration) mode. This time may be adjusted up to 75 seconds.
1. Push any measurement key.
2. Push the MR/DOWN key repeatedly until the display shows Auto oFF. To quickly scroll through all stored memory records, hold down MR/DOWN.
3. Push the CAL/MCLR key to begin. The CAL icon displays above the 15 SEC
display.
26
Operation
4. Push UP/MS or MR/DOWN to change the time. The maximum time is 75
seconds.
5. Push the CAL/MCLR key to accept the new auto shut off time.
3.20 User mode calibration Linc
The Linc the user does not have a user standard solution to calibrate the meter. This ensures more accurate measurements. When the Linc function is used, the User mode is linked to another standard solution. For example: If User and KCl are linked, a KCl standard solution is used to calibrate the instrument.
Note: When a “Linc” is established for the User mode, the Linc applies to all measurement modes using the User solution selection.
TM
function allows for calibration when the meter is in the User mode and
TM1
function

3.20.1 Calibrate meter for User mode

To calibrate the meter for the User mode:
1. Push the COND key, the MIN/SALT key, or the TDS key.
2. Calibrate the meter using a standard solution (see section 4.4 on page 31).
3. Select the User mode (see section 3.11 on page 19).
4. Set the calibration Linc.

3.20.2 Set User mode calibration Linc

The Linc function sets the calibration offset factor of a standard solution to the User solution mode. The Linc stays intact in future calibrations until it is canceled (see
section 3.20.3 on page 28).
Follow the steps below to set the KCl, NaCl or 442 calibration factor to the User solution mode.
1. Push a measurement key to Linc (i.e., COND, RES, MIN/SALT or TDS).
2. Select the User mode (see section 3.11 on page 19).
1
Trademark of Myron L Company
27
Operation
3. Push the MR/DOWN key until Linc displays.
4. Push the CAL/MCLR key. SEL displays with the User icon.
Note: Any additional display of KCl, NaCl or 442 icons indicates a Linc between the additional solution and the User solution. If none of the solution selection icons are displayed, nothing is linked to the User mode.
5. Push UP/MS or MR/DOWN to select a standard solution to link to the User mode calibration constant.
6. Push the CAL/MCLR key to accept the setting. The User mode now uses the calibration offset constant that was created here.
Note: To exit without changing the setting, push any measurement key.

3.20.3 Cancel User mode calibration Linc

To cancel the User mode calibration Linc:
Note: The MP series meter must be in User linked mode to cancel the “Linc.”
1. Push a (linked) measurement key such as COND, RES, MIN/SALT, or TDS. Two solutions are displayed on the left side of the display: User and another, such as KCl.
2. Push the MR/DOWN key until Linc displays.
3. Push the CAL/MCLR key. SEL, User and the linked solution appear on the
display.
28
Operation
4. Push the MR/DOWN key until User is the only solution icon that displays.
5. Push the CAL/MCLR key. The User mode Linc is now canceled.

3.21 Download stored data

The MP-Dock accessory package (HMPDOCK) allows the user to download stored test data to a PC or spreadsheet. The MP-Dock receives power through the USB port, and requires no external power source. The data is transferred through the Infrared (IR) data port on the bottom of the MP meter (Figure 6) to the MP-Dock, and then to the PC.
The MP Datalink software, which is included with the MP-Dock, operates on Windows 2000 and XP, and Macintosh OS9.2 and OSX-based operating systems.
For the latest instructions on communication port selection and data download, refer to the Hach Company MP-Dock User Manual.
1 Infrared data port
Figure 6 MP meter—bottom view
29
Operation
30

Section 4 Calibration

4.1 Calibration intervals

The MP Series meters are designed to not require frequent calibration. Calibration is recommended about once per month with conductivity or TDS solutions. Check the calibration with pH solution twice per month. Some applications may require calibration frequencies to differ from these suggested guidelines.

4.2 Calibration limits

The MP Series meters have built in calibration limits. A nominal “FAC” value is an ideal value stored by the factory. Attempts to calibrate too far (±10% or ±1 pH unit) from this value causes the displayed value to be replaced with “FAC”. If the CAL/MCLR key is pushed, the value is accepted, and the original default factory calibration for this measurement is shown. The need to calibrate so far out that “FAC” appears indicates a procedural problem, incorrect standard solution, a very dirty cell cup or an aging pH/ORP sensor.

4.3 Calibration records

To minimize calibration efforts, keep records. If the calibration adjustments are minimal, calibration can occur less often. Record the following information:
Record changes in conductivity in percentages.
Record changes in pH calibration in pH units.
Conductivity cell calibration is purposely limited to
indicate damage, not drift.
Calibration changes are limited to
end of the sensor’s lifetime and replacement is recommended.
±1 pH unit. Changes beyond that indicate the
±10%. Changes beyond that

4.4 Calibrate the meter

1. Push the measurement key for the parameter to be calibrated.
2. Push CAL/MCLR.
3. Measuring continues. The CAL icon is on. This indicates that calibration can
occur now.
4. Push UP/MS or MR/DOWN to change the reading to the known value.
5. Calibration for each of the four solution types can be performed in either
conductivity, mineral/salt or TDS mode.
Note: The number of steps to calibrate depends on what is to be calibrated.
31
Calibration
Parameter KCl, NaCl or 442 User
COND Gain only Temperature compensation, then Gain RES Done in conductivity Done in conductivity or TDS
TDS Gain only
MIN/SALT Gain only
pH 7, acid, and/or base ORP Zero set with pH 7 automatically
Temperature compensation, Ratio, then Gain
Temperature compensation, Ratio, then Gain
6. Push the CAL/MCLR key to accept the new calibration value. The meter accepts the value and presents the next value for adjustment. If there are no more adjustments, the meter exits CAL mode.
Note: In the CAL mode, the CAL/MCLR key becomes an ACCEPT key. To bypass a calibration step, push the CAL/MCLR key to accept the present value.

