SAFETY INSTRUCTIONS
WARNING
Warning: Identifies conditions or practices that
could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that
could result in damage to the function generator or
to other objects or property.
DANGER High Voltage
Attention: Refer to the Manual
Protective Conductor Terminal
Earth (Ground) Terminal
DANGER Hot Surface
SAFETY INSTRUCTIONS
This chapter contains important safety instructions
that should be followed when operating and
storing the function generator. Read the following
before any operation to ensure your safety and to
keep the function generator in the best condition.
Safety Symbols
These safety symbols may appear in this manual or on the
instrument.
3
AFG-2000 Series User Manual
Double Insulated
Do not dispose electronic equipment as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased.
General
Guideline
CAUTION
Do not place heavy objects on the instrument.
Do not place flammable objects on the
instrument.
Avoid severe impact or rough handling that
may damage the function generator.
Avoid discharges of static electricity on or near
the function generator.
Use only mating connectors, not bare wires, for
the terminals.
The instrument should only be disassembled by
a qualified technician.
(Measurement categories) EN 61010-1:2010 specifies the
measurement categories and their requirements as follows. The
instrument falls under category II.
Measurement category IV is for measurement performed at the
source of a low-voltage installation.
Measurement category III is for measurement performed in a
building installation.
Measurement category II is for measurement performed on
circuits directly connected to a low voltage installation.
Measurement category I is for measurements performed on
circuits not directly connected to Mains.
Power Supply
WARNING
AC Input voltage: 100 ~ 240V AC, 50 ~ 60Hz.
Connect the protective grounding conductor of
the AC power cord to an earth ground to
prevent electric shock.
Safety Guidelines
4
SAFETY INSTRUCTIONS
Fuse
WARNING
Fuse type: F1A/250V.
Only qualified technicians should replace the
fuse.
To ensure fire protection, replace the fuse only
with the specified type and rating.
Disconnect the power cord and all test leads
before replacing the fuse.
Make sure the cause of fuse blowout is fixed
before replacing the fuse.
Cleaning the
function
generator
Disconnect the power cord before cleaning the
function generator.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid
into the function generator.
Do not use chemicals containing harsh products
such as benzene, toluene, xylene, and acetone.
Operation
Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
and avoid strong magnetic fields.
(Pollution Degree) EN 61010-1:2010 specifies pollution degrees and
their requirements as follows. The function generator falls under
degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric
strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused
by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to
condensation which is expected. In such conditions, equipment
is normally protected against exposure to direct sunlight,
5
AFG-2000 Series User Manual
precipitation, and full wind pressure, but neither temperature
nor humidity is controlled.
Do not dispose this instrument as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased. Please make sure
discarded electrical waste is properly recycled to
reduce environmental impact.
6
SAFETY INSTRUCTIONS
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the function generator in the United Kingdom, make sure the
power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
As the colours of the wires in main leads may not correspond with the
coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth
terminal marked with either the letter E, the earth symbol or coloured
Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is
marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal
marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact
the supplier.
This cable/appliance should be protected by a suitably rated and approved
HBC mains fuse: refer to the rating information on the equipment and/or
user instructions for details. As a guide, a cable of 0.75mm2 should be
protected by a 3A or 5A fuse. Larger conductors would normally require
13A types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a
live socket is extremely hazardous. If a cable or plug is deemed hazardous,
turn off the mains power and remove the cable, any fuses and fuse
assemblies. All hazardous wiring must be immediately destroyed and
replaced in accordance to the above standard.
7
AFG-2000 Series User Manual
Model name
AFG-2005
AFG-2105
AFG-2012
AFG-2112
AFG-2025
AFG-2125
Frequency Range
0.1Hz~5MHz
0.1Hz~12MHz
0.1Hz~25MHz
Output waveform
Sine, Square, Ramp, Noise, ARB
Amplitude range
0.1Hz~20MHz
1 mVpp to 10 Vpp (into 50Ω)
2 mVpp to 20 Vpp (open-circuit)
20MHzHz~25MHz
1 mVpp to 5 Vpp (into 50Ω)
2 mVpp to 10 Vpp (open-circuit)
Variable Offset
Variable Duty
SYNC (TTL) output
Save/Recall
Sweep operation
———
AM
———
FM
———
FSK
———
Frequency Counter
———
GETTING STARTED
The Getting started chapter introduces the
function generator’s main features, appearance
and introduces a quick instructional summary of
some of the basic functions. For comprehensive
operation instructions, please see the operation
chapter.
