GW Instek PPH-1503 User Manual

Programmable High Precision DC Power Supply
PPH-1503
User Manual
ISO-9001 CERTIFIED MANUFACTURER
Copyright Statement
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
TABLE OF CONTENTS
Table of Contents
SAFETY INSTRUCTIONS ................................................... 5
OVERVIEW ...................................................................... 11
Introduction ................................................... 11
Key Features ................................................... 13
Operating Principals ....................................... 15
Front Panel ..................................................... 17
Rear Panel ...................................................... 23
Constant Voltage/Constant Current Crossover
Characteristics................................................ 25
GETTING STARTED ......................................................... 27
Start Up .......................................................... 27
DVM and Load Connection ............................. 28
Turning the Output On/Off............................. 29
BASIC OPERATION ......................................................... 31
Basic Power Supply Functions ........................ 31
DVM ............................................................... 41
Pulse Current Measurement ........................... 43
Long integration ............................................. 49
Current Sink Function ..................................... 54
External Relay Control .................................... 56
Save/Recall ..................................................................... 59
Save Settings .................................................. 59
Recall Settings ................................................ 60
Restore Factory Default Settings ..................... 63
3
PPH-1503 User Manual
System Settings .............................................................. 64
System Information ........................................ 64
Utilty Settings ................................................. 65
REMOTE CONTROL ........................................................ 67
Remote Control .............................................. 67
Command Syntax ............................................ 76
Command List ................................................ 80
Command Details ........................................... 85
SCPI Status Registers SCPI ........................... 121
Errors ........................................................... 129
APPENDIX ..................................................................... 133
Replacing the Fuse ....................................... 133
Specifications ............................................... 134
Optional Accessories .................................... 136
Declaration of Conformity ............................ 137
4
SAFETY INSTRUCTIONS
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the PSW or to other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal

SAFETY INSTRUCTIONS

This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to insure your safety and to keep the instrument in the best possible condition.

Safety Symbols

These symbols may appear in the manual or on the instrument.
5
PPH-1503 User Manual
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
6
SAFETY INSTRUCTIONS
General Guideline
CAUTION
Do not place any heavy object on the unit. Avoid severe impact or rough handling that
leads to damaging the unit.
Do not discharge static electricity to the unit. Do not block the cooling fan opening. Do not perform measurements on circuits that
are directly connected to mains power.
Do not disassemble the PSW unless you are
qualified.
(Measurement categories) EN 61010-1:2001 specifies the measurement categories and their requirements as follows. The PPH-1503 falls under category I.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed on
circuits not directly connected to Mains.
Power Supply
WARNING
AC Input voltage range: 90VAC~264VAC Frequency: 50Hz/60Hz To avoid electrical shock connect the protective
grounding conductor of the AC power cord to an earth ground.

Safety Guidelines

7
PPH-1503 User Manual
Fuse
WARNING
Fuse type: T2.0A/250V To prevent fire, replace the fuse only with the
specified type and rating.
Disconnect the power cord before replacing the
fuse.
Make sure the cause of fuse blowout is fixed
before replacing the fuse.
Cleaning the power supply
Disconnect the power cord before cleaning the
oscilloscope.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid into the oscilloscope.
Do not use chemicals containing harsh products
such as benzene, toluene, xylene, and acetone.
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: < 80% Altitude: < 2000m Temperature: 0°C to 40°C
(Pollution Degree) EN 61010-1:2001 specifies pollution degrees and their requirements as follows. The PPH-1503 falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
8
SAFETY INSTRUCTIONS
Storage environment
Location: Indoor Relative Humidity: < 70% Temperature: -10°C to 70°C
9
PPH-1503 User Manual
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)

Power cord for the United Kingdom

When using the power supply in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code:
As the colours of the wires in main leads may not correspond with the coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with either the letter E, the earth symbol or coloured
Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, a cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable, any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced in accordance to the above standard.
10
OVERVIEW

Overview

The PPH-1503 is a portable high-speed programmable DC power supply with flexible operating configurations. In addition to the basic power supply functionality, it is also able to measure pulse current and the average current over long periods of time.
The PPH-1503 is designed for testing the power consumption of battery powered wireless communication devices (e.g. cell phones). Such devices often have large load variations within a short time span. The high precision power supply has excellent voltage stability during pulsed loads and is capable of simultaneously measuring the pulse current, even for very short pulses. In addition, the power supply is able to sink current, allowing it to simulate the characterics of a discharged rechargeable battery for testing chargers and charge control circuits.
OVERVIEW
This chapter contains a brief introduction to PPH-1503, the main features, as well as an overview of the front and rear panel. Use the Getting Started chapter on page 27 to for start up instructions and how to setup the appropriate operation environment.

Introduction

11
PPH-1503 User Manual
Basic Power Supply Function
The PPH-1503 works as a conventional power supply with automatic CC/CV crossover. Parameters such as the output voltage, current, read back refresh rate, data sampling period, power-on status, OVP and current range can be configured using the control panel. The voltage and current settings and the actual voltage/current are displayed on the LCD. For details, see page 31.
Pulse Current Measurement Function
The PPH-1503 can measure the change in instantaneous current and the current of extremely short pulses. The readback refresh rate, data sampling period, trigger delay and trigger level can be set by the front panel keys and is displayed on the LCD. For details see page 43.
Current Measurement over Long Periods
This function can measure the average current of one or more pulses. The readback refresh rate, trigger mode, and trigger timeout and trigger level settings are controled by the front panel keys and is displayed on the LCD display. For details, see page 49.
Current Sink Features
When the voltage of an external power source is greater than the high-speed power supply output, the system will automatically work as an electronic load to sink current. For details, see page 54.
Digital Volt Meter
The PPH-1503 has a DVM function that can measure DC voltages in the range of 0~20VDC. For details, see page 41.
Remote Control
To meet the various needs of customers, the PPH­1503 is designed for USB, GPIB and LAN remote control. For details, see page 67.
12
OVERVIEW
Additional Features
The PPH-1503 has external relay control signals for customers. The relay control signals are synced to the pulse current measurement feature. For details, see page 56.
Features
Low noise: Thermostatically controled fan. Compact, lightweight. 3.5 inch TFT display.
Operation
Constant voltage and constant current operation
(CV/CC).
Output on/off control. Front and Rear output control key. Digital panel control. 5 groups of save/recall settings and 10
automatically generated power-on settings.
Digital voltage and current settings. Software calibration. Alarm buzzer. Key lock function.
Protection Features
Reverse polarity protection. Overvoltage and overcurrent protection
(OVP/Trip).
Overtemperature protection (OTP).

Key Features

13
PPH-1503 User Manual
Interface
USB remote control. GPIB remote control. LAN remote control.
14
OVERVIEW
Overview
The PPH-1503 mainly consists of the follow components:
AC to DC Switching power supply DC to DC Buck converter circuit Precision output control circuit
The block diagram below shows a function description of each of the circuits. The following page will show detailed decriptions of each component.
Block Diagram
AC INPUT
AC TO DC
SWITCH POWER
DC TO DC BUCK CIRCUIT U201,Q201,Q202
L202
LINEAR
CONTROL
Q306~Q308,U303
ON/OFF SWITCH
Q309,Q310, Q315,Q316
AUXILIARY
POWER SUPPLY
U101,T101, Q102~Q105
V-SET
I-SET
AND
VOLTAGE
COMPARATOR
U301A
VOLTAGE
SAMPLING
CIRCUIT
U401,U402
CURRENT
COMPARATOR
U301B
CURRENT SAMPLING
CIRCUIT
U430
SAMPLE
RESISTANCE
R366,R366A R367,R367A
OUTPUT
Swtiching Power Supply
AC power is converted to 24VDC by the switch mode power supply module.
DC Down Conversion
To reduce the input voltage to 24VDC (slightly higher than the settable voltage) the U201 Buck IC is used in conjunction with two power MOSFETS (Q201/Q202) and an inductor (L202).