4.5 Exit calibration mode

When the CAL icon turns off, calibration is complete. To exit calibration mode when the CAL icon is still on, push any measurement key. This cancels any changes not accepted and exits the CAL mode. When CAL mode for pH is exited after the second buffer, the meter enters the same gain for the third buffer.

4.6 Calibrate conductivity, mineral/salt or TDS

To make sure calibration is accurate, follow the items listed below.
1. Clean oily films or organic material from the conductivity cell with foaming cleaner or mild acid.
2. Do not scrub inside the conductivity cell.
3. Rinse the conductivity cup with pure water after taking measurements.
4. Rinse the conductivity cup three times with the standard solution to be used for
calibration (KCl, NaCl, or 442).
Note: Failure to rinse can cause crystals to form in the cup and contaminate future samples.
5. Fill the conductivity cup with same standard.
6. Push the COND key, the MIN/SALT key or the TDS key.
7. Push the CAL/MCLR key. The CAL icon appears on the display.
8. Push the UP/MS key or the MR/DOWN key to adjust to the standard value, or
hold down the key to adjust rapidly.
9. Push the CAL/MCLR key once to confirm the new value and end the calibration sequence for this solution type.
10. To calibrate another solution type, change solution type (e.g., KCl, NaCl, or
442) and repeat this procedure.
32
Calibration

4.7 Calibrate resistivity

Resistivity is the reciprocal of conductivity. Resistivity is automatically calibrated based on the solution type used during a conductivity calibration.

4.8 Reset factory calibration—conductivity, mineral/salt or TDS

If calibration is suspect or known to be incorrect, and no standard solution is available, it is possible to replace the calibrated value with the original factory value for that solution. This ideal factory (FAC) value is the same for all MP Series meters, and it returns to a known state without solution in the cup.
The FAC internal electronic calibration is not intended to replace calibration with conductivity standard solutions.
1. Push the COND key, the MIN/SALT key or the TDS key.
2. Push the CAL/MCLR key twice in COND or three times in TDS.
Note: In the User mode, push the CAL/MCLR key twice in the COND mode and three times in the TDS mode or the MIN/SALT
and ratio adjustments.)
3. Push the UP/MS key until the FAC icon appears.
4. Push the CAL/MCLR key to accept the factory calibration setting.
5. If another solution needs to be reset, select another solution type and repeat
the procedure.
mode. (This bypasses temperature correction

4.9 pH calibration

Note: Always zero-out the MP Series meter with a pH 7 buffer solution before calibrating with acid or base buffers such as pH 4 or pH 10 solutions.
To perform a pH calibration:
1. Rinse the sensor cups three times with a pH 7 buffer solution.
2. Fill both sensor cups with pH 7 buffer solution.
3. Push the pH key to verify the pH calibration. If the display shows 7.00, skip the
pH zero calibration and proceed to section 4.10 on page 34.
33
Calibration
4. Push the CAL/MCLR key to enter calibration mode. The CAL, BUFFER and 7
icons appear. The value displayed is for the uncalibrated sensor.
Note: If a wrong buffer is added (outside of pH 6-8), 7 and BUFFER will flash and the
meter does not adjust. The uncalibrated pH value that shows in step 4 assists in determining the accuracy of the pH sensor. If the pH reading is below pH 6 or above pH 8 with pH 7 buffer solution, the sensor cup needs more rinses, or the pH sensor is defective and needs to be replaced.
5. Push UP/MS or MR/DOWN until the display reads 7.00.
Note: Attempted calibration of > 1 pH point from factory calibration causes the FAC icon to appear. This means that either sensor replacement (see Section 6 on page 41) or a fresh buffer solution is needed. Push the CAL/MCLR key to accept the preset factory value.
6. Push the CAL/MCLR key to accept the new value. The pH zero calibration is now complete. It is recommended that the user performs
the multiple point pH calibration (see section 4.10). If the user does not wish to continue, push any measurement key to exit.

4.10 Set multiple point pH calibrations

Important Note: Acid or base solution can be used for the second point calibration
and then use the other solution for the third point. To verify that a buffer is in the
sensor cup, the display shows either the Acd icon or the bAS icon.
Note: If the Acd icon or the bAS icon flash, fill the sensor cup with either an acid or base solution to resolve the error.
1. Push the CAL/MCLR key twice while in the pH measurement mode to complete the pH zero calibration or verify the pH 7 buffer. The CAL, BUFFER and Acd or bAS icons are displayed.
2. Rinse the sensor cups three times with acid or base buffer solution.
3. Fill both sensor cups again with the same solution.
4. Push UP/MS or MR/DOWN until the display agrees with the buffer value.
5. Push the CAL/MCLR key to accept the second point of calibration. The display
indicates the next type of buffer to be used. The two point calibration is complete now. The user can continue with the third
point of the calibration or exit the calibration process. Push any measurement
34
Calibration
key to exit. If the user exits, the gain value accepted for the buffer is used for both acid and base measurements.
6. Rinse the sensor cup three times with the third buffer solution.
7. Fill the sensor cups again with the same solution.
8. Push UP/MS or MR/DOWN until the display agrees with the buffer value.
9. Push the CAL/MCLR key to accept the third point of calibration. The calibration
procedure is now complete.
Note: Fill the pH/ORP sensor cup with pH storage solution and replace the protective sensor cap when the meter is not in use. Do not allow the cup to dry.