Main Features
8
GETTING STARTED
ARB
USB Interface
Performance
DDS technology using an FPGA provides high
resolution waveforms
25MHz DDS (Direct Digital Synthesis) signal
output series
0.1Hz resolution
Full Function Arbitrary Waveform Capability
20 MSa/s sample rate
10 MHz repetition rate
4 k-point waveform length
10-bit amplitude resolution
Ten 4k waveform memories
Features
Sine, Square, Ramp, Noise
Int/Ext AM, FM, FSK modulation
Modulation/sweep signal output
Save/recall 10 groups of setting memories
Output overload protection
ARB (Arbitrary Waveform) can be edited with
PC software
Interface
USB interface as standard
3.5 inch LCD
9
AFG-2000 Series User Manual
ARB
OUTPUT
50
W
50
W
SYNC
MAIN
OUTPUT
POWER
Save/Recall INT/EXT Hop LIN/LOG
Shape DEP/DEV Rate Start/Stop Gate
FUNC
FREQ
AMPL
OFST
DUTY
Point
Value
0
7 8 9
4 5 6
1 2 3
/
Hz/Vpp kHz/Vrms MHz/dBm % Shift
AM FM FSK Sweep Count
OUTPUT
Enter
Arbitrary Function Generator
AFG-2125
LCD Display Number
pad
Scroll Wheel
MAIN
output port
ARB keysPower
button
Output
control key
Enter key
Arrow keys SYNC output port
Operation
keys
Function keys
High-Z/50Ω
ARB
OUTPUT
50
W
50
W
SYNC
MAIN
OUTPUT
POWER
Save/Recall
FUNC
FREQ
AMPL
OFST
DUTY
Point
Value
0
7 8 9
4 5 6
1 2 3
/
Hz/Vpp kHz/Vrms MHz/dBm % Shift
OUTPUT
Enter
Arbitrary Function Generator
AFG-2025
LCD Display Number
pad
Scroll Wheel
MAIN
output port
ARB keysPower
button
Output
control key
Enter key
Arrow keys SYNC output port
Operation
keys
Function keys
High-Z/50Ω
Panel Overview
AFG-2105/2112/2125 Front Panel
AFG-2005/2012/2025 Front Panel
10
GETTING STARTED
LCD display
3.5 inch, 3 color LCD display.
Keypad
0
/
321
4
7859
6
The digital keypad is used to
enter values and parameters. The
keypad is often used in
conjunction with the selection
keys and variable knob.
Scroll Wheel
The scroll wheel is used to edit
values and parameters in steps of
1 digit. Used in conjunction with
the arrow keys.
DecreaseIncrease
Arrow keys
Used to select digits when editing
parameters.
Output ports
OUTPUT
50
W
50
W
SYNC
MAIN
OUTPUT
SYNC output port (50Ω
impedance).
Main output port (50Ω
impedance).
Enter key
Enter
Used to confirm input values.
Power button
POWER
Turns the instrument power
on/off.
Output control
key
OUTPUT
Turns the output on/off.
Output
Impedance
Shift
+
High Z/50Ω
OUTPUT
Toggles the output impedance
between 50Ω and High-Z.
Operation keys
Hz/Vpp
Selects Hz or Vpp units.
11
AFG-2000 Series User Manual
Shift
+
Save/Recall
Hz/Vpp
Saves or recalls waveforms from
memory.
kHz/Vrms
Selects kHz or Vrms units.
Shift
+
INT/EXT
kHz/Vrms
Sets the source to internal or
external for the modulation and
FSK functions*.
MHz/dBm
Selects MHz or dBm units.
Shift
+
Hop
MHz/dBm
Sets the ―Hop‖ frequency for FSK
modulation*.
%
Selects % units.
Shift
+
LIN/LOG
%
Sets the sweep to linear or
logarithmic*.
Shift
The shift key is used to select the
secondary functions on the
operation keys.
AM
The AM key is used to turn AM
modulation on/off*.
Shift
+
Shape
AM
Selects the modulation
waveform*.
FM
The FM key is used to turn FM
modulation on/off*.
Shift
+
DEP/DEV
FM
Selects the modulation depth or
the frequency deviation*.
FSK
Selects FSK modulation*.
Shift
+
Rate
FSK
Sets the AM, FM, FSK modulation
and sweep function rate*
Sweep
Selects the Sweep function*.
12
GETTING STARTED
Shift
+
Start/Stop
Sweep
Sets the Start or Stop frequency*.
Count
Turns the frequency counter
on/off*.
Shift
+
Gate
Count
Sets the frequency counter gate
time*.
ARB edit keys
ARB
Value
Point
Arbitrary waveform editing keys.
The Point key sets the ARB point
numbers.
The Value key sets the amplitude
value of the selected point.
Function keys
FUNC
The FUNC key is used to select
the output waveform type:
Sine, Square, Ramp, Noise, ARB.
FREQ
Sets the frequency of the selected
waveform.