Operating Principals

15
PPH-1503 User Manual
Linear Output Circuit (Linear Regulator)
The Q306, Q307 dividers reduce the heat on a single component. The U303, U301, U403, U401 and U402 components form a control circuit to achieve accurate output.
Auxillary Power Supply
The independent auxillary DC power supply is achieved with the U101, T101 and Q102~Q105 components.
16

Front Panel

 










 

PPH-1503
LCD display
Numberpad and
secondary function keys
Arrow keys
DVM inputs
Front panel output
terminals
Function keys
Power button
Display
Voltmeter Indicators
Displays the output voltage with up to 5 digits of resolution. The default units are Volts (V).
15.000 v
Ammeter Indicator
Displays the output current with up to 5 digits of resolution, depending on the current range (5A/5mA). The current range is selectable between A and mA.
5.0000 A
or
5.0000 mA
OVERVIEW
17
PPH-1503 User Manual
Setting Display
Displays the voltage and current settings.
V-Set
15.000 V
I-Set
5.0000 A
Parameter Settings Display
Displays the parameter settings. For details on setting parameters, see page 20. The following figure shows the F1 parameter settings (V AND I), for example.
Status Display
Display the current status of the instrument.
Status
Output mode
CV mode:
C V
CC mode:
C C
Overvoltage protection
Enabled:
O.V.P
Disabled:
O.V.P
Alarm
Enabled:
BEEP
Disabled:
BEEP
Key lock
Locked:
LOCK
Unlocked:
LOCK
CurrRange: 5 A LimMode: Limit
PowOnSetup: RST OutputRelay: One
O.V.P: Off RecallSetup: ---
IntRate: 1.00 PLC AverRead[1][2]: 1
18
OVERVIEW
Remote connectivity
Local mode:
RMT
Remote:
GPIB LAN
USB
Output
Switching the Output Source
Front:
FRONT
Rear:
REAR
Output State
On:
ON
Off:
OFF
Function Display
Displays the unit functions. There four functions:
F1: Basic power supply function (V AND I); F2: Digital Voltmeter function (DVM); F3: Pulse current meter function (PULSE); F4: Long integration current measurement function
The basic power supply function is shown below. (The active function is shown in yellow.)
Function Keys
Menu key
Menu key to enter or exit from system settings.
Voltage Setting key
Press the Voltage key to set the voltage settings. See page 34 for operation details.
19
PPH-1503 User Manual
Current Setting key
Press the Current key to set the current settings. See page 35 for operation details.
Front and Rear output toggle key

Front and rear output toggle switch. The key will be lit when the output is set to the rear outputs.
Rear panel output:


Output key
The Output key turns the output on or off. The Output key will lightup when the output is on.
On:
LOCK key

The Lock key is used to disable all the panel keys except for the Output key. Pressing the Lock key for at least 2 seconds will turn the panel lock on or off. The Lock key can also be used to exit from remote control mode. When the panel lock is active the Lock key will light up.
Locked


20
OVERVIEW
Numberpad








a. The numberpad is used to enter various parameters and values. The Clear key can be used to clear set parameters.
b. F1/F2/F3/F4 function short cuts. Press any of the function short cuts when in the main menu to enter the corresponding function interface. F1: Basic power supply function F2: Digital voltmeter function F3: Pulse current meter function F4: Long integration current measurement function.
c. H/L/A Pulse current measurement shortcut keys. These short cut keys only work in the Pulse current measurement main menu.
H: High measurement mode L: Low measurement mode A: Average measurement mode
Directional keys and Enter key

The directional keys are used for parameter and menu selection as well for fine adjustment of the current/voltage settings.
The Enter key is used to confirm the selection of any settings or parameters and to exit after a setting is complete.
21
PPH-1503 User Manual
Power Button
Turns the power on or off. On: Off:
Terminals
Output Terminals (SOURCE)

Output source terminals.
Voltage Feedback Terminals (SENSE)


Sense terminals.
Voltmeter Terminals (DVM)


Digital voltmeter input terminals.
22

Rear Panel

AC power socket
and fuse
Rear panel outputs and
DVM inputs
External
control port
Lan port
USB port
GPIB port
Heat sink fan
Terminals
AC input socket and line fuse
The AC input: 90~264VAC, 50Hz/60Hz Fuse: 2A slow-blow type. See page
133 for details.
USB port
USB device port for remote control. See page 67 for details.
GPIB port
GPIB slave port for remote control. Abides to IEEE488.1 (SCPI) protocol. See page 69 for details.
LAN port
LAN port for remote control. See page 72 for details.
OVERVIEW
23
PPH-1503 User Manual
Output interface
A total of 8 ports: 2 positive output terminals, 2 negative output terminals, a Sense+ terminal, a Sense- terminal, a DVM- input terminal and a DVM+ input terminal. Refer to the printed label under the terminals for the specific order of the terminals.
Relay control interface
A total of 5 ports: A +5V input terminal, a ground terminal and 2 terminals for relay control. See page 56 for relay control details.
24
OVERVIEW
Background
The unit will switch automatically between constant voltage and constant current according to changes in the load.
CV mode
When the load current is less than the current setting, the unit operates in constant voltage mode, changing the current level according to the load but maintaining the set voltage level until the current reaches the set current level. The status indicator will show CV on the LCD when in CV mode.
Constant Current Mode
When the output current reaches the set current level, the unit switches operation to constant current mode. The status indicator will show CC on the LCD display. In CC mode, the current level is maintained and the voltage level is limited to less than the set voltage level to limit the output power from an overload. When the current drops below the set current level, the unit will revert back to CV mode.

Constant Voltage/Constant Current Crossover Characteristics

25
PPH-1503 User Manual
Diagram
Vmax
Imax
Constant
Voltage
Constant
Current
Vout
Iout
26
GETTING STARTED
Checking the AC Voltage
Before the power is turned on, confirm that the input power supply meets the following conditions: 90-264VAC, 50Hz/60Hz
Connecting the AC power cord
The fuse is a 2A slow-blow fuse. Confirm that the fuse is of the correct type and rating before connecting the power cord.
Turning the power on
Press the power button. The LCD will display the line frequency of the AC power supply.
Turning the power off
To turn the power off, press the power button again.

GETTING STARTED

This chapter describes the start up procedures and the preparation that is necessary before operating the power supply.

Start Up

27
PPH-1503 User Manual
Recommended Cables
Model
Specification
Usage GTL-117
10A
Front panel DVM input
GTL-204A
10A
Front panel Source terminal
GTL-203A
3A
Front panel Sense terminal
Front panel wiring
Use the GTL-204A cables for the front panel source connections.

Use the GTL-203A cables for the sense connections.


Use the GTL-117 cables for the DVM connections.


Rear panel connections
Rotate the screws counter clockwise to loosen the ports.
Insert the wires into the appropriate terminal according to the labels printed under the terminals.

DVM and Load Connection

28
GETTING STARTED
Screw the terminals in a clockwise direction to tighten.
Note
For safety considerations, please keep in mind that the front panel and rear panel terminals are physically connected.
Wire Gauge
Load wires must have enough current capacityto minimize cable loss and load line impedance. Voltage drop across a wire should not excess 0.5V. The following list is the wire current rating at 450A/cm2.
Wire Size(AWG)
Maximum Current (A)
20
2.5 18
4
16
6
14
10 12
16
Panel Operation
Press the Output key to turn the output on. The Output key will light-up when the output is on.
The status display on the LCD display will also turn on.
OFF
ON

Turning the Output On/Off

29
PPH-1503 User Manual
When the output is turned on, pressing the Output key again will turn the output off. When the output is off, the Output key will no longer be illuminated and the status on the LCD display will revert back to the OFF status.
Automatic Output Shut Down
Any of the following actions will cause the output to be automatically shut down:
If any of the setups are recalled. OVP/OTP protection is tripped. OCP protection is tripped.
30
BASIC OPERATION
Description
The PPH-1503 operates as a generic power supply with the ability to display different current ranges. The output can be toggled between the front and rear outputs using the Rear key. When the Rear key is lit, it indicates that the rear panel output is active and that the front panel output is off.
Parameter Description
IntRate
The data sampling period derived from the number of power line cycles.
The setting range is:
0.1PLC to 10.00PLC (power line cycles)
1PLC = 16.7ms(60Hz)/20ms(50Hz).
*PLC stands for power line cycles.

BASIC OPERATION

This chapter describes how to set various functions.

Basic Power Supply Functions

31
PPH-1503 User Manual
AverRead[1][2]
Readback refresh rate. This will display the average number count. The settings for parameters[1][2] are shared, the remaining parameters [3][4] are set in their corresponding menus.
[1] Power Supply functions
[2] DWM function.
[3] Pulse current measurement
[4] Long integration current measurement
CurrRange
The current range selection has three settings: 5A, 5mA and Auto.
The 5mA range only accepts a current setting 1A or less. If the 5mA range is selected and if the current setting is greater than 1A, the setting value is automatically reduced to 1A.
32
BASIC OPERATION
LimMode
Current limiting mode
There are 4 settings for the current limiting mode:
Limit, Trip, LimitRelay, Trip Relay.
The Limit settings will limit the current. When the current reaches the setting value, the current remains constant, as in CC mode.
The Trip setting will turn the output off when the current limit has been reached.
See page 56 for details on the Limit Relay and Trip Relay settings.
PowoOnSetup
Power on settings have 11 settings:
Rst/SAV0 ~ SAV9
See page 60 for further details.
RelayControl
The relay control settings have 2 configurations:
Zero/One
See page 56 for further details.
O.V.P
The overvoltage settings have a setting range of 0.05 to 15.20V or OFF.
RecallSetup
There are 6 sets of save/recall memories.
Rst/ SAV0 to SAV4
33
PPH-1503 User Manual
Output Range
Voltage
0.000V~15.000V
Current
0.0000A~3.0000A (0V~15V)
0.0000A~5.0000A (0V~9V)
Parameter Settings
Voltage
Press the Voltage key and the voltage setting on the LCD is activated. A yellow dot appears under the current digit.
V-Set
00.500 V
(a)Use the number pad (keys: 0~9, Clear) to set the voltage value and then press the Enter key.
To enter 12.345V:



The input dialog box appears on the LCD:
12.345
34
BASIC OPERATION
(b)Step Setting:
Press the left and right
arrow keys ( , ) to
fine tune the voltage setting at the digit level. The selected digit will have a yellow dot directly underneath. Press the up and down arrow keys ( , ) to adjust the selected digit. Press the Voltage key again to finish and exit setting the voltage.

