Gyrozen 416 Service manual

Low Speed Centrifuge, 416 – Service manual
# C01DC00304-11
1. Model: Low speed Centrifuge, 416
2. Manufacturer: Gyrozen Co., Ltd.
3. Service: BMS Co., Ltd. Technical Service Team, 82-2-3471-8171
Gyrozen Co., Ltd Customer Service Team, 82-70-8620-5350
1
Low Speed Centrifuge, 416 – Service manual
CONFIDENTIAL
1 Operating Instruction ...............................................................................................................4
1.1 About this manual ..............................................................................................................4
1.2 Safety Label and safety precaution ...............................................................................4
1.2.1 Safety label ...............................................................................................................4
1.2.2 Safety Precaution .....................................................................................................5
2. Installation ...................................................................................................................................6
2.1 Unpacking ..............................................................................................................................6
2.2 Location ..................................................................................................................................7
2.3 Supply the power .................................................................................................................7
2.4 Power on and opening the door.......................................................................................7
3. Device Information.....................................................................................................................8
3.1 Special qualities ....................................................................................................................8
3.2 Technical Specifications ......................................................................................................8
3.3 Outer Description .................................................................................................................9
3.4 Operating Function of Control Panel ...............................................................................9
3.5 Operating System.............................................................................................................. 10
3.6 Main Controller Board...................................................................................................... 11
3.7 Display Controller Board.................................................................................................. 11
4. Disassembling .......................................................................................................................... 12
4.1 Front panel and controller board................................................................................... 12
4.2 Door Assembly................................................................................................................... 13
4.3 Motor cover & Chamber.................................................................................................. 13
4.4 Door lock assembly........................................................................................................... 14
4.5 Motor assembly................................................................................................................. 14
4.6 Anti-vibration rubber ....................................................................................................... 15
2
Low Speed Centrifuge, 416 – Service manual
4.7 Imbalance sensor assembly............................................................................................. 15
4.8 RPM sensor assembly....................................................................................................... 16
5. Service Mode and Adjustment ............................................................................................ 17
5.1 Transition into service mode .................................................................................... 17
5.2 Handling values ........................................................................................................... 17
5.3 Procedure for Imbalance adjustment...................................................................... 18
5.4 Door lock ass’y adjustment....................................................................................... 19
6. Error code and Troubleshooting ........................................................................................ 20
6.1 Error code..................................................................................................................... 20
6.2 Troubleshooting........................................................................................................... 22
7. Maintenance ........................................................................................................................... 23
7.1 Cleaning and disinfection ........................................................................................ 23
7.2 Device test for centrifuge ........................................................................................ 24
7.2.1 Validation for actual RPM................................................................................... 24
7.2.2 Validation of Motor performance ................................................................... 24
8. Parts Information .................................................................................................................... 25
8.1 Assembly Drawing .................................................................................................... 25
8.2 Part List......................................................................................................................... 30
3
Low Speed Centrifuge, 416 – Service manual
1. Operating Instruction
1.1 About this manual
This service manual should be used by specialized engineers authorized by Gyrozen
Co., Ltd.
Any repairing work operated by non-authorized personnel cannot be protected and
guaranteed.
This service manual aims to find possible errors quickly and fix them properly.
Refer to the user’s manual for detailed operation of Centrifuge.
Do not copy or reprint without approval
1.2 Safety Label and Safety precaution
1.2.1 Safety label
The labels attached to the device give information for safety.
Label Information Label Information
Attention label to show
risk and warning
Attention label to warn
electric shock
4
Low Speed Centrifuge, 416 – Service manual
Make sure to
1.2.2 Safety Precautions
Supply proper voltage power according to device’s power requirement.
Let all repairing works done by authorized or qualified personnel.
Use rotors or accessories which are approved by Gyrozen.
Not try to open the lid and or move the device while the rotor is running.
Operate the centrifuge with a rotor properly attached and secured to the shaft.
Not use flammable, hazardous, explosive, or corrosive materials.
NOTE: When it is required to use toxic, radioactive materials or pathogenic micro-
organisms, which belong to the Risk Group II of WHO: “Laboratory Bio-safety Manual,”
should follow the regulation guidelines from WHO.
http://www.who.int/csr/resources/publications/biosafety/Labbiosafety.pdf)
Keep away hazardous materials farther than 30 cm (12 in) from the device during
centrifugation, as recommended in IEC standards 61010-2-020.
Keep the rpm or rcf under its maximum speed in the case that the density of sample
materials is greater than 1.2 g/ml to avoid rotor failure.
Load samples symmetrically in the rotor diagonally to make balance between the tubes.
Balance the load on the rotor totally to prevent the damage to the device even by using
several water-filled tubes.
Place device on a flat, level, rigid and stable surface.
Disconnect power supply prior to maintenance and service work to avoid electrical shock.
Use proper disinfection procedures when centrifuging bio hazardous compounds.
In Blackout
When a blackout takes place while the device is running, the door does not open. And the
rotor speed begins to decrease at natural level. Even if the power turns on before the rotor
stops completely, the rotor does not return to the original speed, but decreases more rapidly
with buzzer sound.