4.11 ORP calibration

The ORP electrodes rarely give false readings unless there is a problem in the reference electrode. For this reason, and because the calibration solutions for ORP are highly reactive and potentially hazardous, the MP meter has an electronic ORP calibration. This causes the zero point on the reference electrode to be set whenever the pH 7 calibration is done.

4.12 Temperature calibration

Temperature calibration is not necessary in the MP series meters.
35
Calibration
36

Section 5 Maintenance

Practice the following care and maintenance of the MP series handheld meters:
Rinse with clean water after each use
Always fill the pH/ORP sensor cup with Hach pH storage solution and replace
the protective cap when not in use.
Avoid solvents
Avoid drops. Shock damage can damage the meter and void the warranty

5.1 Temperature extremes

Solutions in excess of 71 ºC (160 ºF) should not be placed in the sensor cups. This activity can damage the meter. The pH sensor can fracture if the meter’s temperature falls below 0 ºC (32 ºF). Take care not to exceed operating temperatures.
Note: Do not leave an MP series meter in a vehicle or a storage shed on a hot day. This activity can subject the meter to temperatures in excess of 66 ºC (150 ºF) and void the warranty.

5.2 Battery replacement

NOTICE
If the meter is not completely dry before you open the meter, damage to the internal electronics of the meter can occur.
Perform the following steps to replace the battery:
1. Dry the meter completely.
2. Remove the four screws from the base of the meter.
3. Open the meter carefully.
4. Take care to detach the battery from the circuit board.
5. Replace the battery with a new 9V alkaline battery.
6. Replace the bottom housing, ensuring that the sealing gasket is installed in the
groove of the top half of the case.
7. Replace the screws; tighten evenly and securely. Do not overtighten.
Note: All data stored in memory and all calibration settings are protected during power loss or battery replacement. Loss of time and date can occur, however, if the battery is removed for more than 3 minutes (180 seconds).
37
Maintenance

5.2.1 Maintain the conductivity cup

Rinse the conductivity cup with clean water after taking measurements to prevent buildup on the electrodes. Do not scrub the cup. For oily films, add a few drops of foaming, non-abrasive cleaner or isopropyl alcohol, then rinse.
Note: When sampling low-conductivity solutions, make sure the pH/ORP sensor cap is well-seated so that solution does not wash from the pH/ORP sensor cup into the conductivity cup.

5.2.2 Maintain the pH/ORP sensor cup

Keep the pH/ORP sensor cup hydrated with Hach pH Storage Solution. Before replacing the pH/ORP sensor cap, rinse and fill the sensor cup with the storage solution. Never use distilled water to store the sensor cup.

5.3 pH/ORP sensor replacement

Complete installation instructions are included with each replacement sensor. Tools required include a #2 Phillips screwdriver and 1/4-inch wrench.
Note: When the pH/ORP sensor is replaced, it is also a good time to replace the battery.

5.4 Clean the sensors

Perform these procedures to clean the various sensors.

5.4.1 Clean the conductivity/resistivity/TDS sensor

Keep the conductivity cell cup (Figure 7) as clean as possible.
Note: Flush with clean water after use to prevent buildup on electrodes.
When a dirty sample is left in the cup, a film forms. This film reduces accuracy. To clean a visible film of oil, dirt or scale that is in the cell cup or on the electrode:
1. Use isopropyl alcohol or a foaming, non-abrasive household cleaner. Hach acid electrode cleaning solution may also be used less frequently.
2. Pour any of these solutions in the cell cup and allow it to soak for no more than five minutes.
3. Use a cotton swab to gently clean the electrodes.
4. Rinse out the cleaning solution.
38
Maintenance
Figure 7 Model MP-6 sensor cups
1 Temperature sensor 5 pH glass electrode 2 Conductivity cell (built-in electrodes) 6 Reference junction under glass pH
bulb
3 ORP electrode 7 pH/ORP sensor protective cap 4 pH/ORP sensor (replaceable)

5.4.2 Clean the pH/ORP sensor

The pH/ORP sensor in the MP series meters is non-refillable and features a porous liquid junction. It should not be allowed to dry out. If it does dry out, the sensor can sometimes be restored by following the steps below.
1. Clean the sensor cup with isopropyl alcohol.
2. Rinse well. Do not scrub or wipe the pH/ORP sensor.
3. Follow the hot solution method described below: a. Pour a hot salt solution ~60 ºC (140 ºF), such as pH storage solution in the
sensor cup.
b. Allow the liquid to cool. c. Retest.
4. If the hot solution method does not work, follow the Deionized (DI) water
method below:
a. Pour DI water into the sensor cup.
39
Maintenance
b. Allow to stand for no more than four hours (more standing time can deplete
the reference solution and damage the glass bulb).
c. Retest.
5. If neither of the above methods are successful, the sensor must be replaced.
5.4.2.1 Drifting test results
A film on the pH sensor bulb or the reference can cause drifting. Use isopropyl alcohol to clean the glass bulb.
Note: The sensor bulb is very thin and delicate. Do not scrub the pH/ORP sensor.
To clean the sensor:
1. Use isopropyl alcohol or a foaming, non-abrasive household cleaner. Hach
acid electrode cleaning solution may also be used less frequently.
2. Pour any of these solutions in the cell cup and allow it to soak for no more than
five minutes.
3. Use a cotton swab to gently clean the electrodes.
4. Rinse out the cleaning solution.
5. Fill the sensor cup with Hach pH storage solution before the pH/ORP sensor
cap is replaced.
5.4.2.2 Solutions that damage the pH/ORP sensor
Samples that contain chlorine, sulfur or ammonia can damage the pH electrode. Rinse the sensor thoroughly with clean water immediately after any measurement of these liquids.
Samples that reduce (add an electron to) silver, such as cyanide will attack the reference electrode.
Leaving alkaline solutions in the pH sensor cup for long periods of time can damage the sensor.
40