Current
Press the Current key and the current setting on the LCD is activated. A yellow dot appears under the current digit.
I-Set
2.000 A
(a)Use the number pad (keys: 0~9, Clear) to set the current value and then press the Enter key.
To enter 1.2345A:



The input dialog box appears on the LCD:
1.2345
35
PPH-1503 User Manual
Press the left and right
arrow keys ( , ) to
fine tune the current setting at the digit level. The selected digit will have a yellow dot directly underneath. Press the up and down arrow keys ( , ) to adjust the selected digit. Press the Current key again to finish and exit setting the current.

IntRate
Use the arrow keys to select IntRate and press Enter. Input the parameters and press Enter to save.
Range: 0.1 to 10.00
AverRead[1][2]
Use the arrow keys to select
AverRead[1][2] and press Enter. Input
the parameters and press Enter to save.
Range: 1 to 10
CurrRange
Use the arrow keys to select
CurrRange and press Enter to go the
the CurrentRange menu. Input the current range using the up and down arrow keys. Press Enter to save.
36
BASIC OPERATION
LimMode
Use the arrow keys to select LimMode and press Enter to go to the Current Lim menu. Select the Current Lim mode using the up and down arrow keys. Press Enter to save.
See page 56 for further details.
PowOnSetup
Use the arrow keys to select
PowOnSetup and press Enter to go to
the Power On Setup menu. Use the up and down arrow keys to select the power on setting. Press Enter to save.
See page 60 for further details.
RelayControl
Use the arrow keys to select
RelayControl and press Enter. Use the
up and down arrow keys to set the type of relay control. Press Enter to save.
See page 56 for further details.
O.V.P
Use the arrow keys to select O.V.P and press Enter. Input the OVP setting and press Enter to save.
RecallSetup
Use the arrow keys to select
RecallSetup and press Enter to go to
the Recall Setup menu. Use the arrow keys select a stored setup. Press Enter to confirm the recall.
See page 60 for further details.
37
PPH-1503 User Manual
Note:
1. The
Clear
key can be used to clear numbers that
have already been entered.
2. Voltage and current parameter values use stepped input values. All other numerical parameters can use the number pad to enter parameter values.
Operation
REAR / FRONT
After setting all the parameters the Rear key can be used to set the output to the front or rear output terminals.
Pressing the Rear key will toggle the output between the front and rear terminals.
When set to rear, the Rear key will light up and REAR will appear in the status bar on on the LCD.
When set to front, the Rear key will not be lit and FRONT will appear in the status bar on the LCD.

REAR

FRONT
38
BASIC OPERATION
Output
Press the Output key to turn the output on. When the output is on, the Output key will light up and ON (in green) will be shown in the status bar on the LCD.
When the output is off, the Output key will not be lit and OFF (in red) will be shown in the status bar on the LCD.
ON
OFF
Status Description
CV/CC
These two icons represent the output states of the power supply:
CV appears in yellow when the power supply is in constant voltage mode.
CC appears in red when the power supply is in constant current mode.
C V C C
O.V.P
OVP will appear in yellow when the OVP has not been tripped.
The OVP icon will turn red when the OVP has tripped.
When the OVP protection has not been activated, it will be greyed-out.
O.V.P
O.V.P
O.V.P
39
PPH-1503 User Manual
BEEP
When the beeper setting is activated, the BEEP icon will be shown in yellow.
When the beeper settings is turned off, it will be greyed­out.
BEEP
BEEP
LOCK
When the panel lock is activated, the lock icon will be shown in red.
When the panel lock is turned off, the lock icon is greyed-out.
LOCK
LOCK
RMT
In the remote control display area, RMT will be shown in grey when reomote control is disabled.
When GPIB remote control is active, a red GPIB icon is shown. When LAN remote control is active, a red LAN icon is shown and when USB remote control is active, a red USB icon is shown.
RMT
GPIB
LAN
USB
40
BASIC OPERATION
REAR/ FRONT
When the output is set for the rear panel terminals, REAR is displayed in yellow.
When the output is set for the front panel, FRONT is displayed in yellow.
REAR
FRONT
ON/OFF
When the output is off, OFF in displayed.
When the output is on, ON is displayed.
OFF
ON
Description
The PPH-1503 has a separate digital voltmeter with a measurement range of 0~+20VDC. When using the voltage meter, the power supply must be properly grounded.
Parameter Description
Intrate
Sets the reading rate of DVM measurements based on the number of PLCs.
0.01PLC~10.00PLC.
1PLC=16.7ms(60Hz)/20ms(50Hz).
*PLC stands for Power Line Cycle
DVM
41
PPH-1503 User Manual
AverRead[1][2]
The number of samples used to calculate the average. The AverRead setting for the power supply functions[1] and the DVM functions[2] are shared.
Note: [1]: Power supply functions
[2]: DVM function
[3]: Pulse current measurement function
[4]: Long Integration current measurement function
RecallSetup
There are 6 sets of save/recall memories.
Rst/ SAV0 to SAV4
Parameter Setttings
IntRate
Use the arrow keys to select IntRate and press Enter. Input the parameters and press Enter to save.
Range: 0.01 to 10.00
AverRead[1][2]
Use the arrow keys to select
AverRead[1][2] and press Enter. Input
the parameters and press Enter to save.
Range: 1 to 10
42
BASIC OPERATION
Recall Setup
Use the arrow keys to select
RecallSetup and press Enter. Use the
arrow keys select a stored setup. Press Enter to confirm the recall.
See page 60 for further details.
Operation
The unit switches to DWM mode automatically when the DVM inputs are used. Using the DVM
meter function doesn’t affect the
operation of the power supply. The DMV function works with the output on or off.
Connection
For the connection details for the front and rear terminals, please page 28.
Description
Changes in the load current allows us to measure the pulse current.
There are three ways that pulse current can be measured:
1. Measuring the peak current over a single cycle (High Measurement).
2. Measuring the trough current over a single cycle (Low Measurement).
3. Measuring the average current over a single cycle (Average Measurement).

Pulse Current Measurement

43
PPH-1503 User Manual
The high and average measurements are triggered by the rising edge of the pulse current are performed for the time specified for the measurement.
Low measurement is triggered by the falling edge of the pulse current.
Note: Pulse current measurement is only valid up
to 5A.
Parameter Description
IntTime
Integration Time. The integration measurement time
can be set to automatic or to one of the manual settings (High Time, Low Time, Aver Time).
When the integration
measurement time is set to automatic mode, the system will measure the peaks and troughs of the pulse current and will automatically set an appropriate intergration time. The average integration time is the time of all the accumulated peaks and troughs. After the setting the integration time to automatic, the setting will apply to all subsequent pulse measurements, unless the automatic integration mode is applied again or the integration time is manully set. The automatic
44
BASIC OPERATION
Integration time can automatically detect pulses in the 80uS to 833mS range.
The manual time range setting is
33uS to 833333uS. The default units are in microseconds (uS). The unit will automatically round down the last two digits to 00, 33 or 66 micoseconds. For example, for a value of 65.999, it will be rounded down to 33 and 66.01 will be rounded down to 66.
TrigDelay
Trigger Delay When a pulse is detected, there
will be a 25us code execution delay time. The trigger delay settings are used to filter out the current overshoot. Measurement will begin from after the trigger delay time. The trigger delay setting range is: 0~0.10000S, with a resolution of 0.00001S. The setting units are in seconds.
AverRead[3]
Average Reading Count: Reads
back the average number of displayed values.
This parameter is only applicable
for pulse current measurement. The average number range can set from 1 ~ 100 with a resolution of 1.
45
PPH-1503 User Manual
TrigLeve[3]
Trigger Level. To avoid false pulse
measurements, the trigger level can be set close to the current amplitude. All noise and transient currents that are below the trigger level will be ignored. The trigger level has a setting range of 0 to 5A, with a resolution of 5mA. The setting unit for the trigger level is in amps (A). This setting is only valid for pulse measurements.
RecallSetup
Recalls stored settings. A total of 6 settngs can be recalled:
RST/SAV0 to SAV4
Parameter Settings
IntTime
Use the arrow keys to select IntTime, press Enter and then use the arrow keys to select the type of integration time that you want to set (Hight Time, Low Time, Aver Time). Press Enter again to set the integration time. After inputing the integration time, press Enter to return to the pulse current measurement menu. If
Auto Time was selected, press Enter to
return to the pulse current measurement menu.
Example:
High Time 33uS: IntTime

Hight Time

using the
46
BASIC OPERATION
numberpad enter 33

.
The time range can be set between 33uS and 833333uS. The setting units are in microseconds (uS).
TrigDelay
Use the arrow keys to select TrigDelay, press Enter and input the delay. Press Enter again to confirm.
The TrigDelay has a settable range of 0~0.10000S. The setting units are in seconds (S).
AverRead[3]
Use the arrow keys to select
AverRead[3], and then press Enter. Input
the AverRead number and then press Enter again to confirm.
The AverRead setting has a settable range of 1~00.
TrigLeve[3]
Use the arrow keys to select TrigLeve[3] and press Enter. Input the trigger level and press Enter again to confirm.
The TrigLeve parameter has a settable range of 0~5.000A. The setting units are in amperes (A).
RecallSetup
Use the arrow keys to select RecallSetup and press Enter to go the Recall Setup menu. Use the arrow keys to a setup. Press the Enter key to confirm. See page 60 for further details.
47
Panel Operation
Output
Press the Ouput key. When the Output key is lit, pulse current measurement is active.
When no pulse current is detected, NO PULSE will be dispalyed in red on the LCD screen. The unit will wait until the next pulse is detected.
The measurement settings can be edited during measurement. The H, L, A keys on the keypad can be used to perform fast-switching between measurement modes.

