Warning: Identifies conditions or practices that
could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that
could result in damage to the instrument or to
other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Frame or Chassis Terminal
Earth (ground) Terminal
SAFETY INSTRUCTIONS
This chapter contains important safety
instructions that you must follow during
operation and storage. Read the following before
any operation to ensure your safety and to keep
the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the
instrument.
4
SAFETY INSTRUCTIONS
Do not dispose electronic equipment as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased.
General
Guideline
CAUTION
Do not place any heavy object on the
instrument.
Avoid severe impact or rough handling that
leads to damaging the instrument.
Do not discharge static electricity to the
instrument.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening.
Do not disassemble the GPT-9600 unless you are
qualified.
(Measurement categories) EN 61010-1:2010 specifies the
measurement categories and their requirements as follows. The
GPT-9600 does not fall under category II, III or IV.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Power Supply
WARNING
AC Input voltage range:
100-120/220-240VAC ±10%
Frequency: 50Hz/60Hz
Fuse: T 4A 250V
To avoid electrical shock connect the protective
grounding conductor of the AC power cord to
an earth ground.
Safety Guidelines
5
GPT-9600 Series User Manual
Cleaning the
GPT-9600
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Operation
Environment
Location: Indoor, no direct sunlight, dust free,
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees
and their requirements as follows. The GPT-9600 falls under degree
2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric
strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused
by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to
condensation which is expected. In such conditions, equipment
is normally protected against exposure to direct sunlight,
precipitation, and full wind pressure, but neither temperature
nor humidity is controlled.
Storage
environment
Location: Indoor
Temperature: -10°C to 70°C
Relative Humidity: ≤ 85% (no condensation)
Disposal
Do not dispose this instrument as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased. Please make sure
discarded electrical waste is properly recycled to
reduce environmental impact.
6
SAFETY INSTRUCTIONS
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the safety tester in the United Kingdom, make sure the
power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
As the colours of the wires in main leads may not correspond with
the coloured marking identified in your plug/appliance, proceed
as follows:
The wire which is coloured Green & Yellow must be connected to
the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal
which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the
terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment
or contact the supplier.
This cable/appliance should be protected by a suitably rated and
approved HBC mains fuse: refer to the rating information on the
equipment and/or user instructions for details. As a guide, a cable
of 0.75mm2 should be protected by a 3A or 5A fuse. Larger
conductors would normally require 13A types, depending on the
connection method used.
Any exposed wiring from a cable, plug or connection that is
engaged in a live socket is extremely hazardous. If a cable or plug is
deemed hazardous, turn off the mains power and remove the cable,
any fuses and fuse assemblies. All hazardous wiring must be
immediately destroyed and replaced in accordance to the above
standard.
7
GPT-9600 Series User Manual
AC / DC Withstanding Voltage /
Insulation Resistance
Tester
GPT-9603
POWER
STARTSTOP
REMOTE
ESCFIELD
PASSFAILREADY TEST
HIGH VOLTAGE
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
RETURN
EDIT SAVE UTILITY
GPT-9600 Series Overview ................................................. 9
Series lineup ................................................................................................ 9
Model Overview ......................................................................................... 9
Main Features ............................................................................................. 9
Accessories and Package Contents ........................................................ 10
This chapter describes the safety tester in a
nutshell, including its main features and front /
rear panel introduction. After going through the
overview, please read the safety considerations in
the Set Up chapter.
The GPT-9600 Series Safety Testers are AC/DC withstanding
voltage and insulation resistance safety testers.
The GPT-9603 is an AC/DC withstanding and insulation resistance
safety tester.
The GPT-9602 is an AC/DC withstanding safety tester.
The GPT-9612 is an AC withstanding and insulation resistance
safety tester.
The GPT-9601 is purely an AC withstanding safety tester.
Note: Throughout this user manual, the terms ACW, DCW and IR
refer to AC Withstanding, DC Withstanding and Insulation
Resistance, respectively. GPT-9600 refers to any of the GPT-96XX
models.
Model Overview
Main Features
9
GPT-9600 Series User Manual
Features
Timer control
Safety discharge
Over temperature, voltage and current
Remote control start/stop interface terminal
Signal I/O port for pass/fail/test monitoring
and start/stop control/interlock
Standard Accessories
Part number
Description
N/A
GPT-96XX unit.
N/A
User manual CD
N/A
Quick start guide
GHT-114 x1
Test lead
Region dependent
Power cord
N/A
Remote terminal male plug
N/A
Interlock key
N/A
CTC (Calibration Traceable
Certificate)
Optional Accessories
Part number
Description
GHT-205
High voltage test probe
GHT-113
High voltage test pistol
Accessories and Package Contents
Check the package contents to make sure all accessories and
optional items are included before using the GPT-9600.
10
GETTING STARTED
GRA-417
Rack adapter panel
(19”, 4U)
Note
Keep the packaging, including the box, polystyrene
foam and plastic envelopes should the need arise
to return the unit to GW Instek.