Section 6 Troubleshooting

Symptom Possible cause Action
No display even though measurement key is pushed
Inaccurate pH reading
No response to pH changes (MP-6 and MP-6p models)
Meter does not adjust down to pH 7 (MP-6 and MP-6p models)
pH readings drift or respond slowly to change or FAC displays repeatedly
Unstable conductivity, TDS or resistivity readings
Meter cannot calibrate Conductivity or TDS
Resistivity reading is much lower than expected
Battery is weak or not connected.
pH calibration is needed (see section 4.9 on page 33)
Cross-contamination from residual buffers or samples in sensor cup
Calibration with expired pH buffers
Sensor bulb is cracked or an electromechanical short is caused by an internal crack.
pH sensor has lost KCl
Temporary condition due to memory of solution in pH sensor cup for long periods
Bulb dirty or dried out Reference junction clogged
or coated Dirty electrodes Clean the cell cup and the
Test samples that are greater than 1 M
Film or deposits on electrodes
Contamination from previous samples or from the pH sensor cup
Carbon dioxide in the test sample
Check the connections or replace the battery (see
section 5.2 on page 37).
Recalibrate the meter.
Rinse the sensor cup.
Recalibrate using fresh buffers.
Replace the pH/ORP sensor (see section 5.3 on page 38).
Clean and restore the sensor (see section 5.4 on page 38) and recalibrate. If there is no improvement, replace the pH/ORP sensor (see
section 5.3 on page 38).
Clean and restore the sensor (see section 5.4 on page 38) and recalibrate. If there is no improvement, replace the pH/ORP sensor (refer to
section 5.3 on page 38).
electrodes (see section 5.4 on
page 38).
Minimize the test sample exposure to air (see
section 3.6 on page 17).
Clean the cell cup and the electrodes (see section 5.4 on
page 38
).
Rinse the sensor cup more thoroughly before measurement.
Make sure the pH cap is snugly in place (see section 5.4 on page 38).
41
Troubleshooting
42