HIGH

LOW

AVER
Note
The LCD display will indicate which measurement mode is active with an orange line under the active mode for HIGH, LOW or AVER.

HIGH
NO PULSE
PPH-1503 User Manual
48
BASIC OPERATION
Description
The long current integration measurement function measures the mean (average) current over a single or multiple current pulses. The long integration time period must be a full period or integer multiples of a complete period of the measured pulse current. The Long integration measurement calculates the whole integration time as an integer number of integration cycles. An integration cycle is the line cycle period plus the data processing time. For example, if the line frequency is 60Hz, then a single integration cycle is 16.7mS, if the frequency is 50Hz, then a single integration cycle is must be 20mS. Long integraton is one of the methods to extend A/D circuits to exceed beyond their maximum integration time. The A /D conversion circuits can measure a pulse of up to 833 mS. Long integration measurement extends the A/D integration time to achieve a longer pulse measurement. This can extend the measurement time for long integration to a maximum of 60S.
Note: When this feature is used, the current range is
set to 5A.
Long integration
49
Parameter Description
IntTime
Integration time The integration time can be set
manually or automatically by the operator. For manual settings, the integration time can be set to a maximum of 60 seconds. For a line frequency of 60Hz the minimum integration time is 850mS with a step resolution of 16.7mS. For a line frequency of 50Hz, the minimum integration time is 840mS with a step resolution of 20mS.
When the integration time is set to Auto
Time, the system will automatically measure the time between two adjacent rising edges and an appropriate integration time is set for the peak and trough. If there are more than two pulses, the integration time must be set manually.
TrigEdge
Trigger edge Pulse edges are used to trigger long
integration measurement. Regardless of whether a rising or falling edge is used as a trigger, a pulse must furst be detected before measurement can start. Measurement can also start without an edge trigger. When Trig On Neither is selected, measurement starts as soon as the output is turned on.
PPH-1503 User Manual
50
BASIC OPERATION
Timeout
Pulse timeout When long integration measurement is
selected and the unit doesn’t detect a
pulse after a certain amount of time (pulse timeout time), the “No Pulse” message will be displayed on the LCD. This function is only applicable if rising or falling edge is selected as the edge trigger; the Trig On Neither trigger setting has no pulse timeout. The pulse timeout has a range of 1~63 seconds.
TrigLeve[4]
Trigger level. When the rising or falling edge trigger is
selected for long integration current measurement, a pulse must first be detected. The trigger level refers to minimum pulse level required for a pulse to be detected. For example if the trigger level is set to 2A, pulses that are
≤2A will be detected. Pulses <2A will be
ignored. The trigger level range is 0~5A. This setting only applies to long current integration measurements.
RecallSetup
Recalls pre-saved setups. A total of 6 setups can be recalled: RST/SAV0 ~ SAV4. See page 60 for details.
51
Parameter Settings
IntTime
Use the arrow keys to select IntTime then press Enter. Use the arrow keys to select a time setting.
If SetTime was selected, press Enter and then enter the long current integration time. Press Enter to save the setting and return to the long integration measurement menu.
If AutoTime was selected, press Enter to confirm and to go back to the long integration measurement menu.
For manually set integration times, if the set time is not an integer multiple of the integration cycle time, the system will automatically round down to closest maximum integer multiple that can be set.The time range is 850mS to 60S (50Hz) and 840mS to 60S (60Hz). The default unit is seconds (S).
TrigEdge
Use the arrow keys to select TrigEdge and then press Enter. Use the arrow keys to select the type of trigger and press Enter to confirm the trigger settings and to return to the long integration measurement menu.
Timeout
Use the arrow keys to select Timeout and then press Enter. Enter the timeout settings and press Enter again to confirm and to return back to the long integration measurement menu. The time range is 1~63S. The default unit is seconds (S).
PPH-1503 User Manual
52
BASIC OPERATION
TrigLeve[4]
Use the arrow keys to select TrigLeve[4] and then press Enter. Key in te trigger level setting and press Enter again to confirm and to return back to the long integration measurement menu. The trigger level range is 0~5A. The default is amps (A.)
RecallSetup
Use the arrow keys to select RecallSetup and press Enter to go to the Recall Setup menu. Use the arrow keys to select a saved setup and press Enter again to confirm. See page
60 for details.
Operation
Output
Press the Output key. When the Output key is lit, pulse current measurement is active.
When no pulse current is detected, NO PULSE will be dispalyed in red on the LCD screen. The unit will wait until the next pulse is detected.
NO PULSE
53
PPH-1503 User Manual
Function Description
When the test circuit is an active circuit, and the manifested voltage in the test circuit is greater than the output voltage of the power supply, the power supply will automatically disipate current from the external power supply. When this function is in the normal operating state, the power supply outputs the setting voltage, which is equivalent to a constant voltage load rather than constant current load.
The current disipation from the power supply output flows from the positive terminal out to the negative terminal. The amount of current sunk is not controlled from the power supply.
Connection
Connect the positive terminal of the external power supply to the positive terminal on the high-speed power supply. Connect the negative terminal of the external power supply to the negative terminal on the high-speed power supply.

Current Sink Function

54
Conditions
To protect the high-speed power supply when operating as a current sink, the following two conditions must be met:
1. Ensure that the voltage of the external power
supply is greater than the output of the high­speed power supply voltage by 0.3V~2.5V. The voltage difference depends on the high-speed power supply voltage output and the load conditions.
2. To ensure that the high-speed power supply
output voltage is within the range of 0~5V, the current draw cannot exceed 2A. For output voltgages between 5V~15V, the current draw must be reduced by 0.1A for each 1V increase over 5V. See the formula in the table below for the details.
High-speed Power Supply Output Voltage
Maximum Dissipation Current
0~5V
2A
5V~15V
2A-((0.1A/V)*(output voltage - 5V))
BASIC OPERATION
55
PPH-1503 User Manual
Function Description
When the Relay control feature is turned on, it is synced to the current limit of the power suppy. The external relay control is divided into two different types, a limit relay and a trip relay.
The limit relay is used in conjuction with CC mode. When the constant current setting value is reached, the relay control signal will go high and will return back to the low level when the current level goes back below the constant current setting.
The trigger relay is used in conjuction with CC mode. When the constant current setting value is reached, the relay control signal will go high and the output is disabled. When the output goes back on and the current is less than the current setting value, the relay control signal will back to the low level.
Rear Panel Control Interface
The rear panel control interface has five terminals, +5V, IN(software upgrade), OUT (outputs the control signals) and GND (connected to the chasis ground or earth ground), respectively.
Wiring Method
A thin screwdriver or similar tool will need to be inserted into the release mechanism (highlighted in orange in the figure above) to open the terminals. Insert an exposed wire into the terminal and release the mechanism to lock the wire into place.

External Relay Control

56
Schematic Diagram for Relay Control
Limit Relay:
Current
Limit setting
on
off
Relay
Trip Relay:
Current
Limit setting
on
off
Relay
Output turns off due to Trip Mode
User turns output on,and trip condition is corrected.Otherwise,output will trip again.
BASIC OPERATION
57
PPH-1503 User Manual
External Relay Connection
There are two ways to connect an external relay to the unit:
1. Using the the +5VDC relay output to drive an
external relay. Ensure the current doesn’t
exceed 150mA.
+5VDC
OUT
OUT
IN
GND
Power Supply Relay Control
Relay
Internal Source +5VDC
±5%
Protection Diode
Warning: Do not short the 5VDC terminal to the chasis, earth or to the control port GND, otherwise it may damage the unit.
2. Using an external power source to drive the
external relay. The voltage of the source cannot exceed 15V and the current cannot exceed 150mA.
+5VDC
OUT
OUT
IN
GND
Power Supply Relay Control
Relay
Internal
Source +5VDC
±5%
Protection Diode
External Source
15VDC Max
58
SAVE/RECALL
Description
Five groups of system settings are available.
Parameter data
Listed below are the settings that are available for each group (Rst is shown as an example).
Voltage: 00.500V
CurrRange: 5A
Current: 2.0000A
IntRate: 1.00PLC
OutputState: Off
AverRead[1][2]: 1
DispType: Actual V and I
O.V.P: Off
GPIBAddr: 16
LimMode: Limit
GPIBFormat: Exponential
RelayControl: Zero
HighTime: 33us
AverRead[3]: 1
LowTime: 33us
TrigDelay: 0.10000
AverTime: 33us
TrigLevel[3]: 0.000A
IntTime: 1.000s
TrigEdge: Rising
Timeout: 16s
TrigLevel[4]: 0.000A
Operation
Press the Menu key to enter the main menu interface.
Use the up and down arrow to select the Save Setup option.
Press Enter to go to the Save Setup menu.