11
GPT-9600 Series User Manual
PASS/FAIL indicators
Navigation keysFunction keys
HIGH VOLTAGE
indicator
RETURN
terminal
REMOTE terminal
STOP button
START button
POWER
button
Display
Scroll wheel
Configuration keys
READY indicator
TEST indicator
HIGH VOLTAGE
output terminal
AC / DC Withstanding Voltage /
Insulation Resistance
Tester
GPT-9603
POWER
STARTSTOP
REMOTE
ESCFIELD
PASSFAILREADY TEST
HIGH VOLTAGE
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
RETURN
EDIT SAVE UTILITY
Display
240 X 48 dot matrix display (LCD)
Function keys
The function keys correspond to the soft-keys
directly above on the main display.
Pass/Fail
indicators
PASS
FAIL
The PASS and FAIL indicators
light up upon a PASS or FAIL test
result.
Navigation keys
ESC key
ESC
The ESC key is used to exit out of
a menu or cancel a setting.
FIELD key
FIELD
The FIELD key is navigate to the
next menu item when the tester is
either in the EDIT status or in one
of the Utility menus.
Appearance
GPT-9603/9602/9612/9601
Front Panel
12
GETTING STARTED
Directional
arrow keys
The directional arrow keys are
only used for calibration. They are
not used during normal operation
by the operator. The directional
arrow keys can essentially be
ignored by the operator.
READY indicator
READY
The READY indicator is lit when
the tester is ready to begin testing.
TEST indicator
TEST
The TEST indicator is lit when a
test is on. The START button is
used to put the tester into TEST
status from READY status.
HIGH VOLTAGE
indicator
HIGH VOLTAGE
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
The HIGH VOLTAGE indicator
will flash when the output
terminal is active. Only after the
test has finished or stopped will
the indicator turn off.
HIGH VOLTAGE
output terminal
HIGH VOLTAGE
CAUTION
5.0 kVAC MAX.
6.0 kVDC MAX.
The HIGH VOLTAGE terminal
output is used for outputting the
testing voltage. The terminal is
recessed for safety. This terminal
is used in conjunction with the
RETURN terminal.
WARNING
USE EXTREME CAUTION.
Do not touch the HIGH VOLTAGE terminal
during testing.
13
GPT-9600 Series User Manual
RETURN terminal
RETURN
The RETURN terminal is used for
IR, DCW and ACW tests.
Scroll wheel
The scroll wheel is used to edit
parameter values or menu
settings.
UTILITY key
UTILITY
Used to enter the TEST Utility or
Common Utility menu.
EDIT key
EDIT
Used to enter the EDIT status. The
EDIT status allows you to select
the test mode and the test
parameters.
SAVE key
SAVE
The SAVE key is used to save test
parameters when in the EDIT
status or to save utility settings.
REMOTE
terminal
REMOTE
The REMOTE terminal is used to
connect to a remote controller.
STOP button
STOP
The STOP button is used to
stop/cancel tests. The STOP
button will also put the safety
tester in the READY status to
begin testing.
14
GETTING STARTED
START button
START
The START button is used to start
tests.
The START button can be used to
start tests when the tester is in the
READY status. Pressing the
START button will put the tester
in the TEST status.
POWER switch
POWER
Turns the power on. The safety
tester will always start up with the
last test setting from when the
instrument was last powered
down.
15
GPT-9600 Series User Manual
Fan
Fuse holderLine voltage
SIGNAL I/O
GND
TO AVOID ELECTRIC SHOCK THE POWER CORD
PROTECTIVE GROUNDING CONDUCTOR MUST BE
ONLY WITH SPECIFIED TYPE AND RATED FUSE.
NO OPERATOR SERVICEABLE COMPONENTS INSIDE.
DO NOT REMOVE COVERS. REFER SERVICING TO
FOR CONTINUED FIRE PROTECTION. REPLACE
CONNECTED TO GROUND.
QUALIFIED PERSONNEL.
WARNING
SER. NO. LB
SIGNAL I / O
CALIBRATION
USE ONLY
REMOVED FROM THE INSTRUMENT
BEFORE REPLACING THE FUSE
AC
MAX.
400VA
POWER
ENSURE THE POWER IS
LINE VOLTAGE
100 120V
( 50 / 60Hz )
250V
T 4A
FUSE
220 240V
GND
Calibration port
SIGNAL I/O port
SIGNAL I / O
The SIGNAL I/O port is used to
monitor the tester status (PASS,
FAIL, TEST) and input (START/
STOP signals). It is also used with
the Interlock key.
Calibration port
CALIBRATION
USE ONLY
The Calibration port is used for
calibration purposes only.
Calibration is not supported for
end-users. Please see your
distributor for details.
Fan/Fan Vents
Exhaust fan. Allow enough room
for the fan to vent. Do not block the
fan openings.
GND
GND
Connect the GND (ground)
terminal to the earth ground.