Section 7 Contact Information

HACH Company World Headquarters
P.O. Box 389 Loveland, Colorado 80539-0389 U.S.A. Tel (800) 227-HACH (800) 227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com
Repair Service in Latin America, the Caribbean, the Far East, Indian Subcontinent, Africa, Europe, or the Middle East:
Hach Company World Headquarters, P.O. Box 389 Loveland, Colorado, 80539-0389 U.S.A. Tel +001 (970) 669-3050 Fax +001 (970) 669-2932 intl@hach.com
HACH LANGE LTD
Unit 1, Chestnut Road Western Industrial Estate IRL-Dublin 12 Tel. +353(0)1 46 02 5 22 Fax +353(0)1 4 50 93 37 info@hach-lange.ie www.hach-lange.ie
HACH LANGE FRANCE S.A.S.
33, Rue du Ballon F-93165 Noisy Le Grand Tél. +33 (0)1 48 15 68 70 Fax +33 (0)1 48 15 80 00 info@hach-lange.fr www.hach-lange.fr
HACH LANGE APS
Åkandevej 21 DK-2700 Brønshøj Tel. +45 36 77 29 11 Fax +45 36 77 49 11 info@hach-lange.dk www.hach-lange.dk
Repair Service in the United States:
HACH Company Ames Service 100 Dayton Avenue Ames, Iowa 50010 Tel (800) 227-4224 (U.S.A. only) Fax (515) 232-3835
HACH LANGE GMBH
Willstätterstraße 11 D-40549 Düsseldorf Tel. +49 (0)2 11 52 88-320 Fax +49 (0)2 11 52 88-210 info@hach-lange.de www.hach-lange.de
HACH LANGE GMBH
Hütteldorferstr. 299/Top 6 A-1140 Wien Tel. +43 (0)1 9 12 16 92 Fax +43 (0)1 9 12 16 92-99 info@hach-lange.at www.hach-lange.at
HACH LANGE SA
Motstraat 54 B-2800 Mechelen Tél. +32 (0)15 42 35 00 Fax +32 (0)15 41 61 20 info@hach-lange.be www.hach-lange.be
HACH LANGE AB
Vinthundsvägen 159A SE-128 62 Sköndal Tel. +46 (0)8 7 98 05 00 Fax +46 (0)8 7 98 05 30 info@hach-lange.se www.hach-lange.se
Repair Service in Canada:
Hach Sales & Service Canada Ltd. 1313 Border Street, Unit 34 Winnipeg, Manitoba R3H 0X4 Tel (800) 665-7635 (Canada only) Tel (204) 632-5598 Fax (204) 694-5134 canada@hach.com
HACH LANGE LTD
Pacific Way Salford GB-Manchester, M50 1DL Tel. +44 (0)161 872 14 87 Fax +44 (0)161 848 73 24 info@hach-lange.co.uk www.hach-lange.co.uk
DR. BRUNO LANGE AG
Juchstrasse 1 CH-8604 Hegnau Tel. +41(0)44 9 45 66 10 Fax +41(0)44 9 45 66 76 info@hach-lange.ch www.hach-lange.ch
DR. LANGE NEDERLAND B.V.
Laan van Westroijen 2a NL-4003 AZ Tiel Tel. +31(0)344 63 11 30 Fax +31(0)344 63 11 50 info@hach-lange.nl www.hach-lange.nl
HACH LANGE S.R.L.
Via Riccione, 14 I-20156 Milano Tel. +39 02 39 23 14-1 Fax +39 02 39 23 14-39 info@hach-lange.it www.hach-lange.it
43
Contact Information
HACH LANGE S.L.U.
Edif. Arteaga Centrum C/Larrauri, 1C- 2ª Pl. E-48160 Derio/Vizcaya Tel. +34 94 657 33 88 Fax +34 94 657 33 97 info@hach-lange.es www.hach-lange.es
HACH LANGE S.R.O.
Lešanská 2a/1176 CZ-141 00 Praha 4 Tel. +420 272 12 45 45 Fax +420 272 12 45 46 info@hach-lange.cz www.hach-lange.cz
HACH LANGE
8, Kr. Sarafov str. BG-1164 Sofia Tel. +359 (0)2 963 44 54 Fax +359 (0)2 866 04 47 info@hach-lange.bg www.hach-lange.bg
ÇÁCH LANGE E.Ð.Å.
БхлЯдпт 27 GR-115 27 БиЮнб Фзл. +30 210 7777038 Fax +30 210 7777976 info@hach-lange.gr www.hach-lange.gr
HACH LANGE LDA
Av. do Forte nº8 Fracção M P-2790-072 Carnaxide Tel. +351 214 253 420 Fax +351 214 253 429 info@hach-lange.pt www.hach-lange.pt
HACH LANGE KFT.
Hegyalja út 7-13. H-1016 Budapest Tel. +36 (06)1 225 7783 Fax +36 (06)1 225 7784 info@hach-lange.hu www.hach-lange.hu
HACH LANGE SU ANALÝZ SÝSTEMLERÝ LTD.ÞTÝ.
Hilal Mah. 75. Sokak Arman Plaza No: 9/A TR-06550 Çankaya/ANKARA Tel. +90 (0)312 440 98 98 Fax +90 (0)312 442 11 01 bilgi@hach-lange.com.tr www.hach-lange.com.tr
HACH LANGE E.P.E.
27, Avlidos str GR-115 27 Athens Tel. +30 210 7777038 Fax +30 210 7777976 info@hach-lange.gr www.hach-lange.gr
HACH LANGE SP.ZO.O.
ul. Opolska 143 a PL-52-013 Wroc³aw Tel. +48 (0)71 342 10-83 Fax +48 (0)71 342 10-79 info@hach-lange.pl www.hach-lange.pl
HACH LANGE S.R.L.
Str. Leonida, nr. 13 Sector 2 RO-020555 Bucuresti Tel. +40 (0) 21 201 92 43 Fax +40 (0) 21 201 92 43 info@hach-lange.ro www.hach-lange.ro
HACH LANGE D.O.O.
Fajfarjeva 15 SI-1230 Domžale Tel. +386 (0)59 051 000 Fax +386 (0)59 051 010 info@hach-lange.si www.hach-lange.si
44
Section 8 Replacement parts and
accessories

8.1 Replacement parts

Description Item No.
pH/ORP sensor HMPSENS 9V alkaline battery 00024Q

8.2 Accessories

Description Item No.
MP-Dock (facilitates data download to PC or spreadsheet) HMPDOCK

8.3 Consumables

Description Quantity Item No.
Buffer solution, pH 4.01 50 mL 2283426 Buffer solution, pH 4.01 500 mL 2283449 Buffer solution, pH 4.01 4L 2283456 Buffer solution, pH 4.01 20 L 2283461 Buffer solution, pH 7.00 50 mL 2283526 Buffer solution, pH 7.00 500 mL 2283549 Buffer solution, pH 7.00 4 L 2283556 Buffer solution, pH 7.00 20 L 2283561 Buffer solution, pH 10.01 50 mL 2283626 Buffer solution, pH 10.01 500 mL 2283649 Buffer solution, pH 10.01 4 L 2283656 Buffer solution, pH 10.01 20 L 2283661 pH electrode storage solution, 500 mL 500 mL 2756549 pH electrode storage solution, 50 mL 50 mL 2756526
0.001M KCl Conductivity Standard Solution, 148 µS/cm 500 mL 2974249
0.001M KCl Conductivity Standard Solution, 148 µS/cm 50 mL 2974226
0.01M KCl Conductivity Standard Solution, 1413 µS/cm 500 mL 2974349
0.01M KCl Conductivity Standard Solution, 1413 µS/cm 50 mL 2974326
0.1M KCl Conductivity Standard Solution, 12.88 mS/cm 500 mL 2974449
0.1M KCl Conductivity Standard Solution, 12.88 mS/cm 50 mL 2974426
45
Replacement parts and accessories
8.3 Consumables (continued)
Description Quantity Item No.
442-30 Natural Water™1 TDS Standard Solution, 30 ppm
442-30 Natural Water TDS Standard Solution, 30 ppm 50 mL 2974526 442-300 Natural Water TDS Standard Solution,
300 ppm 442-300 Natural Water TDS Standard Solution,
300 ppm 442-1000 Natural Water TDS Standard Solution,
1000 ppm 442-1000 Natural Water TDS Standard Solution,
1000 ppm 442-3000 Natural Water TDS Standard Solution,
3000 ppm 442-3000 Natural Water TDS Standard Solution,
3000 ppm 100 µS/cm NaCl Conductivity Standard Solution 500 mL 2971849 100 µS/cm NaCl Conductivity Standard Solution 50 mL 2971826 1000 µS/cm NaCl Conductivity Standard Solution 500 mL 1440049 1000 µS/cm NaCl Conductivity Standard Solution 50 mL 1440026 10,000 µS/cm NaCl Conductivity Standard Solution 500 mL 2972249 10,000 µS/cm NaCl Conductivity Standard Solution 50 mL 2972226
18.00 mS/cm NaCl Conductivity Standard Solution 500 mL 2307449
18.00 mS/cm NaCl Conductivity Standard Solution 50 mL 2307426
1
Trademark of Myron L Company
500 mL 2974549
500 mL 2974649
50 mL 2974626
500 mL 2974749
50 mL 2974726
500 mL 2974849
50 mL 2974826