SAVE/RECALL

Save Settings

59
PPH-1503 User Manual
Use the left and right arrow keys to select the desired save memory. There are five selections: SAV0, SAV1, SAV2, SAV3, SAV4.
Press the Enter key to save the settings and return to the main interface.

Result
The current settings on the unit will be saved to one of the memory locations (SAV0~SAV4)
Description
There a total of 6 different memory settings that can be recalled: Rst, SAV0, SAV1, SAV2, SAV3, SAV4, SAV5.
Operation
There are two methods to recall the setup settings.
Method 1:
Use the arrow keys to select Recall Setup via F1, F2, F3 or F4 from the display interface

Press the Enter key to enter the Recall Setup interface.

Use the left and right arrow keys to select a setup to recall (Rst, SAV0 ~ SAV4).
Press Enter to confirm and to return to the main interface.


Recall Settings

60
SAVE/RECALL
Method 2:
Press the Menu key
Use the up/down arrow keys to select Recall Setup.
Press Enter to enter the Recall Setup interface.

Use the left and right arrow keys to select which setting to recall.
Press the Enter key to confirm the selection.

Power On Settings
In the main menu, the interface parameter settings area shows PowOnSetup settings. There are 11
PowOnSetup settings to choose from, Rst,
SAV0~SAV4 and SAV5~SAV9.
The main difference between SAV0~SAV4 and SAV5~SAV9 is that SAV0~SAV4 are user saved settings and don’t contain the Power On/Off state (Output State is always off) while the SAV5~SAV9 contain the Power On/Off state. The SAV5~SAV9 settings are synced from corresponding SAV0~SAV4 settings and are identical but for the the Power On/Off state. The Power On/Off states simply indicate if the Output State can be on or off.
The relationship between SAV0~SAV4 and SAV5~SAV9 is as follows:
SAV0 SAV5
SAV1 SAV6
61
PPH-1503 User Manual
SAV2 SAV7
SAV3 SAV8
SAV4 SAV9
62
SAVE/RECALL
Description
The system can retrieve the factory default settings by loading the Rst setting. This setting cannot be modified.
Operation
There are two methods to retrieve the factory default settings. Please see the Recall Settings sections for instructions (page 60).

Restore Factory Default Settings

63
PPH-1503 User Manual
Description
The System Information menu can be used to view the system information or to perform system operations such as set the buzzer function, backlight display brightness or set to the factory conditions.
System Information Items
System Version
View the system software version.
Serial Number
View the machine serial number.
Calibration Unit
Calibration menu. Factory use only.
Utility
System setting items: Buzzer setttings, backlight brightness settings and factory restore.
Operation
Press the Menu key and select System Information. Press the Enter key to enter the System Information menu.

SYSTEM SETTINGS

System Information

64
SYSTEM SETTINGS
Description
There are two utility settings: buzzer settings and backlight brightness settings.
Setting Information
Beep
Sets the when the buzzer is turn on.
BackLight
Adjust the LCD brightness.
Buzzer Operation
In the Utility menu, use the up and down arrows to select Beep.
Press Enter and then set the buzzer state to on or off. When the buzzer is set to on, Beep On will be displayed.

Beep
On
Press the Menu key to exit and return to the main interface. The buzzer status will be displayed on the LCD.
BEEP
Backlight Brightness Adjustment
In the Utility menu, use the up and down arrow keys to select BackLight.
Press Enter to toggle the backlight brightness level. The brightness level is displayed under BackLight. There are three brightness levels: High, Middle, Low.

BackLight
Middle
Press the Menu key to exit and return to the main interface.

Utilty Settings

65
PPH-1503 User Manual
Restore to Factory Settings
In the Utility menu, use the up and down arrow keys to select In factory reset, then press the Enter key to restore to the factory settings. This function is only for factory use.
66
REMOTE CONTROL
Description
The PPH-1503 can be connected via USB using the USB Communications Device (CDC) class.
Interface
Rear panel USB slave port.
Installing the Driver
Before connecting the unit to the USB port of the PC, make sure the approriate driver has been installed. The driver is available from the GW Instek website. When the unit has successfully connected to the PC via USB, USB will be displayed in the status bar in red.
USB
The front panel keys are automatically locked when the unit is in remote mode.
LOCK

REMOTE CONTROL

Remote Control

USB
67
PPH-1503 User Manual
COM port settings
The following settings should be set to the following:
Baud rate: 115200 or less Parity: None Data bits: 8 Stop bits: 1 Data overflow control: None
Function Check
Perform the following query:
*IDN?
The unit will return the manufacturer, model, serial number and software version.
GW INSTEK, PPH-1503, SN: xxxxxxxx, Vx.xx
Disabling Remote Control Mode
Send a remote command to exit from
remote control mode from the PC or long-press the unlock key on the front panel to exit from the remote control mode. The RMT icon in the status bar will become grey when you exit from remote control mode.
The LOCK icon in the status bar will
also turn grey.
Unplug the USB cable from the rear
panel.
RMT
LOCK
Note: USB devices are hot-plug devices. You can directly remove the cable and exit.
68
REMOTE CONTROL
Description
The GPIB remote control can be set from the interface menu. The communication data format, compatibility settings and and address must all be configured before using GPIB remote control.
Interface
Rear panel GPIB port.
Connection
When the unit has been successfully connected via GPIB, GPIB will appear in the status bar in red. The panel key will also be locked.
GPIB
The front panel keys are automatically locked when the unit is in remote mode.
LOCK
Communication Data Format
There are four data formats to select from: Exponential, 2DPS, 3DPS and 4DPS.
Steps
A. Press the Menu key to enter the main menu.
B. Use the up and down arrow keys to select Interface.
C. Press Enter to enter the Interface menu.

D. Use the up and down arrow keys to select GPIB.
E. Press Enter to enter the GPIB menu.


GPIB

69
PPH-1503 User Manual
F. Use the up and down arrows to select Output Format.
G. Press the Enter key to toggle between the different output formats.

H. Press the Menu key to exit and return to the main menu.
Output Formats
There are two different output formats to select from: KEITHLEY 2303 and FLUKE PM2811.
Steps
Follow steps A~E in the previous section, above.
F. Use the up and down arrow keys to select Output Type.
G. Press the Enter key to toggle between each of the output formats.

H. Press the Menu key to exit and return to the main menu.
Setting the GPIB Address
Configuring the GPIB address for connection to a PC.
Steps
Follow steps A~E in the previous section, above.
F. Use the up and down arrow keys to select Primary Address.
G. Press the Enter key and then set the GPIB address. Press the Enter key again to confirm. The address range is 1~30.

70
REMOTE CONTROL
H. Press the Menu key to exit and return to the main menu.
Exiting from Remote Control Mode
Send a remote command to exit from
remote control mode from the PC or long-press the unlock key on the front panel to exit from the remote control mode. The RMT icon in the status bar will becomes greyed-out when you exit from remote control mode.
The LOCK icon becomes greyed-out
when the panel becomes unlocked.
Unplug the connection from the rear
panel.
RMT
LOCK
71
PPH-1503 User Manual
Description
When using the LAN interface a number of settings must be turned on.
IP Mode
The IP address can be configured using either DHCP, Auto IP or Manual IP. Using DHCP to get an IP address automatically assigned. The system will use AUTO IP to obtain an automatically generated IP address to avoid IP address conflicts.
Manu IP
A. Press Menu to enter the main menu.
B. Use the up and down arrow keys to select Interface.
C. Press Enter to enter the Interface menu.

D. Use the up and down arrow key to select LAN.
E. Press Enter to enter the LAN menu.

F. Use the up and down arrow keys to select IP Mode.
G. Press Enter to select Manu IP.

H. Use the up and down arrow key to select the appropriate parameters.
I. Press Enter and then configure each of the parameters.

LAN
72
REMOTE CONTROL
J. Press Enter to confirm each of the configurations.

K. Press the Menu key to exit and return to the main menu.
Parameter Settings:
IP Address: IP address range: 1.0.0.0 to
223.255.255.255 (excluding 127.nnn.nnn.nnn).
Subnet Mask: Subnet Mask Range: 1.0.0.0 to
255.255.255.255.
Gateway: Gateway range: 1.0.0.0 to 223.255.255.255 (excluding 127.nnn.nnn.nnn).
DNS Servers: DNS Server range: 1.0.0.0 to
223.255.255.255 (excluding 127.nnn.nnn.nnn).
DHCP
Follow steps A~F in the previous section, Manun IP, above.
G. Press Enter to select DHCP. The unit will be assigned an IP address, subnet mask, the default gateway and other network parameters from the DHCP server. The corresponding parameters will be shown in the parameter area. Use the arrow keys to view the settings (When an IP address is being assigned, a circular scanning icon will appear).