GPT-9603/9602/9612/9601 Rear Panel
16
GETTING STARTED
Line voltage input
and fuse holder
Line voltage input:
100-120/220-240VAC
Line Voltage Fuse: T 4A 250V
17
Set Up
Background
The GPT-9600 accepts line voltages of 100-120V
and 220-240V at 50Hz or 60Hz.
Steps
1. Connect the power cord to
the AC voltage input.
2. If the power cord does not
have an earth ground,
ensure the ground
terminal is connected to an
earth ground.
GND
Warning
Ensure the power cord is connected to an earth
ground. Failure could be harmful to the operator
and instrument.
3.Press the Power button.
POWER
4. When the unit is powering up, all the LED
indicators will light. Check to make sure all 5
LED indicators are working.
5. Check to make sure the System Self Test passes
without errors.
Power Up
GPT-9600 Series User Manual
18
GETTING STARTED
S Y S T E M S
S TE FLT E
F i r m w a r e C h e c i n gk
H a r d w a r e C h e c i n gk.
S y s t e m C h e c k i g . .n.
MT I E = 0 0R
MT I E = 0 0R. .
MT I E = 0 0R
MT I E = 0 0R
After the System Self Test completes, the tester will
go into READY status and be ready to operate.
H = 2 2 . 0 m A
m AL 0=0 0.7 4 S = 1T
AER
D Y
CWDA CWR IDO EMT M EI
READY Status
1 00
k VAm
19
GPT-9600 Series User Manual
Background
The GPT-9600 Series are a high voltage
instruments that output dangerous voltages.
The following section describes precautions and
procedures that must be followed to ensure a
safe work environment.
WARNING
The GPT-96XX generates voltages in excess of
5kVAC or 6kVDC. Follow all safety precautions,
warnings and directions given in the following
section when using the instrument.
1. Only technically qualified personnel should be
allowed to operate the safety tester.
2. The operating workplace must be fully isolated,
especially when the instrument is in operation.
The instrument should be clearly labeled with
appropriate warning signage.
3. The operator should not wear any conductive
materials, jewelry, badges, or other items, such
wrist watches.
4. The operator should wear insulation gloves for
high voltage protection.
5. Ensure the earth ground of the line voltage is
properly grounded.
6. Ensure any devices that are adversely affected
by magnetic fields are not placed near the
tester.
Workplace Precautions
20
GETTING STARTED
Background
The GPT-9600 Series are high voltage
instruments that output dangerous voltages.
The following section describes precautions and
procedures that must be followed to ensure that
the testers are operated in a safe manner.
WARNING
The GPT-96XX generates voltages of up to 5kVAC
or 6kVDC. Follow all safety precautions, warnings
and directions given in the following section when
using the instrument.
1. Never touch the safety tester, lead wires,
terminals, probes and other connected
equipment when the tester is testing.
2. Do not turn the safety tester on and off quickly
or repeatedly. When turning the power off,
please allow a few moments before turning the
power back on. This will allow the protection
circuits to properly initialize.
Do not turn the power off when a test is
running, unless in an emergency.
3. Only use those test leads supplied with the
instrument. Leads with inappropriate gauges
can be dangerous to both the operator and the
instrument.
4. Do not short the HIGH VOLTAGE terminal
with ground. Doing so could charge the chassis
to dangerously high voltages.
5. Ensure the earth ground of the line voltage is
properly grounded.
Operating Precautions
21
GPT-9600 Series User Manual
6. Only connect the test leads to the HIGH
VOLTAGE terminals before the start of a test.
Keep the test leads disconnected at all other
times.
7. Always press the STOP button when pausing
testing.
8. Do not leave the safety tester unattended.
Always turn the power off when leaving the
testing area.
9. When remotely controlling the safety tester,
ensure adequate safety measures are in place to
prevent:
Inadvertent output of the test voltage.
Accidental contact with the instrument during
testing. Ensure that the instrument and DUT are
fully isolated when the instrument is remotely
controlled.
10. Ensure an adequate discharge time for the
DUT.
When DCW or IR tests are performed, the DUT,
test leads and probes become highly charged.
The GPT-96XX has discharge circuitry to
discharge the DUT after each test. The time
required for a DUT to discharge depends on the
DUT and test voltage.
Never disconnect the safety tester before a
discharge is completed.
22
GETTING STARTED
Background
The GPT-9600 Series are high voltage devices
and as such, daily safety checks should be
made to ensure safe operation.
1. Ensure all test leads are not broken and are free
from defects such as cracks or splitting.
2. Ensure the safety tester is always connected to
an earth ground.
3. Test the safety tester operation with a low
voltage/current output:
Ensure the safety tester generates a FAIL
judgment when the HIGH VOLTAGE and
RETURN terminals are shorted (using the
lowest voltage/current as the testing
parameters).
WARNING
Do not use high voltages/currents when the HIGH
VOLTAGE and RETURN terminals are shorted. It
may result in damage to the instrument.
Basic Safety Checks
23