8.4 Recommended cleaning consumables

Description Quantity Item No.
Isopropyl alcohol 100 mL 1227642 Isopropyl alcohol prep pads 200/pk 2938200 Cotton swabs 100/pkg 2554300 Acid electrode cleaning solution 50 mL 2975126 Acid electrode cleaning solution 500 mL 2975149
46

Section 9 Limited Warranty

Hach Company warrants to the original purchaser against any defects that are due to faulty material or workmanship for a period of two years from date of shipment unless otherwise noted in the product manual.
In the event that a defect is discovered during the warranty period, Hach Company agrees that, at its option, it will repair or replace the defective product or refund the purchase price excluding original shipping and handling charges. Any product repaired or replaced under this warranty will be warranted only for the remainder of the original product warranty period.
This warranty does not apply to consumable products such as chemical reagents; or consumable components of a product, such as, but not limited to, lamps and tubing.
Contact Hach Company or your distributor to initiate warranty support. Products may not be returned without authorization from Hach Company.
Warranty Limitations
This warranty does not cover:
Damage caused by acts of God, natural disaster, labor unrest, acts of war (declared or undeclared), terrorism, civil strife or acts of any governmental jurisdiction
Damage caused by misuse, neglect, accident or improper application or installation
Damage caused by any repair or attempted repair not authorized by Hach Company
Any product not used in accordance with the instructions furnished by Hach Company
Freight charges to return merchandise to Hach Company
Freight charges on expedited or express shipment of warranted parts or product
Travel fees associated with on-site warranty repair
This warranty contains the sole express warranty made by Hach Company in connection with its products. All implied warranties, including without limitation, the warranties of merchantability and fitness for a particular purpose, are expressly disclaimed.
Some states within the United States do not allow the disclaimer of implied warranties and if this is true in your state the above limitation may not apply to you. This warranty gives you specific rights, and you may also have other rights that vary from state to state.
This warranty constitutes the final, complete, and exclusive statement of warranty terms and no person is authorized to make any other warranties or representations on behalf of Hach Company.
Limitation of Remedies
The remedies of repair, replacement or refund of purchase price as stated above are the exclusive remedies for the breach of this warranty. On the basis of strict liability or under any other legal theory, in no event shall Hach Company be liable for any incidental or consequential damages of any kind for breach of warranty or negligence.
47
Limited Warranty
48

Appendix A Temperature compensation

Electrical conductivity indicates solution concentration and ionization of the dissolved material. Because temperature greatly affects ionization, conductivity measurements are temperature dependent, and are normally corrected to read what they would be at 25 ºC.

A.1 Compensation to 25 ºC

The MP series handheld meters includes temperature compensation to 25 ºC.
Temperature compensation can be set to KCl, NaCl or 442 solutions or tailored for special measurements or applications.

A.2 Changes in temperature compensation

Most conductivity meters approximate the temperature characteristics of solutions, and assume a constant value, such as, 2%/ºC. In fact, KCl temperature compensation changes with concentration and temperature in a non-linear fashion. Other solutions change even more. The MP series handheld meters use compensations that change with concentration and temperature instead of single
average values (Figure 8).
Figure 8
49
Temperature compensation

A.3 Graph of comparative error

In the range of 1000 µS, the error of using a KCl temperature compensation on a solution that should be calculated as NaCl or 442 is shown in the chart below
(Figure 9).
Figure 9
To measure natural water-based solutions to 1%, users must alter the internal temperature compensation to the more suitable, pre-loaded 442 values or stay close to 25 ºC.

A.4 Other solutions

A salt solution like sea water or liquid fertilizer acts like NaCl. The NaCl solution compensation provides the greatest accuracy for these solutions.
Many solutions differ greatly from KCl, NaCl or 442. A sugar solution, a silicate, or a calcium salt at a high or low temperature may require a User value to provide readings close to the true compensated conductivity. This is determined experimentally.
The chosen solution characteristic should closely match the sample being tested to achieve accuracy of ±1%.
50

Appendix B Conductivity conversion

B.1 How conductivity conversion works

When the effect of temperature is removed, the corrected conductivity depends on the concentration (TDS). Temperature compensation of the conductivity of a solution is performed automatically by the meter’s internal processor with data derived from chemical tables. Any dissolved salt at a known temperature has a known ratio of conductivity to concentration. Tables of conversion ratios referenced to 25 ºC have been published by chemists for decades.

