Hach Intellical MTC30101, Intellical MTC30103 User Manual

User Manual
DOC022.52.80026
Refillable ORP/Redox Probe: Model MTC30101 or MTC30103
Safety information
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
Specifications
Note: Specifications are subject to change without notice.
Specifications Details
Probe type Digital combination electrode with a refillable Ag/AgCl reference and a built-
in temperature sensor
Range ±1200 mV
Resolution 0.1 mV
Temperature accuracy ±0.3 °C (±0.54 °F)
Operating temperature range 0 to 80 °C (32 to 176 °F)
Storage temperature range 5 to 40 °C (41 to 104 °F)
Junction Ceramic
Reference potential versus Standard Hydrogen Electrode
207 mV at 25 °C
Fill solution 3M KCl with saturated AgCl
Reference type Ag/AgCl (3 M KCl)
Minimum sample depth 20 mm
Dimensions Diameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Cable connection M12 digital output and connector compatible with HQd meters
Product overview
The MTC301 series probe is a refillable, combination oxidation reduction potential (ORP/Redox) probe with a built-in temperature sensor (Figure 1). The MTC30101 or MTC30103 probe is available with a 1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The probe measures absolute mV values in wastewater, drinking water and general applications. The probe measures ORP/Redox in wastewater, drinking water and general applications. A 59 mL bottle of reference electrolyte filling solution (3M KCl solution saturated with AgCl) is included with the probe.
1
Figure 1 Probe overview
1 Probe soaker bottle 4 Protective tape and filing-hole
2 Reference junction, platinum and temperature sensor 5 1 or 3 meter (3.28 or 9.84 ft) cable
3 Filling-hole cap 6 Probe soaker bottle holder
Preparation for use
Prepare the probe for use before calibration or sample measurement.
1. Turn the probe soaker bottle cap counter-clockwise to loosen the cap.
2. Remove the soaker bottle from the probe.
3. Rinse the reference junctions and electrode with deionized water thoroughly to
remove the viscous storage/filling solution completely. Blot dry with a lint-free cloth.
4. Remove the protective tape from the filling hole before initial use (refer to Figure 1 on page 2). Dispose of the protective tape.
5. Add filling solution to the probe as necessary (refer to Fill the probe on page 11). The filling solution must be above the standard solution or sample level during measurement (Figure 2).
6. Make sure that the filling hole is open during measurement for the proper flow of the filling solution.
Figure 2 Measurement method
1 Container 3 Filling solution level
2 Probe body 4 Standard solution or sample level
2
Calibration
Before calibration:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
It is not necessary to recalibrate when moving a calibrated probe from one HQd meter to another if the additional meter is configured to use the same calibration options.
To view the current calibration, push , select View Probe Data, then select View Current Calibration.
If any two probes are connected, push the UP or DOWN arrow to change to the single display mode in order to show the Calibrate option.
Prepare the probe for use (refer to Preparation for use on page 2).
Calibration notes:
Additional standards can be selected in the Calibration Options.
Do not dilute ORP/Redox standards. Use fresh ORP/Redox standard for calibration.
ZoBell’s redox potential is temperature dependent. The HQd calibration routine factors in this temperature dependency allowing accurate calibrations within the temperature range of 0 to 30 °C (32 to 86 °F). Light’s solution should be read at 25 °C (77 °F). Custom ORP/Redox calibration solution values and temperature are user­defined.
The calibration is recorded in the probe and the data log. The calibration is also sent to a PC, printer or flash memory stick if connected.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a calibration error occurs, refer to Troubleshooting on page 12.
Calibration procedure:
1. Connect the
probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn on the meter.
2. Push Calibrate. The
display shows the ORP/Redox standard solution that is necessary for calibration.
3. Add the fresh ORP/Redox standard solution to a beaker or an appropriate container.
4. Rinse the probe with deionized water. Blot dry with a lint­free cloth.
5. Put the probe in the standard solution so that the temperature sensor is completely submerged. Stir gently. Shake the probe from side to side in the sample to refresh the reference junction.
3
6. Push Read. Stir gently. The display will show "Stabilizing" and a progress bar as the probe stabilizes in the sample. The display shows the standard solution value and the mV offset when the reading is stable.
7. Push Done to view the calibration summary.
8. Push Store to accept the calibration and go back to the measurement mode.
Sample measurement
Before measurement:
The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is attached.
If complete traceability is necessary, enter a sample ID and operator ID before measurement. Refer to the HQd meter manual for more information.
Regular calibration is required for the best measurement accuracy (refer to Calibration on page 3).
Prepare the probe for use (refer to Preparation for use on page 2).
Make sure that the platinum disc is clean and smooth (refer to Clean the probe on page 10).
Measurement notes:
Data is automatically stored in the data log when Press to Read or Interval is selected in the Measurement Mode. When Continuous is selected, data will only be stored when Store is selected.
Air bubbles under the sensor tip when submerged can cause slow response or error in measurement. If bubbles are present, gently shake the probe until bubbles are removed.
If a measurement error occurs, refer to Troubleshooting on page 12.
Measurement—direct method procedure:
4
1. Connect the probe to the meter. Make sure that the cable locking nut is securely connected to the meter. Turn the meter on.
2. To significantly reduce the stabilization time for reducing-type samples, put the platinum disc in Reducing Solution for ORP Electrodes for 3-10 minutes before the initial sample measurement.
3. Rinse the probe with the sample.
4. Put the probe in the sample and stir gently. Make sure that the reference junctions are completely submerged. Do not put the probe on the bottom or sides of the container. Shake the probe from side to side in the sample to refresh the reference junction.
5. Push Read. The display will show "Stabilizing" and a progress bar as the probe stabilizes in the sample. The display will show the lock icon when the reading stabilizes. If necessary for the application, record the sample pH and temperature.
6. Repeat steps 3-6 for additional measurements.
7. When measurements are done, store the probe (refer to
Storage
on page 11).
Measurement—conversion to SHE reference procedure:
For some applications, it is customary to report redox potential readings relative to the standard hydrogen electrode (SHE), also called normal hydrogen electrode (NHE). To do this, select the value in Table 1 that corresponds to the temperature of the solution measured. Substitute that value Eref into the equation and solve for Eh:
Eh = E + E
ref
where:
Eh = oxidation reduction potential of the sample relative to the SHE
E = potential developed by the ORP/Redox electrode
E
ref
= potential developed by the reference electrode portion relative to the SHE
(Table 1).
Table 1 shows the potentials, E
ref
, developed by the reference electrode portion relative
to the SHE at various temperatures.
Table 1 Standard potential of reference electrode
Temperature (°C) Electrode potential in mV (E
ref
)
80 163.1
75 167.7
5
Loading...
+ 11 hidden pages