H. Press the Menu key to exit and return to the main menu.
73
PPH-1503 User Manual
Auto IP
Follow steps A~F in the previous section, Manu IP, above.
G. Press the Enter key and select Auto IP. The device will automatically obtain an IP address and subnet address mask based on the current network configuration.The unit will set the IP address in the range of
169.254.0.1 to 169.254.255.254 with a subnet of 255.255.0.0. The parameters will be displayed in the parameter area. Use the arrow keys to view the parameters.

H. Press the Menu key to exit and return to the main menu.
PC Operation
1. Enter the IP address into Microsoft Internet Explorer (IE). After entering the IP address you will be shown the Welcome screen which displays the instrument information. The page also provides three links: Welcome Page, Browser Web Control and View & Modify Configuration (network settings).
74
REMOTE CONTROL
2. Click on “Browser Web Control” to execute commands through the browser, as shown below.
3. Press the “View & Modify Configuration” icon to enter the Modify Config menu, as shown below.
4. Click “Modify Config” to enter the network
configuration setting menu, as shown below. Use
the mouse to click on “Save and Restart” to change
the remote settings for the PPH-1503.
75
PPH-1503 User Manual
Note
Click “Undo Edits” to cancel all the edited settings.
Click “Factory Defaults” to restore to the factory
default settings.
Exiting from Remote Control Mode
Send a remote command to exit from
remote control mode from the PC or long-press the unlock key on the front panel to exit from the remote control mode. The RMT icon in the status bar will become greyed-out when you exit from remote control mode.
The LOCK icon will be greyed-out
when the panel lock is disabled.
Unplug the connection from the rear
panel.
RMT
LOCK
Note: Hot-swappable LAN devices can be directly disconnected to exit.

Command Syntax

The commands that are used with the PPH-1503 meet IEEE488.2 and SCPI standards.
SCPI Commands Overview SCPI
Command Format
SCPI is an ASCII based command language designed for test and
measurement instruments. SCPI commands uses a hierarchical structure
(tree system), and is divided into different subsystems. Each subsystem
76
REMOTE CONTROL
is defined by a different root keyword. Each command consists of a root
keyword and one or more hierarchial key words separated by a colon “:”
and followed by a parameter. There is always a space between the
keywords and the parameters. Any commands followed by a question
mark (?) are queries.
For Example:
:SYSTem:BEEPer:STATe {0|1|OFF|ON}
:SYSTem:BEEPer:STATe?
SYSTem is the root level keyword and BEEPer and STATe are the
secondary and tertiary level keywords. All all levels have a “:”
separating each keyword. Parameters are enclosed in “{ }”.
The commands SYSTem:BEEPer:STATe has {0|1|OFF|ON } as
parameters. The parameters are separated with a space.
SYSTem:BEEPer:STATe? Indicates that the command is a query.
In addition some commands have multiple parameters that are usually
separated by a comma “,”. For example: :STATus:QUEue:ENABle (-
110:-222, -220).
Symbol Desription
SCPI commands have the following convential symbols. These symbols
are not commands but are used to describe the command parameters.
1. Curly Brackets { }
Curly Bracket enclose command string parameters, for example:
{ OFF|ON }
2. Verical Bars
77
PPH-1503 User Manual
Vertical bars are used to separate one or more optional parameters. Only
one command can be selected; With the following two parameters,
{ON|OFF} only ON or OFF can be selected.
3. Square Brackets [ ]
The contents inside square brackets represent keywords or parameters
that can be omitted when executing a command. For example: For the
commands :OUTPut[:STATe] {ON|OFF}, [STATe] can be omitted.
4. Angle Brackets
The parameters in angle brackets must be substituted with a valid
parameter. For example: For the command :DISPlay: CONTrast
<brightness>, <brightness> must be use a numerical value instead such
as, :DISPlay:CONTrast 1
Parameter Types
The commands have a number of different parameter categories. How
the parameters are set depend on the parameter categories.
1. Boolean
Commands parameter that have to states “OFF” and “ON”, for example,
DISPlay:FOCUs {ON|OFF}. “ON” will turn on the focus display
function, while “OFF” will turn it off.
2. Consectutive Integers
78
REMOTE CONTROL
Parameters that use consecutive integers, for example: For the
command :DISPlay:CONTrast <brightness>, <brightness> is an integer
value with a range of 1~3.
3. Continuous Real Number
Parameter that must be a continuous real number can have any value
within the effective range and accuracy. For example: The command
CURRent {<current>|MINimum|MAXimum}, is used to set the current
value for the current operating channel. <current> can be any value
within the setting range of the current channel.
4. Discrete
For discrete parameters, only those values that are listed can be used.
For example: The *RCL{0|1|2|3|4|5} command can only use 0, 1, 2, 3, 4,
5.
5. ASCII Strings
ASCII string parameters must use a combination of ASCII characters in a
string. For example: For the command :MODE <name>, <name> must
be an ASCII string.
Command Abbreviations
The syntax for SCPI commands contain a combination of upper and
lower case letters. The upper case letters in a command represent the
short form of that command.
Commands are not case sensitive and can used in both upper and
lower case. Note, however, to use the short form of the command, only
the capital letter part of the command can be used (no other abbreviation
can be used). For example:
79
PPH-1503 User Manual
:FETCh?
Page 86
:FETCh:ARRay?
Page 86
:READ?
Page 86
:READ:ARRay?
Page 86
:MEASure[:<function>]?
Page 87
:MEASure:ARRay[:<function>]?
Page 87
:DISPlay:ENABle <b>
Page 88
:DISPlay:ENABle?
Page 88
:DISPlay[:WINDow[1]]:TEXT:STATe <b>
Page 88
:DISPlay[:WINDow[1]]:TEXT:STATe?
Page 89
:DISPlay[:WINDow[1]]:TEXT:DATA <a>
Page 89
:MEASure:CURRent?
Can be abbreviated to:
:MEAS:CURR
Command Terminators
When sending a command to the function generator, the command must be terminated with a <new line> character. The IEEE-4888 EOI can also be used as a <new line> character. A command can also be terminated using a carriage return + <new line> character. The command path will always be reset back to the root level after a command has been terminated.
Return values are terminated with 0x0A.

Command List

Measurement Instructions

Display Functions

80
REMOTE CONTROL
:DISPlay[:WINDow[1]]:TEXT:DATA?
Page 89
DISPlay:CONTrast<NRf>
Page 89
:FORMat[:DATA] <type>
Page 90
:FORMat[:DATA]?
Page 90
:FORMat:BORDer <name>
Page 90
:FORMat:BORDer?
Page 90
:OUTPut[:STATe] <b>
Page 91
:OUTPut[:STATe]?
Page 91
:OUTPut:RELay <name>
Page 91
:OUTPut:RELay?
Page 91
:OUTPut:OVP:STATe <b>
Page 92
:OUTPut:OVP:STATe?
Page 92
:OUTPut:OVP <value>
Page 92
:OUTPut:OVP?
Page 92
:[SOURce]:CURRent[:LIMit][:VALue] <NRf>
Page 93
:[SOURce]:CURRent[:LIMit][:VALue]?
Page 93
:[SOURce]:CURRent[:LIMit]:TYPE <name>
Page 93
:[SOURce]:CURRent[:LIMit]:TYPE?
Page 94
:[SOURce]:CURRent[:LIMit]:STATe?
Page 94
:[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <n>
Page 94
:[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Page 94

Data Format Commands

Output Commands

Source Commands

81
PPH-1503 User Manual
:SENSe[1]:FUNCtion <name>
Page 95
:SENSe[1]:FUNCtion?
Page 95
:SENSe[1]:NPLCycles <n>
Page 95
:SENSe[1]:NPLCycles?
Page 95
:SENSe[1]:AVERage <NRf>
Page 96
:SENSe[1]:AVERage?
Page 96
:SENSe[1]:CURRent[:DC]:RANGe[:UPPer] <n>
Page 96
:SENSe[1]:CURRent[:DC]:RANGe[:UPPer]?
Page 96
:SENSe[1]:CURRent[:DC]:RANGe:AUTO <b>
Page 97
:SENSe[1]:CURRent[:DC]:RANGe:AUTO?
Page 97
:SENSe[1]:PCURrent:AVERage <NRf>
Page 97
:SENSe[1]:PCURrent:AVERage?
Page 98
:SENSe[1]:PCURrent:MODE <name>
Page 98
:SENSe[1]:PCURrent:MODE?
Page 98
:SENSe[1]:PCURrent:TIME:AUTO
Page 98
:SENSe[1]:PCURrent:TIME:HIGH <NRf>
Page 99
:SENSe[1]:PCURrent:TIME:HIGH?
Page 99
:SENSe[1]:PCURrent:TIME:LOW <NRf>
Page 99
:SENSe[1]:PCURrent:TIME:LOW?
Page 99
:SENSe[1]:PCURrent:TIME:AVERage <NRf>
Page 100
:SENSe[1]:PCURrent:TIME:AVERage?
Page 100
:SENSe[1]:PCURrent:SYNChronize[:STATe] <b>
Page 100
:SENSe[1]:PCURrent:SYNChronize[:STATe]?
Page 101
:SENSe[1]:PCURrent:SYNChronize:DELay <NRf>
Page 101
:SENSe[1]:PCURrent:SYNChronize:DELay?
Page 101
:SENSe[1]:PCURrent:SYNChronize:TLEVel<NRf>
Page 101
:SENSe[1]:PCURrent:SYNChronize:TLEVel?
Page 102