B.2 Solution characteristics

Real-world applications have to measure a wide range of materials and mixtures of electrolyte solutions. To address this problem, industry applications tend to use the characteristics of a standard material as a model for their solution, such as KCl, which is favored by chemists for its stability.
Users who deal with sea water, etc., use NaCl as the model for their concentration calculations. Users who deal with freshwater work with mixtures including sulfates, carbonates and chlorides. These are modeled in the 442 standard solutions.
The meter contains algorithms for these three most commonly referenced compounds. The solution type in use is shown on the left side of the display. In addition to KCl, NaCl, and 442, a User choice is available. The User mode allows the user to enter the temperature compensation and TDS ratio by hand. This increases the accuracy of readings for a specific solution. That value remains a constant for all measurements, and should be reset for different dilutions or temperatures.
51
Conductivity conversion
52

Appendix C Temperature compensation and TDS derivation

The MP series handheld meters contain internal algorithms for characteristics of the three most commonly referenced compounds. The selected solution type is shown on the left of the display. In addition to KCl, NaCl and 442, a User choice is available. The User mode allows the user to enter the temperature compensation and the TDS conversion ratio of a unique solution.

C.1 Conductivity characteristics

When taking conductivity measurements, the Solution Selection determines the characteristic assumed as the instrument reports what a measured conductivity would be if it were at 25 °C. The characteristic is represented by the temperature compensation, expressed in %/°C.
If a solution of 100 µS at 25 °C increases to 122 µS at 35 °C, then a 22% increase has occurred over this change of 10 °C. The solution is then said to have a temperature compensation of 2.2%/°C.
Temperature compensation always varies among solutions because it is dependent on their individual ionization activity, temperature and concentration. This is why the MP meters feature mathematically generated models for known salt characteristics that also vary with concentration and temperature.

C.2 Temperature compensation of unknown solutions

The user may need to find the corrected conductivity of a solution that differs from the three standard salts. In order to enter a custom fixed temperature compensation for a limited measurement range, enter a specific value through the User function. The temperature compensation can be determined by two different methods.

C.2.1 Find temperature compensation by calculation

1. Heat or cool a sample of the solution to 25 ºC, and measure its conductivity.
2. Heat or cool the solution to a typical temperature where it is normally
measured.
3. Select the User function.
4. Set the temperature compensation to 0%/ºC (see section 3.12.1 on page 20).
5. Measure the new conductivity and the new temperature.
6. Divide the percentage decrease or percentage increase by the 25 ºC value.
7. Divide that result by the temperature difference.

C.2.2 Find temperature compensation by adjustment

1. Heat or cool a sample of the solution to 25 ºC, and measure its conductivity.
2. Heat or cool the solution to a typical temperature where it is normally
measured.
53
Temperature compensation and TDS derivation
3. Set the temperature compensation to an expected value (see section 3.12 on
page 20).
4. See if the compensated value is the same as the 25 ºC value.
5. If the value is not the same, raise or lower the temperature compensation and
measure again until the 25 ºC value is read.

C.3 TDS ratio of unknown solutions

When the effect of temperature is removed, the compensated conductivity varies with the concentration (TDS). The ratio of TDS to compensated conductivity for any solution also varies with concentration. The ratio is set during calibration in the User mode (see section 3.13 on page 20). Measure the TDS of an unknown solution by evaporation and weighing. Then measure the conductivity of the solution, with the now-known TDS, and calculate the ratio. The next time this solution is measured, the ratio is known.
54

Appendix D Additional information on pH and ORP (MP-6 and MP-6p models)

D.1 pH

D.1.1 pH as an indicator

pH measures the acidity or alkalinity of an aqueous solution. Another way to describe pH is as the hydrogen ion activity of a solution.
pH measures the effective, not the total acidity of a solution. A 4% solution of acetic acid (pH 4, vinegar) can be quite palatable, but a 4% solution of sulfuric acid (pH 0) is a violent poison. pH provides the needed quantitative information by expressing the degree of activity of an acid or base.
In a solution of one known component, pH indicates concentration directly. Very dilute solutions may be very slow reading because very few ions take time to accumulate.

D.1.2 pH units

The acidity or alkalinity of a solution measures the relative availability of hydrogen
+
) and hydroxide (OH-) ions. An increase in H+ ions increases acidity, while an
(H increase in OH
pH is defined as the negative logarithm of hydrogen ion concentration. Where
+
concentration falls below 10-7 mol/liter, solutions are less acidic than neutral, and
H therefore are alkaline. A concentration of 10 ions than OH
-
ions increases alkalinity.
-9
-
ions and is an alkaline solution of pH 9.
mol/liter of H+ has 100 times less H+

D.1.3 pH sensor

The active part of a pH sensor is a thin glass surface that is selectively receptive to hydrogen ions. Available hydrogen ions in a solution accumulate on this surface and a charge builds up across the glass interface. The voltage can be measured with a high impedance voltmeter circuit.
The glass surface encloses a captured solution of potassium chloride that holds an electrode of silver wire coated with silver chloride. This is the most inert connection possible from a metal to an electrolyte. It can still produce an offset voltage, but using the same materials to connect to the solution on the other side of the membrane cancels the two equal offsets.
The other electrode, also called the reference junction, allows the junction fluid to contact the test solution, without significant migration of liquids, through the plug material.
The pH/ORP sensor in the MP series meters (MP-6 and MP-6p) (Figure 10) is a single construction in an easy-to-replace package. The sensor body holds an oversize solution supply for long life. The reference junction is a wick that is porous so that it can provide a stable, low, permeable interface. It is located under the glass pH sensing electrode.
55
Additional information on pH and ORP (MP-6 and MP-6p models)
.
Figure 10 pH/ORP sensor construction
1 Glass surface 5 KCl solution
+
ions 6 Glass
2 H 3 Junction plug 7 Electrode wires 4 Platinum button

D.1.4 Sources of error

D.1.4.1 Reference junction

The most common sensor problem is a clogged junction because a sensor is allowed to dry out. The symptom is a drift in the zero setting at pH 7. This explains why MP series meter does not allow more than one pH unit of offset during calibration.