AMPL
Sets the amplitude of the selected
waveform.
OFST
The OFST sets the DC offset for
the selected waveform.
DUTY
The DUTY key sets the duty cycle
of square and ramp waveforms.
*indicates functions/features for the AFG-2105/2112/2125 only.
13
AFG-2000 Series User Manual
NO OPERATOR SERVICEABLE COMPONENTS INSIDE.
DO NOT REMOVE COVERS. REFER SERVICING TO
TO AVOID ELECTRIC SHOCK THE POWER CORD PROTECTIVE
GROUNDING CONDUCTOR MUST BE CONNECTED TO GROUND.
WARNING
QUALIFIED PERSONNEL.
SER.NO. LABEL
AC 100-240V
50-60Hz 25VA
OUTPUT INPUT
MOD Counter
Trigger MOD
MOD input
Trigger input
MOD output
Type B USB portPower socket
Counter input
NO OPERATOR SERVICEABLE COMPONENTS INSIDE.
DO NOT REMOVE COVERS. REFER SERVICING TO
TO AVOID ELECTRIC SHOCK THE POWER CORD PROTECTIVE
GROUNDING CONDUCTOR MUST BE CONNECTED TO GROUND.
WARNING
QUALIFIED PERSONNEL.
SER.NO. LABEL
AC 100-240V
50-60Hz 25VA
Type B USB portPower socket
MOD output
OUTPUTINPUT
MODCounter
TriggerMOD
Modulation output port.
Counter input
Counter input port.
MOD input
Modulation input port.
Trigger input
Trigger input port.
Type B USB port
The type B USB port is used to
connect the function generator to a
PC for remote control.
Power Socket
Input
AC 100-240V
50-60Hz 25VA
Power input: 100~240V AC
50~60Hz.
AFG-2105/2112/2125 Rear Panel
AFG-2005/2012/2025 Rear Panel
14
Display
Waveform type
Counter settings
USB icon
Frequency display
Secondary parameter
display
Modulation, sweep,
counter menu
Waveform type
Press the function key to cycle through different
output waveforms.
Counter settings
Gate time counter settings*.
USB icon
Shows the USB interface status.
Frequency
Display
Displays the main waveform frequency settings.
Secondary
parameter display
Displays secondary waveform parameters and
settings.
Modulation,
sweep, counter
menu
Displays the modulation, sweep and counter
functions as well as the modulating waveform and
source*.
*indicates functions/features for the AFG-2105/2112/2125 only.
GETTING STARTED
15
AFG-2000 Series User Manual
Background
This section describes how adjust the handle and
power up the function generator.
Adjusting the
stand
Pull out the handle
sideways and rotate
it.
ARB
OUTPUT
50W
50W
SYNC
MAIN
OUTPUT
POWER
Save/Recall INT/EXT Hop LIN/LOG
Shape DEP/DEV Rate Start/Stop Gate
FUNC
FREQ
AMPL
OFST
DUTY
Point
Value
0
7 8 9
4 5 6
1 2 3
/
Hz/Vpp kHz/Vrms MHz/dBm % Shift
AM FM FSK Sweep Count
OUTPUT
Enter
Arbitrary Function Generator
AFG-2125
Place the AFG
horizontally.
Place the handle
upright to tilt the
stand.
Place the handle
vertically to hand
carry.
Setting up the Function Generator
16
GETTING STARTED
Power Up
1. Connect the power cord to
the socket on the rear
panel.
2. Press the power button on
the front panel.
POWER
3. The instrument will turn on and load the
default settings (see page 32 for default
settings).
The function generator is now ready to be used.
17
AFG-2000 Series User Manual
How to use the Digital Inputs ............................................ 19
Selecting a Waveform ........................................................ 21
Sine Wave ............................................................................. 21
This chapter lists operation shortcuts and default factory settings.
Use this chapter as a handy reference for instrument functions. This
chapter is to be used as a quick reference; for detailed explanations
on parameters, settings and limitations, please see the operation
chapter (page 34) or specifications (page 137).
18
AFG-2000 Series User Manual
Background
The AFG-2000 has three main types of digital
inputs: the number pad, arrow keys and the scroll
wheel. The following instructions will show you
how to use the digital inputs to edit parameters.
1. First select the
function that must
be edited pressing
one of the function
or ARB keys. The
selected function
will flash.
ARB
OUTPUT
50
W
50
W
SYNC
MAIN
OUTPUT
POWER
Save/Recall
FUNC
FREQ
AMPL
OFST
DUTY
Point
Value
0
789
456
123
/
Hz/VppkHz/VrmsMHz/dBm%Shift
OUTPUT
Enter
Arbitrary Function Generator
AFG-2025
ARB keys
Function keys
2. To edit a parameter, use the
arrow keys to move the
cursor to the digit that
needs to be edited.
cursor
How to use the Digital Inputs
19
AFG-2000 Series User Manual
3. Use the scroll wheel to
increment the parameter by
the resolution of the digit
under the cursor.