Door opening
The door is closed/opened automatically by a door lock unit operated by a solenoid, and it
will not be opened while the rotor is running at all. Even if the door is opened accidentally,
a door limit switch senses it instantly to make the rotor speed decrease.
Device vibration
If the rotor loses balance while running by any reason, it invokes vibration on the device
itself. In this case the Imbalance sensor senses it and makes the rotor begin to decrease with
preset level issuing Imbalance Error warning. This safety function protects the device from
damage during operation.
5
Low Speed Centrifuge, 416 – Service manual
2. Installation
2.1 Unpacking
1) Check if the box contains such parts as;
① Power cable, 1ea
② Emergency door open tool, 1ea
③ User’s manual, 1ea
④ Rotor (Optional)
⑤ Rotor coupling device
2) Open the box and lift out the device carefully together with the safety padding.
3) Remove the safety padding and vinyl wrap.
4) Remove the motor fixing bolts
There are 2 screws placed at the bottom of the device. They protect the imbalance
sensor from any damage caused by some accidental moving of the motor during
delivery.
And they should be removed before the installation of the device as follows;
① Check if a long bolt(A came out from bottom of the device. As it is longer than
the leg pieces, the device cannot stand horizontally.
② Check also another screw around the long one.
Bottom side
B A
Unscrew and remove the 2 screws with a provided wrench.
5) Place the device on the flat surface.
6
Low Speed Centrifuge, 416 – Service manual
2.2 Location
1) Install the device at the solid and flat floor or table. If you place the centrifuge at
the slope, the axis of rotation is possibly changed because of the rotor weight.
2) Install the device keeping a distance of 30cm at least from the wall. The distance is
needed for the air circulation around the device.
3) Install the device at the place with appropriate temperature and humidity. These
conditions have to be maintained constantly as soon as possible.
4) Install the device at the place without any kinds of corrosive gases.
2.3 Supply the power
1) 416 model uses 110V or 220V. Check proper voltage of the device and connect to
adequate power outlet.
2) If the power input is more than +/- 10% of the recommended voltage or
fluctuating frequently, it may affect some functions of the device. In that case it is
recommended to use AVR (Automatic Voltage Regulator).
3) If you want to use the device under the other voltage range, please contact us for
safe usage.
2.4 Power On and opening the door
1) Turn on the device by the switch on the right side of the machine.
2) Press the ‘Door’ button to open the door.
If it does not work (door not open), use the emergency door open tool (as the
figure)
7
Low Speed Centrifuge, 416 – Service manual
3. Device Information
3.1 Special qualities
High safety and low noise
Fixed angle rotor and Swing-out Rotor available
Simultaneous display of rpm and rcf speed
Automatic Alarm function for Imbalance, Door open, Speed trouble
High tech AC Induction Motor adopted
3.2 Technical Specifications
Max. RPM(Fixed Angle/Swing out) 4,000/4,000 rpm
Max. RCF(Fixed Angle/Swing out) 2,700/2,826 g
Max. capacity(Fixed Angle/Swing) 16 x 15 ㎖/4 x 100 ㎖
Run time ≤ 99 min 59 sec or continuous
Acceleration time ≤ 20 sec
Deceleration time ≤ 25 sec
Program memory 10
Power supply 220V, 50/60Hz
Power requirement 300 VA
Dimension (WxDxH) mm 375 x 480 x 260
Weight 19.5 Kg
CE Certification Yes
8
Low Speed Centrifuge, 416 – Service manual
3.3 Outer Description
Main Body
3.4 Operating Function of Control Panel
Angle Rotor,
GRA-S-15-16
Swing Rotor, GRS-S-mw-2
Swing Rotor,
GRS-S-50-4
1) RPM/RCF: to show the RPM/RCF display
2) TIME: to show pre-set time up to 99 min 59 sec (00: continuous)
3) SOFT START & STOP: to accelerate or decelerate slowly.
4) Door: to open the door.
5) RPM/RCF: to switch the RPM/RCF display
6) Up & down arrow: Uses to change input data.
7) TIME: to set test time up to 99 min 59 sec (00: continuous)
8) START/STOP: Uses to start & stop operation.
9) PROG: to save a set of setting values or call the saved setting values.
* If you press the arrow button longer than 3 seconds, the numbers change rapidly and
the set-up is achieved faster.
9
Low Speed Centrifuge, 416 – Service manual
3.5 Operating System
Main board
1) DSPIC 30F4011, MICOM; It controls all devices.
2) SMPS; distributes inlet power to each part as appropriate form.
3) Inverter; transforms the single phase power to the 3 phase for running the AC
induction motor.
4) IPM FSBS5CH60; controls the AC induction motor.
5) Solenoid; open and close the door lid automatically.
6) Temp sensor; measure the temperature of motor at its surface and issue E3 error if
too high(above 110℃).
7) Firmware program; is used to update the firmware with notebook and interfacing
connector.
10