Readback Commands

82
REMOTE CONTROL
:SENSe[1]:LINTegration:TIME <NRf>
Page 102
:SENSe[1]:LINTegration:TIME
Page 102
:SENSe[1]:LINTegration:TIME:AUTO
Page 102
:SENSe[1]:LINTegration:TLEVel <NRf>
Page 102
:SENSe[1]:LINTegration:TLEVel
Page 103
:SENSe[1]:LINTegration:TEDGe <name>
Page 103
:SENSe[1]:LINTegration:TEDGe
Page 103
:SENSe[1]:LINTegration:TimeOUT <NRf>
Page 103
:SENSe[1]:LINTegration:TimeOUT
Page 104
:SENSe[1]:LINTegration:SEARch <b>
Page 104
:SENSe[1]:LINTegration:SEARch?
Page 104
:SENSe[1]:LINTegration:FAST <b>
Page 104
:SENSe[1]:LINTegration:FAST?
Page 105
:STATus:PRESet
Page 105
:STATus:OPERation[:EVENt]?
Page 105
:STATus:OPERation:CONDition?
Page 105
:STATus:OPERation:ENABle <NRf>
Page 106
:STATus:OPERation:ENABle?
Page 106
:STATus:MEASurement[:EVENt]?
Page 106
:STATus:MEASurement:ENABle <NRf>
Page 106
:STATus:MEASurement:ENABle?
Page 107
:STATus:MEASurement:CONDition?
Page 107
:STATus:QUEStionable[:EVENt]?
Page 107
:STATus:QUEStionable:CONDition?
Page 107
:STATus:QUEStionable:ENABle <NRf>
Page 108

Status Commands

83
PPH-1503 User Manual
:STATus:QUEStionable:ENABle?
Page 108
:STATus:QUEue[:NEXT]?
Page 108
:STATus:QUEue:ENABle <list>
Page 108
:STATus:QUEue:ENABle?
Page 109
:STATus:QUEue:DISable <list>
Page 109
:STATus:QUEue:DISable?
Page 110
:STATus:QUEue:CLEar
Page 110
:SYStem:LOCal
Page 117
:SYStem:REMote
Page 116
:SYSTem:COMMunicate:LAN:DHCP[:STATe] <b>
Page 112
:SYSTem:COMMunicate:LAN:DHCP[:STATe]?
Page 112
:SYSTem:COMMunicate:LAN:IPADdress <IP address>
Page 113
:SYSTem:COMMunicate:LAN:IPADdress?
Page 113
:SYSTem:COMMunicate:LAN:AUTOip[:STATe] <b>
Page 113
:SYSTem:COMMunicate:LAN:AUTOip[:STATe]?
Page 114
:SYSTem:COMMunicate:LAN:SMASk <Mask>
Page 114
:SYSTem:COMMunicate:LAN:SMASk?
Page 114
:SYSTem:COMMunicate:LAN:GATEway <IP address>
Page 114
:SYSTem:COMMunicate:LAN:GATEway?
Page 115
:SYSTem:COMMunicate:LAN:DNS <Address>
Page 115
:SYSTem:COMMunicate:LAN:DNS?
Page 115
:SYSTem:COMMunicate:LAN:MANualip[:STATe] <b>
Page 116
:SYSTem:COMMunicate:LAN:MANualip[:STATe]?
Page 116
:SYSTem:COMMunicate:LAN:APPLy
Page 116
:SYSTem:VERSion?
Page 110
:SYSTem:ERRor?
Page 110

System Commands

84
REMOTE CONTROL
:SYSTem:CLEar
Page 111
:SYSTem:LFRequnecy?
Page 111
:SYSTem:POSetup <name>
Page 111
:SYSTem:POSetup?
Page 112
:SYSTem:BEEPer:STATe {0|1|OFF|ON}
Page 116
:SYSTem:BEEPer:STATe?
Page 117
*IDN?
Page 118
*RST
Page 118
*TST?
Page 118
*WAI
Page 119
*CLS
Page 128
*ESE<Enable Value>
Page 127
*ESE?
Page 127
*ESR?
Page 127
*OPC
Page 128
*OPC?
Page 128
*SRE <Enable Values>
Page 126
*SRE?
Page 126
*STB?
Page 127
*TRG
Page 119
*SAV <NRf>
Page 119
*RCL <NRf>
Page 119

System Related Commands

IEEE488.2 Common Commands IEEE488.2

Command Details

85
PPH-1503 User Manual
Command
:FETCh?
Function
Returns the last readback value.
Response Time
Maximum: 16ms.
Example
:FETCh?
Returns the last readback value.
Command
:FETCh:ARRay?
Function
Returns the last array readback values.
Response Time
Maxium: 16ms
Example
:FETCh:ARRay?
Returns the last array readback values
Command
:READ?
Function
Triggers a read operation and returns the read values.
Response time
Maximum: 32ms
Example
:READ?
Returns the read values.
Command
:READ:ARRay?
Function
Triggers a new array. Returns the read array values.
Response time
Max: 32ms
Example
:READ:ARRay?
Triggers a new array. Returns the read array values.

Measurement Commands

86
REMOTE CONTROL
Command
:MEASure[:<function>]?
Function
Performs a “READ?”query on the specified measurement function.
Description
<function>
CURRent[:DC]Measures the current.
VOLTage[:DC]Measures the voltage.
PCURrentMeasures the pulse current.
DVMeterMeasures the DVM input.
LINTegrationLong integration current measurement.
For pulse current and long integration current measurement, if there is no pulse, test for the timeout time.
Response time
Maximum: 32ms
Example
:MEASure: CURRent?
Sets pulse current as the measurement type and reads back the pulse current value.
Command
:MEASure:ARRay[:<function>]?
Function
Performs a “READ:ARRay?”query on the specified
measurement function.
Description
<function>
CURRent[:DC]: Measures the current.
VOLTage[:DC]: Measures the voltage.
PCURrent: Measures the pulse current.
DVMeter: Measures the DMV input.
LINTegration: Long integration current measurement
87
PPH-1503 User Manual
For pulse current and long integration current measurement, if there is no pulse, test for the timeout time.
Response time
Maximum: 32ms
Example
:MEASure:ARRay:PCURrent?
Sets the measurement type to pulse current array measurement and returns the read array value.
Command
:DISPlay:ENABle <b>
Function
Turn the LCD display on or off.
Description
b
0/OFF: Turns the display off.
1/ON: Turns the display on.
Example
:DISPlay:ENABle ON
Turns the LCD display on.
Command
:DISPlay:ENABle?
Function
Queries the state of the display.
Example
:DISPlay:ENABle?
Returns the state of the display.
Command
:DISPlay[:WINDow[1]]:TEXT:STATe <b>
Function
Enable or disable text message mode.
Description
<b>
0/OFF: Disable text message mode.
1/ON: Enable text message mode.
Example
:DISPlay: TEXT:STATe ON
Enables text message mode.

Display Commands

88
REMOTE CONTROL
Command
:DISPlay[:WINDow[1]]:TEXT:STATe?
Function
Returns the state of the text message mode.
Example
:DISPlay:TEXT:STATe?
Returns the state of the text message mode.
Command
:DISPlay[:WINDow[1]]:TEXT:DATA <a>
Function
Defines the ASCII text for display information “a”.
Description
<a>
ASCII string of block of up to 32 characters. Any character over 32 will be truncated. The characters are not case-sensitive.
Used when “:DISPlay:TEXT:STATe ON” is executed.
Example
:DISPlay:TEXT:DATA txt
The ASCII text is set to “txt”.
Command
:DISPlay[:WINDow[1]]:TEXT:DATA?
Function
Returns the text message that was set.
Example
:DISPlay:TEXT:DATA?
Returns the text message that was set.
Command
DISPlay:CONTrast < NRf >
Function
Sets backlight display brightness.
Description
<NRf>
1Weak
2Medium
3Strong
89
PPH-1503 User Manual
Example
DISPlay:CONTrast 3
Sets the backlight to the brightest.
Command
:FORMat[:DATA] <type>
Function
Sets the data format.
Description
<type>
ASCiiASCII format.
SREalIEEE754 single precision format.
DREalIEEE754 double precision format.
Example
:FORMat:DATA SREal
Sets the format to IEEE754 double precision format.
Command
:FORMat[:DATA]?
Function
Queries the data format.
Example
:FORMat:DATA?
Returns the data format.
Command
:FORMat:BORDer <name>
Function
Sets the byte order.
Description
name
NORMal: normal binary byte order.
SWAPped: reverse binary byte order.
Example
:FORMat:BORDer NORMal
Set the data format to the “Normal” binary byte
order.
Command
:FORMat:BORDer?