D.1.4.2 Sensitivity errors

Sensitivity is the receptiveness of the glass surface. A film on the surface can diminish the sensitivity and cause a long response time.
56
Additional information on pH and ORP (MP-6 and MP-6p models)

D.1.5 Temperature compensation

pH sensor glass changes sensitivity slightly with temperature. When the solution is further from pH 7, this effect increases. For example, a pH of 11 at 40 ºC is off by
0.2 units. The MP series meter senses the sensor cup temperature and
compensates the reading.

D.2 Oxidation Reduction Potential/Redox (ORP)

D.2.1 ORP as an indicator

ORP measures the ratio of oxidizing activity to reducing activity in a solution. It is the potential of a solution to give up electrons (to oxidize other things) or gain electrons (reduce).
Similar to acidity and alkalinity, one aspect increases at the expense of the other. Therefore, a single voltage is called the Oxidation-Reduction Potential, and a positive voltage shows a solution that wants to steal electrons (an oxidizing agent). For example, chlorinated water shows a positive ORP value.

D.2.2 ORP units

ORP is measured in millivolts, with no correction for solution temperature. Like pH, it is not a measurement of concentration directly, but of activity level. In a solution of only one active component, ORP indicates concentration. Also, similar to pH, a very dilute solution takes time to accumulate a readable charge.

D.2.3 ORP sensor

An ORP sensor uses a small platinum surface to accumulate charge without reacting chemically. That charge is measured relative to the solution, so the solution “ground” voltage comes from a reference junction. Figure 10 shows the platinum button in a glass sleeve. The same reference is used for both the pH and the ORP sensors. Both pH and ORP indicates 0 for a neutral solution. Calibration at zero corrects the error in the reference junction.
A zero calibration solution for ORP is not practical, so the MP series meters use the offset value determined during calibration to 7 in pH calibration (pH 7 = 0 mV). Sensitivity of the ORP surface is fixed so there is no gain adjustment.

D.2.4 Sources of error

Sources of error are similar to pH. Even though the platinum surface does not break like the glass pH surface, its protective glass sleeve can be broken. A surface film will slow the response time and diminish sensitivity.
57
Additional information on pH and ORP (MP-6 and MP-6p models)
58
Index
B
battery replacement ......................37
C
calibrate
conductivity, mineral/salt or TDS
................................ 32
ORP ..................................... 35
pH ........................................ 33
pH in multiple points ............... 34
resistivity ............................... 33
temperature ........................... 35
calibrate meter for user mode ........27
calibrate the meter ........................31
calibration
intervals
limits ..................................... 31
records ................................. 31
cancel User mode calibration Linc ..28
cell check ....................................25
change the temperature compensation
factor ..........................................20
Change the user-selected conductivity/TDS ratio
check the cell ...............................25
clean the conductivity/resistivity/TDS
sensor .........................................38
clean the pH/ORP sensor ..............39
cleaning consumables ..................45
clear a single record .....................22
clear all records ............................22
company contact information .........43
conductivity and pH/ORP sensor cups
................................ 31
...................20
....................................................13
conductivity characteristics ............53
conductivity conversion .................51
conductivity cup
clean the cup
consumables ...............................45
......................... 38
D
data port ......................................29
default settings .............................25
description of the display ...............15
disable temperature compensation .20
download data .............................29
drifting test results ........................40
E
exit calibration mode .....................32
extreme temperatures ...................37
F
features common to all models ......12
find temperature compensation
by adjustment
by calculation ........................ 53
........................ 53
G
general product information ...........12
I
items on the display ......................15
K
keypad descriptions ......................16
keypads ......................................16
L
Linc function ................................27
M
maintain the conductivity cup .........38
maintain the pH/ORP sensor cup ...38 measure
conductivity
mineral/salt ........................... 18
ORP/Redox ........................... 18
pH ........................................ 18
resistivity ............................... 17
TDS ..................................... 18
measure resistivity ........................17
memory settings ...........................21
MP-Dock .....................................29
........................... 17
O
ORP
as an indicator
Hach ORP sensor .................. 57
sources of error ..................... 57
....................... 57
59
Index
units .....................................57
other solutions ............................. 50
overview of the meters ................. 12
P
pH
as an indicator
sensor .................................. 55
sources of error ...................... 56
units .....................................55
pH/ORP sensor replacement ........ 38
power on/off the meter ................. 15
.......................55
R
replacement parts and accessories 45
reset factory calibration ................ 33
return to factory settings ............... 25
S
safety information ........................ 11
select a solution .......................... 19
select the user mode .................... 19
sensor
clean
.................................... 38
sensor cup
hydrate
.................................38
sensor replacement ..................... 38
set the date ................................. 23
set the time ................................. 22
set user mode calibration Linc ....... 27
solution characteristics ................. 51
solution types .............................. 19
solutions that damage the pH/ORP
sensor .........................................40
specifications .................................7
store a value in the memory ...........21
T
take a measurement .....................17
TDS ratio of unknown solutions ......54
temperature compensation
compensation to 25ºC
temperature compensation and TDS
derivation ....................................53
temperature compensation of unknown
solutions ......................................53
temperature extremes ...................37
temperature format .......................24
time and date ...............................22
troubleshooting ............................41
turn on/off the meter .....................15
............ 49
U
use of hazard information ..............11
user mode ...................................19
user mode calibration Linc function .27
user mode features .......................12
user solution ................................19
V
view the memory recall .................22
view the records ...........................22
W
warranty ......................................47
60
Loading...