In the example above, the
scroll wheel will increment
the parameter in 0.1 volt
increments.
Clockwise increases the
value, counterclockwise
decreases the value.
4. Press the Enter key to
confirm the new parameter
value.
Enter
5. Alternatively, the number
pad can be used to set the
value of the selected
parameter.
0
/
321
4
7859
6
6. To finish editing with the
number pad, select the unit
with one of the unit keys.
(Hz, kHz, MHz, Vpp,
Vrms, dBm, %)
Hz/Vpp
kHz/Vrms
MHz/dBm
%
20
QUICK REFERENCE
Example: Sine Wave, 10kHz, 1Vpp, 2Vdc
Output
50
W
MAIN
1. Press the FUNC
key repeatedly to
select the Sine
wave.
FUNC
→
2.Press FREQ > 1 > 0
> kHz.
FREQ
1
0
kHz/Vrms
3.Press AMPL > 1 >
Vpp.
AMPL
1
Hz/Vpp
4.Press OFST > 2 >
Vpp.
OFST
2
Hz/Vpp
5. Press the OUTPUT
key.
OUTPUT
Example: Square Wave, 10kHz, 3Vpp, 75% duty cycle
Output
50
W
MAIN
1. Press the FUNC
key repeatedly to
select the Square
wave.
FUNC
→
2.Press FREQ > 1 > 0
> kHz.
FREQ
1
0
kHz/Vrms
Selecting a Waveform
Sine Wave
Square Wave
21
3.Press AMPL > 3 >
Vpp.
AMPL
3
Hz/Vpp
4.Press DUTY > 7 > 5
> %.
DUTY
7
5
%
5. Press the output
key.
OUTPUT
Ramp Wave
Example: Ramp Wave, 10kHz, 3Vpp, 25% symmetry
Output
50
W
MAIN
1. Press the FUNC
key repeatedly to
select the Ramp
wave.
FUNC
→
2.Press FREQ > 1 > 0
> kHz.
FREQ
1
0
kHz/Vrms
3.Press AMPL > 3 >
Vpp.
AMPL
3
Hz/Vpp
4.Press DUTY > 2 > 5
> %.
DUTY
2
5
%
5. Press the OUTPUT
key.
OUTPUT
AFG-2000 Series User Manual
22
ARB
Example: 2 ARB points, 10 kHz, 1Vpp.
Output
50
W
MAIN
1. Press the FUNC
key repeatedly to
select the ARB
wave.
FUNC
→
2.Press FREQ > 1 > 0
> kHz.
FREQ
1
0
kHz/Vrms
3.Press AMPL > 1 >
Vpp.
AMPL
1
Hz/Vpp
4.Press Point > 0 >
Enter.
Point
0
Enter
5.Press Value > 5 > 1
>1 > Enter.
Value
511
Enter
6.Press Point > 1 >
Enter.
Point
1
Enter
7.Press Value > ± > 5
> 1 >1 > Enter.
(-511)
Value
/
511
Enter
8. Press the OUTPUT
key.
OUTPUT
ARB - Points
QUICK REFERENCE
23
AFG-2000 Series User Manual
Example: AM modulation. 100Hz modulating square wave. 1 Vpp,
1kHz Sine wave carrier. 70% modulation depth. Internal source
signal.
Output
50
W
MAIN
1. Press the FUNC
key repeatedly to
select the Sine
wave.
FUNC
→
2.Press FREQ > 1 >
kHz.
FREQ
1
kHz/Vrms
3.Press AMPL > 1 >
Vpp.
AMPL
1
Hz/Vpp
4.Press AM.
AM
5.Press Shift >
INT/EXT > select
INT source.
Shift
+
INT/EXT
kHz/Vrms
6. Press Shift > Shape
repeatedly to select
the Square wave.
Shift
+
Shape
AM
→
7. Press Shift > Rate
> 1 > 0 > 0 > Hz.
Shift
+
Rate
FSK
100
Hz/Vpp
Modulation
AM (2100 series only)
24
QUICK REFERENCE
8. Press Shift >
DEP/DEV> 7 > 0 >
%.
Shift
+
DEP/DEV
FM
7
0
%
9. Press the OUTPUT
key.
OUTPUT
10. Press AM again to
deselect the AM
function.
AM
Example: FM modulation. 100Hz modulating square wave. 1Vpp,
1kHz Sine wave carrier. 100 Hz frequency deviation. Internal
Source.