Measurement Commands

90
REMOTE CONTROL
Function
Queries the binary byte order.
Example
:FORMat:BORDer
Returns the binary byte order.
Command
:OUTPut[:STATe] <b>
Function
Turns the output on or off.
Description
<b>
0/OFF: Turn off the output
1/ON: Turn on the output
Example
:OUTPut:STATe ON
Turns on the output.
Command
:OUTPut[:STATe]?
Function
Queries the output state.
Example
:OUTPut:STATe?
Returns the output state.
Command
:OUTPut:RELay <name>
Function
Turns the external relay control signal on or off.
Description
<name>
ZEROOff
ONEOn
Example
:OUTPut:RELay ONE
Turn the relay signal on.
Command
:OUTPut:RELay?
Function
Queries the state of the output relay.

Output Commands

91
PPH-1503 User Manual
Example
:OUTPut:RELay?
Returns the state of the output relay.
Command
:OUTPut:OVP:STATe <b>
Function
Turns OVP protection on/off
Desciption
<b>
0/OFFTurns OVP off.
1/ONTurns OVP on.
Example
:OUTPut:OVP:STATe ON
Turn on OVP.
Command
:OUTPut:OVP:STATe?
Function
Queries the status of the OVP function.
Example
:OUTPut:OVP:STATe?
Returns the status of the OVP function.
Command
:OUTPut:OVP <value>
Function
Sets the OVP level.
Desciption
<value>
0.00-15.00
Example
:OUTPut:OVP 10.05
Sets the OVP voltage to 10.05V.
Command
:OUTPut:OVP?
Function
Queries the OVP voltage level.
Example
:OUTPut:OVP?
Returns the OVP voltage level.
92
REMOTE CONTROL
Command
:[SOURce]:CURRent[:LIMit][:VALue] <NRf>
Description
Sets the current level.
NRf
0.0000-5.0000
Example
:SOURce:CURRent 1.0005
Sets the current level to 1.0005A.
Command
:[SOURce]:CURRent[:LIMit][:VALue]?
Description
Queries the current limit level.
Example
:SOURce:CURRent?
Returns the current limit level.
Command
:[SOURce]:CURRent[:LIMit]:TYPE <name>
Function
Sets the current limit mode.
Description
<name>
LIMit: General limiting mode
TRIP: Output shutdown mode
LIMRELAY|LIMITRELAYGeneral limiting mode and external relay output control mode.
TRIPRELAY: Ouput shutdown mode and external relay output control mode.
Example
:SOURce:CURRent:TYPE LIMITRELAY
Sets the current limit mode to LIMITRELAY.

Source Commands

93
PPH-1503 User Manual
Command
:[SOURce]:CURRent[:LIMit]:TYPE?
Function
Queries the current limiting mode.
Example
:SOURce:CURRent:TYPE
Returns the current limiting mode.
Command
:[SOURce]:CURRent[:LIMit]:STATe?
Function
Queries the current limit state. Returns 0 if the current limit has not been reached, returns 1 if the current limit has been reached.
Example
:SOURce:CURRent:STATe?
Returns the current limit state.
Command
:[SOURce]:VOLTage[:LEVel][:IMMediate][:AMP Litude] < NRf >
Function
Sets the output voltage amplitude.
Description
<NRf>
0.000-15.000
Example
:SOURce:VOLTage 5.321
Sets the output voltage to 5.321V.
Command
:[SOURce]:VOLTage[:LEVel][:IMMediate] [:AMPLitude]?
Function
Queries the output voltage setting.
Example
:SOURce:VOLTage?
Returns the output voltage setting.
94
REMOTE CONTROL
Command
:SENSe[1]:FUNCtion <name>
Function
Selects the type of measurement function: voltage, current, pulse, long integration and DVM measurement.
Description
name
VOLTage: Voltage measurement.
CURRent: Current measurement.
PCURrent: Pulse current
measurement.
LINTegration: Long integration measurement.
DVMeter: DVM input measurement.
Example
:SENSe:FUNCtion VOLTage
Selects “Voltage” as the measurement type.
Command
:SENSe[1]:FUNCtion?
Function
Queries the type of measurement function.
Response time
Maximum: 16ms
Example
:SENSe:FUNCtion?
Returns the type of measurement function.
Command
:SENSe[1]:NPLCycles <n>
Function
Sets the number of PLCs for the integration rate for voltage, current and DVM measurements.
Description
<n>
0.01-10.00
Example
:SENSe:NPLCycles 0.10
Sets the number of PLCs to 0.1.

Readback Commands

95
PPH-1503 User Manual
Command
:SENSe[1]:NPLCycles?
Function
Returns the number of power line cycles used for the integration rate.
Example
:SENSe:NPLCycles?
Returns the number of PLCs.
Command
:SENSe[1]:AVERage <NRf>
Function
Sets the averaging number for the voltage, current and DVM measurements.
Description
<NRf>
1-10
Example
:SENSe:AVERage 3
Sets the averaging number to 3.
Command
:SENSe[1]:AVERage?
Function
Returns the average number.
Examle
:SENSe:AVERage?
Returns the average number.
Command
:SENSe[1]:CURRent[:DC]:RANGe[:UPPer] <n>
Description
Sets the current measurement range.
Description
<n>
MINlow range
MAX:high range
Example
:SENSe:CURRent:RANGe MIN
Sets the current range to low.
Command
:SENSe[1]:CURRent[:DC]:RANGe[:UPPer]?
96
REMOTE CONTROL
Function
Queries the current measurement range
Example
:SENSe:CURRent:RANGe?
Returns the current measurement range.
Command
:SENSe[1]:CURRent[:DC]:RANGe:AUTO <b>
Function
Turns the automatic range function.
Description
<b>
0/OFFTurn off.
1/ONTurn on.
Example
:SENSe:CURRent:RANGe:AUTO ON
Turns the automatic range function on.
Command
:SENSe[1]:CURRent[:DC]:RANGe:AUTO?
Function
Queries the state of the automatic range function.
Example
:SENSe:CURRent:RANGe:AUTO?
Returns the status of the automatic range function.
Command
:SENSe[1]:PCURrent:AVERage <NRf>
Function
Sets the averaging number for pulse current measurements.
Description
NRf
1-100 or 1-5000(pulse current digitization)
Example
:SENSe:PCURrent:AVERage 5
Sets the average number to 5.
97
PPH-1503 User Manual
Command
:SENSe[1]:PCURrent:AVERage?
Function
Returns the average number for pulse current measurement.
Example
:SENSe:PCURrent:AVERage?
Returns the average number.
Command
:SENSe[1]:PCURrent:MODE <name>
Function
Sets the pulse current measurement mode.
Description
Name
HIGH: High pulse mode (trigger on the rising edge).
LOW: Low pulse mode (trigger on the falling edge)
AVERage: Average pulse measurement.
Example
:SENSe:PCURrent:MODE HIGH
Sets the pulse current measurement mode to HIGH mode.
Command
:SENSe[1]:PCURrent:MODE?
Function
Queries the pulse current measurement mode.
Example
:SENSe:PCURrent:MODE?
Returns the pulse current measurement mode.
Command
:SENSe[1]:PCURrent:TIME:AUTO
Function
Sets the pulse current integetration time to automatic.
Example
:SENSe:PCURrent:TIME:AUTO
Sets the pulse current integetration time to automatic.
98
REMOTE CONTROL
Command
:SENSe[1]:PCURrent:TIME:HIGH <NRf>
Function
Sets the integration time for high pulse measurement.
Description
<NRf>
33.3~ 833333Step resolution of 33.3.
Example
:SENSe:PCURrent:TIME:HIGH 0.000233
Sets the integration time for high pulse measurement to 233uS.
Command
:SENSe[1]:PCURrent:TIME:HIGH?
Function
Queries integration time for high pulse measurement.
Example
:SENSe:PCURrent:TIME:HIGH?
Returns the integration time for high pulses.
Command
:SENSe[1]:PCURrent:TIME:LOW <NRf>
Function
Sets the integration time for low pulse measurement.
Description
<NRf>
33.3-833333Step resolution of 33.3
Example
:SENSe:PCURrent:TIME:LOW 0.000233
Sets the integration time for low pulse measurement to 233uS.
Command
:SENSe[1]:PCURrent:TIME:LOW?
Function
Returns the integration time for low pulse measurement.
99
PPH-1503 User Manual
Example
:SENSe:PCURrent:TIME:LOW?
Returns the integration time for low pulse measurement.
Command
:SENSe[1]:PCURrent:TIME:AVERage <NRf>
Function
Sets the integration time for the average pulse measurement.
Description
NRf
33-833333, step resolution of 33.3
Example
:SENSe:PCURrent:TIME:AVERage 0.000233
Sets the integration time for average pulse measurement to 233 microseconds.
Command
:SENSe[1]:PCURrent:TIME:AVERage?
Function
Returns the integration time for the average measurement.
Example
:SENSe:PCURrent:TIME:AVERage?
Returns the integration time for the average measurement.
Command
:SENSe[1]:PCURrent:SYNChronize[:STATe] <b>
Function
Sets the triggering option for pulse current measurement.
Description
<b>
0 /OFF: Digital trigger mode.
1/ON: Pulse level trigger mode.
Example
:SENSe:PCURrent:SYNChronize ON
The trigger mode is set to the pulse level trigger.
100
Loading...